SERVICE MANUAL
742
SUBSYSTEM
INSTALLATION 2 >
OPTIMUS 50/65/80
9890 000 02012 FAULT FINDING 3 >
REPLACEMENT 4 >
PROGRAMMINGS
ADJUSTMENTS
ACCEPTANCE
SERVICE INFORMATION
358/92
DMC Hamburg
SCHEMATIC DRAWINGS Z1-
Printed in Hamburg
Federal Republic of Germany
© 1996 Philips Medizin Systame
ALL RIGHTS RESERVED
Release: 2
List of pages and drawings (LOPAD) Manual Ordeř No: 4512 103 58276
PRB-XRD products
Z0-1 (b/96.0)
Z1-1.1 © (96.0) A3
Z1-1.2 (96.0) A3
Z1-2.1 (b/96.1) A3
71-2.2 (b/96.1) A3
Z1-2.3 (a/95.0) A3
Z1-3.2 (b/96.1) A3
Z1-3.2 (c/96.0) A3
Z1-3.3 (a/96.0) A3
Z1-4.1 (96.0) A3
Z1-4.2 (96.0) A3
Z1-5.1 (a/96.0) A3
Z1-6 (a/96.3) A3
Z1-11.1 (96.0) A3
Z1-11.2 (94.0) A4
Z1-12 (a/96.0) A3
Z1-13.2 (a/96.1) A3
Z1-14.1 (a/96.0) A3
Z1-142 (b/95.0) A3
Z1-15.1 (96.0) A3
Z0-2 (b/96.0)
Z2-1.0 (94.1) A4
Z2-1.1 (b/96.0) A3
Z2-1.2 (b/96.0) A3
Z2-1.3 (a/96.0) A3
72-2.1 (а/96.0) АЗ
72-22 (a/96.0) A3
72-23 (94.0) АЗ
72-3 (b/96.1) АЗ
72-3 (c/96.0) A3
Z2-5.1 (a/96.0) A3
Z2-5.2 (a/96.2) A3
72-5.3 (95.0) АЗ
72-12 (a/96.0) АЗ
72-13 (b/96.0) АЗ
72-14.1 (a/96.0) АЗ
Z2-14.2 (a/96.0) A3
Z2-14.3 (a/96.0) A3
Z2-15.1 (96.0) A3
LIST OF PAGES AND DRAWINGS Module code number: 4512 982 00751
TEXT: DRAWINGS:
1. Productinformation ................ 1 Mechanicaldimensions ....................... 2-1
2. Compatibility ....................... 2 Overlayer for room layout ..................... 7—2
3. Mechanical data .................... 3 Connection of generator ..................,.... 2-3
4. Environmental data ................. 3 Operating panel .....................,........ 2-5
5. Electrical data ...................... 4 Connection diagram .................,........ Z-7.1
6. Tools ............................,. 7 Connectiondiağgam .......................... Ζ-7.2
7. Traceable items .................... 7 Connection diagram ...............,..,.,..... 2-7.3
8. Preparation ........................ 7 Earthing diagram .........,.............,..,.. 2-7.4
9. Planned maintenance ............... 8 Legend for earthing and cabling diagram ㆍ. Z-20.1
1. Product information
The OPTIMUS family of generators for radiography is based on computer-controlled converter technology.
The converter operates in the non-audible frequency range.
Applicational options are essentially achieved by releasing software modules using customized PAL ICs.
Control between the internal function units (FUs) and the external online equipment takes place via a CAN bus.
Safety-relevant signals are transferred directly on the so-called signal bus.
Units without any CAN interface are operated via the "Adapter 4 auxiliary units” option.
1.1. Applications
~ Radiography
— Tomography
1.2. Options
Low-speed rotor control 9890 000 02201/2
High-speed rotor control 9890 000 02211/2
Automatic input of tomo times 9890 000 02221
Tomo density control (TDC) 9890 000 02231 *)
VARIO FOCUS 9890 000 02271
Mains transformer 400 ...480 V, 50/60 Hz,
also for 400 V mains supply without neutral lead N 9890 000 02301
with taps for 400 / 440 / 460 / 480 V
Adapter for 4 aux. units WA 9890 000 02311
Option rack 9890 000 02321
Extension set for one additional tube 9890 000 02341
Tube extension WG 9890 000 02381
Operating panel 9890 000 02401
Operating desk data cable, 10 m, 20 m, 30 m 9890 000 02411/21/31
Stand for operating panel 9890 000 02441
Wall mounting of operating panel 9890 000 02451
26 V DC / 230 V AC Adapter 9890 000 02461
Surge Arrester WN 9890 000 02471
Handswitch for OPTIMUS 9890 000 02491
Patient data organizer (PDO) 9890 000 02551 *)
Area dose calculator | 9890 000 02561 *)
APR extension 9890 000 02571 *)
Decade cabie set 9803 704 20102 14 х4 т, top decade — AMP decade
Mains transformer 190...390V, 50/60 Hz 9803 720 81002 max. 50 kW!
with taps for 190 / 200 / 207 / 220 / 230 / 240 /
250 / 343 / 380 / 390 V
*) These options are possible only in conjunction with the generator firmware Release 2.
2. Compatibility
2.1. Generator components
.of the X-ray generator family OPTIMUS 9890 000 02001
2.2. Tubes
Recommended standard tubes:
RO 17 50
SRO 25 50
SRO 33 100
3. Mechanical data
Installation dimensions and weights: see drawing Z—1
Transport data:
Case 1:
Contents: generator cabinet, operating panel 178 226 210 82 84
Case 2:
1-tube version 73 100
77 67 80
2-tube version 88 115
Contents: H.V. generator
4. Environmental data
according to PMS standard UXW 13600
5. Electrical data
| 50 kW | 65 kW | 80 kW
|
Power data: | |
Mains voltage 3 x 400V +10% (= 880 V-5%, 415 V +6%)
Mains resistance
ー at400V < 300 mQ < 200 MQ
— at 440/460V < 350 mQ < 240 mQ
— at 480 V = 400 mQ < 300 mQ
— at 480 V, valid for DOD only = 300 mg < 240 MQ | < 180 MQ
3 phase WYE
LI 400 V
440 V / 460 V / 480 V
with mains transformer
9890 000 02301.
X-ray Neutral not required if the
L2 installation mains transformer
9890 000 02301 is ordered.
L3
190 V...343 V with
(N) external mains transformer
+ 9803 720 81002 (max. 50 kW).
3 phase DELTA,
balanced earth
L1
Mains transformer
9890 000 02301 is required.
3 phase DELTA,
grounded or floating
Li
Mains transformer
9890 000 02301 is required.
400 V / 440 V / 460 V / 480 V
e Check that the sequence of phases in the wall junction box is correct corresponding to the designations Li, L2, LS.
357J94
— Ra is dependent upon the length of lead Iz between main distributor and wall junction box
and upon the cross section selected, so:
Аз = Вх Reu Rou from table
The resistances already consider the go and return line so that the calculation can be based on simple cable lengths.
6. Tools
Traceable items
X-ray generator (generator cabinet, operating panel)
H.V. generator
1
The items are assigned the same series number when delivered ex factory.
8. Preparation
Connection of the generator: see drawing Z-3
Operating panel: see drawing 2-5
Connection diagram: see drawing Z—7.1/2/3
Earthing diagram: see drawing 2-7.4
Legend: see drawing Z-20.1
8.2. Cables |
~ H.V. cables 9806 402 6xx02
plugs: 03/03
length: 6 ...30m in steps of 2m
capacity: 155 pF/m
diameter: 16.5 mm
The cable length is indicated at the 9th and 10th digits of the code No.
— Operating desk
cable set 10m 9890 000 02411
20m 9890 000 02421
30 m 9890 000 02431
8.3. Manpower
At least two persons to insert the H.V. generator, which weights about 110 kg, in the generator cabinet. (
9. Planned maintenance
The technical documentation for carrying out maintenance work in compliance with the applicable regulations are
available at the responsible authority of Philips Medical Systems.
The importance of having maintenance implemented is pointed out to the operator in the operating instructions.
It must be guaranteed that the person carrying out maintenance work knows about the respective national regulations
“and that this person observes these regulations throughout all steps of maintenance work. ©- - - -
2 = J s |
>~ 23 A
A| UNF >2
一 一 一一 |
ÀAll ID11% | /7
Al Lol 中 和 >
V % グVA
>4 AEs
> č
GZ
Y A4 À %ζ
2G ク乡 4ク
ή 7 тс
7 * A
G
/7
2 3 7
ン {
ク/ 9 ~
AS sh: 8 1%
{
IN С SK
PRB-XRD products
= 70 2 トー —2 =70
1) 3
3) Filler panel
weight : 210 kg
Schr
94-09-22
u00614
A2/A4&
Y With no ofher
cabinets beside them
E= Control cabinet
C= Operating desk
PRB-XRD products
グロ = | ZA À ググ グ
Schr.
A4 96-02-06
generator
PRB-XRD produccts
Sie.
23
00751
A4 94-09-22
982
PRB, 4512
Connection of generator
4512 982 00751 (94.0) 2—3
© 1994 Philips Medizin Systeme
ALL RIGHTS RESERVED
(1) Low profile control desk
CE 3
4 202
q
O [ B 88Ἡ
=
430
480
PRB-XRD products
>) οἱ の 4
Lg
430 |
Sie/Wi.
|
238 —
982
|
| Е TI >
4512
„P İ 400
|『
PRB
ーー の 11 mm
cable passage
Operating panel
4512 982 00751 (a/96.0) 2-5
© 1994 Philips Medizin Systema
ALL RIGHTS RESERVED
ME
Patient data organizer
Stand [s]
1 1
1
pi 88 _ C200x1 #24
1 1 09 C300x2 nti nn ninna nn
+ 上
고 |
basic om EZ139X5 | serial data cable [D9f COMI of РС | # 40 Recommended for connecting the service-PC
EDX! 1
Emergency off
supply 4
1
1
1
1
1
products
1
4
'
— 4 1. tube housing
M
PRB-XRD
1— — cables.
sm à に e а me боны со м зо
о aq
EwG ktte ーー
στ
tube housing
_EwG K1t6 一 一“上 —
L tube hausing
stator cable Il i
2 tubes
Lâ me | .
i
LA ww 1.
ーー metano crias mia
96-03-14
ub0616
A2/A3
о а мо о о о о оно
Connection diagram
4512 982 00751 (b/96.0)
@ Philips Medizin Systeme Z-7.1
C
Patient data organizer
EP GWB Stand Is]
a
1 !
015 EZX46 . data の 5 İÜ C300X1 +47
i |
EZX20 - να gs _ [ ceo PET
| 09 C300x2 nt 그 그
bas IC gm Ez139x5 İ serial data cable
EZX4: 1! 1
1
Emergency aff |1
ЕР, |
nen の τι
ENK23 に ゴイ T2 T2
ENKZ:5.—İT3 73 power supply 3x400V
earth leakage =
mains fuses orcuit breaker ST N N max. SA
I
. f '
AK mains feeder Lt u u
トー ゴ Surge
WN larrester (option)
1 ΤΠ Τι
eu — — — | A neutral conductar N is nat
ar 一 一 一 一一 | required if the mains transformer
perso — — — — — + 9890 000 02301 is ordered
EAA о ーー ーー ーー
supply
PRB-XRD products
EX1100
< cit
2T
—
stator cable 」
nN
us
2.
w一
2
“a
a=
3
3
3
x
o
m
»
E
k
R
|
|
|
|
|
|
|
|
|
|i
|
— — — —HY
cbla -一 一 一 一 一 一 一 一 一 一 一 一 一 一一 一
Lilo
EWG K14:6 — ml. TT ou cm 3 x.131mm<sCreened sms am ee meme cao sa coe mam Stator cable |, m m m m ie ad
EWG K12:2 == Um |
EWG K12:4 一 i L
stator cable il N
EWG K12:6 一 一 -一 人 tome一-一 一 了 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一一 =—
2 tubes
al
_ 쓰 . 64642 TTT
A OE OEE EO A A A A A A HA AA
1...
A
room decade |
A A A A O KA oe ee GR a m —
A EWGX3 A べべ . ....
Doom decade i Ço ...............
人ムへ EwGXS man «ООО 그 ㅡ . ㅡ . ㅡ. 60009 unit contrat! 그 그 그 그 그 그 그 그 고 그 一
LA. EWGX6 n nn KB neee ee ue
Sche.
96-03-14
ub0616
A2/A3
Connection diagram
4512 982 00751 (b/ 96.0} 2-7.1
© Philips Medizin Systeme
Stand [s]
EWG K112— — — 1
上
1
EtwG Ktt4 ーー て t
トー
1
— + L tube housing
1
E1wG K1t6 トー 一 wl — A — — be3x 131mm? screened stator cable | 1
i
Li _l
62\6 kt2 ーー 느
E2WG Ku — — L L tube hausing
E2WG ei — ul md, 3x 13 mm screened 。 _ stator cable İl
tT | “o
E2WG kz2 — L | |
E2WG K124—
— 너 | _ |
iL nte housing
ο smfseee вен on 의 의 으으 으으 의 의 의의 よー
E A rom ο... 그
= La mes Iİ | «........... AT ΗΕ. 그
m la me a ll キー remeeee i 0 그 그 그 의 의 의 의 이오 o
Aw|e gm ООО αςunit control ΝΕ ΕΕ”... 圖
Am —
Lâ wwe | | &........... 그 ㅡ ㅡ . ㅡ Cooling ο 0 ΜΜ —
(A
we me LL e o o tubesupervision nu |
PRB-XRD products
ion o
的 — SSM CN lla 그
Bucky/toma |
96-01-14
Le 出 一 一 一 一 一 一 一 一 一 一 ㅡ 페 lll +
LA 2 ee 2 了 q
Lamas To el release do A | free attachable
[A sare 1. ㅡ še TT TTT a
— — — — £— — — format contacts bucky 1 L. burky/tomo okay). |, i
Rİ terei contista? Lute —
———— bm LL ON --).)͵. 5
————4 tamatimes 3 ο 그 그 그 그 그 그 의 이 의 의 i
σπα μι ON 4
= El o APR extension (22RGDV ep ΓΗ = o
AM ο es eterna patient size compensation 一 一 一 二
一 二 二 di
device interface 2 1 a
DAMA И eat ООО A
+ me
νο ld 2 그 그 그 으으 으으 으으 으으 의 이이 의 .
PRB-XRD products
es A „i p free attachable
La twa 1... ㅡ ㅡ & ㅡ e O +
— — — — K — — — — — — — — — — — ㅡ farma? contacts bycky 1 L. bucky tomo ka) i _ a
ee TTT i
La mu Jo go BM TT +
A İk motes ge TD 二
At 그
aa TT i
Bucky/tomo
for examination units without CAN-interface
Schr.
96-03-14
ub0616
A2/A3
Connection diagram
4512 982 00751 (b/96.0)
@ Philips Medizin Systeme 2-7.3
E; Stand (sì
ii
1 I
|
1
1 ,
P i
I +
|
_— cured
in supply cable 。
t
1 4
|
EL 一 一 一 一 一一
“a i Surge arrester
1 earth th ofa hospital
ital! МЕХ
i tect
ective ital к -一一 一二 一(option)
option 一 一 一 一一 |
{
—@ ceiling suspension
! の 六 -一 一 一 一 一 一 一一 一
"V PRKT —Ø ceiling rails
! -全 一 一 一 一 一 一一
» 8.4mm2
一 八 fable
ÖN m ee ee ee n
ーの
PRB-XRD products
ELO 一 一 一 一 一 一 一 一 一 一 一一 —©
| À 一 一 一 一 一 一 一 一 一 一 一一 3 이 (一 一一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一一—2 tube assembly 1, 2, 3
|
i
了 一 一 一 一 一 一 一 一 一 一一二» 8.4mm
一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一一 -一 -一 ーの
famo ge 学生 ee
5
A A ee 一 —2
一 一 = mm“ gre !
一 一 一 一 一 一 一 一 一一-man 一 一 一 一 了了 了 一 一 一 一 一 一 一 一 一 一 一 一
| V —— 一 = Amme grey __ 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一一 ーの
一 一 一 一一—2 metal can of tube, 2,3
E 工 @一 一 一 一 一 一 一 一 一 一 一 一一 8002. — oy
Sche.
96-03-14
ub0616
A2/A3
Earthing diagram
4512 982 00751 (b/96.0)
© Philips Medizin Systeme 〆-7.4
OPTIMUS 50/65/80 INSTALLATION
e Mount the wall junction box at the place where the generator is intended to be installed.
(See drawing Z-3 in section 1 and manual UNIT 4512 103 75380 for wall junction boxes).
* If necessary, install the optional Surge Arrester WN inside the wall junction box.
To do this see Surge Arrester documentation.
* If applicable, mount the filler panels of the generator to the wall junction box.
* Have the mains cable present at the clinic connected to mains terminal MEX by a person who is authorized for this
job.
« Switch off the mains supply present at the clinic and make sure that it cannot be switched on again by anyone who is
not authorized to do so.
Incase a tilting angle of greater than 45° has been exceeded, the setting-to-work of the generator can be started not before
a waiting time of about 8 hours has passed. Otherwise the H.V. generator may be destroyed by electrical sparkover!
* Unpack generator cabinet E.
142H94
+ Take the two transport bars from the rear side of the cabinet.
* Lift the H.V. generator into the generator cabinet with the transport bars.
The 4 connecting bolts GX1001 to 1004 must point towards the front of the generator cabinet.
— ZX12 一 一一 一 G100X15 Route the cables along the front and left-hand edge
一 ZX35 一一 一 G100X14 of the H.V. generator and fix them!
e Fold the two earthing angles of the H.V. generator outward and screw it on to the members of the cabinet.
