Professional Documents
Culture Documents
00/0
Printed in Great Britain Pergamon Press Ltd
Al~tract--The concentration of aqueous ozone can best be determined by the decolorization of indigo
trisulfonate (600 nm, pH below 4) whenever the ozone cannot be measured directly by its u.v. absorp-
tion. The method is stoichiometric and extremely fast. The change of absorbance vs ozone added is
-2.0 _ 0.1 x 104 M -1 cm -~ and is independent of the concentration of aqueous ozone in the range
0.005-30 mg 1- t. The precision of the analysis is 2% or 3/~g 1-1 for low concentrations if a spectropho-
tometer or a good filter instrument is used. Visual methods can be used to measure 0.01 mg 1-1 ozone.
Secondary oxidants produced by ozone in natural water, including hydrogen peroxide or chlorite, do
not interfere; chlorine can be masked. The reagent solution is stable for 3 months. The method is
recommended for kinetic measurements, for studies of ozonation processes and for visual field methods.
so~ method; the u.v. method was applied only as a more con-
0 H~Nr--~ venient secondary standard.
s o , . I", . . , ~IIl , ~ =| L "'1~j~' ' ~ ~" s o , indigo trisuffomc acid
2.2 Instrumentation
o [, o.e2oooo .-' cm-' 1 Generally, the absorption measurements for the indigo
method were performed at 600 nm on a Beckman DK 2A
1 ~'03
spectrophotometer which was equipped with a thermo-
stated cell compartment maintained at 25°C. Cells of
0 0.5-5 cm path length were used. A VITATRON photo-
sc~../,~.~k..~.O isatin sutfonic acid meter equipped with a 4 cm cell and a 595 nm interference
2x II ".T r and corres~ndimg products
filter of 15 nm bandwidth was also used. In experiments to
"¢'----"'"
(so~) I~.,-0.0 M-' cm-' I test the indigo method as a visual field method, pairs of
graduated glass cylinders were used (volume 250 ml. height
etc 25 cm).
Fig. 1. Indigo trisulfonate and its ozonation product. 2.3 Glassware
Best reproducibilities were achieved when the glassware
n.Ri -I ozone o~ded used for handling of aqueous ozone was conditioned by
Ascff repetitive use for the same procedures.
0.~ 2.4 Experimental procedures for the indigo method used with
a spectrophotometer
During this study, many varied procedures were tested.
O~
Only illustrative examples can be given here:
Example for the determination of aqueous ozone in the
concentration range 0.3-15 mg 1- t : Series of 100 ml volu-
0,~
metric flasks were prepared. To each flask of the series
10 ml of 0.5 M Phosphate Buffer Reagent for pH 2, 1 ml of
1 mM Indigo Reagent and, in case of the measurements of
02 high concentrations of ozone, about 20 ml of distilled water
i-
were added. Some of these flasks were used to determine
[]11 ,, . the standard curve. For this, individual flasks of the series
0.0
4O0 500 600 700 nm were dosed with different amounts of diluted stock sol-
utions of aqueous ozone containing about 2 m g l - ' ozone
Fig. 2. Absorption spectra of Indigo Reagent after stepwise e.g. 0, 5, 10... 25 mt, at random to minimize systematic
ozonation (Indigo trisulfonate added: 10/tM). time drifts of ozone concentrations. This solution was
dosed with a glass pipette whose tip was placed below the
surface of the reagent solution while stirring with a Teflon
covered magnetic bar. The concentration of the ozone in
Gurr Ltd (London) product; Phosphate Buffer Reagent for the diluted stock solution applied was calibrated simul-
pH 2: 28g N a H 2 P O 4 ' H 2 0 and 35g H s P O , (85V.) dis- taneously by measuring its u.v. absorbance at 258 nm in a
solved in t 1. dist. water. Other reagents were of commer- 5 cm u.v. cell. Finally, distilled water was added to the
cial pro analysis grade. 100 ml mark. During this operation the stirrer was tempor-
Aqueous ozone. Stock solutions of aqueous ozone were arily raised above the mark. The absorbance of the residual
prepared by continuously bubbling ozonated oxygen indigo present in these series of test solutions was
(about 4%) through a gas-washing bottle into distilled measured at 600 nm in a 5 cm cell. The values were plotted
water chilled to 2°C. This stock solution generally con- vs the amount of the ozone added. A typical standard
