Professional Documents
Culture Documents
™
REF 220-0318
D-10
Hemoglobin
Testing
System
L20015501
October 2004
D-10 Hemoglobin Testing System
Bio-Rad Laboratories
Diagnostic Group
4000 Alfred Nobel Drive
Hercules, California
94547
USA
CAUTION
Please read through and familiarize yourself with the contents of the service
manual.
All personnel using the D-10 should wear standard laboratory safety apparatus
(e.g. safety glasses, laboratory coat and gloves).
All reference material (i.e. standards and controls) and patient samples should be
considered as biohazardous material and must be handled with caution.
Waste material contains sample waste plus reagents: treat and dispose of as
directed by your laboratory safety guidelines or by local ordinance.
• Principles of Operation
• Technical Specifications
• Installation and De-installation
• Service Software
• Maintenance Procedures
• Parts Replacement Procedures
• Troubleshooting Guide
Also included are electrical and mechanical drawings along with the spare parts
list.
Table of Contents
1 Introduction
1.1 General Description 1-1
1.2 High Performance Liquid Chromatography 1-1
1.2.1 Definition 1-1
1.2.2 General Principle 1-1
1.3 HPLC Applied to the D-10 1-2
1.3.1 General Principle 1-2
1.3.2 Stationary Phase: the Cartridge 1-3
1.3.3 Mobile Phase: the Buffers 1-3
1.3.4 Separation 1-3
1.3.5 Bio-Rad D-10 Hemoglobin A1c 1-4
1.3.6 Bio-Rad D-10 Dual Program 1-4
1.4 Operating Principles 1-5
1.4.1 Theory of Operation 1-5
1.4.1.1 Overview 1-5
1.4.1.2 Sample Preparation 1-5
1.4.1.2.1 Functions 1-5
1.4.1.2.2 Description of the Sample Fluid Pathway 1-6
1.4.1.2.3 Operation of the Sampling Components 1-7
1.4.1.3 Sample Analysis 1-8
1.4.1.3.1 Functions 1-8
1.4.1.3.2 Description of the Sample Analysis Components 1-8
1.4.1.3.3 Operation of the Sample Analysis Components 1-9
1.4.1.4 General Fluidic Diagram 1-14
1.4.1.5 Sample Arm 1-14
1.4.1.5.1 Functions 1-14
1.4.1.5.2 Components 1-16
1.4.1.5.3 Operation of the Sample Arm Components 1-17
1.4.1.6 Rack Handling Functions 1-19
1.4.1.6.1 Functions 1-19
1.4.1.6.2 Description of the Rack Handling Components 1-19
1.4.1.6.3 Operation of the Rack Handling System 1-20
1.4.1.7 User Interface 1-21
1.4.1.8 Device Control/Processor Architecture 1-21
1.4.1.9 Power Supplies 1-22
1.4.2 Physical Organization of the Device:
Mechanical, Electrical and Flow Architecture 1-23
1.4.2.1 Modular Architecture 1-23
1.4.2.2 List of Modules 1-23
1.4.2.3 Electrical Architecture 1-24
1.5 Safety Symbols 1-24
2 Specifications
2.1 Physical Specifications 2-1
2.1.1 Dimensions 2-1
2.1.2 Weight 2-1
2.2 Technical Specifications 2-1
2.2.1 General Specifications 2-1
2.2.2 Throughput 2-1
2.2.3 Buffer Bottle Volumes 2-1
2.2.4 Waste Bottle Volume 2-1
2.2.5 Sample Rack 2-2
2.2.6 Sample Preparation 2-2
2.2.7 Probe Movements 2-3
2.2.8 High Pressure Pump/Proportioning Valve 2-3
2.2.9 Pressure Sensor 2-4
2.2.10 Injection Valve 2-4
2.2.11 Degasser Module 2-4
2.2.12 Cartridge Module 2-5
2.2.13 Detector 2-5
2.2.14 Tubing and Material in Contact with Liquid 2-6
2.2.15 Rack Door 2-6
2.2.16 Front Cover 2-6
2.2.17 Onboard PC 2-7
2.2.18 User Interface 2-7
2.2.19 Printer 2-7
2.2.20 Barcode Reader 2-7
2.2.21 Micro-controller Automation 2-7