Screening cap
earthing angles 一~
199H96
184H95
@ Screw in the 2 feet for the unit at the bottom of the desk.
$
150H94
» Push the insert strips under the keyboard foil. Press the
angulated, protruding end of each insert strip into the
housing of the desk (3).
e Screw off the cable cover at the rear side of the desk..
° Provide drag relief for the supply and data cables with the clamp present on the desk.
e Screwon the cable cover. Make sure that the cable strain relief device of the release switch (1 cable tie) remains under
the cover,
s Set the 4 dowels supplied into the floor (drill bit: 10 mm).
* Screw on the desk stand with 4 screws (13 mm across flats) and washers.
+ Route the supply and data cables from the bottom to the top in the desk stand and provide the cables with drag relief.
Cable ends including plugs should protrude beyond the edge of the desk by about 500 mm.
* Assign the RGDV buttons 1...8 with the desired symbols as described under 3.1.
» Screwon the cable cover. Make sure that the cable strain relief device of the release switch (1 cable tie) remains under
the cover.
e Attach the operating panel on the stand.
— 4 nuts
* Set the dowels supplied into the wall (drill bit: 8 mm).
e Provide drag relief for the supply and data cables in the
wall frame. Cable ends including plugs should protrude
beyond the edge of the desk by about 500 mm.
* Mount the release switch as described under 3.1. Wall fastening (4x) Ball-head bolt (4x)
Cable passage
189H96
* Assign the RGDV buttons 1...8 with the desired symbols
as described under 3.1.
4. Electrical connection
4.1. Earthing
See Z-7.4 "Earthing diagram" in section 1.
e Measure the internal mains resistance at the terminal MEX or WNX1100 (option Surge Arrester) with a suitable
measuring instrument.
L1-L2: RE me
L1-L3: RE... mQ
IN ma
Required max. mains resistance at generator input (without Surge Arrester WN):
190 V * 40 mQ -
220 V * 130 mQ 60 mQ -
240 V* 160 mQ 80 mQ -
400 / 440/ 460/ 480 V 500 mQ 300/ 350/ 350/400 mQ | 200/240/240/300 mQ
2 Connect the mains cable of the generator to terminal MEX:L1/L2/L3 at the wall taking care that the pnase sequence
is correct.
lf the optional Surge Arrester WN is fitted, connect the cables at that point up to terminal WNX1100.
screening clamp
ーー | cable tie
0 0 00809 0 0
>) >
tube1 tube 2 198Η96
EZXI or
4.4.1. Room decade cable EWGX2 for the first room and
EWGX3 for the second room
+ Connect the door switches at the generator.
If none present, link pins 8 - 10. o
RMDROT| &
L ᾱ- door switches
--—-= screen
RM DR OV 8
143H94
b CM TH SW
Po26v| 8 一
HHS lamp
L ©- _ 24V/ max. 5W
=== screen
190H96
For examination units which are provided with a CAN system interface.
+ EZX 23 - signalbus
— The Bucky decades only have to be assigned if any of the following inputs are to be used:
1-2 format contacts Switch the external measuring fields ON/OFF
3-4 tomo mode bucky — tomo switchover
5-6 tomo ready tomo condition met
9-10 bucky ready bucky condition met
— The inputs are only activated by SW programming (see 9.5).
— After activation via the SW any missing inputs must be simulated by jumpers.
Example:
Format contacts on WAX11/12:1-2.
The outer measuring fields can only be selected in the closed state.
¢ At the junior/extremity measuring chamber withdraw pins 101-102-103 or A-D-H for measuring field selection.
These measuring chambers have only one measuring field. The terminal for the left-hand field is used in other
configurations for switching over intensification and must not be connected up here.
AMPLIMAT cables which are provided with a 3-PLUS-connector can be connected up using the adapter connectors
supplied.
> Mark the H.V. cables at the generator and the tube end with the correct polarity.
s FixtheH.V. cables on the left-hand side of the wail junction box on the middle rail for providing drag relief for the cables.
The short ends of the H.V. cables which are going to the H.V. generator must be routed in downward direction in this
area.
The free cable lengths including plugs should be about 1.5 m.
« Twist the H.V. cables counter-clockwise by one turn and connect them to the H.V. generator.
The twisting of the cables provides that the H.V. cables can be put into a loop when the cabinet is placed against the
wail.
The H.V. sockets should always be filled with some oil. At least the lower half of the plugs must be wet with oil.
Do not fit a silicone washer.
Do not rub them with silicone.
Notes
— The union nuts of the high-voltage connectors must be tightened up to ensure good electrical contact for screening.
— Only high-voltage connectors which have threaded flange halves may be used. Older high-voltage cables still have
connectors where the flange halves are kept together with a spring washer.
In such cases the modification kit 4512 103 80852 will be required.
5. Hardware programming
ello.1
sw controlled switch over
x4
6. Switching on the generator X2
» Call the installation program with xrgscope or xrgscope led for PCs with LCD screen.
The following menu line appears:
e Select "OPTIMUS”.
The following menu line appears:
General information:
— Button F1 <help> Call heip/cancel help.
— Button F2 <transmit> Store screen contents/data set in the generator = transmit to generator.
— Button F3 <Save> Store data set on disk; the path desired can be selected.
— Button F4 <load> Load data set from disk.
~ Button Esc Commands one step back; can be used repeatedly.
— Fields with I Select the possible range of values with the RETURN button = l.
The data are specified by the generator as fixed values.
— Fields with [...] Input of data via the keyboard.
Error numbers which appear at the beginning of the programming procedure must be erased from the screen with the
RETURN button.
Note
— Current data files, for instance, for online help, tube types, APR programming are available in BBS.
Product: Generatoren Hamburg
Download area: OPTIMUS
— If you call the installation program with xrgscope ? the possible starting parameters for the service program will be
listed.
8. Setting-to-work overview
This overview shows in what order the programming of a generator should take place.
The methods of programming are described in the following sections.
— Generator ON
9. Configuration
9.1. Date and time
e Select menu ”Program/ Date and Time”.
e Enter the respective local data.
9.3. Tubes
Note
During this procedure the CAN interface on EZX43 must be disconnected if present (THORAVISION or Bucky TH with
bucky controller).
+ Select menu "Program/ Tubes/ Tube 1...3/ Tube 1...3 Data Set”.
s From the list select the respective combination of tube type and housing type and press <Enter>.
+ Reset the generator with the ON button at the desk or button S1 on PCB EZ139.
Then the data which have been configured up to now are read by the processor when the system is started.
Caution!
Wrong programming could destroy the rotor control unit.
+ Select menu "Program/ Tubes/ Tube 1...3/ Tube 1...3 Speed Selection”.
e For each tube connected, program the maximum working voltage which is indicated on the data label:
Max. Tube Voltage Limit: default: 150 kV range: 40...150 kV
Adaptation of the tube is up to this limit.
If older tubes are to be operated on this generator, it is urgently recommended that the maximum kV used in practical
operation so far be specified instead of the theoretically possible value.
After adaptation of a tube the upper kV limit is displayed for each focus of each tube under:
Adapted to [kV]: e.g. 125
All the other limit programmings are performed by the generator automatically and do not usually have to be observed.
L[m] single length for 155 pF/m cable for 200 pF/m cable
14 4.085 4.400
16 4.240 4.600
18 4.395 4.800
20 4.550 5.000
22 4.705 5.200
24 4.860 一
26 5.015 一
28 5.170 一
30 5.325 一
The high-voltage cables type 9806 402 6xx02 currently being supplied have a capacitance of 155 pF/m.
+ Select menu *Program/ Dose Rate Control/ AMPLIMAT/ Chamber 1...5/ Data Set 1...5”,
The number of the chamber corresponds to the specified unit number of the dose measuring unit.
The choice between automatic and manual DRC processing is possible when an authorized hardware key is inserted in
the PC.
Automatic is selected as default and must be used for the initial programming.
Access manual DRC processing by pressing the Esc key.
The manual mode is suitable for:
— Copying complete programming to other measuring chambers
— Setting the basic density
— Changing the desk-displayed names of the programmed film-screen combinations
— Creating backups of the DRC programmings
+ Select the desired data from the files offered for the following programming steps.
The files are part of the installation software.
— Select the programming field with the cursor and enter <Return>.
— Enter the desired file from the list offered.
~ Select the desired data as required.
Based on the combination of the components entered, the processor calculates the switch-off dose, kV correction and
RLF compensation and creates a name for the film/screen combination, e.g. ”B400”.
Since the data selected are not directly stored in the generator, it is recommended that they be entered in the following
table.
Film: onoganangasansanonaaas
Screen:
Data Set 1
Cassette: 20000000000000000006000
Film: seepesaeoopoeewamnonoo
Screen:
Data Set 2
Chamber: ο ΟΣΗ
Cassette:
Sys.corr.: 000006000000000000000 ㅁ 2 20000000000000000000000
Film:
Screen:
Data Set 3
Chamber:
Cassette:
Sys.corr.:
Corr. factor: 0000000000000000000000
Film:
Screen: 0000000000000000000006
Data Set 4
This monitoring does not take effect in the following cases, irrespective of programming:
— Using film/screen combinations with high speed in AEC technique.
— Exposure time in TDC technique is lower than 1 s.
* Program the data set A and B of RGDV 1.....8 for all exam./aux. units desired.
Data set A:
Bucky Controller 1 may only be programmed for RGDV 1...4 and Bucky Controller 2 may only be programmed for
RGDV 5...8.
An RGDV must not be assigned a "Mounted radiographical controller" and a "Release circuit adaptation unit" together.
Data Set B:
Il
mAs steps: Step width in 25, 12 or 6%.
mA steps: Step width in 25, 12 or 6%.
time steps: ” exposure time step width in 25, 12 or 6%.
Density steps: Step width in 25, 12 or 6%.
Density correction (6% steps): -8 ...0...+8 correction steps.
For correction see chapter 13.
Underexposure display
(non-automatic techniques): Yes Underexposure is also indicated with techniques without AMPLIMAT.
No = e.g. tomo
Tube overload protection: On = Overload protection active (default): red = exposure not possible
Off = Exposures are possible irrespective of load status.
desk display tube load
green 100 %
green — yellow 100 %
yellow 80 %
yellow — red 64 %
red 0%
« Select menu "Program/ Registration Devices/ RGDV Interface Assignment/ Bucky/Tomo 1WA...2WA’.
There must be no programming here if the diagnostic unit is connected up via the CAN interface.
* Assign the format and ready contacts of the decade connector WAX11 or WAX12 to a bucky or tomography RGDV.
Refer to Z1-15.1.
ー Decade Bucky1 (X11) See following table.
- Decade Bucky 2 (X12) Program the functions as for the first Bucky decade.
But both the tomo mode switch and the tomo RGDV may not be activated twice.
Tomo mode switch: disable = Input ’tomo mode” is not activated. Changeover
Bucky/tomography not possible via the examination unit.
enable = Input tomo mode’ is activated. Remote changeover
Bucky/tomography possible.
Bucky and tomo RGDV must be defined.
Programmings in ( ) relate to the option "Automatic tomographic time input” via assembly UP of HDH.
The generator must be in the READY state, i.e. the green ring at the desk must be illuminated!
e Reset generator.
+ Select appropriately programmed auxiliary for the respective tube to be break in.
+ Select large focus.
e Release exposures according to the following table.
The exposure run must always be made at one kV level without any repeated start-up of the tube, i.e. the PREP button
must always remain pressed during the run.
In case an error message occurs during the adaptation procedure, reset the generator and repeat the adaptation for this
particular focus.
* Check whether the upper KV limit for data adaptation in the menu "Program/ Tubes/ Tube Limits” is programmed
according to the tube connected.
lf older tubes are to be operated on this generator, it is urgently recommended that the maximum KV used in practical
operation so far be specified instead of the theoretical possible value.
* Repeat the adaptation procedure for each additional focus and tube.
Do not try to adapt VARIOFOCUS as middle focus.
VARIOFOCUS is a combination of both small and large focus.
The kV-dependent mAs limits can be accessed via the menu "Program/ Application Limits/ Thoravision Limits”. They
are activated only in conjunction with an on-line THORAVISION unit.
A change may only be made if instructed to do so by the service centre.
Ona single RGDV button either up to 80 APRs can be programmed directly (10 pages of 8 each) or up to 250 APRs via
menus.
The initial data sets are called ### APR name ### and they have the same exposure parameters.
They can be directly assigned or via menu and submenu levels to registration devices RGDV 1...8.
In case "Test APR’ is displayed after selection of a registration device, at least this particular registration device has not
been assigned to any APRs.
12.1. Language
The language for operating instructions is selected in this menu.
s Select menu ”Program/ Human interface/ Select Language”.
e Select the desired language:
— English
German
French
~ Spanish
Certain characters can be generated at the PC only via the decimal code. To do so, press the "Alt" key on the PC and
enter the numerical code.
~].-
~
ol-
of
alo!
lol
оо
<|
ml
slelrnl
alelu
人 |上 oo 和
w]
al
мы
al
м
ola
slolai
ol]
чо
оо
Nl
ni
oj
oj
a]
Фо
ol
ol
ol
©
>
vl v
?
ol
^う
Ol
@
om]
Of
bleml
bla
A
Pl
>|
B
ul
σα
wy
ο
60]
ol
oO]
оо|
D
al
σι
OF
E
m)
πι
nm] my
F
отм
nf
πι
G
工 | G&G]
Ω|
TO
H
τι
| と| 一
| こと| 一
elo
2 | Z| 5”
с
г
M
s
=
N
o| z
z
o| z
o
о
vo
P
o| v
molu
мощ
a
oj mo
R
m
S
ale
ao
о
T
3
U
¡| <p Cc
NI <| xi sl <| c
zl <| c
<
W
Zl
X
|
X
<| x
Y
NI <|
<
z
NÍ
BÍN|
[
\ >| 一 | -| —|
]
A
y y y y y
z z z 之
{ a é . { Alt +123
| é ů ň 1 Alt +124
} ů é C } Alt +125
- B - ~ ~ Alt +126
A Alt +127
A 4 A A а Alt +160
u u = a 1 Alt +161
+ + + + ó Ait +162
£ £ £ £ ú Alt +163
. . . . ň Alt +164
§ § § § 5 Alt +167
Il It tt Il é Alt +168
| | | | ® Alt +169
= = = = - Alt +170
= = = = % Alt+171
Alt +172
АК +173
Alt +174
Alt +175
5 o ° o Alt +176
+ + + + = Alt +177
2 2 2 2 E Alt +178
A A A A L Alt+192
Á Á Á Á 1 Alt +193
Á Á Á Á T Alt +194
A A A A k Alt +195
A A A A — Alt+196
Á A A A + Alt +197
Æ Æ Æ Æ a Alt +198
¢ С ¢ С А Alt +199
E E E E IL Alt 4200
E E E E F Alt +201
E E E È dk. Alt +202
E E E E F Alt +203
1 i i i E Alt +204
i [ | | = Alt +205
1 1 1 1 dL Alt +206
| i | İ я Alt +207
6 Alt +208
N N N N D Alt +209
О 0 0 0 E Alt +210
Ô Ò о Ô Ě Alt +211
Ó Ó Ó Ó È Alt +212
e $ ç ç Ρ Alt +231
è è è è b Alt +232
é é é é ύ Alt +233
é é é é Ù Alt +234
8 @ é é Ù Alt +235
i ὶ i i ý Alt +236
i i i i Ý Alt +237
i i i î 3 Alt +238
i i ϊ 1 ‘ Alt +239
a Alt +240
A fi A ñ + Alt +241
(の č č à = Alt +242
6 6 6 6 34 Alt +243
6 6 ô 6 Alt +244
6 6 6 6 Alt +245
6 6 6 č + Alt +246
= Alt +247
σ o o o ° Alt +248
ù ù ù ù Alt +249
ú ú ú ú Alt +250
a a a | Alt +251
a ů ů ů 3 Alt +252
y ý ý ý 2 Alt +253
5 Alt +254
Alt +255
« Select menu "Program/ Human interface/ RGDV related assignment/ RGDV 1...8/ Predefined assignment’.
* Select with TAB and cursor-down key one of the files listed up, e.g. "ar65eng.tdl",
Meaning: a = APR data file
r = radiography
65 = version (month/year)
m = mono focus tube
p = pediatrics
V = with VARIOFOCUS settings for RO 1750, SRO 2550, SRO 33 100
V9 = with VARIOFOCUS settings for SRO 0950
eng = language; en(g) = English, de(u) = German,
es(p) = Spanish, fr(a) = French
* Select one of the applications listed up, e.g. "Bucky”, and load the data file.
Applications:
— Bucky bucky
— Wallstd bucky at wallstand
— Free . free cassette
~ Tomo LT/HDH linear tomography with units HDH, BTS2, BTS4 (group)
~ Tomo LIN linear tomography with units HDH, BTS2, BTS4 (paging)
— Тото ВТС tomography with unit Bucky DIAGNOST TC
— Тото ВТН tomography with unit Bucky DIAGNOST TH
Extension GR = Group; APRs are divided into groups (menu technique),
Extension PA = Paging; APRs are assigned directly to an application.
Registration devices 1 2
RGDV 1...8
3 4 у
OF πι О
OL ||22
o. ALajo
«co / APR Assignment/ Assign APR
(... / APR Deassign / Deassign APR )
* Ifrequired, change the location suggested in the display. Otherwise the next vacant location is assigned.
» Select with the cursor from one of the windows a menu to be assigned with submenus.
» Select menu "Program/ Human interface/ RGDV related assignment/ RGDV 1...8/ Manual assignment/ Menu
assignment/ Assign menu’.
+ Ifrequired, change the location suggested in the display. Otherwise the next vacant location is assigned.
e Select with the cursor from one of the windows a menu or submenu.
If no menu layer is desired, proceed to assign APR.
