GE Healthcare

BRIVO CT315/325
PERIODIC MAINTENANCE
Operating Documentation

5306511-1EN
Rev 2
© 2009-2010, General Electric Company,
All Rights Reserved.
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

警告 • 本维修手册仅提供英文版本。
(ZH-CN) • 如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
• 未详细阅读和完全理解本维修手册之前,不得进行维修。
• 忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他形式的
伤害。

• THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


WARNING
• IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE
(EN) OTHER THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO
PROVIDE TRANSLATION SERVICES.

Attention
• DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS
SERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE
SERVICE PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC
SHOCK, MECHANICAL OR OTHER HAZARDS.

ATTENTION • Ce manuel de service n'est disponible qu'en anglais.


• Si le technicien du client a besoin de ce manuel dans une autre langue
(FR) que l'anglais, c'est au client qu'il incombe de le faire traduire.
• Ne pas tenter d'intervenir sur les équipements tant que le manuel service
n'a pas été consulté et compris
• Le non-respect de cet avertissement peut entraîner chez le technicien,
l'opérateur ou le patient des blessures dues à des dangers électriques,
mécaniques ou autres.

警告 • このサービスマニュアルには英語版しかありません。
(JA) • サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業は
その業者の責任で行うものとさせていただきます。
• このサービスマニュアルを熟読し理解せずに、装置のサービスを行わないで
ください。
• この警告に従わない場合、サービスを担当される方、操作員あるいは患者さ
んが、感電や機械的又はその他の危険により負傷する可能性があります。

ATENÇÃO • Este manual de assistência técnica só se encontra disponível em inglês.


• Se qualquer outro serviço de assistência técnica, que não a gems,
(PT-BR) solicitar estes manuais noutro idioma, é da responsabilidade do cliente
fornecer os serviços de tradução.
• Não tente reparar o equipamento sem ter consultado ecompreendido este
manual de assistência técnica.
• O não cumprimento deste aviso pode por em perigo a segurança do
técnico, operador ou paciente devido a choques elétricos, mecânicos ou
outros.

ATTENTION
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE

ОСТОРОЖНО! • Данное руководство по обслуживанию предлагается только на


английском языке.
(RU) • Если сервисному персоналу клиента необходимо руководство не на
английском, а на каком-то другом языке, клиенту следует
самостоятельно обеспечить перевод.
• Перед обслуживанием оборудования обязательно обратитесь к
данному руководству и поймите изложенные в нем сведения.
• Несоблюдение требований данного предупреждения может привести
к тому, что специалист по обслуживанию, оператор или пациент
получат удар электрическим током, механическую травму или другое
повреждение.

ATENCION • Este manual de servicio sólo existe en inglés.


• Si el encargado de mantenimiento de un cliente necesita un idioma que
(ES) no sea el inglés, el cliente deberá encargarse de la traducción del
manual.
• No se deberá dar servicio técnico al equipo, sin haber consultado y
comprendido este manual de servicio.
• La no observancia del presente aviso puede dar lugar a que el proveedor
de servicios, el operador o el paciente sufran lesiones provocadas por
causas eléctricas, mecánicas o de otra naturaleza.

DIKKAT • Bu servis kilavuzunun sadece ingilizcesi mevcuttur.


• Eğer müşteri teknisyeni bu kilavuzu ingilizce dişinda bir başka lisandan
(TR) talep ederse, bunu tercüme ettirmek müşteriye düşer.
• Servis kilavuzunu okuyup anlamadan ekipmanlara müdahale etmeyiniz.
• Bu uyariya uyulmamasi, elektrik, mekanik veya diğer tehlikelerden dolayi
teknisyen, operatör veya hastanin yaralanmasina yol açabilir.

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In Shipment” on ALL
copies of the freight or express bill BEFORE delivery is accepted or “signed for” by a GE representative or hospital receiving
agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event,
within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company
will not pay a claim for damage if an inspection is not requested within this 14 day period.
To file a report:
Fill out a report on http://egems.med.ge.com/edq/home.jsp
Contact the local service coordinator.

ATTENTION
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT


All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be
performed by licensed electrical contractors. In addition, electrical feeds into the Power Distribution Unit shall be performed
by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations and testing shall
be performed by qualified GE Medical personnel. The products involved (and the accompanying electrical installations) are
highly sophisticated, and special engineering competence is required. In performing all electrical work on these products,
GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the
requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party

Attention
service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.

IMPORTANT! . . . X-RAY PROTECTION


X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein contained should be thoroughly
read and understood by everyone who will use the equipment before you attempt to place this equipment in operation. The
General Electric Company, Medical Systems Group, will be glad to assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the useful beam, no practical
design of equipment can provide complete protection. Nor can any practical design compel the operator to take adequate
precautions to prevent the possibility of any persons carelessly exposing themselves or others to radiation.
It is important that anyone having anything to do with x-radiation be properly trained and fully acquainted with the
recommendations of the National Council on Radiation Protection and Measurements as published in NCRP Reports
available from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and of the
International Commission on Radiation Protection, and take adequate steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Medical Systems Group, its agents, and
representatives have no responsibility for injury or damage which may result from improper use of the equipment.
Various protective materials and devices are available. It is urged that such materials or devices be used.

LITHIUM BATTERY CAUTIONARY STATEMENTS

CAUTION
Danger of explosion if battery is incorrectly replaced. Replace only with the same or equivalent
type recommended by the manufacturer. Discard used batteries according to the manufacturer’s
instructions.

OMISSIONS & ERRORS


Customers, please contact your GE Sales or Service representatives.
GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in this publication.

ATTENTION
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE

ATTENTION
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

REVISION HISTORY
REV Date Primary Reason for Change
1 07/22/2009 Initial release.

