User’s Guide
Agilent Technologies
Notices
© Agilent Technologies, Inc. 2011–2013 Warranty Safety Notices
No part of this manual may be reproduced in The material contained in this document is
any form or by any means (including elec- provided “as is,” and is subject to change,
tronic storage and retrieval or translation
without notice, in future editions. Further,
CAUTION
into a foreign language) without prior agree-
to the maximum extent permitted by the
ment and written consent from Agilent A CAUTION notice denotes a haz-
applicable law, Agilent disclaims all war-
Technologies, Inc. as governed by United
ranties, either express or implied, with ard. It calls attention to an operat-
States and international copyright laws.
regard to this manual and any information ing procedure, practice, or the likes
Manual Part Number contained herein, including but not limited of that, if not correctly performed
to the implied warranties of merchantabil-
U1231-90026 or adhered to, could result in dam-
ity and fitness for a particular purpose.
Agilent shall not be liable for errors or for age to the product or loss of impor-
Edition incidental or consequential damages in tant data. Do not proceed beyond a
Seventh Edition, September 20, 2013 connection with the furnishing, use, or CAUTION notice until the indicated
performance of this document or of any conditions are fully understood and
Agilent Technologies, Inc. information contained herein. Should Agi-
5301, Stevens Creek Blvd. met.
lent and the user have a separate written
Santa Clara, CA 95051 USA
agreement with warranty terms covering
the material in this document that conflict
with these terms, the warranty terms in WA R N I N G
the separate agreement shall control.
A WARNING notice denotes a
Technology Licenses
hazard. It calls attention to an
The hardware and or software described in operating procedure, practice, or
this document are furnished under a license
the likes of that, if not correctly
and may be used or copied only in accor-
dance with the terms of such license. performed or adhered to, could
result in personal injury or death.
Restricted Rights Legend Do not proceed beyond a WARN-
U.S. Government Restricted Rights. Soft- ING notice until the indicated
ware and technical data rights granted to conditions are fully understood
the federal government include only those and met.
rights customarily provided to end user cus-
tomers. Agilent provides this customary
commercial license in Software and techni-
cal data pursuant to FAR 12.211 (Technical
Data) and 12.212 (Computer Software) and,
for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial
Items) and DFARS 227.7202-3 (Rights in
Commercial Computer Software or Com-
puter Software Documentation).
• Do not use the meter if it is damaged. Before you use the meter,
WA R N I N G
inspect the case. Look for cracks or missing plastic. Pay
particular attention to the insulation surrounding the connectors.
• Inspect the test leads for damaged insulation or exposed metal.
Check the test leads for continuity. Replace damaged test leads
before you use the meter.
• Do not operate the meter around explosive gas, vapor, or wet
environments.
• Do not apply more than the rated voltage (as marked on the
meter) between terminals, or between terminal and earth ground.
Pollution degree 2
If you are unable to search for the respective DoC, please contact your
NOTE local Agilent representative.
2 Making Measurements
Measuring AC Voltage 32
Measuring DC Voltage 34
Measuring Resistance 39
Testing for Continuity 41
Testing Diodes 45
Measuring Capacitance 49
Measuring Temperature 51
Measuring AC or DC Current 56
Measuring Frequency 61
3 Multimeter Features
Detecting AC Voltage Presence (Vsense) 66
Making Relative Measurements (Null) 68
Capturing Maximum and Minimum Values (MaxMin) 70
Freezing the Display (Trig Hold-Log and Auto Hold-Log) 72
Trig Hold-Log operation 72
Auto Hold-Log operation 72
Recalling Previously Recorded Readings (Recall) 74
Making Scale Transfers (Scale) 76
1
Introduction
About This Manual 2
Documentation map 2
Safety notes 2
Preparing Your Multimeter 3
Checking the shipment 3
Installing the batteries 3
Turning on your multimeter 6
Automatic Power-Off (APO) 6
Enabling the backlight 7
Enabling the flashlight 7
Selecting the range 9
Alerts and warnings during measurement 10
Adjusting the tilt stand 11
Connecting the IR-USB cable 11
Power-on options 13
Your Multimeter in Brief 15
Dimensions 15
Overview 17
Rotary switch 19
Keypad 22
Display screen 24
Input terminals 28
Cleaning Your Multimeter 30
Agilent Technologies 1
1 Introduction
About This Manual
Documentation map
The following manuals and software are available for your
multimeter. For the very latest version, please visit our
website at: http://www.agilent.com/find/hhTechLib.
