You are on page 1of 48

Chapter 7

Calibrations
General information about the calibrations . . . . . . . . . . . . . . . . . . . . . 7: 2 The I2C-bus . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . 7: 2 New pressure transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7: 3 I2C
Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7: 3 Calibrations . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . 7: 4 Why is calibration needed? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7: 4 Calibrating
the AD-systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7: 5 Calibration errors . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . 7: 5 Calibration errors - Pressure transducers . . . . . . . . . . . . . . . . . . 7: 5 Test equipment . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7: 6 Calibration of AD-converters . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . 7: 8 Location of testpoints for AD converter calibration . . . . . . . . . 7: 9 Calibration of
linear transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7:10 Selection diagram . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . 7:11 Venous pressure transducer calibration . . . . . . . . . . . . . . . . . . . . . . 7:12 Arterial
pressure transducer calibration . . . . . . . . . . . . . . . . . . . . . . 7:14 Priming detector calibration . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . 7:16 Occlusion adjustment, blood pump . . . . . . . . . . . . . . . . . . . . . . . . 7:17
Adjustment instruction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7:17 Blood leak detector calibration . . . . . .
. . . . . . . . . . . . . . . . . . . . . . 7:18 Variable flow calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7:20
Conductivity transducer calibration . . . . . . . . . . . . . . . . . . . . . . . . . 7:22 Dialysis pressure transducer
calibration . . . . . . . . . . . . . . . . . . . . . 7:24 Degassing pressure transducer calibration . . . . . . . . . . . . . . . . .
. . 7:26 HPG pressure transducer calibration . . . . . . . . . . . . . . . . . . . . . . . 7:28 pH transducer calibration . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7:30 Pressure regulator calibration, PR 1 . . . . . . . . . . . . . . . . . . . . . .
. . 7:32 Pressure regulator calibration, PR 2 . . . . . . . . . . . . . . . . . . . . . . . . 7:33 UF cell calibration . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7:34 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . 7:34 Prerequisite for the complete UF cell calibration . . . . . . . . . . . . 7:35 UF cell calibration:
Complete mode . . . . . . . . . . . . . . . . . . . . . . . . 7:36 During the UF cell calibration: . . . . . . . . . . . . . . . . . . .
. . . . . . 7:37 When the automatic UF cell calibration is finished: . . . . . . . . . . 7:38 If the calibration is out
of limits . . . . . . . . . . . . . . . . . . . . . . . . . 7:38 UF cell calibration: Intern mode . . . . . . . . . . . . . . . . . . . . . . .
. . . . 7:39 When the internal UF cell calibration is ready: . . . . . . . . . . . . . 7:39 UF cell calibration: Verify
mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 7:40 During the verification of the UF cell calibration: . . . . . . . . .
. . 7:41 When the verification of the UF cell calibration is finished: ... 7:42 BPM test . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7:43 Test method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. 7:43 BPM sub-tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7:43 Check of air leakage . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 7:43 Check of inflation speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7:44 Check of pressure transducer calibration and measured pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . 7:44
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:1
General information about the calibrations
During the calibration the compensation values (offset and coefficient) are calculated by AK 96
dialysis machine and stored in a small (128 bytes) E2PROM (Electrically Erasable
Programmable Read-Only Memory) placed on the actual transducer’s board.
In the beginning of the functional check the AK 96 dialysis machine will check and read all
transducer E2PROMs, and place the values in RAM to be used for the compensation.
The I2C-bus The AK 96 dialysis machine uses a separate bus – the I2C-bus – to communicate
with the transducers. The E2PROM on a transducer board contains an address – information
about the board’s physical position on the transducer motherboard (FM I/O board), physically
placed on top of the UF-measuring cell:
Transducer board
FM I/O board
UF-measuring cell

The arrangement ensures that transducers can not be interchanged without a complaint (= a
technical error) from the machine. The address – the ID-number – for a given transducer must
correspond to its physical position (the “slot”) on the transducer motherboard (FM I/O board).
The communication between a CPU and a set of transducers takes place through the I2Cbus. One
I2C-bus can handle eight E2PROMs (also called I2C-PROMs):
Transducer
Transducer
Transducer board
board
board
E
2
PROM
E
2
PROM
E
2
PROM
I2C bus

CPU
7:2 AK 96 - Calibrations HCEN12212 Revision.05.2008
There are four I2c buses, three of them are connected to transducers and the fourth is for
communication between CPU P and CPU W. Here is an overview:
CPU C:
RTM
Not Used Regulator
Cond A
HPG Not Used Degassing
Not Used Run Time
temp.
Temp A
Pressure Meter
PT-100 at heater.
E
2
PROM E
2
PROM E
2
PROM E
2
PROM E
2
PROM E
2
PROM E
2
PROM E
2
PROM
Address
0 1 2 3 4 5 6 7 CPU C
I2C bus A (I2CCA)
Not Used Venous
UF-cell
Cond B
Not Used A/D Conv Not Used Not Used Pressure
Channel 1
Temp B
MM 95 CPU C
Channel 2
ref. volt.
E
2
PROM E
2
PROM E
2
PROM E
2
PROM E
2
PROM E
2
PROM E
2
PROM E
2
PROM
Address
0 1 2 3 4 5 6 7 CPU C
I2C bus B (I2CCB)

CPU P:
I2C bus A (I2CPA) not used!
Blood
Not Used pH Cond P
Diascan Dialysis
Venous
Arterial leak
Temp P
pressure
pressure
pressure
E
2
PROM E
2
PROM E
2
PROM E
2
PROM E
2
PROM E
2
PROM E
2
PROM E
2
PROM
Address
0 1 2 3 4 5 6 7 CPU P
I2C bus B (I2CPB)

