Professional Documents
Culture Documents
CHAPTER 2 Basic Equipment Components and Image Production - 2011 - Clinical Ultrasound Third Edition
CHAPTER 2 Basic Equipment Components and Image Production - 2011 - Clinical Ultrasound Third Edition
DOPPLER ULTRASOUND 26
Continuous wave Doppler (CW) 26
Pulsed wave Doppler (PW) 26
High pass filtering 28
Aliasing 28
Doppler frequency estimation and display 28
Colour flow imaging 28
Limitations of colour flow imaging 29
Power Doppler 30
INTRODUCTION
The clinical use of ultrasound is an interactive process, which
requires the user to manipulate the ultrasound transducer in contact
with the patient’s skin surface, while viewing and interpreting the
image in real-time. Before the examination, the user must select the
transducer, operating mode and system settings most relevant to
the investigation and thereafter make regular adjustments to the
system controls to optimise the diagnostic quality of the image. To
make informed choices of system components and settings, it is
essential to have an understanding of their function and how
making changes is likely to affect the resulting image. This chapter
describes the main components of the ultrasound imaging system
and explains how they operate to produce ultrasound images and
other useful diagnostic information. The aim of the chapter is to
provide a basic understanding of system function, which will help
users to produce optimum quality diagnostic information when
scanning. The intention is that these explanations should be acces-
sible to all without any detailed knowledge of physics or Figure 2.1 An ultrasound imaging system in use. (Courtesy of
technology. Siemens Healthcare.)
16
Principles of B-mode image formation
A The transmit pulse travels along the ultrasound beam into tissues.
Echoes generated at interfaces return to the transducer
Organ boundary Gas echo Transducer Transmit Returning
pulse echo
Tissue echoes
Cystic area
Depth
Received
echo signal
0 Time
B The time of arrival of echoes after transmission increases
with the depth of the reflector
Figure 2.2 A B-mode image using a curvilinear array Linear array transducer
transducer.
Backing layer
A Electrodes
ultrasound images are displayed as they are formed, resulting in a
real-time display of moving images.
Focal length F
The finite travel time of the ultrasound pulses and echoes to and
from the reflecting target results in a fundamental limitation of
ultrasound imaging systems. The time to form each image line is Focal region
the ‘go and return’ time to the maximum depth. Large imaging
depths result in long line times. When coupled with a large number
B Transducer Acoustic lens
of image lines, this leads to a long frame time and low image repeat
rate or frame rate. If a high frame rate is needed to image a rapidly Figure 2.5 Transducer construction. A: A cross-section through
moving target, e.g. the heart, it may be necessary to restrict the an ultrasound transducer. B: The beam can be focused by adding
imaged depth and/or the width of the image. an acoustic lens to the transducer face, resulting in a narrower
beam in the focal region.
Figure 2.6 Common types of array transducer shape and image format.
B
A
19
CHAPTER 2 • Basic equipment, components and image production
adjacent positions within the tissues, as described earlier. For most As the array elements are less than a wavelength wide, when
applications, this movement is achieved electronically using a excited by the electrical pulse waveform, they produce a wave that
transducer, which has an array of elements (typically 128 or 256). is roughly circular in shape (in the image plane). If the electrical
Array transducers are available in a variety of shapes, giving a pulse is applied to all elements simultaneously, as illustrated in
choice of image formats for different applications. Figure 2.6 shows Figure 2.9A, the circular waves merge to form a flat wavefront and
a range of common types of array transducers for transcutaneous a collimated beam in the straight ahead direction. If element 1 is
imaging. Sector transducers are used where access to the target fired first, followed after a very short delay by element 2, then
tissues is through a narrow acoustic window, e.g. approaching the element 3 and so on, by the time the last element is fired, the circular
heart through an intercostal space between the ribs. Linear and waves from the others have travelled various distances into the
curved array transducers are used where a wide anterior field of tissue and they merge to form a wavefront that propagates at an
view is needed and access is not limited by regions of bone or gas, angle to the straight ahead direction (Fig. 2.9B).
e.g. in the upper abdomen, neck or peripheral vasculature. Endo- When the echoes are received, similar delays are applied before
cavity transducers use small transducer arrays mounted on a suit- they are added together. By steadily increasing the delay between
able stem for insertion into body cavities as shown in Figure 2.7. elements on each successive pulse-echo cycle, the beam is steered
These can bring the transducer into closer proximity to the target, further away from the straight ahead position. Angles of up to ±45°
allowing the use of higher frequencies and avoiding intervening can be achieved by this technique.
obstructions such as gas or bone.