Check the manual revision on the first page of each manual.
• User’s Guide. This manual.
• Quick Start Guide. Printed copy for outdoor use,
included with shipment.
• Service Guide. Free download at the Agilent website.
• Agilent GUI Data Logger Software, Help, and Quick
Start Guide. Free download at the Agilent website.
Safety notes
Safety notes are used throughout this manual (see the
“Safety Notices” section for format examples). Familiarize
yourself with each of the notes and its meaning before
operating your multimeter.
More pertinent safety notes for using this product are
located under the “Safety Considerations” section.
Do not proceed beyond a safety notice until the indicated
conditions are fully understood and met.
Before you proceed with the batteries installation, remove all cable
CAUTION
connections to the terminals and ensure that the rotary switch is at the
OFF position. Use only the battery type specified in the “Product
Characteristics” on page 102.
Full capacity
2/3 capacity
1/3 capacity
[1] Batteries change advised. Always use the specified battery type listed in
page 102.
When the HELP mode is enabled, the multimeter flashes the international
NOTE Morse code distress signal (· · · — — — · · ·) repeatedly. Use this option to
send a visual alert or notification of a distress in progress.
3 Press or Null
to cycle between the HELP, dEMo, or
Recall
Range
Auto
The range is fixed for diode tests, temperature, VZLOW, and AC/DC mV
NOTE measurements.
To PC (host)
IR-USB cable
Tilt stand
Ensure that the Agilent logo on the U1173A IR- USB cable
(purchased separately) connected to the multimeter is facing
up. Firmly push the IR head into the multimeter’s IR
communication port until it snaps into place (see
Figure 1- 3).
Power-on options
Some options can be selected only while you turn the
multimeter on. These power- on options are listed in the
table below.
To select a power- on option, press and hold the specified
key in Table 1- 2 while turning the rotary switch from OFF
to any other position. Power- on options remain selected
until the multimeter is turned off.
Key Description
Key Description
changes.
Range
• Press or
Null
Recall Autoto cycle between the flashlight’s
HELP mode, dEMo mode, or flashlight mode. To learn
more about these modes, see “HELP and dEMo
modes” on page 8.
• Press and hold for more than 1 second to toggle the
flashlight on or off (for any of the modes — HELP, dEMo,
or flashlight mode).
• Press and hold for more than 1 second to exit this
Esc Shift
mode.
Dimensions
Front view
86 mm
52 mm
169 mm
Overview
Front panel
The front panel parts of your multimeter are described in
this section. Click the respective “Learn more” pages in
Table 1- 3 for more information on each part.
2 Keypad page 22
3 Terminals page 28
Rear panel
The rear panel parts of your multimeter are described in
this section. Click the respective “Learn more” pages in
Table 1- 4 for more information on each part.
4 Flashlight page 7
Rotary switch
The measurement functions for each rotary switch position
are described in Table 1- 5 on page 20. Turning the rotary
switch changes the measurement function and resets all
other measurement options.
Click the respective “Learn more” pages in Table 1- 5 for
more information on each function.
Remove the test leads from the measuring source or target before
WA R N I N G
changing the rotary switch position.
Cycles between
page 32 and
Hz Hz Hz AC V AC Hz
page 62
DC V N/A page 34
Cycles between
• Short ( ) continuity, page 39 and
Resistance measurement (Ω)
• Open ( ) continuity[1], or page 41
• Resistance measurement (Ω)
U1233A/U1232A:
Cycles between
• AC A,
• AC Hz, or
Hz U1233A/U1232A: DC A • DC A
Hz Hz AUX
U1231A: Clamp-on AC A (Aux) U1231A:
Cycles between
• Clamp-on DC A (Aux), page 57 and
• AC Hz, or page 62
• Clamp-on AC A (Aux)
U1233A/U1232A:
Cycles between
U1233A/U1232A: DC μA
Hz Hz • AC μA,
AUX U1231A: Auxiliary temperature
• AC Hz, or
measurement
• DC μA
U1231A: N/A
[1] The open continuity test option must be enabled through the Setup menu. To learn more, see “Enable open continuity test
by default” on page 99. The open continuity test option is disabled by default.
