Professional Documents
Culture Documents
2233000-100 Rev 18 MINItrace Service Manual - Maintenance
2233000-100 Rev 18 MINItrace Service Manual - Maintenance
MINItrace
Service Manual – Maintenance
Class A
Direction 2233000-100
Revision 18
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
REVISION HISTORY
17 May 28, 2019 Sec 3-4-2, 3-5, 4, 7-3-6-2: List of Cyclotron Service Tools and references removed. [PCR 17-
108]
Sec 3-5-6: Wording improved.
Sec 3-8: Reference information for consumables list added. [PCR 16-206]
Sec 3-9-2, 3-9-11: Tables updated. [PCR 16-195]
Sec 7: New information added for GCU:s with the new external PS solution. [PCR 16-045]
Sec 7-3: The term "Hyperterminal" is replaced by "a terminal communication software
(terminal emulation)". Any specific information regarding Hyperterminal is rewritten to fit
the use of any terminal. [PCR 16-076]
Sec 7-3-1: Figure 7-20 removed. [PCR 17-024, 18-024]
Sec 7-3-1-3: Table 7-17 updated. [PCR 17-024]
Sec 7-3-8: Table 7-24 updated. [PCR 17-024]
Sec 10-6-1: Table 10-5 updated. [PCR 17-226]
Sec 10-7-1: Information added to use the slot in the door when opening the magnet yoke.
Fig 10-6, 10-7: Layout changed to point out the slot in the door.
Sec 11-4: Updated with PSARC model PS/GEMS02A. [PCR 18-024]
Sec 11-12: Table 11-11 updated. [PCR 18-024]
Sec 12-2-2-1, 12-2-2-2: Replaced with reference to Sec 10-7-1 (duplicate information).
Sec 12-2-3: Updated with PSMC model MPS 9700. [PCR 17-024]
Sec 12-2-4: Table 12-1 updated. [PCR 17-024]
Sec 19-2: System requirements removed. [PCR 14-309]
App A: Obsolete links removed. [PCR 16-238]
16 March 31, 2015 Sec 3-7: Cyclotron job safety section updated. [PCR 14-263]
Sec 3-10-2: Warning and note added to sequence #8. [PCR 14-263]
Sec 9-2-2: Restructured. EVR-CA and CRA112 added in addition to EVR (TIDUR). [PCR 15-
029]
Sec 9-5-4: New section added for replacing temperature controllers. [PCR 15-029]
Revision history 3
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
14 July 11, 2014 Sec 3-9-3: New values added. [PCR 14-109]
Sec 3-9-17: Values updated.
Sec 7-3-2-4: Backup procedure for MSS ver. 2 added. [PCR 14-148]
Sec 9-5-3: New “Target water cooling” section added.
Sec 10-9-11-1: Oil drainage updated. [PCR 13-079]
Ch 11: Note added about DD4 being removed from S/N 28791. [PCR 14-109, 14-148]
Fig 11-4, Fig 11-9, Fig 11-10 and Fig 11-11: Updated with note and new measurements.
[PCR 14-109, 14-148]
Sec 11-4: New section with PSARC model PS/GEMS-01 added. [PCR 13-263]
Sec 11-9-4-2: Step tables updated. [PCR 14-109, 14-148]
Sec 11-9-8-1: Values updated.
Sec 11-9-8-2: Values updated.
Sec 13-6-12: New cable part numbers added. [PCR 13-137]
Sec 15-6-1: Spare parts removed and replaced with reference to MINItrace Spare Parts
Catalog (dir. 2380006-100).
Sec. 15-6-3: New “Targets” section added. [PCR 14-148]
Sec. 15-6-7: New “Water cooling” section added. [PCR 14-148]
Sec 15-7: New cable part number added. [PCR 13-137]
Ch 19: New chapter with information on the MSS software version 2.0 released 2014.
[PCR 14-148]
New appendix B: General radiation safety aspects. [PCR 13-158]
New appendix C: MINItrace radiation safety aspects. [PCR 13-158]
The previous chapters 19 and 20 moved to new appendix D. (Information on the MSS
software version that was discontinued in 2014.) Values updated in tables D-8, D-19, D-
22 and D-28.
13 September 23, Ch 6: Updated with more information about the control system structure. [PCR 13-157]
2013 Sec 17-7: Ladder and climbing safety updated. [PCR 13-200]
4 Revision history
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
12 September 4, All chapters: Reference added to LOTO procedures section. [PCR 13-129]
2013 Ch 2: New chapter with safety information. [PCR 13-129]
Sec 2-1: Reference to Safety Risk Assessment added. [PCR 13-129]
Sec 3-1: Reference added to MT Operator Guide for PM schedule. [PCR 08-198]
Sec 3-2: Safety information added. [PCR 13-129]
Sec 3-2-1: Reference to Safety Risk Assessment added. [PCR 13-129]
Table 3-1: Ladder added. [PCR 13-129]
Table 3-6, 3-7, 3-11, 3-12: LOTO added/updated. [PCR 13-129]
Sec 3-9-1: Fall hazard warning added. [PCR 13-129]
Sec 3-9-2: Pinch hazard warning added in seq. 7. [PCR 13-129]
Sec 3-9-13: LOTO added/updated. [PCR 13-129]
Sec 3-9-14: LOTO added/updated. [PCR 13-129]
Sec 7-2: Updated [PCR 13-157]
Sec 10-1: Warnings added [PCR 13-158]
Sec 10-9-15: Heavy object-warning updated. [PCR 12-082]
Sec 13-1: Updated [PCR 13-157]
Sec 13-2: Updated [PCR 13-157]
Sec 13-6-1: List added with spare parts requiring safety test. [PCR 13-129]
Sec 15-4, 15-5: Helium cooling figures updated. [PCR 12-337]
Sec 15-6-6: Heavy object-warning added. [PCR 12-082]
Sec 17-1: Fall hazard warning added. [PCR 13-129]
Sec 17-2: Pinch hazard warning updated. [PCR 13-129]
Sec 17-7: Ladder and climbing safety section added. [PCR 13-129]
Sec 18-5: Spare parts note added. [PCR 13-129]
Sec 18-5-2: Warning updated. [PCR 13-129]
Sec 18-5-4: Warning updated. [PCR 13-129]
New appendix A: New Safety Risk Assessment appendix added. [PCR 13-129]
10 October 19, Sec 2-8: Test of safety interlocks and warning lights added. [PCR 09-231]
2010 Sec 12-2-4-6: Test panel updated with TPSU door interlock.
Ch 17: Extensive tests of the safety interlock system added (radiation safety, RFPG interlocks
and warning lights). [PCR 09-219, 09-231, 09-232]
Revision history 5
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
6 December 8, Ch 2:
2008 Sec 2-8-16: Table for SCU web interface values added.
Sec 2-8-17: He cooling pressure changed. [PCR 08-196]
Ch 10:
Sec 10-9-6, step 49: Note updated.
Ch 12:
Sec 12-2-3: Reference changed. [PCR 08-243]
Sec 12-2-4-6: Table 12-9 and 12-11 updated.
Sec 12-2-4-6, 12-6-10-3: Log files information added.
Sec 12-2-4-7–8: Tube controller changed to software. “Flaps”, “delta”, “tuning motors”
corrected.
Sec 12-6-9-4: Updated with MINItrace data.
Sec 12-6-9, 12-6-10: Screen voltage to 750 V. [PCR 08-243]
Sec 12-6-10: Battery recycling instruction updated. [PCR 08-191]
Sec 12-6-10-7: Updated with MINItrace data.
Sec 12-6-11: Information on low-pass filters added. [PCR 08-243];
RF tuning instruction updated. [PCR 08-243];
SCU/RFPG tuning: “at 35 kV” added;
Table 12-24 updated.
6 Revision history
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
5 September 2, General
2008 Spare Part Booklet renamed to Spare Parts Catalog (same dir. no.).
New layout.
Ch 2:
Sec 2-5-2: Picture caa000-38030-00 replaced.
Sec 2-5-6: LOTO procedures for MINItrace added (SNO 5145415-100).
Sec 2-7: Clarification made: The customer is responsible for providing all consumables and
parts needed for planned maintenance, unless otherwise stipulated in contract agreement.
Sec 2-8-14: Table updated, grounding (SNO 5136522-100).
Sec 2-8-15: RF Board table updated (SNO 5136522-100).
Sec 2-8-16: PM instruction for SCU Gen II added.
Ch 12:
Sec 12-2: Restructured section.
Sec 12-2-2: Introduction to SCU Gen I and Gen II.
Sec 12-2-3: Information on SCU Gen I.
Sec 12-2-4: Information on SCU Gen II added.
Sec 12-6-9: Board replacement in SCU Gen I added.
Sec 12-6-10: Instructions for replacing spare parts in SCU Gen II.
Sec 12-6-11: Instructions for replacing SCU Gen I with SCU Gen II.
Ch 14:
Sec 14-5, 14-6: Target maintenance and assembly removed.
Revision history 7
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
3 June 14, 2005 General: The words “FRU” and “Renewal Part” changed to “Spare Part”.
Minor layout improvements.
Page 16: Internet link to CDL (GE Common Documentation Library) introduced [SPR 04–202].
Ch 2:
Page 22: (Section 2-3-2) LOTO Spare Parts (three kits) added to Table 2 “On site tools not
included in product”.
Page 29–56: (Section 2-5) New LOTO Spare Parts (three kits) section is introduced, describing
three LOTO kits [SPR 03–196] (main reason for manual upgrade!).
Ch 9:
Page 190: (Section 9-6-1) The “Reprogramming the VCU” instructions corrected (lines 25–30
added).
Ch 12:
Page 302: (Section 12-6-8) Resistor color coding corrected in Illustration 85 “New compo-
nents on RF Circuit Board, Spare Part 2378410”.
8 Revision history
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Revision history 9
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
0E February 8, Radiation Safety information moved to direction 2232998 MINItrace Service Manual –
2001 Health and Safety.
Ch 2 Preventive Maintenance: Updated.
Ch 3 Renewal parts: Illustrations updated.
Ch 8 Cooling Water System: Updated.
Ch 10 Ion Source: Updated.
Ch 13 Extraction System: Updated.
Ch 14 Targets and Radiochemistry System: Updated.
10 Revision history
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Revision history 11
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Manuals are also available for viewing and downloading via Internet.
Please regularly check the GE Common Documentation Library (CDL) website for updates:
1 Click Radiopharmacy.
2 Select product.
4 Click Search.
URL:
http://apps.gehealthcare.com/servlet/ClientServlet?REQ=Enter%2bDocumentation%2bLibrary
TABLE OF CONTENTS
1 INTRODUCTION ........................................................................................................ 25
1-1 Contents of this manual ................................................................................................................ 25
1-2 Associated documents .................................................................................................................. 25
1-3 Drawings ............................................................................................................................................... 25
1-4 Typographical conventions ......................................................................................................... 26
Table of Contents 15
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
16 Table of Contents
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Table of Contents 17
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
18 Table of Contents
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Table of Contents 19
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
20 Table of Contents
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Table of Contents 21
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
22 Table of Contents
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Table of Contents 23
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
24 Table of Contents
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
1 INTRODUCTION
• MINItrace Qilin
• a diagnostic guide.
Note!
The illustrated spare parts list is found in MINItrace Spare Parts Catalog (dir. 2380006-100).
This manual is intended for use by trained service engineers who are required to perform
maintenance on the MINItrace cyclotron system.
1-3 Drawings
For Drawings and Parts Lists, refer to MINItrace Service Manual – Drawings and Parts Lists
(dir. 2233001-100).
1 Introduction 25
1-3 Drawings
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
2 Software menus and buttons are presented in bold text throughout. Thus: Maintenance
menu.
3 The names of operational modes are presented in italic text throughout. Thus: OFF,
EMPTY, Hot, Cold
26 1 Introduction
1-4 Typographical conventions
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
2 SAFETY INFORMATION
• Only personnel with appropriate training and experience should service this system.
• Do not attempt to service the system unless all applicable manuals have been consulted
and are understood.
• The system should be serviced in strict compliance with the safety precautions, warnings,
and instructions contained in this manual, and in any other documentation specific to
the system. Otherwise, personal injury might occur due to electric shock, mechanical or
other hazards.
• Do not attempt to service the system unless relevant safety measures (e.g. Lock-Out
and Tag-Out) are implemented to block any energy sources that could be potential safety
hazards.
• The manufacturer or vendor of this system makes no representation that the act of
reading this manual renders the reader qualified to service the system.
2 Safety information 27
2-1 Risk Assessment (RA)
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
3 PERIODIC MAINTENANCE
CAUTION! Hazards
Read the health and safety information in MINItrace Qilin Operator Guide (dir. 2232995)
and MINItrace Service Manual – Health and Safety (dir. 2232998-100) (especially the Job
Safety Analysis (JSA)) before performing any maintenance on the MINItrace system.
Note!
Do not attempt to lift more than 18.2 kg (40 lbs) by yourself.
Note!
Your system might not include all the subsystems mentioned in this chapter.
• Do not attempt to service the system unless all applicable manuals have been consulted
and are understood.
• The system should be serviced in strict compliance with the safety precautions, warnings,
and instructions contained in this manual, and in any other documentation specific to
the system. Otherwise, personal injury might occur due to electric shock, mechanical or
other hazards.
• Do not attempt to service the system unless relevant safety measures (e.g. Lock-Out
and Tag-Out) are implemented to block any energy sources that could be potential safety
hazards.
3 Periodic maintenance 29
3-2 Safety information
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
• The manufacturer or vendor of this system makes no representation that the act of
reading this manual renders the reader qualified to service the system.
• a list of suitable on-site tools (see Section 3-4 On-site tools on page 31).
• comprehensive information on three LOTO kits and LOTO procedures for the MINItrace
system (see Section 3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
on page 35).
• a sequence list (with empty tables for notes, measurement values, etc.) to facilitate the
documentation of the planned maintenance (see Section 3-9 Sequence list for MINItrace
Planned Maintenance on page 85).
Note!
Copy pages 85–124 to get a complete and comprehensive documentation of the planned
maintenance.
30 3 Periodic maintenance
3-3 Contents of this chapter
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Hardware
Target dump
Collimator tool
3 Periodic maintenance 31
3-4 On-site tools
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Hardware
Lead glass
10:
Lead container for target foils
13:
Bench Radiation Shield (BRS)
32 3 Periodic maintenance
3-4 On-site tools
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Hardware
3 Periodic maintenance 33
3-4 On-site tools
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Hardware
Extension board, RF
Software
CD: MINItrace Service System version x.x Floppy disks: MINItrace system installa-
tion floppy disks
CD: Master System Application Software CD: Master System Application Software
release PC Lynx x.x release x.x
OR
DVD: SUSE Linux version x.x Hard drive: Solaris version x.x
34 3 Periodic maintenance
3-4 On-site tools
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Description
Service laptop
Full European Electrical Loto Kit See Section 3-5-1 Full European Electrical
LOTO kit (spare part 2334312) content on
page 36.
Set of 10 LOTO tags, English See Section 3-5-3 Set of 10 LOTO tags, En-
glish (spare part 2355715) on page 53.
Radiopharmacy valve LOTO kit See Section 3-5-4 Radiopharmacy valve LO-
TO kit (spare part 2398878) content on
page 53.
For spare part numbers, refer to the spare parts listed for each subsystems in MINItrace Spare
Parts Catalog (dir. 2380006-100).
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
LOTO kits are orderable through the following numbers:
• Full European Electrical LOTO kit, can be used World Wide (W.W.): spare part number
2334312.
Tags are to be ordered with Spare part number 2334312 (Full European Electrical LOTO
kit because the Electrical kit does not contain tags.
It is recommended that every Field Engineer has personal LOTO kits (one Electrical LOTO kit
and one valve LOTO kit).
3 Periodic maintenance 35
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
3-5-1 Full European Electrical LOTO kit (spare part 2334312) content
The Full European Electrical LOTO kit contains a number of items, shown in Figure 3-1 and
Table 3-3 below.
Table 3-3: Full european electrical LOTO kit content piece by piece
Lock -
36 3 Periodic maintenance
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Non-labeled Lock-Out -
Hasps.
3 Periodic maintenance 37
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Set of tags -
38 3 Periodic maintenance
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Bags -
3-5-2 Full European Electrical LOTO kit (spare part 2334312) user instructions
Instructions on how to use the Full European Electrical LOTO kit is provided below.
3 Periodic maintenance 39
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
40 3 Periodic maintenance
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
3 Periodic maintenance 41
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
42 3 Periodic maintenance
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
3 Periodic maintenance 43
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
44 3 Periodic maintenance
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
3 Periodic maintenance 45
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
46 3 Periodic maintenance
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
3 Periodic maintenance 47
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
48 3 Periodic maintenance
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
3 Periodic maintenance 49
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
50 3 Periodic maintenance
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
3 Periodic maintenance 51
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
52 3 Periodic maintenance
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
5 Tags TG027E
3 Periodic maintenance 53
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
1 - Padlock -
3-5-5 Radiopharmacy valve LOTO kit (spare part 2398878) user instructions
Instructions on how to use the Radiopharmacy valve LOTO kit is provided below.
54 3 Periodic maintenance
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
3 Periodic maintenance 55
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
56 3 Periodic maintenance
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
3-5-5-3 Instruction for Cablo-10 (Plastic coated steel wire locking device)
3 Periodic maintenance 57
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
58 3 Periodic maintenance
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
3-5-6-1 General
To be authorized to perform GE LOTO procedures you must:
This procedure sets the basic requirements established by the standard and the norms of GE
Healthcare for the energy-blocking (LOTO) and the isolation of energy required to realize any
sort of installation, de-installation, maintenance or repair of this equipment.
It is the obligation of each employee to follow these and other safety instructions regarding
the operation of this equipment or works close to these sites.
WARNING!
The following personnel are authorized to lock and tag the equipment, providing they have
received the LOTO authorized training and demonstrated knowledge annually:
• Support engineers
All non-authorized employees of GE Healthcare and the customer site will be considered as
affected employees.
The person who applies the LOTO device must communicate it verbally to all employees the
date at which the locking and tagging took place, as well as the date when the equipment
might start to work again.
3 Periodic maintenance 59
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
• Inform all the staff that the equipment is to be taken out of operation.
• Find out the energies of the equipment and the subsystem. See Table 3-5.
• Identify those points where these energies exist in the system and the magnitude
of them.
See Section 3-5-6-3 Cyclotron subsystems – energy types and locations on page 61.
5 Apply LOTO.
6 Free any remaining energy in the circuit/component/system that has been subject to
LOTO.
Electric E
Mechanic M
Pneumatic P
Gaseous G
Hydraulic H
Water W
60 3 Periodic maintenance
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Targets & Radiochem- Targets Gas Gas Supply Panel LOTO on Gas Supply Panel,
istry System Identification: G see Table 3-13
11C and 15O Gas Gas Supply Panel LOTO on Gas Supply Panel,
System Identification: G see Table 3-13
Helium Cooling Gas He-gas at Gas Sup- LOTO on Gas Supply Panel,
System Identification: G ply Panel see Table 3-13
Ion source & Central PSARC Electricity PSARC CB in CCAB LOTO on PSARC CB in
region Live hazardous volt- CCAB, see Table 3-11
ages up to 400 V AC
Identification: E
Ion source Gas Gas Supply Panel LOTO on Gas Supply Panel,
Identification: G see Table 3-13
Water Cooling Sys- CSC Electricity CSC CB in CCAB LOTO on CSC CB in CCAB,
tem Live hazardous volt- see Table 3-9
ages up to 400 V AC
Identification: E
3 Periodic maintenance 61
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Note!
The customer is responsible for the Mains Distribution Panel, the gas supply panel and the
compressed air panel. Therefore, these units can be designed in several different ways. In some
cases, a suitable LOTO kit needs to be created by the local GE FE together with the customer.
The pictures below only show examples of how LOTO devices are used.
62 3 Periodic maintenance
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
3 Periodic maintenance 63
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
64 3 Periodic maintenance
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
3 Periodic maintenance 65
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
First Name
Last Name
Function
Modality
66 3 Periodic maintenance
3-5 LOTO (Lock-Out and Tag-Out) kits, LOTO tags, LOTO instructions
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
• Personal whole body dosimeter (worn at collar or in a location where highest exposure
is likely to occur)
• Personal alarm dosimeter (worn in close proximity to the personal whole body dosimeter)
• Safety shoes
• Shoe covers
• Protective gloves
• Room ventilation
• Adequate lighting
• Calibrated contamination monitors (for checking feet, hands, clothing and objects for
potential contamination)
• Calibrated radiation survey meters (to conduct surveys upon entering bunker or radiation
shield of self-shielded cyclotron.
3 Periodic maintenance 67
3-6 Cyclotron job safety
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Example: Your electronic dosimeter displays 120 μSv/h, which means 2 μSv/min. If you spend
2 minutes in that particular area, your additional body dose will be 4 μSv.
• Risk levels:
Red: Do not proceed until all safety precautions have been completed.
• Hazard
• Cause
• PPE (appropriate personal protection equipment, i.e. personal film dosimeter, personal
electronic dosimeter, personal finger dosimeter, safety shoes, shoe covers, full body lab
coat, safety glasses or face shield, protective gloves, etc.)
• Control measures
Note!
The safety analysis in Table 3-16 is only a means of assistance, not an exhaustive analysis of
all possible situations.
68 3 Periodic maintenance
3-6 Cyclotron job safety
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
General advice
• Be informed and instructed on any customer emergency and site-specific safety
procedures.
• Service work should be delayed as long as practical from the last production run with
beam on to minimize dose rates in work areas.
• Wait at least two hours after the last production run with beam on before entering the
bunker.
• Planned maintenance should be requested from the customer to provide at least 24 hours
down time from the last production run with beam on prior to beginning service work
in the cyclotron vacuum chamber.
• For 18F- targets only: During the waiting time, do a dual filling/empty cycle with 16O water
and half an hour dry target. It eliminates any residual activity in the targets and delivery
lines.
• Radiation level surveys must be conducted when entering the bunker or radiation shield
of self-shielded cyclotron.
• Service work should not be performed when initial survey indicates dose rates in excess
of 2 mSv/hr at the position where work will be performed. If dose rates exceed 2 mSv/h
in the working area, work should be done by a senior engineer or escalated to a senior
engineer/local RSO for guidance prior to proceeding. Additional waiting time, removal
of components (for example, target assemblies) by the customer, or other shielding
arrangement should be considered prior to work.
• Always have a second skilled person in the vicinity to keep track of your own actions.
A customer representative should be available within the building/work areas when
service is performed.
• Keep track of dose registered on personal dosimeter and record accumulated dose.
• After working in cyclotron areas, use contamination monitor to check feet, hands, clothing
and any objects removed prior to leaving restricted areas.
• All radioactive waste generated from service activities is the responsibility of the customer
for disposal in accordance with federal, state and/or local regulations.
3 Periodic maintenance 69
3-6 Cyclotron job safety
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Disconnect all targets us- OK to proceed None Refer to the Service System
ing the service laptop with task instructions. Ensure that
pressure in targets is com-
pletely released.
Measure radiation level in Do not proceed Radia- Ionization ef- See General advice above.
the room until all safety tion fect Always request approval
precautions from the customer radia-
have been tion safety officer.
completed
70 3 Periodic maintenance
3-6 Cyclotron job safety
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Switch off water valves to Proceed with Radia- Potential radia- The task will take 2 min to
targets on manifold caution tion tion in the cy- accomplish.
clotron room Do not start any job if the
dose rate is too high. Al-
ways request approval
from the customer radia-
tion safety officer.
Use the door pendant con- Proceed with Pinch Mechanical Take care when opening
trol to open the Integrated caution movement door. Ensure safety shoes
Radiation Shield are worn.
Measure radiation level in Do not proceed Radia- Ionization ef- Measure dose rate in tar-
targets area until all safety tion fect gets area.
precautions
have been
completed
Decide if the radiation level Do not proceed Radia- Ionization ef- Do not start any job if the
is low enough to start work until all safety tion fect dose rate is too high.
precautions Consider:
have been - the present radiation
completed - the time required to com-
plete task
- your personal accumulat-
ed dose
3 Periodic maintenance 71
3-6 Cyclotron job safety
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Remove each mounted Do not proceed Radia- Ionization ef- Time: Minimize any person-
target until all safety tion fect al contact with the targets.
precautions Distance: Always verify
Note! have been head and body position are
The task of removal the completed as far as possible away
targets should normally be from unshielded targets.
performed by the cus- Shielding: Store the targets
tomer’s trained on site engi- immediately in an appropri-
neer as this is an operator ately shielded area/contain-
procedure. er.
Action:
Apply lead shield in front of
targets.
Action:
Remove all connections
(beam current probe, heli-
um, gas and media, cooling
water).
Note: High finger dose ex-
posure!
Action:
Remove target, put it in the
BRS.
Open the vacuum chamber Proceed with Pinch Mechanical Take care when opening
door caution movement door. Make sure that safety
shoes are worn.
72 3 Periodic maintenance
3-6 Cyclotron job safety
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Do not proceed Radia- Target foil parti- Check for presence of tar-
until all safety tion cles get foil particles in the
precautions chamber.
have been
completed Possible radiation sce-
nario:
Question: How do you find
out if (by any chance) tar-
get foil particles (frag-
ments) are present in the
chamber?
Answer: Target foil frag-
ments will be revealed by
high radiation levels at
certain spots. Radiation
levels above 1 mSv/h
(1000 μSv/h) 10 cm in front
of your electronic dosime-
ter indicates target foil
fragments. Use long
tweezers to manipulate
target foils. Move target
foils to lead container.
Check magnet door and Do not proceed Radia- Ionization ef- Continue monitoring radia-
compressed air cylinder until all safety tion fect tion level.
precautions
have been
completed
Unmount and service ion Do not proceed Radia- Ionization ef- Always monitor radiation
source until all safety tion fect levels before removal and
precautions ensure PPE requirements
have been are rigidly observed.
completed
Removing the ion source
will take ~5 mins.
Ion source service will take
~30 min, but with a lower
radiation level (it is done
outside radioactive area).
3 Periodic maintenance 73
3-6 Cyclotron job safety
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Remount ion source Do not proceed Radia- Ionization ef- Shielding, distance: Al-
until all safety tion fect ways verify your head and
precautions body position are as far as
have been possible away from collima-
completed tors.
Check and correct dis- Do not proceed Radia- Ionization ef- Shielding, distance: Al-
tances in central region until all safety tion fect ways verify your head and
precautions body position are as far as
have been possible away from collima-
completed tors.
Check and tighten stem Do not proceed Radia- Ionization ef- Shielding, distance: Al-
screws until all safety tion fect ways verify your head and
precautions body position are as far as
have been possible away from collima-
completed tors.
Check finger contacts and Do not proceed Radia- Ionization ef- Shielding, distance: Al-
feed-through until all safety tion fect ways verify your head and
precautions body position are as far as
have been possible away from collima-
completed tors.
Remove extraction holders Do not proceed Radia- Ionization ef- Shielding, distance: Al-
until all safety tion fect ways verify your head and
precautions body position are as far as
have been possible away from collima-
completed tors.
74 3 Periodic maintenance
3-6 Cyclotron job safety
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Check extraction foils and Proceed with Radia- Ionization ef- Ensure PPE requirements
replace bad ones caution tion fect are rigidly observed.
Check and calibrate the Proceed with Radia- Ionization ef- Shielding, distance: Al-
extraction system caution tion fect ways verify your head and
body position are as far as
possible away from collima-
tors.
Check collimators Do not proceed Radia- Ionization ef- Time: Work fast to mini-
until all safety tion fect mize exposure.
precautions
have been Checking and measuring
completed collimator openings will
take ~5 minutes.
Measure beam current Do not proceed Radia- Ionization ef- Time, distance: Measure
readout in targets, collima- until all safety tion fect fast and use long test
tors, extraction and flip-in precautions probes.
probe have been
completed Measure beam resistance
readout on flip-in probe
and extraction revolvers
will take ~5 min.
3 Periodic maintenance 75
3-6 Cyclotron job safety
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Clean vacuum chamber Proceed with Radia- Ionization ef- Remove all tools from the
and O-ring caution tion fect vacuum tank. Check that
the area is tidy. Use a vacu-
um cleaner, if necessary.
Close vacuum chamber Proceed with Mechani- Tools left inside Check and verify carefully
caution cal can damage that nothing is left inside
system the chamber.
Perform service of diffusion Proceed with Thermal, Harmful fumes Mandatory 2 hour delay to
pump caution chemical if hot oil ex- ensure sufficient cooling
posed to ambi- time.
ent air
Perform service of rough- Proceed with Thermal Roughing Mandatory 2 hour delay to
ing pump caution pump can ensure sufficient cooling
reach up to time.
70–80°C
Empty the compressed air Proceed with Com- Ensure compressed air
moisture trap caution pressed supply is locked out before
air removing water trap.
76 3 Periodic maintenance
3-6 Cyclotron job safety
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Clean targets Do not proceed Radia- Ionization ef- Time: Minimize any person-
until all safety tion fect al contact with the targets.
Note! precautions Distance: Always verify
The task of target mainte- have been head and body position are
nance should normally be completed as far as possible away
performed by the cus- from unshielded targets.
tomer’s trained on site engi- Shielding: Use bench radi-
neer as this is an operator ation shield and long
procedure. tweezers to manipulate
target foils.
Note!
When practical, use a spare
target assembly which has
decayed to minimize dose
during repair and mainte-
nance.
3 Periodic maintenance 77
3-6 Cyclotron job safety
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Replace the helium cooling Proceed with Radia- He pipes con- Measure the radiation level
compressor membrane caution tion tamination before start working.
Clean helium cooling filter Proceed with Radia- Filter contami- As above, verify filter con-
caution tion nation tamination level. See
above.
Check for water and heli- Proceed with Radia- Ionization ef- Verify radiation level in the
um leaks caution tion fect working area. See above.
78 3 Periodic maintenance
3-6 Cyclotron job safety
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Remount targets Do not proceed Radia- Ionization ef- Time: Minimize any person-
until all safety tion fect al contact with the targets.
Note! precautions Distance: Always verify
The task of remounting the have been head and body position are
targets should normally be completed as far as possible away
performed by the cus- from unshielded targets.
tomer’s trained on site engi- Verify target radiation level
neer as this is an operator and handle accordingly.
procedure.
Note!
These targets do not con-
tain radioactive target foils,
as they have not yet been
used in production.
Change 1/16" delivery lines Do not proceed Chemi- 18F– remain- Verify target radiation level
to hot cell until all safety cal, radia- ings in lines and handle accordingly.
precautions tion Manipulate carefully the
have been delivery lines. They could
completed contain rests of 18F– and
contaminate your clothes
or skin.
3 Periodic maintenance 79
3-6 Cyclotron job safety
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Check for water leaks in all Proceed with Electrical Accidental con- Use insulated tools and
system, verify conductivity, caution tact with live or good working procedures
pressure and flow charged com- to avoid electrical shock.
ponents LOTO: Always turn off and
Lock-Out the power supply
before servicing the sys-
tem. The secondary cooling
system (water pump,
PSMC, RFPG) contains
lethal voltages of up to the
mains voltage 400 VAC.
Check all voltages and Proceed with Electrical Accidental con- Use insulated tools and
currents caution tact with live or good working procedures
charged com- to avoid electrical shock –
ponents ground the TPSU before
touching the internal parts.
Check fans and fuses Proceed with Electrical Accidental con- Use insulated tools and
caution tact with live or follow working procedures
charged com- to avoid electrical shock.
ponents
80 3 Periodic maintenance
3-6 Cyclotron job safety
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Fill targets, turn on beam OK to proceed None When using the service
and check all path (flip-in with task laptop, always remember
probe, gas handling, ion to connect and select a
source parameters, extrac- target before running the
tion current, targets cur- beam. In fact, it is possible
rent) to run beam from the ser-
vice system without con-
necting and selecting the
target, but this will damage
the target.
3-6-7 RAs/JSAs
For a comprehensive list of RAs, see Appendix A Risk Assessment (RA).
3-8 Consumables
The customer is responsible for providing all consumables and parts needed for planned
maintenance, unless otherwise stipulated in contract agreement.
3 Periodic maintenance 81
3-8 Consumables
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
To obtain a list of consumables, contact the local GE Healthcare service representative and
refer to DOC1731876 (internal GE document).
Qty Description
Qty Description
1 Anode tube
2 Cathode
4 Anode insulator
8 Washer
2 Cathode screw
1 O-ring
82 3 Periodic maintenance
3-8 Consumables
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Qty Description
2 CFOIL graphite 24 x 69
1 Foil ass'y
Examination gloves
If applicable:
Qty Description
Qty Description
Qty Description
3 Periodic maintenance 83
3-8 Consumables
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Qty Description
84 3 Periodic maintenance
3-8 Consumables
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
PM
(Planned Maintenance)
CHECKLIST
for MINItrace systems
Planned maintenance is to be performed by qualified service personnel only!
CUSTOMER:
...........................................
DATE OF VISIT:
...........................................
PERFORMED BY:
...........................................
(FE signature)
3 Periodic maintenance 85
3-9 Sequence list for MINItrace Planned Maintenance
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
F-1 Working on parts on top of the radiation shield or standing on a ladder involves potential
General Panel Operation
fall hazard.
F-1 General Panel Operation
1. Turn on the Water Cooling Unit power with the front panel switch.
Use secured ladder
1. Turnandon safety harness.
the Water Cooling Unit power with the front panel switch.
2. See Illustration F-1.Press the Main menu button to display the following list of selec-
tions:
2. See Illustration F-1.Press the Main menu button to display the following list of selec-
For instructions on using ladder and safety harness, refer to Section 17-7 Ladder and climbing
tions:
TABLE safety
F-1 on page 564.
TABLE F-1
MODE SELECTtargets, open magnet
3-9-2 Shutdown, stop vacuum, disconnect/remove Mode Select
door Menu
MODE SELECT Mode Select Menu
RUNNING CONDITIONS Display Running Conditions
# Location System Action
RUNNING CONDITIONS Display Running Conditions
PUMPS CONTROL Pump Control Menu
PUMPS CONTROL Pump Control Menu
1 Control Master PRESSURE
Shutdown the MINItrace andDisplay Pressure
logout from Value
the Master.
PRESSURE Display Pressure Value
2 CSC Vacuum CONDUCTIVITY
Record the vacuum pressureDisplay Conductivity
before PM start. Value
CONDUCTIVITY
Refer to MINItrace Qilin Display
Operator Conductivity
Guide Value
(dir. 2232995).
TEMPERATURE Display Temperature Value
See
TEMPERATURE Chapter 10 Vacuum system for details.
Display Temperature Value
PARAMETERS SETTINGS Set Parameters
PARAMETERS
The A2SETTINGS Set Parameters
set point determines when the Penning gauge
DATE/TIME SETTING Set Date and Time
B1 is activated.
DATE/TIME SETTING
Make Set and
sure that the A2 low Datehigh
andset
Timepoints are:
3. Press
6.0 x 10the (low)
-3 or Cursor key to select an item from the main menu.
3. Press 8.0thex 10or-3 Cursor key to select an item from the main menu.
(high)
4. Press the Enter key to display the corresponding item’s menu or values.
4. Press the Enter key to display the corresponding item’s menu or values.
• If the menu screen has a + in the bottom right corner, information exists below
• If the the
menu screen.
screen has a + in the bottom right corner, information exists below
the screen.
• If the menu screen has a + in the upper right corner, information exists above
86 3 Periodic maintenance • If the the
menu screen.
screen has a + in the upper right corner, information exists above
the screen.
3-9 Sequence list for MINItrace Planned Maintenance
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Date: Signature:
5 Service laptop Targets Disconnect all targets by using the Service laptop and
the MINItrace Service System software as follows:
– Select a target menu in the MSS, for example 18F- Tar-
get.
– Click DISCONNECT TARGET.
6 CSC Cooling water Switch off water valves to the targets on Water Manifold.
7 CSC Radiation shield Use the door pendant control (hand hold device located
in the CSC) to open the radiation shield. See Figure 10-5.
WARNING! Pinch hazard
3 Periodic maintenance 87
3-9 Sequence list for MINItrace Planned Maintenance
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Note!
The target assembly, upon removal, must be immediately
stored in an appropriately shielded area/container.
Magnet door
Date: Signature:
88 3 Periodic maintenance
3-9 Sequence list for MINItrace Planned Maintenance
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Table 3-18: Values valid on MINItrace systems up to and including S/N 28790
N 1.0 [mm]
Date: Signature:
3 Periodic maintenance 89
3-9 Sequence list for MINItrace Planned Maintenance
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Table 3-19: Values valid on MINItrace systems from S/N 28791 and forward
L N/A N/A
M 30.6 mm [mm]
N N/A N/A
Date: Signature:
90 3 Periodic maintenance
3-9 Sequence list for MINItrace Planned Maintenance
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
3-9-4 RF
RF stem screws OK
RF finger contacts OK
Date: Signature:
3 Periodic maintenance 91
3-9 Sequence list for MINItrace Planned Maintenance
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
3-9-5 Extraction
16 Cyclotron Extraction Remove single extraction foil holder and (if applicable)
dual extraction foil holder.
WARNING! Radiation
Date: Signature:
92 3 Periodic maintenance
3-9 Sequence list for MINItrace Planned Maintenance
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Date: Signature:
Date: Signature:
3 Periodic maintenance 93
3-9 Sequence list for MINItrace Planned Maintenance
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Date: Signature:
Date: Signature:
94 3 Periodic maintenance
3-9 Sequence list for MINItrace Planned Maintenance
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Date: Signature:
31 No tools left!
3 Periodic maintenance 95
3-9 Sequence list for MINItrace Planned Maintenance
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
96 3 Periodic maintenance
3-9 Sequence list for MINItrace Planned Maintenance
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
[#36] PUMPS
Rotary pump
Diffusion pump
Pump compartment Motor compartment
Oil type
Oil level
Oil filled
Pump cleaned
Cooling water
Date: Signature:
Backing Pirani
Chamber Pirani
Penning gauge
Date: Signature:
Date: Signature:
3 Periodic maintenance 97
3-9 Sequence list for MINItrace Planned Maintenance
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
41 CSC Cooling water sys- Reinstall the targets and open the water
tem valves for the targets.
WARNING! Radioactive parts
98 3 Periodic maintenance
3-9 Sequence list for MINItrace Planned Maintenance
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
3 Periodic maintenance 99
3-9 Sequence list for MINItrace Planned Maintenance
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Membrane
Gasket
Compressor function
Date: Signature:
Cooling water
Date: Signature:
Target Water leaks Valve function Vacuum tight Media tight check
(OK if < 3 psi/minute)
Target 1
Target 2
Target 3
Target 4
Target 5
(dummy target)
Target 6
(if applicable)
Date: Signature:
[#57] pump activat- [#59] Time from pump activated [#59] High vacuum Remarks
ed at: to Green DP LED stops flashing and illuminates valve opens
steady (YES/NO)
Date: Signature:
A1 (Pirani A) x10-
A2 (Pirani B) x10- 1
B1 (Penning) x10-
Date: Signature:
61 CSC Cooling water sys- The Cooling Water System should already be
tem running, see #57.
BP1
BP3
Connections in CSC
PSMC
RFPG
Date: Signature:
RFPG
Magnet
RF1
RF2
Target dual
PSMC
Vacuum
Ion source
Date: Signature:
73 Open CCAB
77 Close CCAB
Note!
CCAB contains either a GCU Gen I or a GCU Gen II.
-15 V ±0.3 V
+15 V ±0.3 V
+24 V ±0.6 V
+5 V ±0.1 V
+9 V ±0.1 V
Date: Signature:
GND_IO CHASSIS ±5 V
Date: Signature:
Current
Voltage
Ripple
AC current R (L1)
S (L2)
T (L3)
AC voltage R (L1)
S (L2)
T (L3)
Date: Signature:
1 Check that the ground cable is securely connected to both the ground stick and the
chassis.
2 Check that the ground cable is in good condition and not damaged.
3 Use an Ohm-meter and measure between the tip of the ground stick and the chassis
(not the screw that holds the cable to the chassis). The resistance should be less than
1 Ω.
102 PS Room Service laptop Use the MSS Magnet System menu to start
the PSMC and set the Magnet Current to
135 A.
RF FORWARD VOLTAGE
RF REFLECTED VOLTAGE
FORWARD POWER
REFLECTED POWER
ANODE VOLTAGE
ANODE CURRENT
GRID VOLTAGE
GRID CURRENT
SCREEN VOLTAGE
SCREEN CURRENT
HEATER VOLTAGE
Date: Signature:
RF FORWARD VOLTAGE
RF REFLECTED VOLTAGE
FORWARD POWER
REFLECTED POWER
ANODE VOLTAGE
ANODE CURRENT
GRID VOLTAGE
GRID CURRENT
SCREEN VOLTAGE
SCREEN CURRENT
HEATER VOLTAGE
Date: Signature:
If specifically asked for (by for example OnLine Center personnel), please record all values
related to the RF system in the table on the opposite page.
Note!
This is not a regular Planned Maintenance issue! Disregard this if a complete set of values is
not specifically asked for!
For details on RF board TP (Test points), see Section13-5-1-2 Extended RFPG SCU Measurement
Module and RF board readout on page 444.
RF FORWARD VOLTAGE
RF REFLECTED VOLTAGE
FORWARD POWER
REFLECTED POWER
ANODE VOLTAGE
ANODE CURRENT
GRID VOLTAGE
GRID CURRENT
SCREEN VOLTAGE
SCREEN CURRENT
HEATER VOLTAGE
RF board
RF DET (TP24)
DVR (TP15)
FV (TP16)
AM2 (TP3)
AM1 (TP2)
AM3 (TP21)
DVR OK (TP1)
FP (TP17)
RV (TP18)
RP (TP19)
RF ENABLE (TP5)
PHASE (TP20)
+5 V (TP11)
-5 V (TP14)
+15 V (TP7)
+12 V (TP8)
-12 V (TP10)
-15 V (TP9)
GND (TP13)
+24 V (TP12)
Date: Signature:
SCU Gen II has a front display used for showing RF values. SCU Gen II can also be connected
to a service laptop where the values can be monitored in a web browser (SCU web interface).
It is also possible to create log files in the SCU web interface.
For more information on using the front display and the SCU web interface, see Section 13-2-4
SCU Gen II on page 390.
114 PS Room Service laptop, MSS Use the MSS Magnet System menu to start
the PSMC and set the Magnet Current to
135 A.
115 SCU Web interface, Fill in the OFF columns in both tables below.
MSS
117 SCU Web interface, Fill in the STANDBY columns in both tables
MSS below.
119 SCU Web interface, Fill in the NORMAL, 0 hour columns in both
MSS tables below.
Wait 1 hour
121 SCU Web interface, Fill in the NORMAL, 1 hour columns in both
MSS tables below.
124 SCU Web interface Download the Statistics log, the Five-seconds
log and the Milli-seconds logs.
125 Service laptop Save the log files in the backup folder in the
service laptop, C:\backup\scu.
126 Service laptop Empty the log folders. Log files cannot be
recorded if the memory is full.
Note!
Make sure the log files are backed up before
deleting them.
[#115, #117, #119, #121] RFPG VALUES IN THE SCU WEB INTERFACE AND MSS
0 hour 1 hour
DEE REF IN
DEE 1
MTCH PH 90
DPA FWD
DPA REFL
FWD PWR
REFL PWR
ANODE V
ANODE C
GRID V
GRID C
SCREEN V
SCREEN C
HEATER V
0 hour 1 hour
FORWARD POWER
REFLECTED POWER
PHASE ERROR
Date: Signature:
127 Gas Room Media supply Check gas cylinder pressure and
output pressure.
GAS CYLINDERS
He (5.5) 2.48 MPa (360 psi) ±0.1 MPa 18F-F- (operation gas)
(14 psi)
He (5.5) 0.5 ±0.2 MPa (73 ±29 psi) 18F-F- (carrier gas)
He (5.5) 0.16 ±0.01 MPa (23.5 ±1.5 psi) All (cooling gas)
Date: Signature:
200
400
600
800
1000
1200
1400
1600
1800
2000
180
160
140
120
100
80
60
40
20
0 200 400 600 800 1000 1200 1400 1600 1800 2000
Ion source current (IIS )[mA]
Flip-in probe OK
Target filled
Date: Signature:
134 Service laptop Safety interlock sys- Check the function of the radiation safety
Cyclotron tem system, RFPG interlocks, RF inhibit external
CCAB Warning lights interlocks, and warning lights.
For instructions, see Chapter 18 Safety sys-
tem.
Note!
Make sure that the terminal screws are tight-
ened thoroughly after the test.
RFPG interlocks OK
Warning lights OK
Date: Signature:
135 Service laptop Configuration Note record configuration values and make
a back-up.
4 SPARE PARTS
The spare parts are listed in MINItrace Spare Parts Catalog (dir. 2380006-100).
To order a spare part, quote the part number, the name and the required quantity. Address
the order to the nearest GE Healthcare service representative.
Procedures for disassembly and reassembly of the MINItrace system and spare parts are
detailed in other chapters in this volume.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
This column contains fault This column may contain pos- This column suggests ways of This column
messages shown on the Mas- sible causes to the fault in- solving the problem. might contain
ter System screen. curred. spare part data
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Database fault in error han- Database error (sqlca.sqlcode) See Informix manuals for ex-
dler, database access fault planation.
Database fault in error han- Database error (sqlca.sqlcode) See Informix manuals for ex-
dler, database fault planation.
Database fault in main menu, Database error (sqlca.sqlcode) See Informix manuals for ex-
database access fault planation.
Database fault in main menu, Database error (sqlca.sqlcode) See Informix manuals for ex-
database fault planation.
Database fault in network Database error (sqlca.sqlcode) See Informix manuals for ex-
sender, database access fault planation.
Database fault in network Database error (sqlca.sqlcode) See Informix manuals for ex-
sender, database access fault planation.
Database fault in parameter Database error (sqlca.sqlcode) See Informix manuals for ex-
editor, database access fault planation.
Database fault in parameter Database error (sqlca.sqlcode) See Informix manuals for ex-
editor, database fault planation.
Database fault in production, Database error (sqlca.sqlcode) See Informix manuals for ex-
database access fault planation.
Database fault in production, Database error (sqlca.sqlcode) See Informix manuals for ex-
database fault planation.
