You are on page 1of 9

Instruction Manual

For All Corning® Hot Plates, Stirrers,


Stirrer/Hot Plates with Digital Displays, and
External Temperature Controller (6795PR)

Corning Cat. No.


230V 230V
Model No. Product Top Plate Size 120V 100V (Non-Europe) (Europe)
PC-400D Hot Plate 5 x 7 in. 6795-400D 6797-400D 6798-400D 6796-400D
(12.7 x 17.8 cm)
PC-410D Stirrer 5 x 7 in. 6795-410D 6797-410D 6798-410D 6796-410D
(12.7 x 17.8 cm)
PC-420D Stirrer/ 5 x 7 in. 6795-420D 6797-420D 6798-420D 6796-420D
Hot Plate (12.7 x 17.8 cm)
PC-600D Hot Plate 10 x 10 in. 6795-600D 6797-600D 6798-600D 6796-600D
(25.4 x 25.4 cm)
PC-610D Stirrer 10 x 10 in. 6795-610D 6797-610D 6798-610D 6796-610D
(25.4 x 25.4 cm)
PC-620D Stirrer/ 10 x 10 in. 6795-620D 6797-620D 6798-620D 6796-620D
Hot Plate (25.4 x 25.4 cm)
6795PR External All Models 6795PR
Temperature
Controller

1.800.561.8187 www. .com information@itm.com


About This Manual Table of Contents
This manual is designed to assist you in the optimal usage of your Corning®
hot plate, stirrer, stirrer/hot plate, or external temperature controller. 1.0 Safety Information. . . . . . . . . . . . . . . . . . . . . . . 2
The manual contains English, German, French, Italian, Spanish, and Polish
versions and are also available at 2.0 Operating Conditions. . . . . . . . . . . . . . . . . . . . 3

Product Voltages 3.0 Product Controls and Indicators . . . . . . . . . . 4


Hot plates, stirrers, and stirrer/hot plates are available in different voltages.
Before ­initial use, check that the unit you received is the correct voltage for 4.0 Product Connections. . . . . . . . . . . . . . . . . . . . . 4
your location.
5.0 Stirring Instructions. . . . . . . . . . . . . . . . . . . . . . 5
Warranty Registration
Register your product warranty online at 6.0 Heating Instructions. . . . . . . . . . . . . . . . . . . . . 6

7.0 Heating Operation . . . . . . . . . . . . . . . . . . . . . . 7


7.1 Principles . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
7.2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

8.0 Repair. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

9.0 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

10.0 Optional Accessories. . . . . . . . . . . . . . . . . . . . 11

11.0 Replacement Parts. . . . . . . . . . . . . . . . . . . . . . 12

12.0 Technical Specifications . . . . . . . . . . . . . . . . 12

13.0 Size and Dimensions. . . . . . . . . . . . . . . . . . . . 13

14.0 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . 13

15.0 Product Disposal. . . . . . . . . . . . . . . . . . . . . . . . 14

16.0 Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

1.800.561.8187 www. .com information@itm.com


1.0 Safety Information Product Damage
This instruction manual contains important operating and maintenance w Keep the product dry and clean.
instructions that must be read, understood, and followed by the product w Do not immerse the product for cleaning.
user. Failure to use this product according to this instruction manual may w The ceramic top may break if impacted.
degrade or defeat the protection normally provided by this product. Read
w The maximum gross weight placed on the top surface must not exceed
this instruction manual prior to product use and keep this instruction
25 lbs. (11 kg).
­manual for future reference.
w These units are not explosion- or spark-proof.
Product Symbols
w Do not heat or stir volatile or flammable materials.
Caution – Risk of Danger: Cautions there is material in the w Do not operate this product near volatile or flammable materials.
instruction manual which must be read, understood, and
w Do not use this product with a metal vessel.
followed in order to preserve product safety features.

