You are on page 1of 2

Chapter 7 · The Polyethylene History – A. Bellare, M.

Spector
45 7 7

7 The Polyethylene History


A. Bellare, M. Spector

Summary debris not only damages the surface of the PE component


but is also known to elicit a biological response that often
Our understanding of the wear behavior of polyethylene results in bone resorption. This bone loss (referred to as
(PE) components has deepened over the past few years as osteolysis) can eventually lead to loosening of the pros-
the adverse effects of gamma irradiation (in air) steril- thetic device.The location and size of PE particle-induced
ization have become understood.This understanding has osteolytic lesions often greatly complicate revision
led to methods to improve the wear performance of the surgery. Work in recent years has focused on processing
polymer using cross-linking. However, it will still be parameters that serve as the determinants of the resis-
several years before the clinical benefits of new methods tance of PE to wear. The reduction of the amount of wear
of processing PE are clear and the benefit-risk ratio is debris from,and surface damage to,PE would prolong the
established. lifetime of such prostheses.
The objective of this chapter is to review the history
of the use of polyethylene in total joint arthroplasty, as
Introduction a basis for understanding the methods being employed
to improve its performance. There are several prior
Ultra-high-molecular-weight polyethylene (UHMWPE) reviews [26, 27] of this subject that can be accessed for
is one of the principal materials employed in total knee useful reference.
arthroplasty. While the lubrication and friction of the
metal-on-PE articulation provides the low-friction
arthroplasty that Sir John Charnley sought,the wear of PE Polyethylene Molecular Structure
yields particulate debris that potentiates an osteolytic
response, which remains a significant problem. The often UHMWPE has a very low frictional coefficient against
rapid and extensive destruction of bone attributable to PE metal and ceramics and is therefore used as a bearing
wear particles is so dramatic, and so challenges revision surface for joint replacement prostheses. Moreover, the
arthroplasty, that it has commanded the most attention wear resistance of UHMWPE is greater than that of other
in recent years.However,the actual incidence of this prob- polymers investigated for this application. Low strength
lem remains somewhat in question. This point notwith- and creep, however, present potential problems.
standing, the prevalence of PE wear particle-induced The term polyethylene refers to plastics formed from
osteolysis is great enough to warrant changes in how the the polymerization of ethylene gas. The possibilities for
material is processed so as to improve its resistance to structural variation of molecules formed by this simple
wear.Extrinsic factors that contribute to the wear of poly- repeating unit for different molecular weight (e.g., crys-
ethylene are also being addressed: prosthetic designs that tallinity, branching, and cross-linking) are so numerous
reduce stresses in the polymer; prosthetic designs and and dramatic, with such a wide range of attainable prop-
manufacturing processes that reduce the number of erties, that the term polyethylene refers to a wide array of
particles released from modular junctions, which can materials. The earliest type of polyethylene was made by
participate in three-body wear of polyethylene; and reacting ethylene at high (20 000–30 000 pounds per
materials that may allow the production of more scratch- square inch) pressure and temperatures of 200°–400°C
resistant metallic counterfaces. with oxygen as catalyst. Such material is referred to as
It is well known that PE components of total joint low-density polyethylene. A great amount of polyethylene
replacement prostheses undergo processes that produce is produced now by newer,low-pressure techniques using
PE wear debris due to the articulation of the harder aluminum-titanium (Ziegler) catalysts. This is called lin-
metallic component, usually a cobalt-chromium alloy, ear polyethylene due to the linearity of its molecules, in
against the softer PE component. The generation of wear contrast to the branched molecules produced by high-
46 II . Past Failures

