You are on page 1of 12

G a s t r o i n t e s t i n a l I m a g i n g • Pe r s p e c t i ve

McCollough et al.
In Defense of Body CT

Gastrointestinal Imaging
Perspective
Downloaded from www.ajronline.org by 200.145.126.224 on 12/04/18 from IP address 200.145.126.224. Copyright ARRS. For personal use only; all rights reserved

FOCUS ON:

In Defense of Body CT
Cynthia H. McCollough1 OBJECTIVE. Rapid technical developments and an expanding list of applications that have
Luís Guimarães supplanted less accurate or more invasive diagnostic tests have led to a dramatic increase in the
Joel G. Fletcher use of body CT in medical practice since its introduction in 1975. Our purpose here is to discuss
medical justification of the small potential risk associated with the ionizing radiation used in
McCollough CH, Guimarães L, Fletcher JG CT and to provide perspectives on practice-specific decisions that can maximize overall patient
benefit. In addition, we review available dose management and optimization techniques.
CONCLUSION. Dose reduction strategies described in this article must be well under-
stood and properly used, but also require broad-based practice strategies that extend beyond
the CT scanner console and default, generic manufacturer settings. In the final analysis, phy-
sicians must request the imaging examination that best addresses the specific medical ques-
tion without allowing worries about radiation to dissuade them or their patients from ob-
taining needed CT examinations. Ongoing efforts to ensure that CT examinations are both
medically justified and optimally performed must continue, and education must be provided
to the medical community and general public that put both the potential risks—and bene-
fits—of CT examinations into proper perspective.

R
apid technical developments and The radiation dose associated with a CT
an expanding list of applications examination (~ 1–14 mSv) is comparable to
that have supplanted less accu- the annual dose received from naturally oc-
rate or more invasive diagnostic curring sources of radiation, such as radon
tests have led to a dramatic increase in the and cosmic radiation (1–10 mSv) [4]. More
use of body CT in medical practice since its importantly, conservative estimations of po-
introduction in 1975. Our purpose here is to tential risk (i.e., any required assumptions are
discuss medical justification of the small po- made toward the direction of overestimating
tential risk associated with the ionizing ra- risk rather than underestimating it) show that
diation used in CT and to provide perspec- the potential risk of dying from undergoing a
Keywords: ACR appropriateness criteria, body CT, tives on practice-specific decisions that can CT examination is less than that of drowning
cancer, CT, gastrointestinal imaging, oncologic imaging,
radiation dose, radiation risk
maximize overall patient benefit. In addition, or of a pedestrian dying from being struck
we review available dose management and by any form of ground transportation, both
DOI:10.2214/AJR.09.2754 optimization techniques. of which most Americans consider to be an
extremely unlikely event. Table 1 provides a
Received March 17, 2009; accepted without revision
Considerations of Risk comparison of the statistical odds of dying
March 23, 2009.
In 2006, the estimated number of CT ex- from an abdominopelvic CT examination
The Calouste Gulbenkian Foundation (Lisbon, Portugal) aminations performed in the United States relative to other causes of death [5–11].
provided salary support for L. Guimarães. was approximately 62 million, up from 46 Estimates of the risk of cancer induction
1
million in 2000 and 13 million in 1990 [1]. in humans from exposure to ionizing radia-
All authors: Department of Radiology, Mayo Clinic, 200
First St. SW, East-2 Mayo Bldg., Rochester, MN 55905.
This increased use of CT is largely because tion come from epidemiological studies of
Address correspondence to C. H. McCollough of the tremendous contributions of increas- various exposed cohorts, the largest of which
(mccollough.cynthia@mayo.edu). ingly powerful CT methods to modern health is the survivors of the 1945 atomic bombings
care. However, in spite of measurable health in Japan. Risk is estimated by looking at the
AJR 2009; 193:28–39
benefits, the media has given considerable at- expected numbers of cancers in a specific
0361–803X/09/1931–28 tention to the very small potential health risk population and the actual numbers observed
associated with the ionizing radiation from a in the exposed cohort. The National Acade-
© American Roentgen Ray Society CT examination [1–3]. mies of Science have published a series of

28 AJR:193, July 2009


In Defense of Body CT

TABLE 1:  Estimated Lifetime Risk of Death From Various Sources agreement provides reassurance that results
from the bomb survivor cohort can be ex-
Cause of Death Estimated No. of Deaths per 1,000 Individuals
pected to translate relatively well to a more
Cancer [38] 228 current, westernized population.
Motor vehicle accident 11.9 The assumption that the risk of cancer in-
Downloaded from www.ajronline.org by 200.145.126.224 on 12/04/18 from IP address 200.145.126.224. Copyright ARRS. For personal use only; all rights reserved

Radon in home
creases in a linear fashion from low doses to
high doses, and that there is no threshold dose
Average U.S. exposure 3
below which radiation is not carcinogenic, is
High exposure (1–3%) 21 referred to as the linear-no-threshold (LNT)
Arsenic in drinking water model of radiation risk. In Figures 1 and 2,
2.5 µg/L (U.S. estimated average) 1
the linear fit of the point estimates is based
on the LNT model. However, the error bars
50 µg/L (acceptable limit before 2006) 13 associated with the point estimates are suffi-
Radiation-induced fatal cancer ciently large that a relative risk of 1.0 (black
Routine abdominopelvic CT 0.5 horizontal line) cannot be excluded, even up
Single phase, ~ 10-mSv effective dose
to a dose of over 0.2 Sv (200 mSv). That is,
in the absence of applying the LNT hypothe-
Annual dose limit for a radiation worker sis (i.e., the fitted solid line), there is not con-
10 mSv (recommended yearly average) 0.5 vincing statistical evidence of increased risk,
50 mSv (limit in a single year) 2.5 particularly for doses below 100 mSv. This
is consistent with the 2006 BEIR VII report
Pedestrian accident 1.6
[13], which notes that “At doses of 100 mSv
Drowning 0.9 or less, statistical limitations make it difficult
Bicycling 0.2 to evaluate cancer risk in humans.” Thus,
Lightning strike 0.013 even though for the purpose of ensuring pub-
Note—Adapted with permission from [97].
lic health and patient safety we choose to as-
sume that a risk does exist, it is essential to
reports concerning the health risks from ex- the results of a large study of British radia- keep in mind that it has not been scientifical-
posure to low levels of ionizing radiation tion workers. The cohort included 174,541 ly demonstrated that there is any risk from
called the Biological Effects of Ionizing Ra- individuals who received an occupational ex- radiation doses below 100 mSv.
diation (BEIR) reports. These reports form posure between 1976 and 2001. In red, the The most troubling aspect of many media
the foundation of current estimates of radia- data from the BEIR VII report are given. reports is the underlying implication that be-
tion risk. In Figures 1 and 2, the relative risk Very good agreement is shown between data cause CT use is increasing, death rates due
of radiation-induced leukemia (Fig. 1) and from the British radiation worker study and to radiation-induced cancers are not far be-
solid neoplasms (Fig. 2) is shown as a func- the data in the BEIR VII report, which relies hind. Such analyses fail to realize two criti-
tion of effective (whole-body equivalent) heavily—though not exclusively—on data cal aspects relating to potential risks of CT:
dose. These figures [12] summarize (in blue) from the Japanese bomb survivors. This nontransferability of risk and mortality re-

4.0 1.4
3.5
1.3
3.0
Relative Risk

Relative Risk

2.5 1.2
2.0
1.5 1.1
1.0
1.0
0.5
0.0 0.9
0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.0 0.1 0.2 0.3 0.4 0.5 0.6
Dose (Sv) Dose (Sv)
NRRW point estimates NRRW linear fit NRRW point estimates NRRW linear fit NRRW linear fit lower
NRRW linear fit lower NRRW linear fit upper BEIR VII all solid NRRW linear fit upper BEIR VII all solid upper
NRRW BEIR leukaemia mortality BEIR VII lower BEIR VII all solid lower
NRRW VII upper