* Select menu ”Program/ Human interface! RGDV related assignment’ RGDV 1...8/ Manual assignment/ APR
assignment/ Assign APR”.
* Select from one of the windows an initial APR or an APR which has not been assigned.
Initial APR: *### APR name ###
Not assigned APR: e.g. "Thorax ap”
+ If required, change the location suggested in the display. Otherwise the next vacant location is assigned.
Only after a reset of the generator the menus, submenus and APRs are displayed on the desk.
* = The basic setting of this data can also be performed from the desk ("Reset” + APR). Refer to the operator’s manual.
Shifting:
+ Select menu "Program/ Human interface/ RGDV related assignment/ RGDV 1...8/ Manual assignment Move menu”.
Re-naming: (
« Select menu "Program/ Human interface/ RGDV related assignment/ RGDV 1...8/ Manual assignment/ Rename |
menu”.
If only because of the limited scope for labelling it is recommended that a maximum of two assignments be selected per
APR extension.
If only one APR extension is to be connected up to 8 APRs and if there is only one WA assembly, the latter assembly must
be programmed as 2WA by closing the soldering jumper W1 on its back panel.
* Define assignment of the extended RGDV and APR keys and enter in the tables below:
— For the first APR extension select two RGDVs.
— Read out the respective number of the APRs to be assigned.
To do this select the corresponding APR at the control desk and establish the respective number using the menu
“Program/ Human Interface/ APR Data Set/ Select APR Data Set”.
Device Interface 2
Assignment for RGDV... Assignment for RGDV...
s Inthe corresponding menus "Human Interface/ RGDV related Assignments/ RGDV 1...8/ External APR Assignments/
Device Interface 1...2” save the numbers determined.
Desired density
New switch-off dose = —— x Old switch-off dose
Measured density
The switch-off dose can be set on the PC even without a hardware key.
To do so, call up the automatic DRC programming, repeat all the selections and change the correction factor for switch-off
dose accordingly. Each time this programming is called up all the selections must be repeated.
Organ-dependent correction:
Select the APR to be changed on the desk.
Select menu "Program/ Human interface/ APR data set’ Select APR data set”.
Confirm the APR number displayed with <Transmit>.
Select menu "Program/ Human interface/ APR data set/ Change APR data set”.
Exposure data density: —16 … +16 = correction in steps of 6%.
The number of correction steps must be matched to the programmed step length of the desk display.
Example:
The desk display has been programmed to the R20 series (=12%) in the menu "Program/ Registration devices/
RGDV 1...8/ Data Set B/ Density steps”. To be able to display a density correction of +1 for a certain APR two
corrections steps (2x6% = 12%) must be programmed under this APR.
Select the next APR on the desk, select it in the programming menu and change it.
Etc.
+ Save the new parameters as default values. To do this press the “Reset” button and the corresponding APR button.
The asterisk in the APR name as an indication of overwritten data disappears.
Exposure data O: anatomical mAs product based on the screen-film combination used.
!
Il
In the APR standard files supplied the following APR is programmed to kV-mA technique
(language: German/English/French/Spanish):
~ Dens axis F / dens axis F / densaxis F / atlas F
The APR is marked with ”F”.
The mAs value is based on a 400-type screen-film combination and must be adapted to the combinations actually used.
If, for example, the 200-type combination is used, the mAs value must be doubled.
If the TDC option is installed, the preferable technique for all exposures is the one where the exposure time is the
determining factor. TDC is not restricted to tomography applications.
e Select menu "Program/ Human interface/ APR Data Set/ Select APR Data Set”,
+ Note the number of APR data set ”x", for instance, 100.
Changing of the number of APR data set "y" to be moved/copied to the number of APR data set "x"
» Select APR data set ”y” on the desk.
e Select menu “Program/ Human interface/ APR Data Set/ Select APR Data Set”.
Transmit with F2.
+ Select menu "Program/ Human interface/ APR Data Set/ Change APR Data Set”,
¢ Replace the number of APR data set ”y” with the number of APR data set”x” in the input mask, for instance, nnn > 100.
« Select menu "Program/ Human interface/ RGDV related assignment/ RGDV 1...8/ Manual assignment/ APR deassign/
Deassign APR”.
An APR which has been deleted is no longer displayed on the desk but remains stored in the generator. It can be
re-activated according to 12.3.3..
For U.S.A.:
After completition of setting-to-work, the system must be tested for H.H.S. compliance according the PM.S..
comprehensive compliance testing workbook (code No. 4535 800 2035.).
To save the configuration data use the "Configuration Backup” disk supplied.
+ Save the complete SW programming of the generator using the menu "Accept/ Backup/ CU Complete’
on the floppy
disk. .
* Recommendation:
In addition, save the APR programmings individually for each RGDV using the menu "Accept/ Backup/ RGDV related
Assignments/ RGDV 1...8/ APR Assignments” on floppy disk.
File name: apr_bak#.tdl # = RGDV-number
Note
In a backup of the APR programmes all the customized assignments of film-screen combinations will be lost. If APR
programmes are loaded into the generator using the Restore command, it is always the first film-screen combination
which is assigned to a measuring chamber as the default (data set 1 of chamber 1... 5).
To restore the customized APR assignment, it is absolutely essential that you make a note of which other film-screen
combinations (data set 2... 5 of chamber 1 ... 5) are assigned to which APRs.
We recommend creating the information on the PC as a simple text file in the following sequence:
- RGDV1... 8
— Menu name and, where applicable, submenu name
~ APR
Data Set 2...5 and/or name of the film-screen combination, e.g. G400
and storing it on the backup floppy disk.
e Recommendation:
In addition, save the programmings for film/screen combinations using the menu "Program/ Dose Rate Control
AMPLIMAT/ Chamber 1...5/ Data Set 1...5” (manual processing) and with the SAVE function (F3 key) on floppy disk.
Recommended file name: drc##.tdl ## = Chamber and Data Set Number
* Provide the floppy disk with the serial number of the generator.
e Keep the floppy disk in the service documentation.
21. Labels
All lables become visible by swiveling out the label bracket simply by hand and without any tool. The bracket is located
at the top left corner of the front side of the cabinet, visibly marked by an”? (for information) and text "Certified Component
Lables Here”. If you swivel the label bracket 90 degrees to the right the following labels will appear at its bottom side:
~ X-Ray Control: ー type designation
- serial No.
— name and address of manufacturer
~ DHHS certification statement (if necessary)
ー date of manufacture
— X-Ray H.V. Generator: — type designation
— serial No.
— name and address of manufacturer
— DHHS certification statement (if necessary) |
- date of manufacture (
— Technical Data label with UL/CSA classification (if necessary)
1 OPTIMUS 65 ( | En CD Ji )
4B0V 5A 1454
> 480ν 4a 120 A
H.V. generator
1 OPTIMUS 80 Kul )
—
[ 3— 58/60 н2 | (9) 1
400 SA 190 A
4BØV 4A 160 A
1 OPTIMUS 100 65
"
?
identification
ot
2 | = PHILIPS |||
MADE IN GERMANY
7
400ν
482V
SA
4A "m | 2328
192 A
Generator 一 一- - ] J
PHILIPS-nameplate | plate
Rating ss == TT cabinet te
upe 4890 000 0200x배부
| RB type Produktname
5 OPTIMUS xxx D-22335 Hamburg/Germany 8 s/n Fobr・ Nr
、 code Code Nr.
Regulation- dependant labels
(—-X-RAY CONTROL- Philips
Medical Syst ]
type 9890 020 0200x | DeeepmnidM gwtewialens ~
SİM ΧΧ ΧΧΧΧ Réntgenstrasse 24 9
D OPTIMUS xxx D-22335 Hamburg/Germany
CERTIFICATION This product complies with the
Per formance Standard under the Radiation Control HHS
for Health and Safety Act of 1968, applicable С 때 O DD
一 Lot date of manufacture. Manufactured: MONAT JAHR |
| 350/60 HZ | WARNING:
49gv SA 230 A THIS X-RAY UNIT MAY BE DANGEROUS
Mains transformer [iNX-RAY HV CONTROL ~ Philips Medical Systems | ) 7 一 ~ 48gV
EOUIPMĚNT I АА 198 A TO PATIENT AND OPERATOR UNLESS SAFE
1 1 EXPOSURE FACTORS AND OPERATING | HHS
(option)
190...390V
7 SR ie o 200 02003
CERTIFICATION
mo
This product complies with the
ο
RSntgensfrosse 24
DO E
KW SHOCK AN
CLASSIFIED BY UNDERWRITERS LABORATORIES INC
Gi
(R)
LR37323C |
UL
CSA
INSTRUCTIONS ARE OBSERVED.
Certified
Performance Standard under the Radiation Control HHS 12 Component
for Health and Safety Act of 1968, applicable Г DN Labels Here
8 3 ん Lat date of manufacture. Manufactured: MONAT JAHR J 3— 58/68 н2 | CD hı
4 の 9V ЗА 260 A
Schr.
г N 480V 4 A 220 A
/ / 350/0 nz | C) | 7 = X-RAY EQUIPMENT =
400V sa 1450 100 CLASSIFIED BY Date LABORATORIES INC (R)
' o 7 4894 44 120 A WITH RESPECT TD ELECTRI (CA .
— RAY EQUIPMENTS É UL KW SHOCK AND MECHANICAL HAZARDS ONLY LuR37323c JUL
96-03-14
a~so/eoHz | CD li
A3/A3
ADELINA
山
FAULT FINDING
TEXT
Contents ........................
1. eee 3--0.1
Tools --υ-υ-------- 3-1
a
Notes ............,...........,.....,...44.,
K K K S O K R K O K K K K KKK 3-1
N
Strategy .......................
0.00. 3-1
9
DRAWINGS
Central rack, Service aid ............................
1... eserse 3Z—1
Tools
mb,
.
~ mAs meter
Multimeter
PC incl. 3.5” FDD, HW-dongle, serial interface cable, free RAM = 590 KB
2. Notes
Caution!
After the generator has been switched off, hazardous voltages are still applied to the d.c. intermediate
circuits of the converter, the rotor control and the mA control.
These voltages are usually discharged within 1 minute to values which are no longer dangerous.
3. Strategy
There are 3 categories of errors:
— The generator cannot be switched on at all or only for a short time.
See => 5. "Initialization phase of the generator”
=> 6. "Switch-on not possible”
— The generator can be switched on but no error numbers are displayed on the operating desk.
For fault finding use the service PC.
See => 4. "Connecting the service PC”
=> 5, "Initialization phase of the generator”
=> 7. "Error numbers”
— Error messages are displayed on the desk.
For fault finding use the service PC.
See => 4. "Connecting the service PC”
=> 7. "Error numbers”
4. Service-PC
4.1. Connection
GND — 5 5 ~ GND
(9--pole, female) (9-pole, male)
4.2. Operation
e insert the floppy disk with the service program in the PC.
。 Call the program with xrgscope or with xrgscope led for PC's with LCD screen.
Note
— Current data files, for instance, for online help, tube types, APA programming are available in BBS.
Product: Generatoren Hamburg
Download area: OPTIMUS
— If you call the installation program with xrgscope ? the possible starting parameters for the service program will be
listed.
Г —— APR Deassign
Select Menu
Deassign APR
r—— Menu Assignment
Select Menu
Assign Menu
ーーーー Delete Menu
= Move Menu
Rename Menu
External APR Assignments
Device Interface 1
———- Device Interface 2
APR modifiable by User
— Adjust
-— Tube Adaptation
—— CAN Auto Configuration [w-8 ]
ーー Area Exposure Product [w-—8 ]
Specific Yield of Tubes
Specific Yield of Tubes 1 ... 3
Add Filter Correction Tables
2 mm Al
1 mm Al + 0,1 mm CU
1 mm Al + 0.2 mm CU
Wedge Filter Correction Table
Wedge 1
Wedge 2
Finger Wedge
-— Dose Rate Control [4 ]
TDC AMPLIMAT (Don’t change the values without order from DMC Hamburg!)
—— Boost Adaptation [w—#]
m Accept
ーー Backup [w-3 ]
(7 RGDV related Assignments
RGDV1..8
APR Assignmenis
— CU Complete
ーー Restore [4 ]
~~ RGDV related Assignments
RGDV1.. 8
L APR Assignments
— CU Complete
—— İnspect
-— Exposure Counter
— Type of Tube 1 ....3 m)
— Faultfind (w— |
—— Power ON Results
— Options
— Internal CAN Configuration
— SW/HW-Versions
ーー Logging Table
— Error Log
Error Log Index
Select Error Detail
Error Log Detail
—> Error Detail of CU
Error Info
Program Trace
一 一> Error Detail of kV
Error Info
Error Trace
Error HW Set Values
Error Read HW Values
Error Specific Information
—-> Error Detail of mA
Error Info
Actual Status
Status Trace
HW Set Values
Read HW Values
Specific Information
Error Log Clear
r X-Ray Log
Tube Temperature Supervision Logging
Tube Temperature Supervision Temperature Log
Tube Temperature Supervision Load Log
Tube Temperature Supervision Error Log
Dose Rate Control Logging
Read Actual Status
None Automatic Technique Calculation
AEC
——— AEC Calculation
—— AEC Trace
7- 7DC Calculation
TDC Trace
— CU Trace
ーー Quit
FU-kV [8]
— Faultfind
—— Power ON Results
L_ Read Configuration
|... Logging Tables
E Read Trace
Read Error (Only at current error with contents)
Error: Info
Error: Trace
Error: HW Set Values
Error: Read HW Values
Error: Specific information
-— Functional Test
Г Test Watchdog
— Test DAC-ADC
-— Test Converter
L Switch Error Handling
— Monitoring
~ Measure Temperatures
— kV Measurements
m Converter Measurements
ーー Read 8 Bit Port
— Read 16 Bit Port
— Read Memory
— Quit
FU-mA [κ-- ]
— Program
L_ Read Focus Limits
— Faultfind
—— Logging Tables
Read Actual Status
Read Status Trace
Read Error (Only at current error with contents)
Error: Info
Error: Status Trace
Error: HW Set Values
Error: Read HW Values
Error: Specific Information
—— Functional Test
L Test Watchdog
r— Monitoring
Read k Nominal
Read I; Actual
Read Intermed Circuit Voltage
Read |, Measurement Trace
Read 8 Bit Port
Read 16 Bit Port
Read Memory
—— Read I, Corrections
—— Adaptation Results
Select Adaptation Table for Reading
Read Previousiy Selected Adaptation Table
—— Quit
In case data get lost they can easily be restored in the CMOS areas after the error source has been eliminated.
Saving of data:
s Select menu ”Accept/ Backup/ CU Complete”.
+ Store the data on floppy disk "Generator configuration data” found in the service documentation.
Default file name: cubackup.tdl
Recommended file name: s/n of the generator, e.g. 960007. idl
File size: approx. 250 kB
Transfer time: approx. 6 mins.
e Recommendation: (
In addition, save the APR programming individually for each RGDV via the menu “Accept/ Backup/ RGDV related
Assignments/ RGDV 1...8/ APR Assignment” on floppy disk.
File name: apr_bak#.tdi #= RGDV - number
Assignment of film/screen combinations to the individuai APRs is not saved in this procedure!
e Recommendation:
In addition, save the programmings for the film/screen combinations via the menu ”Program/ Dose Rate Control/
AMPLIMAT/ Chamber 1...5/ Data Set 1...5” (manual processing) and store them with the SAVE function (F3 key) on
floppy disk.
Recommended file name: drc##.tdi ## = chamber and data set number
Assignment of film/screen combinations to the individual APRs is not saved in this procedure!
Restoring of data:
Note
During this procedure the CAN interface on EZ X43 must be disconnected if present (THORAVISION or Bucky TH with
bucky controller). (
Most of the programmings and logging tables can also be stored via the SAVE-function (button F3) of XRG SCOPE.
— For service use, only keep the latest version of the backup.
Pulling-up of ENK 2
Selftest of ...
control desk C: All display elements are switched on for a short moment.
central unit EZ139:
kV control EZ130: voltage E is measured in the d.c. intermediate circuit.
mA control EZ119:
basic interface EZ150:
rotor control EY:
universal I/O EWA/B 102:
Indicating device: The red status LED of the associated printed-circuit board or assembly is illuminated.
When the selftests have successfully been completed, the status LED’s are blinking.
The central unit establishes connection to each functional unit via the CAN bus.
Indicating device: The red status LED of the associated printed-circuit board or assembly grows dark.
|
The generator is internally ready.
l
The external ready circuits are checked (unit ready, door contact closed, thermal contact of the tube closed, tube not
overloaded).
Test APR No APR data have yet been loaded onto the selected
RGDV.
81 kV Ready status.
0 4
An APR with AEC technique has been selected.
sku axial
Schädel ax. 4 craneo axial
22-2
7. Error numbers
Analysed errors:
— These errors are indicated by 4 digits.
— The first two digits indicate the functional unit FU reporting the error.
00xx = CU-functional unit is concerned
02xx = kV-functional unit is concerned
03xx = maA-functional unit is concerned
etc.
Displayed errors:
— These errors are indicated on the display of the operating desk for the customer.
— The customer must call the service.
The customer can inform the service about the respective error nurnber and the service can order the spare parts |
needed at an early stage of the maintenance procedure.
— In case an error of this category occurs frequently within a certain period of time, a displayed error can be generated.
00BS APR: The RGDV of the APR is not ready for operation
00CM CAN: FU #1H sent event and did not answer RTR
00M0 Unable to initialize FU(s) #1H, #2H, #3H, #4H, #5H, #6H
00M3 No response at all from FU(s) #1H, #2H, #3H, #4H, #5H, #6H
00Mx ER_error$code$MC
O0TB TTS: Tube Supervision Error from FU kV; thermal switch of tube housing okay?
