You are on page 1of 6

ARTICLE IN PRESS

American Journal of Infection Control 000 (2020) 1−6

Contents lists available at ScienceDirect

American Journal of Infection Control


journal homepage: www.ajicjournal.org

Major article

Impact of air-conditioner outlet layout on the upward airflow induced by


forced air warming in operating rooms
Kazuhiro Shirozu MD, PhD a,*, Hidekazu Setoguchi MD, PhD b, Kenzo Araki MD c, Taichi Ando MD d,
Ken Yamaura MD, PhD e
a
Department of Anesthesiology and Critical Care Medicine, Kyushu University Hospital, Fukuoka, Japan
b
Operating Rooms, Kyushu University Hospital, Fukuoka, Japan
c
Department of Anesthesiology, Tagawa municipal hospital, Kyushu University Hospital, Fukuoka, Japan
d
Department of Anesthesiology and Critical Care Medicine, Kyushu University Hospital, Fukuoka, Japan
e
Department of Anesthesiology and Critical Care Medicine, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan

Key Words: A B S T R A C T
Forced air warming system
Cleanliness Background: Previously, we found that an upward air current in the head area, induced by forced air warm-
Particle counter
ing (FAW), was completely counteracted by downward laminar airflow. However, this study did not include
Surgical site infection
any consideration of the air-conditioner outlet layout (ACOL); hence, its impact remains unclear.
Anesthesia
Laminar airflow Methods: This study was performed in 2 operating rooms (ORs)—ISO classes 5 and 6, which are denoted as
OR-5 and OR-6, respectively. Both ORs have distinct ACOLs. The cleanliness, or the number or ratio of shifting
artificial particles was evaluated.
Results: During the first 5 minutes after particles generation, significantly more particles shifted into the sur-
gical field in OR-5 when compared to OR-6 (13,587 [4,341-15,913] and 106 [41-338] particles/cubic foot, P <
.0001). Notably, FAW did not increase the number of shifting particles in OR-6. The laminar airflow system
fully counteracted the upward airflow caused by FAW in OR-6, where the ACOL covered the operating bed.
However, this did not occur in OR-5, where the ACOL did not fully cover the operating bed.
Conclusions: Regardless of cleanliness ability of OR, an ACOL that fully covers the operating bed can prevent
upward airflow in the head area and reduce the number of artificial particles shifting into the surgical field,
which are typically caused by FAW.
© 2020 Association for Professionals in Infection Control and Epidemiology, Inc. Published by Elsevier Inc. All
rights reserved.

BACKGROUND However, the previous study had the following limitations: (1) the
air-conditioner outlet in the operating room fully covered the operat-
Actively warming patients during surgery, by methods such as ing bed, including the head area; (2) the FAW temperature was set at
forced air warming (FAW), provides clear benefits.1,2 On the other 38°C, so the results may not be applicable to relevant higher temper-
hand, FAW generates upward airflow in the head area, which poten- atures; and (3) the surgical lights were moved away from the operat-
tially causes contamination. We have previously reported that the ing field.
presence of downward laminar airflow (LAF) completely counteracts As compared to conductive fabric, Kumar et al. reported that an
this, which enables FAW to be used with negligible contamination.3 FAW system set at 43°C significantly mobilized artificial bubbles that
were generated in head area, to move over the anesthesia drape, and
into the surgical field.4 However, this study did not mention the air-
* Address correspondence to Kazuhiro Shirozu, MD, PhD, Lecturer, Department of conditioner outlet layout (ACOL) in the operating room, which has
Anesthesiology and Critical Care Medicine, Kyushu University Hospital, 3-1-1 Maida-
recently been highlighted as an area of great importance in such
shi, Higashi-ku, Fukuoka 812-8582, Japan.
E-mail address: shiron@kuaccm.med.kyushu-u.ac.jp (K. Shirozu). studies. In one study, fewer microbes were detected in the sterile
Conflicts of interest: None to report.Contribution: All authors helped review the field of an operating room (OR) when a single large diffuser was
literature and write the manuscript, and approved the final version of the manuscript. used, as compared to using a multi-diffuser array or 4-way throw dif-
Funding: The authors received support from an educational research fund of the Oper-
fuser.5 In addition, Kumar et al. reported that an FAW set at a higher
ating Rooms in Kyushu University Hospital.

