You are on page 1of 130

GE Healthcare

Brivo CT315/325
Installation
Operating Documentation

5306510-1EN
Rev 7
© 2009~2012, General Electric Company,
All Rights Reserved.
Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

 本维修手册仅提供英文版本。
警告  如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
(ZH-CN)
 未详细阅读和完全理解本维修手册之前,不得进行维修。
 忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他形式的伤害。

• THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


WARNING
• IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE
(EN) OTHER THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO
PROVIDE TRANSLATION SERVICES.

Attention
• DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS
SERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE
SERVICE PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC
SHOCK, MECHANICAL OR OTHER HAZARDS.

ATTENTION • Ce manuel de service n'est disponible qu'en anglais.


• Si le technicien du client a besoin de ce manuel dans une autre langue
(FR) que l'anglais, c'est au client qu'il incombe de le faire traduire.
• Ne pas tenter d'intervenir sur les équipements tant que le manuel service
n'a pas été consulté et compris
• Le non-respect de cet avertissement peut entraîner chez le technicien,
l'opérateur ou le patient des blessures dues à des dangers électriques,
mécaniques ou autres.

警告 • このサービスマニュアルには英語版しかありません。
(JA) • サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業は
その業者の責任で行うものとさせていただきます。
• このサービスマニュアルを熟読し理解せずに、装置のサービスを行わないで
ください。
• この警告に従わない場合、サービスを担当される方、操作員あるいは患者さ
んが、感電や機械的又はその他の危険により負傷する可能性があります。

ATENÇÃO • Este manual de assistência técnica só se encontra disponível em inglês.


• Se qualquer outro serviço de assistência técnica, que não a gems,
(PT-BR) solicitar estes manuais noutro idioma, é da responsabilidade do cliente
fornecer os serviços de tradução.
• Não tente reparar o equipamento sem ter consultado ecompreendido este
manual de assistência técnica.
• O não cumprimento deste aviso pode por em perigo a segurança do
técnico, operador ou paciente devido a choques elétricos, mecânicos ou
outros.

ATTENTION
Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION

ОСТОРОЖНО! • Данное руководство по обслуживанию предлагается только на


английском языке.
(RU) • Если сервисному персоналу клиента необходимо руководство не на
английском, а на каком-то другом языке, клиенту следует
самостоятельно обеспечить перевод.
• Перед обслуживанием оборудования обязательно обратитесь к
данному руководству и поймите изложенные в нем сведения.
• Несоблюдение требований данного предупреждения может привести
к тому, что специалист по обслуживанию, оператор или пациент
получат удар электрическим током, механическую травму или другое
повреждение.

ATENCION • Este manual de servicio sólo existe en inglés.


• Si el encargado de mantenimiento de un cliente necesita un idioma que
(ES) no sea el inglés, el cliente deberá encargarse de la traducción del
manual.
• No se deberá dar servicio técnico al equipo, sin haber consultado y
comprendido este manual de servicio.
• La no observancia del presente aviso puede dar lugar a que el proveedor
de servicios, el operador o el paciente sufran lesiones provocadas por
causas eléctricas, mecánicas o de otra naturaleza.

DIKKAT • Bu servis kilavuzunun sadece ingilizcesi mevcuttur.


• Eğer müşteri teknisyeni bu kilavuzu ingilizce dişinda bir başka lisandan
(TR) talep ederse, bunu tercüme ettirmek müşteriye düşer.
• Servis kilavuzunu okuyup anlamadan ekipmanlara müdahale etmeyiniz.
• Bu uyariya uyulmamasi, elektrik, mekanik veya diğer tehlikelerden dolayi
teknisyen, operatör veya hastanin yaralanmasina yol açabilir.

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In Shipment” on ALL
copies of the freight or express bill BEFORE delivery is accepted or “signed for” by a GE representative or hospital receiving
agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event,
within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company
will not pay a claim for damage if an inspection is not requested within this 14 day period.
To file a report:
Fill out a report on http://egems.med.ge.com/edq/home.jsp
Contact the local service coordinator.

ATTENTION
Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT


All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be
performed by licensed electrical contractors. In addition, electrical feeds into the Power Distribution Unit shall be performed
by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations and testing shall
be performed by qualified GE Medical personnel. The products involved (and the accompanying electrical installations) are
highly sophisticated, and special engineering competence is required. In performing all electrical work on these products,
GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the
requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party

Attention
service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.

IMPORTANT! . . . X-RAY PROTECTION


X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein contained should be thoroughly
read and understood by everyone who will use the equipment before you attempt to place this equipment in operation. The
General Electric Company, Medical Systems Group, will be glad to assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the useful beam, no practical
design of equipment can provide complete protection. Nor can any practical design compel the operator to take adequate
precautions to prevent the possibility of any persons carelessly exposing themselves or others to radiation.
It is important that anyone having anything to do with x-radiation be properly trained and fully acquainted with the
recommendations of the National Council on Radiation Protection and Measurements as published in NCRP Reports
available from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and of the
International Commission on Radiation Protection, and take adequate steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Medical Systems Group, its agents, and
representatives have no responsibility for injury or damage which may result from improper use of the equipment.
Various protective materials and devices are available. It is urged that such materials or devices be used.

LITHIUM BATTERY CAUTIONARY STATEMENTS

CAUTION
Danger of explosion if battery is incorrectly replaced. Replace only with the same or equivalent
type recommended by the manufacturer. Discard used batteries according to the manufacturer’s
instructions.

OMISSIONS & ERRORS


Customers, please contact your GE Sales or Service representatives.
GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in this publication.

ATTENTION
Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION

ATTENTION
Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

REVISION HISTORY
REV Date Primary Reason for Change
7 03/08/2012 Updated the rating plates of system components for IEC3. (Sec. 9)
Added the Gantry rating plates for India SKD. (Sec. 9)
Updated the rating plate of software collector. (Sec. 4)
Added the screen illustration of “Hardware Settings” software version 2.2x and a notice for selecting initial laser po-
sition angle. (Sec.4)
Added the screen illustrations of “Language Settings” and “Preference Settings” for software version 2.2x. (Sec. 4,
Sec. 8)
Updated Table 4-1, Illustration 4-11, 4-8 and 4-2 (Added GND cable for OC). (Sec. 4)

Revision History
Added a note to indicate that there is no Pinch Point Label. (Sec. 8)
Added the Read Docs Caution label, Weight Maximum Cation label, PDU Movement Caution Label. and Type B Ap-
plied Part Label. (Sec. 8)
Updated the storage temperature and humidity. (Appendix A)

6 01/18/2011 Added warning for neutral line comments. (Sec. 4)


Added cable connector size to system power/signal cables list for cable positioning. (Sec. 4)
Updated SCB Film from 5337069 to 5397664. (Sec. 8)
Updated Table Caution Labels. (Sec. 8)
Added Rating Plate Comparison Label for Spanish. (Sec. 8)
Updated Phantom Holder from 5334729 to 5395684. (Sec. 9)
Updated Coronal Head Holder_Option from 2299548 to 2201850. (Sec. 9)
Added figures for Axial Head Holder_Standard (5334730) and Axial Head Holder_Option (2115996-3). (Sec. 9)

5 07/01/2010 Added Multi-language warnings for French, Japanese, Brazil, Russian, Spanish and Turkish.
Updated Tools and Test Equipment. (Sec. 2)
Added notice for environment check. (Sec. 2)
Updated Power Frequency from 50/60 1Hz to 50/60 %2Hz. (Sec. 2)
Updated Gantry Dollies installation method. (Sec. 3)
Updated Gantry Top Cover Removal. (Sec. 3)
Added packaging position of Gantry and Table connection washer. (Sec. 3)
Added Gantry/Table anchor adjuster JIG. (Sec. 3)
Updated gantry level specification from "+/- 0.5mm/1000mm" to "+/- 0.2 degree". (Sec. 3)
Updated distance between the table base and gound from no less than 5mm to no less than 17mm. (Sec. 3)
Added Adapter installation and cable connection. (Sec. 3, Sec. 4)
Updated system cable interconnection figure and system power/signal cable list. (Sec. 4)
Updated OC cable interconnection. (Sec. 4)
Updated PDU power cables connection (high voltage input and output direction). (Sec. 4)
Added E-stop terminal to PDU for Adapter. (Sec. 4)
Added Adapter check for ground continuity check. (Sec. 4)
Added warning for system power. (Sec. 4)

4 03/03/2010 Updated Gantry Front/Rear Cover installation/removal process. (Sec. 3)


Added appropriate language caution labels. (Sec. 8)
Updated storage temperature from -10C~50C to -30C~50C. (Appendix)

3 11/16/2009 Added contents of labels and emblems. (Sec. 8)


Updated System Component Identification. (Sec. 9)

REVISION HISTORY
Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
2 11/04/2009 Updated manual to close SPRGEBct05986.
Updated tools and test Equipments supplied by GEHW. (Sec. 2)
Updated Illustration 3-7 Distance between two transportation supports. (Sec. 3)
Updated bubble level position for Gantry Left/Right level adjustment. (Sec. 3)
Updated spec from 5mm to 5mm~15mm for distance of gantry base and ground. (Sec. 3)
Added figures for anchor adjuster lock tool. (Sec. 3)
Deleted 3-2-6 Gantry Verification. (Sec. 3)
Updated Table anchor position. (Sec. 3)
Updated Illustration 4-1 System Cable Interconnection. (Sec. 4)
Updated Table 4-1 System Power/Signal Cables for adding cable length. (Sec. 4)
Updated Indication Light / Switch Connection (System X-ray-ON Light). (Sec. 4)
Added step “Remove the azimuth lock pin.” before system power on and switches description of PDU. (Sec. 4)
Deleted 4-7 Leveling of Table. (Sec. 4)
Updated figures for DICOM Carame Configuration and PostScript Camera Configuration. (Sec. 4)
Updated scan parameters of Table and Gantry Alignment. (Sec. 5)
Updated spec to ‘no less than 5mm’ for distance of table base and ground. (Sec. 5)
Updated QA phantom diameter from 266mm to 214mm. (Sec. 7)
Added DFOV and Recon for High Contrast Scan List. (Sec. 7)
Updated Image Performance Check List of System Test. (Sec. 7)
Added Scan Control Box Film. (Sec. 9)
Updated Table label position. (Sec. 9)

1 07/17/2009 Initial release.

REVISION HISTORY
Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

TABLE OF CONTENTS

SECTION PAGE

SECTION 1 - INSTALLATION STEERING GUIDE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1


1-1 OPERATION FLOWCHART . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-2 OPERATION FOR X-RAY EXPOSURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2

SECTION 2 - SITE PREPARATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1


2-1 SITE PREPARATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-2 TOOLS AND TEST EQUIPMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2-2-1 Items Supplied By GEHW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2-2-2 Items Supplied By Field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2-3 ENVIRONMENT CHECK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
2-4 POWER LINE VOLTAGE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6

Table of Contents
2-5 ACCESSORY CHECK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
2-6 TORQUE CONVERSION TABLES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
2-6-1 [Conversion: kgf•cm -> N•m] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
2-6-2 [Conversion: N•m -> kgf•cm] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
2-6-3 [Conversion: N•m -> lbf•ft] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9

SECTION 3 - POSITIONING SUBSYSTEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1


3-1 FLOOR AND ROOM PREPARATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3-2 GANTRY INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3-2-1 Gantry Dollies Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3-2-2 Drill Gantry Anchor Hole . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
3-2-3 Gantry Level Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
3-2-4 Gantry Anchoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
3-3 TABLE INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
3-4 OC INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
3-5 ADAPTER INSTALLATION (IF NO ADAPTER, PLEASE IGNORE THIS SECTION) . 3-18

SECTION 4 - WIRING AND POWER-ON. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1


4-1 SYSTEM POWER, GROUND AND SIGNAL CABLE CONNECTION . . . . . . . . . . . . . 4-1
4-2 CABLE INTER-CONNECTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4-2-1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4-2-2 System Cable Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4-2-3 System Cable Checklists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4-2-4 OC Cable Interconnection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
4-2-5 Adapter Cables Connection (If No Adapter, Please Ignore This Section) . . . 4-10
4-2-6 PDU Cables Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
4-2-7 Gantry / Table Cable Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
4-3 INDICATION LIGHT / SWITCH CONNECTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
4-3-1 System X-ray-ON Light and Scan Room Door Safety Switch . . . . . . . . . . . . 4-13
4-4 GROUND CONTINUITY CHECK. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
4-5 CONTROL PANEL CONNECTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
4-6 SYSTEM POWER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
4-7 PRODUCT KEY / OPTION SOFTWARE/PRODUCT PROTOCOL INSTALLATION . 4-19
4-8 EMERGENCY OFF TEST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-23

i
Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

TABLE OF CONTENTS

SECTION PAGE
4-9 MONITOR SET-UP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
4-9-1 Keys Assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
4-9-2 Lock of OSD Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
4-9-3 Monitor Gamma Value. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
4-9-4 Monitor Color Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
4-9-5 Contrast and Brightness (Minor Adjustment). . . . . . . . . . . . . . . . . . . . . . . . . 4-25
4-10 JEDI Date & Time Correction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
4-11 SYSTEM CONFIGURATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-27
4-11-1 Hardware Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-27
4-11-2 System Setting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-27
4-12 CAMERA SETTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
4-13 CONNECTION THE SYSTEM TO THE NETWORK . . . . . . . . . . . . . . . . . . . . . . . . . . 4-38
4-13-1 Hardware Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-38
4-13-2 Remote Host Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-38

SECTION 5 - GANTRY AND TABLE INTEGRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1


5-1 TABLE TO GANTRY ALIGNMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5-2 ANCHORING OF TABLE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
5-2-1 Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
5-2-2 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
5-2-3 Table Anchoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
5-3 ROTATION OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
5-4 CRADLE ABSOLUTE POSITION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7

SECTION 6 - TUBE ALIGNMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1


6-1 PLANE OF ROTATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6-2 DETECTOR BEAM-ON-WINDOW. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6-3 Q-CAL CHANNEL RATIO (ONLY FOR bRIVO ct325). . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6-4 ISO CENTER ALIGNMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6-5 FILTER CENTER ALIGNMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1

SECTION 7 - SYSTEM TEST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1


7-1 POSITIONING LIGHTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7-2 INTERCOMMUNICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7-3 TABLE LATCH CHECK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
7-4 SCAN TEST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
7-5 CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
7-5-1 Hilight Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
7-5-2 Air and Phantom Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
7-6 CT NUMBER ADJUSTMENT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
7-7 IMAGE PERFORMANCE CHECK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-6

ii
Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

TABLE OF CONTENTS

SECTION PAGE

SECTION 8 - INSTALLATION COMPLETION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1


8-1 COVER INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
8-1-1 Gantry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
8-1-2 Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
8-2 ATTACHING LABELS AND EMBLEMS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
8-3 SYSTEM VERIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7
8-4 LANGUAGE SELECTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7
8-5 BACK-UP SYSTEM STATE DATA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
8-5-1 Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
8-5-2 Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
8-6 SYSTEM AND ROOM CLEANING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10

Table of Contents
SECTION 9 - SYSTEM COMPONENT IDENTIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
9-1 COMPONENT IDENTIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1

iii
Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION

iv
Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

SECTION 1 - INSTALLATION STEERING GUIDE

1-1 OPERATION FLOWCHART


The following is the typical installation and check/adjustment procedure.