00Ux ER_error$code$SC
02HC W Z nominal value out of range: + 1 % + 0.2; 3 detections within 30 ms; duty cycle range 3 %...30 %
O2HE W kV value during standby too large: > 3 kV for > 400 ms after PŘEP
02HF E kV value during standby too large: > 4 kV for > 400 ms after PREP
02HH Ε kV actual value out of range: 20 kV > U > 170 kV; 3 detections within 30 ms
02HI W E value during standby out of range: 470 V > E > 780 V; 3 detections within 30 ms
02HJ E E value during standby out of range: 450 V > E > 800 V; 3 detections within 30 ms
02HK W E value during high tension out of range: 400 V > E > 780 V; 3 detections within 30 ms
02HL E E value during high tension out of range: 350 V > E > 800 V; 3 detections within 30 ms
02HM W converter 1 temperature out of range: 0 °C > T > 85 °C; 3 detections within 30 ms
02HN E converter 1 temperature out of range: 0 °C > T > 90 °C; 3 detections within 30 ms
02HO W converter 2 temperature out of range: 0 °C > T > 85 °C; 3 detections within 30 ms
02HP E converter 2 temperature out of range: 0 °C > T > 90 °C; 3 detections within 30 ms
02HQ W high tension tank temperature out of range: 0 °C > T > 80 °C; 3 detections within 30 ms
02HR E high tension tank temperature out of range: 0 °C > T > 85 °C; 3 detections within 30 ms
02HS W divider test cathode out of range: 45.5 kV > U > 50.5 kV; 3 detections within 30 ms
O2HT E divider test cathode out of range: 43 kV > U > 53 kV; 3 detections within 30 ms
02HU W divider test anode out of range: 45.5 kV > U > 50.5 kV; 3 detections within 30 ms
02HV E divider test anode out of range: 43 kV > U > 53 kV; 3 detections within 30 ms
03PB W le out of tolerance: + 10 % (lp > 5 mA, exp. time s 44 ms) or £ 3% (le > 5 mA, exp. time > 44 ms)
03PC E I, out of tolerance: + 30 % (le > 5 mA, exp. time > 44 ms)
08DD error when offset out of range in APR data structure while surging APR
08SA error when requesting message object to CAN_tx_task from scan task
08SB error when requesting message object to CAN_tx_task from test task
8. Power supply
ENF1 released.
Phase supervision
a) Without mains adaptation transformer:
— Phase L1 is missing: Mains contactors ENK2 and ENK1 cannot be activated.
— Phase L2 is missing: |The generator can be switched on but does not go into the READY state.
The filament-circuit supply is missing.
There is an error message from function unit kV.
— Phase L3 is missing: o "ON” circuit without supply voltage.
Fault tracing:
The generator cannot be brought into the READY state (e.g. no desk display).
Check the low-voltage supply.
— ENF1 released:
Ground fault/short-circuit of one/several phase(s).
Check ENK2 and, if necessary, the contacts of ENK1.
Check the leads and the mains adaptation transformer.
Have contacts ENK2 or ENK1 dropped out?
Check visually. Be careful when doing so since the unit is still connected to mains.
The damping resistors are unsoldered because the converter was active and ENK1 was not switched on although
activated by the software.
Probably termination of exposure.
This procedure can only happen once since the generator cannot go into STANDBY when intermediate-circuit voltage E
is missing.
In case intermediate-circuit voltage E is present, ENK1 is activated by the software of the kV-control and remains
activated for the complete time the unit is in operation.
In case of high impedance or when the tolerance of the symmetry resistors of the intermediate-circuit capacitor battery
is too large, capacitors may be destroyed by overvoltage. In case ENK1 has already been activated, ENF1 will probably
release.
The release of ENF2 switches the generator off since the supply voltage for the "ON circuit and, consequently, the supply
voltage of contactors ENK2 and ENK‘1 is interrupted.
9. Converter
— Short-circuit on one of the two capacitors (in case both capacitors are concerned, ENF1 will be released):
Low resonant-circuit frequency.
The IGTBs can break because of overcurrent.
Overvoltage at the resonant capacitor which is intact.
kV overswing.
DC short-circuit current possible because of resonant current which has not yet died off.
— Short-circuit:
IGBTs defective. DC short-circuit current causes the release of ENF1. The resistors of the protective wiring might be
destroyed in advance.
The NTC resistor for temperature measurement is supervised via the software with respect to logical values.
The valid temperature range is between these error conditions.
Open/shorted measuring circuits or any values going beyond the temperature limits will cause an error message.
¢ Before the ENF1 is pushed back to the on position check if all contacts of ENK1 1-2, 3-4 and 5-6 are open in the
non-energized condition of the relay. If not, replace the relay before you switch on ENF1 and proceed with other test
activities.
The first thing to look at will be the emitter-collector / emitter-gate impedance at every IGBT 1 to 4. If all 4 of a kV power
unit are not 0 Ohms (50 kW) and none of the 8 of a double converter generator is on 0 Ohms one should not suspect the
power unit(s) (so far).
Are there any damages on the driver PCB(’s)?
e Checkthe snubber diodes V 500/501 /502/503 for short-circuit. If one has a short-circuit some of the resistors linked
to the damaged diode(s) must also be open or have some overheat characteristics.
The second step should be the measurement of the rectifier(s) EQV5 (E2QV5). It could have been damaged from
overvoltage (surge). Look for short-circuits and, after the next switch-on, for error codes O2HI and/or 02HJ (E_value out |
of range = DC power supply) in the error log index.
O2HI = 470V> E value > 780 V in standby = 30 ms
02HJ = 450V> E value > 800 V —dto—
* Remove the driver PCB(’s) to look at the current tracks for short-circuit (insulation damaged?)
e Check all 4 DC capacitors for short-circuit. Are the DC symmetry resistors R1 + R11 ok (47 kQ)?
e Switch on ENF1.
With switch-on the converter DC supply will be charged via the dumping (spring) resistors EN R1, R2 and R3. If there
is still any kind of short-circuit in the machine we could not measure with a (low voltage) Q-meter and/or there is a part
in the generator which fails when the AC or DC increases a certain level, one or two of the spring resistors might become
very hot and will open.
If it does not happen, measure the converter DC supply at ENK2 41(+) and 42(-). It should have a value between 480 V
and 750 V.
If the generator is in a stable standby condition, proceed with the converter driver test without converter DC supply.
Caution! QX1101
Before this driver test can be carried out the kV power 20 X1101
unit(s) must be disconnected from the mains supply Q X1102
(leads of unit(s) EQ/E2Q to ENK1 :1, :3, :5). 20 Χ1102
QX1103
This safety measure is also valid for the chopper test to
2Q X1103
guarantee that the measurements can be carried out
without any risks involved.
I 000
1 3 56
X1
EN100
174H95
The KV driver test is software controlled via PC. Due to the missing PREP and exposure requests the signals EN_X_C/
and CTRL X C/ have to be set low-active at the backpanel at locations X76 and X74 (see drawing Z2-5.1/2).
Caution!
Do not forget to remove these connections after the test. Otherwise kV will start immediately with the PREP
command in normal application mode.
* Check whether the gate voltage is about -14.2 V +0.3V against emitter for every IGBT.
e Check the +15 V supply for the IGBT drivers now. Drivers 1 and 2 are supplied by chopper 1 while drivers 3 and 4
are supplied by chopper 2. The common zero point is the emitter.
OPTIMUS
147H95
08890 801 00 97
JAMOd AX
00103
001 D3
HH
ie
O OP
POA
V505
[ [
y
i
2028그
cova —]
OO
$023
exl_| EU Leu
q= 6469 8159 LES 9169 61569 $150 61599 ZISO A.
1X £0G3 2059
vosy
so 그
SLUSH 9068
coca
3-34
3-35
"6980 801 CH9y
FAULT FINDING
173H96
Jomod AX
001 003
001 03 Ζ0ΥΧ
007
102X 002X
ced DO Ley
σουχ ecu ¿cu
a ]ux
a ex
err Torv
Στρ. 00%
ZOSA
cosa | 1192 | 0159 | 6052 | 8062
7
0052
(b/96.1) E
2053 #062
8054
TOEY ip
1054
or
00TY
1158 9054
σ0εΧ rosa (9198
= 60S4| | 1062 | 9052 | 5093 | pog3 | —
005A
cOLX
a 品
EM 3 TOTX OOTX
3
OPTIMUS 50/65/80
9
x
[1668 934 DI
CC] 084 сен ビコ
b981E
SPAM UO € |BAd}
OPTIMUS
OPTIMUS 50/65/80 3
WOl} LEX PUB LM
[011409
YALA
FAULT FINDING OPTIMUS 50/65/80
The range of the control signal is + 3.7 V + 0.2 V for the on condition and + 1.2 V + 0.2 V for the off condition at the
specified measuring point against generator ground (see drawings of principals and PCB layout).
The range of the driver signal (gate against emitter) is — 14.2 V + 0.3 V for the off condition and + 13.5 V + 0.3 V for
the on condition.
« Select menu "FU kV/ Faultfind/ Functional Test/ Test Converter” at the service PC.
The question [power supply mains — E disconnected ?:] will come up.
Answer with ‘yes’ (type Return twice) and transmit with [F2].
If the test takes longer than 10 minutes it may happen, that the test will be denied by the kV. control. This happens if the
DC voltage = E-value increases 5 V (the DC capacitors are slowly charged by the + 15 V ofthe drivers). Then short-circuit
the DC at collector C1 and emitter E2.
Do not establish a constant short-circuit to avoid a big problem after the testi
The test itself is short. The pulse time is 2.5 sec long, but the PC screen says [completed] after 5 seconds.
kV_control sends pulses for 5 seconds, but the hardware timer on the kV_control inhibits more pulses after 2.5 ms. Within
“this time the actual kV have to be on the nominal value.
176H95 | ;
175195
Tube data must be loaded as a data set from floppy disk via PC and central unit CU to into function unit mA.
The procedures described below cannot be carried out before the complete data set for the tube housing assembly is
present in central unit CU.
Before the tube adaptation can be started, tube conditioning must be implemented.
With the present generator release the conditioning must be implemented manually.
To measure this total value an exposure must be released with 40 kV without filament current. The current measured
is the correction value for all standard exposures (4 mA, measuring circuit current depending on the kV).
As opposed to the standby filament current value of the predecessor versions of the generator, the standby filament
current value of the OPTIMUS generator is not fixed.
It is determined for each focus individually. A 40 kV exposure must be released with the focus to be measured while all
other foci are switched off.
The filament current must be changed until an emission current of 100 pA is obtained.
The associated filament current value is the individual standby filament current (1% to be substracted so that the
fluoroscopic current of any of the other foci is not affected).
The adaptation program takes place fully automatically.
Based on 120 single exposures for each focus a data field is created in the CMOS of function unit mA. The adjustments
for all other exposures are interpolated from this data field during operation.
During the adaptation procedure any limit values such as maximum filament current, maximum kV, maximum tube load,
maximum output, current of the generator etc. are taken into account.
Boost adaptation
Boost time determination (positive boosting).
With the predecessor versions of the generator, a calculated boost current was added to the exposure filament current
fora fixed time of 400 ms.
With OPTIMUS generators the boost current is also fixed but with a variable time.
The amount of the boost current is the sum of the maximum filament current (of the respective filament) plus 2000 mA.
To determine the time values an exposure must be started in the kV isowatt point (determined from tube and generator
parameters).
As soon as the 100% kV value is reached, the maximum filament current plus 2000 mA is adjusted by function unit mA.
The emission current is measured every 2 ms until the maximum tube current or the maximum possible generator current
is reached.
In case this procedure takes too long (warming up of the tube), the measurement is continued with a second exposure
after a sufficient period of time has passed.
The measurement starts again at the value obtained last.
An innovation of the OPTIMUS generator is the determination of the negative boosting (blanking of the filament
current).
The measurement is started with the same kV isowatt exposure which is used for the determination of the positive boost
time but with maximum filament current.
As soon as the 100% kV value is reached, the maximum filament current of the filament circuit is abruptly reduced to
500 mA.
Every 2 ms the emission current is measured until a value of 100 uA is obtained.
The values for the blanking times are required for techniques such as, for instance, cine.
A filament current value of 500 m must not be exceeded for otherwise the output to supply a gridswitch box (which might
be present) is too low.
The following procedure takes place after the generator has been switched on:
Function unit mA initializes itself and afterwards establishes connection with central unit CU via CAN.
For 3 seconds all foci are boosted with their respective specified maximum filament current. Then blanking of the filament
current (500 mA) takes place for a variable period of time (derived from negative boost adaptation) to bring the filament
current to the STANDBY value.
The change of the filament current value upon a change of the focus which was the usual routine for the predecessor
versions of the OPTIMUS generator does no longer take place — all STANDBY values remain constant.
During operation the following procedure takes place after the release of PREP:
— The filament current is raised from the individual STANDBY filament current which depends on the focus to the boost
current.
The switch-on time of the boost current results from the difference between STANDBY and intermediate filament current.
— The intermediate filament current is a calculated value. It is calculated in such a way that the filament current and thus
the filament temperature is brought to exposure level when the boost current is switched off for 50 ms directly at the
end of the preparation phase (RQ_SN_X/ already active) and/or directly before high voltage is switched on (RQ_SN_X/
active when the patient is quiet).
— Atthe end of exposure the filament current is reduced to the minimum value of 500 mA (negative boosting) for a short
time.
At the same time the temperature of the filament abruptly drops to a level which corresponds to the level of the
intermediate filament current.
Afterwards heating takes place with the intermediate filament current.
Now the tube would be ready for the next exposure with the same preparation.
~ In case preparation is released, negative boosting takes place until heating can go on with the STANDBY filament
current.
All the intelligent assemblies/pc boards communicate via the CAN bus. There they are connected in parallel to the two
lines CAN_L (low) and CAN_H (high).
Symptoms of errors:
— The generator is inoperable.
The red LED of one or more of the assemblies/pc boards is flashing.
Parameter settings on the control desk are accepted and displayed with a considerable delay.
in the error memory there are several entries which in the code begin with 00C (apart from 00CJ) or the error description
contains a reference to signal conflicts.
Error localization:
— Entries in the error memory clearly draw attention to the fact that the assembly and pcb are not communicating properly —
or not at all.
— Control measurement of CAN levels with an oscilloscope during data transmission and in the quiescent state.
Data transmission is triggered by pressing any desk button.
If the levels are outside the tolerance or are not symmetrical, the CAN driver of an assembly/pcb is faulty.
Since all the users are connected to the bus in parallel, the troublemaker can only be found by disconnecting one user
after another.
Disconnection may only take place with the generator switched off.
(= GI ——| CAN_H
LL
RE tt +++ AAA 7
difference
greater than 1.5V
during CAN traffic
A
=2,5V
=2,5V
Fr ed ye pe p ера
= V ty NP PE
一 一 一 一 一一 一 一 一 一 一 站
一 一 二
一 一 二
一 一 一 一 一 一 一 一 一 一 一 一| ground CAN_L
197H96
On the control desk an incorrect exposure is indicated if an exposure cannot be terminated according to the parameters
set. Frequent causes of underexposure are the following:
— Tomography time of the unit does not coincide with the exposure time of the generator. Permissible tolerance: £ 10%
— Measuring chamber incorrectly programmed, not connected or faulty.
Check the following: RGDV programming
Programming of AMPLIMAT sensitivity
Programming of EZ150 Basic Interface (Gain, 15 V/40 V supply)
Programming of screen-film combination (Data Sets 1...5)
— The APR selected is not matched to the technique used or the screen-film combination.
Check the following: — APR programming
м
The standard APRs supplied have parameters which are generally matched to a 400—type screen-film combination. If
the standard APRs are used, the exposure parameters will have to be changed according to the speed of the screen-film
combination actually used.
This also applies if an automatic technique is programmed as the preferred technique. In automatic techniques the mAs
and ms-parameters are used for Fault Exposure Detection.
AEC/AECF limits:
— Backup time AEC: Exposure time based on 10 times the mAs of the respective manual
technique (kV—mAs). 4 s after overriding.
— Backup time AECF: 10 times the exposure time of the respective manual technique (KV-mAs).
TDC limits:
— Fault Exposure Detection: < 10...4% dose for 10 times the sample time
10 x sample time
Dose minimum = x 40% nominal dose
exposure time (corr.)
In the event of a fault the exposure is aborted after approx. 11 times Sample Time. If the Fault Exposure Detection fails
to respond in the event of a fault, shutdown takes place after reaching the backup time or the max. mAs product.
The switch—off voltage should be at least 1.2 V to guarantee good TDC regulation. Program the higher gain factor on
EZ150 BASIC INTERFACE (= 4512 108 05964) if necessary.
Programming possibilities:
Menu “Faultfind/ Logging Table/ X-Ray Log/ Dose Rate Control Logging/ ...