https://doi.org/10.1016/j.ajic.2020.06.202
0196-6553/© 2020 Association for Professionals in Infection Control and Epidemiology, Inc. Published by Elsevier Inc. All rights reserved.
ARTICLE IN PRESS
2 K. Shirozu et al. / American Journal of Infection Control 00 (2020) 1−6

temperature (43°C) disrupted the ventilation airflow over the surgi- FAW device, with the warming temperature set to 43°C. The air-con-
cal site.4 Refaie et al. found that surgical lights could further disturb ditioning temperature in the OR was set to 25°C.
LAF.6 Given these previous studies, we hypothesized that surgical
field contamination was affected by the temperature setting of the Particles shifting from the head area into the surgical field
FAW, the OR ACOL, and the presence and location of any surgical
lights. Particles were counted in the surgical field based on general sur-
The specific purpose of this study was to examine the effects of gery use assumptions, which included holding the head area in paral-
ACOL on the possibility of contamination into surgical field induced lel and using 2 particle counters (AeroTrak 9110, TSI Incorporated,
by FAW, at the highest relevant temperature setting, 43°C, and in the Shoreview, MN) at 30-second intervals for a total of 8 minutes. Suc-
presence of a surgical light. We compared the FAW effects between tion sampling was performed at 1 cubic foot/min to measure the
class 5 (352 < maximum particles (size 30.5 mm)/m3 < 3,520, former quantity of dust particles with a grain size of at least 0.5 mm. A fabric
classification class 100) and class 6 (3,520 < maximum particles (size cover was used above the under-warming blanket to investigate the
3 0.5 mm)/m3 < 35,200, former classification class 1,000) ORs with prevention of the direct upward airflow caused by FAW, as well as
different ACOLs (Table 1). The primary objective was to investigate any subsequent affects.
the effect of ACOL on the number or ratio of artificially generated par-
ticles that were shifted from the head area into the surgical field by
Three-dimensional measurements of airflow direction and speed
FAW. Additionally, we investigated how this particle movement was
affected by placing a cover above the under-warming blanket.
The airflow caused by FAW or LAF, either in the presence or
absence of surgical lights above the operating table, were evaluated
by a 3-dimensional (3D) measurement of airflow direction and speed.
METHODS
A 3D ultrasonic anemometer (WA-790; Sonic Corporation, Tokyo,
Japan) was used to measure the direction and speed of the airflow
Experimental operating room circumstance
within a total of 168 points with 300 mm intervals. These intervals
consist of 7 points on the x-axis, 6 points on the y-axis (with the head
This was a simulated study on manikins and not on human sub-
located at point Y6), and 4 points on the z-axis. The axes and mea-
jects. The study was performed in 2 ORs at Kyushu University Hospi-
surement points are defined and denoted in Fig. 1c and d.
tal, Fukuoka, Japan, where all ORs are equipped with a unidirectional
vertical LAF system (Shinko Air Handling Unit; Shinko Industrial Co.
Ltd., Osaka, Japan). The ORs used in this study were designed as ISO Cleanliness assessment
class 5 or 6, denoted as OR-5 or OR-6, respectively. The OR-5 used
had a height of 3 m, a floor space of 48.36 m2, and receives an air sup- To investigate the OR cleanliness, a cleanliness recovery test was
ply of 33,131 m3/h, resulting in 228 air changes per hour. This large performed in which dust was artificially generated in the surgical
ventilation volume enables the OR-5 to achieve high levels of cleanli- field, with the air-conditioner turned off. The dust removal process
ness. However, the large ACOL in this room dose not fully cover the was then subsequently assessed after the air-conditioner was turned
operating table (Fig 1a and Table 1). In contrast, the OR-6 used had a on.7 Using a particle counter (KC-03A; RION Co., Ltd., Tokyo, Japan),
height of 3 m, a floor space of 34.22 m2, and receives an air supply of suction sampling was performed at 1 cubic foot/min to measure the
10,102 m3/h, resulting in 98 air changes per hour. Here, the wide air- quantity of dust particles at least 0.5 mm in size. Either the air-condi-
conditioner outlet covers the entire operating table (Fig 1b and tioner or FAW was turned on with an initial dust particle generation,
Table 1). The air-conditioning outlet in each OR contains several pan- and changes in the dust particle quantity were measured at 1-minute
els, which includes a final point-of-use high-efficiency particulate air intervals until the particle concentration decreased below 10/cubic
filter on each panel. foot. A total of 6 measurement points were taken at the centre and
An upper-body manikin was placed in a supine position on the edges of the operating table in the Y3 cross-section (Supplemental 1a
operating table and covered with a surgical drape with an under- and b), at heights of 800 and 1,500 mm.
warming blanket placed underneath it. Artificial particles were gen-
erated at the head area of the manikin by rubbing a typical surgical Statistical analysis
towel 5 times, for a total of about 2 seconds. An anesthesia drape was
clipped to intravenous poles and raised 150 cm above the floor, and Further analysis was needed to review the number of particles
the height of the operating bed was set as 100 cm above the floor. A shifting from the head area into the surgical field under FAW, within
3M Bair Hugger (Model 750; 3M Company, Maplewood, MN) and 5 minutes of generation. Power analysis (a = 0.05, b = 0.20) indicated
under-warming blanket (Model 585; 3M Company) comprised the that a subject sample size fewer than 3 per group was needed to