3 System Description
3.1 System Overview 3-1
3.2 Frame 3-2
3.3 Modules 3-2
3.3.1 Syringe Module 3-2
3.3.2 High Pressure Module 3-5
3.3.3 High Pressure Pump 3-8
3.3.4 Cartridge Module 3-9
3.3.5 Detector Module 3-11
3.3.5.1 Description 3-11
3.3.5.2 Detector Module Evolution 3-13
3.3.5.2.1 Identification 3-13
3.3.5.2.2 Reference Photodiode 3-13
3.3.5.2.3 Detector Board Evolution 3-14
3.3.5.2.4 Jumper Configuration 3-15
3.3.5.2.5 Part Replacement 3-16
4 Installation Procedure
4.1 Installation Requirements 4-1
4.2 Installation 4-1
4.2.1 Unpacking and Inspection 4-1
4.2.2 Set Up 4-1
4.2.3 Barcode Labels 4-1
4.2.4 Performing Test Runs 4-1
4.3 De-installation 4-2
4.3.1 Decontamination 4-2
4.3.1.1 Clean/Decontaminate Sampling Fluid Path 4-2
4.3.1.2 Clean the Dilution Well 4-2
4.3.1.3 Clean the Sample Probe 4-2
4.3.1.4 Clean the Internal Waste Bottle 4-2
4.3.1.5 Repeat Clean/Decontaminate Sampling Fluid Path 4-2
4.3.2 Storage Fluid in the Instrument 4-2
4.3.3 Sample Arm Locking and System Shutdown 4-3
4.3.4 Packing 4-3
4.3.4.1 Removal of the External Components 4-3
4.3.4.2 Packing Accessories 4-3
4.3.4.3 Packing the Waste Tank 4-4
4.3.4.4 Packing the D-10 4-4
5 Software Overview
5.1 D-10 Application 5-1
5.2 User Software 5-1
5.3 *.ini Files 5-1
5.3.1 'Dia.ini' File 5-2
5.3.2 'instrument.ini' File 5-2
5.3.2.1 High-Pressure Pump Parameters 5-2
5.3.2.1.1 Flow Rate Calibration Parameters 5-2
5.3.2.1.2 High-Pressure Sensor Calibration Parameters 5-3
6 Replacement Procedures
6.1 List of Procedures 6-1
6.2 Front Cover Assembly 6-5
6.3 Thermal Printer 6-6
6.4 Buffer Bottle Support 6-8
6.5 Syringe Module 6-10
6.6 Electronic Module 6-13
6.7 Cartridge Module 6-17
6.8 Sample Probe 6-21
6.9 Dilution/Wash Station 6-21
6.10 Syringe 3-Way Valve 6-22
6.11 1mL Motorized Syringe 6-26
6.12 Thermal Fuse 6-28
6.13 Degasser Module 6-29
6.14 Vacuum Pump 6-33
6.15 Interface Board 6-35
6.16 Rack Door Module 6-37
6.17 Proportioning Valve 6-39
6.18 High Pressure Pump 6-42
6.19 7-Port, 3-Position HP Valve + 23µL Sample Loop 6-45
6.20 Pressure Sensor 6-49
6.21 Column Heater Coil 6-53
6.22 Z Position Sensor 6-56
6.23 CCFL Backlight 6-60
6.24 Power Supplies 1 & 2 6-62
6.25 Motherboard (PCM-5896) 6-65
6.26 LCD Interface Board 6-71
6.27 LCD Display 6-73
6.28 Waste Manifold 1 6-74
6.29 Waste Manifold 2 6-76
6.30 Hard Disk 6-78
7 Adjustments
7.1 List of Procedures 7-1
7.2 Detector Adjustment 7-2
7.2.1 Preparation 7-2
7.2.2 Adjustment Trimmers and Test Point Locations 7-2
7.2.3 Procedure 7-3
7.3 Automatic WC/FC Determination 7-5
7.3.1 Preparation 7-8
7.3.2 Automatic Tests 7-11
7.3.3 Returning the Instrument to Operating Condition 7-13
7.4 WC/FC Verification 7-14
7.4.1 Preparation 7-15
7.4.2 Procedures 7-16
7.4.2.1 Static Absorption Measurement (18 minutes) 7-16
7.4.2.2 Dilamex+ Gradient Test (~14 minutes) 7-17
7.4.2.3 Step Gradient Test (36 minutes) 7-18
7.4.2.4 Continuous Gradient Test (~12 minutes) 7-20
7.4.2.5 Entering the WC and FC Values 7-21
7.4.3 Automatic Sequence 7-21
7.4.4 Returning the Instrument to Operating Condition 7-22
7.5 Pressure Sensor Verification 7-22
7.5.1 Preparation 7-22
7.5.2 Procedure 7-23
7.6 Pressure Sensor Calibration (Field Procedure) 7-24
7.6.1 Preparation 7-24