CPU P

I2C bus CPU W New pressure transducers Since a new


pressure transducer has an ID from it’s calibration in production, a technical error will appear
when it is installed.
Therefore always perform an I2C scanning when a new transducer has been installed.
I2C Errors The error codes for the I2C buses are found as C CC 008 xxx for CPU C (I2CCA and
I2CCB) and as P PTRA for CPU P (I2CPB).
If all error codes for one bus are issued it is most likely that one of the transducer has an error
that takes “down” the whole bus. The best solution to the problem is to disconnect one transducer
at the time until only one error code is issued. Then the disconnected transducer is the faulty one
and should be exchanged for a new one. If error codes for more than one I2C bus is issued the
error is probably due to some defect on the FM I/O board.
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:3
Calibrations The different transducer are calibrated in different ways. The pressure transducers
are calibrated in three steps (0 mmHg, -500 mmHg and +500 mmHg) while e.g. the conductivity
cells are calibrated in two steps (0.0 mS and 14.0 mS)
The calibration of the UF-cell differs radically from the rest, and requires a so called calibration
station.
Why is calibration needed? No transducer is ideal – all transducers have two basic errors: a
sensitivity error and an offset error:
Output signal (+)
Ideal
Offset error
Real
Input (-)
Input (+) Sensitivy error
Output signal (-)

To compensate for these two, the machine must calculate an offset value which will bring the line
to cross through the zero point, and a calibration coefficient – a factor, typically between 0.90 –
1.10 which will correct the slope of the line so it matches the ideal.
No pressure
+500mmHg
E
2
PROM
-500mmHg

Note
Never try to fit a another transducer board to a transducer since there is an analog trimming made with a resistor
(laser trimmed) during manufacturing.
Note
7:4 AK 96 - Calibrations HCEN12212 Revision.05.2008
Calibrating the AD-systems Since all transducers are read by an AD-system (MUX +
AD-converter), it is necessary to calibrate these before calibration of any transducer. There are
four different AD-converters in the AK 96 dialysis machine according to page 7: 8
An AD calibration is very simple: It involves measuring the voltage across a resistor (a reference
voltage), and entering the reading in the respective calibration menus.
Calibration errors If the machine finds that one of the calculations generated an offset or a
coefficient which is outside its limits, it will generate a calibration error. There are several
possibilities:
Calibration error Reason for the error
Set value and measured value are too separated
If a value, far away from the reading on the reference instrument is entered. Typically, entering a positive
pressure, when the actual pressure is negative.
Set value is too near values entered earlier
The calibration achieves highest accuracy, if the measuring points are far apart.
A three-point calibration with zero, -15mm Hg and
+18 mmHg will generate a poor result.

Calibration errors - Pressure transducers If the AK 96 dialysis machine finds, during the pressure
transducer calibration procedure, that a calculated offset or calibration coefficient is outside
certain limits, a calibration error appears. These limits are:
• Offset: ±20 mmHg
• Calibration Coefficient: 0.9 – 1.1
• Correlation Coefficient: Low limit 0.95
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:5
Test equipment
Following calibration instructions are based on the availability of some test equipment, listed
below:
Test equipment Tolerance Manufacturer Order no
Conductivity measurement
±0.1 mS/cm MESA, IBP
Pressure measurement ±2 mmHg to ±200 mmHg ±1% beyond ±200 mmHg If BPM is installed: ±0.8 mmHg"
MESA, IBP, Druck
UF calibration box Gambro K21908002
Temperature measurement:
±0.2 oC MESA, IBP
Digital Voltmeter ±0.5% Fluke
Blood leak Calibration Cover
N/A Gambro K40169001
Gauge pin kit for blood pump rollers (including three different pins)
T = 1.1 (Stop pin = 1.1x2x0.7 = 1.5 mm) T = 1.6 (Stop pin = 1.6x2x0.7 = 2.2 mm) T = 2.0 (Stop pin = 2.0x2x0.7 = 2.8 mm) T =
wall thickness of the blood tubing
Gambro K40158001
Gauge pin kit for blood pump rollers (to be ordered separately)
T = 1.1 (Stop pin = 1.1x2x0.7 = 1.5 mm) T = 1.6 (Stop pin = 1.6x2x0.7 = 2.2 mm) T = 2.0 (Stop pin = 2.0x2x0.7 = 2.8 mm) T =
wall thickness of the blood tubing
Gambro K40158002 K40158003 K40158004
Calibration tube kit N/A Gambro K13983002
Tube for calibration of priming detector
N/A
BPM test equipment N/A Gambro K22151001
BPM Cuff N/A Gambro 110350
pH buffers ± 0.07 pH units pH 7- 100810001
pH 9 - 100810002
Special tool For silicone tube connectors Gambro K12755001
7:6 AK 96 - Calibrations HCEN12212 Revision.05.2008
This page is intentionally left blank
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:7
Calibration of AD-converters
1. Start the AK 96 dialysis machine, enter the Service menu and select
Calibrations.
2. Choose AD-converter.
When selecting “AD-converter” following display appears:
SELECT AD-CONVERTER CPU-C CPU-P CPU-W MM95
3. Select AD-converter to calibrate with the Cursor keys and press the
Select key to enter the calibration.
CPU X AD LOW REF SET HIGH xx.xx mV xx.xx xx.xx xx.xx
4. Measure the voltage (using a 4.5 digit voltmeter), between following
testpoints or across the resistor:
CPU Testpoint
Ref
Tetspoint Circuit board Limits
(mV)
CPU-C TP 6 TP 5 Control Driver
board
45.00 - 55.00
CPU-P TP 1 TP 2 Protective Driver
board
45.00 - 55.00
CPU-W TP 3 TP 4 Protective CPU
board
4500 - 5500
MM 95 R 24 R 24 MM I/O board 45.00 - 55.00

5. Set the “AD REF VOLTAGE” on the display to the value shown on
the voltmeter and press the Select key.
6. Press the lit UF Start/Stop button to load the value into the E2PROM.
CPU X AD-REF VOLTAGE STORED
7. Press the Back key to exit.
7:8 AK 96 - Calibrations HCEN12212 Revision.05.2008
Location of testpoints for AD converter calibration
TP 6
AD ref voltage range: AD ref voltage range:
45.00 to 55.00 mV 45.00 to 55.00 mV
PVREF
TP1 TP 5
TP2
CPU C A/D ref
Z0VAD
Control driver board Protective driver board
AD ref voltage range: 4.500 to 5.500 V
TP3 WREF
CPU W PROM
CPU W SafeRAM
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:9
MM I/O AD ref voltage range: 45.00 to 55.00 mV
(TP1) (TP2) TP4 Z0VRW
R24
pH plug
Protective CPU board MM I/O board
Calibration of linear transducers
1. Start the AK 96 dialysis machine, enter the Service menu and select
Calibrations.
2. Choose Transducers.
3. Select Extern or Internal flow support.
Note
• Select Extern for no flow (pumps stopped).
• Select Internal and the AK 96 maintains a flow during the calibration.
Note
Support condition for different calibrations
Calibration Flow support condition
Venous pressure transducer CPU P Internal or External
Venous pressure transducer CPU C Internal or External
Arterial pressure transducer Internal or External
Priming detector Internal or External
Blood leak detector Internal
Conductivity cell Internal
Dialysis pressure transducer External
Degassing pressure transducer External
HPG pressure transducer External
pH transducer External
Run time meter Internal or External

The machine enters the Selection diagram. Refer to page 7:11.