tained a steady-state concentration of ozone of 40 mg 1- ~. curve is presented in Fig. 3(B). The determination of the
Lower concentrations were achieved by dilution with dis- ozone in the unknown samples is thereafter performed
tilled water. Whenever not required for later use, this water similarly by dosing specific volumes of these samples to
was acidified with phosphoric acid to pH 2. When the others of the prepared flasks of the series. In order to be
bottle was kept closed the concentration of ozone de- within the range of the amount of Indigo Reagent used the
creased less than 0.2~ per min (Hoign6 & Bader, 1978). size of the sample was selected to contain an estimated
The aqueous ozone was transferred into the individual test 10--40#g ozone. For example, when the concentration of
bottles with glass pipettes or by plastic dispensers. The ozone was expected to be 0.5 mg 1-1 20-80 ml were
concentration of the diluted ozone water was determined injected and when the concentration was very high, such as
before and after the samples were withdrawn. These cali- 15 mg 1-t. 1-2.5 ml were dosed. The volumetric flasks were
brations of the stock solution were performed by measur- filled again with distilled water to the I00 ml mark. The
ing the u.v. absorbance of ozone at 258 rim. They were residual absorption at 600 nm was compared with the stan-
based on a molar absorptivity of 2900 M - * c m - * which dard curve and the original concentration of ozone was
was determined in other studies by a direct comparison calculated.
with the iodine method, here considered as the primary Comparable procedures were adapted to determine
standard (Kilpatrick & Herrick, 1965; Hoign6 & Bader, ozone in other concentration ranges. In cases of higher
1976). For these calibrations t h e samples of aqueous ozone concentrations, higher amounts of the Indigo Reagent were
was acidified by sulfuric acid to give 0.3-0.5 M acid. About dosed. In this case the determination of the residual
0.2 g o f potassium iodide were added. The liberated iodine absorbance of the Indigo Reagent was determined using
was titrated with 0.1 M sodium thiosulfate using starch 0.5 cm cells. An example of a standard curve is shown in
indicator. Fig. 3(A). In this case the amounts of the Indigo Reagent
Care was taken that the aqueous ozone was always and of the ozone applied were about 10 times higher than
handled in exactly the same way, whether it was trans- those in the example shown in Fig. 2. When lower concen-
ferred into samples containing tndigo g ~ t or inlo the trations of ozone were determined, less Indigo Reagent was
u.v. cells or into flasks for the iodide t i t r a t i ~ , All concen- used. Figure 3(C) gives an example in which the concen-
trations of ozone are therefore finally based on the iodide trations were 4 times lower than those for the example in
Determination of ozone in water by the indigo method 451
5iS
A4©m
0.04
08' ~, (A)
0.03
0.02
06 •~
"\ 001
0A ~ I I t I~'lk ! IL
(200 5 IO )5 20 25
Added [ 031/zg I-)
0.2
Fig. 4. Standard curve absorbance vs ozone added to
Indigo Reagent. Measurement with filterphotometer. Inter-
O0 I I i ,.I
I 2 3 4~ ference filter 595 nm. Cone. of Indigo Reagent 0.6 M. (Con-
Added[03] n ~ f' centrations are based on final volume.)
o8~.\
- "\.\ (B)
glass cylinders A and B. Cylinder A was filled to its 25 cm
mark with the sample of aqueous ozone. Cylinder B was
filled with distilled water. Cylinder B was then emptied
into an Erlenmeyer flask and refilled with this reference
solution until the color intensity, when observed from top,
matched that of cylinder A. A standard curve was obtained
04 •~e by repeating such measurements using different known
amounts of ozone. The height of the reference solution in
cylinder B was plotted vs the concentration of the aqueous
02 ozone applied in cylinder A. A standard curve is given in
Fig. 5. Thereafter, the concentration of ozone in the
samples was determined by comparing the height of sol-
(20 I
o.i
I
0.2
I
0.3
"N ~
0'~ ution in the reference cylinder with the heights given in the
Added [ 03] mg I standard curve. The size of the sample was taken into
IlOOem account for the calculation of the concentration.
