You are on page 1of 16

White Paper | Spring 2015

Breast Tomosynthesis
The Use of Breast Tomosynthesis in
a Clinical Setting
Table of Contents

Introduction.......................................................................................................................................................1
Superiority of Hologic 3D Mammography to 2D...................................................................................1
Improved Sensitivity and Reduced Recalls..............................................................................................1
Greater Performance Using Two-view Tomosynthesis.............................................................. 2
Performance in Different Breast Compositions and Lesion Types................................................ 2
Performance in Calcifications, Masses and Distortions............................................................. 2
Performance in Invasive and Noninvasive Cancers................................................................... 2
Performance in Fatty and Dense Breasts.......................................................................................3
Performance Compared to Ultrasound...........................................................................................4
Performance in the Evaluation of Symptomatic Patients........................................................... 5
Clinical Considerations in Implementing Breast Tomosynthesis................................................... 6
One-view Versus Two-view Tomosynthesis.................................................................................. 6
Benefits of Having Both Tomo and 2D Images in All Views......................................................7
Patient Selection and Management................................................................................................ 8
Use for Screening or Diagnostic Imaging..................................................................................... 8
Use for Women with Dense Breasts............................................................................................... 8
Reducing Patient Dose in 3D Mammography.....................................................................................10
Reading Time...................................................................................................................................................1 1
Advances in Hologic 3D Mammography................................................................................................1 1
3D Guided Biopsy................................................................................................................................1 1
Contrast-enhanced Breast Imaging...............................................................................................12
Conclusions....................................................................................................................................................12
Glossary...........................................................................................................................................................12
References...................................................................................................................................................... 13
Breast Tomosynthesis
The Use of Breast Tomosynthesis in a Clinical Setting
Andrew Smith, Ph.D.
Vice President – Imaging Science, Hologic, Inc.

Introduction decreased callback rates, help in localizing increased diagnostic accuracy and signifi-
structures in the breast, and improved cantly reduced recall rates for non-cancer
Since the United States (U.S.) Food and lesion and margin visibility. In 2013 the FDA cases. These results were consistent with
Drug Administration’s (FDA) approval of the approved a new mode for the Hologic 3D those of an independent reader study in
first commercial systems in 2000, digital mammography system, whereby C-View™ which University of Pittsburgh researchers
mammography has become an accepted software generates the 2D image from the found a 7% improvement in the area under
standard of care in breast cancer screening 3D dataset directly, avoiding the need for the receiver operating characteristics (ROC)
and diagnosis and has paved the way for a separate 2D exposure and essentially curve for 2D plus tomo compared to 2D
the newest groundbreaking technology halving the radiation dose associated alone.2 The FDA advisory panel considered
in this arena – breast tomosynthesis, also with the combo mode procedure. all three reader studies and voted that
referred to as tomosynthesis, or simply Hologic’s clinical data demonstrated
This white paper provides detailed
“tomo.” both the effectiveness and safety of 3D
information about the performance of
Breast tomosynthesis is a screening and Hologic 3D mammography technology now mammography.
diagnostic modality that acquires images that it has been evaluated in large-scale
of a breast at multiple angles during a Improved Sensitivity and
screening trials and is in routine clinical
short scan. The individual images are use. It also looks at the performance of
Reduced Recalls
then reconstructed into a series of thin, 3D mammography in different breast The performance of 3D mammography has
high-resolution slices typically 1 mm thick, composition and lesion types, discusses been evaluated in a large screening trial
which can be displayed individually or in several issues to consider when introducing from Oslo, Norway. Results have been
a dynamic ciné mode. A tomosynthesis this technology into clinical practice, presented from the first three months and
dataset greatly reduces detection and provides a summary of some of the the first year of the two-year trial.3-5 In the
challenges associated with overlapping advanced applications for this modality. one-year evaluation of 12,631 screening ex-
structures in the breast, which is the
aminations, in which participants were im-
primary drawback of conventional 2D Superiority of Hologic 3D aged with both 2D and 3D mammography,
analog and digital mammography. Mammography to 2D the researchers reported that the detection
This technology has been available in rate for invasive cancers increased 40%,
The performance of 3D mammography
Europe and other countries recognizing the overall cancer detection rate increased
has been evaluated in a number of venues,
the CE mark since 2008. In February 27% and the false positive rates decreased
including the clinical trials in support of
2011, the Hologic Selenia® Dimensions® by 15% for examinations employing 2D and
the FDA submissions, and more recently,
3D mammography™ system was the first 3D mammography compared to 2D mam-
in Europe in large screening trials and in
commercial system approved by the mography alone. These results were seen
U.S. sites that monitored performance
FDA. The system is approved for use across all breast densities.
before and after the introduction of 3D
in the same clinical indications as 2D
mammography into routine clinical practice. The results of the Oslo trial are
mammography, including breast cancer
Hologic 3D mammography has been shown summarized below:
screening, diagnosis, and intervention.
to be superior to 2D mammography. • Invasive cancer detection increased 40%.
With the Hologic 3D mammography™
system, a combined examination of Hologic conducted a large multi-center • Cancer detection increased 27%.
3D and 2D imaging, known as combo clinical trial comparing the performance • False positives decreased 15%.
mode, takes only seconds longer than of combo mode to that of 2D digital mam- Another large population-based screen-
a conventional two dimensional digital mography alone in support of its FDA ing trial conducted in Italy involved over
mammogram. In clinical use, Hologic 3D submission.1 The two reader studies for this 7,000 women and reported a 51% increase
mammography offers significant benefits, trial found that the addition of tomo to in cancer detection with the use of tomo-
including increased cancer detection, digital mammography both significantly synthesis.6

