CXDI APPLICATIONS SETUP
ADDENDUM FOR VERSI 3
4. SYSTEM/ SETUP MENU / CUSTOMIZE DISPLAY
There no longer is a setting to display the tray button on the screen when category screen
is used. The choice is now category or tray and is not switch able in the user mode.
Page | of 62. SYSTEM {/ SETUP MENU / SYSTEM SETUP / ANNOTATION
a =
-
Select “FREE ANNOTATION” and the following menu will appear: Press ‘ENABLE? to
see other selections available.
EE ANNOTATION
Page 2 of 6‘The font size can be set as a default from 8 to 144. The user can also change it when
entering free text onto the image. The default arrow width can be set to thin, medium, or
thick. The user can also change arrow width when entering arrows.
To enter a list of most often used text, click on “SET” next to ‘TEXT LIST”
EXPIRATION
AWANEIGHTS
Wi0 WEIGHTS
STANDING
WT BEARING.
POST-OP RECOVERY RM
‘Once these are entered, the user ¢ this when they select the list in the
user mode:
Page 3 of 6Optional Seting
Srnec
Breer) eee
aN eco
Oe eae meee
i eceraetiste er
oe ear
fe FT nea ae
“COMMON ANNOTATION”: Annotation Settings to be displayed on ANNOTATION
Screen in QA Mode: There are 3 choices, Fixed Annotation, Free Annotation, and Multi
text_If Free Annotation is selected, the user can still modify Fixed Annotation as well.
Wendy Taylor fs
fer ECC) Check image“FREE ANNOTATION”: Automatic display of keyboard gives the choice of adding text
through list or using the keyboard:
Cretan
| |
Page 5 of 63. SYSTEM / EDIT EXPOSURE MODE
Masking must first be enabled by service in the service/setup mode. Then each exposure
mode must be re-programmed with the “COLLIMATION MASK” setting enabled under
the ‘CORRECT PROCEDURE’ tab.
Page 6 of 6‘The operator's name of the currently logged in user is displayed in the ‘OPERATOR’S NAME"
field of “STUDY INFO’ during exposure. This replaces the previous list of operator's in
‘STUDY INFO’ which the user selected from when entering patient demographics for an exam.
The operator's nameinitials are still output in the DICOM header and can be displayed on the
image as previously done. Bear in mind if the display is on the image, initials should be used
instead of the whole name. The logged in operator’s name is also output to the audit log and
cannot be changed during exposure or when reproducing i
Sree oy
Renee
a
Cerrar
Lewgt tea
ke Lower | hole Lower
Exveniy
‘When the logged in user desires to log out, press the user name button appearing on the bottom
left of the main menu:
Page 9 of 10rte
Press OK and the login screen will appear for the next user to log
Login
ee
rer)
Service will set up the other functions if requested by the customer.