All array transducers make use of one or both of two methods to
scan the beam through the tissues. The beam can be electronically Other beam steering applications
stepped to different positions along the array or steered in different
directions. Beam steering is used with some linear arrays to form a trapezoidal
field of view (Fig. 2.6). Here, a half sector is added to each side of
The stepped array the normal rectangular field of view using beam steering with a
group of elements at each end of the array.
Beam steering is used by some manufacturers with linear array
Figure 2.8 illustrates the stepped array principle for a linear array
transducers to generate images of a target from a number of differ-
transducer. As explained in Chapter 1, in order to produce a rela-
ent directions. The direction of the beams from the linear array may
tively parallel-sided ultrasound beam, the active transducer width
be steered in up to nine different directions on successive image
or aperture must be at least 10 times the wavelength of the ultra-
frames, giving a so-called compound image. This technique helps
sound wave. As each element in a typical array is only about 1
to give more complete images of curved organ or lesion boundaries
wavelength wide, the electrical pulse waveform is applied simulta-
and helps to average out image speckle. However, the frame rate
neously to a group of say 10 adjacent elements, e.g. elements 1–10.
is reduced, as each complete compound image takes longer to form.
The echo signals received back at these elements can be added
Beam steering is used with pulsed Doppler systems (see ‘Doppler
together also to form the first B-mode line. When this first pulse-
ultrasound’ section, below) to improve the angle at which the
echo cycle is complete, the beam position is stepped along the array
Doppler beam interrogates vessels that lie parallel to the skin
by applying the pulse to elements 2–11 and adding the returned
surface.
echoes from these. By continuing this process, the beam can be
stepped along the whole array to form a large number of closely
spaced B-mode lines.
The stepped array process is used with linear array transducers
and with curved array transducers, which are simply an array
of elements mounted onto a curved surface to give a curved field
of view. Flat wavefront
Transmit in straight ahead
The steered array pulse direction
Time delays
Matching layer Acoustic lens
Line 1
Wave travels at an
Line 2 angle to the straight
Transmit ahead direction
pulse
Line 3
Transducer element array
Active aperture
Array elements Curved wave from first element
A Cut-away of a linear array transducer, B The active aperture is stepped
showing array elements, matching along the array on each B If each element is fired in turn after a short delay, a wave is formed
layer and cylindrical lens pulse–echo cycle which propagates at an angle to the straight ahead direction
Figure 2.8 The linear, stepped array transducer. Figure 2.9 The steered array principle.
20
Transducers and beam forming
Time delays
Transmit Focus
pulse
Focal depth F
A Transmit focus: if the outer elements are fired first with increasing delays a a
towards the central elements, a converging wavefront is formed which F
produces a focused transmit beam F
Time delays
Focal points on image line A The ratio of aperture a to focal depth B A higher ratio of aperture a to
F is restricted for sector transducers focal depth F can be maintained
Receive leading to weak focusing in linear arrays by increasing the
signal active aperture at larger depths
Current receive focus
Figure 2.11 Aperture control and focusing.
Array elements
of the transducer and so focusing is less effective at large depths,
B Receive focus: in reception, delays are applied to the signals received at leading to poorer resolution in these regions of the image.
each element before they are added together. The delays are calculated
for each target depth so that the receive focus tracks the current receive
depth, producing a much narrower receive beam
The beam former
Figure 2.10 Electronic focusing. In modern ultrasound systems, the beam manipulation techniques
described above are all implemented using digital techniques in a
component of the ultrasound system known as the beam former.1
The beam former is programmed to activate a chosen group of ele-
ments to transmit a focused pulse. Echo signals are then received
Electronic focusing by the elements within the active aperture. These are converted into
digital format and stored temporarily as a data series for each
Focusing of the beam in the scan plane can be achieved in both element. The beam former then uses the stored element data to
stepped and steered arrays by controlling the timing of signals to create B-mode lines by applying the techniques described above.
and from the active elements. In Figure 2.10A, the outer elements By using more than one beam former, the same stored element
of the active area are fired first, followed after small delays by signals can be used to form several adjacent B-mode lines at the
consecutive element pairs towards the centre. The circular waves same time, reducing the time needed to form each image.
from the individual elements add together to form a concave wave-
front, which converges to a focus, reducing the width of the beam. Elevation focusing
Increasing the delay time between elements reduces the distance of
the focus from the transducer.