At=,, I
o.o,. ~ ICl 2.6 Experimental procedures applied for other spectrometric
analysts of ozone
For some tests, ozone had to be determined by the direct
0.03 measurement of its u.v. absorption at 258 nm even in
natural waters. In these waters, which sometimes exhibited
significant background absorption at 258 nm, the measured
absorbance was corrected by determining the background
002 absorbance after immediately destroying the residual
ozone with an approx. 10-fold excess of aqueous nitrite or
butene-ol.
The ACYK method was carried out as described by
oo
Masschelein & Fransolet (1977). The methods based on
I I • i~ h• Approx., A 2~~,.
6 0 Ohm
O. 02 0.04 0,06 (208 O I- ='4
Added [031 mg I-' i
25 • -~o. eo
Fig. 3. Standard curves o f absorbance (A) at 600 nm vs
ozone added to Indigo Reagent. Measurement with 2o .!o.
s t ~ r o p h o t o m e t e r . Conc. of Indigo Reagent added: (A):
0.1 raM; (B) 10 #M; (C) 2.5/~M. (Cone. of Indigo Reagent
and ozone are based on final volumes.)
15~ ~0.48
,e ax -~o32
,
w.R. 15/4--D
452 H. BADERand J. HO1GNE
o-tolidine and DPD were performed as described in Stan- the variances is thereby expected to be due to the
dard Methodsfor the Examination of Water and Wastewater variance in ozone doses.
I1975).
The effect of the rate of mixing of the aqueous
ozone with the Indigo Reagent solution was tested in
special series. Only a small difference of the sensitivity
3. RESULTS
factor (2°; between matched pairs of extended series
3.1 Absorption spectra of the Indigo Reagent of simultaneously prepared samples) could be found
The Indigo Reagent shows an absorption maxi- between normally and extremely slowly stirred
samples.
mum at 600 nm. The apparent absorption coefficient,
based on the amount of total mass of commercial 3.3 Measurements with filter photometers
(non-purified) reagent added is 16,500 M - 1 cm- 1. The
shape of the absorption curve and )~,x do not change Measurements of the residual absorbance of the
with ozone addition (see Fig. 2). Indigo Reagent with a photometer equipped with a
595 nm filter gave results and variances comparable
3.2 Sensitivity and precision of the indigo method as to those determined with the spectrophotometer. An
used with spectrophotometers example of a standard curve is given in Fig. 4. The
slope of this curve is 85°,0 of that expected from
The absorption of samples containing the Indigo
measurements using a spectrophotometer at 600nm,
Reagent decreased linearly with the amount of ozone
probably because the 595 nm filter does not quite
added over all the concentration ranges investigated.
match the maximum of the absorption spectrum and
From the slopes of standard curves the photometric
the filter band-width of about 15nm exceeds the
sensitivity of the analytical method may be deduced.
maximum of the absorption band of indigo trisulfo-
Values are presented in Table 1. The absorbance
change at 600nm is 2.0 + 0.! x 104cm -1 per mol 1-1 nate (c.f. Fig. 2).
of added ozone for all concentrations tested. Lack of 3.4 Visual methods
precision may be due to uncertainties in the calibra-
An example of results found by the visual method is
tion of aqueous ozone applied for the determination
presented in Fig. 5. An incremental ozone dose of
of the standard curves. We estimate that this uncer-
10 pg 1- 1 can be easily detected when 25 cm cylinders
tainty can cover a range of 5% when series performed
are used.
at different concentration ranges are considered. The
precision of the slope of single standard curves (about 3.5 Stoichiometricfactor
8 points measured) was generally within this range.
Extrapolation of the standard curves to the inter-
Neither deterioration of the absorbance of the Indigo
cept on the absissa shows that only 80% of the com-
Reagent solution to 80% of its original value during
mercial indigo product applied is effective in consum-
extended storage (3 months) nor a partial decoloriza-
ing ozone. This calculation is based on the assump-
tion by a preliminary ozonation affected these calibra-
tions that the potassium indigo trisulfonate applied
tions.
was 100`0/0 pure and that 1 mol ozone added would
The error of the ozone determination decreases
decolorize 1 tool of indigo. Above pH 4 the stoichio-
with the relative amount of decolorization and
metric factor often decreases and it seems to become
increases in case of very low concentration ranges.
somewhat erratic.