../ Read Actual Status”: Technique and parameters of the last exposure
.../ AEC/ AEC Calculation”: Data of the selected APR with AEC or AECF
.../ AEC/ AEC Trace”: Control values of the last AEC exposure
../ TDC/ TDC Calculation’: Data of the selected APR with TDC
../ TDC/ TDC Trace”: Control values of the last TDC exposure
Adjustment possibilities:
CV1_GND_OL converter power part 1 ground overload (generator basis = 4512 104 70203/70602)
EZ130X1:A7-EZX24:20-EQ100X1:20
CV2_GND_OL converter power part 2 ground overload (generator basis = 4512 104 70203/70602)
EZ130X1:A28-EZX34:20-E2Q100X1:20
EX ON exposure on
EWA100X2:A9-EWAX14:7—
part of: exon old world
FD
C CH2 central field measuring chamber 2
EZ150X1:A4-EZX22:12—
15V, Ri= 220 Q
FD
L CH3 left field measuring chamber 3
EZ150X1:C9-EZX31:11-
15V, Rj=220Q
GND
MN ON mains on
C300X1:6-EZX46:6-EZX47:2-EN100X1:2-EZX44:14-
PO_12V + 12 V supply
EN100X1:1-EZX47:1-EZX46:7-C300X1:7ー
PO_26V + 26 V supply
EZ102X2:DBZ28-EZ119X2:AC14-EZ130X2:AC14-EZ139X2:AC14
—EZ150X2:AC14-EZX1:5-EZX2:3-EZX3:9-EZX11:1-EZX17:1-EZX18:1-EWAX1:4-
-—EWAX2:4-EWAX3:4-EWAX4:4-EWAX41:1-EWAX23:9-EWAX24:5-EWA100X2:A14-EWA100X2:C14=
-EQ100X2:1-E20100X2:1--
RD PR X ready preparing for X-ray or not ready preparing for X-ray (low active)
NR_PR_X/ EZ139X1:A3-EZX23:3-EZX45:4-EZX46:4-C300X1:4- -EWAX51:4-EWAX52:4-EWA100X2:A24-
driven by CU
measuring point: EZX83
part of: signal bus
SW SF CF 2 switch side field to central field bucky measuring chamber 2 (EWA or EWB) or 4 (EWB)
EWAX12:1-EWA100X1:A20-
TB
2 RT tube 2 selection check
EWGX11:3-EZX11:3-EZ130X1:A10
0V/5V, low active
TO PG 1 tomo program 1
EWA100X1:A13-EWAX22:1—
TO PG 2 tomo program 2
EWA100X1:C13-EWAX22:2—
TO PG 3 tomo program 3
EWA100X1:A14-EWAX22:3—
TO PG 4 tomo program 4
EWA100X1:C14-EWAX22:4-
το ρα 6 tomo program 6
EWA100X1:C15-EWAX22:6—
TO PG 7 tomo program 7
EWA100X1:A16-EWAX22:7—
TO PG 8 tomo program 8
EWA100X1:C16-EWAX22:8-
X_ACT_S/ X-Ray active signal, kV > 75% nominal value or ‘fluoroscopy technique’ high tension on
EZ119X2:A8-EZ130X2:A8-EZ139X2:A8-EZ150X2:A8-EZX52:5-
OV/5V
measuring point EZX77
HTON (high tension on) or FLON (fluoroscopy high tension on) signal
part of: XS/XRG bus, controls X_ACT/ status
3Z-1
Aqua UG
XZ060 8OL CLS? UO 6X
O
с. а
SS
더 >
c ©
+ Q)
Cn
ay
Us
SHH0JS Cx T © 8x
wop(づ 4X
snnis QLH 5140}; СОН $1; СН 5141$ (60 내 © 0084 ASL- İCİ SH
AS* İÇİ 9H
008X ASL+ | €H
10500 (@) [9 a SX Agee
oul
cH
T Es LX IG
25
/ (3% (Эх anna 3504340
(95.0)
| pa | 430 [λ/νς ζ]
0380 ἤ a 95 JW so(Jx T Ox
40 AO JW (e VI 1H MGE,
030013 νο GX
I NY LH AV Ci
IS
1Η ΑΝ Ce
OPTIMUS
COLZI
LE900n
1450/"4DS 72-50-56 yy/ZY
OPTIMUS 50/65/80 REPLACEMENT
REPLACEMENT
TEXT
1. H.V.generator ...............................
1 eee 4-1
Operating panel rei Ko 4-1
3. Printed-circuitboards ино иена ensee 4—2
1. H.V. generator
The H.V. generator is a traceable item and is therefore labelled as follows:
— type number
~ serial number .
combined label
— manufacturer
— HHS certification
— code number
H.V. generators have a serial number which has the following meaning:
Example:
96 07 005
Meaning:
96 = year of manufacture, e.g. 1996
01 = power class, e.g. 50 kW, 1 tube
005 = consecutive number
Power classes:
01 = 50 kW, 1 tube
02 = 50 kW, 2 tubes
03 = 65/80 kW, 1 tube
04 = 65/80 kW, 2 tubes
With the H.V. generator for replacement a separate label will be delivered. This must be affixed to the label bracket
on the top left corner of the generator cabinet. See drawing 2Z-1 Labelling.
The new type number, code number and serial number must be entered on the master card for the generator.
2. Operating panel
The new type number, code number and serial number must be entered on the master card for the generator.
Please, send a copy of the corrected master card to the address mentioned above.
OPTIMUS (96.2) E 4 一1
OPTIMUS_50/85/80_4_962E © 1996 Philips Medizin Systeme
AIL RIGHTS RESERVED
REPLACEMENT OPTIMUS 50/65/80
3. Printed-circuit boards (
HW SW programming Adaptation
B R
PC programming via XRGScope of tube emarks
— To attend to:
X4, X10, X42, X44, X52
EN 100 Power ON
circuit
see Z1-15.1
— address
EWA Back panel W1...W3
— ground
W11...W13
R226 EE A“ | W20D ov
co LS 4 E W3 -sv
may be adjusted q x a“
for SOkW version a supply voltage
with resistor in ο ο for chamber
parallel で amplifier
X1 x1 x1 X1
F3 038 더
w2
3.15A 5 =:
| E wi b
T chamber contrat type
ON = transistor driver [V1...V59 existing)
F1 F2 z W
5 121 Σ G4
AD |w3 GFF = CMOS driver (SMO)
321 = 2 QD We
Law: | EE -_ lan |ws
[mw2 |
321
S
E 03 o o о
gain E개: o
porn,
SV controlled switch aver
y2 , 引 0 2 Em | 2
14
ReL21 ReL22
D4 4512 113 18025 18026
D5 4512 113 18035 18036
PCB programming
central rack EZ
OPTIMUS (b/96.2) 57-1
O Philips Medizin Sys teme
Operating desk € High speed rotor control Y 9890 000 02211 High speed rotor contrat Y 9890 000 02212
C 300 ου Y 100 RocoHs Y 100 Roco HS
4512 108 0824x
. 4512 108 0826x ЕЕ 4512 108 0870x
è
xl] XAT X12 X13 X14
| x3 x4 X2 X1 x2
02 Ly 02
С 」
ana, coo 1ina [Rotor contrat
1 5 02 ji
和 wi a i ㆍ yz м? |3 A
χε 1
>x
[ео
a
1
x
F
和 w2 a i i w3 W3 iğ i
E
x1
|
7 |
02 = 4512 113 2051) (4512 114 20514) I
|
I
I
I
X6
i
o 더 1
| ——
X44 FO=0+0=0+
1 4
OFF Phe X1 X43 X43 X44 X45 X42 X46 X47
3
D2 = 4512 113 20401 (4512 114 20401 02 = 4512 113 22301 14512 114 22301)
compatible: 4512 113 22301 (4512 114 22301
X2
Schr.
D10
96-06-05
PCB programming
Options
OPTIMUS (b/96.1) 5Z-2
© Philips Medizin Systeme
OPTIMUS 50/65/80 ADJUSTMENTS
ADJUSTMENTS
TEXT
SID, diaphragm aperture and type of filters are supplied by the diagnostic unit, where they are also adjusted.
In the generator default values are given for the specific yield of a tube and filter correction.
These default values can be found as reference files on floppy disk in order to recreate the original settings if need
be.
Reference files:
ref yiel.tdl specific yield of tube
ref 2al.tdl filter 2 mm Al
ref_01cu.tdl filter 1 mm Al + 0.1 mm Cu
ref. O2cu.tdl filter 1 mm Al + 0.2 mm Cu
The specific yield curve relates to tungsten anodes and 2.5 mm primary filters.
Display on the desk is in: [cGycm?].
e Determine area dose at the following kV settings and compare it with the respective value displayed on the desk,
Example:
— displayed area exposure product: 8.8 cGyem?
— measured dose: 890 uGy = 0.089 cGy
— calculated area exposure product: measured dose x exposed area
= 0.089 cGy x 100 cm?
8.9 cGycm?
It
8.9—8.8
— difference in %: x 100
= 1.12 %
8.9
e If there are any deviations of over 5% it is recommended that the yield curve be corrected in accordance with
the procedure described in 1.2.
Principle:
For each kV specified a dose measurement is taken under the same conditions. If the distance between the focus
and the measuring cell deviates from 1 m, all the dose values must be corrected with the square of distance (unit
of measurement is [m]). Dividing the dose values by the mAs product set gives the respective current yield.
Procedure:
e Measure dose at each kV checkpoint and use it to calculate specific yield.
The values determined must be greater at higher kVs settings and produce a characteristic with a slight curve
on the graph. If considerable fluctuations are detected, the measurements must be repeated at the points in
question.
Range: 0.00 ... 400.00 uGy/mAs
The values can only be stored in the generator if they are within the range specified and rise uniformly with kV.
{
Specific yield
default
yield [uGy/mAs] | 11.0 | 20.8 | 32.2 | 45.7 | 58.0 | 73.5 | 88.7 | 106.1 | 124.4 | 143.6 | 163.1 | 181.8
125.0 124.4
100.0 106.1
88.7
75.0
MT
50.0 58.0
25.0 Ta
32.2
20.8
11.0
0.0 - [KV]
40 50 60 70 80 90 100 110 120 130 140 150
200H96
e Correct the default values of the specific yield for all the kV checkpoints using the menu "Adjust/ Area Exposure
Product/ Specific Yield of Tube 1...3” with the factor determined and save with <Transmit>.
e Save the specific yield curve with the SAVE function of XRGSCOPE (F3 key) on the backup disk.
Recommended file name: act yiel.tdl
10 mAs
0.15
collimation 10 x 10 cm atthe height of the measuring cell
no filter
Principle:
At otherwise identical settings the dose is determined for the kV values specified with and without filter. The ratio
of dose values with/without filter produces the respective current correction factor.
Procedure:
Accept measured dose values (not the corrected ones!) for the respective kV checkpoints from yield
measurement or measure them again if any changes have been made to the test-setup or settings.
Move the filter to be checked into the radiation beam.
Filter correction — 2 mm Al
MEX AAA
default factor 0.25 0.37 0.5 0.65
=
new factor
0.9
0.8
0.7
0.66
0.6
0.56
0.5
0.47
0.4
0.39
0.3
0.1:
0.0 [kV]
40 50 60 70 80 90 100 110 120 130 140
301H96
e Read out the default values of the filter tables for each kV checkpoint, correct with the factor determined and,
write back into the generator with <Transmit>. '
Menu ”Adjust/ Area Exposure Product/ Add Filter Correction Tables/...
… 2 mm AL”
… 1 mm AL+0.1mm CU”
… 1 mm AL+0.2mm CU”
» Save the specific correction tables with the SAVE function of XRGSCOPE (F3 key) on the backup disk.
Recommended file names:
act_2al.tdl filter 2 mm Al
act_01cu.tdl — filter 1 mm Al + 0.1 mm Cu
act_O2cu.idl filter 1 mm Al + 0,2 mm Cu
ACCEPTANCE
1 Preface ーー 7-1
5 ExposureCounter .........................................
a 7-3
1. Preface
The national rules for accepting a X-ray system are very different. Therefore in the following is given an example
for checking the generator in the U.S.A.
OPTIMUS generators are factory—calibrated and checked for compliance with the parameter readout tolerances
as Stated in the relevant Operator’s Manuals.
Provided these generators are installed and set fo work in accordance with the Installation Manuals only the
following limited field compliance testing is required.
2. Test equipment
~ Keithley voltage divider model No. 35080 with filter packs 32867C, 5C, 9C or equivalent.
— Oscilloscope (storage)
— Digital mA, mAs meter,
Notes
Do not start test until generator has been switched on for at least one hour.
Direct (invasive) kVp measurements on OPTIMUS generators with HV divider tanks normally available to the field
service organization are not permitted.
Measurements of kV using instruments other than the Keithley instrumentmay lead to larger measuring tolerances.
The causes are to be found in the specific frequency response and transient response of each test instrument.
3. Setup
+ Switch off generator and also switch off main disconnect breaker to system.
+ Connect digital mA meter as per instructions in the relevant Service Manual.
e Set up the Keithley voltage divider complete with the appropriate filter as per Keithley Instructions Manual
No. 3294 OIM.
+ Connect the oscilloscope to the Keithley divider.
Note
Make sure that the oscilloscope has been calibrated with the aid of the Keithley divider as described in the Keithley
Instructions Manual before starting any testing (par. 3.6. Internal calibration).
e Calculate rejection limits based on the exposure parameter "Specification Limits” shown in the table below.
The "Specification Limits” are based on the actual tolerances as listed in the generator Operator’s Manuals.
These "Specification Limits” must be restricted to include the actual measuring instrument error. See also
section 6, par. 3.1. of "Comprehensive Compliance Testing Manual” No. 4535 800 2034. regarding how to
calculate rejection limits.
4, Test
+ Switch the system on.
* Measure the mains voltage on ENF1.
Reference voltage: Mains voltage programmed + 10%
LILA
L2-L3 .........
technique 80 mAs +3% +0.5mAs | 77.1... 82.9mAs | ........... MAS | ............. mAs
Owing to an offset current in the measuring circuit of the HV generator the measured values for mA/mAs must be
adjusted using the following formulas:
U [kV]
lcorrected [MA] = Imeasured [MA] — Offset = 0.2... 0.75 mA
Reale. MQ]
5. Exposure Counter
Before handing over the generator
to the customer, read the Exposure Counter via the "Accept/inspect/ Exposure
Counter” menu and record the figure in the table.
The count cannot be changed, so it is recommended that also whenever the tube or the CU PCB is being replaced,
or whenever the entire CU PCB programming is being deleted, the count be recorded in the system logbook and/or
in the following table.
Implemented
FCO No. Remarks
Date Signature
Implemented
FCO No. Remarks
Date Signature
(95.0) E Checklist
© 1995 Philips Medizin Systeme Checklist
ALL RIGHTS RESERVED
OPTIMUS 50/65/80 Section Z
SERVICE MANUAL
Drawings
Schematic diagrams
Basis
mA control Z1-4.1
H.V. generator | Z1-4.2
Options
| 나
| |
|
|
qu
I | | ]
Service PE — — 4x5 | I I I "mM "
I I ㅣ |
EZ139 EZ02 62130 . EZ119 62150
~ 만 I Low voltage supply | kV Control | mA Control i Basic interface
OL nen AE
DRC rot ο χι I I | I
I | | i
X1/X2 xx2 xi x2 xt x2 xt χι xx x1/x2 XX2
7 Fu ζ 2 I ノノ External XS-bus
ググ
eo の πε 2 za τή . _ | | İ Zu ~"
System progranii pu Y 70 Y
ctrl Ena 42% dl LS I XRE b | | O O m a Ἴ
| | | to =
[ ジンクッ
Y termal x
の> || EZx6OEzx20Ezx3OEZx7 の
ES Ezxt | EZX34 EZX24 EZX35 EZXİZ | bens のM
W . units
0 | + da L | αμα EY
Rotor contra Function
RC unit
| (tow or high speed) a
| woe u
7
release
swifch
2hooxa rerC 200x1 자 || 1002
> | > i K1 I
o Emergency OFF Ex 100x1 3..400...480V T x1100 a
5 i
$ на [Surge arrester] ㆍ REN I
SE 50/60Hz me ee EZXS > | re a Block dia ram
三 5 1 Г $ 1 Г 5 1
' 5 E * E E E E
로 일 이 | pi: oS te Ly きき る3 1 Lo gf:
dii
+: ses Бе а δὲν dåd 33
ра18 BEİ
SB EE
İİİ
きる き 3 る
FE)
3 8 |
& fia
ŘE
238 18δ ὃ ἅ ἓ 2222
ー
11111) trip tiro 7217 17771117 1717 1127 1117
E
'
|
[
102 102 102
Universal Universal
|
Universal
|
1/0 interface 1/0 interface 1/0 interface
_
SG
EZX45 자 2
X51 CAN/XS-Bus 7
>
X52a
EZX45 ~ 2
NN
W Option rack
AV/A3 96-01-24 Schr.