Illustration 1-1 Installation Flow

Installation Procedures (Estimated time to complete by two field engineers)

Carrying System and Site Preparation 2H

Steering Guide
Installation
Unpack system packaging 4H

Positioning Subsystem and Wiring 4H

Power On

Gantry / Table Integration 1H

X-ray Tube Alignment 3H

System Alignment and Test 6H

Installation Completing

(H: Hour)

SECTION 1 - INSTALLATION STEERING GUIDE 1-1


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

1-2 OPERATION FOR X-RAY EXPOSURE

Operation Reference Section X-Ray

Plane of Rotation 6-1 Yes

Detector Beam on Window 6-2 Yes

Q-cal Channel Ratio 6-3 Yes

ISO Center Alignment 6-4 Yes

Filter Center Alignment 6-5 Yes

Table to Gantry Alignment 5-1 Yes

Scan Test 7-4 Yes

Calibration 7-5 Yes

CT Number Adjustment 7-6 Yes

Image Performance Check 7-7 Yes

1-2 SECTION 1 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

SECTION 2 - SITE PREPARATION

CAUTION
All steps outlined in the Preinstallation manual must be complete before proceeding with
these installation procedures.

2-1 SITE PREPARATION


Before installation work begins, check the site to verify the completion of duct work, raceway runs, and contractor
installed wiring, according to specifications. Review the Preinstallation Manual and the installation layout plans and
use them to check for compliance in all areas.
Room preparation should include the following before equipment installation begins:

Site Preparation
1. Conduit runs, floor, and/or wall duct in place.
2. Contractor supplied wiring pulled and tagged.
3. Electrical power for system installed and checked. Main circuit breakers OFF and tagged.
4. Room air conditioning completed and checked.
5. Doorways and hallways checked for equipment clearance before delivery. If necessary, arrangements made
for riggers.
6. Power available for hand drills, vacuum cleaners and test equipment.
7. Arrangements made for room lockup, tool lockup, and equipment security until it is unpacked.
8. Telephone extension in control room.

SECTION 2 - INSTALLATION STEERING GUIDE 2-1


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

2-2 TOOLS AND TEST EQUIPMENT


The following tools and test equipment must be obtained by the installing personnel before starting to install the
system.

2-2-1 Items Supplied By GEHW


YES NO

____ ____ Gantry Service Tool (Cover Dolly)


____ ____ Phantom Holder Assy
____ ____ QA Phantom (2144715)
____ ____Alignment JIG holder for X-ray beam alignment procedure (5371128, only for single slice system)
____ ____Alignment JIG (5371129, only for single slice system)
____ ____ 6 Gantry Anchor Bolts and 6 Table Anchor Bolts
____ ____ Insulating Oil, for filling cable receptacles
____ ____ Hoist-2-PC, Long Tube
____ ____ Hoist-2-PC ARM
____ ____ Gantry Center Weight Rear Adjustment Sheet (x2)
____ ____ Gantry Side Weight Adjustment Sheet (x2)
____ ____ Patient Access Assembly

2-2 SECTION 2 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

2-2-2 Items Supplied By Field


YES NO

____ ____ Covers and/or plastic sheeting with tie ends, to protect equipment from dust
____ ____ Standard service person tool kit (as per pole policy)

____ ____ Thermometer for room temperature checks (0C - 100C)


____ ____ Clamp-on Volt-Ammeter, Scale 2/20 Amps, 1% Accuracy
____ ____ Digital multimeter, FLUKE Model 87 series or equivalent
____ ____ Wiring tools, lugs, and solderless terminals with crimping tools for AWG. #14 through #4 unit with
markers and tags
____ ____ Vacuum cleaner
____ ____ Extension cords
____ ____ Flashlights / Torch

Site Preparation
____ ____ Greenlee punches for conduit fittings
____ ____ Saber saw/ Hack saw with metal blades to cut duct covers
____ ____ GAFCHROMIC Film XR-QA (5321742)
____ ____ Hexagonal key wrench and spanner set (Metric Size)
____ ____ Cable cutter (KLEIN 63050)
____ ____ Wire clamp crimping tool for crimps up to 50mm size
____ ____ Wire clamp crimping tool type 59239 (AMP)
____ ____ Torpedo Level: Sears Industrial: 9" (9-39829)
FUJI SEIMITSU Level: 524A 1DIV= 0.5mm/M
____ ____ Hammer
____ ____ Electric hammer drill and selection of bits including masonry bits for Table/Gantry anchor bolts
12mm drill bit (2203081: HILTI, TE-CX 12/17 or equivalent)
____ ____ Densitometer, Sensitometer (Option)
____ ____ Dial Gauge

SECTION 2 - INSTALLATION STEERING GUIDE 2-3


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

2-2-2 Items Supplied By Field (Continued)


YES NO

____ ____ Tube hoist assembly (2268140)


____ ____ Dranetz line disturbance meter
____ ____ Buzzer for ringing out wires and cables (i.e., continuity check of wires)
____ ____ Spray can of “WD-40” or other light oil.
____ ____ Clinometer
____ ____ HV Bleeder varian HV-1 3 wire (S96300) or GE type (C1515A)
____ ____ HV Bleeder cable x 2 (5503805) or GE (C1665A)
____ ____ Earth Resistance/Bonding test meter
____ ____ Tube/Detector alignment dial gauge with magnetic base
____ ____ ESD portable anti static kit (46-194427P231)
____ ____ Scope isolator (to measure 750V DC) 
Tektronix 6902 - region/country level only
____ ____ Grease (Shell Alvania type or equivalent, 91337153)
____ ____ Grease gun
____ ____ XG extender board (C653319)
____ ____ Safety Glasses
____ ____ Gantry/Table Anchor Adjuster JIG (5389135)

2-4 SECTION 2 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

2-3 ENVIRONMENT CHECK


During the installation and start-up of the system you must take continuous temperature checks to make sure the
environment meets specifications. Excessive temperature, coupled with reduced air flow from clogged filters, may
overstress components and cause failures.

NOTICE
Component freezing occurs if CT system is exposed to temperatures below 0C for a period
longer than two days. Allow a minimum of 12 hours for the CT system to adjust to ambient
room temperature, prior to installation. (It’s better to keep the CT system in transportation
packaging for 12 hours for acclimatization.)

Specifications
During installation, use the temperature/humidity recorder and a thermometer to record the temperature and relative
humidity in the scan room. Make sure the data you collect meets the following temperature and humidity specifica-
tions.

Site Preparation
GANTRY / SCAN ROOM:
• Temperature : 20C ~ 28 C (at least 5 C/hr gradient)
• Relative Humidity : 30 ~ 70 % RH
OPERATOR’S CONTROL / EQUIPMENT ROOM:
• Temperature : 15C ~ 30C (at least 5 C/hr gradient)
• Relative Humidity : 20 ~ 80 % RH
• Air conditioning of the building may be sufficient.

SECTION 2 - INSTALLATION STEERING GUIDE 2-5


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

2-4 POWER LINE VOLTAGE

Instrument
Use the same instrument (Dranetz, etc.) to monitor line voltage during this check that you used during the preinstal-
lation check. The recommended line analyzer is the Dranetz Model 606-3 (3 channel) with the 101 frequency option,
or Dranetz Model line monitor plug in modules. You can use a similar analyzer, as long as it can monitor and report
mean voltage, surges, sags, short term impulses, and frequency drift.

Check Using an Instrument


Monitor the input power lines 380V to the PDU, check in particular:
• Verify that the current line voltage equals the voltage measured during the preinstallation check.
• Verify that mean line voltage does not fall outside of equipment specifications during the day to day op-
eration of other site equipment during the installation period.
• Look for any line voltage disturbances and check for a correlation between disturbances and system
malfunctions such as “Aborts” or component failures.
From the printouts, ensure that the power supply meets the following criteria:
• The system shall operate on an electrical power supply with 380V (phase to phase line voltage), and
220V (phase to neutral) nominal input voltages.
• Daily Voltage Variation: +10% to -5% from nominal steady state (50/60 Hz)

• Frequency: 50/60 2% Hz

2-5 ACCESSORY CHECK


Check if all accessories to be required have been shipped with the system.
Use the Packing Check List shipped with the accessory package.

2-6 SECTION 2 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

2-6 TORQUE CONVERSION TABLES


Use the following conversion formula and table to change a unit suitable for a torque wrench you have.

2-6-1 [Conversion: kgf•cm -> N•m]

Conversion Formula:
xx (N•m) = xxxx (kgf•cm) x 0.098
Conversion Table:

1 kgf•cm = 0.098 N•m


Conversion: kgf•cm -> N•m
kgf•cm 0 1 2 3 4 5 6 7 8 9

Site Preparation
N•m
0 0 0.0981 0.1961 0.2942 0.3923 0.4903 0.5884 0.6865 0.7845 0.8826
10 0.9807 1.0787 1.1768 1.2748 1.3729 1.4710 1.5690 1.6671 1.7652 1.8632
20 1.9613 2.0594 2.1574 2.2555 2.3536 2.4516 2.5497 2.6478 2.7458 2.8439
30 2.9420 3.0400 3.1381 3.2361 3.3342 3.4323 3.5303 3.6284 3.7265 3.8245
40 3.9226 4.0207 4.1187 4.2168 4.3149 4.4129 4.5110 4.6091 4.7071 4.8052
50 4.9033 5.0013 5.0994 5.1974 5.2955 5.3936 5.4916 5.5897 5.6878 5.7858
60 5.8840 5.9820 6.0800 6.1781 6.2762 6.3742 6.4723 6.5704 6.6684 6.7665
70 6.8646 6.9626 7.0607 7.1587 7.2568 7.3549 7.4529 7.5510 7.6491 7.7474
80 7.8452 7.9433 8.0413 8.1394 8.2375 8.3355 8.4336 8.5317 8.6297 8.7278
90 8.8260 8.9239 9.0220 9.1200 9.2181 9.3162 9.4142 9.5123 9.6104 9.7084
100 9.8065

SECTION 2 - INSTALLATION STEERING GUIDE 2-7


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

2-6 TORQUE CONVERSION TABLES (CONTINUED)

2-6-2 [Conversion: N•m -> kgf•cm]

Conversion Formula:
xx (kgf•cm) = xxxx (N•m) x 10.2
Conversion Table:

1N•m = 10.2 kgf•cm


Conversion: N•m -> kgf•cm
N•m 0 1 2 3 4 5 6 7 8 9
kgf•cm
0 0 10 20 31 41 51 61 71 82 92
10 102 112 122 133 143 153 163 173 184 194
20 204 214 224 235 245 255 265 275 286 296
30 306 316 326 337 347 357 367 377 388 398
40 408 418 428 439 449 459 469 479 490 500
50 510 520 530 541 551 561 571 581 592 602
60 612 622 632 643 653 663 673 683 694 704
70 714 724 734 745 755 765 775 785 796 806
80 816 826 836 847 857 867 877 887 898 908
90 918 928 938 949 959 969 979 989 1000 1010
100 1020

2-8 SECTION 2 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

2-6 TORQUE CONVERSION TABLES (CONTINUED)

2-6-3 [Conversion: N•m -> lbf•ft]

Conversion Formula:
xx (lbf•ft) = xxxx (N•m) x 0.737
xx (N•m) = xxxx (lbf•ft) x 1.356
Conversion Table:

1N•m=0.737 lbf•ft

Conversion: N•m -> lbf•ft

N•m 0 1 2 3 4 5 6 7 8 9

Site Preparation
lbf•ft

0 0 0.737 1.474 2.211 2.948 3.685 4.422 5.159 5.896 6.633

10 7.370 8.107 8.844 9.581 10.318 11.055 11.792 12.529 13.266 14.003

20 14.740 15.477 16.214 16.951 17.688 18.425 19.162 19.899 20.636 21.373

30 22.110 22.847 23.584 24.321 25.058 25.795 26.532 27.269 28.006 28.743

40 29.480 30.217 30.954 31.691 32.428 33.165 33.902 34.639 35.376 36.113

50 36.850 37.587 38.324 39.061 39.798 40.535 41.272 42.009 42.746 43.483

60 44.220 44.957 45.694 46.431 47.168 47.905 48.642 49.379 50.116 50.853

70 51.590 52.327 53.064 53.801 54.538 55.275 56.012 56.749 57.486 58.223

80 58.960 59.697 60.434 61.171 61.908 62.645 63.382 64.119 64.856 65.593

90 66.330 67.067 67.804 68.541 69.278 70.015 70.752 71.489 72.226 72.963

100 73.700

SECTION 2 - INSTALLATION STEERING GUIDE 2-9


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION

2-10 SECTION 2 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

SECTION 3 - POSITIONING SUBSYSTEM

Note
Before you start the installation, make sure the site preparation complies with conditions and in-
structions found in the pre-installation manual. Failure to comply will result in excessive installation
delay and potential increased, un-recoverable installation costs. This product is designed to meet
specific mechanical installation standards, which should be reviewed prior to installing this system.

3-1 FLOOR AND ROOM PREPARATION


The Gantry and Table must rest on solid concrete or other basic flooring, NOT on resilient tile or carpeting which
may slowly yield over a period of time and disturb the Table to Gantry alignment.
The floor structure must be capable of withstanding both the combined weight of the Table and Gantry and the
individual area loading of these components. Refer to Section 2, Room Layouts, of the PREINSTALLATION

Positioning
Subsystem
manual.
Method and Placement of anchoring must not reduce the structural strength or integrity of the floor.

SECTION 3 - INSTALLATION STEERING GUIDE 3-1


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

3-2 GANTRY INSTALLATION

3-2-1 Gantry Dollies Installation


1. Remove all the transportation packaging.
Note
The Gantry Side Covers had been removed before shipped from manufactory.

Illustration 3-1 Remove Gantry Transportation Packaging

all the transportation packaging removal


(with top, front and rear covers)

2. Remove the rest of the gantry covers.


• Top Covers
• Front and Rear Covers

3-2 SECTION 3 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

3-2 GANTRY INSTALLATION (CONTINUED)

Gantry Top Cover Removal


a. Disconnect the fan power cable connector that connected to the gantry rear cover.
b. Unscrew two screws that secure the top cover.

Illustration 3-2 Top Cover Removal

Top Cover_L

Positioning
Subsystem
Associated Bracket Plug (x2)
Screw (x2)

c. Hold the cover bracket to horizontally draw the top cover outward to release plugs from the both sides of
the top cover bracket. (See Illustration 3-2)
d. “Crash” is heard when plugs were released, hold the cover bracket to lift the Top Cover upward until it can
be take out from the Mounting Bracket.

Illustration 3-3 Top Cover Removal

Mounting Bracket

e. Repeat the above steps a. through d. for the other Top Cover.

SECTION 3 - INSTALLATION STEERING GUIDE 3-3


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

3-2 GANTRY INSTALLATION (CONTINUED)

Gantry Front/Rear Cover Removal

CAUTION
Possible injury!
If the cover dolly fall over, they can injure persons. Therefore, after using the cover dolly,
keep them in a safe place and/or secure them from falling over even when the customer or
other persons accidentally hit the cover dolly.