Table 1
Comparison of basic capability between OR-5 and OR-6

ISO classification Class 5 OR Class 6 OR

Former classification Class 100 Class 1,000


Cleanliness 352 < maximum particles 3,520 < maximum particles
(size 3 0.5 mm)/m3 < 3,520 (size 3 0.5 mm)/m3< 35,200
Height (m) 3 3
floor space (m2) 48.36 34.22
air supply (m3/h) 33,131 10,102
air change (per hour) 228 98
ACOL Non-fully covering Entirely covering the operating table
the operating table
ACOL, air-conditioner outlet layout.
ARTICLE IN PRESS
K. Shirozu et al. / American Journal of Infection Control 00 (2020) 1−6 3

Fig 1. Experimental set-up and airflow measurement points. (a) Top view of the OR-5, with the patterned portion indicating the air-conditioner outlet. (b) Top view of the OR-6,
with the patterned portion indicating the air-conditioner outlet. (c) Top view illustrating the measurement points. The floor was marked with tape every 300 mm. There were 7
measuring points on the x-axis (X1-X7; X4 was the centre), and 6 measuring points on the y-axis (Y1-Y6). (d) The view from the foot of the operating bed. There were 4 measuring
points on the z-axis (800, 1100, 1400, and 2900 mm from the floor). All measuring points were 300 mm apart.