Database fault in show status, Database error (sqlca.sqlcode) See Informix manuals for ex-
database access fault planation.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Database fault in show status, Database error (sqlca.sqlcode) See Informix manuals for ex-
database fault planation.
Database fault in start unit, Database error (sqlca.sqlcode) See Informix manuals for ex-
database access fault planation.
Database fault in start unit, Database error (sqlca.sqlcode) See Informix manuals for ex-
database fault planation.
Database fault in supplies, Database error (sqlca.sqlcode) See Informix manuals for ex-
database access fault planation.
Database fault in supplies, Database error (sqlca.sqlcode) See Informix manuals for ex-
database fault planation.
Database fault in system sta- Database error (sqlca.sqlcode) See Informix manuals for ex-
tus, database access fault planation.
Database fault in system sta- Database error (sqlca.sqlcode) See Informix manuals for ex-
tus, database fault planation.
Fault at safety interlocks, I/O One of the VME I/O boards in Print errorlog and check num- I/O board
board is missing the GCU is not responding. ber for board 0=DIO_1,
1=DIO_2, 3=AI, 4=AO, 5=SMB
Then check select LED on I/O
board should be flashing or on
if working. Try to change
board.
Fault at safety interlocks, The door to the cyclotron vault Close the door.
safety door is open or the radiation shield is open.
Fault at startup, I/O board is One of the VME I/O boards in Print errorlog and check num- I/O board
missing the GCU is not responding. ber for board 0=DIO_1,
1=DIO_2, 3=AI, 4=AO, 5=SMB
Then check select LED on I/O
board should be flashing or on
if working. Try to change
board.
Fault during foilchange, out of All foils are used up. Time for service. Change to Foils
foils new foils.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Fault for balance, regulation The balance motor has not Check the motor. Motor-unit, DCM
fault reached position within board
120 seconds (old_mo-
tor_speed).
Fault in BCA system, I/O board One of the VME I/O boards in Print errorlog and check num- I/O boards
is missing the GCU is not responding. ber for board 0=DIO_1,
1=DIO_2, 3=AI, 4=AO, 5=SMB
Then check select LED on I/O
board should be flashing or on
if working. Try to change
board.
Fault in BCA system, probe Probe not responding for Check that the magnet is on.
fault 2 seconds.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Fault in BCA system, too high Target current too high, rela-
beam current on target 1 tive to maximum current con-
figured from the Service Sys-
tem. Average of 10 readings.
Fault in BCA system, too high Target current too high, rela-
beam current on target 2 tive to maximum current con-
figured from the Service Sys-
tem. Average of 10 readings.
Fault in BCA system, too high Target current too high, rela-
beam current on target 3 tive to maximum current con-
figured from the Service Sys-
tem. Average of 10 readings.
Fault in BCA system, too high Target current too high, rela-
beam current on target 4 tive to maximum current con-
figured from the Service Sys-
tem. Average of 10 readings.
Fault in BCA system, too high Target current too high, rela-
beam current on target 5 tive to maximum current con-
figured from the Service Sys-
tem. Average of 10 readings.
Fault in BCA system, too high Target current too high, rela-
beam current on target 6 tive to maximum current con-
figured from the Service Sys-
tem. Average of 10 readings.
Fault in beam checking, beam The difference between the foil Check vacuum and foils. Foils
losses too high from foil to current and target current
target must be less than the configu-
ration value “Loss foil/tar-
get[%]”.
Fault in beam checking, beam The difference between the Check vacuum and foils. Foils
losses too high from probe to probe current and foil current
foil must be less than the configu-
ration value “Loss
probe/foil[%]”.
Fault in beam starter, system Buttons are pressed too often Take it easy.
is still busy at the Service System.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Fault in cooling system, I/O One of the VME I/O boards in Print errorlog and check num- I/O board
board is missing the GCU is not responding. ber for board 0=DIO_1,
1=DIO_2, 3=AI, 4=AO, 5=SMB
Then check select LED on I/O
board should be flashing or on
if working. Try to change
board.
Fault in Ion source, gas flow Flow is outside set flow ±20%, Try again. If still problems, Flowreg
fault or unstable for 20 seconds. check gas supply. Next check
flow regulator.
Fault in Ion source, gas off Illegal to set ion source current Turn on the gas before setting
fault when the gas is switched off. current.
Fault in Ion source, I/O board One of the VME I/O boards in Print errorlog and check num- I/O board
is missing the GCU is not responding. ber for board 0=DIO_1,
1=DIO_2, 3=AI, 4=AO, 5=SMB
Then check select LED on I/O
board should be flashing or on
if working. Try to change
board.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Fault in Ion source, in bad Illegal to set ion source current Turn on the ion source before
state when the ion source is setting current.
switched off.
Fault in Ion source, magnet Magnet is off or current Check the PSMC.
current too low < 100 A.
Fault in Ion source, PSARC off Illegal to set ion source current Turn on the ion source before
when the ion source is setting current.
switched off.
Fault in Ion source, set ARC It is not possible to set the arc Use values within the config-
current is outside limits current above the configured ured limits.
limit.
Fault in Ion source, vacuum Check the vacuum system Vacuum compo-
fault (VCU). nents
Fault in LTF system, I/O board One of the VME I/O boards in Print errorlog and check num- I/O board
is missing the GCU is not responding. ber for board 0=DIO_1,
1=DIO_2, 3=AI, 4=AO, 5=SMB
Then check select LED on I/O
board should be flashing or on
if working. Try to change
board.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Fault in magnet system, I/O One of the VME I/O boards in Print errorlog and check num- I/O board
board is missing the GCU is not responding. ber for board 0=DIO_1,
1=DIO_2, 3=AI, 4=AO, 5=SMB
Then check select LED on I/O
board should be flashing or on
if working. Try to change
board.
Fault in magnet system, PSMC Problems with the serial com- Check the cables, report to the
communication fault munication to the PSMC. service engineer.
Fault in motor system, balance Motor has not reached posi- Check the motor.
motor fault tion within 60 s.
Fault in motor system, busy Motor is already calibrating, Wait and try again.
calibrating wait.
Fault in motor system, calibra- Motor has not reached posi- Motor-unit, DCM
tion fault tion within 60 seconds. board
Fault in motor system, foil 1 Motor has not reached posi- Check the motor.
motor fault tion within 60 s.
Fault in motor system, foil 2 Motor has not reached posi- Check the motor.
motor fault tion within 60 s.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Fault in motor system, I/O One of the VME I/O boards in Print errorlog and check num- I/O board
board is missing the GCU is not responding. ber for board 0=DIO_1,
1=DIO_2, 3=AI, 4=AO, 5=SMB
Then check select LED on I/O
board should be flashing or on
if working. Try to change
board.
Fault in motor system, initial- Trying to use a non-existing Check the config parameters,
ization fault motor. The GCS configuration using the Service System.
is not complete.
Fault in motor system, inter- DCM board power error when Check LED for +15 and -15 on DCM board
lock fault running a motor. the DCM board.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Fault in RF system, cannot set Standby-ready/cavity-water- RFPG shuts down from stand-
normal mode flow/magnet-on/vacuum- by ready to off when normal
ok/door error, or Dee-volts-ok mode is started. Refer to
signal is not received from the “cannot set standby mode”.
RF within 100 seconds.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Fault in RF system, cannot set Waterflow is not OK Mains power for the RFPG is
standby mode (DI_RF_WFN), or standby- missing.
ready signal is not received
from the RF within 90 seconds.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Fault in RF system, Dee volt- DI_RF_ERRORGoing to stand- DEE VOLTS not OK within
age fault by. 60 secs.
Fault in RF system, I/O board One of the VME I/O boards in Print errorlog and check num- I/O board
is missing the GCU is not responding. ber for board 0=DIO_1,
1=DIO_2, 3=AI, 4=AO, 5=SMB
Then check select LED on I/O
board should be flashing or on
if working. Try to change
board.
Fault in RF system, initializa- The GCS configuration is not Check the config parameters,
tion fault complete. using the Service System.
Fault in RF system, safety door The accelerator room door or Close safety door.
is open radiation shield is open.
Fault in RF system, spark latch The RF system is signaling Reset the RFPG by switching
up SPARKS too long to the control off its mains power for a few
system. seconds. Run the RF system at
reduced Dee voltage in order
to condition vacuum and sta-
bilize RF action. If the system
persists in sparking search for
possible cause within the RF
cavity.
Fault in RF system, start fault Flaps have moved through No Dee voltage has been
whole search area without found. Equivalent to “Fault in
succeeding. RF system, Dee voltage fault”.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Fault in start beam for target The door to the cyclotron vault Close the door.
1, Safety door open or the radiation shield is open.
Fault in start beam for target The door to the cyclotron vault Close the door.
2, Safety door open or the radiation shield is open.
Fault in start beam for target The door to the cyclotron vault Close the door.
3, Safety door open or the radiation shield is open.
Fault in start beam for target The door to the cyclotron vault Close the door.
4, Safety door open or the radiation shield is open.
Fault in start beam for target The door to the cyclotron vault Close the door.
5, Safety door open or the radiation shield is open.
Fault in start beam for target The door to the cyclotron vault Close the door.
6, Safety door open or the radiation shield is open.
Fault in target 1, initialization The GCS configuration is not Check the config parameters,
fault complete. using the Service System.
Fault in target 2, initialization The GCS configuration is not Check the config parameters,
fault complete. using the Service System.
Fault in target 3, initialization The GCS configuration is not Check the config parameters,
fault complete. using the Service System.
Fault in target 4, initialization The GCS configuration is not Check the config parameters,
fault complete. using the Service System.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Fault in target 5, initialization The GCS configuration is not Check the config parameters,
fault complete. using the Service System.
Fault in target 6, initialization The GCS configuration is not Check the config parameters,
fault complete. using the Service System.
Fault in target system, He Helium compressor is not on. Check Circuit breaker in the CB
compressor fault CB in PDU off. PDU.
Fault in target system, He The He-pressure is not reach- Check He gasbottle pressure.
cooling pressure fault ing 18 psi within 80 seconds.
Fault in target system, He flow He flow not OK. Check if there is any flow in Flow-meter
1 fault the flowmeter.
Fault in target system, He flow He flow not OK. Check if there is any flow in Flow-meter
2 fault the flowmeter.
Fault in target system, I/O One of the VME I/O boards in Print errorlog and check num- I/O board
board is missing the GCU is not responding. ber for board 0=DIO_1,
1=DIO_2, 3=AI, 4=AO, 5=SMB
Then check select LED on I/O
board should be flashing or on
if working. Try to change
board.
Fault in target system, target Target waterflow too low. Check the flowswitch. if too Water flow
waterflow 1 fault low flow check that the valves switch
on the manifold is open. Check
that the coolig system is run-
ning.
Fault in target system, target Target waterflow too low. Check the flowswitch. if too Water flow
waterflow 2 fault low flow check that the valves switch
on the manifold is open. Check
that the coolig system is run-
ning.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Fault in vacuum system, I/O One of the VME I/O boards in Print errorlog and check num- I/O board
board is missing the GCU is not responding. ber for board 0=DIO_1,
1=DIO_2, 3=AI, 4=AO, 5=SMB
Then check select LED on I/O
board should be flashing or on
if working. Try to change
board.
Fault in vacuum system, mag- Magnet door open. Close the magnet door.
net door is open
Fault in vacuum system, vacu- Vacuum not OK. Check the vacuum system (on Vacuum compo-
um fault the VCU). nents
Internal communication fault Error when creating a commu- Print errorlog (for further inves-
in error handler, SW fault 11 nication socket. Check access tigations). Error from cre-
rights in the socket directory, ate_socket (errno).
should be “prw-rw-r-- 1 MINI-
trac MINItrac”.
Internal communication fault Error when creating a commu- Print errorlog (for further inves-
in network GCS receiver, SW nication socket. Error from tigations). Check access rights
fault 11 create_socket (errno). in the socket directory, should
be “prw-rw-r-- 1 MINItrac
MINItrac”.
Internal communication fault Error when creating a commu- Print errorlog (for further inves-
in network GCS receiver, SW nication socket. Error from tigations). Check access rights
fault 11 create_socket (errno). in the socket directory, should
be “prw-rw-r-- 1 MINItrac
MINItrac”.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Internal communication fault Error when creating a commu- Print errorlog (for further inves-
in network receiver, SW nication socket. Error from tigations). Check access rights
fault 11 create_socket (errno). in the socket directory, should
be “prw-rw-r-- 1 MINItrac
MINItrac”.
Internal communication fault Error when creating a commu- Print errorlog (for further inves-
in parameter editor, SW nication socket. Check access tigations). Error from cre-
fault 11 rights in the socket directory, ate_socket (errno).
should be “prw-rw-r-- 1 MINI-
trac MINItrac”.
Internal communication fault Error when creating a commu- Print errorlog (for further inves-
in show status, SW fault 11 nication socket. Error from tigations). Check access rights
create_socket (errno). in the socket directory, should
be “prw-rw-r-- 1 MINItrac
MINItrac”.
Internal communication fault Error when creating a commu- Print errorlog (for further inves-
in start unit, SW fault 11 nication socket. Error from tigations). Check access rights
create_socket (errno). in the socket directory, should
be “prw-rw-r-- 1 MINItrac
MINItrac”.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Master internal fault in main Could not start a subtask (er- Report to service engineer or
menu, cannot execute pro- rno). system admnistrator and try
gram again.
Master internal fault in main File report/report.txt fault (sta- Check the directory /ex-
menu, report file could not be tus from xv_set, or 0). port/home/MINItrac/report.
read
Master internal fault in mainte- The target or chemistry sys- Check the status of the target
nance, target or chemistry is tem is disconnected. or chemistry system. If OK,
disconnected connect using the Master
maintenance function.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Master internal fault in produc- Not possible to start a subtask Report to service engineer or
tion, cannot execute program (0). system admnistrator and try
again.
Master internal fault in produc- Not possible to start a second Report to service engineer or
tion, SW fault 4 program. (0) system admnistrator and try
again.
Master internal fault in start Timeout on s_alive or _install Check the network.
unit, no Ethernet response (unit).
Master internal fault in start Wrong version number on GCS The wrong software version is
unit, wrong program version (unit). installed, check with system
administrator.
Master internal fault in start Target configuration Check the Master and GCS
unit, wrong target type not matching between Master configuration, they must
and GCS (unit). match.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Network communication fault GCSport is not defined (unit). Check file /etc/services, which
in network receiver, etherport should contain definition of
environment undefined GCSport.
Network communication fault Error from create_inet_socket, Print errorlog and report to
in network receiver, network connect (errno). service engineer.
connection problem
Network communication fault Error from create_inet_socket, Print errorlog and report to
in network receiver, SW setsockopt (errno). service engineer.
fault 13
Network communication fault Error when creating a network Print errorlog (for further inves-
in network receiver, SW fault 9 communication socket. Error tigations). Check access rights
from create_inet_socket (er- in the socket directory, should
rno). be “prw-rw-r-- 1 MINItrac
MINItrac”.
Network communication fault GCSport is not defined (unit). Check file /etc/services, which
in network GCS receiver, should contain definition of
etherport environment unde- GCSport.
fined
Network communication fault Not possible to find the ad- Print errorlog (for further inves-
in network GCS receiver, net- dress located in the TFTP tigations). Change the address
work connection problem bootstrap PROM. Error from in the NVRAM with a CONSOLE
create_inet_socket, connect or change the TFTP PROM.
(errno).
Network communication fault Error from create_inet_socket, Print errorlog and report to
in network GCS receiver, SW setsockopt (errno). service engineer.
fault 13
Network communication fault Error when creating a network Print errorlog (for further inves-
in network GCS receiver, SW communication socket. Error tigations). Check access rights
fault 9 from create_inet_socket (er- in the socket directory, should
rno). be “prw-rw-r-- 1 MINItrac
MINItrac”.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Network communication fault The hostname that is request- Print the errorlog, contact the
in network GCS receiver, the ed does not exist (errno). system administrator, and
network host is not defined check the file /etc/hosts for
“gemsacs”.
Network communication fault GCSport is not defined (unit). Check file /etc/services, which
in network receiver, etherport should contain definition of
environment undefined GCSport.
Network communication fault Problems with the network Print errorlog (for further inves-
in network sender, communi- communication. Error from tigations) Check cables. Use
cation broken send_inet_sig, (errno). the “ping” command: “ping
gemsacs”, which should give
result “gemsacs is alive”.
Network communication fault GCSport or is not defined (unit). Check file /etc/services, which
in network sender, etherport should contain definition of
environment undefined GCSport.
Network communication fault Error when creating a network Print errorlog (for further inves-
in network sender, SW fault 9 communication socket. Error tigations). Check access rights
from create_inet_socket (er- in the socket directory, should
rno). be “prw-rw-r-- 1 MINItrac
MINItrac”.
Network communication fault The host name that is request- Print the errorlog, contact the
in network sender, the net- ed does not exist. (errno). system administrator, and
work host is not defined check the file /etc/hosts for
“gemsacs”.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Serious fault in cooling system, The secondary cooling system. Check if the pump is running Pump
cooling pump is not running on the water cooling system.
Serious fault in cooling system, Check the secondary cooling Check the secondary cooling
cooling system in local mode system, it should be set to re- system, it should be set to re-
mote mode. mote mode.
Serious fault in cooling system, Cooling water temperature Check water temperature on
cooling water temperature too too high. the water system. If to high
high check that the primary water
system is running.
Serious fault in Ethernet com- Master communication bro- Check power and cables.
munication, communication ken.
is broken
Warning at safety interlocks, External interlock number 1 is Check if the external interlock
safety 1 fault activated. is activated.
Warning at safety interlocks, External interlock number 2 is Check if the external interlock
safety 2 fault activated. is activated.
Warning at safety interlocks, External interlock number 3 is Check if the external interlock
safety 3 fault activated. is activated.
Warning at safety interlocks, External interlock number 4 is Check if the external interlock
safety 4 fault activated. is activated.
Warning at safety interlocks, External interlock number 5 is Check if the external interlock
safety 5 fault activated. is activated.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Warning for warning signs, I/O One of the VME I/O boards in Print errorlog and check num- I/O board
board is missing the GCU is not responding. ber for board 0=DIO_1,
1=DIO_2, 3=AI, 4=AO, 5=SMB
Then check select LED on I/O
board should be flashing or on
if working. Try to change
board.
Warning in BCA system, initial- The GCS configuration is not Check the config parameters,
ization fault complete. using the Service System.
Warning in BCA system, sys- Buttons are pressed too often Take it easy.
tem is still busy at the Service System.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Warning in cooling system, The GCS configuration is not Check the config parameters,
initialization fault complete. using the Service System.
Warning in cooling system, Cooling water pressure too Check pressure on the meter
low pressure low. on the cooling water system,
if necessary fill water, also
check for water leaks.
Warning in cooling system, Cooling water level too low. Check water level in cooling
low water level water system, if necessary fill
water, also check for water
leaks.
Warning in cooling system, not The PSMC is on and/or the Turn off the PSMC and/or
allowed to switch off PSarc is on and/or the RF is in PSARC. If necessary in local
standby. mode.
Warning in cooling system, Buttons are pressed too often Take it easy.
system is still busy at the Service System.
Warning in Ethernet communi- The GCS configuration is not Check the config parameters,
cation, initialization fault complete. using the Service System.
Warning in Ethernet communi- The GCS configuration is not Check the config parameters,
cation, initialization fault complete. using the Service System.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Warning in foil supervising, (Target+collimators) current is Check if foils are broken. Foils
extraction foil break less than foil current for con-
secutive measuring.
Warning in global system, ini- The GCS configuration is not Check the config parameters,
tialization fault complete. using the Service System.
Warning in global system, Buttons are pressed too often Take it easy.
system is still busy at the Service System.
Warning in Ion source, initial- The GCS configuration is not Check the config parameters,
ization fault complete. using the Service System.
Warning in Ion source, ion Cathode wearing > configured Time for service. cathode
source life time exceeded lifetime. Only warning, nothing
is stopped.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Warning in Ion source, system Buttons are pressed too often Take it easy.
is still busy at the Service System.
Warning in LTF system, step- The GCU contains one stepper Try again.
per motor is busy motor board. If a second
command is sent to the step-
per motor module while it is
busy, the first command will
be finished and the second will
be ignored.
Warning in magnet system, The GCS configuration is not Check the config parameters,
initialization fault complete. using the Service System.
Warning in magnet system, Problems with the serial com- Check the cables, report to the
PSMC communication fault munication to the PSMC. service engineer.
Warning in magnet system, Buttons are pressed too often Take it easy.
system is still busy at the Service System.
Warning in main menu, the The second connected server Check that the correct server
second server is not logged in is not logged in. name is defined in the databas
table parameters. Check that
the server is logged in with
OpenWindows™ started.
Check that the original Master
server is included in the xhost
list (enter command “xhost” to
see the list. If not included, en-
ter “xhost ‘servername’” where
‘servername’ is the name of
the original Master server
(petxxx)).
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Warning in maintenance The receiving unit is not ready. Verify the hotcell door, the
function, receiving unit is not FDG system status and maybe
ready other things as well.
Warning in motor system, Buttons are pressed too often Take it easy.
system is still busy at the Service System.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Warning in network signal The Master is trying to connect Exit and try again later.
distributor, Service System when the GCS is in local mode.
control
Warning in production, receiv- The receiving unit is not ready. Verify the hotcell door, the
ing unit is not ready FDG system status and maybe
other things as well.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Warning in service system GCS has read an unexpected Check cables, and the serial
communication, service sys- character. port parameters.
tem communication fault
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Warning in status handler, Buttons are pressed too often Take it easy.
system is still busy at the Service System.
Warning in target 1, LTF motor The difference between the Try again. If still problems
is not moving true- and calculated stepper check that the motor in the
motor position has twice ex- LTF is running.
ceeded 20 µl.
Warning in target 1, stepper A move command has been Wait and try again.
motor is busy sent to a motor that is already
moving. This command is ig-
nored, the previous command
continues.
Warning in target 1, target is It is not possible to use a dis- Check the target, and if OK
disconnected connected target. connect using the mainte-
nance function “Connect” at
the Master.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Warning in target 1, target Target pressure is outside lim- Check configuration parame-
pressure fault its, where limits depend on ters.
target and action and configu-
ration parameters.
Warning in target 2, LTF motor The difference between the Try again. If still problems
is not moving true- and calculated stepper check that the motor in the
motor position has twice ex- LTF is running.
ceeded 20 µl.
Warning in target 2, stepper A move command has been Wait and try again.
motor is busy sent to a motor that is already
moving. This command is ig-
nored, the previous command
continues.
Warning in target 2, target is It is not possible to use a dis- Check the target, and if OK
disconnected connected target. connect using the mainte-
nance function “Connect” at
the Master.
Warning in target 2, target Target pressure is outside lim- Check configuration parame-
pressure fault its, where limits depend on ters.
target and action and configu-
ration parameters.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Warning in target 3, LTF motor The difference between the Try again. If still problems
is not moving true- and calculated stepper check that the motor in the
motor position has twice ex- LTF is running.
ceeded 20 µl.
Warning in target 3, stepper A move command has been Wait and try again.
motor is busy sent to a motor that is already
moving. This command is ig-
nored, the previous command
continues.
Warning in target 3, target is It is not possible to use a dis- Check the target, and if OK
disconnected connected target. connect using the mainte-
nance function “Connect” at
the Master.
Warning in target 3, target Target pressure is outside lim- Check configuration parame-
pressure fault its, where limits depend on ters.
target and action and configu-
ration parameters.
Warning in target 4, LTF motor The difference between the Try again. If still problems
is not moving true- and calculated stepper check that the motor in the
motor position has twice ex- LTF is running.
ceeded 20 µl.
Warning in target 4, stepper A move command has been Wait and try again.
motor is busy sent to a motor that is already
moving. This command is ig-
nored, the previous command
continues.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Warning in target 4, target is It is not possible to use a dis- Check the target, and if OK
disconnected connected target. connect using the mainte-
nance function “Connect” at
the Master.
Warning in target 4, target Target pressure is outside lim- Check configuration parame-
pressure fault its, where limits depend on ters.
target and action and configu-
ration parameters.
Warning in target 5, LTF motor The difference between the Try again. If still problems
is not moving true- and calculated stepper check that the motor in the
motor position has twice ex- LTF is running.
ceeded 20 µl.
Warning in target 5, stepper A move command has been Wait and try again.
motor is busy sent to a motor that is already
moving. This command is ig-
nored, the previous command
continues.
Warning in target 5, target is It is not possible to use a dis- Check the target, and if OK
disconnected connected target. connect using the mainte-
nance function “Connect” at
the Master.
Warning in target 5, target Target pressure is outside lim- Check configuration parame-
pressure fault its, where limits depend on ters.
target and action and configu-
ration parameters.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Warning in target 6, LTF motor The difference between the Try again. If still problems
is not moving true- and calculated stepper check that the motor in the
motor position has twice ex- LTF is running.
ceeded 20 µl.
Warning in target 6, stepper A move command has been Wait and try again.
motor is busy sent to a motor that is already
moving. This command is ig-
nored, the previous command
continues.
Warning in target 6, target is It is not possible to use a dis- Check the target, and if OK
disconnected connected target. connect using the mainte-
nance function “Connect” at
the Master.
Warning in target 6, target Target pressure is outside lim- Check configuration parame-
pressure fault its, where limits depend on ters.
target and action and configu-
ration parameters.
Warning in target system, ini- The GCS configuration is not Check the config parameters,
tialization fault complete. using the Service System.
Error message Possible cause (parameter) Possible solution (Action) Related spare
parts
Warning in target system, too It is not possible to select more Do not select more than two
many selected targets than two targets. targets.
Warning in vacuum system, The GCS configuration is not Check the config parameters,
initialization fault complete. using the Service System.
Power monitor error One or several internal volt- Check the voltages on the test
ages in the control unit are points and via the CONSOLE
outside the specification port, and document the val-
(±10%). ues.
Turn the mains switch off/on.
If the Power monitor error re-
mains, the error needs more
investigation. Call service.
Ground fault error Short-circuit to chassis (tempo- Press the Ground fault reset
rary or permanent), for exam- button on the front panel. If
ple, in 24 V-driven parts the Ground fault error re-
(probes, sensors, etc.) mains, the error is permanent
and needs more investigation.
Call service.
6-1 Introduction
The cyclotron system is computer-controlled.
The control system comprises three primary units: the Master System, the General Control
Unit (GCU) and the MINItrace Service System (MSS).
• The Master System exercises operational control over the tracer production, runs and
provides the primary interface between the MINItrace system and the operator.
• The General Control Unit (GCU) provides the control interface between the Master
System and the accelerator.
• The MINItrace Service System (MSS) provides a dedicated maintenance terminal for
system service.
The Master System and the GCS communicate over an ethernet local area network, using
RJ45 cables. The MSS is connected to the GCS by a serial link.
• keeps databases, for example, for batch reports, media information, error messages,
network log.
The GCS
• contains no hard disk but loads the software from the Master system
• communicates with both the Master system (Ethernet) and the MSS (serial).
The MSS
• requires knowledge about the hardware and manual verification of all actions
See MINItrace Spare Parts Catalog (dir. 2380006-100), section MT207 CCAB, and Chapter 7
General control unit for more information on the General Control System.
CAUTION!
7-1 Introduction
The purpose of the General Control Unit (GCU) is to regulate and control the MINItrace cyclotron
and peripherals.
There are two different types of GCU, called Gen I and Gen II in the manual.
• GCU Gen I is based on a PowerPC™ CPU and VME bus. See Section 7-2 GCU – type Gen I
on page 165.
• GCU Gen II is based on an ETX CPU module and PCI bus. See Section 7-3 GCU – type
Gen II on page 202.
Note!
There are two GCU Gen II models:
When replacing a GCU with internal PS it must be upgraded to a GCU with an external PS.
Figure 7-2 shows the upper part of two CCAB equipped with the GCU Gen I and GCU Gen II.
Gen I Gen II
• serial ports
1 The VME bus follows the VME bus specification ANSI/IEEE STD1014-1987, IEC821 and IEC297.
The GCU is mounted in a 19" rack unit, which contains three fans, VME bus backplane, I/O
backplane, CPU, different I/O boards and two power supplies.
From the mains power inlet, the two power supplies generate all the power requirements for
the CPU–board and the I/O-boards. All signals between the I/O backplane and the VME
backplane are galvanically isolated with opto couplers for noise reduction.
The GCU is connected to the Master System via an Ethernet local area network. The GCS
(General Control System) software is downloaded from the Master System at start-up, and
anytime the RESET button is pressed.
Lock-out and tag-out the control cabinet circuit breaker at the MDP, before any service
is performed on the GCU.
• Analog In board, AI
WARNING!
Apply LOTO as applicable in accordance with GE regulations and local restrictions. For
LOTO procedures, see Section3-5-6 LOTO procedures for the MINItrace system on page59.
For Lock-Out and Tag-Out (LOTO) procedures of the equipment, see Section 3-5-6 LOTO
procedures for the MINItrace system on page 59.
For spare part numbers, refer to the spare parts listed for each subsystem in MINItrace Spare
Parts Catalog (dir. 2380006-100).
1 Turn off the GCU by switching OFF the GCU-CB on the back of the CB panel in the CAB.
The fans stop.
5 Mount the lower front panel and fasten the retaining screws.
7 The software starts to be downloaded from the Master via the Ethernet local area network
to the CPU board.
Figure 7-6: GCU backplane -15 V/+15 V/+24 V/+5 V measuring points
The connectors P3–P22 are connected via cables to the different sub-systems and are
receiving/sending control signals.
Connector Connected to
P4 PSMC
P6 Not used
P7 Not used
P8 Not used
Connector Connected to
P9 RFPG/SKG
P10 RFPG/SKH
P11 RSARC/P1
P13 CSC/P7
P14 CSC/P2
P15 CSC/P3
P16 PSMC
P17 CSC/P5
P18 CSC/P18
P19 TSC/P1
P20 TSC/P2
P2-6 AI board
P2-8 AO board
P2-12 CI board
1 Loosen the cables on the rear and loosen the screws on the front.
7-2-5-1 Introduction
The CPU board is a Motorola™ MVME2301, which is based on the PowerPC 603 32-bit
microprocessor. Please refer to Motorola's web-page for details on the CPU board.
• information about the CPU board front panel LEDs and switches.
Positions Description
J16 VMEbus System Controller Selection At system power-up or reset, the MVME2300SC deter-
mines whether it is system controller by its position on
the bus; or not.
J15 Flash Bank Selection The setting of J15 determines whether the board fetches
its reset vector from bank A or from bank B. Both banks
contain the on-board firmware,
Figure 7-12: Connectors LEDs and switches on CPU board front panel
Switches Description
RST Reset switch. Reset all GCU Devices and download software from
the Master.
LEDs Description
CPU Green LED indicates that signal line on processor bus is active.
Connectors Description
To reach the PLCC Flash Devices you need to remove the TPMC866 Serial Interface Module,
see Figure 7-13.
7-2-5-6 Configuration
Introduction
The CPU board must be configured as described below for use in the General Control Unit
(GCU).
The two Flash-memories XU1 and XU2 (on the MVME2301 board mounted in the MINItrace
General Control Unit, GCU) contain a boot strap program, used to download the application
program from a specified Unix node on the network (normally the MINItrace master
workstation).
The bootstrap program uses a standard protocol (TFTP, Trivial File Transfer Protocol) to onnect
to the master and download a HEX file.
TFTP allows users to transfer files to and from a remote machine. It uses a well known port
number defined in /etc/services to connect to the tftpd daemon process on the remote host.
(In the bootstrap program the port number is hardcoded to 69.)
To be able to connect to the Master and fetch the GCS application file the GCU CPU board
has to be set up with correct network addresses (own address + Master address) and path
to the object file. This information is stored in non-volatile RAM. When the bootstrap program
starts up the current parameter setting is sent to the serial port. If a carriage return is received
within 5 seconds an interactive setup procedure is entered. If not, the TFTP boot sequence
starts.
1 Connect the RJ45 4 pair category 5 cable to DEBUG-port on the CPU board, see Figure 7-3
[9600 baud, 8 data bits, 1 stop bit, no parity].
2 Reset the CPU. (The J15 jumper must have jumper closed at 2 and 3, see Table 7-2.) The
bootstrap program starts. The current parameter settings are listed at the console screen.
You should now get a menu with the following layout on the next page:
If any if these parameters need to be changed, press m. This will list a number of questions
to answer. The current setting of the parameter asked for is shown within braces, just press
<Enter> to select this value.
When all parameters are defined correctly and c is pressed, the GCS software will automatically
be loaded from the Master system at reset and power on.
The values shown for “virtual io” may vary between different versions of the GCS software.
When the text “Root suspended” is shown, the GCS software is running, and the Master may
be started.
For the GCS to properly connect to the Master system when running the application, the
IP-adress must be defined in the GCS configuration using the MSS software.
Optoisolated outputs
V C (+24 V) 24 V 10%
(T a = 50°C) (T a =50°C)
Optoisolated inputs
V C (+24 V) 24 V 10%
V IL at V C =24 V Max 6 V
Power requirements
+5 V supply (VCC)
Voltage 4.75–5.25 V
Voltage 21.6–26.4 V
Current (without output loads and all inputs connected Typ 400 mA Max 470 mA
to +24 V common)
The correct positions for DIO #1are listed in Table 7-2. The correct positions for DIO #2 are
listed in Table 7-6.
Table 7-5: DIO #1, switches and jumpers for the GCU
Positions Description
SWITCH 1 ON
SWITCH 2 ON
SWITCH 3 ON A10
SWITCH 4 ON A11
SWITCH 5 ON A12
SWITCH 6 ON A13
SWITCH 7 ON A14
SWITCH 8 ON A15
SW2 INTERRUPT REQUEST LEVEL SHUNT 1 THRU SHUNT 8 ARE USED TO SELECT ONE IN-
TERRUPT REQUEST LEVEL OR NONE. A CLOSED SHUNT
MEANS THAT THIS LEVEL IS SELECTED, ALL OTHER
SHUNTS MUST BE OPEN.
SW3 INTERRUPT PRIORITY LEVEL SWITCH 1 THRU SWITCH 3 SHOULD BE SET TO THE SAME
BINARY VALUE AS THE SELECTED INTERRUPT REQUEST
LEVEL. FOR THE DIO BOARD, THIS VALUE IS ZERO.
Positions Description
SW4 INTERRUPT VECTOR WORD SWITCH 1 THRU SWITCH 8 ARE USED TO SET A WORD
THAT DISTINGUISHES THE DIO BOARD FROM OTHER
BOARDS AT THE SAME IRQ LEVEL ON SIMULTANEOUS
INTERRUPT REQUESTS.
SWITCH 1 ON D00
SWITCH 2 ON D01
SWITCH 3 ON D02
SWITCH 4 ON D03
SWITCH 5 ON D04
SWITCH 6 ON D05
SWITCH 7 ON D06
SWITCH 8 ON D07
SW5 WATCH DOG ENABLE SHUNT 1–2 AND SHUNT 2–3 ARE USED TO SELECT
SOURCE OF RESET FOR THE DIO OUTPUT LATCHES.
Table 7-6: DIO #2, switches and jumpers for the GCU
Positions Description
SWITCH 1 ON
SWITCH 2 ON
SWITCH 3 OFF A10
SWITCH 4 ON A11
SWITCH 5 ON A12
SWITCH 6 ON A13
SWITCH 7 ON A14
SWITCH 8 ON A15
SW2 INTERRUPT REQUEST LEVEL SHUNT 1 THRU SHUNT 8 ARE USED TO SELECT ONE IN-
TERRUPT REQUEST LEVEL OR NONE. A CLOSED SHUNT
MEANS THAT THIS LEVEL IS SE-LECTED, ALL OTHER
SHUNTS MUST BE OPEN.
Positions Description
SW3 INTERRUPT PRIORITY LEVEL SWITCH 1 THRU SWITCH 3 SHOULD BE SET TO THE SAME
BINARY VALUE AS THE SELECTED INTERRUPT REQUEST
LEVEL. FOR THE DIO BOARD, THIS VALUE IS ZERO.
SW4 INTERRUPT VECTOR WORD SWITCH 1 THRU SWITCH 8 ARE USED TO SET A WORD
THAT DISTINGUISHES THE DIO BOARD FROM OTHER
BOARDS AT THE SAME IRQ LEVEL ON SIMULTANEOUS
INTERRUPT REQUESTS.
SWITCH 1 ON D00
SWITCH 2 ON D01
SWITCH 3 ON D02
SWITCH 4 ON D03
SWITCH 5 ON D04
SWITCH 6 ON D05
SWITCH 7 ON D06
SWITCH 8 ON D07
SW5 WATCH DOG ENABLE SHUNT 1–2 AND SHUNT 2–3 ARE USED TO SELECT
SOURCE OF RESET FOR THE DIO OUTPUT LATCHES.
All input and output ports are galvanically isolated (optoisolated) from the VME bus.
All +5 Volt supplies are connected via the P1 connector and all +24 Volt supplies are connected
via the P2 connector. Every byte-wide output or input port may have their own +24 Volt supply
and can thus be galvanically isolated from each other.
• Gas pressures
• Vacuum level
• Motor positions
• Beam extraction
• Phase load
7-2-7-1 Specifications
Table 7-7: Analog board specifications
Analog Inputs
Input impedance 1 MΩ
Power requirements
Max current 1 A
The correct positions for use in the GCU are listed in Table 7-8.
Positions Description
SW1 BASE ADDRESS SWITCH 1 THRU SWITCH 8 ARE USED TO DETERMINE THE BASE ADDRESS FOR THE
AI BOARD.
OFF= LOGIC 1, ON = LOGIC 0.
SWITCH 1 ON
SWITCH 2 ON
SWITCH 3 OFF A10
SWITCH 4 OFF A11
SWITCH 5 ON A12
SWITCH 6 ON A13
SWITCH 7 ON A14
SWITCH 8 ON A15
Positions Description
SW2 INTERRUPT REQUEST SHUNT 1 THRU SHUNT 8 ARE USED TO SELECT ONE INTERRUPT REQUEST LEVEL
LEVEL OR NONE. A CLOSED SHUNT MEANS THAT THIS LEVEL IS SELECTED, ALL OTHER
SHUNTS MUST BE OPEN.
SW3 INTERRUPT PRIORITY LEV- SWITCH 1 THRU SWITCH 3 SHOULD BE SET TO THE SAME BINARY VALUE AS THE
EL SELECTED INTERRUPT REQUEST LEVEL. FOR THE AI BOARD, THIS VALUE IS ZERO.
SW4 INTERRUPT VECTOR SWITCH 1 THRU SWITCH 8 ARE USED TO SET A WORD THAT DISTINGUISHES THE
WORD AI BOARD FROM OTHER BOARDS AT THE SAME IRQ LEVEL ON SIMULTANEOUS IN-
TERRUPT REQUESTS.
SWITCH 1 ON D00
SWITCH 2 ON D01
SWITCH 3 ON D02
SWITCH 4 ON D03
SWITCH 5 ON D04
SWITCH 6 ON D05
SWITCH 7 ON D06
SWITCH 8 ON D07
A differential amplifier (IC111) provides a single ended output to the A/D-converter (ADC, IC
112). The ADC is a standard 574-type with 12-bit resolution. The ADC has a +10 Volt reference
signal inside.
The ADC output is connected to the VME bus interface via three quad optocouplers.
A DC/DC converter (IC118) uses the +5 Volt supplied from the VME bus via the P1 connector
to produce +15 Volts and -15 Volts. Via a voltage regulator (IC119), +5 Volt is then derived
from the +15 Volt output of IC 118.
There is also an opto isolated “POWER OK” signal to the VME bus interface which senses if
the +15 Volt or -15 Volt on the board falls below +10 Volt or -10 Volt respectively
7-2-8-1 Specifications
Table 7-9: Analog Out board specifications
Analog Outputs
Number of outputs 8
Number of outputs 2
Number of outputs 2
Power requirements
Max current 1 A
The correct positions for use in the GCU are listed in Table 7-10.
Positions Description
SWITCH 1 ON
SWITCH 2 ON
SWITCH 3 OFF A10
SWITCH 4 ON A11
SWITCH 5 OFF A12
SWITCH 6 ON A13
SWITCH 7 ON A14
SWITCH 8 ON A15
SW2 INTERRUPT REQUEST LEVEL SHUNT 1 THRU SHUNT 8 ARE USED TO SELECT ONE IN-
TERRUPT REQUEST LEVEL OR NONE. A CLOSED SHUNT
MEANS THAT THIS LEVEL IS SELECTED, ALL OTHER
SHUNTS MUST BE OPEN.
SW3 INTERRUPT PRIORITY LEVEL SWITCH 1 THRU SWITCH 3 SHOULD BE SET TO THE SAME
BINARY VALUE AS THE SELECTED INTERRUPT REQUEST
LEVEL. FOR THE AI BOARD, THIS VALUE IS ZERO.
Positions Description
SW4 INTERRUPT VECTOR WORD SWITCH 1 THRU SWITCH 8 ARE USED TO SET A WORD
THAT DISTINGUISHES THE AI BOARD FROM OTHER
BOARDS AT THE SAME IRQ LEVEL ON SIMULTANEOUS
INTERRUPT REQUESTS.
SWITCH 1 ON D00
SWITCH 2 ON D01
SWITCH 3 ON D02
SWITCH 4 ON D03
SWITCH 5 ON D04
SWITCH 6 ON D05
SWITCH 7 ON D06
SWITCH 8 ON D07
J101 SHUNT 1–2 AND SHUNT 2–3 ARE USED TO SET THE AO
BOARD OUTPUT VOLTAGE. A VOLTAGE RANGE IS SELECT-
ED BY CLOSING THE CORRESPONDING SHUNT.
The output voltage range for all analog outputs is strappable for ±10 Volts or 0 to +10 Volts.
The output impedances are 200 Ω and the outputs should be loaded with about 1 MΩ to give
full output with 12 bit accuracy. There are also two buffered +10 Volt outputs and two –10 Volt
outputs. All outputs are protected against EMI with RC filters and diodes.
A DC/DC converter (IC115) uses the +5 Volt supplied from the VME bus via the P1 connector
to produce +15 Volts and -15 Volts. Via a voltage regulator (IC118), +5 Volt is then derived
from the +15 Volt output of IC 115.
There is also an opto isolated POWER OK signal to the VME bus interface which indicates if
the +15 Volt or -15 Volt voltages on the board fall below +10 V or -10 V respectively.
7-2-9-1 Specifications
Table 7-11: Beam current analyzer specifications
Number of inputs 32
Number of inputs 5
V C =24 V
V IL Max 6 V
V IH Min 21.5 V
I IL at V I =6 V Max -10 mA Typ -8 mA
I IH at V I =21.5 V Min -0.8 mA Typ -1 mA
Analog Outputs
Number of outputs 2
Output 1 voltage with ±20 µA input ±10 V
Output 2 voltage with ±200 µA input ±10 V
Accuracy ±1% ±0.01 µA
Impedance output 1 and 2 typ 200 W
POWER OK Output
Power requirements
+5 V supply
Voltage 4.5–5.5 V
Each current-to-voltage converter uses one of four operational amplifiers contained in a quad
operational amplifier. All together eight quad operational amplifiers are used (IC 1-8).
All current inputs are protected against EMI with RC filters and diodes.
Two multiplexers (IC10, 11) select one of the 32 voltage converters for output. The output
from the multiplexers are amplified (IC 9). Output 1 gives ±10 V out for ±200 µA in and Output
2 gives ±10 V out for ±20 µA in. The outputs are read by a 12-bit ADC to achieve high accuracy
(±0.01 µA).
The five channel select inputs are optoisolated and use 24 V signals with negative logic.
A DC/DC converter (IC15) produces ±15 V from the single +5 V supply of the VME bus.
There is also an opto isolated POWER OK signal with open collector output, which indicates
if the +15 V or -15 V voltages on the board fall below +10 V or -10 V respectively.
Number of DC motors
Input characteristics
Output characteristics
Power requirements
Temperature
Mechanical dimensions
Depth with front panel and connectors 169.93 ± 0.4 mm= 6,690±0.015 inch
The DCM power supply provides the ±15 V voltages. A voltage regulator uses the +15 V to
produce a stabilized +5 V supply.
One analog and three digital signals are used to control the eight output voltages.
The “flying capacitor” technique is used, the eight flying capacitors must be updated once
per second.
The motor drive current can be read out as an analog signal (multiplexed).
There is also an open collector POWER OK signal to the VME bus interface which indicates if
the +15 V or -15 V voltages on the board fall below +10 V or -10 V respectively.
IS MAINT The Ion source runs from the MSS. It is not allowed to run RF under
these circumstances.
BEAM ALLOW Radiation shield doors are closed and its Ok to run beam.
7-2-12 Disassembly/reassembly
The purpose of this section is to detail the removal and reinstallation of the rack-mounted
circuit boards within the General Control Unit.
CAUTION!
The GCU boards and externally connected devices can be damaged if the boards are
connected or disconnected with power on.
Always turn off the VME power supply and peripheral power supplies before connecting
or disconnecting the GCU boards at the backplane.
For Lock-Out and Tag-Out (LOTO) procedures of the equipment, see Section 3-5-6 LOTO
procedures for the MINItrace system on page 59.
7-2-12-1 Tools
One tool is required for the removal and re-installation of the rack-mounted circuit boards:
2 Press the upper handle upwards and the lower handle downwards until the board comes
loose from its connector.
1 Check that jumpers and switches are correctly set for the board, refer to the previous
sections. For the CPU board, check that two PLCC flash memories are present on the
board.
2 Insert the board, with the component side facing to the right, into the upper and lower
guide rails.
3 Press the upper and lower handles simultaneously until the board is completely inserted.
Do not use excessive force. Difficulty in mounting may be due to misalignment between
the board and the guide rails.
4 Tighten the two retaining screws on the front panel of the board.
When loading software from the master, the CPU will automatically test communications
with the different boards.
4 RUN, SCON and STATUS LEDs should be lighted on the CPU board (STATUS LED only
faintly lighted).
This GCU type is designed around an ETX CPU module which communicates with other
modules via a PCI bus. It also has integrated Ethernet interface, PCI bus interface, and serial,
USB, and monitor ports which are accessible via connectors on the front and rear panels.
All electronics is mounted on a built-in I/O board and a connector board. There are no
replaceable circuit boards.
• Digital I/O
• Analog I/O
• DC motors
It is also equipped with ports for connecting external control and monitoring equipment, for
example the MINItrace Service System (MSS).