NOTICE
Potential for front and rear covers damage.
Front and rear covers removal and installation can be safely accomplished by one person
using the dollies provided with the system. Failure to use these dollies will significantly in-
crease the likelihood of damage to the covers. Do not lean covers against walls.

Special Equipment
• Gantry Cover dolly
f. Remove the mylar window from the front and rear covers.

Illustration 3-4 Gantry Front and Rear Covers


Right View

Front Cover Rear Cover

Holes for Dolly


Studs Installation

Cover Mounting Bolt (x4)

g. Install the four dolly studs onto the Gantry Front/Rear Cover.

3-4 SECTION 3 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

3-2 GANTRY INSTALLATION (CONTINUED)


h. Install the cover dollies to the dolly studs.
Note
Please pay attention to the direction of the cover dollies when removing the front cover.

Positioning
Subsystem
Short Side
Long Side

i. Unscrew the two Mounting Bolts of the Front/Rear Cover.


j. Slightly disengage the Gantry Front/Rear Cover and disconnect the cable connectors. (See Illustration 3-6)
For the Front Cover:
- FMPhone RL1 ---- 5321620 (to front cover) & 5321399 (to TGP)
- S-DISP RL1 ---- 5321607 (to front cover) & 5321391 (to TGP)
For the Rear Cover:
- RMPhone RL1 ---- 5321619 (to rear cover) & 5321398 (to TGP)
- X-DISP RL1 ---- 5321618 (to rear cover) & 5321395 (to TGP)
- Fan RL3 ---- 5321537

NOTICE
For Gantry Front/Rear Cover securing, magnet is attached on the Gantry HLA Frame, please
be careful to separate the Gantry Front/Rear Cover from the Gantry HLA and avoid falling
down. (See Illustration 3-5)

SECTION 3 - INSTALLATION STEERING GUIDE 3-5


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

3-2 GANTRY INSTALLATION (CONTINUED)

Illustration 3-5 Gantry Front/Rear Cover Securing


Front Cover Rear Cover

Magnet

k. Remove the Gantry Front/Rear Cover away from the Gantry.

Illustration 3-6 Gantry Cover Dollies

Dolly Stud (x4)

Front Cover Rear Cover

FMPhone RL1 & S-DISP RL1 RMPhone RL1 & X-DISP RL1 & Fan RL3

Plug (x2)

Mounting Bolt Hole (x2)

3-6 SECTION 3 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

3-2 GANTRY INSTALLATION (CONTINUED)


3. Assemble the gantry dollies to the Gantry and delivery the system to the scan room.
a. Remove four fork brackets from the Gantry HLA by unscrewing its mounting screw.

Positioning
Subsystem
Fork Bracket

b. Install four fork brackets to the Gantry Dollies.

Fork Bracket

Mounting Screw (x2)

SECTION 3 - INSTALLATION STEERING GUIDE 3-7


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

3-2 GANTRY INSTALLATION (CONTINUED)


c. Assemble Gantry Dollies to the Gantry HLA by screwing its three mounting screws.

Illustration 3-7 Gantry Dollies Installation

Mounting Screw (x3)

Transportation Support

3-8 SECTION 3 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

3-2-2 Drill Gantry Anchor Hole


The following parts are provided with the system.
• Gantry Standard Anchor Kit
Anchor Bolts (5214921)
Washer (2105873)
Anchor Nut
Leveling Adjuster
Leveling Pad
Note
The other anchor can also be used, instead of the anchor kit shipped with the system. In this case,
you can use the other anchoring procedures that are not described in this manual, but the minimum
tensile load strength of the anchor MUST meet the specification below.

NOTICE
The minimum tensile load strength of the anchor must be 13000N.

Positioning
Subsystem
4. Remove the two transportation supports from the gantry base by unscrewing four bolts.
5. To fix gantry position for anchoring.
Method 1 (Recommended)
a. Move the Gantry to the appropriate position of the scan room and lower the Gantry on the floor.
b. Scribe a mark at each of the gantry’s anchor hole locations, and then move the Gantry away.
c. Align two transportation supports’ holes to meet marks on the floor.
Method 2
a. Position the two transportation supports directly on the floor.
b. Confirm the distance between the two transportation supports and make marks.

Illustration 3-8 Distance between Two Transportation Supports

520
134
2.8

938

1238

SECTION 3 - INSTALLATION STEERING GUIDE 3-9


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

3-2-2 Drill Gantry Anchor Hole (Continued)


6. Drill the holes on the floor.
a. Attach a piece of tape to the 12mm concrete drill bit (2203081: HILTI, TE-CX 12/17 or equivalent) in order
to assure a minimum of 100mm depth of anchor hole.

Illustration 3-9 12 mm Concrete Drill Bit Preparation

100mm or more
Tape
b. Use electric drill with the bit prepared in the previous step, so that the lower edge of the tape just touch the
floor surface.
c. Clear debris of the hole several times while drilling to prevent the drill bit from binding.

NOTICE
During drilling, use the transportation supports always keep the drill bit perpendicular to the
floor.

Or During drilling, not use the transportation supports. With one person drilling the hole,
the second person can visually insure perpendicularity throughout the drilling operation.

3-10 SECTION 3 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

3-2-2 Drill Gantry Anchor Hole (Continued)

NOTICE
If any obstruction is found under the proper anchoring position, stop drilling and use the al-
ternate position. Use the alternate positions only in the case where its corresponding prop-
er position is not possible. (See Illustration 3-10)

Illustration 3-10 Gantry Anchor Positions (Proper and Alternative)

Proper Anchor Position

Alternative Anchor Position

Positioning
Subsystem
d. Use the vacuum cleaner to suck up the concrete dust from the hole. Also vacuum dust from the surrounding
area when finished.
e. Hammer the anchor bolt into the hole made in step b.

Illustration 3-11 Anchor Bolt

7. Locate the four leveling pads, and let each of anchor bolts go through the leveling pad respectively.
8. Position the Gantry over the anchor bolts appropriately.
9. Lower the Gantry to close to the floor and position each of leveling pads beneath its associated adjuster.

SECTION 3 - INSTALLATION STEERING GUIDE 3-11


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

3-2-3 Gantry Level Adjustment


10. Place the bubble level at the positions shown as Illustration 3-12.

Level Specification: +/- 0.2 degree

Illustration 3-12 Gantry Adjuster and Bubble Level Position


Front View Top View

Adjuster (Proper)
Alternative
Bubble Level Position
Bubble Level Position

11. Level the Gantry using the Gantry/Table Anchor Adjuster JIG.

NOTICE
The distance between the Gantry stationary base and floor is 5~15mm.
To Adjust Leveling Adjuster To Lock Adjuster Lock Ring

3-12 SECTION 3 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

3-2-4 Gantry Anchoring


12. Tighten the anchor nut to the specified torque to fix the anchor bolts into the floor.

Torque: 50 N•m

NOTICE
If specified torque cannot be achieved, another anchor point must be used. See Illustration
3-10 for alternative anchor position.
13. Loosen all of the anchor nuts, then re-tighten the anchor nuts securely (less than 50 N•m).

NOTICE
Do not tighten the anchor nuts to the specified torque (50 N•m), because the anchor bolts,
tighten in previous step, has been fixed securely into the floor. So, to over-tighten nuts will
damage the frame of the Gantry and also bring an out-of-level Gantry.

Illustration 3-13 Gantry Anchor Assembly

Positioning
Subsystem
Anchor Bolt
Anchor Bolt
Anchor Washer

Leveling Adjuster Nut

Adjuster Lock Ring

Anchor Washer

Gantry Stationary Base


Leveling Adjuster
Leveling Pad
5mm~15mm Adjuster Lock Ring
90mm
100mm

SECTION 3 - INSTALLATION STEERING GUIDE 3-13


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

3-3 TABLE INSTALLATION

Table Dollies Removal


1. Remove all the transportation packaging, except dollies, from the Table.
2. Use the dollies to raise the Table to be convenient for sliding to the ground along the packaging box side cover.

Illustration 3-14 Remove Table Transportation Packaging

3. Delivery the Table to the scan room and close to the Gantry.
4. Use the dollies to lower the Table and set the four level adjusters until the table keeps level. (The distance be-
tween the table base and ground is no less than 17mm.)

Illustration 3-15 Table Adjuster Setting

Proper Anchor Position


Alternative Anchor Position

196

762
no less than 17mm

3-14 SECTION 3 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

3-3 TABLE INSTALLATION (CONTINUED)


5. Align the Table with the Gantry to meet mounting holes.

Illustration 3-16 Align Table with Gantry

Positioning
Subsystem
Mounting Holes (x2)

6. Loosen caster support bolts to lower the table until the adjusters rest on the floor. Note that caster support
bolts must be evenly loosened.
7. Remove the caster side frame and longitudinal frame as shown in Illustration 3-17; remove the two caster side
frames by unscrewing its four support bolts, then remove the caster longitudinal frames by unscrewing its four
table base mounting bolts.

Illustration 3-17 Table Dolly Removal

Caster Side Frame


Caster Longitudinal Frame

CAUTION
After removing the dolly and before anchoring, the table lacks balance. Therefore, do not
place anything onto the cradle until the table is fixed with the anchors.

SECTION 3 - INSTALLATION STEERING GUIDE 3-15


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

3-3 TABLE INSTALLATION (CONTINUED)


8. Temporarily tighten the table mounting bolts to the Gantry base with the washers (5321489) shipped with
the table base cover packaging securely.

CAUTION
Before installing anchors the table might be lack in balance when the cradle moves in or out.
Therefore, the table mounting bolts must be tighten securely.

Illustration 3-18 Install the Table to the Gantry

Mounting Bolt with Washer (x2)

3-16 SECTION 3 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

3-4 OC INSTALLATION
1. Remove all the transportation packaging, place the OC so that the operator can see the patient on the table
from the operator console.
2. Place the Scan Control Box, Monitor, keyboard, and mouse on the OC table as shown in Illustration 3-19. This
Illustration is an example.

Illustration 3-19 Position of OC

Monitor
Table (Option)

Scan Control Box

Positioning
Subsystem
Keyboard Mouse

Host

SECTION 3 - INSTALLATION STEERING GUIDE 3-17


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

3-5 ADAPTER INSTALLATION (IF NO ADAPTER, PLEASE IGNORE THIS SECTION)


1. Remove all the transportation packaging, delivery the Adapter to the scan room or the control room and close
to the PDB.
2. Verify that there is enough clearance around the Adapter for service access and leave at least 10cm between
Adapter and walls for air circulating.
3. Turn the four adjusters located near the four casters until they contact the floor, so that the Adapter does not
slide.
4. If site specifications require seismic mounting, use the Adapter seismic brackets that were shipped with the
Adapter. Refer to Illustration 3-20 for hole placement.

Illustration 3-20 Seismic Adapter Mounting Hole Locations

Seismic Bolt Anchor


Locations, both Sides

5. Remove the Adapter top and front covers.


6. Set the voltage in the Adapter as follows:
a. Check the supplied voltage at Power Distribution Box.
Nine kinds of input rating voltages (200V, 220V, 240V, 380V, 400V, 420V, 440V, 460V and 480V) can be
selected.
b. Remove the protection plate from the transformer terminal, then set the Adapter input rating voltage by
changing voltage selection cable as shown in Illustration 3-21.

3-18 SECTION 3 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

3-5 ADAPTER INSTALLATION (CONTINUED)

Illustration 3-21 Example for 380V Input Voltage

Positioning
Subsystem
6 5 4 3 2 1H3 6 5 4 3 2 1H2 6 5 4 3 2 1H1

For other inputs rating voltage, please select appropriate transformer terminal to connect according to the
following specification.

3 Phases Input 200V 220V 240V 380V 400V 420V 440V 460V 480V
Phase 1 1H1 1--6 1--5 1--4 2--6 2--5 2--4 3--6 3--5 3--4
Phase 2 1H2 1--6 1--5 1--4 2--6 2--5 2--4 3--6 3--5 3--4
Phase 3 1H3 1--6 1--5 1--4 2--6 2--5 2--4 3--6 3--5 3--4

SECTION 3 - INSTALLATION STEERING GUIDE 3-19


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION

3-20 SECTION 3 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

SECTION 4 - WIRING AND POWER-ON

4-1 SYSTEM POWER, GROUND AND SIGNAL CABLE CONNECTION


This section describes the sequence of system cabling and power-on test. Complete as much of the Contractor
supplied wiring as possible before starting the system cabling. Some Contractor supplied wiring, for example
ground cables, must be pulled along with the system cables. In addition, some raceway work, such as cutting and
fitting covers at equipment locations, must be finished after completing the cabling and placing the equipment in
position.
All system cables use a system component designator for identification. The end of each system cable has a label
near the connector that identifies the CABLE NUMBER and CONNECTOR NUMBER.

Wiring and Power-On

SECTION 4 - INSTALLATION STEERING GUIDE 4-1


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-2 CABLE INTER-CONNECTION

4-2-1 Introduction

WARNING
ELECTROCUTION ! TURN OFF THE SWITCH OF THE POWER DISTRIBUTION BOX AND TAG
IT TO VERIFY THAT ALL POWER IS DISCONNECTED TO THE SYSTEM.

When performing system and ground wiring, ensure that all wiring conforms to the information contained in this
manual, in particular to the tables and illustrations contained in this section. The tables in this section list:
• All contractor supplied cables
• System furnished cables
• GEHW Supplied cable Part Numbers
The illustrations provide detailed information on the actual connector and terminal strip locations. Illustrations
include :
• System Interconnect Diagram
• OC Cable Connection
• Gantry Cable Connection
1. Verify that all ground wiring and other contractor wiring are complete to the point of equipment placement.
2. Check for reasonable cable routing. Don’t forget to leave enough cable take-up for equipment maintenance,
etc. Perform as neatly a job as possible.
Note
Keep signal and control cables away from power cables and power wiring. When laying cables in
a raceway, locate the signal cables in a separate section of the raceway.
3. Verify that both ends of each cable are clean cut and properly located before terminating and connecting them.
4. Ensure that all connections are tight. Use suitable tools. Check and tighten all visible connections, especially
ground connections.

NOTICE
Attach a tie-wrap to each system cable within 30 mm of each terminal end.

4-2 SECTION 4 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-2-2 System Cable Guidelines

NOTICE
Check all the ground connections on the system for tightness and proper termination.
Check ground connection at the power supply transformer in a hospital. Check ground from
transformer to its earth connection.
Use the following standard identification in this Installation Manual.

Component Designator Component


Basic System SG Scanning Gantry
PT Patient Table
OC Operator Console
ADAPTER (Option) Adapter

Wiring and Power-On


Customer Supplied PDB Power Distribution Box

Illustration 4-1 System Cable Interconnection

TO POWER SOURCE

PDB Primary System X-RAY ON DOOR SAFETY


Power Emergency LIGHT SWITCH
Disconnect OFF

Adapter (Option)

240VAC, 5A
24VDC, 200mA or MORE

Gantry TGP CN11 TABLE PS RL1 Table


PDU 24V TABLE RL2
Gantry and Table GND
PDU OC POWER

DTB J3
TGP CN5

Connector
UPS (Option) To SCB J1 SCB J1 To TGP
OC POWER IN

HOST CON
OC NIC

OC SCB Fix Cable CONTRACTOR OR


USER SUPPLIED
NIC USB/COM

ETHERNET PHONE LINE


TRANCEIVER FOR INSITE

SECTION 4 - INSTALLATION STEERING GUIDE 4-3


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-2-3 System Cable Checklists


Table 4-1 is the inter-connect lists for power/signal cables. For the system inter-connect diagram, see Illustration
4-2.