4. Select transducer to calibrate with the Cursor keys and press the
Select key to enter the calibration.
5. Follow the instructions for each calibration.
Note
If the calibration is to be aborted without saving the new calibration values: Press Back key.
Note
7:10 AK 96 - Calibrations HCEN12212 Revision.05.2008
Selection diagram
VENOUS PRESSURE TRANSDUCER (P)
RUN TIME METER
VENOUS PRESSURE TRANSDUCER (C) pH TRANSDUCER ARTERIAL PRESSURE TRANSDUCER
HPG PRESSURE TRANSDUCER
PRIMING DETECTOR
DEGASSING PRESSURE TRANSDUCER
BLOOD LEAK DETECTOR
DIALYSIS PRESSURE TRANSDUCER
CONDUCTIVITY A TRANSDUCER
CONDUCTIVITY A B P TRANSDUCER
CONDUCTIVITY B TRANSDUCER CONDUCTIVITY P TRANSDUCER
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:11
2. Zero set the reference instrument, wait for stable reading.
3. Start the AK 96 dialysis machine, enter the Service menu and select
Calibrations.
4. Select Transducers, External and VENOUS PRESSURE
TRANSDUCER CPU P from the selection diagram.
5. Press the Select key to confirm the selected transducer.
6. Apply zero pressure by opening the three clamps A, B and C on the calibration tube set. Set
the “REF SET” value on the AK 96 to 0 mmHg.
7. Press the Select key to enter the setting.
8. Close clamp A. Apply a pressure of approx. -500 mmHg, shown on
the reference instrument. Close clamp B.
VEN PR P LOW REF SET HIGH -500 -550 -500 550
9. Set the “REF SET” value on the AK 96 to the reading on the
reference instrument.
10. Press the Select key to enter the setting.
7:12 AK 96 - Calibrations HCEN12212 Revision.05.2008

Venous pressure transducer calibration


1. Connect the Gambro reference instrument (together with a
calibration tube set) to the venous pressure transducer (blue) on the front of the AK 96 dialysis
machine, according to following figure.
C
D
X
AB
11. The machine now enters the sensitivity calibration.
Press the Select key to continue.
12. Open clamp A and B, then close clamp A again. Apply a pressure
of approx. +500 mmHg, shown on the reference instrument. Close clamp B.
VEN PR P LOW REF SET HIGH 500 -550 +500 550
13. Set the “REF SET” value on the AK 96 to the reading on the
reference instrument.
14. Press the Select key to enter the setting.
15. Press the lit UF Start/Stop button to load the calibration values into
the E2PROM.
16. The calibration is completed when the information display shows:
EEPROM STORAGE COMPLETED Press BACK to exit
17. Press the Back key to exit.
18. Continue with the venous pressure transducer calibration for CPU C. Follow this instruction
except for paragraph 4, where VENOUS PRESSURE TRANSDUCER CPU C shall be selected
instead of VENOUS PRESSURE TRANSDUCER CPU P.
Note
If the calibration is to be aborted without saving the new calibration values: Press Back key.
Note
Note
If a calibration error occurs:
• Try to do the calibration again.
• If the error occurs a second time: Change the transducer and the transducer board. Report the error message on the
error display to your Gambro service representative.
Note
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:13
2. Zero set the reference instrument, wait for stable reading.
3. Start the AK 96 dialysis machine, enter the Service menu and select
Calibrations.
4. Select Transducers, External and ARTERIAL PRESSURE
TRANSDUCER CPU P from the selection diagram.
5. Press the Select key to confirm the selected transducer.
6. Apply zero pressure by opening the three clamps A, B and C on the calibration tube set. Set
the “REF SET” value on the AK 96 to 0 mmHg.
7. Press the Select key to enter the setting.
8. Close clamp A. Apply a pressure of approx. -500 mmHg, shown on
the reference instrument. Close clamp B.
ART PR LOW REF SET HIGH -500 -550 -500 550
9. Set the “REF SET” value on the AK 96 to the reading on the
reference instrument.
10. Press the Select key to enter the setting.
7:14 AK 96 - Calibrations HCEN12212 Revision.05.2008

Arterial pressure transducer calibration


1. Connect the Gambro reference instrument (together with a
calibration tube set) to the arterial pressure transducer (red) on the front of the AK 96 dialysis
machine, according to following figure.
C
D
X
AB
11. The machine now enters the sensitivity calibration.
Press the Select key to continue.
12. Open clamp A and B, then close clamp A again. Apply a pressure
of approx. +500 mmHg, shown on the reference instrument. Close clamp B.
ART PR LOW REF SET HIGH 500 -550 +500 550
13. Set the “REF SET” value on the AK 96 to the reading on the
reference instrument.
14. Press the Select key to enter the setting.
15. Press the lit UF Start/Stop button to load the calibration values into
the E2PROM.
16. The calibration is completed when the information display shows:
EEPROM STORAGE COMPLETED Press BACK to exit
17. Press the Back key to exit.
Note
If the calibration is to be aborted without saving the new calibration values: Press Back key.
Note
Note
If a calibration error occurs:
• Try to do the calibration again.
• If the error occurs a second time: Change the transducer and the transducer board. Report the error message on the
error display to your Gambro service representative.
Note
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:15
Priming detector calibration
1. Start the AK 96 dialysis machine, enter the Service menu and select
Calibrations.
2. Select Transducer, External and PRIMING DETECTOR from
the selection diagram.
3. Press the Select key to start calibration routine.
4. The display shows:
INSERT TUBE Press SELECT to confirm
5. Insert an empty tube in the priming detector and press the Select key.
Use the normal tube from the blood line.
6. The calibration is now done automatically and the calibration factor
is written into the E2PROM.
7. The calibration is completed when the information display shows:
EEPROM STORAGE COMPLETED Press BACK to exit
8. Press the Back key to exit.
Note
If the calibration is to be aborted without saving the new calibration values: Press Back key.
Note
Note
If an technical error is detected, the calibration must be restarted, or the priming detector replaced.
Note
7:16 AK 96 - Calibrations HCEN12212 Revision.05.2008
• The GO pin should barely pass within the calibrating area.
• The STOP pin is NOT allowed to pass in any point within the calibrating area.
Note
Put grease on the GO and STOP pins when not used and store gage in plastic bag to prevent it from corroding.
Note
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:17