Fig. 1—Trends in relative risk for leukemia (and 90% confidence intervals) Fig. 2—Trends in relative risk all malignant neoplasms excluding leukemia (and
excluding chronic lymphocytic leukemia in over 170,000 radiation workers by 90% confidence intervals) in over 170,000 radiation workers by lifetime radiation
lifetime radiation dose. The black horizontal line indicates a relative risk of 1 (i.e., dose. The black horizontal line indicates a relative risk of 1 (i.e., no increase in risk).
no increase in risk). NRRW = National Registry for Radiation Workers; BEIR VII = (NRRW = National Registry for Radiation Workers; BEIR VII = Biological Effects of
Biological Effects of Ionizing Radiation VII [13]). (Reprinted with permission from Ionizing Radiation VII [13]). (Reprinted with permission from Muirhead CR, O’Hagan
Muirhead CR, O’Hagan JA, Haylock RG, et al. Br J Cancer 2009; 100: 206–212) JA, Haylock RG, et al. Br J Cancer 2009; 100: 206–212)

AJR:193, July 2009 29


McCollough et al.

duction through disease identification and Physics Society recommends against nation (e.g., delayed or inaccurate diagnoses
treatment. Nontransferability of risk means quantitative estimation of health risks or treatment) must exceed the potential risk
that a medically related dose given to one in- below an individual dose of 5 rem [50 associated with the examination. Medical
dividual does not transfer the potential risk mSv] in one year or a lifetime dose of justification includes a consideration of evi-
of cancer induction to those who have never 10 rem [100 mSv] above that received dence-based recommendations for relevant
clinical scenarios and an understanding of
Downloaded from www.ajronline.org by 200.145.126.224 on 12/04/18 from IP address 200.145.126.224. Copyright ARRS. For personal use only; all rights reserved

undergone CT. Thus, the focus must be on from natural sources.… Below 5–10
those individuals undergoing CT—not on rem (50–100 mSv)…risks of health ef- the risk of disease for each patient. If the
all Americans as was reported by the me- fects are either too small to be observed health risks and likelihood of a disease are
dia subsequent to the release of report 160 or are nonexistent. high, increased risk from radiation and IV
by the National Council on Radiation Pro- contrast medium is justified if CT can detect
tection [14], which documented a sixfold in- Thus, our purpose is to counter the alarm- the disease (e.g., the source of infection in
crease in the dose delivered to members of ist statements being made in both general hospitalized patients with sepsis). Addition-
the U.S. public from medical sources. This and professional forums regarding the po- ally, increased benefits from higher tube cur-
increase was averaged over all Americans, tential dangers to public health from the in- rents (higher radiation dose) are justified
even though medical doses were delivered creased use of CT. We address this topic in when they permit diagnostic-quality images
to a much smaller—and more elderly—seg- terms of the two fundamental principles of to be obtained (e.g., in morbidly obese pa-
ment of the U. S. population. radiation protection as applied to medicine: tients). Low-dose techniques are justified
Second, the small potential risk of cancer justification and optimization. when the patient is asymptomatic or when
induction must be considered in the context image quality does not require discrimina-
of the potential incremental (survival) bene- Justification tion between structures with soft-tissue at-
fit from undergoing CT. From Table 1, it can The field of radiation protection, which tenuation (e.g., CT colonography [CTC], re-
be seen that the lifetime risk of a fatal can- seeks to minimize the radiation dose levels to peat CT for renal stone disease). Unjustified
cer from all causes is 22.8%, and the lifetime exposed persons, embraces three guiding prin- exams include those for which CT is not in-
potential risk of a fatal cancer from the radi- ciples when applied to medical exposures: dicated (e.g., defensive medicine), for which
ation associated with a body CT examination alternative imaging would be more appropri-
is approximately 0.05%. However, in a pa- 1. Justification: The examination or pro- ate (e.g., MRI for perianal fistulas), or for
tient with a known cancer, the risk of cancer cedure must be medically indicated. which unnecessary series are acquired (e.g.,
death is already much higher than average. 2. Optimization: The examination or pro- routine delayed images in the absence of liv-
In those patients, CT is used to stage cancer cedure must use doses that are as low as er/renal masses).
with an aim toward cure or extending life. reasonably achievable (ALARA), with- Justification should also take into account
As an example, cure from colorectal cancer out compromising the diagnostic task. potential alternatives—such as ultrasound,
is not possible unless hepatic metastases can 3. Limitation: In medicine, upper limits MRI, or optical imaging—as well as urgency
be diagnosed and treated. In these patients, to dose levels are typical only for oc- and clinical availability. Once the determi-
the use of CT is a critical means of reducing cupationally exposed individuals (i.e., nation is made that an appropriate CT exam-
mortality. Thus, the benefit-to-risk ratio for the radiologist or technologist). Lim- ination can benefit the patient, CT parame-
any patient will be driven by the benefit and its are rarely established for medical- ters should be optimized and dose reduction
appropriateness of the CT examination. ly necessary examinations or proce- techniques used to perform the diagnostic
Although informing the public of poten- dures. One example in which patient task at the lowest appropriate level of radia-
tial health risks—even small risks—is ap- dose limits have been established is tion dose. These strategies are discussed in
propriate, journalistic responsibility should screening mammography. However, the Optimization section, which appears lat-
ensure that the data are presented in a man- when a screening mammogram, physi- er in this article.
ner that puts the risk into perspective. Stating cal examination, or patient symptoms
that a CT scan is the equivalent of 600 chest indicate the need for diagnostic mam- Symptomatic Patients
x-rays may be an accurate estimate; however, mography, no dose limits are applied. The American College of Radiology
such statements imply that a CT scan deliv- The philosophy of the U.S. Food and (ACR) Appropriateness Criteria [16, 17] and
ers “a lot” of radiation simply because 600 of Drug Administration (FDA) is not to others have provided evidence-based guide-
anything seems like a relatively large num- establish dose limits because, as with lines to help physicians recommend an ap-
ber. The implication is that CT is a high-dose any medicine or medical intervention, propriate imaging test. Table 2 delineates the
examination and presents a substantial risk the medical practitioner must be able to indications for which the ACR Gastrointesti-
to the recipient. Quite the opposite is true. tailor the examination to the particular nal (GI) Expert Panel considers CT the most
In their position statement on radiation risk patient and medical concern. appropriate imaging option [18]. To illustrate
[15], the Health Physics Society, a nonprof- their erudite considerations of justification
it scientific professional organization whose Maximizing Benefit-to-Risk Ratio and optimization, we consider two clinical
mission is excellence in the science and prac- The CT examination should be performed scenarios: small-bowel obstruction (SBO)
tice of radiation safety, stated the following: only when the radiation dose is deemed to be and suspected hepatic metastases.
justified by the potential clinical benefit to SBO—SBO accounts for 20% of all acute
In accordance with current knowl- the patient. An alternative way of stating this surgical admissions, potentially resulting
edge of radiation health risks, the Health is that the risk of not performing the exami- in bowel ischemia, bowel strangulation, or

30 AJR:193, July 2009


In Defense of Body CT

TABLE 2:  Clinical Scenarios for Which the American College of Radiology Gastrointestinal Imaging Expert Panel [17]
Consider CT the Most Appropriate Imaging Technique

Variants of General Clinical Scenario for Which CT


Is the Most Appropriate Examination Is Not the Most Appropriate Examination (Most Appropriate
General Clinical Scenario (CT Appropriateness Rank)a Modality and Its Appropriateness Rank; CT Appropriateness Rank)a
Downloaded from www.ajronline.org by 200.145.126.224 on 12/04/18 from IP address 200.145.126.224. Copyright ARRS. For personal use only; all rights reserved

