Eoheeieen Ti Lead mode: five lead or three leads optional 2.Accuracy: + 1% or + Ibpm, take big 3. Resolution: 1 BPM (stroke / min) Pee 1 Methods of measurement: RA-LL impedance method 2. Range of measurement: adults: 0 to 120BrPM children / newborn: 0~150 BrPM In the range from 7BrPM to 120BrPM, the measurement error should be + 2BrPM. Other range errors are not defined 2. Resolution: 1% 3. Accuracy: within the range of 70% to 100%, the measurement error shall be + 4%; ‘The measurement error is not defined in the range of 096 to 699 Piece 1 Measuring range: 30=240bpm 2. Resolution: 1bpm 3. Accuracy: + 36pm Eootcroran) tion method 1. Measurement technique: oscil 2. Measurement parameters: systolic/ diastolic / average pressure 3, Measurement mode: manual, continuous, automatic 4. Continuous measurement mode measurement time: 5 minutes PMc once) 1, Channels: dual channel 2. Measuring range: 0-50 DEG C 3. Resolution: 0.1 DEG C 4, Accuracy: 0.1€ (excluding sensor error) 1, Measurement: infrared radiation absorption technology 2. Measuring range: 0~ 99mmHg 3. Accuracy 0 to 40mmHg:2mmHg 41 to 76mmHg:5% 77 to 99mmHe:10% 4, Resolution: lmmHg 1. Channels: dual channel 2. Pressure names: APT, PA, CVP, RAP, LAP, ICP, P1, P2 3, Range of measurement and alarm ranged from ART:0 to 300mmHg PA:-6 to 120mmHg, CVP/RAP/LAP/ICP:-10 to 40mmHg P1/P2:-10 to 300mmHg 4, Pressure sensor Sensitivity: 5uV/V/mmH impedance range: 300-3000, resolution: ImmHg ‘Accuracy: + 2% or + 1mmHg, take big update time: about 1 minutes Modular Patient Monitor 3