You are on page 1of 112

1

Chrom-Ed Book Series

Raymond P. W. Scott

LIQUID CHROMATOGRAPHY

DETECTORS

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
2

COPYRIGHT @2003 by LIBRARY4SCIENCE, LLC


ALL RIGHTS RESERVED

Neither this book or any part may be reduced or transmitted in any


form or by any means, electronic or mechanical , including
photocopying, microfilming, and recording or by any information
storage and retrieved system without permission in writing from the
publisher except as permitted by the in-user license agreement.

World Wide Web


http://www.library4science.com/

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
3

Contents
Introduction .............................................................................................................1
Detector Specifications.....................................................................................2
Dispersion in Connecting Tubes................................................................4
Low Dispersion Tubing...............................................................................7
Dispersion in the Detector Sensor Volume.......................................... 10
Apparent Dispersion from Detector Sensor Volume....................... 13
Dispersion Resulting from the Detector Time Constant................... 16
LC Detectors Based on Refractive Index Measurement............................ 19
The Refractive Index Detector .................................................................... 20
The Angle of Deviation Method............................................................. 21
The Fresnel Method .................................................................................. 23
The Christiansen Effect Detector ........................................................... 26
The Interferometer Detector ................................................................... 29
The Thermal Lens Detector .................................................................... 33
The Dielectric Constant Detector........................................................... 35
The UV Detectors ............................................................................................... 40
The UV Absorption Detectors..................................................................... 40
The Fixed Wavelength UV Detector..................................................... 42
The Multi–Wavelength UV Detector ......................................................... 48
The Multi–Wavelength Dispersive UV Detector ................................ 49
The Diode Array Detector ........................................................................... 52
The Fluorescence Detector........................................................................... 57
The Single Wavelength Excitation Fluorescence Detector .............. 59
The Multi Wavelength Fluorescence Detector......................................... 62
Transport Detectors............................................................................................ 66
The Moving Wire Detector .......................................................................... 67
The Chain Detector........................................................................................ 69
The Modified Moving Wire Detector......................................................... 70
The Disc Detector........................................................................................... 76
The Evaporative Light Scattering Detector............................................. 78
Liquid Light Scattering Detectors.............................................................. 81
The Low Angle Laser Light Scattering Detector.............................. 83
The Multiple Angle Laser Light Scattering (MALLS) Detector.... 85
The Electrical Conductivity Detector......................................................... 88
The Electrochemical Detector ..................................................................... 93
Electrode Configurations.......................................................................... 94
Electrode Construction............................................................................. 96
The Multi–Electrode Array Detector.......................................................100
References...........................................................................................................105
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
4

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
1

Introduction

Although chromatography was discovered late in the 1890s its


development was almost negligible until the 1940s and this was largely
due to the lack of an inline sensitive detector. The first, effective inline
liquid chromatography (LC) detectors were the refractive index detector
reported by Tiselius and Claesson (1) in 1942 and the conductivity
detector described by Martin and Randall (2) in 1951. These two devices
should have evoked a growth in LC development, but, in the early fifties,
gas chromatography (GC) was invented which completely eclipsed the
development of LC. It was not until the early 1960s that the renaissance
of LC took place, initially based on the use of the refractive index of
Tiselius and Claesson. Although a significant number of GC detectors
were developed over two or three years, the development of LC
detectors was much slower, largely due to the fact that low
concentrations of solute in a liquid do not change the properties of a
liquid nearly as much as they do a gas. In fact, the development of LC
detectors was gradual and arduous.

In a similar way to the development of GC there has been a continuous


interaction between improved detector performance and improved
column performance. Initially, separations monitored by detectors with
improved sensitivity permitted a precise column theory to be developed
and experimentally substantiated. This allowed new columns to be
designed with reduced dispersion and higher efficiencies. The improved
efficiencies, however, produced small volume peaks, small, that is,
compared with the volume of the detector sensor and the dispersion that
took place in the conduits of the detector system..

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
2

As a consequence, the ultimate efficiency obtainable from the column


was determined by the geometry of the fluid conduits of the detector
and not its sensitivity. This provoked detector redesign, with smaller
sensor volumes, different geometry and shorter connecting tubes
between the column and sensor. In turn, these modifications allowed
much smaller particles to be used in the column resulting in even lower
column dispersion and higher efficiencies. In this way, just as in GC,
detector design and column design have interacted over the years to a
point where the performance of LC columns are now commensurate
with those of GC columns.

Unfortunately, even today, there is no LC detector that has an equivalent


performance to the flame ionization detector (FID) used in GC. In
general, LC detectors have sensitivities of two to three orders of
magnitude less than their GC counterparts and linear dynamic ranges
one to two orders of magnitude lower. Only highly specific LC detectors
have sensitivities that can approach those of GC detectors.
Detector Specifications

Detector specifications are like those for GC detectors and are listed as
follows,

1. Dynamic Range
2. Response Index or Linearity
3. Linear Dynamic range
4. Detector Response
5. Detector Noise Level
6. Detector Sensitivity or Minimum Detectable
Concentration
7. Total System Dispersion
8. Sensor Dimensions
9. Detector Time Constant
10. Pressure Sensitivity
11. Flow Sensitivity
12. Operating Temperature Range

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
3

In general the specifications are the same for both GC and LC detectors
with the exception of detector dispersion. Although, detector dispersion
has a minimal effect on the resolution in GC separations, detector
dispersion can actually destroy a separation achieved in an LC column if
the system is not designed correctly. Dealing with the other
specifications, the dynamic range and linear dynamic range are the same
as those defined in book 4. The response index, the measure of detector
linearity, can also be determined in exactly the same way, either by the
incremental method of calibration, or the logarithmic dilution method. In
the logarithmic method of calibration, mobile phase, now a liquid, is
passed continuously through an enclosed stirred vessel containing a
known mass of solute, the eluent passing directly into the detector. The
logarithm of the detector output is plotted against the logarithm of the
calculated solute concentration and the magnitude of the response index
determined from the slope of the curve in the manner described in book
4.

The response, noise and sensitivity are measured in exactly the same way
as for GC detectors. Pressure sensitivity and pressure tolerance have a
more important significance in LC as in multidimensional LC, the
detector may be situated between two or more columns and thus must
tolerate pressures up to the input pressure (e.g., several thousand p.s.i).
Pressure sensitivity and flow sensitivity are also more important in LC
due to the relatively high pressures involved and the sensitivity of many
sensors to pressure changes (e.g., the refractive index detector and the
UV detector). However, LC columns have a high impedance to flow and
so pressure pulses are often smoothed out in the column and do not
reach the detector. Dispersion that takes place in a column is very
important and will be dealt with in some detail.

Dispersion in Detector Sensors

There are three sources of dispersion in LC detector sensors,

1. Dispersion from Connecting Tubes(Newtonian)


2. Dispersion from Sensor Cell Volume (Newtonian)
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
4

3. Dispersion from Sensor Cell Volume ( Dilution)

Each of these sources of dispersion are controllable by careful sensor


design and employing appropriate cell geometry.
Dispersion in Connecting Tubes

The dispersion that takes place in an open tube results from the parabolic
velocity profile that occurs under conditions of Newtonian flow, (i.e.
when the velocity is significantly below that which produces turbulence).
Under condition of Newtonian flow, the distribution of fluid velocity
across the tube adopts a parabolic profile as shown in figure 1.

The velocity at the walls is virtually zero and that at the center a
maximum. This situation is depicted diagramatically in Figure 1.

Newtonian Flow

Tube Walls

Mobile Phase
Velocity

Parabolic Velocity
Profile

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
5

Figure 1. The Parabolic Velocity Profile of a Solute Band Passing


Through a Tube
Due to the relatively high velocity at the center of the tube and the very
low velocity at the walls, the center of the band of solute passing down
the tube will move ahead of that situated at the walls. The resulting effect
of band dispersion is depicted in figure 2.

Tube Walls
Initial Band Dispersed Band
Width

Figure 2. Band Dispersion Resulting from Newtonian Flow

The dispersion in open tubes was examined by Golay (3) and Atwood
and Golay (4) and experimentally by Scott and Kucera (5) and
Lochmuller and Sumner (6). The variance per unit length of an open
tube (H) according to Golay is given by
2D m r 2u
H = +
u 24 D m
where (Dm) is the diffusivity of the solute in the mobile phase,
(u) is the linear velocity of the mobile phase,
and (r) is the radius of the tube.

At relatively high velocities (i.e., at velocities much greater than the


optimum velocity of the tube, which will usually be true for all
connecting tubes)
r 2u
H =
24 D m

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
6

Furthermore, Q = p r 2u

where (Q) is the flow rate through the tube.

Q
Thus, H =
24 p D m
Now, (H) is the variance per unit length of the tube but a more useful
parameter to the analyst is the volume variance (sv2). This can be
derived using the relationship predicted by the Plate Theory (see book
6).
2

s 2v =
( V0 ) 2
=
( p r2 l ) =
p 2r 4l 2
n n n
where (Vo) is the volume of the tube
and (l) is the length of the tube
l p r4 l Q
, consequently s 2v 2 4
Now H = = p r lH =
n 24 D m
Thus, expression for the volume standard deviation (s v(l) ) for tubes of
different length is
Ê pl Q ˆ 0.5 2
sv = Á ˜ r (1)
Ë 24 Dm ¯

Employing equation (1) it is possible to calculate the value of (s v(l) ) for


a range of cylindrical connecting tubes of different radii and different
lengths.

Table 1 Standard Deviation of Connecting Tubes of Different


Sizes

Connecting Tubes for Liquid Chromatography


Standard Deviation of Tube Dispersion
Tube Diameter l=1 cm l=2 cm l=5 cm l=10 cm l=15 cm
0.001 in, 0.00254 cm 22.3 nl 31.5 nl 49.9 nl 70.5 nl 86.4 nl
0.002 in, 0.00508 cm 47.6 nl 67.3 nl 106.4 nl 150 nl 184.4 nl
0.003 in, 0.00762 cm 107 nl 151.3 nl 239.2 nl 0.34 ml 0.41 ml
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
7

0.005 in, 0.01270 cm 298 nl 421 nl 0.67 ml 0.94 ml 1.15 ml


0.010 in, 0.02540 cm 1.19 ml 1.68 ml 2.66 ml 3.76 ml 4.61 ml

(Dm) is taken as 2 x 10-5 cm2sec-1 and the flow rate at 0.5 ml/min. All
values are fairly typical for the normal operation of the chromatographic
system near optimum conditions.

It is seen from table 1 that the effect of dispersion in connecting tubes is


large due to the very low diffusivity of solutes in liquids. It will be shown
in book 8 that for the successful use of microbore columns (columns less
than 2 mm I.D.) tube dispersion needs to reduced to about 80 nl. Again
assuming that to minimize the chance of tube blocking, the limiting
minimum I.D. for the connecting tube is made to be 0.003 in (and tubes
of this diameter will still easily block) then the connecting tube must be
less than 1 cm long. It is clear that the length of the connecting tube
between column and detector must be reduced to an absolute minimum.
If the tubing diameter is reduced further and the column diameter is
increased, then longer tubing lengths may be possible. Alternatively, the
resolution of the early peaks can be sacrificed in favor of later eluting
peaks which will also allow longer connecting tubes to be used. These
techniques to reduce the effect of connecting tube dispersion in LC are
common with most manufacturers. The simple solution of designing the
chromatographic system such that the detector sensor is situated very
close to the end of the column does not appear to be considered a
practical option.

Low Dispersion Tubing

In order to avoid dispersion in mobile phase conduits a number of


attempts to design low dispersion tubing has been reported. The first
attempt was by Halasz et al. (8), who crimped and bent the tube into
different shapes to interrupt the Newtonian flow and introduce radial
flow within the tube. His devices had limited success and the tubes had a
tendency to block very easily.
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
8

In 1978 Tijssen (9), developed a theory to describe the radial flow that
was induced into coiled tubes by the continual change in direction of the
fluid as it flowed round the spirals (his theory will be considered in detail
in Book 9). Tijssen found that by coiling the tubes significantly reduced
dispersion, particularly at high flow rates However, the coils were a little
clumsy to form as the radius of the coil was required to be less than 3
times the internal radius of the tube for optimum performance. A more
practical system was introduced by Katz and Scott (10), who developed
a serpentine form of connecting tube that met the requirement that the
radius of the serpentine bends (a/2 in the diagram) was less that 3 times
that of the internal radius of the tube. A diagram of a serpentine tube is
shown in figure 3.

Figure 3 Low Dispersion Tubing

During passage through the tube, the direction of mobile phase flow
changed by 180o as it passed from one serpentine bend to another.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
9

0.5
Variance per Unit Length ( ml 2 /cm) Straight Tube 0.007 in !I.D.

1.0

1.5
Serpentine Tube 0.10 in I.D.

0 1.0 2.0 3.0 4.0


Flow rate (ml/min)

Figure 4 Graphs of Peak Variance against Flow Rate for Straight


and Serpentine Tubes
This violent change in direction resulted in extensive radial flow which
aided radial transfer and greatly reduced the dispersion. This effect is
clearly shown by the curves relating the variance against flow rate for
straight and serpentine tubes shown in figure 4. It is seen that at high
flow rates, the dispersion is reduced by over an order of magnitude by
the serpentine tubing relative to the dispersion that occurred in the
straight tube.

Despite the apparent advantages, low dispersion serpentine tubing


appears to have been employed in only one commercial LC detector. It
should be pointed out that any conduit system that has low dispersion
will also provide very fast heat transfer rates. Serpentine tubing has been
also used in commercial column ovens to heat the mobile phase rapidly
to the column oven temperature before it enters the column. The
serpentine tubing allows effective heat exchange with a minimum of heat
exchanger volume to distort the concentration profile of the solvent

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
10

gradient. The different forms of dispersion profiles that are obtained from
various types of connecting tubes used in LC are shown in figure 4.

Dispersion Peak from Dispersion Peak from


Serpentine Tube Coiled Tube

Dispersion Peak from Dispersion Peak from


Straight Tube Straight Tube
(0.25mm I.D.) (0.18mm I.D.)

Figure 4 Dispersion Profiles from Different Types of Tube

These dispersion curves were obtained using a low dispersion UV


detector (cell volume, 1.4 ml) and a sample valve with a 1 ml internal loop.
All tubes were of the same length and carried the same mobile phase at a
flow rate of 2 ml/min. employed. The peaks were recorded on a high
speed recorder. The peak from the serpentine tubing is seen to be
symmetrical and has the smallest width. The peak from the coiled tube,
although still very symmetrical is the widest at the points of inflexion of
all four peaks. The peak from the straight tube 0.25 mm I.D. is grossly
asymmetrical and has an extremely wide base width. The width and
asymmetry is reduced using a tube with an I.D. of 0.18 mm but serious
asymmetry remains. Where the design of the chromatograph precludes a
close proximity between the column and the detector, the use of low
dispersion serpentine tubing may be a satisfactory alternative.

Dispersion in the Detector Sensor Volume

The finite nature of the detector sensor volume can cause peak
dispersion and contribute to the peak variance by two processes. Firstly
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
11

there will be dispersion resulting from the Newtonian flow of fluid


through the cell in much the same manner as the flow of a viscous fluid
through an open tube. This will furnish a variance similar in form to that
predicted by Golay but, as the tube length is small and the tube length to
radius ratio much larger than that from a connecting tube, a different
equation is necessary to describe the dispersion effect.

Secondly, there will be a peak spreading which results from the finite
volume of the sensor. If the sensor has a significant volume, the
concentration measured will not be that entering the detector cell but the
average concentration throughout the cell. Thus, the true profile of the
peak can not be monitored. If the sensor volume is significantly smaller
than the peak volume the effect will merely give the peak an apparent
dispersion. However, if the sensor volume becomes of the same order of
magnitude as the peak volume, then the peak profile will be distorted
and resolution will be lost. In the extreme case two peaks could coexist
in the sensor at one time and only a single peak will be represented.

The effect of viscous flow on dispersion will first be considered.