Occlusion adjustment, blood pump


• Check surface on the roller at the pump rotor, if it is damaged exchange the unit.
• The roller unit should easily go back to upper end position when it has been pressed down to
end position, exchange unit if it gets stuck in any position.
Note
If the blood lines/infusion lines have a different wall thickness, the occlusion of the blood pump must be adjusted to
the right size.
Note

Use the gauge pins (stop/go) available in the kit K40158001. The kit includes following three
different gauge pins: T = 1.1 (Stop pin = 1.1x2x0.7 = 1.5mm) T = 1.6 (Stop pin = 1.6x2x0.7 =
2.2mm) T = 2.0 (Stop pin = 2.0x2x0.7 = 2.8mm)
T = wall thickness of the blood tubing
Use following formula to calculate which gauge pin to use: 2 x wall thickness x 0,7 (mm)
Adjustment instruction Check the range between the blood pump rollers and the path according
to the picture below. Adjust range between the blood pump rollers and the path if it is necessary.
Blood leak detector calibration
1. Make sure there is no fluid in the AK 96 monitor. If there is fluid in the monitor, press
Rinse/Disinfection-button to drain the machine.
2. Remove the blood leak detector cover and clean the inside of the
blood leak housing.
3. Mount a blood leak calibration cover (K40169001).
4. Start the AK 96 dialysis machine, enter the Service menu and select
Calibrations.
5. Select Transducer, Intern and BLOOD LEAK DETECTOR
from the selection diagram.
6. Place the A-pickup tube in the concentrate container.
7. Press the Select key to enter the Blood leak calibration routine.
Let the machine stabilize in conductivity and temperature a few minutes. Wait for green fluid
path.
8. Make sure the filter is in horizontal position and set “REF SET” to 0.
9. Press the Select key when the BL LEAK value is stable.
10. Turn the calibration filter to vertical position and set “REF SET”
to 100.
11. When the BL LEAK value is stable, press the Select key to enter the
setting.
12. The machine now enters the sensitivity calibration.
Press the Select key to continue.
13. Press the lit UF Start/Stop button to store the calibration values.
14. The calibration is completed when the information display shows:
EEPROM STORAGE COMPLETED Press BACK to exit
7:18 AK 96 - Calibrations HCEN12212 Revision.05.2008
15. Press the Back key to exit.
16. Drain the AK 96 dialysis machine and remount the standard blood
leak cover.
Note
If the calibration is to be aborted without saving the new calibration values: Press Back key.
Note
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:19
Variable flow calibration
The variable flow calibration will set the number of steps needed to obtain a flow of 500 ml/min.
The value will then be used during functional check and when entering service mode as the flow
always should be 500 ml/min in functional check and service mode.
1. Start the AK 96 dialysis machine, enter the Service menu and select
Calibrations.
2. Place A-pickup tube in the concentrate container.
3. Select Other / Variable flow to initiate the calibration.
4. Following text will be displayed during the calibration:
STEPS: XXX FLOW: ---- ml/min Please wait. Press BACK to abort
STEPS: the position of the stepper motor FLOW: the average dialysis flow
5. When the AK 96 dialysis machine has found a new calibration value it will automatically store
the value in the Safe-RAM. Following text will automatically appear:
ADJUSTING VARIABLE DIALYSIS FLOW Please wait XX seconds
6. At the end of the calibration the new calibration data is displayed:
VAR. FLOW MOTOR POSITION STORED NEW VALUE: XXX OLD VALUE: YYY
7. Press the Back key to finish the calibration.
7:20 AK 96 - Calibrations HCEN12212 Revision.05.2008
Note
During the variable flow calibration it is possible to supervise the conductivity and temperature.
If variable flow is not installed following attention will be displayed:

VARIABLE FLOW IS NOT INSTALLED Check preset


Note
Note
If the calibration is to be aborted without saving the new calibration values: Press Back key.
Note
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:21
Conductivity transducer calibration
The calibration is a "two point calibration". The two points are 0.0 mS/cm (RO-water) and 14.0
mS/cm (with concentrate).
WARNING
After completing the conductivity calibrations, the conductivity value must be confirmed by the hospital laboratory.
WARNING
Note
• Before conductivity calibration, Chemical cleaning must be performed, one with Citric acid and one with
Hypochlorite.
• For machines with variable flow: Make sure the variable flow is calibrated before performing the conductivity cell
calibration. Refer to page: 7:20
Note

1. Start the AK 96 dialysis machine, enter the Service menu and select
Calibrations.
2. Select Transducer, Intern.
3. Select the conductivity transducer to be calibrated from the selection
diagram. The possible cond. cell calibrations are:
• CONDUCTIVITY A TRANSDUCER
• CONDUCTIVITY B TRANSDUCER
• CONDUCTIVITY P TRANSDUCER
• CONDUCTIVITY ABP TRANSDUCERS
4. Press the Fluid bypass button.
5. Connect the Gambro reference instrument to the dialyzer connectors.
6. Press the Fluid bypass button.
7. Connect both A- and B-concentrate connectors to the AK 96.
8. Let the AK 96 and the reference instrument stabilize for at least 30 minutes. Make sure that
the machine is not in bypass, i.e. the Fluid bypass button is not lit.
Note
There must be flow through the measuring cell of the reference instrument during the conductivity cell calibration.
Note