Acute diffuse abdominal pain Adult postoperative patient presenting with fever (8) Pregnant patient (US of abdomen, 8; 5)
and fever or suspected
abdominal abscess
Adult postoperative patient presenting with persistent
fever and no abscess seen on CT within the last 7 d (8)
Adult patient presenting with fever, nonlocalizing
abdominal pain, and no history of recent surgery (8)
Acute pancreatitis Patient presenting with severe abdominal pain and Patient presenting with severe abdominal pain of unknown cause;
elevated amylase lipase level; 48 h after presentation, no first episode of pancreatitis suspected (US of abdomen, 8, 6)
improvement or degradation (assume no prior imaging) (8)
Patient presenting with severe abdominal pain, elevated Patient presenting with severe abdominal pain, elevated amylase
amylase lipase level, fever, and elevated WBC count (9) lipase level, no fever or evidence of fluid loss at admission, and
clinical score pending (US of abdomen, 8, 7)
Patient presenting with severe abdominal pain, elevated
amylase lipase level, hemoconcentration, oliguria, and
tachycardia (9)
Blunt abdominal trauma Stable patient (8) Unstable patient (US screening for hemoperitoneum, 7, 4)
Stable patient presenting with hematuria > 35 RBC per
high-power field (8)
Crohn’s disease Adult initially presenting with abdominal pain, fever, or Child (age < 14 y) with known Crohn’s disease presenting with
diarrhea; Crohn's disease suspected (8) stable, mild symptoms (US of abdomen and pelvis, 6, 5)
Child (age < 14 y) initially presenting with abdominal pain,
fever, or diarrhea; Crohn’s disease suspected (8)
Adult with known Crohn’s disease presenting with fever,
increasing pain, leukocytosis, and so on (8)
Child (age < 14 y) with known Crohn’s disease presenting
with fever, increasing pain, leukocytosis, and so on (8)
Adult with known Crohn’s disease presenting with stable,
mild symptoms (7)
Dysphagia Patient presenting with oropharyngeal dysphagia with an
attributable cause (x-ray barium swallow modified, 8; CT not
mentioned)
Patient presenting with unexplained oropharyngeal dysphagia
(x-ray pharynx dynamic and static imaging, 8; CT not mentioned)
Jaundice Otherwise healthy patient presenting without pain and Patient presenting with acute abdominal pain and at least one of the
with one or more of the following: weight loss, fatigue, or following: fever, history of biliary surgery, or known cholelithiasis
anorexia; duration of symptoms is > 3 mo (9) (US of abdomen, 9, 7)
Patient who cannot tolerate undergoing a radical surgical Patient presenting with clinical findings and laboratory results that
procedure presenting without pain and with one or more make mechanical obstruction unlikely (US of abdomen, 8, 5)
of the following: weight loss, fatigue, or anorexia; duration
of symptoms is > 3 mo (9)
Patient presenting with confusing clinical picture and not described
in previous scenarios (US of abdomen, 8, 7)
Left lower quadrant pain Older patient presenting with clinical findings typical of Woman of childbearing age (transabdominal US with graded
diverticulitis (8) compression, 8, 7)
Patient presenting with acute, severe pain with or without
fever (9)
Patient presenting with chronic, intermittent, or low-grade
pain (8)
Obese patient (8)
(Table 2 continues on next page)

AJR:193, July 2009 31


McCollough et al.

TABLE 2:  Clinical Scenarios for Which the American College of Radiology Gastrointestinal Imaging Expert Panel [17]
Consider CT the Most Appropriate Imaging Technique (continued)
Variants of General Clinical Scenario for Which CT
Is the Most Appropriate Examination Is Not the Most Appropriate Examination (Most Appropriate
General Clinical Scenario (CT Appropriateness Rank)a Modality and Its Appropriateness Rank; CT Appropriateness Rank)a
Downloaded from www.ajronline.org by 200.145.126.224 on 12/04/18 from IP address 200.145.126.224. Copyright ARRS. For personal use only; all rights reserved

Liver lesion characterization Patient presenting with indeterminate mass on Patient presenting with findings typical of benign mass on initial
initial imaging > 1 cm, no suspicion or evidence of imaging, no history of malignancy (no imaging at this time, 8, 4)
extrahepatic malignancy or liver disease (CT or MRI
depending on availability, 8)
Patient presenting with indeterminate mass on initial Patient with known history of extrahepatic malignancy presenting
imaging > 1 cm, known or suspected liver disease with initial imaging findings typical of benign mass (no imaging at
associated with a high risk of hepatocellular carcinoma this time, 8, 5)
(e.g., chronic hepatitis, cirrhosis, hemochromatosis)
(CT or MRI depending on availability, 8)
Patient presenting with initial imaging findings typical of malignant
hepatic mass (no imaging at this time, 7, 6)
Patient with known history of extrahepatic malignancy presenting
with initial imaging findings typical of indeterminate solitary mass
> 1 cm (percutaneous biopsy liver, 8, 7)
Patient presenting with initial imaging finding of small lesion < 1 cm
(no imaging at this time, 8, 5)
Palpable abdominal mass Patient presenting with a palpable abdominal mass (8)
Pretreatment staging of Patient with rectal cancer presenting with a large Patient presenting with rectal cancer, small or superficial
colorectal cancer lesion (8) (endorectal US, 8, 6)
Patient with colon cancer other than cancer of rectum (8)
Right lower quadrant pain Adult patient presenting with fever, leukocytosis, and Pregnant woman presenting with fever and leukocytosis (US of
classic presentation clinically for appendicitis (8) abdomen, right lower quadrant, 8, 6)
Adult or adolescent patient presenting with fever, Child (age < 14 y) presenting with fever, leukocytosis, possible
leukocytosis; possible appendicitis, atypical appendicitis, atypical presentation (US of abdomen, right lower
presentation (8) quadrant, 8, 7)
Right upper quadrant pain Patient presenting with fever, elevated WBC count, and positive
Murphy sign (US of abdomen, 9, 5)
Patient presenting with suspected acalculous cholecystitis (NUC
cholescintigraphy, 8, 6)
Patient presenting with no fever, normal WBC count (US of
abdomen, 8, 7)
Patient presenting with no fever and normal WBC count; US shows
only gallstones (NUC cholescintigraphy, 8, 6)
Hospitalized patient presenting with fever, elevated WBC count,
and positive Murphy sign (US of abdomen, 9, 7)
Suspected liver metastasis Patient presenting for initial imaging test after detection
of primary tumor (8)
Patient presenting for surveillance imaging after treatment
of primary tumor (8)
Patient with high suspicion of malignancy based on
abnormal surveillance findings on US, CT, or MRI in
portal venous phase (8, MRI and percutaneous biopsy
have the same score)
Patient with high suspicion of benignancy based on
abnormal surveillance findings on US, CT, or MRI in
portal venous phase (8, MRI has the same score)
Suspected SBO Suspected complete or high-grade partial SBO (8)
Suspected intermittent or low-grade SBO (7, small-bowel
follow-through and enteroclysis have the same score)
Note—Adapted with permission of the American College of Radiology. No other representation of this material is authorized without expressed, written permission
from the American College of Radiology. Refer to the ACR website at www.acr.org/ac for the most current and complete version of the ACR Appropriateness Criteria.
US = ultrasound, SBO = small-bowel obstruction, NUC = radionuclide.
a Appropriateness rank ranges from 1 to 9, with 9 being most appropriate and 1, least appropriate.

32 AJR:193, July 2009


In Defense of Body CT

death, particularly if diagnosis is delayed. tions because their risk for disease is much In the past (before MDCT), this wide
Because of its diagnostic accuracy for high- lower than that of symptomatic patients [34]. range of considerations had minimal con-
grade SBO (> 90%) and its ability to iden- CTC is one screening examination that has sequences on the resultant CT examination.
tify the cause of obstruction, CT is recom- undergone extensive scrutiny. The justifica- However, now, with the advanced level of
mended as the first-line test in the initial tion for using CTC as a screening examina- MDCT technology available at even small
Downloaded from www.ajronline.org by 200.145.126.224 on 12/04/18 from IP address 200.145.126.224. Copyright ARRS. For personal use only; all rights reserved