Dispersion in Detector Sensors Resulting from Newtonian Flow
Most sensor volumes are cylindrical in shape, are relatively short in
length, and have a relatively small length-to-diameter ratio. The small
length-to-diameter ratio is in conflict with the premises assumed in the
development of the Golay equation for dispersion in an open tube.
Atwood and Golay (11) extended the theory of dispersion in open tubes
to tubes having small length-to-diameter ratio. The theory is complex
and not relevant here as, if appropriate cell design is employed, the
dispersion from viscous sources will be negligible. Nevertheless, the
effect on solute profiles is shown in figure 5.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
12

n=30

10
1 3
Sample Concentration

0.3
0.1

0.03

0.01

0 1.0 2.0 3.0


Normalized Elution Volume V/V T

Figure 5 Elution Curves Presented as a Function of the


Normalized Tube Length
It is seen that serious peak distortion can occur but as the length of the
sensor cell is increased the distortion is reduced, but the dispersion
increases. However, if the conduits to the cell are appropriately designed
to produce secondary flow in the cell, then the parabolic velocity profile
is destroyed, and the dispersion and peak distortion eliminated. The
manner of entry of the mobile phase from the connecting conduits are,
consequently, designed to produce this secondary flow and the manner
in which this is achieved is shown in figure 6.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
13

Optical
Window

Inlet from Column

Exit
Mobile Phase

To Waste

Detector
Cell

Optical
Window

Figure 6 The Design of a Modern Absorption Cell

The Newtonian flow is distorted by the manner in which the inlet and
outlet conduits are connected to and from the cell. Mobile phase enters
the cell at an angle that is directed at the cell window. It follows, that the
mobile phase flow has to virtually reverse its direction to pass through
the cell producing a swirling action which introduces strong radial flow
and disrupts the Newtonian flow. The effect also occurs at the exit end of
the cell. The flow along the axis of the cell now must reverse its direction
to pass out of the port which is accomplished by attaching the exit
conduit at an angle to the axis of the cell. Employing this type of entry
and exit connections eliminates dispersion resulting from viscous flow.
Apparent Dispersion from Detector Sensor Volume

The detector can only respond to the average value of the solute
concentration throughout the sensor cell. At the extreme, if the sensor
cell volume was large enough to contain two closely eluted peaks the
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
14

response would appear as a single peak, albeit very distorted in shape.


This extreme condition rarely occurs, but serious peak distortion and loss
of resolution can quite often happen. This will be evident when the
sensor volume is of the same order of magnitude as the peak volume.
The problem can be particularly severe when columns of small diameter
are being used. The situation is depicted in figure 7.

1.25E+00 Column Length 3 cm


Column Diameter 3 mm
Particle Diameter 3 mm
Column Efficiency 5000 plates
Concentration in Arbitrary Units

1.00E+00 k' of Solute 2

7.50E-01

5.00E-01

2.50E-01

0.00E+00
-0.02 -0.01 0 0.01
Volume Flow in ml
Cell Volume
2.5 ml

Figure 7 Effect of Sensor Volume on Detector Output

Consider the elution profile of a peak eluted from a column 3 cm long, 3


mm I.D. packed with particles 3 m in diameter as shown in figure 7. If
the peak is eluted at a (k') of 2, from figure 7 it is seen that the peak
width at the base is about 14 ml wide. The sensor cell volume is 2.5 ml
and the portion of the peak in the cell is depicted in the figure. The
detector will obviously respond to the mean concentration of the slice
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
15

contained in the 2.5 ml sensor volume. It is also clear that, if the sensor
volume is increased, a larger part of the peak will be contained in the
cell. As a consequence, the output will be an average value of an even
larger portion of the peak which will produce serious peak distortion.
The effect of a finite sensor volume can be easily simulated with a
relatively simple computer program and the output from such a program
is shown in figure 8.

Column Length 15 cm
Column Diameter 1 mm
Particle Diameter 5 mm
(k') of first eluted peak 1.0
Solute Concentration (normalized)

0.1 ml
1.0 ml
2.0 ml
3.0 ml

5.0 ml

Volume Flow of Mobile Phase

Figure 8 The Effect of Detector Sensor Volume on the Resolution


of Two Solutes

The example given, is, by far, not the worst case scenario, but is a
condition where the detector sensing volume has a very serious effect on
the peak profile and, consequently, the resolution. The small bore column
produces peaks having relatively small volumes and which are
commensurate with the volume of the sensing cell. From figure 8, it is
seen that even a sensor volume as small as 1 ml will have a significant
effect on the peak width and the maximum resolution will not be
obtained from the column. It is clear that the sensor cell volume must be
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
16

no greater than 2 ml if the column performance is not to be denigrated to


a significant extent. To emphasize the effect of cell volume, it should be
noted that the results from a sensor cell having a volume of 5 ml are
virtually useless. Unfortunately, many commercially available detectors
have sensor volumes as great as, if not greater than 5 ml. If small bore
columns are to be employed, such sensor volumes must be carefully
avoided.
Dispersion Resulting from the Detector Time Constant.

In addition to the sources of dispersion so far discussed, the peak can


appear to be further dispersed by the combined time constant of the
sensor and its associated electronics. It must be emphasized that the time
constant of the system can not actually disperse an eluted peak, but its
effect of it on the sensor measurement can produce an apparent peak
dispersion. Thus the term appear is used as the solvent profile itself is not
changed, only the profile as presented on the recorder or printer. The
effect of the detector time constant can be calculated and the results
from such a calculation are shown in figure 9.
T' = 0 sec

T' = 0.6 sec


Detector Response

T' = 1.5 sec

Time

Figure 9. Peak Profiles Demonstrating Distortion Resulting from


Detector Time Constant
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
17

The undistorted peak, monitored by a detecting system with a zero time


constant, is about 4 seconds wide. An LC column operating at a flow
rate of 1 ml/min. and having a peak base-width of 4 seconds would
represent a peak with a volume of about 67 ml. It follow, that the peaks
depicted would represent those eluted fairly late in the chromatogram.
However, despite the late elution, the distortion is still quite severe. To
avoid distortion of the early peaks the time constant would need to be at
least an order of magnitude less. Scott et al. (12) measured the time
constants of two photocells and their results are shown in figure 10.

Cadmium Sulphide Photocell


Log
Decay
Curve

Decay
Normalized Curve
Curve

0 5 10

IP-28 Photomultiplier

Log
Log
Decay
Decay
Curve
Curve

Normalized
Curve
Decay
Curve

0 0.25 0.5
Time (seconds)

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
18

Figure 10. The Response Curves of Two Photocells

The output each photocell to fast transient changes in incident light


intensity was monitored with a high speed recorder. The curves for the
cadmium sulfide photocell, figure 10 (chosen as an old type, sensor with
a very slow response) is shown at the top of the figure. From the slope
of the log curve, the time constant was calculated to be about 2.5
seconds. Such an extremely slow response would be impractical for
modern chromatographic systems (i.e., two or more peaks could elute
within the period of the time constant). The result of the slow response
of the cadmium sulfide sensor, would be to cause the peaks merge into a
single distorted peak.
The performance of the photomultiplier (representing a sensor with a fast
response) is shown in the lower curves of figure 10. The time constant,
determined from the slope of the log curve, was only 40 milliseconds. A
response time of 40 milliseconds is acceptable for most LC separations.
Nevertheless in fast LC separations, solutes can be eluted in less than 100
milliseconds in which case an even faster response might be necessary.

Contemporary sensors and electronic systems use fast solid state sensors
and solid state electronic components. Thus, most commercial detector
systems are sufficiently fast for the vast majority of chromatography
applications. As a general rule, the overall time constant of an LC
detecting system should be less than 50 milliseconds. For specially very
fast separations, a lower value of 15 milliseconds may be necessary. Fast
sensors and electronics will respond to high frequency noise so the
chromatographic system must be designed to reduce short term noise.
This may involve magnetic screening to reduce the effect of stray, low-
frequency electromagnetic fields from nearby power supplies and any
high energy consuming laboratory equipment.

In general, as the peaks in LC separations can be extremely small all


sources of dispersion must be taken into account. It follows that in the
design of the chromatograph, careful steps must be taken to minimize

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
19

the effect from such dispersion sources and to ensure the integrity of the
separation is maintained.

LC Detectors Based on Refractive Index Measurement

LC detectors range from those that are exclusively non specific (i.e., bulk
property detectors, e.g., the refractive index detector) through those that
are partially specific (i.e. partial solute property detectors, e.g., the UV
detectors) to the totally specific detectors (i.e., solute property detectors,
e.g., the fluorescence detector). In general, the sensitivity increases
progressively as the detector becomes more specific, the highest
sensitivities being obtained from the specific detectors.

Refractive index is a bulk property of the column eluent and so detection


depends on the solute modifying the overall refractive index of the
mobile phase sufficiently to provide a signal twice that of the noise. Bulk
property detectors have an inherently limited sensitivity irrespective of
the instrumental technique that is used. Consider an hypothetical bulk
property detector that monitors the density of the eluent leaving the
column. Assume it is required to detect the concentration of a dense
material, such as carbon tetrachloride (specific gravity 1.595), at a level
of 1 mg/ml in n-heptane (specific gravity 0.684).

Let the change in density resulting from the presence of the solute at a
concentration of 10-6 g/ml be (D d). It follows, that to a first
approximation,
X s (d1 - d 2 )
Dd =
d1

where (d1) is the density of the solute, carbon tetrachloride,


(d2) is the density of the mobile phase, n-heptane
and (Xs) is the concentration of the solute to be detected.

Thus for the example given,

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
20

(1.595 - 0.684 ) x 10 -6
Dd =
1.59
= 5.71 x 10 -7
The coefficient of cubical expansion of n-heptane is about 1.6 x 10-3 per
oC. The temperature (Dq) that would produce a change equivalent to the

presence of carbon tetrachloride at a concentration of 10-6 g/ml can,


therefore, be calculated.

5.71 x 10 -7 o
Thus, Dq = C
-3
1.6 x 10

= 3.6 x10 -4 o C

Therefore, to detect a concentration of one part per million of carbon


tetrachloride (at a signal to noise ratio of two), then the temperature
variation must be maintained below 1.8 x 10-4 oC. Such temperature
stability is extremely difficult to maintain and, thus, temperature control
will limit the sensitivity obtainable from the detector. Even the heat of
adsorption and desorption of the solute on the stationary phase can
produce temperature changes of this order of magnitude.

Similarly, the density of the contents of the cell will change with pressure
and, if there is a significant pressure drop across the cell, also with flow
rate. These stability problems apply to all bulk property detectors and,
thus, bulk property detectors in general will all have a limited sensitivity
(on average for most compounds, this will be about 10-6 g/ml). In
addition, even to achieve this sensitivity, the sensor must always be
operated under very carefully controlled conditions.

The Refractive Index Detector

One of the first on-line detectors to be developed was the refractive


index detector originally described by Tiselius and Claesson (14) in 1942.
Despite its limited sensitivity, this detector can be very useful for

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
21

detecting those compounds that are nonionic, do not adsorb in the UV,
and do not fluoresce.

Since 1942, there have been many types of refractive index detectors
introduced and a number of different optical systems utilized. Only those
in common use or having particular interest will be described here.
The Angle of Deviation Method

When a monochromatic ray of light passes from one isotropic medium,


(A), to another, (B), it changes its wave velocity and direction. The
change in direction is called the refraction and the relationship between
the angle of incidence and the angle of refraction is given by Snell's law,
namely,
n sin (i )
n' B = B =
nA sin (r)

where (i) is the angle of incident light in medium (A),


(r) is the angle of refractive light in medium (B),
(nA) is the refractive index of medium (A),
(nB) is the refractive index of medium (B),
and (n'B) is the refractive index of medium (B) relative to that of
medium (A).

Refractive index is a dimensionless constant that normally decreases with


increasing temperature. The reported values are usually taken at 20o or
25oC and are mean values measured for the two sodium lines. If the
mobile phase is allowed to flow through a hollow prism and a ray of
light passes through the prism it will be diverged from its original path
and can be focused onto a photocell. If the refractive index of the mobile
phase changes, due to the presence of a solute, the angle of deviation of
the transmitted light will also alter and the amount of light falling on the
photocell will change. A number of manufacturers have employed the
angle of deviation method in refractive index detector design.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
22

Sample
Mirror Mask Light
Lens Source

Sensor Recorder
Zero Amplifier and
Adjust Power Supply
Reference

Figure 11. The Refractive Index Detector Based on the Angle of


Deviation Method of Measurement
A diagram of a simple refractive index detector that is based on the
angle of deviation method of measurement is shown in figure 11. The
differential refractometer monitors the deflection of a light beam caused
by the difference in refractive index between the contents of the sample
cell and those of the reference cell. A light beam from an incandescent
lamp is confined to the region of the cell by an optical mask. A lens
collimates the light beam through both the sample and reference cells to
a plane mirror. The mirror reflects the beam back through the sample
and reference cells to a lens which focuses it onto a photocell.

In fact, it is the location of the beam, rather than its intensity, that
changes with the refractive index difference between the contents of the
two cells. As the position of focus of the beam on the photoelectric cell
changes, the electrical output changes which is electronically modified to
provide a signal proportional to the concentration of solute in the cell.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
23

sucrose
glucose
maltose

xylose lactose

maltotriose

Time

Figure 12. Chromatogram from an RI Detector Based on the Angle


of Deviation Method of Measurement
An example of a separation monitored by a refractive index (RI) detector
for a typical sugar application is shown in figure 12.
The Fresnel Method

The relationship between the reflectance from an interface between two


transparent media and their respective refractive indices is given by
Fresnel's equation,
1 È sin 2 (i - r) tan 2 (i - r) ˘
R = Í 2 + ˙
2 Î sin (i + r) tan 2 (i + r) ˚

where (R) is the ratio of the intensity of the reflected light to that of the
incident light and the other symbols have the meanings previously
assigned to them.
sin(i) n
Now, = 1
sin(r) n 2

where (n1) is the refractive index of medium (1),


and (n2) is the refractive index of medium (2).

Consequently, if medium (2) represents the column eluent, any change in


(n2) will change (R) (i.e., DR) and, thus, measurement of (DR) will detect

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
24

changes in solute concentration. The first to utilize this principal of


detection was in the construction of a practical detector was Conlon (4).
Conlon's device is now obsolete but it illustrates the principle of the
Fresnel method of detection very simply. A diagram of Conlon's
detector is shown in figure 13.