2.0 Operating Conditions


Caution – Hot Surface: Cautions that the top plate is too Corning hot plates, stirrers, and stirrer/hot plates are designed to p
­ rovide
hot to touch. safe functioning under the following conditions:
w Indoor use
w Altitude up to 6,500 feet (2,000 meters)
Indicates that the unit is plugged into power supply.
w Ambient temperatures of 0°C to 30°C
w Product should be placed on a flat surface at least 12 in. (30.5 cm) from
Indicates that the accessory external temperature walls, 48 in. (122 cm) from ceilings, and 12 in. (30.5 cm) from other hot
­controller is properly plugged into the unit. plates if using multiple units.
w Maximum relative humidity of 80% for temperatures up to 31°C, decreasing
linearly to 50% relative humidity at 40°C.
Warnings
w Pollution Degree 2: Any foreign matter that may accumulate on or within
Personal Injury the product during normal use is not electrically conductive.
w Do not use this product in a manner other than as stated in the Operating w Installation Category II: Product is designed for connection to an e
­ lectrical
Conditions section as the protection provided by the ­equipment may be branch circuit inside a building with main supply voltage fluctuations not
impaired. exceeding ±10% of the nominal voltage.
w This product is designed for use in laboratory environments by persons
knowledgeable in safe laboratory practices.
w Always wear safety glasses and other appropriate protective equipment
when operating this product.
Electric Shock
w This product must be connected to a grounded power outlet for safe
functioning.
w Use only the power cord supplied with the product.
w The power cord is the device available for full disconnect from mains input.
w Position the product for use so the power cord can be easily disconnected
without having to move the product.
w Disconnect the power cord before moving or cleaning the unit.

2 3

1.800.561.8187 www. .com information@itm.com


3.0 Product Controls and Indicators To Connect the External Temperature Controller
1. Turn the Stir Control Knob and Heat Control Knob to the OFF p
­ osition.
1 2. Disconnect power cord.
7 6 3. Insert temperature controller connector into the input connector.
4. Reconnect power cord.
3 5 5. Product is now ready for use with the External Temperature Controller.
Repeat the above process when disconnecting the Temperature Controller.
2 4

5.0 Stirring Instructions


for PC-410D, PC-420D, PC-610D, PC-620D Models
1. Fill the vessel with solution to be stirred.
1. Power Indicator: Illuminates at all times when the product is ­properly
connected to input power. 2. Place the stir bar into the vessel.
2. Stir Control Knob: Turn the knob all the way counterclockwise to turn off 3. Place the vessel in the center of the top surface.
­stirring function. Turn the knob clockwise to set desired stirring speed. 4. Turn Stir Control Knob until the Stirring Speed Display shows the
3. Stirring Speed Display: Shows the speed set for stirring desired speed. The speed setting can be adjusted according to the table
below.
4. Heat Control Knob: Turn the knob all the way counterclockwise to turn
off heating function. Turn the knob clockwise to set the desired heating Operating Range (rpm) Adjustable Increment (rpm)
temperature. 60 - 100 5
5. Heating Temperature Display: Shows the temperature set for heating. 100 - 200 10
6. Hot Top Indicator: Illuminates when the temperature of the top is too 200 - 400 20
hot to touch (greater than ~60°C). 400 - 1,150 50
7. Temperature Probe In-Use Indicator: Illuminates when the external
w Flashing Display: The number will FLASH when the actual stirring
­temperature probe is connected to the unit.
speed is not at set speed.
w Constant Display: The number will not flash when the actual stirring
4.0 Product Connections
speed is at the set speed. The number will remain constantly ON when
1. Power Cord Input: Connect the supplied power cord into this ­connector. the actual stirring speed is at the set speed.
2. External Temperature Controller Input: Connect the optional External 5. When desired stirring is complete, turn the Stir Control Knob to the OFF
Temperature Controller (Corning Cat. No. 6795PR) into this connector. position. Allow stir bar to cease rotation before removing the vessel from
the unit.
6. Heating and stirring can be performed simultaneously with the PC-420D
and PC-620D models, which offer both heating and stirring functions.
NOTE: The viscosity of the stirred material will affect the ability of the
stir bar to remain coupled to the rotating ring magnet. Materials of high
viscosity must be stirred at slower speed settings.
2 1