From the mains power inlet, an integrated power supply generates the basic power
requirements for the GCU. DC/DC converters on the I/O board generate the necessary internal
voltages.
The GCU is connected to the Master System via an Ethernet local area network. Any new GCS
software stored on the Master System will be downloaded to the GCU at start-up, or whenever
the GCU is reset.
Before disassembling the GCU, remove the mains power cable from the mains power
inlet at the rear of the GCU/external PS.
7-3-1 Description
The GCU is located in electronics cabinet CCAB. The GCU consists of a rack, front panel, I/O
board, connector board, power supply and an IEC320 power inlet filter with fuses.
IS MAINT Green Ion source is run from the MSS. RF cannot be run under these condi-
tions.
BEAM ALLOW Green Radiation shield doors are closed and it is OK to run beam
System Activity Green The application software on the IDE flash disc is accessed (Flashing)
Note!
For information on error conditions indicated by LEDs, see Section 7-3-5 Troubleshooting on
page 221.
Button Function
Ground fault reset Resets the ground fault state LED indicator
System reset Reboots the GCU. The firmware will NOT be reset or implement-
ed.
CAUTION!
Before resetting a ground fault, investigate and correct the error causing the fault. (See
Section 7-3-5 Troubleshooting on page 221.)
1 If the service laptop has no serial port, use the USB-to-RS232 adapter supplied.
2 For manufacturing purposes.
The MSS connector on the front panel is used for connecting the MINItrace Service System
(MSS) to the GCS. A shielded, straight cable with 9-pole D-Sub connector should be used. If
the service laptop has no serial port, use the USB-to-RS232 adapter supplied.
There are service locations on the GCU and the Cyclotron Support Cabinet (CSC). They are all
equipped with identical MSS connectors. It makes it possible to control the GCU from either
of these service locations. However, only one of the service locations can be used at the time
to connect a MSS to a specific unit.
Note!
It is not possible to communicate with the GCS from several service locations at the same time.
The CONSOLE connector can be used for direct communication with the GCU hardware, for
example by connecting a laptop and using a terminal communication software (terminal
emulation).
The two USB connectors for standard USB devices (for example a keyboard or mouse) and
the MONITOR connector are used for manufacturing purposes only.
Note!
The below figure shows a GCU with internal PS. For a GCU with external PS, the mains power
switch and fuses are located on the external PS.
The test points are used for measuring internal voltages as listed in Table 7-16. The test points
are short-circuit protected by 1 kΩ serial resistors.
Note!
For information on voltage tolerances, see Section 7-3-5 Troubleshooting on page 221.
Table 7-17 lists the GCU rear panel connectors with cable connection points.
Connector Connected to
P1 External PS in CCAB
(only on GCU with
external PS)
P3 CSC/MSS
P9 RFPG/SKG
P10 RFPG/SKH
P11 PSARC/J1
P12 MDC/PSARC
P13 CSC/P7
P14 CSC/P2
P15 CSC/P3
Connector Connected to
P17 CSC/P5
P18 CSC/P8
P19 TSC/P1
P20 TSC/P2
J8 MSB/TS
(only on GCU with in- J8 is the mains power switch. Contains fuses.
ternal PS)
All connectors are D-Sub connectors of different types, except P1 (Molex) and P25 (RJ-45).
For pin configuration, see Section 7-3-8 Interconnect data on page 234.
Old GCU with internal PS: New GCU with external PS:
The rating label is located at the rear of the The rating label is located on the external PS
GCU. plate.
The control system is designed around the ETX CPU module. Besides the main processor it
also has integrated Ethernet interface, PCI bus interface, and serial, USB, and monitor ports.
These interfaces and ports are accessible via connectors on the front and rear panels. The
ETX module communicates with other modules on the I/O board via a PCI bus.
The FPGA contains a major part of the system specific logic functions, for example motor
control logics, digital I/O and analog I/O. These functions are used to control and monitor the
accelerator subsystems via connectors on the rear panel. The subsequent sections provide
more information on the I/O functions.
The CPLD is used to control the reset and boot sequences of the GCU.
Note!
The below figure shows a GCU with internal PS. For a GCU with external PS, the mains power
switch and fuses are located on the external PS.
Note!
Voltage supply shown with thicker lines.
• Gas pressures
• Vacuum level
• Phase load
• Motor positions:
- Beam extraction
The Analog Out module sets the Dee voltage, extraction motor speed and the Ion Source
Power Supply Current in the accelerator.
The switched power supply provides 24 V DC, ±15 V DC and 5 V DC for the GCU. DC/DC
converters inside the GCU provide the other necessary internal voltages; 1.2, 2.5, 3.3 and
isolated 5 V DC. See Figure 7-22.
7-3-2 Operation
The GCU hardware is initialized by the local BIOS before the boot process is started. The GCU
operating system, located on the IDE flash disc, is then loaded into the memory by the BIOS.
The operating system checks for any program updates on the Master System before starting
the application program, which sets the GCU in its operational default state.
Resets and boots the GCU completely. Also implements any new application software
or firmware for the GCU. Also called cold reset.
Reset of the application software. The firmware will not be reset; any new firmware will
not be implemented. Also called warm reset.
Occurs at some pre-defined error conditions detected by the control unit. Same type of
reset as when pressing the System Reset button.
1 The 24V, +15V, -15V and 5V LEDs are lit if the power supply delivers power at these
voltages. The actual voltages are not necessarily within specifications.
2 The System Boot LED is lit during loading of the FPGA firmware.
3 The Power OK LED is lit when the ETX module has booted and all system voltages are
within the specifications.
4 The Init Ready LED is lit when the system has completed all initialization procedures.
CAUTION!
Do not change a parameter value unless the consequences are fully understood.
Changing IP addresses
The default IP addresses in a new GCU are:
GCS 192.0.4.40
Note!
The default netmask is 255.255.255.0.
These are normally used when the Master System is not connected to a local network.
However, when connected to a local network with other addresses, the default addresses
are not applicable and have to be replaced.
Note!
The IP addresses of the Master System and GCU must be in the same sub-net.
1 Connect the laptop to the CONSOLE port using a male, straight, 9-pole, D-Sub serial
cable.
4 A terminal software is a program in Windows for communicating with other devices over
the serial port of your computer. By running a terminal you can communicate directly
with the CPU in the GCU. When you start the terminal application for the first time, you
will be asked to set the communication parameters.
Data bits: 8
Stop bits: 1
Parity: None
5 Press Enter until the log-in prompt appears and log in as follows:
Type set_ip
Note!
The new addresses used above are only examples.
The GCU can now connect to the Master System when reset, and check for new software.
Note!
To maintain maximum long-term performance of the accelerator, shut down the system in a
controlled way from the Master System.
If the mains power to the GCU needs to be turned off, we recommend shutting down the
accelerator in a controlled way from the Master System. If not, the accelerator subsystems
might be exposed to unnecessary wear.
To shut down the MINItrace system from the Master System, see MINItrace Qilin Operator
Guide (dir. 2232995).
When the system is shut down, the GCU can be turned off at the Mains Switch (located on
the PS for a GCU with external PS, or on the GCU for a GCU with internal PS). All GCU outputs
will then automatically be deactivated.
The new GCU software is a part of the complete MINItrace application software, which is
distributed on a CD included in the software upgrade kit. The software upgrade is made on
the Master System.
During an GCU software upgrade, the GCU configuration file with the parameters settings
might be over-written by a default version. Therefore, the GCU configuration should be backed
up before the upgrade to facilitate reinstalling the old configuration after the upgrade.
However, if it is obvious in the release note supplied that the configuration file in the GCU will
not be affected, it is normally not necessary to back up and reinstall the configuration.
Note!
In general, the GCU configuration should be backed up:
The procedure details depend on the software version on the service laptop:
Note!
If the laptop has no serial port(s), use the USB-to-RS232 adapter(s) supplied.
1 Start MSS.
3 Click Export.
4 In the dialog, type a suitable time stamp after Configuration for easy identification.
5 Click Save.
b Click Config.
10 Click Exit.
The configuration file is now saved on the MSS as GEM_CONF.DAT in the folder
C:\MINItrace\MSS.
After the installation, cycle mains power on the GCU to boot the GCU.
During the boot procedure the new software on the Master System will be loaded to the GCU.
The release note supplied should tell if the configuration file has been replaced.
Note!
If the laptop has no serial port(s), use the USB-to-RS232 adapter(s) supplied.
1 Start MSS.
3 Click Import.
5 Click Open.
The old configuration file saved on the MSS laptop reflects the actual system setup, such as
installed components, parameter settings, etc.
b Click Config.
9 Click Exit.
7-3-3 Specifications
For pin configuration, see Section 7-3-8 Interconnect data on page 234.
Digital I/O
Optoisolated Inputs
VC (+24 V) 24 V ±10%
Optoisolated Outputs
VC (+24 V) 24 V ±10%
Analog I/O
Analog Inputs
Input impedance 1 MΩ
Analog Outputs
Number of outputs 8
DC motor driver
Number of DC motors
Output characteristics
Number of inputs 16
Power requirements
Dimensions
Weight 7.5 kg
Environment
Ambient temperature
Control system room 18–25°C
Directives This product is marked with the CE symbol in accordance with the
following directives:
• EMCD 2014/30/EU concerning electromagnetic compatibility.
This product is marked with the GOST (Russia only) and EAC sym-
bols.
The product meets all applicable requirements and technical reg-
ulations in the Customs Union member countries (Russia, Belarus,
Kazakhstan, Armenia, and Kyrgyzstan).
Safety standards EN 61010-1 ed. 3:2010, concerning safety requirements for electri-
cal equipment for measurement, control and laboratory use.
• Pollution degree: 2
• Installation category: 2
• Group 1
This product uses RF (radio frequency) energy only for its in-
ternal function.
To reduce the risk of electrical shock, do not remove covers. Refer servicing to qualified
personnel.
For Lock-Out and Tag-Out (LOTO) procedures of the equipment, see Section 3-5-6 LOTO
procedures for the MINItrace system on page 59.
7-3-4-1 Cleaning
When necessary, wipe the panel using a cloth moistened with mild detergent.
Required tools:
1 Open the rear door on CCAB using the cabinet door tool.
2 Measure the voltages on the test points at the rear of the GCU (see Table 7-21).
3 Check that they are within the specifications. If not, contact Online Center.
7-3-5 Troubleshooting
This section describes troubleshooting of the GCU based on error messages, LED indicator
status and other measurements.
RAD SH OK ON Radiation shield doors closed OFF Radiation shield doors not closed
(micro switch, signal A)
IS MAINT OFF Ion source is not run from the ON Ion source is run from the MSS.
MSS RF cannot be run then.
MAGN D OK ON Magnet door is closed (magnet OFF Magnet door not closed
switch, signal B)
BEAM ALLOW ON Radiation shield doors are closed OFF Radiation shield doors are not
and it is OK to run beam closed and/or beam cannot be
run
RCAV W OK ON Water cooling is OK for RF cavity OFF Water cooling not OK for RF
and it is OK to run RF cavity
PSARC P ON ON Power for PSARC is turned on OFF Power for PSARC is turned off
Eth Activity FLASHING Communication through Ether- OFF No communication through Eth-
net interface (LAN). ernet interface (LAN).
Init Ready ON System initialization completed OFF System initialization not complet-
ed
System Activi- FLASHING Accessing the application soft- OFF No communication with the IDE
ty ware located on the IDE flash flash disc
disc
System Boot FLASHING The GCU boots up (a few flashes OFF GCU does not boot if no flashes
during the booting process) during the booting process
Ground fault OFF No ground fault is detected ON Ground fault detected and not
reset
If all LEDs are off, check the fuses at the mains inlet of the GCU. See Section 7-3-6
Disassembly/reassembly on page 226 for instructions.
If the potential difference exceeds the accepted value, the error handler in the GCU:
• turns on the Ground fault LED (will be lit until the Ground fault reset button is pressed)
Corrective action: Press the Ground fault reset button on the front panel. If the RF
still is turned off and the Ground fault LED turned on, the error
is permanent and needs more investigation. Check the voltages.
Note!
This is not a calibrated measurement device but gives valuable information for trouble-shooting
the GCU.
Data bits: 8
Stop bits: 1
Parity: None
3 Press Enter until the log-in prompt appears and log in as follows:
4 Type show_voltages and press Enter. A list of the voltages will be shown on the screen.
See also Table 7-23.
5 Check that they are within the specifications. If not, call Online Center.
• The version information of the application file gcsd is shown in the MSS. The application
has a major and minor version number. See also Section7-3-2-4 Upgrading GCU software
on page 213.
• The version information of the firmware is checked by the application at start. The
firmware has a major and minor version number.
7-3-6 Disassembly/reassembly
The purpose of this section is to detail the removal and reinstallation of the fuses and of a
complete GCU.
To reduce the risk of electrical shock, do not remove covers. Refer servicing to qualified
personnel.
Lock-out and tag-out the control cabinet circuit breaker at the MDP, before any service
is performed on the GCU.
CAUTION!
Shut down the system from the Master System before turning off the power to the GCU.
CAUTION!
The GCU can be damaged if cables are connected or disconnected with power on.
For Lock-Out and Tag-Out (LOTO) procedures of the equipment, see Section 3-5-6 LOTO
procedures for the MINItrace system on page 59.
Note!
There are two GCU Gen II models:
When replacing a GCU with internal PS it must be upgraded to a GCU with an external PS.
7-3-6-2 Tools
• General FE service tools
2 Press OFF on the VCU control panel located on the cyclotron. Wait 30 minutes to allow
the vacuum system to cool.
3 Open the rear door on CCAB using the cabinet door tool.
6 Disconnect the IEC 320 mains cable from the mains power inlet (2 or 6).
7 For a GCU with external PS: Disconnect the DC cable (4) from the rear of the GCU.
1
1
0
4
5
Description Location
CAUTION!
Call service if the new fuses blow immediately when turning on the system.
There are two fuses located at the mains power inlet. The mains power inlet is placed on the
GCU or external PS, see Figure 7-23.
1 Shut down the GCU. See Section 7-3-6-3 Shutting down power to the GCU and PS on
page 227.
Note!
The below figure shows a GCU with internal PS, where the fuses are found on the GCU.
For a GCU with external PS, the fuses are located on the external PS.
3 Replace the old fuses with new ones (rating T4AH 250 V).
5 Turn on the power as described in Section 7-3-6-7 Turning on the power to the GCU and
PS on page 233.
6 Close and lock the rear door on CCAB using the hex wrench.
• For instructions on upgrading from GCU Gen I to Gen II, refer to Customer Information
DOC0759837CI.
• For instructions on rebuilding a GCU Gen II with internal PS into a GCU with external PS,
refer to DOC1719487.
Note!
There are two GCU Gen II models:
When replacing a GCU with internal PS it must be upgraded to a GCU with an external PS.
Note!
Before removing the old GCU, back up the configuration file in the GCU. The new GCU only
contains a default version of the configuration file. By transferring the old file to the new GCU,
the file does not have to be changed again. For instructions, see Section 7-3-2-2 Configuration
of the GCU on page 211.
3 Remove the four retaining screws on the front panel of the GCU using a screw driver
(see Figure 7-25).
5 If the GCU has an external PS and if the PS is not working properly, remove the PS
according to the instructions in Section Removing an external PS on page 232.
Note!
Pay attention to the labels on the cables and the marking on the connector board at the
rear of the GCU!
5 Turn on the power as described in Section 7-3-6-7 Turning on the power to the GCU and
PS on page 233.
6 Close and lock the rear door on CCAB using the cabinet door tool.
Note!
There are two GCU Gen II models:
When replacing a GCU with internal PS it must be upgraded to a GCU with an external PS.
1 Remove the old GCU (see Section Removing the GCU on page 230).
2 Install a new GCU (see Section Installing the GCU on page 230).
Figure 7-26:
1: External PS fitted on the side rails of the cabinet.
2: DC connector on GCU.
1 2
Removing an external PS
To remove an external PS:
1 Shut down the power to the GCU and PS, see Section 7-3-6-3 Shutting down power to
the GCU and PS on page 227.
2 Remove the four screws that fastens the PS to the cabinet stand:
Installing an external PS
Viewed from the back of the cabinet, install the PS on the left hand side of the cabinet, as
follows:
1 Mount the PS plate with two M5 screws to the top rail, positioned according to the figure
below.
2 Drill a 4.5 mm hole in the lower rail at the plate’s lower left cutout. Place a 2 mm washer
under the plate and fasten the PS to the lower rail with a M5 screw.
3 Turn on the power as described in Section 7-3-6-7 Turning on the power to the GCU and
PS on page 233.
2 For a GCU with external PS: Connect the DC cable running from the PS to the Connector
for DC cable (4) on the GCU.
5 Close the rear door on CCAB using the cabinet door tool.
6 Check that the GCU is working properly, as decribed in Section 7-3-6-8 Basic function
check on page 234.
• The GCU starts the boot process. It can be followed by the LED indicator sequence. See
Section 7-3-2-1 System startup, reset and booting on page 210 LED indicator sequence
during reset and boot process. If there is new application software or firmware on the
Master System, it will be downloaded to the GCU.
Note!
If other IP-addresses than the standard MINItrace addresses are used on the site, the
IP-addresses in the new GCU must be changed manually. For instructions see
Section 7-3-2-2 Configuration of the GCU on page 211.
• When the boot process is completed, check the status of the LED indicators. See
Table 7-13.
Spare part
For spare part numbers, refer to the spare parts listed for each subsystem in MINItrace Spare
Parts Catalog (dir. 2380006-100).
Rear connectors
1, 2 24V supply
3 +15V supply
4 -15V supply
5 +5V sense
13 GND 5V sense
Shield Chassis
1 Not Used 1
2 TX1 2
3 RX1 3
4 Not Used 4
5 GND_COM1COM2 5
6 Not Used 6
7 Shorted to pin 8 7
8 Shorted to pin 7 8
9 Not Used 9
Shield Chassis
1 Not Used 1
2 RX3 2
3 TX3 3
4 Not Used 4
5 GND 5
6 Not Used 6
7 Shorted to pin 8 7
8 Shorted to pin 7 8
9 Not Used 9
Shield Chassis
1 Not Used 1
2 RS485 - 2
3 Not Used 3
4 Not Used 4
5 GND 5
6 RS485 + 6
Shield Chassis
1 Not Used 1
2 RS485 - 2
3 Not Used 3
4 Not Used 4
5 GND 5
6 RS485 + 6
Shield Chassis
1 Not Used 1
2 RS485 - 2
3 Not Used 3
4 Not Used 4
5 GND 5
6 RS485 + 6
GCU/P9 RFPG/SKG
Shield Chassis
1 Chassis 1
2 STANDBY 2
3 NORMAL MODE 3
4 RESET 4
5 FIX MODE 5
6 PULSE MODE 6
7 FREQ LOAD 7
8 RF INHIBIT 8
9 STANDBY READY 9
10 STANDBY ERROR 10
11 START MODE 11
12 TUNING MODE 12
13 DEE VOLTS OK 13
14 SPARK* 14
16 GN Grid normal 16
17 SN Screen normal 17
18 EN EHT Normal 18
19 V+24V_IO2 19
20 HN Heater Normal 20
27 FREQ DATA 27
28 DPA TEMP ok 28
29 FREQ CLOCK 29
30 Spare 30
31 Controller fault* 31
34 RF standby 34
35 Not Used 35
36 Not Used 36
37 GND_IO_2 37
GCU/P10 RFPG/SKH
Shield Chassis
1 Chassis 1
2 GND_AIO 2
3 DVR_RTN 3
4 Spare -AI33_-AI34_-AI35 4
5 Spare –AI35 5
6 FP_RTN 6
7 RP_RTN 7
10 DVR SIG 10
11 Spare –AI33 11
12 Spare –AI34 12
13 FP SIG 13
14 RP SIG 14
GCU/P11 PSARC/P1
Shield Chassis
1 Not Used 1
2 Not Used 2
8 PSARC VOLTAGE 8
10 Not Used 10
11 Not Used 11
12 PSARC HV ON 12
18 +15VOLT PSARC 18
19 GND PSARC 19
21 PSARC CURRENT 21
22 GND PSARC 22
23 GND PSARC 23
24 PSARC NO FAULTS 24
25 Not Used 25
Shield Chassis
1 Chassis 1
2 Not Used 2
3 V+24V_IO 3
6 Not Used 6
7 V+24V_IO 7
GCU/P13 CSC/P7
Shield Chassis
1 Chassis 1
2 V+24V_IO 2
3 V+24V_IO 3
4 Not Used 4
5 Not Used 5
6 +PRESSURE 13N 6
7 +PRESSURE 18F 7
8 +PRESSURE F2 8
9 +PRESSURE DUAL 9
14 GND_IO 14
15 GND_IO 15
16 V+24V_IO 16
17 V+24V_IO 17
18 V+24V_IO 18
19 V+24V_IO 19
20 V+24V_IO 20
21 V+24V_IO 21
22 GND_IO 22
23 Not Used 23
24 GND_IO 24
25 Not Used 25
GCU/P14 CSC/P2
Shield Chassis
1 Chassis 1
2 Pirani 2+ 2
3 Pirani 1+ 3
4 Penning + 4
7 FOIL 1 CURRENT 7
8 GND_BCA 8
9 TARGET 5 CURRENT 9
10 TARGET 3 CURRENT 10
11 TARGET 1 CURRENT 11
12 GND_IO 12
13 V+15V_CI 13
14 -10VOLT (AO) 14
16 -10VOLT (AO) 16
18 -10VOLT (AO) 18
25 FOIL 2 CURRENT 25
26 PROBE CURRENT 26
27 TARGET 6 CURRENT 27
28 TARGET 4 CURRENT 28
29 TARGET 2 CURRENT 29
30 - He Pressure 30
31 + He Pressure 31
32 Not Used 32
33 +10VOLT (AO) 33
34 +10VOLT (AO) 34
35 +10VOLT (AO) 35
36 Not Used 36
37 GND RFPG 37
GCU/P15 CSC/P3
Shield Chassis
1 Chassis 1
2 GND_IO 2
3 BEAM_ALLOWED 3
4 Radiation interlock 4* 4
5 Radiation interlock 2* 5
7 GND_IO 7
8 VACUUM OK 8
10 Target_Dual_Waterflow ok 10
14 GND_IO 14
15 GND_IO 15
16 Radiation interlock 5* 16
17 Radiation interlock 3* 17
18 Radiation interlock 1* 18
19 MAGNET TEMPERATURE OK 19
20 WATER COOLING OK 20
21 GND_IO 21
22 GND_IO 22
23 GND_IO 23
24 GND_IO 24
Shield Chassis
1 MAGNET TEMPERATURE OK 1
Shield Chassis
1 MAGNET TEMPERATURE OK 15
4 Not Used
7 Not Used
8 Not Used
9 Not Used
5 shorted to pin 12
GCU/P17 CSC/P5
Shield Chassis
1 Chassis 1
3 GND_IO 3
4 GND_IO 4
5 DUAL_MOTOR 2 5
6 DUAL_MOTOR 1 6
7 SINGLE_FOIL_MOTOR 7
12 V+24V_IO 12
13 V+24V_IO 13
14 HELIUM OUT 14
15 TARGET 6 COOLING 15
16 TARGET 4 COOLING 16
17 TARGET 2 COOLING 17
18 He COMPRESSOR ON/OFF 18
19 HE Compressor ON 19
21 GND_IO 21
22 GND_IO 22
23 DUAL_MOTOR 2 23
24 DUAL_MOTOR 1 24
25 SINGLE_FOIL_MOTOR 25
26 GND_IO 26
31 V+24V_IO 31
32 HELIUM IN 32
33 HELIUM LOOP 33
34 TARGET 5 COOLING 34
35 TARGET 3 COOLING 35
36 TARGET 1 COOLING 36
37 GND_IO 37
GCU/P18 CSC/P8
Shield Chassis
1 Chassis 1
2 Not Used 2
9 OPEN SPARE V4 9
10 OPEN SPARE V2 10
18 V+24V_IO 18
19 V+24V_IO 19
20 Not Used 20
21 Not Used 21
22 RF ON (RFPG on) 22
23 PSMC ON 23
28 OPEN SPARE V3 28
37 V+24V_IO 37
GCU/P19 TSC/P1
Shield Chassis
1 Chassis 1
2 GND_IO 2
3 GND_IO 3
4 DI_SPARE1 4
5 DO_SPARE2 5
6 +Pressure 11C 6
7 11C_TEMP_OK 7
8 11C_VALVE7 8
9 11C_VALVE5 9
10 11C_VALVE3 10
11 11C_VALVE1 11
12 V+24V_IO 12
13 V+24V_IO 13
14 GND_IO 14
15 GND_IO 15
16 DI_SPARE2 16
17 DO_SPARE3 17
18 DO_SPARE1 18
19 11C_DI_NOT_Used 19
20 11C_FURN_ON 20
21 11C_VALVE6 21
22 11C_VALVE4 22
23 11C_VALVE2 23
24 V+24V_IO 24
25 V+24V_IO 25
GCU/P20 TSC/P2
Shield Chassis
1 Chassis 1
2 GND 2
3 SPARE4_FURN_ON 3
4 SPARE2_FURN_ON 4
5 DI_SPARE6 5
6 DI_SPARE4 6
7 DO_SPARE6 7
8 DO_SPARE4 8
9 15O_DI_NOT_USED 9
10 15O_VALVE3 10
11 15O_VALVE1_NOT_USED 11
12 V+24V_IO 12
13 V+24V_IO 13
14 GND_IO 14
15 GND_IO 15
16 SPARE3_FURN_ON 16
17 SPARE1_FURN_ON 17
18 DI_SPARE5 18
19 DI_SPARE3 19
20 DO_SPARE5 20
21 +PRESSURE_15O 21
22 15O_VALVE4 22
23 15O_VALVE2 23
24 V+24V_IO 24
25 V+24V_IO 25
Shield Chassis
1 Chassis 1
2 GND_IO 2
3 H2O Interlock* 3
4 H2O_DI_NOT_USED 4
5 H2O_TEMP_OK 5
6 H2O_FURN_ON 6
7 H2O_VALVE1_NOT_USED 7
8 V+24V_IO 8
9 GND_IO 9
10 H2O Interlock* 10
11 H2O_DI_NOT_USED 11
12 H2O_TEMP_OK 12
13 H2O_FURN_ON 13
14 H2O_VALVE1_NOT_USED 14
15 V+24V_IO 15
Shield Chassis
1 Chassis 1
7 SPARE_DI58 7
8 SPARE_DI46 8
9 SPARE_DO95 9
10 SPARE_DI30 10
11 SPARE_DO23 11
12 SPARE_DO1 12
13 V+24V_IO 13
14 GND_IO 14
20 SPARE_DI57 20
21 SPARE_DO96 21
22 SPARE_DI32 22
23 SPARE_DO22 23
24 SPARE_DO12 24
25 V+24V_IO 25
Shield Chassis
1 Chassis 1
10 SW1B TARGET 2 10
11 SW1B TARGET 1 11
12 SW1C TARGET 2 12
13 SW1C TARGET 1 13
1 Bi-directional pair A + 1
2 Bi-directional pair A - 2
3 Bi-directional pair B + 3
4 Bi-directional pair C + 4
5 Bi-directional pair C - 5
6 Bi-directional pair B - 6
7 Not Used 7
8 Chassis 8
Front connectors
Shield Chassis
1 Not Used 1
2 TX2 2
3 RX2 3
4 Shorted to pin 6 4
5 GND_COM1COM2 5
6 Shorted to pin 4 6
7 Shorted to pin 8 7
8 Shorted to pin 7 8
9 Not Used 9
Shield Chassis
1 Not Used 1
2 TX1 2
3 RX1 3
4 Shorted to pin 6 4
5 GND_COM1COM2 5
6 Shorted to pin 4 6
7 Shorted to pin 8 7
8 Shorted to pin 7 8
9 Not Used 9
• The secondary cooling unit comprises a water manifold, distributing cooling water to a
number of subsystems. See Chapter 9 Cooling water system for details.
• The Vacuum Control Unit (VCU) with control keys and display. See Chapter 10 Vacuum
system for details.
• A connector to the radiation shield door control. See Chapter 17 Radiation shield for
details.
• A Compressed air moisture trap is housed in the CSC. See Section 8-3 Compressed air
moisture trap on page 256 and Section 8-7-3 Empty the compressed air moisture trap
on page 261 for details.
• The Power Distribution Unit (PDU) contains Mains switch, fuses, CBs, contactors and
support power for vacuum pumps, helium cooling pump and radiation shield door
actuators. See Section 8-4 Power distribution (CSC electronics) on page 256 for details.
• The Customer Interface Connector (CIC) on the top of the CSC allows the customer to
add interconnects etc. See MINItrace Qilin Site Planning Guide (dir. 2232993-100) for
details.
The moisture trap is to be emptied during PM (Planned Maintenance). See Section 8-7-3 Empty
the compressed air moisture trap on page 261.
• The MDP, Mains Distribution Panel, is supported by the customer. It shall include two, or
more, Circuit Breakers (CBs) for the outgoing lines.
• The CCAB, Control Cabinet, has a Transformer from which the voltages are 400 Volts
(phase-to-phase) for the outlet to the CSC and the RFPG and internal to the GCU and the
PSMC. From the transformer there is also a 3×208 V-support for the PSARC. All the different
outlets from the transformer are secured by circuit breakers (CBs).
• The RFPG has two different power inlets, one 400–480 V coming from the MDP and an
other, 400 V coming from the CCAB.
• The CSC, Cyclotron Support Cabinet, has a 400 V-inlet from the CCAB. The outlets are
400 V and 230 V (one phase) for pumps and valves.
The units in the power distribution system contains and provides lethal voltages of up to
mains voltage, 400 V AC.
Lock-out and tag-out the mains circuit breakers in the MDP before servicing the power
distribution system. Verify with a multi-voltmeter.
CAUTION!
If you apply power and run the pumps with the phases swapped, this will damage the
system.
Consider this when installing or maintaining the power support to the MINItrace subsystem.
Always check that the pumps run in the correct direction. If a pump run in wrong direction,
correct the fault at the right place. Use the drawings for guidance (L1, L2, L3, N and PE
must be connected to the correct terminal).
When applying LOCK-OUT and TAG-OUT to the main CBs in the MDP, the power for
the Vacuum system will be turned off as well. But the Vacuum system needs to cool
down for at least 1 hour before turned OFF, so please plan your time.
• If any of the CBs in the Control Cabinet or RFPG are turned OFF, instead of the Main CBs
in the MDP, please observe this:
The RFPG gets its power input from two different units. One from the MDP and an
other from the Control Cabinet. Only when both of these inputs are turned off the
RFPG is secure for servicing. Verify with a multimeter that the power is OFF.
For more information about maintenance on each unit, please refer to corresponding chapter
in this manual or MINItrace Service Manual – OEM Manuals (dir. 2233002-100).
• If the main transformer in the CCAB, Control Cabinet, is going to be replaced it’s important
that the voltage tapping on the new unit is the same as the original one. On the top of
the transformer the voltage inputs are marked 400 and 480 V.
8-4-5 Specifications
The mains input voltages can be 400 V AC or 480 V AC, 50 or 60 Hz.
RFPG 25 kVA
Total 50 kVA
In the CCAB (Control cabinet) there are seven CBs (Circuit Breakers) to turn off the CSC, PSMC,
GCU, PSARC and one of the lines to the RFPG. See Figure 8-5.
9-1-1 Specifications
The external primary cooling system is connected to a heat exchanger, which is part of the
closed loop of the secondary cooling system. The primary cooling system supply’s inlet water
at a temperature of not more than +13°C and provides temperature regulation of the
secondary cooling system within ±1°C.
A lower temperature of the cooling medium with a corresponding decrease in flow can be
acceptable if the problems with condensation are handled accordingly.
Cooling capacity 30 kW
Flow 50 liters/min
F20 RFPG 15
F22 RF 1 10–12
F23 RF 2 10–12
F26 PSMC 4
F27 VACUUM 4
Figure 9-1: Cooling water components in the CSC (Cyclotron Support Cabinet)
The left red diode turns on if the The green diode turns on if the tem- The right red diode turns on if the
temperature is too high. perature is OK. temperature is too low.
3 Turn the “sun” knob to the following value: Value on water temperature digital display
(green figures right above the controller) - 20.
(Example: If the value on water temperature display is 25, turn knob to 25-20 = +5.
If the value on water temperature display is 17, turn knob to 17-20 = -3.)
4 Remove the black cover from the hole to the left of the “moon” knob.
5 Use a small screwdriver and adjust the potentiometer until the green LED lights up. See
Figure 9-5.
6 Put the “sun” knob back to 0. The regulator should now regulate the water temperature
to 20°C.
7 The white potentiometer at the upper left corner should be adjusted to give 10–20 pause
between shunt motor motions. Check that the system is not oscillating.
Minus on the control unit is lit when the motor will open towards the heat exchanger, plus is
lit when the motor will close. Green light indicates that the temperature is ok.
If the temperature is stable in the range 19–21º and the green 0 indicator on EVR-CA is lit,
the temperature is ok. If the temperature is unstable and/or the red - or + indicator is lit, it
needs to be adjusted.
4 Adjust knob P until the temperature is stable in the range 19–21ºC and the green 0
indicator is lit.
Make sure that the temperature is stable in the range 19–21ºC. The display on CRA112 shows
the read temperature by default.
To change set-temperature
The set-temperature should be 20ºC.
1 Make sure that the motor wheel is fully pressed in order to enable valve control.
2 Roll the wheel until the display shows P-4 (default from factory).
2 Check for damage on all tubing exposed to radiation. Exchange tubing if necessary.
5 Check all water flow switches: Turn off the valve on each circuit on the pressure side.
Check the signal response. Record all parameters of interest, including these:
d Water conductivity.
6 Replace the ion exchanger if necessary (too high conductivity). See Section 9-7 Ion
exchanger and filter on page 280.
7 Disassemble and clean the filter. See Section 9-7 Ion exchanger and filter on page 280.
9-5-1 Tools
• Standard hand tool kit
WARNING! Radiation
Always monitor radiation levels before servicing the water cooling system. Pay particular
attention to the filter and the ion exchanger, where radioactive particles may accumulate.
The secondary cooling system (water pump, CSC, PSMC, RFPG) contains lethal voltages
of up to the mains voltage, 400 V AC.
The increased height of the newer water cooling pump makes it interfere with the piping
above the pump. This calls for a pipe conversion kit if an old pump is to be replaced by a new
pump. See table below for detailed information and the spare parts.
You have an old pump The motor (or any other vi- You also need a pipe con-
2223593 (50 Hz) or tal part) is broken. version kit, since the new
2273907 (60 Hz). motor is slightly bigger
than the old motor.
For spare part numbers, refer to the spare parts listed for the sub systems in MINItrace Spare
Parts Catalog (dir. 2380006-100).
For spare part numbers, refer to the spare parts listed for the sub systems in MINItrace Spare
Parts Catalog (dir. 2380006-100).
15 Replace the old pump with the new one. Mount the
nuts and the clamping rings on the flanges.
2 Mount the terminal at the top end of the DIN rail, on top of the fuse terminal.
3 Connect the 1-meter cable between the top terminal and terminal 3 at the bottom left
in the cabinet.
4 Remove the old cable from F3 on the right side of the fuse terminal.
(This will not be used any longer.)
5 Remove the old shunt valve control unit and mount the new CRA112 shunt valve control
unit in the Cyclotron Support Cabinet.
6 Attach the sensor to the brass T-piece with the hose clamp.
7 Route the cables the same way as the old ones and connect the power cable connector
to the shunt valve motor. Connect the other end of the cable to F3 on the new terminal
and to the free fuse connection in the cabinet.
Note!
It does not matter where it is connected to the new terminal and the free fuse terminal
since this motor runs on 24 V AC.
At the MSS (MINItrace Service System): At the CSC (Cyclotron Support Cabinet):
2 Open the air valve, V70, on the left side of the ion exchanger (Z1).
3 Open the quick connection on the right side of the ion exchanger (see Figure 9-11). Press
the blue ring and pull out the tube.
4 Loosen the tube fitting for the transparent hose at the air valve, V70.
5 Keep the hose end high by tying it to the tube above valve V2. Otherwise, empty the
contents of the hose into the bucket.
7 Loosen the connections at the top of the ion exchanger and keep them for the new ion
exchanger (see illustration to the left).
CAUTION!
Use Teflon thread tape only. Use of other materials may damage plastics in the cooling
system.
1 Wind Teflon thread tape onto the threads of the tube connections.
2 Fasten the tube connections to the ports on the ion exchanger, by hand.
5 Attach the thin hose with quick connection to the ion exchanger.
6 Attach the transparent hose to the connection at air valve V70. Tighten the nut carefully,
do not overtighten as this may damage the hose.
7 Top up the filter circuit by opening valve V50 until water comes out of air valve V70, then
close V70.
9 Open valve V51 just enough for flow meter F10 to rotate.
10 Note that HIGH CONDUCTIVITY alarm may light when the top-up water circulates. If the
ion exchanger is working, the alarm indicator should extinguish within a few minutes.
6 Insert the new cartridge and fasten the filter housing by hand.
8 Top up the filter circuit by opening valve V50 until water comes out of air valve V70, then
close V70.
10 Check that the filter housing is tight. If water leaks, tighten further.
11 Note that HIGH CONDUCTIVITY alarm may light when the top-up water circulates. If the
ion exchanger is working, the alarm indicator should extinguish within a few minutes.
10 VACUUM SYSTEM
The vacuum system is controlled by a separate Vacuum Control Unit (VCU) independent of
the General Control Unit (GCU). The independent GCU ensures safe and automated operation
of the vacuum system. See Chapter 7 General control unit for a detailed description of the
GCU. The VCU is included in the Cyclotron Support Cabinet on the right side of the MINItrace,
see Figure 10-1.
The VCU includes a pressure gauge (TPG 300), displaying reading from different vacuum
system gauges.
During operation, the external surface of the vacuum diffusion pump can reach
temperatures of up to 250°C.
Do not touch!
The arm and pneumatic actuator for the high vacuum valve is a potential pinch hazard.
The lid of the high vacuum valve is a severe pinch hazard when closing.
10-2 Overview
The following description is based on Figure 10-2 and Figure 10-3 below.
PIR 1
BV
RP
In order to cover the pressure range from atmosphere to the 10-7 mbar (10-5 Pa) range, two
vacuum gauges are connected to the vacuum chamber. The Pirani gauge handles the 1 bar
to 10-3 mbar (0.1 MPa to 10-5 Pa) range and the Penning gauge takes over in the high vacuum
range, <10-3 mbar (<0.1 Pa).
A second Pirani gauge monitors the backing pressure of the diffusion pump.
The diffusion pump, which is supplied with a baffle, is the first stage in the vacuum pumping
system. In the second stage, the diffusion pump is backed by a rotary pump. The diffusion
pump is water-cooled, see Chapter 9 Cooling water system.
The vacuum chamber has a rather small volume and is made of aluminum. No iron surfaces
are exposed to vacuum (iron can have a negative influence on humidity).
For further information, refer to Manufacturer’s Manuals (OEM) for the vacuum system.
10-3-1 Specifications
Table 10-1: Vacuum chamber specifications
Material Aluminum
Power requirement
Power 2.4 kW
Weight 30 kg
Manufacturer/type Santovac 5
Power requirements
Power 1 kVA
Weight 33 kg
Fore-vacuum oil
Manufacturer/type Pfeiffer/P3
Volume 1.1 l
CAUTION!
The controls on the VCU are for use by authorized service personnel only, with the
exception of the vacuum readout procedure described.
For more information concerning operation of the VCU keyboard and readout of the pressure
gauge control, please refer to MINItrace Qilin Operator Guide (dir. 2232995) and Manufacturers
operating instruction.
The programming is done on the Balzers TPG 300 pressure gauge controller.
• To enter the SET UP mode, press funct and group key simultaneously
• Please program the parameters correct, or you need to start over from the beginning again.
• funct key Press this key every time there is bold-script-text in the func-key-fields below.
• step key Press this key to change value in the step-key-fields below. This has to be done when the text is in
bold-script.
56 Code Leave as 0
57 Code Leave as 0
58 Code Leave as 0
86 Code Leave as 0
87 Code Leave as 0
88 Code Leave as 0
• The chamber must be vented (see Section10-7-1 Open the vacuum chamber on page295).
The pumps may continue running to maintain the vacuum level in other parts of the
vacuum system.
Before replacing vacuum spare parts, the vacuum system must be closed down and the
compressed air supply must be disconnected (see Section 10-7-2 Shutting down the vacuum
system on page 298). To enable the pumps to cool down the vacuum system should be closed
down at least 4 hours before starting the disassembly work.
Note!
The vacuum pumps continue to run.
VV HVV
DP RV
PIR 1
BV
RP
2 Press the Left Door Open and Right Door Open switches to fully open the radiation
shield.
Personal injury can be caused when the radiation shield doors move.
The door pendant control must be disconnected from the CSC during service on the
cyclotron.
3 Remove the magnet yoke compression screw located in the upper left corner of the
magnet yoke, see Figure 10-6 (1).
4 Open the magnet yoke by hand. Use the slot in the door as shown in Figure 10-6 (2).
The magnet yoke will shut with high force if the magnet door switch is operated or if the
PSMC is accidentally turned on.
1. 2.
3.
WARNING! Radiation
Dust and fragments inside the vacuum chamber can be highly radioactive. The surface
of the vacuum chamber can also be contaminated.
Always use protective clothes, gloves and goggles when working inside the vacuum
chamber.
Always use cleaning tissues moistened with alcohol when cleaning surfaces with a potential
risk of contamination.
Follow federal state and/or local regulations for disposal of radioactive waste.
CAUTION!
Allow the diffusion pump to cool down for a total of 4 hours after shutting down the
vacuum system. Otherwise the pump oil may be oxidized and carbonized on the inside
surfaces of the diffusion pump.
c The rotary pump continues running for 1 hour (green) to maintain vacuum in the
diffusion pump while it is hot.
VV HVV
DP RV
PIR 1
BV
RP
3 Wait another 3 hours for the diffusion pump to cool down to a safe working temperature.
• Cover the O-ring surface with a thin layer of vacuum grease and make sure it is free
from dust and dirt.
3 Carefully check that the vacuum chamber can be closed and that nothing is squeezed
during closing.
4 Carefully close the magnet yoke by hand. Use the slot in the door as shown in Figure 10-7
(1), to reduce the pinch hazard.
1. 2.
5 Tighten the magnet with the compression screw, see Figure 10-7 (2), until the slot between
the yoke and the magnet is almost zero.
The vacuum system emits an oil mist during start-up. This can be harmful to the lungs.
Do not stay in the vicinity of the vacuum system during the first hour of pumping.
1 At the vacuum and cooling control panel – Press the PUMP key. Follow the pump-down
sequence and compare with earlier pressure vs. time records.
Note!
A variation with earlier records can exist due to different factors. High humidity and long
chamber opening time will lengthen the pump-down time. The Pump-down time can
under normal circumstances be 2–3 hours.
a Open the vacuum chamber, as described in Section 10-7-1 Open the vacuum
chamber on page 295.
c Close the vacuum chamber as described in Section 10-7-3 Closing the vacuum
chamber on page 299.
• Alcohol
• De-greasing solvent
• Caustic soda
• GE O-ring set
The pump oil in the diffusion pump and the rotary pump will be radioactive after production
runs.
Always wear protective gloves when replacing pump oil. Do not touch eyes or mouth.
Follow federal, state and/or local regulations when disposing of pump oil.
Cleaning tissue, rubber gloves, cotton tops or other items used for cleaning inside the
vacuum chamber are radioactive.
Dispose of these items in accordance with federal, state and/or other local regulations.
CAUTION! Cleanliness
The cleaner all surfaces inside the vacuum chamber are, as well as all other components
that get in contact with the vacuum system, the better the end pressure will be.
Always use rubber or latex gloves when you work with vacuum components.
For Lock-Out and Tag-Out (LOTO) procedures of the equipment, see Section 3-5-6 LOTO
procedures for the MINItrace system on page 59.
2 The heating element of the diffusion pump is now switched off and the High Vacuum
Valve is closed. The rotary pump will run for 60 minutes during diffusion pump cool-down.
3 Allow the pump oil to run down to the bottom of the diffusion pump, and the pump to
cool down. This takes about 4 hours.
If the pump oil is hot and air is allowed to enter into the diffusion pump, the oil may burn
and produce fumes, which contaminate the interior of the vacuum chamber. In addition,
the pump oil will be destroyed when it reacts with the air.
Harmful fumes are formed when air is mixed with hot pump oil.
4 Unscrew the filler plug/dipstick at the pump fluid filler neck. Check the oil level on the
dipstick: it should be covered up to about 6 mm from the top of the stick (compare with
specifications in the DIF 250 Operating Instructions).
5 If necessary, slowly refill pump oil. The fill volume for the diffusion pump is 0.5 liter (min.
0.45 liter, max. 1.0 liter).
Note!
The pump fluid spreads only slowly within the diffusion pump. If the oil is warm (not hot),
it flows more easily. The actual pump fluid level can be read about 3 minutes after filling.
If the final pressure (<4,0 x 10-5) cannot be obtained any more (and no leaks can be detected
in the vacuum system), it is time to clean the inside of the diffusion pump.
2 Clean the jet system. See Section 10-9-15-3 Disassembly of diffusion pump on page 322,
with Scotch Brite or steel wool and alcohol to remove all the black material on the parts.
When all the parts are shiny use paper and isopropanol to get rid of all dust. Rinse well
with alcohol.
3 Clean the interior of the diffusion pump with soft paper and alcohol.
4 Clean the interior of the baffle with soft paper and alcohol to remove used-oil droplets
inside.
5 Clean all surfaces exposed to O-rings with soft paper and alcohol. Make sure there is no
dust on the surfaces.
6 Reinstall the diffusion pump. See Section 10-9-15-4 Reassembly of diffusion pump on
page 323.
2 Start the vacuum system, see Section 10-7-4 Start up the MINItrace after vacuum system
maintenance on page 300. Note the time.
3 At the TPG 300, check that backing pressure drops after approx. 2 minutes.
4 Verify that it takes 25–35 min before the diffusion pump is hot (LED switches from flashing
to continuous green).
CAUTION!