Light
Source

Photocell

Glass Rod

From
Column

To Waste

Figure 13 A Simple Detector Based on the Fresnel Method of


Refractive Index Measurement

The sensing element consists of a rod prism sealed into a tube through
which the solvent flows. The rod (6.8 mm in diameter and 10 cm long)
is made from a glass rod, bent to the correct optical angle (just slightly
less than the critical angle) and an optical flat is ground on the apex of
the bend (see figure 13). The optical flat is then sealed into the window
of a flow-through cell. The photocell is arranged to be one arm of a
Wheatstone bridge and a reference photocell (not shown) which
monitors light direct from the cell, is situated in another arm of the
bridge.
A commercial refractive index detector working on the Fresnel principle
is shown diagramatically in figure 14. Light from a tungsten lamp passes
through an IR filter (to minimize thermal effects) onto a magnifying
assembly and prism that also splits the beam into two.
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
25

To User
View Port
Flow Cell
Lens/Splitter

Prism
IR Filter
Flat
Mirror

Flat
Focusing Mirror
Lenses
Lamp
Photocell

Figure 14 A Diagram of the Optical System of a Refractive Index


Detector Operating on the Fresnel Method

The two light beams are arranged to pass through the sample and
reference cells respectively. Refracted light from the mobile phase/prism
surface passes through the prism assembly and focused onto two
photocells. The prism is also arranged to reflects some light to an
aperture where the surface of the prism can be observed. The photocell
outputs are electronically processed and passed to either a potentiometric
recorder or a computer data acquisition system. The refractive index
range monitored by the device for a given prism is limited and,
consequently, there are usually three different prisms available to cover
the RI ranges of 1.35–1.4, 1.31–1.44 and 1.40–1.55 respectively. In
figure 15 is shown the separation of a series of polystyrene standards
monitored by this type of detector.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
26

5
Polystyrene
1. MW 1,650,000
2. Mw 480,000
3. MW 180,000 2 4
4. MW 76,000
5. MW 39,000
6. MW 11,600 7
7. MW 2,900 3 6 8
8. MW 580. 1

0 5 10 15
Time (minutes)

Figure 15 The Separation of Some Polystyrene Standards Using a


RI Detector Operating on the Fresnel Method
The separation was carried out by size exclusion on a column packed
with 5 mm particles and operated at a flow rate 0.8 ml/min.
As a result of limited sensitivity and restricted linear dynamic range, the
RI detector is only used for those applications where, for one reason or
another. all other detectors are inappropriate or impractical.
This type of detector does, however, have one particular area of
application for which its characteristics make it particularly suitable and
that is for monitoring the separation of polymers. This is because for
those polymers containing more than six monomer units, the refractive
index is proportional to polymer concentration and independent of its
molecular weight. Consequently, quantitative estimation of each polymer
mixture can be obtained by simple normalization of peak areas and no
individual response factors are required. RI detectors have sensitivities of
about 1 x 10-6 g/ml, a linear dynamic range of about 200 and a response
index (r) lying between 0.97 and 1.03.
The Christiansen Effect Detector

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
27

This procedure for measuring refractive index arose from the work of
Christiansen on crystal filters (14,15). Consider a cell packed with
particulate material having the same refractive index as the mobile phase
passing through. If a beam of light passes through the cell there will be
little of no refraction or scattering. However, if the refractive index of the
mobile phase changes, there will now be a refractive index difference
between the mobile phase and that of the packing. As a consequence
some light will be refracted away from the incident beam and the
intensity of the transmitted light will be attenuated. Thus, if the
transmitted light is focused onto an appropriate photocell, then any
change in refractive index caused by the elution of a solute will produce
scattering and a consequent change in electrical output.
In practice, he optical dispersions of the media are likely to differ, and
consequently the refractive index will only match at one particular
wavelength. As a result the fully transmitted light will be largely
monochromatic. Light of different wavelengths will be proportionally
dispersed depending on the wavelength at which the two media have the
same optical dispersion. Thus, a change in mobile phase refractive will
change both the intensity of the transmitted light and its wavelength.
This device was made by GOW-MAC Inc., who claimed it had a
sensitivity of 1 x 10-6 refractive index units (the maximum that cold be
expected). This would be equivalent to a sensitivity of 9 x 10-6 g/ml of
benzene (refractive index 1.501) eluted in n-heptane (refractive index
1.388). The cell volume was kept to 8 ml (a little large for modern
sensors) which was small enough to work satisfactorily with 4.6 mm I.D.
LC columns. Different cells packed with appropriate materials were
necessary to cover the refractive index range of 1.31 to 1.60. A diagram
of the Christiansen detector is shown in figure 5.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
28

Solid Eluent
Packing from To
Column Waste

Achromat

Lamp

Photocells

Condensor
Prism
Aperture

Reference To
solvent Waste

Figure 16. The Christiansen Effect Detector

In the optical unit there is a pre focused lamp having an adjustable


voltage supply to allow low energy operation when the maximum
sensitivity is not required. The condensing lens, aperture, achromat and
beam splitting prisms are mounted in a single tube which permitted easy
optical alignment prevented contamination from dust. The device
contains two identical and interchangeable cells. The disadvantage of this
detector is that the cells must be changed each time a different mobile
phase is chosen in order to match the refractive index of the packing to
that of the new mobile phase. The refractive indices of the cell packing
can be closely matched to that of the mobile phase by using appropriate
solvent mixtures. In most cases solvent mixing can be achieved without
affecting the chromatographic resolution significantly (e.g. by replacing a
small amount of n-heptane in a mixture with either n-hexane or n-octane
depending on whether the refractive index needs to be increased or
decreased. However a considerable knowledge of the effect of different
solvents on solute retention is necessary to accomplish this procedure
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
29

successfully. As a result of limitations inherent in his type of detector


combined with the general disadvantages of the RI detector per se has
not made the Christiansen Effect Detector very popular.
The Interferometer Detector

The interferometer detector was first developed by Bakken and Stenberg


(16) in 1971. The response of the detector depends on the change in the
effective path length of a beam of light passing through a cell when the
refractive index of its contents changes due to the presence of an eluted
solute. Light that has passed through the cell is focused on a photocell.
Coincidentally a reference beam of light from the same source is focused
on the photocell, interference fringes are produced. The fringes change
as the path length of one light beam changes with reference to the other,
thus, as the concentration of solute increases in the sensor cell during an
elution of a peak, a series of electrical pulses will be generated as each
fringe passes the photocell.
The optical path length (d) of light through the cell depends on the
change in refractive index (Dn), and the path length (l), thus,
d!= Dn l
In addition, the number of fringes (N) which move past a given point on
the photo cell (or the number of cyclic changes of the central portion of
the fringe pattern) is given by,

2Dnl
N =
l
where (l) is the wavelength of the light employed.
The larger the value of (N) for a given (Dn), the more sensitive the
detector will be. It follows that, (l) should be made as large as possible.
However, this procedure for increasing the sensitivity is limited by the
dead volume of the column and the dispersion that can be tolerated
before chromatographic resolution is impaired. A diagram of the simple
optical system originally employed by the authors is shown in figure 17.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
30

Plain
Mirror

From
Column

Plain
Mirror
Lamp

Sensor
Cell
Half
Mirror
To Waste

Photocell

Figure 17 The Original Optical System Used by Bakken and


Stenberg in Their Interferometer Detector

Light from an appropriate source strikes a half silvered mirror and is


divided into two paths. Part of the beam is reflected by a plane mirror
back along the same path and onto a photocell.

Time

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
31

Figure 18 Chromatogram from the Bakken and Stenberg


Interferometer Detector

The other part of the beam passes through the sensor cell to a plane
mirror, where it is reflected back again through the sensor cell to the half
silvered mirror that reflects it onto the photocell. Interference takes place
with the other half of the light beam on the surface of the photocell. The
trace resulting from the elution of 8 ml of dioxane through the cell is
shown in figure 18.

Each peak shown in figure 18 represents the passage of a fringe across


the surface of the photocell. The four interference peaks represents a
single chromatographic peak. The number of fringes will be directly
proportional to the total change in refractive index, which, in turn, will
be proportional to the total amount of solute present. In this form the
detector has limited use, but has been developed into a commercially
viable instrument called the Optilab DSP by Wyatt Technology Inc. A
diagram of the optical system of the Optilab interference detector is
shown in figure 19.

Light from the source is linearly polarized at -45o to the horizontal plane.
Horizontal and vertical polarized light beams are produced and after
passing through the Wollaston prism, one beam passes through the
sample cell and the other beam through the reference cell. The sample
cell beam is horizontally polarized and the reference cell beam is
vertically polarized. After passing through the cells, the beams are
focused onto a second Wollaston prism and then through a quarter-wave
plate which has its fast axis set -45o to the horizontal plane. A beam that
is linearly polarized in the fast axis plane after passing through the plate
will lead another linearly polarized but orthogonal beam by a quarter of
a wavelength. The phase difference results in a circularly polarized beam.
Each of the beams focused on the Wollaston prism consists of two such
perpendicular beams which, after the quarter wave plate, result in two
circularly polarized beams of opposite rotation. These beams will
interfere with each other to yield the original linearly polarized beam. A
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
32

second polarizer is placed at an angle (90 – b) to the first, allowing about


35% of the signal to reach the photocell. A filter transmitting light at 546
nm precedes the photocell.

Wollaston Sample Wollaston


Prism Cell Prism
Lamp
Interference
Filter

Lens Photocell
Polarizer Lens
Analyzer
Mask Reference
Cell Quarterwave
Plate
Courtesy of Wyatt Technology
Figure 19 The Optilab Interference Refractometer Detector
If the sample cell contains a higher concentration of solute than the
reference cell the refractive index will be higher and the interfering
beams will be out of phase. The refractive index difference (Dn) and the
phase difference (Dp) are related by
2pLDn
Dp =
l

where (L) is the length of the cell,


and (l) is the wavelength of the light.

The circularly polarized beams will, therefore, interfere to yield a linearly


Dp
polarized beam which is rotated radians, and the amplitude of the
2
light striking the photocell (Ap) will be given by

Ê D pˆ D pˆ
A p = A o cos Á 90 - b - ˜ = A o cos ÊÁ b - ˜
Ë 2 ¯ Ë 2 ¯

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
33

An extremely high sensitivity is claimed for this system but it is difficult


to interpret the data in terms of minimum detectable concentration The
smallest cell (1.4 ml) is reported to give a sensitivity of about 2 x 10-7 RI
units at a signal-to-noise ratio of two. Consequently, for benzene (RI =
1.501) sensed as a solute in n-heptane (RI=1.388 ) this sensitivity would
represent a minimum detectable concentration of 5.6 x 10-5 g/ml. The
alternative 7 m l cell would decrease the minimum detectable
concentration to about 1 x 10-6 g/ml, similar to that obtained for other
refractive index detectors.

A number of LC detectors have been developed that are either based on


refractive index measurement or function on some physical property of
the mobile phase system that is related to the refractive index. Although
most are not commercially available, they demonstrate the range of
sensing techniques that have been investigated as possible methods of
detection.

The Thermal Lens Detector

If a laser is focused on an absorbing substance, the refractive index of the


material can be modified in such a way that the medium behaves as a
lens. The thermal lens effect was first reported by Gorden et al. (25,26)
in 1964 but since that time the phenomenon has been investigated by a
number of workers. Thermal lens formation results from extremely weak
laser light adsorption The excited-state molecules subsequently decay
back to ground state causing localized temperature increases to occur in
the sample. Since the refractive index of the medium depends on the
temperature, the ensuing spatial variation of refractive index produces an
effect which appears equivalent to the formation of a lens within the
medium.

For most liquids, the temperature coefficient of refractive index is


negative and consequently, the insertion of a liquid in the laser beam
produces a concave lens that results in beam divergence. Buffet and
Momis (27) used the thermal lens effect to develop a small volume
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
34

detector, a diagram of which is shown in figure 20. The device consists


of a heating laser, the light from which is passed directly through the
sample via two lens and a half mirror. Another laser, the probe laser,
passes light in the opposite direction through one lens, through the
sample to the half mirror where the light is reflected onto a photocell.

Half Reference
Mirror Photocell

Heating Probe
Laser Lens Lens Laser

Pin Hole
Mask
Heat Filter
Filter Sample
Cell
Photocell

Figure 20 The Layout of a Thermal Lens Detector

A filter and a pinhole screen is placed between the mirror and the photo-
cell to remove the heating laser light. When an absorbing solute is eluted
from the column through the cell, a thermal lens is produced causing the
probe light to diverge, and the intensity of the light passing through the
pinhole and on to the photocell is reduced.

The cell volume can be as little as a few microliters and, thus, would be
suitable for use with microbore columns. A sensitivity of 10-6 AU has
been claimed for the detector and a linear dynamic range of about three
orders of magnitude. The thermal lens detector is, in fact, a special form
of the refractive index detector and might, therefore, be considered a
universal detector. Nevertheless, like other bulk property detectors, it can
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
35

not be used with gradient elution or flow programming and has


sensitivity that is no better, if as good as, other refractive index
detectors.
The Dielectric Constant Detector

T he refractive index of a substance is a complementary property to the


dielectric constant and in some circumstances is a direct function of it.
For non-polar substances, the relationship between dielectric constant (e)
and refractive index (n) is given by

e!!=!!n2

For semi-polar substances or mixtures of semi-polar substances and non-


polar substances the Lorentz-Lorenz equation applies

e -1 n2 - 1
= 2
e +2 n + 2

However, for polar substances and mixtures of polar and semi-polar


substances the relationship breaks down and no simple functions describe
dielectric constant in terms of refractive index.

The more polar the substance, the larger is its dielectric constant. In
n o r m a l chromatography (as opposed to reversed phase
chromatography) the mobile phase is normally less polar than the solutes
being eluted. Thus, the presence of a solute in the mobile phase will
increase the dielectric constant of the mobile phase. Conversely, in
reversed phase chromatography the solute is usually less polar than the
solvent and the dielectric constant of the mobile phase is reduced by the
presence of a solute. Thus. a device situated at the end of the column
which responds to changes in dielectric constant would act as a
chromatography detector. The sensor often takes the form of a
cylindrical or parallel plate condenser. The volume of the sensor must be
as small as possible to minimize dispersion. In addition, as the sensitivity

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
36

of the device is proportional to the electrical capacity of the sensor, the


capacitor plates must be very close together.

A suitable circuit for use in dielectric constant measurement is an


electrical "bridge", the detector cell being situated in one arm of the
bridge. If the sensor cell has a capacity greater than 100 pF, then a Wein
bridge can be used; however such a cell may well have a fairly large
volume. For smaller capacity cells, the Schering bridge is more
appropriate and a diagram of a Schering bridge is shown in figure 21.

r
Cell
R
C

Ro
Co
C'

Figure 21 The Schering Bridge for the Measurement of Small


Capacities
No capacitor is ideal, all will have some inductance and resistance in
addition to its capacity. In fact, because the plates of the capacitor are
situated in the mobile phase, if uninsulated, it is very likely to have a
significant resistance component. The current though the resistive
component of a conductor is in phase with the applied voltage and the
capacity component lags the applied voltage by 90o. Thus, there are two
components to be balanced before the output of the bridge (across (D))
can be used to monitor the elution of a solute.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
37

The Schering bridge is balanced by the iterative adjustment of (Ro) and


(C'). At balance the following relationships will hold:

C Co r C'
= and =
Ro R R Co

The resistance-component of the cell reduces the bridge sensitivity to


changes in capacity and thus the plates should be well insulated to
prevent conductivity through the mobile phase.

The capacity of the sensor can also be measured by making it one


component of a resistance/capacity or an inductance/capacity oscillator.
The frequency will depend, among other things, on the capacity of the
sensor and, in turn, on the dielectric constant of the material between the
plates. The frequency general can be heterodyned against a reference
oscillator and the frequency difference will then be proportional to the
change in capacity and hence the dielectric constant of the mobile phase.

Poppe and Kunysten (28) described a dielectric constant detector which


included a reference cell for temperature compensation. The cell
consisted of two stainless steel plates 2 cm x 1 cm x 1 mm separated by a
gasket 50 mm thick. The two cells were identical and clamped back to
back, sharing a common electrode.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
38

Figure 22 The Sensor of a Dielectric Constant Detector

The device was reported to have a sensitivity of 10-6 g/ml for


chloroform (e = 4.81) in n-octane. As might be expected, it was found to
be very sensitive to pressure changes in the cell (thought due to plate
deformation) even when constant flow pumps were employed. The first
dielectric constant detector became commercially available in 1979 (29)
and was described by Benningfield and Mowery (30). Several
applications were reported by Bade et al. (31). A diagram of the sensor
is shown in figure 22.
Each cell consisted of a concentric cylinder (inner electrode) inside a
larger cylinder (the outer electrode) which formed the outer wall of the
cell. Both electrodes were made of stainless steel. The two cylinders were
electrically isolated with a cylindrical flow path through the cell. The
inner cylindrical electrodes were 1.26 cm in diameter and 0.625 cm long
separated from the outer cylinder by about 0.009 cm.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
39

Methylene
Dichloride

Trumyristin

Tripalmatin

0 5 10 15
Time (minutes)

Figure 23 The Separation of Some Triglycerides Monitored by a


Dielectric Constant Detector

The linear dynamic range of the detector was reported to be 3.5 x 104 .
The sensitivity was quoted as about 1 x 10-7 g/ml, which would be close
to the theoretical limit for bulk property detectors. An example of the
use of the dielectric constant detector to monitor a separation of
triglycerides is shown in figure 23.

Bulk property detectors have neither the sensitivity nor the linear
dynamic range of solute property detectors and are less frequently used
in modern LC analyses. None can be used satisfactorily with gradient
elution, flow programming or temperature programming and so they
restrict the choice of development. They do have certain unique areas of
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
40

application, some of which have already been mentioned. Their use


probably represents less than 5% of all LC analyses.

The UV Detectors

Although over the years a large number of LC detectors have been


developed and described, the vast majority of all contemporary LC
analyses are carried out using one of four detectors, the UV detector in
one of its forms, the electrical conductivity detector, the fluorescence
detector and the refractive index detector. In addition, some form of the
UV detector probably accounts for 80% of those analyses.

The UV Absorption Detectors

UV absorption detectors respond to those substances that absorb light in


the range 180 to 350 nm. Many (but not all) substances absorb light in
this wavelength range, including those substances having one or more
double bonds (p electrons) and substances having unshared (unbonded)
electrons, e.g. all olefins, all aromatics and compounds, for example,
containing > C = O , > C = S , – N = N – groups. The sensor of a UV
detector consists of a short cylindrical cell having a capacity between 1
ml and 10 ml through which passes the column eluent. UV light is
arranged to pass through the cell and fall on a photo–electric cell (or
array). The output from the photocell passes to a modifying amplifier
and then to a recorder or data acquisition system.