detect a significant difference between OR-5 and OR-6. The Data is Particle shifting in OR-5 and OR-6
presented as median (IQR[range]) or mean § standard deviation.
Additionally, Tukey or Sidak’s multiple comparison post hoc test was As aforementioned, there was no significant difference in the ini-
used in a 2-way analysis of variance (Fig 2, and Supplements 1 and tial number of generated particles in the head area in OR-5 and OR-6
2). All statistical analyses were performed using Prism 6 software either with or without FAW (Table 2). During the first 30 seconds
(GraphPad Software, La Jolla, CA), with P < .05 considered to be statis- after particle generation under FAW working, 5,063 [2,449-9,523]
tically significant. and 8 [3-16] particles/ cubic foot (P < .0001) shifted in OR-5 and OR-
6, respectively. Between 30-60 seconds, 4,140 [1,039-5,155] and 10
[2-14] particles/cubic foot (P < .0001) shifted in OR-5 and OR-6 under
FAW working, respectively (Fig 2a and Table 2). After a total of 5
RESULTS minutes after particle generation, OR-5 and OR-6 exhibited 13,587
[4,341-15,913] and 106 [41-338] (P < .0001) shifted particles/ cubic
Cleanliness in OR-5 and OR-6 foot. In the first 30 seconds after particle generation without FAW
(Fig 2a), 54 [11-116] and 4 [2-6] particles/ cubic foot (P = .99) shifted
Before the air-conditioner was activated, there was no significant in OR-5 and OR-6, respectively. As a check, these results were con-
difference in the number of particles per cubic foot between OR-5 firmed to be quite similar to the shifting ratios, as defined by the par-
and OR-6, regardless of whether FAW was implemented or not (not ticles found in the surgical field at each time point to those in the
shown). The duration between activating the air-conditioner and head area at 30 seconds x 100 (Supplement 2a).
achieving a particle count below 10/ cubic foot was significantly
shorter in OR-5 (3.3 § 0.5 minute) than in OR-6 (6.3 § 1.6 minute) Particle shifting with FAW
without FAW (Supplement 1). The same finding was obtained when
FAW was used in OR-5 and OR-6 by achieving less than 10 particles/ Utilizing FAW in OR-5 increased the number of particles that
cubic foot in 3.0 § 0.0 and 7.7 § 0.8 minutes, respectively. Notably, shifted during the first 30 seconds (5,063 [2,449-9,523] and 54 [11-
there was also no significant difference in the times obtained (Sup- 116] / cubic foot, P < .0001, with and without FAW, respectively) or
plement 1). 30-60 seconds (4,140 [1,039-5,155] and 50 [23-72]/cubic foot, P <
ARTICLE IN PRESS
4 K. Shirozu et al. / American Journal of Infection Control 00 (2020) 1−6

Fig 2. The number of particles shifting into surgical field. (a) A comparison of the number of particles shifting into surgical field from the head area between OR-5 and OR-6, with or
without FAW. Blue or brown represent the cases of the OR-5 or OR-6. Triangles or circles represent the cases of the with or without FAW. (b) The effect of covering the under-warm-
ing blanket on the number of particles shifting in OR-5. Red or purple represent the cases of the without or with FAW. (c) The view of covering above the warming-blanket. The
number of samples in each group was 10. Triangles or Circles indicate the median values. ****P < .0001.

.0001, with and without FAW, respectively) after particle generation, 5 minutes (not shown). Furthermore, there was no significant differ-
as well as the total particles that shifted in 5 minutes (13,587 [4,341- ence in the initial number of particles generated in the head area in
15,913] and 167 [75-349]/cubic foot, P < .0001, with and without OR-5, with or without the use of a covering (Fig 2b). These results
FAW, respectively) into surgical field (Fig 2a and Table 2). In using were also confirmed by the shifting ratio results (Supplement 2b).
FAW in OR-6, the number of particles shifted into the surgical field
did not significantly increase during 0-30, 30-60, or 60-90 seconds 3D airflow direction and speed in OR-5 or 6 under FAW working with
and in 5 minutes after particle generation, which was 106 [41-338] LAF without surgical light
and 69 [52-99] / cubic foot, P = .47, with and without FAW, respec-
tively (Fig 2a and Table 2). These results are also similar to those for From the feet side at Y3 cross section, LAF caused homogenous
the shifting ratio shown in Supplement 2a. Notably, there is no signif- downward airflow (»56 cm/s) around the operating table in OR-6,
icant difference in the initial number of generated particles in the but not in OR-5 (Fig 3).
head area with or without FAW in OR 5 or 6.
3D airflow direction and speed in OR-5 or OR-6 under FAW working
Particle shifting under FAW working with a covering above the under- with surgical light
warming blanket
From the right side at X4 cross section, FAW did not cause airflow
Generally, utilizing a physical cover above the under-warming above the surgical drape or surgical field in OR-5 and OR-6. In head
blanket did not affect in the number of particles in the 0-30, 30-60, or area, upward airflow was detected both in OR-5 (»36 cm/s) and OR-6
60-90 seconds intervals after particle generation. It also did not affect (»39 cm/s) without LAF (Fig 4, upper and supplement 3). In OR-5,
the total particles counted in the surgical field of OR-5 within the first upward airflow (»37 cm/s) in the head area remained under LAF. In

