You are on page 1of 1

Olympus NDT Canada

505, boul. du Parc-Technologique Tel. : (418) 872-1155


Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com

EDDY CURRENT: ECT EXTENDED ASME CALIBRATION TUBE

RECOMMENDED CALIBRATION TUBES


The CT02 ECT calibration tube below provides every defect that is required to perform a good ECT calibration and
build strong phase-to-depth curves. In fact, this tube is not only conforming to the ASME standards. In addition, it
also provides wall thinning defects in order to help users to perform a more accurate calibration.

DIFF (pits)
SP01-XXX

GR10
(internal) GR20 H100a H80 H60 H40 H20 H100b GR40 GR60

ABS (wall thinning)

GR10 GR20 H100a H80 H60 H40 H20 H100b GR40 GR60

ARTIFICIAL DEFECT PARAMETERS IN FUNCTION OF TUBE WT AND OD


Defect Depth (%WT) Diameter Defect Depth (%WT) Diameter

GR10 10 % (internal) - H80 80 % 0.078”

GR20 20 % - H60 60% 0.125”

GR40 40 % - H40 40% 0.188”

GR60 60 % - H20 20% 0.188” (4 holes)

If OD>0.75”, 0.067” If OD>0.75”, 0.125”


H100a Thru H100b Thru
Else: 0.052” Else: 0.078”

VARIABLE CHART IN FUNCTION OF TUBE OD


Value Value Value
Variable Variable Variable
in mm in mm in mm

A 1.25” 32mm D 3.00” 76mm TL 20.50” 521mm

B 1.50” 38mm GLa 1/16” 1.6mm SP_OD OD + 3/4" OD + 19mm

C 3.50” 89mm GLb 1/2" 13mm SP_TCK 3/8” 10mm