The relationship between the intensity of UV light transmitted through a


cell (IT) and the concentration of solute contained by it (c) is given by
Beer's Law.
I T = I o e -klc

or ln (IT) = ln (Io) - kcl

where (Io) is the intensity of the light entering the cell,


(l) is the path length of the cell,

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
41

and (k) is the molar extinction coefficient of the solute for the
specific wavelength of the UV light.

Differentiating,
Ê IT ˆ
d Á l og ˜
Ë Io ¯
= - kl
dc
The sensitivity of the detector, as measured by the transmitted light, will
be directly proportional to the extinction coefficient (k) and the path
length of the cell (l). To increase the sensitivity of the system, (l) must be
extended but there is a limit to which (l) can be increased as the cell
volume and, in particular, the length of the cell must be restricted. This is
necessary to minimize peak dispersion in the sensor and to avoid more
than a small fraction of a peak existing in the cell at any one time This
problem has already been discussed. To restrict peak dispersion, the
radius of the cell must also be reduced as (l) is increased. Thus, less light
will fall on the photo–cell, the signal–to–noise ratio will be reduced and
thus the detector sensitivity or minimum detectable concentration
denigrated. Thus, increasing the detector sensitivity by increasing the
path length has limitations and a well–designed cell involves a careful
compromise between cell radius and length to provide the maximum
sensitivity. Most modern UV detector sensors have path lengths that
range between 1 and 10 mm and internal radii that range from about 0.5
to 2 mm
I
Now, Log T = k' l c = A
Io

where (A) is termed the absorbence


Now (DA) is sometimes employed to define the detector sensitivity
where the value of (DA) is the change in absorbence that provides a
signal-to-noise ratio of two.

Thus D A = k' l D c
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
42

where (D c) is the detector concentration sensitivity or minimum


detectable concentration.

DA
Thus Dc=
k' l
Thus. two detectors, having the same sensitivity defined as the minimum
detectable change in absorbence, will not necessarily have the same
sensitivity with respect to solute concentration. Only if the path lengths
of the two sensors are identical will they also exhibit the same
concentration sensitivity. This can cause some confusion as it would be
expected that two instruments having the same spectroscopic sensitivity
would also have the same chromatographic sensitivity. To compare the
sensitivity of two detectors given in units of absorbence the path lengths
of the cells in each instrument must be taken into account.
UV detectors can be used with gradient elution providing the solvents do
not absorb significantly over the wavelength range that is being used for
detection. In reversed phase chromatography, the solvents usually
employed are water, methanol, acetonitrile and tetrahydrofuran (THF), all
of which are transparent to UV light over the total wavelength range
normally used by UV detectors. In normal phase operation more care is
necessary in solvent selection as many solvents that might be appropriate
as the chromatographic phase system are likely to absorb UV light very
strongly. The n-paraffins, methylene dichloride, aliphatic alcohols and
THF are useful solvents that are transparent in the UV and can be used
with normal distribution systems (e.g. a polar stationary phase such as
silica gel).
The Fixed Wavelength UV Detector

The fixed wavelength UV detector uses light of a single wavelength (or


nearly so) which is produced by a specific type of discharge lamp.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
43

100 253.7
Mercury
Lamp

10

302.2 313.2
100
307.6 Zinc
Relative Emission

Lamp
213.9
10

280.1
277.1
100
326.1 346.6
Cadmium 340.3
Lamp
228.8 325.4
10 293.1 308.1
226.5
288.1 313.2
214.4
283.6

200 250 300 350


Wavelength in nm

Figure 24 Emission Spectra for Three Discharge Lamps

The most popular lamp is the low pressure mercury vapor lamp, which
generates most of its light at a wavelength of 254 nm. Other lamps that
could be used are the low-pressure cadmium lamp which generates the
majority of its light at 225 nm and the low pressure zinc lamp that emits
largely at 214 nm. None of the lamps are strictly monochromatic and
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
44

light of other wavelengths is always present but usually at a significantly


lower intensity. The emission spectra of the mercury, cadmium and zinc
lamps are shown in figure 24. It is seen that to obtain monochromatic
light an appropriate filter would be needed. The low pressure mercury
light source (wavelength 253.7 nm) provides the closest to true
monochromatic light of all three lamps. However, there is light present
of significant intensity below 200 nm, but light of such wavelengths is
generally absorbed by the mobile phase.
The zinc lamp has a major emission line at 213.9 but the emission line at
307.6 is of comparable intensity and a suitable filter would be needed if
detection was required to be exclusively at the lower wavelength. The
cadmium lamp has a major emission line at 228.8 but light is emitted at
both lower wavelengths and at substantially higher wavelengths and so
an appropriate filter would again be desirable. Suitable interference filters
can be quite expensive to construct, which may account for the
unpopularity of these two lamps. They do, however, emit light at
wavelengths which would give an increased sensitivity to substances
such as proteins and peptides, which might make their use worthwhile in
the biotechnology field. A diagram of a typical optical system for a fixed
wavelength UV detector is shown in figure 25. Light from the UV
source is collimated by a suitable lens and passed through both the
sample cell and the reference cell and then on to two photo cells The
cells are cylindrical with quartz windows at either end. The reference cell
compensates for any absorption that mobile phase might have at the
sensing wavelength. The outputs from the two photo cells are passed to
a signal modifying amplifier so that output is linearly related to the
concentration of solute being detected. For reasons already discussed,
modern sensor cells have angular conduits that form a 'Z' shape to
reduce dispersion. The UV sensor can be sensitive to both flow rate and
pressure changes but this instability can be greatly reduced if the sensor
is well thermostatted.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
45

To From
Waste Column
Quartz
Windows

Sample cell

Photo
Cells
Quartz
Lens
Quartz Reference Cell
Low Pressure
To Windows Reference
Mercury Lamp
Waste Flow

Figure 25. The Fixed Wavelength UV Detector

The fixed wavelength UV detector is one of the most commonly used


LC detectors; it is sensitive, linear and relatively inexpensive. Sensitivity
(minimum detectable concentration) can be expected to be about 5 x
10–8 g/ml with a linear dynamic range of about three orders of
magnitude for 0.98 < r < 1.02. The separation of some aromatic
hydrocarbons by exclusion chromatography on a very high efficiency
column (efficiency ca 250,000 theoretical plates) monitored by a fixed
wavelength detector is shown in figure 26. All the solutes are distinctly
resolved despite their having molecular weight differences equivalent to
only two methylene groups. The peaks from such columns are only a
few microliters in width and so a specially reduced volume sensor cell
was necessary to cope with the high efficiencies and allow the
consequent improved resolution to be realized. The molecular weight of

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
46

decyl benzene is 218 and, thus, one methylene group would represent a
differential of only 6.4% of the molecular weight.

Column, length, 10m, I.D., 1 mm, mobile phase tetrahydrofuran, flow-rate 30 ml/min.,
adsorbent Partisil 10. Column efficiency ca. 250,000 theoretical plates. Solutes
benzene, ethyl benzene, butyl benzene, hexyl benzene, octyl benzene and decyl
benzene. Such columns must be packed in short lengths (about 1 m) and
subsequently joined and are thus somewhat tedious to construct.

Figure 26. The Separation of Some Alkyl Benzenes by Exclusion


on a High Efficiency Column.

A very simple fixed wavelength detector suitable for use in preparative


chromatography is shown in figure 27. This detector was invented by
Miller and Strusz (32) and originally manufactured by GOW-MAC
Instruments. As opposed to detectors used for analytical purposes,
detectors for preparative work need to have a very low sensitivity as
sample sizes are large and consequently the solute concentrations are
very high. If analytical detectors are used for preparative work a portion
of the eluent is split from the main stream, diluted with more mobile
phase and then passed through the detector. In practice, this is a rather
awkward procedure. As seen in figure 27 the column eluent passes
through a delivery tube and onto a supporting plate that is usually made
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
47

of fused quartz, so that adequate UV light can reach the photo cell
placed on the other side of the plate.

From From
Column Column

Glass Glass
Plate Plate

UV Lamp

Mask Photo
Flow of Mobile Phase Cell
Over Plate

Figure 27. Fixed Wavelength Detector for Preparative Work

The liquid flows over the plate and the effective path length of the sensor
will be the film thickness which will be unique to the particular solvent
used as the mobile phase. The UV lamp is situated above the upper side
of the plate and the photo cell on the lower side. A reference photo cell
(not shown) is situated close to the lamp and the output used to
compensate for changes in light intensity from variations in lamp
emission. The short path length results in a low sensitivity and the
detector can operate satisfactorily at concentrations as high as 10-2 g/ml
(1% w/w), which is ideal for preparative chromatography. A particular
advantage of this type of sensor is its very low flow impedance and thus
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
48

can easily accommodate the high flow rates used in preparative LC. The
film thickness does depend, among other factors, on the column flow
rate, consequently, precise flow control is necessary for the detector to
perform satisfactorily.

The Multi–Wavelength UV Detector

The multi–wavelength detector employs a light source that emits light


over a wide range of wavelengths. Employing an appropriate optical
system (a prism or diffraction grating), light of a specific wavelength can
be selected for detection purposes. Light of a specific wave length
wavelength might be chosen where a solute has a absorption maximum
to provide maximum sensitivity. Alternatively, the absorption spectra of
an eluted substances could be obtained for identification purposes by
scanning over a range of wavelengths. The latter procedure, however,
differs with the type of multi–wavelength detector being used.
There are two basic types of multi–wavelength detector, the dispersion
detector and the diode array detector, the latter being the more popular.
In fact, very few dispersion instruments are sold today but many are still
used in the field and so their characteristics will be discussed. All
multi–wavelength detectors require a broad emission light source such as
deuterium or the xenon lamp, the deuterium lamp being the most
popular.
The two types of multi-wavelength detectors have important differences.
In the dispersive instrument, the light is dispersed before it enters the
sensor cell and thus virtually monochromatic light passes through the
cell. However, if the incident light is of a wavelength that can excite the
solute and cause fluorescence at another wavelength, then the light
falling on the photo cell will contain the incident light together with any
fluorescent light that may have been generated. It follows, that the light
monitored by the photo cell may not be monochromatic and light of
another wavelength, if present, would impair the linear nature of the
response. This effect would be negligible in most cases but with certain
fluorescent materials the effect could be significant. The diode array
detector operates quite a differently. Light of all wavelengths generated
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
49

by the deuterium lamp is passed through the cell and then dispersed over
an array of diodes. Thus, the absorption at discrete groups of wavelengths
is continuously monitored at each diode. However, light falling on a
discrete diode may not be derived solely from the incident light but may
contain light generated by fluorescence excited by light of a shorter
wavelength. Unfortunately, this effect is exacerbated by the fact that the
cell contents are exposed to light of all wavelengths emitted by the
source and so fluorescence is more likely. Thus, under some
circumstances, measurement of transmitted light may involve fluorescent
light and the absorption spectrum obtained for a substance may be a
degraded form of the true absorption curve.

The ideal multi–wavelength detector would be a combination of both the


dispersion system and the diode array detector. This arrangement would
allow a true monochromatic light beam to pass through the detector and
then the transmitted beam would itself be dispersed again onto a diode
array. Only that diode sensing the wavelength of the incident light would
be used for monitoring the transmission. In this way any fluorescent light
would strike other diodes, the true absorption would be measured and
accurate monochromatic sensing could be obtained.
The Multi–Wavelength Dispersive UV Detector

A diagram of the multi–wavelength dispersive UV detector is shown in


figure 28. Light from the deuterium lamp is collimated by two curved
mirrors onto a holographic diffraction grating. The dispersed light is then
focused by means of a curved mirror, onto a plane mirror and light of a
specific wavelength is selected by appropriately positioning the angle of
the plane mirror. Light of the selected wavelength is then focused by
means of a lens through the flow cell and, consequently, through the
column eluent. The exit beam from the cell is then focused by another
lens onto a photo cell which gives a response that is some function of the
intensity of the transmitted light. The detector is usually fitted with a
scanning facility that allows the spectrum of the solute contained in the
cell to be obtained. There is an inherent similarity between UV spectra of
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
50

widely different types of compounds, and so UV spectra are not very


reliable for the identification of most solutes.

Deuterium
Lamp
Curved
Mirror

Plane
Mirror

To Waste
Photo From
Cell Column

Curved
Mirror

Plane Diffraction
Mirror Grating
Quartz Sample
Lenses Cell

Courtesy of the Perkin Elmer Corporation


Figure 28 The Multi–Wavelength Dispersive UV Detector
The technique can be used, however, to determine the homogeneity of a
peak (e.g., by comparing spectra taken from both sides of the peak.
Both spectra are normalized and either one is subtract one from the
other and the difference is shown to be zero, or the ratio of the two
spectra is calculated and the result shown to be unity.
A common use of multi-wavelength choice is to enhance the sensitivity
of the detector by selecting a wavelength that is characteristically
absorbed by the substance of interest. Conversely, a wavelength can be
chosen that substances of little interest in the mixture do not adsorb and,
thus, make the detector more specific to those substances that do. An
example of the use of the variable wavelength UV in this way is afforded
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
51

by the separation of some carboxylic acids that is monitored by UV


absorption at 210 nm. The separation is shown in figure 29. The
separation of a series of common fatty acids was carried out on a
reversed phase column using water buffered with phosphoric acid as the
mobile phase.
1
2

5 6
7
4 8

Time (minutes)

Column: Spherisorb® Octyl, 25 cm x 4.6 mm I.D., 5 mm particles. Mobile Phase: 0.2


M phosphoric acid. Flow rate 0.8 ml/min. monitored at 210 nm. 1. tartaric acid, 2.
lactic acid, 3. malic acid, 4. formic acid, 5. acetic acid, 6. citric acid, 7. succinic acid, 8.
fumaric acid.
Courtesy of Supelco Inc.

Figure 29. The Separation of Some Carboxylic Acids Monitored by


UV Absorption at 210 nm

Multi-wavelength dispersive detectors has proved extremely useful,


providing adequate sensitivity, versatility and a linear response. As a
result of the need for an optical bench inside the instrument, however, it
is somewhat bulky In addition, it has mechanically operated wavelength
selection and requires a stop/flow procedure to obtain spectra "on-the-
fly". In contrast, the diode array detector has the same advantages but
none of the disadvantages.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
52

The Diode Array Detector

The diode array detector also utilizes a deuterium or xenon lamp that
emits light over the UV spectrum range. Light from the lamp is focused
by means of an achromatic lens through the sample cell and onto a
holographic grating. The dispersed light from the grating is arranged to
fall on a linear diode array.

The resolution of the detector (Dl) will depend on the number of diodes
(n) in the array, and also on the range of wavelengths covered (l2 - l1).

l 2 - l1
Thus Dl =
n

Consequently, the ultimate resolving power of the diode array detector


will depend on the semi–conductor manufacturer and on how narrow
the individual photo cells can be commercially fabricated.

A diagram of a diode array detector is shown in figure 30.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
53

Deuterium
Lamp
Achromatic
Lens System

Outlet

Photo Diode
Array
Shutter

Flow
Cell

Eluent
from
Holographic Column
Grating

Figure 30. The Diode Array Detector


Light from the broad emission source is collimated by an achromatic
lens system so that the total light passes through the detector cell onto a
holographic grating. In this way the sample is subjected to light of all
wavelengths generated by the lamp. The dispersed light from the grating
is allowed to fall onto a diode array. The array may contain many
hundreds of diodes and the output from each diode is regularly sampled
by a computer and stored on a hard disc. At the end of the run, the
output from any diode can be selected and a chromatogram produced
using the UV wavelength that was falling on that particular diode.
During chromatographic development, the output of one diode is
recorded in real time producing a real time chromatogram. It is seen that
by noting the time of a particular peak, a spectrum can be obtained by
recalling from memory the output of all the diodes at that particular
time.
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
54

Ratio A 235 \A 245

Ratio = 2.5

Absorbance 254 nm

I max = 274 nm
P.I. =0

The chlorthalidone was isolated from a sample of tablets and separated by a reverse
phase (C18) on a column 4.6 mm I.D., 3.3 cm long, using a solvent mixture
consisting of 35% methanol, 65% aqueous acetic acid solution (water containing 1%
of acetic acid). The flow rate was 2 ml/min.

