Mobile Viewer
Revision History
This manual has a revision number. This revision number changes whenever the manual is updated due to
software or technical specification change. Contents of this manual are subject to change without prior
notice.
© 2016 -2017 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved.
This manual is based on the maximum configuration. Therefore, some contents may not apply to your
device. If you have any question, please contact our Customer.
Intended Audience
This manual is for medical personnel, biomedical engineers, authorized technicians or service
representatives responsible for troubleshooting, repairing and maintaining the monitors
Illustrations
All illustrations in this manual serve as examples only. They may not necessarily reflect the setup or data
displayed on your patient monitor.
Conventions
Italic text is used in this manual to quote the referenced manuals, chapters, sections and formulas.
Bold text is used to indicate the screen texts and names of hard keys.
DANGER
Indicates an imminent hazard that, if not avoided, will result in death or serious injury.
WARNING
Indicates a potential hazard or unsafe practice that, if not avoided, could result in death or
serious injury.
CAUTION
Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor
personal injury or product/property damage.
NOTE
Provides application tips or other useful information to ensure that you get the most from your
product.
1.2 Dangers
There are no dangers that refer to the product in general. Specific “Danger” statements may be given in
the respective sections of this manual.
WARNING
The equipment should be used by the IT engineers of the hospital or user service engineers of
Mindray or under their instructions. The user of the equipment should receive sufficient
training. Anyone unauthorized or untrained must not perform any operation on it.
To avoid inadvertent disconnection, route all cables in a way to prevent a stumbling hazard.
Wrap and secure excess cabling to avoid risk of entanglement or strangulation by patients or
personnel.
When disposing of the package material, observe applicable local or hospital’s waste control
regulations and keep it out of children’s reach.
If the monitoring network contains wireless devices, the wireless devices should be provided
with CE identifier. Before connecting the equipment to the power line, check that the voltage
and frequency ratings of the power line are the same as those indicated on the equipment’s
label or in this manual.
The Mobile Server cannot directly support the monitoring equipment of other manufacturers.
1.4 Cautions
CAUTION
If the power supply of the hospital is unstable, provide a UPS to the Mobile Server. If the Mobile
Server system is powered off unexpectedly, the system may become abnormal and fail to
operate properly when it is turned on next time or even lead to a more severe consequence.
The Mobile Server system should be maintained every 3 to 6 months. Continuous operation for
a long term may cause the system to operate abnormally.
Always install or carry the equipment properly to avoid damage caused by drop, impact,
strong vibration or other mechanical force.
Before using the Mobile Viewer, you need to first deploy the Mobile Server. The Mobile Server provides
monitoring data to the Mobile Viewer and functions as the central server of Mobile Viewer.
WARNING
The Mobile Viewer is not intended for primary monitoring. There is no audible alarm sound in
the Mobile Viewer, and all the data is delayed. Always refer to your primary monitoring source
for current information, real-time patient data and combine this data with personal clinical
observation of the patient before administering interventions.
Item Requirement
The environment where the Mobile Server is installed should be reasonably free from noises, vibration,
dust, corrosive, flammable and explosive substances.
When the cabinet is positioned, a sufficient space in front and behind should be left for convenient
operation, maintenance and repair. Moreover, to maintain good ventilation, a gap of at least 2 inches (5cm)
should be kept around the equipment.
When the Mobile Server system is moved from one place to another, condensation may occur as a result of
temperature or humidity difference. In this case, never start the system before the condensation
disappears.
NOTE
It is recommended to install the Mobile Server system in the equipment room of the hospital or
medical institution.
Each component of the Mobile Server system must be powered by the specified power source.
To protect the hospital personnel from electric shock, the Mobile Server system (including the host and
displays) must be properly grounded. The Mobile Server system is provided with a 3-wire power cable,
which must be plugged into a properly grounded 3-wire receptacle. If a 3-wire, grounded receptacle is not
available, consult the hospital electrician.
WARNING
Make sure that the operating environment and power source of the Mobile Server system meet
the specific requirements; otherwise, unexpected consequences, e.g. damage to the
equipment, may arise.
Appropriate power supply must be selected according to the setup of the system power
voltage; otherwise, serious damage may be caused to the system.
Never use a 3-wire to 2-wire adapter for the equipment. It indicates a potential hazard or
unsafe practice that, if not avoided, could result in death or serious injury or property damage.
