Signal Conditioning Wheatstone Resistive Bridge Sensors: Application Report
Signal Conditioning Wheatstone Resistive Bridge Sensors: Application Report
ABSTRACT
Resistive elements configured as Wheatstone bridge circuits are used to construct force and
pressure sensors. The resistive elements used to make the bridge change resistance in
response to mechanical strain.
This report discusses the basic concepts of resistive bridge sensors and three circuits
commonly used for signal conditioning their output.
Contents
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2 Signal Conditioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.1 Single Op Amp Differential Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.2 Three Op Amp Instrumentation Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.3 Two Op Amp Instrumentation Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
List of Figures
1 Wheatstone Bridge Sensor and Thevenin Equivalent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2 Single Op Amp Differential Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
3 Sensor Output With No Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
4 Thevenin Equivalent Shows Required Gain is 50% Higher . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
5 Three Op Amp Instrumentation Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
6 Two Op Amp Instrumentation Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1 Introduction
There are two main technologies used to create resistive bridge sensors: strain gauge (or gage)
and integrated circuits.
Strain gages are widely used and have been available for many years. Typically, the strain
gauge is bonded to a rigid structure, and when a force acts upon the structure, the strain gauge
changes resistance. Strain gauge sensors are commonly used for both force and pressure
measurement. For more details refer to The Pressure Strain and Force Handbook, Omega
Engineering, Inc. http://www.omega.com.
More recently, monolithic resistive bridge sensors have become commercially available as
integrated circuits. The bridge circuit is constructed on a silicon die. When a force is exerted on
the die, the resistance changes. Normally this type of sensor is used to measure pressure. For
more information refer to Solid–State Pressure Sensors Handbook, Sensym,
http://www.sensym.com.
1
Figure 1 shows a resistive Wheatstone bridge circuit and its Thevenin equivalent. When an
excitation voltage is applied between Vexc and GND and all resistances are equal, the voltage
at SIG+ and SIG– is 1/2 Vexc.
Wheatstone Bridge Sensor Thevenin Equivalent
R2 ∆ R << R ⇒ RTH ≈ R 2
Vexc
Sig+
R+∆R +
R–∆R VTH+
– Vexc ± Vsig
R2 VTH± =
Sig- Sig+ Sig- 2
R–∆R R+∆R +
VTH– –
Sensors are designed so that when acted upon, opposite resistors in the bridge change
resistance as shown by ±∆R. The voltage at SIG+ and SIG– is then offset from 1/2 Vexc.
In a measurement system, the differential voltage, Vsig = (SIG+) – (SIG–), is the electrical signal
indicating the amount of force or pressure acting upon the sensor. With four active elements, the
output voltage is: Vsig +
Vexc DR . Since ∆R is proportional to the force or pressure, this
R
can be rewritten as Vsig = Vexc × F × S, where F is the force or pressure and S is the sensitivity,
or output voltage, of the sensor in mV per volt of excitation with full scale input, as specified by
the sensor manufacturer.
Wheatstone resistive bridge sensors can be analyzed using Thevenin’s Theorem, where the
circuit is reduced to voltage sources with series resistance. Figure 1 shows the Thevenin
equivalent circuit. ∆R is normally very small in comparison to R so the Thevenin equivalent
series resistance, commonly known as the source resistance, can be taken as equal to R. The
Thevenin voltage sources have a common mode component equal to 1/2 Vexc and a differential
component equal to 1/2 Vsig.
2 Signal Conditioning
Normally full scale output voltages are in the 10 mV to 100 mV range, and need to be amplified
in a data acquisition system. Three circuits commonly used for amplification are the one op amp
differential amplifier, the three op amp instrumentation amplifier, or the two op amp
instrumentation amplifier. These amplifiers amplify the differential input voltage, and reject the
common mode input voltage. They are well-suited for use with Wheatstone bridge sensors.
Depending on design requirements one may be better suited than the other.
All three circuits require resistor matching to achieve good CMRR. Variations of these amplifiers,
which add gain and offset adjustment, can be found in various application literature.
_
R4
R1 VO = [(Sig +) – (Sig –)]x
Sig+ R2
+
990Ω 1010Ω
Sig- = 2.475 V
1010Ω 990Ω
Vsig = 50 m V
Sig+ = 2.525 V
_
500Ω 1kΩ VO = 5.000 V
2.525 V +
150kΩ
SENSOR
Rg _
R4 2Rf
R1 VO = [(Sig +) – (Sig –)]x x +1
+ R2 Rg
Rf
_ R3
R1=R2 and R3=R4
SIG + +
Rg R1
R4
SIG – +
R3
Again, using a resistor pack for R1–R4 and setting the gain using Rg will help achieve better
CMRR performance. This amplifier is not as balanced as the 3 op amp in-amp. A signal at SIG–
passes through both op amps, whereas, only one op amp is seen from SIG+.
Texas Instruments and its subsidiaries (TI) reserve the right to make changes to their products or to discontinue
any product or service without notice, and advise customers to obtain the latest version of relevant information
to verify, before placing orders, that information being relied on is current and complete. All products are sold
subject to the terms and conditions of sale supplied at the time of order acknowledgement, including those
pertaining to warranty, patent infringement, and limitation of liability.
TI warrants performance of its semiconductor products to the specifications applicable at the time of sale in
accordance with TI’s standard warranty. Testing and other quality control techniques are utilized to the extent
TI deems necessary to support this warranty. Specific testing of all parameters of each device is not necessarily
performed, except those mandated by government requirements.
In order to minimize risks associated with the customer’s applications, adequate design and operating
safeguards must be provided by the customer to minimize inherent or procedural hazards.
TI assumes no liability for applications assistance or customer product design. TI does not warrant or represent
that any license, either express or implied, is granted under any patent right, copyright, mask work right, or other
intellectual property right of TI covering or relating to any combination, machine, or process in which such
semiconductor products or services might be or are used. TI’s publication of information regarding any third
party’s products or services does not constitute TI’s approval, warranty or endorsement thereof.