2 12/02/2010 System Deleted X-ray buzzer. (PM0102)

OC Added indication for mainboard battery in illustration 2. (PM0205)

DAS/Detector Updated LED in the Detector Temperature Controller Cycles from "D3" to "DS3". (PM0402)

Revision History
X-Ray Generator Updated dip switch S1 from “T6 switch” to “5th switch” for KV&MA display mode. (PM0606)

PDU Updated power line voltage specification to 50/60Hz. (PM0702)

REVISION HISTORY
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE

REVISION HISTORY
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

CONTENTS

GENERAL
Introduction
Tools
CT PM Schedule
CT PM Report
CT PM PROCEDURE
System
PM0101 Verify Emergency Stop Buttons (Gantry/Scan Control Box)
PM0102 Check X-ray On lights / Operation of Scan Abort
PM0103 Caution Label Check

Contents
PM0104 Image Check (Image Performance)
PM0105 Check Error Logs
PM0106 Perform Filter Curve Test
PM0107 Check Ground Cable Terminals
PM0108 Check DAS Count
PM0109 Image Performance
PM0110 Tube Gas Bubble Inspection
PM0111 Phantom Check
Console
PM0201 Verify Audio Function
PM0202 Inspect FANs
PM0203 Verify Mouse for smooth operation
PM0204 Clean Display monitor / Console exterior
PM0205 Check System clock
PM0206 Check Cables / Power Cables
PM0207 Clean the Dust Proof Filters
Gantry
PM0301 Check the number of Gantry revolutions
PM0302 Clean Mylar ring cover
PM0303 Verify Cables and Hardware are tight
PM0304 Gantry Cover Cleaning
PM0305 Check Positioning Light
PM0306 Check Gantry Anchor
PM0307 Main bearing grease-up
PM0308 Check Drive Belt for wear

CONTENTS 7
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

CONTENTS

CT PM PROCEDURE (CONTINUED)
DAS/Detector
PM0401 Check/Clean Detector face
PM0402 Verify Detector heater control
Table
PM0501 Check Table Cover
PM0502 Gap between Table Cover and Cradle
PM0503 Check Cradle Rail
PM0504 Holder Stability Check
PM0505 Verify Unlatch Function
PM0506 Clean Cradle Tray / Table Cover
PM0507 Check Table anchors
XG
PM0601 Check / Clean Radiator FAN
PM0602 Check X-ray tube oil and HV connector
PM0603 Check HV Tank and HV connector
PM0604 Check Power Cable connection
PM0605 Check Tube overheat safety
PM0606 Check KV and mA
PM0607 Check Parts
PDU
PM0701 Check Power Cable Terminals
PM0702 Check Power Line Voltage

8 CONTENTS
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

GENERAL

INTRODUCTION
BEFORE STARTING A PERIODIC MAINTENANCE, READ THE SAFETY GUIDELINES MANUAL THOROUGHLY
TO PREVENT EQUIPMENT DAMAGE, PERSONAL HARM, HARM TO OTHERS OR DEATH DURING SERVICE.
Recommended PM frequency is 2 times a year for Brivo CT315/CT325 scanners. The frequency may vary due to
local ordinance and the usage of the system, the system availability etc. If you need more PMs, order extra copies
of this document or copy the schedules.
TOOLS
GENERAL TOOLS
• Service Manuals
• Socket Head Wrench Set
• Torque Wrench

General
• Protractor
• Oscilloscope
• Digital Multimeter
• Grease-2188551 (Shell Alvania #2 or equivalent)
• Alcohol (ethyl alcohol-no water addition, or isopropyl disinfectant alcohol)
• CRC or equivalent (WD40, etc...)
• Cloth
• Kim Wipe or equivalent
• Soft tissue
• Glass Cleaner
• Dust Bag
• HEPA Vacuum Cleaner
• Film Holder, Developer, and type52 Polaroid Film (For X-ray alignment, adjustment, not used in the PM)
SPECIAL TOOLS
• HV Divider, HV Cables, and Wrench for HV Cable
• Silicone Oil
• Resurfacer (P9183SU)
• Flexible Abrasive (P9183SV)
• Grease Gun

GENERAL 9
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
blank

10 GENERAL
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SCHEDULE

Recommended PM frequency is 2 times a year for Brivo CT315/CT325 scanners. The frequency may vary due to
local ordinance and the usage of the system, the system availability etc. If you need more PMs, order extra copies
of this document or copy the schedules.

Priority Code

Pr (Priority) - Description
1 Safety and Regulatory
2 Image Quality
3 Procedures that make the system unavailable for scanning
4 Procedures that can be performed while the customer is scanning

Frequency Code

Schedule
Fr (Frequency) - Description
S Semi-Annually (every 6 months)
A Annually (every 12 months)

Safety and Regulatory

ITEM SUB-SYSTEM DESCRIPTION Pr Fr Annual PM Schedule


A B
PM0101 System Verify Emergency OFF buttons (Gantry / Console / PDU) 1 S ↓ ↓
PM0102 System Check X-ray ON lights or buzzer/operation of scan abort 1 S ↓ ↓
PM0103 System Caution Label Check 1 S ↓ ↓
PM0105 System Check Error Logs 1 S ↓ ↓
PM0111 System Phantom Check 1 A ↓
PM0201 O.C Verify Audio Function 1 S ↓ ↓
PM0301 Gantry Check the number of Gantry revolutions 1 S ↓ ↓
PM0501 Table Check Table Cover 1 S ↓ ↓
PM0502 Table Gap between Table Cover and Cradle 1 S ↓ ↓
PM0503 Table Check Cradle Rail 1 S ↓ ↓
PM0504 Table Holder Stability Check 1 S ↓ ↓
PM0505 Table Verify Unlatch Function 1 S ↓ ↓
PM0607 XG Check Parts 1 S ↓ ↓

SCHEDULE 11
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SCHEDULE (CONTINUED)

Image Quality

ITEM SUB-SYSTEM DESCRIPTION Pr Fr Annual PM Schedule


A B
PM0104 System Image Check (Image Performance) 2 S ↓ ↓
PM0302 Gantry Clean Mylar ring cover 2 S ↓ ↓
PM0401 DAS Check/Clean Detector surface 2 S ↓ ↓
Subtotal 1:15 1:20

12 SCHEDULE
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SCHEDULE (CONTINUED)

System Performance

ITEM SUB-SYSTEM DESCRIPTION Pr Fr Annual PM Schedule


A B
PM0106 System Perform Filter Curve Test 3 A 0:15
PM0107 System Check Ground Cable Terminals 3 A 0:05
PM0108 System Check DAS Count 3 A 0:15
PM0109 System Image Performance 2 A 0:50
PM0110 System Tube Gas Bubble Inspection 2 A 0:40

PM0202 O.C Inspect FANs 3 S 0:05 0:05


PM0203 O.C Verify Mouse for smooth operation 3 S 0:05 0:05
PM0204 O.C Clean Display monitor/Console exterior 4 S 0:10 0:10