Allow the diffusion pump to cool down for a total of 4 hours after shutting down the
vacuum system. Otherwise the pump oil may be oxidized and carbonized on the inside
surfaces of the diffusion pump.
There are two different rotary pump models: DUO 20 (newer model) and DUO 016 (older
model). See Figure 10-8.
DUO 20
Gas ballast
valve Filler cap
WARNING!
The rotary pump is hot (70–80°C) after operation. Use protective cloths to move the pump.
DUO 20
Check and refill operating fluid according to instructions in MINItrace Service Manual – OEM
Manuals (dir. 2233002-100).
DUO 016
1 If needed, move the rotary pump for better access to the sight glasses (see Figure 10-8).
2 Look through the sight glass at the end of the pump compartment. Check the color of
the pump oil against the color scale in the MINItrace Service Manual – OEM Manuals
(dir. 2233002-100). Replace the oil if it is contaminated to a higher degree than 5 on the
scale, see Section 10-9-12 Rotary pump oil on page 317).
3 Check the oil levels in the pump and motor compartments. The sight glasses should be
about 3/4 full.
a Open the filler cap at top of pump compartment and refill pump oil.
b Open the filler cap at top of motor compartment and refill motor oil.
5 Check the condition of the oil mist filter: clean with alcohol if there is oil on the filter, see
Section 10-9-10 Oil mist filter on page 314.
Calibration Cleaning
Pirani 2 gauge (Chamber) See Section 10-8-7-1 Pirani See Section 10-8-7-2 Pirani
2 gauge (chamber) calibra- 2 gauge (chamber) cleaning
tion on page 306 on page 306
Pirani 1 gauge (Backing) See Section 10-8-7-3 Pirani See Section 10-8-7-4 Pirani
1 gauge (backing) calibration 1 gauge (backing) cleaning
on page 307 on page 307
3 Loosen four screws on the front panel of the TPG 300 and carefully pull the unit out of
the rack. Do not stretch the cables which are connected at the rear panel.
4 If necessary adjust the ATM screw at the 2 TPR input on the PI 300 module until the
readout is 1.0E03 ±0.2E03 mbar (1.0E05 ±0.2E05 Pa).
5 Press the PUMP key and verify the pumping sequence. Check also that the Pirani 2 (A2)
counts down when the pressure is falling. This will be seen only when the system is
roughing.
6 When the chamber pressure is in the 10E-6 mbar (10E-4 Pa) range, the reading should
remain at 1.0E-04 mbar (1.0E-02 Pa). If not, adjust the HIGH VAC screw.
1 Remove the Pirani 2 gauge, see Section 10-9-6 Pirani 2 gauge (chamber) on page 311.
d Wash and rinse the interior of the gauge head with alcohol. Repeat twice.
2 At the TPG 300, press sensor to select Pirani 1 readout, A1 (backing pressure).
3 Loosen four screws on the front panel of the TPG 300 and carefully pull the unit out of
the rack. Do not stretch the cables which are connected at the rear panel.
4 If necessary adjust the ATM screw at the 1 TPR input on the PI 300 module until the
readout is 1.0E03 ±0.2E03 mbar (1.0E05 ±0.2E05 Pa).
6 Check also that the Pirani 2 (A2) counts down when the pressure is falling. This will be
seen only when the system is roughing.
7 Monitor the readout of the backing pressure as it slowly drops. Compare to earlier record
that the readout is reasonable (< 2x10-1 mbar). It is not necessary to calibrate high
vacuum.
d Wash and rinse the interior of the gauge head with alcohol. Repeat twice.
3 Remount the Pirani 1 gauge and connect the cable, see Section 10-9-7-2 Insertion
procedure on page 312.
4 Follow the procedure in Section 10-8-7-3 Pirani 1 gauge (backing) calibration on page 307
above to calibrate the Pirani 1 gauge.
Jumps in the pressure readout normally indicate that the anode pin and the ionization chamber
in the Penning gauge are dirty.
1 Remove the Penning gauge (see Section 10-9-8 Penning gauge (chamber) on page 312).
2 Clean the ionizing compartment and the center anode pin with Scotch Brite.
6 No calibration is necessary.
The pump oil in the diffusion pump and the rotary pump will be radioactive after production
runs.
Always wear protective gloves when replacing pump oil. Do not touch eyes or mouth.
Follow federal, state and/or local regulations when disposing of pump oil.
Cleaning tissue, rubber gloves, cotton tops or other items used for cleaning inside the
vacuum chamber are radioactive.
Dispose of these items in accordance with federal, state and/or other local regulations.
Always disconnect the compressed air supply at the cyclotron support cabinet before
removing any air-controlled vacuum components.
CAUTION! Cleanliness
The cleaner all surfaces inside the vacuum chamber are, as well as all other components
that get in contact with the vacuum system, the better the end pressure will be.
Always use rubber or latex gloves when you work with vacuum components.
For Lock-Out and Tag-Out (LOTO) procedures of the equipment, see Section 3-5-6 LOTO
procedures for the MINItrace system on page 59.
• Socket wrench: 8, 10 mm
• Lint-free tissues
• SEAL 10 KF
• SEAL 25 KF
• SEAL 250 KF
• SEAL 40 KF
• O-RING SET
Seals and O-rings should be replaced, if required, before reassembling vacuum components.
CAUTION!
Make sure that the diffusion pump has cooled down before venting the vacuum system.
1 Shut down the vacuum system, see Section 10-7-2 Shutting down the vacuum system
on page 298.
2 Check that the diffusion pump has cooled down (4 hours after shutdown). Touch the
heating element at the bottom of the diffusion pump lightly (see Figure 10-10). The
diffusion pump should not be more than “hand warm” before proceeding.
3 Disconnect the cable from the Pirani 1 gauge by turning the connector counter-clockwise
a quarter of a turn.
4 Carefully loosen the wing nuts on the Pirani 1 gauge mounting clamps, and slightly tilt
the gauge to bleed air into the diffusion pump.
CAUTION!
Bleed the diffusion pump carefully in order to avoid contamination in the vacuum system.
5 Check that the Pirani 1 O-ring is clean and undamaged. If required, clean or replace.
Apply a thin layer of vacuum grease to the O-ring.
2 Disconnect the cable from the Pirani 2 gauge by turning the connector counter-clockwise
a quarter of a turn.
3 Loosen the wing nuts on the mounting clamps of the Pirani 1 gauge and remove the
clamps.
6 Clean the O-ring with alcohol and apply a thin layer of vacuum grease.
2 Calibrate the Pirani 2 gauge, see Section 10-8-7-1 Pirani 2 gauge (chamber) calibration
on page 306.
4 Clean the O-ring with alcohol and apply a thin layer of vacuum grease.
2 Calibrate the Pirani 1 gauge, see Section 10-8-7-3 Pirani 1 gauge (backing) calibration
on page 307.
5 Clean the O-ring with alcohol and apply a thin layer of vacuum grease.
2 Disconnect the compressed air supply, see Section10-9-5 Disconnecting the compressed
air supply on page 311.
3 Remove the compressed air tubing from the valve: push the tube fitting towards the
connector and pull out the tube.
4 Using a wrench, loosen and remove the T-connector (backing valve) or elbow connector
(roughing valve), see illustration to the left.
5 Disconnect the cable from the valve by turning the ring on the connector.
6 Loosen the wing nuts on the mounting clamps of both valve ports.
9 Clean the O-rings with alcohol and apply a thin layer of vacuum grease. The O-rings
must be absolutely free from dust.
4 Connect the compressed air line(s): slide the tube into the fitting until it butts against the
inside shoulder.
10-9-10-1 DUO 20
Refer to MINItrace Service Manual – OEM Manuals (dir. 2233002-100).
Removal procedure
1 Shut down and vent the vacuum system, Section 10-7-2 Shutting down the vacuum
system on page 298 and Section 10-9-4 Venting the vacuum system on page 309.
WARNING!
The diffusion pump and the rotary pump are hot during operation. Allow 4 hours cool-down
time before starting any service work.
3 For access to the oil mist filter, carefully move the pump outward without stretching the
vacuum hoses.
8 Check the oil levels in the pump and motor compartment sight glasses. Both sight glasses
should be 1/2 to 3/4 full. If required, refill with vacuum oil (see Section 10-9-12 Rotary
pump oil on page 317).
Insertion procedure
1 Install the new oil mist filter and attach the clamp.
WARNING!
The diffusion pump and the rotary pump are hot during operation. Allow 4 hours cool-down
time before starting any service work.
The pump oil in the diffusion pump and the rotary pump will be radioactive after production
runs.
Always wear protective gloves when replacing pump oil. Do not touch eyes or mouth.
Follow federal, state and/or local regulations when disposing of pump oil.
2 Disconnect the compressed air supply, see Section10-9-5 Disconnecting the compressed
air supply on page 311.
3 Check that the diffusion pump has cooled down: touch the heating element at the bottom
of the diffusion pump lightly.
4 Remove the diffusion pump, see Section10-9-15 Diffusion pump maintenance on page319.
5 Disconnect the backing line: loosen the clamp at the fore vacuum port and remove the
valve assembly (backing valve, roughing valve, Pirani 1 gauge). Carefully place the
assembly on the floor.
6 Place absorbent material under the drain plug to collect spillage during the draining.
8 Open the drain plug and drain the old pump oil.
9 When finished, pour the oil into, for example, an empty oil container.
10 Dispose of the old oil according to federal, state and/or local regulations for radioactive
waste.
3 Open the filler plug/dipstick at the pump fluid filler neck (or fill through the fore vacuum
port).
5 Slowly refill pump fluid until the dip stick is covered to about 6.5 mm from the top (about
0.5 l).
Note!
The pump fluid spreads only slowly within the diffusion pump. If the oil is warm (not hot),
it flows more easily. The actual pump fluid level can be read about 3 minutes after filling.
6 Wipe any dust or residue from the open end of the cross-piece and the fore vacuum
port.
7 Always replace the two O-rings in the pump fluid filler neck and drain plug, see
Figure 10-13. Check the condition of the other O-rings, replace if necessary.
8 Apply a light coat of vacuum grease to the O-ring, and reconnect the metal hose to the
fore vacuum port.
WARNING!
The diffusion pump and the rotary pump are hot during operation. Allow 4 hours cool-down
time before starting any service work.
The pump oil in the diffusion pump and the rotary pump will be radioactive after production
runs.
Always wear protective gloves when replacing pump oil. Do not touch eyes or mouth.
Follow federal, state and/or local regulations when disposing of pump oil.
10-9-12-1 DUO 20
Replace operating fluid according to instructions in MINItrace Service Manual – OEM Manuals
(dir. 2233002-100).
3 Open the drain plug at the end of the pump and drain the pump (see Figure 10-14).
4 Open the drain plug at the motor and drain the motor.
5 Dispose of the old oil according to federal, state and/or local regulations for radioactive
waste.
Filler cap
Filler cap
Sight glass
Sight glass
Drain plug
Drain plug
2 Open the filler cap on the motor compartment and pour in new pump oil. Fill until the
sight glass appears 3/4 full, about 1.1 l.
3 Open the filler cap on the pump compartment and pour in new pump oil. Fill until the
sight glass appears 3/4 full, about 1.7 l.
10-9-13 O-rings
Worn out or damaged O-rings should be replaced where applicable during disassembly and
reassembly work. An O-ring is acceptable if it is elastic and without cracks. Clean the O-ring
and apply a very thin layer of vacuum grease before reassembly.
CAUTION!
Never use metal tools when you remove an O-ring or seal from an aluminum body, there
can be scratches on the surface.
2 Allow the pump to cool down for 4 hours. This is to protect the oil from oxidation and
carbonization on the surfaces of the inside of the diffusion pump.
WARNING!
CAUTION!
Keep all cabling away from the hot surface of the diffusion pump.
3 In the Vacuum Control Unit (VCU), unscrew the screws for the lower alu-panel and switch
off Circuit Breaker CB3 (diffusion pump) and CB1 (rotary pump). Also turn off the Circuit
breaker CSC on the Control Cabinet.
5 Ventilate the diffusion pump by removing the Pirani 1 gauge (backing), see Section 10-9-7
Pirani 1 gauge (backing) on page 312. Tilt the Pirani gauge slowly to let air into the diffusion
pump while preventing contaminants from entering the pump.
b The temperature sensor using a 3 mm hex key to remove the cover and a 7 mm
wrench to remove the sensor.
7 At the cooling water system, close the valves (marked Vacuum) for cooling water to the
vacuum system: first the PRESSURE line (the upper one), and then the RETURN line (the
lower one).
8 At the diffusion pump, disconnect the piping for cooling water. Use a small bucket, or
similar, for waste-water from the piping.
The diffusion pump weighs 30 kg. Two people are needed when lifting it.
1 Place a roller plate or wooden lumber (thickness about 30 mm) underneath the pump
so it cannot fall down when it is loose.
2 At the air manifold, close the valves for compressed air to the diffusion pump valves.
5 Loosen the pump house from the vacuum chamber in these steps:
a Unscrew the two outer, upper M10-screws and replace them with guide pins (GEPS
part no 909262, CAC206-33014-00-4)
d Drag the unit out from its position, and be careful not to scratch the sealing area.
6 Mark the position of the diffusion pump on the baffle before unscrewing the M10-nuts
holding the diffusion pump.
7 Unscrew the clamps that attach the diffusion pump to the baffle crosswise and with
small steps. Take care about the O-ring between the baffle and the pump.
8 If the pump oil has been spread all over inside the vacuum system, you also have to
separate the baffle from the pump house. Mark the position before unscrewing the nuts.
1 Drain the diffusion pump oil, see Section10-9-11 Diffusion pump oil (Santovac) on page315.
9 Remove the locking spring on the shaft and pull out the shaft.
12 Clean the jet system and the inside of the diffusion pump. Please see Section 10-8-3-3
Diffusion pump cleaning procedure on page 303.
2 Carefully assemble the jet system and mount the diffusion pump in reverse order (see
Section 10-9-15-3 Disassembly of diffusion pump on page 322.
3 Fill the diffusion pump with new pump oil; the fill volume is 0.5 l. See Section 10-9-11-2
Refill the diffusion pump on page 316.
2 Make sure that the marks on the connection between the diffusion pump and the baffle
are in line. Then fasten the clamps crosswise and with small steps.
3 Carefully lift the diffusion pump back into position. Use a roller plate or wooden lumber
(about 30 mm) underneath to reach the correct level.
4 Carefully put the pump assembly back in position. Use the guide pins for positioning the
pump house.
7 Connect the tubing for the compressed air on the diffusion pump. Make sure that the
tubing are correctly mounted.
8 Connect the piping for the cooling water on the diffusion pump.
9 Open the valve at the air manifold for the compressed air to the valves for the diffusion
pump.
10 Open the valves at the water manifold: first the RETURN line (lower), and then the
PRESSURE line (upper) for cooling water to the diffusion pump. Make sure that there is
no water leakage.
11 Connect all the electrical wiring for the diffusion pump, including the temperature sensor.
12 Reinstall the cleaned or new Pirani gauge and connect it, see Section 10-9-6-2 Insertion
procedure on page 312.
13 Switch on the Circuit Breakers, CB3 and CB1, in the Vacuum Control Unit (VCU),
• Teflon glide plate (thickness 8 mm) with grooves for packing and knitted copper tube (RF
seal). The plate should be attached to the surface of the pump house to support when
you turn the DP around the upper left fastener point.
• Support washer to keep the gap between DP housing and vacuum chamber in service.
• DP support: A construction of steel tubes and plates to support the DP in service position.
Note!
All removed parts or waste that is radioactive shall be handled and disposed in accordance
with (I/A/W) local laws and restrictions. Please refer to Appendix B General radiation safety
aspects and Appendix C MINItrace radiation safety aspects for further instructions during
work with radioactive equipment/parts.
10-9-16-4 Dismounting
Note!
All numbers inside brackets/parenthesis refers to pos. balloons on drawing CAC206-31001-04-1,
except for if other reference is stated.
Action Check
3. Dismount the cable duct by the radiation shield and cut the inner
stripes/cable ties that retain the cooling tubes for DP (37).
4. Remove the three upper screws (61) (use the hex wrench supplied
with the spare part, see Figure 10-18, item 1) and install stud (see
Figure 10-18, item 8) with nut (see Figure 10-18, item 5) in the middle
of the upper holes (see Figure 10-19). Verify that the stud (item 8) is
securely fastened.
1: Hex wrench
2: Glide plate
5: Nut
6: Screws
7: Screw
8: Stud
9: Washer
10. Support washer
Action Check
5. Install two screws (see Figure 10-18, item 6) in the two remaining
upper holes (see Figure 10-19). Verify that there is about 15 mm
gap/play between the two screw heads (see Figure 10-19) and DP
housing (29).
Action Check
6. Remove the three lower screws (61) and back off nut on upper stud
until DP housing (29) touches the head of upper screws. See Fig-
ure 10-19.
8. Restrain DP (37) and install glide plate (see Figure 10-18, item 2) from
below and upwards (see Figure 10-20).
CAUTION! Be careful not to damage sealing surface on vacuum
chamber (27), O-ring (6), the contact spring (36) and connecting plate
on DP housing (29). Verify that the three rivet nuts enters the lower
screw holes on DP housing (29) so that glide plate (see Figure 10-
20) is mounted in the right position.
Action Check
9. Install the screw with washer (see Figure 10-18, item 7) in the middle
hole and tighten, then remove the stud (see Figure 10-19) and right
upper screw (see Figure 10-19) to make it possible to turn DP (37)
outwards (see Figure 10-21).
Action Check
10. Prepare the support (see Figure 10-21) by placing it on the floor on
the right side of the DP (37). Carefully turn out the DP (37) and place
the support so that the bended plate leans on the DP (37). The DP
(37) is now in service position (see Figure 10-21). DP (37)
heater(s)/element(s) may now be replaced in accordance with en-
closed INFICON instructions (see Figure 10-22).
(For further information on the diffusion pump, see the Inficon diffusion pump PDI 250-W
manual in MINItrace Service Manual – OEM Manuals (dir. 2233002-100). OEM = Original
Manufacturer Equipment)
10-9-16-5 Mounting
Action Check
11. Lift the DP (37) and remove the support (see Figure 10-21). Carefully
turn the DP (37) to upright position and fasten right upper screws
(see Figure 10-19) and stud (Figure 10-19).
12. Remove the screw (see Figure 10-18, item 7) that retains the glide
plate (see Figure 10-18, item 2). Retain the DP (37) and remove the
glide plate (see Figure 10-20).
14. Install the nut (see Figure 10-19) to the stud (Figure 10-19), and turn
it until the DP housing (29) is flush with the vacuum chamber (27).
Mount the three lower screws (61) and tighten.
15. Remove the two upper screws (see Figure 10-19) and install the
original screws (61). Remove the stud (see Figure 10-19) and nut
(Figure 10-19) and reinstall the last of the original screws (61).
Tighten all six screws to the correct torque.
17. Reinstall the cable duct and install new stripes/cable ties at the
cooling tubes for the DP (37).
18. Wipe the MINItrace clean from possible contamination that might
have occurred during implementation of the spare part. Perform a
production test and return the system to operator/customer.
Note!
Dummy Dee 4 (DD4) is not included in systems with serial number 28791 and higher.
When
the ion source produces negative ions (H-). These ions are picked up and accelerated in orbits,
guided by the magnet field and the radiofrequent electric field. The electric power, the ion
source gas and the cooling water (water from the water cooling system is used to remove
heat from the ion source system) is brought to the ion source by means of the ion source
system assembly, see Figure 11-1.
Figure 11-2 shows the ion source assembly in place inside the vacuum chamber.
11-2 Overview
Do not remove the ion source cable when the ion source is powered up.
WARNING! Radiation
Dust and fragments inside the vacuum chamber can be highly radioactive. The surface
of the vacuum chamber can also be contaminated.
Always use protective clothes, gloves and goggles when working inside the vacuum
chamber.
Hydrogen (H2) gas is explosive at certain mixing proportions with air. No open flames!
In case of unexpected pressure drop in the H2 bottle, always find the cause of the leak.
11-2-2 Specifications
Typical beam current levels on the beam probe (“flip” probe) at the center of the cyclotron
are shown in Table 11-1:
500 30–40
1000 80–110
1500 130–170
2000 165–
The plasma column and the cathodes are put in a closed volume in order to maintain the
gas pressure without loading the cyclotron vacuum system excessively. The only opening to
the cyclotron vacuum system is the so called slit, which is a small opening centered around
the cyclotron mid-plane.
Negative ions will basically only exist in the outer “shell” of the plasma (PIG discharge) and
due to that, a concentration of negative ions will appear in the slit opening. The ions can be
extracted from the ion source by a positive potential. In order to define the edge of the plasma
with respect to the slit, a restriction is included, which defines the outer boundary of the
plasma.
Cathode
Restriction
Slit
H
PSARC + Cyclotron
mid plane
Plasma
Cathode
2
H
The position of each slit defines the starting point for the acceleration of the particles. This
must be carefully determined to match the accelerator parameters (particle, magnet field,
RF field and frequency).
The arc current supplied from the PSARC controls the output beam current from the ion
source. The gas flow into the source effects the beam intensity but is normally set to a fix
value.
Across the ion source a potential drop is developed, caused by electron emission from the
cathode surfaces. In the operating range of the PSARC, the ion source will then act as a
negative load resistance for the PSARC.
The RF voltage presents a positive potential once per period. Ions leave the ion source during
this phase and are accelerated in orbits guided by the magnet field and by the shape of the
central region electrodes. The beam time pattern will thus be pulsed with a pulse frequency
equal to the RF frequency.
Ion source
Note!
Only three dummy dees from systems S/N 28791.
PS/GEMS-02A PS/GEMS-01
HIGH VOLTAGE
SCALE MONITORS
POWER HIGH VOLTAGE
ON
2 AMP/2 kV VOLTAGE
ON ON
POWER
PS/PK003R1.3
ENABLE
POLARITY METER DAMPING TREND METER DAMPING
ON
POS NEG
HIGH VOLTAGE
STANDBY
800 290
REMOTE
OFF / RESET PROGRAMMING MILLIAMPERES KILOVOLTS
INTERLOCK ENABLE
CONTROL CONTROL
I mA
O EXTERNAL kV
POWER
For information on model PS/GEMS-02A, see Section11-4-1 Model PS/GEMS–02A on page 337.
For information on model PS/GEMS-01, see Section 11-4-2 Model PS/GEMS–01 on page 339.
The high voltage connector (J1) supplies lethal voltage of up to 3000 V DC, even when
power is off.
Do not attempt to make or remove any high voltage connections to the PSARC until the
load and the PSARC have been discharged (grounded). It may take up to 60 seconds to
discharge the PSARC if it is unloaded.
11-4-1-1 Specifications
Table 11-2: PSARC specifications
Current stability 1%
ON ON
For details on the PSARC refer to the Glassman instruction manual PS/GEMS-02A.
TB1 POWER ON
J3
1
2
J1 3
4 F1
J2
F2
E1
GROUND
Connection Connected to
J3 GCU
The high voltage connector (J1) supplies lethal voltage of up to 3000 V DC, even when
power is off.
Do not attempt to make or remove any high voltage connections to the PSARC until the
load and the PSARC have been discharged (grounded). It may take up to 60 seconds to
discharge the PSARC if it is unloaded.
11-4-2-1 Specifications
Table 11-5: PSARC specifications
Current stability 1%
HIGH VOLTAGE
SCALE MONITORS
ON
2 AMP/2 kV VOLTAGE
POWER
For details on the PSARC refer to the Glassman instruction manual PS/GEMS-01.
Power On
TB1 F1
J3
J1 J2
1
2
3 F2
4
GROUND
Connection Connected to
J3 GCU
Figure 11-10: Ion Source Power Supply (PSARC) front and rear view
FRONT VIEW
REAR VIEW J1
JHV1
TB2
1
2
3
TB1 F1
1 Power On
2 F2 Indicator
3
4
5 F3
E1
The JHV1 connector supplies lethal voltage of up to 3000 V DC, even when power is off.
Do not attempt to make or remove any high voltage connections to the PSARC until the
load and the PSARC have been discharged (grounded). It may take up to 60 seconds to
discharge the PSARC if it is unloaded.
11-4-3-3 Specifications
Constant current control, featuring
Connection Connected to
E1 Protective ground
F1, F2, F3 20 A, 250 V slow blow 208 V Breaks the input power to PSARC.
There is no need of gas flow adjustment but change of the pressure of the H2 gas supply will
allow for changes if that by any reason will be of interest.
The gas control valve is put in the secondary cooling water system section in the right side
cabinet of the radiation shield
The H2 gas supply should be furnished with valves in order to be able to close the supply
during weekends and other longer stops (customer responsibility)
11-5-2 Specifications
Nominal H2 gas supply pressure: 6 bar. This corresponds to a gas flow of about 6 ml/min (at
normal pressure).
11-6 Pipes
The pipes from the gas bottles are made of stainless steel.
Keep track of the H2 usage in order to be sure of a tight system. Possible leaks in the H2 supply
must be found ASAP.
11-8-1 General
The expected planned maintenance interval is at least 3 months, but large variations in system
usage will probably call for a flexibility in maintenance planning.
It is important to note that an ion source may fail suddenly. This is normally related to loose
short circuiting fragments inside the ion source. The problem can normally be cured by a
surge of current into the ion source, burning away the shorting fragment.
The normal deterioration of the ion source refers to cathode wear and widening of the anode
tube slit. The heads of the cathodes (Ta) will loose material during the operation due to
bombardment of ions from the plasma. The lost material will “sputter” onto adjacent surfaces.
The ion source works as long as the cathodes are in place and the cathode heads “cover”
the plasma extension. However, massive loss of cathode material will, depending on the
build-up of sputtering material, slowly change the internal dimensions of the ion source. This
can lead to a drop in output. The risk of short circuits also increases.
The concentration of ions in the anode tube slit also leads to a wear of the slit sides, resulting
in a slow widening of the slit opening. The larger area of the slit results in a lower gas pressure
inside the ion source with a drop in output as a result.
• Functional check
For Lock-Out and Tag-Out (LOTO) procedures of the equipment, see Section 3-5-6 LOTO
procedures for the MINItrace system on page 59.
11-9-1 Drawings
D1 = Dee-electrode 1 (“puller”)
DEE 1
D2 = Dee-electrode 2 (“Dee post”)
DD1 = Dummy Dee 1
DD2 = Dummy Dee 2
DD3 = Dummy Dee 3
DD4 = Dummy Dee 4*
DD1 IS = Ion Source
D1 DD4
DD2 D2
DD3
DEE 2
Serial numbers
28773–
–28772 28790 28791– Edge of lower
Pos. [mm] Note Dee connector
A 1.5 1.3
B 2.0 1.5
C IS K
3.0 2.7 D2 centred on lower Dee connector
D 3.0 2.7 D2 centred on lower Dee connector (Ion Source tool)
IS tube slit opening centered on IS tool CAC203-33003
E
marking Dee edge
ref. pin hole
F 13.2 13.0 Center pin hole (Ø 1.0 mm)
G 20.5 20.0 of cyclotron
H (29.3) (30.9) (Ø 3.0 mm)
I (30.2) (31.6) E
J 20.6 21.4 21.3 IS
Dee edge ref. pin in contact with lower
K
connector
L 4.0 N/A
M 28.8 30.0 30.6
N 1.0 N/A
Flush to sector
L edge*
Flush to L DD4*
sector edge A
D1
DD1
B
IS
F N
Screen
Center pin hole of
G J cyclotron (Ø 3.0 mm)
M
C D
DD2 DD3
(H) (I)
D2
Flush to Flush to
sector edge sector edge * Only applicable to systems up to
and including S/N 28790
Note!
DD4 removed from system S/N 28791.
11-9-2 Tools
• Gloves
• Glasses
• Feeler gauges for the different gaps (drills with suitable diameters can be used)
• Calipers
• Magnifying glass
• Gamma monitor
• If the ion source is seriously damaged, you need a complete ion source assembly.
For spare part numbers, refer to the spare parts listed for each sub systems in MINItrace
Spare Parts Catalog (dir. 2380006-100).
11-9-4-1 Shut down the cyclotron and open the vacuum chamber
Action Check
1. In the MSS MAIN MENU, click on the center of the cyclotron to access
the Ion Source and Gas Handling System menu.
Action Check
Action Check
Note!
The vacuum pumps continues to run.
Note!
The vacuum pumps continues to run.
VV HVV
DP RV
PIR 1
BV
RP
7. Disconnect the services to the ion source on the outside of the vacuum
chamber on the lower right hand side (two water quick connections, gas
quick connection and arc voltage connector).
CAUTION!
Keep disconnected tubing and cabling away from the diffusion pump.
8. Open the vacuum chamber, see Section 10-7-1 Open the vacuum
chamber on page 295.
WARNING! Radiation
Always monitor radiation levels and take all relevant safety precautions to limit the dose
delivered to personnel before starting any work.
Before the ion source can be removed, some of the elements in the central region must be
removed or adjusted in position in order to make necessary space for the removal of the ion
source.
Note!
Numbers printed in bold type refer to item numbers in Figure 11-13 (part of drawing
CAC203-31001) if not otherwise noted.
Action Check
Dwg CAC203-31001-09-1
(detail)
Dwg CAC203-31001-09-1
(detail)
3. Carefully remove the puller electrode 8 and put aside for further
inspection. Do not drop!
Dwg CAC203-31001-09-1
(detail)
Action Check
Note!
DD4 removed from system S/N 28791.
Dwg CAC203-31001-09-1
(detail)
Note!
Handle the sealing surface on the inside of the vacuum chamber
carefully. Avoid scratches.
Handle the ion source assembly with great care. Do not drop!
Dwg CAC203-31002-08-1
(detail)
Note!
Numbers printed in bold type refer to Figure 11-14 (part of drawing CAC203-31002) if not
otherwise noted.
Action Check
CAUTION!
Always use gloves and perform maintenance on a clean work bench with good light con-
ditions.
3. Remove anode tube 42 with internal anode restrictions and put aside for
later inspection.
4. Disconnect the two cable shoes 2 by loosening the fixing screws 36.
Action Check
3. Check the slit of the anode tube 42 with feeler gauge and
magnifying glass. If the opening size is <0.35 mm and the an-
ode tube restrictions are in good shape the anode tube can
be reused.
Action Check
Dwg CAC203-31001-09-1
(detail)
Note!
Always replace parts if you have any doubts about their condi-
tion. Reuse of parts is not recommended but could be a tempo-
rary method to quickly restore operation of the cyclotron.
Action Check
Do not scratch the surfaces. They have a vital function for the
gas tightness of the ion source.
Dwg CAC203-31002-08-1
(detail)
Note!
Wet cleaning with solvent is not necessary if the surfaces are
clean. Inspection with magnifying glass will be useful.
11-9-5-4 Reassembly of the ion source (to end ion source overhaul)
Note!
Numbers printed in bold type refer to Figure 11-14 (part of drawing CAC203-31002) if not
otherwise noted.
CAUTION!
Always use gloves and keep new parts clean and free from grease
Carefully inspect the new components. Do not use components that by any reason are not
in perfect condition.
Action Check
2. Move the cathode assembly back and forth in the upper anode
section 16 in order to center the assembly by use of the recess
for the insulator 20.
3. When fully sure about the centering, pull the screw 18 tight
with cathode 13 and el. connection 19 centered in their groves.
Make screw 18 tight until the electrical connection 19 is fixed
and cannot easily be moved by hand. Check insulator 20
condition afterwards for eventual cracks.
Note!
If you hear a “click-sound” during the actions above there is a
great risk that an insulator has been cracked. Disassembly will
be necessary to ensure conditions of insulators.
Action Check
Note!
Check carefully that connecting copper surfaces are clean.
Note!
Check that electrical connections 19 will not move during this
action, in order to avoid accidental short circuit to ground.
10. Clean the O-ring on the vacuum flange (Item 1, dwg CAC203-
31001, shown in Figure 11-13) and apply a thin coating of
vacuum grease if necessary.
Dwg CAC203-31001-09-1
(detail)
The ion source is now ready for reinstallation into the cyclotron.
Note!
Numbers printed in bold type refer to item numbers in Figure 11-13 (part of drawing
CAC203-31001) if not otherwise noted.
Action Check
Handle the ion source assembly with great care. Do not drop!
2.
Note!
Do not lose the O-ring.
Note!
Handle the sealing surface on the inside of the vacuum chamber
carefully. Avoid scratches.
Action Check
Make sure that the wedge on the top plate goes underneath
the flange on the slot in the middle plate.
Tighten (by hand) the knurled screw (that will pull the ion source
toward the puller) until the stop screw touches the middle
plate.
Tighten the three screws in the rear of the ion source that go
into the vacuum chamber. See Figure 11-12.
The ion source should now be in the same position as before
the removal.
Tighten the three screws in the rear of the ion source that go
into the vacuum chamber. See Figure 11-12.
4.
Note!
Actions 5–17 below and the actions in Section 11-9-7 Upper
Dee connector assembly on page 364 are the most important
adjustments of the central region elements. Failure to meet the
requirements will most probably lead to reduced beam perfor-
mance of the cyclotron.
Insert the center ref. pin 15 (diam. 3.0 mm) through the center
hole in the Dee connector 5, and fully into the reference hole
in the magnet pole.
Action Check
5. Insert the ion source tool (see Figure 11-15, drawing CAC203-
33003) carefully on the center ref. pin 15.
1 Push the tool carefully against the lower Dee connector 5.
2 Make sure that the 1 mm Dee edge ref. pin on the tool
can rest on the edge of the lower Dee connector.
Dwg CAC203-33003-02-3
(detail)
3 With Ion Source Adjustment Tool: Use mild force to push
the tool into position. If the tool does not fit, loosen the
knurled screw and adjust with the stop screw.
Action Check
6. Adjust carefully the angle of the ion source slit surface and the
position of the ion source, until the slit is centered on the middle
reference mark on the tool, and the slit surface is parallel to
and touching the reference surface of the ion source tool (see
Figure 11-12).
7. Tighten the three ion source clamping screws for angle adjust-
ment (item 23, 2 pcs and item 34, drawing CAC203-31002) and
tighten the knurled screw finger-tight (systems with Ion Source
Adjustment Tool) or the ion source mounting screws 16 (sys-
tems without Ion Source Adjustment Tool) carefully to position
slit and ion source tool flush.
Make sure that the ion source tool is not exposed to excessive
force during these actions.
8. Make a final check of the radial position and angle of the ion
source slit.
9. Remove the ion source tool and center ref. pin 15.
10. Install the puller electrode D1 8 using the screw and the
washer – do not tighten.
11. Carefully insert the center ref. pin 15t hrough the ref. hole on
the puller electrode 8 and further into the magnet pole after
passage of the inner Dee connector 5.
12. Move the puller electrode 8 carefully towards the slit surface;
the puller electrode 8 will be guided by the center ref. pin 15.
13. Set the distance A (see Figure 11-12) between the puller D1
and the slit surface by inserting a suitable cylindrical pin or a
feeler gauge. Lock in position by fixing the screw.
16. Make sure that the three (four) dummy Dee electrodes are
positioned according to measures (see Figure 11-12).
Readjust if necessary.
17. Remove the center ref. pin 15 carefully – always use a screw-
driver to create a mild counter force on the base section with
the ref. hole on the puller electrode 8 when retracting the
center ref. pin 15.
Note!
Numbers printed in bold type refer to item numbers in Figure 11-13 (part of drawing
CAC203-31001).
Action Check
Note!
These three parts have probably not been disassembled.
Note!
It is important that the screen 25 edges are flush with or lower
than the electrodes DD1 11 and DD2 13 to form a free passage
for the ions.
Action Check
Action Check
Action Check
Note!
A new ion source or a used ion source which has been exposed to air will
normally provide a higher beam current. This “bonus current” will unfor-
tunately disappear after a couple of hours.
Note!
It is important to know this effect in order to not be fooled on the final
performance after leaving the premises. A “bonus current” of +20% can
be expected.
Note!
Always keep the slit surface parallel to the tool surface.
12 MAGNET SYSTEM
12-1-1 Introduction
The magnet system generates a magnetic field that retains the ion beam in a controlled orbit
during acceleration. The magnet system also focuses the beam.
The magnetic field is generated by the PSMC providing current for the two water cooled
copper coils mounted on poles. The magnet is mounted so that the beam plane is vertical.
One of the side yokes is supported on hinges and can easily be opened for access to the
interior of the system.
Lock-out and tag-out the PSMC circuit breaker at the control cabinet before servicing
the magnet system.
Lock-out and tag-out the PSMC circuit breaker at the control cabinet before servicing
the magnet system.
Physical injury can be caused by tools that are attracted to the cyclotron magnet.
Use non-magnetic tools when working in the vicinity of the cyclotron magnet.
The magnetic field of the cyclotron magnet can interfere with physiological implants such
as pacemakers.
Persons having physiological implants should not enter the cyclotron room.
The magnetic field of the cyclotron magnet can erase the information contained on for
example credit cards or other sensitive equipment.
12-2-1 General
The magnet system is divided into two parts, the magnet coil and the magnet power supply
(PSMC).
The magnet yoke (vacuum chamber) can be opened as a door. The heavy doors movement
is damped by a adjustable hydraulic damper. An interlock switch indicates if the door is open.
The copper coils and the yoke are water-cooled.
The PSMC delivers up to 150 A of current to the main coils and is made by Danfysik. Normally,
the PSMC is remote controlled via a serial line from the General Control System (GCS). The
PSMC is water-cooled.
• Check and adjust the hydraulic door damper. Check the function of the interlock switch.
Look for water leeks.
For Lock-Out and Tag-Out (LOTO) procedures of the equipment, see Section 3-5-6 LOTO
procedures for the MINItrace system on page 59.
• Gloves
There are two set-screws on the damper which are locked by small hex screws. Check the
movement and adjust if needed.
2 Check for damage and leaks on tubings, specially in the area around the hot diffusion
pump.
4 If any of the pipes needs to be replaced, use Swagelok 3/8" (9.5 mm) WP 300 PSIG hose.
The complete PSMC is a spare part. There is no need for replacement of parts inside the PSMC.
There are two different PSMC models: MPS 883 (older model) and MPS 9700 (newer model).
See Figure 12-8.
MPS 9700
3
1
4
2
1 PSMC control unit, front view 3 PSMC control unit, rear view
2 PSMC power unit, front view 4 PSMC power unit, rear view
MPS 883
MPS 9700
The PSMC control unit weighs about 17 kg. Be careful when you move the PSMC control
unit out of and into position.
The PSMC power unit weighs about 22 kg. Two persons, or suitable lifting devices, are
required to lift the PSMC power unit.
3 Turn off the water valves for the PSMC in the Cyclotron Support Cabinet. See Figure 12-5.
4 Lock-out and tag-out the circuit breaker in the Mains Distribution Panel (MDP) for the
Control Cabinet.
5 Disconnect the 400V MAINS cable at the rear of the PSMC control unit.
6 Disconnect the rest of the cables at the rear of the PSMC control unit.
7 Loosen the screws on the front panel of the PSMC control unit.
8 Carefully pull out the PSMC control unit from the cabinet.
9 Remove the plastic cover at the rear of the PSMC power unit.
10 Measure with a multimeter on the magnet output cables. Make sure that there is no
residual current in the PSMC.
12 Disconnect and remove all cables from the rear of the PSMC power unit.
13 Put a bucket on the cabinet floor below the PSMC water connections to catch any water
spillage.
14 Disconnect the water INLET and OUTLET hoses from the rear of the PSMC power unit.
15 Loosen the screws on the front panel of the PSMC power unit.
16 Carefully pull out the PSMC power unit from the cabinet.
MPS 883
3 Turn off the water valves for the PSMC in the Cyclotron Support Cabinet. See Figure 12-5.
4 Lock-out and tag-out the circuit breaker in the Mains Distribution Panel (MDP) for the
Control Cabinet.
5 Remove the cover plate on the front, below the PSMC. The cover plate can interfere with
the water connectors.
7 Remove the plastic cover and disconnect the magnet output connectors. Remember
the locations of the magnet output connectors.
8 Disconnect the PSMC power input connector and the interface connectors.
9 Put a table, or similar, below the front of the PSMC. The PSMC can fall down when you
pull it out from the cabinet.
10 Loosen the screws on the front panel and pull out the PSMC. Be careful, the PSMC is
heavy.
MPS 9700
1 Carefully put the PSMC power unit into the cabinet and push the PSMC power unit into
position.
2 Tighten the four screws on the front panel of the PSMC power unit.
3 Carefully put the PSMC control unit into the cabinet and push the PSMC control unit into
position.
4 Tighten the four screws on the front panel of the PSMC control unit.
5 Connect the water INLET and OUTLET hoses to the rear of the PSMC power unit.
6 Connect all signal cables at the rear of the PSMC units according to the marking.
Note!
The connections ANALOGUE CONTROL X9 and LOCAL CONTROL X10 are not used.
8 Connect the magnet output cables to the PSMC power unit. Make sure the connections
are securely fastened.
11 Turn on the water valves for the PSMC in the Cyclotron Support Cabinet.
12 Turn on the circuit breaker in the Mains Distribution Panel (MDP) for the Control Cabinet.
MPS 883
1 Lift the PSMC in position on a table, or similar, in front of the control cabinet and push it
into the cabinet on its rails. Be careful, the PSMC is heavy.
4 Connect the magnet output connectors. It is important that the hex screws are
well-tighten.
6 Turn on the water valves for the PSMC in the Cyclotron Support Cabinet.
7 Turn on the circuit breaker in the Mains Distribution Panel (MDP) for the Control Cabinet.
1 Remove the PSMC model 883 (see Section 12-2-3-1 Removal of the PSMC on page 377).
5 Install the new slide rails in the CCAB with the screws.
9 Install the PSMC model 9700 (see Section 12-2-3-2 Installation of the PSMC on page 379).
3 Use a hex key and check that the connectors are tight. There should not be any indication
that the cables are loose.
MPS 9700
• The POWER ON LED is on when the magnet is on and everything is normal. The Master
System or the Service System controls the PSMC.
• If the FAULT LED is on, the magnet is off and cannot start up because of an interlock.
If some kind of magnet start-up failure appears during normal operation from the Master
System, change to the Service System. From the Service System, you have more possibilities
to check status and interlocks. To get more information about the internal status and interlocks
of the PSMC, use the PSMC 9700 Service software (available for GE Healthcare internal use
only, refer to service note DOC1974785).
For more information on how to operate the PSMC, refer to the OEM manual DANFYSIK Magnet
Power supply 9700.
MPS 883
• The REMOTE LED and the ON LED are on when the magnet is on and everything is normal.
The Master System or the Service System controls the PSMC.
• If the STANDBY LED is on, the ON LED is off. This mean that the magnet is off and standby.
• If the INTERLOCK LED is on, the magnet is off and cannot start up because of an error.
If some kind of magnet start-up failure appears during normal operation from the Master
System, change to the Service System. From the Service System, you have more possibilities
to check status and interlocks. If it is also impossible to run the magnet from the Service
System, it is possible to start up the magnet system from the PSMC front panel. Do as follows:
2 Press the SET key to enter values for the magnet current.
3 Press the ON key and the START key to start the magnet ramp up procedure.
4 Follow the reading on the PSMC display and verify that the magnet current rises.
For more information on how to operate the PSMC, refer to the OEM manual DANFYSIK Magnet
Power supply 883, System 8800.
For information on how to operate the MINItrace Service System, see Section 19-7-2 Magnet
system on page 617.
MPS 883, System 8800 Magnet Power Supply Danfysik Service/Operating manual
13 RF SYSTEM
There is no control panel fitted to the RFPG. It is designed exclusively for control via the GCU
(General Control Unit). The GCU also handles the RF system interlocks regarding functions
external of the RFPG, and interlocking with other subsystems. RFPG internal interlocks are
handled by the Source and Control Unit (SCU) described further.
Water cooling is provided by the cooling system in the Cyclotron Support Cabinet.
Always lock-out and tag-out mains supply to the RFPG before opening any covers.
If the RF system is operated without all covers and doors in position, or if the RF cable is
not properly connected, RF power might be emitted from the RF system.
Before operating the RF system, check that all screws are fitted on covers and panels,
that all locking screws are fitted at the terminals of the RF feeder cable, and that the
magnet door of the cyclotron is closed.
13 RF system 385
13-1 General description
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
13-2 Overview
The Radio Frequency System comprises:
• The RF cable.
• To pull the H- ions out of the Ion Source in the center of the vacuum chamber.
• To accelerate these H- ions to a high speed as they orbit (with increasing radius) inside
the vacuum chamber.
386 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
To obtain this, a considerable amount of electric power (10 kW at 100.5–101.5 MHz) has to
be fed into the vacuum chamber, and applied to the so-called Dees (there are two Dees,
Figure 13-1). This is accomplished as follows:
• The power is transmitted to the cyclotron by means of the RF cable (a coaxial transmission
cable).
• The power is consumed by the (resonant) load inside the vacuum chamber (the two
Dees).
Figure 13-2 shows the location of these sub-assemblies and Table 13-1 summarizes their
functions.
RFPG
13 RF system 387
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
TAU Final RF power amplifier which delivers its output power to the RF
(and DPA) cavity via the RF feeder cable. The TAU is driven from the DPA.
GSPU Houses power supplies for grid and screen voltages to the tube
within the TAU. Contains control relays for all supplies to the tube.
TPSU Houses power supply for the anode voltage to the tube within the
TAU. Mains power is connected to terminals in the TPSU.
388 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Note!
MINItrace RFPG systems equipped with SCU Gen I can be upgraded with SCU Gen II (see
Section 13-6-12 Replacing SCU Gen I with SCU Gen II on page 478).
The front panel of the SCU must be closed during operation since it contributes to improve
the shielding.
The system controller and the Interface Boards handles only digital signals, whereas the Tube
Controller handles both digital and analog signals.
The SCU includes indicators that aid trouble-shooting and service: LEDs on some of the boards
show status and the Measurement Module shows voltages, currents and power levels within
the SCU.
Note the following for SCU Gen I units in MINItrace RFPG systems manufactured ...
13 RF system 389
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Note!
MINItrace RFPG systems equipped with SCU Gen I can be upgraded with SCU Gen II.
For replacement instructions concerning SCU Gen II, see the following sections:
• Section 13-6-12 Replacing SCU Gen I with SCU Gen II on page 478
390 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Figure 13-6 is a simplified description of the main function sections in the SCU including front
and rear panel interfaces. Table 13-3 contains a brief description of the main functions in the
SCU.
