Professional Documents
Culture Documents
1999 Article
1999 Article
ABSTRACT
An X-ray tube with a Mo target and Zr filter, operated at 45
kV/20 mA, was used to excite samples (5 ~tL deposited on a quartz
support) and the total reflection angle condition was obtained with a
double reflector module built with two 10-cm-long 7-mm-thick quartz
crystals placed 50 ~m apart. A high-resolution spectrometer based on
a Si(Li) detector coupled to a multichannel analyzer was used for
X-ray detection and the spectra were interpreted with the AXIL
software.
The system was calibrated with standard chemical solutions con-
taining Cr, Fe, Cu, Zn, and Pb, and Y was used as an internal standard
to correct eventual geometric errors and high-voltage instabilities of
the X-ray generator. The limits of detection were 19, 9, 5, and 4 n g / m L
for Cr, Fe, Cu, and Zn, respectively, analyzed through characteristic Ks
X-rays, and 7 n g / m L for Pb, through L~ X-rays, considering 50 ~tL
I i = siC i (1)
Nuclear Instrumentation
The detection of the X-rays was made using a Si(Li) semiconductor
detector coupled to a preamplifier, amplifier, and multichalmel pulse
analyzer card inserted in a microcomputer. The detector, manufactured
by EG & G Ortec, model SL 30165, is 3-mm-thick and has an active area
of 30 m m 2, 7.5-gm-thick a Be window, and a distance of 5 m m between
the Be window and the crystal surface, presenting a resolution of 165 eV
to 5.9 keV (Fig. 1).
Excitation System
The excitation system is composed of a fine-focus X-ray tube, with a Mo
target and Zr filter, coupled to a high-voltage generator operated at 45 kV
and 20 mA. The X-ray tower has four radiation exits (two point types and
two line types), at an angle distance of 90~ from each other, which can be
used simultaneousl~ given that one line type is used in the TXRF (Fig. 1).
E ]PA
dilcrit = __99
(2)
For the system in use, with a Mo target X-ray tube and quartz (SiO2)
as the sample support, the parameters are p = 2.5 g/cm 3, Z = 30 elec-
trons/molecule, A = 60.0863 g/mol, and an energy of Mo Kc~ X-ray inci-
dent radiation E = 17.44 keV, yielding a critical angle of 6.4 min. This
angle is adjusted using the micrometer screws of the module reflector.
Stardard Solutions
To perform the calibration of the system, four standard solutions
containing Ca, Cr, Fe, Cu, Zn, Ba, and Pb were prepared at different and
well-known concentrations (ranging from 2 to 30 gg/mL, approximately).
Yttrium was added, always to a final concentration of 8.33 gg/mL, as an
internal standard to eliminate the errors caused by excitation/detection
geometry. The thin film obtained from the sample formed on the quartz
support does not have a regular geometry; therefore, the intensity of the
characteristic X-ray depends on the position of the thin film on the sup-
port. This geometric effect of the sample (8,9) can be corrected by calcu-
lating the relative intensity of each element in relation to this internal
standard [(Eq. (3)].
Preparation of Samples
for Excitation~Detection
An aliquot of 5 gL of each one of the standard solutions was pipeted
on a quartz support and dried under vacuum, forming a thin film (two
replicates for each standard solution). The samples were excited for 500 s
and the X-ray spectra obtained were fitted by the AXIL software. The net
1000
9. - . 80O
o
o
" 600
8
Z 400
uJ
m
~, 20o
)<
0
2 4 6 8 10 12 14 16 18 20
ENERGY (keV)
areas of the characteristic X-ray peaks of the elements that compose the
samples were evaluated.