7.6.2 Procedure 7-25
7.7 Flow Rate Verification 7-27
7.7.1 Preparation 7-27
7.7.2 Volumetric Method 7-28
7.7.3 Gravimetric Method 7-30
7.8 Flow Rate Calibration 7-31
7.8.1 Preparation 7-32
7.8.2 Volumetric Method 7-32
7.8.3 Gravimetric Method 7-35
7.9 Thermal Module Check 7-38
7.9.1 Preparation 7-38
7.9.2 Procedure 7-38
7.10 Thermal Module Calibration 7-39
7.10.1 Preparation 7-40
7.10.2 Interface Board Gain and Offset Measurements 7-40
7.10.3 Sensor Offset Correction Measurement 7-42
8 Troubleshooting
8.1 Error Messages 8-1
8.2 Troubleshooting Guide from Operation Manual 8-3
8.3 Diagnostics and Checks 8-7
8.4 Sensor Check 8-17
8.5 Warning Messages 8-18
9 Maintenance
9.1 Tools/Materials 9-1
9.2 Maintenance Procedures 9-2
9.2.1 Safety 9-2
9.2.2 Preparation 9-2
9.2.3 Dilution/Wash Station 9-3
9.2.4 Tube Holder Cleaning 9-3
9.2.5 Vertical Lead Screw Lubrication (Z and TH movements) 9-5
9.2.6 Flow Cell Cleaning 9-6
9.2.7 Inspection and Cleaning 9-7
9.2.7.1 HPLC Module 9-7
9.2.7.2 Cartridge Module 9-7
9.2.7.3 Syringe Module 9-7
9.2.7.4 Wiring 9-8
9.2.7.5 Barcode Reader 9-8
Appendix B: Diagrams
D-10: Sample Preparation Circuit (HP valve in load position) B-1
D-10: Sample Preparation Circuit (HP valve in inject position) B-2
D-10: HP Circuit (HP valve in load position) B-3
D-10: HP Circuit (HP valve in inject position) B-4
D-10: General Fluidic Diagram B-5
D-10: Electrical Block Diagram, Sheet 1 B-6
D-10: Electrical Block Diagram, Sheet 2 B-7
List of Figures
Number Title Page
Section 1
Section 3
Figure 3-1: Exploded View of the Frame and All Modules 3-1
Figure 3-2: Frame, Drawing 1 3-2
Figure 3-3: Frame, Drawing 2 3-2
Figure 3-4: Syringe Module, Front 3-2
Figure 3-5: Syringe Module, Rear 3-2
Figure 3-6: Syringe Module, Location and Fastening 3-3
Figure 3-7: Syringe Module, Fluidic Connections 3-4
Figure 3-8: Syringe Module Board, Electrical Connections 3-4
Figure 3-9: High Pressure Module 3-5
Figure 3-10: High Pressure Module, Location and Fastening 3-6
Figure 3-11: High Pressure Module, Fluidic Connections 3-6
Figure 3-12: High Pressure Module Board, Electrical Connections 3-7
Figure 3-13: High Pressure Pump, Fluidic Connections 3-8
Figure 3-14: High Pressure Pump, Electrical Parts 3-8
Figure 3-15: Cartridge Module, Internal View 3-9
Figure 3-16: Cartridge Holder 3-9
Figure 3-17: Cartridge Module, Fluidic Connections 3-10
Figure 3-18: Cartridge Module, Electrical Parts 3-10
Figure 3-19: Detector Module, Internal View 1 3-11
Figure 3-20: Detector Module, Internal View 2 3-11
Figure 3-21: Detector Module, Fluidic Connections 3-12
Figure 3-22: Detector Board Connections, Adjustments and Test Points 3-12
Figure 3-23: Detector Module Label 3-13
Section 5
Figure 5-1A: Numeric Keyboard 5-9
Figure 5-1B: Alphanumeric Keyboard 5-9
Figure 5-2A: File Export 5-10
Figure 5-2B: File Import 5-10
Figure 5-3A: File Selection, Save 5-10
Figure 5-3B: File Selection, Open 5-10
Figure 5-4: Example of Sensed Information 5-11
Figure 5-5: Graphic Screen 5-11
Figure 5-6: Password Screen 5-12
Figure 5-7: Main Screen 1 5-13
Figure 5-8: Main Screen 2 5-13
Figure 5-9: D-10 Syringe module (1/6) 5-14
Figure 5-10: D-10 Syringe module (2/6) 5-15
Figure 5-11: D-10 Syringe module (3/6) 5-16
