You are on page 1of 8

GASTROENTEROLOGY 2013;144:918 –925

A Population-based Comparison of Immunochemical Fecal Occult Blood


Tests for Colorectal Cancer Screening
THIBAUT RAGINEL,1,2,3,4 JOSETTE PUVINEL,5 OLIVIER FERRAND,6 VERONIQUE BOUVIER,1,2,3,4 ROMUALD LEVILLAIN,7
ANGELA RUIZ,7 OLIVIER LANTIERI,7 GUY LAUNOY,1,2,3,4 and LYDIA GUITTET1,2,3,4
CLINICAL AT

1
Centre Hospitalier Universitaire (CHU) de Caen, Caen; 2Normandie University, Caen; 3Université de Caen Basse-Normandie (UCBN), Cancers and Preventions,
Caen; 4INSERM U1086, Caen; 5Association Bourbonnaise Interdépartementale de Dépistage des Cancers (ABIDEC), Moulins; 6Action de Dépistage Organisé des
Cancers 18 (ADOC18), Saint-Doulchard; and 7Institut InterRégional pour la Santé (IRSA) de Tours, La Riche, France

racy than the Magstream FIT. ClinicalTrials.gov num-


See related article, Anderson AE et al, on page 526 ber: NCT01251666.
in CGH.
Keywords: Colorectal Neoplasm; Early Detection; Sensitiv-
ity; Specificity.

BACKGROUND & AIMS: Quantitative fecal immuno-


chemical tests (FITs) identify individuals with colorectal
cancer with greater levels of accuracy than guaiac tests. We
compared the performances of 2 FITs in a population
W ith roughly 1 million cases and 500,000 deaths
each year worldwide, colorectal cancer is a major
public health concern, especially in industrialized coun-
undergoing screening for colorectal cancer. METHODS: tries.1 The natural history of this cancer allows for early
We collected fecal samples from 19,797 individuals in detection. Until now, 2 strategies have been shown to
France (age, 50⫺74 y) who participated in a colorectal reduce the mortality associated with this cancer signifi-
cancer screening program, from June 2009 through May cantly: the regular guaiac fecal occult blood test2 and
2011. Samples were analyzed using the Magstream (Fu- once-only flexible sigmoidoscopy.3–5
jirebio Inc, Tokyo, Japan) and OC Sensor (Eiken Chemical Since the beginning of the 1990s, it has been known
Co, Tokyo, Japan) (2 samples each) FITs, as well as the that immunochemical methods allow for improvement of
Hemoccult II guaiac test (SKD, Villepinte, France) (3 sam- sensitivity of fecal occult blood tests.6 Automated quan-
ples each). Colonoscopies were performed for patients titative fecal immunochemical tests (FITs) make adapta-
tion of the cut-off value possible to limit the loss in
with positive results from all 3 tests. The cut-off values for
specificity. Two randomized controlled trials, conducted
levels of hemoglobin in buffer and stools were 55 ng/mL
in The Netherlands on roughly 10,000 average-risk indi-
and 180 ␮g/g for the Magstream and 150 ng/mL and 30
viduals each, found that the OC Sensor FIT (Eiken Chem-
␮g/g for the OC Sensor, respectively. Results from the
ical Co, Tokyo, Japan) had both better screening perfor-
FITs were compared with those from the guaiac test for
mances and higher acceptability than the guaiac
cut-off values for stool samples, positivity rates, and the Hemoccult II test (SKD, Villepinte, France).7,8 One study
receiver operating characteristic curve values. The num- conducted in France on roughly 32,000 subjects found
bers needed to screen and the numbers needed to scope to that the Magstream FIT (Fujirebio Inc, Tokyo, Japan) also
detect an advanced neoplasia (cancer, adenoma ⱖ10 mm, had a better diagnostic accuracy than Hemoccult II.9,10
or high-grade dysplasia) were calculated. RESULTS: A For both Magstream and OC Sensor FITs, it has been
positive test result was found in 1224 participants (6.2%); shown that a gain in both sensitivity and specificity is
1075 (87.8%) underwent a colonoscopy examination. Of possible compared with the guaiac test Hemoccult II,
these, 334 were found to have advanced neoplasia. Con- providing an adequate cut-off value and using a small
sidering the cut-off values associated with the positivity number of samples.9,11–13 These results recently were con-
rate of Hemoccult II (1.6%), the numbers needed to screen firmed in France.14 An indirect comparison based on a
were 239 for Hemoccult II, 166 for a 1-sample Magstream literature review was in favor of slight superiority of OC
FIT, and 129 for a 1-sample OC Sensor FIT; the numbers Sensor over Magstream.15 However, this analysis suffered
needed to scope were 3.3, 2.3, and 1.8, respectively. For the from limitations inherent to comparison between differ-
same false-positive rate as Hemoccult II (0.98%), the true- ent populations.
positive rates for Magstream and OC Sensor FITs were
0.65% and 0.90% respectively, compared with 0.42% for Abbreviations used in this paper: FIT, fecal immunochemical test;
Hemoccult II. The OC Sensor FIT had a greater area under FOBT, fecal occult blood test; FPR, false-positive rate; NNScope, num-
the receiver operating characteristic curve value than the ber needed to scope; NNScreen, number needed to screen; ROC, re-
ceiver operating characteristic; TPR, true-positive rate.
Magstream FIT. CONCLUSIONS: Based on results from © 2013 by the AGA Institute
a large, population-based study, the OC Sensor FIT iden- 0016-5085/$36.00
tifies patients with colorectal cancer with greater accu- http://dx.doi.org/10.1053/j.gastro.2013.01.042
May 2013 COMPARISON OF PERFORMANCES OF 2 FITS 919