Schedule
PM0205 O.C Check System Clock 3 S 0:10 0.10
PM0206 O.C Check Cables / Power Cables 3 S 0:05 0:05
PM0207 O.C Clean the Dust Proof Filters 3 S 0:10 0:10
PM0303 Gantry Verify cables and hardware are tight 3 S 0:20 0:20
PM0304 Gantry Gantry cover cleaning 4 S 0:10 0:10
PM0305 Gantry Check Positioning lights 3 S 0:10 0:10
PM0306 Gantry Check Gantry anchor 3 A 0:05
PM0307 Gantry Main bearing grease-up 3 A 0:20
PM0308 Gantry Check drive belt for wear 3 A 0:05
PM0402 DAS Verify Detector Heater Control 3 S 0:05 0:05
PM0506 Table Clean Cradle tray / Table cover 4 S 0:10 0:10
PM0507 Table Check Table anchors 3 A 0:05
PM0601 XG Check/Clean Radiator FAN 3 S 0:10 0:10
PM0602 XG Check X-ray tube oil and HV connector 3 S 0:20 0:20
PM0603 XG Check HV Tank and HV connector 3 S 0:20 0:20
PM0604 XG Check Power cable connection 3 S 0:05 0:05
PM0605 XG Check X-ray Tube overheat safety 3 S 0:15 0:15
PM0606 XG Check KV and mA 3 A 1:00
PM0701 PDU Check Power Cable Terminals 3 A 0:05
PM0702 PDU Check Power Line Voltage 3 A 0:05
Subtotal 2:50 6:35
Total 4:05 7:55

*1:Option

SCHEDULE 13
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE

14 SCHEDULE
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

PM Report (1/2)

Schedule A/B
Site : Date:
System ID : Dispatch No. :
Check Item Result
Safety and Regulatory
System Verify Emergency OFF buttons (Gantry/Console)
Check X-ray ON lights or buzzer / Operation of scan abort
Caution Label Check
Check Error Logs
Gantry Check the number of Gantry revolutions
Console Verify Audio Function
Table Check Table Cover
Gap between Table Cover and Cradle
Check Cradle Rail

Report
Holder Stability Check
Verify Unlatch Function
XG Check Parts
Image Quality
System Image Check (Image Performance)
Gantry Clean Mylar ring cover
DAS/Detector Check/Clean Detector face

Comment:

Field Engineer / ID:

REPORT 15
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

PM Report (2/2)

Schedule A

Check Item Result


System Performance
Console Inspect FANs
Verify Mouse for smooth operation
Clean Display monitor/Console exterior
Check System Clock
Check Cables / Power Cable Terminals
Clean the Dust Proof Filters
Gantry Verify cables and hardware are tight
Gantry Cover Cleaning
Check Positioning Light
Main bearing grease-up
DAS/Detector Verify Detector heater control
Inspect FANs
Table Clean Cradle tray / Table Cover
XG Check / Clean Radiator FAN
Check X-ray tube oil and HV connector
Check HV Tank and HV connector
Check Power Cable connection
Check Tube overheat safety

Comment:

16 REPORT
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

PM Report (2/2)

Schedule B

Check Item Result


Safety and Regulatory
Phantom Check
System Performance
System Perform Filter Curve Test
Check Ground Cable Terminals
Check DAS Count
Image Performance
Tube Gas Bubble Inspection
Console Inspect FANs
Verify Mouse for smooth operation
Clean Display monitor/Console exterior
Check System Clock

Report
Check Cables / Power Cable Terminals
Clean the Dust Proof Filters
Gantry Verify cables and hardware are tight
Gantry Cover Cleaning
Check Positioning Light
Check Gantry Anchor / Gantry Isolation
Check drive belt for wear
DAS/Detector Verify Detector heater control
Inspect FANs
Table Clean Cradle tray / Table Cover
Check Table anchors / Table Isolation
XG Check / Clean Radiator FAN
Check X-ray tube oil and HV connector
Check HV Tank and HV connector
Check Power Cable connection
Check Tube overheat safety
Check KV and mA
PDU Check Power Cable Terminals
Check Power Line Voltage

Comment:

REPORT 17
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE

18 REPORT
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SYSTEM 1/1 PM-0101


Purpose : Verify Emergency Stop Buttons (Gantry/Scan Control Box) Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
1. Press an emergency stop button on the Gantry front control and display panel, and verify that the power for
Table/Gantry is removed.
2. Turn the power switch of the Operator Console OFF, then ON again. Verify that all systems, including Table/
Gantry, are turned ON.
3. Repeat this check for other emergency stop buttons on the Gantry and Scan Control Box (See illustration
below).
• Two(2) buttons on the Gantry front control and display panel.

System
• Two(2) buttons on the Gantry left/right frame.
• One(1) button on the scan control box.

Illustration 1-1 Emergency Stop Buttons Location

Gantry Front Control and Display Panel Gantry Right Side

Emergency Stop Button (x2)

Scan Control Box


Emergency Stop Button

Emergency Stop Reset Switch


Emergency Stop Button_Right

SYSTEM 19
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SYSTEM 1/1 PM-0102


Purpose : Check X-ray On lights / Operation of Scan Abort Estimated Time:0:05 Rev 2

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION

CAUTION
Do not enter scan room unprotected when x-rays are present.

1. Check the X-ray ON lights for the following :


a. Perform an axial scan.
b. Verify that the X-ray ON lights illuminates on the keyboard and Gantry front panel.
c. Verify that the X-ray room (Gantry room) warning lights connected to the system are operating normally.
2. Check the operating of scan abort for the following :
a. Perform an axial scan.
b. Press the Abort switch on the Operator Console while performing a scan and verify that X-ray exposure is
terminated.

20 SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SYSTEM 1/1 PM-0103


Purpose : Caution Label Check Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
1. Every caution label is securely attached to the Gantry front, Table front end, and Console.
If they have come off, replace them (See illustration below)

Illustration 1-2 Caution Label Location

System
Caution Label

Caution Label

Caution Label

SYSTEM 21
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SYSTEM 1/1 PM-0104


Purpose : Image Check (Image Performance) Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
1. Set the water part of QA phantom to the phantom holder.
2. Perform axial scans and verify that the image has no artifact.

3. Verify that the water CT number and air CT number equal 0 ± 3 and -1000 ± 5 respectively.
4. If the means and standard deviations are out of specifications, or artifacts are present, schedule time to look
into possible problems.

22 SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SYSTEM 1/1 PM-0105


Purpose : Check Error Log Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
Service Lines : Specific and detailed information concerning the error including (but not limited to) : --Additional
problem descriptions cause information. Information indicating what was supposed to happen, what did happen, and
how it was detected (for example, expected value vs. actual value, name of file attempted to open, etc.). This
includes any information concerning abnormal conditions related to the error.
Note
Refer to ERROR CODE AND MESSAGE LIST in the SYSTEM tab of the DIAGNOSTICS Manual.

