Suzuken Kenz Ecg 103 User Manual
Suzuken Kenz Ecg 103 User Manual
> FEED (2) A/D converter and power supply testing. Example ‘An A/D number will be displayed in the upper row on the sTD 200H LCD. If the number is from 100H to 300H, the A/D converter is normal. Power voltage will be displayed in the lower row on the POWER, 234H bee. oo For kattory operation, a number loss than 200H means the battery neads recharging, [Ei Printer Check Program ‘Test printing is performed by holding down the FEED key and chen pressing the power bution, * Test printing will continue until the power button is pressed, 4Troubleshooting Flow Chart This section concerns troubleshooting procedures to correct poor ECG recordings. ECG tracing is unclear Is paper sot property? |__| Set paper properly YES, 1 Repaired? — NO YES Is specified recording paper i recor Ivo L_Use sbecities paver Yes be - Repaired? Jo. Yes 5 Call service engineerAC interference ‘Ave any X-ray, ultrasonic or other electronic instruments nearby? YES NO ‘ Move the electrocardiograph away from tho instruments or turn off instruments. Is electrocardiograph properly grounded? VES 1 ‘Are all electrodes screwed tightly enough to their leads? YES 1 Is dip switch No. 5 of Hum/Drift filter set off? wo Repaired? ke Wo +L sroune properiy. wo Repaired? ve bag LTiahten screw ty. —p| Pepaires? a WoL Set on ap switch No.6 YES Call service engineer. END |_——___ sired? L— Repoites YES 16Vertical wandering of ECG baseline 1s patient moving or talking? a Keep patient stil NO = Repaired? NO ves Are patient's skin and electrodes: Wash skin and electrodes with | clean? [No] either soap and water or cotton saturated with alechol. ves “ wal Repairea? aE | | Is dip switch No. 5 of Hum/Drift 5 Hee ~ Sot on dip owitch No. 5. YES, | | red? | wo Repaired? aE Call service engineer.EOG is disturbed by fine and irregular waves such 2s EMG, etc. | nso? Is patient tense’ 7 NO ——-| Put the patient at ease by gently explaining how simple and harmless the ECG procedure is. See Repaired? I. WoL ees YES Is room temperature too low? ——+| Maintain room temperature at a ves lovel comfortable to patient | nO | Repairea? | wo ere Yes ‘Are patient's skin and electrodes __| Wash skin and electrode with either are clean? NO soap and water or cotton saturated + with alechol. [ves | Repaired? NoL See YES 1 Is each electrode properly attached to the patient? VES Properiy attach electrodes. Repaired? |__| Re YES Is muscle tremor filter on? NO —+{ Set on MF 1 or MF 2. YES Call service engineer. END | —_{ Repaired? e YES 18Paper does not feed. 15 paper magazine propery set” | =| Reset paper magazine properly. YES - ? no L Repaired = ? ——+| Load new paper. Is paper used up’ ves pap NO ? a LARS YES 's specified recording paper used? |= --[ Use specified paper YES) — —_| Ropaired? L_—__——+ NO VES Call service engineer. end |LCD shows no display. |s power bution pressed on? w Press power button on. YES Repaired? no Looe ves poration by batter . — — partion by battory? | YES Is battary fully charged?” |_—+[ Charge fully Is power cable properly ING connected to AC YES eB Power source connector? NO | Connect it properly 1 | wal _Repairea? YES Call service engineer. ||? END pa10 Circuit Block Diagram ay / T / nou L aff t Hof 4 fHouwrnesu] / _. ji ONHOUINS| 1, ‘4 ty 2 a Z oa oy aa oe Pe o| | Bin ‘o. Or a TT nas vos “es aes cnvoe won 1 Jan wane we L = Pe 1 on FF 1 ---- Beis bien em 1 [ ava tare : J met ET oa ee g 4 uous SAS et i sed oi [Thc ' =f ST Ross NORIOr= Specifications and standard accessories SPECIFICATIONS Leeds ‘Standard 12 lead | input impedance ‘over 20 MO | CRM iE ‘over 120.68 Frequency Response 1.05-100 Hz (witin -3 8) Time Gonstant over 32 sec. - Filters Digital fiter A ~3.d8 (60 Hz) oF more Muscle 1 - 3 8 (95/45 Hz) or more Muscle 2 :-5 dB (35/45 Hz) or more Drift" --8d8 (05 H2) or mora Calibration 5, 10, 20 mmimnV in manual mode _| Automatically solscts 5 mm/m\ or 10 mm/m\ | osvimv 10 mm/0.8V co ‘Automatic centering Recerding Mode ‘AUTO 1: Norma automatic recording ‘AUTO 2 : Reverse-recording AUTO 3 Track-ecarding MANUAL Master “Autores record lead and tne rterals. | Recoraing Paper Rll paper, 69 mn » 90 M er 50 mm » 20 M 7 {use han sensitity ecorsing papen Rosording Spoed 25 mr/e0s. and 50 mm/sec. Recording Time Select 3, 4,5,6 7,8, 8 oF 10 seo. in automatic mode. ~ Power Supply ‘AG 100, 120, 220 240V ( 10%) (Fixed before shipping) 1G builtin rechargeable Ni-Cd battery (Gption) Classification (EC) Class 1 type OF Dimensions 205 |W) x 210 (0) «80 (H) mm ‘Weight Approx. 28 kg STANDARD ACCESSORIES Patient cable 1 Pe ECG clip-on electrode (4 pesiset) set Chest electrode, suction for adult (6 pes/set) 1 set ECG electrode cream (70 a) i pe Power cable 7 Tp “Grounding wire pe Fuse 2 pcs ‘Accessories bag TPs Recording paper i 1 roll Instruction manual 1p 22SUZUKEN CO.,LTD. HEAD OFFICE : 8, HIGASHIKATAHA-CHO, HIGASHI-KU, NAGOYA 461-8701, JAPAN TEL (02) 950-6327 FAX(052) 962-7440