1
The performance of 3D mammography tomo MLO showed that the tomo MLO-only a very significant increase in performance
in routine screening practice has also arm performed better than 2D imaging relative to 2D imaging for cases involving
been evaluated in observational stud- alone, but not as well as 2D plus both masses and distortions. For cases involving
ies, in which the changes in performance tomo views. microcalcifications, there was a small, but
measures with and without the use of not statistically significant, improvement in
These results are consistent with other
tomosynthesis in clinical practice were the ROC performance with the addition of
studies, illustrating that MLO-only
reported.7,8,9,10,11 The largest of these 3D imaging.
tomosynthesis is likely to be inferior to
(Friedewald, et al. 2014) reported on the two-view tomo. These study results are
performance of tomosynthesis in 13 sites Performance in Invasive and
explored in greater detail in the discussion Noninvasive Cancers
in the United States, and found that the in- of one-view versus two-view tomosynthesis
troduction of 3D mammography increased From the FDA studies it could be predicted
later in this paper.
the invasive cancer detection rate by 41% that the majority of additional cancers
while reducing the recall rate by 15%, Performance in Different found by 3D mammography will be mass
as well as observing an increase in the Breast Compositions and lesions and not calcification-only cancers
positive predictive value for both recalls because of the much greater improvement
Lesion Types
and biopsies. The other studies reported in the ROC curve performance in the
single-institution results, but with similarly The expanding library of clinical trial reader studies for non-calcifications than
positive results, showing an overall aver- results on the use of tomosynthesis for cases involving calcifications. Thus, it
age cancer detection rate increase of 22% makes it possible to evaluate its perfor- is to be expected that the gain in sensitivity
and an average recall rate reduction of mance in different breast compositions using 3D mammography can be primarily
28%. The Rose study specifically called out and lesion types such as calcifications, attributable to invasive cancers. Recent
the invasive cancer detection rate, which masses and distortions, invasive and results reporting the performance of 3D
showed an increase of 54% with the use of noninvasive cancers, and fatty and dense mammography in screening are showing
2D and 3D mammography. breast tissue. There are also some study exactly this. Skaane4 reported a 40% in-
results demonstrating how the use of to- crease in the detection of invasive cancers
Greater Performance Using mosynthesis may affect the management using 3D mammography, with no increase
Two-view Tomosynthesis of symptomatic patients. in the detection of ductal carcinoma in situ
All of the previously referenced clinical (DCIS). Similarly, Rose10 reported a 53%
studies used two-view mammography for Performance in Calcifications, increase in invasive cancer detection using
both 2D and 3D imaging. As part of the Masses and Distortions 3D mammography, and as with Skaane, no
study submitted to the FDA, another arm The clinical trial data presented as part increase in the detection of noninvasive
was investigated: single-view tomosyn- of Hologic’s FDA submission has been cancers. Ciatto also showed an increase
thesis (MLO) imaging in combination with analyzed by separating the image sets into in cancer detection of about 50% and no
two-view (CC and MLO) 2D imaging. In this calcification and non-calcification cases. increase in the detection of in-situ cancers.6
study, the performance of 2D imaging plus Rafferty et al. found that 2D plus 3D offered

2D 2D+3D

120
Number of cancers detected

100

80

60

40

20

0
fatty dense
Breast Density
The study by Skaane showed improvements in cancer detection in both INCREASED CANCER DETECTION: The 3D reconstructed slice shown on the
fatty and dense breast categories. right reveals a definitive spiculated mass that is only faintly revealed in the
2D image shown on the left. (Diagnosis: Invasive ductal carcinoma)

2
This represents one of the key benefits round one or more years out or when the breasts. Because denser breasts have
of 3D mammography – the potential for mass became palpable, had the tomo scan more structure noise (fibroglandular
earlier detection of invasive cancers – not been performed. tissue) than fatty breasts, it was expected
exactly the cancers that will advance to that 3D mammography would provide
become life-threatening if not detected in Performance in Fatty and improved performance in the denser
time for effective treatment. These tomo-
Dense Breasts breasts; however, clinical data shows
only cancers represent cancers that were The addition of 3D imaging has been that 3D mammography helps in both fatty
missed in 2D imaging and would not have shown to improve the performance of and dense breast groups.3
been found until a successive screening mammography in both fatty and dense
This was reported in the paper by
Haas et al.7 This study looked at the
performance of 3D mammography in
13,000 women undergoing breast cancer
screening. They found that the addition
of 3D imaging reduced recall rates for all
breast density groups, with statistically
significant reductions in recall rates for
scattered fibroglandular (reduction of 25%),
heterogeneously dense (reduction of 39%)
and extremely dense breasts (reduction
of 57%).
Other researchers have reported similar
trends. Rafferty studied the performance
of 3D mammography in women with dense
breasts and found an increase in the recall
rate for cancer cases and a reduction in
the recall rate for non-cancer cases.12 In
2D 3D a separate study, Rafferty found that 2D
plus tomo was significantly better than 2D
mammography alone in ROC performance
for both fatty and dense breasts.13 While
there was a gain in the area under the
ROC curve in both breast density types,
the gain was 2-3 times higher in dense
breasts than it was in fatty breasts. Rafferty
also reported large recall rate reductions
in both fatty and dense breast types.
Philpotts et al. reported on tomosynthesis
visualization of breast cancers as a func-
tion of mammography density.14 They
found that 3D mammography was particu-
larly beneficial for visualizing non-calcified
breast cancers in scattered and heteroge-
neously dense breasts, with about 70% of
cancers in these density categories seen
only or better with tomosynthesis. Patients
with fatty and extremely dense breasts
had cancers seen equally well using
2D 3D tomosynthesis and 2D mammography.
In terms of the detection of invasive
OCCULT IN 2D: The architectural distortion in this breast, while essentially occult in the 2D
lobular carcinoma (ILC), Gandini et al. has
mammograms, is easily visualized in the 3D images. reported that the detection of ILC was
significantly higher using tomosynthesis

3
than digital mammography, especially in
dense breasts.15 Radiologists were twice as
likely to miss an ILC in dense breasts using
digital mammography than when using
tomosynthesis.
These results are as expected. Fatty breasts
often have sufficient parenchyma that
tomosynthesis would be expected to offer
some advantages. However, the even larger
improvement in performance in denser
breasts using tomosynthesis illustrates that
the technology is doing what is expected
from the physics principles – reducing
2D 3D: 23 mm 33 mm 43 mm superimposed parenchyma.