While electronic focusing techniques can be used to produce a beam
The transmitted pulse is normally focused to a depth chosen by
that is very narrow in the direction along the scan plane, focusing
the user via the machine console and has a single value for each
at right angles to the scan plane, the slice thickness direction, often
transmission. In reception, as the depth from which echoes are
relies simply on a fixed acoustic lens as shown in Figure 2.12A. This
being received is known from the time after transmission, dynamic
fixed and relatively weak focusing results in a slice thickness that
control of the echo delays is applied so that the receive focus effec-
is much greater than the in-plane beam width. A large slice thick-
tively tracks the echo source. Focusing is applied for each point
ness can result in poor visualisation of small cysts and blood vessels.
along the line by calculating the travel time to each receive element
Slice thickness can be reduced near to the transducer by the use of
and then aligning the received signals in time before adding them
arrays containing several parallel rows of elements rather than just
together. The resulting receive beam is much narrower than the
one. These are referred to as 1.25D, 1.5D, 1.75D, 2D multi-row or
transmit beam (Fig. 2.10B).
matrix transducers according to the level of technology employed
(and manufacturer terminology). In a 1.25D array, the multiple
Aperture control rows are used simply to give aperture control, reducing the number
of rows used for superficial targets (Fig. 2.12B). The 1.5D array also
employs electronically controlled focusing in elevation but the rows
Focusing is most effective when the aperture width, i.e. the active
are paired to give symmetric delays. The 1.75D array has independ-
length of the transducer, is similar to the focal depth. If the focal
ent control of delays to each row.
depth is much greater than the aperture width, then the focusing
effect is weak. As the depth of the focus is increased by the user,
the active aperture width would normally be increased to maintain 3/4D transducers
effective focusing. For linear array transducers, the aperture can be
increased potentially to the full length of the array, maintaining an The beam manipulation techniques described so far result in a real-
effective focus to large depths (Fig. 2.11). For curved arrays and time, cross-sectional image, which allows the anatomy to be exam-
phased arrays, the maximum aperture is restricted by the geometry ined, slice by slice, as the transducer is moved by the operator. The
21
CHAPTER 2 • Basic equipment, components and image production
3D transducer is used to acquire image data from a 3D volume of The alternative type of transducer that can be used for 3D acquisi-
tissue. Here, the transducer is held in a fixed position by the opera- tion is the square array transducer, which uses a square matrix of
tor as the image plane is swept sideways automatically through the transducer elements, e.g. 128 × 128 (Fig. 2.13B). This type of trans-
target volume and a set of 2D images stored. ducer can be used in phased array mode to produce a sector image
Two types of transducer designs are available to carry out 3D in one plane as described earlier. Using beam steering in the orthog-
volume acquisitions. The most common type contains a curvilinear onal plane, the sector image can then be swept sideways on each
array mounted on a swivel within a small water bath in the trans- successive frame to interrogate a pyramid-shaped 3D volume of
ducer head (Fig. 2.13A). To acquire a 3D volume, the array is tissue.
rotated slowly by an electric motor, sweeping the image plane Both types of transducer can be used to produce continuously
sideways through the tissues. The orientation of the scan plane is updated volume data sets by repeating the volume acquisition
measured by an angle sensor coupled to the motor. The operator process. The repeat rate of the resulting 4D images may be quite
holds the transducer assembly steady as a set of 2D curvilinear low due to the time required to form a set of 2D images. By restrict-
images is acquired. ing the depth of the images and the angle of the acquired pyramid
of data, image update rates of up to 30 per second can be achieved
using a square array transducer.
Perhaps the most widely useful application of 3D transducers is
in the acquisition of a single volume data set from the target tissues
in the form of multiple adjacent planes. If the 3D volume data set
is stored, it can be interrogated and displayed in several ways. The
set of 2D views of the original scan plane being swept through the
tissues can be reviewed. Similar views can be displayed of a set of
sections orthogonal to the original 2D plane. 2D images can also be
created at a range of depths in a plane parallel to the transducer
face (C-scan). This plane is one that is not normally accessible with
a standard 2D transducer. The 3D data set can also be processed
for display in various viewing modes including see through
view and surface rendering, where a 3D surface corresponding to
the skin surface, for example, is segmented from the data and
displayed.2 Surface rendering is most effective where the surface of
interest is adjacent to a region of fluid (e.g. in obstetrics), so that the
surface echo can be detected automatically by the system
processor.