Figure 5-12: D-10 Syringe module (3A/6) 5-16
Figure 5-13: D-10 Syringe module (4/6) 5-17
Figure 5-14: D-10 Syringe module (5/6) 5-17
Figure 5-15: D-10 Syringe module (6/6) 5-19
Figure 5-16: D-10 Degasser module (1/4) 5-20
Figure 5-17: D-10 Degasser module (2/4) 5-21
Figure 5-18: D-10 Degasser module (3/4) 5-21
Figure 5-19: D-10 Degasser module (4/4) 5-22
Figure 5-20: D-10 Detector module (1/4) 5-23
Figure 5-21: D-10 Detector module (2/4) 5-24
Figure 5-22: D-10 Detector module (3/4) 5-25
Figure 5-23: D-10 Detector module (4/4) 5-26
Figure 5-24: D-10 Rack Door module (1/2) 5-27
Figure 5-25: D-10 Rack Door module (2/2) 5-28
Figure 5-26: D-10 Rack Introduction module (1/1) 5-29
Figure 5-27: D-10 Front Cover module (1/2) 5-30
Figure 5-28: D-10 Front Cover module (2/2) 5-30
Figure 5-29: D-10 All Status (1/2) 5-31
Figure 5-30: D-10 All Status (2/2) 5-33
Figure 5-31: D-10 HP module: pump, valve (1/17) 5-34
Figure 5-32: D-10 HP module: pump, valve (2/17) 5-35
Figure 5-33: D-10 HP module: pump, valve (3/17) 5-36
Figure 5-34: D-10 HP module: pump, valve (4/17) 5-37
Figure 5-35: D-10 HP module: pump, valve (5/17) 5-37
Figure 5-36: D-10 HP module: pump, valve (6/17) 5-38
Section 7
Figure 7-1: Detector Board: Adjustment Trimmers and Test Point Locations 7-2
Figure 7-2: WC-FC auto determination kit tubing 7-7
Figure 7-3: WC-FC auto determination kit tubing 7-15
Figure 7-4: HPLC Module Board, Electrical Connections 7-40
Section 9
Figure 9-1: D-10 Sample arm module: X (1/4) 9-4
Figure 9-2: D-10 Sample arm module: TH (2/4) 9-4
Figure 9-3: D-10 Sample arm module 9-4
Figure 9-4: D-10 Sample arm module: Z (1/4) 9-5
Figure 9-5: D-10 Sample arm module: Z (2/4) 9-5
Figure 9-6: Flex Cables on Z Board 9-8
Figure 9-7: Flex Cables on X Board 9-8
Figure 9-8: Liquid Pump 9-9
Figure 9-9: Liquid Pump Components 9-10
Figure 9-10 Correct Pump Head Orientation 9-10
Figure 9-11 Pinch Valve Tubing Connections 9-11
1 INTRODU CTION
The D-10 Hemoglobin Testing System uses the principles of high-performance liquid
chromatography (HPLC). Whole blood samples undergo an automatic two-step dilution
process and are then introduced to the analytical (high-pressure) flow path. Pre-diluted
samples are automatically detected, based upon use of a sample vial adapter in the
sample rack, and the dilution step is omitted. Between sample injections, the sample
probe is rinsed with wash solution to minimize sample carryover.
A single, dual-piston, low-pulsation, HPLC pump with a proportioning valve delivers the
buffer solution to the analytical cartridge and detector. A programmed buffer gradient
of increasing ionic strength delivers the sample to the analytical cartridge, where the
hemoglobins are separated, based upon their ionic interactions with the cartridge
material.
The separated hemoglobins then pass through the filter photometer flow cell where
changes in the absorption are measured at 415nm.
Chemical separation can be accomplished using HPLC. This method uses the fact
that certain compounds have different migration rates, based on the stationary and
mobile phase in use. The extent or degree of separation is mainly determined by the
choice of stationary phase and mobile phase.