REDUCED RECALL RATES: The 2D mammogram reveals what appears to be a spiculated mass laterally in Performance Compared
the right CC view. 3D slices at 23, 33 and 43 mm above the breast platform show that this 2D finding was
superimposed structures, resolved through the use of 3D imaging. to Ultrasound
Of great interest is understanding the
relative performance of ultrasound
compared to tomosynthesis in breast
cancer screening. 3D mammography, like
ultrasound, has a superior performance in
dense breasts relative to 2D mammogra-
phy. However, unlike ultrasound, in which
the recall rate of 2D and ultrasound was
4 times that of 2D mammography alone
as was seen in the ACRIN 6666 trial, 3D
mammography improves sensitivity without
increasing the recall rate.16,4,10 Further clini-
cal research will be needed to identify the
respective roles of 3D mammography and
ultrasound, particularly in screening wom-
en, but it is clear that 3D mammography
can offer improved cancer detection while
simultaneously reducing false positives.
Some recent studies have compared the
relative performance of tomosynthesis and
ultrasound. Aguillar looked at whole breast
handheld physician-performed ultrasound
following breast tomosynthesis and
concluded that adding the ultrasound
exam had little impact in cancer detec-
tion and a low positive predictive value.17
Similarly, Chung concluded that in dense
breasts, tomosynthesis showed better
diagnostic performance and a reduced
2D 3D
benign biopsy rate than breast ultrasound.18

ADDED VALUE FOR CALCIFICATIONS: The 2D mammogram on the left shows right medial
Performance in the Evaluation of
microcalcifications. The 3D reconstructed slice on the right illustrates the associated architectural Symptomatic Patients
distortion only revealed on the CC tomo image and not on the mammogram. (Diagnosis: Ductal
carcinoma in-situ/high grade) The use of 3D mammography in diagnostic
assessment offers the opportunity for both

4
improved performance and a reduction in
the number of X-ray images needed, with
|a resultant reduction in both dose and
exam time.
Zuley et al. found that two-view tomo
significantly improved diagnostic accu-
racy for non-calcified lesions compared
to supplemental mammographic views.19
All 8 radiologists participating in the study
showed improved performance. Because
the number of diagnostic views in the
evaluation of masses or focal asymmetries
can average three or more,20 there is a
clear opportunity to reduce radiation expo-
sure through the use of 3D mammography
in diagnostic evaluations.
Butler et al. had similar conclusions, and
reported that 3D mammography results in 2D 3D
decreased number of images required for
diagnostic cases.21 They further concluded
that this expedites the workup and yields
better patient throughput.
Other researchers such as Svahn have
also shown that the combined diagnostic
performance of digital mammography and
tomosynthesis is superior to either digital
mammography or tomosynthesis alone.22
Several studies have shown that 3D
mammography is superior to 2D mam-
mography in predicting tumor size, demon-
strating margins, extents of lesions, and in
staging:
–R
 afferty et al. reported that 3D 2D 3D
mammography was significantly
better than 2D mammography in 3D MAMMOGRAPHY IN DENSE BREASTS: The cancer in this dense breast is much better visualized in
detecting cancers, particularly those the 3D images than in the 2D mammograms. (Diagnosis: Signet ring cell carcinoma).

exhibiting architectural distortion, and in


characterizing cancer morphology.23 His study concluded that this modality compression views, lowering both
provided critical information for radiation dose and offering the potential
– Moonet al. showed that adding
prospective treatment planning by the to reduce biopsies on non-malignant
tomosynthesis to digital mammography
multi-disciplinary team.25 lesions.28
increased cancer detection and
diagnostic performance in diagnostic – Fornvik et al. found 3D mammography
Clinical Considerations
workup.24 superior to 2D digital mammography
in the assessment of breast tumor size
in Implementing Breast
–M
 ichell et al. showed that 3D Tomosynthesis
and stage.26
mammography is superior to 2D
mammography in predicting the – Meacock et al. found that 3D Clinical research has shown the benefits
histological tumor size because mammography was more accurate than of 3D mammography in screening and
3D mammography demonstrates 2D in tumor size measurement.27 diagnostic indications, as well as in a range
the margins and extents of the of breast compositions and tissue types.
– Tagliafico et al. found that 3D
mammographic lesions more clearly. However, there are a number of clinical
mammography could replace spot

5
Recall Rate Reduction by Breast Density

57%

39%

31%
25%

Predominantly Scattered Heterogenously Extremely dense


fatty fibroglandular dense 2D 3D
In the study by Haas, et al., the addition of 3D mammography decreased VALUE IN FATTY BREASTS: While the 2D mammogram reveals the 12:00
recalls across all breast densities, with significant reductions in denser o’clock mass, the 3D images more accurately characterizes this mass as
breasts. spiculated (Invasive ductal carcinoma).