4 5

1.800.561.8187 www. .com information@itm.com


6.0 Heating Instructions 5. Insert the tip of the External Temperature Probe into the solution.
w The
tip should be located in the center of the vessel and at
Without External Temperature Controller for PC-400D, PC-420D,
approximately one-half of the depth of the solution.
PC-600D, and PC-620D Models
6. Secure the position of the External Temperature Controller by using a
1. Fill the vessel with solution to be heated. ring stand/support rod and clamp.
2. If using a PC-420D or PC-620D model and the stirring function, place the w Ensure
the cable of the External Temperature Controller does not come
stir bar into the vessel. into contact with the heating surface.
3. Place vessel in the center of the top surface. 7. Turn the Heat Control Knob until the Heating Temperature Display
4. Turn Heat Control Knob until the Heating Temperature Display shows shows the desired heating temperature.
the desired temperature. The temperature setting can be adjusted in 5°C w Flashing
Display: The number shown on the Heating Temperature
increments. Display will FLASH when the actual heating temperature is not at the
w Flashing
Display: The number shown on the Heating Temperature set temperature.
Display will FLASH when the actual heating t­ emperature is not at the w Constant Display: The number shown on the Heating Temperature
set temperature. Display will remain constantly ON when the actual heating temperature
w Constant Display: The number shown on the Heating Temperature is at the set temperature.
Display will remain constantly ON when the actual heating temperature w Hot Top Indicator: The Hot Top Indicator will be ON at all times when
is at the set ­temperature. the temperature of the top surface is too hot to touch (greater than
w Hot
Top Indicator: The Hot Top Indicator will be ON at all times when ~60°C).
the temperature of the top surface is too hot to touch (greater than w The Hot Top Indicator will FLASH when the Heat Control Knob is
~60°C). turned OFF but the top surface is still too hot to touch.
w The
Hot Top Indicator will FLASH when the Heat Control Knob is w TheHot Top Indicator will be OFF when the temperature of the top is
turned OFF but the top surface is still too hot to touch. less than ~60°C.
w TheHot Top Indicator will be OFF when the temperature of the top is Caution: The Hot Top Indicator will turn OFF when the power cord is dis-
less than ~60°C. connected from the product even if the temperature of the top s­ urface is
Caution: The Hot Top Indicator will turn OFF when the power cord is still too hot to touch.
­disconnected from the product even if the temperature of the top surface
is still too hot to touch.
7.0 Heating Operation
With External Temperature Controller, Corning Cat. No. 6795PR 7.1 Principles
for PC-400D, PC-420D, PC-600D, PC-620D, and 6795PR Models
The heating element and a temperature sensor are located just beneath
1. Connect the External Temperature Controller to the connector on the the ceramic top surface of the product. The microprocessor-controlled heat,
back of the unit. generated by the heating element, is based upon the sensor temperature
w Temperature
Probe in Use Indicator: This will illuminate when the and the value set on the Heating Temperature Display. When the sensor
External Temperature Controller is properly connected. temperature is not within range of the value set on the display, the display
2. Fill the vessel with solution to be heated. will FLASH. When the sensor temperature is within range, the value displayed
will remain constantly ON.
3. If using a PC-420D or PC-620D model and the stirring function, place the
stir bar into the vessel. The Heating Temperature Display does not indicate the actual temperature
4. Place the vessel in the center of the top surface. of materials placed on top of the product or the actual temperature of the
ceramic top surface.

