Item BATASAN SPESIFIKASI

SPESIFIKASI TEKNIS
1 System Analog system
a. Max. mA (Tube Current) 500mA
b. Max. kV (Tube Voltage) 125kV
c. Floating patient table Yes, 4 way floating
d. Floor mounted tube stand Yes, available
e. Wall bucky stand Yes, available

2 Generator
a. Line Nominal 110-120, 220-230 VAC
b. Line Frequency 50 / 60 Hz
c. Phase Single phase
d. Line Power Requirement 3 kVA
e. kV Range 40~125kV
f. kV Step 1kV step
g. mA Range 10mA s/d min 500mA
h. Timer Range 0.001 sec s/d 10 sec, 38 steps
i. mAs Range 0.1 to 500mAs
j. Max. Power Output 500mA pada 80kV
400mA pada 100kV
320mA pada 125kV

3 Tube
a. Tube Focal Spot Size 1.0/2.0 mm
b. Collimator Manual
c. Collimator lamp timer Min 30 sec
d. Collimator Field Size Min 43x43 cm at 100 cm SID

4 Table
a. Tabletop Inherent Filtration Less than 1.7mmAl eq. at 100kVp
b. Longitudinal Cassette stroke 430mm
c. Longitudinal movement 1,000 (±500)mm
d. Lateral movement 250(±125)mm
e. Lock (Brake) mechanism EM Lock
f. Center indication Sound and LED

5 Tube Stand
a. Tube rotation angle +/-135°
b. Longitudinal movement Max 2,100mm
c. Lateral movement 220mm
d. Vertical movement 1,330mm
e. Lock (Brake) mechanism EM Lock
f. Balance Counter Weight
g. Column Rotation 90° step

6 Bucky Wall Stand


a. Vertical Cassette stroke 1,260mm
b. Lock (Brake) EM Lock
c. Balance Counter Weight