The stationary phase refers to the solid support contained within the cartridge over
which the mobile phase continuously flows.
The mobile phase refers to the buffer that is continuously applied to the cartridge,
acting as a carrier for the sample solution.
As the sample solution flows, with the mobile phase, through the cartridge, the
components of the solution will migrate according to the non-covalent interactions of
the compounds with the stationary phase. The chemical interactions of the stationary
phase, and the sample, with the mobile phase, determine the degree of migration
and separation of the components contained in the sample. For example, samples,
which have stronger interactions with the stationary phase than with the mobile
phase, will elute from the cartridge less quickly, and have a longer retention time.
Samples having weaker interactions with the stationary phase will have a shorter
retention time.
ION-EXCHANGE CHROMATOGRAPHY
Mobile Phase
Hemoglobin
In a gradient elution, different compounds are eluted by the increasing strength of the
salt solution. The sample is injected while a weaker mobile phase is being applied to
the system. The strength of the mobile phase is later increased, in increments, by
raising the salt content, which results in the elution of retained components.
The separation cartridge is a straight, calibrated tube made of stainless steel. It has a
length of 40.7mm and an external diameter of 6.35mm. The stationary phase is
maintained between two porous discs located at each end, in a portion of the
cartridge that has a length of 30mm and an internal diameter of 4mm. The dead
volumes are minimal.
The cartridge is inserted into a cartridge module, which has two Peltier elements
attached to it. This allows thermal regulation of both the cartridge and the mobile
phase.
The molecules of the stationary phase are methacrylate polymers which completely
fill the cartridge. They are non-porous spherical particles with diameters ranging
between 4 and 6µm (see Figure 1-2).
The two buffers used in the D-10 consist of a combination of phosphoric acid
(H3PO4) and three other chemical compounds ('BIS-TRIS', NaH2PO4 and Na2HPO4).
1.3.4 Separation
The gradient between the two buffers is optimized to modify the pH of the mobile
phase. This modification of the pH, and the differences between the ionization
potential of the different hemoglobins, results in the desired separation.
The HbA1c test is used to separate and quantitate the A0 and A1c hemoglobins.
Analysis time is 3 minutes. Figure 1-3 shows the elution gradient for HbA1c.
100
90
80
70
% buffer 2
60
50
40
30
20
10
0
0 20 40 60 80 100 120 140 160 180 200
time (s)
In addition to the A0 and A1c hemoglobins, this kit allows the separation and
quantification of hemoglobins A2, F, S, D and C. Analysis time is 6.5 minutes. Figure
1-4 shows the elution gradient for the Dual Program.
100
90
80
70
% buffer 2
60
50
40
30
20
10
0
0 50 100 150 200 250 300 350 400
time (s)
1.4.1.1 Overview
From a functional viewpoint, the D-10 system is made up of two sub-systems. One
is responsible for sample preparation, while the other performs high-pressure liquid
chromatography. The two systems – sample preparation and chromatography –
have the sample loop in common.
There are two types of samples that can be processed on the D-10:
• whole blood tubes
• prediluted samples in sample vials
Whole blood tubes are placed directly into the sample rack. Prediluted samples are
first placed into a specific rack adapter before being placed into the sample rack.
The rack has 10 sample positions. Once introduced into the system, the samples
are automatically processed.
The entire system is organized into integrated modules that are built into a frame
with an external skin. The back of the system contains external ports for connection
to the reagent bottles. The front of the device includes a door that is normally
locked. This door, along with a door on the right side, provides access for
maintenance operations. The bottom door, at the front of the system, allows access
to the high-pressure chromatography components.
1.4.1.2.1 Functions
All the components that are used to perform these functions are represented in
the "D-10: Sample Preparation Circuit (HP valve in load position)" diagram
located in Appendix B.
Note: the diagram does not represent the movements made by the probe.
Refer to the "D-10: Sample Preparation Circuit (HP valve in load position)"
diagram located in Appendix B.
The dilution/wash station is where whole blood samples are diluted and the
sample probe is washed. The station consists of a well used for the two-step
dilution of samples and a well for washing the probe between samples.
Refer to the "D-10: Sample Preparation Circuit (HP valve in load position)"
diagram located in Appendix B.
