Professional Documents
Culture Documents
1 INTRODUCTION
The transmitter PCB includes circuitry for measuring the signals from the NDIR gas sensor,
converting and linearising into CO2 concentration and supplying an output as either 4-20mA or via
USB to an Excel spreadsheet.
Alphasense 4-20mA digital NDIR transmitters offer convenience and easy maintenance:
• Transmitters are shipped pre-calibrated for immediate use
• Small circuit board size allows smaller sensor housing
• Temperature is measured and reported via the digital output if you wish to temperature
compensate using Excel calculations.
• Range is selected when ordering. Range can be altered by Alphasense or distributor:
0-5000 ppm (indoor air, urban air, cabin air quality)
0-5 % (safety)
0-20% (combustion)
0-100% (industrial process monitoring)
Besides periodic sensor re-calibration by Alphasense or distributor, the transmitter electronics require
no maintenance throughout the lifetime of the sensor. Please read these instructions to ensure
correct installation and use of your CO2 gas sensor with transmitter.
Component Description
J1 Mini B USB socket for connection to PC or similar
J2 2.1mm DC In socket (positive tip)
J3 4-pin Molex plug for power (pins 3,4) and 4-20mA signal (pins 1,2)
J4 Six-pin debugging plug: DO NOT USE
P1 10-turn potentiometer to adjust IR bulb voltage (bulb is inside IRC-A1/ AF)
2 SETUP
Dimensions
The pillars are tapped to accept M3 screws. We recommend a screw length of at least 8mm to
ensure rigid fixing. It is good practice to hold the pillar when screwing into the pillars to stop the pillar
from rotating on the PCB. It may be easier to remove the sensor whilst screwing the circuit board
pillars to your housing. If you remove the pillars, ensure that, if mounting to the lid of your housing,
you include the washer between the pillar and PCB to ensure correct height of the assembly - see
figures 2 and 5.
Figure 2. Mounting pillar configuration for .Figure 3. Mounting pillar configuration for attaching to
attaching to the lid of an enclosure. the base of an enclosure.
(Long pillars may be required Pt: No 034-0005-00)
Figure 4. Side view of mounting to lid of Figure 5. Side view of mounting to the base of an
an enclosure. an enclosure.
Allow 30 minutes after plugging the sensor back into the board for the output to stabilise.
The transmitter is supplied with one 4-pin Molex socket for power and 4-20mA signal, with 200mm of
cable. User sources the optional 2.1mm DC IN socket or USB cable.
WARNING Do not supply mains AC power to this unit: this will destroy the transmitter and
void the warranty.
The transmitter is protected against incorrect polarity and over voltage, and will not function if you
have reversed the power supply.
Your 4-20mA supply must be able to provide 100mA with at least 10 volts to either the DC IN socket
(socket J2) or to pins 3 (ground) and 4 (power in) on the Molex socket.
Although there is an on-board regulator, it is good practice to minimise noise pickup in your power
cable. If the digital or 4-20mA signals are noisy, check your power cable and 4-20mA cable for noise
and ripple.
Ensure that there is a good gas seal between the sensor and the housing during both calibration
check and use.
If the gas sample is pumped, ensure that the flow rate is sufficient. A flow rate of at least 300sccm
(0.3L/m) is the minimum flow rate for pumped gas sampling lines, with 500sccm the recommended
flow rate. Alternatively, the sample gas can diffuse to the front face of the sensor.
If the top dust/ oil filter has become badly contaminated, contact Alphasense for replacement dust
filter.
Replacement dust/ oil filters and spare parts can be ordered by quoting the part numbers below.
5 CALIBRATION
The 4-20mA transmitter is shipped pre-calibrated to the range specified in your purchase order.
If you are using digital data collection, add an additional column to the Excel spreadsheet,
multiplying the transmitter calculated concentration by a scaling factor that corrects for your
determined calibration error. You do not need to correct for zero error, if you have also
applied temperature correction.
If 4-20mA signal, adjust your alarm points to correct for the error.
If you need to recalibrate the [transmitter plus sensor], return the complete unit to Alphasense,
specifying your signal method (4-20 or digital), reported error and gas concentration that you
determined the error and full scale (range).
The USB interface on the NDIR transmitter uses a virtual COM port (VCP) on the PC. The USB
drivers are supplied with the software. These need to be unzipped into a directory. The first time the
board is attached, the PC will attempt to find the drivers. Use the "Search for a suitable driver for my
device" option, specify a location, and then browse to the directory where the ftdi drivers were
unzipped.
WARNING Do not plug the transmitter PCB into your PC without first installing the correct
drivers (“NDIR Transmitter interface V1.10”). Failure to do so could compromise future
installations of USB devices.
Download the file NDIR Transmitter board from the Alphasense Web site. This file is compressed with
WinZip® and will need to be extracted into a temporary directory prior to installation.
This file is located at: http://www.alphasense.com/index.php/safety/application-notes/
Once the downloaded file has been extracted into a temporary directory, either double click the
“Setup.exe” icon or run the setup file using Start|Run from the Windows® Start menu. Once the setup
has started you will be prompted for a installation directory. The default location is C:\Program
Files\Alphasense_NDIR\ but you can change this. The installation program will then ask for a
particular program group, Again, you can change this. We recommend you follow the default settings
during the setup process. The installation software will confirm that the setup is complete.
If the installation fails to complete the setup for any reason, it is recommended that you use the
Windows® Add and Remove function (Start|Control Panel) to uninstall the software and start the
setup again.
Once the software has been installed correctly, you can navigate to C:\Program
Files\Alphasense_NDIR and double click the NDIR.exe icon to start the program. You can also create
a desktop short cut from here.
7 TROUBLESHOOTING
Problem Action
Ensure that you are supplying between 7 and 40VDC to either
the DC IN socket or pins 3 and 4 on the Molex socket.
No 4-20mA signal Ensure the polarity is correct.
Remember that if you are using only the USB cable, the 4-
20mA output is disabled.
Check for noise on your signal and power cables. Shielded
twisted pairs are recommended.
Add screening around PCB if it is located in an electrically
Noisy signal noisy environment.
If you have the IRC-A1 cell in an area where there is excessive
light, return transmitter and sensor to Alphasense to replace
with IRC-AF to reduce light interference.
Check digital or 4-20mA signal error with error specification in
Inaccurate readings
IRC-A1/AF data sheet on the Alphasense web site.
Return transmitter to Alphasense or distributor for calibration
Transmitter does not cover the
to your preferred range. Range will have been specified when
required concentration range.
you ordered.
Using the Excel spreadsheet, select the temperature
Temperature correction is required.
correction option. See CALIBRATION section above.
Software does not load on your PC
Contact Alphasense for assistance.
or Mac.
Test point
1 Reference channel signal
2 Active channel signal
3 Thermistor voltage signal
4 Reference voltage
5 DCI voltage (after internal regulator)
6 DAC (for 4-20mA signal)
8 WARRANTY
Transmitters are warranted for two years. IRC-A1 and IRC-AF are each warranted for two years.
Alphasense Limited
300 Avenue West
Skyline 120
Great Notley
Essex CM77 7AA
UK