9. Press the Select key to enter the calibration routine.


7:22 AK 96 - Calibrations HCEN12212 Revision.05.2008
10. Calibration with RO-water
10.a. Wait until the reading on the reference instrument has stabilized
at 0.0 mS/cm.
10.b. Check that the temperature on the reference instrument is
approx. 37 °C.
10.c. Set REF SET to 00.0.
10.d. Press the Select key to enter the value.
11. Calibration with concentrate
11.a. Place A-pickup tube in the concentrate container.
11.b. The AK 96 will now regulate conductivity to approx.
14 mS/cm. Let the conductivity stabilize for 20 minutes. The green fluid path must be lit.
11.c. Check that the temperature on the reference instrument is
approx. 37 °C.
11.d. Set the “REF. SET” value on the AK 96 display to the same
value shown on the reference instrument.
11.e. Press the Select key to enter the value.
12. The machine now enters the sensitivity calibration.
Press the Select key to continue.
13. Press the lit UF Start/Stop button to load the calibration values into
the E2PROM.
14. The calibration is completed when the information display shows:
EEPROM STORAGE COMPLETED Press BACK to exit
15. Press the Back key to exit.
Note
All the transducers A, B and P are all calibrated in this way.
Note
Note
If the calibration is to be aborted without saving the new calibration values: Press Back key.
Note
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:23
Dialysis pressure transducer calibration
1. Disconnect tube no. 24316.
Insert a plug (from the calibration kit) according to the figure below.
2. Disconnect the angled silicone connector from TAVA. Insert a plug
(from the calibration kit) according to the figure below.
3. Disconnect tube no. 24458 from the top of the deairating chamber.
Connect Gambro reference instrument to the silicone connector according to the figure below.
D
2
3
1
C
B
A
X

4. Zeroset the reference instrument, wait for stable reading.


5. Start the AK 96 dialysis machine, enter the Service menu and select
Calibrations.
6. Select Transducer, External and DIALYSIS PRESSURE
TRANSDUCER from the selection diagram.
7. Press the Select key to confirm the selected transducer.
8. Apply zero pressure by opening the two clamps A, and B on the
calibration tube set. Set the “REF SET” value on the AK 96 to 0 mmHg.
9. Press the Select key to enter the setting.
10. Close clamp A. Apply a pressure of approx. -500 mmHg, shown on
the reference instrument. Close clamp B.
7:24 AK 96 - Calibrations HCEN12212 Revision.05.2008
11. Set the “REF SET” value on the AK 96 to the reading on the
reference instrument.
12. Press the Select key to enter the setting.
13. The machine now enters the sensitivity calibration.
Press the Select key to continue.
14. Open clamp A and B, then close clamp A again. Apply a pressure
of approx. +500 mmHg, shown on the reference instrument. Close clamp B.
15. Set the “REF SET” value on the AK 96 to the reading of the
reference instrument.
16. Press the Select key to enter the setting.
17. Press the lit UF Start/Stop button to load the calibration values into
the E2PROM.
18. The calibration is completed when the information display shows:
EEPROM STORAGE COMPLETED Press BACK to exit
19. Press the Back key to exit.
If a calibration error occurs:
• Try to do the calibration again.
• If the error occurs a second time: Change the transducer and the transducer board. Report the
error message on the error display to your Gambro service representative.
Note
If the calibration is to be aborted without saving the new calibration values: Press Back key.
Note
Note
• Reconnect the fluid path to its original design after the calibration.
• Always perform a Heat disinfection after the fluid path has been opened.
Note
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:25
Degassing pressure transducer calibration
1. Replace the silicone T-connector to the degassing pressure transducer
with a straight silicone connector. Insert a plug (from the calibration kit) according to the figure
below.
2. Disconnect the degassing pressure transducer from the flow pump.
Connect Gambro reference instrument to the silicone connector according to the figure below.
DX
1
2
C
B
A

3. Zeroset the reference instrument, wait for stable reading.


4. Start the AK 96 dialysis machine, enter the Service menu and select
Calibrations.
5. Select Transducer, External and DEGASSING PRESSURE
TRANSDUCER from the selection diagram.
6. Press the Select key to confirm the selected transducer.
7. Apply zero pressure by opening the two clamps A, and B on the
calibration tube set. Set the “REF SET” value on the AK 96 to 0 mmHg.
8. Press the Select key to enter the setting.
9. Close clamp A. Apply a pressure of approx. -500 mmHg, shown on
the reference instrument. Close clamp B.
10. Set the “REF SET” value on the AK 96 to the reading on the
reference instrument.
11. Press the Select key to enter the setting.
7:26 AK 96 - Calibrations HCEN12212 Revision.05.2008
12. The machine now enters the sensitivity calibration.
Press the Select key to continue.
13. Open clamp A and B, then close clamp A again. Apply a pressure
of approx. +500 mmHg, shown on the reference instrument. Close clamp B.
14. Set the “REF SET” value on the AK 96 to the reading of the
reference instrument.
15. Press the Select key to enter the setting.
16. Press the lit UF Start/Stop button to load the calibration values into
the E2PROM.
17. The calibration is completed when the information display shows:
EEPROM STORAGE COMPLETED Press BACK to exit
18. Press the Back key to exit.
If a calibration error occurs:
• Try to do the calibration again.
• If the error occurs a second time: Change the transducer and the transducer board. Report the
error message on the error display to your Gambro service representative.
Note
If the calibration is to be aborted without saving the new calibration values: Press Back key.
Note
Note
• Reconnect the fluid path to its original design after the calibration.
• Always perform a Heat disinfection after the fluid path has been opened.
Note
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:27
HPG pressure transducer calibration
1. Replace the silicone T-connector from the HPG pressure transducer with a straight silicone
connector. Insert a plug (from the calibration kit) according to the figure below.
2. Connect Gambro reference instrument to the angled silicone
connector according to the figure below.
DX
1
2
C
B
A

