URIC ACID FS*
Toos
Cat. No, Kitsz0
40300021 RIAX 20m + RZ1x _20mI + 1x Smistd
19001 99 10025 RISK 8Dmi+R21x 100ml
40300029 RIT GODmi+R21x 200ml
10300 030 6x Smi standard
Method
Enzymatic colorimetric test, "TOOS*
Utic acid is oxidized to Allantoine by Uricase. The generated
Hydrogen Peroxide reacts wih 4-Aminaantipyrine and N-Ethy
NA2-nycroxy-3-sulfopropyl}¢m-tolucin (TOOS) toa blue violet dye.
Reagent preparation and stability
“The reagents and the standard are readysto-use and stable upto the
‘ond ofthe indicated month of expiry, if cantamination is avoided and
Sslored at 2~ 8 °C. The reagents must be protected from light.
Specimen
‘Serum, heparinized or EDTA plasma, urine. For serum / plasma
Avoid hemolysis! Dilute urine + 10 with dit, water,
Components and concentration in the test
Rt: Phosphate buter pH7.0 100 mmalt
Toos ‘mmol
Ascorbate Oridase 21 KUM
R2: Phosphate buter pH70 100 mmatt
‘Aminoantpyrine 0,3 mmol
Potassiummhexacyanoterate (I) 10 mola
Po 21 KU!
Urcase 250 Un
‘Standard: 6 mola (357 umnolt)
Notes
“The reagents contain Sodium Azide (0095 %) as preservative. Do
not swallow! Avoid contact wth skin and mucous membranes!
Normal range ** rene)
‘Serum’
Men: 34-7.0mgldl 200-420 yo
Women: 24-S.7mgidl 140 340 matt
Urine: 250-750 mg!24n 1,5 4.5 mmoli24 n
References
4. Barham, D., Trinder,P., Analyst. 97, (1972), 142 — 145,
2 Thotels, W., Hoffmeister, H., Busch, E.W., Koller, P.U.,
Volimar, J, Dtsch, Med. Wehr. 98, (1973), 980 386
3. Hem, Rd, Cannon, D.C, Winkelnann J: Clinical Chemisty,
Principles and Technics, 2" edition, Harper & Row Publishors
Ine, Hagerstown Marylend 1974, 534
4. Fossall,P., Prencipe, L., Bet G., Cin. Chem, 26, (1980)
22231
+ fuid stable
Targa BT 3000
INSTRUMENT PARAMETER
Test Typ Endpoint with
Starter
‘Serum Starter (Inactive)
Filters (A/B) 546700
Units mg/dl
Test Method With Factor
Test Methodology Toos
Number of washes an
Delay Time (Sec) 0
Inc. Time (Sec) 300/300
Reading Time (Sec) 10
Test Limit (Conc) 20
Reactio Limit (mABS) 0.000
Max ABS Delta (mABS) 0.000
Reagent mAbs Limit (mABS) 100
Reagents A/B (ul) 240/60
Reaction Direction Increasing
Reagent Dilution 1
Initial ABS. (maBs) 0.000
Curve Acceptance 100%
‘Automatic profile (Inactive)
Rerun Test Rat Bik HM — 00:00
Pathological Repetition (Inactive)
Serum Parameters
Tests Name URIC ACID
‘Sample Volume ww 6
Dilution Ratio 18
Min, Max. M. 3.4017.00
Min: Max. F. 2.4015.70
Min. Max. B. (0.000/0.000
Urine Parameters
Tests Name
Urine Volume uy 8
Min, Max. Mt 0.000 / 0,000
Min: Max. F. 0.000 1 0.000
Min, Max. B. 0.000 1 0.000
Auto Dilution 410
Multi Factor 4,00
CALIBRATOR:
15.1097