System

SYSTEM 23
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SYSTEM 1/1 PM-0106


Purpose : Perform Filter Curve Test Estimated Time:0:15 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
Note
Refer to BOW-TIE FILTER CURVE TEST in the GANTRY tab of the FUNCTIONAL CHECK /
ADJUSTMENT Manual.

1. Remove the service window from the Gantry left side cover.
2. Switch OFF the ‘CB3’ switch on the PDU.
3. Remove all covers from the Gantry.
4. Switch ON the ‘CB3’ switch on the PDU.
5. Perform two air stationary scans with the filter installed and removed.
Scan 1 Scan 2
SCAN TIME :3.0 sec SCAN TIME :3.0 sec
THICKNESS :10 mm THICKNESS :10 mm
FOCUS :LARGE FOCUS :LARGE
KV :120 KV KV :120KV
mA :60 mA mA :60 mA
SERIES NO. :1 SERIES NO. :2
(Others) :(defaults) (Others) :(defaults)

FILTER :INSTALLED FILTER :REMOVED


Note
For filter removal, refer to the Component Replacement manual.
(Switch OFF the ‘CB3’ switch prior to removing the filter!)

6. Check the filter curve using VIEW VECTOR. (Check that the curve is smooth)
7. Visually check the filter for discoloration or cracks.
Replace if necessary.

24 SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SYSTEM 1/1 PM-0107


Purpose : Check Ground Cable Terminals Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
Inspect that all the ground cables are securely connected on terminals.

System

SYSTEM 25
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SYSTEM 1/1 PM-0108


Purpose : Check DAS Count Estimated Time:0:15 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
Note
Refer to DAS COUNT CHECK in the X-RAY GENERATOR tab of the FUNCTIONAL CHECK /
ADJUSTMENT Manual.

This check verifies that the x-ray tube dose falls within specifications by checking DAS Counts.
This check should be performed when the x-ray tube is replaced.
1. Perform an axial scan.
AXIAL SCAN
SCAN TIME :1.5 sec
THICKNESS :10 mm
FOCUS :LARGE
KV :120 KV
mA :60 mA
SERIES NO. :1
(Others) :(defaults)

Bow-tie Filter :Installed


2. Plot the active mean and verify that Max of Mean data (Active Views) is within specifications as below:
For Brivo CT315: 350000 ≤ ≤ 700000 (Specification)
For Brivo CT325: 700000 ≤ ≤ 1100000 (Specification)

26 SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SYSTEM 1/1 PM-0109


Purpose : Image Performance Estimated Time:0:50 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
1. Perform Image Performance Check.
Note
Refer to the following items in the SYSTEM tab of the FUNCTIONAL CHECK / ADJUSTMENT
Manual.
• General
• Image Performance
• Quality Assurance

System
2. If the measure value is out of specification, perform the following ;
• KV and mA Check
• Phantom/Highlight Calibration
• CT Number Adjustment
Note
Refer to the following items in the SYSTEM tab and X-RAY GENERATOR tab of the FUNCTIONAL
CHECK / ADJUSTMENT Manual.
• KV Check with HV Divider
• mA Check
• Calibration
• CT Number Adjustment
3. After Calibration or CT Number adjustment, perform Image Performance Check again.

SYSTEM 27
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SYSTEM 1/6 PM-0110


Purpose : Tube Gas Bubble Inspection Estimated Time: 0:40 Rev 2

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
1. Perform Image Performance Check.
Under some circumstance, gas bubble in the tube will cause tube warm up error and it will also lead to image
artifact. Following is some illustrations of artifact.

Streaks
Streak
Smudge

28 SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SYSTEM 2/6 PM-0110


Purpose : Tube Gas Bubble Inspection Estimated Time: 0:40 Rev 2

2. To prevent image artifacts as above and notice bubbles when small, perform Sinogram to further confirm Tube
gas bubble.
a. Go to Service → Diagnostics → Offline Scan, perform an offline air scan.
AXIAL SCAN
KV: 140KV
mA: 60mA
Scan FOV: 43cm
(Others): (default)

System
b. Go to Service -> Diagnostics -> Scan Analysis, select just finished offline scan exam, press Plot VVC.

SYSTEM 29
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SYSTEM 3/6 PM-0110


Purpose : Tube Gas Bubble Inspection Estimated Time: 0:40 Rev 2

c. Select Offset Correction, press OK.

d. Select Channel from Region of Interest, press Plot Now. A normal plot will look clear as below (channel
343).

30 SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SYSTEM 4/6 PM-0110


Purpose : Tube Gas Bubble Inspection Estimated Time: 0:40 Rev 2

e. Following picture will appear on the screen, channel 343 when normal will look level as below.

System
Note
If shadow or white band exists, make sure to select appropriate channel so that the channel could
across shadow or white band.
f. Analysis: If shadow (white band) appears in Sinogram (as shown in following three figures), it means that
there is a high probability that gas bubble exist in tube. When small, bands may not cover all views or look
like streaks across the VVC.
Note
sometimes there is no problem for some air scan data, it is also possible for existing bubble, just
doesn't affect that special scan.

SYSTEM 31
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SYSTEM 5/6 PM-0110


Purpose : Tube Gas Bubble Inspection Estimated Time: 0:40 Rev 2

32 SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SYSTEM 6/6 PM-0110


Purpose : Tube Gas Bubble Inspection Estimated Time: 0:40 Rev 2

3. Take off tube for further inspection (see component replacement manual for detailed procedures), check tube
port and anode end fill plug for gas bubble.

View through the port

System
Gas bubble in tube

View through anode end fill plug

4. Inform Customer.
If gas bubble is confirmed through above inspection, advise customer to order new tube for replacement
because it may lead to image artifact.

SYSTEM 33
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

SYSTEM 1/1 PM-0111


Purpose : Phantom Check Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
Verify visually that phantoms are intact from any damage or scratches which affect the use.

34 SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

OPERATOR CONSOLE 1/1 PM-0201


Purpose : Verify Audio Function Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
1. Verify that the Operator can properly hear sounds in the Scan room through the speaker.
2. Verify that the voice of the Operator can be properly heard in the Scan room through the microphone.
3. If necessary, adjust the intercom volumes on the Scan Control Box (Refer to the CD-ROM manual, Functional
Check & Adjustment, Operator Console).
Note

Operator Console
In case sound from the scan room is noisy, replace the microphone in the Gantry.