3. Zeroset the reference instrument, wait for stable reading.


4. Start the AK 96 dialysis machine, enter the Service menu and select
Calibrations.
5. Select Transducer, External and HPG PRESSURE
TRANSDUCER from the selection diagram.
6. Press the Select key to confirm the selected transducer.
7. Apply zero pressure by opening the two clamps A, and B on the
calibration tube set. Set the “REF SET” value on the AK 96 to 0 mmHg.
8. Press the Select key to enter the setting.
9. Close clamp A. Apply a pressure of approx. -500 mmHg, shown on
the reference instrument. Close clamp B.
10. Set the “REF SET” value on the AK 96 to the reading on the
reference instrument.
11. Press the Select key to enter the setting.
7:28 AK 96 - Calibrations HCEN12212 Revision.05.2008
12. The machine now enters the sensitivity calibration.
Press the Select key to continue.
13. Open clamp A and B, then close clamp A again. Apply a pressure
of approx. +500 mmHg, shown on the reference instrument. Close clamp B.
14. Set the “REF SET” value on the AK 96 to the reading of the
reference instrument.
15. Press the Select key to enter the setting.
16. Press the lit UF Start/Stop button to load the calibration values into
the E2PROM.
17. The calibration is completed when the information display shows:
EEPROM STORAGE COMPLETED Press BACK to exit
18. Press the Back key to exit.
If a calibration error occurs:
• Try to do the calibration again.
• If the error occurs a second time: Change the transducer and the transducer board. Report the
error message on the error display to your Gambro service representative.
Note
If the calibration is to be aborted without saving the new calibration values: Press Back key.
Note
Note
• Reconnect the fluid path to its original design after the calibration.
• Always perform a Heat disinfection after the fluid path has been opened.
Note
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:29
pH transducer calibration
A complete calibration procedure consists of the following steps:
• Offset calibration, with a buffer solution (a solution with a very accurate and stable pH-value)
having a pH-value in the range of pH 6.5-7.5, followed by:
• Sensitivity calibration with a buffer solution having pH-value that differs at least 2 pH units
from the buffer used for offset calibration.
1. Make sure there is no flow or fluid in the AK 96. The inlet water
must be closed.
2. Start the AK 96 dialysis machine, enter the Service menu and select
Calibrations.
3. Select Transducer, External and pH TRANSDUCER from the
selection diagram.
4. Press the Select key to start calibration routine.
5. Remove the pH sensor from the pH sensor housing by loosening the nut holding the pH sensor
and rinse the sensor in distilled water.
6. Check that the “REF SET” on the AK 96 displays pH 7.
7. Put the pH sensor in a cup with pH 7 buffer solution (ordering no.
100810001) which must cover the glass membrane i.e. the level in the cup should be at least 30
mm from the bottom.
8. Stir in order to achieve balance in the buffer solution. Do not let the
pH sensor lean against the bottom of the cup. Wait for 2 minutes.
9. Press the Select key to set the “REF SET” value on the AK 96 and
wait until the calibration display returns.
Note
If an error message is shown after this step, check the buffer solutions, level in the cup and repeat the procedure from
step 1. If there still is an error after 2nd try, replace the pH-sensor.
Note

10. Rinse the pH sensor in distilled water.


11. Check that the “REF SET” on the AK 96 displays pH 9.
12. Put the pH sensor in a cup with pH 9 buffer solution (ordering no.
100810002) which must cover the glass membrane i.e. the level in the cup should be at least 30
mm from the bottom.
13. Stir in order to achieve balance in the buffer solution. The pH sensor must not lean against
the bottom of the cup. Wait for 2 minutes.
7:30 AK 96 - Calibrations HCEN12212 Revision.05.2008
14. Press the Select key and wait. The display shows: Entering sensitivity
calibration.
Note
If an error message is shown after this step, check the buffer solutions, level in the cup and repeat the procedure from
step 1. If there still is an error after 2nd try, replace the pH-sensor.
Note

15. Press the Select key to confirm that the calibration shall continue.
16. When the calibration display returns, press the lit UF Start/Stop button
to load the calibration values into the E2PROM.
17. The calibration is completed when the information display shows:
EEPROM STORAGE COMPLETED Press BACK to exit
18. Press the Back key to exit.
19. Remount the pH-sensor into the pH sensor housing.
Note
If the calibration is to be aborted without saving the new calibration values: Press Back key.
Note
Note
• pH calibration should be performed (according to the Maintenance manual) on a regular basis to ensure optimum
performance.
Note
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:31
Pressure regulator calibration, PR 1
1. Turn off the water supply.
2. Connect Gambro reference instrument just before the INPS pressure
switch, according to following figure.
D
X
PR 2
C
A
INPS Pressure regulator, PR 1
B

3. Close clamp A, B and C.


4. Turn on the water supply.
5. Start the AK 96 dialysis machine.
6. Perform a functional check and wait until the conductivity and the
temperature are stable. The fluid path on the operator’s-panel must be green before the
calibration begins.
7. Open clamp C.
8. Adjust the pressure regulator, PR 1, until the pressure measured at
the reference instrument is 600 mmHg ±75 mmHg.
9. Perform a Drain.
10. Disconnect the Gambro reference instrument.
Note
• Reconnect the fluid path to its original design after the calibration.
• Always perform a Heat disinfection after the fluid path has been opened.
Note
7:32 AK 96 - Calibrations HCEN12212 Revision.05.2008
Pressure regulator calibration, PR 2
1. Turn off the water supply.
2. Disconnect tube no. 24304 from the angled silicone connector (after
the PR 2).
3. Insert a silicone T-connector according to the figure below. Use a
short tube from the calibration kit and connect the T-connector to the angled silicone connector.
4. Connect the reference instrument to the T-connector according to
the figure below.
5. Close clamp A, B and C.
DX
A
Pressure regulator, PR 2
C
INPS
B

6. Zeroset the Reference instrument, wait for stable reading.


7. Turn on the water supply.
8. Start the AK 96 dialysis machine.
9. Perform a functional check and wait until the conductivity and the
temperature are stable. The fluid path on the OP-panel must be green before the calibration
begins.
10. Open clamp C.
11. Adjust the pressure regulator, PR 2, until the pressure measured at
the reference instrument is 130 mmHg ± 10 mmHg.
12. Perform a Drain.
Note
• Reconnect the fluid path to its original design after the calibration.
• Always perform a Heat disinfection after the fluid path has been opened.
Note
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:33
UF cell calibration
General
This section describes the calibration routine of the UF-measuring unit.
The UF-measuring unit consists of two flow measurement channels called channel 1 (Ch1) and
channel 2 (Ch2).
T
12
Cond. Cell P
ZEVA
P
HPG
UF-measuring cell
TAVA
BYVA
DIVA