Keypad
The operation of each key is explained below. Pressing a key
enables a function, displays a related symbol, and emits a
beep. Turning the rotary switch to another position resets
the current operation of the key.
Click the respective “Learn more” pages in Table 1- 6 for
more information on each function.
Null
Sets the Null/Relative mode. Enters the Hold-Log Recall menu.
Recall Max
• While in Null mode, press again to
Null
Recall • Press Min
to jump to the last record. Press
Max
view the stored reference value that has and hold for more than 1 second to
Min
been saved. The display will return to jump to the first record.
Range
normal after 3 seconds. • Press or
Null
Recall to scroll through each
Auto
page 68
• Pressing while the relative value is
Null
Recall record. and
being displayed will cancel the Null mode. • Press for more than 1 second to
Tri g Hold
Auto Log
page 74
store all records into the multimeter’s
non-volatile memory. Press for more
Esc Shift
this mode.
session.
Trig Hold
Freezes and stores the present reading in the Automatically freezes the present reading
Auto Log display ( mode). once the reading is stable ( mode)
• In Trig Hold-Log mode, press to
Tri g Hold
Auto Log • In Auto Hold-Log mode, the reading is
manually trigger the holding of the next updated automatically once the reading is page 72
measured value. stable and the count setting is exceeded.
• Press for more than 1 second to exit
Tri g Hold
Auto Log • Press for more than 1 second to exit
Tri g Hold
Auto Log
Turns the LCD backlight on for 15 seconds Turns the LED flashlight on or off. page 7 and
(default) or off. page 7
Display screen
The display annunciators of your multimeter are described
in this section. See also “Measurement units” on page 26 for
a list of available measurement signs and notations and
“Analog bar graph” on page 27 for a tutorial on the analog
bar graph located at the bottom of your display screen.
Calibration enabled -
Measuring units -
Measurement units
The available signs and notations for each measurement
function in your multimeter are described in Table 1- 8. The
units listed below are applicable to the primary display
measurements of your multimeter.
Sign/Notation Description
33.34 ,
200 , , ,
[1] The analog bar graph display update rate is approximately 33 times/second for
DC voltage, current, and resistance measurements.
Input terminals
The terminal connections for the different measurement
functions of your multimeter are described in the table
below. Observe the rotary switch position of your multimeter
before connecting the test leads to the connector terminals.
Ensure that the probe accessories are connected to the correct input
WA R N I N G
terminals for the selected measurement function before starting any
measurement.
To avoid damaging this device, do not exceed the rated input limit.
CAUTION
Hz
600 Vrms
Hz
AUX
AUX
Hz Hz
600 Vrms
AUX
Hz Hz
2
Making Measurements
Measuring AC Voltage 32
Measuring DC Voltage 34
Measuring AC/DC mV 34
Using VZLOW for Voltage Measurements 36
Measuring Resistance 39
Testing for Continuity 41
Testing Diodes 45
Measuring Capacitance 49
Measuring Temperature 51
Measuring AC or DC Current 57
Measuring Frequency 62
Agilent Technologies 31
2 Making Measurements
Measuring AC Voltage
Measuring AC Voltage
Set up your multimeter to measure AC voltage as shown in
Figure 2- 2. Probe the test points and read the display.
3 Voltage source
AC
2
Hz
Measuring DC Voltage
Set up your multimeter to measure DC voltage as shown in
Figure 2- 4. Probe the test points and read the display.
Measuring AC/DC mV
You can choose to set the multimeter to measure AC or DC
mV at the rotary positions shown below.
• U1233A:
• U1232A: AUX
Hz
• U1231A: AUX
3 Voltage source
DC
Use the VZLOW function to remove ghost or induced voltages from your
NOTE measurements
Ghost voltages are voltages present on a circuit that should not be
energized. They are usually caused by capacitive coupling between
energized wiring and adjacent unused wiring. The VZLOW function can
remove ghost voltages from your measurements by dissipating the
coupling voltage. Use the VZLOW function to reduce the possibility of false
readings in areas where the presence of ghost voltages are suspected.
