Professional Documents
Culture Documents
■■ isinfection
D
The disinfection method recommended in the
product labeling for the OneTouch Verio Pro+
® ™
does not need user calibration coding, it produces were conducted under institutional review board/ethics
reliable readings in fewer steps. committee–approved protocols with the informed
The ergonomic design of the OneTouch Verio Pro+
® ™
consent of the participants. (The intermediate precision
Meter (Figure 1) fits comfortably in the palm of the and disinfection studies did not involve human subjects
hand and allows the user to easily and firmly hold the or samples from subjects.) System accuracy, clinical
meter. When the meter is turned on or a test strip is performance, repeatability, and intermediate precision
inserted, the LCD color display with backlight comes on evaluations were performed in a manner consistent with
automatically, enabling the user to easily read results the guidelines published by the International Organiza-
in low light conditions. The OneTouch Verio Pro+
® ™
tion for Standardization (ISO).¹ Linearity was assessed
Meter is equipped with safety features to control according to Clinical and Laboratory Standards Institute
infection transmitted through blood borne pathogens. (CLSI; formerly NCCLS) EP6-A.C Interference and anti-
These features include a push-button strip ejector coagulants were tested according to CLSI guidelines
that minimizes contact with used test strips and a EP7-A2D and EP14-A2,E respectively.
Table 1.
OneTouch Verio Pro+ BGMS specifications
® ™
Feature Specification
Figure 2.
OneTouch Verio Test Strip
® ™
2
System accuracy
The ISO criterion for acceptable system accuracy is that Table 2.
Ö 95% of BGMS results fall within specified error toler- OneTouch Verio Pro+ BGMS system accuracy (n = 600)
® ™
presented in this section assessed accuracy according A. Overall system accuracy according to ISO and
to two specifications for error tolerance: tighter specifications
• ISO system accuracy specification ISO specification
Individual glucose results are considered accurate
Within ± 15 mg/dL 598 of 600
if they fall within ± 15 mg/dL of the reference result
or ± 20% of reference 99.7%
at glucose concentrations < 75 mg/ dL or within ± 20%
at glucose concentrations Ö 75 mg/dL. Tighter specification
• Tighter system accuracy specification Within ± 12 mg/dL 591 of 600
Individual glucose results are considered accurate or ±15% of reference 98.5%
if they fall within ± 12 mg/dL of the reference result
at glucose concentrations < 80 mg/dL or within ± 15%
at glucose concentrations Ö 80 mg/dL. B. System accuracy according to ISO accuracy intervals
tests (duplicate tests with each of three test strip lots) Glucose < 80 mg/dL Glucose Ö 80 mg/dL
were performed from a single fingertip lancing. ± 10 ± 12
mg/dL mg/dL ± 10% ± 15%
• Comparison testing was performed before and
after OneTouch Verio Pro+ BGMS testing using the
® ™
127/132 128/132 429/468 463/468
YSI 2300 STAT Plus Biochemistry Analyzer (YSI, Inc., 96.2% 97.0% 91.7% 98.9%
™
Results
Bias plot of Verio Pro+ Meter results (n = 600)
™
3
Clinical performance
Methods Figure 4.
OneTouch Verio Pro+ BGMS clinical accuracy
® ™
Results
• Accuracy (Figure 4): Across all sample types, 99.8%
C. Arterial samples (n = 200 results)
(565/566) of OneTouch Verio Pro+ BGMS results
® ™
4
Table 3. Figure 5.
OneTouch Verio Pro+ BGMS clinical performance: OneTouch Verio Pro+ BGMS clinical performance:
® ™ ® ™
zone definitions
Zone A: No effect on clinical action
Zone B: Altered clinical action — little or no effect on clinical outcome
Zone C: Altered clinical action — likely to affect clinical outcome
Zone D: Altered clinical action — could have a significant medical risk
Zone E: Altered clinical action — could have dangerous consequences
5
Hematocrit performance
Methods Figure 6.
OneTouch Verio Pro+ BGMS hematocrit performance:
® ™
Results
tolerance boundaries.
• The largest bias observed for an individual
OneTouch Verio Pro+ BGMS result was –17.8%
® ™
6
Repeatability Intermediate precision
Methods Methods
• A venous blood sample (hematocrit 41.5%) was • The testing procedure involved three control solution
adjusted in the laboratory to five target glucose levels (low, mid, and high), three lots of OneTouch ®
levels ranging from 42 to 330 mg/dL. Verio Test Strips, and 30 OneTouch Verio Pro+ Meters
™ ® ™
• The testing procedure involved three lots of (10 meters per test strip lot).
OneTouch Verio Test Strips and 30 OneTouch
® ™ ®
• Testing was executed in the laboratory by multiple
Verio Pro+ Meters (10 meters per test strip lot).
™
operators over a 10-day period.
• Testing (10 replicates per meter at each glucose
• On each day of testing, duplicate tests were performed
level) was executed by a single operator on one day.
with each meter at each control solution level.
• For each glucose level, the pooled mean glucose,
pooled standard deviation (SD), and pooled • For each control solution level, the mean glucose (all
coefficient of variation (CV) were calculated. test strip lots, meters, and days), SD (pooled for the test
strip lots), and CV were calculated.
Results
Results
• All pooled CV values were ≤2.55% (Table 5),
indicating that from strip to strip, the OneTouch® • The pooled CV values range from 1.53% to 2.11%
Verio Pro+ BGMS provided repeatable results across
™ (Table 6), demonstrating that the OneTouch ®
a wide range of blood glucose levels. Verio Pro+ BGMS provides reproducible test results
™
Table 5.