The aim of this study was therefore to compare tion, if Hemoccult II was not analyzed or not processed and no
directly the screening performances of Magstream and FIT was positive, the subject was asked to repeat the Hemoccult
OC Sensor FITs in the targeted average-risk population, II only, without performing any FIT again.
taking into account the cut-off value and the number The global screening test result was sent to both the patient
and general practitioner, without details on each fecal occult
of samples.
blood test (FOBT) result. If the screening was positive, the
subject was encouraged to undergo a colonoscopy. The colono-
Materials and Methods scopy results were recorded following the usual ongoing process
Study Population and Design for an organized screening program in the 2 areas. The colono-

CLINICAL AT
Between June 2009 and May 2011, there were 19,797 scopy findings were classified according to the most advanced
residents of 2 French counties (Allier and Cher) who fulfilled pathologic lesion found. Advanced neoplasias included high-risk
inclusion criteria for the ongoing biennial organized screening adenomas (adenomas measuring ⱖ10 mm or adenomas with
program (aged 50 –74 y, no digestive symptoms or personal or high-grade dysplasia) and invasive cancers (invasion of malig-
first-degree relatives with a history of colorectal cancer or ad- nant cells beyond the muscularis mucosae). Intramucosal carci-
vanced adenoma, no colonoscopy in the past 5 years) who were nomas were included in high-risk adenomas. Hyperplasic polyps
enrolled in a study comparing Magstream and OC Sensor FITs. were not included as neoplasia. Patients with incomplete colono-
Participation in the study was suggested at the time of central- scopy, no invasive cancer diagnosed, and no further exploration
ized invitation. The general practitioners consulted by the pa- to confirm the absence of any pathologic lesion (double-contrast
tients checked the inclusion criteria and provided the guaiac barium enema or virtual colonoscopy) were considered as having
Hemoccult II test (SKD, France) (routine test) together as Mag- no colonoscopy result.
stream and OC Sensor FITs to all subjects agreeing to join the
study. All participants signed a written informed consent. Statistical Analysis
For the guaiac test, 2 samples per stool on 3 consecutive Three strategies were used to compare the performances
bowel movements were required. For each of the FITs, 1 sample of the FITs: at identical cut-off values of hemoglobin concen-
per test on 2 consecutive bowel movements was required. Only tration in the stools (biochemical point of view), at an identical
the guaiac test was performed on the first stool. Then, for each positivity rate implying a similar number of colonoscopies (en-
of the last 2 stools sampled, subjects were asked to perform all doscopic workload point of view), and for either a similar num-
3 tests on the same stool, but no order between tests was ber of true-positive or false-positive results (patient’s harm or
specified. In Allier county, in accordance with the ongoing pro- benefit point of view).
gram, a specific diet (meatless for 3 days) was recommended. To this purpose, the performances of each FIT were analyzed,
Subjects performed the 3 tests themselves at home (on the kit varying the number of samples analyzed and the positivity cut-
card for Hemoccult II, in OC-Auto sampling bottle for OC off value. For the 2-sample FIT strategy, the test was positive if
Sensor and in New Hemtube for Magstream), recorded the date at least one of the samples contained a concentration of hemo-
of each bowel movement involved, and sent the 3 tests by mail globin greater than the cut-off value. For the 1-sample FIT
to the centralized Institut InterRégional pour la Santé labora- strategy, only the last sample was considered and compared with
tory. The Hemoccult II reading was performed by trained staff the cut-off value. For each FIT, all concentrations of hemoglobin
without rehydration as planned in the national organized greater than the cutoff having determined colonoscopy in the
screening program, and blinded to the results of the FITs. study were assessed as alternative cutoffs. However, the cut-off
Magstream samples were analyzed using the Magstream HT value selected by the manufacturer to define Magstream as a
instrument, and OC Sensor samples were analyzed using the OC qualitative test is smaller than the cut-off value used in our
Sensor Diana instrument. As recommended, crude results of study (80 ␮g/g instead of 180 ␮g/g of hemoglobin in the stools).
Magstream and OC Sensor FITs were standardized by conver- Many of the subjects with one sample of Magstream between 80
sion into concentration of hemoglobin in the stools (see the and 180 ␮g/g of hemoglobin in the stools (positive at the
Supplementary Appendix for conversion details derived from manufacturer’s cut-off value, but negative at the study cut-off
previous analytic studies).16 –18 value) were positive for either Hemoccult II, one of the OC
Screening was considered positive if the Hemoccult II test was Sensor samples, or the other sample of Magstream at the study
positive (at least 1 of the 6 slots was positive), or at least 1 of the cut-off value and were hence offered a colonoscopy. Therefore,
2 samples of OC Sensor reported a concentration of hemoglobin the performances of one sample Magstream at the manufactur-
in the stool greater than 30 ␮g/g (eg, 150 ng/mL hemoglobin in er’s cut-off value were derived from the study data, assuming
the buffer), or at least 1 of the 2 samples of Magstream reported that the small proportion of missing colonoscopies would have
a concentration of hemoglobin in the stool greater than 180 detected targeted lesions at a similar frequency as the performed
␮g/g (eg, 55 ng/mL in the buffer). The study cut-off value for ones (see the Supplementary Appendix). The same extrapolation
Magstream was the cut-off value that led, in a former study, to was performed for 1-sample OC Sensor FIT at the manufactur-
the same number of false-positive results as Hemoccult II, a er’s cut-off value (20 ␮g/g of hemoglobin in the stools).
scenario compatible with financial resources and availability of For each FOBT, the results of colonoscopies of subjects with
endoscopists characterizing health care systems such as the a positive test result were classified as true-positive results (sub-
French one.9 A similar scenario was targeted for OC Sensor, jects with a targeted lesion) or false-positive results (subjects
adopting a cut-off value near the manufacturer’s one, limiting with no targeted lesion). Then, the true-positive rate (TPR) was
the number of positive results for a 2-sample strategy but allow- defined as the proportion of true-positive results among the
ing comparison with other studies.7,12,19,20 subjects who performed the FOBT. The false-positive rate (FPR)
Because Hemoccult II was the routine test in France, if it was was defined as the proportion of false-positive results among the
negative and both FITs were either negative or unable to be subjects who performed the FOBT. In practice, the TPR is
analyzed, the screening was considered to be negative. In addi- directly proportional to the sensitivity of a screening test,
920 RAGINEL ET AL GASTROENTEROLOGY Vol. 144, No. 5