6 7

1.800.561.8187 www. .com information@itm.com


The following table shows a typical difference between the temperature 7.2 Safety
set on the Heating Temperature Display and the actual temperature The microprocessor-controlled system has built-in error routines to detect
­measurement of the ceramic top surface. product operation in some unfavorable conditions. The type of error rou-
Actual Actual tines which may be active in a product varies depending upon the function-
Temp. Set Top Surface Temp. Set Top Surface ality of the model and the operating software version.
on Display (°C) Temp. (°C) on Display (°C) Temp. (°C)
50 45 350 285 If an error routine is engaged, the product will typically shut down. Some
error routines will display an error code number in the left digital location
100 85 400 325
of the Heating Temperature Display when the product is shut down.
150 125 450 365
200 165 500 405
250 205 550 440
A product shut down by an error routine must be disconnected and recon-
300 245
nected to mains input before it can be used again.
(This information was taken using a 6795-420D with no top load in 20°C
ambient conditions).
Using the External Temperature Controller (Corning Cat. No. 6795PR) enables
Caution: Error routines built into these products do not substitute the need
precision temperature control of materials placed in vessels on top of
to use these products per the specified operating conditions and according
the product.
to safe laboratory practices. There are potential failure modes in product
When the External Temperature Controller is connected to the product, the functionality or in the process of use that could result in uncontrolled or
closed loop control process is extended to include temperature input from unexpected heating of the top surface. Reaction plans should be developed
the External Temperature Controller. The microprocessor-controlled heat, and safety precautions put in place based on the worse case scenario that
generated by the heating element, is based upon the sensor temperature any materials placed on the top surface could be subjected to a continuous
located in the tip of the External Temperature Controller and the value set supply of heat, raising the material temperature to levels in excess of 550°C.
on the Heating Temperature Display. When the sensor temperature is not
within range of the value set on the display, the display will FLASH. When
the sensor temperature is within range, the value displayed will remain
8.0 Repair
constantly ON. There are no direct user serviceable components inside this series of
­products. A list of available replacement parts are listed in Section 11.0.
To directly control the temperature of liquid in a vessel on the top surface,
place the External Temperature Controller into the liquid and connect the
Controller to the product. The Heating Temperature Display shows the
temperature of the liquid and can be used to adjust the temperature of
the liquid. 9.0 Maintenance
See Section 10.0 for ordering information.
Power
Caution: The top surface and the vessel used may be at substantially higher Caution: Disconnect power to the product by unplugging the power cord
temperatures than indicated by the setting on the Heating Temp­erature before performing any maintenance or inspection procedures.
Display as the controller regulates the liquid temperature inside the vessel.
w Inspect the power cord regularly and replace if damaged. Use only
replacement power cords available from Corning or Corning a­ uthorized
product distributors.

8 9

1.800.561.8187 www. .com information@itm.com


Ceramic Top Plate
w These products are supplied with a Pyroceram® top that is easy to clean
and highly resistant to scratches, corrosion, and chemical attack.
w The ceramic top may break during use if not properly maintained.
w Keep the ceramic top clean. A nonabrasive cleaner may be used to clean
the ceramic top.
w Inspect the ceramic top for damage during cleaning.
w Discontinue product use if the ceramic top is chipped, etched, or shows
excessive scratching. A replacement top can be ordered (Section 11.0).

­

General
w It is important to keep this product dry and clean.
w Remove minor exterior liquid spills promptly.
w Clean exterior surfaces with a nonabrasive cleaner. Do not reconnect
product to power input until all cleaned surfaces have dried.
w If liquid or wet solid material gets inside the product, immediately
­disconnect the power to the product and discontinue use.
Accessories (Shown with Corning Cat. No. 6795-420D)

10.0 Optional Accessories


Corning Cat. No. Description
6795PR External Temperature Controller for digital display Hot Plates and
Hot Plate/Stirrers. Use with all PC-400D, 420D, 600D, and 620D models.
FEP-coated stainless steel.
440129 Vertical Support Rod. Supplied as two 9 in. (22.86 cm) rods which can be
screwed together. Use with all PC-XXXD series products. Stainless steel.
440140 Boss Head Clamp. Use for connecting the 440141 Holding Rod at a 90° angle
to the 440129 Support Rod. Aluminum.
440141 Holding Rod. Use for holding the 6795PR External Temperature Controller in
position. Aluminum.
6795KIT Universal Accessories Kit includes: 6795PR, 6970SR, and 440129
6795-420KIT Kit includes: 6795-420D Stirrer/Hot Plate, 6795PR, 6970SR, and 440129
6795-620KIT Kit includes: 6795-620D Stirrer/Hot Plate, 6795PR, 6970SR, and 440129
400430 PTFE-coated magnetic stir bar, .39 x 2 in. (1 x 5.1 cm) – recommended size
for all PC-610D and 620D models
401435 PTFE-coated magnetic stir bar, .39 x 1 in. (1 x 2.5 cm) – recommended size
for all PC-410D and 420D models
6970SR Stir bar retriever, polypropylene

