Professional Documents
Culture Documents
John Edwards
Process NMR Associates, LLC, Danbury, CT
Presented at 129th AOAC Annual Meeting and Exposition, September 30, 2015
Los Angeles, CA
NMR Systems
University of Aarhus, Denmark
October 31, 2014
Cryogen-Free, 110V, “Portable” NMR Systems
? ?
Question:
Why is this competitors product so good?
Sample preparation:
Extract with 20:80 D2O:CD3CN
Centrifuge or Settle
Supernatant analyzed.
Answer:
Also contained ~10 wt% Sulfoaildenafil
A Viagra Analog
Typical Herbal Extracts
60 MHz
300 MHz
Comparison of 300 and 60 MHz NMR of ManUp Now Herbal Supplement –acetonitrile-D2O extract
Component Wt% = 100 x (Component Molar Integral / Std Molar Integral) x (Wt Std / Wt Sample) x (Component Mw / Std Mw)
300 MHz
60 MHz
60 MHz
300 MHz
Component Wt% = 100 x (Component Molar Integral / Std Molar Integral) x (Wt Std / Wt Sample) x (Component Mw / Std Mw)
Component Wt% = 100 x (Component Molar Integral / Std Molar Integral) x (Wt Std / Wt Sample) x (Component Mw / Std Mw)
Component Wt% = 100 x (Component Molar Integral / Std Molar Integral) x (Wt Std / Wt Sample) x (Component Mw / Std Mw)
300 MHz NMR – PDE5i Standards
43 MHz Benchtop NMR (Magritek Spinsolve) – PDE5i Standards
43 MHz
60 MHz
300 MHz
43 MHz
60 MHz
300 MHz
43 MHz
60 MHz
300 MHz
300 MHZ
Undeclared Caffeine
43 MHz
60 MHz
Dehydroepiandrosterone
43 MHz
Dehydroepiandrosterone
43 MHz
43 MHz
Anal. Chem., 2014, 86 (23), pp 11897–11904
Acknowledgments
Kristie Adams and Anton Bzhelyansky – USP
Leap Technologies (43 MHz) - Cosa-Xentaur (60 MHz)