NPC Metrology Laboratory
CERTIFICATE OF CALIBRATION
CERTIFICATE NO.: 2019-051-003
for
Insulation Resistance Tester
Metrel Model MI 3210
Serial No.: 17010537
Submitted by:
EURO-ASIA ENERGY SOURCE INCORPORATED
2nd Fir. No. 265 Unit N Cortes Bldg. Cor.
Sct. Fernandez & Tomas Morato Sts.,
Brgy. Laging Handa, Quezon City
This certifies that the instrument was calibrated by comparison against the standards
maintained by the NPC-Central Metrology Laboratory using procedure PES-ESD-PM-012-
WI-006.
The standards used are traceable to the National Metrology Laboratory of the Department of
Science and Technology and to the National Institute of Standards and Technology through
Ark One Solutions, Inc.
The test data in the attached Calibration Report were those obtained during the time of
calibration.
Itis the responsibility of the user to determine the suitability of the instrument for its intended
use.
f
HOMER S. GALANG
Mandger
Testing and Metrology Services Division
Date: 08 July 2019
Page 1 of 3
Quezon Avenue comer BIR Road, Diliman, 1100 Quezon City, Philippines
Tel. Nos. (632) 921-3541 079 Fax No. (632) 921-2468 P.O, Box 10183NPC Metrology Laboratory
CALIBRATION REPORT
2019-051-00;
Page 2 of 3
Reference: WP-ESD-TMD-019-051 Date: July 2, 2019
Description; Tnsulation Resistance Tester Make; “Metrel MI 3310
Serial No: 1701053; TagNo: N/A
1. Resistance Measurement Test Results
Test STD, MEAS, NOMINAL AS FOUND (UUT) ASLEFF WUT)
VOLTAGE INPUT ——UNCERT. TOLERANCE READING DEV. «READING DEW.
wo ona) 10) ta) ota) oi) iy) _ te)
250 1 = 0010 + 008 1400 0.00. +1400 ~~~-0.00
100 + 028 + 8.00 100 ° 100 °
200 4 0570 + 13.00 200 ° 200 °
500 5 £ 004 028 499-001 4.99 0.01
100 + 0285 = 8.00 100 ° 100 °
500 + a5 27.95 499 a 499 a
1000 eo eee 500 000-500-000
100 «0285 + 8.00 100 ° 100 °
500 + 1425 +2795 499 499 a
2500 5 + 00m + 028 5.00 000 5.00 0.00
100 + 028 + 8.00 100 ° 100 °
500 1425 £795 499 a 499 a
5000 so 4 04 280 49.9 “oa 49.9 “oa
100+ 028 + 8.00 100 ° 100 °
500 1425 + 2790 498 “a 498 2
(cay ay (cay _Gy_ _ a)
Te Sopra cil Toca ge Mieco Coac1l MrGoscEd (race
TL, Output Voltage Test Results
UUT MEASUREMENT _ NOMINAL AS FOUND (STD) AS LEFT (STD)
SETTING UNCERTAINTY TOLERANCE READING »=—=DEY. «= READING OY.
wo wo wo w wo w o
266 = 00203 = 1630 ~«—«a65.62—=~—~C«O BSCR gS
317 0.0305 + 3885 —515.63 137 515.63 137
aw) ow aw) aw) aw) aw) a
104 0.0040 = 0.08 1.05 “0.01 1405 “oor
3.05 = oor + oa8 3.05 0.00 3.05 0.00
5.09 + 0.0204 0.28 5.09 0.00 5.09 0.00
102 = 00409, + 054 1039 0.01 10.19 001
Reviewed by, Approved,
Castro Homer $.Galang _
‘Technician Prigeigi Engineer C MSD ManagerNPC Metrology Laboratory
CALIBRATION REPORT
2019-051-003
Page _3 of _3
Reference: WP-ESD-TMD-019-051 Date: July 2, 2019
Description: Insulation Resistance Tester Make: “Metrel MI 3210
Serial No.: 17010537 TagNo: N/A
Test Standards
Description Tag No. Calibration Due Date Calibration Traceability
Meagim Desade RSI, appx s/s/2020 TEDI/NML.
Merci Deeaie ee 2L-MDB-o2 1/23/2020 ITDI/NML
~~ 2L-MDB-03 3/24/2020 TIDI/NML
Meonum Decade Resor, oD 1/24/2020 Mrs
geen emersormga 2L-MDB-05 4/24/2020 TTDI/NML,
— 2L-HRS-o1 12/12/2019 ITDI/NML,
Precision Multimeter, Fluke/ 88464 2L-DM6-03, 9/4/2019 Ark One/NIST, ITDI/NML
KV Meter, Phenix/ KVM 100 2L-KVM-02 11/29/2019 Ark One/NIST, ITDI/NML
Test Conditions:
‘Temperature: 23 + 2 °C Relative Humidity: < 70% RH
Test Methods:
I- Measurement of the preset resistance of Megohm Decade Boxes.
II- Measurement of the UT's voltage output using Digital Multimeter.
Remarks:
Readings are within limits.
Declaration and Notes:
4. This report applies only to the instrument identified above (Make, Model, Tag No., and Serial No.),
and the data contained herein are those that were obtained at the time of calibration.
. This report may be reproduced only in full and with the permission from this Laboratory.
3. The quoted uncertainty of the standard value is obtained at 95% confidence level.
‘Test Performed by:
~ Ballon
Technician’