evaluation of high-grade SBO [17–21] (Ta- tion includes the high mortality from colon facilities, differences in patient preparation,
ble 2). If, on the other hand, clinical symp- cancer, its long preclinical course, and the examination acquisition, and postprocess-
toms suggest a low-grade SBO, the accura- potential for polypectomy to eliminate the ing can have considerable impact on disease
cy of routine CT is much lower [20]. In such progression to invasive cancer [34]. On the conspicuity and, ultimately, on patient care.
circumstances, CT enteroclysis with larger basis of its performance characteristics [35– The development of new CT technolo-
volumes of oral contrast material or fluoros- 37], CTC was endorsed in 2008 as an accept- gies has facilitated the rapid growth of CT
copy with real-time visualization of gut mo- able colorectal cancer screening test by the in medical practice. Faster scanning and im-
tility are likely more helpful in identifying American Cancer Society [38]. The potential proved spatial resolution have led to the in-
points of low-grade obstruction [18]. MR risks of CTC are small, but include cancer re- corporation of a CT scanner in the emergency
enterography may be preferred for pregnant sulting from ionizing radiation, bowel perfo- department as a method for triaging trauma
patients or in practices with greater MR ex- ration, and unnecessary treatment and work- as well as the development of a wide array
perience [22, 23]. up of extracolonic findings. These risks must of organ-specific CT examinations (e.g., CT
Suspected hepatic metastasis—Suspect- be balanced against the anticipated lifetime angiography, enterography, and urography)
ed hepatic metastasis is a frequent indication risk of colorectal cancer (5–6% [39]) and the that guide management decisions. Emerging
for abdominal CT. The detection of hepatic prevalence of advanced colorectal neoplasia CT applications, such as CT cholangiogra-
metastases determines therapeutic decisions, (3–9% [40, 41]). phy and dual-energy CT, will continue to ap-
with early detection and treatment of some Brenner and Georgsson [39], using the pear as radiologists and primary care provid-
metastases now resulting in cure as a result conservative LNT model, estimated the po- ers incorporate new technologic possibilities
of subsequent chemotherapy, resection, or ra- tential risk of radiation-induced malignancy into patient care. Practices can maximize pa-
diofrequency ablation [24–26]. The poten- from CTC to be 0.14% for a 50-year-old adult tient benefit by creating mechanisms that fa-
tial increased risk of additional radiation in and 0.07% for a 70-year-old adult, with these cilitate innovation, establishing common CT
these scenarios is overwhelmed by the poten- risks decreasing further when optimized pro- acquisition protocols and quality programs,
tial benefits of accurate detection and char- tocols are used. They concluded that the ben- and eliminating nonbeneficial (inappropri-
acterization of liver lesions. The GI Expert efit-to-risk ratio for CTC was greater than 1. ate) examinations.
Panel ranked CT with contrast material as the The risk of bowel perforation at screening The establishment of common CT acqui-
most appropriate initial imaging test after the CTC is also extremely low, probably slight- sition and reconstruction protocols, which
detection and treatment of a primary tumor. ly less than that for optical colonoscopy at are tailored to each individual scanner mod-
Multiple other tests (MRI with gadolinium, 0.001–0.02% [42, 43]. Extracolonic CT find- el to deliver comparable image quality (sec-
PET, percutaneous biopsy) were considered ings may be beneficial [44, 45], but may also tion width, spatial resolution, temporal reso-
equally appropriate when a liver lesion was increase financial burden or morbidity [46, lution, image noise, and so on), and contrast
found on surveillance imaging; however, they 47], resulting in ongoing efforts to minimize enhancement protocols (oral, IV, or both) can
are also associated with some level of risk as these effects [48]. improve clinical benefits for patients for sev-
well (e.g., nephrogenic systemic fibrosis, ra- eral reasons. First, standardization reduces
dioisotopes, or infection, respectively) [18]. Practice Decisions That Individualize and variations in the resultant images due to ra-
To maximize patient benefit, CT acquisition Maximize Patient Benefit diologist-, technologist-, or scanner-depend­
parameters and patient preparation should be Each CT examination should be tailored ent factors. The ability to diagnose interval
tailored to the individual patient and indica- and effectively implemented for each patient changes in follow-up examinations is thus
tion (i.e., to detect and characterize liver le- on the basis of clinical history, suspected dis- greatly enhanced.
sions). Portal phase imaging is usually suffi- ease (and pathophysiology), patient size, ra- Second, the use of common examination
cient [27, 28] unless a patient’s primary tumor diologic conspicuity, and morbidities affect- protocols allows knowledge transfer from sub-
is hypervascular (e.g., neuroendocrine tu- ing the use of IV and oral contrast agents, specialized radiologists (e.g., a GI radiologist)
mors) [29, 30], whereas higher-spatial-resolu- while also taking into account the acquisi- to general radiologists elsewhere in the prac-
tion imaging may be warranted in looking for tion capabilities of the specific CT system tice. Creation of standardized protocols re-
some metastases or hepatocellular carcinoma to be used. The complexity and interrelated- quires input from medical physicists to maxi-
(HCC) [31, 32]. Higher tube currents may be ness of these multiple decisions argue for ra- mize image quality and ensure proper use of
needed if thinner sections are chosen to off- diologists to have the central role in guiding dose reduction techniques, and from referring
set increased image noise, and oral contrast and coordinating these decisions. Such lead- clinicians to maximize impact on clinical de-
agents may be needed to maximize the detec- ership maximizes the benefit-to-risk ratio for cision-making. Interdisciplinary collaboration
tion of other metastases in the bowel [33]. patients and involves coordinating knowl- often leads to the conclusion that image qual-
edge transfer and communication among ity and lesion detection are more important
Asymptomatic Patients many members of the patient care team— than lower radiation doses, particularly when
Asymptomatic patients are a unique group including the referring clinician, radiologist, the implications for misdiagnosis can be high
of individuals to consider for CT examina- medical physicist, technologist, and nurse. (e.g., suspected pancreatic cancer or HCC).

AJR:193, July 2009 33


McCollough et al.

Third, standardized organ- or disease-


spe­cific protocols permit scheduling triage
to the appropriate level of scanner technolo-
gy in those practices where such options ex-
ist. For many CT examinations, the difference
Downloaded from www.ajronline.org by 200.145.126.224 on 12/04/18 from IP address 200.145.126.224. Copyright ARRS. For personal use only; all rights reserved

in examination quality is minimally affected


as long as some minimum level of CT tech-
nology is used. However, for some examina-
tions, such as multiphase CT enterography
(Fig. 3) or scanning of an obese patient, pa-
tients should be preferentially scheduled on
higher-detector-row systems (e.g., 64-MDCT
or above) having higher x-ray power capabili-
ties to allow multiphase imaging at the proper A B
time intervals and image widths (< 3 mm) and
Fig. 3—84-year-old man with occult gastrointestinal blood loss and negative upper and lower endoscopies.
low noise levels in large patients. Multiphase CT enterography was performed on 128-MDCT scanner because of its isotropic spatial resolution
Quality programs that track compliance and greater power capability, permitting multiphase scanning of abdomen and pelvis.
with practice-based decisions and imaging A and B, Axial (A) and coronal (B) images from the enteric phase show an enhancing ileal mass (arrow), which
was found to be tubulovillous adenoma at surgical resection.
protocols can be extremely useful in identify-
ing areas for improving patient care. We peri-
odically monitor our general CT practice, as tablish mechanisms for transferring patients data measured and stored at our institution
well as all CTC examinations, for compliance to MRI or ultrasound when those examina- and obtained with the same sets of CTDI
with general guidelines and scanner-specific tions are more appropriate than CT. phantoms, with the same model of CTDI
protocols [49]. Our experience is that these ionization chambers, and with knowledge of
programs can identify areas needing im- Optimization the routine scanning parameters for an ab-
provement, minimize interobserver variabili- The technical aspects of medically justi- dominal CT examination, we calculated the
ty or lack of communication among members fied examinations should be optimized so CTDIvol for routine abdominal examinations
of the health care team, and improve patient that the required level of image quality is from the mid 1980s to 2004. The data shown
education and report timeliness and clarity obtained while keeping dose as low as pos- in Figure 4 support the general statement
[50]. Examples of changes we have made in sible [57]. The radiology community (radi- that dose tends to decrease as the half-val-
our practice as a result of these quality pro- ologists, medical physicists, and manufac- ue layer of the beam spectrum increases, al-
grams include standardizing oral contrast use turers) has worked to implement ALARA though other factors such as geometry and
for a variety of CT indications, and use of principles in CT [58–62], the most basic of detector type also play an important role in
low-dose imaging for the detection of renal which is the required adjustment of tech- determining dose [71]. In 2000, the Europe-
stones in patients with a prior positive CT ex- nique factors based on patient size (attenu- an Commission [72] set their diagnostic ref-
amination showing urolithiasis. ation characteristics) [63, 64]. Appropriate erence value, which is an investigation level
Eliminating nonbeneficial and inappro- dose management in CT means ensuring at the upper 75th percentile of doses used in
priate CT examinations likely represents that the right dose is delivered for the spe- clinical practice, at 35 mGy.
the most important step toward reducing cific patient attenuation and the specific di- Recently, the ACR’s CT Accreditation
CT risk. Evidence-based recommendations agnostic task. Program decreased their reference value
[17] or decision support tools [51] are used from 35 to 25 mGy, reflecting the downward
by many practices to facilitate appropriate Dose per Examination: The Historical View trend in body CT doses over the past decade
referral for CT. These electronic tools can The radiation output from a CT scanner [73]. Newer MDCT systems have eliminat-
significantly reduce the number of low-util- has been measured and reported in a stan- ed the dose inefficiency of the early multi-
ity examinations [51] and seem to have wide dardized manner since the early days of CT detector-row systems (4-slice scanners) for
clinical acceptability [52]. using a parameter known as the CT dose thin images, and automatic exposure control
Nonbeneficial CT examinations ordered index (CTDI) [65–67]. The volume CTDI systems can lower patient dose dramatically
as a result of defensive medicine practices or (CTDIvol) is a similar metric to describe the (20–40%). Even in large patients, dose re-
self-referral are more problematic. Both de- radiation output of a scanner, but takes into ductions can still be achieved using automat-
fensive medicine and self-referral can result account overlap or gaps between consecutive ic exposure control because the dose is dis-
in an increased number of marginally benefi- rotations of the x-ray tube [68–70]. Because tributed more effectively [74]. Finally, in the
cial imaging studies and increase cost as well of this consistency in CT dose metrics, it is 1980s, 10-mm images and multiple breath-
as patient risk [53–55]. Successfully address- relatively straightforward to assess how the holds were standard. Now, a 5-mm image
ing these practices would likely require ac- dose per examination has changed over the width and a single breath-hold is considered
tions by third-party payers or the government past two decades. Our large CT practice has routine, with 2- to 3-mm image widths used
[56]. Because it is unrealistic to expect that operated more than 100 different CT scan- for some applications (e.g., CT enterography)
all patients referred for CT will have entirely ners since the original EMI Mark I head CT and reconstructions as narrow as 1 mm or
appropriate indications, practices should es- scanner was installed in 1973. Using CTDI less used for multiplanar reformations or 3D