10 11

1.800.561.8187 www. .com information@itm.com


11.0 Replacement Parts 13.0 Size and Dimensions
Corning Cat. No. Description Model Usage Model No. Top Plate Size Product Dimensions
411007 Power cord PC-400D/410D/420D, 120V PC-400D/410D/420D 5 x 7 in. 4.25 x 7.75 x 11 in.
410956 Power cord PC-600D/610D/620D, 120V (12.7 x 17.8 cm) (10.8 x 19.7 x 28 cm)
411021 Power cord PC-400D/410D/420D and PC-600D/610D/620D, 100V PC-600D/610D/620D 10 x 10 in. 4.625 x 11 x 15.375 in.
(25.4 x 25.4 cm) (11.75 x 19.7 x 39.05 cm)
410942 Power cord PC-400D/410D/420D and PC-600D/610D/620D,
230V (UK plug)
440124 Power cord PC-400D/410D/420D and PC-600D/610D/620D, 14.0 Troubleshooting
230V (Euro Plug)
Contact Corning Ceramic top plate/ PC-400D/410D/420D, All Voltages w I have a beaker of water on my hot plate and set the temperature for
Customer Service element assembly, 550°C. Why does the display setting blink and not remain constant?
5 x 7 in.
The display will blink at any time when the temperature sensor is not
Contact Corning Ceramic top plate/ PC-600D/610D/620D, All Voltages within range of the set temperature value. The temperature measured
Customer Service element assembly,
by the sensor is a composite of the temperature of the heating element
10 x 10 in.
located beneath the sensor, the ceramic top above the sensor, and the
Contact Corning PC control board —
very small air space around the sensor. Water requires a ­substantial
Customer Service
amount of heat in order to boil yet remains at a constant temperature
440135 Control knob All models
of 100°C for the duration of the boiling process. Although the heating
element is producing maximum heat at the 550°C setting, the water
consumes this heat so quickly during the boiling process that the heat
12.0 Technical Specifications
is unable to raise the temperature measured by the sensor to within the
Power range of the 550°C set value.
120V/60Hz 100V/60Hz 230V/50Hz 230V/50Hz Temp. Stir Range
Model No. Type (6795 models) (6797 models) (6796 models) (6798 models)1 Range2 (rpm) Weight w How long does it take to bring a beaker of water to a boil?
PC-400D Hot 628W/ 548W/ 628W/ 628W/ 5°C-550°C — 6.0 lbs. Using a 600 mL PYREX® beaker with 400 mL of water at 25°C, it takes
Plate 5.3A 5.5A 2.7A 2.7A (41°F-1,022°F) (2.7 kg) approximately 15 minutes to bring the water to a full, rolling boil.
PC-600D Hot 1043W/ 1043W/ 1043W/ 1043W/ 5°C-550°C — 10.0 lbs.
w Can I use a metal tray on top of my Corning® hot plate?
Plate 8.7A 10.5A 4.5A 4.5A (41°F-1,022°F) (4.5 kg)
PC-410D Stirrer 73W/ 78W/ 73W/ 73W/ — 60-1,150 7.0 lbs.
No. The metal will act as a heat sink, and have a high probability of
0.7A 0.8A 0.3A 0.3A (3.2 kg) creating an abnormal heating condition. If an abnormal condition is
PC-610D Stirrer 73W/ 78W/ 73W/ 73W/ — 60-1,150 11.5 lbs.
detected, the product will shut down. A metal vessel will also scratch the
0.7A 0.8A 0.3A 0.3A (5.2 kg) ceramic top plate.
PC-420D Stirrer/ 698W/ 623W/ 698W/ 698W/ 5°C-550°C 60-1,150 7.0 lbs. w The stir bar keeps decoupling. Why and what can I do to stop this?
Hot Plate 5.9A 6.3A 0.3A 0.3A (41°F-1,022°F) (3.2 kg)
These units are programmed to minimize decoupling. However, liquid
PC-620D Stirrer/ 1113W/ 1113W/ 1113W/ 1113W/ 5°C-550°C 60-1,150 11.5 lbs.
Hot Plate 9.3A 11.2A 4.8A 4.8A (41°F-1,022°F) (5.2 kg)
viscosity, stir bar magnetic strength, vessel used, and speed changes
1
Cat. Nos. beginning with 6798 in 230V are non-European configuration models.
can cause decoupling. High viscosity liquids must be stirred at slower
2
The Temperature Range using the External Temperature Controller is 5°C - 200°C (41°F - 392°F). speed settings. The magnetic strength of stir bars can weaken over time
and may need to be replaced. The vessels used need to have thin, flat
bottoms to insure optimal performance. Rapid decreases in stir speed
can cause decoupling as the magnet slows down quicker than the stir bar
and the liquid.
w What size vessel should I use?
The vessels used on the top of a hot plate must not be larger than the
top plate.