whereas the FPR is inversely proportional to its specificity. The In Table 1, 16.5% of participants underwent colorectal
paired design with blinding of details on test results allows one cancer screening for the first time. The median and mean
to draw a receiver operating characteristics (ROC) curve, plotting times between the last bowel movement for FOBT and
the TPR against the FPR. laboratory analysis were approximately 3 and 4 days, re-
The number needed to screen (NNScreen) to detect 1 subject
spectively. The rate of nonanalyzable tests decreased from
with a targeted lesion was calculated as the inverse of the TPR.
The number needed to scope (NNScope) to detect 1 subject with Magstream (4.4%) to Hemoccult II (2.0%) and OC Sensor
a targeted lesion was calculated as the inverse of the proportion (1.6%) (P ⬍ 10⫺3). A total of 1301 (6.6%) subjects had at
of true-positive results among subjects with a positive test (in- least 1 of the 3 tests nonanalyzable or not performed.
CLINICAL AT

verse of the positive predictive value). The analysis was con- Women and older subjects were significantly more fre-
ducted considering, first, advanced neoplasias, and, second, in- quent among this population. The positivity rate of
vasive cancers as targeted lesions. Hemoccult II was 1.6% in both areas. When considering
Proportions were compared with chi-square tests. A sensitivity 2 samples at the study cut-off values and only analyz-
analysis was conducted, first, drawing the ROC curves for detec- able tests, the positivity rates were identical for both
tion of advanced neoplasias excluding all subjects with at least areas for OC Sensor, but higher in Cher than Allier area
one nonanalyzable test, and, second, stratified on study area.
for Magstream. The colonoscopy rate did not differ
Statistical analysis was performed with R 2.12.121 (R Core
Team, Vienna, Austria) for Mac OS 10.6.
between tests (P ⫽ .73).
The study was approved by the local ethics committee (Co- The performances of the 3 FOBTs are compared in
mité de Protection des Personnes) and the French sanitary au- Table 2, for the study conditions and following the man-
thority (Agence Française de Sécurité Sanitaire des Produits de ufacturer’s guidelines. In the study conditions, for both
Santé). All participants provided written informed consent. The FITs and with only 1 sample, the NNscreen was smaller
trial was recorded on clinicaltrials.gov under the identifier than with the guaiac test for detection of invasive cancers,
NCT01251666. All authors had access to the study data and or advanced neoplasias. Despite an increase in the total
reviewed and approved the final manuscript. number of colonoscopies, the NNscope for detection of
these lesions was also smaller than with the guaiac test.
Results For each FIT, adding a second sample at the same cut-off
Figure 1 presents the study flow chart. Among the value decreased the NNscreen, by increasing the number
1224 subjects who were positive for at least one of the of colonoscopies at the expense of an increase in the
tests, an analyzable colonoscopy was performed for 1075 NNscope. The positivity rate of 1-sample Magstream at
subjects (87.8%), with a mean time interval between a the manufacturer cut-off value was similar to that with
positive screening and colonoscopy of 71 days (median, 56 1-sample OC Sensor at the study cut-off value, with a
days). No colonoscopy complications were reported. A similar positive predictive value.
total of 334 participants were diagnosed with advanced Comparison of TPR and FPR for detection of advanced
neoplasia at colonoscopy. neoplasias according to the concentration of hemoglobin