34 AJR:193, July 2009


In Defense of Body CT

tomic dimensions [63, 64, 79, 80]. For body


50
CT, a reduction in mAs of a factor of 4–5
45 from adult techniques is generally accept-
able in infants, whereas for obese patients, an
40 increase of at least a factor of 2 is required
European Commission 2000 reference value
Downloaded from www.ajronline.org by 200.145.126.224 on 12/04/18 from IP address 200.145.126.224. Copyright ARRS. For personal use only; all rights reserved

[79]. To achieve sufficient exposure levels for


35
obese patients, either the rotation time or the
CTDIvol (mGy)

30 tube potential may need to be increased.


American College of Radiology 2008 reference value
25 Angular and Longitudinal (x, y, z)
Tube Current (mA) Modulation
20
Angular mA modulation deals with the
15 variations in x-ray attenuation in the scan
plane (e.g., in the anteroposterior versus lat-
10 10-mm section width 7−10 mm 1−5 mm eral direction). The mA is varied as the x-ray
tube rotates according to the patient attenua-
5
tion at the same level. Attenuation informa-
0 tion is determined from the CT radiograph
Picker 1200 Picker 1200 Imatron EBCT GE 9800 GE HiSpeed GE Siemens (e.g., scout image) or from the previous 180°
(3.5-mm AI) (5.2-mm AI) (10-mm AI) (7.3-mm AI) (7.1-mm AI) LightSpeed 16 Sensation 64
(8.3-mm AI) (9.5-mm AI) projection. Longitudinal (z) mA modulation
takes into account variations in attenuation
Fig. 4—Bar graph shows historical trends in radiation dose at Mayo Clinic Rochester as measured using among different regions (e.g., shoulders vs
volume CT dose index (CTDIvol) on standard phantoms during routine scanner testing. Note decreased dose abdomen) by varying the mA along the pa-
over time even with the use of much smaller slice thicknesses over the past decade. EBCT = electron beam CT,
AI = aluminum. (Adapted with permission from [98]) tient’s long axis. The operator must, howev-
er, prescribe the desired level of image qual-
renderings. Hence, today’s CT systems pro- required in film-based radiography to avoid ity. The paradigms used to accomplish this
vide thinner image widths, improved spatial overexposing the film (i.e., a too-darkened are at present relatively manufacturer-specif-
resolution, and decreased partial volume av- film). Nonetheless, it remains the responsi- ic. Combining angular and longitudinal (x, y,
eraging at a fraction of the scanning time and bility of the CT operator to take patient size z) mA modulation is the most comprehensive
patient dose than the lower-quality, thicker- into account when selecting parameters that approach to CT dose reduction.
slice-width examinations of older scanners. affect radiation dose and image quality, the
most basic of which is the tube current–time Automatic Exposure Control
CT Scanner Accreditation product (mAs) [58, 62]. Modern CT systems adjust the x-ray tube
A CT scanner’s radiation dose levels and For most CT applications, it is common to current in response to variations in patient at-
image quality should fall within manufac- standardize the tube potential (kV) and gan- tenuation [81–84]. Methods of adapting the
turer specifications and comply with appli- try rotation time (in seconds): The fastest tube current to patient attenuation, known ge-
cable regulatory requirements. Regional and rotation time is typically used to minimize nerically as automatic exposure control sys-
national health care payers have begun to re- motion blurring or artifact and the lowest kV tems, are analogous to photo timing in gener-
quire that sites show satisfactory equipment is selected to maximize image contrast, pro- al radiography and have shown reductions in
performance and clinical use through pro- vided that the patient size and tube current dose of about 20–40% when image quality is
grams such as the ACR’s CT Accreditation limits are adequate to provide sufficient mAs appropriately specified. An exception to this
Program, which address both adult and pe- [74–78]. Thus, tube current, which simply trend occurs with obese patients, in which the
diatric practice. A thorough equipment qual- scales the dose rate up or down, is the prima- radiation output of the system is increased to
ity assurance program should be designed ry parameter that is used to take into consid- ensure adequate image quality. In obese pa-
and overseen by a qualified medical physi- eration patient size. tients, however, much of the additional x-ray
cist and should be supported by technolo- Tube current should be adjusted as a func- dose is absorbed by excess adipose tissue.
gists who coordinate protocol development tion of patient size based on the overall atten- Thus, doses to internal organs do not increase
and perform basic image quality checks on uation of the anatomy of interest, as opposed linearly with increases in tube current [85].
a routine basis. to patient weight, which is not a perfect sur- Automatic exposure control is a broad
rogate for attenuation [63, 64, 79]. Important- term that encompasses not only tube current
Fixed Tube Current (Technique Charts) ly, a previous evaluation of images obtained modulation to adapt to changes in attenuation
and Patient Size across a range of patient sizes showed that ra- in a specific patient, but also determining and
Unlike film-based radiographic imaging, diologists do not find the same image noise delivering the right dose for any patient (in-
a CT image never looks overexposed in the level acceptable for all patient sizes [64]. Ra- fant to obese adult) to achieve the diagnostic
sense of being too dark or too light. As a diologists tend to demand lower image noise task. Because the specific implementations of
consequence, CT users are not technical- in children relative to larger patients because automatic exposure control differ by manu-
ly compelled to decrease the tube current– children often lack adipose tissue between or- facturer, users need to take the time to under-
time product (mAs) for small patients, as is gans and tissue planes and have smaller ana- stand how their particular system functions

AJR:193, July 2009 35


McCollough et al.

and to set up the image quality reference ap-


propriately for differing diagnostic tasks (e.g.,
CTC vs routine abdomen or CT enterogra-
phy). In addition, the image noise quality re-
quirements should be adapted for pediatric
Downloaded from www.ajronline.org by 200.145.126.224 on 12/04/18 from IP address 200.145.126.224. Copyright ARRS. For personal use only; all rights reserved

and obese patients because prior studies have


shown that the same level of image noise is
not clinically accepted across large variations
in patient size [64, 74, 80, 86, 87].

