Certificate of Calibration Cae)
Model ADT20 Report Ni NI9O9I316
Description: Multifunction Loop Calibrator Calibr 9/13/2019
Serial Number 31019010118, Due Date: 9/12/2020
Temperature: 23.1°C Humidity: 57% RIL
Status: New Certificate Date: 9/13/2019
Procedure: ADT210;Lab3.0.4; Data Type: As-Left
Data Results: In Tolerance
od Location of Calibration:
Additel Corporation
2900 Saturn Sireet #B
Brea CA 92821
United States
Reference Standards used in this calibration are traceable to the SI Units through National Institute of Standards
and Technology (NIST) or other recognized National Metrology Institutes (NMI's). This calibration is compliant to
ISO/IEC 17025:2017 and ANSI/NCSL 2540-1:R2002.
By default, Additel reports the measurement result and its associated measurement uncertainty. When the
measurement result is outside Additel’s published specifications, it is indicated by a fail condition without taking
the associated measurement uncertainty into consideration. Alternatively, if a compliance decision is desired
taking the measurement uncertainty into account, Additel utilizes the ANSI 2540.3 Handbook Method 6 to make
statements of conformity to specifications.
The results on this calibration report apply only to the item (s) calibrated or testec. Uncertainties of the
measurements are based upon a coverage factor of k=2 providing an approximate 95% confidence level. The
recommended calibration due date is included in the report only if it is specified by the customer,
This report shall not be used by the customer to claim product certification, approval, or endorsement by NVLAP,
NIST, or any agency of the U.S. government. This report shall not be reproduced, except in full without written
approval of Additel Corporation.
Standards Used
Model Description Serial Number Cal Date Due Date
5730A Fluke Multifunction Calibrator 4085502 2/14/2019 2/14/2020
3458A Keysight Digital Multimeter MY45055128 2/5/2019 2/5/2020
Comments: Performed By: A
‘Alvin Wong
ose
Approved By:
Eri
ic Chavier
Page | to 2Report Number: N19091316
Test Results
DC Voltage Measurement
Accuracy: #(0,01%RD + 2mV.
Range Rel DUT | Difference [Specification] Uncertainty
Wy W) (Wy Wy ‘W)
0.0000. 0.000. 0.0000 +0.0020
5.0000 5.000 0.0000. £0.0025
10.0000. 10.000 0.0000
01030 30,0000 30.001 0.0010 +0.0050 2.0E-04
Condition
DC Current Measurement
Accuracy :_+(0.01%RD + uA)
Range, Reference DUT Difference | Speci ition] Uncertainty,
(mA) (mA) (mA) (mA) (mA) (mA) | Condition
0.0000 0.000 0.0000 +0.0020 1.0E-04 pass
“4.0000 | 4.000 | 0.0000} 0.0024 ass
12.0000 | 11.999 | -0.0010 | ¥0.0032 pass
Oto 24 24.0000, 23.998 -0.0020 0.0044 pass
24mA DC Output
Accuracy :_+(0.01%RD + 2uA)
Range DUT | Reference | Difference [Specification] Uncertainty] 5.4,
(mA) (mA) (mA) (mA) (mA) (mA) _
0.000 [0.0001 {0.0001 ~=0.0020 [7.0F-04 [pass
4.000 [3.9999 {0.0001 | +0.0024 | 1.0E-04 [pass
72,000|_12.0003 0.0003 —[~+0.0032_|~3.0:-04_[ pass
01024 24.000 | 24.0008 | 0.0008 | 20.0044 | 6.0E-04 | pass
Page 2 to 2