considerations to be evaluated when de- two-view 2D combined with two-view superiority (Gennaro, Wallis).34,35,36
termining how to introduce this technology tomo imaging, with the single-view tomo
No published study using single-view tomo
to a clinical practice. What configuration providing only half the performance gain
has demonstrated an increase in both
of 2D and tomo views ensures the earliest of two-view tomo.30
sensitivity and specificity compared to 2D
possible detection of breast cancers and
Other data supports this finding: digital mammography.
reduction of unnecessary recalls? How will
these choices affect patient dose? How – Rafferty et al. found that 12% of lesions In addition to the likely loss of sensitivity
should patients be managed in a mixed were better seen on the tomo MLO im- that occurs if only one tomo view is taken,
environment? age, 15% better seen on tomo CC and 9% there are some clinical challenges that
of lesions were visible only on tomo CC.31 arise with single-view tomo imaging.
These considerations are discussed in
Neither the CC nor the MLO view always
more detail below. – Beck et al. found that only about half of
captures all the breast tissue, so both
the lesions were equally well seen on
views in some form are preferred.
One-view Versus Two-view both the MLO and CC view, with 34% of
Tomosynthesis cancers better or only seen on the CC Mixing technologies, such as combining a
The relative performance of one-view view,32 while 7% of lesions were only seen tomo MLO view and a 2D mammography
versus two-view 2D mammography is well on one view. The authors emphasized CC image, might address the tissue
understood. Screening using two views the importance of including the CC view coverage, but creates its own set of
offers an increase in cancer detection and in 3D mammography and concluded that issues. It might be difficult, for example,
a reduction in recall rates compared to obtaining both views is necessary to en- to correlate a suspicious lesion seen in
single-view mammography; the paper by sure that a cancer will be optimally visual- 2D CC with the same lesion in the tomo
Wald et al. estimates the sensitivity gain ized, and the greatest potential benefit MLO, or vice versa. The approach of a
is 24% and recall rate reduction is 15%.29 from tomosynthesis will be derived. tomo MLO and a 2D CC is likely inferior in
clinical performance to performing tomo
Equivalently, single-view tomosynthesis – Similar results were reported by Baker et
imaging in both the MLO and CC view –
(either CC or MLO) is a lower-dose pro- al., who found 8% of lesions were visible
and offers no dose advantage. The reason
cedure compared to two-view tomosyn- only on the tomo CC view and 1.4% only
for this is clear. Just as some cancers in
thesis, but it has been demonstrated to on the tomo MLO.33
the MLO view are better appreciated in
have poorer clinical performance. There These results are also consistent with tomo imaging than in a 2D MLO image,
is considerable evidence that two-view evaluations in which studies comparing some cancers in the CC view are better
tomo has increased sensitivity relative to the ROC performance of two tomo views appreciated in tomo imaging than in a 2D
one-view tomo. This has been illustrated demonstrate superior performance over CC image. Indeed, the recent publication of
in the reader study reported by Rafferty in two-view digital mammography (Michell), a clinical trial in support of tomosynthesis
which the clinical performance of two-view but studies comparing one-view tomo approval for GE Healthcare showed no
2D combined with a single (MLO) tomo to two-view digital mammography have superiority of single-view tomosynthesis
view was inferior to the performance of poorer performance and do not show compared to 2D mammography.37

6
The presentation by Zuley et al., which compared to single-view tomo, in which phy practice and critical to perceiving subtle
looked at the visibility of cancers in the CC sensitivity will suffer. changes that may be associated with a
and the MLO tomo views, concluded that An alternative approach to acquiring two cancer. Obtaining a 2D exam along with
the tomo CC view depicted “substantially tomo views, given a fixed radiation dose, the tomo exam allows direct comparison
more cancers” than the tomo MLO view.38 would be to acquire only one tomo view, of current 2D images with prior 2D images.
Given this body of evidence, if only one but double the dose for that view. This The 2D exam is also useful for the rapid
tomo view is to be acquired, the MLO certainly would lower noise and may result detection of calcifications and perception of
tomo view appears to be the wrong one in a superior image due to the increased their distribution. Segmental and clustered
if maximizing cancer detection is the photon statistics. However, better clinical calcifications are more easily and quickly
clinical goal.39 performance has been seen for two-view appreciated with 2D because they can
An even more challenging situation is tomo than for higher dose single-view traverse multiple tomo slices.
when the exam consists solely of a tomo tomo. The Gennaro 2009 study showed
The tomosynthesis portion of the 2D plus
MLO. It could be difficult to see asymme- that the use of single-view tomo at 2x dose
tomo exam is also critical in optimizing
tries with only one view, and comparison to achieved inferior performance, compared
performance. The tomosynthesis image
2D prior images would also be challenging. to digital mammography, whereas Michell,
reduces structure overlap, minimizing re-
The best clinical performance will likely be who used two tomo views at approximately
calls for overlapped structures and better
seen in protocols that acquire both tomo CC 1x dose each, achieved superior
demonstrates masses and architectural
and MLO image sets. performance.33,35
distortions.
Performing two views uses more radiation
Benefits of Having Both Tomo Because of the value of having both 2D
dose than one view. However, these doses and 2D Images in All Views and 3D views, the original Hologic FDA
are commonly accepted in conventional trials looked at the performance of 3D
There are several reasons why acquiring
mammography, in which two-view mam- imaging when used in combination with
both a 2D and tomo image together are
mography is performed to improve the 2D mammography. The trial demonstrated
useful, especially in screening. It is well
cancer detection rate. Likewise, two-view superior performance with the addition of
known that comparison of current images
tomo is associated with higher sensitiv- tomo, but at the cost of additional radiation
with prior images is standard mammogra-
ity along with reduction in recall rates, as

2D 3D

GREATER PERCEPTION OF EXTENT OF DISEASE: In addition to the subtle


area of architectural distortion best defined on the 3D reconstructed slice
on the right (top arrow), a second spiculated mass is also revealed (bottom
arrow) 21 mm posterior to the primary area of interest. (Diagnosis for both
areas: Invasive ductal carcinoma)
2D 3D
REDUCED NEED FOR WORKUP: 3D mammography demonstrates a
definitive architectural distortion only subtly appreciated on the 2D digital
mammogram, replacing the need for a diagnostic workup that may not fully
or accurately resolve the 2D imaging dilemma.

7
Tomo Results
Study Compared Trial Report
to 2D35

Tomo-MLO compared GE SenoClaire Summary of


Lower
to 2D FFDM Safety and Effectiveness

GE SenoClaire Summary of
Tomo-MLO plus 2D CC Lower
Safety and Effectiveness

GE SenoClaire’s tomosynthesis has not been shown to be superior to 2D,


and depending upon the protocol, it has been shown to be inferior to 2D.
The clinical performance of the GE tomo methods, as measured using area
under the ROC curve, were both lower than 2D.

