Professional Documents
Culture Documents
Expiratory Asynchrony: Hong-Lin Du, MD, Yoshitsugu Yamada, MD
Expiratory Asynchrony: Hong-Lin Du, MD, Yoshitsugu Yamada, MD
Expiratory Asynchrony
Hong-Lin Du, MDa, Yoshitsugu Yamada, MDb,*
a
Clinical Research Department, Newport Medical Instruments, Inc., PO Box 2600,
Newport Beach, CA 92658, USA
b
Department of Anesthesiology, School of Medicine, Yokohama City University,
Yokohama, Japan
* Corresponding author.
E-mail address: yamaday@med.yokohama-cu.ac.jp (Y. Yamada).
1078-5337/05/$ - see front matter Ó 2005 Elsevier Inc. All rights reserved.
doi:10.1016/j.rcc.2005.02.001 respiratorycare.theclinics.com
266 DU & YAMADA
the breath cannot be cycled off in a timely manner by the flow criteria, the
pressure termination criteria allow the ventilator flow to be cycled off
whenever the respiratory system’s recoil force or the expiratory muscle
recruitment creates an airway pressure increment above the set target
pressure. The time criteria act like a safety tool that may be activated in
situations such as massive airway leakage.
Table 1
Expiratory trigger sensitivity (ETS) in intensive care ventilators
Ventilator Expiratory trigger sensitivity
Ventilator pre-selected Adult Star 25% of peak inspiratory flow
fixed ETS Bear 1000 30% of peak inspiratory flow
Bird 8400 25% of peak inspiratory flow
Drager Evita 4 25% of peak inspiratory flow (adult)
6% of peak inspiratory flow
(pediatric and neonate)
Puritan Bennett 7200 5 L/min
Siemens Servo 300 5% of peak inspiratory flow
Siemens Servo 900 25% of peak inspiratory flow
VersaMed iVent 25% of peak inspiratory flow
User-selectable ETS Hamilton Galileo 10% to 40% of peak inspiratory flow
Puritan Bennett 840 1% to 45% of peak inspiratory flow
Cardiopulmonary Venturi 5% to 80% of peak inspiratory flow
Taema Horus 0 to 30 L/min (user-selectable)
Automated ETS Newport e500 5% to 55% of peak inspiratory flow
EXPIRATORY ASYNCHRONY 269
25 % Peak Flow
Airway
Flow
Neural TI
Fig. 1. The airway flow profile is affected by the respiratory time constant (Tau) during pressure
support ventilation. In patients with a shorter respiratory time constant (eg, patients with adult
reparatory distress syndrome), the flow decays at a faster speed and reaches 25% expiratory
trigger threshold earlier (solid line). In patients with a longer respiratory time constant (eg,
patients with COPD), the flow decays at a slower speed and reaches 25% expiratory trigger
threshold later (dotted line).
the flow decay is slower, and it takes longer to reach the preselected ETS
level. The ventilator termination of inspiratory flow could be less dependent
on the ending of patient neural inspiratory effort and more dependent on the
respiratory time constant. Because the respiratory time constant varies
widely among patients, it is understandable why a preselected fixed ETS
causes expiratory asynchrony. Clinical studies and mathematical analysis
have shown that, with a fixed level of ETS, patients with longer time
constants tend to have delayed termination and that patients with shorter
time constants tend to have premature termination [1,2,5,9].
ETS
Controller
Pressure
Support
Breath
Table 2
An example of expiratory trigger sensitivity ranges in relation to the respiratory time constants
for adult applications
Measured respiratory time constant (sec) Range of ETS
\0.8 10% to 35%
0.8 to 1.2 20% to 45%
[1.2 30% to 55%
Abbreviation: ETS, expiratory trigger sensitivity.
272 DU & YAMADA
Fig. 3. Waveforms showing ‘‘run-away’’ of proportional assist ventilation. Note that the
second breath has markedly longer ventilator inspiratory time when compared with the first
breath. The flow and positive pressure in the second breath continue after the end of the
patient’s inspiratory effort. (From Ambrosino N, Rossi A. Proportional assist ventilation
(PAV): a significant advance or a futile struggle between logic and practice. Thorax 2002;
57:272–6; with permission.)
support through flow assist on the basis of the measured P0.1 (an indicator of
the magnitude of patient inspiratory effort) and does not require accurate
measurement of respiratory elastance or airway resistance [18]. The efficacy
of this new breath type has yet to be validated.
