SEFELEC SXS50-56

Markets Served
Electrical safety tester
Subject title Markets served subhead

The electrical safety testers in the SXS series carry out simply and comprehensively the
electrical safety tests required by the principle standards concerned by the Low-Voltage
Directive. The XS is a combination of a dielectric strength tester, a megohmmeter and
an earth continuity controller.

Associated with an FXS rack, it carries out leakage current measurements under
nominal voltage.

Standards EN 61010-1, EN 60335-1, EN 60950, EN 60598-1, EN 60601-1, EN 60204-1

TECHNICAL Voltage reading • “IMAX” detector by Memorisation


CHARACTERISTICS • On digital kilovoltmeter maximum current adjustable • Memorisation of 10 test
directly connected to the from 0.01 mA to 10 mA by parameter configurations
• Dielectric strength tester output terminals steps of 0.01 mA (SXS50-56) (voltage, thresholds, time…)
from 0 to 5 kV AC and from • Specification: • Combination of DELTATEST
0 to 6 kV DC (SXS56 model) +/- (1.5 % + 20 V) and IMAX modes MEGOHMMETER
• Insulation resistance • Display: 600 digits FUNCTION
from 50 kΩ to 200 GΩ (2 TΩ IMIN threshold function
option) Voltage stability • Detects whether the test Measurement range
• Earth continuity from • Better than 1 % for the mains probe is connected correctly • 50 kΩ to 200 GΩ (2 TΩ
1 mΩ to 1 500 mΩ from supply variations of +/- 10 % to the sample option XS-20)
5 to 32 A (SXS50-56) • Adjustable from 0.01 mA to • The insulation resistance is
• Leakage current and 10 mA (SXS50) (UTEST/UMAX) x 200 GΩ
power measurement (with Current
FXS rack) • Short circuit < 13 mA AC DC Voltage Option Specification
• 8 test steps in sequence (SXS50/56) and < 9 mA DC • Positive terminal to ground • +/- (1.5 % + 1 U)
mode (SXS56) for the max. voltage • Residual ripple < 1 % for • Display: 2000 digits
• 10 parameter memories adjustment I < 100 µA (SXS56)
per function • Nominal 10 mA AC (SXS50- Thresholds
• RS232 interface in series 56) and 4 mA DC (SXS56) Fault detection signalling • A High threshold (allowing
• By visual indication (LCD the presence of the
As option : ETHERNET, PLC Current reading screen and LED lights) and element to be measured
or IEEE488-2 interfaces • On shunt inserted into the test sound to be detected) and a Low
circuit • Memorisation of the voltage threshold adjustable from
DIELECTRIC STRENGTH • from 0 to 999 µA and default current on the 50 kΩ to 200 GΩ (2 TΩ option
TEST FUNCTION •from 1.00 to 9.99 mA LCD display unit XS20)
• Specification: • Nominal HV source
Output voltage +/- (2.5 % + 2 U) 1 U = interruption when the Measurement voltage
• 0 to 5 kV AC (50 or 60 Hz). 1µA or 10µA sinusoid goes to zero Configurable in steps of 1V
Limited to 4.2 kV AC with • Resolution: 1000 digits from:
the FXS501 Voltage application time • 10 V to 500 V DC or
• 0 to 6 kV DC (SXS56) Fault detection • Adjustment of the rise, • 20 V to 1000 V DC (option
• Specification • “DELTATEST” detector holding and fall times from XS-26)
+/- (2 % + 50 V) of the The ΔI detection allows 0 to 999 seconds • Specification: +/- (1 % + 2 V)
programmed value between complete insensitivity to the • DC current: 2 mA
100 and 5000 V and for a current due to the resistance
current < 100 µA (SXS50) with and the capacitance of the Measurement time
breakdown detection modes: sample tested. • Adjustable from 0 to 999
ΔI, IMAX or ΔI+IMAX seconds or continuous

Memorisation
• Memorisation of 10 test
parameter configurations
(voltage, thresholds,
SXS50/56 2014 (REV.A)