NOTICE
In China, all cables used to provide system power and ground must be CCC certified.

WARNING
THE CUSTOMER SUPPLY THE NEUTRAL LINE FROM DISTRIBUTION TRANSFORMER TO
POWER DISTRIBUTION BOX CAN NOT BE OPENED. (NO ANY FUSES AND BREAKERS ON
NEUTRAL LINE.)
NO ANY OTHER LOADS ON CT POWER SUPPLY.

Table 4-1 System Power/Signal Cables

Conductors
Cable Connec- cross-sectional
Part Num- Part Num- Rated Actual
No FROM TO Length tor Size Notes
ber ber Area Voltage Voltage
(m) (mm)
2
(mm )
PDU (L1, L2,
- Customer Supplied
L3, N)
1 PDB - - - 22~35 AC 600V AC 380V (3 Power Cables & 1
(or Adapter) Neutral Cable)
-
(L1, L2, L3)
GroundUL1431
2 PDB - Ground - - - 22~35 AC 600V -
Customer Supplied
3* Adapter - PDU 5389328 22 16.50.2 - - - Power Cable
4* Adapter - Ground 5389329 22 - - - - Ground Cable
Adapter Emergency
5* Adapter - PDU 5389330 22 16.50.2 - - -
Stop Cable
TGP,
6 - SCB, J1 5321390 150.4 55.5x17 - - - Signal Cable
CN5
5321623/
7 DTB, J3 - OC NIC 150.2 - - - - Network Cable
5321623-2
5332759/ 350.2
8 PDU, J3 - LCD1/OC 16.50.5 - - - Power Cable
5332759-2 260.2
Table Driver
Gantry
9 - Board 5322528 - - - - - Power Cable
PS RL
24VDC
TGP, 5321396/ Signal Cable for
10 Table, RL2 5322518 - - - - -
CN11 5321396-3 speaker
Gantry,
11 5327740 Table, GND 5326620 - - - - - Ground Cable
GND
PDU
12 - OC 5428144 150.3 - - - - Ground Cable
GND
*: These parts with “*” are used with adapter as option.

4-4 SECTION 4 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

Illustration 4-2 System Interconnect Diagram

GANTRY OC
From PDB
or Adapter PDUBOX WORKSTATION
PDUBOX L1 OC POWER POWER IN
PDUBOX L2 PDU GND OC GROUND
PDUBOX L3
DTB J3 NIC
PDUBOX N
PDUBOX G1 SCB

SCB
TGP CN5 (FIX CABLE)

TABLE
GANTRY PS +24V/GND PS RL1
TGP CN11 RL2

Wiring and Power-On


REAR BASE GROUND FRAME GROUND

POWER CABLE PDUBOX N


GROUND CABLE SIGNAL CABLE

Note
• All units must be isolated from building structures.
• Cable specifications are found in Tables 4-1.
• Numbers within a circle represent cable number.
• Cable No. assigned to each system cable identified in Table 4-1 System Power/Signal Cables.

SECTION 4 - INSTALLATION STEERING GUIDE 4-5


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-2-3 System Interconnect Diagram (Continued)


The system will only support DICOM Laser Camera. The operation console could connect with a DICOM Laser
Camera directly or through a Switch or Hub.

Illustration 4-3 Laser Camera Interconnect Diagram


Note
For detail, refer to the Laser Camera Installation manual.

Laser Camera OC

Network Network (RPA)

Switch/Hub

4-6 SECTION 4 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-2-3 System Interconnect Diagram (Continued)


The system will support InSite connection with broad band.
Network connection like ADSL.
An ADSL modem is necessary.

Illustration 4-4 InSite Modem Interconnect Diagram

OC

Ethernet ADSL modem

Line

Wiring and Power-On


TO PHONE TACK

Illustration 4-5 Advantage Windows Interconnect Diagram

ADVANTAGE WINDOWS OC

RJ45 Cable RJ45 Cable


(10/100/1000BASE-T) (10/100/1000BASE-T)
ETHERNET HUB ETHERNET

SECTION 4 - INSTALLATION STEERING GUIDE 4-7


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-2-4 OC Cable Interconnection


1. Remove the rear little cover of LCD monitor by unscrewing its one screw and route the LCD monitor signal ca-
ble to the Host PC.
2. Route the mouse and keyboard cables to the Host PC.
3. Route the following cables to the Host PC:

Between Units Remarks


Cable No.
From To
5332759 PDU, J3 Host PC and Monitor Power Cable
(Illustration 4-6) (Illustration 4-7 and 4-8)
5321623 DTB, J3 Host Computer, DTB CON Ethernet Cable
5328642 SCB, J1 DB26 TGP Cable (5321390) Connection to TGP BD
SCB, DB25 Host PC, SCB CON Connection to Host Computer

4. Assemble the rear little cover of LCD monitor by screwing its one screw.

Illustration 4-6 OC Power Cable

OC Power Cable
(To Monitor and Host PC)

4-8 SECTION 4 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-2-4 OC Cable Interconnection (Continued)

Illustration 4-7 Monitor Power Cable Connection

Wiring and Power-On


Fixed Bridge

Illustration 4-8 Rear Panel of Host Computer


Power Cord CON

Mouse (green)

Keyboard (purple)

USB(x4)

Ethernet
GND CON
DTB (for Equipment: Printer)
(from PDU)

Graphics Adapter

Scan Control Box CON

SECTION 4 - INSTALLATION STEERING GUIDE 4-9


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-2-5 Adapter Cables Connection (If No Adapter, Please Ignore This Section)
1. Verify that the main breaker of the Power Distribution Box should be OFF.
2. Switch OFF the breaker on the adapter.
3. Route the three power cables and ground cable from PDB to the adapter.
4. Route the ground cable from Adapter to PDU.
5. Route the four cables (3 power cables and 1 N line cable) from adapter to PDU.
6. Route the E-Stop cable from adapter to PDU.

Illustration 4-9 Adapter Cables Connection

Power Switch

Note:Shield of E-Stop Cable must


be fixed on the Adapter Frame.

Power Cables Input Terminal


(PDB to Adapter)

Ground Cables Input/Output Terminal Power Cables Output Terminal


(PDB to Adapter, Adapter to PDU) E-Stop Terminal 3 380V Power Cables, 1 N Line Cable
(Adapter to PDU)

4-10 SECTION 4 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-2-6 PDU Cables Connection


1. Verify that the main breaker of the Power Distribution Box should be OFF.
2. Switch OFF the breakers on the PDU.
3. Remove the PDU Cover by unscrewing its screws.
4. Connect the four cables (3 power cables and 1 N line cable) from PDB or adapter to the PDU.
5. Connect the E-Stop Cables from Adapter to PDU E-Stop Terminal.
a. If system has no adapter, please ignore this step.
b. If system has adapter, remove the short pins from the E-Stop Terminal, then connect the E-Stop Cables.
(Shown as Illustration 4-10)

Illustration 4-10 PDU Power Cables Connection

Wiring and Power-On


Short Pins

380V High Voltage


Input Terminal
(from PDB or Adapter)

Input Power Cable


(from Adapter)

N Line Terminal 380V High Voltage


Output Terminal

6. Connect the ground cable to the Terminal on PDU Box.

Illustration 4-11 Ground Cable

From OC

From PDB or Adapter From PDU

SECTION 4 - INSTALLATION STEERING GUIDE 4-11


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-2-7 Gantry / Table Cable Connection

Illustration 4-12 Gantry and Table Cable Connection

Ground Cable
Single Cable Connector

Power Cable Connector (24V)

1. Connect the Signal and Power Cable Connectors between the Gantry and the Table.
2. Connect the Ground Cable between the Gantry and Table.
3. Use tie-wraps to tie the Cables.

4-12 SECTION 4 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-3 INDICATION LIGHT / SWITCH CONNECTION


The Indication light or switch below can be connected to this CT system.
• X-ray-ON Light
• Scan room door safety switch
• System ready

4-3-1 System X-ray-ON Light and Scan Room Door Safety Switch

System X-ray-ON Light


A light can be connected to the Gantry to indicate when X-ray ON is initiated. System controlled contacts (RL2) are
provided within the Gantry.

NOTICE

Wiring and Power-On


The Terminal Rating (T3-1 and T4-1):
250VAC, 5A and AWG22 to AWG12

Terminals Contact

X-ray ON X-ray OFF


T4 1 1
WT (T3-1 and T4-1)
T3
9 9
5 5

Connect the X-ray ON light to the Teminals T3-1 and T4-1 located at the Gantry rear.

Scan Room Door Safety Switch


When the scan room door is opened, the X-ray exposure is NOT permitted.

NOTICE
The Terminal Rating (T1-1 and T2-1): AWG22 to AWG12.
Note
The following chart describes relationship between system status and contact open / close.

Door: CLOSED Door: OPEN


Terminals (X-ray Exposure: OK (X-ray Exposure: NG Selection
Rotor ON: OK) Rotor OFF: NG)

WT (T1-1 and T2-1) T8-1

T7-1 (default)

Connect the Scan Room Door Safety Switch to the Terminal T1-1 and T2-1 located at the Gantry rear.

SECTION 4 - INSTALLATION STEERING GUIDE 4-13


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-3-1 System X-ray-ON Light and Scan Room Door Safety Switch (Continued)
Note
The example below shows when N.O. (Normal Open; with the switch OPEN at door closed) type
switch is connected. If N.C. (Normal Close; with the switch CLOSED at door closed) type is used,
change connection of the T7-1 to the T8-1.

Illustration 4-13 Indication Lights Connection

T1

T8

+24VDC
㩺㪈㪏 㪮㪟㪫
㩺㪈㪏 㪙㪣㪢
㩺㪈㪏 㪩㪜㪛
㩺㪈㪏 㪰㪜㪣
TGP
Relay 3 㩺㪈㪏 㪙㪩㪥
Relay 2 Relay 1 㩺㪈㪏 㪙㪣㪬
Wire Terminal
㪈㪋
㪂 㪚㪦㪤
㪩㪜㪣㪘㪰㪈 㪐
㪛㪚㪉㪋㪭

㪠㪥㪧㪬㪫㪈 㩿㪛㪦㪦㪩㩷㪪㪮㪠㪫㪚㪟㪄㪛㫆㫆㫉㩷㪺㫃㫆㫊㪼㪀
㪮㫀㫉㪼㩷㪫㪼㫉㫄㫀㫅㪸㫃 㪌 㪈
㪘㪮㪞㪉㪉㪄㪘㪮㪞㪈㪉 㩺㪈㪏 㪩㪜㪛
㩺㪈㪏 㪙㪣㪬 㪈㪊 㪥㪅㪦㪅

㪫㪈㪄㪈 㪫㪈㪄㪉 㩺㪈㪏㩷㪞㪩㪥
㪦㪬㪫㪧㪬㪫㪈 㩿㪙㪘㪰㩷㪣㪘㪤㪧㪀 㪘㪮㪞㪉㪉㪄㪘㪮㪞㪈㪉 㪫㪉㪄㪈
㩺㪈㪏㩷㪞㪩㪥
㪫㪉㪄㪉 㩺㪈㪏㩷㪙㪣㪢 㪚㪦㪤
㪉㪌㪇㪭㪘㪚㩷㪤㪸㫏㪃㩷㪌㪘㩷㪤㪸㫏 㪫㪊㪄㪈 㪫㪊㪄㪉 㪈㪊 㪄

㪩㪜㪣㪘㪰㪉
㪛㪚㪉㪋㪭

㪫㪋㪄㪈 㪫㪋㪄㪉
㪦㪬㪫㪧㪬㪫㪉 㩿㪩㪼㪸㪻㫐㩷㪪㫋㪸㫋㫌㫊㩷㪣㪘㪤㪧㪀 㪘㪮㪞㪉㪉㪄㪘㪮㪞㪈㪉 㪫㪌㪄㪈 㪫㪌㪄㪉
㩺㪈㪏㩷㪮㪟㪫 㪈 㪌

㪫㪍㪄㪈 㪫㪍㪄㪉 㪥㪅㪦㪅㪈㪋 㪂


㪉㪌㪇㪭㪘㪚㩷㪤㪸㫏㪃㩷㪌㪘㩷㪤㪸㫏
㪫㪎㪄㪈 㪫㪎㪄㪉
㩺㪈㪏㩷㪙㪩㪥 㪚㪦㪤
㪫㪏㪄㪈 㪫㪏㪄㪉 㪈㪊 㪄

㪩㪜㪣㪘㪰㪊
㪛㪚㪉㪋㪭

㩺㪈㪏㩷㪰㪜㪣㩷 㪈 㪌

㪥㪅㪦㪅㪈㪋 㪂

4-14 SECTION 4 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-4 GROUND CONTINUITY CHECK

WARNING
VERIFY THAT WALL POWER (IN PDB) IS OFF, TAGGED, AND LOCKED.

Use an earth bonding test meter or equivalent device to measure the DC resistance between the following
components in the system, and verify that each is less than 0.1 ohm. Record resistance values in Table 4-2.

Ground Specification: less than 0.1 


Also check if those ground cables are securely connected at each end.

Wiring and Power-On


Table 4-2 Ground Continuity Check

Resistance () Connection OK?


From To (Check here when (Check here when
complete.) complete.)
GND (PDB) GND (PDU or Adapter_Option)
GND (Adapter_Option) PDU
GND (PDU) GND (GANTRY)
GND (GANTRY) GND (TABLE)

SECTION 4 - INSTALLATION STEERING GUIDE 4-15


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-5 CONTROL PANEL CONNECTION


1. Disconnect two cable connectors from the control panel (J1 and J2).
2. Remove the control panel from the gantry front cover by unscrewing its 4 bolts.

Illustration 4-14 Control Panel Removal


Connector_J2 Connector_J1

Bolt (x4)
S-DISP Dummy

3. Hang the control panel on the support pole of the Gantry.


4. Plug the S-DISP Dummy (5329766) to the control panel J2.
5. Connect the cable connector (5321391, from TGP to S-DISP) to the control panel J1 for its service position.

Illustration 4-15 Place Control Panel to Service Position


J2
J1

Support Pole Front Control Panel Cable (From TGP to Control Panel)

4-16 SECTION 4 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-6 SYSTEM POWER


Measure phase to phase line voltage is 380V and phase to neutral nominal input voltages is 220V on the power
distribution box (hospital supply).

WARNING
DON’T MANUALLY ROTATE THE GANTRY IF “SERVICE” SWITCH IS OFF STATE,
ESPECIALLY WHEN GANTRY IS POWERING ON.

Switching ON the System


• Remove the azimuth lock pin.

Wiring and Power-On


Azimuth Lock Pin

• Power ON the PDB.