13 RF system 391
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Function Description
RF Detector Buffers and converts RF signals from the RF Cavity and the TAU to dc signals for
feedback and monitoring.
The Load Phase Detector monitors the tuning status of the RF cavity and generates
an output signal which is used by the GCU to tune the RF cavity into resonance.
System controller The Embedded Technology Extended (ETX) module is the main controller of the RFPG.
It controls the different modes of RF operation and provides communication with the
SCU via Ethernet, USB and RS-232 interfaces.
I/O manager The Field Programmable Gate Array (FPGA) module handles logic functions for digital
I/O and analog I/O, both internally and to external units. Controls RFPG power supplies.
AD and DA converters Interfaces digital signals from/to the FPGA to/from other parts in the SCU, the GCU
and other units in the RFPG.
For pin assignments and signal names, see Section 13-2-4-9 Interconnection signals data on
page 431.
392 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
SKC TAU/P6
PLD/SKD PS/TB2
SKE TPSU/TB2
SKF GSPU/SKB
SKG GCU/P9
SKH GCU/P10
SKK TAU/P2
SKN TAU/P8
SKP TAU/P7
SKQ TAU/P9
SKR TAU/P10
13 RF system 393
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
TP9 GROUND —
TP22 N/A —
TP23 N/A —
394 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
TP26 GROUND —
TP27 GROUND —
TP29 GROUND —
TP30 GROUND —
The push buttons are used to navigate in the menus using the under-line cursor and execute
commands (see Figure 13-9 and Table 13-7).
Green light on the top LED only. Power to the SCU is turned on.
No light on the other three LEDs.
Red light on all LEDs An error has been detected by the SCU.
The error message is shown on the display.
13 RF system 395
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
396 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
IP restored [<]
13 RF system 397
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
For more information on the I/O signals, see Section Analog In and Analog In (raw) pages on page 405 to Section Dig-
ital Out page on page 412.
For more information on IP and other addresses, see Section SCU Settings page on page 420.
For more information on log files, see Section Log page on page 414.
The web interface has a main page and a number of sub pages. It has three user access
levels providing different functions. When starting the web interface, the ALL USERS access
level is entered automatically. The SERVICE access level provides more functions and is
intended for service purposes. The login password for this level can be requested from GE
Online Center. The ENGINEERING access level is intended for manufacturing purposes only
and is not described in this manual.
Functions overview
Functions on the ALL USERS access level (common for all levels):
• Monitor analog in, digital in, analog out, digital out and interlock signals.
398 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Additional functions on the SERVICE access level (further to the ALL USERS level):
• Change IP addresses, serial numbers, time and reset the hour counter.
The subsequent sections show the functions and the sub pages as they appear on the SERVICE
access level.
If the SCU is not connected to the Ethernet network or if there is no such network (coaxial
cables are used instead), the laptop can be connected directly to the rear of the SCU.
1 Remove the rear cover required to get access to the network switch (CCAB door) or the
SCU (upper right-hand cover).
2 Connect the Ethernet port on the laptop to either the SCU or the switch as follows:
• To the switch using a straight-thru RJ45 cable. Make sure that the SCU is connected
to the network.
• To the LAN socket at the rear of the SCU using a twisted RJ45 cable.
2 Switch on (push upwards) the RFPG circuit breaker at the rear of the CCAB. The SCU and
RFPG starts up which takes about 40–60 s.
3 Check that the Main menu appears on the front display on the SCU (see Figure 13-9).
4 Make sure that the IP address stored in the SCU is 192.0.4.71 by selecting More > IP
address > Read IP on the front display.
Note!
192.0.4.71 is the default address. It can be changed if required for matching another local
network address configuration.
13 RF system 399
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
The Main page in the SCU web interface appears in a few seconds.
If the Main page does not appear, the configuration of network parameters on the laptop
might be wrong. To correct it, see Section Setting the TCP/IP address and LAN parameters in
the PC on page 400.
Note!
Microsoft™ Windows™ XP and Internet Explorer are required in the procedures below.
The dialog examples below are from Windows XP Professional version 2002 SP 2. Other
versions might differ.
400 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
1 In Windows, select Start > Control Panel > Network Connections. Double-click the Local
Area Connections # that is being used.
Alternatively, left-click the network symbol on the right at the bottom of the screen.
1. Click Properties 2. Select Internet Protocol (TCP/IP) 3. Select Use the ... and type
and click Properties 192.0.4.72 and 255.255.255.0.
Click OK.
13 RF system 401
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
If the SCU web interface still does not appear, check the communication to the SCU by pinging
as follows:
If it returns “Request timed out”, the communication is not OK. Check the cable
connections.
402 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Main page
Figure 13-13 shows the Main page when the SCU status is OFF. Figure 13-14 shows status
STANDBY READY and Figure 13-15 shows DEE VOLTAGE OK.
Note!
All other examples on the web interface in this section show how the pages appear when
logged in as SERVICE and the SCU status is DEE VOLTAGE OK, unless stated otherwise.
The menu panel on the left is displayed on all sub pages and makes it easy to navigate through
the listed pages. It is also possible to change access level in this panel.
The upper panel contains refresh rate options. It appears on pages that monitor parameters
values.
The large right panel contains digital indicators that show interlock status and analog
parameters. The interlocks must all be turned green to make it possible to run the RFPG. If
any of them are dark, the cause must be investigated and rectified before the RFPG can be
run.
The analog parameters shown in this panel are described in Section Analog In and Analog In
(raw) pages on page 405.
The lower panel is shown on all sub pages. It shows general information on the SCU and the
system status. The status can be OFF, “Any interlock”, WARMING UP, START, TUNE, DEE VOLT
OK, SPARK, FIX, PULSE or STANDBY READY.
The four indicators on the right in the lower panel monitors the limits for the Spark, Anode
Current, Reflected Power and Screen Current signals. Dark indicator means that a signal is
within the limit. If the indicator is yellow, the signal has reached the limit and will cause a
reduction of the RF output power.
13 RF system 403
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
404 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
RF Inhibit 1 and 2 Normal indication for RF operation RF operation is inhibited by safety relays
within the Control Interface Unit (CIU).
RFPG cannot be set to STANDBY.
Waterflow OK Normal indication for RFPG in STANDBY TAU cooling water fault detected. (Latched
READY or any mode of RF operation interlock. RFPG set to OFF.)
Earth Stick OK Normal indication for RFPG in STANDBY TPSU cover open, TPSU earth stick removed,
READY or any mode of RF operation. or TPSU thermal switch tripped. (Latched in-
terlock. RFPG set to OFF.)
Door OK Normal indication for RFPG in STANDBY TAU door open. (Latched interlock. RFPG set
READY or any mode of RF operation. to OFF.)
ALL PS OK All SCU power supplies are OK One or several power supplies in the SCU
have failed to deliver the required power.
The Analog In (raw) page shows the signals before the conversion and are values in the range
-4096 – +4096.
Figure 13-16 shows an example from running the system in status DEE VOLTAGE OK.
13 RF system 405
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Signal Description
Fwd Pwr S Forward power compensation scale for anode current limiting
Refl OT Log Over-tone amplitude from reflected coupler (+30 dBm = 10 Vpeak RF into SKR)
Fwd OT Log Over-tone amplitude from forward coupler (+30 dBm = 10 Vpeak RF into SKQ)
Fwd Log Amplitude from forward coupler (+30 dBm = 10 Vpeak RF into SKQ)
Refl Log Amplitude from reflected coupler (+30 dBm = 10 Vpeak RF into SKR)
406 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Signal Description
Lim Sum Turns positive if any of Lim Refl, Lim Scrn or Lim Anode are negative
Other Mod Input RF modulator input voltage (proportional to SKC voltage). Maximum is normally
7–8 V.
Extra Input Volt meter for INPUT banana jacks on rear panel (max. ±10 V)
Relative humidi- SCU internal relative humidity. Normally about half the room humidity due to higher
ty internal temperature.
13 RF system 407
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Digital In page
The Digital In page shows the status of the SCU digital input signals. The signals shown
underneath DC and PS are interlocks. The RF system can only run normal operation if all
interlock indicators are turned green.
The indicators can be either green or dark. Their indications are described in Table 13-10.
Fix Mode GCU request for RF FIX mode FIX mode is not requested
Pulse Mode GCU request for RF PULSE mode PULSE mode is not requested
Normal Mode GCU request for RF NORMAL mode NORMAL mode is not requested
408 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Earth Stick OK Safety switches within TPSU OK, Either TPSU cover open, TPSU earth stick
Normal indication for RFPG in STANDBY removed, or TPSU thermal switch tripped.
READY or any mode of RF operation. (Latched interlock. RFPG set to OFF.)
Waterflow OK Cooling water flow (TAU) OK. TAU cooling water fault detected.
Normal indication for RFPG in STANDBY (Latched interlock. RFPG set to OFF.)
READY or any mode of RF operation.
Airflow OK TAU cooling fan operating OK. Either TAU cooling fan is OFF or not
Normal indication for RFPG in OFF, working as expected, or TAU door open.
STANDBY READY or any mode of RF oper- (Latched interlock. RFPG set to OFF.)
ation.
+24V OK External +24 V power supply OK External +24 V power delivery failure
ALL PS OK All SCU power supplies OK One or several SCU power supplies have
failed to deliver the required power
13 RF system 409
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Signal Description
Dee DVR1 Sign Dee voltage 1 read. Output 0–10 V to GCU (10 V = 4096).
Multi Multi Fwd Internal multiplier for forward power compensation for anode current limiting
Multi Refl Internal multiplier for reflected power compensation for anode current limiting. Not
used.
410 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Signal Description
Other Phase Load Load tuning phase error. Output ±10 V to GCU (-10 V = -4096, 10 V = 4096).
DPA Fwd DPA Forward power. Output 0–10 V to GCU (10 V = 4096).
Dee Comp Internally connected to comparator for limiting Dee pulses (10 V = 4096)
Dee Lim Internally connected to modulator for limiting Dee pulses (10 V = 4096)
Spark Comp Internally connected to comparator for spark detection (10 V = 4096)
13 RF system 411
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
The indicators can be either green or dark. Their indications are described in Table 13-12.
Mode Standby Ready RF is in STANDBY READY. (Signal to GCU). RF is not in STANDBY READY
Dee Volts OK In NORMAL mode, Dee voltage is above Condition not fulfilled
the limit set by Dee voltage OK configura-
tion. (Signal to GCU).
Start Mode In NORMAL mode, Dee voltage is below Condition not fulfilled
the limit set by Dee voltage Detected
configuration. Pulsed RF power. (Signal
to GCU).
Tuning Mode In NORMAL mode, Dee voltage is above Condition not fulfilled
the limit set by Dee voltage Detected
configuration and not Dee Volts OK.
Continuous RF power. (Signal to GCU).
412 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
DC AFN TAU cooling fan operating OK. (Signal to Air flow interlock detected.
GCU). (Latched interlock. RFPG set to OFF.)
WFN Cooling water operating OK. (Signal to Water flow interlock detected.
GCU). (Latched interlock. RFPG set to OFF.)
ESOK Safety switches within TPSU OK. (Signal TPSU cover opened, TPSU earth stick re-
to GCU). moved, or thermal switch tripped.
(Latched interlock. RFPG set to OFF.)
Error General Fault All PS OK. (Signal to GCU). All PS not OK.
(Latched interlock. RFPG set to OFF.)
Cont Fault Controller OK. (Signal to GCU). Alarm from internal controller watch dog.
Controller not OK.
(Latched interlock. RFPG set to OFF.)
Standby Error No Standby Error. (Signal to GCU). STANDBY READY not reached within time
limit set by configuration
Power Fan ON Command to GSPU to turn ON the fan Command to GSPU to turn OFF the fan
Heater ON Command to GSPU to turn ON the heater Command to GSPU to turn OFF the heater
EHT Con ON Command to GSPU to turn ON contactors Command to GSPU to turn OFF contac-
in the TPSU tors in the TPSU
Screen ON Command to GSPU to turn ON screen Command to GSPU to turn OFF screen
grid power supply grid power supply
Grid ON Command to GSPU to turn ON grid power Command to GSPU to turn OFF grid
supply power supply
13 RF system 413
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Note!
Flickering if RF mode is PULSE.
Ext Loop Sel The external feedback loop is selected The internal feedback loop is selected
which regulates the Dee voltage of the which regulates the output power of the
cyclotron. (Internal signal.) RFPG
90 LM Internal signal to set load match phase Internal signal to set load match phase
detector to 90º offset detector to 0º offset
Log page
On the Log page it is possible to create and view log files. Most of the log files contain digital
in and out and analog in and out signals in the SCU. These signals are described previously
under the respective pages. The log files are collected during time spans of different lengths
and sample times.
The log files that can be obtained are described in Table 13-13.
414 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Statistic Logs the signals status every fif- Operation modes, limit conditions statistic.log
teen minutes. and analog in signals (as presen- When the buffer is full, the con-
tation values). Header with serial tent is copied to old_statistic.log.
numbers of the SCU, RFPG and
cyclotron, version number of the
SCU software, and type of sys-
tem (MINItrace).
Event Logs a message when a specific Event message, time stamp. event.log
event occurs (error, interlock, etc.) The time stamp facilitates trou- When the buffer is full, the con-
ble-shooting by comparing to the tent is copied to old_event.log.
Milliseconds log.
Long Starts logging the signals when See the Statistic log. longTerm.log
term test a Long term test starts (see Sec- Only one file exist.
tion Long term test page on
page 424).
Five sec- Logs the signals status in a tem- Digital in, digital out, analog in SLOW_date_time
onds porary buffer every five seconds and analog out signals
for the past four hours.
The log file is then created manu-
ally by clicking Save log.
Millisec- Logs the signals in a temporary Digital in, digital out, analog in FAST_date_time
onds buffer every millisecond for the and analog out signals
past ten seconds.
The log file is then created manu-
ally by clicking Save log.
Sweep Starts logging the signal status Analog in signals (as presentation Sweep_date_time
when a Sweep run starts (see values) and the frequency. The
SectionSweep page on page423) same header as in the Statistic
log.
13 RF system 415
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
A new window shows the log files available. The example below shows Milliseconds log
files.
3 To save the log file on the laptop, select File > Save As and choose a suitable location.
If the SCU file manager with the Log On As dialog does not appear, the network
parameters on the laptop might be wrong. To correct it, see Section Setting the TCP/IP
address and LAN parameters in the PC on page 400.
2 In the Log On As dialog, type service as username and the password (can be requested
from GE Online Center).
3 In the file manager, click log. The folders for the different log file types appear.
4 Click the desired folder to view the log files. The files can now be copied to other locations,
deleted, etc.
Note!
The log folders should be emptied regularly. Log files cannot be recorded if the memory
is full. Make sure the log files are backed up before deleting them.
416 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Config page
The Config page shows configuration parameters for the RF system.
On the ALL USERS access level all parameters are read-only, that is none of the parameters
can be changed.
When logged in as SERVICE, some of the values can be changed. These values are shown
with black color. The read-only values have gray color.
CAUTION!
Do not change a parameter value unless the consequences are fully understood.
The parameter values for each system are unique. However, they can be adjusted if the
system performance tends to deviate from the specification.
For guidance on setting the configuration parameters, see Section 13-6-12-7 Setting the
configuration in a new SCU Gen II on page 489.
13 RF system 417
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Parameter Description
Dee 1 Pickup Factor1 Ratio between Dee 1 voltage and Dee 1 voltage pickup (factory set)
Dee 1 Cable Factor1 Ratio between output and input of coaxial cable between pickup 1 and SCU (factory
set)
Set for 35kV Dee 11 Control voltage from GCU at 35 kV setting Dee 1 (factory set)
Fwd Coupling Factor1 Ratio of RFPG forward output coupler (factory set)
Refl Coupling Factor1 Ratio of RFPG reflected output coupler (factory set)
Dee Voltage Detected1 Switch over point between START and TUNING mode
Dee Voltage Close1 Switch over point between internal and external loop
Dee Voltage OK1 Maximum allowed percentage below Dee voltage set to reach Dee Voltage OK
418 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Parameter Description
Modulator in alarm Modulator level that enables interlock if ACL, RVL or SCL
Limit Fwd Pwr Multipl Allowed additional anode current for every 8 kW forward power
Limit Refl Pwr Multipl Allowed subtracted anode current for every 8 kW reflected power
Standby Timeout Issues Standby Error if not STANDBY READY within the set time
Normal Mode Timeout Issues Normal mode Error if not Dee voltage within the set time
Time Base Cycle time used for sampling all internal parameters
RF Enable Pulses Number of Time Base units for RF Enable during pulse
RF Disable Pulses Number of Time Base units for RF Disable during pulse
Tuning Mode Delay Number of Time Base units until TUNING mode is generated
Spark Delay Number of Time Base units until SPARK mode is generated
Spark Restart Wait Number of Time Base units until restart of RF after SPARK mode
Temperature Max Upper interlock limit for temperature in SCU. If set to 0 interlock is disabled.
Temperature Min Lower interlock limit for temperature in SCU. If set to 0 interlock is disabled.
Humidity Max Upper interlock limit for humidity in SCU. If set to 0 interlock is disabled.
Humidity Min Lower interlock limit for humidity in SCU. If set to 0 interlock is disabled.
1 The value can be changed. All other values are read-only. The consequences must be fully understood before
changing a value!
13 RF system 419
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
420 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Set IP address Click to implement new addresses entered in the empty fields.
The SCU will reboot which takes about 30–60 s.
Note!
The network addresses should not be changed unless required to match another local network
configuration.
SCU serial number Shows the present SCU serial number. Can not be changed.
Set Serial numbers Click to implement new numbers entered in the fields.
Note!
After replacement of the battery or the entire SCU, the RFPG and cyclotron serial numbers should
be set according to the systems at the site.
Set Time and date Click to implement new time and date entered in the empty fields.
Hour counter Number of hours the has been turned on since the last reset of the
hour counter.
Reset hour counter Resets the hour counter. Should be made after installation of a new
tube.
Reboot SCU Starts rebooting the SCU which takes about 30–60 s.
13 RF system 421
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Test page
The Test page makes it possible to run the RF system without regulation (forward power loop
or Dee voltage loop). It is also possible to simulate interlocks in the hard-wired interlock circuits
RF inhibit 1 and 2, respectively, and override the TPSU door software interlock. These functions
can be used to test the safety system during installation and maintenance of the cyclotron.
CAUTION!
The consequences must be fully understood before using these functions. For more
information, contact GE Online Center.
The Regulation indicator is normally lit (green color), meaning that the RF system regulation
is turned on. To turn off the regulation, click Turn on/off.
The Test relay 1 and Test relay 2 indicators are normally off (gray color), meaning that the
relays do not break the interlock circuits RF Inhibit 1 and 2, respectively.
To simulate an interlock in circuit RF Inhibit 1, click Turn on/off next to Test relay 1. The
indicator is lit (turns green). For RF Inhibit 2, click Turn on/off next to Test relay 2.
The TPSU Door intl indicator is normally off (gray color), meaning that the software override
of the TPSU door interlock is turned off. When turned on, the software interlock system believes
that the TPSU panel is in place even though it is removed. The hardware interlock system is
not affected.
To turn on the software interlock override, click Turn on/off. The indicator is lit (turns yellow).
The TPSU Door intl is turned off automatically after 15 minutes, if not turned off manually
before.
422 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Sweep page
The Sweep page makes it possible to run the RF system and log its behaviour while sweeping
the frequency, phase or Dee voltage.
The sweep run requires a start and stop value for the selected parameter, number of steps
and the time per step. Any written comments will be stored in the log file.
Note!
A sweep run can be started only if the system is in STANDBY READY mode.
When the run is finished, click Download file. A new window appears where the log file
Sweep_date and time is available, see Table 13-13. The file contains analog in signals (as
presentation values) and the frequency.
13 RF system 423
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
The Long term test does not require any input parameters. Any written comments will be
stored in the log file.
Note!
A Long term test can be run only if the system is in STANDBY READY mode.
To start the Long term test, click Start. The green Test Running indicator is lit.
To finish the test, click Stop. However, the test is aborted in advance if the system goes to
OFF mode or when the test has run for 24 h.
When the run is finished, click Download file. A new window appears where the log file
longTerm.log is available. Note that there is only one version. The file is a copy of the
Statistic.log, see Table 13-13.
Furthermore, the software monitors cooling and safety switches. It also checks the levels of
the various voltages at all times and if a fault level is detected at any time (including time
during the power-up sequence), the voltages are removed from the tube amplifier and a fault
signal is indicated.
424 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Sequence description
The power-up sequence makes sure that the power supplies are energized in the correct
order and with the correct timing (Figure 13-26).
When all power supplies have been applied to the tube successfully, the SCU switches to
STANDBY READY state.
If the start-up sequence has not finished within 120 seconds, the SCU generates an alarm
STANDBY ERROR. If so, there might also be other alarms which indicate the cause of this error.
During the opposite procedure, that is when the tube amplifier is powered down, the power
supplies are turned off in a particular sequence. The TAU fan runs for 5 minutes after power
has been switched off.
Alarm enabling
Before fan and power supplies have been switched on, there are alarms activated from the
air switch and power supply monitors. To avoid interrupting the power-up sequence, the
alarms are disabled until the units have been switched on.
Alarm processing
If an alarm occurs during the start-up sequence or during operation, the power supplies are
turned off in a controlled way. The fault information is delivered to the SCU and GCU.
The fault signals require a reset signal to be sent from the GCU for the fault indications and
interlocks to be removed.
13 RF system 425
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Status indicators
The status indicators in the SCU web interface for interlocks and power supplies related to
the power amplifier tube are described in Section Digital Out page on page 412.
Status check
The SCU continuously checks all the status signals which determine whether the RFPG may
be started or not.
The SCU first checks if there is an external interlock RF INHIBIT from the safety relays in the
Control Interface Unit (CIU). If so, the SCU will not allow the RFPG to be activated to an operation
status, not even to STANDBY.
Note!
RF INHIBIT will disable the RFPG not only via the SCU but also via a double relay system.
The SCU also checks the status of signals monitored inside the SCU and signals from the
Driver Power Amplifier (DPA).
If no faults or interlocks have been found, the SCU will accept a request STANDBY from the
GCU.
Start to STANDBY
Upon request STANDBY from the GCU, the software in the SCU starts the power-up sequence
of the tube amplifier power supplies (see Section 13-2-4-7 Power-up sequence of the power
amplifier tube on page 424).
When all power supplies have been started and the tube is ready for operation, the SCU
delivers STANDBY READY to the GCU.
If a fault occurs during the start-up sequence or later, the software generates an alarm and
the SCU aborts the start-up sequence.
RF operation modes
The RFPG must be in state STANDBY READY before any RF mode can be run. The RF system
can operate in three modes – NORMAL, PULSE and FIX:
• NORMAL is the automatic mode which is used for normal production. Only NORMAL
utilizes Dee voltage feedback (EXTERNAL LOOP).
• PULSE mode is used for service and troubleshooting; tuning must be done manually by
the operator.
• FIX mode is used for service and troubleshooting; tuning must be done manually by the
operator.
426 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
The System Controller is directly engaged in control of the RF modes and it works in close
association with the frequency source and modulator.
The RF operation sequence for NORMAL mode from STANDBY READY is described in detail in
Figure 13-27 and Table 13-16.
The sequence for PULSE and FIX modes is described in Table 13-17.
NORMAL mode
The following actions take place when NORMAL mode is started:
1 START mode
(Simultaneously, the RF frequency scans across the tuning range in order to find
resonance.)
The SCU continues to generate pulsed RF power until the signal DET LIMIT changes state.
It indicates that a Dee voltage above the detection level has been found. Consequently,
the SCU goes to TUNING mode
2 TUNING mode
RFPG delivers continuous RF to the RF Cavity and the GCU controls the frequency in order
to match the RF Cavity.
When the RF cavity is being tuned closer to resonance, the Dee voltage feedback signal
approaches the value of the Dee voltage reference. The ratio of Dee voltage to Dee
voltage reference is computed. When this ratio becomes 60%, the SCU switches from
internal loop to external loop.
From this point, the RFPG operates with Dee voltage feedback. In the previous stages of
NORMAL mode, a fraction of the RFPG output power was feedback.
3 DEE VOLTS OK
When the tuning of the RF Cavity has progressed so far that the Dee voltage has reached
about 98% of the Dee voltage reference, the SCU generates DEE VOLTS OK.
4 ERROR
There is a timer set at 60 seconds, which runs during all three stages of the NORMAL
mode start procedure. If DEE VOLTS OK is not achieved within 60 seconds from NORMAL
mode was started, an alarm ERROR (ERROR IN NORMAL MODE) is generated. This signal
is used by the GCU.
FIX mode
FIX mode is very simple:
Upon the command FIX from the GCU the SCU generates continuous RF power to the DPA.
13 RF system 427
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
PULSE mode
PULSE mode is similar to FIX but the SCU generates pulsed RF power to the DPA.
SPARK
If the Dee voltage disappears once it has been detected (DET LIMIT has been present), the
SCU initiates a pause, during which RF output is switched off in order to allow the RF Cavity
to recover after a possible flashover. The duration of the pause is about 1 second. During the
pause, status SPARK is indicated on the SCU web interface. The RF is then restarted
automatically.
Spark detection
If a spark occurs, DET LIMIT disappears and the following things happen almost simultaneously:
If the Dee voltage has not returned within the Spark Delay time, the alarm SPARK will be
shown on the SCU web interface.
SPARK can be initiated not only by flashovers (arcing) in the RF cavity, but by any kind of
unexpected RF disturbances which cause the Dee voltage to fall to a low level.
RF output power
Detection
limit
t
1 ms 9 ms 0.2–1 s
RF mode
STANDBY READY
START
TUNING
DEE VOLTS OK
SPARK
Time t0 t1 t2 t3 t4 t5 t6 t7 t8
428 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
t1 NORMAL mode command line from GCU START Frequency selected. Internal loop feedback.
goes low. RF driving pulses start (RFPG). mode Pulse level determined by safe operation of
RFPG, that is RVL and ACL may be limiting.
t1–t2 Fine tuning starts scan (GCU). START Resonator voltage below detection limit.
mode Resonator voltage below detection limit but
Fine tuning motion improves resonator tun- increasing.
ing (GCU). START
mode
t2 Pulsing stopped (RFPG). TUNING Resonator voltage above detection limit. DET
mode LIMIT has changed state.
t2–t3 RF output continuous (RFPG). Tuning be- TUNING PHASE LOAD signal developed in RFPG con-
comes feedback controlled (GCU). mode trols the RF frequency via GCU Internal loop.
t3–t4 RF drive power increased due to feedback TUNING RF drive power increases until detected RF
action (RFPG). mode is the same value as DEE VOLT REF input.
t4 Resonator voltage within ±1% of demanded DEE Detected RF signal within ±1% of DEE VOLT
value (RFPG). VOLTS REF input SIGNAL.
OK
Driving stopped if necessary (RFPG). Decision made within RFPG after an ad-
justable delay as to whether RF drive is to be
stopped.
t5–t6 RF drive off for period of SPARK pause (RFPG). SPARK SPARK pause about 1 sec. determined within
RFPG.
13 RF system 429
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
t6–t7 RF drive power increased TUNING RF drive power increases until detected RF
(RFPG). mode is within ±1% of DEE VOLT REF input.
t7 Resonator voltage within ±1% (RFPG). DEE Detected RF signal within ±1% of DEE
VOLTS VOLT REF input. signal.
OK
t7–t8 Normal RF operation with full resonator DEE Detected RF signal within ±1% of DEE VOLT
voltage (RFPG). VOLTS REF input signal.
OK
Table 13-17: Control system sequence for PULSE and FIX modes
t1 PULSE mode goes low. PULSE PULSE mode command received (GCS). Con-
trol loop is closed using forward amplitude
feedback and is inhibited from switching to
the detected RF feedback. Pulsing of RF drive
is enabled.
t3 FIX mode goes low. FIX FIX mode command received (GCS). Control
loop is closed using forward amplitude
feedback and is inhibited from switching to
the detected RF feedback. Pulsing is inhibited.
430 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Abbreviations:
AI = Analog In
AO = Analog Out
DI = Digital In
DO = Digital Out
PLD/SKD PS / TB2
5 0V(24V) 5 PWR 0V
2 Not used 2
3 Not used 3
4 Not connected 4
7 Not connected 7
9 SI- 9 DI
13 RF system 431
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
12 Not connected 12
13 Not connected 13
432 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
SKG GCU / P9
2 Standby# 2 DI (24V)
3 Normal_Mode# 3 DI (24V)
4 Reset# 4 DI (24V)
5 Fix_Mode# 5 DI (24V)
6 Pulse_Mode# 6 DI (24V)
7 Freq_Load 7 DI (24V)
8 RF_Inhibit_1 8 DI (24V)
9 Standby_Ready# 9 DO
(24V)
10 Standby_Error 10 DO
(24V)
11 Start_mode# 11 DO
(24V)
12 Tuning_Mode# 12 DO
(24V)
13 Dee_Volts_OK# 13 DO
(24V)
14 Spark 14 DO
(24V)
15 Error# 15 DO
(24V)
13 RF system 433
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
27 FREQ_CLK 27 DI (24V)
28 Not connected 28 DO
(24V)
29 FREQ_DATA 29 DI (24V)
30 Not connected 30 DO
(24V)
31 Cont_Fault 31 DO
(24V)
32 General_Fault 32 DO
(24V)
34 RF_Inhibit_2 34 DI (24V)
35 Not connected 35
36 Not connected 36
37 HTL(0V) 37 GND
434 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
SKM TAU/Water
flow switch
4 Not used 4
5 Screen_Current_Test 5 AO
8 Phase_Load_Return 8 GND
10 DVR_Signal 10 AO
11 Anode_Current_Test 11 AO
13 FP_Signal 13 AO
13 RF system 435
13-2 Overview
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
SKK TAU / P2
2 Not connected 2
5 Not connected 5
12 Not connected 12
SKR RF_REFL TAU / P10 RF IN Coax cable (from RFPG output coupler)
436 13 RF system
13-2 Overview
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
13-3-1 General
The Tube Amplifier Unit, TAU, operates two resonant circuits, the Grid circuit and the Anode
Circuit, both of them tuned in resonance with the operation frequency.
Furthermore, there is the parameter Phase Set which can be set from the MSS Configuration
page. Phase Set defines the phase at which the tuning of the RF Cavity is in balance.
• The RF system cannot deliver full RF output power and the DPA Reflected Power reads
a high value (>10 W). (Displayed on RFPG Measurement Module (SCU Gen I, or the front
display or SCU web interface (SCU Gen II).)
Tuning procedure
You need a flat head screwdriver, about 4 mm wide, and with a blade length of minimum
90 mm.
4 On the front panel of the TAU, remove the screw marked “Grid Tuning”.
5 Insert the screwdriver straight through the hole in the panel. Find the tuning screw, the
end of which is located about 85 mm from the panel.
6 Very slowly turn the screwdriver any direction while observing the display of the
Measurement Module (SCU Gen I, or the front display or SCU web interface (SCU Gen II).
If the value shown tend to increase, turn the other direction. The proper grid circuit tuning
is the point where the DPA Reflected Power is minimum. DPA Reflected Power should be
adjusted for <5 W, and it is normally possible to achieve about 2 W when minimum.
Note!
5 W is shown as 0.5 on the Measurement Module (SCU Gen I)!
7 Should there be difficulties to find a well defined minimum of DPA reflected power,
decrease the Dee voltage level and try adjusting again.
13 RF system 437
13-3 Overhaul of Tube Amplifier and RF system tuning
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
• The Reflected Power reads a high value and the RF system might not be able to deliver
the required Dee voltage due to Reflected Voltage Limiting.
If there is a less severe mismatch the Reflected Power will still read comparatively high and
furthermore the Forward Power will read higher than normally for a fixed voltage.
The optimum setting of the phase is somewhat dependent upon the tuning state of the anode
circuit. Hence if the anode circuit has been retuned, Phase Set should be checked afterwards.
Tuning procedure
1 Start the RF system in NORMAL mode from the MSS.
RF parameter Displayed on
Phase MSS
438 13 RF system
13-3 Overhaul of Tube Amplifier and RF system tuning
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
9 If the reflected power is lower, Phase Set was changed the right way. Basically the best
Phase Set is the setting which offers minimum Forward Power and zero Reflected Power.
However, also watch Anode Current and DPA Forward Power. Sometimes these
parameters will vary with Phase Set even though Forward Power and Reflected Power
do not change much.
10 Repeat 4–9.
Note!
If Anode Current and DPA Forward Power still read too high the anode circuit needs tuning.
See Section 13-3-4 Anode circuit on page 439.
• The RF system cannot deliver full RF output power and the Anode Current is high. Anode
Current Limit may be lit intermittently or constantly.
Tuning the anode circuit is effected by moving the anode shorting plate. Moving the anode
shorting plate 1 mm will change the resonance frequency of the anode circuit by about
0.1 MHz. Moving it upwards means lowering the resonance frequency.
Note!
Try optimizing Phase Set before starting to retune the anode circuit.
See Section 13-3-3 Phase set on page 438. (Setting the phase is more easily performed than
tuning the anode circuit and may sometimes create good enough a function.)
Tuning procedure
1 Start the RF system in NORMAL mode from the MSS.
13 RF system 439
13-3 Overhaul of Tube Amplifier and RF system tuning
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
RF parameter Displayed on
Phase MSS
5 Shut down the RFPG. See Section 13-6-2 Shutting down the RFPG on page 447.
Disconnect the cable at P17. Undo the 18 screws around the edge of the cover. Repeatedly
undo each screw about a turn at a time until the cover comes off.
8 Measure and record the position of the connector clamp at the top of the anode circuit.
10 Carefully measure and record the space from the plate on which the tube socket is
mounted to the anode shorting plate.
11 By turning the four screws on top of the RFPG two turns clockwise or anti-clockwise, the
anode shorting plate moves 2 mm up or down.
12 Tighten the screws on the connector clamp. Ensure the clamp is restored to the previous
distance from the anode shorting plate.
Should the plate connector tend to become stretched, there are two screws at the back
of the connector clamp which will enable adjustment.
13 Put back the upper front cover. Tighten the screws about a turn at a time repeatedly
until the screws are fully tightened.
16 At the mains distribution panel, switch on mains for RFPG high voltage section.
17 At the rear of the Control Cabinet, switch on the circuit breaker marked “RFPG”.
440 13 RF system
13-3 Overhaul of Tube Amplifier and RF system tuning
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
20 If the anode current is lower, the adjustment was successful. If the readings are within
the limits indicated above and the “anode current limit” is not lit intermittently no further
adjustment is needed.
Should the readings be higher or “anode current limit” be lit more constantly, the anode
shorting plate should be adjusted the opposite direction from the original position.
Again repeat steps 4–19, but turn the screws on top of the RFPG the other way.
Note!
Phase set may need some readjustment after the anode circuit has been tuned.
Note!
This instruction only applies to SCU Gen I.
The planned maintenance protocol is found in Section 3-9-15 RFPG: SCU Gen I Measurement
Module readout (STANDBY/NORMAL mode) on page 110.
The RFPG SCU is the upper left unit in the RFPG. See Figure 13-28 below.
13 RF system 441
13-5 Planned maintenance
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
RFPG SCU
1 Use the MSS RF System menu to set the RF system in STANDBY mode.
2 Use the MSS RF System menu to set the RF system in NORMAL mode.
442 13 RF system
13-5 Planned maintenance
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
3 Use the MSS RF System menu to set the RF system in STANDBY mode.
4 Use the MSS RF System menu to set the RF system in NORMAL mode.
RF FORWARD VOLTAGE
RF REFLECTED VOLTAGE
FORWARD POWER
REFLECTED POWER
ANODE VOLTAGE
ANODE CURRENT
GRID VOLTAGE
GRID CURRENT
SCREEN VOLTAGE
SCREEN CURRENT
HEATER VOLTAGE
13 RF system 443
13-5 Planned maintenance
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Note!
This is not a regular Planned Maintenance issue! Disregard this if a complete set of values is
not specifically asked for!
Regarding the RF board readout: Find the appropriate test points on the RF board, use a volt
meter and record the values in the table in Section 3-9-15 RFPG: SCU Gen I Measurement
Module readout (STANDBY/NORMAL mode) on page 110.
444 13 RF system
13-5 Planned maintenance
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
RF DET (TP24)
DVR (TP15)
FV (TP16)
AM2 (TP3)
AM1 (TP2)
AM3 (TP21)
DVR OK (TP2)
FP (TP17)
RV (TP18)
RP (TP19)
RF ENABLE (TP5)
PHASE (TP20)
+5V (TP11)
-5V (TP14)
+15V (TP7)
+12V (TP8)
-12V (TP10)
-15V (TP9)
GND (TP13)
+24V (TP12)
13 RF system 445
13-5 Planned maintenance
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
446 13 RF system
13-5 Planned maintenance
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
WARNING!
Always lock-out and tag-out power to the RFPG before opening any covers.
For Lock-Out and Tag-Out (LOTO) procedures of the equipment, see Section 3-5-6 LOTO
procedures for the MINItrace system on page 59.
For spare part numbers, refer to MINItrace Spare Parts Catalog (dir. 2380006-100).
- Contactor (TPSU)
- High voltage transformers 7.8kV 4A
- Relay 240V (RLY1,2) RFPG
- Door switch RFPG
- RFPG Contactor kit
- Circuit breaker 10A 1PHASE Type K
- High volts. rectifier 12 PCS
- Contactor
- CAC223-31011 EHT Assembly, MT
- Circuit breaker 3PH 40-100A SW1 RFPG
- GSPU Assy
- Anode current monitor board
- Tube controller board
- GCU I/O Backplane board
When other spare parts involving the RFPG, GCU, Control interface board, and external
interface box have been replaced, it is recommended to do the test.
a Switch off and apply LOTO on the RFPG circuit breaker at the rear of the Control
Cabinet.
b Switch off and apply LOTO on the RFPG circuit breaker at the Mains Distribution
Panel.
13 RF system 447
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
WARNING!
If the rear panel of the TPSU is opened, the earth stick (ground hook) inside the panel
must be hooked on to one end of the high voltage fuse F1.
Always turn off power to the RFPG before opening any covers.
Note!
Make sure that the front TAU cover and the upper rear covers of the RFPG are in place when
the RFPG is on to fully ensure RF functionality.
2 wrenches 7/8"
2 clamps
1 wrench 8 mm
1 wrench 10 mm
1 hex key 4 mm
1 hex key 5 mm
1 flat-head screwdriver 10 mm
1 flat-head screwdriver 5 mm
1 Phillips screwdriver No 2
adhesive tape
3 Attach the earth stick (ground hook) to one end of the HV fuse F1.
4 On the water manifold of the cooling system, shut the valves for water to and from the
RFPG.
448 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
6 Loosen the two water tubes marked IN and OUT at the top rear of the TAU. Be prepared
with a vessel and some absorbing material to collect any remaining water from the
loose tubing.
If required, it is also possible to clamp off the tubes to prevent any remaining water from
leaking out.
Note!
Mark the tubes to avoid confusion when reconnecting.
b Unscrew each of the 18 screws around the edge of the cover about one turn at a
time until the cover comes off.
10 Measure and record the position of the connector clamp at the top of the anode circuit.
(Note the distance from the clamp to the anode shorting plate).
11 Undo the screws and nuts on both sides of the connector clamp and remove the outer
half of the clamp.
12 Undo and remove the clamp at the bottom of the anode circuit.
13 Loosen the screw on the anode inductor halves and remove the anode inductor.
14 Remove the anode voltage lead (the small copper clamp) from the power tube.
15 Carefully loosen the water couplings on the power tube using two 7/8" wrenches.
Turn the nuts on the power tube itself. The brass couplings on the nylon tubes should
be held in a fixed position during loosening.
Be prepared with some absorbing material to remove any small amounts of water that
may come out of the water tubing.
16 Carefully pull the power tube upwards. Jiggling the tube back and forth a little will make
the tube come out of its socket more easily.
Note!
Wear gloves while handling the power tube to prevent it from getting stained.
13 RF system 449
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
19 Put the dismounted tube in a safe place, preferably in the manufacturers box.
450 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
1 Dismount the spark gap bracket (see above: Dwg CAC222-31011-03-3, pos 57).
Note!
On some systems there is no spark gap bracket to dismount.
2 Fasten the screws (3) in the threaded holes on each side of the socket. Use the right hand
screw to keep the shorting arm off the socket. (Not shown above).
Note!
Wear gloves while handling the power tube to prevent it from getting stained.
The cooling lines should face the open front of the TAU. The cooling line marked “IN”
should face to the right and “OUT” to the left.
5 Put the rubber pad (2) and the tool plate (4) on top of the power tube and mount the
washers and the nuts (1, 5). See Figure 13-31.
13 RF system 451
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
6 Push the power tube downwards by tightening the nuts slightly. Jiggle the power tube
to gently force it into the socket and then tighten the nuts further. Repeat until the power
tube is fully inserted in the socket. See Figure 13-32.
Note!
It is important that the power tube is inserted straight into the socket. Make sure to tighten
the nuts evenly on both sides.
b Use a small piece of adhesive tape to attach the end of the Teflon to the power
tube.
c Wind the Teflon tightly around the power tube. The winding should be up against
the metal ring at the bottom of the tube’s metal jacket.
d Use a small piece of adhesive tape to attach the outer end of the Teflon.
452 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Use two 7/8" wrenches. Turn the nuts on the power tube itself. The brass couplings on
the nylon tubes should be held in a fixed position during tightening.
11 Reconnect the water tubing at the back of the TAU. Do not mix up IN and OUT. Refer to
“Water System Schematic” on drawing CAC222-31001-xx-1.
13 At the water manifold of the Cooling Water System, open the valves for water from and
to the RFPG.
• The water connections on the power tube need careful inspection. Drops of water
may need several minutes to appear if there is a small leak.
• Run cooling water for half an hour to ensure there is no leakage within the TAU
before completing the reassembly.
b Adjust the capacitor ring to a position as shown in Figure 13-32 (15 mm).
c Adjust the anode inductor halves. The openings should be the same at the front
and at the rear.
d Tighten the clamp. Make sure that the clamp act on all “fingers” of the anode inductor
and thus bringing all of them in good contact with the capacitor ring.
19 Remount the outer half of the connector clamp at the top of the anode inductor:
a Carefully restore the connector clamp to the position noted before disassembly.
20 Make sure the shorting arm has not been affected by the power tube replacement:
a Check that the spring loaded arm can move freely towards the power tube.
b Make sure that the arm touches both the anode (the metal ring on the power tube)
and the screen grid (the outer spring collet on the socket).
13 RF system 453
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
21 Restore the spark gap bracket to its original position, see Figure 13-33.
Note!
On some systems there is no spark gap bracket to remount.
22 Adjust the gap from domed nut of the spark gap bracket to the capacitor ring to 4 mm,
see Figure 13-33.
Note!
On some systems there is no spark gap to adjust.
• Reinstall the upper front cover. Tighten the screws about a turn at a time until all
screws are fully tightened.
25 At the back of the TPSU, restore the earth stick to its holder.
27 After replacing the power tube, log the reading of the hours counter on the rear of the
Source and Control Unit (SCU).
2 At the rear of the Control Cabinet, switch on the circuit breaker marked “RFPG”.
454 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
3 On the front of PS Assy in the TAU, pull out the plunger of the door switch SW1.
CAUTION!
The RFPG can now be switched on without the TAU front panel in position. Do not
short-circuit the terminals of the power supplies on the PS Assy which are capable to
deliver high current.
6 In order to verify proper function of the new power tube, the DC parameters of the tube
need to be checked as follows:
a By means of a DVM, make sure the DC-voltage at the terminals of the heater power
supply (PS2) on PS Assy is 7.7 V ±100 mV.
b Record the dc-parameter readings on RFPG Measurement Module (SCU Gen I), or
front display or SCU web interface (SCU Gen II):
Screen current 23 mA ±3 mA
9 At the MSS, start the magnet system and start the RF system to STANDBY.
11 Click NORMAL.
13 RF system 455
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
14 Record the RF parameter readings on RFPG Measurement Module (SCU Gen I), or front
display or SCU web interface (SCU Gen II):
15 By increasing Dee voltage to about 39 kV, verify that the RFPG can deliver its specified
output 10 kW.
Note!
Sparks may occur initially when running on high Dee voltage.
16 If the RFPG proves not to be able to deliver its output power at any power level, see
Section 13-3 Overhaul of Tube Amplifier and RF system tuning on page 437.
456 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
If inhaled, beryllium oxide might cause irritation to the lungs. In contact with the skin,
beryllium oxide might cause allergic responses, infections or mechanical injury.
Some components in the DPA contain beryllium oxide ceramic (see Figure 13-34). If these
components are damaged when changing DPAs, beryllium oxide might escape and cause
health problems. Use protective goggles and gloves when handling the DPA. Do not remove
the cover of the DPA.
Always handle used parts in accordance with federal, state and/or local regulations.
13 RF system 457
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
458 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
13 RF system 459
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
460 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
• P21
• P22
• P23
14. Gently pull the DPA out. Make sure the coils
at the back are duly undone from the DPA
unit!
13 RF system 461
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
For spare part number, refer to the spare parts listed for each subsystem in MINItrace Spare
Parts Catalog (dir. 2380006-100).
The PETtrace system includes a GSPU which is similar but not identical and thus has another
part number! Note the following:
For Lock-Out and Tag-Out (LOTO) procedures of the equipment, see Section 3-5-6 LOTO
procedures for the MINItrace system on page 59.
13-6-8-1 Introduction
The anode current measurement board within the TPSU of the MINItrace RFPG has been
replaced by a new design (Q2 2003). The new design is named anode current monitor board.
This board is compatible with the old one, but its design is more robust.
For spare part number, refer to the spare parts listed for each subsystem in MINItrace Spare
Parts Catalog (dir. 2380006-100).
13-6-8-2 Disassembly
Always lock-out and tag-out mains supply to the RFPG before opening any covers.
462 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Action Check
1. Turn off MINItrace in the Master System, or turn off the RF system in the
MINItrace Service System (MSS).
3. Detach the lower rear panel of the RFPG by turning the screws ¼ turn
counter clockwise.