Adjusting Tube Potential (kV) Based on


Patient Size
Several investigators have studied the use
of lower-tube-potential (kV) CT to improve A B
image quality or reduce radiation dose. The
Fig. 5—Lowering tube potential in smaller patients is a means of improving image contrast while
physics principle behind lower-kV imaging simultaneously reducing the radiation dose. 1-year-old male infant being treated for pelvic sarcoma who
is that the attenuation coefficient of iodine underwent two contrast-enhanced CT examinations.
increases as photon energy decreases toward A, Axial contrast-enhanced 3-mm slice at level of right portal vein obtained using 120 kV and volume CT dose
the k-edge energy of 33 keV. In CT exami- index (CTDIvol) of 3.5 mGy shows left hydronephrosis that subsequently resolved after resection of pelvic tumor
(not shown).
nations involving the use of iodinated con- B, Follow-up CT scan obtained 9 months after A. Axial contrast-enhanced 3-mm slice at level of right portal vein
trast medium, the superior enhancement of shows similar image quality using 80 kV and CTDIvol of 1.69 mGy (50% dose reduction).
iodine at lower tube potentials improves the
conspicuity of hypervascular or hypovascu- will be born healthy and not develop child- served with the use of in-plane shielding,
lar abnormalities. However, images obtained hood cancer changes by only 0.5%, from which has been shown to significantly alter
using lower tube potentials tend to be noisi- 95.93% to 95.43% [89–91]. CT attenuation accuracy [95, 96].
er, mainly because of the higher absorption Pediatric patients—When a CT examina-
of low-energy photons by the patient, unless tion is deemed to be needed in a pediatric Conclusion
the tube current is adequately adjusted [58, patient, the scanning protocol must be spe- In recent years, the media have focused
75, 77, 78, 82–84, 88]. This continues to be cifically tailored for children [92] using dose on the potential danger of radiation expo-
an area of active investigation, particularly in reduction features such as tube current mod- sure from CT while ignoring the poten-
pediatric CT (Fig. 5). ulation [74, 81], a child-specific bowtie filter tial for great individual benefit. The atten-
and scanning field of view, and a weight- or tion drawn to CT dose issues resulted in
Examples of Patient-Specific Dose size-based technique chart to determine the increased awareness in the radiology com-
Reduction Strategies appropriate kV, mAs, or both [93, 94]. For munity of the need to carefully adapt scan-
Pregnant patients—Imaging of the preg- pediatric patients, the noise level does not in- ning parameters to the individual patient and
nant patient includes the need to consider the crease significantly with the decrease of kV. examination. However, an increased level of
radiation risk to the embryo or fetus. Com- Because of this, the dose reduction or image patient anxiety has also ensued, occasionally
mon indications for CT in a pregnant pa- quality improvement from low-kV imaging resulting in a refusal to undergo a medical-
tient include suspected pulmonary embo- is more significant in pediatric patients than ly indicated examination. Reliable data are
lism, appendicitis, or urinary tract calculi. in adult patients [63, 78, 93, 94]. It is essen- not available as to the frequency of this oc-
For scanning regions outside the abdomen tial, however, that a weight- or size-based currence; however, anecdotal reports—such
and pelvis, such as for suspected pulmonary kV–mAs technique chart be used to avoid as an 84-year-old patient scheduled for aor-
embolism, the dose to the fetus is very low excessive noise levels and avoid motion arti- tic dissection repair who refused presurgical
(< 0.1 mGy) because only scattered radia- facts or long scanning times. Thus, the low- CT for planning stent placement because of
tion reaches the fetus. For all indications, the er-kV scanning protocol must be carefully “that cancer-causing stuff”—are alarming.
scanning volume should be restricted to the evaluated by radiologists and physicists for The benefits of CT—including the avoidance
necessary anatomy, and dual-pass (with and every type of pediatric examination. of exploratory or negative surgeries—can-
without contrast medium) studies should be Selective in-plane shielding—Products are not simply be dismissed in the introductory
avoided, if possible [89]. Fetal dose can be commercially available to selectively shield statements of articles that go on to empha-
reduced with careful selection of the scan- radiation-sensitive tissues and organs during size the potential risks associated with ioniz-
ning parameters (e.g., use of wider beam CT. However, investigations that consider ing radiation. Neither can examinations such
collimation, higher pitch, lower mAs, lower both image quality and dose have shown that as MRI or ultrasound be endorsed simply
kV, or reduced scanning range). It is impor- tube current modulation or simply reduced because they avoid ionizing radiation. Tru-
tant, however, to realize that for a routine bi- tube current can achieve greater dose reduc- ly mitigating the potential risk of CT means
phase CT examination of the abdomen and tions relative to in-plane shielding for the maximizing individual patient benefit, work-
pelvis, the probability of birthing a healthy same resultant image noise. In addition, de- ing with providers to ensure appropriate re-
baby decreases by only 0.1%, from 96.00% creasing the tube current does not introduce ferral, and individualizing patient prepara-
to 95.90%, and the probability that the baby the beam-hardening and streak artifacts ob- tion and CT acquisition and postprocessing

36 AJR:193, July 2009


In Defense of Body CT

techniques to maximize disease conspicuity 10. Smith AH, Hopenhayn-Rich C, Bates MN, et al. 2007; 189:1128–1134
and appropriate therapeutic decisions. Cancer risks from arsenic in drinking water. Envi- 24. Curley SA. Outcomes after surgical treatment of
Dose reduction strategies described in this ron Health Perspect 1992; 97:259–267 colorectal cancer liver metastases. Semin Oncol
article must be well understood and properly 11. Subcommittee on Arsenic in Drinking Water, Na- 2005; 32[6 suppl 9]:S109–S111
used, but also require broad-based practice tional Research Council. Arsenic in drinking wa- 25. Sofocleous CT, Nascimento RG, Gonen M, et al.
Downloaded from www.ajronline.org by 200.145.126.224 on 12/04/18 from IP address 200.145.126.224. Copyright ARRS. For personal use only; all rights reserved