• Switch ON the Adapter power switch.
• Power ON Gantry

13 14
5 6
3 4 R3
1 2 R2
K11 R1
A1 A2
RMT J8

TS1 F1 TS2
CB1 NZMB1 A80 13 14
L1 1 2 5 6 L1
F2
L2 3 4 3 4 L2
L3 5 6 1 2 L3
F3
N K1
RMT J8 A1 A2

GND

CB6 D10 CB5 D4


1 2 220V 1 2
RMT J5
115V 1 2
SR 115V
CB4 D16
T1
CB7 D10
1 2
OC 220V
CB3 D16
1 2
Sv 220V (Axial Drive)
CB2 D6
1 2
Gantry P.S 220V

NEUTRAL
K2 N10
1 2
P.S 24V Table 24V

RMT J1 RMT J8
A1
K2
N A2

SECTION 4 - INSTALLATION STEERING GUIDE 4-17


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-6 SYSTEM POWER (CONTINUED)


- Switch ON ‘CB1’ on PDU (Centralized control power)
- Switch ON ‘CB2’ on PDU (Gantry PS)
- Switch ON ‘CB3’ on PDU (Axial Drive)
- Switch ON ‘CB4’ on PDU (Slip Ring)
- Switch ON ‘CB5’ on PDU (RMT)
- Switch ON ‘CB6’ on PDU (Control CB4 and CB5)
- Switch ON ‘CB7’ on PDU (OC)
- Switch ON Safety Loop (XG Power) switch on PDU.
• Power ON OC
- Switch ON the main switch of the console at the backside.
- Push power on button in front of console.

Switching OFF the System

Click the Shutdown Icon shutdown , located on the upper left of the screen. Then, click [OK] button in response to
“Attention: Shut down the system.”.
The system is automatically switched OFF.
If you want to switch off the gantry. Please switch off ‘CB2’ and ‘CB4’, and then switch off ‘CB1’.

NOTICE
Please make sure tube temperature is not too high when you decide to switch off ‘CB4’ or
you may need special care the tube to avoid being scald.

4-18 SECTION 4 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-7 PRODUCT KEY / OPTION SOFTWARE/PRODUCT PROTOCOL INSTALLATION


The system is shipped with a standard of option software. So, at installation, you must configure the system using
the system Product Key DVD-RAM, then install the option software if necessary using the option DVD-RAM.
To install the system Product key DVD-RAM and option DVD-RAM(s), refer to Functional Check/Adjustment,
Introduction, Installing an Option Newly to the Installed-base System.
1. Install the Option DVD-RAM(s).
Includes one option software, depending on the option kit
a. Application shutdown.
b. On the desktop menu, click on Install Options button located on the upper right of the screen.
c. The ‘CT Software Options‘ screen appears.

Wiring and Power-On


d. Click on Install.

SECTION 4 - INSTALLATION STEERING GUIDE 4-19


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-7 PRODUCT KEY / OPTION SOFTWARE/PRODUCT PROTOCOL INSTALLATION (CONTINUED)


Note
If you click on an option in the list, the option is highlighted and “Install“ button is disabled. In this
case, enter the cursor in the option screen, then click on the right button of the mouse to enable the
“Install“ button.
e. The ‘Select mechanism’ window appears, select Permenent.
• Permanent: selected when a customer purchased the option formally.
• Flex Trial: selected when a customer uses the option within a fixed time period as a trial. (The message
that notices the expiration date will be displayed over 10 days before the expiration of a Flex Trial.)

f. The ‘Select Device’ window appears, select DVD.


• DVD: selected when the Option key DVD is used at installation.
• Manual: selected when the licence string is used at installation. The licence string is a unique number
with 25 characters that you get from your OLC.

4-20 SECTION 4 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-7 PRODUCT KEY / OPTION SOFTWARE/PRODUCT PROTOCOL INSTALLATION (CONTINUED)


g. The following screen appears. Insert the option key DVD into the DVD drive, then click on OK.

h. The ‘CT Software Options‘ screen appears.

Wiring and Power-On

i. Select the Option to be installed from the “Available Options” table.


ii. Click on Install.
The option loading is started, then the message window appears after completion. The option(s) is
displayed in the “Installed Options” table.
iii. Click on OK.
iv. Click on Quit.
v. When other option(s) is installed, repeat all steps above.

SECTION 4 - INSTALLATION STEERING GUIDE 4-21


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-7 PRODUCT KEY / OPTION SOFTWARE/PRODUCT PROTOCOL INSTALLATION (CONTINUED)

GE HANGWEI MEDICAL SYSTEMS CO., LTD.


No. 2 NORTH YONG CHANG STREET
BEIJING ECONOMIC-TECHNOLOGICAL DEVELOPMENT ZONE
BEIJING, 100176, P.R.CHINA

DESCRIPTION: xxxxxxxx
MODEL NO. xxxxxxx
SERIAL NO. xxxxxxxxxx
MANUFACTURED: MONTH YEAR

2. Install Product Protocol.


a. Insert the Protocol disk into DVD-RAM drive in the Host PC.
b. Select Service  Utility  Selective Save.
c. In Type column, select Restore.
In Filesets column, select Protocol.
d. Click on Start to restore the protocols.
e. After restore completed, click on quit to exit.

4-22 SECTION 4 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-8 EMERGENCY OFF TEST


After this test, press the reset button on the OC to resume the system.
1. Power ON the system.
2. Move the cradle IN from its home position.
3. Press one of the emergency stop switches on the Gantry.
4. Verify that power is removed from the Gantry and Table. Verify the following:
• LEDs on the TGP board are extinguished.
• Gantry display panel is extinguished.
5. Press one of the cradle drive buttons, to verify the emergency stop disables the cradle drive.
6. Move the cradle to the home position by hand, to verify the emergency stop released the cradle clutch.
7. Make sure the cradle latches securely in the home position.

Wiring and Power-On


8. Press the Reset button on the TGP board or use a pencil to press the emergency stop reset switch on the SCB.
Make sure the cradle drive now work.

Illustration 4-16 Location of Emergency Stop Reset Switch

Emergency Stop Reset Switch

SECTION 4 - INSTALLATION STEERING GUIDE 4-23


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-9 MONITOR SET-UP


Note
The monitor is normally shipped with adjustment except Contrast and Brightness. Therefore,
Contrast and Brightness MUST be adjusted to the level an user desires at Installation.

4-9-1 Keys Assignment

Illustration 4-17 OSD Menu

Exit

Power Switch

ITEM Main Function Action

Menu Activate OSD


Always Jump to next line

Up Slide Controller Increase value

Selection Point To previous selection

Command "Enter" key

Down Slide Controller Decrease value

Selection Point To subsequent selection

Exit Except "Exit OSD" One menu level upwards (settings are retained)
Menu

In "Exit OSD" Menu Return to main menu (settings are retained)

4-9-2 Lock of OSD Menu

Keys Action

1 time Exit key and 3 times UP key Lock or unlock OSD

4-24 SECTION 4 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-9-3 Monitor Gamma Value


The gamma value for Monitor display should be set 1.8, 2.0 and 2.2 (default).

Recommending keeping the default value.

Gamma 2.2 Contrast 128


(Recommended)
Brightness 128

Backlight 254

Color Temperature Color 1: 6500K

4-9-4 Monitor Color Temperature


This menu allows you to adjust the white in the video.

Wiring and Power-On


4-9-5 Contrast and Brightness (Minor Adjustment)
Recommending adopting the factory default level of contrast and brightness, no need to adjust them unless you
think it is necessary assuredly to be adjusted.
The adjustable scope of contrast and brightness and backlight is from 0 to 255.
Note
Any adjustment of contrast and brightness besides default value will change the gamma curves.

Contrast
The contrast level adjustment processes as:
1. Opening OSD when it is unlocked via pressing 'Menu' key
2. When Brightness is selected, the scroll bar of contrast comes out in 1st line in right area.
3. Selecting 'contrast' via pressing 'Up' key.
4. Then adjusting 'Up' key to increase or 'Down' key to decrease contrast level to desirable one.
5. Pressing 'Menu' key to exit item, then pressing 'Up' key to enable it, two selectable changes 'Reject Changes'
and 'Accept Changes' come out in right area. Pressing 'Menu' key again to highlight 'Accept Changes', and
executing via 'Up' key.

Brightness
The brightness level adjustment processes as:
1. Also selecting Brightness as above, the scroll bar of brightness comes out in 2nd line in right area.
2. Selecting 'brightness' via pressing 'Up' key, then 'Menu' key.
3. Then adjusting 'Up' key to increase or 'Down' key to decrease brightness level to desirable one.
4. Pressing 'Menu' key to exit item, then pressing 'Up' key to enable it, two selectable changes 'Reject Changes'
and 'Accept Changes' come out in right area. Pressing 'Menu' key again to highlight 'Accept Changes', and
executing via 'Up' key.

SECTION 4 - INSTALLATION STEERING GUIDE 4-25


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-10 JEDI DATE & TIME CORRECTION


Note
The purpose of this procedure is to make the JEDI and System time/date the same.
1. Verify that the Generator test screen appears. If it does not, click on Service icon to display the Service Menu
desktop, then select Diagnostics  Generator Test .
2. Select Set JEDI time, then click on Apply.
3. Compare the JEDI time/date displayed and system’s one. The system’s time/date are displayed in the left
column on the monitor.
If these time/date were not same, set the JEDI time/date to the system’s one:
(If system time must be adjusted, click Date Setting from desktop menu located on the upper right of the
screen, adjust the time, click Set then Quit.)
a. Click on SYS, then Get time.
b. Immediately click on JEDI, then Set time.
c. Verify that the message “Time was set” appears, then click on Close.
d. Click on Close to terminate the set-JEDI-time window.

4-26 SECTION 4 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-11 SYSTEM CONFIGURATION

4-11-1 Hardware Connection


For cable connection, see Illustration 4-2.

4-11-2 System Setting


1. To display the desktop menu:
Click on Service icon to display the Service Menu desktop, then select Utilities  Shutdown  Application
Shutdown.
2. On the desktop menu, click on Reconfig button located on the upper right of the screen.

NOTICE
The characters of alphanumerics, underscore, and period ONLY can be used. If other

Wiring and Power-On


characters would be used, the system will not start up. (Other characters are rejected to
input even if they are input.)This system configuration menu appears.
3. Select System to display System Setting screen:
Use this screen, perform configuration: Hospital Name, Time Zone and Machine #.

SECTION 4 - INSTALLATION STEERING GUIDE 4-27


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-11-2 System Setting (Continued)


4. Select Preference to display Preference Settings screen:
Using this screen, perform configuration: Time Format, Units for Patient Weight, Date Format, etc..

Software Version: 1.65 Software Version: 2.2x

5. Select Lanuage to display Language Settings screen:


Using this screen, perform configuration: Language and Default AutoVoice Language.

Software Version: 1.65 Software Version: 2.2x

4-28 SECTION 4 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-11-2 System Setting (Continued)


6. Select Hardware to display Hardware Settings screen:
Using this screen, perform configuration: Hardware Diagnosis, Keyboard and Start Angle for Head Scan.

NOTICE
The following window is displayed on initial configuration. 
For the software version 2.2x, under Initial Laser Position Angle field, select the correspond-
ing initial laser position angle according to gantry mode number. If the wrong configuration
is selected on initial configuration, you must Load-From-Cold to re-select this configuration.
• If Gantry mode number is 5323923-1 or 5324907-1, please select 0.
• If Gantry mode number is 5323923 or 5324907, please select 90.

Wiring and Power-On


Software Version: 1.65 Software Version: 2.2x

SECTION 4 - INSTALLATION STEERING GUIDE 4-29


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-11-2 System Setting (Continued)


7. Select Network to display Network Settings screen:

le
amp
Ex

4-30 SECTION 4 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-11-2 System Setting (Continued)


a. Click on Network on.
The check box becomes red.
b. OC Parameters Setting:
i. Enter Host Name.
Host Name identifies the name of this CT. This name must be unique for the site, and will appear in
UNIX prompts and in a few other places. If you set up a network connection to the site, the OC
computer will appear on the site’s network as <Host Name>.
ii. Enter IP Address.
The IP address is administrated by a network administrator. Ask a network administrator to assign an
IP address.
Note

Wiring and Power-On


IP address is appointed exclusive network information.
iii. Enter Net Mask.(if required.)
The Net Mask is to divide the IP address between the network bits and the host bits. The Net Mask is
determined by a network administrator.
c. If this system is not connected to a “Site Subnet” then;
• Verify all other fields are set to the defaults:

OC - Net Mask Gateway - Host Name Gateway - IP Address

255.255.255.0 gw 192.168.10.1

• You should now skip to Step 8 (Initiate Reconfiguration).


d. If this system is connected to a “Site Subnet”, then continue as follows:
Gateway Parameters Setting:
• Enter the Gateway Parameters (IP Address). If possible consult the site’s network administrator before
configuring the interface to the site’s network. The site administrator should provide the Gateway IP
Address.
e. If customer wants to synchronize the system time to their NTP server, then select Enable Network Time Protocol,
and enter the Primary Server IP address.
8. Initiate Reconfiguration:
a. Click on Accept after completing the configuration.
“Installing Product sw” window appears.
The system configuration procedures are started, then automatically login as root. It takes approx. 3 minutes.

SECTION 4 - INSTALLATION STEERING GUIDE 4-31


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-12 CAMERA SETTING


1. On the desktop menu, click on Startup button located on the upper right of the screen to launch Application.
2. On the service desktop menu, select Configuration  Install Camera.
3. The Install Camera window appears, along with a warning message pop-up box, to remind you that all filming
queues must be empty before you begin to update or delete a camera.
First, you must click OK in the warning message box.

Illustration 4-18 Warning

Illustration 4-19 Camera Installation Window

4. The Graphical User Interface displayed shows a list of cameras installed. (See Illustration 4-19)
5. Click on Add button to add a new camera. (See Illustration 4-19)

4-32 SECTION 4 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-12 CAMERA SETTING (CONTINUED)


6. A dialog window for the camera type (DICOM/PostScript) appears.

Illustration 4-20 Dialog Box for Camera Type

7. DICOM is selected.

Wiring and Power-On


a. A list of camera models appears (See Illustration 4-21). Select the appropriate model from the list and click
Select. Clicking Select presets all the parameters to that models except the network parameters. (See
Illustration 4-22)
Selection of a different camera model clears the Image Quality parameters, because these are camera
manufacture dependent.

Illustration 4-21 Camera Model Dialog (DICOM)

SECTION 4 - INSTALLATION STEERING GUIDE 4-33


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-12 CAMERA SETTING (CONTINUED)


b. Enter the Network Parameters.
• Device Name: A unique name used to identify the camera.
• Host Name: DICOM print server host name, as defined by the hospital.
• IP Address: DICOM print server IP address, ad defined by the hospital.
• Application Title: DICOM print server application entity title, as defined by the print server. You should
consult the manufacturers DICOM Conformance Statement.
Note
The Application Entity Title for the camera may be site specific. Make sure that you check with the
Camera Manufacturer’s Representative and the hospital network administrator to ensure you are
using the correct AE Title for the destination DICOM Print Camera.
• TCP/IP Listen Port: DICOM print server TCP/IP listen port, as defined by the server. You should consult
the manufacturers DICOM Conformance Statement.
• Comments: Optional comments used by the DICOM print server.