Tomo CC Tomo MLO


TWO-VIEW TOMO OPTIMIZES THE CHANCE FOR CANCER DETECTION:
A lesion can be seen in the tomo CC view, but it is not apparent in any of
the tomo MLO slices. (The central MLO slice is shown above.) A number
of researchers have concluded that two-view tomo improves radiologist
performance over single-view tomo.

dose to the patient due to the essentially replace every 2D system with tomo-capable women per day can be accommodated on
double exposures of both tomo and 2D. units. It is likely that many facilities will machines dedicated to screening use. This
Using Hologic C-View software, it is pos- implement tomosynthesis mammography might be a consideration in situations where
sible to generate a 2D image directly from in phases, beginning with one or two a limited number of tomo-capable systems
the 3D dataset, obviating the need for 2D systems initially, similar to the pattern seen are available.
exposures, and providing the clinical ben- in the transition from analog to digital
efit of both 2D and tomo at essentially the mammography. During this implementation Use for Women with Dense Breasts
same radiation dose as a 2D exam alone. phase, facilities will need to develop criteria 3D mammography has been shown to
Additional information about this process for determining which patients will receive have value in both fatty and dense breasts,
is provided in the following section. tomo exams, as well as processes to ensure but it has a greater impact for women with
efficient patient management in a mixed dense breasts. Therefore, if a practice
Both the 2D and tomo images in an exam
environment. does not have enough systems to screen
are valuable. In conclusion:
Determining which patients should receive all women, it is reasonable to reserve 3D
•T
 he 2D image is useful for comparison
tomo exams is not a straightforward issue, mammography for women with dense
to priors.
and there is not a single solution that breasts. However, as 3D mammography
•T
 he 2D image allows for quick reading offers a benefit in both fatty and dense
will fit every situation. Each facility must
of microcalcifications. breasts, the eventual goal should be to
consider what is known about the benefits
•T
 he tomo image reduces structure of tomosynthesis and make decisions screen all women using tomosynthesis
overlap and better demonstrates imaging.
based on their unique requirements and
masses.
implementation strategy. Some potential
•U
 sing both 2D and tomo in both the considerations are outlined below.
Reducing Patient Dose in
CC and MLO views maximizes clinical 3D Mammography
performance. Use for Screening or
One area in which extensive research and
•U
 sing C-View software allows the Diagnostic Imaging
development efforts have been focused
generation of the 2D images with no 3D mammography has shown value in is the creation of a 2D image generated
additional radiation beyond the tomo a diagnostic evaluation of a symptomatic from a 3D dataset. This method provides
exposures. breast. It also can be used as a screen- a 2D image for use during image review,
Patient Selection and Management ing tool to improve sensitivity and reduce but does not require an X-ray exposure
recalls. Therefore, either or both indica- to generate the 2D image, as it is created
Most breast imaging centers have multiple tions are acceptable uses of the technol- directly from the 3D slices. In November
digital mammography systems, and it may ogy. Since diagnostic procedures often 2011, Hologic announced the commercial
not be economically feasible to immediately take longer than screening exams, more

8
LESION SEEN ONLY ON 3D CC IMAGING

No lesion is apparent on the standard 2D screening views, above.

The lesion cannot be seen on the 3D MLO slices, shown here at every 10 mm.

The lesion is easily seen on the 3D CC slice 21, but not on the 2D CC, even in retrospect.

9
release and CE mark of its C-View Wallace et al. studied the performance of tomosynthesis with C-View 2D images
synthesized 2D image reconstruction C-View 2D images in a reader study, and found comparable results regarding positive
algorithm that eliminates the need for using ROC analysis, they concluded that predictive values and cancer detection rates.43
a conventional 2D mammogram as a generated 2D mammograms with tomosyn-
In the Oslo trial, tomosynthesis
component of a tomosynthesis screening thesis allowed similar interpretive perfor-
with C-View 2D images showed
procedure. C-View technology became mance to standard digital mammography
comparable cancer detection and
available in the U.S. in 2013. in combination with tomosynthesis, and
positive predictive value as tomo
therefore, may be an acceptable alternative
This approach provides the advantage with digital mammography.
for screening.41 Zuley et al. studied the per-
of reducing the number of exposures,
leading to shorter exam times, increased formance of C-View 2D images in a reader Reading Time
patient comfort due to reduced time study, and concluded that generated 2D
alone or in combination with tomosynthesis, Breast tomosynthesis involves the genera-
under compression and reduced patient
is comparable in performance to standard tion of considerably more images than
dose. This software allows screening with
2D mammography alone or in combination standard 2D digital mammography, as each
3D mammography at the same dose as
with tomosynthesis, and may eliminate the single 2D digital image is now replaced
conventional digital mammography.40
need for acquired 2D images as part of a with perhaps 50 or more slices. As a result,
The performance of the generated 2D routine clinical study.42 the time to perform the evaluation of these
image has been evaluated in the clinical images has been a topic of interest. Initial
trial in support of the C-View software FDA The performance of C-View software in
results indicated that the reading time
submission. The clinical trial demonstrated screening has perhaps been most exten-
approximately doubled from 49 seconds
that: sively measured in the Skaane screening
to 92 seconds for a 4-view bilateral exam.3
trial. Using tomosynthesis plus C-View
• 3D mammography with C-View 2D images As readers gain experience, it now appears
2D images, Skaane reported an increase
is superior to 2D alone for all breast types. that the increase in reading time using 3D
in cancer detection compared to 2D
mammography is shrinking. The group
• 3D mammography with C-View 2D imaging. A direct comparison between the
from Oslo reports that after reading 2,000
images is superior to 2D alone in performance of tomosynthesis with digital
examinations, the 3D mammography read-
reducing recall rates. mammography to the performance of
ing time dropped 40% from initial values

2D C-View 2D
C-View images eliminate the need for additional exposures and keep the dose
for 3D mammography exams comparable to that of a conventional 2D digital
mammography exam.

The Affirm™ breast biopsy guidance system used with the Hologic Selenia
Dimensions 3D mammography system enables 3D guidance for biopsy
procedures.

