Professional Documents
Culture Documents
A Smart Pill For Drug Delivery With Sensing Capabilities: Member, IEEE Member, IEEE
A Smart Pill For Drug Delivery With Sensing Capabilities: Member, IEEE Member, IEEE
R. Goffredo1* Student Member, IEEE, D.Accoto1Member IEEE, M. Santonico2, G. Pennazza2 and E. Guglielmelli1
Member, IEEE
1362
The impedance meter was firstly tested by measuring a set of value expected from simulation. Gas production in the
known resistances, within the range 100Ω - 10kΩ. The electrolytic reservoir can be estimated using Faraday’s law:
relative estimated error between the measured value and the 𝑚=( ∙ )
𝑄 𝑀
(2)
real one is lower than 6%. The absolute (∆R) and relative 𝐹 𝑧
errors (R%) were calculated according to the following Where m is the mass of gases developed at the electrodes, Q
∆𝑅 is the electrical charge, F is Faraday’s constant (96500 C
equations: ∆𝑅 = 𝑅𝑡 − 𝑅𝑀 and 𝑅% = , where Rt is the 𝑀
𝑅𝑀 mol-1), , is the mass to charge ratio of the electrolyte. For
theoretical resistance and RM is the measured resistance. 𝑧
Error is approximately linear with the measured impedance. the considered reaction, 22 µmol of gases are produced by
Thus, the absolute error can be evaluated from the measured 29 µmol of electrons, corresponding to a total charge
impedance values. To this purpose, a software correction has Q =2.83 C. Since the electrical charge Q is the time integral
𝑡
been introduced in order to increase the accuracy of the of current I (𝑄 = ∫0 𝐼(𝜏)𝑑𝜏), the amount of gases produced
measurement. Using Matlab (R2012a, Mathworks) curve can be evaluated by monitoring the current through the
fitting tools, the error has been linearly fitted to compensate electrodes. To this purpose, an interface was created in
the measurements. The corresponding correction factor has Labview (National Instrument) running on a PC equipped
been implemented in the firmware. The post-acquisition
with a DAQ NI USB-6009 ADC/DAC converter card. An
correction is implemented on small subsets of impedances
analog-to-digital line recorded the current through the pump
from 10Ω to 10kΩ. The error is acceptable (< 6%) over an
extended range of impedances without the need for extra when a constant 3V voltage was applied to the electrodes.
circuitry devoted to the measurements outside the optimal Current flowing through the electrolytic cell is influenced
range (1 kΩ - 10 kΩ) around Zcalib. A second set of measures by the concentration of ions and increases over time (Fig.5).
were conducted using a circuit comprising a variable
capacitance (0.1nF, 10nF, 100nF) and a 2.2 kΩ resistance,
connected in series and in parallel. Real and imaginary parts
have been correctly measured (error < 6%) using the same
correction factor previously defined.
III. MODELING AND TESTING
A. Drug delivery system
The pressure in the electrolytic solution reservoir needed to
deliver the desired drug volume Vd was evaluated in
simulation (Comsol Multiphysics, adopting the large Fig.5 Current at the electrodes vs. time
deformations solver).The parameters used in simulation The amount of gas produced, Vgas, is proportional to the
were those of the fabricated system. Membrane surface: 40 electrical charge Q according to (2) and can be calculated
mm2; thickness: 0.1 mm; material: PDMS (Young’s following logical flow diagram shown in Fig 6.
modulus of 800 kPa and Poisson’s ratio of 0.49). The
membrane is clamped at its edge and uniformly loaded by a
pressure P. Results show that the pressure needed to
displace a volume Vd = 60 μl is P = 30 kPa. The
Fig. 6 Key steps for gas production measurement.
corresponding maximum membrane displacement is 2.5
mm (Fig. 4). The number of moles of electrons produced in 120 sec, ne-,
𝑄
is ne-= = 5.5 μmol, corresponding to 7.3 μmol of gas (164
𝐹
μl of gas in STP conditions).
B. Impedance sensor
The impedance sensor has been preliminarily tested in vitro
measuring the impedance of a wadding stripe soaked with
NaCl solution at different concentrations (20 - 350 mg of
NaCl in 100 ml of water). Two electrodes of a naso-gastric
probe (Sandhill, mod. ZANBG-44), used in medical
practice for esophageal impedance/pH monitoring, have
Fig. 4 Results of the simulation of the membrane deformation been connected to the sensor and inserted in the cotton
By approximating the mixture of O2 and H2 to an ideal gas, support. Figure 7 shows the sensor response during in vitro
it is straightforward to calculate the amount of gas that at test. The signal is proportional to the tissue impedance for
the pressure P produces the volume displacement Vd. Such concentrations between 0.2% P/V and 0.02% P/V. Out of
quantity is 22 µmol, corresponding to 502 µl of gases in that range, the signal reaches the saturation value.