Block diagram
Expansions
u00615
ring core i . | ー ー хз
with é winding | X33 VO 8 8—J9 K2
|| >
Ya NA 一 一 A し V3
| | vg る| R6 5 iz
| 5 X2:1 X2:2 X49 X45 X14
V5
I At K2 Ki =
MEX | ὅ ὁ LE < LLI LA | RC1
ti || ke-
1 13 E Lo
C2 v7
ο
=,1004 AVS Nb
Stand ts)
12© |t 2 3
OPTIMUS
3~400V すき 13 | |
50, 60Hz C $1 v 100
max. 5A N | -V Power on
9 Ť X46 x 47 X1:2 xt circuit
EN
|
MN EM OF
| 站 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一一 ニーーーー
に ーーー、」 İma ne mn le mn mn mm me mme me me nue mu mu me = mn me = me me mue me = 一 一 小 一 一-一 一 一一
Y Y Y у Y Y Y Y
ие в м 3 м
ZA AAA \ 一/ MVL 一一 一 一 一
96-06-06
3 e 一 上 一 | C
> (option)
_Desk Power supply
OPTIMUS (b/96.1) 21-2.1
© Phitips Medizin Systeme
Wall junction box |
|
MEX | F125A Remarks: Mnemonic's
uo Гы olar
| >
Г = 3 = “ E | ーー 1) A filter may AC OV XG mains supply OV X-ray generator
be inserted here. AC 230V L1 mains supply 230V AC phase 1
は | 2opel > 10044 10004 EZX13:3 AC 230V 12 mains supply 230V AC phase 2
Mains supply 1 23 | 2% Be
3 ~ 400V,,.480V—> | | 가짜 kV Power unit 一 [ um | © GND 12v ground +12V supply
L3
30, 60Hz oO ole R3 > É EN/EZQ L MN EM OF mains power emergency off
| | 一 33 | 34 MNON/ mains on
N
|| | 41
나 42 PO 12V +12V supply
1 Connected according 가 PW ON NG DC supply relay power on -
PE F to mains voltage. Discharging cincuit κ PW ON PO DC supply relay power on +
【
| っ
I
|
| X3:1
I x33 ΤΊ V al 873 K2 |
| A Yk k Sw
| | νο S|R6 5 5 12
I X3:2 > X2:1 X2:2 X49 ×15 xt
| V5 tre JIX1205
o MEX || ez [É2 JE 4 96 = SL Thermo fuse
= o ο ο LA
V2 İÇİ TX1206 1
= τηE I 1 B b INv7
=
JF A1 K2 A1 K1 RC1
S E |
| =100 ya 本 ve =hs Е кз Kİ,| АLa АLA
Stand (5) 13
cn [TZS 2
| N No
3-400V 0 —> i >
50, 60HZ “7 i 2 a3 .
max. SA N |
51 7 100
| Y v1 Power on
| X46 x 47 X42 xi circuit
| MN EM OF EN
ーーーーーーーーーーーーーーーーーーーーーー
ゴーーーゴーーー|-ーーーーー キーーーーーーーーーーーーーーーー トーー ニー オーーー
I EZX43:1
| LAE 230V 너 のEzx476 の EZx47.7 OEZX47:2 Gal EZX47:9 > só 40
| Lac ov x6 EZX13:2
A EZX44:14 EZX44:6
| lic 230v [2 52X33 Emergency EZX4:1
M
|| 1 OFF EZX4:2;
RESET SW/ РМ №№ SIPHID PW ON PO
| I 그
E | м 42 23 | EZX44:1C 2 EZX151C 20 Serei QEZx 优 EZX4e6 >EZX46:7
| ~ ソノ
| \ーーン V
q> | F3SA Mm TEМ2 İm I EZ139 EZ130
° | GNDİ2V ММО MNOM/ PO 12v Central unit kV Control
LA |
28 | v. | v |
2 ив в N L3 N ,
ma .
Lin
NS À EYA WG 、
EWN Adaptation EZ119 E7102
gs Rotor control Tube extension Second 2h NA ZION c mA | Low Ka K1 Power supply
(option) low supply
voltage (option) ontrol voltage
supply A Z with mains transformer
ains Transtorme
OPTIMUS (b/96.1 (option) ; | ;
Ll ll C 71-2.2
© Phitips Medizin Systeme OFF ON Desk
PO 26V SW PO 26V 1
EZX 131
Power supply
EN
EZX13:2 u E
. 8 a
B EB 5 $ Е
heat sink s 8 Я 5 5
5 <
Е
>
= u
§ a
8 時
A e
= a
62119 EZ130 EZ139 EZ150 g 二 の =
a č U =
10 日 8 E 5 E
3 5
=
=
= <2 Β =2 5
= É2
E —
3g de 2 2 5 a 5 = 5 p
be 230V LA « 196V = © 同 + en a 巨 = =
a 5 a < a e E g
= = E 5
= £ a 3 a 与 È o = S
E 들 a a a & a i 5 2 5
1
8 a
8 a
F
2
F
>= 3< 8
=
=고 5
D
8
o
5
©
EZX21
EZX22
EZX31
. EZX17 EZX19 ΕΖΧ32 EZX や 1
時SNS
26vV HB EZX1 EZX2 EZX3 EZX11 EZX18 EZX20 EZX41 EZX35 EZX44 EZX46 EZX51
E > 30kHz E
heat sink
PO 26V SW
3= LSV SNS В
に 60103
©
C42
EX SNS
Regulator
SV SNS 개 driver
{PWM}
R3
8 à & È
e 를 上 X2:A 亿
a
>
x2:0BZ26
Mnemonics
=
ov oc
a s5V SNS SV sense
Ev NS 15V sense
로a
+26V SNS 26V sense
9 AC OV XG imains supply OV X-ray generator
LA AC 230V 11 mains supply 230V AC phase 1
mil NG 15V -16V supply
<m
No PO 5V «SV supply
LS PD SV +15V supply
roe
PO 26V1
PO 26V SW
srt
+26V supply optional
+26V supply switthed
Low V voltage9 p
power
W
5 Up р
lу
POWERFAI L/ power failure reset 、
OPTIMUS
si se
(a/95.0) mak test points for
Z1-2.3
© Philips Medizin Systeme TP DRV2 internal signals
TP SON
TP VOD
voc
+480...750V
Ti
galvanically decoupled
= E power supply řar
~ drivers 1 ond 2
E PO CVI
X6 ОКН | ΙΕΡΟ (νὰ)
|
[10 Ak г} jcriver 1 ks04..507 R506...508
т .
τος
+ CV GND
x201 (Cv2 GNO} EZ130
Enitter 2 V502 V503 kV contrat
] №1506 > mafie [C501
X8 1220
med
X+27
H-3 D X4:16 Î ~
iyi lord
2-3) R29 X300
トーーーベ ーー Gute 3 xa e CV EN)
KCV2 EMI
1-3/
-3/ > x13 ピコ Xİ —— ia
12-34) R509...511 KCV2 İLA
3x39R
X301 X42 Cvt NO OL
Emitter3 ICV2 END OU
x9 R504 R50
№505 È AE ENG CV
ENGCV2) J
tee de X117 — [driver 4 V34 图 R24 | C19
2-4) R31 X400 I 4.7n
— ——» inte 4
kV power E Q
11-47 >
xt R32
1 }
< 4512 108 08622
прок
L 12-87 | 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一一
X10 X401
Emitter 4 ㅣ 65/80kW connection | 50kW cannectian
X24
11es pes EO r KV power unit fir Ps SO KW, 『 一 πι Measure
ー dre a
X31Lrd mounted in the front of the generator. I | X100 NX 100
上 上 上 Ep kW, E2Q | |
mounted on the redr side o
Low
EZ102
voltage Supply
下 !
| 2
Pp57 fesi blower U아이 . any information
Il in brackets is vi valid for E20 .
|
| .
1 DU Atos
¡SS
も>
heat sink
x53 Sion VIVA, o
Li V5
u a 3 x |
M 」 | 480... 750V |
nemonics ob L “
Ost.
ENG CV1 E value converter power part 1 DC supply negative Ev1ID/ converter power part 1 identification 113 IGBT3 power part 1
3. ” 00050. и 12 — 12 = 2 в т ES \ |
60Hz κ | |
ENGCV2 ・ ・ ㆍ . 2. + * Cv2 10/7 a . “2 ㆍ 1137 k
EPO vi
ΑΊ/Α3 96-05-06
E value converter pawer part i DC supply positive CV10L/ converter power part
4 overload MA IGBTé power portt
ve É ” CV2 O ” +2 ㆍ πω x1103 |
Cv1 EN/ converter power part © enable CVITM converter power nart 1 temperature 121 lb 000) 1 イー |
Discharging af
WE o ти + >2 < 24 (GBT1 power part2 OC circuit after
(VA GND © converter power part 4 ground 14 122 XY X2] Χ N21 22) X23 Х12| X16
switch-off of
vb0274
TI
galvanicalty decoupled
= E power supply Far
~ drivers t and 2
]
E PO CV1
X6 Soz | | - E PO CV21
|
r ms > Xt14 ーー 7 ¡river 4
I ・
m R506...508
[504.507 |.
3439/7
12-
x
12 イ TY
> Gate 1
vi Da
X44 4
(1-47 > -
{12-17} R1 ICV2 OAs}
一 一和 CV1 OND
x01 R2 {CV2 GND}
oe
Emitter 1 V500 E» (Vi DB/
1ČV2 DBA
×102 NN500C < dI №501
24-
V34 [SÍR24 == 019
- 90
O Gate é I 100
kV power EQ
2 4512 108 09623] (E20)
Woes Até iS
し 112-441 개제 => => 128 12 108 A AC DG GE IU es ASINO SANA O NC RODE S| P
]
X401
È Emitter 4 .
| 65/80kW connection | 50kW cannection
| .
X402
< } X1001 x 1004 +
|
lez — — |
| EG
二
| H.V. generator
La
u [1
sera 5 内
> 1 2 |
|
Mnemonics
Power supply x1102
Ost,
WED «2
LE
we TH er. |
1227 IGBT2
(GBT2 power power p part 2
263 pare Le
CVA GND OL converter power part 1 ground overload Hi 23 der +2
X |Ť 2
CV2GNDOL aa ни IGBT1 power part1 2 37 16873 power par
, 。
Eh xi | || 42 kV power unit
1 2, 16572 power part 1 Е ん IGAT4 power part 2 R1 1831 IR2 ]
kz EMC filter 200 | 7132
OPTIMUS (c/96.0) |
© Philips Medizin Systeme Power supply
Not connected
for grounded
3. 400 … 480 ——2> diz]
delta mains 50, 60Hz
ur E20
Mnemonics
kV
power unit AvHT high tension actual value
EQ, E20 AV HT AN high tension actual value anode side
AV CT CA high tension actual value cathade side
CAN H generatar CAN high active
Start CANL generator CAN Low active
Selftast CM TH common for thermal sensor
CM TH SW © common for tube housing switch
CAN-announcement CTRLXC/ control X-ray command
Reridy CVIEN converter power part 1 enable
Watch dog error CV2 EN a ㅋㅋ 2 ㆍ
CV1 GND Converter power part 1 ground
Cv2 END a ” “2 ㆍ
Evi D converter power part 1 identification
Power supply v2 0 2
Ne so OL converter power part 1 ground averload
. 2 a a ㆍ . ㆍ ㆍ
Tube extension { CVIOL converter power port 1 overload
W2 0 a чо.
oe ーー CV1 TM converter power part 1 temperature
(vz TM ㆍ "42 ㆍ
T
ENG CV1 E value converter power part 1 DC suppty negative
o I ENG CV2 .. ㆍ ㆍ a Qe ㆍ .
E POCVI E value converter power part 1 DC supply positive
-- — EPOCV2 - + ・ =. 2- » ・
EN X C/ enable X-ray command
.
kV power unit ' EDGE HT AN high tension anode side actual value
EQ 7 НТ AN GND high tension anode side ground
EN100 K2 HTCA high tension cathode side actual value
HT CA GNO high tension cathode side ground
— — E Power ON
E . contactor 2 power part 2
0 О > - power pari
E kV brain
id, 16811 power part 1
EB
tay IGBT2 power port 4
== 1-3 IGBT3 power power part
L
5 43 par 4
=> Tube 1
r extension 114, IGBTS power part 1
+ Em
一 L — 3 2-1
BV 10871 power part 2
8
virtual ground
kV power unit
Ost./Schr.
E28 (option)
A1/A3 96-01-19
|
supply q x3 ーー X5, X6
Low e
va0267
| kV-control
OPTIMUS (a/96.0] 21-3.3
© Philips Medizin Systeme
Start
Selftest
ES output
mA brain EZL1
EZX1001 Grid switch of tube
from back panel
D18
4512 108 05984
D22 onwards.
RAM
Mnemanics
Use fuses with AC OV XG mains supply OY X-ray generator
ceramic housing AC 230V L2 mains supply 230Y AC phase 2
only ! CAN H generator CAN high
CANE generator CAN Law
CAN DF O
CAN DF 4
CTRL X CZ contral X-ray command
ENXC/ enable X-ray commund
FI TF11 filament transformer 1 line 1
FI TF12 » ㆍ 4.2
galvanic decoupling
STOP X C/ stop X-ray command
H.V. generator regulator/ TB CU FR NG tube current frequency negative
measuring circuit TB CU FR PO tube current frequency positive
PWM driver X ACT S/ X-ray active signal
supervision
г => Supervision
EZ119
EZX13:3
1
Power supply filter
EZX13:2 1
EZK122
|
EZX12:3
EZX 4.
= № supervision
EZX12:5
| 圖
96-06-05
v00323
AWA3
mA control
OPTIMUS (96.0) Z1-4.1
© Philips Medizin Systeme
>
K1
: = K1, KZ and H.V.plug 2 are present
Tube on only with 2-tube version.
K2 K2 tube 2
oto >
—ы——_
| R2 K1 fube 1
i Dre | 一 ゆめ
| 1 6 104
| C2 Tİ
65/80kw 50kw | E
connection connection | e 5 gls 人 FÅ P
기
二
È M (2. 22020
R22 Кая ER"18 L_ L2 xi
T TB CU FR PO Baye
———e mA control
Zv7 Eze ER25 Ave =023 U7f SER ven TE CU FR NG 나 EZ119
1001 «1008 = = R31 Г 7 9 7 1 | X21 | Γι EZX35:14
kV power unit г -E 1 23 İ İ
Ea | | | 0 1 | |
7002. pts mA/mAs measurements are to be 1
corrected by an offset value, V6,V8,V9N 3622 : 13 χω] HT AN EZX35:2 、
See section “Acceptance” | ” DD AR X xml. HT AN END A > (TY DN
$ rt | cor ld ddA dd |x23 一 ЕЮ Anses (7) tez
: R27 C11 less [em çiz 6251526224 external
| R12 P MET Por © o H.V, changeover
switch GWB
kV contrat
L& хи 1 HT CA ae m £7130
4003 1003 Tě は E ES НТ СА GND >.
kV power unit | | R R14 Lee и 1 1 |x22 -上 Ezx359
C28 L xd TP HT GND Es >
{ rom [1004 PA
op 『 RIS Te PET ТР
OPTIMUS
R29 AVS
ED ya xt) TP HT SG EZX35:12
25 »|
|| dei
р 一二 二 二 二二 和 100
| pe
| R1
1 ・
| 4 ta) Ki |
[$ т 1, №.
=
D
7 3
一 一一 圖|
EZXI2:1
EA FL TEA Fu | el A
solo T3 12 =K2 \ tube 2 加
回 , 52022 FLTF12 х52 X65 Es Er
EZX42:3 | LU x Хе T = | |
mA control 12 | Mnemonics
EZ19 EZX42:4 FI TF2 1 хе xal a 518
< Π 圖 1] É 月 TF11 filoment transformer
1 tine 1
pS EL TF2 2 X55 X62 4_8 BTF12 - o 1-2
P a | = κ А ТЕ? 1 - -2+1
EZX12:6 154 хе = FLTF2 2 。 。 2-2
HT AN high tension anade side actual value
HT AN GMD high tension unade side ground
HT CA high tension cathode side actuat value
XM:7 ~
Ost./Schr.
voltage9
Lowsupply 、
— BET< PO 15V supply md HT CA GND high tension cathode side ground
EZ102 EZX35:15 NG 15V XK: トー -tsy . NG ὃν «EV supply
PO SY BV supply
| EZ 102 70260. 7
- —
FX23
NS 84 m SW/
RESET > CANL AZ O EZX71 EZXS2
3
Fa2 > CANH X2:C3 。 52002 A CANA xs
¿sí e 13 1000 0 eg System interface i
τρ 内
X2Al 國
В x
DI [ :internal
LR ET 163 ROSNX/ > SIRLAR, AZE > b| | extensions
à EZX85 wa CTRL X/ ~ STP XC X2:A7 pm EZX75
みつ Cm x ACT? > EN X C/ X27 ュー GEZE
бо νο 038 XACT S/ X2:A8 。 СЕ
Signal bus< = JO y $
o o
Function key
lo i
ο
ος i
o 3
o 28) s D4
BA 3 Flash Mnemonics
マン 3 WI ROM 9 (Start μμ o :
~ or e Selftest CAN H generator . CAN high
it active
「 EZX42 EZX43 Flash ro [FR Дика [|@ 出 호 É| _CAN- annauncemen qui , generator CAN Low active ;
Le ο RX CAN 2 BX CAN 1 pyppucarian e
Reud X signal us X-ray request commani
$= 7 了
06 SCANL > EPROM 05 Nie CTRL x C7 control Xeray command
™ Flash OR BV OV dose rate (signal reference of image intensifier
System CAN j ο o AS
ο. HA,
yl S—SCANH
CAN GND ++ programmed
accord ROM
or status X6 OR BY SG signal of image
dase rate signal neferel intensifier
mage
9ο Og ~ > 047/05 y EPROM OR FQ NG dose rate signal pulses! negative
kg 92 ΜΑΗ. SCANPO > OR FO PG dose rate signal (pulses! positive
人 — DR TV NT dose rate of TY chain signal negative, fluoro
POWERFAL/ OR TV PT dose rate of TV chain signal positive, fluoro
51 OS BV OV dase (signal ramp! reference af image intensifier
» DS Bv SG dose signul ramp of image intensifier signal
. RESET KEY, | DS MC ον dase {signal ramp} reference of selected measuring chamber
| DS MC SG dase signnt ramp of selected measuring chamber
W2 DSR data set rendy.
A DTR data herniinat ready
I EN X/ signat bus enable X-ray
ㆍ Poe 013/025 ENXC/ enable X-ray command
| G4 ・ MN ON: mains on
' T W NG 15V -15V supply
| 1 interface contrat Y NR PR X/ nat ready preparing for X-ray [7139
i POV +V supply
' Basic interface Pas DS ME SG PO OV sOV supply
| EZ 150. baci DS MC ov PO SV EN sum
measuring chamber PO 12v +42V supply
PO 15V +15V supply
PG 26V +26V supply
EZX6 A DRC Interface POWERFAILY power faiyre
DRC usfrequency
pused 217 DS BY Ov AT Bus RO MN ON ready mains power en
217 DS Bv SG IMUX,ADC) RESET C/ reset Fu
input u ei 2108 OR BV OV RESET KEY CU reset
. - 2:18 DRBYS5 | RESET SW/ system reset
TV es 2:[19 DR TY NT RQ SN X/ request synchronizaHon
È RX CAN system CAN optional
ZAS 中 TFT RX CAN system CAN 1 optional
RX CAN2 system CAN 2 optional
RXD receive deta
は 。 SCAN GND system CAN bus ground
| 2620 00004 | Main SCANH system CAN high active
a computer S CANL system CAN tow active
| lo eS Lot used S CAN PO system CAN supply
29 RXD (386 SX} STOP X C/ stop X-ray command
a
。 이 ni
| Service PC «—» x5 [A η + 03 Tx transmit data
Schr.