Figure 1. Study flow chart.


May 2013 COMPARISON OF PERFORMANCES OF 2 FITS 921

Table 1. Demographic Characteristics and Test Results of Included Subjects


All participants Excluding subjects with at least 1 nonanalyzable test

Allier n (%) Cher n (%) P value Allier n (%) Cher n (%) P value
Sex .14 .19
Men 6649 (44.6) 2125 (43.4) 6266 (45.0) 2002 (43.9)
Women 8252 (55.4) 2771 (56.6) 7665 (55.0) 2563 (56.1)
Age, y ⬍10⫺3 ⬍10⫺3
50–54 2984 (20.0) 920 (18.8) 2828 (20.3) 865 (19.0)

CLINICAL AT
55–59 3226 (21.7) 1051 (21.5) 3045 (21.9) 982 (21.5)
60–64 3591 (24.1) 1340 (27.4) 3380 (24.3) 1265 (27.7)
65–69 2408 (16.2) 843 (17.2) 2240 (16.1) 777 (17.0)
70–74 2692 (18.1) 742 (15.2) 2438 (17.5) 676 (14.8)
Screening round ⬍10⫺2 ⬍10⫺3
First 2526 (17.0) 741 (15.1) 2365 (17.0) 675 (14.8)
Subsequent 12,375 (83.1) 4155 (84.9) 11,566 (83.0) 3890 (85.2)
Test results
Hemoccult II ⬍10⫺3 .98
Positive 238 (1.6) 78 (1.6) 225 (1.6) 74 (1.6)
Negative 14,428 (96.8) 4663 (95.2) 13,706 (98.4) 4491 (98.4)
Nonanalyzable 209 (1.4) 148 (3.0) — —
Not performed 26 (0.2) 7 (0.1) — —
Magstreama ⬍10⫺2 ⬍.05
Positive 451 (3.0) 181 (3.7) 444 (3.2) 173 (3.8)
Negative 13,747 (92.3) 4539 (92.7) 13,487 (96.8) 4392 (96.2)
Nonanalyzable 592 (4.0) 148 (3.0) — —
Not performed 111 (0.7) 28 (0.6)
OC Sensorb .01 .69
Positive 594 (4.0) 207 (4.2) 558 (4.0) 189 (4.1)
Negative 14,038 (94.2) 4633 (94.6) 13,373 (96.0) 4376 (95.9)
Nonanalyzable 217 (1.5) 43 (0.9) — —
Not performed 52 (0.3) 13 (0.3)

NOTE. Characteristics of the population and positivity rates for each test (study cut-off value and for 2-sample FITs).
aTwo-sample Magstream at a cut-off value of 55 ng/mL Hb in the buffer (⬃180 ␮g/g in the stools).
bTwo-sample OC Sensor at a cut-off value of 150 ng/mL Hb in the buffer (30 ␮g/g in the stools).