STP conditions. Then the sensor was tested in vivo on a GERD patient, with
Then, gas production was measured on the system the approval of the University Ethical Committee.
developed to compare its performance with the maximum
1363
Fig. 9 The impedancemeter tested (left); large size PCB (middle),
miniaturized electronics (right)
power consumption (electrolytic pump) and
biocompatibility, assured by a proper selection of materials
Fig. 7 Sensor response to different NaCl concentrations (e.g. PDMS membrane in contact with the drug; saline
The diagnosis on the patient had been previously made with electrolytic solution). The drug delivery system has been
traditional ambulatory naso-gastric probe. Compliance with experimentally validated in lab. It can be activated using
the standard IEC60601-1 (0.1mA in normal conditions, information about the electrical properties of the tissue in
0.5mA in single fault condition) was verified before in vivo contact with the pill. The impedance sensor is capable of
testing. At the end of the medical exam, the probe was monitoring in body conditions as it has been calibrated on
connected to the developed electronics and, as for in vitro known impedance values of the tissue.
test, a couple of electrodes used to read the impedance. The Furthermore, miniaturization of the sensing system makes it
electrodes impedance is normally negligible but the suitable for integration on-board the smart pill. The capsule
interruption of contact between the electrodes and the tissue for local drug delivery triggered by impedance measurement
causes sharp increases in the value of impedance. During the is a novel tool for the treatment of local active pathologies.
test impedance values between 600-700 Ω were measured,
as shown in Fig. 8 (mean value: 645Ω; standard deviation: ACKNOWLEDGEMENT
31.8 Ω, 4.9% of the mean value, data acquisition every 500 The research was carried out in collaboration with the
ms). Measured impedance was in agreement with literature Policlinico Universitario Campus Bio-Medico di Roma.
data related to GERD patients [11]. Moreover the developed
sensor provided measurements in good agreement with REFERENCES
ambulatory instrumentation. The possibility of measuring a
[1] SB Tiwari, AR Rajabi-Siahboomi “Extended-release oral drug delivery
wider range of impedance values makes the sensor a good technologies: monolithic matrix systems”, Drug delivery systems, 2008
Springer.
[2] J. L. Gonzalez-Guillaumin, D.C. Sadowski, K. V. I. S. Kaler
“Ingestible Capsule for Impedance and pH monitoring in the esophagus” ,
IEEE Trans. on Biomedical Engineering Vol. 54,2007
[3] J. F. Colombel, et all. “Infliximab, Azathioprine, or Combination
Therapy for Crohn’s Disease”, The New England Journal of Medicine,
April 15, 2010
A
[4] J. Marcial Portilla, Unyoung Kim “ Low-power, Self-contained
reciprocating micropump through electrolysis and catalyst-driven
B recombination toward drug delivery applications” MEMS 2013, Taipei,
Taiwan, January 20-24 2013.
[5] C. G. Cameron, M S. Freund “Electrolytic actuators: Alternative, high
perfom-, ance material-based devices”, PNAS, 2002 vol. 99 7827–7831
[6] D. Accoto, V. Mattioli, P.Valdastri, A. Menciassi, P. Dario “A
Fig. 8 Oesophageal impedance profile on a period of 93 s.
miniaturized drug-delivery system for intra-corporal use” Sensors and
A: 645Ω, impedance mean value measured with the proposed system.
Microsystems: Proc. of the 8th Italian Conference, Trento, Italy, 12-14
B: 600 Ω, impedance mean value measured with ambulatory
February 2003
instrumentation (Sleuth, Sandhill).
[7] P. Bogonez-Franco, R. Bragos, A. Bayes-Genis , J. Rosell –Ferrer
candidate for impedance measurement along the whole “Implantable Bioimpedance Monitoring using ZigBee”,31st Annual
gastrointestinal tract. The sensor tested on large scale, was International Conference IEEE/EMBS, 2009
miniaturized by scaling the dimensions, for further [8] Pei-Ju Chao, Eng-Yen Huang, Kuo-Sheng Cheng, Yu-Jie Huangl
integration on the capsule. Electrodes are mounted on the “Electrical Impedance Spectroscopy as Electrical Biopsy for Monitoring
capsule surface (fig.1). The microcontroller chosen is Radiation Sequelae of Intestine in Rats”, BioMed Research International,
August 2013.
PIC12f1822; the on-board battery is activated by a magnetic [9] D. Ummarino, S.Salvatore, B. Hauser, A. Staiano, Y.Vandenplas,
reed. Data read during navigation are stored in an EPROM “Esophageal impedance baseline according to different time intervals”.
memory. Figure 9 shows the device (left), large size PCB European J. of Medical Research 2012.
(middle) used for in vitro and in vivo testing and its [10] B. F. Kessing, A. J. Bredenoord, P. W. Weijenborg, G.J.M.
miniaturized version (right). Hemmink, C. M. Loots, and A.J.P.M. Smout, “Esophageal acid exposure
decreases intraluminal baseline impedance levels”,American Journal of
Gastroenterology, 2011.
IV. DISCUSSION AND CONCLUSIONS [11] L. Majer, V. Stopjakovà, E. Vavrinsky, “Wireless measurement
system for non-invasive biomedical monitoring of psychophysiological
In this work a novel drug delivery system has been proposed. process” Journal of electrical engineering, VOL. 60, NO. 2, 2009,57-68
The designed system can be integrated on a smart pill for [12] MSPS,12-Bit Impedance Converter, Network Analyzer AD5933,
oral administration thanks to its small dimensions, low Analog Devices Datasheet.
1364