10
and is now ~60 seconds.45 They conclude be performed if a lesion cannot be located biopsy can be performed as long as the
that 3D mammography interpretation time using standard biopsy imaging methods? system supports tomosynthesis-guided
is acceptable for high-volume screening. Many lesions found with 3D mammography needle localizations.48 The ability to
Other researchers report that the reading can in retrospect be located and biopsied perform tomosynthesis imaging using a
time for 3D mammography is about 50% using ultrasound or stereotactic guidance. biopsy paddle and fast tomosynthesis
longer than for 2D mammography.46 But subtle lesions sometimes can only reconstructions may facilitate biopsies
be identified using 3D imaging.47 This using tomosynthesis.
Advances in Hologic 3D requires that biopsy systems employ
Mammography imaging and localization using 3D. The Contrast-enhanced
Hologic 3D mammography system offers Breast Imaging
The growing adoption of 3D mammography
an interventional add-on device that
in clinical use creates an opportunity for Contrast-enhanced breast imaging is a
utilizes 3D imaging for lesion identification
technological evolutions that may be procedure that images the distribution of
and targeting. Using this device, a single
useful in streamlining workflow, improving an iodinated contrast agent using either
tomo scan is performed, the lesion is
diagnostic accuracy and expanding clinical 2D or 3D X-ray imaging technologies. This
targeted and the x,y,z location of the lesion
applications. Some of the recent advances technology is in its early evaluation stage,
calculated directly from the 3D image.
and ongoing efforts in these areas are but may offer some advantages relative
discussed in the following sections. Advantages of this procedure compared to contrast breast MRI in terms of reduced
to stereotactic biopsy include improved cost, comparable care to patients for whom
3D Guided Biopsy visibility of lesions that are occult in 2D MRI is contraindicated, and access to
The ultimate diagnosis of a breast cancer imaging, faster lesion targeting, fewer patients in areas where MRI systems are
lesion is made using biopsy tissue X-ray exposures, and reduced patient not available.49 Contrast-enhanced breast
sampling. The ability for 3D mammography procedure time. Even without the add- imaging combines functional information
to identify lesions not readily visible with on biopsy device, there is a method from the distribution of the contrast agent
2D digital mammography or ultrasound whereby lesions can be targeted under and morphological information from the
has created a problem – how can a biopsy tomosynthesis guidance. An open surgical X-ray images. Hologic has received FDA

2D 2D contrast 3D 3D contrast

CONTRAST IMAGING: This study of 2D and 3D iodine contrast mammography was acquired under a single compression. The proven cancer in the subareolar
breast (horizontal arrow) is not visible on the enhanced 2D mammogram except for the clips placed at biopsy, but is easily seen on the 2D and 3D dual energy
contrast images. Contrast imaging led to the detection of an additional cancer in the far medial breast (downward arrow.) The 3D image shows the irregular
shape of the lesion, making it highly likely that the lesion is malignant.

11
approval and CE mark for a dual modality superior performance in the detection of diagnostic benefits, including enhanced
system, which is capable of imaging the masses and architectural distortions and performance in assessing tumor size and
functional 2D contrast uptake and the mor- equivalent or slightly better performance in stage and more clearly demonstrating
phological 3D image in rapid sequence, the detection of microcalcifications in using margins and extent of lesions.
and combining these two image sets into a 2D plus tomo imaging compared to 2D Important new applications involving 3D
single fused study. In the fused study, the alone. Acquisition of both the CC and MLO mammography include contrast-enhanced
2D contrast image can identify potential views in 2D and tomo provided statistically imaging for patients when access to breast
lesions based on their physiological state significant superior performance compared MRI is limited or contraindicated, and meth-
that causes increased contrast agent to 2D alone; however, use of only the MLO ods of biopsying lesions under 3D image
uptake. The standard 3D image can then tomo with both the 2D CC and MLO tomo guidance.
be overlaid on the 2D contrast image and views also provided better performance
provide morphological information on the compared to 2D alone – just not as good
lesion, such as improved visibility of associ- as acquiring both CC and MLO tomo views.
ated spiculations. Finally, it was demonstrated that the addi-
tion of 3D imaging to conventional 2D imag-
Conclusions ing provides improved performance in both
fatty and dense breasts, compared to 2D
Breast tomosynthesis is an exciting
alone, with the performance gain in dense
technology that is revolutionizing breast
breasts higher than in fatty breasts.
imaging. It has demonstrated value in both
screening and diagnostic evaluations. With the use of the C-View generated 2D
The improvements in clinical performance, image, cancer detection and other clinical
compared to 2D mammography, are benefits of 3D mammography are available
significant. Multiple peer-reviewed clinical at comparable radiation dose to standard
publications report that the use of two-view 2D digital imaging and at about half the
tomo in screening offers both improved dose of 2D plus 3D imaging.
cancer detection rates and reduced call- There is a growing body of evidence that
back rates compared to 2D alone. 3D mammography has the potential to
Clinical studies using the Hologic 3D reduce the number of exposures needed
mammography system have demonstrated for diagnostic imaging and provide other

Glossary
2D Conventional digital mammography. Also known as FFDM.

A technology involving limited angle tomography acquisition and reconstruction. Also referred to as
3D
digital breast tomosynthesis, DBT, 3D tomosynthesis, tomosynthesis and tomo.
Same as recall rate. The percentage of women recalled from screening for further assessment.
Callback rate
In mammography screening, the majority of recalled cases are false positives.
An imaging mode whereby both a 3D and 2D digital mammography image set are acquired in one
Combo mode
breast compression.

C-View™ 2D image generated from the 3D reconstructions.

The percentage of women recalled from screening for further assessment. In mammography
Recall rate
screening, the majority of recalled cases are false positives.

ROC Receiver Operating Characteristics

Sensitivity The measure of how many cancers are detected.

Specificity The measure of how many non-cancers are correctly identified.

Synthesized or generated 2D A method of creating a 2D image from a reconstruction of a 3D dataset. See C-View.