in the stools is provided in Supplementary Figure 1. For with either 1 or 2 samples (providing a higher cut-off
each FIT, the TPR and FPR decreased as the cut-off value, value), for both the Magstream and OC Sensor FITs.
expressed as the concentration of hemoglobin in the When fixing the positivity rate at a higher level (3.0%),
stools, increased. Furthermore, adding a second sample of for both the Magstream and OC Sensor the NNScreen
the same FIT at the same cut-off value increased both the decreased, and the NNscope increased compared with a
TPR and the FPR. In the range of cut-off values in which 1.6% positivity rate, while staying largely better than
performances of both FITs were evaluable, the TPR of with the guaiac test. Both the NNscope and NNscreen
Magstream was greater than that of OC Sensor at the observed with the OC Sensor were better than with the
same cut-off value and using the same number of sam- Magstream.
ples. However, at the same time, the FPR of Magstream
Figure 2 compares the ROC curves associated with the
also was greater than that of OC Sensor.
guaiac test, Magstream, or OC Sensor FITs for the detec-
In Table 3, for a similar number of colonoscopies, the
tion of advanced neoplasias. Reading Figure 2 vertically,
NNscreen for detection of advanced neoplasias was de-
creased by 30% using 1-sample Magstream FIT instead of for a similar FPR of 9.8‰, the TPR increased from
the guaiac test (166.4 instead of 238.5). At the same time, Hemoccult II (TPR, 4.2‰), to 1-sample Magstream (7.3‰)
the NNscope also was decreased from 3.3 to 2.3. For the and 1-sample OC Sensor (9.7‰). Reading Figure 2 hori-
same total number of colonoscopies, both the NNscreen zontally, for a similar TPR of 4.2‰, the FPR decreased
and the NNscope associated with 1-sample OC Sensor from Hemoccult II (FPR, 9.8‰) to 1-sample Magstream
FIT were smaller than with 1-sample Magstream. Findings (5.1‰) and 1-sample OC Sensor (2.8‰). Globally, the area
were similar if the targeted lesion was invasive cancer only. below the curve of OC Sensor was greater than that of
For both the Magstream and OC Sensor, the same num- Magstream. Similar findings were observed for invasive
ber of positive subjects could result from use of 1 sample, cancers. In the range of positivity rates evaluable for both
or from use of 2 samples at a higher cut-off value. Then, number of samples given the study design, the perfor-
for a similar number of colonoscopies, the NNscope and mances for detection of advanced adenomas observed
NNscreen for detection of advanced neoplasias were close with either 1- or 2-sample FITs (at different cut-off values)
922 RAGINEL ET AL GASTROENTEROLOGY Vol. 144, No. 5

Table 2. Comparison of Clinical Performances of Hemoccult II, Magstream, and OC Sensor Fecal Occult Blood Tests in Study
Conditions, and According to Manufacturer Guidelines (Extrapolated)
Manufacturer cut-off
Study cut-off valuea value

Magstream OC Sensor Magstream OC Sensor


FIT FIT FIT FIT
Guaiac
test 1 sample 2 samples 1 sample 2 samples 1 sample 1 sample
CLINICAL AT

Cut-off value (␮g hemoglobin/g of stools) NA 180 180 30 30 80 20


Cut-off value (ng hemoglobin/mL in the buffer) NA 55 55 150 150 20 100
Positives 316 390 632 551 801 548 694
(positivity rate, %) (1.60) (1.97) (3.19) (2.78) (4.05) (2.77) (3.51)
Colonoscopies 277 344 554 488 712 484 615
(colonoscopy rate, %) (87.7) (88.2) (87.7) (88.6) (88.9) (88.3) (88.6)
Advanced neoplasias 83 146 210 225 290 209 275
(positive predictive value, %) (30.0) (42.4) (37.9) (46.1) (40.7) (43.2) (44.7)
(true-positive rate, %) (0.42) (0.74) (1.06) (1.14) (1.46) (1.06) (1.39)
Number needed to scope 3.3 2.4 2.6 2.2 2.5 2.3 2.2
Number needed to screen 238.5 135.6 94.3 88.0 68.3 94.7 72.0
Invasive cancers 19 28 36 37 44 37 44
(positive predictive value, %) (6.9) (8.1) (6.5) (7.6) (6.2) (7.6) (7.2)
(true-positive rate, %) (0.10) (0.14) (0.18) (0.19) (0.22) (0.19) (0.22)
Number needed to scope 14.6 12.3 15.1 13.2 16.2 13.1 14.0
Number needed to screen 1041.9 707.0 549.9 535.1 449.9 535.1 449.9
False positives 194 198 344 263 422 275 340
(false-positive rate, %) (0.98) (1.00) (1.74) (1.33) (2.13) (1.39) (1.72)
aStudy cut-off value for Magstream is designed to equal the same number of false-positive results as Hemoccult II, as derived from prior studies.
Study cut-off value for OC Sensor is designed to ensure comparability with other studies, while limiting the colonoscopy workload.

were very similar. An increase in TPR was possible by suggest that the OC Sensor outperforms Magstream for
using 2 samples, at the expense of increased FPR. both the detection of invasive cancers alone or ad-
Results of the sensitivity analysis (available on request) vanced neoplasias as a whole; the 2 tests provide a
did not change the findings. substantial improvement in cancer screening in com-
parison with Hemoccult II. Strikingly, for the same
Discussion number of positive results, 1-sample OC Sensor nearly
Our results, directly comparing the performances doubled the number of advanced neoplasias detected
of the 2 FITs in an average-risk population, strongly compared with the 3-sample guaiac test. For the same

Table 3. Comparison of Clinical Performances of Magstream and OC Sensor FITs for 2 Positivity Rates
1.6% Positivity rate (same as Hemoccult II) 3% Positivity rate