Courtesy of the Perkin Elmer Corporation


Figure 31. Dual Channel Plot from a Diode Array Detector
Confirming Peak Purity

The diode array detector can be used in a number of unique ways and
an example of the use of a diode array detector to verify the purity of a
given solute is shown in figure 31. The chromatogram monitored at 274
nm is shown in the lower part of figure 31.As a diode array detector was
employed, it was possible to ratio the output from the detector at
different wavelengths and plot the ratio simultaneously with the
chromatogram monitored at 274 nm. If the peak was pure, the ratio of
the adsorption at the two wavelengths (those selected were 225 and 245
nm) would remain constant throughout the elution of the entire peak.
The upper diagram in figure 31 shows this ratio plotted on the same time
scale and it is seen that a clean rectangular peak is observed which
unambiguously confirms the purity of the peak for chlorthalidone. The
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
55

wavelength chosen to provide the confirming ratio will depend on the


UV adsorption characteristics of the substance concerned, relative to
those of the most likely impurities to be present, consequently the
wavelengths must be chosen with some circumspection.

Anthracene

The separation was carried out on a column 3 cm long, 4.6 mm in diameter and
packed with a C18 reversed phase on particles 3 m in diameter.
Courtesy of the Perkin Elmer Corporation

Figure 32. The Separation of Some Aromatic Hydrocarbons

Another interesting example of the use of the diode array detector to


confirm the integrity of an eluted peak is afforded by the separation of
the series of hydrocarbons shown in figure 32.The separation appears to
be satisfactory and all the peaks appear to represent individual solutes;
without further evidence, it would be reasonable to assume that all the
250 nm
peaks were pure. However, by plotting the adsorption ratio, , for
255 nm
the anthracene peak it becomes apparent that the peak tail contains an
impurity as the clean rectangular shape of the peak top is not shown. The
absorption ratio peaks are shown in figure 33.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
56

The ratio peaks depicted in figure 33 clearly indicate the presence of an


impurity by the sloping top of the anthracene peak. This is further
confirmed by the difference in the spectra for the leading and trailing
edges of the peak.

250/255 nm Ratio

Irregular top to peak


showing presence
of impurity

Figure 33 Curves Relating the Adsorption Ratio, 250 nm , and


255 nm
Time
Spectra taken at the leading and trailing edge of the anthracene peak are
shown superimposed in figure 34. Further work identified the impurity
as t-butyl benzene at a level of about 5%.

Spectra of Trailing
Edge of Peak
Adsorption Units

Spectra of Leading
Edge of Peak

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
57

Figure 34 Superimposed Spectra Taken at the Leading and


Trailing Edges of the Anthracene Peak

Another example of the use of the diode array detector to demonstrate


peak purity is shown in figure 35.

Impure Pure
mAU mAU

a b

200 240 280 320 190 230 270 310


(nm) (nm)
b
a d
c
e

Time (minutes)

Figure 35 Diode Array Spectra Demonstrating Peak Purity


A chromatogram is shown containing five peaks and spectra have been
taken of peak (a) and peak (b) halfway up the rising side of each peak, at
the top of each peak and halfway down the trailing edge of each peak.
The spectra are also included in the figure 35. The impure and pure peak
are unambiguously identified illustrating the value of this type of
detecting system for analytical purposes.
The performance of both types of multi-wavelength detectors are very
similar and typical sensitivities would be about 1 x 10-7g/ml (significantly
less than the fixed wavelength detector) with a linear dynamic range of
about 5000 and a response index lying between 0.97 and 1.03.

The Fluorescence Detector

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
58

When light is emitted by molecules that are excited by electromagnetic


radiation, the phenomenon is termed photoluminescence. If the release of
electro-magnetic energy is immediate, or stops on the removal of the
excitation radiation, the substance is said to be fluorescent. If, however,
the release of energy is delayed, or persists after the removal of the
exciting radiation, then the substance is said to be phosphorescent.
Fluorescence has been shown to be extremely useful as a detection
process and detectors based on fluorescent measurement have provided
some of the highest sensitivities available in LC.

When a molecule adsorbs light, a transition to a higher electronic state


takes place and this absorption is highly specific for the molecules
concerned; radiation of a specific wavelength or energy is only absorbed
by a particular molecular structure. If electrons are raised to an upper
excited single state, due to absorption of light energy, and the excess
energy is not immediately dissipated by collision with other molecules or
by other means, light will be emitted at a lower frequency as the electron
returns to its ground state and the substance is said to fluoresce. As some
energy is always lost before emission occurs then, in contrast to Raman
scattering, the wavelength of the fluorescent light is always greater than
the incident light.

Detection techniques based on fluorescence affords greater sensitivity to


sample concentration, but less sensitivity to instrument instability, (e.g.
sensor temperature and pressure). This is due to the fluorescent light
being measured against a very low light background (i.e., against a very
low noise level). This is opposite to light absorption measurements where
the signal is superimposed on a strong background signal carrying a high
noise level. Unfortunately, relatively few compounds fluoresce in a
practical range of wavelengths. However, some compounds, including
products from foods, drugs, dye intermediates etc., do exhibit
fluorescence and can be monitored by fluorescent means. In addition,
many substances can be made to fluoresce by forming appropriate
derivatives.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
59

The optical system of most fluorescent detectors is arranged such that


the fluorescent light is viewed at an angle to the exciting incident light
beam. This minimizes the amount of incident light that can interfere with
the fluorescent signal. Under such circumstances, the fluorescent signal is
viewed against a an almost black background and thus, furnishes the
maximum signal to noise. A filter can be used to reduce the background
light still further by the removal of any stray scattered incident light.

The fluorescence signal (If) is given by

I f = f I o 1 – e -k c l
( )
where (f) is the quantum yields (the ratio of the number of photons
emitted and the number of photons absorbed),
(Io) is the intensity of the incident light,
(c) is the concentration of the solute,
(k) is the molar absorbence,
(l) is the path length of the cell.

Fluorescence detectors can be simple or complex, the simplest consists of


a single wavelength excitation source and a sensor that monitors
fluorescent light of all wavelengths. For certain samples, this form of
fluorescence detector can be very sensitive and relatively inexpensive.
However, employing excitation light of a single wavelength and only a
broad emission wavelength, it is not very versatile. Conversely, the
fluorescence spectrometer fitted with a small sensor cell is far more
complex but with both selectable excitation wavelengths and emission
wavelengths is extremely versatile. In addition, excitation and emission
spectra can be obtained as required.

The Single Wavelength Excitation Fluorescence Detector

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
60

The single wavelength excitation fluorescence detector is probably the


most sensitive LC detector that is available, but is achieved by forfeiting
versatility. A diagram of a simple form of the fluorescence detector is
shown in figure 36. The excitation light is normally provided by a low
pressure mercury lamp which is comparatively inexpensive and provides
relatively high intensity UV light at 253.7 nm. Many substances that
fluoresce will be excited by light of this wavelength.
Photo Quartz
Cell Window
Quartz
Window
Lens

Excitation
Light
Fluorescent From UV Light
Light Column To Waste Source

Figure 36. The Single Wavelength Excitation Fluorescent Detector


The excitation light is focused by a quartz lens through the cell. A
second lens, set normal to the incident light, focuses the fluorescent light
onto a photo cell. A fixed wavelength fluorescence detector will have a
sensitivity (minimum detectable concentration at an excitation
wavelength of 254 nm) of about 1 x 10-9 g/ml and a linear dynamic
range of about 500 with a response index of 0.96 < r <1.04.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
61

11
2

10
4 6

12
3 8

9
1 14
7
13 15

Column: 2 Pecosphere™–5C C18 (150 mm x 4.6 mm) in series. Mobile Phase: 90%
acetonitrile/10% water. Flow rate: 2.0 ml/min. Detector Fluorescence (Excitation 254
nm total emission sensed). Sample: 20 ml of NBS Standard.
1. Naphthalene 2. Fluorene 3. Acenaphthene
4. Phenanthrene 5. Anthracene 6. Fluoranthracene
7. Pyrene 8. Benzo(a)anthracene 9. Chrysene
10. Benzo(b)fluoranthene 11. Benzo(k)fluoranthene 12. Benzo(k)fluoranthene
13. Dibenz(a,h)anthracene 14.Indeno(1,2,3,cd)pyrene 15. Benzo(ghi)perylene

Courtesy of the Perkin Elmer Corporation


Figure 37. Separation of the Priority Pollutants Monitored by the
Simple Fluorescence Detector

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
62

An example of a separation monitored by a simple fluorescence detector


is the separation of a mixture of priority pollutants shown in figure 37.
The excitation light was approximately monochromatic at 254 nm and
all wavelengths of the fluorescent light was sensed by the photo cell.
There are some compromises between the expensive fluorescence
spectrometer detector and the single wavelength excitation fluorescence
detector. Some have a single monochromators that select the wavelength
of the excitation light, others employ a single monochromator to select
the emission wavelength or provide emission spectra.
The Multi Wavelength Fluorescence Detector

The multi wavelength fluorescence detector contains two


monochromators, one to select the excitation wavelength and the second
to select the fluorescence wavelength or produce a fluorescence
spectrum A diagram of the multi wavelength fluorescence detector is
shown in figure 38.
The detector comprises a fluorescent spectrometer fitted with suitable
absorption cell that is sufficiently small so as not to degrade the
resolution of an LC column. There are two distinctly different light paths
one for the excitation light and one for the emitted light. The different
light paths are depicted separately, the excitation light in dark blue and
the emitted light in light blue.

The broad band excitation source (usually a deuterium lamp) is placed at


the focal point of an ellipsoidal mirror (shown at the top left hand corner
of the diagram). The resulting parallel light beam falls on a toroidal
mirror that focuses it onto the grating on the left-hand side of the
diagram. This grating selects the wavelength of the excitation light. Light
of the selected wavelength passes to a spherical mirror and then to a
ellipsoidal mirror (shown at the base of the diagram) which focuses it
onto the sample. The excitation light path is in dark blue and is situated
on the left-hand side of the diagram. A beam splitter is situated between
the spherical mirror and the ellipsoidal mirror (in the center of the

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
63

diagram) which reflects a portion of the incident light onto another


toroidal mirror which focuses it onto the reference photo cell.

Excitation Toroidal
Light Source Mirror Photo
Ellipsoidal Cell
Mirror Spherical
Mirror
Toroidal
Mirror

Spherical
Mirror

Beam
Splitter Reference
Cell

Grating that Selects the


Wavelength of the
Excitation Light
Grating that Selects the
Wavelength of the
Ellipsoidal Sample Emitted Light
Mirrors Cell

Figure 38. The Fluorescence Spectrometer Detector

The path of the fluorescent light is in light blue and is largely on the right
hand side of the diagram. Fluorescent light from the cell is focused by an
ellipsoidal mirror on to a spherical mirror which then focuses the light
onto a grating situated (seen at about center right of the figure). The
grating selects the specific wavelength of the fluorescent light to be
monitored. Light of the selected wavelength passes to a photoelectric cell
which monitors its intensity.
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
64

The instrument is quite complex but is extremely versatile. The use of


the detector to optimize both the excitation light and the fluorescence
light to provide high selectivity for the Fluoropa derivative of neomycin
is shown in figure 39. It is a very good example of the selection of a
specific excitation light wavelength and the complementary emission
light wavelength to provide maximum sensitivity.

Neomycin

0 2 4 6
Time (minutes)

Column: Supercosil LC–8, 15 cm x 4.6 mm, 5 m m particles: Mobile Phase:


tetrahydrofuran : 0.0056M sodium sulfate/0.007M acetic acid/0.01M pentane
sulfonate, 3:97. Flow rate: 1.75 ml/min. Post Column reagent: 1L 0.4M boric
acid/0.38M potassium hydroxide containing 6 ml 40% Brij–35, 4 ml mercaptoethanol,
0.8g o-phthalaldehyde. Flow rate 0.4 ml/min. Mixer 5 cm x 4.6 mm column packed
with glass beads. Reactor 10 ft x 0.5 mm knitted Teflon capillary tubing. Reaction
Temperature 40oC. Sample 20 ml of a mobile phase extract of a commercial sample.
Excitation wavelength 365 nm;emission wavelength 418 nm.

Courtesy of Supelco Inc.

Figure 39. Detection of Neomycin OPA Derivative at an Excitation


Wavelength of 365 nm and an Emission Wavelength of 418 nm
Optimizing excitation and emission light wavelengths to obtain
maximum sensitivity for a complex mixture can become quite involved
as shown by the separation of some priority pollutants depicted in figure
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
65

40. The separation was carried out on a column 25 cm long, 4.6 mm in


diameter and packed with a C18 reversed phase.

5
Fluorescence Intensity

2
1

9 11
8 10 14
4

12 13
3 6
7
15

Time in Seconds

1 Naphthalene 9 Chrysene
2 Acenaphthene 10 Benzo(b)fluoranthene
3 Fluorene 11 Benzo(k)fluoranthene
4 Phenanthrene 12 Benzo(a)pyrene
5 Anthracene 13 Dibenz (a,h)anthracene
6 Fluoranthene 14 Benzo(ghi)perylene
7 Pyrene 15 Indeno(123-cd)pyrene
8 Benz(a)anthracene
Fluorescence Detector Program
Time Wavelength of Wavelength of
(seconds) Excitation Light Emitted Light
0 280 nm 340 nm
220 290 nm 320 nm
340 250 nm 385 nm
510 260 nm 420 nm
720 265 nm 380 nm
1050 290 nm 430 nm
1620 300 nm 500 nm

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
66

Figure 40. Separation of a Series of Priority Pollutants with


Programmed Fluorescence Detection
The mobile phase was programmed from a 93% acetonitrile, 7% water
to 99% acetonitrile, 1% water over a period of 30 minutes. The gradient
was linear and the flow rate was 1.3 ml/min. The separation illustrates the
clever use of wavelength programming to obtain the maximum
sensitivity. During development both the wavelength of the excitation
light and that of the emission light were changed to provide maximum
sensitivity for the particular solute.
The detector can provide fluorescence or excitation spectra by arresting
the flow of mobile phase when the solute resides in the detecting cell and
scanning either the excitation or fluorescent light. (This is the same
technique as that used to provide UV spectra with the variable
wavelength UV detector). As a consequence, it is possible to obtain
excitation spectra at any chosen fluorescent wavelength or fluorescent
spectra at any chosen excitation wavelength. Thus, even with relatively
poor spectroscopic resolution many hundreds of spectra can be
produced, any or all of which (despite many spectra being very similar)
can be used to confirm the identify a compound.

Transport Detectors

A transport detector consists of a carrier such as a metal chain, wire or


disc that passes continually through the column eluent extracting a
sample of the mobile phase containing the solute as a thin film adhering
to its surface. The mobile phase is eliminated by evaporation leaving the
solute as a coating on the carrier. The carrier is then scanned by an
appropriate sensing technique to monitor the residual solute. For
example, an FID could be used to sense the pyrolysis products of the
solute by heating the carrier and most of the pyrolysis products
containing carbon would be detected. The system is obviously restricted
to those solutes that are involatile and, in addition, the solvents used for
the mobile phase must be volatile (and extremely pure). The former
condition is usually met in LC, otherwise the analysis would probably be

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
67

carried out by GC. The latter condition is normally easy to achieve as


there is a wide choice of solvents readily available for LC.
The system appears to be ideal but the early models had some
disadvantages. The instruments were bulky, expensive and some
incorporated a 90strontium source (for the argon detector) all of which
were unpopular. In addition, as a result of the basic design, the
anticipated high sensitivity was not realized and the apparatus was
clumsy and difficult to operate. However, as it was a universal detector
and was unaffected by the solvents used, it was readily accepted by the
soap and cosmetic industry.
The Moving Wire Detector

The original wire transport detector was developed by James et al. (33)
and subsequently manufactured and marketed by Pye Unicam. A
diagram of the Pye Unicam detector is shown in figure 41.