4
Energized wiring
primary display. Pressing for the third time will restart the auto
Esc Shift
Auto
identification
Press Esc Shift to lock the initial signal identification Press Esc Shift to lock the initial signal identification
Press Esc Shift to exchange the signal indication to AC V Press Esc Shift to exchange the signal indication to DC V
to restart Auto
Identification
Prolonged use of the VZLOW function will consume the capacity of the
NOTE battery-under-test.
Measuring Resistance
Set up your multimeter to measure resistance as shown in
Figure 2- 9. Probe the test points and read the display.
resistance at the probe tips will be subtracted from all future display
readings.
• Because the multimeter’s test current flows through all possible paths
NOTE between the probe tips, the measured value of a resistor in a circuit is
often different from the resistor’s rated value.
• The resistance function can produce enough voltage to forward-bias
silicon diodes or transistor junctions, causing them to conduct. If this is
suspected, press to apply a lower current in the next higher range.
Range
Auto
3 Resistor
60.00 kΩ <2.3 ± 1 kΩ
600.0 kΩ <23 ± 10 kΩ
6.000 MΩ <131 ± 60 kΩ
60.00 MΩ <131 ± 60 kΩ
Short detected,
beeper sounds
Circuit open
Open detected,
beeper sounds
Circuit
complete
• You can set the beeper to sound and the backlight to flash as a
NOTE continuity indication whether the circuit-under-test is less than (short)
or more than or equal to (open) the threshold resistance.
• The continuity function detects intermittent shorts and opens lasting
as short as 1 ms. A brief short or open causes the multimeter to emit a
short beep and flash.
• You can enable or disable the audible and visual alert via the Setup
menu. See “Changing the continuity test alerts” on page 93 for more
information on the audible and visual alert options.
4
OFF
(open)
ON
(closed )
3
Esc cShift
Testing Diodes
Set up your multimeter to test diodes as shown in
Figure 2- 14. Probe the test points and read the display.
If the beeper is enabled during diode test, the multimeter will beep briefly
NOTE for a normal junction and sound continuously for a shorted junction, below
0.050 V. See “Changing the beep frequency” on page 87 to disable the
beeper.
Measuring Capacitance
Set up your multimeter to measure capacitance as shown in
Figure 2- 17. Probe the test points and read the display.
3 Capacitance
Measuring Temperature
For the U1233A, set up your multimeter to measure
temperature as shown in Figure 2- 19. For the U1231A and
U1232A, refer to Figure 2- 20. Probe the test points and read
the display.
The option to toggle between °C and °F is only available for the U1233A
NOTE model.
Always set the temperature unit display per the official requirements
CAUTION
and in compliance with the national laws of your region.
Esc Shift
Heat source
AUX
Ambient temperature
Available options:
• °C - Temperature measured in °C only.
Range
• °C°F - During temperature measurements, press Auto to
switch between °C and °F.
• °F - Temperature measured in °F only.
Range
• °F°C - During temperature measurements, press Auto to
switch between °F and °C.
4 Press to save the changes. Press and hold
Tri g Hold
Auto Log Esc Shift until
the multimeter restarts
Measuring AC or DC Current
Set up your multimeter to measure AC or DC current as
shown in Figure 2- 24 and Figure 2- 25. Probe the test points
and read the display.
• To measure current, you must open the circuit under test, then place
NOTE the multimeter in series with the circuit.
• Insert the black test lead into the COM terminal. Insert the red test lead
in an input appropriate for the measurement range.
• Set the positive input terminal to the A terminal and set the rotary switch
position to Hz for currents above 600 μA.
• Set the positive input terminal to the mA terminal and set the rotary switch
position to Hz for currents below 600 μA.
Reversing the leads will produce a negative reading, but will not damage
NOTE the multimeter.
Placing the probes across (in parallel with) a powered circuit when a
CAUTION
lead is plugged into a current terminal can damage the circuit you are
testing and blow the multimeter's fuse. This happens because the
resistance through the multimeter's current terminals is very low,
resulting in a short circuit.