Table 6.
OneTouch Verio Pro+ BGMS repeatability:
® ™
mean glucose SD CV
(mg/dL) (mg/dL) (%) mean glucose SD CV
(mg/dL) (mg/dL) (%)
43
42.9 1.09
1.09 2.55
2.55 38 0.80 2.11
97.4
97 1.83
1.83 1.88
1.88
128.6 2.44 1.90 117 1.79 1.53
129
193.7 2.44
3.91 1.90
2.02 339 5.94 1.75
343.2 7.48 2.18
194 3.90 2.01
343 7.48 2.18
7
Linearity Effect of common anticoagulants
Methods Methods
• Venous blood samples collected from each of nine • A laboratory study was performed to evaluate
donors (hematocrit range 37.1% to 43.3%) were the potential matrix effects of four common
adjusted in the laboratory to eight glucose levels anticoagulants (lithium heparin, sodium heparin,
ranging from approximately 20 to 700 mg/dL. potassium EDTA, and sodium citrate) on OneTouch ®
with weights proportional to the inverse of the YSI Meters and three OneTouch Verio Test Strip lots.
® ™
• Weighted linear regression analysis: slope = 0.99; evaluated for accuracy according to the ISO
y-intercept = 2.2 mg/dL; r² = 0.997 (n = 3,456). specification described above.
Results
• At least 99.6% accuracy was achieved with each
anticoagulant across the entire glucose range tested
(Table 7).
Table 7.
OneTouch Verio Pro+ BGMS accuracy (per ISO
® ™
8
Interference
Methods Table 8.
OneTouch Verio Pro+ BGMS: Maximum allowed
® ™
9
Disinfection Conclusions
The accuracy, hematocrit performance, precision, and
Methods linearity of the OneTouch Verio Pro+ Blood Glucose
® ™
• The U.S. Food and Drug Administration requires Monitoring System have been demonstrated in labora-
that the efficacy of disinfectants used to clean and tory and clinical studies. In both the system accuracy
disinfect the exterior surfaces of blood glucose and the clinical performance studies, accuracy levels
monitoring systems be validated against hepatitis B exceeded the widely recognized ISO criteria. The per-
virus utilizing the actual materials that make up the formance at low glucose concentrations was consistent
components of the system.H with the high level of performance across all glucose
concentrations. The OneTouch Verio Pro+ BGMS
® ™
the bleach solution was tested against surrogate all meters using OneTouch Verio Test Strips would
® ™
hepatitis B virus on three lots of the three materials show similar levels of performance.
that make up the contactable surfaces of the meter.
The materials were wiped with bleach solution; after
a one-minute exposure time, the materials were
tested for viral titer (along with positive and negative
controls).
• A separate procedure evaluated the durability
of OneTouch Verio Pro+ Meters subjected to the
® ™
disinfection method.
Results
• The disinfection method recommended in the
OneTouch Verio Pro+ BGMS product labeling was
® ™
Meter.
• Functionality and appearance of the OneTouch ®
tion method.
* OneTouch Verio System Meters include OneTouch Verio , OneTouch Verio Pro, OneTouch Verio IQ, and OneTouch Verio Pro+.
® ™ ® ™ ® ™ ® ™ ® ™
10
References
1. International Organization for Standardization. In vitro
diagnostic test systems—requirements for blood-glucose
monitoring systems for self-testing in managing diabetes
mellitus. ISO International Standard 15197:2003.
2. NCCLS. Evaluation of the Linearity of Quantitative
Measurement Procedures: A Statistical Approach;
Approved Guideline. NCCLS document EP6-A (ISBN
1-56238-498-8). NCCLS, 940 West Valley Road, Suite
1400, Wayne, Pennsylvania 19087-1898, USA, 2003.
3. Clinical and Laboratory Standards Institute. Interference
Testing in Clinical Chemistry; Approved Guideline—Second
Edition. Clinical and Laboratory Standards Institute
document EP7-A2 (ISBN 1-56238-584-4). Clinical and
Laboratory Standards Institute, 940 West Valley Road,
Suite 1400, Wayne, Pennsylvania 19087-1898, USA,
2005.
4. Clinical and Laboratory Standards Institute. Evaluation
of Matrix Effects; Approved Guideline—Second Edition.
Clinical and Laboratory Standards Institute document
EP14-A2 (ISBN 1-56238-561-5). Clinical and Laboratory
Standards Institute, 940 West Valley Road, Suite 1400,
Wayne, Pennsylvania 19087-1898, USA, 2005.
5. Parkes JL, Slatin SL, Pardo S, Ginsberg BH. A new
consensus error grid to evaluate the clinical significance
of inaccuracies in the measurement of blood glucose.
Diabetes Care. 2000;23(8):1143– 1148.
6. Jacobs, DS, et al. d-Xylose absorption test. In Laboratory
Test Handbook, Lexi-Comp Inc., Hudson, Ohio, USA, 1996,
pp. 369–371.
7. U.S. Food and Drug Administration Letter Dated
September 17, 2010. Letter to Manufacturers of Blood
Glucose Monitoring Systems Listed with the FDA. http://www.
fda.gov/MedicalDevices/ProductsandMedicalProcedures/
InVitroDiagnostics/ucm227935.htm.
11
www.LifeScan.com
© 2012 LifeScan, Inc. Milpitas, California, 95035 6/12 AW 3088095