Relative Intensities
As an illustration, Fig. 2 shows the X-ray spectrum of one standard
sample, where the concentration of the all elements is 8.33 ~tg/mL. With
the intensities of the characteristic X-rays obtained using standard sam-
ples, the relative intensities Ri were calculated using Eq. (3). It has to be
pointed out that the relative intensities Ri is not only the ratio between
the intensities of the measured element Ii and that of the internal stan-
dard Iy, because the concentration of the internal standard Cy (8,9) has to
be taken into account. In other words, using Eq. (1), the following equa-
tion can be written:
Ii _ Si Ci
Iy Sy Cy
or
I-Li C y = s--LC i
Ir Sy
10 I I I I I I
R= 0,2EBOZn I]
~8 Zn r2 = 09993
v
[]
Cu R = Q2256Q
Z
~6 r2= 0.981
z
m
uJ R= 0.1422Fe
Fe r2 = Q9923
LU
OE
R=0.074t.Cr i
r2 = 99981
I r I I
5 1~ is 9 2s 30 35
CONCENTRATION (~g/mL)
I I "" I I I I
~ 18 ~k~ LDO=054e6.Z2-31.860Z+4686
v
Z
O14
I-
I-
14.
O8
I-
$8-
,--I
4-
2~
Cr IVh Fe Co Ni Cu Zn
0 ! I I I I I I
23 24 25 ;6 27 28 29 30
ATOMIC NUMBER(Z)
Fig. 4. Limits of detection (ng/mL) for some elements in liquid samples
using 50 ~tL aliquots deposited on quartz supports and analyzed for 1000 s.
I-L Cy
R i = [y
and because
Si si (3)
Sy
the result is
R i = SiC i (4)
Limits of Detection
The limits of detection LODi (10,11) were calculated using Eq. (5) for
a analysis time of 1000 s and the results obtained are shown in Fig. 4. For
this, the background intensities Ii(BG) (i.e., the areas under the peaks of
the elements of interest), were obtained through the fitting of the spectra
by the AXIL software, the intensity Ir for the element Y was taken as one
obtained with the concentration of 8.33 ~tg/mL:
use of 50 gL in the analysis, the LODi would be 1.9, 0.9, 0.5, 0.4, and
0.7 n g / m 3, respectively.
ACKNOWLEDGMENTS
The authors thank Funda~io de Amparo ~ Pesquisa do Estado de
Silo Paulo, Silo Paulo State Foundation for Support to Research (Procs.
95/0018-7 and 96/5818-4) for the financial support.
REFERENCES
1. Y. Yoneda and T. Horiuchi, Optical flats for use in X-ray spectrochemical micro-
analysis, Rev. Sci. Instrum. 42, 1069 (1971).
2. H. Aiginger and P. Wobrauschek, A method for quantitative X-ray fluorescence
analysis in the nanogram region, Nucl. Instrum. Methods 114, 157-158 (1974).
3. P. Wobrauschek and H. Aiginger, Total reflection fluorescence spectrometric determi-
nation of elements in ng amounts, Anal. Chem. 47, 852-855 (1975).
4. C. Koopmann and A. Prange, Multielement determination in sediments from the
German Wadden Sea--investigations on sample preparation techniques, Spectrochim.
Acta 46B, 1395-1402 (1991).
5. V. H. Poblete, Ph.D. Thesis, Facultad de Ciencias F~sicas y Matematicas, Universidad
de Chile, Santiago, Chile (1996), Chap. III, p. 48.
6. H. Aiginger, Historical development and principles of total reflection X-ray fluores-
cence analysis (TXRF), Spectrochim. Acta 46B, 1313-1321 (1991).
7. A. Prange and H. Schwenke, Trace element analysis using total-reflection X-ray flu-
orescence spectrometry, Adv. X-Ray Anal. 35, 899-923 (1992).
8. W. Ladisich, R. Rieder, and P. Wobrauschek, Quantitative total reflection X-ray fluo-
rescence analysis with monoenergetic excitation, X-ray Spectrom. 23, 173-177 (1994).
9. R. Klockemk/imper and A. von Bohlen, Elemental analysis of environmental samples
by total reflection fluorescence: a review, X-ray Spectrom. 25, 156-162 (1996).
10. L. A. Curie, Limits for quantitative detection and quantitative determination, Anal.
Chem. 40, 586-593 (1968).
11. W. Ladisich, R. Rieder, P. Wobrauschek, and H. Aiginger, Total reflection X-ray fluo-
rescence analysis with monoenergetic excitation using rotating anode X-ray tubes,
Nucl. Instrum. Methods Phys. Res. 330A, 501-506 (1993).