You are on page 1of 1
Test record FO-U8-118/mpr Filter drying ROMMELAG ENGINEERING Machine number 321202 1. Description of test or reference to test instruction: ‘The purpose of the filter drying is to dry the air filters after the steam sterilisation in order that they remain permeable to air. The dry weight of the filters is measured, Then a sterilisation is carried out after which the filters are dried and then re-weighed ‘The residual humidity is determined from the weighing results. 2. Test plan: 3. Measuring devi Manufacturer: Sartorius Type: MSE2202S —_Serial-No.: 33505848 4. Anticipated results: ‘Acceptance criteria: Maximum residual humidity <= 5 % 5. Deviation: Residual humidity of filters: Pos. 25 fiter 1335723 Pos. 30.1 filer 1335723, Pos, 30.2 filter 1336723, Pos. 30.3 filter 1336723, Pos. 40 filter 1335723 6. Appendix (number of sheets Aseptic diagram No:,932181-202-0101 Valid certificate Lot-No.: 12983 M Date: Name / Signature: [FO_U0800-118-mpr-gb] Createdby | Checkedby | Releasedby: revoo_321200 K Stepanek | _Preautschy | _K. Stepanek Date: 2017-06-09 Rev.00 Created by’ Checked by | Releasedby: Dok votes K. Stepanek HR. Fre K. Stepanek Date: 2016-10-01 Rev.0t Page Toft