Image plane
swept sideways
through tissues
Acoustic window
Coupling fluid
Array
Gear
Motor
Position
sensing device
Cables
Housing
A A 3D volume can be imaged using a 2D curvilinear B The square array transducer has a square matrix
transducer which is rotated by an electric motor of elements, which can steer the beam to image
a pyramidal volume
B-mode system
Transmit beam former processor
Receive/
digitise
Channel
Post-processing
memory
Receive beam
Cine memory
former
Display
Echo signal
Image formation
processing
Transducer
Echo size
Echo size
Time/depth Time/depth
A Echo signals received at the transducer B TGC is used to compensate for attenuation C The user can adjust the TGC
become weaker with increasing so that echoes from similar targets are the via slide controls on the
depth/time of arrival due to attenuation same size regardless of target depth console
0 dB Tissue–gas interface
-20 dB
Large tissue–tissue
-40 dB interfaces
Echo size
Echo size
-60 dB
Tissue scattering
-80 dB
-100 dB Blood
Time/depth Time/depth
A Echo sizes from different target types. B Echo signals before compression C Echo signals after compression
A 60 dB range of echo strengths must
be displayed to include tissue interfaces
and scattering
Harmonic imaging
2f0 Frequency
The processing described so far has assumed that the transmitted Clutter can be suppressed by removing
pulse consists of a few cycles of the chosen transmit frequency, e.g. the fundamental frequencies and forming
5 MHz, and that the returned echoes are simply miniature versions the image from the first harmonic
of the same thing. However, at the relatively high pulse amplitudes Figure 2.17 Harmonic imaging.
transmitted by modern ultrasound imaging systems, propagation
through soft tissues is non-linear and the transmitted pulse becomes When an image is formed from the original fundamental fre-
distorted as it travels into the tissue. The effect is strongest in the quency that was transmitted, it may contain clutter due to weak
centre of the ultrasound beam where the amplitude is highest. echoes from beam side lobes or multiple reflections (reverberation)
Distortion of the waveform is associated with the generation of of the pulse between strongly reflecting features (Fig. 2.17). As
harmonics or multiples of the original fundamental frequency that harmonic generation is only significant where the pulse amplitude
was transmitted. is high, the clutter can be reduced by removing the fundamental
24
Image processing
Coded pulse
Echo amplitude
waveform
RF signal
1 1 0 1
Phase
Time
A Undemodulated RF echo signal B Demodulated echo signal
Figure 2.18 Pulse coding. The phase of the elongated transmit Figure 2.19 Demodulation.
pulse is alternated according to a chosen digital code. The same
code is embedded in a digital filter applied to returned echoes to
pinpoint their time of arrival.
B-mode instrumentation and processing
• Time gain compensation (TGC) is used to compensate for
frequency from the returned echoes and forming an image from the increasing attenuation of echoes with depth.
harmonic frequency. This is known as harmonic imaging.4 As • The dynamic range of echoes must be compressed to display
the energy at the harmonic frequency is always less than that at the reflected and scattered echoes simultaneously.
fundamental, the system penetration may be reduced to some • Harmonic imaging may be used to reduce reverberations and
extent when operating in harmonic mode. clutter in the image.
To form a harmonic image, the transducer must be able to trans- • Pulse coding may be used to increase penetration for a given
mit the fundamental frequency, but also receive the second har- frequency.
monic (twice the fundamental). Hence a transducer with a wide • Echoes must be demodulated to remove the transmit frequency
frequency response is required. The harmonic image can be formed before image formation.
by applying a frequency-selective filter to the returned echoes, • Post-processing is applied as the image is read out from the
which rejects the fundamental frequency but allows through the image memory.
harmonic frequency.5
Harmonic images can also be produced using a method called
pulse inversion. In this method, two pulse-echo cycles are used to
form each B-mode line. In the first cycle, a short imaging pulse is
transmitted and echoes received as normal and stored. Then a its harmonics. This is referred to as the radio-frequency or RF signal.
second pulse, which is an inverted version of the first, is transmitted The brightness value that is displayed at each point in the image is
along the same path and echoes received. The second line of echoes related to the amplitude or envelope of the RF signal. The final
is subtracted from the first. In low amplitude parts of the beam, processing step in forming the B-mode image is to remove the
where there is little distortion of the transmitted pulse, the lines of radio-frequency element of the signal to leave the envelope signal.
echoes cancel out. In the high-amplitude regions, the lines of echoes This process is referred to as demodulation (Fig. 2.19).
generated from the distorted normal and inverted pulses contain
energy at the fundamental and second harmonic frequencies. The Image memory
fundamental parts cancel out leaving the second harmonic parts to
form the image.