Illustration 4-22 DICOM Camera Configuration

4-34 SECTION 4 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-12 CAMERA SETTING (CONTINUED)


c. Medium Type specifies the type of film being used. Currently, only Blue Film and Clear Film are support-
ed.
d. Set Destination to the final location for film output: either Processor or Magazine.
e. Film Orientation selects film orientation. Only Portrait is currently supported.
f. Set the Magnification Type. This parameter selects the algorithm used to interpolate pixels for proper film
resolution. Set this parameter after consulting the camera manufacture to ensure optimal image quality.
Choices are described below:
• None: No interpolation. This option is not supported by all cameras.
• Replicate: Adjacent pixels are interpolated. This can result in “pixelized” images. This algorithm is not
normal preferred.

Wiring and Power-On


• Bilinear: A 1st order interpolation of pixels. Results in images usually described as blurred. This algo-
rithm is not normal perferred.
• Cubic: A 3rd order interpolation. Used with a large number of possible formulations. Camera manufac-
tures define parameters called “smoothing type” to set coefficients used in this algorithm. The imple-
mentation of these “smoothing coefficients” is camera dependent.
g. Select the Standard Film Formats. Select the film format by choosing ON in the pull down menu box lo-
cated below each selection. (See Illustration 4-22) Valid film formats are set by the camera manufacture.
Example: IMATION doesn’t support 1x2, 2x4 or 4x4; AGFA doesn’t support 4x2.
The DICOM print convention designates film formats by column and row. (e.g. 12 on 1 film is 4x3)
h. After the camera has been configured, click the Advanced button. This creates the camera device file for
you automatically and pops up the Advanced Parameters screen. (See Illustration 4-23)

Illustration 4-23 Camera Attribute Dialog

SECTION 4 - INSTALLATION STEERING GUIDE 4-35


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-12 CAMERA SETTING (CONTINUED)


i. Advanced camera parameters control the image quality of films.
Note
For more information on the proper settings for these parameters, please see the Camera’s DICOM
Print Device Conformance Statement or the Camera Manufacturer’s Representative.
You may need to refer to a copy of the Conformance Statement as you are working with the Camera
Manufacturer’s Representative, to correctly set up the DICOM Print Camera I/Q and Time-out
Settings.
• Configuration: This parameter is camera manufacturer dependent as is typically used to specify the
image contrast. The Configuration field may be up to 1024 characters long. The field will scroll
automatically as text is entered. To review your entry, simply click and hold the middle mouse button,
while the cursor is in the field, and drag the mouse towards the right (or left) as needed.
Recommended Configuration Setting Values:
Agfa Drystar (MG3000) Perception_LUT=KANAMORI (100)
Imation Dryview (8700) LUT=0, 7
Kodak Laser Printer 190 CS434\CN0\PD1.20
• Minimum and Maximum Density: Used to set brightness of the images on film. The range of values
is 0~4095, although the valid range for a specific camera is manufacture dependent. 
For Maximum Density, input the correct value into the text box. 
For Minimum Density, set it to ON and input the correct value in the text box.
Recommended Minimum and Maximum Density Starting Values:
Agfa Drystar (MG3000) Min: 20 or 23; Max: 300
Imation Dryview (8700) Min: (Blank); Max: 300
Kodak Laser Printer 190 Min: 20; Max: 300
• Smoothing Type: Set it to ON, and input the selected value. This parameter is used when the
magnification type is CUBIC. It represents the coefficient for the image resolution algorithm. This
parameter is camera manufacturer dependent, and should be re-verified with your radiology department.
Recommended Smoothing Type Starting Values and Ranges:
Agfa Drystar (MG3000) Start Value: 140; Range: 137~150
Imation Dryview (8700) Start Value: 3; Range: 3~13
Kodak Laser Printer 190 Start Value: Enhanced; Range: Normal
• Empty Density: This parameter sets the density for empty film viewports. Typically, BLACK is used but
WHITE is an available option. The minimum and maximum density values are used as the
representation.
• Border Density: This sets the density for the border used around the film viewports. Typically, BLACK
is used but WHITE is an available option. The minimum and maximum density values are used as the
representation.
• Film Size: Allows the system to specify a particular film size, if selected.
• Trim: YES produces a white (clear) box surrounding each image.
• Priority: This sets the print priority.
j. If you have completed entry of advanced parameters, click Done.
Note
You can determine camera information by looking at the contents of the following files:
For DICOM Print Camera: /usr/g/ctuser/app-defaults/devices/name.cfg, name.cfg is the cam-
era device name from the printer configuration GUI.
4-36 SECTION 4 - INSTALLATION STEERING GUIDE
Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-12 CAMERA SETTING (CONTINUED)


8. PostScript is selected.
a. Enter the Network Parameters.

Illustration 4-24 PostScript Camera Configuration

Wiring and Power-On


• Device Name: A unique name used to identify the camera.
• Host Name: PostScript print server host name, as defined by the hospital.
• IP Address: PostScript print server IP address, ad defined by the hospital.
• Queue Name: Keep system default value.
• Comments: Optional comments used by the PostScript print server.
b. Select the Standard Film Formats. Keep system default value. (See Illustration 4-24) Valid film formats
are set by the camera manufacturer.
c. If you have completed entry of parameters, click Ok.

SECTION 4 - INSTALLATION STEERING GUIDE 4-37


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

4-13 CONNECTION THE SYSTEM TO THE NETWORK

4-13-1 Hardware Connection


Connect a 10Base-T cable to the Ethernet port of Host PC.

4-13-2 Remote Host Setting


1. Select ImageWorks  Network  Select remote host.
2. “Remote host parameter” windows appears.
3. Click Add for new network entry, then enter the appropriate parameters:

Name of the devise the CT scanner is connected to. Ask a network


Host Name
administrator.
Address of the devise the CT scanner is connected to. Ask a network
Network address
administrator.
Network protocol For DICOM devises, select DICOM.
When the CT is connected to the Advantage Windows, enter 4006.
Port number
When connected to other devises, ask a network specialist.
Comment Any comment can be entered.

4. Click Save.
5. Verify that it is available to transmit the image data from the CT scanner.
6. Verify that it is available to execute Query/Retrieve from the DICOM workstation side.

4-38 SECTION 4 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

SECTION 5 - GANTRY AND TABLE INTEGRATION

5-1 TABLE TO GANTRY ALIGNMENT


The following procedure adjusts the Table position in the right-left direction so that the Table Center line and the
scan center are aligned. See Illustration 5-1.
After completing the adjustment, anchor the Table.

Illustration 5-1 Table Right-Left Position


GANTRY

Scan Center

Gantry and Table


Integration
TABLE

Right-Left
Table Center Line Direction

WARNING
ROTATION HAZARD!
WHEN YOU PRESS THE ‘POSN LIGHT’ BUTTONS, THE GANTRY QUICKLY ROTATES TO
THE 0 DEG. POSITION. IF IT IS NECESSARY TO PLACE ANY PART OF YOUR BODY INSIDE
OR NEAR THE GANTRY, SWITCH OFF THE ‘CB3’ SWITCH FIRST.
1. Align ISO height.
a. Advance the cradle into the Gantry.
b. Perform an axial scan of the cradle using the following technique:
Scan Parameters: Axial, 3 sec, 120kV, 60mA, 10mm thickness, 43cm DFOV (other default)

SECTION 5 - INSTALLATION STEERING GUIDE 5-1


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

5-1 TABLE TO GANTRY ALIGNMENT (CONTINUED)


c. Check on the image of distance between the top surface of cradle and gantry ISO center shall be 135cm ±
5mm. If not, adjust the table leveling adjuster to make sure the distance is in spec.

135mm

2. Perform an axial scan of the cradle at least 3 points (the front of cradle, the middle of cradle, the end of cradle)
using the following technique:
Scan Parameters: Axial, 3 sec, 120kV, 60mA, 10mm thickness, 43cm DFOV (other default)
3. Check whether the table center line and the scan center are aligned:
Verify that the difference in the distances (|1-2|) is within 2mm. If not, proceed to next step.

1 2

4. Adjust the Table right-left position:


a. Loosen the Table base two mounting bolts. (See Illustration 5-2)
b. Move the Table to center the phantom holder in the scan plane, according to step 3.
c. Tighten the Table base mounting bolts securely.

5-2 SECTION 5 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

5-1 TABLE TO GANTRY ALIGNMENT (CONTINUED)


d. Return to step 2.

Illustration 5-2 Table and Gantry Mounting Bolts

Gantry Base

Table Base Mounting Bolt (x2)

Gantry and Table


Integration

SECTION 5 - INSTALLATION STEERING GUIDE 5-3


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

5-2 ANCHORING OF TABLE

NOTICE
Be sure to have the appropriate hospital personnel approve the location of the table/Gantry
before starting to drill any anchor holes.

DANGER
CRASH HAZARD!! THE TABLE AND GANTRY MUST BE ANCHORED.

5-2-1 Precautions
• Complete the Table to Gantry Alignment before anchoring the Table.
• Prior to anchoring, after drilling the anchoring holes, check that the Gantry and Table are tightly fixed
together by two bolts.

5-2-2 General
The following hardwares are provided with the system.
• Table Standard Anchor kit:
Anchor bolts (5339079)
Washers (5321489)
Anchor nuts (5321490)
Leveling adjuster (5321487)
Leveling Pad (5321488)

5-2-3 Table Anchoring


This anchoring procedure is used for the anchor kit shipped with the system.
Note
The other anchor can also be used, instead of the anchor kit shipped with the system. In this case,
you can use the other anchoring procedures that are not described in this manual, but the minimum
tensile load strength of the anchor MUST meet the specification below.

NOTICE
The minimum Tensile load strength of the anchor must be 13000 N.

5-4 SECTION 5 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

5-2 ANCHORING OF TABLE (CONTINUED)


1. Attach a piece of tape to the 12 mm concrete drill bit (2203081: HILTI, TE-CX 12/17 or equivalent) in order to
assure a minimum of 100 mm depth of anchor hole.
Note
The concrete drill bit length is 250mm is recommend. The total length for electric drill and concrete
drill bit is less than 600mm is recommend for fit the space of the table.

Illustration 5-3 12 mm Concrete Drill Bit Preparation

100mm
Tape
2. Make holes at the anchorposition, using electric drill with the bit prepared inthe previous step, so that the lower

Gantry and Table


edge of the tape just touch the floor surface.

Integration
NOTICE

During drilling, always keep the drill bit perpendicular to the floor. With one person drilling
the hole, the second person can visually insure perpendicularity throughout the drilling
operation. Clear the hole of debris several times while drilling to prevent the drill bit from
binding.
Note
If any obstruction is found under the proper anchoring position, stop drilling and use the alternate
position (see illustration 5-4). Use the alternate positions only when the case its corresponding
proper position is not possible.

Illustration 5-4 Table Anchor Positions (Proper and Alternative)

PROPER ANCHOR POSITION

ALTERNATIVE ANCHOR POSITION

SECTION 5 - INSTALLATION STEERING GUIDE 5-5


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

5-2-3 Table Anchoring (Continued)


3. Use the vacuum cleaner to suck up the concrete dust from the hole. Also vacuum dust from the surrounding
area when finished.
4. Place the bubble level at the positions.
5. Level the Table using the Gantry/Table Anchor Adjuster JIG. (Refer to Gantry anchor adjustment)

Level Specification: +/- 0.2 degree

Illustration 5-5 Table Adjuster and Bubble Level Position

Position of Bubble Level

6. Install the Anchor. (Refer to Gantry Anchoring.)

Illustration 5-6 Table Anchor Assembly

Anchor Bolt

Nut
Anchor Washer
no less than 17mm

Adjuster Lock Ring


90mm
100mm

Leveling Adjuster

7. Verify that the Table to Gantry Alignment should be correct.


(Refer to Section 5-1, TABLE TO GANTRY ALIGNMENT.)

5-6 SECTION 5 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

5-3 ROTATION OPERATION


Refer to:
Functional Check/Adjustment, Gantry, Rotational Operation, ROTATION SPEED.

5-4 CRADLE ABSOLUTE POSITION


Refer to
Functional Check/Adjustment, Table, Longitudinal Function, CRADLE ABSOLUTE POSITION.

Gantry and Table


Integration

SECTION 5 - INSTALLATION STEERING GUIDE 5-7


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION

5-8 SECTION 5 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

SECTION 6 - TUBE ALIGNMENT

6-1 PLANE OF ROTATION


Refer to:
Functional Check/Adjustment, System, X-Ray Alignment, PLANE OF ROTATION.

6-2 DETECTOR BEAM-ON-WINDOW


Refer to:
Functional Check/Adjustment, System, X-Ray Alignment, DETECTOR BEAM-ON-WINDOW.

Tube Alignment
6-3 Q-CAL CHANNEL RATIO (ONLY FOR BRIVO CT325)
Refer to:
Functional Check/Adjustment, System, X-Ray Alignment, Q-CAL CHANNEL RATIO.

6-4 ISO CENTER ALIGNMENT


Refer to:
Functional Check/Adjustment, System, X-Ray Alignment, ISO CENTER ALIGNMENT.

6-5 FILTER CENTER ALIGNMENT


Refer to:
Functional Check/Adjustment, System, X-Ray Alignment, FILTER CENTER ALIGNMENT.

SECTION 6 - INSTALLATION STEERING GUIDE 6-1


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION

6-2 SECTION 6 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

SECTION 7 - SYSTEM TEST

7-1 POSITIONING LIGHTS


Refer to:
Functional Check/Adjustment, Gantry, Table and Gantry, POSITIONING LIGHTS.

7-2 INTERCOMMUNICATION
This includes the following main tests:
• OC speaker test, including tests for Auto Voice, X-ray ON, and Alert
• Gantry speaker test, including Auto Voice

System Alignment
1. Select [Service]  [Diagnostics]  [H/W Diag (Shutdown)]  [Interactive Test]  [Audio Test].
The Audio test screen appears.

and Test
2. To return to [Interactive Test Menu], click on [Back].
Note
To exit from [H/W Diag (Shutdown)] menu, the system must be rebooted so that the [Exit] button
must be used to shut down the system.

OC Speaker - Auto Voice


3. Click on [Show Player] for Auto Voice.
4. Click on [Play]. The voice message, “Breathe in ...” will be heard from the OC speaker.
5. While playing auto voice message, move the L and R volume sliders for Auto voice so that the message will
become loud or faint.
Note
When the [Play Back] button is ON, a sound via the Gantry microphone (patient voice) will NOT be
heard from the OC speaker. This function is one for recording, so that the [Play Back] button MUST
be OFF when exiting from this test if it is performed.
6. Click on [File] -> [Exit] to exit from this test.

SECTION 7 - INSTALLATION STEERING GUIDE 7-1


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

7-3 TABLE LATCH CHECK


To perform the following procedure, a person should ride on the cradle, or some substitutes should be placed on
the cradle.
1. Move the cradle to the Out limit position by the ‘OUT’ button.
2. While moving the cradle in the In direction by continuously pushing the ‘IN’ button, perform the following:
a. Push the latch switch (Free). See Illustration 
Verify that the cradle stops moving, and that the Gantry display goes blink; this indicating that the cradle is
unlatched.
b. Push the latch switch again (Latch).
c. Release the ‘IN’ button, and then push it.
Verify that the cradle starts moving and the Gantry display changes to normal (displaying a position in mil-
limeter).
3. Repeat the above same check for the latch switch on the opposite side.