3 Voltage source
LOAD
AC
Voltage source
LOAD
DC
2
Hz
3 Voltage source
LOAD
DC
Voltage source
LOAD
AC
2
Hz
Measuring Frequency
Your multimeter allows simultaneous monitoring of real-
time voltage or current with frequency measurements.
Table 2- 2 highlights the functions allowing frequency
measurements in your multimeter.
Legend
Hz Hz Hz
Hz Hz Hz
AUX
Hz Hz
90%
50%
10%
+ Width – Width
Period
4 Voltage source
AC
3
Esc Shift
2
Hz
3
Multimeter Features
Detecting AC Voltage Presence (Vsense) 66
Making Relative Measurements (Null) 68
Capturing Maximum and Minimum Values (MaxMin) 70
Freezing the Display (Trig Hold-Log and Auto Hold-Log) 72
Trig Hold-Log operation 72
Auto Hold-Log operation 72
Recalling Previously Recorded Readings (Recall) 74
Making Scale Transfers (Scale) 76
Agilent Technologies 65
3 Multimeter Features
Detecting AC Voltage Presence (Vsense)
You are advised to test on a known live circuit within the rated AC
WA R N I N G
voltage range of this product before and after each use to ensure that
the Vsense detector works.
Voltage could still be present even if there is no Vsense alert
indication. Do not rely on Vsense detector with shielded wire. Never
touch live voltage or conductor without the necessary insulation
protection or power off the voltage source.
The Vsense detector may be affected by differences in socket design,
insulation thickness, and insulation type.
You are advised to measure voltage by using test leads through the
CAUTION
VZLOW, AC V, or DC V function after using the Vsense function, even if
there is no alert indication.
• Place the top of the multimeter (with the Vsense indicator) close to a
NOTE conductor when sensing for AC voltages (as low as 24 V in the Hi.SE
setting).
• The low sensitivity setting can be used on flush mounted wall sockets
or outlets and various power strips or cords.
• The high sensitivity setting allows for AC voltage sensing on other
styles of recessed power connectors or sockets where the actual AC
voltage is recessed within the connector itself.
2 AC
Voltage source
>1s
1
Esc Shift
Press and hold Esc Shift for more than 1 second to disable the
Vsense function.
Null can be set for both auto and manual range settings, but not in the
NOTE case of an overload.
open (nulls the test lead capacitance), shorted (nulls the test
lead resistance), or across a desired null value circuit.
Press Null
Recallto enable
the Null function
Press again to
Null
Recall
1 Press Max
Min to enable the MaxMin operation.
2 Press again to cycle through the Max, Min, Avg, or
Max
Min
function.
• Changing the range manually will also restart the recording session.
NOTE
• If an overload is recorded, the averaging function will be stopped. is
shown in place of the average value.
• The APO (auto power-off) function is disabled when MaxMin is
enabled.
• The Auto Hold-Log operation monitors the input signal and updates the
NOTE display and, if enabled, emits a beep whenever a new stable
measurement is detected. The multimeter is ready to update the
display once the variation of the reading exceed the setting of the
variation count.
• The annunciator will flash while attempting to acquire a stable
reading.
If the reading value is unable to reach a stable state (when exceeding the
NOTE preset variation), the reading value will not be updated.
second. Records stored through this method remains saved even when
the multimeter is turned OFF or if the battery is replaced.
stored entry.
ii Press to view the next stored entry or press
Range
Auto
Null
Recall
3 Press Null
Recall for more than 1 second to exit the Recall
menu.
4
Multimeter Setup Options
Using the Setup Menu 80
Editing numerical values 81
Setup Menu Summary 82
Setup Menu Items 84
Changing the variation count 84
Enabling and changing the Smooth refresh rate 85
Enabling and changing the voltage alert level 86
Changing the beep frequency 87
Changing the auto power-off (APO) timeout 88
Changing the LCD backlight timeout 89
Adjusting the LCD backlight intensity 90
Enabling the LED flashlight timeout 91
Adjusting the LED flashlight intensity 92
Changing the minimum measurable frequency 93
Changing the continuity test alerts 93
Changing the power-on greeting tone 95
Resetting the Setup items 96
Changing the scale conversion value 96
Enable the AC/DC mV measurement 98
Enable open continuity test by default 99
Changing the temperature unit 99
Agilent Technologies 79
4 Multimeter Setup Options
Using the Setup Menu
Legend Description
Max
Press or
Min
at each menu item to change the
preset settings. The menu item (in the secondary
display) will flash to indicate that you can now
Max change the menu item values.