strategies that extend beyond the CT scan- ter. Washington, DC: National Academies Press, Radiofrequency ablation in the management of
ner console and default, generic manufactur- 1999 liver metastases from breast cancer. AJR 2007;
er settings. In the final analysis, physicians 12. Muirhead CR, O’Hagan JA, Haylock RG, et al. 189:883–889
must request the imaging examination that Mortality and cancer incidence following occupa- 26. Wolpin BM, Mayer RJ. Systemic treatment of col-
best addresses the specific medical question tional radiation exposure: third analysis of the orectal cancer. Gastroenterology 2008; 134: 1296–
without allowing worries about radiation to National Registry for Radiation Workers. Br J 1310
dissuade them or their patients from obtain- Cancer 2009; 100:206–212 27. Francis IR, Cohan RH, McNulty NJ, et al. Multi-
ing or undergoing the needed CT examina- 13. Committee to Assess Health Risks from Exposure detector CT of the liver and hepatic neoplasms:
tions. Ongoing efforts to ensure that CT ex- to Low Level of Ionizing Radiation, National Re- effect of multiphasic imaging on tumor conspicu-
aminations are both medically justified and search Council. Health risks from exposure to low ity and vascular enhancement. AJR 2003; 180:
optimally performed must continue, and levels of ionizing radiation: BEIR VII Phase 2. 1217–1224
education must be provided to the medical Washington, DC: National Academies Press, 2006 28. Sheafor DH, Frederick MG, Paulson EK, Keogan
community and general public that put both 14. National Council on Radiation Protection and MT, DeLong DM, Nelson RC. Comparison of un-
the potential risks—and benefits—of CT ex- Measurements. Ionizing radiation exposure of the enhanced, hepatic arterial-dominant, and portal
aminations in proper perspective. population of the United States. Bethesda, MD: venous–dominant phase helical CT for the detec-
National Council on Radiation Protection and tion of liver metastases in women with breast car-
Acknowledgments Measurements, 2009: report no. 160 cinoma. AJR 1999; 172:961–968
We thank Kris Nunez for her help with 15. Health Physics Society Website. Radiation risk in 29. Miller FH, Butler RS, Hoff FL, Fitzgerald SW,
manuscript preparation and Jeremy McBride perspective: position statement of the Health Nemcek AA Jr, Gore RM. Using triphasic helical
for his suggestions. Physics Society. www.hps.org/documents/radia- CT to detect focal hepatic lesions in patients with
tionrisk.pdf. Adopted 1996. Revised 2004. Ac- neoplasms. AJR 1998; 171:643–649
References cessed April 10, 2009 30. Silverman PM. Liver metastases: imaging consid-
1. Brenner DJ, Hall EJ. Computed tomography: an 16. The Royal College of Radiologists. Making the erations for protocol development with multislice
increasing source of radiation exposure. N Engl J best use of clinical radiology services: referral CT (MSCT). Cancer Imaging 2006; 6:175–181
Med 2007; 357:2277–2284 guidelines, 6th ed. London, UK: The Royal Col- 31. Fletcher JG, Wiersema MJ, Farrell MA, et al.
2. Brenner DJ, Elliston CD, Hall EJ, Berdon WE. lege of Radiologists, 2007 Pancreatic malignancy: value of arterial, pancre-
Estimated risks of radiation-induced fatal cancer 17. American College of Radiology Website. ACR atic, and hepatic phase imaging with multi-detec-
from pediatric CT. AJR 2001; 176:289–296 Appropriateness Criteria: October 2008 version. tor row CT. Radiology 2003; 229:81–90
3. Rogers LF. Radiation exposure in CT: why so www.acr.org/SecondaryMainMenuCategories/ 32. Kawata S, Murakami T, Kim T, et al. Multidetec-
high? (editorial) AJR 2001; 177:277 quality_safety/app_criteria.aspx. Published Octo- tor CT: diagnostic impact of slice thickness on
4. National Council on Radiation Protection and ber 2008. Accessed March 9, 2009 detection of hypervascular hepatocellular carci-
Measurements. Ionizing radiation exposure of the 18. American College of Radiology Website. Gastro- noma. AJR 2002; 179:61–66
population of the United States. Bethesda, MD: intestinal imaging. www.acr.org/SecondaryMain- 33. Kim S, Ha H, Park S, et al. Gastrointestinal tract
National Council on Radiation Protection and MenuCategories/quality_safety/app_criteria/pdf/ metastasis from primary lung cancer: CT findings
Measurements, 1987: Report no. 93 ExpertPanelonGastrointestinalImaging.aspx. Ac- and clinicopathologic features. AJR 2009 (in press)
5. Committee on Passive Smoking, National Re- cessed March 9, 2009 34. Obuchowski NA, Graham RJ, Baker ME, Powell
search Council. Environmental tobacco smoke: 19. Frager D, Medwid SW, Baer JW, Mollinelli B, KA. Ten criteria for effective screening: their ap-
measuring exposures and assessing health effects. Friedman M. CT of small-bowel obstruction: val- plication to multislice CT screening for pulmo-
Washington, DC: National Academies Press, 1986 ue in establishing the diagnosis and determining nary and colorectal cancers. AJR 2001; 176:
6. U.S. Food and Drug Administration Website. the degree and cause. AJR 1994; 162:37–41 1357–1362
Whole body scanning using computed tomogra- 20. Maglinte DD, Gage SN, Harmon BH, et al. Ob- 35. Johnson CD, Chen M-H, Toledano AY, et al. Ac-
phy (CT). www.fda.gov/cdrh/ct/risks.html. Pub- struction of the small intestine: accuracy and role curacy of CT colonography for detection of large
lished 2002. Accessed March 9, 2009 of CT in diagnosis. Radiology 1993; 188:61–64 adenomas and cancers. N Engl J Med 2008;
7. Jemal A, Siegel R, Ward E, et al. Cancer statistics, 21. Megibow AJ, Balthazar EJ, Cho KC, Medwid SW, 359:1207–1217
2008. CA Cancer J Clin 2008; 58:71–96 Birnbaum BA, Noz ME. Bowel obstruction: eval- 36. Graser A, Stieber P, Nagel D, et al. Comparison of
8. National Safety Council Website. Odds of death uation with CT. Radiology 1991; 180:313–318 CT colonography, colonoscopy, sigmoidoscopy
due to injury, United States, 2005. www.nsc.org/ 22. Beall DP, Fortman BJ, Lawler BC, Regan F. Im- and faecal occult blood tests for the detection of
research/odds.aspx. Published 2005. Accessed aging bowel obstruction: a comparison between advanced adenoma in an average risk population.
March 9, 2009 fast magnetic resonance imaging and helical com- Gut 2009; 58:241–248
9. Nero AV, Schwehr MB, Nazaroff WW, Revzan puted tomography. Clin Radiol 2002; 57:719–724 37. Pickhardt PJ, Choi JR, Hwang I, et al. Computed
KL. Distribution of airborne radon-222 concen- 23. Jaffe TA, Miller CM, Merkle EM. Practice pat- tomographic virtual colonoscopy to screen for
trations in U.S. homes. Science 1986; 234:992– terns in imaging of the pregnant patient with ab- colorectal neoplasia in asymptomatic adults. N
997 dominal pain: a survey of academic centers. AJR Engl J Med 2003; 349:2191–2200

AJR:193, July 2009 37


McCollough et al.

38. Levin B, Lieberman DA, McFarland B, et al.; 51. Sistrom CL, Dang PA, Weilburg JB, Dreyer KJ, Standardized methods for measuring diagnostic
American Cancer Society Colorectal Cancer Ad- Rosenthal DI, Thrall JH. Effect of computerized x-ray exposures. New York, NY: AAPM, 1990
visory Group, US Multi-Society Task Force, and order entry with integrated decision support on 66. Jucius RA, Kambic GX. Radiation dosimetry in
American College of Radiology Colon Cancer the growth of outpatient procedure volumes: sev- computed tomography: application of optical in-
Committee. Screening and surveillance for the en-year time series analysis. Radiology 2009; strumentation in medicine. VI. Proc Soc Photo
Downloaded from www.ajronline.org by 200.145.126.224 on 12/04/18 from IP address 200.145.126.224. Copyright ARRS. For personal use only; all rights reserved