The syringe assembly consists of a calibrated syringe with the plunger connected
to the flange nut for the stepper motor lead screw. A slotted optical sensor is used
to signal that the syringe is in the position where the filling volume is 0 (home
position). When the probe is placed into a sample tube, the sample is aspirated
by the syringe through the sample loop of the 7-port, 3-position valve. The sample
is then diluted in the dilution well. The outlet of the well is closed by means of the
pinch valve associated with the well.
The two-step dilution is done in several stages. Refer to Section 2.2.6 for sample
and diluent volumes. After the first dilution, a portion of the dilution is aspirated
out of the well by the probe. The dilution well is drained by suctioning the liquid,
using the liquid pump, while the pinch valve associated with the dilution well is
open. The other pinch valve is closed during this time. The pinch valve closes and
the sample is diluted once again in the well. The diluted sample is aspirated,
using the syringe, into the probe and then through the sample loop.
The last step in sample processing consists of washing the sample probe, in
order to avoid carryover, before moving on to the next sample. The washing
sequence is as follows:
1. The dilution well is emptied, by suctioning from the liquid pump, through
the pinch valve associated with the dilution well.
2. The dilution well is flushed by dispensing wash solution into the well and
then draining it off.
3. The inside and outside of the probe are washed in the probe wash well by
dispensing the wash solution and then suctioning it off (using the liquid
pump) through the pinch valve associated with the probe wash well.
Note: during evacuation of liquid, one pinch valve is open at a time; the other
remains closed.
Note for Prediluted Samples: sample vials are used for prediluted samples. In this
case, the sample is aspirated directly from the sample vial to fill the sample loop.
Only the probe is washed. Sample vials are identified based on height and the
presence of a magnet on the adapter.
1.4.1.3.1 Functions
All the components used for these functions are represented on the "D-10: High
Pressure Circuit (HP valve in load position)” diagram located in Appendix B.
Refer to the "D-10: High Pressure Circuit (HP valve in load position)” diagram
located in Appendix B.
The degasser is a sealed chamber through which three flow circuits pass:
• circuit associated with reagent 1
• circuit associated with reagent 2
• circuit associated with the wash solution
At the end of each circuit is an optical sensor, designed to detect bubbles at the
output of the circuit.
The proportioning valve is a 3-way valve with the inlets connected to reagents 1
and 2 and the outlet connected to the inlet of the high-pressure pump. The
normally open inlet is connected to reagent 1.
The high-pressure pump has pistons that are driven by a camshaft/reducer set in
motion by a DC motor. This pump is also used in the Bio-Rad DuoFlow system.
An optical encoder is fitted on the motor shaft and is used to regulate the speed.
An optical sensor and a slotted disk interrupter provide a rotation reference at the
shaft output (cam sensor). The flow section of the pump includes the following:
• an inlet connector divided in two, one to each piston. A check valve is fitted
on each piston inlet.
• an outlet connector that joins the outlets of each piston. A check valve is
fitted on each piston outlet as well.
• an inlet and outlet connector for filling the piston backwash circuit.
The high-pressure circuit goes through the motorized 7-port, 3-position high-
pressure valve. The Inject position is used to place the sample loop into the high-
pressure circuit. The valve is motorized by a stepper motor, associated with
optical sensors, which ensure that the valve is positioned correctly.
The next component of the high-pressure circuit is the analytical cartridge, which
is at the core of hemoglobin separation. A coil of stainless steel tubing provides
the connection to the cartridge. The cartridge and tubing coil are fitted in a
cartridge module heated to 35°C by Peltier effect modules. The temperature is
measured by a Pt100 ohm temperature sensor. A thermal fuse keeps the system
protected against overheating.
The cartridge outlet circuit goes through the detector. The circuit includes:
• a high-pressure flow cell
• a light source consisting of a high-luminosity LED emitting in the blue range
• a focusing lens
• photoelectric sensors for measuring the incident light, along with the light
absorbed through an interference filter centered on 415nm
• a printed circuit board for processing the signals from the detector
At the detector outlet, the liquid flows to the external waste bottle through a
backpressure coil.
Refer to the "D-10: High Pressure Circuit (HP valve in load position)" diagram
located in Appendix B.
The vacuum created in the degasser extracts a portion of the dissolved gas in the
reagents as they pass through.