time…)
SEFELEC SXS50-56
Electrical safety tester

EARTH CONTINUITY Example of a sequence Operator Storage temperature


FUNCTION carrying out a dielectric test • By safety loop suppressing • -10° C to + 60° C
with parameter memory n°1, the generation of High
Measurement range followed by an insulation Voltage or current Overvoltage category
• 0.001 Ω to 1.500 Ω test with parameter memory • Red green light • CATII
• Possibility of voltage n°0, then a Pause then an
dropping display (Norm earth continuity test with Equipment under test Degree of pollution
EN60204 ) parameter memory n° 5 • Immediate fault detection •2
repeated on 10 different and nominal HV transformer
Specification points (Multiple Continuity): voltage interruption Category
• +/-(2.5 % + 10 mΩ) • Short circuit of the output • Category 1 (earth leakage
• Display: 1500 digits L1: RIGID 1 terminals and discharge protection)
L2: MEGOHM 0 of the capacitances tested
Thresholds L3: PAUSE (voltage option DC, t < 1 sec.
• A High threshold and a L4: POINTS 10 per μF)
low threshold adjustable L6: CONTINUITY 5
from 0.001 Ω to 1.500 Ω
• Possibility of adjusting the The WINPASS-MX software on a GENERAL
threshold in volts (Norm PC allows an infinite number of CHARACTERISTICS
EN60204) test steps to be carried out.
Presentation
AC measurement current INTERFACE • Table device
• 5 to 32 A AC by • Device supplied with the • Metallic housing
graduations of 0.5 A RS232 interface
• Specification: Dimensions
+/- (1 % + 0.5 A) CONTROL SOFTWARE • Height 131 mm
• Possibility of applying the • Width 440 mm
current gradually from 5 Application software • Depth 450 mm
A to the maximum value • WINPASS-MX: powerful
selected for the test. software for controlling the Weight
• Voltage in open circuit: devices in the SXS series • 27 kg
< 6 or < 12 V AC sinusoidal • Specific on-demand
• Frequency: mains supply software Mains supply
(50 or 60 Hz) • 230 V or 115 V +/- 15 %
PROTECTIONS Instrument single phase, 47 - 63 Hz
CURRENT LEAKAGE By time-delay fuse • Consumption: 70 to 600 VA
MEASUREMENT according to the test
See the FXS501 technical
sheet Operating temperature
• 0° C to + 45° C
SEQUENCE FUNCTION
8 test lines sequenced
automatically among:
Dielectric, Insulation,
Continuity, Leakage, Pause,
Multiple continuity.
Each test function is
associated to a parameter
memory number.
Storage of 10 sequence
configurations.
SXS50/56 2014 (REV.A)
SEFELEC SXS50-56
Electrical safety tester

OPTIONS ACCESSORIES TE81-XS C0193


2-wire continuity test probe Housing equipped with 6
XS-02 TE54 with remote control and international mains sockets
PLC interface, available Test probe (dielectric + PASS/FAIL lights (for multiple
contacts: insulation) continuity tests)
• START
• PASS and FAIL
• ENDOFTEST
• FAULT

XS-03 CO160-XS
Input output 0-10 volts: CO183 Red/green lamp
• Input 0-10 volts to control 2-wire cord with PI001-00-L02
high voltage. crocodile clips HV pistol with remote control
• Output 0-10 volts to read - length 2M
the current or the voltage.

XS-06 PI001-00-LXX
IEEE488-2 interface (Talker - HV pistol with remote control
Listener) - max length 10M KRXS
19" rack kit
XS-08 CO184
PLC Option 2-wire banana cord PI001-05-L02
Input 0-10 volts probe with retractable 2 HV pistols 50 VA + return
tips remote control - length 2M
XS-100
Ethernet Interface

XS-10
AC-DC (6kV DC - 10 mA
short circuit) SXS50

XS-20 C0177 C0200


Insulation resistance Fixed HV cord Housing equipped with
measurement up to 2 TΩ mains socket

XS-22
Display of resistance in
MΩ x km

XS-26 CO175
Insulation resistance Standard XS return cord
measurement under 20 to
1000 V DC by steps of 1 V

SOFTWARE
WINPASS-MX software for
electro-domestic application
SXS50/56 2014 (REV.A)
SEFELEC SXS50-56
Electrical safety tester

SAFETY CAGES CALIBRATION BOXES XS91-4


Calibration kit for 30A
CA001-00-00+CO007-00-L05 XS91-1 continuity
Safety cage for SXS50/56 Calibration kit for 50VA
dielectric up to 10kV DC

SD51
Fault simulator for XS 50 VA

CA002-00-00+CO007-00-L05
Safety cage for SXS50/56

XS91-3
Calibration kit for insulation

REM
Resistance standard with
high value (from 1 MΩ to
1 TΩ)

Eaton
Sefelec sas
19 rue des Campanules
F-77185 Lognes
France

Head Office +33 (0)1 64 11 83 40


Fax +33 (0)1 60 17 35 01
South-east contact +33 (0)4 74 60 79 78
South-west contact +33 (0)5 56 80 53 22

After-sales service +33 (0)1 64 11 83 48


SXS50/56 2014 (REV.A)

Fax +33 (0)1 60 17 85 64

© 2014 Eaton - Sefelec sas Eaton is a registered trademark.


All Rights Reserved
Printed in France All other registered trademarks are commercial
trademarks of their respective owners.