Ch1 is the channel used to measure the main flow to the dialyzer and Ch2 is used to measure the
flow from the dialyzer, i.e. the sum of the main flow and the UF flow. The UF flow is the
difference between Ch2 and Ch1 flows. UF flow is also called differential flow.
The UF cell calibration routine can be run in 3 different modes. The modes are Complete, Intern
and Verify.
• Select Complete for an automatic UF cell calibration.
• Select Intern for an internal calibration (taration but updating the E2PROM instead of RAM).
This is done without the external UF calibration equipment connected to the AK 96 dialysis
machine.
• Select Verify for a verification of the result from the latest calibration.
7:34 AK 96 - Calibrations HCEN12212 Revision.05.2008
Prerequisite for the complete UF cell calibration
• Perform both a disinfection and a cleaning decalcification to remove both organic and calcium
precipitates, according to the AK 96 Operator’s Manual.
• For machines with variable flow: Make sure the variable flow is calibrated before performing
the UF cell calibration. Refer to page: 7:20
• The Complete calibration routine requires that the external UF calibration station (ordering no.
K21908002) is connected to the AK 96, see fig. below.
AK 96 High Level
V.ref.
Low Level
UFVA
Flow in/out
Connection for the UF calibration cable
24 V DC from power adapter
Dialysis fluid nipples Drain nipple

The equipment contains a reference volume, a flow directing valve (UFVA) and electronics. The
UF calibration station controls the directing valve by measuring low level and high level of the
reference volume and depending on the selected UF mode the valve will become closed or
opened when low or high level is detected.
Low Level fill up. The reference is filled just over the first
level transducer. Low Level drain. The reference is emptied just below the
first Level transducer. High Level fill up. The reference is filled to just over the
second level transducer. High level drain. The reference is emptied just below the
second Level transducer.
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:35
UF cell calibration: Complete mode
Note
• Make sure that the AK 96 dialysis machine is complete shut down. All electrical connections must be done before
the power for AK 96 is turned on.
• Be aware of that there is no main switch on the UF calibration box.
Note

1. Connect the UF calibration station to the AK 96 external


communication port. Use the UF calibration cable, K21876001.
2. Connect the UF calibration unit to the mains via the power adapter.
3. Start the AK 96 dialysis machine with the main switch and the
ON/OFF button.
4. Place B-pickup tube in the concentrate container.
Note
If the machine is equipped with a UFD-kit, let the AK 96 pass FCH (wait for green fluid path) before entering
Service mode.
Note

5. Enter the Service menu.


6. Select Calibrations/Transducers/UF/Complete.
7. Connect the dialysis fluid tubes to the UF calibration box.
8. Enter the Reference volume with the Display Up/Down keys (The
Reference volume is labelled on the UF Calibration unit)
9. Press the Select key to confirm.
10. After the Reference volume is confirmed the UF cell calibration runs
automatically, until the calibration is finished.
7:36 AK 96 - Calibrations HCEN12212 Revision.05.2008
During the UF cell calibration: The time display shows the time since the calibration was started.
The information displays also shows the different stages during the calibration:
• E2PROM test: The E2PROMS are tested incase there is a communication error with the UF
cell.
• Fluid preparation: The AK 96 dialysis machine awaits that the fluid is correct in temperature
and conductivity (37 °C and 14.4 mS on the B Cell)
• Low and High level fill up: During this period the reference volume is filled and emptied, this is
done to check that the tool is working properly.
• UF stabilization: During this period the fluid path is heated up so there is no temperature drift in
the components.
• Main flow offset calibration: During this period the offset on both channels is measured and
saved in RAM memory. The test is performed in the same way as during taration.
• Ch1 flow calibration: The valves are set so the flow will go through the Ch1 until the reference
volume is filled. Since the reference volume is known the flow can be calculated. This value is
used to calculate a new coefficient for the Ch1 so the measured volume match the actual volume.
The new coefficient for Ch1 is saved in RAM and used in the next test.
• Differential flow calibration: The new coefficient (Ch1) is used when the Ch2 coefficient is
calculated by running the same flow trough both channels. i.e. the same as differential flow in
taration.
• Verifying calibration of UF-cell: The verification is done by draining the reference volume
through the Ch2. Since it is a known volume the accuracy of the UF measuring system is
measured.
• Verifying calibration of Isolated UF: The reference volume is then filled through Ch2, the flow
is backwards through the cell, negative value on Ch2. When it is filled it is emptied just through
the Ch2 and the volume is measured by Ch2 only and the UF cell is checked for accuracy during
Isolated UF.
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:37
When the automatic UF cell calibration is finished: 1. Press the Back key.
After the calibration AK 96 continues to store the calculated values. When the storage is
completed, the verification results are displayed.
First the display for UF Volume (for 3 seconds):
UF VOL REF VOL VOL DIFF 1006.5 1007.3 -0.28 %
Then the display for Isolated UF Volume (for 3 seconds):
ISO VOL REF VOL VOL DIFF 1007.5 1007.3 +0.28 %
Then the previous display is showed again for 3 seconds and so on.
When ready, press the Back key to exit the Complete UF cell calibration.
Note
Perform a Rinse before leaving the AK 96 dialysis machine.
Note

If the calibration is out of limits If the calibration is out of limits then it will be displayed as
calibration failed.
The maximum limits are:
• UF volume deviation ±1.0 %
• Isolated UF volume ±1.0 %
Note
The other values such as offset and coefficients can be found in the GXL menu for UF Calibration logging: For Ch1
and Ch2 Offset ±500 Coefficient 0.92-1.02
Note
Note
The equipment is turned off in the following order: 1. Press AK 96 ON/OFF button.
2. Turn off the AK 96 main switch.
3. Disconnect the UF calibration box from the power socket.
Note
7:38 AK 96 - Calibrations HCEN12212 Revision.05.2008
UF cell calibration: Intern mode
Note
Select Intern for an internal calibration (taration but updating the E2PROM instead of RAM). This is done without
the external UF calibration equipment connected to the AK 96 dialysis machine.
Note

1. Start the AK 96 with the main switch and the ON/OFF button.
2. Place B-pickup tube in the concentrate container.
3. Enter the Service menu
4. Select Calibrations/Transducers/UF/Intern.
The internal UF cell calibration runs automatically, until the calibration is finished.
When the internal UF cell calibration is ready: After the calibration, AK 96 displays:
INTERNAL UF CALIBRATION READY SELECT for info, BACK to save
Press the Select key to show more information.
Press the Back key to save the calibrations values.
When the storage is completed, AK 96 displays:
EEPROM STORAGE COMPLETED Press BACK to exit
When ready, press the Back key to exit the internal UF cell calibration.
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:39
UF cell calibration: Verify mode
Note
• Make sure that the AK 96 dialysis machine is complete shut down. All electrical connections must be done before
the power for AK 96 dialysis machine is turned on.
• Be aware of that there is no main switch on the UF calibration box.
Note