OPERATOR CONSOLE 31
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

OPERATOR CONSOLE 1/1 PM-0202


Purpose : Inspect FANs Estimated Time:0:05 Rev 2

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
1. Verify that the power to the Console is OFF.
2. Remove the service window from the Gantry left side cover.
3. Switch OFF ‘CB7’ switch on the PDU.
4. Remove the Host PC side cover.
5. Switch ON ‘CB7’ switch on the PDU.
6. Turn ON power on the Host PC.
7. Using a flashlight, or by feeling for the proper airflow, check that the CPU and system main fans are operating
normally. (See illustration 1)
8. FANs should also be free of dirt and should operate quietly.
If any FAN is defective, replace the unit.

Illustration 1 OC Fan

System Main Fan

32 OPERATOR
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

OPERATOR CONSOLE 1/1 PM-0203


Purpose : Verify Mouse for smooth operation Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
1. Verify that the power to the Operator Console is ON.
2. Display the cursor on the monitor.
3. Verify that the mouse movement is smooth.
If not smooth, replace it.

Operator Console

OPERATOR CONSOLE 33
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

OPERATOR CONSOLE 1/1 PM-0204


Purpose : Clean Display Monitor / Console Exterior Estimated Time:0:10 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
1. Verify that the power to the Operator Console is OFF.
2. Clean the LCD monitor screen using Glass Cleaner, etc.
Clean also the exterior of the console, including the keyboard and scan control box.

34 OPERATOR
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

OPERATOR CONSOLE 1/1 PM-0205


Purpose : Check System Clock Estimated Time:0:10 Rev 2

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
Check system date time is correct. If not, process the following steps to replace mainboard battery and to adjust
system clock
1. Verify Operator Console Power Switch "CB7" is OFF in PDU.
2. Remove the computer cover.
3. Replace the mainboard battery wtih model "CR2302".

Operator Console
Illustration 2 Mainboard Battery

Battery

OPERATOR CONSOLE 35
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

OPERATOR CONSOLE 1/1 PM-0206


Purpose : Check Cables / Power Cables Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
1. Verify that the power to the Operator Console is OFF.
2. Remove the service window from the Gantry left side cover.
3. Switch OFF ‘CB7’ switch on the PDU.
4. Verify that all the cables to the Host PC are properly installed/secured.
5. Verify that the power cables (AC 220 V) are securely connected at the Host PC and LCD monitor.

Illustration 3 Power Cable Terminal

Power Cable to
PC and LCD
Power Cable
from PDU

36 OPERATOR
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

OPERATOR CONSOLE 1/1 PM-0207


Purpose : Clean the Dust Proof Filters Estimated Time:0:10 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
1. Power OFF the OC power.
2. Remove the chassis side cover.
3. Press the locking lever of the front cover to remove the front cover.

Illustration 3-1 Remove the Host Front Cover

Operator Console
Locking Lever (x3)

4. Remove the dust proof filter from the front cover.

Illustration 3-2 Remove the Dust Proof Filter

OPERATOR CONSOLE 37
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE

38 OPERATOR
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

GANTRY 1/1 PM-0301


Purpose : Check the number of Gantry revolutions Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
Note
Refer to SLIP RINGS AND BRUSHES in the GANTRY tab of the FUNCTIONAL CHECK /
ADJUSTMENT Manual.

Understand and follow the Safety Guidelines Manual.


1. Read the gantry revolutions from service → Utilities → getstat to get the gantry revolutions.

Gantry
2. Based on the number of revolutions, follow the table shown hereafter:
3. After cleaning the slip ring or the brush, please be sure that the contact between the slip ring and the brush is
ok. In order to further optimize the contact, please keep the gantry rotates up to more than 300 revolutions
without workload.

Gantry Revolutions Check Items


Every 300K Clean Signal/Power Slip Ring
Wear Check and Clean Signal/Power Brush
Every 1000K Brush Debris Cleaning
(or 12 Months)

4. END

GANTRY 39
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

GANTRY 1/1 PM-0302


Purpose : Scan Window Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION

CAUTION
Use proper procedures to avoid blood-borne pathogens.
The maintenace tasks which follow include procedures concerned with the cleaning of
equipment (Table Tray, Gantry Cover, etc.). Refer to local/company guidelines for those
procedures, especially where body fluids or other possible contaminates are present.

Understand and follow the Safety Guidelines Manual.


1. Check the Scan Window for the following :
• Check that no scratches or damage are present on the Scan Window.
• Check that the ring surface is clean, free of contrast agent or smudge; if not, clean it.
• Contrast agent can be wiped clean using warm water.
• Check that the ring is properly installed in the Gantry opening.
• Check scan window’s joint position.

2. END

40 GANTRY
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

GANTRY 1/1 PM-0303


Purpose : Verify Cables and Hardware are tight Estimated Time:0:20 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
Understand and follow the Safety Guidelines Manual.
1. Remove the Gantry left side cover.
2. Switch OFF the ‘CB2’, ‘CB4’ and ‘XG Power’ switches on the PDU.
3. Remove all covers from the Gantry.
4. Rotate the gantry in the CW direction by hand and make sure all cables, harness, and other hardware are firmly
attached.

Gantry
Note
If any components are loose, tighten it to the specified torque.
Refer to TORQUE VALUE FOR SCREW TIGHTENING in the INTRODUCTION tab of the
FUNCTIONAL CHECK / ADJUSTMENT Manual.

5. END

GANTRY 41
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

GANTRY 1/1 PM-0304


Purpose : Gantry Cover Cleaning Estimated Time:0:10 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
Use a cloth and a mild detergent or alcohol to clean the covers.

CAUTION
Use proper procedures to avoid blood-borne pathogens.
The maintenace tasks which follow include procedures concerned with the cleaning of
equipment (Table Tray, Gantry Cover, etc.). Refer to local/company guidelines for those
procedures, especially where body fluids or other possible contaminates are present.

CAUTION
Do not look directly into the laser behind the alignment light window. The laser can case eye
damage.

NOTICE
Do not attach cleaning detergent around Microphone.

NOTICE
AVOID EQUIPMENT DAMAGE!
Do not bend or deform the film of the alignment light window.

42 GANTRY
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

GANTRY 1/1 PM-0305


Purpose : Check Positioning Lights Estimated Time:0:10 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION

WARNING
ROTATION HAZARD!
WHEN YOU PRESS THE ‘POSN LIGHT’ BUTTON, THE GANTRY TUBE ROTATES TO
APPROXIMATELY THREE O’CLOCK POSITION. IF IT IS NECESSARY TO PLACE ANY PART
OF YOUR BODY INSIDE OR NEAR THE GANTRY, SWITCH OFF THE ‘CB3’ SWITCH FIRST.