Oxygen Detector
Evaporator Oven
Oven
From
Column Hydrogen
FID
Coating Nitrogen Nitrogen
Block
Pyrolysis
Oven

Nitrogen
Flow Restrictors

Feed Collection
Spool Spool
Cleaner Oven

Figure 41 The Pye Unicam Moving Wire detector


The carrier wire from a spool passes through an oven at 750˚C and any
residual lubricants remaining on the wire after drawing are burnt off.
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
68

After passing round a pulley the wire then goes through a coating block
where it is coated with a thin film of mobile phase. The wire then passes
through an evaporator at about 105˚C (or a temperature appropriate for
the solvent(s) employed). In the evaporation chamber a nitrogen stream
is allowed to flow counter-current to the wire movement to improve the
rate of solvent evaporation. The wire carrying the solvent-free solute
then enters (via a restriction) a pyrolysis tube that is maintained at about
500˚C. The pyrolysis tube has a nitrogen supply entering at either end
which sweeps the tube contents, including the pyrolysis products that are
formed, out through a center tube into the FID.
The FID was a standard detector and the device functioned well and, as
expected, was completely independent of the nature of the solvents used
for the mobile phase. The detector sensitivity was disappointing, and
found to be little better than the average refractive index detector viz. 5
x 10-6 g/ml. The poor sensitivity resulted from excessive noise (not a
weak signal) which may have been caused by high boiling impurities in
the solvents, fluctuations in the nitrogen flow and/or irregularities in the
pyrolysis process. The linear dynamic range was also found to be less
than two orders of magnitude. However, the device did establish the
viability of the transport method for LC detection.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
69

Synchronous Gold Chain


Motor

Evaporation
Tunnel

LC Column Coating
Gauze Block
Jet FID
Air Inlet

Figure 42. The Chain Detector


The Chain Detector

About the same time as James et al. developed the wire transport
detector, Haahti and Nikkari (34) described a similar device, simpler in
design, that employed a chain loop in place of the wire as a carrier. A
diagram of their apparatus is shown in figure 42. The carrier, a gold
chain driven by a synchronous motor, passes through a coating block
where it is wetted with the column eluent. The chain then enters an
evaporator tunnel, is heated, and the solvent volatilized leaving the solute
deposited on the chain. The chain then exits the tunnel into the actual
flame of an FID.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
70

Surfactant

Mineral Oil

Ethyl n-Heptane
Alcohol

Sample: mineral oil and a surfactant, solvent: n-heptane/ethyl alcohol, column: 2 x 300
mm, column packing: silica gel, flow rate: 0.7 ml/min., chart speed: 24 cm/min.,
evaporator temperature: 150˚C, nitrogen flow: 30 ml/min., hydrogen flow rate: 25
ml/min., oxygen flow rate: 30 ml/min.

Figure 42. Chromatogram Obtained from the Chain Detector


Combustion takes place, ions are produced in the expected manner, and
the ion current processed in the usual way. Unfortunately, due to the
occlusion of local, high concentrations of solute between the links of the
chain, the detector output was extremely noisy and thus the system
exhibited a relatively poor sensitivity. A chromatogram obtained with the
chain detector is shown in figure 42. The noise spikes on each peak are
obvious, which, besides affecting the overall sensitivity of the detector,
makes quantitative analysis extremely difficult.

The Modified Moving Wire Detector


This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
71

In the early 1960s, a number of attempts were made to improve the


performance of the transport detector. In 1966 Karmen (35) introduced
an aspirating system to draw the pyrolysis products into the hydrogen
flame detector. Later in 1970, Scott and Lawrence (36) developed the
Karman's system further and introduced a modified form of the original
moving wire detector.

The sensitivity of the original detector, in addition to being degraded by


the high noise level, was also limited by the proportion of the pyrolysis
products that entered the FID. Excepting synthetic polymers, (which
often quantitatively produced monomers) many compounds yielded only
a few percent of volatile compounds on pyrolysis. Thus the FID could
only sense a very small fraction of the products from the solute
deposited on the wire. If the solutes, were completely combusted in
oxygen or air, however, then all the carbon in the solute would be
converted to carbon dioxide. Furthermore, if the carbon dioxide was
then reduced to methane by mixing it with hydrogen and passing it over
a nickel catalyst, the carbon dioxide would be quantitatively converted to
methane which could be detected by the FID. This procedure would
increase the sensitivity of the detector to those substances that gave poor
yields on pyrolysis and, in addition, increase the linear dynamic range
and possibly provide a predictable response.

A diagram of the original moving wire detector modified in this way is


shown in figure 43. The hydrogen lines entering the detector body were
enlarged to reduce the flow impedance and permit the satisfactory
operation of the aspirator. The detector was connected to a 2 in. length
of 1/2 in I.D. thin walled stainless steel tube filled with about 2 g of nickel
catalyst and then to the aspirator. The wide tube was closed with a loose
plug of quartz wool and .

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
72

Air Detector
Evaporator Nickel
Oven Catalyst
Oven
From Molecular
Column Entrainer
FID Hydrogen/Argon
Coating Oxygen Oxygen
Block

Oxygen Flow Restrictors

Collection
Feed
Oxidation Spool
Spool
Oven Cleaner Oven

Figure 43 The Modified Moving Wire Detector


The nickel catalyst was prepared by absorbing a saturated solution of
nickel nitrate onto 20/40 BS mesh brick dust, decomposing the nitrate at
500˚C for 3 hours followed by reduction of the nickel oxide so produced
to metal in a stream of hydrogen at 250˚C. The jet/venturi aspirator
(supplied commercially as a molecular entrainer) was placed in line with
the hydrogen flow to the detector.

To improve the aspirating efficiency, the gas used was a mixture of


hydrogen and argon, and with this mixture, the jet/venturi pressure drop
continuously sucked the combustion gasses into the hydrogen stream. In
figure 43 the reduced pressure side of the aspirator is shown connected
to the side limb of the oxidation tube and the two tube system used in
the original moving wire detector, was replaced by a single tube. The
oxygen or air is fed into the center of the tube and, thus, provided both
the evaporator flow and the oxidation flow. All tubes were constructed
of quartz. The linear dynamic range of the system was determined to be
about four orders of magnitude a considerable improvement on the
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
73

pyrolysis instrument. The response index for a series of compounds of


different chemical types ranged from 0.96 to 1.04. and the actual
response was found to be proportional to the carbon content of the
solutes.

Solvent 1 2 3 4 5 7 8 9 10 11 12

Injection

Figure 44 The Separation of Blood Liquids Employing


Incremental Gradient Elution and Monitored by the Modified
Moving Wire Detector

However, due to the limited number of compounds that were tested this
relationship should be assumed only with caution. A chromatogram of
blood lipids obtained by incremental gradient elution and monitored by
the modified detector is shown in figure 44. As incremental gradient
elution involves a program of 12 solvents ranging from hydrocarbons,
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
74

chlorinated hydrocarbons, nitro-paraffins, esters, ketones and alcohols.


This separation illustrates the versatility that is provided by this detector
for solvent selection.

Van Dijk (37), attempted to improve the sensitivity of the detector by


using a spray procedure for coating the wire. The column eluent entered
an atomizer and spray from the nozzle was directed onto the wire. The
spray partly concentrated the solute by evaporation and also increase the
load on the wire. A linear dynamic range of about 3 x 103 was claimed
for the system and a sensitivity increase of 50 over the original wire
transport detector (probably about 3 x 10-6 g/ml).
Yang et al. (38) also used a thermal spray for coating the wire and also
claimed an increased sensitivity. A heated chamber (through which the
conduit from the column passed) was placed above a moving stainless
steel belt. The solvent was rapidly brought to its boiling point resulting in
spray leaving the exit of the conduit and coating the belt. Yang et al.
also employed a photo-ionization detector and an electron capture
detector as alternatives to the FID.

Compton and Purdy (39) redesigned the Pye Unicam FID by inserting a
rubidium silicate glass bead above the flame and thus made its response
specific and changed it into a nitrogen phosphorus detector. Stolyhwo et
al. [40] employed metal spirals wound on wire and stranded wire to
increase the surface area of the carrier to increase the proportion of the
column eluent taken into the detector. A detectable mass of 100 ng of
triolei was claimed but the actual concentration sensitivity was not
reported. Pretorious and Van Rensburg (41) tried to increase the carrier
take-up by coating the wire with sodium silicate, kaolin and copper
kaolin. some improvement to appears to have been realized. Slais and
Krejei (42) replaced the normal FID with the NPD detector and used it
to detect chlorine compounds. They mixed the combustion gases with
hydrogen and passed the mixture directly into the NPD. At a column
flow rate of 0.37 ml/min., the sensitivity of the detector was stated to be
about 3 x 10-7 g/sec, which is equivalent, in concentration units, to about
1.6 x 10-6 g/ml.
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
75

The moving wire detector has also been modified by Dugger (43) for
radioactivity detection (e.g., detection of 14 carbon labeled compounds).
To detect 14 carbon compounds, the solute on the wire was oxidized to
carbon dioxide and the radioactive gas passed to a Geiger-Muller tube.
To detect tritium, the tritiated water produced on combustion was passed
over heated iron to reduce it to hydrogen and tritium, which was then
also passed to a Geiger-Muller tube.

More recently De Peña et al. (44, 45) have developed the transport
detector still further. A photograph of their detector, produced by
Scientific Detectors Ltd., is shown in figure 45.

Figure 45. The Uni-Mass Transport Detector


The major novel feature of this transport detector is the use of an
oxidized titanium tape as the carrier which has a greatly improved
loading capacity and readily wets with both dispersive and highly polar
solvents (e.g., n-heptane and water). In addition, the coating, evaporating
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
76

and pyrolysis system has been made much more compact and the argon
detector (which is used as the sensor) is made physically very close to
the pyrolyser (ca 2-3 cm) A photograph of the internal layout of the
detector is shown in figure 46.

Figure 46. The Internal Lay-Out of the Uni-Mass Detector

The performance of this detector remains to be established. It has found


great interest in biotechnology field as it responds to all substances and
the relative responses to different compounds appear to be within a
factor of 5 and thus all components are detected.

This detector should have a theoretical limit of detection of about 1x10-9


g/ml (i.e. similar to the fluorescence detector) but this sensitivity has not
been realized yet.
The Disc Detector

The disc detector was originally described by Dubsky (46) who used a
rotating gauze disc as the carrier. A diagram of the device is shown in
figure 47. The rotating disc has a perimeter made of wire gauze and the
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
77

column exit is situated just above the gauze so that the eluent flows
through the gauze coating it with a film of mobile phase. The excess of
mobile phase is collected in a container situated below the disc. An
infrared lamp, or some other appropriate heater, is placed a little ahead
of the point of coating, in the direction of rotation of the disc which
evaporates the solvent leaving the solute coated on the gauze.

Gauze
Disc
Column Heater

Electrodes

Synchronous
Motor
Flame FID

Figure 47. The Disc Detector


The FID is placed diametrically opposite to the point of coating and the
flame jet is situated beneath the gauze in such a manner that the flame
itself is in contact with the gauze.

The FID electrodes are placed just above the gauze, directly over the
flame. The ion current is collected by the electrodes and amplified in the
usual way and fed to a recorder or data acquisition system. This
arrangement is simple compared with the conventional wire transport
detector but although a tenfold increase in sensitivity is claimed, this is
difficult to confirm from the published data given. Szakusito and
Robinson (47) reported that the metal gauze disc carrier was a source of
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
78

excessive noise. Specifically, significant "spikes" were caused by local


concentrations of solute accumulating at the intersections of the wire
mesh during evaporation. These authors employed an alumina disc, 4.5
in diameter, with the edge tapered to 0.25 mm thick. The thin edge was
used for coating and detection. It appeared that there a significant
reduction in noise, but again, sensitivities were not given in terms that
would allow comparison with other detectors. It appeared that in
continual use, the pores of the alumina could become blocked with
residue from incompletely combusted solutes or mobile phase
components and so the life of the alumina disc may well be limited. The
disc does appear to be a simpler transport system than the wire or chain
but its reliability and sensitivity remain to be established.

The Evaporative Light Scattering Detector

Nebulizer Gas Eluent from Column

Nebulizer

Photomultiplier
Optical Fibers
Light Source
Light Trap

Fan

Exhaust

Courtesy of Polymer Laboratories Inc.


Figure 48. A Diagram of the Evaporative Light Scattering
Detector

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
79

The evaporative light scattering detector, as its name implies, utilizes a


spray that continuously atomizes the column eluent into small droplets.
These droplets are allowed to evaporate, leaving the solutes as fine
particulate matter suspended in the atomizing gas.

The atomizing gas can be air or, if necessary, an inert gas. The
suspended particles pass through a light beam and the scattered light
viewed at 45o to the incident light beam by means of a pair of optical
fibers. The scattered light transmitted through the fibers is sensed by a
photomultiplier and the output electronically processed and passed either
to a computer data acquisition system or to a potentiometric recorder. A
diagram of the light scattering detector is shown in figure 48.

Eluent and
Atomizing Gas

Atomized Stream
of Particles

Collimated To Photocell
Light Beam

Courtesy of Polymer Laboratories Inc.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
80

Figure 49 The Sensor of a Commercial Evaporative Light


Scattering Detector
Theoretically, the detector responds to all solutes that are not volatile and
as the light dispersion is largely Raleigh scattering, the response should
be proportional to the mass of solute present; as a consequence, it is
sometimes referred to as the mass detector. For a linear response, the
droplet size must be carefully controlled as it also determines the particle
size of the dried solutes. Detector sensitivity is claimed to be between 10
and 20 ng of solute. However, in these terms it is difficult to compare
with other detectors. The great advantage of the detector, however, is its
catholic response and that its output is linearly related to the mass of
solute present. However, the magnitude of the response does vary
widely between different substances.
A diagram of the sensor of the evaporative light scattering detector
manufactured by Polymer Laboratories is shown in figure 49.

Courtesy of Polymer Laboratories Inc.


Peak Compound Mass (mg) Retent.Time(min).

1 cholesterol ester 5 0.717


2 triglyceride 18 1.746
3 cholesterol 10 4.687
4 unknown 10 8.860
5 phosphatidyl choline 10 10.028
6 phosphatidyl - 17.390
ethanolamine

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
81

Figure 50. The Separation of Some Lipid Class Materials


Monitored by a Evaporative Light Scattering Detector
The eluent is atomized in a stream of nitrogen and the finely divided
spray passes down a heated chamber during which time the solvent is
evaporated. The removal of the solvent produces a stream of particles
which then pass through a collimated beam of light. The scattered light
at an angle to the incident light is focused onto a photomultiplier tube
and the output is processed in an appropriate manner electronically. The
device is fairly compact and relatively simple to operate. An example of
the results obtained when used for monitoring a general lipid class
analysis is shown in figure 50. The minimum detectable mass estimated
from this chromatogram appeared to be about 10 ng of solute. To some
extent, this detector provides a replacement for the transport detector as
it detects all substances irrespective of their optical or electrical
properties.

Liquid Light Scattering Detectors

Light scattering detectors differ from evaporative light scattering


detectors in that they respond to the light scattered by a polymer or
large molecular weight substance present in the column eluent itself. The
scattering is measured as it passes through an appropriate sensor cell
while illuminated by a high intensity beam of light. The high intensity
light source is achieved by the use of a laser (light amplification by the
stimulated emission of radiation) that also generates the light at the
appropriate wavelength for measurement. There are two forms of the
detector: the low angle laser light scattering (LALLS) detector and the
multiple angle laser light scattering (MALLS) detector. Both devices
are commonly used but the multiple angle laser light scattering detector
is more versatile as it provides molecular dimensions as well as the
molecular weight of the eluted solute.

In the LALLS detector, the scattered light is measured at a very small


angle to the incident light (virtually 0o), and consequently the signal can
be affected by scattering from contaminating particulate matter present
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
82

as impurities in the eluent. This can cause significant noise, and thus,
reduce the detector sensitivity. Discussions on the subject have been
given by Wyatt (48) and some early experimental results reported by D.
T. Phillips(49).

The ratio of the intensity of the light scattered at an angle (f), (If) to the
intensity of the incident light (Io), for Rayleigh light scattering is given by
If
= a w Rf
Io

where, (a) is the attenuation constant,


(w) is a function of the refractive index,
and (Rf) is Rayleigh's constant

If
Thus, Rf =
a w Io

Now the molecular weight (Mw) of the solute is related to the Rayleigh
factor by the following expression,
Rf
Mw =
c(K - 2A 2 Rf)

where (c) is the concentration of the solute,


(A2) is a function of polymer-polymer interactions,
and (K) is the polymer optical constant.

Substituting for (Rf)


If
a w Io If
Mw = =
Ê
cÁ K - 2B 2
If ˆ
˜
(
c a w I o K - 2B 2 I f )
Ë a w Io ¯
where
2 p 2 h2
K =
Ê dhˆ 2
l4 N Á ˜
Ë dc ¯

where (h) is the solvent refractive index,


This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
83

(l) is the wavelength of the light in vacuum,


and (N) is Avogadro's number.