Ensure that there is no high voltage by touching the earth stick (ground hook) of the TPSU
to the terminals of the capacitor C3 and to both sides of the high voltage fuse F1. Hook it
then on to the right hand side of the high voltage fuse F1. Keep the earth stick in this position
while work is going on within the RFPG.
4. NOTE!
Install a short-circuiting wire across the terminals of the high voltage
capacitor C3, in order to fully eliminate the unlikely risk of the capacitor
recharging while leads are disconnected at the circuit board.
Wire is furnished as part of the anode current monitor board spare part.
5. At the anode current monitor board within the TPSU, dismount the pro-
tective plate. Loosen by depressing the little hook on the top of the nylon
PCB supports (see Figure 13-35).
Disconnect all leads from TS1, TS2, TS3 and TS4. Label to leads to avoid
confusion when reconnecting.
TS 1 is hard to access while the board is installed and the lead connected
there might have to be disconnected after the board has been loosened.
In order to dismount the board, unscrew two screws and four nylon PCB
supports. Detach the board from the RFPG.
13-6-8-3 Assembly
Action Check
6. On the new anode current monitor board remove the protective plate.
Loosen by depressing the little hook on the top of the nylon PCB supports.
Unscrew the nuts holding the PCB-supports. See Figure 13-35.
13 RF system 463
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Action Check
• Fix the board by fitting washers and screws at the corners at the
left hand side of the board.
• Fit the threaded nylon PCB-supports (part of the new board) to the
other four mounting holes. Note: No washers to be used here.
• Put the protective plate onto the ends of the nylon PCB-supports
and push inwards until it locks in position.
12. Turn on mains for the RFPG at the Mains Distribution Panel.
Turn on the mains circuit breaker at the rear of the Control Cabinet.
Figure 13-35: Anode current monitor board protective plate with nylon PCB support
464 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
For spare part numbers, refer to the spare parts listed for each subsystem in MINItrace Spare
Parts Catalog (dir. 2380006-100).
• A coding strip makes it impossible to insert the RF board in the wrong position.
Figure 13-37: Low pass filters connected to inputs SKQ and SKR on the SCU
See Section 13-6-9-1 Instructions for the new RF circuit board and LP filters on page 466 for
instructions on how to implement the new RF circuit board and LP Filters.
13 RF system 465
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Note!
All removed parts or waste that is radioactive shall be handled and disposed I/A/W local laws
and restrictions. Please refer to Appendix B General radiation safety aspects and Appendix C
MINItrace radiation safety aspects for further instructions during work with radioactive
equipment/parts.
Action Check
1. Turn off and apply LOTO on the RFPG mains power circuit breakers at
the rear of CCAB and the MDP.
For Lock-Out and Tag-Out (LOTO) procedures of the equipment, see
Section 3-5-6 LOTO procedures for the MINItrace system on page 59.
2. Insert low pass filters in series with the cable to inputs SKQ and SKR, see
Figure 13-36.
(If installed earlier, replace the old filters with the two new filters.)
Ground yourself by an ESD field kit. Use the Card ejector to detach the
old RF board, place it in an antistatic bag. Take the new board out of its
antistatic bag and install it in the same position as the board you re-
moved.
Note!
The old RF circuit board should be disposed I/A/W local laws and restric-
tions.
466 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Action Check
Note!
If you do not obtain Dee voltage, you have to perform troubleshooting.
Anode current
(Ampere)
3.00
2.80
2.60
2.40
2.20
2.00
1.80
1.60
1.40
1.20
1.00
0 2 4 6 8 10 12 FP (kW)
13 RF system 467
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Action Check
6. Set Dee voltage to 35 kV. The reading shall now be: 34.6–35.4 kV, about
8 kW FP, about 2.25 Amp anode current and less than 60 mA screen
current at the RFPG MM.
If the screen current is too high, check the anode voltage, preferably this
should be at least 7.5 kV.
If you do not obtain 34.6 kV or if you read anode currents above 2.3 Amp
with 8 kW FP you have to adjust the clamp downwards. If the screen
current is too high and anode current is less than 2.2 Amp at 8 kW FP,
try to adjust the clamp upwards. If anode current is above 2.2 Amp and
the screen current is above 60 mA you are in trouble, if you cannot set
a slightly higher anode voltage. Preferably the anode voltage should not
be above 8 kV at 35 kV Dee voltage.
7. Wipe the MINItrace clean from possible contamination that might have
occurred during implementation of spare part. Perform a production
test (verify that you read at least 34.6 kV of Dee voltage) and return
system to operator/customer.
1 Turn off the RF system, see Section 13-6-2 Shutting down the RFPG on page 447.
2 Removal of outer covers, see Section 13-6-3 Removing covers on page 448:
13-6-10-1 Tools
The following tools are required for servicing the SCU:
468 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
2 Pull the circuit board remover lever (located at the lower end of the circuit board).
3 When the circuit board is loose, pull the circuit board out of the SCU.
13-6-10-3 Reassembly
1 Apply suitable equipment to prevent static voltage damage to circuit boards.
2 Switch on the RFPG circuit breaker at the rear of the Control Cabinet (CCAB).
3 Check that the four LEDs on the RF board (the left-most circuit board in the SCU rack)
indicate that the supply voltages (+15V, –15V, +5V, +24V) are present.
13 RF system 469
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
7 Check the values shown on the Measurement Module of the SCU. The allowed deviations
are as follows:
d. SCREEN CURRENT 23 mA ± 3 mA
8 When STANDBY is indicated, use the arrow buttons to set the Dee voltage to 35 kV.
9 Click NORMAL.
Upgrading from SCU Gen I to SCU Gen II is described in Section 13-6-12 Replacing SCU Gen I
with SCU Gen II on page 478.
470 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
13-6-11-2 Tools
The following tools are required for servicing the SCU:
If the SCU cannot be accessed for backing up the files before the replacement, it is important
that up-to-date configuration values are available from previous back-ups.
2 Connect a straight-thru network cable between the laptop and a LAN switch. Use a
twisted cable if you connect directly to the rear of the SCU.
3 Create the following back-up folder on the laptop for storing the configuration files:
C:\backup\scu.
If the SCU file manager with the Log On As dialog does not appear, the network
parameters on the laptop might be wrong. To correct it, see Section Setting the TCP/IP
address and LAN parameters in the PC on page 400.
6 In the Log On As dialog, type service as username and the password (can be requested
from GE Online Center).
13 RF system 471
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
7 Copy the files cycloMT.cfg and scuMT.cfg from the Config folder to the back-up folder
C:\backup\scu on the laptop.
CycloMT.cfg will be used in the new SCU, scuMT.cfg is stored for backup only.
CAUTION!
Do not copy scuMT.cfg to another SCU. ScuMT.cfg is unique for each SCU and cannot be
used in other SCUs.
8 We also recommend copying the log files in the Log folder to the back-up folder. The
files are useful for future trouble-shooting and service.
Note!
The log folders should be emptied regularly. Log files cannot be recorded if the memory
is full. Make sure the log files are backed up before deleting them.
CAUTION!
Do not copy scuMT.cfg to another SCU. ScuMT.cfg is unique for each SCU and cannot be
used in other SCUs.
After replacing the SCU, restore the backed up configuration files as follows:
1 Open the SCU file manager as described in step 2–5 in the previous section.
3 Copy the backed up version of cycloMT.cfg from the back-up folder to the Config folder.
4 To activate the configuration file, re-boot from the SCU web interface (see Section SCU
Settings page on page 420) or cycle the power to the SCU.
1 Turn off the RF system, see Section 13-6-2 Shutting down the RFPG on page 447.
3 Unscrew the six retaining screws on the SCU front panel using the small screwdriver
(see Figure 13-41).
472 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
4 Disconnect the front display cable behind the panel. The front panel should not be left
hanging in the cable only.
6 To remove the battery from the holder, press the retainer spring aside using the small
screwdriver (see Figure 13-42).
Retainer spring
7 Insert the new battery in the holder with the plus pole facing up.
13 RF system 473
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
10 Recycle or dispose of the old battery in accordance with local or country laws in order
to minimize potential effects on the environment and human health.
Note!
The time and date settings will be lost after the battery replacement. To set time and date,
see Section SCU Settings page on page 420.
For spare part number, refer to the spare parts list for sub system MT 225 SCU in MINItrace
Spare Parts Catalog (dir. 2380006-100).
If the SCU file manager with the Log On As dialog does not appear, the network
parameters on the laptop might be wrong. To correct it, see Section Setting the TCP/IP
address and LAN parameters in the PC on page 400.
4 In the Log On As dialog, type service as username and the password (can be requested
from GE Online Center).
5 If there is an nk.bin file on the top level (root) in the SCU file manager, delete the file.
6 Copy the nk.bin file on the CD to the top level in the SCU file manager.
474 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
2 Type telnet 192.0.4.71 and click OK. The Telnet window appears.
4 Type the password and press Enter. The password can be requested from GE Online
Center.
5 At the prompt, type copy “\storage card\scu\nk.bin” “\storage card”. Press Enter.
2 Check that the SCU version number shown on the display is the new one.
1 Turn off the RF system, see Section 13-6-2 Shutting down the RFPG on page 447.
2 Remove the RFPG upper right rear cover, see Section13-6-3 Removing covers on page448.
13 RF system 475
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
3 Unscrew the retaining screws for all connectors at the rear of the SCU using the
screwdriver (see Figure 13-44).
5 Unscrew the four retaining screws at the front of the SCU (see Figure 13-44).
7 When discarding an SCU, follow the federal, state and/or local regulations for disposal
of electrical and electronic equipment.
2 Fasten the SCU at the front of the cabinet with the four retaining screws.
Note!
Make sure that the cable labels match the markings on the rear panel.
2 Switch on the RFPG circuit breaker at the rear of the Control Cabinet (CCAB).
476 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
3 Make sure that the SCU boots up as usual. See also Section Turning on power to the SCU
on page 399 and the subsequent sections.
7 Check the following values on the front display of the SCU or on the web interface. The
allowed deviations are as follows:
d. SCREEN CURRENT 23 mA ± 3 mA
8 When STANDBY is indicated, use the arrow buttons to set the Dee Voltage to 35 kV.
9 Click NORMAL.
3 Download the Statistics log, the Milliseconds log and the Five seconds log.
4 Save the log files in the backup folder on the laptop, C:\backup\scu.
Note!
Do not overwrite ScuMT.cfg-files that belong to other SCUs.
13 RF system 477
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
SCU Gen II replaces SCU Gen I in case of obsolete spare parts for SCU Gen I.
• Tie wraps
13-6-12-2 How to use the adapter cables in MINItrace NSCU upgrade kit
There are two different versions of SCU Gen I. One version has Edac 516 connectors on the
rear panel, the other one has D-Sub connectors (Figure 13-46).
The adapter cables in the upgrade kit should ONLY be used if SCU Gen I has Edac 516
connectors. This is described further on in the upgrade procedure in Section Connecting Edac
516 – D-Sub adapter cables on page 483.
478 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
1 IMPORTANT! If SCU Gen I is functioning, record all values shown on the Measurement
Module on the old SCU before turning it off (Figure 13-47). The values will be used when
setting the configuration parameter in SCU Gen II. If the SCU is not working, use the most
recently documented values.
13 RF system 479
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
4 Switch off the RFPG circuit breaker at the rear of the CCAB (Figure 13-48).
5 Apply LOTO on the RFPG, see Section 3-5-6 LOTO procedures for the MINItrace system
on page 59.
7 Disconnect all cables from the rear panel on the SCU. If necessary, use a screwdriver to
loosen retaining screws (Figure 13-49).
Note!
Any low-pass filters on position SKQ and SKR shall NOT be moved to the new SCU
(Figure 13-50)!
Note!
The illustration shows a unit with D-Sub connectors. There are also units with
square-shaped Edac 516 connectors (Figure 13-46).
480 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
8 On the front panel, unscrew the two upper screws first (Figure 13-49). This is important
because the SCU does not rest on rails!
9 Unscrew the two lower screws while pressing against the front panel with the hand to
prevent the SCU from hanging down (Figure 13-49).
13 RF system 481
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
1 Move the existing cage nuts on the cabinet front and add the four new nuts included in
the kit.
2 Insert the SCU in the lower position and fasten it using the retaining screws.
3 Fasten the blind plate in the upper position using the screws and washers included in
the kit.
482 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
3 Carefully cut the tie wraps using the cutting pliers (Figure 13-52).
CAUTION!
If SCU Gen I had D-Sub connectors, ignore this section and go to Section Connecting the
cables on page 486.
13 RF system 483
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
CAUTION!
Wrong connection of the cables might damage the SCU or other parts.
2 Tighten the Edac 516 retaining screws fully while pressing the connectors towards each
other (Figure 13-54). Pay attention to the keying in the connectors.
1 New cable part numbers introduced 2014. The new numbers are shown in boldface and the old numbers in
brackets.
Note!
The D-Sub cables to SKG and SKH should also be connected to the rear panel on the new
SCU. However, the ground cable (green/yellow) will not be reused.
484 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
13 RF system 485
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
2 Attach all connectors to the rear panel (Figure 13-55). If necessary, use a screw driver to
tighten the D-Sub connectors.
Note!
Do NOT move any low-pass filters from the old SCU to SCU Gen II (Figure 13-50). SCU Gen II
shall NOT have low-pass filters.
CAUTION!
Wrong connection of the cables might damage the SCU or other parts.
486 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
Connect the RJ45 cable (CAB225-31286) between the LAN socket on the SCU and a free
socket in the network switch in CCAB (Figure 13-56).
• Route the cable through the holes in the cabinet roofs and in the cable duct.
4 Bundle the cables using tie wraps and fit them into the cable duct again.
Do not pull the cables too hard! Use suitable exit holes along the metal cable duct.
If the existing holes are not sufficient, make new holes using a saw.
5 Make sure that no cable risks being pinched against sharp edges on the cable duct.
13 RF system 487
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
6 Attach the covers to the cable duct. If required, secure the covers with tie wraps.
If the SCU is not connected to the Ethernet network or if there is no such network (coaxial
cables instead), the laptop can be connected directly to the rear of the SCU.
1 Remove the rear cover required to get access to the network switch (CCAB door) or the
SCU (upper right-hand cover).
2 Connect the Ethernet port on the laptop to either the network switch or the SCU as
follows:
• To the network switch using a straight-thru RJ45 cable. Make sure that the SCU is
connected to the network.
• To the LAN socket at the rear of the SCU using a twisted RJ45 cable.
4 Switch on (push upwards) the RFPG circuit breaker at the rear of the CCAB. The SCU and
RFPG starts up which takes about 40–60 seconds.
5 Check that the Main menu appears on the front display on the SCU (see Figure 13-9).
6 Make sure that the IP address stored in the SCU is 192.0.4.71 by selecting More > IP
address > Read IP on the front display.
Note!
192.0.4.71 is the default address. It can be changed if required for matching another local
network address configuration.
488 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
The Main page in the SCU web interface appears in a few seconds.
If the Main page does not appear, the configuration of network parameters on the laptop
might be wrong. To correct it, see Section Setting the TCP/IP address and LAN parameters in
the PC on page 400.
Introduction
Due to small differences in calibration of electronics and signal filtering, the new SCU (SCU
Gen II) might measure Dee voltages and RF power slightly different than the old SCU (SCU
Gen I). (This is similar to when replacing the RF board in SCU Gen I.) The difference might cause
that the real Dee voltages used for beam irradiation changes a few percent compared to the
SCU Gen I. The beam properties, such as ion source efficiency and beam width, then also
change slightly. Beam tuning might then be necessary to reach the same working point as
before. If so, the expected tuning is to update Dee voltage ~±2 kV.
It is recommended to use cyclotron data from the near past as reference when upgrading
to SCU Gen II. Such data can be found in production log files, production reports and PM
protocols. Primarily, the RF parameters and beam properties are recommended to be used
for comparison. Additional beam tuning might be necessary. However, the upgrade can also
be done without history references.
Preparation
Prepare the following before starting the configuration:
• Get the following parameters from the configuration page accelerator in the MSS:
Parameter Value
Dee_scale
Parameter Number
SCU IP address1
1 If other than standard address for standard addresses of cyclotron network 192.0.4.XX, the default address will
do.
• Make sure there is a back-up folder for configuration files, C:\backup\scu, on the laptop.
13 RF system 489
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Parameter Value/calculation
Settings
1 Log on to the SERVICE level in the SCU web interface.
a IP address
SCU/RFPG tuning
Adjust and tune the RFPG as follows (internal loop adjustment):
1 Update the Config page in the SCU web interface (see Section Config page on page 417)
with the calculated values in Table 13-22.
7 Enter the Fwd Pwr value in the field for Fwd Power at 35kV.
9 Record the values for the RF parameters and beam properties and compare with the
values at 35 kV recorded just before the SCU replacement.
If the parameter values for the working system could not be recorded before the SCU
replacement, choose a suitable Dee voltage setting from a previous PM.
490 13 RF system
13-6 Disassembly/reassembly procedures
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
1 Run at 35 kV.
• Fill in the values from the SCU web interface Main page in Table 13-23.
2 Go to STANDBY READY.
• Fill in the values from the SCU web interface Main page in Table 13-23.
3 Download the Statistics log, the Milliseconds log and the Five seconds log.
4 Save the log files in the backup folder on the laptop, C:\backup\scu.
DEE Ref In kV
DEE 1 kV
Mtch ph 90 º
DPA Fwd W
DPA Refl W
Fwd Pwr kW
Refl Pwr kW
Anode V kV
Anode C A
Grid V V
Grid C mA
Screen V V
Screen C mA
Heater V V
13 RF system 491
13-6 Disassembly/reassembly procedures
MINItrace SERVICE MANUAL – MAINTENANCE GE HEALTHCARE
DIRECTION 2233000-100 REVISION 18
Note!
Do not overwrite ScuMT.cfg-files that belong to other SCUs.
For instructions, see Section Backing up the configuration files on page 471.
Whenever changing the SCU configuration, make a back-up and store it appropriately.
Beam verification
Make a suitable test irradiation from the service laptop or the Master station.
If applicable data exist, compare the ion source output and beam characteristics (beam width,
transmission properties) with the situation just before the SCU replacement. Change the Dee
voltages if necessary.
492 13 RF system
13-9 Document references
GE HEALTHCARE MINItrace SERVICE MANUAL – MAINTENANCE
REVISION 18 DIRECTION 2233000-100
14 EXTRACTION SYSTEM
One or two (optional) foil holder assemblies guided by a beam orbit-shaped rail provides a
flexible extraction mechanism. The extraction system has two extraction modes:
• Single-beam extraction: one beam is guided to one out of five target positions.
• Dual-beam extraction (optional): two beams are guided simultaneously to two out of six
target positions (Note that the second beam is always directed onto target number six).
14-2 Overview
Figure 14-1: Extraction system
The extraction units move along a common curved guide plate. Each extraction unit carries
a foil holder. The guide plate is shaped in such a way that the position of the foil, regardless
of which beam exit is chosen, follows the curvature of the beam orbit at the final energy.
The single extraction drive unit moves the single extraction foil holder along the guide plate,
guiding the extracted beam to one of five exit ports.
The dual extraction drive units move the dual extraction foil holder along the guide plate, and
radially, guiding the extracted beam to the sixth exit port.
The drive units, which include motor and position potentiometer, are placed on the outside
of the vacuum chamber. The drive shafts penetrate the vacuum chamber wall by a double
vacuum seal arrangement.
Microswitches define the operating range of the extraction system. The GCU (General Control
Unit) is used to (via the MSS, MINItrace Service System) manually recalibrate the extraction
positioning after maintenance. The microswitches are used as references. Centering of the
beam in each target is performed by the GCU. Current readings from the beam collimators
are converted to fine-adjustment signals to the drive unit.
The foil carousel and the extraction foils are radioactive after production runs.
Always handle used parts in accordance with federal, state and/or local regulations.
14-2-2 Specifications
—
The drive unit contains motor, two limit sensors and a potentiometer for indication of the
position. The foil holder itself carries two extraction foils and can be flipped in position for
each foil by a brackets mounted on the guide plate near both end positions.
The software in the GCS controls the movement of foil holder arm. The movement can be
calibrated and the foil position can be flipped from the MSS.
The foil holder can be mounted in position or taken out in a few seconds.
For instructions about how to load new foils, see Section 14-8-4 Dual extraction unit on
page 503.
Always wear protective clothing, foot protection (shoes with steel toe cap), gloves and
goggles.
Always handle used parts in accordance with federal, state and/or local regulations.
14-3-3 Specifications
—
One drive unit, together with a spring, moves the foil holder sideways. An other drive unit,
also together with a spring, moves the foil holder in or out.
The drive units contains motor, two limit sensors and a potentiometer for indication of the
position. The dual extraction foil holder only contains one extraction foil. The software in the
GCS controls the movement of foil holder. From the MSS the movement can be calibrated.
The foil holder can be mounted in position or taken out in a few seconds.
For instructions about how to load new foils, see Section 14-8-3 Replacing the single extraction
unit foils on page 499.
Always wear protective clothing, foot protection (shoes with steel toe cap), gloves and
goggles.
Always handle used parts in accordance with federal, state and/or local regulations.
14-4-3 Specifications
—
14-5-3 Specifications
—
14-8 Disassembly/reassembly
Always wear protective clothing, foot protection (shoes with steel toe cap), gloves and
goggles.
Always handle used parts in accordance with federal, state and/or local regulations.
For Lock-Out and Tag-Out (LOTO) procedures of the equipment, see Section 3-5-6 LOTO
procedures for the MINItrace system on page 59.
• Ohm-meter
• Tweezers
• Gloves
• Lead glass
2 See instructions in Section 10-7-1 Open the vacuum chamber on page 295.
2 Check that no tools or cleaning tissues are left in the vacuum chamber or on the magnet
yoke.
3 Wipe off the chamber O-ring with your finger (gloves on!).
5 Close the vacuum chamber, see the instructions provided in Section 10-7-3 Closing the
vacuum chamber on page 299.
14-8-3-1 Tools
Screwdriver
Plastic gloves
Note!
The foils are extremely fragile and can only be cut with a pressing motion NOT sawing.
1 2 3 4
3 Place the foils in the gap between the retainer and the holder using a forceps.
5 Turn the foil assembly upside down, as shown in Figure 14-7 (B). Then loosen the screw
to let the foils align with the foil holder edge.
A B
6 Tighten the screw and then remove the excessive foil material outside of the foilholder.
2 Check that the foil holder is not grounded by using a standard Ohm-meter:
d Check that the resistance between foil holder and vacuum chamber is at least
500 kΩ.
14-8-4-1 General
The two halves of the dual extraction foil holder are held together by knurled brass knobs
and a threaded steel rod. The knobs are machined to fit over the raised center section of the
holder. A C-clip fits into the groove in the recess at both ends of the assembly. When you
loosen the knurled knob on the threaded end of the rod, the C-clip tension spreads the two
halves of the holder, creating a gap to hold the foil.
Remove the foil holder by pulling it out of the guide tube. It is kept in place by a spring-loaded
screw. Place the foil holder in the container.
3 Use the forceps to pull the old foil out of the holder.
Place the used extraction foil in a disposal bag for radioactive waste.
Cut the foils by pressing a sharp razor blade against the cardboard or plastic plate.
CAUTION!
The extraction foils are extremely fragile. Handle with care. Do NOT saw the blade back
and forth.
Cut one 9 mm long piece of foil from the strip for the dual extraction holder. (Refer to drawing
CAC204-34077-xx-4 for foil dimensions.)
9 mm x 16 mm piece of foil
Carefully insert the new extraction foil into the foil holder:
1 If necessary, use the tip of the screwdriver blade to lift the corner of the foil.
2 Slide the back part of a pair of forceps beneath the center of the foil.
3 Insert the foil into the gap and center it in the opening.
4 Use the edge of the metal bar to gently even the foil with the edge of the holder.
Squeeze the two halves of the foil holder together, then tighten the knob.
Make a final check of the extraction foils in the foil holder. Use an optical magnifier in to
see if there are cracks in the extraction foils.
2 Check that the foil holder is not grounded by using a standard Ohm-meter:
d Check that the resistance between foil holder and vacuum chamber is at least
500 kΩ.
f Verify that the resistance between the cable shoe and the vacuum chamber still is
20 kΩ.
1 From the Extraction page, in the MSS set all drive units (Extraction motor 1, Extraction
motor 2 and Balance) to 50%.
7 Select Motor 2.
Calibration is finished.
The target media is brought to the target through tubes from the Target Support Cabinet
(TSC).
After production, the radioactive media is brought to the hot cell(s) through tubes.
Note!
The 18F- Nb 25 target only requires water cooling.
Note that the helium cooling compressor is a part of the target system.
• 11CO
2
• 15O
• 13N
The gas targets comprises a target body, a front flange and a helium cooling flange. The
liquid targets comprises a rear flange with a silver body insert, a helium cooling flange and
a front flange (18F- and 13N targets, 3.9 mm, 6 mm and 12 mm) or a rear flange, a niobium
body and a front flange (18F- Nb 25 targets). The front flange will guide the target into correct
position of the cyclotron target insulated connection bayonet. The bayonet mechanism
assures rapid and simple mounting and dismounting of the target.
All target services, cooling water and high speed recirculation helium gas for foil cooling are
delivered to the target through quick connections.
Note!
The 18F- Nb 25 target only requires water cooling.
Note!
The 18F- Nb 25 target requires no helium cooling.
The helium cooling system is a closed system for cooling of the target foils. A high-flow helium
jet flows between the foils of each target. Heat from the foils is transferred to the helium gas.
In a heat exchanger, the helium gas heat is dissipated to a cooling water circuit.
The system includes the piping and valves for helium gas distribution, a heat exchanger, and
a compressor to make the helium gas to circulate in the system at high speed.
3 Helium cooling assembly, located beneath the magnet. See Figure 15-3.
A block diagram of the helium cooling system is shown in Figure 15-4. Figure 15-3 shows the
components on the helium cooling assembly.
The helium flow is divided into two circuits: target 1 to 5 and target 6. The compressor
maintains a high gas flow through the targets. A pressure transducer (23) monitors the
pressure of the cooled gas. The targets are connected to the helium cooling system by solenoid
valves during beam operation.
A filter in the return line from the targets removes water from the helium. When necessary,
helium is refilled from the gas bottle.
The GCU controls the solenoid valves and monitors the pressure signal via interface relays
for the solenoid valves in the CSC.
WARNING!
Vent the helium tubing before servicing the helium cooling compressor.
The helium cooling tubing and the filter can contain fragments of radioactiv foils which
should be removed from the workspace as soon as possible to reduce hand/finger dose.
Dispose of the foils in accordance with federal, state and/or other local regulations for
radioactive waste.
See also Section 15-5-1-2 Helium cooling system leak check on page 514.
For Lock-Out and Tag-Out (LOTO) procedures of the equipment, see Section 3-5-6 LOTO
procedures for the MINItrace system on page 59.
Note!
If the system is only equipped with 18F- Nb 25 targets, maintenace on the helium cooling system
is not necessary.
15-5-1-1 Filter
1 Empty any water contents of the filter body by opening the bottom screw slightly.
2 Check if the filter body is dirty. If necessary, unscrew the filter body and clean or replace
the filter.
3 If the pressure falls under 16 psi, the system is refilled from the helium gas bottle (see
Figure 15-4). The time between each refill should not be shorter than 2 minutes.
4 If the time between each refill is shorter than 2 minutes, check the helium cooling system
for leaks.
A set of 8 uniquely marked spares with every active product line are run during installation.
The lines run through a plastic conduit with shallow bends, to facilitate the future replacement
of spares.
When the system is running short of spare transfer lines (Tefzel tubes), a new set of lines must
be inserted. This procedure is a part of Planned Maintenance.
The lines run through a plastic conduit with shallow bends, to facilitate the replacement of
spares.
Note!
Remember to mark the ends of each line before you pull the lines.
1 Tape the end of the polyurethane or Tefzel tubing, to prevent any debris from entering
it while you push/pull it to its destination.
2 Fish a line through the conduits between the cyclotron room and the hot lab.
Note!
Do not use excessive force! Avoid pulling in one single Tefzel tube, since it may be deformed
(reduced inner diameter).
4 Decide which line(s) to you will use after PM, and label both ends with a piece of masking
tape.
5 Run 8 spares, for a total of 9 lines for each target on the system.
Hot lab Hot lab conduit 1/16" Tefzel Refer to the system pipe work drawing.
Run 8 spares, for a total of 9 lines for each target on the
system.
Following a production run, the target and associated tubing contain radioactive gas.
For Lock-Out and Tag-Out (LOTO) procedures of the equipment, see Section 3-5-6 LOTO
procedures for the MINItrace system on page 59.
2 Insert the tube into the connector until it passes the shoulder inside the tip, and extends
beyond the outer tip of the connector by 1 to 2 mm.
3 Screw the threaded end of the connector into the threaded hole on the target or coupling.
1. 2.
15-6-3 Targets
For disassembly and reassembly instructions for targets, refer to MINItrace Qilin Operator
Guide (dir. 2232995).
CAUTION!
There will always be some water spillage when opening up the water cooling circuits.
Always keep a bucket and a cloth nearby to catch any spilled water.
Preparations
Always wear protective clothing, foot protection (shoes with steel toe cap), gloves and
goggles.
Always handle used parts in accordance with federal, state and/or local regulations.
1 Close the water valves for TARG S, TARG and TARG D (dual systems only) in the CSC
cabinet.
3 Remove the target from the cyclotron system. If necessary, place the target in a lead
container to decay before starting to work on it.
The quick connectors are sealed with Loctite 242 or similar, so some force might be
necessary when unscrewing them.
CAUTION!
2 Apply Loctite 242 or similar to the threads of the new quick connectors and screw them
into the rear flange by hand.
Note!
• Make sure the quick couplings do not protrude more than 17.5 mm from the rear
flange.
• Make sure to orientate the quick couplings so that the opening devices are easy to
access.
Finishing off
If the water cooling tubes are to be replaces as well, proceed to Section 15-6-7 Target water
cooling on page 521.
1 Attach the target to the cyclotron system and connect the water cooling tubes to the
rear flange quick connectors.
2 Open the water valves for TARG S, TARG and TARG D (dual systems only) in the CSC
cabinet.
3 Start the water-cooling by selecting the applicable target in the MSS. Check for water
leaks, especially around the couplings. Tighten further, if necessary.
– Master internal fault in start unit, wrong process-id at the Master system
– Master internal fault in start unit, wrong chem type at the Master system
If you do a reconfiguration from the MSS (Service System) you need to press “Save to GCS”
on the CONFIG PARAMETERS menu, otherwise you do not save the new configuration.
For spare part numbers, refer to MINItrace Spare Parts Catalog (dir. 2380006-100).
1 Make sure the target is disconnected, this is done from the MSS.
3 Remove the two compressed air tubes by pressing the blue shoulders and pull out the
tube.
If you, for some reason, need to disassemble the Valve assy, please mark the position of the
brass-coupling as its possible to reassemble it the wrong way.
1 Attach the valve assy, but do not tighten the screws, the unit should be a little bit loose
so it can be aligned with the syringe.
3 Connect the two compressed air tubes. Push them in and verify that the blue shoulder
comes out a few millimeters.
Also the syringe fill and empty times can be adjusted. For 1 ml the fill time should be about
10 seconds and the empty time about 4 seconds. Too fast filling could pop the needle out of
the vial and too slow filling means insufficient water to the target. For more details about
fill/empty times, see MINItrace Qilin Installation Manual (dir. 2232994-100).
The valve panel is located down at the left-hand side of the targets, inside the shield doors.
If any of the Valve assys need to be replaced, this is the procedure:
3 Remove the two compressed air tubes by pressing the blue shoulders and pull out the
tube.
On the Valve Panel you also find the PEEK filters. Refer to MINItrace Qilin Operator Guide
(dir. 2232995).
CAUTION!
There will always be some water spillage when opening up the water cooling circuits.
Always keep a bucket and a cloth nearby to catch any spilled water.
15-6-7-1 Tubes
Preparations
1 Close the water valves for TARG S, TARG and TARG D (dual systems only) in the CSC
cabinet.
3 Remove all radioactive targets from the cyclotron system and place in a lead container
to avoid extra dose.
2 Disconnect the tubes from the manifold by pushing down the ring on the connector and
pulling out the tube.
3 Plug the connectors on the manifold using the supplied stop plugs.
4 Mount the new quick connections on the new tubes by inserting the tube into the ferrule
nut and then attaching a quick connector fitting. Use a wrench or similar tool to tighten
the coupling firmly.
6 Route the new tubes behind the retaining screws in the shield plastic.
Note!
Make sure to leave enough slack to move the manifold for service access.
8 Mount the check valve on the return line (OUT), see Section 15-6-7-2 Check valve on
page 522.
9 Remove the plugs from the manifold and connect the new tubes according to labelling.
Finishing off
If the target rear flange quick connectors are to be replaces as well, proceed to Section 15-6-3-1
Target rear flange quick connectors on page 516.
1 Attach the target(s) to the cyclotron system and connect the water cooling tubes to the
rear flange quick connectors.
2 Open the water valves for TARG S, TARG and TARG D (dual systems only) in the CSC
cabinet.
3 Start the water-cooling by selecting the applicable target in the MSS. Check for water
leaks, especially around the couplings. Tighten further, if necessary.
3 Remove all radioactive targets from the cyclotron system and place in a lead container
to avoid extra dose.
1 Unscrew the old check valve from the target water cooling line.
2 Pay attention to the marking on the valve and make sure to mount the new check valve
with the arrow pointing towards the manifold. Tighten firmly.
Finishing off
1 Attach the target(s) to the cyclotron system and connect the water cooling tubes to the
rear flange quick connectors.
2 Open the water valves for TARG S, TARG and TARG D (dual systems only) in the CSC
cabinet.
3 Start the water-cooling by selecting the applicable target in the MSS. Check for water
leaks, especially around the couplings. Tighten further, if necessary.
The helium cooling compressor weighs 25 kg. Two people are needed when lifting it.
15-6-9-1 Tools
• Standard hand tool kit
• Torque wrench
WARNING!
Vent the helium tubing before servicing the helium cooling compressor.
The helium cooling tubing and the filter can contain fragments of radioactiv foils which
should be removed from the workspace as soon as possible to reduce hand/finger dose.
Dispose of the foils in accordance with federal, state and/or other local regulations for
radioactive waste.
1 Shut down the cyclotron from the MINItrace Service System (MSS).
2 Close the helium supply used for the helium cooling system.
3 Switch off the circuit breaker for the helium compressor at the Cyclotron Support Cabinet
(CB4).
4 Vent the helium lines to remove pressure by click EMPTY in the He-cooling-box from the
MSS.
5 Mark the tubes to the helium cooling compressor and disconnect the tubes from the
compressor.
6 Remove the five cover retaining screws and lift off the top cover.
8 Remove the filter/muffler elements 3 and the filter element 4. Wash the filter elements
with isopropanol.
9 Loosen four socket cap screws and remove the pump head 5.
10 Loosen two Phillips screws and remove the retainer plate 10.
3 Place the retainer plate 10 over the center of the diaphragm. Fasten the Phillips screws
and tighten to 3.4 Nm (30 lbf*in).
4 Mount the pump head 5. Fasten the socket cap screws and tighten to 10.2–11.3 Nm
(90–100 lbf*in).
6 Place the gasket 2 and top cover 1 on the pump head. Fasten the cover retaining screws.
8 Fill the helium cooling system from the MSS as described in the next section.
If the helium cooling circuit is filled with only air or just a portion of helium there is a possibility
that the air will become radioactive ozone. The ozone will damage O-rings and membrane
in the helium cooling circuit. The purpose of the helium filling routine is therefore to make
sure that there is only helium in the circuit.
5 Repeat the Flush-Fill-Empty-sequence three times, but end with a Fill, to make sure that
the helium-system is free from air.
6 Repeat steps 2–7 for all target positions to make sure that all targets are filled with helium
only.
7 Select one target and check that the pressure is OK (see Section 15-5-1-2 Helium cooling
system leak check on page 514).
8 Unselect targets.
• Assembly instruction
All removed parts or waste that is radioactive must be handled and disposed of according
to federal, state, or local laws and restrictions.
Action Check
Note!
Keep the following in mind before, during and after work with radioactive equipment.
• Plan your work; in other words think first then take the actions.
• Measure the dose rate and calculate how long you can work.
• Stay as short time as possible in the radiation area, to reduce the accumulated dose.
• Use protection if possible, however you shall always wear protective goggles, gloves,
TLD ring, TLC badge and personal radiation monitor.
• After performed actions, compare actual pick up of dose with estimated dose.
1 If dose rate and working conditions are satisfying. Continue with work as described
in op. 2 to last op.
2 If dose rate and working conditions are not satisfying, apply necessary safety
precautions before further action. Then continue with work as described in op. 2 to
final op.
Action Check
2. Shut off the helium and water-cooling supplies for the target
cooling.
6. Dismount the blind flange on the position where the new target
is going to be mounted.
7. Check that the sealing area is clean, and that the O-ring surface
is free from scratches.
9. Mount the collimator (3) on the rails with screws (14) and
washers (16).
10. Adjust the collimators with the tool, see Section 16-3-1-3 Install
the collimator and resize the opening on page 550.
11. Slide the shield onto the target and fasten cover with nylon
screw. The screw facing upward when the connectors are
facing towards you.
Action Check
13. Disconnect all tubes and wires to the LTF panel. Mark them
with destination.
14. Dismount the LTF panel from the cabinet. Note that all media
to the LTF panel must be shut off, helium, compressed air etc.
Put the panel on a suitable plane surface.
15. Fasten the dispenser assys on The LTF panel with three hex
screws and washers.
Action Check
16.
Action Check
17.
Action Check
Action Check
Action Check
19.
Fasten the 2-way valve assy on the target panel with the wing
screw. Fasten connectors for trap assy and gently push the
assy on place. Note The 3-way valve is positioned on the panel
on older machines. On newer machines it is positioned in the
hotcell. Fasten the PEEK cross with two screws and distances
on the panel. Fasten the connectors to the check valves push
the valve in place, note that the arrow on the valve should face
uppwards. The tubes are connected according to CAC208-
31018 and instructions below.
Action Check
21. Put the LTF panel back on its place in the cabinet. Start with
connect the marked wires and tubes you disconnected before
loosing the LTF panel.
Action Check
22. Make the tube connections between the valve panel and the
LTF. All destination valve connectors are clearly marked with
the same label names as the source connectors. Do not push
all the tubes/lines through at once. Push and connect one
tube/line at a time to prevent mistakes.
1 Choose an LTF source connection and push its tube
through the TSC shield penetration.
2 Cut a small piece, about 5–10 mm, at the end of the tube.
There may be some dirt from the shielding inside it.
Make the Tefzel lines between the LTF panel and the cor-
responding valve panel connector. Refer to drawing
CAC600-17000.
5 Cut a small piece, about 5–10 mm, at the end of the tube.
There may be some dirt from the shielding inside it.
Make the PEEK line connections between the LTF and the valve
panel. Refer to drawing CAC208-31018.
Do not bend the PEEK lines; they may be cracked.
23.
Add one water valve and one helium valve to the target cooling
manifold, on the target position you have chosen. Connect the
signal cable from the valves on the connection panel. Refer to
wiring diagram CAC200-54001.
Action Check
24. Connect the tubes to water and to the helium on the valves.
Mount the connectors on the target side on the tubes. The
water tubes shall be 40 mm shorter. Note the connectors with
backing valve are for the water. Make sure that the flow direc-
tion is correct. Pressure is on the outer connections, return on
the inner connections on each block. The black tubes are for
helium and the blue is for water.
25. Lead the PEEK tubes from the valve panel in the raceway up
to the target. Connect with the PEEK connectors. Finger-tighten.
Note the length on 0.75 PEEK tube from trap assy to target shall
be 3 m long.
26. Connect the target current sensor on the connection panel and
lead it in the raceway up to the target. Use the screws as retain-
ers. Refer to wiring diagram CAC200-54001.
27. Put cable 5457134 (904783) in the cable duct on top of the cy-
clotron. Connect to CSC connection board P11 and LTF P3.
Action Check
Action Check
5 Fill helium cooling circuit, shut off valve. Wait 10 min check
pressure drop.
Helium cooling Oilless diaphragm compressor GAST See the GAST Operation & Mainte-
compressor nance Technical Manual in MINItrace
Service Manual – OEM Manuals
(dir. 2233002-100).
16 BEAM DIAGNOSTICS
16-1 Introduction
The beam diagnostic system monitors the beam current at different positions in the cyclotron
and the target system in order to control the beam from the ion source to the target. The
beam diagnostic system consists of the following elements:
• The internal flip-in type of probe at a small radius, close to the center of the cyclotron
that gives information about the beam current in the IN-position.
• The extraction foils, single and optional dual extraction, where the current from the
stripped-off electrons are monitored
• The up/down collimators at each beam exit which monitors the beam position by giving
feedback to the diagnostic system. They also define the beam boundaries.
• The target bodies which monitors the beam current. A filter-box outside the vacuum
chamber filters the beam current signal from unwanted RF.
• The Beam Current Analyzer (BCA), located in the GCU, where the current from the different
readings is analyzed.
The flip-in probe, the extraction foils and the beam collimators are radioactive after
production runs.
Always handle used parts in accordance with federal, state and/or local regulations.
A coil drives the mechanism for the probe motion. To activate the probe an external current
source drives the coil and together with the magnet field, this generates a force on which
brings the probe into the beam path. Consequently, the probe will not work if the main magnet
field is not activated.
A signal to the GCU monitors the IN-position of the probe. The probe will automatically fall
back into OUT-position when the current through the coil disappears.
The beam current information coming from the probe verifies that the beam current is at the
expected value and sufficient for isotope production. A satisfactory beam current level on
the probe verifies that most of the accelerator subsystems work normally and that the
condition of the ion source is good.
16-2-3 Collimators
At each of the six beam exits on the cyclotron vacuum chamber, a pair of beam collimators
is mounted. The beam collimators have two functions:
The size of the hole in the collimator defines the outer boundary of the beam and
consequently also the size of the beam spot hitting the target foils. The intercepted beam
will be picked up by the collimators and represents a small portion of the total target
current.
The collimators are insulated from ground. The BCA (Beam Current Analyzer) measures
possible unbalance in intercepted beam currents on the upper and lower collimator and
also corrects the position of the beam until the collimators read the same value.
16-2-4 Targets
The targets are insulated from ground in order to enable measuring of the isotope beam
current. Also see Chapter 15 Targets and radiochemistry system.
For a detailed description of the BCA, see Chapter 7 General control unit.
16-3-1 Collimators
As a part of Planned Maintenance, the beam collimator openings are measured (vertically
and horizontally) and adjusted if needed.
• all collimators
CAUTION!
Always wear clean cotton or non-powdered examination gloves when working in the
chamber. Take care NOT to scratch the interior surfaces of the chamber.
WARNING! Radioactivity
Always measure the radioactivity when you open the vacuum chamber door, to know
how long you can stay in the chamber area. Follow the dose limit regulations in the
country.
2 Shut off the system (if on) and open the shield doors.
3 Locate the ion source tool box (often stored in the empty TSC cabinet) and take out the
paper burner target and the collimator adjustment tool.
4 Remove the first target from the target mount, and replace it with the paper burner
target. Leave the cooling lines attached to the original target.
5 Attach the target’s current sensing line to the paper burner target. Make sure the paper
burner is firmly seated in the bayonet mount.
WARNING! Radioactivity
You can only leave targets in place in the shield area if you have not bombarded it yet.
If you have already bombarded the target, you must remove it and store it inside lead
shielding.
6 Close the shield doors and press PUMP to start the vacuum system.
7 Attach the service laptop to the MSS port on the CSC panel. Go to the MAGNET SYSTEM
page and click ON to turn on the magnet.
8 Set the current to 132 Amps. (Type in a value, do not use the arrows.)
Password: scxscx
11 Set the Foil holder 1 position to the value shown for the corresponding target position
in the Configuration Parameters window.
12 When the RF SYSTEM standby goes to the Ready state, click on RF NORMAL.
Wait for Dee volts OK to stabilize. (No sparking allowed during this test.) If necessary,
return to Section 13-3 Overhaul of Tube Amplifier and RF system tuning on page 437 and
condition the system.
Make sure the RF SYSTEM window still shows Dee volt, ok.
14 Select ION SOURCE (IS ON) and set the Ion source to 100 mA. Wait for Dee volts to stabilize.
15 Adjust the ion source current to about 4 microamps on the probe, because the paper
burner does NOT have cooling water lines.
16 Make the paper targets out of business cards while waiting for the Dee volts to stabilize:
c Draw a horizontal line 19.5 mm above the bottom edge of the card.
d Draw a vertical line 19.5 mm from the right edge of the card.
e Make about 6 cards per system target, including the dummy port.
WARNING! Radioactivity
Wear gloves and protective eye wear, and use forceps to install the target papers. Use
tools to shift the paper burner target to a new port, and to adjust the collimator blades.
Do not touch the collimator blades or bombarded paper burner target with your (gloved)
hands.
Click the MAGNET UP and DOWN arrow keys to optimize the foil 1 current reading.
19 Adjust the foil holder positions in small steps (10 to 20%) while watching the waveform
on the BEAM CONTROL page.
a Click the Foil holder 1 pos UP arrow to shift all current to the upper collimator.
b Record the upper foil 1 position value shown in the Foil 1 number: field.
c Click the Foil holder 1 pos DOWN arrow shift all current to the lower collimator.
d Record the lower foil 1 position value shown in the Foil 1 number: field.
WARNING!
Take care! The magnet is still ON. Remove all credit cards and mechanical watches. Use
forceps to insert and remove the paper targets.
22 Write the target number, burn trial number and date along the top of the card.
23 Open the left shield door and insert the paper target into the bottom of the burner with
the cross facing the beam.
26 Click IS ON.
Read the probe current; make sure RF Frequency is the same as before. Wait for the Dee
volts to stabilize.
You should see all the beam current on the upper collimator.
29 Type/enter the foil 1 position value you recorded for the lower collimator value into the
Foil 1 number: field, to switch the current to the lower collimator.
30 Type the upper collimator value, then type the lower collimator value into the field 3 or
4 times for each value, to sweep the beam back and forth between the collimators.
You may need to experiment with the number of times you enter the values. The number
of times needed to sweep the beam across the collimator blades will increase as the
paper thickness increases. Try not to over burn (blacken) the paper. It is better to under
burn than over burn the paper target.