Following the various processing steps described above, the B-mode
lines are written into the image memory to form the image as an
Pulse coding array of pixels similar to an image in a digital camera. The dimen-
sions of the image in the memory array are typically 1200 × 1200
The maximum depth from which echoes can be received is deter- pixels, with each pixel capable of storing up to 256 different grey
mined by the attenuation of the pulse and returning echoes by the levels (8 bits). As for a digital camera, the system is capable of
tissues they pass through. Higher frequencies give improved image storing many images (over 2000 for some models) and successive
resolution but suffer more rapid attenuation and hence reduced frames are stored in a cine memory until the memory is full. The
penetration of tissues. For a given frequency, penetration can be first image stored is then overwritten so that the last frames are
improved by using a longer transmitted pulse, making its echoes always stored. The chosen image is read out from the image
easier to detect. However, the increased pulse length degrades memory in a raster pattern which is synchronised to that of the
the axial resolution of the system. Pulse coding techniques allow display monitor.
the use of long transmit pulses while maintaining good axial The image can be written to the image memory and read out in
resolution.6 a number of different ways. If the image memory is being con-
The long transmit pulse is divided into time segments in which stantly updated with new images and the last image to be formed
the phase of the wave is alternated according to a predetermined is always read out to the display, a real-time, moving image is
code (Fig. 2.18). When the echoes are received, they are passed displayed. In freeze mode, new images are not written into the
through a digital filter which is programmed to search for the code. memory and the last image stored is read out repeatedly to the
When a match is detected, the time at which it occurs gives the display monitor. In cine loop mode, the writing process is again
effective time of arrival of the echo, restoring the axial resolution. frozen but the whole sequence of stored images is read out in a
Pulse coding allows greater penetration to be achieved at higher repeated cycle showing the last 30 seconds or so of the real-time
frequencies. image.
Receiver fT
amplifier fR
Doppler shift fD
A frequency fD B
Figure 2.22 Continuous wave Doppler. A: System block diagram. B: The demodulation process.
Transmit
Gate frequency Delay
Sample
fT
volume
Blood vessel
Low pass filter
Receiver Mixer
(fT + fR) Sample Angle cursor
amplifier fR (fT × fR)
(fT – fR)
Low pass
filter
Transducer
Sample volume
Doppler shift
frequency fD Figure 2.24 The pulsed Doppler sample volume and angle
cursor are adjusted by the operator to align with the vessel of
Figure 2.23 A pulsed wave Doppler system block diagram. interest.
PRF, pulse repetition frequency.
of the pulse is then received from increasing depths, following each of a cursor in the centre of the sample volume to indicate the direc-
transmission. The returned signals are processed as for CW Doppler tion of blood flow to the system.
to extract the Doppler shift frequencies. Doppler signals from a The process for extracting the Doppler shift frequency is similar
specified blood vessel are then isolated for further processing by in principle to that used for CW Doppler. However, the received
the use of a time gate, which allows through only those signals signal now consists of one brief sample of the Doppler frequency
corresponding to the depth of interest (Fig. 2.23). The position of from the depth corresponding to the target vessel, for each trans-
the time gate (the sample volume) is indicated on the B-mode image mitted pulse (Fig. 2.25). These can be thought of as a set of samples
by cursors (Fig. 2.24). The user adjusts the position of the Doppler of the equivalent CW Doppler shift frequency. The value of each
beam and the depth (or range) of the sample volume along it, so sample is determined by the phase relationship between the cor-
that the sample volume is located over the vessel of interest. As this responding fR sample and the reference transmit frequency fT. The
is a pulse-echo technique, the same transducer or active aperture continuous Doppler shift frequency signal is constructed from the
can be used to transmit and receive. The user also adjusts the angle samples by low pass filtering.7
27
CHAPTER 2 • Basic equipment, components and image production
Aliasing
The sampled nature of the reconstructed pulsed Doppler signal
The pulsed Doppler gating process gives a sample leads to a major limitation of pulsed Doppler systems at high blood
of the Doppler shift frequency for each pulse echo cycle flow velocities. Each transmitted pulse leads to one sample of the
Doppler shift signal. Hence the number of samples produced per
second, the sample rate, is the same as the number of Doppler
pulses transmitted per second, the pulse repetition frequency (PRF).