Magstream OC Sensor Magstream OC Sensor


FIT FIT FIT FIT
Guaiac
test 1 sample 2 samples 1 sample 2 samples 2 samples 2 samples
Cut-off value (␮g hemoglobin/g of stools) NA 206 234 68 116 187 48
Cut-off value (ng hemoglobin/mL in the buffer) NA 63 72 340 580 57 240
Positives 316 312 330 316 317 599 595
(positivity rate, %) (1.60) (1.58) (1.67) (1.60) (1.60) (3.03) (3.01)
Colonoscopies 277 273 295 282 285 526 523
(colonoscopy rate, %) (87.7) (87.5) (89.4) (89.2) (89.9) (87.8) (87.9)
Advanced neoplasias 83 119 123 154 160 203 241
(positive predictive value, %) (30.0) (43.6) (41.7) (54.6) (56.1) (38.6) (46.1)
(true positive rate, %) (0.42) (0.60) (0.62) (0.78) (0.81) (1.03) (1.22)
Number needed to scope 3.3 2.3 2.4 1.8 1.8 2.6 2.2
Number needed to screen 238.5 166.4 161.0 128.6 123.7 97.5 82.1
Invasive cancers 19 24 20 31 33 34 37
(positive predictive value, %) (6.9) (8.8) (6.8) (11.0) (11.6) (6.5) (7.1)
(true-positive rate, %) (0.10) (0.12) (0.10) (0.16) (0.17) (0.17) (0.19)
Number needed to scope 14.6 11.4 14.8 9.1 8.6 15.5 14.1
Number needed to screen 1041.9 824.9 989.9 638.6 599.9 582.3 535.1
False positives 194 154 172 128 125 323 282
(false-positive rate, %) (0.98) (0.78) (0.87) (0.65) (0.63) (1.63) (1.42)
May 2013 COMPARISON OF PERFORMANCES OF 2 FITS 923

similar imputation for 2 samples of FIT at the manufac-


turer cut-off value.
In comparison with Hemoccult II, Magstream FIT
appeared less specific than expected based on our pre-
vious study.9 This could be explained by changes in
automat (Magstream HT instead of Magstream 1000)
and buffer (Magstream NewHemtube instead of Mag-
stream Hemtube) for the Magstream test, or an appar-

CLINICAL AT
ent increase in the specificity of Hemoccult II owing to
dietary restrictions (although debated25) in Allier. In
addition, the present study included subjects with pre-
vious rounds of Hemoccult II, which may have an
impact on its specificity.
The superiority of OC Sensor over Magstream con-
firms previous evidence based on an indirect compari-
son relying on a literature review,15 and laboratory
Figure 2. ROC curves for Hemoccult II, Magstream, and OC Sensor
tests, according to the number of samples of immunochemical tests. analyses.17,18 However, this study suggests the superi-
ority of the OC Sensor test, directly comparing the 2
tests in a real-life scenario. The joint increase in sensi-
tivity and the decrease in specificity of Magstream for
number of unnecessary colonoscopies (false-positive re- the detection of advanced neoplasias compared with
sults), the OC Sensor detected more advanced neopla- OC Sensor at the same cut-off value expressed in con-
sias than Magstream. centration of hemoglobin in the stools may be ex-
Our study had several limitations. Only subjects posi- plained by the poorer precision of the Magstream mea-
tive for at least one of the tests were explored by colono- surement because a much higher overlap in crude
scopy, preventing an estimation of sensitivity and speci- measurements between different concentrations of he-
ficity of each test. Although a minority of studies moglobin in the feces had been observed for Magstream
evaluating FOBT implied an endoscopy (colonoscopy or than OC Sensor in our in-laboratory analysis.17 Fur-
sigmoidoscopy) for all subjects,22,23 this limitation is thermore, OC Sensor is based on latex agglutination
shared by most of the studies comparing the performance immunoturbidimetry, whereas Magstream is based on
of colorectal cancer screening programs, inasmuch as it is magnetic particle agglutination. Our results confirm
evaluated in the targeted population.7–9,14 Moreover, in those of other studies that have shown that qualitative
our study, plotting TPR against FPR provided a compre- FITs using the same cut-off value expressed in concen-
hensive version of the ROC curve, with direct and practi- tration of hemoglobin in the stools may be associated
cal interpretation. Indeed, TPR in participants is the prod- with different clinical characteristics (adapted from
uct of prevalence of the lesion, the sensitivity of the test, Brenner et al26), supporting the need for clinical trials,
and the colonoscopy rate of positive subjects, and FPR is at least in high-risk subjects, to evaluate future new
the product of 1 minus the prevalence, 1 minus the FITs.
specificity, and colonoscopy rate. The paired design guar- The superiority of OC Sensor compared with Mag-
antees that the prevalence is identical whatever the tests, stream concerned both the detection of invasive cancers
whereas the delivery of positive screening results without and high-risk adenomas, hence offering the opportunity
specification on which FOBT(s) was(ere) positive guaran- to increase both the reduction in colorectal cancer–spe-
tees that the colonoscopy rate does not depend on the cific mortality and incidence, although not shown. In
test. Thus, comparing TPR or FPR between tests is equiv- addition, the rate of nonanalyzable tests was higher with
alent to comparing sensitivity and specificity as in relative Magstream compared with OC Sensor test. This was ex-
ROC curves.9,24 plained both by failure to correctly paste the identifying
As in other studies, we evaluated indicators such as barcode due to the rounded form of the collecting tube, and
positivity rate, detection rate among participants, and introduction of too much stool in the tube. These difficulties
positive predictive values, which reflect the performance are consistent with previous findings.18 In our study, how-
of the tests. To allow for standardization of our results, we ever, and because Hemoccult II is the routine test in France,
transformed the crude value given by the FIT automats nonanalyzable FITs were not repeated, hence overestimating
into concentration of hemoglobin in the stools, but our the rate of nonanalyzable tests.
conversion equation for Magstream was not perfect.17 As suggested by other studies,9,13,14 for moderate posi-
Finally, our evaluation of the performances of 1-sample tivity rates we found that both Magstream and OC Sensor
FIT at the manufacturer’s cut-off value implied the im- FITs led to identical performances with either 1 of 2
putation of only 13% of the total number of expected samples (provided a different cut-off value is used),
colonoscopies, resulting in reliable results. Too many un- reflecting equivalence at the population level, rather
available colonoscopies prevented us from performing a than at the level of the subject.10 Differences in bleed-
924 RAGINEL ET AL GASTROENTEROLOGY Vol. 144, No. 5