Table 2
Comparison of initial generated number in head area or shifting particle number into surgical field between OR-5 and OR-6

FAW Class 5 OR Class 6 OR

+ +

Initial generated number in head area 77,912 [44,926-92,766] 53,956 [43,818-95,696] 68,400 [23,396-10,1857] 86,823 [60,682-96,492]
Total number during 5 min in surgical field 13,587 [4,341-15,913] 167 [74.5-349] 106 [41-338] 69 [52-99]
During the first 30 s 5,063 [2,449-9,523] 54 [11-116] 8 [3-16] 4 [2-6]
Between 30-60 s 4,140 [1,039-5,155] 50 [23-72] 10 [2-14] 7 [0-12]
NOTE. Median (IQR[range]), number/cubic foot.
ARTICLE IN PRESS
K. Shirozu et al. / American Journal of Infection Control 00 (2020) 1−6 5

Fig 3. Airflow direction and speed at the Y3 cross section. Airflow direction and speed at the Y3 cross section in OR-5 (left) or OR-6 (right) under FAW working with LAF without sur-
gical lights. Airflow speeds are represented by the numeric values enclosed in the squares and by the length and colour of the arrows. The arrows in the figures indicated the mean
direction and speed for 10 s.

OR-6, upward airflow in the head area was completely counteracted fact, even when FAW was introduced, the number of particles shifting
by LAF (downward airflow: »55 cm/s) (Fig 4, lower and supple- did not increase when the air-conditioner outlet fully covered the
ment 4). operating bed (Fig 2a).
Additionally, we hypothesized that upward airflow, induced by
DISCUSSION FAW, could be prevented by covering the under-warming blanket,
which would inhibit direct flow. However, this had little actual effect
To our knowledge, the ACOL effects on upward airflow induced by on the number of shifting particles (Fig 2b). Thus, indicating that
FAW has not been previously investigated. Generally, in an OR with a upward airflow around the head area might be due primarily to the
wide air-conditioner outlet that fully covers the operating bed, signif- discharge of warmed air underneath the surgical drape, rather than
icantly fewer particles shifted into the surgical field from the head direct airflow from the small pores of the under-warming blanket. It
area when compared with an OR with an air-conditioner outlet that is also possible that covering the under-warming blanket was insuffi-
did not fully cover the operating bed, despite extremely high air cient in this work because the fabric cover used had air permeability.
exchange times (Fig 2a and Table 1). Our results indicate that OR The investigation of airflow and speed (Fig. 3 and 4) clearly
cleanliness is determined primarily by the amount of ventilation air revealed why the shifting particle shifting was so dramatically
and OR-5 is superior to OR-6 (Supplement 1). However, ACOL is more affected by ACOL. Specifically, air-conditioner outlets that fully cov-
important for preventing particle shifting that is induced by FAW. In ered the operating table, such as that in OR-6, can generate a homo-
geneous downward airflow over a wide range that can counteract
the upward airflow in the head area caused by FAW. This effectively
prevents any particles from shifting into the surgical field. Contrarily,
if there is a defect in the outlet, as in OR-5, the downward airflow in
the surgical field becomes obscured, and subsequently it cannot pre-
vent particle shifting.
Previously, Belani et al. used artificially generated bubbles to ana-
lyse their movement from the head area into the surgical field. He
concluded that FAW caused surgical field contamination, while con-
ductive patient warming devices were safe for surgical use because
they do not impact the ventilation airflow.4 Additionally, by observ-
ing helium soap bubbles, McGovern et al. found that FAW caused
convectional air currents to develop between the surgeon’s body and
the operating table, which transported the floor-level air upward and
into the surgical site.8 These authors also recommended a conductive
warming device be used instead of FAW. Importantly, these reports
did not investigate the airflow around the operating bed or the ACOL,
and thus the results might be affected by ACOL. In recent study, we
showed that sanitation quality was kept under FAW working even in
non-LAF equipped OR. However, the upward FAW-induced airflow in
the head area was not counteracted.9 Although, the global guidelines
of the world health organization for SSI prevention indicate that the
use of LAF was not necessarily recommended for patients undergoing
total arthroplasty surgery, warranting further research on these
issues.
Notably, this study has certain limitations, including the lack of
uniformity in the number of generated particles. To address this, we
Fig 4. Airflow direction and speed at the X4 cross section. Airflow direction and speed
analysed the ratio of shifting particles, and the results were almost
at the Y3 cross section with (upper) or without LAF (lower) caused by FAW with surgi-
cal lights. The red trapezoids represent surgical lights. Airflow speeds are represented identical to the results obtained based on the direct particle counting.
by the numeric values enclosed in the squares and by the length and colour of the As a further limitation, we did not examine the effects of surgical
arrows. staff, surgical devices, or other factors that may influence OR airflow.
ARTICLE IN PRESS
6 K. Shirozu et al. / American Journal of Infection Control 00 (2020) 1−6