Min
Max
Press Min
or again to switch between two values,
to cycle through multiple values from a list, or to
decrease or increase a numerical value.
position the cursor over the first numerical digit (most right
digit).
Next, use the and
Null
Recall
Range
Auto to move the cursor to the other
numerical digit(s).
• Press Null
Recall to move the cursor to the left, and
• Press Range
Auto to move the cursor to the right.
When the cursor is positioned over a digit, use the Max
Min and
keys to change the numerical digit.
• Press Max
Min to increment the digit, and
• Press to decrement the digit.
When you have completed your changes, save the new
numerical value by pressing . (Or alternatively, if you
Tri g Hold
Auto Log
Set the multimeter’s Auto Hold-Log variation count from page 72 and
(001 to 999) counts
1 count to 999 counts. Default is 50 counts. page 84
(3.2, 3.4, 3.8, 4.2) kHz Set the multimeter’s beep frequency (3.2 kHz, 3.4 kHz,
page 87
or –.– (off) 3.8 kHz, 4.2 kHz, or off). Default is 3.8 kHz.
Set the LCD backlight brightness (Lo, 02, 03, ME, 05, 06, or page 7 and
Lo, 02, 03, ME, 05, 06, or Hi
Hi). Default is Hi. page 90
Set the LED flashlight timeout period from 1 to 99 seconds
page 7 and
(01 to 99) s or disabled (1 minute, 39 seconds). Enable the LED flashlight timeout
page 91
by selecting E (enabled). Default is disabled (15.d).
Set the LED flashlight brightness (Lo, 02, 03, ME, 05, 06, or page 7 and
Lo, 02, 03, ME, 05, 06, or Hi
Hi). Default is Hi. page 92
Set the minimum measurement frequency (0.5 Hz or page 62 and
(0.5 or 5.0) Hz
5.0 Hz). Default is 0.5 Hz. page 93
Press Press
Press Press
Press Press
Press Press
NOTE (see page 13). This method, however, is temporary and Smooth will be
turned off when the multimeter’s power is cycled.
Press Press
Press Press
Press Press
Press Press
030.d V
V(oltage Alert) (1 to 660).(d or E) V
(disabled)
Press Press
Press Press
Press Press
Press Press
Press Press
Press Press
Press Press
Press Press
(15.E) minutes
A(PO) (01 to 99).(d or E) minutes
(enabled)
Press Press
Press Press
Press Press
Press Press
(15.E) seconds
b(acklight) (01 to 99).(d or E) seconds
(enabled)
Press Press
Press Press
Press Press
Press Press
Press Press
Press Press
Press Press
(15.d) seconds
t(orchlight) (01 to 99).(d or E) seconds
(disabled)
Press Press
Press Press
Press Press
Press Press
Press Press
Press Press
Press Press
Press Press
Press Press
Press Press
Press Press
Press Press
Press Press
Press Press
Press Press
multimeter will beep once and return to the first Setup item.
Press Press
Press
Press
Press Press
Press Press
Hz
• U1231A: AUX
mV on or oFF oFF
• When this Setup item is enabled, the original functions of the rotary
NOTE switch positions shown above are disabled and replaced by AC/DC mV
measurements.
• For AC/DC mV measurements, the measurement range is fixed at
600 mV and the input impedance is typically 10 MΩ.
• Press to switch between DC mV, AC mV, and frequency
Esc Shift
measurements.
Press Press
Press Press
Always set the temperature unit display per the official requirements
CAUTION
and in compliance with the national laws of your region.
Press Press
Press Press
Press Press
5
Characteristics and Specifications
Product Characteristics 102
Specification Assumptions 103
Measurement Category 104
Measurement category definition 104
Electrical Specifications 105
DC specifications 105
AC specifications 108
Capacitance specifications 109
Temperature specifications 110
Frequency specifications 111
Frequency sensitivity specifications 111
Scale transfer (mV) 112
Display update rate (approximate) 113
Product Characteristics
Product characteristics specified in the table below are applicable for both
NOTE U1231A, U1232A, and U1233A models unless stated otherwise.