The proportioning valve selects the flow of reagent 1 (normally open position) or
reagent 2 when the valve is active. The valve is synchronized with the high-
pressure pump. As the pump piston pulls back to allow reagent to fill the pump
head, the valve proportions reagents 1 and 2 for optimal hemoglobin separation.
The synchronization is achieved using signals supplied by the cam sensor and
the motor encoder. These signals are associated with two pump parameters:
• delay between the cam signal and the start of the suction phase, called
"Waiting Count" (WC, parameter ECAMFILLSHIFT1)
• duration of the suction phase, called "Filling Count" (FC, parameter
EFILLDURATION)
The high-pressure pump ensures that liquid flows through the cartridge-detector
chromatography circuit. Data, supplied by the motor encoder, the cam signal and
the signal sent by the high-pressure sensor, drives the pump to achieve a
constant flow. The front of the high-pressure pump has an inlet purge valve in
parallel with the pump inlet. The port connection has two functions:
• manually primes the inlet circuits, either for reagent 1 (proportioning valve
not active) or for reagent 2 (proportioning valve active)
• manually primes the pump by injecting liquid through the pump pistons to
remove any air bubbles that may be present in the pump
The pressure sensor is a strain gauge sensor, which is used to monitor the
pressure in the circuit. Pump flow is raised slightly based on the pressure signal
from the sensor.
The cartridge module is an aluminum unit that is heated to 35°C with two Peltier
effect modules fitted in series. A thermal fuse, also in series with the Peltier effect
modules, is used to open the Peltier effect module power circuit if there is any
overheating. A Pt100 ohm temperature sensor, sunk into the aluminum unit
between the stainless steel tubing coil and the cartridge, measures the
temperature. The length of the coil is sufficient to raise the temperature of the
reagents close to the cartridge temperature.
The cartridge module has a front door to access the high-pressure circuit and
easily remove or install the cartridge. An arrow on the cartridge holder indicates
how the cartridge is to be installed. Arrows on both the holder and the cartridge
face the direction of flow. The cartridge holder fits securely in the cartridge
module when the door is closed and maintains the fluid pathway. The outlet of the
cartridge module is connected directly to the inlet of the detector flow cell.
Silicon photodiode
Plano-convex lens
The optical bench, containing the detector and the flow cell, has a light-tight cover
to avoid interference from external light. The flow cell has two integrated PEEK
tubes directly connected to the fluid pathway for the liquid inlet (blue tube) and
outlet (orange tube). The inlet tube is connected to the cartridge module and
enters the smaller side of the conical shaped flow cell. The outlet tube, from the
larger side of the conical flow cell, is connected to the external waste via a
backpressure coil. This coil maintains the backpressure in the flow cell to
suppress bubble formation.
The light source is a high-luminosity LED emitting in the blue range. The incident
light falls on the flow cell from the smaller side. This optical inlet side of the cell
has a plano-convex lens as the only geometric optical component. It is positioned
in the axis of the flow cell.
On the side of the flow cell optical outlet, there is a 415nm interference filter and a
silicon photodiode for measuring the sample absorbance. The reference
measurement is taken by a silicon photodiode placed close to the optical axis
between the source and the flow cell.
The sensors and the light source are connected to a printed circuit board placed
outside the optical bench. A removable metal cover protects the detector PCB.
For further details on the detector board, refer to diagram D-10 Detector Board
Diagram, Figure 1-6.
V out:
V in supply DC/DC converter Analog digital
supply potentiometer
LED source
analog power supply
LED source control
gain
Load adjustment
Resistance
Silicon
photodetector Data out
Sigma/delta 20 bits A/D
converter
Sample
sensor
offset
adjustment
to interface
3rd stage:A/D voltage reference
1st stage: I to V converter 2nd stage:amplifier board
conversion source
gain
Load adjustment
Resistance
Silicon
photodetector Sigma/delta 20 bits A/D Data out
converter
Reference
sensor digital
potentiometer
offset
adjustment
The supply voltage and the reference voltage required for the converters are
produced on the detector board.
1.4.1.5.1 Functions
The sample arm is a functional module represented in the "D-10, Sample Arm
Module, Drawing 1 (front view)" (Figure 1-7) and in the "D-10, Sample Arm
Module, Drawing 2 (rear view)" (Figure 1-8). The arm supports the sampling
probe, tube holder, barcode reader and magnetic sensor. The functions of the
sample arm are:
• lateral left/right movement (X-axis movement)
present
positions the probe above the dilution well
positions the probe above the probe wash well
• vertical up/down movement (Z-axis movement)
the sample type detected
positions the probe in the high (home) position to allow X-axis
movements
• vertical up/down movement of the tube holder system (TH-axis movement)
holds the tube when the probe is removed from the tube
positions the tube holder system to enable X-axis movements
The barcode reader reads any barcode present on the samples once the rack is
introduced to the system.