Thus, a basic relationship can be developed between the molecular


weight of the scattering material, the intensity of the scattered light and
the physical properties of the materials and equipment being involved.
However, constants are in used, the magnitude of which are difficult to
determine. In practice a simple graphical procedure is used to determine
the molecular weight of the solute without the need to determine all the
pertinent constants. Rearranging the equation

1 c(K - 2A 2 Rf) cK
= = - 2cA 2
Mw Rf Rf

cK 1
or = 2cA 2 +
Rf Mw

Now (c), (K), and (Rf) are either known or can all be calculated from
Ê cK ˆ
known data and light scattering measurements; thus, by plotting Á ˜
Ë Rf ¯
against (c) a straight line will be produced with the intercept being
Ê 1 ˆ
Á ˜.
Ë Mw ¯

The Low Angle Laser Light Scattering Detector

The optical system of the low angle laser light scattering detector
produced by LDC Analytical of the Thermo Instruments Corporation is
shown diagramatically in figure 51. To conserve space, a folding prism is
used that allows the device to be contained to a reasonable size yet
accommodate the length of the laser generator. Light from the laser
passes through a diverging lens, through a chopper and then through the
folding prism. On leaving the prism the beam passes through some
measuring attenuators and a calibrating attenuator shutter and then
through the cell. An annular mask is situated between the cell and the
relay lens and only allows light scattered in the cell at a low angle to pass
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
84

to the relay lens. Between the annular shutter and the relay lens is a
safety attenuator that ensures that none of the potentially damaging laser
light can reach the photomultiplier. The scattered light is focused through
a field stop onto the forward detector lens. Between the field stop and
the forward detector lens is a prism that allows the scattered light to be
viewed through a microscope.
Diverging Lens
Chopper
LASER
Folding Prism
Photomultiplier
Rear
Measuring Forward Detector
Attenuators Annulus Detector Lens
Mask Lens Sensor
Apertures Aperture

Cell
Windows
Opal
Condensing Diffuser
Lens Relay
Lens Viewing
Calibrating
Safety Microscope
Attenuator/ Analyzer Red
Shutter Attenuator Field Filter Polarizer Filter
Stop Holder

Courtesy of LDC Analytical, Thermo Instruments Corporation.

Figure 51 Optical Diagram of a Low Angle Laser Light Scattering


Detector
Between the forward detector lens and the rear detector lens is A filter
holder and an analyzer/polarizer. Finally the light is focused through a
sensor aperture to an opal diffuser that spreads the scattered light
through a red filter and onto the photo- multiplier.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
85

The device is frequently used with a refractive index detector in series to


coincidentally measure the refractive index of the eluent. This is
necessary to calculate (K) from the refractive index. As discussed, the
molecular weight of a solute is determined from the intercept of the
cK
graph relating ( ) to the solute concentration (c) as shown in figure
Rf
52. The concentration calculated from the refractive index detector
employing the data from prior calibration.

cK cK
Rf 2A 2 Rf

1
Mw

Figure 52 Determination of Molecular Weight from Low Angle


Light Scattering Measurements
The detector sensitivity appears to be similar to that of the refractive
index detector and with about the same linearity. However, the greatest
advantage of this detector is that it can provide molecular weight data
for extremely large molecules.
The Multiple Angle Laser Light Scattering (MALLS) Detector

The multiple angle laser light scattering detector differs from the low
angle device, in that scattering measurements are made at a number of
different angles, none of which are close to the incident light. This
significantly reduces problems associated with light scattering from
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
86

particulate contaminants. Data taken at different angles to the incident


1/ 2
light allows the root-mean-square (rms) of the molecular radius r 2
to be calculated in addition to the molecular weight of the substance. The
relationship that is used is as follows:
cK 1/ 2
= a r2 sin (q )2 + b M w
Rf
Theory can provide explicit functions for (a) and (b) but these constants
are obtained in practice by the use of calibrating standards of known
molecular weights and molecular radii. In addition, a photocell will not
have precisely the same response to low light intensities and calibration
procedures are also necessary to account for their different responses.

The number of different angles of measurement differs with different


instruments, and some measure the scattered light intensity at 16
different angles. In general, the more data points taken at different
angles, the more precise the results will be. A diagram of a (MALLS)
detector system which measures the light scattered at three different
angles is shown in figure 53.

Sensor 2

Sensor 1 Sensor 3

LASER
Beam
LASER

Glass Cell

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
87

Courtesy of Wyatt Technology Corporation


Figure 53 The Multiple Angle Laser Light Scattering Detector
(miniDawn®)
This device (the miniDawn®), manufactured by Wyatt Technology
Corporation, contains no mirrors, prisms or moving parts and the light
paths are direct and not "folded". Light passes from the laser
(wavelength 690 nm) directly through the sensor cell. Light scattered
from the center of the cell passes through three narrow channels to three
different photocells, set at 45o and 90o and 135o to the incident light.
Thus scattered light is continuously sampled at three different angles
during the passage of the solute through the cell.
A continuous analog output is provided from the 90o sensor and all the
sensors are sampled every 2 sec. The molecular weight range extends
from 10 3 to 10 6 Daltons and the rms radii from 10 to 50 nm. The
total cell volume is about 3 ml and the scattering volume is 0.02 ml. The
detector has a sensitivity, defined in terms of the minimum detectable
excess Rayleigh ratio of 5 x 10-8 cm-1. This is difficult to translate into
normal concentration units but appears to be equivalent to a minimum
detectable concentration of about 10-6 g/ml.

Kc
R(q )
Slope Gives a Value
for the rms Molecular
Radius

Intercept gives a
Value for the
Molecular Weight

0 0.5 1.0
sin 2( q/2)

Figure 54. Calibration Curves


This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
88

The relationship between the intensity of the scattered light, the


scattering angle and the molecular properties are as follows:
cK 1
= 2cA 2 + where P(q) describes the
Rf M w P(f)
dependence of the scattered light on the angle of scatter and the other
symbols have the meanings previously attributed to them.
The relationship between the angle of scattering, (q), the molecular
weight and the rms molecular radius of the solute is obtained using the
appropriate equation. Employing suitable reference materials, graphs of
the form shown in figure 54 can be constructed to evaluate constants (a)
and (b) and thus permit the measurement of the molecular weight and
molecular radius of unknown substances.

The Electrical Conductivity Detector

The electrical conductivity detector measures the conductivity of the


mobile phase. There is usually background conductivity which must be
backed-off by suitable electronic adjustments. If the mobile phase
contains buffers, the detector gives a base signal that completely
overwhelms that from any solute usually making detection impossible.
Thus, the electrical conductivity detector a bulk property detector. and
senses all ions whether they are from a solute or from the mobile phase.
In order to prevent polarization of the sensing electrodes, AC voltages
must be used and so it is the impedance not the resistance of the
electrode system that is actually measured. From a physical chemistry
stand point the conductivity of a solution is more important than its
resistance. However, it is the resistance (impedance) of the electrode
system that determines the current across it.

The resistance (R) of any conductor varies directly as its length (L) and
inversely as its cross sectional area (a).
rL
Thus, R =
a
where (r) specific resistance of the conductor.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
89

The specific resistance of a conductor is the resistance across two


opposite faces of a 1 cm cube made of the conductor material. The
specific conductance (k) of a solute is the reciprocal of the specific
resistance, i.e.,
1
k =
r
Thus the conductance of a given solution (C) is given by,
1 ka
C = = ohms -1
R L
The conduction of all the ions produced by 1 g equivalent of an
electrolyte at any particular concentration can be evaluated by imagining
two large parallel electrodes, 1 cm apart, and the whole of the solution
placed between them. The conductivity of the system is called the
equivalent conductance (L).
If 1 g equivalent of the electrolyte is dissolved in (v) ml of solution; then
it follows that this will cover (v) cm2 of electrode area. It follows that in
the above equation (a) becomes (v) and (l) is unity and
1000k
L = ohms -1 cm 2
c
where (c) is the electrolyte concentration in gram equivalents /liter.

Lc
or k =
1000
Thus,
1 L ca
C = = ohms -1
R 1000 L
Thus, the cell resistance is inversely proportional to the electrolyte
concentration (c), the equivalent conductance of the electrolyte and the
cell geometry [(a) and (L)]. Thus the cell output (which is, in fact
proportional to the resistance change of the cell) must be modified to
provide an output linearly related to concentration.

The basic sensor, being so simple, has changed little from that first
described (1). The sensor volume has been reduced but still consists of
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
90

two electrodes situated in the column eluent, the resistance (or strictly the
impedance) between which is measured by an appropriate electronic
circuit. Figure 55 shows three simple forms of the conductivity sensor
although many other geometric forms have been described in the
literature. The basic sensor is, indeed, very simple, the top design is easily
constructed and can have an effective sensing volume of only a few
microliters.

Electrodes
From
Column To Waste

PTFE Insulating Sensor PTFE Waste


Tube Volume Tube

From To Waste
Column

Stainless Steel Stainless Steel


Tube. (Electrode Tube. (Electrode
Grounded) to Amplifier)

Sensor
Volume PTFE Waste
PTFE Insulating Tube
Tube

From
Column To
Stainless Steel Waste
Tubes. (Isolated
Electrodes)

Figure 55. The Electrical Conductivity Detector


The center sensors are made from short lengths of stainless steel tubing
(about 0.020 in I.D. and 1/16 in O.D). The two tubes slide into a PTFE
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
91

tube and the space (inside the PTFE tube) between the two tubes is
sensing volume. The lower sensor is very similar, but two lengths of
stainless steel tube are employed, both of which are isolated from any
other metal connection by PTFE tubing. The two electrodes are, thus,
electrically isolated (this can be advanatgeous for certain types of
electronic measuring circuits). These sensors have sensing volumes of
only a few nanoliters and are, thus, often used in detectors for capillary
electrophoresis.
The electrical system consists of a frequency generator (1,000-5,000 Hz)
that provides an AC potential across the cell. The sensor is placed in one
arm of a Wheatstone bridge as shown in figure 56.

Sensing Cell

Out-of-Balance
Signal

AC Supply

If a single sensor is employed, it is situated in one arm of the bridge; if, however, a
reference sensor is also used, then the two sensors are situated in opposite arms of the
bridge. In either case the out–of–balance signal can be passed to a precision rectifier
and the output either handled by an analog circuit or sampled directly by the
computer. Sometimes a variable resistance is situated in one of the other arms of the
bridge and is used for zero adjustment to back-off any signal arising from ions
contained by the mobile phase.

Figure 56 One Form of the Wheatstone Bridge Used with the


Electrical Conductivity Detector
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
92

The out-of-balance signal is rectified and the DC signal either passed to a


nonlinear amplifier or a computer data acquisition system. The non-
linear amplifier is necessary to render the output linearly related to ion
concentration. If the output is passed directly to the computer, then the
linearization can be carried out with appropriate software and the
chromatogram is presented on the printer.
An example of a separation monitored with a conductivity detector
useing an ion supression technique (see book 15) is shown in figure 57.

Time 10 Minutes

Courtesy of Dionex Inc.

A proprietary ion exchange column, IonPacCS12, was used and the mobile phase
consisted of a 20 nM methanesulphonic acid solution in water. A flow rate of 1
ml/min was employed and the sample volume was 25 ml.

1. Lithium 2. Sodium 3. Ammonium


4. Potassium 5. Magnesium 6. Calcium

Figure 57. Determination of Alkali and Alkaline Earth Cations


The electrical conductivity detector has a sensitivity (minimum detectable
concentration) of about 5 x 10-9 g/ml and a linear dynamic range of
about 200 where 0.97 < r < 1.03. It is used in probably over 95% of all

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
93

analyses involving ion exchange procedures to separate inorganic and


organic ions.
The Electrochemical Detector

The electrochemical detector responds to substances that are either


oxidizable or reducible and the electrical output is an electron flow
generated by a reaction that takes place at the surface of the electrodes. If
the reaction proceeds to completion (exhausting all the reactant) the
current becomes zero and the total charge generated will be proportional
to the total mass of material that has been reacted. This process is called
coulometric detection. If, however, the mobile phase is flowing past the
electrodes, the reacting solute will be continuously replaced as the peak
passes through the detector. All the time there is solute present between
the electrodes, a current will be maintained, albeit varying in magnitude.
Until relatively recently, this procedure was that most common
employed in electrochemical detection and is called amperometric
detection.

The electrochemical detector requires three electrodes, the working


electrode (where the oxidation or reduction takes place), the auxiliary
electrode and the reference electrode (which compensates for any
changes in the background conductivity of the mobile phase). The
processes taking place at the electrode surface can be very complex;
nevertheless, the dominant reaction can be broadly described as follows.
At the actual electrode surface the reaction is extremely rapid and
proceeds almost to completion. This results in the layer close to the
electrode being virtually depleted of reactant. As a consequence, a
concentration gradient is established between the electrode surface and
the bulk of the solution. This concentration gradient causes solute to
diffuse into the depleted zone at a rate proportional to the solute
concentration in the bulk of the mobile phase. Thus, the current
generated at the electrode surface will be determined by the rate at
which the solute reaches the electrode and consequently, as the process
is diffusion controlled, will depend on solute concentration and the

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
94

magnitude of solute diffusivity. The response of the detector or current


(i) is described by the following equation,
i = n F A K T c ua

where (n) is the number of electrons per molecule involved in the


reaction,
(F) is the Faraday Constant,
(A) is the area of the working electrode,
(KT) is the limiting Mass Transfer Coefficient,
(c) is the solute concentration,
(u) is the linear velocity of the mobile phase over the surface
of the electrode,
and (a) is a constant usually taking a value between 1/3 and 1/2.

It is seen that the current (i) (and, thus, the sensitivity), can be raised by
either increasing the electrode area, increasing the transfer coefficient or
increasing the velocity of the mobile phase past the electrodes. It would
appear that increasing the electrode area would be the easiest, however,
increasing the electrode surface area while maintaining an amperometric
response also increases the noise, often to such an extent that there is an
overall reduction in detector sensitivity. Weber and Purdy (50) and
Hanekamp et al. (51) showed that under certain conditions a reduction
in the sensor size produces a significant increase in signal-to-noise and,
thus an increase in sensitivity. Higher flow rates will also increase the rate
of solute transfer and the sensitivity which would be an added advantage
to miniaturization. However, the sensor will be very flow sensitive and
thus the flow rate must be kept very constant and the detector would
not be amenable to flow programming development. Miniaturization
would also reduce peak dispersion in the sensor.
Electrode Configurations

The electrodes can take a number of different geometric forms which


have been described in some detail by Poppe (52). Some examples of
different electrode configurations are shown in figure 57. Examples (A)
and (B) are two common forms of thin-layer cells. (A) has the working
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
95

electrode sealed into the cell wall with the reference and auxiliary
electrodes situated down-stream to the working electrode. (B) is similar
to (A) but with the auxiliary electrode sealed in the wall of the tube
opposite the working electrode and, again, the reference electrode down-
stream. (C) is a wall-jet electrode where the eluent is allowed to impinge
directly onto the working electrode which is situated opposite the jet.
This arrangement not only increases the value of (u) (the velocity of the
liquid passing over the electrode and thus the transfer coefficient (KT))
but also scrubbs the surface of the working electrode helping to reduce
the need for frequent cleaning. (D) and (E) are two examples of
cylindrical electrodes; in (D) the working electrode is in the form of a rod
stretching across the diameter of the sensor cell and in (E) the working
electrode comprises an annular ring set in the cell wall.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
96

A B
Auxilary Reference
Auxilary and Reference Electrode Electrodes
Electrodes Downstream Downstream

Working Electrode Working Electrode


C D
Flow from
Column Auxilary and Auxilary and
Working
Reference Reference
Electrode
Electrodes Electrodes
Downstream Downstream

Working Electrode

Auxilary and
Working Reference
Electrode E
Electrodes
Downstream

Figure 57. Different Electrode Configurations


In both cases the auxiliary and the reference electrodes are situated
down-stream to the working electrode.
Electrode Construction