To reconstruct a sine wave of a given frequency, it is necessary to
have at least two samples for each cycle of the wave, i.e. the PRF
must be at least 2fD. If the Doppler frequency for flow towards the
transducer (forward flow) increases to the point where the PRF is
The Doppler frequency is reconstructed
by further smoothing and filtering
too low, it will appear instead as an aliased, reverse flow frequency,
shifted downwards by an amount equal to the PRF.
Figure 2.25 Pulsed Doppler demodulation. Aliasing is most likely to be a problem when the Doppler shift
frequency is high (high blood flow rates and high transmit fre-
quency) and the PRF is low (due to large target depth).
A PRF
clock
Velocity
Transmit
Gate frequency
fT
Receiver Mixer
amplifier fR (fT × fR)
B Mean
frequency
Image
Autocorrelator Variance Display
Power memory
Time along the line. To remove the large echoes from stationary and
slowly moving tissues, a technique called delay line cancellation is
then used rather than filtering. Each line of echoes in the ensemble
10 ms is stored digitally and compared with the next one by subtracting
intervals one from the other. This removes echoes from stationary targets,
Figure 2.27 A: A sonogram from the carotid artery. B: To which do not change, while preserving signals from those that are
construct the sonogram, the velocity distributions in consecutive moving, as they do not cancel out. Modern systems use more elabo-
10 or 20 ms samples are displayed as adjacent vertical lines. rate versions of this technique to give enhanced separation of sta-
The amount of blood flowing at each velocity is represented on tionary echoes.
a greyscale. Estimation of the Doppler shift frequency is most commonly
carried out by a technique called autocorrelation.7 In essence, each
line of Doppler signals generated by each pulse-echo cycle in the
ensemble is compared with the one before it. The rate at which
the phase of the signal, at each point along the line, changes in the
interval between one line in the ensemble and the next leads to an
distribution within a stationary sample volume placed within a
estimate of the Doppler frequency at that point in the line. By
blood vessel. Colour flow imaging systems give a real-time 2D
extending this process over the ensemble of pulses, an estimate of
image of blood flow patterns in blood vessels as a colour overlay
the mean frequency is obtained for each point along the line. Mean
on a B-mode image. Here the colour of the display, at each point in
frequency is represented in the image on a colour scale, commonly
the image, represents the mean blood flow velocity at a given time.
using red for flow towards the transducer and blue for flow away
Colour flow imaging systems can be seen as an extension of the
(Fig. 2.29). Low frequencies are displayed as dark shades of red or
time gating approach from a single sample volume to multiple
blue leading, with increasing frequency, to lighter shades through
adjacent sample volumes along the Doppler beam. The beam is also
to white. The display normally includes a colour wedge showing
swept through the region of interest as in B-mode image formation,
the range of available shades of red and blue next to a correspond-
to form a 2D colour map. However, the time restrictions imposed
ing velocity scale. The velocity scale indicates the component of
by real-time 2D image formation require different techniques to be
blood flow velocity in the direction of the Doppler beam (i.e. not
used for removal of stationary tissue echoes and for Doppler fre-
corrected for Doppler angle).
quency estimation. Whereas frequency analysis in the sonogram is
The autocorrelator can also give an estimate of the variability of
performed on a 10 or 20 ms sample of the Doppler signal, each line
the Doppler frequency during the ensemble for each point along
in the colour flow map must be formed in about 1 ms to allow real-
the line. This is expressed in terms of the variance of the signal and
time display.
is used to display an alternative form of colour map which can
For each line in the colour flow image, the beam is held in the
indicate the degree of disturbance in the blood flow in the vessel.
same position for a number of pulse-echo cycles (typically eight).
The number of pulses transmitted in this sequence is referred to as
the ensemble length. As illustrated in Figure 2.28, the received Limitations of colour flow imaging
echoes are mixed with the transmit frequency and filtered to extract
the Doppler shift frequency as for CW and pulsed Doppler systems. As the mean frequency values displayed in each line of the colour
No gating is applied as Doppler signals are required from all depths flow image are estimated from the set of samples obtained from
29
CHAPTER 2 • Basic equipment, components and image production
30