‘SURVEYING DIVISION
AK PT. ASABA
USER NAME:
PT. Hutama Karya
Reference No.
Date issued
Certificate of Calibration
TOTAL STATION
MKS1909215
18 September 2019
Instrument name: TOPCON/ GTS-102N
Setial No. 2P4641
Inspection date: 16 September 2019
Inspectorname : Albert Paken Lumalan
This certifies that the above named instrument has been calibrated by PT. ASABA - Authorized
SOKKIA Service Center with standard traceable to Japanese surveying manufacture Association-
ISIMA-103-2002, according to SOKKIA Service Manual.
(1) Theodolite
No. item Result [Max jolerance
j VJ 0,01 mm [Tess than 0.24 mm
1 | Telescope | Max. deviation of telescope axis FEO oT es 4
Perpendicularity of telescope axis and i 5
2 | Mechanism | horizontal axis (Collimation) eid | cee eee
Horizontality of horizontal axis (Inclination) 2see. | Tess than 10 see
3__| WEncoder | H Eccentricity see. [Tess than 12 see
Vertical eccentricity Sisee. | less than 15 sec.
4 | Vertical cirele_ Vertical constant difference 2 sec. | less than 20 sec.
Vertical zero point calculation 2-sec. | less than 20 sec.
| Automatic | Compensator error (+3 min) 2sec. | less than 10 sec
compensator | Compensator eror 3 min) 2sec. | less than 10 see.
@) Distance Meter
(2-1) Laser power
No. | Test item Result Tolerance Standard
1_] Laser power 45uW, + 14W class 1 1EC60825-1
(2-2) Accuracy _ Target: APOL
No. | Nominal distance | Measured distance Result Tolerance (@(22ppmxD)mm
1 10,692 m. 10,690 m 2mm| Tess than = 2.022 mm
2 57334 57332 m. 2mm Tess than = 2,115 mm.
2 169.103 m 169,100 m 3mm Jess than + 2.338 mm
3 754,460 m 754,456 mi, 4mm Tess than = 3,508 mm
“The inspection of EDM complies with Japanese Standard JSIMA-102 :2006
Nate
PI-ASABA sot reponse nsunen oo ene
Caiman intra s6(oa) nowt
Résommandod ot cation
op
Approved By:
RICKY ONG
TECHNICAL MANAGER
PLASABA
JL, Ir, H. Juanda No.7, Jakarta - Indonesia | Telp. +6221 - 352 2977, 380 0404 Fax : +6221 - 3522972
SAE Le SUT)J\ PT. ASABA
Gem SURVEYING DIVISION
a
i
USER NAME:
PL: Hutama Karya Reference No: MKS1909214
Date issued: 18 September 2019
Certificate of Calibration
AUTOMATIC WATERPASS
Instrument name : NIKON/AZ 28
Serial No. 765591
Inspection date 16 September 2019
Inspector name Albert Paken Lumalan,
This certifies that the above named instrument has been calibrated by PT. ASABA - Authorized
SOKKIA Service Center with standard traceable to Japanese surveying manufacture Association-
JSIMA-103-2002, according to SOKKIA Service Manual.
Wea
No. Item Result Max. tolerance
A: Line of sight horizontality 2sec less than 10 second
2 Sighting axis deviation 0.2mm less than 0.5 mm
3 ‘Compensator range 17 min More than 16 min
4 Compensator accuracy 2sec Less than 10 sec
=
=
tee s T.
ea bese aacenes a Approved By
eee ae
TECIINICAL MANAGER
Pr aSaBa
JL. Ir. H, Juanda No.7, Jakarta - Indonesia | Telp. +6221
REE RU OReC}
382 2977, 380 0404 Fax : +6221 - 352 2972