Illustration 7-1 Latch Switch Location

One Latch Switch


per side (x2)

7-2 SECTION 7 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

7-4 SCAN TEST


This ’Scan Test’ procedure assumes that all the subsystems (Operator Console, Gantry, Table, DAS/Detector, and
X-ray generator) are functioning normally.
In this procedure refer to the Operator manual for detailed information.

Warm Up Scan
1. Make sure that nothing is in the scan area in the Gantry room.
2. Click on the ‘Tube Warm up’ icon to start the x-ray tube warm-up program.
3. Press the blinking [Start Scan] button on the OC keyboard to start the warm-up scan.
It will take approx. 1.5 minutes to complete scans.
Make sure that all warm up scans complete without error.
4. Click on [OK] button to return to the initial screen.

System Alignment
Scoutview Scan
5. Perform a series of scout scans to check scout scan operations:

and Test
a. Perform a scout scan. (Any scan parameters can be set.)
b. Lay the QA phantom horizontally on the cradle.
Perform an AP (0) scout scan on the phantom.

Verify that the phantom diameter (In-Out direction) on the image is 214 mm  3 mm.
c. Set the QA phantom on the phantom holder.
Perform an LT (90) scout scan on the phantom.
d. Use the [Landmark] button on the Gantry control panel to perform the phantom positioning so that
positioning light and center of the phantom are aligned.
e. Select [Suspension] for Autovoice.
f. Press the [Landmark] button and verify that:
• The cradle can NOT be moved.
• The cradle position is the same value as the external landmark.
g. Click on Confirm, then verify that the [Move to Scan] button blinks.
h. Press the [Move to Scan] button, then press the [Stop Move] button before the cradle reaches the set
position. Verify that the cradle stops moving and the [Move to Scan] button blinks again.
i. Press the [Move to Scan] button and move the cradle to the set position. Verify that the [Start Scan] button
blinks.
j. Perform the scan and verify that the scan can complete with no error.

SECTION 7 - INSTALLATION STEERING GUIDE 7-3


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

7-4 SCAN TEST (CONTINUED)

Axial Scan Sequence


6. Perform a series of axial scans to check axial scan operations:
a. Set the QA phantom on the phantom holder.
b. Use the [Landmark] button on the Gantry control panel to perform the phantom positioning so that
positioning light and center of the phantom are aligned.
c. Select [Inspiration] for Autovoice.
d. Use the following scan parameters: (Other parameters can be set to any value.)
Thickness: :10 mm
kV :120 kV
mA :150 mA
Interval :10 mm
Scan time :3 sec
Position :Feet First
e. Click on Confirm, then verify that the [Start Scan] button blinks.
f. Press the [Start Scan] button and verify that:
• You can hear and understand auto voice for scan start.
• The scan starts.
• Both X-ray ON indicators on the OC keyboard and Gantry panel is lit and X-ray ON alarm starts to sound.
g. Press the [Pause] button during scanning and verify that:
• The scan stops at next ISD and the [Confirm] button is changed to the [Resume] button.
• You can hear and understand auto voice for scan end.
h. Click on Resume, then verify that the [Start Scan] button blinks again.
i. Press the [Start Scan] again to continue scanning.
j. Press the [Stop Scan] button and verify that:
• The scan stops IMMEDIATELY and the [Confirm] button is changed to the [Resume] button.
• Both X-ray ON indicators on the OC keyboard and Gantry panel is OFF and X-ray ON alarm stops to
sound.
• You can hear and understand auto voice for scan end.
k. Click on End Exam and verify that the rotation of the Gantry and rotor in the X-ray tube stops to rotate.

7-4 SECTION 7 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

7-4 SCAN TEST (CONTINUED)

Axial Scan
7. Perform a series of axial scans to check axial scan operations:
a. Set the QA phantom on the phantom holder.
b. Use the following scan parameters: (Other parameters can be set to any value.)

Acquisition No. 1 2 3 4 5
Scan Time [sec] 1.0 1.5 2.0 3.0 5.0

Thickness :10 mm


kV :120 kV
mA :150 mA
Interval :10 mm

System Alignment
Position :Feet First
No. of Images :2

and Test
c. Verify that each scan completes with no error.

7-5 CALIBRATION

7-5-1 Hilight Calibration


Refer to:
Functional Check/Adjustment, System, System Calibration, HILIGHT CALIBRATION.

7-5-2 Air and Phantom Calibration


Refer to:
Functional Check/Adjustment, System, System Calibration, AIR AND PHANTOM CALIBRATION.

7-6 CT NUMBER ADJUSTMENT


Refer to:
Functional Check/Adjustment, System, System Calibration, CT NUMBER ADJUSTMENT.

SECTION 7 - INSTALLATION STEERING GUIDE 7-5


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

7-7 IMAGE PERFORMANCE CHECK


1. Acquire the scans listed in Table 7-1.
Select ‘Abdomen’ or any anatomy of interest for ‘Infant.’
The phantoms to be scanned should be centered.
2. For each slice, verify the following:
• Visually check each image for rings, bands, streaks, shadings or artifacts near the image center.
• Position a circular ROI at the locations listed below (see also Illustration 7-2) and write down the data on
Table 7-1.
- Mc-Mean CT number of a center ROI.
- SD-Standard deviation of a center ROI.
- Mo-Mean CT number of a ROI 10 cm off center (Y = 0).

• The following ROI sizes (area) are used to measure ROI data.

FOV ROI AREA


25 cm 674.11 mm2
43 cm 1994.30 mm2

• To display a ROI data;


i. Position a cursor (here, a circle cursor).
ii. Press the [ROI] key.
• Verify that the data falls within the specifications shown in Table 7-2 to 7-3.
Note
The CT# of water will change according to its temperature by a factor of -0.4/C.

High Contrast Spatial Resolution and Artifact Check Using QA (Quality Assurance) Phantom
3. The 1.0 mm pattern (see Illustration 7-3) must be clearly visible by the following techniques:

Table 7-1 Scans List

Kind of Scan SFOV DFOV kV Time mA THK Interval Recon


(cm) (cm) (sec) (mm) (mm)
Axial Head 25 10 120 2.0 150 10 - PFRM
Helical Head 25 10 120 2.0 150 10 10 PFRM
Axial Large 43 10 120 1.5 130 10 - PFRM
Helical Large 43 10 120 1.5 130 10 10 PFRM

Check if any artifact is observed on the image or not.

7-6 SECTION 7 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

7-7 IMAGE PERFORMANCE CHECK (CONTINUED)

Illustration 7-2 ROI Locations

+Y ROI Circle
(Area of 674.11 mm2 or 1994.30 mm2)
Image

-X +X
Mc (0.0 cm)
Mo (10.0 cm)
SD (0.0 cm)

System Alignment
and Test
-Y

Illustration 7-3 QA Phantom

1.0 mm Pattern

SECTION 7 - INSTALLATION STEERING GUIDE 7-7


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION

7-8 SECTION 7 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

For Brivo CT315

Table 7-2 Image Performance Check

FOV*1 TIME THK SPECIFICATION DATA


PTM kV mA Description
(cm) (sec) (mm) MC SD MO MC SD MO

Water part of QA phontom Large 120 1 200 10 0±4 <4 MC ± 3

Water part of QA phontom Large 120 1 20 10 - < 15 - -

Water part of QA phontom Large 140 1 160 10 0±4 <4 _ -

Water part of QA phontom Small 120 1 200 10 0±4 < 5.5 MC ± 2

Water part of QA phontom Small 120 1 20 10 - < 20 _ -

Water part of QA phontom Small 120 1.5 190 7 0±4 < 5.5 _ -

Water part of QA phontom Small 120 2 200 5 0±4 < 5.5 _ -

Water part of QA phontom Small 120 3 200 3 0±4 <6 _ -

Water part of QA phontom Small 120 3 200 2 0±4 <8 _ -

Water part of QA phontom Small 120 5 200 1 0±4 <9 _ -

Water part of QA phontom Small 140 1 40 10 0±4 < 10 _ -

Water part of QA phontom Small 120 1.5 150 10 - < 3.5 _ - Axial scan

Water part of QA phontom Small 120 1.5 150 10 - < 3.5 _ - Helical scan
Water part of QA phontom Small 120 2 200 10 - <6 _ -
*1: Large: 43 cm - Set the water part of QA phantom 70mm left from ISO center.
Small: 25 cm - Set the water part of QA phantom 40mm left from ISO center.

7-8 INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

For Brivo CT325

Table 7-3 Image Performance Check

SPECIFICATION

FOV*1 TIME THK 2i mode DATA


PTM kV mA Mode 1i mode Description
(cm) (sec) (mm) A B |Delta (A– SD ratio
MC MC MC B)| (A/B) MC
SD SD SD SD
Water part of
QA phontom
Large 120 1.5 140 10 2i 04 < 5.5 04 < 5.5 5 1 ± 0.1

Water part of
QA phontom
Large 120 1.5 140 10 1i 04 < 5.5 -

Water part of
QA phontom
Large 120 1.5 200 7 2i 04 < 5.5 04 < 5.5 5 1 ± 0.1

Water part of
QA phontom
Large 140 1.5 120 10 2i 04 04 5

Water part of
Large 140 1.5 120 7 2i 04 04 5
QA phontom
Water part of
QA phontom
Large 120 1.5 40 10 2i 04 04 5

Water part of
QA phontom
Large 120 1.5 60 7 2i 04 04 5

Water part of
Large 120 1.5 80 5 2i 04 04 5
QA phontom
Water part of
QA phontom
Large 120 1.5 100 3 2i 04 04 5

Water part of
QA phontom
Large 120 1.5 100 1 2i 0±4 0±4 5

Water part of
Large 140 1.5 40 10 1i 04
QA phontom
Water part of
QA phontom
Large 140 1.5 70 6 1i 04

Water part of
QA phontom
Large 140 1.5 80 4 1i 04

Water part of
QA phontom
Large 140 1.5 100 2 1i 04

*1: Large: 43 cm - Set the water part of QA phantom 70mm left from ISO center.
Small: 25 cm - Set the water part of QA phantom 40mm left from ISO center.

INSTALLATION STEERING GUIDE 7-9


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

Table 7-3 Image Performance Check (Continued)

SPECIFICATION

FOV*1 TIME THK 2i mode DATA


PTM kV mA Mode 1i mode Description
(cm) (sec) (mm) A B |Delta SD ratio
(A–B)| (A/B)
MC SD MC SD MC SD MC SD
Water part of
QA phontom
Small 120 1.5 140 10 2i 04 < 5.5 04 < 5.5 5 1 ± 0.1

Water part of
QA phontom
Small 120 1.5 140 10 1i 04 < 5.5

Water part of
QA phontom
Small 120 1.5 40 10 2i 04 04 5

Water part of
QA phontom
Small 120 1.5 60 7 2i 04 04 5

Water part of
QA phontom
Small 120 1.5 70 5 2i 04 04 5

Water part of
QA phontom
Small 120 1.5 80 3 2i 04 04 5

Water part of
QA phontom
Small 120 1.5 90 2 2i 04 04 5

Water part of
QA phontom
Small 120 1.5 100 1 2i 04 04 5

Water part of
QA phontom
Small 140 1.5 40 10 1i 04

Water part of
QA phontom
Small 140 1.5 70 6 1i 04

Water part of
Small 140 1.5 80 4 1i 04
QA phontom
Water part of
QA phontom
Small 140 1.5 100 2 1i 04

*1: Large: 43 cm - Set the water part of QA phantom 70mm left from ISO center.
Small: 25 cm - Set the water part of QA phantom 40mm left from ISO center.

7-10 INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION

7-11 INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

SECTION 8 - INSTALLATION COMPLETION

8-1 COVER INSTALLATION


Refer to Component Replacement manual for detailed installation procedure.

8-1-1 Gantry
Install the following covers in the reverse order of removal.
• Front and Rear Covers
• Top Covers
• Left and Right side Covers

NOTICE

Completion
Installation

When installing the Gantry front cover, first cover the front edge of the cradle with a cloth
to prevent the front cover from scratches or damage.

CAUTION
If the service angles (cover stand) fall over, they can injure persons. Therefore, after using
the service angles, keep them in a safe place and/or secure them from falling over even when
the customer or other persons accidentally hit the service angles.

8-1-2 Table

Install the following covers:


• Left and Right Base Covers
• Bottom Cover
• Speaker Cover

SECTION 8 - INSTALLATION STEERING GUIDE 8-1


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

8-2 ATTACHING LABELS AND EMBLEMS


1. Attach the Scan Control Box Film (5397664) shipped with OC onto the SCB.
(Choose the appropriate language)

Illustration 8-1 Attaching the SCB Film

2. Check and attach appropriate language caution labels.


Before attaching label, thoroughly clean the machine surface where to attach the label.
Refer to the table below for proper location and attaching method.

NOTICE
In order to avoid any violation of each National Regulation, the appropriate language caution
labels MUST be attached.

Note: BRIVO CT315/CT325 dose not have Pinch Point Caution Label.
Please ignore it, if this label is delivered in the product package.

8-2 SECTION 8 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

8-2 ATTACHING LABELS AND EMBLEMS (CONTINUED)

Description PN Illustration Location Qty


Read Docs Cau- 5442200- Attach on gantry front 4
tion Label EN cover, table, operator
console and Adapter
! CAUTION (Option).
The specific language
AVOID INJURY. caution labels should be
Read and understand
information in manuals applied next to the
before operating product. corresponding English
labels.
Do NOT overlap any
portion of the English
label.
Weight Maxi- 5442206- Attach on Table. 1
mum caution EN

Completion
The specific language

Installation
label caution labels should be
! CAUTION applied next to the
corresponding English
AVOID INJURY. labels.
Do Not Exceed Table Do NOT overlap any
>180kg Maximum Capacity portion of the English
of 180 kg (397 lb).
label.

PDU Movement 5442205- Attach on Adapter. 2


Caution Label EN The specific language
caution labels should be
! CAUTION applied next to the
corresponding English
PDU CAN MOVE AND
labels.
DAMAGE CABLES.
Do NOT overlap any
Do not lean on or move
portion of the English
when connected to power.
label.

Type B Applied 5428394 Attach on Table. 1


Part Label

Cellular Phone 2152854 Attach on to p of the 1


Caution Label Host PC, Chinese label
was attached in manu-
factory before ship ping,
attach appropriate lan-
Attach appropriate
language label
guage label beside Chi-
beside Chinese
label
nese label

SECTION 8 - INSTALLATION STEERING GUIDE 8-3


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

8-2 ATTACHING LABELS AND EMBLEMS (CONTINUED)

Description PN Illustration Location Qty


Laser Caution 2351331 Attach on g antry front 1
Label cover, English and Chi-
nese labels were
attached in manufactory
before shipping, attach
appropriate language
label beside English
label.
5159035

Table Pinch 5409638 Attach on table top cov- 2


Point Caution (2199652) ers, Chinese label was
Label attached in manufactory
before shipping, attach
5401918
appropriate language
(5129092)
label beisde Chinese
label, each side of table
has one.

8-4 SECTION 8 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

8-2 ATTACHING LABELS AND EMBLEMS (CONTINUED)

Description PN Illustration Location Qty


Cradle Latch 5401919 Attach on side of table 2
Caution Label (2210375) top covers, Chinese
label was attached in
ϡৃ⿏ࡼ
⌏ࡼᑞᵓ
manufactory before
shipping. Attach appro-
ৃ⿏ࡼ

priate language la bel


beside Chinese label,
each side of t able has
one.