1. Connect the UF calibration box to the AK 96 external communication


port. Use the UF calibration cable, K21876001.
2. Connect the UF calibration unit to the mains via the power adapter.
3. Start the AK 96 with the main switch and the ON/OFF button.
4. Place B-pickup tube in the concentrate container.
Note
If the machine is equipped with a UFD-kit, let the AK 96 pass FCH (wait for green fluid path) before entering
Service mode.
Note

5. Enter the Service menu.


6. Select Calibrations/Transducers/UF/Verify.
7. Connect the dialysis fluid tubes to the UF calibration box.
8. Enter the Reference volume with the Display Up/Down keys (The
Reference volume is labelled on the UF Calibration unit)
9. Press the Select key to confirm.
10. After the Reference volume is confirmed the UF cell verification runs
automatically, until the it is finished.
7:40 AK 96 - Calibrations HCEN12212 Revision.05.2008
During the verification of the UF cell calibration: The time display shows the time since the
calibration was started.
The information displays also shows the different stages during the calibration:
• E2PROM test: The E2PROMS are tested incase there is a communication error with the UF
cell.
• Fluid preparation: The AK 96 awaits that the fluid is correct in temperature and conductivity
(37 °C and 14.4 mS on the B Cell)
• Low and High level fill up: During this period the reference volume is filled and emptied, this is
done to check that the tool is working properly.
• UF stabilization: During this period the fluid path is heated up so there is no temperature drift in
the components.
• Main flow offset calibration: During this period the offset on both channels is measured and
saved in RAM memory. The test is performed in the same way as during taration.
• Ch1 flow calibration: The valves are set so the flow will go through the Ch1 until the reference
volume is filled. Since the reference volume is known the flow can be calculated. This value is
used to calculate a new coefficient for the Ch1 so the measured volume match the actual volume.
The new coefficient for Ch1 is saved in RAM and used in the next test.
• Differential flow calibration: The new coefficient (Ch1) is used when the Ch2 coefficient is
calculated by running the same flow trough both channels. i.e. the same as differential flow in
taration.
• Verifying calibration of UF-cell: The verification is done by draining the reference volume
through the Ch2. Since it is a known volume the accuracy of the UF measuring system is
measured.
• Verifying calibration of Isolated UF: The reference volume is then filled through Ch2, the flow
is backwards through the cell, negative value on Ch2. When it is filled it is emptied just through
the Ch2 and the volume is measured by Ch2 only and the UF cell is checked for accuracy during
Isolated UF.
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:41
When the verification of the UF cell calibration is finished: When the verification is completed,
AK 96 displays the results:
First the display for UF Volume (for 3 seconds):
UF VOL REF VOL VOL DIFF 1006.5 1007.3 -0.28 %
Then the display for Isolated UF Volume (for 3 seconds):
ISO VOL REF VOL VOL DIFF 1007.5 1007.3 +0.28 %
Then the previous display is showed again for 3 seconds and so on.
When ready, press the Back key to exit the verification of the UF cell calibration.
Note
Perform a Rinse before leaving the AK 96 dialysis machine.
Note
Note
The equipment is turned off in the following order: 1. Press AK 96 ON/OFF button.
2. Turn off the AK 96 main switch.
3. Disconnect the UF calibration box from the power socket.
Note
7:42 AK 96 - Calibrations HCEN12212 Revision.05.2008
BPM test
Test method
• Attach BPM cuff, 110350, tight to the BPM test equipment, K22151001.
• Start the AK 96 dialysis machine with the main switch and the ON/OFF button.
• Enter the Service menu.
• Select Diagnose/BM/BPM.
BPM sub-tests
Following sub-tests are included:
• Air leakage test: measures that the equipment is airtight.
• Inflation speed test: measures the time to fill the cuff with air.
• Static pressure: Check of pressure transducer calibration
Each of the tests can be carried out separately but the above order for carrying out the sub-tests is
recommended.
Check that the results from the sub-tests not exceed following limits:
• Air leakage: maximum 18 mmHg/ 3 min
• Inflation speed: 2 -11 s
• Static pressure: Maximum deviation from reference to be within ± 3 mm Hg. 50 ± 3 mmHg 150
± 3 mmHg 250 ± 3 mmHg
Check of air leakage By entering this test mode the air leakage for the BPM module, cuff hose
and cuff will be tested automatically. Before entering the test mode cuff and cuff hose to be
connected to the AK 96. Cuff to be wrapped tight to the BPM test equipment.
BPM test equipment is used to simulate the patient arm circumference. The air leakage is only
tested for 90 sec but the value is recalculated for 3 minutes. If the displayed value is above 18
mm Hg / 3 minutes, this indicates that there is a possible air leakage in the system.
It is of course also possible to perform this test without the cuff and cuff hose connected but then
the test will only check for air leakage within the AK 96. In this case replace the cuff and cuff
hose by a tube, which is closed.
HCEN12212 Revision.05.2008 AK 96 - Calibrations 7:43
Check of inflation speed By entering this test mode the capacity for the pump within the module
will be tested automatically. Before entering the test mode cuff and cuff hose to be connected to
the AK 96 dialysis machine.
Cuff to be wrapped tight to the BPM test equipment. If the displayed time is above 11 seconds,
this test indicates that the pump is worn-out.
Check of pressure transducer calibration and measured pressure By entering this test mode it will
be possible to check the actual pressure value for the BPM pressure on the AK 96 display.
Before entering the test mode, pressure transducer calibration tube from kit K13983002 to be
connected to the BPM connector on the AK 96. No cuff and cuff hose to be connected.
Note
The pressure transducer calibration tube kit must be drained before being connected, to make sure that no fluids or
particles enters the BPM connector or tubing’s. This might otherwise damage or destroy the sensitive BPM sensors.
Note

When “Measured pressure” is displayed it is possible to start applying the external reference
pressure.
Note
Pressures shall be applied very carefully so that no over pressures occur.
Note

To check the actual pressure value, external pressure reference instrument to be connected and
external pressure of 250, 150, 50 mm Hg to be applied. Check deviation from reference pressure
instrument for displayed value.
7:44 AK 96 - Calibrations HCEN12212 Revision.05.2008

You might also like