WARNING! Radioactivity
Wear gloves and protective eyewear when you open the target shield door and remove
the paper. Always neasure the activity when you open the shield door and vacuum
chamber doors.
32 Use forceps to remove the paper from the paper burner target.
• The burn mark on a liquid target should be centered horizontally, but slightly lower
than center on the vertical axis.
• The burn mark on a gas target should be centered directly over the crosshair.
WARNING! Radioactivity
The paper is radioactive (carbon isotope). Do not touch the paper with bare hands. Store
the papers in a lead container for at least 2 hours. If possible, leave the paper in the
container over night.
34 Repeat the test two or three more times at the current target location, to check for
consistency, before moving on to the next target port.
• IF the beam is too far above or below the horizontal line, proceed to Section 16-3-1-2
Reposition the collimator opening on page 549.
• IF the beam position changes each time you repeat the test, call support.
35 Test each occupied target port on the system, including position five (proton target). Use
a tool to shift the Paper Burner from one position to the next.
WARNING! Radioactivity
Remember, the paper burner target becomes increasingly radioactive the longer you
use it. Always use a tool to move the paper burner from one target port to another. When
you finish the test, store the paper burner in a lead container.
Liquid targets are pressurized with helium to increase the boiling point of the target liquid.
The gas forms a bubble at the top of the target chamber. The collimator is positioned to direct
the beam below the center point of the target, to avoid this bubble.
On gas targets, the collimator is positioned to direct the beam to the center of the target.
1 Measure the collimator opening with calipers to determine the factory setting. The
standard collimator diameter is about 6 mm, but could be as large as 8 mm.
• IF the top of the liquid target burn mark -or- the center of the gas target burn mark
falls below the horizontal line, raise the entire collimator.
• IF the bottom of the burn mark -or- the center of the gas target burn mark falls
above the horizontal line, lower the entire collimator.
a Measure the distance between the actual and optimal burn mark location.
4 Use metric drill bits to reposition the collimator while maintaining the original diameter.
Remember to keep the drill bits perpendicular to the angled target port when you insert
them into the collimator opening to make the adjustments. (See Figure 16-2 and the
example, below.)
Example:
The burn mark is 2 mm too high, and the collimator diameter is 6 mm. To shift the
collimator down 1 mm:
b Insert a 7 mm drill bit into the opening, move the bottom blade against the bit and
tighten into place.
d Insert a 6 mm drill into the opening and move the top blade against the bit.
WARNING! Radioactivity
Wear gloves and protective eyewear when you open the target shield door and remove
the paper. Always neasure the activity when you open the shield door and vacuum
chamber doors.
2 Click SYSTEM OFF on the BEAM CONTROL page to shutdown the cyclotron in an orderly
manner.
4 See Figure 16-6. Attach the collimator adjustment tool to the bayonet mount.
5 Use a 3 mm hex wrench to loosely fasten the collimator blades into place.
6 Use calipers to measure interior diameter of the collimator. Start with the standard
collimator opening of 6 mm in diameter.
7 Slide the center pin in and out of the tool while you make the adjustment.
• Slide the pin into the tool to check the center of the collimator blade opening.
8 After you size the collimator opening, remove the collimator adjustment tool, and replace
it with the paper burner target.
10 Place a new paper target into the paper burner, and close the shield door.
14 Set the current to 132 Amps. (Type in a value, do not use the arrows.)
Make sure current goes up to 150, stays there, then goes back to the set value. Otherwise
the beam may not reach the target.
Password: scxscx
18 Click IS ON.
3 Open the vacuum chamber. See Section 10-7-1 Open the vacuum chamber on page 295.
WARNING! Radiation
Internal parts of the vacuum chamber are radioactive after production runs.
Always handle used parts in accordance with federal, state and/or local regulations.
5 Remove the two upper and the two lower screws and washers. See Figure 16-7.
Note!
If you are replacing tantalum collimators with graphite collimators, the isolators also need to
be replaced. See Section 16-3-1-5 Replacing isolators (graphite collimators) on page 554.
2 Fasten the new collimators with the two upper and the two lower screws and washers.
Note!
Be careful not to overtighten the screws when fastening graphite collimators. The graphite
collimators are brittle and might break.
3 Check with a multimeter, adjusted to measure resistance, that the collimator is connected
to the GCU. The multimeter will show 20 kΩ if things are correctly connected.
4 Keep the multimeter connected as in the previous step. Check on the MSS screen that
there is a vertical bar coming up on the diagnostic page.
5 Close the vacuum chamber, refer to the instructions provided in Section 10-7-3 Closing
the vacuum chamber on page 299.
Start by removing the collimators for the dual option, if fitted on the system.
Continue with replacing the single option isolators before fitting new dual option isolators.
Note!
It might be necessary to remove one of the screws that holds the L-profile to the chamber
wall to get better access to the screw that holds the rail.
Note!
Be careful when tightening the screw on the new isolator. The isolators are brittle.
Note!
Be careful when tightening the screw on the new isolator. The isolators are brittle.
7 Remount the screws that holds the L-profile to the chamber wall, if removed.
Note!
Be careful when tightening the screw on the new isolator. The isolators are brittle.
Note!
Be careful when tightening the screw on the new isolator. The isolators are brittle.
6 When both upper and lower rails are fitted, check that the upper and lower rail does not
collide. See Figure 16-11. Adjust if necessary.
1 Move (by hand) the coil until the probe is in its PROBE IN position, and let it return by itself.
2 When the probe has find back to its PROBE OUT position, measure the distance to
38–40 mm between the pole valley and the probe, see Figure 16-12.
If the distance is out of tolerance, loosen carefully the two (one on each side) allen screws
and turn the coil while the probe is kept at the correct distance. Tighten the screws and check
the function again.
Check that the foil holder is not grounded by using a standard Ohm-meter:
4 Check that the resistance between foil holder and vacuum chamber is at least 500 kΩ.
a Connect the negative probe of the multi-meter to ground and the positive probe to
a target.
c At the MSS–Read the current for Target 1, Target 2, Target 3, Target 4, Target 5,
Target 6 and check that the MSS shows the same values, that is ~10 µA.
17 RADIATION SHIELD
The roller bearing hinges are of virtually non-friction type and the necessary force to drive
the door is low and limited to avoid any risk of pinch hazard.
Working on parts on top of the radiation shield or standing on a ladder involves potential
fall hazard.
For instructions on using ladder and safety harness, refer to Section 17-7 Ladder and climbing
safety on page 564.
2 Press the Left Door Open and Right Door Open switches to fully open the radiation
shield. Press the close-switches for closing the doors.
Personal injury can be caused when the radiation shield doors move.
The door pendant control must be disconnected from the CSC during service on the
cyclotron.
Note!
A foldable combination ladder is included in the MINItrace delivery from October 2013.
The ladder should be found hanging close to the cyclotron.
Note!
Maximum allowed load on the ladder: 150 kg
3 Make sure that the ladder step (second step from the top) is fully secured in the adapters.
Note!
The topmost step must not be used for securing the ladder.
• Use the two safety wires on top of the shield to anchor the harness, see Figure 17-6.
• Hold on to the ladder when reaching for a safety wire to anchor the harness.
• When anchoring the harness, always use the closest securing point located in a straight
line of pull compared to where you are working.
18 SAFETY SYSTEM
The cyclotron includes a number of hard-wired interlock systems to guarantee safe system
operation for the personnel.
The drawing principal diagram over safety interlock and other CIB signals (CAC000-51001)
includes the following safety interlocks:
• RF inhibit
Machine protection is provided by an extensive use of interlock signals. The interlock functions
are designed to protect the equipment in most operating situations and in cases of
malfunction.
The cyclotron system is however only a part of a total PET system and some of the safety
systems must be provided by the customer. These safety systems may include facility services
such as, ventilation, fire protection and customer-supplied power and water systems
connected to the cyclotron.
As the cyclotron systems are sold in a wide range of countries with different regulations, it is
not possible to recommend safety systems applicable everywhere.
Note!
It is assumed that all personnel working with the equipment has participated in adequate
training to acquire the necessary knowledge of the risks involved and how to plan the work to
reduce the possible risks.
Figure 18-1: Interlock signals from magnet door and shield doors
• Two door micro switches (left and right door) on the radiation shield indicate if the shield
doors are open or closed, see Figure 18-2. If the system is run from the Master System,
there is no power from the PSARC or RFPG, or to the ion source gas valve if the doors
are open.
• Magnet door micro switch. Indicates if the magnet door is open. There is no power to
PSARC and RF is inhibited if the magnet door is open.
• Signals from the GCU to the BEAM ON/PSMC ON warning light on the top of the cyclotron,
see Figure 18-2. These signals could also be used by the customer for several warning
signs.
• A software controlled maintenance signal from the MSS (MINItrace Service System) allows
full operation of either PSARC or RFPG for service purposes.
• A Start PSARC signal from the GCU, generated by software. Allows main power to the
PSARC.
• Five extra interlock signals to the GCU board. Could be used by the customer for various
devices. They can also generates messages on the Master screen.
The shield door switches are located below the warning light.
Figure 18-3: Magnet door switch and ion source gas valve
• Warning lights
Note!
The RFPG safety interlocks must be tested when a number of selected RFPG-related spare parts
have been replaced, see Section 13-6-1 Spare parts requiring safety interlock test on page 447.
18-5-2 Preparation
WARNING!
Apply LOTO as applicable in accordance with GE regulations and local restrictions. For
LOTO procedures, see Section3-5-6 LOTO procedures for the MINItrace system on page59.
1 Disconnect any external safety interlock system before starting the test.
2 • Connect the External interlock switch box (see Figure 18-4) to the Customer Interface
Connector on top of the CSC.
The procedures are different for SCU Gen II and SCU Gen I (see Figure 18-5).
2 Verify that the PSARC P ON LED on the GCU front panel is lit (green).
4 Verify that the LEDs RAD SH OK, BEAM ALLOW and PSARC P ON on the GCU front panel
are turned off.
5 Verify that both the gas flow and PSARC shut off.
6 Verify that the error messages “IS: Safety Door open”, “IS: PSARC fault”, “IS: HV Relay fault”
and “Safety: Safety Door open” appear on the MSS.
2 Verify that the PSARC P ON LED on the GCU front panel is lit (green).
4 Verify that the LEDs BEAM ALLOW and PSARC P ON on the GCU front panel are turned
off.
5 Verify that both the gas flow and PSARC shut off.
6 Verify that the error messages “IS: Safety Door open”, “IS: PSARC fault”, “IS: HV Relay fault”
and “Safety: Safety Door open” appear on the MSS.
SCU Gen II: On the Main page in the SCU web interface, check that RF Inhibit 1 and/or
RF Inhibit 2 are lit (green).
SCU Gen I: On Interface board 1, check that LED 2 (RF INHIBIT) is turned off.
9 SCU Gen II: On the Main page in the SCU web interface, check that RF Inhibit 1 and/or
RF Inhibit 2 are turned off.
SCU Gen I: On Interface board 1, check that LED 2 (RF INHIBIT) is lit.
2 Verify that the PSARC P ON LED on the GCU front panel is lit (green).
4 Verify that the LEDs RAD SH OK and PSARC P ON on the GCU front panel are turned off.
5 Verify that both the gas flow and PSARC shut off.
6 Verify that the error messages “IS: PSARC fault” and “IS: HV Relay fault” appear on the
MSS.
SCU Gen II: On the Main page in the SCU web interface, check that RF Inhibit 1 and/or
RF Inhibit 2 are lit (green).
SCU Gen I: On Interface board 1, check that LED 2 (RF INHIBIT) is turned off.
9 SCU Gen II: On the Main page in the SCU web interface, check that RF Inhibit 1 and/or
RF Inhibit 2 are turned off.
SCU Gen I: On Interface board 1, check that LED 2 (RF INHIBIT) is lit.
4 Verify that the IS MAINT LED on the GCU front panel is lit (green).
5 On the Ion source page, under SYSTEM INTERLOCK indicators, verify that RF is yellow
and the rest are green.
6 Verify that it is not possible to turn on the ion source (ion source gas is though possible).
8 Verify that the RF shuts down and error messages “RF can not set standby mode” and
“Safety: Safety Door open” appear on the MSS.
11 Re-connect any customer safety system that was disconnected before the test.
12 If there are no customer safety system to re-connect, the External interlock switch box
can remain in the Customer Interface Connector.
• Verify that SW1 and SW2 on the External interlock switch box are in position CLOSE.
1 On the Beam control page in the MSS, click SYSTEM OFF to make sure that the magnet,
ion source and RF are turned off.
a RF STANDBY
The system should reach Standby Ready after about 1–2 min.
b RF NORMAL
c PROBE IN
d GAS ON
e IS ON
9 Verify that the upper warning light (beam ON) also is lit.
3 Check that the switches for the front panel, rear panel and earth stick moves freely and
do not risk getting stuck.
4 Make sure that the earth cables to the panels are attached.
The indicators are lit only when all three switches are pressed in (see Figure 18-6).
230 V mains voltage is present on the panel microswitches, terminals, circuit breakers,
and contactors behind the RFPG front panel. Components behind the rear panel operate
lethal high voltage.
Two qualified persons are required when working with the RFPG turned on and the panels
off. Use electricity-proof rubber gloves.
Do not touch high voltage components even if the RFPG is turned off.
WARNING!
Make sure that the RFPG is connected to mains, the TPSU SW 1 is ON and the RFPG state is
OFF.
2 Check that the indicator is turned off (Earth Stick OK for SCU Gen II or LED 5 (EARTH
ALARM) for SCU Gen I).
2 Check that the indicator is turned off (Earth Stick OK for SCU Gen II or LED 5 (EARTH
ALARM) for SCU Gen I).
3 Check that the indicator is turned off (Earth Stick OK for SCU Gen II or LED 5 (EARTH
ALARM) for SCU Gen I).
4 Remove the rear panel again and put the earth stick back in its holder.
230 V mains voltage is present on the panel microswitches, terminals, circuit breakers,
and contactors behind the RFPG front panel. Components behind the rear panel operate
lethal high voltage.
Two qualified persons are required when working with the RFPG turned on and the panels
off. Use electricity-proof rubber gloves.
Do not touch high voltage components even if the RFPG is turned off.
WARNING!
1 On the Main page in the SCU web interface, check that Earth Stick OK is lit (green).
4 On the Test page, click Turn on/off for TPSU Door intl. (You must be logged in as SERVICE.)
The processor now believes that the panel is in place eventhough it has been removed.
6 Start Standby.
9 On the Main page in the SCU web interface, check that SCU status is OFF.
(The system status is now OFF. The figure below is only an example showing system
status DEE VOLTAGE OK.)
11 On the Test page, click Turn on/off for TPSU Door intl. The indicator is turned off and
the interlock activated.
CAUTION!
Use suitable devices for protection against electrostatic discharge (ESD) when handling
the circuit boards.
5 On the Tube controller board, check that LED 5 (EARTH ALARM) is turned off.
The processor now believes that the panel is in place even though it has been removed.
8 Start Standby.
The procedures are different for SCU Gen II and SCU Gen I (see Figure 18-10).
6 On the water manifold, close one of the water circuits for the RF cavity.
7 SCU Gen II: On the Main page in the SCU web interface, check that RF Inhibit 1 and/or
RF Inhibit 2 are turned off.
SCU Gen I: On Interface board 1, check that LED 2 (RF INHIBIT) is lit.
4 On the Test page, click Turn on/off for Test relay 1. (You must be logged in as SERVICE.)
6 On the Main page in the SCU web interface, check that the SCU status is OFF.
(The system status is now OFF. The figure below is only an example showing system
status DEE VOLTAGE OK.)
CAUTION!
Use suitable devices for protection against electrostatic discharge (ESD) when handling
the circuit boards.
2 On the water manifold, close one of the water circuits for the RF cavity.
The processor in the RFPG now believes that all conditions for start are met eventhough
the water circuit still is shut off.
The fan and possibly the heater power supply will start.
9 Check that the RFPG does not reach STANDBY READY and that after about 1½ minute
all power supplies are turned off (the fan runs).
10 On the MSS RF page, check that heater power supply or grid power supply is yellow.
17 On the water manifold, open the water circuit for the RF cavity.
19-1 Introduction
This chapter describes the MINItrace Service System from version 2.x (MSS) software version
for Microsoft™ Windows™ 7 released in 2014.
Note!
The MSS software version that was discontinued in 2014 is described in Appendix D The
MINItrace Service System (MSS) software.
MINItrace Service System (MSS) is a software running on a laptop. MSS makes it possible for
the service engineer to manually control the accelerator, the targets and the radiochemistry
subsystems. MSS is intended for setting and monitoring parameters, trouble-shooting,
configuration, and for user administration.
• Software installation
• Software description
Note!
Before 2014, the laptop was not included in all cyclotron deliveries.
MSS controls the General Control System (GCS) via commands transferred over serial cables.
19-2-1 Introduction
The following sections describe how to connect and set up the laptop.
The service laptop can be connected to the General Control System (GCS) or the Cyclotron
Support Cabinet (CSC).
For details regarding the operation of the service laptop, refer to the accompanying operator
manual.
Installation of the MINItrace Service System (MSS) software is described in Section 19-3 MSS
software installation on page 594.
1 Install the USB-to-RS232 adapter using the instructions and CD supplied with the adapter.
If there is no driver for Microsoft Windows 7 supplied, download the driver from the
adapter manufacturer's web site.
The adapter should be connected to a free USB port on the service laptop when requested.
4 The installed port will appear as USB-to-Serial Com. Port (COM1). Any printer or parallel
ports can be ignored.
6 Click the Port Settings tab. Make sure tha the settings are correct: Baud Rate 9600, Data
bits 8, Parity None, Stop bits 1 and Flow control None, see Figure 19-2.
3 Under Select a power plan, select Balanced (or Dell is preselected), and click Change
plan settings.
Equipment
Cables to use:
• USB-to-RS232 adapter
Hardware connections
1 Connect the USB-to-RS232 adapter and the RS232 cable between a USB port on the
service laptop and the GCU port on the GCU or CSC.
Note!
If the service laptop is equipped with a serial port, the adapter is not required.
Software connection
It is assumed that MINItrace Service System is installed.
1 Check that the service laptop is connected as described in Section Hardware connections
on page 593.
3 Select LOCAL mode. Check system condition on the Master System and select Service
System control in the Maintenance menu. If the Master System is down or not connected,
LOCAL mode is automatically active.
Note!
The system returns to the Master within 6 minutes if connection with the Service System
is not established. See also below.
Unsuccessful connection
The following happens after selecting Service System control at the Master System if
connection with the Service System is not established, or if connection is interrupted:
1 minute Heating of any oven is turned off and air cooling of oven starts.
1 In the MSS main window, click Navigation > Exit to close the application.
3 The Master System automatically regains control within a few seconds, if connected.
The MSS software is pre-installed on the service laptop delivered with the system. The MSS
software is also included in the delivery on a CD.
This section describes how to install the MSS software on the laptop; the first time on an
empty laptop and to upgrade an existing installation.
Note!
Leaving the service laptop unattended a short period during the installation might abort or
cancel the installation procedure due to security functions in the operating system.
Note!
If the previous MSS version has been used before this installation, back up the configuration
parameters on a file before disconnecting MSS (see Section D-3-12 Configuration parameters
menu on page 741). The file will be a useful reference when setting up the configuration
parameters with MINItrace Service System.
3 The installation program should start automatically. Click Run setup.exe in the dialog
that appears.
4 Installation of Microsoft .NET Framework 3.5 SP1 takes place, if this add-on is not already
installed on the laptop.
5 The database installation starts. Click Accept in the license agreement dialog.
7 Wait until the black terminal window is closed (might take a few minutes).
8 When the MSS Setup dialog appears, follow the instructions and click Next.
9 Choose Complete installation. Click Next. The installation might take a few minutes.
4 The installation program should start automatically. Click Run setup.exe in the dialog
that appears.
5 When the MSS Setup dialog appears, follow the instructions and click Next.
6 Choose Complete installation. Click Next. The installation might take a few minutes.
Note!
It is possible to magnify the screen in Windows 7 Professional. Click the Start icon and select
All Programs>Accessories>Ease of Access>Magnifier.
Log view Open a more detailed action log with several filtering
options.
Task Description
Administration Create user accounts, set up system options, and export and import
default configuration values
Configuration Configure GCS and targets, and export and import configuration
values
Click Action log to select messages for User events (= user actions) and/or Xcs events (=
system actions).
To filter the messages according to severity, select Info, Warning, Error or Fatal. By selecting
Debug all messages are shown.
19-5 Administration
In the Administration pane it is possible to create, modify or delete user accounts.
When logging in the first time, there is only one user: SysAdmin (Administrator). The default
password is pw.
The user SysAdmin is allowed to administrate users, set system options and export/import
default configurations.
Login name User's login name (6–12 characters, A-Z, a–z, 0–9. Space is not
allowed.)
Password User's password (6–12 characters, A-Z, a–z, 0–9. Space is not
allowed.)
Account expire date None, or enter a date when the user account expires
Password expire date None, or enter a date when the user password expires
Failed login attempts Shows number of incorrect login attempts. The counter is reset
after a correct login.
There are four categories with different task access levels. See Table 19-1. A user can have
a combination of categories assigned to get increased access rights.
System Administrator X — — —
Click Save to save any changes or click Cancel to quit without saving.
• In COM port settings, the GCS communication parameters are set. See Section 19-2-2
Initial communication and system setup on page 590.
19-6 Configuration
Before running the cyclotron, a number of parameters must be set in the Configuration task,
for example, the positions of the targets and the host address for the Master System. This is
normally done at the factory.
Note!
The configuration parameters given in the following sections are typical values and might vary
between sites.
The user logged in must have access to the Configuration task (see Table 19-1).
• GCS (accelerator)
• Target
• Export
• Import
Option Description
Next to the bottom right corner of the pane there are four buttons:
Button Description
Default Restores the default GCS configuration. The values become valid
when clicking Save.
Save Opens a window where the updated parameter values can be saved
to the GCS with name and date label.
• Magnet
• RF
• Ion source
• Extraction
• Foil diary
• Global
19-6-1-1 Magnet
Increasing current ≈ 40 mA
19-6-1-2 RF
Scan start pos (frequency) Scanning start position for frequency 70%
Dee scale factor Scaling factor, conversion of Dee voltage, read signal into true kilo- 244
volts, shown to the operator
Phase offset 2
Operation time Time that the ion source has been used [Ah]
Lifetime Maximum length of time that the ion source can be used 1000 Ah
Max arc current Maximum current of the ion source power supply 2000 mA
19-6-1-4 Extraction
I probe startup Maximum current for the flip-in probe at ion beam startup 15 µA
Probe/foil loss max Maximum ion beam loss between flip-in probe and active foil 80%
Foil/target loss max Maximum ion beam loss between active foil and active target 20%
19-6-1-6 Global
• Target position 1–6 on cyclotron. Choice of new target type and choice of tracers –
indication/entry fields for the target system parameters.
• Click Change to open the Select target and Select tracers lists. The desired target and
tracer for the target position can be selected from these lists. The content of the Select
tracers list depends on the target selected.
• Click Save to assign the selected target and tracers to the target position.
The current parameter values are shown in the left column and changes are made in the Edit
column.
At the bottom right corner there are four buttons as described below.
Button Description
In the Configuration target menu for 18F- target, the target parameters can be changed.
18F- 18F-12 mm
3.9 mm and Gen II
6 mm 18F- Nb 25
Ltf id 2 or 3 2 or 3
Min pressure1 Minimum pressure allowed for the high pressure 250 psi 250 psi
Set pressure Minimum allowed low pressure 035 psi 035 psi
Delta pressure Minimum pressure increase allowed during filling 10 psi 8 psi2
16 psi3
18F- 18F- 12 mm
3.9 mm and Gen II
6 mm 18F- Nb 25
Max leak pressure Maximum variation of target pressure during leak check 95% 95%
Dry time Time to dry the target 020 min 020 min
1 This value is used to detect a possible leak in the target during production. There is a restriction mounted in the
high pressure supply line. This causes the pressure to drop even if the leak is very small.
2 18F- 12 mm Gen II
3 18F- Nb 25
4 This value is typically set 10 μA higher than the maximum allowed beam on target. This is because the beam
otherwise is turned off if the target current would temporarily higher (slight overshoot in the regulation or similar).
For example, the 18F- Nb 25 target is maximally allowed for 50 μA. Still the value is configured to 60 μA to prevent
a temporary current level above 50 μA to shut the system down.
In the Configuration target menu for 13N target, the target parameters can be changed.
Ltf id 1
Max leak pressure Maximum variation of target pressure during leak 20%
check
In the Configuration target menu for 11CO2 target, tracers can be selected and the target
parameters be changed.
Max leak pressure Maximum variation of target pressure during leak 20 psi
check
Receiving unit intl di- GCS interlocks for direct production 1–5
rect
In the Configuration target menu for 15O target, tracers can be selected and the target
parameters be changed.
Max leak pressure Maximum variation of target pressure during leak 20 psi
check
Configuration – Import is used for importing a previously exported configuration file to the
GCS.
The following options are available by clicking the respective part of the cyclotron image:
• Magnet system
• Ion source (and gas handling system) (click the center of the cyclotron)
• Motor calibration
• Beam control
• Target 1–6
When clicking a subsystem, the corresponding window is opened. Several subsystem windows
can be open at the same time, but all subsystems cannot be run simultaneously due to
conflicts.
Option Function
Target number 1–6 Controls the corresponding target and chemistry system.
Magnet Controls the magnet system, including the magnet power supply.
Extraction and Diag- Controls the extraction and diagnostic system, including the flip-in
nostics probe and the different foil options.
Ion source Controls the ion source, including the ion source power supply and
ion source gas.
• Magnet control and indicator section – indicator and entry fields for current and voltage.
Also includes buttons for simple control of magnet power, and a status indicator.
• Buttons:
- Reset – resets magnet error, provided the error has been corrected.
- System Off – turns off all subsystems (not vacuum and cooling).
Set Entry field/Button Enter the new current value and click Set.
Ramps the current first to 150 A, waits 30 s
and then ramps current to entered value
(max. current and wait time configurable).
Magnet water flow Green/Yellow Yellow if the water flow (water manifold) to the magnet is
missing.
Internal interlock Green/Yellow Yellow if any of the following internal interlock fails:
Freewheel diode overtemp, Low water Flow, Door open, Diode
heatsink overtemp, Chassis overtemp, IGBT heatsink overtemp,
H.F. diode overtemp, Switch reg. DCCT failure, IGBT driver failure,
Switch regulator supply failure, IGBT driver failure, or Overcur-
rent.
Internal warning Green/Yellow Yellow if any of the following internal warning fails:
AC undervoltage, Excessive current ripple, or Ground leak cur-
rent.
The interlocks can, for example, be used on hotcell doors for controlled delivery of tracers.
The interlocks have to be configured for each tracer type. If the configuration parameter is
0, the interlocks can be defined by the customer.
19-7-3 RF system
The RF system pulls out the ions from the ion source and accelerates the particles by the so
called “Dees”. The Dees also form a resonating circuit which is tuned by motor-driven
capacitors, called “flaps”.
To make sure that the ion source is off, a maintenance relay will be set when entering this
menu, and reset when closing this window.
• State and mode section – status indicators and buttons for control of the RF system:
- Normal – starts the RF system with Dee voltage as entered in the Dee voltage
section.
- Fix – sets the RF system in TEST mode with continuous output proportional to the
Dee voltage reference (Set Dee voltage).
Note!
TEST mode – the flaps must be adjusted manually.
- Pulse – sets the RF system in TEST mode with pulsed output, with amplitude
proportional to the Dee voltage reference (Set Dee voltage).
Note!
TEST mode – the flaps must be adjusted manually.
• Dee voltage section – indicator and entry fields for the Dee voltage.
• Phase and frequency section – indicator fields for the frequency and phase.
• Configuration parameters section – indicator fields showing default values for Dee
voltage, phase and frequency start position. The default values can not be changed from
this menu.
• Buttons:
This section also includes system interlock indicators. The system interlock indicators are
green under normal conditions, and change to yellow if an interlock condition occurs. These
interlocks are only relevant for this window. If any of the system interlock requirements is not
fulfilled, the corresponding system interlock shows the causing error and the RF system can
not be turned on.
RFPG
Standby ready Green/Yellow Yellow when RFPG is off. Yellow if error occurs or interlock is active
during start-up or operation; RFPG goes to OFF.
Standby error Green/Yellow Yellow if an error or time-out (120 s) occur upon STANDBY request.
Grid power supply Green/Yellow Yellow if a grid power supply error occurs. RFPG goes to OFF.
Screen power supply Green/Yellow Yellow if a screen power supply error occurs. RFPG goes to OFF.
Anode power supply Green/Yellow Yellow if overcurrent or overvoltage error in anode PS. RFPG goes
to OFF.
Heater power supply Green/Yellow Yellow if a heater power supply error occurs.
Waterflow Green/Yellow Yellow if the internal RFPG water flow is missing or the water flow is
too low.
Anode current limit Green/Yellow Yellow if momentary limit of RFPG output power because anode
current rises above allowed level. Can occur when tuning at start-
up.
TAU air flow Green/Yellow Yellow if fans in the TAU are not OK. RFPG goes to OFF.
TAU front cover Green/Yellow Yellow if the front cover of the TAU is not closed. RFPG goes to OFF.
TPSU safety switches Green/Yellow Yellow if TPSU cover panel is loose, grounding stick is not in holder,
HV transformer temperatureis > 140°C. RFPG goes to OFF.
SCU processor Green/Yellow Yellow if a SCU processor fault occurs. RFPG goes to OFF.
SCU power supply Green/Yellow Yellow if the SCU power supply is not OK. RFPG goes to OFF.
Reflected voltage Green/Yellow Yellow if output power limiting caused by RF cavity tuning error (can
limit occur occasionally).
RF cavity
Water flow Green/Yellow Yellow if RF cavity water flow is not OK. (If any of the two separate
water circuits is not OK.)
System interlock
Vacuum ok Green/Yellow Yellow if vacuum not OK. (Not below Penning LOW LEVEL set point,
4×10-5 mbar)
Set Dee Entry field/Button Typ. 35 kV Enter Dee voltage (kV) and click Set Dee.
(NORMAL mode)
This section shows the RF Forward and Reflected power. The values are updated continuously.
Reflected power Indicator field Typ. 0–0.4 kW Shows RF power that has been reflected
from the resonator. This value can reach up
to 1 kW if the resonator is not properly tuned.
To be able to control the frequency manually, the RFPG must be in OFF, STANDBY, FIX or PULSE
mode.
Set (Phase offset) Entry field/Button Type a value and click Set. The phase
changes to the new value.
Frequency Indicator field and Typ. 35–65% Shows the frequency position.
meter
Set (Frequency) Entry field/Button Type a value and click Set. The frequency
changes to the new value.
Save to configuration Button Save the set value for frequency start posi-
tion to configuration area.
The Ion Source Power Supply (PSARC) and the gas handing system should be turned off (closed)
when closing the window.
To be sure that the RF system is off, a maintenance relay will be set when entering this window
and reset when closing the window.
• Ion source section – indicator and entry fields for current, voltage and gas flow. Buttons
for simple control of the Ion Source Power Supply, and status indicators for subsystems
and system interlocks indicators are also included.
• Buttons:
- Reset – resets ion source errors, provided the error has been corrected.
- System Off – turns off all subsystems (not Vacuum & Cooling).
Gas status Indicator field Proton/Off Shows the gas flow, read out from the ion
source mass flow controller.
Gas On/Off Indicator/Button Gas On/Gas Off Turns/shows the ion source gas flow on/off.
Set (arc current) Entry field/Button Enter the new current value and click Set.
The PSARC then changes the arc current.
Cathode op time Indicator field Shows how long time [Ah] the ion source
cathodes have been operating.
Catodes replaced Button Sets the cathode operation time to zero, after
confirmation.
The lower part of the window shows the ion source and gas handling system status. The ion
source and gas handling status is shown by green or yellow indicators. If an error or interlock
occur the corresponding indicator changes from green to yellow.
The window also includes system interlock indicators. The system interlock indicators are
green under normal condition, and change to yellow if an interlock condition occurs. If any
of the system interlock requirements is not fulfilled, the corresponding system interlock shows
the causing error and the ion source can not be turned on. These interlocks are only relevant
for this window.
Status
Water flow Green/Yellow Yellow if ion source water flow is not OK.
System interlock
Vacuum ok Green/Yellow Yellow if vacuum is not OK. Chamber pressure is not OK.
The diagnostic system measures the particle beam current at different sections of the
accelerator. At this security level, no beam may be produced, so when testing this system an
external current source, such as a multi-meter in resistance measurement mode, has to be
used.
• Beam current section – indicator fields and meters for probe current, foil 1 current, foil 2
current, and target 1 to 6 current. Includes indicator fields for the four beam collimators.
Also includes buttons for simple control of the flip-in probe, and a status indicator.
• Extraction section – indicator field, entry field and indicator meter for foil holder position.
Also includes indicator and entry fields for foil number and foil wear, and buttons for
simple control of the foils.
• Buttons:
- Reset – resets BCA power errors, provided the error has been corrected.
- System Off – turns off all subsystems (not Vacuum & Cooling).
(Probe status)
Probe Out Button Will move the flip-in probe to inactive posi-
tion and allow beam extraction.
Note!
Collimator current indicates stray-beam when irradiating realated target. A well-tuned MINItrace
cyclotron has approximately the same value on each collimator pair.
Set (Balance) Entry field 0–100% Sets balance position to entered value.
Foil holder 1 pos Indicator field and 0–100% Shows foil holder 1 position.
meter
Foil holder 2 pos Indicator field and 0–100% Shows foil holder 2 position.
meter
Set (foil holder posi- Entry field 0–100% Enter the new position [%] and click Set. The
tion) foil holder moves to the new position.
Max foil wear Indicator field mAh Shows maximum allowed foil wear.
Set (Max foil wear) Entry field mAh Sends the entered maximum foil wear, in
mAh, to the GCS. The max. foil wear is the
same for all foils. The present maximum al-
lowed foil wear value is shown.
Foil number Indicator field 1 or 2 Shows active foil for foil holder 1.
Foil 1 (foil holder 1) Indicator field 0–[operator config- Number of mAh that the foil has been used.
ured] mAh
Foil 2 (foil holder 1) Indicator field 0–[operator config- Number of mAh that the foil has been used.
ured] mAh
Foil 1 (foil holder 2) Indicator field 0–[operator config- Number of mAh that the foil has been used.
ured] mAh
Foils replaced Button Sets wear of both foils to zero (foil holder 1)
and resets the foil diary kept by the GCS, after
the user has confirmed replacement.
The Reset button resets a BCA power error, provided the cause of the error has been corrected.
The vacuum system maintains the evacuated state of the cyclotron vacuum chamber.
• Vacuum system section – indicator fields for chamber and backing pressure, status
indicators for subsystem state and interlocks, and a reset button.
• Buttons:
Chamber pressure Indicator field Typ. Shows the cyclotron vacuum chamber
7.0×10-7 mbar pressure.
Backing pressure Indicator field Typ. Shows the cyclotron vacuum backing pres-
1.0×10-2 mbar sure.
Chamber closed Indicator Green/Yellow Yellow if the vacuum chamber is not closed.
Reset vacuum error Button Resets vacuum errors, provided the errors
have been corrected.
This section shows the cooling system status. If an error (or interlock) occurs the indicator
changes from green to yellow.
WARNING!
Make sure that no personnel is present in the cyclotron room before entering this menu.
Check that the beam will be directed into a target before turning on the accelerator.
To open this window, the user needs to be logged in as user access category Advanced
Service Operator (see Table 19-1). Otherwise, Beam control in the Subsystem control task
will be prohibited.
• Vacuum section – an on/off-button and status indicator for vacuum, and indicator field
for chamber pressure.
• Magnet section – an on/off-button and status indicator for magnet, and indicator and
entry field for current.
• RF section – buttons and indicator fields for RF modes, function modes, and Dee voltage
and frequency.
• Ion source section – an on/off-button and status indicator for PSARC, indicator and entry
fields for arc current and voltage. Also includes on/off buttons and status indicator for
gas.
• GCS interlock section – indicators for external interlocks. There are also indicators for
cooling and the Vault door.
• Beam current section – indicator fields and meters for current on probe, foil 1 and 2, and
targets 1–6, and collimators. Also includes an in/out-buttons and status indicator for
probe.
• Extraction section – indicator field, indicator meter and entry field for foil holders and
balance positions, and indicator field for foil number. Also includes a button for changing
foil.
• Configuration section – indicator fields for magnet current, Dee voltages, and frequency,
and entry and indicator fields for foil holder, balance positions and a button for saving
foil positions to the GCS.
• Buttons:
- Save all to GCS – saves the set parameters (entry fields): Dee voltage, foil positions
and magnet current, to GCS. When saving the parameters the ion source must be
OFF and RF must be in STANDBY mode.
- Close – if beam is ON, user will be informed that beam will be shut off when exiting
the Beam control window.
• Indicators:
Vacuum
Chamber pressure Indicator field Typ. 7×10-7 Shows the chamber pressure.
mbar
Targets
Table 19-28: Magnet, RF system and Ion source sections – indicators and controls
Magnet
Set Entry field/Button Enter the new current value and click Set.
Ramps the current first to 150 A, waits 30 s
and then ramps current to entered value
(max. current and wait time configurable).
RF
Function mode Indicator StartMode/ See Section 19-7-3 RF system on page 619.
TuningMode/
DeeVoltOk/
DeeVoltNotOk/
Sparks/
StandbyError/
StandbyReady/
WarmUp/
ErrInNormMode/
Off
Set Dee V Entry field/Button Enter Dee voltage (kV) and click Set.
Set frequency Entry field/Button Enter the new start frequency and click Set.
Ion source
Gas On/Off Indicator/Button Gas On/Gas Off Turns/shows the ion source gas flow on/off.
Set (arc current) Entry field/Button Enter the new current value and click Set.
The PSARC then changes the arc current.
Gas status Indicator field Proton/Off Shows the gas flow, read out from the ion
source mass flow controller.
Probe Indicator field and 0–100 µA Shows the beam current on the flip-in probe.
meter
Foil 1 Indicator field and 0–100 µA Shows the beam current on Foil 1.
meter
Foil 2 Indicator field and 0–100 µA Shows the beam current on Foil 2.
meter
Target 1 Indicator field and 0–100 µA Shows the beam current on target 1.
meter
Target 2 Indicator field and 0–100 µA Shows the beam current on target 2.
meter
Target 3 Indicator field and 0–100 µA Shows the beam current on target 3.
meter
Target 4 Indicator field and 0–100 µA Shows the beam current on target 4.
meter
Target 5 Indicator field and 0–100 µA Shows the beam current on target 5.
meter
Target 6 Indicator field and 0–100 µA Shows the beam current on target 6.
meter
Coll 1U Indicator field and 0–10 µA Shows the beam current on Collimator 1U.
meter
Coll 1L Indicator field and 0–10 µA Shows the beam current on Collimator 1L.
meter
Coll 2U Indicator field and 0–10 µA Shows the beam current on Collimator 2U.
meter
Coll 2L Indicator field and 0–10 µA Shows the beam current on Collimator 2L.
meter
Set (Balance) Entry field 0–100% Sets balance position to entered value.
Foil holder 1 pos Indicator field and 0–100% Shows foil holder 1 position.
meter
Foil holder 2 pos Indicator field and 0–100% Shows foil holder 2 position.
meter
Set (foil holder posi- Entry field 0–100% Enter the new position [%] and click Set. The
tion) foil holder moves to the new position.
Foil number Indicator field 1 or 2 Shows active foil for foil holder 1.
Change foil 1/2 Button Change foil 1 to foil 2. When foil 2 is reached,
the GCS will not allow another foil change
until foils replaced has been confirmed.
Configuration
Magnet current Indicator field ≈ 134 A Shows configured value for magnet
current.
Dee voltage Indicator field ≈ 35 kV Shows configured value for Dee voltage.
Frequency Indicator field ≈ 70% Shows configured value for start frequen-
cy.
Target 1 pos. Indicator field ≈ 15% Shows configured value, target 1 foil
holder 1 position.
Target 2 pos. Indicator field ≈ 30% Shows configured value, target 2 foil
holder 1 position.
Target 3 pos. Indicator field ≈ 45% Shows configured value, target 3 foil
holder 1 position.
Target 4 pos. Indicator field ≈ 60% Shows configured value, target 4 foil
holder 1 position.
Target 5 pos. Indicator field ≈ 75% Shows configured value, target 5 foil
holder 1 position.
Target 6 pos. Indicator field ≈ 50% Shows configured value, target 6 foil
holder 2 position.
Set new Entry field For each target, entered value sets foil
holder position for later saving.
Balance pos. Indicator field ≈ 5% Shows configured value for balance po-
sition. (5% is park position.)
Set new Entry field Entered value sets foil balance position
for later saving.
If the Cooling on indicator is green, the cooling pump is running. If it is yellow, the pump is
off.
Figure 19-53: GCS and safety door interlocks and cooling section
The position of the servo within the limits, is monitored by means of a variable voltage from
a potentiometer on the servo motor unit. The control system calibrates this position information
and normalizes the actual signal from the servo motor unit so that position 0% is indicated
when the servo is at the lower limit and 100% at the higher limit.
• Foil holder 1/2 position section – indicator fields and meters for the two foil holders, and
drive current indicator fields.
• Balance position section – indicator field and meters for the balance position, and drive
current indicator field.
• Foil number indication – indicator for active foil for foil holder 1.
• Buttons:
Calibrate Button Calibrate Foil holder Calibrates the selected motor. (If no motor
1 motor has been previously calibrated, this function
Calibrate Foil holder automatically calibrates all motors.)
2 motor
Calibrate Balance
motor
Foil holder 1 position Indicator field and 0–100% Shows the Foil holder 1 position.
meter
Foil holder 2 position Indicator field and 0–100% Shows the Foil holder 2 position.
meter
Balance position Indicator field and 0–100% Shows the balance position.
meter
Foil number Indicator field 1 or 2 Shows the active foil for foil holder 1.
19-7-9 18F- Nb 25 and 18F- (3.9 mm, 6 mm and 12 mm Gen II) targets
In the 18F- target, radioactive 18F-F is produced from 18O-H2O.
Figure 19-56: 18F- target control window (3.9 mm, 6 mm and 12 mm Gen II targets)
The 18F-target control and 18F- Nb 25 target control windows contain the following:
• Target and vacuum section – indicator fields for the target type, the target position, the
vacuum chamber pressure and the target pressure.
• 18F- 3.9 mm, 6 mm and 12 mm Gen II targets only: Helium cooling section – buttons for
simple control of the helium cooling, the helium and the water cooling valves, a status
indicator and control button for the helium compressor, and an indication of the helium
pressure.
• 18F- extraction section – buttons for simple control of the valves, and buttons and
indications for the pumps.
• Buttons:
- Fill H2-16O – fills the target with 16O water according to the filling parameters.
- Fill H2-18O – fills the target with 18O water according to the filling parameters.
- Dry to Outlet 1 – He-gas flows through the target for a preset time to dry the target,
target open to outlet 1.
- Dry to Outlet 2 – He-gas flows through the target for a preset time to dry the target,
target open to outlet 2.
- Stop dry – stops drying target before the preset time is reached.
Target type and tar- Text Target 18F- or Shows selected target and position number.
get position N Target 18F- Nb 25
at target position N
Vacuum Indicator field ≈ 7.0 E-7 mbar Shows the vacuum chamber pressure.
Note!
18F- Nb 25 targets are water cooled only and has, therefore, no helium cooling section.
The water and helium cooling status indicators are either green or yellow. If an error or
interlock occur, the corresponding indicator changes from green to yellow.
Table 19-34: 18F- water and helium section – cooling controls and indicators
Water cooling
Water cooling Indicator On/Off Yellow if water flow is not on or too low.
Green/Yellow
(Water cooling valve) Indicator and button Open/Close Shows and controls status of the helium and
Green/Yellow water cooling valves. These valves are con-
trolled by the same signal and always have
the same status.
To change status on the valves, click one of
the valves. Green = Open.
(He cooling valve) Indicator and button Open/Close Shows and controls status of the helium and
Green/Yellow water cooling valves. These valves are con-
trolled by the same signal and always have
the same status.
To change status on the valves, click one of
the valves. Green = Open.
Indicator and button Open/Close Shows and controls state of the valve 1a.
1a
Green/Yellow
Indicator and button Open/Close Shows and controls state of the valve 1b.
1b
Green/Yellow
Indicator and button Open/Close Shows and controls state of the valve 2a.
2a
Green/Yellow
• Target and vacuum section – indicator fields for the target type, the target position, the
vacuum chamber pressure and the target pressure.
• Water and helium cooling section – buttons for simple control of the helium cooling, the
helium and the water cooling valves, a status indicator for the helium compressor, and
an indication of the helium pressure. See Section 19-7-9-2 18F- water and helium cooling
section on page 649.
• 13N target extraction section – buttons for simple control of the valves, and buttons and
• GCS interlock section – indicators for external subsystem states and interlocks.
• Buttons:
- Connect On/Off – Off = disconnect when target needs service. On = connects target
after service.
- Dry – He-gas flows through the target for a present time to dry the target.
- Stop dry – stops drying target before the preset time is reached.
Target type and tar- Text Target 13N at target Shows selected target and position number.
get position N position N
Vacuum Indicator field ≈ 7.0 E-7 mbar Shows the vacuum chamber pressure.
Indicator and button Open/Close Shows and controls state of the valve 1.
1
Green/Yellow
Indicator and button Open/Close Shows and controls state of the valve 2.
2
Green/Yellow
• Target and vacuum section – indicator field for the target type, the target position, the
vacuum chamber pressure and the target pressure.
• Water and helium cooling section – buttons for simple control of the helium cooling, the
helium and the water cooling valves, a status indicator for the helium compressor, and
an indication of the helium pressure. See Section 19-7-9-2 18F- water and helium cooling
section on page 649.
• 11CO production section – indicator field for the oven temperatures. It also includes
2
buttons for simple control of the valves and the oven, and status indicators for the oven.
• Buttons:
- Connect On/Off – Off = disconnect when target needs service. On = connects target
after service.
Table 19-38: 11CO2 target and vacuum section – controls and indicators
Target type and posi- Text Target 11CO2 at tar- Shows selected target and position number.
tion get position N
Vacuum Indicator field Typ. 7.0E-7 mbar Shows the chamber pressure.
