Analysis of heavy-metal
contaminants in soils
and sludges
Figure 1. Spectrum
obtained using a
0 molybdenum
10 12 14 16 secondary target,
Energy (keV)
which gives optimum
excitation for lead
Performance Accuracy
The Epsilon 5 software features a very The accuracy of the results is presented
powerful deconvolution algorithm, in Table 3. The calibration RMS value is a
which analyzes the sample spectrum and statistical comparison (1 sigma) of the
determines the net intensities of certified chemical concentrations of the
element peaks, even when they overlap standards with the concentrations
one another. calculated by regression in the
The accuracy with which this is carried calibration procedure. What this means
out is essential to trace element analysis. for the accuracy of determination within
individual samples is presented in Table 3.
Planetary ball mill
WT-M
WT-L K = 0.2441
RMS = 10.9753
Correl. = 0.9999
Rel. Err. = 5.7368E-03
0
0 2000 4000 6000
Cchem. (Zn ppm)
Figure 4. Calibration
graph for Zn
* The LLDs quoted are typical for soil samples. LLDs for individual samples vary according to sample
composition.
CdK
LLD (100s)
Conclusions
I(counts/ch)
0.44 ppm 80 kV
accurate and precise and the method benefits from simple, essentially hazard-
2.05 ppm 40 kV
0
22.5 23.0 23.5 24.0
Energy (keV) free, sample preparation. The stability of the Epsilon 5 is such that individual
calibrations can be used for months. Time-consuming re-standardizations are
unnecessary and the resultant data are highly consistent over time.
Although diligent care has been used to ensure that the information herein is accurate, nothing contained herein can be construed to imply any
representation or warrantee as to the accuracy, currency or completeness of this information. The content hereof is subject to change without
Equipment Configuration
Epsilon 5 energy-dispersive X-ray fluorescence spectrometer and controlling software.
further notice.Please contact us for the latest version of this document or further information. © PANalytical B.V. [2009]
X-ray tube: Type: side-window tube
Anode: Gd
Window: Be (300 μm)
Rating: 25 – 100 kV, 0.5 – 24 mA,
maximum power 600 W
Internal water cooling
Detector type: PAN-32: Ge X-ray detector
Printed in The Netherlands on 50% recycled, chlorine-free paper. 9498 707 26712 PN4837