Although PAV has attracted clinicians and researchers as a promising
addition to more conventional modes of mechanical ventilation, the request
of measuring respiratory elastance and airway resistance has been
practically challenging. At bedside, it is impractical to sedate a patient to
obtain accurate measurements of respiratory elastance and airway re-
sistance in a repeated manner. This is especially true for patients in the
weaning process. It is this group of patients who might benefit the most
from the use of PAV. To improve the usability of PAV at the bedside,
Younes et al [19,20] recently proposed a method for the noninvasive
estimation of inspiratory resistance and passive elastance. Farre et al [21]
have proposed that the forced oscillation technique applied by the
ventilator during noninvasive PAV could be useful in estimating ventilator
resistance. These techniques may turn PAV into a practically useful mode,
but evaluations of these techniques in terms of simplicity, reproducibility,
reliability, and robustness under bedside applications have yet to be
conducted.
274 DU & YAMADA
Airway leakage
Expiratory asynchrony during PAV may occur if there is airway leakage.
Unlike other breathing modes whereby leaks cause less ventilator assistance,
during PAV a small or medium leak at the airway (eg, leakage through an
uncuffed endotracheal tube) may result in greater and longer assistance from
the ventilator because the ventilator pressure is generated in proportion to
all gases delivered regardless of whether the delivered gases go into the
patient or leak out.
Medium or large leaks at airway may cause expiratory asynchrony during
PAV due to the leak’s effect on the measurements of respiratory elastance
and airway resistance. In children under mechanical ventilation, Main et al
[22] demonstrated that leaks larger than 20% resulted in gross overesti-
mation of respiratory elastance and airway resistance. The overestimation of
these parameters can contribute to expiratory asynchrony or run-away in
PAV.
0.3
0.2
0.1
0
0 0.2 0.4 0.6 0.8 1.0 1.2 0 0.2 0.4 0.6 0.8 1.0 1.2
Respiratory time constant τ (s)
System control delay: 5ms 20ms 40ms
Fig. 4. Ventilator inspiratory termination delay time at various levels of respiratory time
constant and assist gains during proportional-assist ventilation. Ventilator inspiratory
termination delay time is defined as the interval from the end of the patient inspiratory effort
to the end of the ventilator inspiratory flow. (From Du HL, Ohtsuji M, Shigeta M, et al.
Expiratory asynchrony in proportional assist ventilation. Am J Respir Crit Care Med
2002;165:972–7; with permission.)
276 DU & YAMADA
Neural
Neural TI Neural TE
Timing
The average neural expiratory time was 950 milliseconds. Out of this
expiratory time, the ventilator mandatory flow delivery continued for
500 milliseconds after the end of the neural inspiratory time. This means
that the ventilator and patient were in asynchrony for 53% of the neural
expiratory phase.
Obviously, when SIMV is used in infants, there is an unacceptable
amount of expiratory asynchrony.
There has been no study on expiratory asynchrony in infants during PSV
or PAV using a reliable method of estimating onset timings of neural
inspiration and expiration. The patient-ventilator synchrony during PSV
and PAV relies heavily on the flow profile at the airway opening, which is
a result of Pmus, respiratory elastance, and airway resistance. Because infants
usually have a weak Pmus and high airway resistance, the flow profile at
airway opening in infants is not as representative of the patient neural effort
profile as it is in adults. Therefore, it can be assumed that there are more
chances of expiratory asynchrony with these two modes in infants than in
adults. The fact that most infants have airway leakage [27] further increases
the chances of expiratory asynchrony during PSV and PAV.
Summary
Expiratory asynchrony is a universal phenomenon, and expiratory
synchrony occurs only by chance. Expiratory asynchrony exists in all
breath modes and has a significant impact on the patient’s work of
breathing and the weaning process. Advancements in ventilator designs and
basic physiologic science could lead to the improvement of the expiratory
asynchrony.
References
[1] Parthasarathy S, Jubran A, Tobin MJ. Cycling of inspiratory and expiratory muscle groups
with the ventilator in airflow limitation. Am J Respir Crit Care Med 1998;158:1471–8.
[2] Jubran A, Van de Graaff WB, Tobin MJ. Variability of patient-ventilator interaction with
pressure support ventilation in patients with chronic obstructive pulmonary disease. Am J
Respir Crit Care Med 1995;152:129–36.
[3] Greenough A, Wood S, Morley CJ, et al. Pancuronium prevents pneumothoraces in
ventilated premature babies who actively expire against positive pressure ventilation. Lancet
1984;1:1–13.
[4] Lipscomb AP, Thorburn RJ, Reynolds EO, et al. Pneumothorax and cerebral haemorrhage
in preterm infants. Lancet 1981;1:414–6.
[5] Nava S, Bruschi C, Fracchia C, et al. Patient-ventilator interaction and inspiratory effort
during pressure support ventilation in patients with different pathologies. Eur Respir J 1997;
10:177–83.