12 13

1.800.561.8187 www. .com information@itm.com


w Corning sells replacement parts. Can I do the repairs myself instead of 16.0 Warranty
sending it in to Corning’s equipment repair department?
Corning Incorporated (Corning) warrants that this product will be free from
When repairs are completed by Corning or a Corning authorized repair defects in material and workmanship for a period of two (2) years from
facility, the performance and safety of the product will be verified before date of purchase. CORNING DISCLAIMS ALL OTHER WARRANTIES WHETHER
being returned to you. We do sell replacement parts so customers can EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTIES OF
complete repairs themselves. It is recommended that only people know­ MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE. Corning’s
ledgeable in electronics complete those repairs. There is no warranty or sole obligation shall be to repair or replace, at its option, any product or
return on equipment replacement parts. part thereof that proves defective in material or workmanship within
the warranty period, provided the purchaser notifies Corning of any such
15.0 Product Disposal defect. Corning is not liable for any incidental or consequential damages,
commercial loss or any other damages from the use of this product.
According to Directive 2012/19/EU of the European Parliament and Council
of 4th July 2012 on waste and electronic equipment (WEEE) as amended, This warranty is valid only if the product is used for its intended purpose
Corning® Hot Plates, Stirrers, and Stirrer/Hot Plates are marked with the and within the guidelines specified in the supplied instruction manual.
crossed-out wheeled bin and must not be disposed of with domestic waste. This warranty does not cover damage caused by accident, neglect, misuse,
improper service, natural forces, or other causes not arising from defects
Consequently, the buyer shall follow the instructions for reuse and recycling
in original material or workmanship. This warranty does not cover motor
of waste electronic and electrical equipment (WEEE) provided with the
brushes, fuses, light bulbs, batteries, or damage to paint or finish. Claims
products and available at the following link:
for transit damage should be filed with the transportation carrier.
In the event this product fails within the specified period of time because
of a defect in material or workmanship, contact Corning Customer Service

Corning’s Customer Service team will help arrange local service where
available, coordinate a return authorization number, or provide a return
authorization number and shipping instructions. Products received without
proper authorization will be returned. All items returned for service should
be sent postage prepaid in the original packaging or other suitable car-
ton, padded to avoid damage. Corning will not be responsible for damage
incurred by improper packaging. Corning may elect for onsite service for
larger equipment.
Some states do not allow limitation on the length of implied warranties
or the exclusion or limitation of incidental or consequential damages. This
warranty gives you specific legal rights. You may have other rights which
vary from state to state.
No individual may accept for, or on behalf of Corning, any other obligation
of liability, or extend the period of this warranty.
For your reference, make a note of the model and serial number, date of
purchase, and supplier here.

Model No.________________________________________________________________

Serial No. _________________________________________________________________

Date Purchased ____________________________________________________________

Supplier __________________________________________________________________

14 15

1.800.561.8187 www. .com information@itm.com