ing pattern between lesions (intensity and intermit- average-risk population according to positivity threshold and
tence) underlies this finding. When limitation of the number of samples. Int J Cancer 2009;125:1127–1133.
10. Guittet L, Bouvier V, Guillaume E, et al. Colorectal cancer screen-
colonoscopy workload is targeted, we would recom-
ing: why immunochemical faecal occult blood test performs as
mend the use of one sample of FIT. Alternatively, when well with either one or two samples. Dig Liv Disease 2012;
high positivity rates are affordable for the health care 44:694 – 699.
system, using 2 samples of FIT may increase the detec- 11. van Rossum LG, van Rijn AF, Laheij RJ, et al. Cutoff determines
tion rate of advanced neoplasias at an acceptable num- the performance of a semi-quantitative immunochemical faecal
ber needed to scope. occult blood test in a colorectal cancer screening programme. Br J
Cancer 2009;101:1274 –1281.
Concerning the performance of FITs at moderate pos-
CLINICAL AT

12. Hol L, Wilschut JA, van Ballegooijen M, et al. Screening for colo-
itivity rates, our results are clearly in favor of 1-sample OC rectal cancer: random comparison of guaiac and immunochemical
Sensor. Selection of the cut-off value then could be ad- faecal occult blood testing at different cut-off levels. Br J Cancer
justed by politic decision makers to fit the possibility of 2009;100:1103–1110.
endoscopy. The data presented in this article provide 13. Park DI, Ryu S, Kim YH, et al. Comparison of guaiac-based and
results for up to a 2.8% positivity rate for 1-sample OC quantitative immunochemical fecal occult blood testing in a pop-
Sensor (150 ng/mL in the buffer cut-off value which is ulation at average risk undergoing colorectal cancer screening.
Am J Gastroenterol 2010;105:2017–2025.
equivalent to 30 ␮g/g hemoglobin in the stools), whereas
14. Faivre J, Dancourt V, Manfredi S, et al. Positivity rates and perfor-
some studies provide results for lower cut-off values.11–13,20 mances of immunochemical faecal occult blood tests at different
However, although OC Sensor is more accurate than Mag- cut-off levels within a colorectal cancer screening programme. Dig
stream, it is also actually more expensive. Even if the costs of Liver Dis 2012;44:700 –704.
colorectal cancer screening programs are determined mainly 15. Guittet L, Bailly L, Bouvier V, et al. Indirect comparison of perfor-
by the number of generated colonoscopies,27 for a fixed mances of two quantitative immunochemical faecal occult blood
tests in a population with average colorectal cancer risk. J Med
positivity rate, OC Sensor would be more effective (higher
Screen 2011;18:76 – 81.
sensitivity), less deleterious (higher specificity), but also more 16. Allison JE, Fraser CG, Halloran SP, et al. Comparing fecal immu-
costly. Cost-effectiveness analyses are required to guide fur- nochemical testing: improved standardization is needed. Gastro-
ther decisions. enterology 2012;142:422– 424.
17. Guittet L, Guillaume E, Levillain R, et al. Analytical comparison of
Supplementary Material three quantitative immunochemical fecal occult blood tests for
colorectal cancer screening. Cancer Epidemiol Biomarkers Prev
Note: To access the supplementary material 2011;20:1492–1501.
accompanying this article, visit the online version of 18. Lamph SA, Bennitt WE, Brannon CR, et al. Evaluation report:
Gastroenterology at www.gastrojournal.org, and at http:// immunochemical faecal occult blood tests. Centre for Evidence-
based Purchasing 2009, Report no. 09042.
dx.doi.org/10.1053/j.gastro.2013.01.042.
19. Castiglione G, Visioli CB, Ciatto S, et al. Sensitivity of latex agglu-
tination faecal occult blood test in the Florence District population-
References based colorectal cancer screening programme. Br J Cancer 2007;
1. Parkin DM, Bray F, Ferlay L, et al. Global cancer statistics, 2002. 96:1750 –1754.
CA Cancer J Clin 2005;55:74 –108. 20. Grazzini G, Visioli CB, Zorzi M, et al. Immunochemical faecal
2. Hewitson P, Watson Glasziou P, et al. Cochrane systematic review of occult blood test: number of samples and positivity cutoff. What is