However, the relative standard deviations calculated
from extended series were generally about + 2% when 3.6 Rates of reactions
the concentration of ozone was above 0.1 mg 1-1 and
For the application of the Indigo Method for kin-
when the residual absorbance was in the region of
10-60% of the initial absorbance. The main part of etic measurements it is important to note that all the
ozone is consumed by the reagent immediately upon
mixing. This was observed on a 1 s time scale even in
systems where the concentration of ozone decreases
Table 1. Typical sensitivity values of the Indigo Method by fast concurrent reactions, e.g. 20% s-1, and even
when the concentrations of both the indigo trisulfo-
Range of ozone
co dose AA* nate and the ozone were, at the time of mixing, as low
(/~M) (mg 1-1) (M - l c m - l) a s 10 - 7 M {Hoign6 & Bader, 1981). That means that
the rate constant for the reaction of ozone with indigo
100 0.2-3.5 (2.02 _ 0.10)' 104 trisulfonate must be higher than 107M - is-1.
10 0.02-0.3 (2.02 _ 0.10)" 104
2.5 0.02q?.l (1.99 +_0.10)" 104 Aqueous indigo trisulfonate which is preozonized
until just decolorized consumes further ozone only
Co = cone. of potassium indigo trisulfonate applied, slowly. The apparent reaction-rate constant of t h e s e
based on final volume. ozonolytic products is about 2500 M - ts- 1 (Hoign6 &
AA = C h a n g e o f a b s o r b a n c e ( c m - 1) at 600 n m m o l - 1
Bader, 1981). Based on this, we may assume that t h e
1- t added o z o n e .
* Estimated range of accuracy of calibrations = indigo method will not be significantly disturbed by
+ 0 . 1 . 1 0 ' * M - l cm -1. short temporary local overdoses of ozone.
Determination of ozone in water by the indigo method 453
03oddN
~nm
-- As.~ O25(mg I"~)
~7t
O.
-- aCVK A~" o~
DPD - - - o-Tolidine A=~'~'~ O~15
o61- .y ....... om A~'L" oz~
/
I_ // IACVK
O.5b.. . . . . -~ \ .............
]"'x,,x~" . - ' " Indigo
0.41'- "" k."~,~
0.3
0.2 I-
t ""
" " --..¢. o-Tolidine
~ "~ ~ .~......
N
water from Greifensee--a highly eutrophic lake found for this reagent when the absorbancc of the
exhibited the maximum rate of change encountered in disulfonate was measured at 610nm, where this re-
our survey studies (Fig. 7). Such post-decolorizations agent shows its maximum absorption (Hoign6 &
are, however, very small when purified drinking Bader, 19801.
waters are considered.
Table 4. Sensitivities of different colorimetric methods have the disadvantage of being non-specific; iodine is
for the determination of aqueous ozone expressed as also formed by secondary oxidants such as produced
AA (change of absorbance per 1 mol 1-t ozone added by ozonation of water below pH 6-7 that contains
per I cm)
organic solutes. Moreover, the DPD method shows a
Wavelength AA large time drift and therefore cannot be recommended
Method (nm) (M- 1 cm- t) Ref. for extended series of measurements as required for
kinetic studies.
u.v. of 03 258 +2900 a
Indigo 600 - 20,000 b 4.4 Dynamic range of the Indigo Method
ACVK 550 - 2500 c
Orthotolidine 440 + 32,000 d The ozone dose must be adjusted to decolorize
DPD 510 + 15,000 e 20-903/o of the Indigo Reagent. For one fixed concen-
Iodide/triiodide 352 +25,000 f tration of the reagent and one predetermined relative
a--Kilpatrick (1965) and Hoign6 & Bader (1976). ratio of volume of reagent to volume of aqueous
b---this study. ozone, the dynamic range is therefore restricted to a
c--own measurement; method of Masschelein (1977). factor of about 4.5.
d--own measurement and Sulzer (1958). In non-automated systems, preliminary tests will
e--own measurement, method of Palin (1975). help to locate the approximate range of concen-
f---talc, from data of Shechter (1973).