Gantry
CAUTION
Damage to visual function !!
Never look directly into the positioning lights. The lights can severely harm the eyes.

Understand and follow the Safety Guidelines Manual.


1. Push the ‘POSN Light’ button.
2. Verify that the positioning lights are turned ON.
Note
(If the positioning lights are not turned ON, push the reset switch on the TGP Board once and
perform the above)
3. Check if positioning lights are properly adjusted.
4. Verify that the positioning lights are turned OFF automatically after approximately 1 minute.

5. END

GANTRY 43
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

GANTRY 1/1 PM-0306


Purpose : Check Gantry anchor Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
Understand and follow the Safety Guidelines Manual.
1. Inspect that the Gantry is firmly anchored to the floor by four anchor bolts and that no cracks exist on the
insulating plate for each anchor bolt.
Note
If any anchor bolts are loose, tighten it to the specified torque.

Specified Torque : 50Nm

Illustration 1 Gantry Anchor Position

Anchor Bolt

Nut

Anchor Washer

Adjuster Lock Ring

Gantry Base

2. Verify that the Gantry is in no contact with the floor.

3. END

44 GANTRY
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

GANTRY 1/2 PM-0307


Purpose : Main Bearing Grease-Up Estimated Time:0:20 Rev 1

SPECIAL TOOLS/SUPPLIES

NOTICE
To ensure the quality of the lubricant, please order the lubricant from GPO. Otherwise,
unqualified lubricant may cause damage to bearing.
• Grease Gun (U0018BF)
• Grease Cartridge : Shell Alvania Grease #2 (2188551)

PREREQUISITES
• None

TASK DESCRIPTION
Understand and follow the Safety Guidelines Manual.
1. Remove the service window cover from the Gantry left side cover.

Gantry
1. Switch OFF the ‘CB2’, ‘CB4’ and ‘XG Power’ switches on the PDU.
2. Remove all covers from the Gantry.
Note
New grease must be used. Do NOT reuse a grease cartridge; open a brand new grease cartridge
every time you perform this operation.
3. Input the grease (SHELL ALVANIA grease #2) into the grease gun.

Illustration 2 Slip Ring Main Bearing Grease-Up

GANTRY 45
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

GANTRY 2/2 PM-0307


Purpose : Main Bearing Grease-Up Estimated Time:0:20 Rev 1

4. Attach the grease gun to the bearing grease nipple (see illustration 2).
5. Switch ON the ‘CB2’ switch.
6. Inject the grease (one cartridge) while rotating the Gantry slowly by hand. Any grease which leaks between the
shaft of main bearing and housing must be wiped clean.

NOTICE
Slow pumping operation is required. If pumped fast, new grease will be soon ejected so that
new grease can NOT reach the main bearing.
7. Switch ON the ‘CB4’ and ‘XG Power’ switches on the PDU.
8. Rotate the Gantry for approximately 10 minutes using the toggle switches on the TGP board and verify that the
Gantry rotation is smooth.
9. Switch OFF the ‘CB2’, ‘CB4’ and ‘XG Power’ switches on the PDU.
10. Wipe clean any grease which has leaked between the bearing and the housing.
11. Restore the Gantry to the original configuration.

12. END

46 GANTRY
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

GANTRY 1/1 PM-0308


Purpose : Check Drive Belt for Wear Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION

WARNING
ROTATION HAZARD!
WHEN YOU PRESS THE ‘POSN LIGHT’ BUTTONS, THE GANTRY QUICKLY ROTATES TO
THE -14 DEG. POSITION. IF IT IS NECESSARY TO PLACE ANY PART OF YOUR BODY INSIDE
OR NEAR THE GANTRY, SWITCH OFF THE ‘CB3’ SWITCH FIRST.

Understand and follow the Safety Guidelines Manual.

Gantry
1. Remove the Gantry left side cover.
1. Switch OFF the ‘CB2’, ‘CB4’ and ‘XG Power’ switches on the PDU.
2. Remove all covers from the Gantry except front cover.
3. Switch ON the ‘CB2’ switch.
4. Rotate the Gantry by hand (CW) and inspect condition of drive belt. Check for cracks, wear, or fraying. Make
sure to check the entire drive belt.
5. Replace drive belt if necessary.

6. END

GANTRY 47
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE

48 GANTRY
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

DAS / DETECTOR 1/1 PM-0401


Purpose : Check / Clean Detector Surface Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION

WARNING
ROTATION HAZARD!
WHEN YOU PRESS THE ‘POSN LIGHT’ BUTTONS, THE GANTRY QUICKLY ROTATES TO
THE -14 DEG. POSITION. IF IT IS NECESSARY TO PLACE ANY PART OF YOUR BODY INSIDE
OR NEAR THE GANTRY, SWITCH OFF THE ‘CB3’ SWITCH FIRST.

DAS/Detector
Understand and follow the Safety Guidelines Manual.
1. Remove the Gantry left side cover.
2. Switch OFF the ‘CB2’, ‘CB4’ and ‘XG Power’ switches on the PDU.
3. Remove all covers from the Gantry except rear cover.
4. Clean dust from the Detector surface using a vacuum cleaner, and use a cloth moistened with alcohol to clean
it.

5. END

DAS/DETECTOR 49
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

DAS / DETECTOR 1/1 PM-0402


Purpose : Verify Detector Heater Control Estimated Time:0:05 Rev 2

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION

WARNING
ROTATION HAZARD!
WHEN YOU PRESS THE ‘POSN LIGHT’ BUTTONS, THE GANTRY QUICKLY ROTATES TO
THE -14 DEG. POSITION. IF IT IS NECESSARY TO PLACE ANY PART OF YOUR BODY INSIDE
OR NEAR THE GANTRY, SWITCH OFF THE ‘CB3’ SWITCH FIRST.

Understand and follow the Safety Guidelines Manual.


1. Remove the Gantry left side cover.
2. Switch OFF the ‘CB3’ and ‘XG Power’ switches on the PDU.
3. Remove the Gantry right side cover.
4. Remove the Detector Temperature Controller Box cover.
5. Verify that the “DS3” LED in the Detector temperature controller cycles, blinks, on and off.
6. Touch the Detector surface with a hand to verify that the temperature is approximately 36 degrees Celsius.