The choice of electrode construction material is restricted due to the


need for mechanically ruggeness and long term stability. The most
common material is carbon paste made from a mixture of graphite and
some suitable dielectric substance. This material has the disadvantage that
it is soluble in some solvents, although, using special waxes or polymers
as dielectric binders to contain the graphite, helps reduce the solubility
problem. Vitreous or 'glassy' carbon is an excellent electrode material
particularly if organic solvents are to be used and is probably the most
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
97

popular contemporary electrode material. Glassy carbon is produced by


slowly baking a suitable resin at elevated temperatures until it is
carbonized and then heating it to a very high temperature to cause
vitriation. Vitreous carbon is relatively pure, mechanically strong, has
good electrical properties and can be readily cleaned mechanically. It also
performs particularly well when operated at a negative potential. Glassy
carbon electrodes are preferable to carbon paste electrodes due to their
inherent resistance to solvents.
Electrochemical detection imposes certain restrictions on the type of
chromatography that is employed and the mobile phase that is used. The
detecting system requires a conducting mobile phase and thus must
contain water. Thus, the majority of 'normal phase' systems are not
usable. In addtion, very high solvent concentrations may render it
insufficiently conducting. Reversed phase chromatography, however, is
ideally suited to electrochemical detection. Nevertheless, certain
precautions must be taken for its effective use. The mobile phase must be
completely oxygen free which can be removed by bubbling helium
through the solvent reservoir. It is also important to remove oxygen
from the sample before making an injection. The solvents must also be
free of metal ions or serious base line instability will result. Non-aqueous
solvents such as pure acetonitrile can be employed but certain salts like
tetrabutyl-ammoniumhexafluorophosphate must be added to render the
solvent conducting.
Basic Electrochemical Detector Electronics
A simplified form of the circuit that is in general use is shown in figure
58. The amplifiers can best be considered as FET operational amplifiers
such as the TLO81. The auxiliary electrode is held at a fixed potential by
the first amplifier, the voltage being selected by the potentiometer (P)
that is connected to a regulated power supply. The current flowing
through the working electrode is processed by the second amplifier and
the output fed to the recorder or data acquisition system. The
electrochemical detector as described is extremely sensitive but has
certain disadvantages.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
98

Regulated Power
Supply

Amp.1 A

P
R W

Amp.2 Feed Back


Resistanc
e
Output

Figure 58 Basic Circuit Used with Electrochemical Detectors


The mobile phase must be extremely pure, oxygen free and devoid of
metal ions. A more serious problem arises from the adsorption of the
oxidation or reduction products on the working electrode surface. As a
result, the electrode system must be frequently calibrated to ensure
accurate quantitative analysis. In addition, the detector must be regularly
dissembled and cleaned (usually by a mechanical abrasion). Many
attempts have been made to avoid this contamination problem but,
although it has been reduced, it has not been completely eliminated
(particularly in the amperometric form of operation). Due to potentially
low sensing volume the detector is very suitable for use with small bore
columns. An example of the use of the electrochemical detector to
monitor the separation of a series of catacholamines is shown in figure
59. It is seen that the detector is very sensitive but its value in g/ml
(which would allow some comparison with other detecting methods) can
not be evaluated from the data.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
99

8 9 6

1
4 100 pA
10
11
6 13
2 7
16
5
3
12
17
18
15

14

0 5 15 20
Time (Minutes)

Courtesy of the Perkin Elmer Corporation


Column: HC-3 C18 (100 mm x 4.6 mm); mobile phase: aqueous solution of 100 nM
formic acid , 0.35 nM octane sulphonic acid, 1.0 nM citric acid 0.10 nM EDTA, 5%
acetonitrile, 0.25% v/v diethylamine, pH to 3.10 with potassium hydroxide; flow rate 1
ml/min; detection: oxidative amperometric with glassy carbon electrode at 100 mV
potential vs. Ag/AgCl electrode.

1. 3,4 dihydroxymendelic acid 200 pg 10. dopamine 200 pg


2. L-dopa 600 pg 11. metanephrine 400 pg
3. vanillymendelic acid 400 pg 12.3,4-dihydroxyphenylacetic acid
200 pg
4. norepinephrin 200 pg 13. N-methyl dopamine 400 pg
5. a-methyl dopa 600 pg 14. tyramine 1 ng
6. 3-methoxy,4-hydroxyphenylglycol 15. 5-hydroxyindole-3-acetic acid
400 pg 200 pg
7. epinephrine 200 pg 16. 3-methoxytyramine 400 pg
8. 3,4-dihydroxybenzylamine 200 pg 17. 5-hydroxytryptamine 200 pg
9. normetanephrin 400 pg 18. homovanillic acid 400 pg

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
100

Figure 59. The Separation of Some Catacholamines Monitored by


an Electrochemical Detector.

The column appears to be packed with 3 micron particles which should


provide an efficiency of ca. 16,000 theoretical plates. This high efficiency
means the peaks would have a very small volume, thus, even with the
small peak masses, the peak concentrations could be relatively high.
Quoting sensitivities in terms of the mass contained in each peak will
always give an enhanced impression of detector sensitivity when a high
efficiency column is used.
The Multi–Electrode Array Detector

The success of the electrode array as an LC detector is probably due to


the development of the porous carbon electrode. This electrode is made
of porous graphitic carbon, which has a very high surface area, is
mechanically robust and, more important, is permeable to the mobile
phase. As a consequence, flow through electrodes can be constructed.
The material ideal for electrochemical detection in a number of ways. As
the surface area is greatly in excess of that required for efficient
electrochemical reaction, it can be severely contaminated before it fails to
function. In fact, as much as 95% of the surface can be contaminated
before it requires cleaning. When the electrode becomes sufficiently
contaminated to require cleaning (which, according to the manufacturers,
may occur between one and three years of continual use), the
contamination can be rapidly removed by flushing with nitric acid.

The porous graphitic carbon electrode facilitates the construction of


electrode arrays. In use, the large surface area of the porous electrode
ensures that 100% of the eluted material is reacted. Thus, the
electrochemical reaction is no longer amperometric, but now
coulometric. This is an important difference and makes the array system
practical. The electrode system is shown diagramatically in figure 60.
Each electrode unit has a central porous carbon electrode, on either side
of which is situated a reference electrode and a auxiliary electrode. As
the pressure drop across the porous electrode is relatively small, these
electrode units can be connected in series forming an array. Up to 16
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
101

units can be placed in series and these arrays are commercially available.
However, a sensor system that contains as many as 80 electrodes in the
space of a few millimeters has also been constructed (53).

Counter Counter
Electrodes Electrodes

Reference Reference
Electrodes Electrodes
Porous Working
Electrodes

Courtesy ESA Inc.

Figure 60. The Coulometric Electrode System Employing Porous


Graphitic Carbon Electrodes
A progressively greater potential is applied sequentially to the electrodes
of each consecutive unit. This results in all the solutes migrating through
the array until each reaches the unit that has the required potential to
permit its oxidation or reduction.

A comparison between amperometric and coulometric detection is


shown in figure 61. In amperometric detection only a small part of the
solute is reacted and so the remainder can proceed to the next electrode
system and be detected again. In this way each of the units will detect all
the solutes and the graph shown in the upper part of the figure is
produced. It is seen that there is no discrimination by the different
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
102

electrode voltages. Coulometric detection results in the total solute being


reacted and thus it will be detected by that unit that has the required
potential and not be sensed by other units.

Retention
Time Electrode Voltage

Retention
Time Electrode Voltage

Courtesy of the Analyst.

Figure 61 Three Dimensional Graphs Demonstrating the


Difference between Amperometric and Coulometric Detection
Employing an Electrode Array
It follows that distinct peaks will be produced only at those units with the
appropriate potentials. The result is a three dimensional graph similar to
that shown in the lower part of the figure. It is seen that each unit
produces a peak at a potential that is characteristic of the solute being
detected. Although the voltage increases progressively from one unit to
the next, reaction is not completed at one unit only. This is because
reaction will not take place at a specific potential, but over a narrow
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
103

range of potentials. In practice the signal is usually detected over three


contiguous electrode units. For example, the first will oxidize a very
small amount of the solute, the second unit will be the dominant unit and
oxidize the majority of the solute, while the third unit will oxidize the
small remaining quantity of solute.

6 7 8
3 4 5
1 2 Electrode
Units

6 7 8
5
2 3 4
1 Electrode
Units

Courtesy of the Analyst.

Figure 62. Use of the Electrochemical Array to Confirm


Compound Identity
This produces a characteristic pattern of peaks for a particular solute.
The ratio of peak height for the three contiguous electrode units that
sense the substance will be different for different substances although
they may be reacted at the same three electrodes. This is obviously a
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
104

method of confirming the identity of the solute and is demonstrated in


figure 62. The upper graph shows the reference chromatogram of two
standards each showing specific retention times and a specific peak
pattern as provided by the electrode array detector. The lower graph is
for a similar sample and, although the retention times for the pair of
solutes is very similar, the pattern given by the electrode array detector
clearly shows that the second compound eluted is not the same as that of
the second standard.
27 29
30 21 28 24 11
17 6 7 8

9 16
12
23
26 10 14
25

0 10 20
2
22 19 15
20 5 Retention Time (minutes) 4 18
3 1 13

Mobile Phase 1% Methanol to 40% Methanol in a Phosphate (0.1 mol l-1 buffer with
ion pairing (pH 3.4)
1. Dihydroxyphenylacetic acid 16. Metenephrine
2. Dihydroxyphenylethylene glycol 17. Methoxyhydroxyphenyl glycol
3. L-Dopa 18. Methoxytyramine
4. Dopamine 19. N-methylserotonin
5. Epinephrine 20. Norepinephrim
6. Guanine 21. Normetenephrine
7. Guanosine 22. Salsolinol
8. Homovanallic acid 23. Octopamine
9. Hydroxybenzoic acid 24. Seratomin
10. Hydroxyindoleacetic acid 25. Tryptophan
11. Hydroxyphenylacetic acid 26. Tyrosine
12. Hydroxyphenyllactic acid 27. Uric Acid
13. Hydroxytryptophan 28. Vanillic acid
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
105

14. Kynurenine 29. Vanylmandelic acid


15. Melatonin 30. Xanthine
Courtesy of the Analyst.
Figure 11 The Separation of 30 Neuroactive Substances Monitored
by an Electrochemical Array

The electrode array detector also gives improved apparent


chromatographic resolution in a similar way to that of the diode array
detector. Two peaks that have not been chromatographically resolved
and are eluted together can still be shown as two peaks that are resolved
electrochemically and can be quantitatively estimated. Another
advantage is that high oxidation potentials can be used without the high
background currents and noise that usually accompany such operating
conditions. The electrodes that are operating at high voltages are
"buffered" by the previous electrodes operating at lower voltages which
results in reduced background currents and noise.
Another example of the application of the detector to the separation of a
number of neuroactive substances (54)is shown in figure 63. It is seen
that for certain applications the electrochemical array detector can be
extremely useful.Nevertheless, in order to use the detector, the solutes
must be amenable to electrochemical reaction and capable of being
separated using a mobile phase that will conduct an ion current

References
1. A. Tiselius and D. Claesson, Arkiv. Kemi Minearl. Geol.,
15B(No. 18)(1942)
2. A, J. P. Martin and S. S. Randall, Biochem. J., 49(1951)293
4. M. J. E. Golay, "Gas Chromatography 1958", (Ed. D. H. Desty),
Butterworths, London (1958)36.
5. J. G. Atwood and M. J. E. Golay, J. Chromatogr., 218(1981)97.
6. R. P. W. Scott and P. Kucera, J. Chromatogr. Sci.,18(1971)641.
7. C. H. Lochmuller and M. Sumner, J. Chromatogr.
Sci.,18(1980)159.
8. I. Halasz, H. O. Gerlach, K. F. Gutlich and P. Walkling, US Patent
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
106

3,820,660, (1974).
9. R. Tijssen, Separ, Sci. Technol., 13(1978)681.
10. E. D. Katz and R. P. W. Scott, J. Chromatogr., 268(1983)169.
11. J. G. Atwood and M. J. E. Golay, J. Chromatogr., 218(1981)97.
12. R. P. W. Scott, P Kucera and M. Munroe, J. Chromatogr.,
186(1979)475.
13. A. Tiselius and D. Claesson, Ark. Kemi. Mineral. Geol.
15B(No.18)(1942).
14. C. Christiansen, Ann. Phys. Chem., 3(1884)298.
15. C. Christiansen, Handbook of Chemical Microscopy (Ed. E. M.
Chanot and C. W. Mason), John Wiley and Sons, New York,
Vol. 1((1958)101 and 189.
16. M. Bakken and V. J. Stenberg, J. Chromatogr. Sci., 9(1971)603.
17. 7. J. P. Gorden, R. C. C. Leite, R. S. Moore, S. P. S. Posto, J. R.
Whinnery, Bull. Am. Phys. Soc., (2) 9(1964)501.
18. J. P. Gorden, R. C. C. Leite, R. S. Moore, S. P. S. Posto, J. R.
Whinnery, J. Appl. Phys., 36(1965)3.
19. C. E. Buffet and M. D. Momis, Anal. Chem. 54(1982)1824.
20. R. A. Grant, J. Appl. Chem. 8(1959)136.
21. M. Poppe and J. Kunysten, J. Chromatogr. Sci. 10(1972)16A.
22. L. V. Benningfield Jr., Pittsburgh Conference on Analytical
Chemistry and Applied Spectroscopy, March 5-9(1979)paper 123.
23. L. V. Benningfield Jr. and R. A. Mowery Jr., J. Chromatogr. Sci.,
19(1981)115.
24. 14. R. K. Bade, L. V. Benningfield Jr., R. A. Mowery and E. N.
Fuller, Am. Lab.13(10)(19810130.
25. J. P. Gorden, R. C. C. Leite, R. S. Moore, S. P. S. Posto, J. R.
Whinnery, Bull. Am. Phys. Soc., (2) 9(1964)501.
26. J. P. Gorden, R. C. C. Leite, R. S. Moore, S. P. S. Posto, J. R.
Whinnery, J. Appl. Phys., 36(1965)3.
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
107

27. C. E. Buffet and M. D. Momis, Anal. Chem. 54(1982)1824.


28. M. Poppe and J. Kunysten, J. Chromatogr. Sci. 10(1972)16A.
29. L. V. Benningfield Jr., Pittsburgh Conference on Analytical
Chemistry and Applied Spectroscopy,March 5-9(1979)p 123.
30. L. V. Benningfield Jr. and R. A. Mowery Jr., J. Chromatogr. Sci.,
19(1981)115.
31. R. K. Bade, L. V. Benningfield Jr., R. A. Mowery and E. N.
Fuller, Am. Lab.13(10)(19810130.
32. J. M. Miller and R. Strusz, Am. Lab. Jan(1970)29
33. A. T. James, J. R. Ravenhill and R. P. W. Scott, Chem. Ind.
(1964)746.
34. E. O. A. Haahti and T. Nikkari, Acta. Chem. Scand.,
17(1963)2565.
35. A. Karmen, Anal. Chem. 38(1966)286.
36. R. P. W. Scott and J. F. Lawrence, J. Chromatogr. Sci.,8(1970)65.
37.12. 14. L. M. Van Dijk, J. Chromatogr. Sci. 10(1972)31.
38. L. Yang, G. J. Ferguson and M. L. Vertal, Anal.
Chem.,56(1984)2632.
39. B. J. Compton and W. C. Purdy, J. Chromatogr., 169(1979)39.
40. A. Stolyhwo, O. S. Privet and W. L. Erdahl, J. Chromatogr. Sci.
8(1970)65.
41 V. Pretorius and J. F. van Rensburg, J Chromatogr. Sci.,
11(1973)263.
42. K. Slais and M Krejci, J. Chromatogr., 91(1974)181.
43. N. A. Dugger, US Patent 528,343.
44. De Peña
45 De Peña
46. H. Dubsky, J. Chromatogr.,71(1972)395.
47. J. J. Szakusito and R. E. Robinson, Anal. Chem., 46(1974)1648.
48. P. J. Wyatt, Applied Optics 7(196801879.
This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.
108

49. D. T. Phillips, Nature 221(1969)1257.


50. S. G. Weber and W. C. Purdy, Ind. Eng. Chem. Prod. Res. Dev.,
20(1981).
51. H. G. Hanekamp, P. Boss and R. W. Frei, Trends in Anal. Chem,
1(1982)135.
52. H. Poppe, Instrumentation for High-Performance Liquid
Chromatography, (Ed. J. F. K. Huber) Elsevier, Amsterdam and
New York (1978).
53. A. Atsushi, T. Matsue and I. Uchida, Anal Chem., 64 (1992)44.
54. C. N. Svendsen, Analyst, 118(Feb)(1993)123.

This eBook is protected by Copyright law and International Treaties. All rights are reserved. This book is covered by an End User
Licensee Agreement (EULA). The full EULA may be seen at http://www.library4science.com/eula.html.

You might also like