NPC Metrology Laboratory CERTIFICATE OF CALIBRATION CERTIFICATE NO.: 2019-051-003 for Insulation Resistance Tester Metrel Model MI 3210 Serial No.: 17010537 Submitted by: EURO-ASIA ENERGY SOURCE INCORPORATED 2nd Fir. No. 265 Unit N Cortes Bldg. Cor. Sct. Fernandez & Tomas Morato Sts., Brgy. Laging Handa, Quezon City This certifies that the instrument was calibrated by comparison against the standards maintained by the NPC-Central Metrology Laboratory using procedure PES-ESD-PM-012- WI-006. The standards used are traceable to the National Metrology Laboratory of the Department of Science and Technology and to the National Institute of Standards and Technology through Ark One Solutions, Inc. The test data in the attached Calibration Report were those obtained during the time of calibration. Itis the responsibility of the user to determine the suitability of the instrument for its intended use. f HOMER S. GALANG Mandger Testing and Metrology Services Division Date: 08 July 2019 Page 1 of 3 Quezon Avenue comer BIR Road, Diliman, 1100 Quezon City, Philippines Tel. Nos. (632) 921-3541 079 Fax No. (632) 921-2468 P.O, Box 10183 NPC Metrology Laboratory CALIBRATION REPORT 2019-051-00; Page 2 of 3 Reference: WP-ESD-TMD-019-051 Date: July 2, 2019 Description; Tnsulation Resistance Tester Make; “Metrel MI 3310 Serial No: 1701053; TagNo: N/A 1. Resistance Measurement Test Results Test STD, MEAS, NOMINAL AS FOUND (UUT) ASLEFF WUT) VOLTAGE INPUT ——UNCERT. TOLERANCE READING DEV. «READING DEW. wo ona) 10) ta) ota) oi) iy) _ te) 250 1 = 0010 + 008 1400 0.00. +1400 ~~~-0.00 100 + 028 + 8.00 100 ° 100 ° 200 4 0570 + 13.00 200 ° 200 ° 500 5 £ 004 028 499-001 4.99 0.01 100 + 0285 = 8.00 100 ° 100 ° 500 + a5 27.95 499 a 499 a 1000 eo eee 500 000-500-000 100 «0285 + 8.00 100 ° 100 ° 500 + 1425 +2795 499 499 a 2500 5 + 00m + 028 5.00 000 5.00 0.00 100 + 028 + 8.00 100 ° 100 ° 500 1425 £795 499 a 499 a 5000 so 4 04 280 49.9 “oa 49.9 “oa 100+ 028 + 8.00 100 ° 100 ° 500 1425 + 2790 498 “a 498 2 (cay ay (cay _Gy_ _ a) Te Sopra cil Toca ge Mieco Coac1l MrGoscEd (race TL, Output Voltage Test Results UUT MEASUREMENT _ NOMINAL AS FOUND (STD) AS LEFT (STD) SETTING UNCERTAINTY TOLERANCE READING »=—=DEY. «= READING OY. wo wo wo w wo w o 266 = 00203 = 1630 ~«—«a65.62—=~—~C«O BSCR gS 317 0.0305 + 3885 —515.63 137 515.63 137 aw) ow aw) aw) aw) aw) a 104 0.0040 = 0.08 1.05 “0.01 1405 “oor 3.05 = oor + oa8 3.05 0.00 3.05 0.00 5.09 + 0.0204 0.28 5.09 0.00 5.09 0.00 102 = 00409, + 054 1039 0.01 10.19 001 Reviewed by, Approved, Castro Homer $.Galang _ ‘Technician Prigeigi Engineer C MSD Manager NPC Metrology Laboratory CALIBRATION REPORT 2019-051-003 Page _3 of _3 Reference: WP-ESD-TMD-019-051 Date: July 2, 2019 Description: Insulation Resistance Tester Make: “Metrel MI 3210 Serial No.: 17010537 TagNo: N/A Test Standards Description Tag No. Calibration Due Date Calibration Traceability Meagim Desade RSI, appx s/s/2020 TEDI/NML. Merci Deeaie ee 2L-MDB-o2 1/23/2020 ITDI/NML ~~ 2L-MDB-03 3/24/2020 TIDI/NML Meonum Decade Resor, oD 1/24/2020 Mrs geen emersormga 2L-MDB-05 4/24/2020 TTDI/NML, — 2L-HRS-o1 12/12/2019 ITDI/NML, Precision Multimeter, Fluke/ 88464 2L-DM6-03, 9/4/2019 Ark One/NIST, ITDI/NML KV Meter, Phenix/ KVM 100 2L-KVM-02 11/29/2019 Ark One/NIST, ITDI/NML Test Conditions: ‘Temperature: 23 + 2 °C Relative Humidity: < 70% RH Test Methods: I- Measurement of the preset resistance of Megohm Decade Boxes. II- Measurement of the UT's voltage output using Digital Multimeter. Remarks: Readings are within limits. Declaration and Notes: 4. This report applies only to the instrument identified above (Make, Model, Tag No., and Serial No.), and the data contained herein are those that were obtained at the time of calibration. . This report may be reproduced only in full and with the permission from this Laboratory. 3. The quoted uncertainty of the standard value is obtained at 95% confidence level. ‘Test Performed by: ~ Ballon Technician’

You might also like