7. END

50 DAS/DETECTOR
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

TABLE 1/1 PM-0501


Purpose : Check Table Cover Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
Check that the Table Cover is not deformed and not damaged. Also check that the Table Cover does not interfere
with the cradle.

Table

TABLE 53
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

TABLE 1/1 PM-0502


Purpose : Gap between Table Cover and Cradle Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
Check that the position of the table cover and the cradle are parallel, and the gap between them is 7mm or less.
Note
Refer to PARALLEL SETTING OF CRADLE in the TABLE tab of the FUNCTIONAL CHECK /
ADJUSTMENT Manual.

54 TABLE
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

TABLE 1/2 PM-0503


Purpose : Check Cradle Rail Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
1. Check the cradle for the following
• Check that the Velcro tape of the cradle is securely attached. If the Velcro tape has come off, replace
the cradle.
• Check that the attachment rails are not damaged or cracked. If damaged, replace them.
• Check that the cradle rails are not damaged or cracked. If damaged, replace the cradle as an assembly.
• Set the body strap to the attachment rails, then pull the strap by the force of approx. 15 kgf, using a spring
balance. Verify that the strap do NOT be removed from the rail. If not, replace the strap, attachment rails
or cradle.

Table

TABLE 55
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

TABLE 2/2 PM-0503


Purpose : Check Cradle Rail Estimated Time:0:05 Rev 1

Illustration 1 Cradle Rail with Attachment Rail

Velcro Tapes

Cradle
Attachment Rails
(Both Side)

Cradle Rails

Illustration 2 Body Strap Attachment

Body Strap

Approximately 15kgf Cradle Rail

Attachment Rail
Cradle

56 TABLE
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

TABLE 1/1 PM-0504


Purpose : Holder Stability Check Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
1. Verify that the following holders is not loose by hand, when they are attached to the cradle edge.
• Cradle Extender
• Phantom Holder

Table

TABLE 57
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

TABLE 1/1 PM-0505


Purpose : Verify Unlatch Function Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
1. Move the cradle “IN” by 500mm from out limit.
2. Push the Latch SW.
3. Verify that cradle position LED on the Gantry display blinks, and cradle does not move “IN” when cradle “IN”
SW is pressed.
4. Push the Latch SW again.
5. Verify that cradle position LED on the Gantry display is on, and cradle does move “IN” when cradle “IN” SW is
pressed.
6. Repeat the same check for the opposite unlatch switch.

Illustration 3 Unlatch Switch Location

One Unlatch Switch


Per Each Side (x2)

58 TABLE
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

TABLE 1/1 PM-0506


Purpose : Clean Cradle tray / Table Cover Estimated Time:0:10 Rev 1

SPECIAL TOOLS/SUPPLIES
• Vacuum Cleaner

PREREQUISITES
• None

TASK DESCRIPTION
1. Clean table covers and cradle.

CAUTION
Use proper procedures to avoid blood-borne pathogens.
The maintenace tasks which follow include procedures concerned with the cleaning of
equipment (Table Tray, Gantry Cover, etc.). Refer to local/company guidelines for those
procedures, especially where body fluids or other possible contaminates are present.

Table
NOTICE
AVOID EQUIPMENT DAMAGE!
Do not use chloric detergent.

a. Clean dust from the Table top cover (tray) using a vacuum cleaner.
b. Clean the Table covers.
c. Use a cloth and a mile detergent or alcohol to clean the covers.

TABLE 59
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

TABLE 1/1 PM-0507


Purpose : Check Table Anchors Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
1. Inspect that the Table is firmly anchored to the floor, by four bolts and that no cracks exist on the insulating
plate for each anchor bolt.
Note
If any anchor bolts are loose, tighten it to the specified torque.

Specified Torque : 50Nm

Illustration 4 Proper Anchor Position

Anchor Bolt

Nut

Anchor Washer

Adjuster Lock Ring

2. Verify that the Table is in no contact with the floor.

60 TABLE
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

X-RAY GENERATOR 1/1 PM-0601


Purpose : Check/Clean Radiator FAN Estimated Time:0:10 Rev 1

SPECIAL TOOLS/SUPPLIES
• Vacuum Cleaner

PREREQUISITES
• None

TASK DESCRIPTION

WARNING
ROTATION HAZARD!
WHEN YOU PRESS THE ‘POSN LIGHT’ BUTTONS, THE GANTRY QUICKLY ROTATES TO
THE -14 DEG. POSITION. IF IT IS NECESSARY TO PLACE ANY PART OF YOUR BODY INSIDE
OR NEAR THE GANTRY, SWITCH OFF THE ‘CB3’ SWITCH FIRST.

X-Ray Generator
Understand and follow the Safety Guidelines Manual.
1. Remove the Gantry left side cover.
2. Switch OFF the “CB3” and “XG Power” switches on the PDU.
3. Remove the Gantry right side cover.
4. Verify that the radiator FANs for the X-ray tube cooling are operating normally, and not blocked with dust.
Vacuum clean if necessary.

5. END

X-RAY GENERATOR 65
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

X-RAY GENERATOR 1/1 PM-0602


Purpose : Check X-ray Tube Oil and HV Connector Estimated Time:0:20 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION

CAUTION
Shock hazard.
Wait more than five minutes following the last scan before performing the following
procedure.

Understand and follow the Safety Guidelines Manual.


1. Remove the Gantry left side cover.
2. Switch OFF the ‘CB2’, ‘CB4’ and ‘XG Power’ switches on the PDU.
3. Remove all covers from the Gantry except rear cover
4. Position the X-ray tube to the 3 o’clock position (as viewed from the Table).
5. Turn OFF the OC power and switch OFF ‘CB1’ Switch for the main system.
6. Remove the HV connectors and verify the following:
• The oil in the HV wells is pure.
• The amount of the oil has not reduced, if reduced, replenish it with fresh oil (silicone oil).
• No air exists inside the tube. (Verify this by peeping, looking, through the window.)
7. Check the HV (high voltage) connectors for the following:
• No cracks exist on the candle stick surface.
• The connector pins are clean.
• The rubber packing is not damaged.

8. END

66 X-RAY
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

X-RAY GENERATOR 1/1 PM-0603


Purpose : Check HV Tank Oil and HV Connector Estimated Time:0:20 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION

CAUTION
Shock hazard.
Wait more than five minutes following the last scan before performing the following
procedure.