BIOS
a
»
ua0634
=
Low voltage +
| power supply Central unit
. EZ 102
$1 | $2 | $3 | 54 | $51 56 | 57 | 58
PO Sv 05 s
RQ SN X/
46 bit RO PRX
micro controller NR PR X/
RESET/
olphonumeric display
NR PR X/ NR PR X/ EN x/
RO PR X RO PR X
| 80150
PO 12V
L XRG bus
EZX46
X
RD MN ON/
X2:1
X22
X18 | &v, 20kHz
Qutput specification
X2:3 MNOM |
0 2X 4Vere + 10% / 20kHz +20% X
1 AV, 20KHZ
trapezoidal,
X24
ゃ 2 x 2Vaf 4 10% / 20kHz +20% X2:5
Power supply
trapezoidal for
X2:6 alphanumeric display
PO 26V 0 +40... +44V X2:7
| 29, 20Η
EZX20 0 +14,25 ... +15.75V X2:8
Î 2V, 20kHz
L + +4.95... +5.25V X2:9
x49
X2:10
Ost./Schr.
X1:11
X2:11
EZX6 μα] Input specification Х2:12
X113
e +20... +30V X2:13
X2:4
xt7
C300
96-01-19
X2:17
x2:21 CPU
v00275
X2:22
A1/A3
C200 X2:23
x123
Remarks:
Power supply X2:26 “SV
Desk panel 4) Pull-up resistor and
Unks (300X3:1-6/3-7/5-9
from CPU 4512 108 08244 onwards,
Operating panelC
OPTIMUS (96.0)
© Philips Medizin Systeme
Z1-11.1
25 MAS
Ost.
A4 94-06-17
OPTIMUS Z1-11.2
il
X17
x2:03 Tlly ¥ 站Ii [|| у M RO/SRO tubes
069 OOM
BE 一 V12 ma7 1004 X43 Τί κ -- II, ll Ha,
[ “一 ШТ Г.
№ R44 = C15
Х2:01 =
十 = KZ
à 一 一一 表
——e 522
(50) .
Mnemonics
| Acceleratian , break AC CH Gcceleration power check
galvanic decounting fas0.08/t2s| Confina 也 = 0..60s 名 sean
‘eax che:
y X 2.6 PH SK Lob H Hel Phase check BRT break tube
END ground
OPTIMUS
Km
一一 X30) AU H H L Acceleration
g3 Lx0..25V
, H a 12,5... 45V vi
Basic interface AT _RCOM kro ” e EZx51:09 asic interface
EZ150 ke e в EZt50
PR po sv lxto7 sy
EZX5t11 XtB
VER (32
Eu
Low voltage supply ERE ao Ίχιπ I
Ost./Schr.
EZ102 o 1 L
C14
EZXG108 N hgty lxtosÈ Tw =15V
96-03-13
EYA
va0301
AVA3
Low speed
Rotor control
OPTIMUS (a/96.0)
© Philips Medizin Systeme 21-12
너트 к 43 ps IGBT - Modut
> be
Power supply 2 -BF Lo. rc 9 e 5 E 1 € terminal EX1100
3. 400¥ 9 —— pays? ==
from ENE3 tube e Pension RO/SRO tubes
Po 8 8 L2
Nm ku Filter and Cau u us xes2) u Ping Core υ y
rectifier = L1 TI [ [| y i | v
エー v M XTN Π M
ow | w へ καν LT | м, MW
a
E/RZ9 2014 LE LE LE ses . I 3 windings
Ἐν
Rectifier 2 8 4 =
E =— A ~
下 Ne
N x45
NTC
X4:1 re
R24
Uref=+5V BK]
x44 © | Stat
END © | Selftest
STATUS “Sm © | CAN-amouncement
.
Mnemonics O | |Ready
© | | Watch dog error
CANH generator CAN high active
CANL generator CAN low active
RC ST 2/ X3:1 Tube extensi
NG negative ube extension,
Pa positive u RE ST.3/ X32 Tubes over of stator contactors
RCST2/ | rotor control stator 2 Computer
3/ 3 R34 X5 Confrot
RESETC/ reserFu
! Driver
R36 WWE SA with | № 1..2
can only be measured galvanic IGBT modul
via 1 kHz filter wi decoupling
made: 이 open
W2
Hi
1
3
address:
23 START-HI № READY -C x2
mt sl | o STARTER 3Ø REAUY-GND X2:2% Lot used
X2:5...10
r 13 CAN H
+2 CAN L
110 RESET CZ
xt
paz Y > IV
do dem
」 1004
ーー
8 Je Codinging [Rot +
|Foto” control
TE » ~15V W2 7
il i
x12 ,
Los w3 A i
IP
上 7 Les 3
Ost./Schr.
X13
Y
、 Jo EY100
A1/A3 96-03-13
Rotor control HS
4512 108 0870.
EY
va0302
High speed
rotor control
9890 000 02212
OPTIMUS (0/96.1)
© Philips Medizin Systeme 21-13.2
Tube extension for 2 tubes Tube extensions for 3 tubes
Hardware ammi
Roam 2 Room 3 cable
kV control Basis interface High speed kV control Basis interface High speed
EZ10 E7150 rotor control rotor control Tube 2 Tube 3
EZ130 | Tube 2+3
1 Tube 3
| ΕΖΧΊ 7 CEzx46 EY100X3 EZX1 EZX16 ? ΕΥΊΟΟΧ3 Tube 2
1 ww Ş
T sees tink Xp X11 Soldering link 시 control (tube 3) Ш
K1 K2 KS K3 K1 K2 K1
111 K2 κ «κ KM
1 KIZ
control (tube 3) X11
7 emu
!
Soldering
o
link 1
|
!
УЕ)
KI K2 KS K3 K41 K12
X2 Room decade 1
EZX1 MI $ =Xe Room decade 1 X2 X1 с A
Room decade SDA K3 Room decode EM
<——+ lxs À KS TÁ×13| 3rd一fube in一一coom 2 XB ? X3
”~ 人 一 Room decade 3
© A Room decade 2
κι Room decade 2
X5 、 、
Cooling unit control 1
EZX2 Mu 45 Cooling unit 1 X5 Xi K2
EZX2
Cooling unit 《一 一 K2 Cooling unit pu K2 | | X6 Cooling unit contrat 3
E- Cooling unit 2
| 은 2%6 Cooling unit control 2 |
X8 sa
| x8 —— Tube supervisian 1
EZX3 7 |$ | Tube supervision 1 | x8 X7 o 14
EZX3 å
Tube supervision O—— Tube supervision >: XI tl x Tube supervision 3
은 si Tube supervision 2
OPTIMUS
M + Tube supervision 2
Rotor control i As 一一 一人 Statar supply 1 Rotor control a #6 K11 o 7 一~ 一 m K11 一 Stator supply 1
K12
их Statar supply 2 esa t ly 2
ator supply E ASS Statařor supplyly 3
X61
X61
X616 X51 5 SK
ENF3:13 Ž30V CT K1 ENF3:13 230V… 1Х51 K1 ? X63
$ X63 ? X63
im jm X6
x52 X65 xe2) |그 그 X52 X6 X62 xs2 xe xe2l エゴ ーーーーーー |
| м ke | | к Ig |
ENX 1101 N X57 X67 || fube1
ube tube2
ube | |
ENX101 由一 [457 X67 | X57 x67İ | | fube1
ube1 tube2/
[tube2/3 |
X68 | x68 x68] |
H.V. generator ! EG
К, X82 |
に cr H.V. chan gre | | |... HV. generator!
KS | χε tube? KS | yaz jo; 一 一 一 一 一 一一 |
X72
© — ©
X72 X81 TT TIT N Tg OS x72 = Ta 1002 |
Ost./Schr.
X83
X83 X83 η 3;tube? de
|tube3 |||
X84)
X84 X84 |
96-03-13
Tube extension
1WG 2WG m changeover sw
Tube extension Tube extension
1) Connect here if
vn0292
Α2/Α3
+
근n
NJ
z
Ирим5 роомало Бызпоц адл4 : MS TO HL
DUAS PDgIJgAO 5ulsnOM 29 이 AS TO HL
z z
1 PPOTJaA0 5MjsnOM aqny 170 HL
BOO1ddAO Buisnoy agny д 9JE 5
7011
uunyad E Dans
90420 2 agny
SUE gi €/Z 서 0005 soynys LI İğ et
(INTO
Ea /€ 8L LI zel
za 1281 Y ZA
ZX3N MS
LXIN MS
X NT AS A 9 JE 5
UO 2 100 000002 42145 A MIS
NO Z MS
10 | 100 600003 4215 NÜ MI MS | Addns voypys LI DM
` M Bum 401145 Ц AIS 1044003 40104
MIS || Ll
A 8U 401045 AIS U
Wy
LI fus
пам 391035 DIS ソ
Z 4IDJUOI 4000 0004 Z 13 90 He
し 430403 Joop woos | 13
Α0 4594001
80 048 fx
4201053
4000
soup
4004
woos
AO 30 WH EYE anı sy
19 40 Ha
£ 개
Z J01DIS 1041002 20404 7015 20 EK |
2 Touoydo Addins A9Z Lex
£ Tuoyde Addns Age
Z A9Z Ud TEX
L AIT Od
Aiddns A9Z ASZ Od : EX
Yi0uoudp Addns Ag Z ASL Od EZ aw E
L imuoydo Addis ASL L ASI
Addns
Od baa Brox
ASL ASL Od ZİN
uo asnsodxa NOX3
va SA O— — — À se at
L4 4204003 Z ¿an Bunonz Ta UDIJDJS MJOM
LST ох JO 001120105 DIA
ΗΤΤΕΣ] 111310
A Tar UT EL juawabpaymMouyay
σι σι
SH Ed 《过 Ly
11 4904003 ¿run Bunoo) LD a
MMS ÖLASNOY BON) UDUAYOZ MS HL WD
Buisnoy agny vowwo>
3
HL WD
UOYDSIPU! UO} DIpDY 40 LOUKHOI ASW
IX SIX:
3S, JIYOLZUŞ|
=
TX
|
venia ee
DI 154
17060) 1
o
GO Ta
MS NIT
ZA Od Fe
. ASH Er
nů ně
SERVICE MANUAL
Drawings
Wirings
Cabinet E Z2-1.0
Cabinet wiring E, 50 KW Z2-1.1
Cabinet wiring E, 65/80/100 KW Z2-1.2
Earthing diagram Z2—1.3
Option rack W
Tube extension WG/
230V/24V Adapter WR
Basis rack Z
Power supply N
Rotor control Y
OPTIMUS
2 nd converter 2 Q
Converter Q
H. V. generator G
Sie./Ost.
A4 94-10-24
22 1.0
OPTIMUS
Cabinet E
OPTIMUS (94.1) 之2 一 1.0
© 1994 Philips Medizin Systeme
ALL RIGHTS RESERVED
FRONT ПГ
= |
| a x4101
| R noz
A
sk
ll
Z X47
|
i
Cabinet earth
M 100 |
Z X13:2 |
System
İNS a
N (i Le ‘i I |
i
| | O A su? L2 :
хх О = 9200 хи |
apa F3 | 5100 x15
712 |K2 ひら し czoo x13 El X24
T1 T2 T3 |
|| [ооо
eta
Q100 X1 ea x12
(2 |]
Î
| |
YA100 X13 or Y200 X3
YA100 X12 or Y200 X2
YA100 X11 or Y200 X1
ul
х
YA100 X14 or Y200 X4 |
G100 X4 ==
Vo
|
fl
X1101 yů
хит
o
М КЛ:
1 = a | |
= C300 X1 operating panel
X42 X 46 X47
X4
N K1:3 =
NN
S;
K2:42
K2:41 Termination resistor CAN oa | X44
pp me!
. | X51 x52 |
N K4:5 sed 200 11 + N100 x1
(Y ノ YA100 “十 | eee Termination resistor CAN/XS
{mo option present)
Q Converter Programming plug
(desk present)
x17
ο
0100 x2 -| x20
w
A [200 X1
X5 ! operating panel
< Backpanel Eb = F +
oo
一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一 一一
X6
YA100
4 XŤ X2 x3 X4 ха
N F3:
N X1 <— high speed
G x1002 6X1003
A or
Z x51
Top of H.V. generator low speed >
x1100 Y
rotor control
9 9
EF)
| || 200
Y100 UG Xu7
ome
N F3:T1 fee
X16
X1
N F3:T2 == X12
・ T ZIT 2 Ν Ε3ΙΤ3 κο Χ13
3 windings | O 00 ©
Ν Χ1102 == X14
Schr.
: ‘ ТЕ
« ry LJ)
100 Y
96-02-28
O C13:1 A C3:1 x | |
O |
x15 | | i
dI
ub0780
A3/A3
의 Cabinet wiring
x1003 X1002
|
m
Z x12
50kW
OPTIMUS (b/96.0) 22-11
© Philips Medizin Systeme
| Netz
Cabinet earth)
front 一 一一 一 Pear
LT 12 13 o xttot and 2a x01
| Q x1102 and 20 x1102
| | | | | A x1103 and 20 x 1103
| cl [2 13 © © 5 |
o o |
| x1
6100 X15 es 7 X7
Z X47 =
| |
6 windings X19 264
| F1 ot
A
내 100 :
i:
1
|
| Z X13:2 System
|
|
+
X34
9 N F3
X13
X7
fem
! 20100 X1
| gado!
| 一 一 一 一 | の X13:3 G100 X14
C300 X1 operafing panel
| » X42 či ri -
! Н | b O 5-8 2 Termination resistor CAN End Ger !
| x| x © F=a200 X4 and 20200 x11
lle ΕΞ 11
|
| | İN
XS1
N100 In
Y100 X1 === 9 E Termination resistor CAN/XS
F2 112 KZ O P0200 X13 and 20200 x13 (no option present)
IN. plug
| T1 T2 T3
: N [| | 000
(desk present)
[201 00 X2
I E x1101
π | | Y208 X3 ο er x1
쁘 Y200 X2
Y200 X1 Г Backpanel (Ek) a =
operating panel
Y200 X4 ;
X1101 Y
же та N k11 ©) E13 i |
ΧΊ
| ×2 x1102 N K2:42
Z X24 N 13 ©) N K2:41
Z x17 X1103 lf
№ КБ веры) 200 |
Q Converter
100
C3 C13
200 100 X46 X47
O-
ON
DN
O
3 Ὃ
10 O1
INI
tu
O O Y ROTO HS
G X1002 G X1001 X4101
NKT1 B x 13
Top of H.V. generator X1100 Y LM X1102
K
© NK1T3 = K
I X 1103
(E de) N Kt5 (©) 200
Schr.
- 100 ..
96-02-28
ο) db
| | T x35
圖 ol ls
| (3 X15 |
| |
ub0781
E
A3/A3
| a οἱ [ο ο |
G X1004 G X1003
Cabinet wiring
x1001 1003 1002 1004
Z x12
65/80/100/kW
(b/96.0) 22-12
© Philips Medizin Systeme
Pear E
TT TADO (0
EW Option rack
6 windings
OL”
Backpanel
e
C Operating panel
7101007777770
Protective earth centre
EY Rotor control
Tube assembly
OL
È Q Converter
Metal can
of tube
Schr.
G H.v. generator
96-02-28
un0782
A3/A3
Earthing diagram
. . Е
cer m
ミミ ヒマ ヒビ
NE NE
MEMU MU
SETA 一 9
を :Zツ ーー N m
UL
LU — SG
一 KtA2
2 -一
leu
D ZOLHX
<
ELIZA
Aad
DEY
RC1
<a CLI
K1
Va
1
11123
|
cita 113
©
(4744 001
У:
#143 —
OS pe €:£X 00L
— 9:24
229
て11 コ ーー E O+— To
Γή
Ox
レー し EX 00L
uia O
ーー ZZ
e]
L1
O
ONA
OFF
Ех TN.
R3
PG)
МА
ея TA.
R2
— €
MZA TA
R1
ーー M
(a/96.0)
© Philips Medizin Systeme
OPTIMUS
S8L00n
IUS 92-10-96 EV/EV
у
N
Z2-2.2
W
一 一
ах
うら MN —=
se
co
=
Paや
上
ce ram
GERGER
ο ου 5ο
DE N° Yu
sz —O m
EZ O N
UZM +) 5
EL4d
στα ) 2143
K1
klid
1438 人
T:ZX 00L
e — O) 2 m OT EEX 00L
E eme, ETA
am
— 0) 9 “Ora
Mİ) E
一 Ot L'EX 008
Dem ZEN
F1
G OL
ONA
OFFV
ED TA.
R3
TYTN
— S1X
МА
TO EAN
IVI
PEA
WU TN
R1
po LIDI
(a/ 96.0)
© Philips Medizin Systeme
OPTIMUS
| 98/0Dn
342$ 92-10-96 EV/EV
front rear
T
X1101. GN F471
!
X1102
X1103
Ge
Φρα
FAT2
Ν F1:T3
|
|
| | El KI
| |
| |
| | 0 0
| y LL
0 Befor instatling NT cancel the following connections:
NT X1201
xm2 DI NT хм
3301 si” NT X1202
x1202 N NF+T1-NF2L1
N FITZ - N F2:L2
NT X1103 ンン Ем X1206
г x NK
N K2:A1 - aa
№100 eХ2:1
Ost.