Additionally, we did not measure colony forming units or the direct References
impact on SSIs. Further study is needed.
1. Kurz A, Sessler DI, Lenhardt R. Perioperative normothermia to reduce the incidence
of surgical-wound infection and shorten hospitalization. Study of Wound Infection
CONCLUSIONS and Temperature Group. N Engl J Med.. 1996;334:1209–1215.
2. Melling AC, Ali B, Scott EM, Leaper DJ. Effects of preoperative warming on the inci-
Overall, we conclude that the ACOL and the position relationship dence of wound infection after clean surgery: a randomised controlled trial. Lancet.
2001;358:876–880.
between ACOL and operating bed, is an important factor in prevent-
3. Shirozu K, Kai T, Setoguchi H, Ayagaki N, Hoka S. Effects of forced air warming on
ing particles from shifting into the surgical field. Additionally, FAW airflow around the operating table. Anesthesiology. 2018;128:79–84.
can be used safely to warm the patient during the operation, without 4. Belani KG, Albrecht M, McGovern PD, Reed M, Nachtsheim C. Patient warming
excess heat: the effects on orthopedic operating room ventilation performance.
causing surgical field contamination, in the case where the ACOL is
Anesth Analg. 2013;117:406–411.
set directly and completely above the area experiencing upward air- 5. Wagner JA, Greeley DG, Gormley TC, Markel TA. Comparison of operating room air
flow induced by FAW. distribution systems using the environmental quality indicator method of dynamic
simulated surgical procedures. Am J Infect Control. 2019;47:e1–e6.
6. Refaie R, Rushton P, McGovern P, et al. The effect of operating lights on laminar
Acknowledgment flow: an experimental study using neutrally buoyant helium bubbles. Bone Joint J.
2017;99-B:1061–1066.
I would like to express our deepest appreciation to Mr. Tetsuya 7. Kai T, Ayagaki N, Setoguchi H. Influence of the arrangement of surgical light axes on
the air environment in operating rooms. J Healthc Eng. 2019;2019: 4861273.
Kai, who work in Kyushu Medical Hospital for his assistance in this 8. McGovern PD, Albrecht M, Belani KG, et al. Forced-air warming and ultra-clean ven-
study. tilation do not mix: an investigation of theatre ventilation, patient warming and
joint replacement infection in orthopaedics. J Bone Joint Surg Br. 2011;93:1537–
1544.
SUPPLEMENTARY MATERIALS 9. Shirozu K, Takamori S, Setoguchi H, Yamaura K. Effects of forced air warming sys-
tems on the airflow and sanitation quality of operating rooms with non-laminar air-
Supplementary material associated with this article can be found flow systems. [e-pub ahead of print]. Perioper Care Oper Room Manag. https://doi.
org/10.1016/j.pcorm.2020.100119. Accessed July 18, 2020.
in the online version at https://doi.org/10.1016/j.ajic.2020.06.202.

You might also like