tration before specifying the analytical procedure with
respect to ratios of volumes to be used. During our
parison. The most straightforward method is the di- studies, we covered a region of measurement of
rect measurement of the absorbance of aqueous ozone 0.005-30 mg l-1 of aqueous ozone by merely adjust-
at 258 nm. This absorption occurs, however, in a ing the relative volume of aqueous ozone and of
region where many types of other solutes interfere: in Indigo Reagent applied. This entire range exhibits
natural waters, e.g. humic materials often cause a sig- constant sensitivity. It covers the requirements gener-
nificant background absorption which varies with ally encountered in ozone applications.
ozonation. In addition, the instrumental measurement
must be performed at the time at which the concen- 4.5 Calibrations
tration of ozone is of interest. Moreover, due to the In this study we used the u.v. absorbance of ozone
low molar absorptivity of ozone, the sensitivity of this in distilled water at 258 nm as a secondary standard
method is relatively low. This direct method, however, to calibrate the analytical standard curves. At the
must still be recommended as the best defined and present time we still recommend such calibrations,
useful reference method for the calibration of ozone although different commercial products of indigo tri-
concentrations whenever the limitations of sensitivity sulfonate tested so far have shown rather comparable
and background absorptions are not relevant. sensitivity factors (Hoign6 & Bader, 1980). Sensitivity
In contrast, the Indigo Reagent absorbs light in a factors can, however, only be exchanged between
spectral region where only a few other substances laboratories which perform the measurements on cali-
interfere. Natural waters, for example, have generally brated spectrophotometers at 600 nm. Filter instru-
no background absorption in the 600 nm region. The ments will require individual calibrations.
sensitivity, when expressed as the change in absorb-
ance per added ozone, is quite high. In addition, 5. CONCLUSIONS
instrumental sensitivities are generally high at 600 nm
and even the sensitivity of the human eye is relatively The Indigo method covers the range of concen-
high at this wavelength. For most types of waters trations between 0.005 and 30 rag l-t. The method is
measurements can be performed even hours after the relatively straightforward and does not require any
ozonated water has been added to the reagent sol- difficult or time-consuming operations. It can be per-
ution. formed with the classical instrumentation of water-
The sensitivity of the indigo method is about 8 work laboratories. The Indigo Reagent is stable: it
times higher than that of the ACVK method. In ad- can be stored for months as a reagent which is
dition, measurements at 600nm are preferable to "ready to use". The absorption to be measured at
measurements at 550 nm where more aqueous solutes 600 nm is situated in a range where solutes of natural
may interfere. waters do not interfere and where even simple visual
The o-tolidine method exhibits a sensitivity 1.6 methods can be used for measurement. The change
times higher than the indigo method. Its non-specific of absorbance reflects the amount of ozone added
response to many kinds of oxidants, its low wave- with high and constant sensitivity factor of AA =
length region of absorption, and its temporal instabi- - 2 0 , 0 0 0 c m - 1 ( m o l . l - l ) - I ozone added. All stan-
lity, however, disqualify this method when compared dard curves are linear. The sensitivity does not vary
with the others (Sulzer, 1958; Hofman & Stern, 1969), with ozone concentration, with small changes of tem-
The DPD method for the determination of ozone is perature of reaction, or with the chemical compo-
based on a preliminary oxidation of iodide. However, sition of the water whenever the pH is below 4. For
all methods based on oxidations of iodide to iodine many types of waters the instrumental or visual
456 H. BADERand J. HOI,GNF'~
measurements can be performed even a few hours Hoigne J. & Bader H. (1976) The role of hydroxyl radical
after the ozone has reacted with the Indigo Reagent. reactions in ozonation processes in aqueous solutions.
The method is relatively selective. Secondary oxi- Water Res. 10, 377--386.
Hoign6 J. & Bader H. (1978) Ozone and hydroxyl-radical
dation products do generally not interfere. The reac- initiated oxidations of organic and organometallic trace
tion of ozone with the Indigo Reagent is so fast that it impurities in water. Orqanometals and Organometalloids:
can be applied to stop and measure even very fast Occurrence and .late in the Environment. (Edited by
ozonation reactions. The method has been well tested Brinckman F. E. & Bellama J. M.), Am. Chem. Soc.