POWER SUPPLY
Battery type:
• 4 × 1.5 V AAA Alkaline battery (ANSI/NEDA 24A or IEC LR03), or
• 4 × 1.5 V AAA Zinc Chloride battery (ANSI/NEDA 24D or IEC R03)
Battery life:
• 500 hours typical (based on new Alkaline batteries (1000 mAH) for DC
voltage measurement, with backlight and flashlight disabled)
• Low battery indicator will flash when the battery voltage drops below 4.4 V
(approximately)
POWER CONSUMPTION
450 mVA maximum (with backlight and flashlight enabled)
FUSE
10 × 38 mm 11 A/1000 V fast-acting fuse
DISPLAY
Liquid crystal display (LCD) (with maximum reading of 6600 counts)
FLASHLIGHT
Cool white LED (5500 K typical); luminous intensity from 2240 mcd to 5600 mcd
OPERATING ENVIRONMENT
• Operating temperature from –10 °C to 55 °C, 0% to 80% RH
• Full accuracy up to 80% RH for temperatures up to 30 °C, decreasing linearly
to 50% RH at 55 °C
• Altitude up to 2000 meters
• Pollution degree 2
STORAGE COMPLIANCE
–40 °C to 60 °C, 0% to 80% RH without batteries
SAFETY COMPLIANCE
• IEC 61010:1-2010/EN 61010-1:2010
• USA: UL 61010-1 (3rd Edition)
• Canada: CSA C22.2 No. 61010-1:2012
MEASUREMENT CATEGORY
CAT III 600 V
DIMENSIONS (W × H × D)
86 × 169 × 52 mm
WEIGHT
• U1232A and U1233A: 371 grams (with batteries and holster)
• U1231A: 365 grams (with batteries and holster)
WARRANTY
Please refer to http://www.agilent.com/go/warranty_terms
• Three years for product
• Three months for product’s standard accessories, unless otherwise specified
• Please take note that for the product, the warranty does not cover:
• Damage from contamination
• Normal wear and tear of mechanical components
• Manuals, fuses, and standard disposable batteries
CALIBRATION CYCLE
One year
Specification Assumptions
• Accuracy is given as ±(% of reading + counts of least
significant digit) at 23 °C ± 5 °C, with relative humidity
less than 80% RH.
• AC V and AC A specifications are AC coupled, true rms
and are valid from 5% of range to 100% of range.
Measurement Category
The Agilent U1231A/U1232A/U1233A Handheld Digital
Multimeters have a safety rating of CAT III, 600 V.
Electrical Specifications
Specification assumptions are given on page 103.
NOTE
DC specifications
Table 5-1 DC specifications
600 V
0.1 V 2% + 3 2% + 3 2% + 3 - - 3 kΩ
(VZLOW)[2]
Notes for DC voltage specifications:
1 The accuracy of the 600 mV range is specified after the Null function is used to subtract the thermal effect (by shorting the
test leads).
2 For VZLOW (low input impedance) measurements, autoranging is disabled and the multimeter’s range is set to 600 V in the
manual ranging mode.
100 nA
60 MΩ [5] 0.01 MΩ 1.5% + 3 1.5% + 3 1.5% + 3 - -
//10 MΩ
1 Overload protection: 600 Vrms for short circuits with <0.3 A current.
2 Maximum open voltage is <+3 V
3 Built-in buzzer beeps when the resistance measured is less than 23 Ω ± 10 Ω. The multimeter can capture intermittent
measurements longer than 1 ms.
4 The accuracy of the 600 Ω to 6 kΩ range is specified after the Null function is used to subtract the test lead resistance and
thermal effect (by shorting the test leads).
5 For the ranges of 6 MΩ and 60 MΩ, the RH is specified for <60%.
1 Overload protection: 600 Vrms for short circuits with <0.3 A current.
2 Built-in buzzer beeps continuously when the voltage measured is less than 50 mV and beeps once for forward-biased
diode or semiconductor junctions measured between 0.3 V and 0.8 V (0.3 V ≤ reading ≤ 0.8 V).