Item Requirement
2.5 Specification
2. Select the driver that needed to be installed, and select Install Driver.
3. After the installation, the driver status changes to “The driver has been installed successfully!”, as
shown in Figure 3-3.
2. Select the desired installation location for the server, and then select Install.
NOTE
Considering the Windows permission for user account control, we recommend installing the
server in a disk other than system disk.
NOTE
Figure 3-6 is for illustration only. Please set CMS according to actual conditions.
4. After the configuration is completed, the installation program will check whether the configuration is
correct.
NOTE
"Errors reported here must be corrected before the service can be started." is displayed in the
last line, indicating that, if an error is reported after this line, the error must be solved before
the service is started, as shown below. If only this line is displayed, the configuration is correct.
5. After the configuration is completed, you can select whether to start the service. If you select Yes, the
service will be started directly; if you select No, you can start the service manually in the Start menu
later.
NOTE
Restart the Mobile Server is needed after changing the settings.
3. Enter the IP address, select Add to add the IP address; select an IP address in the lower list, and select
Delete on the right to delete this IP address.
1. Select the IP address for listening by the server in IP Address. If All is selected, all the local IP
addresses will be listened to.
2. Enter the HTTP port number for listening by the server in Port. The default value is 80.
3. Enter the HTTPS port number for listening by the server in SSL Port. The default value is 443.
NOTE
The working directory of the current server is displayed in ServerRoot and cannot be changed.
1. Select the server language in Language. You can select English, simplified Chinese, etc.
2. Select the client user authentication mode in Authentication. If Local is selected, the user uses the
local account of computer to log in; if Domain is selected, Domain Name also needs to be
configured, and the user uses the domain account of the domain which the computer joined to log
in.
3. Select the ECG standard in ECG Standard. Two standards are available: AHA and IEC.
NOTE
The error of module log4cplus is reported because Apache fails to detect the module depth
correctly. This error does not affect use of the Mobile Server service.
Use the domain account of the domain that the computer joined.
2. Select Manage, or select Control Panel →System and Security →Administrative Tools →
Computer Management.
3. Select System Tools → Local Users and Groups → Users in menu tree in the left.
NOTE
The first time you start the service, please insert the USB dongle to the computer and import
the license from USB dongle.
If the server is not configured when the service is installed, it is required to configure the server
using Mobile Server Manager before the service is started. This management tool can be found
in Mobile Server of the Start menu. After configuration, use Test in the Start menu to check
whether the configuration is correct.
4. In the Recovery tab, set the operations for the first failure, second failure and subsequent failures of
setting all to restart the service.
5. Select OK.
If the USB dongle is not inserted, the page will prompt that Cannot find USB dongle! Please insert the
USB dongle into the computer installed Mobile Server and refresh the page.
3. Select Import License, if the page displays as shown below, and there exist some key files in
C:\License folder, the import is successful. Otherwise, please retry.
3.10 Uninstalling
1. Find the Mobile Server menu in the Start menu, and select Uninstall.
2. Select Next.
1. Search Mindray Mobile Viewer in Google Play Store, and select INSTALL.
2. Get the installation file of Mobile Viewer from service provider, and copy the file to the equipment,
find this file in the file manager of equipment, select it, and select Install.
NOTE
This application needs the access permission of storage space. Authorize it in permission
The second installation method requires enabling "Allow installation of apps from unknown
sources" in Android.
4.3 Settings
1. Select Setup on the login screen to enter the Setup screen.
NOTE
If you select the Log-Out button, this login will be canceled.
2. Enter the server domain name or IP address in IP address, e.g., www.mindray.com or 10.48.11.1.
3. Enter the number of port for communicating with the server in Port. The default value is 443.
4. If encrypted communication is not adopted, cancel selecting Use SSL. The default port number for
not using encrypted communication is 80.
NOTE
If encrypted communication is adopted and login failure is prompted, clear the data and cache
of this application from system settings or uninstall and reinstall this application, and try it
again.
Buffer Time indicates the cache time for viewing real-time waveform, including options such as
0s, 2s, 4s, and 6s. If 0s is selected, there is no cache, and the waveform is displayed immediately
after being obtained from the server; if 2s is selected, the waveform will be displayed after the
waveform of 2s is obtained from the server.
2. After setting, select Save. If the system prompts Monitor data setting updated, the settings have
been saved.
2. Select the desired language, and then select Save to switch the screen to the corresponding
language.