X-Ray Generator
Understand and follow the Safety Guidelines Manual.
1. Remove the Gantry left side cover.
2. Switch OFF the ‘CB2’, ‘CB4’ and ‘XG Power’ switches on the PDU.
3. Remove all covers from the Gantry except rear cover
4. Position the HV Tank to the 3 o’clock position (as viewed from the Table).
5. Turn OFF the OC power and switch OFF ‘CB1’ Switch for the main system.
6. Remove the HV connectors and verify the following:
• No oil has leaked from the HV tank, check for air bubbles in the viewing port.
• The oil in the well is pure.
• The amount of the oil in the HV well has not reduced, if reduced, replenish it with fresh oil (silicone oil).
7. Check the HV (high voltage) connectors for the following:
• No cracks exist on the candle stick surface.
• The connector pins are clean.
• The rubber packing is not damaged.

8. END

X-RAY GENERATOR 67
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

X-RAY GENERATOR 1/1 PM-0604


Purpose : Check Power Cable Connection Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION

CAUTION
Shock hazard.
Wait more than five minutes following the last scan before performing the following
procedure.

Understand and follow the Safety Guidelines Manual.


1. Remove the Gantry left side cover.
2. Switch OFF the ‘CB2’, ‘CB4’ and ‘XG Power’ switches on the PDU.
3. Remove all covers from the Gantry except rear cover
4. Position the HV Tank to the 3 o’clock position (as viewed from the Table).
5. Turn OFF the OC power and switch OFF ‘CB1’ Switch for the main system.
6. Verify the following:
• Cables are securely connected between the HV tank and the inverter.
• The power cables and GND cable are securely fastened in the terminal block connector at the Gantry
rear base.

7. END

68 X-RAY
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

X-RAY GENERATOR 1/1 PM-0605


Purpose : Check X-ray Tube Overheat Safety Estimated Time:0:15 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION

WARNING
ROTATION HAZARD!
WHEN YOU PRESS THE ‘POSN LIGHT’ BUTTONS, THE GANTRY QUICKLY ROTATES TO
THE -14 DEG. POSITION. IF IT IS NECESSARY TO PLACE ANY PART OF YOUR BODY INSIDE
OR NEAR THE GANTRY, SWITCH OFF THE ‘CB3’ SWITCH FIRST.

X-Ray Generator
Understand and follow the Safety Guidelines Manual.
Check for overheat error by performing the following:
1. Remove the service window cover from the gantry left side cover.
2. Switch OFF the ‘CB3’ switch at the PDU. (The other switches should be set to ON)
3. Remove the both side covers from the Gantry.
4. Disconnect the rotor connector.
5. Switch ON the ‘CB3’ switch.
6. Try to perform a scan.
7. Verify that a scan can not be performed and that the “XG ERROR” message is reported on the monitor status
screen.
Note
Refer to X-RAY TUBE OVERHEAT SAFETY in the SYSTEM tab of the FUNCTIONAL CHECK /
ADJUSTMENT Manual.

8. END

X-RAY GENERATOR 69
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

X-RAY GENERATOR 1/1 PM-0606


Purpose : Check KV and mA Estimated Time:1:00 Rev 2

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
Understand and follow the Safety Guidelines Manual.
1. Preparation:
a. Remove the left side cover from the Gantry.
b. Set the QA Phantom to the phantom holder.
c. Move the cradle to position the phantom at the Detector window.
(This is to protect the Detector by reducing x-ray dose to it during the check procedure)
d. Perform Warm-up scans, if the x-ray tube is cold.
e. Wait five minutes or more following the last scan prior to proceeding to the next step.
2. Set the 5th switch of dip switch S1 to ON on the TGP Board.
(KV & MA display mode on Gantry Display)
3. Perform scans of KV - MA stations specified in Table 1.
Verify that the KV and MA values displayed on the Gantry Display is within the following specifications:
KV: ± 5 kV
MA: ± 20 mA
a. Select Service Menu → Diagnostics → Offline Scan.
b. Select Stationary from ‘Technic - [Scan Type]’.
c. Set the scan parameters according to Table 1.
(KV and MA values are displayed in real-time during an exposure, and ‘0’ is displayed after the exposure is
finished)
4. Set the 5th switch of dip switch S1 to OFF on the TGP Board.

5. END

Table 1 KV - mA Stations

Time Thickness KV mA
1.5 or 2.0 sec 5 mm 120 20 40 60 100 160 200
(Display*) 2012 4012 6012 10012 16012 20012
140 20 40 60 100 160 -
(Display*) 2014 4014 6014 10014 16014
Stationary Scan (Present Degree); (Other scan parameters): (defaults)
*: Example, KV=120KV, display will be ‘2012’. That means mA and KV/10 value.

70 X-RAY
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

X-RAY GENERATOR 1/1 PM-0607


Purpose : Check Parts Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION
Understand and follow the Safety Guidelines Manual.

Parts Estimated Life


RAM on the KV 8 years Replace Nonvolatile RAM on the KV Control Board
Control Board 1. Upload database

X-Ray Generator
2. Battery replacement
3. Download database

1. Replace the RAM on the KV Control Board, if necessary.

X-RAY GENERATOR 71
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE

72 X-RAY
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

PDU 1/1 PM-0701


Purpose : Check Power Cable Terminals Estimated Time:0:05 Rev 1

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION

WARNING
ELECTRIC SHOCK HAZARD !!
DANGEROUS VOLTAGES EXIST WITHIN THE UNIT, USE EXTREME CARE TO AVOID
INJURY TO PERSONNEL AND EQUIPMENT.

CAUTION

PDU
Leave all power to the system OFF and tagged until power is called for in specific steps.

Understand and follow the Safety Guidelines Manual.


1. Turn OFF system power in the Power Distribution Box.
2. Verify that all the power cables are securely connected to their terminals within the PDU.

3. END

PDU 73
Brivo CT315
Brivo CT325
GE HEALTHCARE PERIODIC MAINTENANCE
REV 2 5306511-1EN

PDU 1/1 PM-0702


Purpose : Check Power Line Voltage Estimated Time:0:05 Rev 2

SPECIAL TOOLS/SUPPLIES
• None

PREREQUISITES
• None

TASK DESCRIPTION

WARNING
ELECTRIC SHOCK HAZARD !!
DANGEROUS VOLTAGES EXIST WITHIN THE UNIT, USE EXTREME CARE TO AVOID
INJURY TO PERSONNEL AND EQUIPMENT.

Understand and follow the Safety Guidelines Manual.


1. Measure the incoming power line voltage:
380V (phase to phase line voltage)
220V (phae to neutral) nominal input voltage
2. Verify power line voltage falls within the following specifications: +10 % ~ -5 % for 50/60 Hz

3. END

74 PDU