12
References: 12 
Rafferty EA, Niklason L. FFDM versus FFDM Moon WK, Cho N, Bae Min Sun et al. Value
24 

with tomosynthesis for women with radiologi- of additional digital breast tomosynthesis
1
 afferty EA, Park JM, Philpotts LE et al.
R cally dense breasts: an enriched retrospective combined with digital mammography in a
Assessing radiologist performance using reader study. Radiological Society of North diagnostic setting. Radiological Society of
combined digital mammography and America annual meeting, Chicago, IL, 2011. North America annual meeting, Chicago, IL,
breast tomosynthesis compared with digital 2013.
13 
Rafferty EA, Niklason L, Smith A. Comparison
mammography alone: results of a multicenter, of FFDM with breast tomosynthesis to FFDM Michell M, Iqbal A, Wasan R, et al.
25 

multireader trial. Radiology Jan; 266(1): 104-13, alone: performance in fatty and dense breasts. A comparison of the accuracy of digital breast
2013. Tomosynthesis Imaging Symposium, Duke tomosynthesis with two dimension digital
2 
Gur D, Bandos AI, Rockette HE, et al. Is an University, 2009. mammography in measurement of breast
ROC-type response truly always better than tumour size. Radiological Society of North
14 
Philpotts LE, Raghu M, Geisel JL, et al.
a binary response in observer performance America annual meeting, Chicago, IL, 2010.
Tomosynthesis in Breast Cancer Visualization
studies? Acad Radiol. (17)5:639-645, 2010. as a Function of Mammographic Density. Fornvik D, Zackrisson S, Ljunberg O, et
26 

3 
Skaane P, Osteras BH, Ebeb EB, Gullien Rl. Radiological Society of North America annual al. Breast Tomosynthesis: Accuracy of
Comparison of Digital Mammography (FFDM) meeting, Chicago, IL, 2013. tumor measurement compared with digital
and FFDM Plus Digital Breast Tomosynthesis in mammography and ultrasonography. Acta
15 
Gandini G, Fonio P, Bergamasco L, et al.
Mammography Screening for Cancer Detection Radiol 3; 240-247, 2010.
Comparative study with digital mammography
according to Breast Parenchyma Density. (DM) combined with digital breast tomosyn- Meacock LM, et al. The accuracy of
27 

Radiological Society of North America annual thesis (DBT) for detection of invasive lobular breast cancer size measurement: Digital
meeting, Chicago, IL, 2014. carcinoma (ILC). Radiological Society of North breast tomosynthesis (DBT vs. 2D digital
4
 kaane P, Bandos AI, Gullien R, Eben EB, et al.
S America annual meeting, Chicago, IL, 2013. mammography (DM)). European College of
Comparison of digital mammography alone Radiology annual meeting, Vienna, Austria,
16
 erg WA, Blume JD, Cormack JB, et al.
B
and digital mammography plus tomosynthesis 2010.
Combined screening with ultrasound and
in a population-based screening program. mammography vs. mammography alone in Tagliafico A, Astengo D, Cavagnetto F, et al.
28 

Radiology, Apr; 267(1):47-56, 2013. women with elevated risk of breast cancer. One-to-one comparison between digital
5
 kaane P, Bandos AI, Gullien R et al. Prospec-
S JAMA May 14; 299(18):2151-63, 2008. spot compression view and digital breast
tive trial comparing full-field digital mammog- tomosynthesis. Eur Radiol. Mar,22(3):539-44,
17 
Aguillar VL, Ferreira VVC, Endo E, et al. Whole
raphy (FFDM) versus combined FFDM and 2012.
breast US after Screening Breast Tomosynthe-
tomosynthesis in a population-based screen- sis: Initial Experience. Radiological Society of Wald NJ, Murphy P, Major P, et al. UKCCCR
29 

ing programme using independent double North America annual meeting, Chicago, IL, multicentre randomized controlled trial of one
reading with arbitration. Eur Radiol 2013 Aug; 2014. and two-view mammography in breast cancer
23(8):2061-71. screening. BMJ 311:1189-93, 1995.
18
 hung J, Cha ES, Lee JE, et al. Digital Breast
C
6 
Ciatto S, Houssami N, Bernardi D, et al. Tomosynthesis and Breast Ultrasound: Rafferty EA, Park JM, Philpotts LE, et al.
30 

Integration of 3D digital mammography with Additional Roles in Dense Breasts with Diagnostic accuracy and recall rates for digital
tomosynthesis for population breast-cancer Category 0 at Conventional Digital mammography and digital mammography
screening (STORM): a prospective comparison Mammography. Radiological Society of North combined with one-view and two-view
study. Lancet Oncology. Jun; 14(7):583-9, 2013. America annual meeting, Chicago, IL, 2014. tomosynthesis: results of an enriched reader
7 
Haas BM, Kalra V, Geisel J, et al. Comparison of study. AJR Am J Roentgenol. Feb; 202(2):273-
19 
Zuley ML, Bandos AI, Ganott MA, et al. Digital
Tomosynthesis Plus Digital Mammography and 281, 2014.
breast tomosynthesis versus supplemental
Digital Mammography Alone for Breast Cancer diagnostic mammographic images for Rafferty E, Niklason L, Jameson-Meehan
31 

Screening. Radiology 2013 Dec;269(3):694- evaluation of non-calcified breast lesions. L. Breast tomosynthesis: one view or two?
700. Epub 2013 Jul 30. Radiology, Jan; 266(1): 89-95, 2013. Radiological Society of North America annual
8
 ainiero M, Lourenco A, Barry-Brooks M.
M meeting, Chicago, IL, 2006.
20 
Harvey JA, Nicholson BT, Cohen MA. Finding
Breast Cancer Screening Pre and Post- early invasive breast cancers: a practical Beck N, Butler R, Durand M et al. One-
32 

tomosynthesis: Comparison of Recall Rate, approach. Radiology, July 2008; 248:61-76. View Versus Two-View Tomosynthesis: A
Biopsy Positive Predictive Value, and Cancer Comparison of Breast Cancer Visibility in
Detection Rate. Radiological Society of North
21 
Butler RS, Crenshaw J, Kalra V, et al.
the Mediolateral Oblique and Craniocaudal
America annual meeting, Chicago, IL, 2013. How tomosynthesis optimizes patient work
Views. American Roentgen Ray Society annual
up, throughput, and resource utilization.
9 
Mitra N, McCarthy AM, Kontos D, et al. meeting, Washington, D.C. April 2013.
Radiological Society of North America annual
Implementing Digital Breast Tomosynthesis meeting, Chicago, IL, 2013. Baker JA, Lo JY. Breast tomosynthesis: state-of-
33 