[6] Younes M, Kun J, Webster K, et al. Response of ventilator-dependent patients to delayed
opening of exhalation valve. Am J Respir Crit Care Med 2002;166:21–30.
[7] Van de Graaff WB, Gordey K, Dornseif SE, et al. Pressure support: changes in ventilatory
pattern and components of the work of breathing. Chest 1991;100:1082–9.
[8] Tobin MJ, Yang KL, Jubran A, et al. Interrelationship of breath components in neighboring
breaths of normal eupneic subjects. Am J Respir Crit Care Med 1995;152:1967–76.
[9] Yamada Y, Du HL. Analysis of the mechanisms of expiratory asynchrony in pressure
support ventilation: a mathematical approach. J Appl Physiol 2000;88:2143–50.
[10] Branson RD, Campbell RS. Pressure support ventilation, patient-ventilator synchrony, and
ventilator algorithm [editorial]. Respir Care 1998;43:1045–7.
[11] Kacmarek RM. NIPPV: patient-ventilator synchrony, the difference between success and
failure? Intensive Care Med 1999;25:645–7.
[12] Du HL, Amato MBP, Yamada Y. Automation of expiratory trigger sensitivity in pressure
support ventilation. Respir Care Clin North Am 2001;7:503–17.
[13] Younes M. Proportional assist ventilation: a new approach to ventilatory support. Am Rev
Respir Dis 1992;145:114–20.
[14] Younes M, Puddy A, Roberts D, et al. Proportional assist ventilation: results of an initial
clinical trial. Am Rev Respir Dis 1992;145:121–9.
[15] Giannouli E, Webster K, Roberts D, et al. Response of ventilator-dependent patients to
different levels of pressure support and proportional assist. Am J Respir Crit Care Med 1999;
159:1716–25.
[16] Ranieri VM, Grasso S, Mascia L, et al. Effects of proportional assist ventilation on
inspiratory muscle effort in patients with chronic obstructive pulmonary disease and acute
respiratory failure. Anesthesiology 1997;86:79–91.
[17] Navalesi P, Hernandez P, Wongsa A, et al. Proportional assist ventilation in acute
respiratory failure: effects on breathing pattern and inspiratory effort. Am J Respir Crit Care
Med 1996;154:1330–8.
280 DU & YAMADA
[18] Du HL. Breathing apparatus and method. United States Patent no. 6,622,726. September 23,
2003.
[19] Younes M, Kun J, Masiowski B, et al. A method for noninvasive determination of inspi-
ratory resistance during proportional assist ventilation. Am J Respir Crit Care Med 2001;
163:829–39.
[20] Younes M, Webster K, Kun J, et al. A method for measuring passive elastance during
proportional assist ventilation. Am J Respir Crit Care Med 2001;164:50–60.
[21] Farre R, Mancini M, Rotger M, et al. Oscillatory resistance measured during noninvasive
proportional assist ventilation. Am J Respir Crit Care Med 2001;164:790–4.
[22] Main E, Castle R, Stocks J, et al. The influence of endotracheal tube leak on the assessment
of respiratory function in ventilated children. Intensive Care Med 2001;27:1788–97.
[23] Du HL, Ohtsuji M, Shigeta M, et al. Expiratory asynchrony in proportional assist
ventilation. Am J Respir Crit Care Med 2002;165:972–7.
[24] Younes M, Bilan D, Jung D, et al. An apparatus for altering the mechanical load of the
respiratory system. J Appl Physiol 1987;62:2491–9.
[25] Younes M. Patient-ventilator interaction with pressure-assisted modalities of ventilatory
support. Semin Respir Med 1993;14:299–322.
[26] Beck J, Tucci M, Emeriaud G, et al. Prolonged neural expiratory time induced by mechanical
ventilation in infants. Pediatr Res 2004;55:1–8.
[27] Bernstein G, Knodel E, Heldt G. Airway leak size in neonates and autocycling of three flow-
triggered ventilators. Crit Care Med 1995;23:1739–44.
[28] Parthasarathy S, Jubran A, Tobin MJ. Assessment of neural inspiratory time in ventilator-
supported patients. Am J Respir Crit Care Med 2000;162:546–52.
[29] Milic-Emili J, Zin WA. Relationship between neuromuscular respiratory drive and
ventilatory output. In: Fishman AP, editor. Handbook of physiology. Section 3: the
respiratory system, vol. III. Bethesda (MD): American Physiological Society; 1986.
p. 631–46.
[30] Mognoni P, Saibene F, Sant’Ambrogio G, et al. Dynamics of the maximal contraction of the
respiratory muscles. Respir Physiol 1963;4:193–202.