early detection of colorectal cancer and adeno- 251:147–155 Opt Instrum Eng 1977; 127:286–295
matous polyps, 2008: a joint guideline from the 52. Rosenthal DI, Weilburg JB, Schultz T, et al. Radi- 67. Shope TB, Gagne RM, Johnson GC. A method for
American Cancer Society, the US Multi-Society ology order entry with decision support: initial describing the doses delivered by transmission x-
Task Force on Colorectal Cancer, and the Ameri- clinical experience. J Am Coll Radiol 2006; 3: ray computed tomography. Med Phys 1981; 8:
can College of Radiology. CA Cancer J Clin 2008; 799–806 488–495
58:130–160 53. Hillman BJ, Joseph CA, Mabry MR, Sunshine 68. American Association of Physicists in Medicine.
39. Brenner DJ, Georgsson MA. Mass screening with JH, Kennedy SD, Noether M. Frequency and costs The measurement, reporting, and management of
CT colonography: should the radiation exposure of diagnostic imaging in office practice: a com- radiation dose in CT: report of AAPM Task Group
be of concern? Gastroenterology 2005; 129:328– parison of self-referring and radiologist-referring 23 of the Diagnostic Imaging Council CT Com-
337 physicians. N Engl J Med 1990; 323:1604–1608 mittee. College Park, MD: AAPM, 2008: AAPM
40. Lieberman D, Moravec M, Holub J, Michaels L, 54. Hillman BJ, Olson GT, Griffith PE, et al. Physi- report 96
Eisen G. Polyp size and advanced histology in pa- cians’ utilization and charges for outpatient diag- 69. International Electrotechnical Commission. Medi-
tients undergoing colonoscopy screening: impli- nostic imaging in a Medicare population. JAMA cal Electrical Equipment. Particular requirements
cations for CT colonography. Gastroenterology 1992; 268:2050–2054 for the safety of x-ray equipment for computed to-
2008; 135:1100–1105 55. Studdert DM, Mello MM, Sage WM, et al. Defen- mography. Geneva, Switzerland: Inter­national
41. Kim DH, Pickhardt PJ, Taylor AJ. Characteristics sive medicine among high-risk specialist physi- Electrotechnical Commission (IEC) Central Of-
of advanced adenomas detected at CT colono- cians in a volatile malpractice environment. fice, 2002: IEC publication no. 60601-2-44 Ed.
graphic screening: implications for appropriate JAMA 2005; 293:2609–2617 2.1; part 2-44
polyp size thresholds for polypectomy versus sur- 56. Dunnick NR, Applegate KE, Arenson RL. The 70. McNitt-Gray MF. AAPM/RSNA physics tutorial
veillance. AJR 2007; 188:940–944 inappropriate use of imaging studies: a report of for residents: topics in CT—radiation dose in CT.
42. Limburg PJ, Fletcher JG. Making sense of CT the 2004 Intersociety Conference. J Am Coll Ra- RadioGraphics 2002; 22:1541–1553
colonography–related complication rates. Gastro- diol 2005; 2:401–406 71. McCollough CH, Zink FE, Morin RL. Radiation
enterology 2006;131:2023-2024; discussion, 2024 57. McCollough CH, Primak AN, Braun N, Kofler J, dosimetry for electron beam CT. Radiology 1994;
43. Pickhardt PJ. Incidence of colonic perforation at Yu L, Christner J. Strategies for reducing radia- 192:637–643
CT colonography: review of existing data and im- tion dose in CT. Radiol Clin North Am 2009; 72. European Commission. European guidelines on
plications for screening of asymptomatic adults. 47:27–40 quality criteria for computed tomography (EUR
Radiology 2006; 239:313–316 58. Food and Drug Administration. FDA public 16262 EN). Luxembourg: European Commission,
44. Xiong T, Richardson M, Woodroffe R, Halligan health notification: reducing radiation risk from 2000
S, Morton D, Lilford RJ. Incidental lesions found computed tomography for pediatric and small 73. McCollough C, Branham T, Herlihy V, et al. Ra-
on CT colonography: their nature and frequency. adult patients. Pediatr Radiol 2002; 32:314–316 diation doses from the ACR CT accreditation pro-
Br J Radiol 2005; 78:22–29 59. Frush DP, Donnelly LF, Rosen NS. Computed to- gram: review of data since program inception
45. Hassan C, Pickhardt PJ, Laghi A, et al. Computed mography and radiation risks: what pediatric and proposals for new reference values and pass/
tomographic colonography to screen for colo- health care providers should know. Pediatrics fail limits. Oak Brook, IL: RSNA, 2006
rectal cancer, extracolonic cancer, and aortic an- 2003; 112:951–957 74. McCollough CH, Bruesewitz MR, Kofler JM Jr.
eurysm: model simulation with cost-effectiveness 60. Task Group on Control of Radiation Dose in CT dose reduction and dose management tools:
analysis. Arch Intern Med 2008; 168:696–705 Computed Tomography. Managing patient dose in overview of available options. RadioGraphics
46. Gluecker TM, Johnson CD, Wilson LA, et al. Ex- computed tomography: a report of the Interna- 2006; 26:503–512
tracolonic findings at CT colonography: evalua- tional Commission on Radiological Protection. 75. Funama Y, Awai K, Nakayama Y, et al. Radiation
tion of prevalence and cost in a screening popula- Ann ICRP 2000; 30:7–45 dose reduction without degradation of low-con-
tion. Gastroenterology 2003; 124:911–916 61. Kalra MK, Maher MM, Toth TL, et al. Strategies trast detectability at abdominal multisection CT
47. Pickhardt PJ, Hanson ME, Vanness DJ, et al. Un- for CT radiation dose optimization. Radiology with a low-tube voltage technique: phantom study.
suspected extracolonic findings at screening CT 2004; 230:619–628 Radiology 2005; 237:905–910
colonography: clinical and economic impact. Ra- 62. Linton OW, Mettler FA Jr. National conference on 76. Huda W, Ravenel JG, Scalzetti EM. How do ra-
diology 2008; 249:151–159 dose reduction in CT, with an emphasis on pediat- diographic techniques affect image quality and
48. Berland LL. Incidental extracolonic findings on ric patients. AJR 2003; 181:321–329 patient doses in CT? Semin Ultrasound CT MR
CT colonography: the impending deluge and its 63. Boone JM, Geraghty EM, Seibert JA, Wootton- 2002; 23:411–422
implications. J Am Coll Radiol 2009; 6:14–20 Gorges SL. Dose reduction in pediatric CT: a ra- 77. Nakayama Y, Awai K, Funama Y, et al. Abdomi-
49. American College of Radiology, National Radiol- tional approach. Radiology 2003; 228:352–360 nal CT with low tube voltage: preliminary obser-
ogy Data Registry Website. CT colonography reg- 64. Wilting JE, Zwartkruis A, van Leeuwen MS, vations about radiation dose, contrast en-
istry. https://nrdr.acr.org/portal/CTC/Main/page. Timmer J, Kamphuis AG, Feldberg M. A rational hancement, image quality, and noise. Radiology
aspx. Accessed March 9, 2009 approach to dose reduction in CT: individualized 2005; 237:945–951
50. Blackmore CC. Defining quality in radiology. J scan protocols. Eur Radiol 2001; 11:2627–2632 78. Siegel MJ, Schmidt B, Bradley D, Suess C, Hilde-
Am Coll Radiol 2007; 4:217–223 65. American Association of Physicists in Medicine. bolt C. Radiation dose and image quality in pedi-

38 AJR:193, July 2009


In Defense of Body CT

atric CT: effect of technical factors and phantom gen, Germany: European Journal of Medical 92. Frush DP, Donnelly LF. Helical CT in children:
size and shape. Radiology 2004; 233:515–522 Physics, 2002:43–53 technical considerations and body applications.
79. McCollough CH, Zink FE, Kofler J, Matsumoto 86. Kalra MK, Maher MM, D’Souza RV, et al. Detec- Radiology 1998; 209:37–48
JS, Thomas KB, Hoffman AD. Dose optimization tion of urinary tract stones at low-radiation-dose 93. Hollingsworth C, Frush DP, Cross M, Lucaya J. He-
in CT: creation, implementation and clinical ac- CT with z-axis automatic tube current modula- lical CT of the body: a survey of techniques used
Downloaded from www.ajronline.org by 200.145.126.224 on 12/04/18 from IP address 200.145.126.224. Copyright ARRS. For personal use only; all rights reserved

ceptance of size-based technique charts. (abstr) tion: phantom and clinical studies. Radiology for pediatric patients. AJR 2003; 180:401–406
Radiology 2002;225(P):591 2005; 235:523–529 94. Paterson A, Frush DP. Dose reduction in paediat-
80. Kalra MK, Maher MM, Toth TL, et al. Techniques 87. Nyman U, Ahl TL, Kristiansson M, Nilsson L, ric MDCT: general principles. Clin Radiol 2007;
and applications of automatic tube current modu- Wettemark S. Patient-circumference-adapted dose 62:507–517
lation for CT. Radiology 2004; 233:649–657 regulation in body computed tomography: a practi- 95. Geleijns J, Salvadó Artells M, Veldkamp WJ, Ló-
81. Gies M, Kalender WA, Wolf H, Suess C, Madsen cal and flexible formula. Acta Radiol 2005; pez Tortosa M, Calzado Cantera A. Quantitative
M. Dose reduction in CT by anatomically adapted 46:396–406 assessment of selective in-plane shielding of tis-
tube current modulation: simulation studies. Med 88. Huda W, Scalzetti EM, Levin G. Technique fac- sues in computed tomography through evaluation
Phys 1999; 26:2235–2247 tors and image quality as functions of patient of absorbed dose and image quality. Eur Radiol
82. Haaga JR, Miraldi F, MacIntyre W, LiPuma JP, weight at abdominal CT. Radiology 2000; 217: 2006; 16:2334–2340
Bryan PJ, Wiesen E. The effect of mAs variation 430–435 96. Leswick DA, Hunt MM, Webster ST, Fladeland
upon computed tomography image quality as 89. Wagner LK, Huda W. When a pregnant woman DA. Thyroid shields versus z-axis automatic tube
evaluated by in vivo and in vitro studies. Radiolo- with suspected appendicitis is referred for a CT current modulation for dose reduction at neck CT.
gy 1981; 138:449–454 scan, what should a radiologist do to minimize Radiology 2008; 249:572–580
83. Kalender WA, Wolf H, Suess C. Dose reduction in potential radiation risks? Pediatr Radiol 2004; 97. Gerber TC, Carr JJ, Arai AE, et al. Ionizing radia-
CT by anatomically adapted tube current modula- 34:589–590 tion in cardiac imaging: a science advisory from
tion: phantom measurements. Med Phys 1999; 90. McCollough CH, Schueler BA, Atwell TD, et al. the American Heart Association Committee on
26:2248–2253 Radiation exposure and pregnancy: when should Cardiac Imaging of the Council on Clinical Car-
84. McCollough CH. Automatic exposure control in we be concerned? RadioGraphics 2007; 27:909– diology and Committee on Cardiovascular Imag-
CT: are we done yet? Radiology 2005; 237:755–756 917; discussion, 917–908 ing and Intervention of the Council on Cardiovas-
85. Schmidt B, Kalender WA. A fast voxel-based 91. Wagner LK, Lester RG, Saldana LR. Exposure of cular Radiology and Intervention. Circulation
Monte Carlo method for scanner- and patient- the pregnant patient to diagnostic radiations: a 2009; 119:1056–1065
specific dose calculations in computed tomogra- guide to medical management. Madison, WI: 98. McCollough CH. CT dose: how to measure, how
phy. In: Guerra AD, ed. Physica medica. Erlan- Medical Physics Publishing, 1997 to reduce. Health Phys 2008; 95:508–517

AJR:193, July 2009 39

You might also like