X1201 Ge N F2:L1 |
| Jo $ © G N X1101 - N
X1202 "8==NF212 ㅣ x Ix ”
94-10-07
X1203 = N F2:L3 Е? 1 | K2 F3
X1204 _Gk=sN X1101 I 2 2
I
×1205 Bj==N100 X2:1 |
N T X1206 “GE N K2:41 ㆍ
v00339
A3/A3
NT X1204 μα
Mains transformer
OPTIMUS 194.0) 22-23
© Philips Medizin Systeme
St s2
で Î N K2:4!
TS
| i N K2:42
i i
ido 4 sels XS | 1
[23 c e == XII xi2 =で
χιοο XiOi X| x300 x301
Å eve 77777 Å fm v3:E TN ws xi っ ХА ==
ViiG s! | v3:6 Sa xi
x4 — | Net! (er x2l L
LI Er) 고
一 一一 n |
52
| | © 一 一 一一
~~
DN x2
y le LX >. NOLI ibi) x22 =£
EG È | /
fps 一 一 ~ LI (DİK)
Pr
\
ΙΟΟΣΧΙΟΙ L _ 100:X 30! \ a D O O
X
| V | CE © | | ` Mest!
es
(vio)
а Li (vio)
1
ОЕО |
OEO | O
+
O
一
— x5 X25 ==
A
1 | |
一
| | |
cp
| peo
| || L.
i ㅣ レン |
C3:2 | [ct3:2 A — + |
OEO |
|
One |
|
ol lol! )
|
X 3
GB) GB) , | joMeLsMI (ての)
la (gm)
so V D V 4 C | D Y zajem (51)
> 100:x200 100:X400 ,
ノ i
I00:X 201 а I00:X401 | /
/ î V3:0 tra rela V5:A
i 1 ーーーーーーー- ^ 100 - Le |
/ .
È y
V2:G VAE
{G XI003/D
A3/A3
(G X1002)
0/20
kV power unit
KE
i Luvi
1
I
i X21 he
!
| LI (bm) _| /
|
| 一 一一
x22 la
q UN e.
Ч
| /
| / 一 ~ Li (Bl
| /
、 200
\
ΟΟ:ΧΙΟΙ L- =
Pa
N
o|o : es — |
\ X24 ===
x O |
OFO ーー
nus
I
I
| a
‘om
L. —
1
E
+ |
1
a
U
I
I
O ) ХЗ
GtB) | 圖 e Mi (ra)
V2 V4 | D /
.- 2SIE este MU (DIO
И ах! |
100:x200 100:x400 ノ
G E
x400 X401
V4:G
Ost.
V4:E
他
100 X31l (rd)
96-05-02
|
vc0309
(G XIOO3 ノ D
A3/A3
{G XI002)
0/20
kV power unit
Fa ou LOM X81EZİ—RESET Sw
ον L 013
PO 40v to PO 40Y £ CH4 X82 [EN x/
№ 5У NG 15V wp PR x
SIGN CHI SIGN CHA X83 ビー /
RF OV CH3 RF OV CHE. X84 EERO SN χ/
X85[CT-Cm x/
X86[ET-X ACT/
Measuring chamber 5 XE,
x41
PO 400V
FOR CHS
REL CHS cos Tube extension System CAN
PO 40V UDE X42 X43
NG 5V X16 CAN 2
SIGN CHS RC ST 2/ SCAN L
RF OV CHS RC ST 3/ SÓN GND
5 CAN PO
Rotor control 1
X51
{} fod
fed
ilah
ni,
Room detade Cooling unit control DRC input fl fey
x1 X2 X61 „ků
a 4 E $ OR FG Pd | ya
DS BY OV o.fol DR FO NG fæ
[ 8 [ 6 OR BY OV д DS BV SG 圖
è PO 26V 3 OR TV KT DR BV SG
SW UN EX $ $ LE”So; OR TV PT 2
PO 26v 5 5 V107 İvme [voz V329 |v5t İv70 İvst an x
$ 6 À À Ruin À À 7 i ἤ
cM Sw uni vo [№99 İv77 vs. [veo İv76
$ CUCT12 4 у Y a İP
RM DR CT 3 PO Sv 3 L 上 L L B mi
Ost./Schr.
cf
LT 8 pufv 一 一 - H_3 vii
RM DR ov ㅠㅜЦ ГВ
O, pte, se lume) es v75, |va3. |v134 |v132 İvso
TTTT DER oN
A1/A3 96-01-25
vd0314
Back panel
Basis rack Z
OPTIMUS (a/96.0) Z2-5.1
© Philips Medizin Systeme
kV power unit EG
X24
ny 4 Tube extension
12 H.V. generator
"2 x11
X12
War 113 PO 26V
Was
E NG Cv1
Hé Aretintd $45 21
Poti TB 27
Cvi TM
TB 3/
FTF12 一 | $ A tr22
E TB 2 RT/ 18 3 RT/
ντο CV?D/
으아
EVA NI OL
01600
CVA ENZ
Cv165 D/ Power supply
XB X16 X15
AC 230V L1 AC 230V Li AC 230V LI
AC OV XG AC OV XG AC OV XG
AC 230V 12
kV power unit E2Q H.V, generator
X34 x35
ウシ 21 ran CA GND
23 122
23 AN GND
E Vacs 124 TP HT
TP HT GN)
E PD (V2
AZ Si (v2 D/ T
8 oon u
FR
TB CU FR NG
Cv2 GND OL 15
002 60
CV2 EN
CV2 65 10/
[ CM TH $
」 . 9
Function programming plug for X44 THOL Sw/ 6
with / without cM TH Sw 3
operating panel C 8
-- PO 26V | 3
1 {È
x 53 ]4\64 ]4111 14069 |ν
we Ave XAO BANA
CANL
>
2
X72E H-CANH ー he
RESET С, i CAN H X73 reser ©/
STOP X Cs $ 8 oe o. X74
に CTRLx kV power unit EG/E20
X ACT S/ ὁ > Х75Е-- 5ТОР х t/ X17 X18
vs 59 №49 №74 ×76 xc/
ХУТЕЕ--х т sy PO 26V PO 26V
Rotor control 2 주 4 È
X151 må 一|
RC ON/
CAN L Operating panel
CAN H Termination resistor for X52 x8 比
PO SV O+—PO 26V i 60 266 1 PO 26V 1
o
PO BV ~ 3
NG BV 130R бо X5 X6
o O -
Ost./Séhr.
5 Gİ
ο . -
‘ne L
96-08-27
va0315
A1/A3
Back panel
Basis rack Z
OPTIMUS (0/96.2)
O Philips Medizin Systeme Z2-5.2
Basic interface Central unit CU kV control mA control Low voltage supply
150 139 130 X119
x1 EN X/ x1
102
X1 X1
CA CA CA fez
x1
PQ 4004 —+0-——O GO PO 0V
08 2
117 —0 ен!
о? о Lo 2 οἷ--- RESET sw/ H2/—+o 2 of nz обо
Fou cHt—to 3 of mic AL 230 LI odo
Ra su x/ =—Lo 3 ol MR PR X/ 13 — 2043
FDC CH —to * ol mca ramixe —to * of crat x, ne do den arm —İe$
FORM —o > of— FoR GZ AC OV XG 一 [oo
ROM2x/ —to > ob x act ENG CVi—t+o 5 o| E PO tv
REL CM —to 5 oL ret az pov —po-S-o cui rm to 5 o| evi, area lodo
SIGN CHI 一 +-o 7 of— sign на çk
REMIX o 7 ob— Ex tria Lo 7 ol eva
RF OV CHI —o ® ot— pF ov cH2 обо cvs ano —po- Lo TF2 4 —Loo
FnLcH3 一 to 7 of—~ FOL CHa обо
o9o Cv165 9アーー ト o 9 oo cyt ens
FOC CHS —+o 1 of 501 014 0! ol POWERFAL/ ta3rT/ o © of Tezer, a 722 —Loo
FDR CH3 —to ob A odo
ole OM TH sw —o Tol ma sw/
REL CHI —+0 È ot ва 0% oo MI de lol ma οξο
saz 一 Ho 8 of sion ne обо
oo 183 —o Pol mz
RE OVCH3 —İo of RF ov (ης o* 6 sır — Фо обо
FOL CHS to 5 o обо
eran tito Bo РМ ON P0 —+o * oİ— pw oN NG/ o
POT —İo “o SCANL —to © of — s canny rca —to 6 of Hr ca o обо
FOR CHS to 17 6 обо
SCAN GAO 一 To 7 ムトー scANPO HT AN 一jo 7 e+ HT AN GND
REL CHS —+0 $ o ово o Bo odo
SEN CHS —to бо odo
о Зо TpHrss 一 lo 9 っ し TP HT GND
RF OV CHS —to 2° of pa rov os oo odo
0210 оо oo odo
EUCT12 026 Eu cri о22 0 gt 一 lo % of — 124 odo
020 odo
oo 2zu—o ol 22
oo oo 237 —+o ** of 23 scova loto
RC RDZ οὓς odo
RC ONY 020 1244 —to 2 of 124
RC ST 3/ o É o RE ST 2/ о 26 о ENG cv2 —tO0 26 of— E pocvz at 230v 12 =—Lo #0
027 5 обо
oo va Tm —o 2 od. ev2 py
RM DR OV 08 0 RM OR CT о28 о cv2 aL —to 28 of CV2 END OL Lo Bo
CM SW 3 o odo
Sw UN EX οσο Cv2 GND —to-22o I
oo обо cvz 65107 —o 9 o上 cvz en © Sol
oily oo 18 cure Pa POWERFAL/ обо
ο ϱ χ
Lo 32 038 0 о 32 о o ol ta wu rR NG [31 pozsvsw odo
№17 _ №21 [vat [vio №22 V28 {Ve N37 13
소소 E À
3 V131 İv15
L l ㄴㄴ .İ
Mm 조조
X2 X2 XZ
CA X2 X2
СА CA ㄷ , DBZ
Ρ0 ον ——0——© のーー で Oo やーーで
PO sy 一 +っ
っ 2 2 vio
CN to 3 dt cant 6661
cano 3 ol— cant CANH io 3 of — cane CANH 一 ro3o CANE
LT. oo ο---ο O 。
creLxC/ ーーo 5 oト ー msET UR XU —o È ot RESET C/ CTRLx ソ ー オ o 5 oト上 esir CTR.X CZ —o É ol— RESET
NX —o 7 OL stop x cv NX —o A
7 o| stop x cy NX —o 7 o — stm x cy ВХ си —to 7 o|— stop x cy
ο ὃ οἷ--χλιτς/ o! o x ACT S/ o 8 of x acrsy °F ot ХАСТ 5/
0° 0 dol
o9o o%o o%o
front side обо οὓς о бо о бо ot
PO 15V dodo. ti oo oo
MG bv —to-Zo. ono
x
8
x
s
>
ο8
13 12 72 Bo |
8 8 8
agi —
『 ο) e ς 29
ola,
— ss 一 T «İT
p => e c
> Bic Бо
i
EN So
OS MC Ov 7 아주 DS MC SG DS MC ov —Lo a ol ais MC SG o > o
serene PC
oso] bold
yso7o| dy os bv ov —+o 7 of — ps ev ss οἵο oo
обо or uv ov —+o 8 of — pe ev 56 о Зо lo BY
oo Lol
DR TV NT Lo "9 of pp ty pT о Во To Vo
E
УВ
о 20 о| „VT pr ra Ni — Lo 2° od m Fa Po o% o Wo
020 Bol.
C- う
оо 020 oo
boe
C%
02 0 020 oo 020
020 ος 020 о Фо Роу
020
p TI
otto oo oo oo
оо 050 050 o Šob 一 上 sv
обо
oo ово ово oo
o27 っ ote 010 do
ote
0280 oo 0280
020 020 ozo odo PO 26V
020 оо
обо oo обо обо .
~ oo oo ο οἱ -- dol
oo
= oor πι co > PO 26V1 Log
a
vi
a 402 №65
LAY
9の
a
m
<
=
v00362
ul
=
=
=
N
=
x>
Så
δὲ
ロ
a
<
Back panel
Basis rack Z
No
=z
je 之
=
=
<
o
=
=
=
N
Pé
we
OPTIMUS (95.0)
© Philips Medizin Systeme
Z2-5.3
EZ X51
X16
PO ===
x14
EN F3:T1
x12
EN F3:T2 — LT
EN F3:T3
EN X1102
100
e
=
나
vu
г .
A stator cable fo | |
a intermediate connection strip
고 or fube extension
ον
N
<Я YA
à< &>
Low speed rofor control
| =O =:
」 {when tube extension is presenti
EN F3:T1 EN ERT?
1 GNO NS x3 |
O 7 ο ο ο ο O
i EN F3:T3 X41
| EN x1102 |
| | |
и v L3 N |
ㅣ
x1 x2 хз X4 i "О o
1
I
I
! O
| O
I
| xe
I a
i | |
I
1
| [ο ο oo ㅇㅇ oo co ㅇㅇ
1
1
1
I
I
! vi
i
1
ㅣ
| P VW V U N
1 | o o o o |
!
I
I
|
i
ZOO |
1
1
x5 x6 x7 xe 1 O O
N | N
1
i i I
Y100 X43 Y100 X44 I
I
| X46 X47
Schr.
O fo NG app К T 5 3 O
i vzoo xs 。 A Ó E
Stator cable to a ノ ノ ノ
intermediate connection strip i
ub0787
A3/A3
ar tube extension Y I . Y
! High speed rotor control
|
OPTIMUS (b/96.0) |
© Philips Medizin Systeme 72-13
+ + E È
XI XT 2 3 4 5 6 7 8 9 X12
Y100 X3 (HS)
y (LS)
x13 κά X15
K411
K113 K11:A4
f Ne NE KIZIAI | | К 5 が
6 由 à
à ο οιi 1
©! ©! O 和 K1tA2
|
X44
| K12:A2
K12:1
K11 A
K12 | DE _/
12 12 JÓ “^
© ®© © O © © © © O © X43
re = 5 Le Luz
y
Rotor control unit
52 62 64 68 82 84
Ost./Schr.
x51 57 61 63 65 67 69 72 81 83
| ή | ς) ( 1 | |] | |
| G K2:1 G K2:2
96-02-28
G K1:1 GK12
NF3:L3 N X1101
va0340
A3/A3
WG
Tube extension
OPTIMUS la/96.0) Z2-14.1
© Phitips Medizin Systeme
front rear
ーー .
Z X11 Z x1 Z x2 Z x3 |
| a [|
| x14 Г
OOOOO TOOOOO
| 735
| i WG Ù ペー ° KM K12
| “0 68 κα ooooolooooo
| 57 61 67 =
| | | LO gy bo Eun
N F3:L3 | | G K2:1 İç K2:2
i N X101 GK G K1:2
j WG x57 WG X51 | EZ x16 Y100 .
|| or or
| Y100 X3 YA100 W
| x î
|
! | \6 + |
| .
|
İN |
|
|
| WE x7
| x16 NE
WB X1
| re X4
V eee
WE X14
| YA
| i l <— high speed 100
G Y200 Y100 ar
low speed—>
rotor control
Top of H.V. generator x2
| | | 1626
O_O
2 X46 X47 Tİ T
Ko 7 WG X68 121102 | 165
Schr.
OO © 0 TIP
O== WG X62
1 WG KT a
- To - TRE |L--------=-------------1
i
96-02-27
LY) 1-ーーーーーーーーーーーーーーーーーーーーーーー
2
O==WG X67 |
KT Om WG x61 » |
1 i
ua0784
A3/A3
Tube extension
πμ
2WG x7 I
| {housing decade 1)
i
{housing decade 2] + x12 4 x2 4 х5 È x8
| (housing decade 3}
i
ps x |
|
14 x1 4 5 6 7 8 9 XI ΧΊ 2 3
X12 | 4 5 6 7 8 9 X12
圖
|
Ì
|
|
1
|
I
|
I
2WG X13 |
TWG X13
1WG X15
ys κά |
|
|
X13 X14 X15 I X13 X14 X15
|
|
I
|
|
|
|
|
l
|
|
I
|
|
! iz z 13
|
00600000000 O I
00060000000
x41 |
|
Al A1 X4 1 o
|
KT K12
X42 !
| K11 K12 LD 一
A2 A2 |
SP
| A2 O
O O 9 + 9 O O X43 P © © OO © © Y O X43
Ì
|
—
u fv Jw | U У W
ες...) 1
|
| rel) = cp.
2W6 K11 stator supply 2 I WG K11 stator supply 1 stator supply 3
|
|
|
62 64 68
|
62 64
X51 52 57 61 63 5 82
Ì
84
Ost./Schr.
> 67 69 1
A
V | |
| 52 57 61 63 65 67 72 81 +
Г | 4 |
Av | m (Е.
|
Ì
|
I WG a ーー
96-02-28
3
A3/A3
TV ロロ 2 WG
|
|
| IWG/2WG
Tube extension
OPTIMUS {a/96.0)
© Philips Medizin Systeme Z2-14.3
front rear mara
j | 2. Adapter | Adapter |
X41 X41
[È WA X42 | Z X19
X42 x42 |
: | om 2WA X41
i
| x51 Г х51 |
ro x52 u X45 ㆍ
x52 ‚ х52
| | x51
x
1
W 2WA WAV - i ー
‚| 1
00” X41 | | i
|| [ez x19 ! 」 i
X42
一 — — i «| | |
X45 1 X45 を E»
|| xs WA X51 ji o | WA X51 1!
maZ X45 : i E 000 힐
JUUL | bo
x52 i |
| 1
= |
WA
Adapter 4 auxil units
OPTIMUS (96.0) 22-15.1
O Philips Medizin Systeme