The document summarizes the analysis of lactulose as per the USP monograph. A Thermo Scientific Dionex UltiMate 3000 LC system was used to analyze lactulose on a Thermo Scientific Hypersil Gold amino HPLC column. The results for relative retention times and resolution of lactulose and related compounds exceeded the criteria in the USP monograph.
The document summarizes the analysis of lactulose as per the USP monograph. A Thermo Scientific Dionex UltiMate 3000 LC system was used to analyze lactulose on a Thermo Scientific Hypersil Gold amino HPLC column. The results for relative retention times and resolution of lactulose and related compounds exceeded the criteria in the USP monograph.
The document summarizes the analysis of lactulose as per the USP monograph. A Thermo Scientific Dionex UltiMate 3000 LC system was used to analyze lactulose on a Thermo Scientific Hypersil Gold amino HPLC column. The results for relative retention times and resolution of lactulose and related compounds exceeded the criteria in the USP monograph.
Title Related Compound analysis of Lactulose as per USP monograph
Sodium Lactate standard Chromatogram
1 Procedure
Description The Thermo Scientific Dionex UltiMate 3000 LC system is applied
for the analysis of Lactulose. The separation was performed on a Thermo Scientific Hypersil Gold amino HPLC column.
The results obtained for RRT and Resolution exceeded the
criteria stated in the USP monograph. The results obtained are as follows: USP acceptance Sr. No. Parameter Obtained Results criteria Relative retention time for Epilactose 1 0.85 min 0.92 min with respect to Lactulose peak