Table Extender 2198038 English and Chinese 1


Caution Label label were attached in

Completion
Installation
manufactory before
shipping. Attach appro-
priate language label
beside English label or
on upside surface as
pointed by the arrow.

P9235RM COR- 2198037 Attach appropriate lan- 1


ONAL H.H Cau- guage label beside Eng-
tion Label lish label.
(Option)

SECTION 8 - INSTALLATION STEERING GUIDE 8-5


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

8-2 ATTACHING LABELS AND EMBLEMS (CONTINUED)

Description PN Illustration Location Qty


Rating Plate 5411812 Select and attach 3
Comparison (5411392-SPN) Rating plate comparison appropriate language
Label (5411392-FRE) English Spanish labels to the following
(5411392-CRO) Desc/Description Descripción
Model Modelo N° positions shown as
(5411392-ITA) Source Alimentación
S/N˄ᑣ߫ো˅ Nº SERIE below.
(5411392-GER) Manufactured Fecha de fabricacion
˄⫳ѻ᮹ᳳ˅
(5411392-POR) Volts Volts
Potencia momentánea
Momentary Power
Continuous Power Potencia continuo
Momentáneo

p le
Momentary
Continuous Continua
m
E xa

Three copies of the


same rating plate
labels are a ttached
respectively on Gan -
try, OC and Apdapter.
Rating plate comparison
English Spanish
Desc/Description Descripción
Model Modelo N°
Source Alimentación
S/N˄ᑣ߫ো˅ Nº SERIE
Manufactured Fecha de fabricacion
˄⫳ѻ᮹ᳳ˅
Volts Volts
Momentary Power Potencia momentánea
Continuous Power Potencia continuo
Momentary Momentáneo
Continuous Continua

Rating Plate
for Gantry

Rating Plate
for OC
Rating plate comparison
English Spanish
Desc/Description Descripción
Model Modelo N°
Source Alimentación
S/N˄ᑣ߫ো˅ Nº SERIE
Manufactured Fecha de fabricacion
˄⫳ѻ᮹ᳳ˅
Volts Volts
Momentary Power Potencia momentánea
Continuous Power Potencia continuo
Momentary Momentáneo
Continuous Continua

Attach appropriate
language label
beside Chinese
label

Rating Plate
for Adapter

Rating plate comparison


English Spanish
Desc/Description Descripción
Model Modelo N°
Source Alimentación
S/N˄ᑣ߫ো˅ Nº SERIE
Manufactured Fecha de fabricacion
˄⫳ѻ᮹ᳳ˅
Volts Volts
Momentary Power Potencia momentánea
Continuous Power Potencia continuo
Momentary Momentáneo
Continuous Continua

Front Rear

8-6 SECTION 8 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

8-3 SYSTEM VERIFICATION


Click on Service icon to display the Service desktop, then select Configuration  Verify Options.

8-4 LANGUAGE SELECTION


The system is shipped with English language. If other language is required to display, appropriate language must
be selected.
1. To display the desktop menu:
Click on Service icon to display the Service Menu desktop, then select Utilities  Shutdown  Application
Shutdown.
2. On the desktop menu, click on Reconfig button located on the upper right of the screen.
The System configuration menu appears.
3. Select Preferences.

Completion
Installation
Software Version: 1.65 Software Version: 2.2x

SECTION 8 - INSTALLATION STEERING GUIDE 8-7


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

8-4 LANGUAGE SELECTION (CONTINUED)


4. Select the Language for this site.
5. Select the Default AutoVoice Language for this site.

Software Version: 1.65 Software Version: 2.2x

6. Initiate Reconfiguration:
a. Click on Accept after completing the configuration.
“Installing Product sw” window appears.
The System configuration procedures are started, then automatically login as root. It takes approx. 3
minutes.

8-8 SECTION 8 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

8-5 BACK-UP SYSTEM STATE DATA


The basic system data (System State data) must be saved in advance in case the System Hard Disk Drive in the
OC is broken.
Note
The following will be saved as a system state data.
- Option key information
- Log files (log for success/failure when saving/restoring software sets)
- Scan protocol files
- IOS preferences
- Calibration files
- Autovoice files
- Next patient exam number
- User preferences
- InSite information
- Tube usage information
- Crosstalk A, B channel data (for dual slice system)
- TnT Definition File log (for dual slice system)

Completion
Installation
- Q cal channel ratio (for dual slice system)

8-5-1 Preparation
• One DVD-RAM

8-5-2 Procedures
Method I
1. Verify that the system is powered ON. If it is not, switch ON the main switch of the OC. The system is started
up, then the main screen appears.
2. Insert the DVD-RAM into DVD drive or plus USB disk to PC.
3. Click on Service menu Uilities.
4. Select Save System State.
5. “Do you want to save system state?” appears, click on Yes.
6. “Please choose the storage device for saving system state!” appears, click on DVD or USB to continue.
7. “This is an Optical Disk containing a file system.” appears. Click on Continue.
Note
At this time, you can cancel system state data saving procedure. If ‘Continue’ is selected, this pro-
cedure can NOT be canceled.
8. “Warning: Saving system state will erase the previous contents” appears. Click on Okay.
9. “System state saved successfully” appears. Click on Okay.

SECTION 8 - INSTALLATION STEERING GUIDE 8-9


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

8-5 BACK-UP SYSTEM STATE DATA (CONTINUED)


Method II
1. Click on Service icon to display the Service Menu desktop, then select Utilities -> Shutdown -> Application
Shutdown.
2. Click on Savestate.
3. “Do you want to save system state?” pops up. Click on Yes.
4. Insert the DVD-RAM into the DVD-RAM drive or plus the USB disk to Host PC.
5. “Please choose storage device for saving system state!” appears, select DVD or USB to saving system state.
6. “This is an Optical Disk containing a file system.” appears. Click on Continue.
Note
At this time, you can cancel system-state-data saving procedure. If ‘Continue‘ is selected, this
procedure can NOT be canceled.
7. ‘Warning: Saving System state will erase the previous contents.‘ appears. Click on Okay.
8. ‘System state saved successfully‘ appears. Click on Continue.

8-6 SYSTEM AND ROOM CLEANING


1. Check scanner, equipment, and control rooms for compliance to temperature and humidity control
requirements.
2. Clean rooms and exterior panels of components.
3. Vacuum clean all electronics cabinets and equipment dust filters. GE recommends a weekly vacuuming of the
dust filters at first since new installation sites experience excessive dust and dirt. Lengthen the interval
between maintenance over a period of time, depending on the individual site environment.

8-10 SECTION 8 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

SECTION 9 - SYSTEM COMPONENT IDENTIFICATION

9-1 COMPONENT IDENTIFICATION


This system consists of the following main and sub compatible components. Each component can be identified by
its rating plate. Refer to the Renewal Parts Manual for other system components/parts.
Brivo CT315/CT325 system has 14 Locator Cards, cover subsystem of Gantry, Table, OC and PDU, not including
options Locator Cards.

Table 9-1 System Component Identification

System Component
COMPONENT PART NUMBER LOCATION OF NAME PLATE LOCATOR

identification
CARDS
IDENTIFIED
Operator Console 5325157 Console Rating Plates Example comply with YES
IEC60601-1:1988+A1+A2 and IEC60601-
1:2005

SECTION 9 - INSTALLATION STEERING GUIDE 9-1


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

Table 9-1 System Component Identification (Continued)

COMPONENT PART NUMBER LOCATION OF NAME PLATE LOCATOR


CARDS
IDENTIFIED
Table 5322519 YES
Rating Plate
WEEE Label

Cradle Tray_Rear

1105mm 5mm

9-2 SECTION 9 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

9-1 COMPONENT IDENTIFICATION (CONTINUED)

Table 9-1 System Component Identification (Continued)

COMPO- PART NUMBER LOCATION OF NAME PLATE LOCATOR


NENT CARDS
IDENTIFIED
Gantry 5323923 (BRIVO CT315) YES
5324907 (BRIVO CT325)
5402419 (India BRIVO
CT315)
5402422 (India BRIVO 80mm 30mm

CT325) SFDA

System Component
System
Rating Plate

identification
PDU
Gantry

Gantry Rating Plates Example comply


with IEC 60601-1: 1988+A1+A2 and
IEC 60601-1: 2005

Gantry Rating Plate for India BRIVO CT

MADE BY
GE MEDICAL SYSTEMS(INDIA) PVT LTD.
#122, PART1, EPIP, WHITEFIELD,
BANGALORE-560066, INDIA

MADE FOR
GE HANGWEI MEDICAL SYSTEMS CO. LTD.
NO.2, NORTH YONG CHANG STREET,
ECONOMIC TECHNOLOGICAL DEVELOPMENT ZONE,
BEIJING, CHINA-100176

DESC : GANTRY
MODEL : xxxxxxx
S/N :
MANUFACTURED :
VOLTS : 380V 3N~ 50/60Hz
KVA MOMENTARY : 40KVA
KVA CONTINUOUS : 15KVA

SECTION 9 - INSTALLATION STEERING GUIDE 9-3


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

9-1 COMPONENT IDENTIFICATION (CONTINUED)

COMPO- PART NUMBER LOCATION OF NAME PLATE LOCATOR


NENT CARDS
IDENTIFIED
PDU 5324901 (380V) PDU Rating Plates Example comply with YES
IEC60601-1:1988+A1+A2
and IEC60601-1:2005

9-4 SECTION 9 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

Table 9-1 System Component Identification (Continued)

COMPONENT PART NUMBER LOCATION OF NAME PLATE LOCATOR


CARDS
IDENTIFIED
X-ray Tube 2232785-2 YES

System Component
identification
Generator 2227720 YES
SW KV Control 2214614-19 YES
JEDI Programmed

SECTION 9 - INSTALLATION STEERING GUIDE 9-5


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

9-1 COMPONENT IDENTIFICATION (CONTINUED)

Table 9-1 System Component Identification (Continued)

COMPONENT PART NUMBER LOCATION OF NAME PLATE LOCATOR


CARDS
IDENTIFIED
HV Tank 2186730 YES

Power Unit 2227721 N/A NO

9-6 SECTION 9 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

9-1 COMPONENT IDENTIFICATION (CONTINUED)

Table 9-1 System Component Identification (Continued)

COMPONENT PART NUMBER LOCATION OF NAME PLATE LOCATOR


CARDS
IDENTIFIED
Collimator 2259785-4 (Brivo CT315) YES
2320783-2 (Brivo CT325)

System Component
identification

SECTION 9 - INSTALLATION STEERING GUIDE 9-7


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

9-1 COMPONENT IDENTIFICATION (CONTINUED)

Table 9-1 System Component Identification (Continued)

COMPONENT PART NUMBER LOCATION OF NAME PLATE LOCATOR


CARDS
IDENTIFIED
Auxiliary unit 2213293 N/A NO
GDAS Assy 5142027-2 (Brivo CT315) YES
5142027 (Brivo CT325)

Detector 2245151 (Brivo CT315) YES


2321043 (Brivo CT325)

9-8 SECTION 9 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

9-1 COMPONENT IDENTIFICATION (CONTINUED)

COMPONENT PART NUMBER LOCATION OF NAME PLATE LOCATOR CARDS


IDENTIFIED
Cradle Extender 2137478 YES
Assy

System Component
19 INCH LCD 5336642 YES

identification
Monitor ASSY 5327908
P9235RM 2201805 YES
CORONAL H.H
(Option)

Axial Head 5334730 N/A NO


Holder (Standard)
2115996-3 YES
(Option)

Phantom Holder 5395684 NO


QA Phantom
QA phantom 2144715 NO
Phantom
Holder

SECTION 9 - INSTALLATION STEERING GUIDE 9-9


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION

9-10 SECTION 9 - INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

APPENDIX A - SYMBOLS AND CLASSIFICATION

Symbol Publication Description


417-5032 Alternating Current

335-1 Three-phase Alternating Current


3

335-1 Three-phase Alternating Current with neutral


conductor
3N

Direct Current

Appendix
417-5019 Protective Earth (Ground)

348 Attention, consult ACCOMPANYING DOCUMENTS

417-5008 OFF (Power: disconnection from the mains)

417-5007 ON (Power: connection to the mains)

Warning, HIGH VOLTAGE

Emergency Stop

INSTALLATION STEERING GUIDE 1


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

Symbol Publication Description

Type B

417-5339 X-ray Source Assembly Emitting

417-5009 Standby

Start

Table Set

Abort

Intercom

(on Operator Console)


Power On: light on
Standby: light off

Warning sign :
Radiation of Laser Apparatus

2 INSTALLATION STEERING GUIDE


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

Symbol Description

Microphone (Mic)

Contrast

Brightness

System storage prior to installation:

Appendix
30°C
Maintain storage temperature between 0C and +30C

0 °C

System storage prior to installation:


Maintain non-condensing storage
humidity below 70%
DO NOT store system longer than 90 days

Humidity 0-70%
excluding Condensation

Air Pressure System storage and shipment:


Maintain Air Pressure between 700 and 1060hPa
700-1060hPa

INSTALLATION STEERING GUIDE 3


Brivo CT315
Brivo CT325
GE HEALTHCARE INSTALLATION
REV 7 5306510-1EN

CLASS 1 EQUIPMENT
Any permanently installed equipment containing operator or patient accessible surfaces must provide backup
protection against electric shock,in case the BASIC INSULATION fails. In addition to BASIC INSULATION,Class1
equipment contains a direct connection to a PROTECTIVE(EARTH) CONDUCTOR which prevents shocks when a
person touches a broken piece of equipment or touches two different equipment surfaces simultaneously.

TYPE B EQUIPMENT
CLASS I, II, or III EQUIPMENT or EQUIPMENT with INTERNAL ELECTRICAL POWER SOURCES provide an
adequate degree of protection against electric shock arising from (allowable) LEAKAGE CURRENTS or a
breakdown in the reliability of the protective earth connection.

ORDINARY EQUIPMENT
Enclosed EQUIPMENT without protection against the ingress of water.

OPERATION 0f EQUIPMENT
CONTINUOUS OPERATION WITH INTERMITTENT LOADING.
Operation in which EQUIPMENT is connected continuously to the SUPPLY MAINS.The stated permissible loading
time is so short that the long term on-load operating temperature is not attained.The ensuing interval in loading is,
however, not sufficiently long for cooling down to the long term no-load operating temperature.
EQUIPMENT not suitable for use in the presence of a FLAMMABLE ANESTHETIC MIXTURE WITH AIR or WITH
OXYGEN or NITROUS OXIDE

CLEANING
The system is NOT WATERPROOF. It is NOT designed to protect internal components against the ingress of
liquids.Clean external system surfaces(Gantry,table,consoles and accessories)with a soft cloth dipped in hot water
and wrung DAMP/DRY. (NOT dripping!) IF NECESSARY, use only mild (dish washing liquid) soap to remove dirt.

NOTICE
Avoid damage to equipment! Some “spray and wipe” cleaners etch and permanently cloud
clear plastic surfaces!! Use only warm water and mild liquid soap to clean surfaces.

4 INSTALLATION STEERING GUIDE


www.gehealthcare.com

Imagination at work

You might also like