3 Open voltage for diode: <+3 V DC
4 The maximum display for diode measurements is 2100 counts.
1 Overload protection for 60 μA to 600 μA range: 600 Vrms for short circuits with <0.3 A current.
2 Overload protection for 6 A to 10 A range: 11 A/1000 V; 10 × 38 mm fast-acting fuse.
3 Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals >10 A to 20 A
for 30 seconds maximum. After measuring currents >10 A, cool down the multimeter for twice the duration of the
measured time before proceeding with low current measurements.
4 DC current range of 0.6 mA to 1 mA is not measureable on the U1232A and U1233A models.
AC specifications
Table 5-2 AC specifications
1 Overload protection: 600 Vrms. For millivolt measurements, 600 Vrms for short circuits with <0.3 A current.
2 Input impedance: 10 MΩ (nominal) in parallel with <100 pF.
3 VZLOW input impedance: 3 kΩ (nominal).
60 μA[2] 0.01 μA 1.5% + 3 - <2.5 V
Capacitance specifications
Table 5-3 Capacitance specifications
1 Overload protection: 600 Vrms for short circuits with <0.3 A current.
2 The accuracy of for all ranges is specified based on a film capacitor or better, and after the Null function is used to
subtract the residual values (by opening the test leads).
3 The maximum display is 1200 counts.
Temperature specifications
Table 5-4 Temperature specifications
Accuracy
Thermal type Range Resolution
U1233A
1 The specification above is specified after the multimeter has been left stationary in the same operating environment for
1 hour at least. If the unit is exposed during storage in a high humidity (condensing) environment, ensure that the
multimeter has been in the same operating environment for 2 hours at least.
2 The accuracy does not include the tolerance of the thermocouple probe.
3 Do not allow the temperature sensor to contact a surface that is energized above 30 Vrms or 60 V DC. Such voltages
poses a shock hazard.
4 Ensure that the ambient temperature is stable within ±1 ºC and that the Null function is used to reduce the test lead’s
thermal effect and temperature offset. Before using the Null function, set the multimeter to measure temperature
without ambient compensation ( ) and keep the thermocouple probe as close to the multimeter as possible (avoid
contact with any surface that has a different temperature from the ambient temperature).
5 When measuring temperature with respect to any temperature calibrator, try to set both the calibrator and multimeter
with an external reference (without internal ambient compensation). If both the calibrator and multimeter are set with
internal reference (with internal ambient compensation), some deviations may show between the readings of the
calibrator and multimeter, due to differences in ambient compensation between the calibrator and multimeter. Keeping
the multimeter close to the output terminal of calibrator will help reduce the deviation.
6 The temperature calculation is specified according to the safety standards of EN/IEC-60548-1 and NIST175.
Frequency specifications
Table 5-5 Frequency specifications
1 Overload protection: 600 V; input signal is <20,000,000 V × Hz (product of voltage and frequency).
Table 5-6 Frequency sensitivity and trigger level specifications for voltage measurements
600 V 50 V 50 V 50 V
1 Maximum input for specified accuracy, refer to “AC specifications” on page 108.
60 μA 30 μA 30 μA
600 μA 30 μA 30 μA
6A 0.5 A 0.5 A
10 A 0.5 A 0.5 A
1 Maximum input for specified accuracy, refer to “AC specifications” on page 108.
Accuracy
Range Resolution
U1231A U1232A U1233A
1 Overload protection: 600 Vrms for short circuits with <0.3 A current.
2 The accuracy of the DC 600 mV range is specified after the Null function is used to subtract the thermal effect (by
shorting the test leads).
3 Input impedance: 10 MΩ (typical)
Times/second
Function
U1231A U1232A U1233A
AC V (V or mV) 5 5 5
DC V (V or mV) 5 5 5
AC V/DC V (VZLOW) 1 1 1
Ω 5 5 5
Diode 5 5 5
DC A (μA, mA, or A) - 5 5
AC A (μA, mA, or A) - 5 5
Contact us
To obtain service, warranty, or technical
assistance, contact us at the following
phone or fax numbers:
United States:
(tel) 800 829 4444 (fax) 800 829 4433
Canada:
(tel) 877 894 4414 (fax) 800 746 4866
China:
(tel) 800 810 0189 (fax) 800 820 2816