colorectal cancer screening using the fecal occult blood test (hemoc- the best strategy for colorectal cancer screening? Br J Cancer
cult): an update. Am J Gastroenterol 2008;103:1541–1549. 2009;100:259 –265.
3. Atkin WS, Edwards R, Kralj-Hans I, et al. Once-only flexible sig- 21. R Development Core Team (2010). R: a language and environ-
moidoscopy screening in prevention of colorectal cancer: a multi- ment for statistical computing. Vienna, Austria: R Foundation for
centre randomised controlled trial. Lancet 2010;375:1624 – Statistical Computing. Available at: http://www.R-project.org/.
1633. 22. Allison JE, Sakoda LC, Levin TR, et al. Screening for colorectal
4. Segnan N, Armaroli P, Bonelli L, et al. Once-only sigmoidoscopy in neoplasms with new fecal occult blood tests: update on per-
colorectal cancer screening: follow-up findings of the Italian Ran- formance characteristics. J Natl Cancer Inst 2007;99:1462–
domized Controlled Trial—SCORE. J Natl Cancer Inst 2011;103: 1470.
1310 –1322.
23. Morikawa T, Kato J, Yamaji Y, et al. A comparison of the
5. Schoen RE, Pinsky PF, Weissfeld JL, et al. PLCO Project Team.
immunochemical fecal occult blood test and total colonoscopy
Colorectal-cancer incidence and mortality with screening flexible
in the asymptomatic population. Gastroenterology 2005;129:
sigmoidoscopy. N Engl J Med 2012;366:2345–2357.
422– 428.
6. Allison JE, Tekawa IS, Ransom LJ, et al. A comparison of fecal
24. Cheng H, Macaluso M. Comparison of the accuracy of two tests
occult-blood tests for colorectal-cancer screening. N Engl J Med
1996;334:155–159. with a confirmatory procedure limited to positive results. Epidemi-
7. van Rossum LG, van Rijn AF, Laheij RJ, et al. Random comparison ology 1997;8:104 –106.
of guaiac and immunochemical fecal occult blood tests for colo- 25. Konrad G. Dietary interventions for fecal occult blood test screen-
rectal cancer in a screening population. Gastroenterology 2008; ing: systematic review of the literature. Can Fam Physician 2010;
135:82–90. 56:229 –238.
8. Hol L, van Leerdam ME, van Ballegooijen M, et al. Screening for 26. Brenner H, Haug G, Hundt S. Inter-test agreement and quantitative
colorectal cancer: randomised trial comparing guaiac-based and cross-validation of immunochromatographical fecal occult blood
immunochemical faecal occult blood testing and flexible sigmoid- tests. Int J Cancer 2010;127:1643–1649.
oscopy. Gut 2010;59:62– 68. 27. Berchi C, Bouvier V, Réaud JM, et al. Cost-effectiveness analysis
9. Guittet L, Bouvier V, Mariotte N, et al. Performance of immuno- of two strategies for mass screening for colorectal cancer in
chemical faecal occult blood test in colorectal cancer screening in France. Health Econ 2004;13:227–238.
May 2013 COMPARISON OF PERFORMANCES OF 2 FITS 925

laboratory, and to the research team Cancers and Preventions who


Received August 30, 2012. Accepted January 22, 2013.
were involved in the study.
Reprint requests
Conflicts of interest
Address requests for reprints to: Lydia Guittet, MD, PhD, Inserm
The authors disclose no conflicts.
U1086, Cancers and Preventions, CHU Caen, Centre François
Baclesse, Avenue du Général Harris, 14000 Caen, France. e-mail:
Funding
guittet-l@chu-caen.fr; fax: 33 (0)2 31 45 86 30.
The study was funded by the French National Institute for
Acknowledgments Cancer (Institut National du Cancer), and the French National
The authors are grateful to all subjects who agreed to join the League against Cancer (Ligue Nationale Contre le Cancer); the

CLINICAL AT
study, to the general practitioners who included the patients, to the automat analyzers were provided on loan by Eiken (OC Sensor
endoscopists who performed the colonoscopies, to the members of Diana automat) and Fujirebio (Magstream HT automat)
the 2 regional associations in charge of colorectal cancer screening, manufacturers. The sponsors and manufacturers had no role in
ADOC18 and ABIDEC, to the Institut InterRégional pour la Santé the study design or data analysis.

You might also like