(DBT) in a Screening Population: PPV1 as a the-art and review of the literature. Academic
Measure of Outcome. Radiological Society of
22 
Savhn T, Andersson I, Chakraborty D, et al.
Radiology 18:10. 1298-1310, 2011.
North America annual meeting, Chicago, IL, The diagnostic accuracy of dual-view digital
2013. mammography, single-view tomosynthesis Michell MJ, Iqbal A, Wasan R, et al. Phase I trial
34 

and a dual-view combination of breast to determine the performance of digital breast


10 
Rose SL, Tidwell AL, Bujnoch LJ, et al. tomosynthesis and digital mammography in tomosynthesis versus two dimension digital
Implementation of breast tomosynthesis in a a free-response observer performance study. and film-screen mammography. Radiological
routine screening practice: an observational Radiat Prot Dosimetry. 139(1-3): 113-117(2010). Society of North America annual meeting,
study. AJR Am J Roentgenol. 2013 Jun; Chicago, IL, 2010.
200(6):1401-8, 2013.
23 
Rafferty EA, Saksena MA, Halpern EF, et
al. Comparison of lesion detection and Gennaro G, Toledano A, di Maggio C, et al.
35 
11
 riedewald SM, Rafferty EA, Rose SL, et al.
F characterization in invasive cancers using Digital breast tomosynthesis versus digital
Breast Cancer Screening Using Tomosynthesis breast tomosynthesis versus conventional mammography: a clinical performance study.
in Combination with Digital Mammography. mammography. Radiological Society of North Eur Radiol 20:1545-1553, 2009.
JAMA 2014;311(24):2499-2507. America annual meeting, Chicago IL, 2013.

13
 allis MG, Moa E, Zanca F, et al. Two-view
W
36 
Zuley M, Guo B, Catullo V, et al. “Comparison
41 
 onnenschein M. The use of breast
S
46 

and single-view tomosynthesis versus full- of Two-dimensional Synthesized Mammograms tomosynthesis with C-View images and
field digital mammography: high-resolution versus Original Digital Mammograms Alone tomosynthesis guided biopsy in daily practice:
x-ray imaging observer study. Radiology, and in Combination with Tomosynthesis our experience in Switzerland. Symposium
Mar;262(3):788-96, 2012. Images.” Radiology. 2014 Jun;271(3):664-71. SY23, European College of Radiology annual
meeting, Vienna, Austria, 2014.
GE SenoClaire Summary of Safety and
37 
Skaane P, Bandos A, Eben E, et al. “Two-View
42 

Effectiveness, August 26, 2014. http:// Digital Breast Tomosynthesis Screening with  chrading S, Distelmaier M, Dirrichs T, et al.
S
47 

www.accessdata.fda.gov/cdrh_docs/pdf13/ Synthetically Reconstructed Projection Images: Digital Breast Tomosynthesis-guided Vacuum-


P130020b.pdf. Comparison with Digital Breast Tomosynthesis assisted Breast Biopsy: Initial Experiences and
with Full-Field Digital Mammographic Images” Comparison with Prone Stereotactic Vacuum-
Zuley ML, Koo JJ, Plecha DM, et al. Analysis
38 

Radiology. 2014 Jun; 271(3):655-63. assisted Biopsy. Radiology, Ahead of Print,


of Cancers Missed on Digital Breast
10.1148/radiol.14141397, 2014.
Tomosynthesis. Radiological Society of North Bernardi D, Pellegrini M, Valentini M, et al.
43 

America annual meeting, Chicago IL, 2014. The STORM II Trial: Interim Comparison of  reer PE, Niell B, Rafferty EA. Preoperative
F
48 

Screen-reading Strategies in Population Breast Tomosynthesis-guided Needle Localization of


http://www.fda.gov/downloads/AdvisoryCom-
39 

Screening. Radiological Society of North Mammographically and Sonographically Occult


mittees/CommitteesMeetingMaterials/Medi-
America annual meeting, Chicago, IL, 2014. Breast Lesions. Radiology, Ahead of Print,
calDevices/ MedicalDevicesAdvisoryCommit-
10.1148/radiol.14140515, 2015.
tee/RadiologicalDevicesPanel/UCM325901.pdf. Ekseth U, Jahr G, Izadi M, et al. Trends in Time
44 

to Interpretation of Tomosynthesis Based Yang TL, Chiang CL, Chou CP. Contrast-
49 
Wallace LP, Rathfon GY, Ganott MA et al.
40 

Screening Examinations with Increasing enhanced breast tomosynthesis versus


Diagnostic Accuracy of Combination Synthetic
Experience. Radiological Society of North dynamic contrast-enhanced breast MRI in
Mammograms with Tomosynthesis vs. Combi-
America annual meeting, Chicago, IL, 2013. the diagnosis of suspicious breast lesions on
nation FFDM with Tomosynthesis. Radiological
mammograms. Radiological Society of North
Society of North America annual meeting, Dang PA, Freer PE, Humphrey KL, et al.
45 

America annual meeting, Chicago IL, 2013.


Chicago, IL, 2013. Addition of tomosynthesis to conventional
digital mammography: effect on image
interpretation time of screening examinations.
Radiology, Jan; 270(1):49-56, 2014.

Hologic, Inc. WP-00060 REV 005 ©2015 Hologic, Inc. Hologic, Affirm, C-View, Dimensions, Hologic 3D
35 Crosby Drive mammography, 3D mammography and Selenia are trademarks and/or registered trademarks of
Bedford, MA 01730-1401 USA Hologic, Inc. and/or its subsidiaries in the United States and/or other countries. This information is
Tel: +1.781.999.7300 intended for medical professionals in the U.S. and other markets and is not intended as a product
Sales: +1.781.999.7453 solicitation or promotion where such activities are prohibited. Because Hologic materials are
Fax: +1.781.280.0668 distributed through websites, eBroadcasts and tradeshows, it is not always possible to control where
www.hologic.com
such materials appear. For specific information on what products are available for sale in a particular
country, please contact your local Hologic representative or write to womenshealth@hologic.com.

You might also like