(10-7)
The oven status is also shown with different colors on the column and the heating elements,
as follows:
• IDLE – all grey (this status only appears at control system start-up).
Valves
Indicator and button Open/Close Shows and controls state of valve 7. Valve
7
Green/Yellow for air-cooling of the oven.
Oven
• Target and vacuum section – indicator field for the target type, the target position, the
vacuum chamber pressure and the target pressure.
• Water and helium cooling section – buttons for simple control of the helium cooling, the
helium and the water cooling valves, a status indicator for the helium compressor, and
an indication of the helium pressure. See Section 19-7-9-2 18F- water and helium cooling
section on page 649.
• 15O production section – indicator field for the oven temperatures. It also includes buttons
for simple control of the valves and the oven, and status indicators for the oven.
• Buttons:
- Connect On/Off – Off = disconnect when target needs service. On = connects target
after service.
Table 19-40: 15O target and vacuum section – controls and indicators
Target type and posi- Text Target 15O at target Shows selected target and position number.
tion position N
Vacuum Indicator field 7.0E-9 mbar (10-9 ) Shows the chamber pressure.
In the 15O-production section the valve status is shown either green indicating that the
corresponding valve is open or yellow indicating that the valve is closed. The oven status is
also shown with different colors on the column and the heating elements, see
Section 19-7-11-2 11CO2 production section on page 658.
The oven status is also shown with different colors on the column and the heating elements,
as follows:
Valves
Oven
• Target and vacuum section – indicator field for the target type, the target position, the
vacuum chamber pressure and the target pressure.
• Water and helium cooling section – buttons for simple control of the helium cooling, the
helium and the water cooling valves, a status indicator for the helium compressor, and
an indication of the helium pressure. See Section 19-7-9-2 18F- water and helium cooling
section on page 649.
• Buttons:
- Connect On/Off – Off = disconnect when target needs service. On = connects target
after service.
Table 19-42: Dummy target and vacuum section – controls and indicators
Target
Target type and tar- Text Target Proton at tar- Shows selected target and position number.
get position N get position N
Vacuum
Vacuum Indicator field 7.0 E-7 mbar Shows the vacuum chamber pressure.
19-7-14 Exit
To exit MSS, click Exit before switching off and disconnecting the service laptop. See
Section 19-4 Main window on page 596.
Before starting any installation or service operations, read and understand all applicable RAs.
All RAs are available for GE personnel in GenSuite. Non-GE personnel can request RAs from
the local GE service representative.
Available RAs:
Note!
It is assumed that all personnel working with the equipment has participated in adequate
training to acquire the necessary knowledge of the risks involved and how to plan the work to
reduce the possible risks.
A function of the neutrons is to keep the protons together. If the number of the neutrons
versus the number of protons is well balanced, the isotope is stable. 12C is an example of a
stable nucleus. If there is excess or deficiency of neutrons, the nucleus is unstable and will
eventually release a particle or two at the same time. This is called radioactive decay. 11C
(deficiency of neutrons) and 14C (excess of neutrons) are examples of unstable isotopes. These
isotopes are called radioactive isotopes.
A nucleus that is unstable has excess energy. This energy is released through radioactive
decay and the energy is emitted as radiation. Radiation hitting human tissue might cause
significant injuries, depending on the type of the radiation, amount of energy, etc.
The atomic nuclei are also called nuclides. Radioactive nuclides are called radio nuclides
which is the correct name of a radioactive element.
The alpha particle only reaches a few centimeters in air, and is completely stopped by the
human skin. Consequently, alpha particles are harmless as long as they are kept outside the
human body. If alpha particles get into the body, they deposit their entire energy in a small
tissue volume and may cause significant injuries. Therefore, alpha radiation is the most
dangerous if the radioactive substance gets into the body.
In case of deficiency of neutrons, a beta plus (ß+) particle is emitted. This particle is positively
charged and is called positron. 11C and 18F are examples of radio nuclides emitting ß+-radiation.
A ß+-particle reacting with a ß--particle yields gamma radiation with an energy of 511 keV.
This is the PET tracer and PET scanner key process. See Table B-4.
The energy in gamma radiation is carried by electromagnetic waves as photons. The shorter
wavelength, the higher is the energy of the photons. Gamma radiation, which has the shortest
wavelength of all, therefore has the largest energy content and power of penetration.
60Coand 137Cs are examples of beta decaying radio nuclides, where the gamma radiation
usually is used.
B-1-4 Radioactivity
The intensity of the radiation from a radioactive source depends on the number of nuclei
decaying per time unit. The more decays the more radiation is emitted.
1 mCi = 37 MBq
B-2-1-1 Dose
The unit of activity, Bq, does not say anything about the risks caused by a radiation source.
To describe the risks, a quantity is needed which tells about the amount of radiation energy
absorbed by tissues, and the extent of biological harm involved. This quantity is known as
radiation dose, or merely dose in everyday language. The basic unit of radiation dose is
sievert (Sv). The most commonly used unit is mSv.
The local organization is responsible for providing adequate training so that the personnel
fully understand the used monitoring instrumentation, and that the instrumentation is
calibrated on a regular basis.
B-2-2-2 TLD
The Thermal Luminescence Dosimeter (TLD) resembles the film dosimeter and are rather
small and easy to carry. However, the TLD contains three tablets made of a radiation sensitive
material. The radiation causes a change in the energy state of the electrons in this material.
The charged energy can be released in the form of light pulses and measured in a reader
unit. The TLD is more sensitive and accurate than a film dosimeter and it is immediately usable
after reading.
The main intention with radiation safety work is to limit the dose exposure to personnel in
general. However, the national regulation specifies maximum annual dose limits. Allowable
maximum dose rates are not specified but it is recommendable for the local organization to
establish rules to guide the personnel in the daily work.
The following information is referring to European and US guide lines and may not be
applicable in all countries. Find the accurate specifications in the national regulations.
Note!
The quantity is effective dose equivalent, see Section B-2-1-4 Dose equivalent on page 671.
Note!
Low energy gamma exposure to be
considered.
Table B-3: GE Healthcare dose exposure limits (constraint levels from 2010)
Note!
Low energy gamma exposure to be
considered.
When the decision is made to commence the service action, the following check list will be
an example of good practice in the work planning:
• Define the main position of your body, hands and head during the service intervention.
• Measure the dose rate in those positions with and without low energy filter.
• Calculate the dose exposure for the different areas and take final decision about
commencement of work.
Note!
Also see the job safety analysis section. It provides a possible radiation scenario for each main
PM task. Please consult that column to be able to distinguish between a normal and an abnormal
dose rate.
B-3-2-1 Preparation
• Always wear required PPE (personal protection equipment) such as safety shoes (note
that a safety shoe steel cap can not become radioactive), shoe covers, full body lab coat,
safety glasses (with side shields) or face shield, protective gloves, and always wear
required personal radiation measuring devices (see AppendixC MINItrace radiation safety
aspects).
• Always get approval for entry from responsible radiation safety officer on site.
• Never enter an area where risks of radiation hazards exist until the risks have been
clarified.
• After isotope production, wait for at least two hours before entering the cyclotron room.
Always let the system cool down as long time as possible before start of work on the
system.
• When working with open radiation sources (e.g. liquids) use hot cells with proper
ventilation.
• Plan the work action carefully, prepare with adequate tools and arrange possible
protective shielding.
• Never machine or otherwise treat components and material coming from a cyclotron
room.
• Remove contaminated clothes and put them into plastic bags. If a dosimeter by any
chance indicates that an article of clothing is contaminated, mark and leave that article
of clothing on site (treat that article of clothing the same way as an old but active target,
that is, in an area dedicated for radioactive parts.) Check it at the next convenient
opportunity.
• If the skin is contaminated, clean the area immediately with soap and water.
• Use the digital personal dose-monitoring instrument and read the actual pick up of dose.
• Record your exposure of the day in a way established by your local GE Service
organization and/or in accordance to national/local laws and regulations.
Lead is an excellent material for stopping gamma radiation. The same holds for steel and
concrete, even though not as effective as lead. Lead is often used in small vial containers, in
Hot Cell walls and in bench-top shielding. Concrete is most commonly used in cyclotron vault
walls.
Note!
It is assumed that you are familiar with the content of the Appendix B General radiation safety
aspects.
Note!
It is assumed that all personnel working with the equipment has participated in adequate
training to acquire the necessary knowledge of the risks involved and how to plan the work to
reduce the possible risks.
• ion source
• extraction system
The targets where the nuclear reactions take place during irradiation are connected to the
vacuum chamber.
The targets are the main source of radiation during irradiation (requires massive radiation
shielding) and after terminated irradiation when the targets are heavily radioactive by induced
radioactivity. Most of the components close to the accelerated orbiting ion beam will pick up
radiation.
All radioactive subsystems are located in the cyclotron room, which is provided with all
necessary safety systems to eliminate risks.
• Hydrogen/ 2 MeV
Note!
The neutron production will only exist during cyclotron beam operation.
The plastic material in front of the targets moderates the energy of the neutrons, but it does
not stop them. The neutrons are finally captured in the vault walls (or the optional radiation
shield).
Liquids 18F 110 min The 18F isotope has a relatively long half-life and proba-
bility for high radioactivity (1–4 Ci total target content)
calls for special attention. Droplets of 18F fluoride can be
a very strong point radioactive source and is also at-
tached very hard to unprotected skin. Always carry pro-
tecting gloves and check for possible contamination
thoroughly.
13N 10.5 min The maximum radioactivity of 13N isotope in the target
will be in the range of 100–150 mCi.
Gases 11C 20.5 min The target can contain up to 1 Ci of radioactivity for both
(11C or 15O, de- isotopes.
pending on 15O 2 min Possible leaks will raise the activity level in the surround-
option) ing air and does not present radiation hazards in form
of point sources.
Note that the 15O gas from the target may be trans-
formed into 15O water vapor (by means of an oven).
Maximum level of radioactivity is 50 mCi.
The active content of the target will disappear in about 1 minute after delivery. Some residual
isotopes will always be residing in the target after the delivery. The amount of residual activity
will vary, depending on several parameters such as target condition, target filter condition,
blocked transfer lines etc.
Under extreme circumstances (clogged transfer line) all produced isotopes may still be left
in the target. Normal conditions for an 18F target will be 3–5% residual activity. Most of this
activity will leave the target on a second rinse. Therefore, the rinsing water coming from a
target may contain radioactivity and has to be handled as a normal target delivery (target
content to be transferred to a hot cell or another shielded area).
The target is contained inside the integrated radiation shielding and will not present any
hazards to the environment as long as the shield is sealed.
• In case of a single target foil rupture, radioactivity (and possibly ruptured target foils
fragments) might mix with the helium cooling gas. The helium gas (as well as target foil
fragments) will spread throughout the helium cooling system piping and helium cooling
compressor heads, situated at the bottom of the TSC.
The worst condition scenario is related to a carbon-11 production, where 1000 mCi of
activity could reside inside the target in the end of a production. A single Target Foil
rupture during these conditions could push activity in the range of 200 mCi back into
the helium cooling compressors, resulting in an estimated dose rate of about 1 mSv/h
at a distance of 1 m from the helium cooling compressors.
• In case of leaks or ruptures of the target foils, the most likely is a double target foil
rupture. During such a failure, the target content (and possibly ruptured target foils
fragments) will rush into the Vacuum Chamber and reside there until the operator decides
to take further actions. Consequently, there is a low probability that radioactivity will be
pushed into the helium cooling compressors.
• In case of an LTF (Liquid Target Filler) valve malfunction in the Target Support Cabinet,
radioactivity may spread in the LTF mechanisms (valves, pipes, syringes).
Labels indicate that the TSC and the helium cooling compressor is a potential radiation hazard.
Always wear Personal Alarm Dosimeter to get prompt notice of radiation!
Under normal conditions the maximum dose rates will only be present during delivery of the
products. Nevertheless, the safety strategy should always consider problems in the transfer
of the radioactive product such as a blockage of the line.
The worst case scenario is that the total activity will be distributed evenly in the whole system
volume, making the transport line a corresponding “radioactive line source” with a distributed
activity proportional to the transfer line volume per length unit.
C-2-3-4 Leakage
Leakage in the transfer lines will be avoided by using tubing and piping with adequate quality
and minimization or prohibition of connection and joints.
To ensure full safety for possible leaks, the transfer lines can be put in a conduit ending in a
safe environment. The conduit will pick up the radioactivity and guide it to a safe place (hot
cell or other shielded area).
Possible leaks in a gas target will raise the activity level in the surrounding air and does not
present radiation hazards in form of point sources.
Besides the radioactive substance from the target, note the following:
• Activity as aerosol (very small “droplets” of 18F fluoride, 15H2O or 13NH3) might appear
in the helium gas evacuating from product vial in the hot cell.
• The helium gas, used to push the radioactive substance from the target to the hot cell,
is not radioactive.
Wait for at least two hours before entering the cyclotron room after isotope production.
The low energy gamma radiation may reach very high levels. Therefore, it is necessary to
closely monitor radiation levels when entering the work area.
C-2-5 Waste
The MINItrace waste gas system consists of a waste container (or another secluded room of
some kind). Some sites use a delay pipe. The output of the waste gas system can be directly
connected to a ventilation system if required.
Always search for possible contamination and treat spill accordingly. If a dosimeter by any
chance indicates that an article of clothing is contaminated, mark and leave that article of
clothing on site (treat that article of clothing the same way as an old but active target, that
is, in an area dedicated for radioactive parts.) Check it at the next convenient opportunity.
Manufacturer Model
C-3-4 Digital dose monitor (dose rate, accumulated dose, energy discrimination)
A lot of manufacturers offer these instruments, see table below. For information on standard
personal radiation monitors, see also Table C-2.
Manufacturer Model
FAG FH 40F1
Note that some digital dose monitors have an energy discrimination facility to enable
separation of low energy gamma radiation (affects skin dose) from high energy gamma
radiation (affects body dose).
Note!
For all common purposes, a standard personal radiation monitor as described in Section C-3-3
Personal radiation monitor on page 685 will be sufficient.
External maintenance staff must not enter any classified area in the users premises
without approval from the radiation safety officer on site.
In order to reduce the radiation from the target foils, a small lead container, with wall thickness
40–50 mm, with removable lid is recommended.
Note!
It is strongly recommended to store the target foils separately. Contamination of the target
foils is in the order of magnitudes higher than any other possible waste from the cyclotron
system.
Hot cell
The hot cell should provide adequate radiation attenuation in order to provide safe working
conditions for staff on the outside of the hot cell. The specification of the hot cell depends on
what type and amount of isotope is going to be processed.
A hot cell usually is furnished with a transparent window (high dense glass) for inspection
purposes. Sometimes there are manipulators for interactive work inside the hot cell.
In the process to specify a hot cell at least the following areas of interest should be considered:
• Controlled ventilation
• Isotope delivery interlocking (verification system that hot cell is closed before delivery of
isotopes)
• Radioactive emission
D-1 Introduction
This chapter describes the MINItrace Service System (MSS) software version for Microsoft™
Windows™ XP which was discontinued 2014.
Note!
The next version of the service system software has a different structure and user interface,
and is described in Chapter 19 MINItrace Service System version 2.x (MSS) software.
The MSS consists of the MSS software package running on a laptop, called service laptop.
The laptop can be connected to the General Control System (GCS).
The MSS makes it possible for the service engineer to control the accelerator, the different
targets and the radiochemistry subsystems. It is intended for trouble-shooting, and for running
the accelerator in a flexible way to set and monitor different parameters. The software controls
the GCS via commands transferred over serial cables.
The user interface is a graphical package which shows a schematic picture of the accelerator,
the targets and the radiochemistry subsystems, and all related parameters.
The MSS software and communication cables are included in the cyclotron delivery. The
laptop should be provided by the customer for cyclotron systems delivered with the MSS
software version that was discontinued 2014.
The following sections describe the service laptop and its connections to the cyclotron, the
MSS software and its structure.
D-2 Connect and disconnect the service laptop to the cyclotron system
The service laptop used for the MSS should have Microsoft™ Windows™ 2000 or Windows™
XP installed. It should also have a 3.5" drive, external or internal, to allow for software
installation.
A service laptop that is equipped with USB ports for communication also requires the
USB-to-RS232 adapter supplied for connection to the control system.
For installation of the MSS software on the laptop, see Section D-4 Software installation on
page 750.
For details regarding the operation of the laptop, refer to the accompanying operator manual.
D-2-1-1 Equipment
The equipment required:
• USB-to-RS232 adapter
• GCU Gen I: Port P3 on the rear panel, or the MSS port on the CSC
• GCU Gen II: The MSS port on the front panel or on the CSC
Note!
If the service laptop is equipped with a serial port, the adapter is not required.
3 Select LOCAL mode. Check system condition on the Master System and select Local in
the Maintenance menu. If the Master System is down or not connected, LOCAL mode is
automatically active.
Note!
The system returns to the Master System within 6 minutes if connection with the MSS is
not established. See also below.
4 Click the MSS icon on the laptop desktop to start the MSS software.
Unsuccessful connection
The following happens after selecting Local at the Master System, if connection with MSS is
not established, or if connection is interrupted:
1 minute Oven heating turned off, air cooling of oven OC1A starts, which is audible.
Communication setup
The GCS is controlled by commands transferred via serial communication.
When starting the MSS software, the service laptop is automatically configured as in Table D-1
below.
Parity No
Databits 8
Stop bit 1
1 Go to the main menu and click MAIN MENU in the active menu.
4 The Master System automatically regains control within a few seconds, if connected.
To select a function, click the desired button. Entry fields are provided for input of numerical
values. To enter a value, set the cursor to the entry field, enter the number and press RETURN.
Indicator fields provide readout values for different parameters.
• Ion Source and Gas Handling System menu (click the center of the cyclotron)
• Exit
Each menu choice will replace the Main menu with the related submenu. Only one menu can
be active at a time.
Magnet Controls the magnet system, including the Magnet Power Supply.
Ion source Controls the ion source, including the Ion Source Power Supply and
ion source gas.
Extraction, Diagnos- Controls the extraction and diagnostic system, including the flip-
tics in probe and the different foil options.
Target Number 1–6 Controls the corresponding target and chemistry system.
The Magnet Power Supply (PSMC) can be on when leaving the page.
• Control and indicator window – indicator and entry fields for current and voltage. Also
includes buttons for simple control of magnet power, and a status indicator.
• Message window – shows messages, see Section D-3-1-1 Message window on page 696.
• Safety window – “LEDs” for indication of external interlocks, see Section D-3-2-3 Safety
window on page 699.
2 MAIN MENU, exit from the Magnet System menu and return to the Main menu.
Set Current Entry field Move the cursor to this field, enter a new
current value and press RETURN. Ramps the
current first to 150 A, waits 60 s and then
ramps current to entered value (max. cur-
rent and wait time configurable).
Magnet State Indicator/white ON/OFF Shows PSMC state. Signal from contactor
inside PSMC.
The magnet status window also includes a RESET button which resets magnet error, provided
the error has been corrected.
magnet water flow Green/yellow Yellow if the water flow (water manifold) to the magnet is missing.
internal interlock Green/yellow Yellow if some of the following internal interlock fails:
Freewheel diode overtemp, Low water Flow, Door open, Diode
heatsink overtemp, Chassis overtemp, IGBT heatsink overtemp, H.F.
diode overtemp, Switch reg. DCCT failure, IGBT driver failure, Switch
regulator supply failure, IGBT driver failure and Overcurrent.
internal warning Green/yellow Yellow if some of the following internal warning fails:
AC undervoltage, Excessive current ripple and Ground leak current.
To make sure that the ion source is off, a maintenance relay will be set when entering this
menu, and resets when leaving this menu.
• Status window – “LEDs” for indication of subsystems and interlocks. Also includes System
Interlock indicators.
• Message window – shows messages, see Section D-3-1-1 Message window on page 696.
• Safety window – “LEDs” for indication of external interlocks, see Section D-3-2-3 Safety
window on page 699.
• Dee voltage control and indicator window – indicator and entry fields for the Dee voltages,
and two status indicators.
• Frequency and RF power indicator window – indicator fields and meters for the frequency
and phase and indicator fields for the RF power.
• Configuration value window – indicator fields showing default values for Dee voltage,
phase and frequency start position.
1 NORMAL, starts the RF system with Dee voltage as entered in the Dee voltage window.
4 FIX, sets the RF system in TEST mode with continuous output proportional to the Dee
voltage reference (Set Dee voltage).
Note!
TEST mode – the frequency must be adjusted manually.
5 PULSE, sets the RF system in TEST mode with pulsed output, with amplitude proportional
to the Dee voltage reference (Set Dee voltage).
Note!
TEST mode – the frequency must be adjusted manually.
7 MAIN MENU, exit from the RF system menu and return to the Main menu.
This window also includes system interlock indicators. The system interlock indicators are
green under normal condition, and change to yellow if an interlock condition occurs. These
interlocks are only relevant for this menu. If any of the system interlock requirements is not
fulfilled, the corresponding system interlock will show the causing error and the RF system
cannot be turned on.
RF CAVITY
water flow Green/yellow Yellow if RF cavity water flow is not OK. (If any of the two separate
water circuits is not OK.)
RFPG STATUS
standby ready Green/yellow Yellow when RFPG off. Yellow if error occurs or interlock is active
during start-up or operation; RFPG goes to OFF.
standby error Green/yellow Yellow if time-out (120 sec) upon STANDBY request.
grid power supply Green/yellow Yellow if a grid power supply error occurs. RFPG goes to OFF.
screen power supply Green/yellow Yellow if a screen power supply error occurs. RFPG goes to OFF.
anode power supply Green/yellow Yellow if overcurrent or overvoltage error in anode PS. RFPG goes
to OFF.
heater power supply Green/yellow Yellow if a heater power supply error occurs.
water flow Green/yellow Yellow if the internal RFPG water flow is missing or the water flow is
to low.
anode current limit Green/yellow Yellow if momentary limitation of RFPG output power because anode
current exceeds the permissible level. Can occur when tuning at
start-up.
TAU air flow Green/yellow Yellow if fan in the TAU are not OK. RFPG goes to OFF.
TAU front cover Green/yellow Yellow if the front cover of the TAU is not closed. RFPG goes to OFF.
TPSU safety switches Green/yellow Yellow if TPSU cover panel loose, grounding stick not in holder, HV
transformer temperature > 140°C. RFPG goes to OFF.
SCU processor Green/yellow Yellow if a SCU processor fault occurs. RFPG goes to OFF.
SCU power supply Green/yellow Yellow if the SCU power supply is not OK. RFPG goes to OFF.
refl. voltage limit Green/yellow Yellow if output power limiting caused by RF Cavity tuning error (may
occur occasionally).
SYSTEM INTERLOCK
Green/yellow Yellow if vacuum not OK. (Not below Penning LOW LEVEL set point,
4×10-5 mbar)
Set Dee Voltage Entry field Typ. 35 kV Place cursor in field, enter Dee voltage (kV)
(Normal Mode) and press RETURN.
RF Function Mode Indicator/white start mode In normal mode, the RF system looks for
(scans) resonance.
To be able to control the frequency manually the RFPG must be in OFF, STANDBY, FIX or PULSE
mode, see previous section.
In the frequency control and RF Power Indicator window, indicators are shown with both an
indicator meter and a value or just shown with a value.
Note: 0% defined frequency = 100.5 MHz; 100% defined frequency = 101.5 MHz
Frequency Indicator field and Typ. 35–65% Shows the frequency position.
meter
Set Frequency Entry field Move the cursor to this field, enter a value
and press RETURN. The frequency changes
to the new value
Button to save fre- Save the “set value” for frequency start posi-
quency tion to config area.
Phase Indicator field and Shows the RF resonator status. Showed value
meter should be close to “set phase offset” when
the resonator is properly tuned. In pulse
mode this indicator is put out of operation.
Set Phase offset Entry field Move the cursor to this field, enter a value
and press RETURN. The phase changes to
the new value
Refl. power Indicator field Typ. 0–0.4 kW Shows RF power that has been reflected
from the resonator. This indicated value can
reach up to 1 kW if the resonator is not
properly tuned.
The Ion Source Power Supply (PSARC) and the gas handing system should be turned off (closed)
when leaving the menu.
To be sure that RF system is off, a maintenance relay will be set, when entering this menu
and reset when leaving this menu.
The Ion Source and Gas Handling System menu consists of these windows:
• Control and indicator window – indicator and entry fields for current, voltage and gas
flow. Also includes buttons for simple control of the Ion Source Power Supply and the
gas handling system. Includes status indicators for both systems.
• Status window – “LEDs” for indication of subsystem states and interlocks. Also includes
system interlocks indicators.
• Message window – shows messages, see Section D-3-1-1 Message window on page 696.
• Safety window – “LEDs” for indication of external interlocks, see Section D-3-2-3 Safety
window on page 699.
In addition, two buttons are shown in the Ion Source and Gas Handling System menu:
2 MAIN MENU, exit from the Ion Source menu and return to the Main menu. Checks also
if the gas flow and the PSARC are turned off, otherwise a message will be shown.
D-3-4-1 Ion source and gas handling control and indicator window
Figure D-15: Ion source and gas handling control and indicator window
Table D-8: Ion source and gas handling controls and indicators
PSARC Status Indicator/white ON/OFF Shows PSARC state. Feedback from contac-
tor in CAB 3.
Set Current Entry field Place cursor in field, enter new current value
and press return. The PSARC then changes
the arc current.
Cathode operation Indicator field Shows how long time [Ah] the ion source
time cathodes have been operating.
This window also includes system interlock indicators. The system interlock indicators are
green under normal condition, and change to yellow if an interlock condition occurs. If any
of the System Interlock requirements is not fulfilled, the corresponding System Interlock will
show the causing error and the ion source cannot be turned on. These Interlocks are only
relevant for this menu.
The Ion Source and Gas Handling Status window includes a RESET button which resets ion
source and gas handling errors, provided the errors are rectified.
Figure D-16: Ion source and gas handling system status window
Table D-9: Ion source and gas handling system status indicators
SYSTEM INTERLOCK
STATUS
water flow Green/yellow Yellow if ion source water flow is not OK.
The diagnostic system measures the particle beam current at different sections of the
accelerator. At this security level no beam may be produced, so when testing this system an
external current source, such as a multi-meter in resistance measurement mode, has to be
used.
• Diagnostic System window – indicator fields and meters for probe current, foil 1 current,
foil 2 current, and target 1 to 6 current. Includes indicator fields for the four beam
collimators. Also includes buttons for simple control of the flip-in probe, and a status
indicator.
• Extraction System window – indicator field, entry field and indicator meter for foil holder
position. Also includes indicator and entry fields for foil number and foil wear, and buttons
for simple control of the foils.
• Message window – shows messages, see Section D-3-1-1 Message window on page 696.
• Safety window – “LEDs” for indication of external interlocks, see Section D-3-2-3 Safety
window on page 699.
In addition, two buttons are shown in the Diagnostic and Extraction System menu:
2 MAIN MENU, exit from the Diagnostic and Extraction System menu and return to the
Main menu.
See Table D-10 for explanation of diagnostic system controls and indicators.
Probe Indicator field and 0–200 µA Normally 0 µA. Shows current 0 µA if an ex-
meter ternal source is connected to the probe.
Foil 1 Indicator field and 0–200 µA Normally 0 µA. Shows current 0 µA if an ex-
meter ternal source is connected to the foil.
Foil 2 Indicator field and 0–200 µA Normally 0 µA. Shows current 0 µA if an ex-
meter ternal source is connected to the dump
Target 1 Indicator field and 0–200 µA Normally 0 µA. Shows current 0 µA if an ex-
meter ternal source is connected to target 1.
Target 2 Indicator field and 0–200 µA Normally 0 µA. Shows current 0 µA if an ex-
meter ternal source is connected to target 2.
Target 3 Indicator field and 0–200 µA Normally 0 µA. Shows current 0 µA if an ex-
meter ternal source is connected to target 3.
Target 4 Indicator field and 0–200 µA Normally 0 µA. Shows current 0 µA if an ex-
meter ternal source is connected to target 4.
Target 5 Indicator field and 0–200 µA Normally 0 µA. Shows current 0 µA if an ex-
meter ternal source is connected to target 5.
Target 6 Indicator field and 0–200 µA Normally 0 µA. Shows current 0 µA if an ex-
meter ternal source is connected to target 6.
Probe Status Indicator/white IN/OUT Shows the position of the flip-in probe.
Note!
Collimator current indicates stray-beam when irradiating realated target. A well-tuned machine
will have approximately the same value on each collimator pair.
Foil holder 1 position Indicator field and 0–100% Shows foil holder 1 position.
meter
Foil holder 2 position Indicator field and 0–100% Shows foil holder 2 position.
meter
Set foil holder posi- Entry field Place cursor in field, enter new position [%]
tion and press return. The foil holder moves to
the new position.
Max. foil wear Set Entry field Sends the entered max. foil wear, in mAh, to
the GCS. The max. foil wear is the same for
all foils. The present max. foil wear value is
shown.
Present wear Foil 1 Indicator field 0-operator config- Number of mAh that the foil has been used.
ured mAh
Present wear Foil 2 Indicator field 0-operator config- Number of mAh that the foil has been used.
ured mAh
Foil number Indicator field 1 or 2 Shows which foil that is active in the foil
holder.
The Diagnostic and Extraction Status window includes a RESET button which resets a BCA
power error, provided the cause of the error has been corrected.
The vacuum system maintains the evacuated state of the cyclotron vacuum chamber.
• Vacuum control and indicator window – indicator fields for chamber and backing pressure,
and a status indicator.
• Cooling control and indicator window – buttons for simple control of the Cooling system
and a status indicator.
• Vacuum Status window – “LEDs” for indication of subsystem state and interlocks. Also
includes System Interlock indicators.
• Message window – shows messages, see Section D-3-1-1 Message window on page 696.
In addition, two buttons are shown in the Vacuum and Cooling System menu:
2 MAIN MENU, exit from the Vacuum and Cooling System menu and return to the Main
menu.
Chamber pressure Indicator field Typical 7.0x10-7 Shows the cyclotron vacuum chamber
mbar pressure.
Backing pressure Indicator field Typical 1.0x10-2 Shows the cyclotron vacuum backing pres-
mbar sure.
The Vacuum System Status window includes a RESET button which resets vacuum errors,
provided the errors are rectified.
This window shows the cooling water system status. The status is shown by indicators which
are either green or yellow. If an error or interlock occurs the corresponding indicator changes
from green to yellow.
STATUS
The position of the servo within the limits, is monitored by means of a variable voltage from
a potentiometer on the servo motor unit. The control system shall be capable of calibrating
this position information, that is it shall normalize the actual signal from the servo motor unit
so that position 0% is indicated when the servo is at the lower limit and 100% at the higher
limit.
• Control and indicator window – provides numerical values and indicator meters for foil
holders and Balance positions. Also includes buttons for calibration of the motors, a
status indicator and current indicators.
• Message window – shows messages, see Section D-3-1-1 Message window on page 696.
2 MAIN MENU, exit from the Motor Calibration menu and return to the Main menu.
Foil holder 1, 2 posi- Indicator field and 0–100% Shows the foil holder 1, 2 position.
tion meter
Balance position Indicator field and 0–100% Shows the balance position.
meter
Foil number Indicator field 1 or 2 Shows the active foil for foil holder 1.
The Motor Status window includes a RESET button which resets a motor error, provided the
cause of the error has been corrected.
WARNING!
Make sure that no personnel is present in the cyclotron room before entering this menu.
Check that the beam will be directed into a target before turning on the accelerator.
To enter this menu, enter the correct password and click the OK button. If the password is
incorrect, a message “WRONG PASSWORD, try again?” is written in the Message window.
The CANCEL button can be selected to exit from the Beam Control menu and return to the
Main menu.
• Vacuum Indicator window – indicator field for chamber pressure and a status indicator.
• Target Pressure window – one indicator field for the target pressures.
• Magnet control and indicator window – indicator and entry field for current, and a status
indicator.
• RF system control and indicator window – indicator and entry fields for Dee voltages,
Frequency and status indicators.
• Beam control and indicator window – indications and entry fields for probe current, foil
1 current, foil 2 current, target 1 to 6 current and collimator currents.
• Extraction control and indicator window – indicator field, indicator meter and entry field
for foil holder and balance positions, and indicator fields for foil number.
• Ion Source control and indicator window – indicator and entry fields for current and
voltage. Also includes a status indicator.
• Configuration Parameters window – indicator fields for magnet current and Dee voltages,
and entry and indicator fields for foil holder and balance positions.
• Message window – shows messages, see Section D-3-1-1 Message window on page 696.
• Safety window contains “LEDs” for indication of external interlocks, see Section D-3-2-3
Safety window on page 699.
13 Save new target pos, saves the foil positions for the different target positions to the GCS.
14 Save all to GCS, saves the set parameters (entry fields): Dee voltage, magnet current,
frequency and foil positions, to GCS. When saving the parameters the ion source must
be OFF and RF must be in STANDBY mode.
16 MAIN MENU:
a If the beam is ON, the user must verify that the beam can remain ON before returning
to the Main menu.
b If the beam is OFF, sets the RF system in STANDBY mode, shuts down the ion source
and returns to the Main menu.
If the set parameters have not been stored explicitly, the user will be prompted to
store parameters before exiting this page.
VACUUM
Chamber pressure Indicator field Typ 7x10-7 mbar Shows the chamber pressure.
COOLING
TARGET
D-3-8-2 Magnet, RF system and ion source indicator and control windows
Figure D-31: Magnet, RF system and ion source indicator and control windows
Table D-19: Magnet, RF system and ion source indicator and control windows
MAGNET
RF SYSTEM
Set Dee V Entry field Place cursor in field, enter Dee voltage (kV)
and press RETURN.
Set Freq Entry field Place cursor in field, enter start frequency
and press RETURN.
ION SOURCE
Set Entry field Place cursor in field, enter new current value
and press return. The PSARC then changes
the arc current.
Probe current Indicator field and 0–100 µA Shows the beam current on the flip-in probe.
meter
Foil 1 current Indicator field and 0–100 µA Shows the beam current on foil 1.
meter
Foil 2 current Indicator field and 0–100 µA Shows the beam current on foil 2.
meter
Target 1 current Indicator field and 0–100 µA Shows the beam current on target 1.
meter
Target 2 current Indicator field and 0–100 µA Shows the beam current on target 2.
meter
Target 3 current Indicator field and 0–100 µA Shows the beam current on target 3.
meter
Target 4 current Indicator field and 0–100 µA Shows the beam current on target 4.
meter
Target 5 current Indicator field and 0–100 µA Shows the beam current on target 5.
meter
Target 6 current Indicator field and 0–100 µA Shows the beam current on target 6.
meter
Coll 1U Indicator field and 0–10 µA Shows the beam current on collimator 1U.
meter
Coll 1L Indicator field and 0–10 µA Shows the beam current on collimator 1L.
meter
Coll 2U Indicator field and 0–10 µA Shows the beam current on collimator 2U.
meter
Coll 2L Indicator field and 0–10 µA Shows the beam current on collimator 2L.
meter
Set (Balance) Entry field 0–100% Sets balance position to entered value
Foil holder 1 position Indicator field and 0–100% Shows foil holder 1 position.
meter
Foil holder 2 position Indicator field and 0–100% Shows foil holder 2 position.
meter
Set (foil holder pos.) Entry field 0–100% Place cursor in field, enter new position [%]
and press return. The foil holder moves to
the new position.
Foil number Indicator field 1or 2 Shows active foil for foil holder 1.
CONFIGURATION PARAMETERS
Magnet Current Indicator field ≈ 134 A Shows default value for proton magnet cur-
rent.
Dee Voltage Indicator field ≈ 35 kV Shows default value for Dee voltage.
Frequency Indicator field ≈ 70% Shows default value for start frequency.
Tar.1, pos. Indicator field ≈ 15% Shows default value, target 1 foil holder 1
position.
Tar.2, pos. Indicator field ≈ 30% Shows default value, target 2 foil holder 1
position.
Tar.3, pos. Indicator field ≈ 45% Shows default value, target 3 foil holder 1
position.
Tar.4, pos. Indicator field ≈ 60% Shows default value, target 4 foil holder 1
position.
Tar.5, pos. Indicator field ≈ 75% Shows default value, target 5 foil holder 1
position.
Tar.6, pos. Indicator field ≈ 50% Shows default value, target 6 foil holder 2
position.
Set new Entry field For each target, entered value will set foil
holder position for later saving.
Balance, pos. Indicator field ≈ 5% Shows default value for balance position.
(5% is park position.)
Set new Entry field Entered value will set foil balance position
for later saving.
• Target and Vacuum section – indicator fields for the target type, the target position, the
vacuum chamber pressure and the target pressure.
• Water and Helium cooling section – buttons for simple control of the helium cooling, the
helium and the water cooling valves, a status indicator for the helium compressor, a
bargraph for the helium pressure and a “LED” for indication of a flow error. Also includes
a Water cooling Status window which shows a “LED” for indication of an water flow error.
• Message window – shows messages, see Section D-3-1-1 Message window on page 696.
• Safety window – “LEDs” for indication of external interlocks, see Section D-3-2-3 Safety
window on page 699.
5 FILL TARGET WITH H218O – fills the target with enriched 18O-water.
9 DRY TARGET TO OUTLET1 – He-gas flows through the target for a present time to dry
the target, target open to outlet 1.
10 DRY TARGET TO OUTLET2 – He-gas flows through the target for a present time to dry
the target, target open to outlet 2.
11 STOP DRYING – stops drying target before the preset time is reached.
12 MAIN MENU – exits from the 18F--target menu and return to the Main menu.
TARGET
Target position Indicator field pos: 1–6 Shows the slot number in which this target
is installed.
VACUUM
Vacuum Indicator field ≈ 7.0 E-7 mbar Shows the vacuum chamber pressure.
At the top of this section the water cooling status indicators is shown. This indicator is either
green or yellow. If an error or interlock occur, the corresponding indicator changes from green
to yellow.
Water cooling
The remaining part of this section shows Water and Helium cooling controls and indicators.
He cooling
He Cooling valve Indicator and button OPEN/CLOSE Shows and controls status of the helium and
Green/yellow water cooling valves. These valves are con-
trolled by the same signal and always have
the same status.
To change status on the valves, use the
Water Cooling valve mouse and click one of the valves. Green =
Open.
• Target and Vacuum section – indicator fields for the target type, the target position, the
vacuum chamber pressure and the target pressure. See Section D-3-9-1 Target and
vacuum section on page 732.
• Water and Helium cooling section – buttons for simple control of the helium cooling, the
Helium and the water cooling valves, a status indicator for the helium compressor and
an indicator meter for the helium pressure. Also includes a Water Cooling Status window
which shows a “LED” for indication of an water flow error, see Section D-3-9-2 Water
and helium cooling section on page 733.
• Message window – shows messages, see Section D-3-1-1 Message window on page 696.
• 13N extraction section – buttons for simple control of the valves. Safety window contains
“LEDs” for indication of external interlocks, see SectionD-3-2-3 Safety window on page699.
7 DRY TARGET – He-gas flows through the target for a present time to dry the target.
8 STOP DRYING – stops drying target before the preset time is reached.
9 MAIN MENU – exits from the 13N-target menu and return to the Main menu.
• Target and Vacuum section – indicator fields for the target type, the target position and
the vacuum chamber pressure. See Section D-3-9-1 Target and vacuum section on
page 732.
• Water and Helium cooling section – buttons for simple control of the helium cooling, the
helium and the water cooling valves, a status indicator for the helium compressor, an
indicator meter for the helium pressure and a “LED” for indication of a flow error. Also
includes a Water Cooling Status window which shows a “LED” for indication of an water
flow error, see Section D-3-9-2 Water and helium cooling section on page 733.
• Message window – shows messages, see Section D-3-1-1 Message window on page 696.
• Safety window – “LEDs” for indication of external interlocks, see Section D-3-2-3 Safety
window on page 699.
5 MAIN MENU, exit from the Proton-target menu and return to the Main menu.
A Password window is opened when Config parameters is selected. To gain access to the
configuration pages you must enter the correct password and click the OK button. If the
wrong password is entered, the message “WRONG PASSWORD, try again?” is shown in the
Message window.
• Magnet System, RF System, Extraction System, Ion Source System, Global and Foil diary
Configuration windows – contain indication/entry fields for the accelerator system
parameters.
• Message window – shows messages. These messages can be acknowledged with the
ACK, YES or NO buttons when shown, otherwise they disappear within about 10 seconds.
See Section D-3-1-1 Message window on page 696.
In addition, six buttons are shown in the Config Parameters, Accelerator menu:
6 MAIN MENU, exit from the Config Parameters, Accelerator menu and return to the Main
menu.
Magnet system
increasing current ≈ 40 mA
RF system
scan start pos (frequency) Scanning start position for frequency 70%
phase set 2
is operation time Time that this ion source has been used [Ah]
is life length Maximum length of time that this ion source can be used 1000 Ah
max arc current Maximum current of the Ion Source Power Supply 2000 A
Extraction system
i probe start up Maximum current for the flip-in probe at ion beam start 15 µA
up
probe/foil loss max Maximum Ion Beam loss between flip-in probe and active 80%
foil
foil/target loss max Maximum Ion Beam loss between active foil and active 20%
target
Global
Foil diary
• Target position, Choice of new target type and Current type windows – contain
indication/entry fields for the target system parameters.
• Message window – shows messages. These messages can be acknowledged with the
ACK, YES or NO buttons when shown, otherwise they disappear within about 10 seconds.
See Section D-3-1-1 Message window on page 696.
In addition, three buttons are shown in the Config Parameters, Target menu:
3 MAIN MENU, exit from the Config Parameters, Target menu and return to the Main menu.
Target position
15O
13N
18F-
dummy
none
Choise of tracers
Target position, Target type, Target address and parameters GCS windows – contain
indication/entry fields for the 13N Config parameters.
Message window – shows messages. These messages can be acknowledged with the ACK,
YES or NO buttons when shown, otherwise they disappear within about 10 seconds. See
Section D-3-1-1 Message window on page 696.
In addition, six buttons are shown in the Config Parameters, 13N menu:
6 MAIN MENU, exit from the Config Parameters, 13N menu and return to the Main menu.
ltf id 1
Parameters GCS
max leak pressure ± tolerance on change of target pressure at leak check 20 psi
• Target position, Target type, Target address and parameters GCS windows – contain
indication/entry fields for the 18F- Config parameters.
• Message window – shows messages. These messages can be acknowledged with the
ACK, YES or NO buttons when shown, otherwise they disappear within about 10 seconds.
See Section D-3-1-1 Message window on page 696.
In addition, six buttons are shown in the Config Parameters, 18F- menu:
6 MAIN MENU, exit from the Config Parameters, 18F- menu and return to the Main menu.
LTF id 2 or 3 2 or 3
Parameters GCS
min pressure1 Minimum pressure allowed for the high 250 psi 250 psi
pressure
delta pressure Min. press. incr. allowed as target gets filled 10 psi 8 psi
max leak pressure Pressure must not drop below this level dur- 95% 95%
ing leak check
1 Note: This value is used to detect a possible leak in the target during production. There is a restriction mounted
in the high pressure supplyline. This will cause the pressure to drop even if the leak is very small.
2 Note: This value is typically set 10 µA higher than the maximum allowed beam on target. This is due to the fact
that the beam will otherwise be turned off in case the target current would temporarily be at a higher value (slight
overshoot in the regulation or similar).
For example the 18F- Gen II target is maximally allowed for 50 µA, still the value is configured to 60µA to prevent
a temporary current level above 50 µA to shut the system down.
To enter the Beam Control menu, password 1, and Configuration Parameters menu,
password 2, the user must enter the correct password.
D-3-14 Exit
To return to the Master System from the laptop PC, click Exit before switching off and
disconnecting the laptop PC – see Chapter 1 Introduction.
To install the MSS software, follow the instructions in Setup Instructions MSS included in the
software package.
The files located in the EYES directory are installed from the EyeOnLine license floppy disk.
For information about the EyeOnLine program see Section D-5 MSS software structure on
page 752.
C:\MINITRAC\EYES files:
• EYEONLIN.EXE
• EOL_S.EXE
• MET_SHEL.EXE
• EYECOM.EXE
• SYSTEM08.FNT
• SYSTEM16.FNT
• INSTALL2.BAT
• RECALL2.BAT
C:\MINITRAC\MSS files:
• MINI0***.SLB (*** = 000, 001, 004, 005, 006, 007, 020, 021, 022, 110, 111 and 112)
• MINI.INI
• MINI.PAR
• PASSWORD.DAT
• GEM_CONF.DAT
• MAIN_MSS.EXE
• MSS.BAT
• NONE.BAT
The file ERROR.LOG will be created when the MSS software are running.
D-5-1 Introduction
The MSS software runs as a separate program on the service laptop connected to the GCS.
The MSS software is written in C, using EyesCream™ and Asynch Manager as tools. EyeOnLine
is used for communication between the application program and EyesCream. Asynch Manager
is used for serial communication.
First level Gives access to the accelerator, the different targets and the radio-
chemistry system. Each subsystem may be controlled independently
of the others.
Subsystem parameters may be read and set, with the normal inter-
lock system always active. At this level no beam can be produced.
This level contains a menu for each accelerator subsystem and for
each target, including connected chemistry processes.
Third level Accessible only when entering a correct password. The user can
change the passwords.
Gives access to the configuration parameters defined for later use
at production, using the Master system. The installation positions
of the six different Targets, all connected hardware addresses and
the host address for the Master, is also defined at this level.
All parameters must be stored before the GCS can work properly.
The parameters may also be stored in a file.
All set parameters are input fields. When clicking Save to GCS, the new parameter values are
transferred to the GCS.
All commands are entered by clicking with the mouse. Messages from GCS is queued.
File Description
File Description
NONE.BAT Allows the user to look at different menus without being con-
nected to the system.
ERROR.LOG At runtime, all messages shown are logged in this file, together
with time for program start and program exit.
All messages are added, which after a time of using the system
might create a too large file. In this case, it may be deleted or
renamed. If the file does not exist at program start it is created.
File Description
File Description
File Description
File Description
File Description
The MSS software can be run in two different modes. These modes are chosen by clicking
the desired icon on the desktop as decribed below.
• NONE – only to look at different menus without being connected to the GCS.
www.gehealthcare.com
imagination at work