pressure processes. The linear polymers can be used to was Montell (Formerly Himont), which produced the
make high-density polyethylene by means of the higher Hi-Fax 1900H, a resin that has different structure and
degree of crystallinity attained with the regularly shaped properties compared to the Hoechst resins [45]. How-
molecules. Typically, there is no great difference in mole- ever, Hi-Fax is no longer available, and GUR 1050 and
cular weight between the low- and high-density varieties, GUR 1020 remain the only grades of polyethylene used
e.g., 100 000–500 000. However, if the low-pressure in orthopedic implants.
process is used to make extremely long molecules, i.e., The as-synthesized polyethylene resin particles are
UHMWPE, the result is remarkably different. This mate- approximately 100 µm, but can be submicrometer in size
rial, with a molecular weight between 1 and 10 million, is as well. The broad size distribution of GUR 4150 (the dig-
less crystalline and less dense than high-density polyeth- it “4” refers to the country code, USA, which was the
ylene and has exceptional mechanical properties. It is ex- nomenclature used for GUR resins.) powder particles
tremely tough and remarkably wear resistant; a 0.357 have been measured by Pienkowski et al. [35, 36]. Each
magnum bullet fired from 25 feet bounces back from a powder particle contains 10- 30-µm diameter aggregates
l-inch thick slab of UHMWPE. The material is used in comprising approximately 1-µm diameter nodules con-
very demanding applications (e.g., ore chutes in mining nected to each other by fibrils.Olley et al.[34] have shown
equipment) and is by far the most successful polymer that voids or defects remain along the resin boundaries
used in total joint replacements. It far outperforms the even after the powder is "fully” consolidated into bulk
7 various acrylics, fluorocarbons, polyacetals, polyamides, components. The likely reason for the presence of defects
and polyesters which were tried for such purposes. is the high viscosity associated with the ultra-high mole-
cular weight of the polyethylene that is required for high
wear resistance. The incomplete consolidation of high-
Processing of Polyethylene molecular-weight polyethylene resin compared with low-
molecular-weight polyethylenes, however, is not a major
Implants of polyethylene are usually manufactured by the concern. Gul et al. showed that there was no correlation
machining of components from bulk stock fabricated between the degree of consolidation of UHMWPE pow-
from as-synthesized polyethylene powder using mainly der particles and the rate of generation of particulate
ram extrusion or compression molding. These processes wear debris under the processing conditions that they
involve application of heat and pressure to consolidate the used [23].
powder into bulk components, followed by machining of The nascent UHMWPE powder contains extended-
the implant components, packaging, and sterilization. chain crystals (thick lamellae) as well as thin lamellae
Various grades of UHMWPE resins have been avail- [16]. The high melting temperature of 141°C observed us-
able for orthopedic implant application, primarily from ing a differential scanning calorimeter suggests that the
Ruhrchemie AG, which later changed its name to powder contains mostly extended-chain crystals, as pre-
Hoechst, and is currently called Ticona. The early sent in high-pressure crystallized UHMWPE. However, a
UHMWPE resin used by John Charnley was called RCH- study utilizing morphological, chemical, and molecular
1000 for (R)uhr (CH)emie. This resin is similar to the techniques indicated that a dual lamellar structure exist-
current GUR 1020 UHMWPE resin. RCH-1000 was clas- ed.It is postulated that the fibrils of the polyethylene resin
sified as a form of HDPE (high-density polyethylene), powder contain thick, extended-chain crystalline lamel-
which is why earlier papers refer to UHMWPE as HDPE lae, while 20-nm thick lamellae (such as those present in
[26]. Later, the orthopedic grade of polyethylene was bulk components manufactured using molding or ram
called CHIRULEN. Since the 1990s, the UHMWPE resin extrusion) exist in the spherical domains [16].It is unclear
used in implants has been called GUR or (G)ranular why the powder morphology contains fibrils connecting
(U)HMWPE (R)uhrchemie. Common examples of poly- spherical domains (sometimes referred to as the “cauli-
ethylene resins used today are GUR 1050 and GUR 1020. flower” morphology). The presence of fibrils in the
The numbers following GUR refer to the following: the powder suggests that the as-synthesized powder has
first digit refers to approximate or loose density (1); the UHMWPE macromolecules trapped in a low-entangle-
second digit refers to presence (1) or absence (0) of cal- ment, aligned state compared with melt-crystallized
cium stearate, which has been used as a lubricant to as- UHMWPE. This low entanglement would assist in con-
sist processing; the third digit refers to molecular weight solidation of powder during molding or ram extrusion
(2=2 million g/mole and 5=5 million g/mole), and the processes.A highly entangled state would make it harder
fourth digit (0) refers to the resin grade. Calcium to consolidate the powder, since it would require the
stearate is no longer added to orthopedic-grade poly- chains from powder particles to disentangle and then re-
ethylene, since reports showed increased levels of oxi- entangle with the chains of the adjacent powder particles.
dation and fusion defects associated with calcium The most common processes used to consolidate
stearate [24, 40, 43, 46]. Another source of UHMWPE polyethylene powder particles into bulk stock are com-

You might also like