4.4 Login
Enter User name and Password on the login screen, select Login to perform the login operation. If the
verification succeeds, the system goes to the home screen My Patients.
NOTE
You will enter the demo mode if you use demo as User Name and demo as Password.
4.5 Log-Out
Select Log-Out on the setup screen to log out.
1. Select the icon on the toolbar of My Patients screen to enter the Patients Management screen.
All the patients are displayed by grouping of departments.
2. Select the check box on the right of a patient to follow/unfollow this patient. A maximum of 32
patients can be followed; select the position to the left of the check box of a patient, and view the
real-time monitoring parameter and waveform of this patient. For details, see Monitor Single
Patient.
3. Select the My Patients tab. This tab displays the list of currently followed patients. Slide leftward on a
patient to display . You can select to unfollow this patient.
Select to enter the Patients Management screen. Select to enter the Setup screen.
After selecting patients to follow, you can return to the My Patients home screen to view the parameter
and alarm information of followed patients.
In the area of each patient, the patient's personal information (the bed No., name, etc.) is displayed on the
upper left side, and the alarm message of this patient is on the right; the parameter area is at the lower
part. You can drag leftward and rightward in the parameter area to view more parameters. If a parameter
has a background color and blinks, this parameter is in the alarm status currently. If the parameter is in the
hollow style, this parameter is unreliable (As shown below, HR of the first patient is unreliable).
The latest alarm message is displayed in the upper right corner of each patient's area. Select the alarm
message to display the alarm message box and show more alarm messages. If there is “√” before a
message, this alarm has been acknowledged; if time is displayed on the right of the message, the time is
the time when the alarm is triggered.
The View Bed screen contains the alarm area, parameter area,and waveform area from top down.
Alarm area: It displays the latest alarm message. You can select to view more alarms and alarm details.
Parameter area: It displays the real-time parameters. You can drag leftward and rightward to view
more parameters. If a parameter has a background color and blinks, this parameter is in the alarm
status currently. If the parameter is in the hollow style, this parameter is unreliable.
Waveform area: It displays the real-time waveform. Select to freeze the waveform and change
the waveform background to gray. You can drag the waveform leftward to view earlier waveforms.
Select this button again to cancel the freezing operation.
Events of this patient can be reviewed on this screen. When Event Level is selected for the first time,
events will be sorted in the descending order of priority; if it is selected again, events will be sorted in the
ascending order of priority. When Event Time is selected for the first time, events will be sorted in the
descending order of time; if it is selected again, events will be sorted in the ascending order of time.
Select an event to view details of this event. It is the same as the operation of viewing alarm details.
Config the server IP address or domain name, port in Mobile Viewer Setup.
Input user name and password in login view of Mobile Viewer, select Login.
Considering the Windows permission for user account control, please uninstall the server and
reinstall it under a directory of a non-system disk.
Find the Mobile Server menu in the Start menu, select Test to check the server configuration. If
"Syntax OK" is prompted, the server configuration is correct.
1. If the configuration is correct, try to double-select to open bin\httpd.exe in the installation directory.
3. Access http://127.0.0.1:port/index.html in the browser (port is the port in Apache Settings of Mobile
Server). If "It Works!" is displayed, httpd.exe can run normally.
5. If the problem cannot be fixed yet, contact the service personnel for support.
To perform the above operation, right-click the service and select Run as administrator.
Information Is Displayed
Please try the following operations to check whether the USB dongle or license is available:
If the license is not imported, please check whether the USB dongle is inserted into the Mobile Server
computer.
If the license is imported, please check whether there are key files in C:\License.
Open the installatin folder of Mobile Server, open the file bin\FrontendLog.log, check whether the
USB dongle or license is detected.
Check whether the server IP address and port number in Mobile Viewer settings are configured
correctly.
If the network connection is normal, check whether the user name or password is matched.
If encrypted communication is adopted and login failure is prompted, clear the data and cache of this
application from system settings or uninstall and reinstall this application, and try it again.
If multiple users use the same one account, contact your administrator to check if the local account or
domain account of the server is locked. Accounts are locked when other users made several attempts
to log in using a wrong password, If this is the cause of the problem, we sugguest that the
administrator creates accounts for every user or change the account-lock security policy.
Follow patients. For how to follow patients, see 4.6.1 Following/Unfollowing Patients.
As initialization of the Mobile Server takes about one minute, please wait patiently.