Symp. Ser. 82, pp. 292 313.
and applied for the determination of ozone in many
Hoign6 J. & Bader H. (1979a) Ozonation of water: Selec-
different types of waters. Unexpected difficulties have tivity and rate of oxidation of solutes. Ozone Sci. Enqng.
not arisen. More practical experiences with this 1, 73 85.
method are now being accumulated in many water- Hoign6 J. & Bader H. (1979b). Ozonation of water: "Oxi-
works and laboratories. dation-Competion values" of different types of waters
used in Switzerland. Ozone Sci. En,qno. !, 357 372.
However, this Indigo M e t h o d cannot be used as a Hoign6 J. & Bader H. (1980) Bestimmung yon Ozon und
primary standard. At the present time the reagent Chlordioxid in Wasser mit der lndigo-Methode. Vom
must still be calibrated. A standard curve can be pro- Wasser. 55, 261--280.
duced using an aqueous solution of ozone in distilled Hoign6 J. & Bader H. (1981) Rate Constants of direct
reactions of ozone with organic compounds in water: In
water which is calibrated by the direct u.v. absorption
preparation.
measured at 258 nm or by the classical iodometric Kilpatrick M. L. & Herrick C, D. (1965) The decompo-
procedure. It is, however, quite possible that a wider sition of ozone in aqueous solution. J. Am. chem. Soc. 78,
application of the indigo m e t h o d will stimulate cen- 1784-1889.
tral national laboratories or a chemical company to Masschelein W. J. & Fransolet G. (1977) Spectrophoto-
metric determination of residual ozone in water with
market an indigo reagent with a calibrated sensitivity ACVK. J. Am. Wat. Wks. Ass. 69, 461-462.
factor for the determination of ozone and chlorine Palin A. T. (1975) Current DPD method for residual halo-
dioxide (see Hoignb & Bader, 1980). gen compounds and ozone in water. J. Am. Wat. Wks
Ass. 67, 32-33.
Acknowledgements--We thank Professor Werner Stumm Preisler P. W., Hill E. S., Loeffel R. G. & Shaffer Ph.A.
for his continuous encouragement, and Dr Jiirg Zobrist (1959) Oxidation reduction potentials, ionization con-
and Dr Joan Davis for the valuable reviewing and dis- stants and semiquinone formation of indigo solfonates
cussions. Andreas Locher is thanked for his assistance in and their reduction products. J. Am. chem. Soc. 81,
performing many of the experiments during his apprentice- 1991-1995.
ship stage. Schweizerisches Lebensmittelbuch (1972) Trinkwasser und
Mineralwasser. Kap. 27, p. 8.
REFERENCES Shechter H. (1973) Spectrophotometric method for deter-
mination of ozone in aqueous solutions. Water Res 7,
Dorta-Schaeppi Y. & Treadwell W. D. (1949) Ueber eine 729-739.
kolorimetrische Bestimmung von hochverdiinntem Ozon Standard Methods for the Examination of Water and
in Sauerstoff. Heir. chim. Acta 32, 356-364. Wastewater (1975) 14th Edition. APHA-AWWA-WPCF.
Dr~iger H. & Dr~iger B. (1966) Anzeigepr~iparat f'tir Sulzer F. (1958) Ueber das Verhalten w~isseriger Ozonl6-
PriifriShrchen zur Anzeige von Ozon und Verfahren sungen. Schweiz. Z. Hydrol. 20, 16-29.
zu seiner Herstellung. German Patent No. 1'204'854 Thenard A. & Thenard P. (1872) M6moire sur Faction de
(26.5.1966). l'ozone sur le sulfate d'indigo et I'acide ars6nieux. C. r.
Fonrobert E. (1916) Chemie in Einzeldarstellungen, Vol. IX, Acad. Sci. 75, 458-465.
p. 231. Das Ozon, Verlag Ferd. Enke, Stuttgart. Verein Deutscher Ingenieure, VDI, 2468 (1977) Messen
Hofmann P. & Stern P. (1969) Photometrische Bestim- gasf6rmiger Immissionen. Messen der Ozon-Konzentra-
mung niedriger Ozonkonzentrationen in W~issern. Ana- tion. Manuelles photometrisches Verfahren. Indigosul-
lyt. Chim. Acta 47, 113-120. fons~iure-Verfahren.