The magnetic sensor detects the magnets present on any sample vial adapters in
the rack and indicates to the system that the sample is prediluted.
Z and TH
movement
TH lead screw
Tube
holder
X movement
Z lead screw
TH axis
sensor
device for
driving the X home sensor
X pulley sampling arm
(1 of 2) X belt
1.4.1.5.2 Components
Refer to figures "D-10, Sample Arm Module, Drawing 1 (front view)" (Figure 1-7)
and "D-10, Sample Arm Module, Drawing 2 (rear view)" (Figure 1-8). The sample
arm assembly contains a fixed component that is mounted on the frame. The
fixed component holds all of the parts needed for X-axis movements and a mobile
subassembly, which travels along the X-axis and includes all the components for
movement along axes Z and TH.
The mobile subassembly is moved along the X-axis by a stepper motor with a
pulley and toothed belt. A device for driving the sampling arm is fitted on the belt.
Two aluminum rails, containing metal glides with a plastic surface finish, are used
to guide the mobile subassembly along the X-axis. A slotted optical sensor (X-
axis sensor) associated with a slotted disk fitted on the motor shaft (motor
encoder) monitors motor rotation. Another slotted optical sensor with a sliding flag
is used to locate the initialization position of the X-axis movement (X-axis
initialization or home sensor). The mobile subassembly contains components for
both probe and tube holder movement along the Z-axis.
Refer to figures "D-10, Sample Arm Module, Drawing 1 (front view)" (Figure 1-7)
and "D-10, Sample Arm Module, Drawing 2 (rear view)" (Figure 1-8).
Each X-axis probe position (dilution well, probe wash well, tube sample position)
corresponds to a motor position where the slot in the slotted disk is placed in front
of the sensor. The initialization position for the X-axis is the probe position above
the dilution well.
Counting the optical signals from the X-axis sensor (associated with the slotted
disk) controls the arm movements. Frequent returns to the home position are
placed in between the arm movements for safety. If the arm is already initialized
during the initialization movement, a short movement is made to change the
status of the X-axis movement sensors and check if they are operating correctly.
For Z-axis movements, both the motor steps and the signals from the Z-axis
sensor (associated with the slotted disk) are counted. For a given tube height, the
sample position on the Z-axis is determined and the movement takes place by
counting steps and slots. The Z-axis initialization position is reached when the
slots of the slotted disk are positioned in front of the Z-axis sensor and when the
Z-axis initialization sensor is covered. If the arm is already initialized during the
initialization movement, a short movement is made to change the status of the Z-
axis movement sensors and check if they are operating correctly.
For TH-axis movements, both the motor steps and the signals from the TH-axis
sensor (associated with the slotted disk) are counted. When a tube stops the TH
during downward movement, a flag is engaged in the slot of the TH tube height
sensor making it possible to stop the motor. The number of steps completed by
the motor is used to determine the height of the tube. Using this method for
measuring tube height determines if the sample is a primary or prediluted tube.
The probe depth is adjusted based on the type of tube detected.
If a sample vial is detected using the tube height measurement, its presence is
confirmed using the Hall effect sensor. If the Hall effect sensor detects the
presence of a magnet, and confirms that the sample is a sample vial, then the
sample is processed. If confirmation does not occur, the tube is ignored.
The TH-axis initialization position is reached when the slots of the slotted disk are
positioned in front of the TH-axis sensor and when the TH-axis initialization
sensor is covered. If the arm is already initialized during TH-axis initialization, a
short movement is made to change the status of the TH-axis movement sensors
and check if they are operating correctly. The movements along the Z and TH
axes are simultaneous when possible.
The barcode reader is used to identify each sample. The identification process
occurs in two stages:
• the sample arm moves over the entire length of the rack once it is
introduced and the tube barcodes are scanned. The sample identifications
are then placed into the software to build the worklist.
• prior to sampling from the tube, the barcode for the sample is re-read, with
the exception of any sample identification manually entered by the user
into the worklist. If the information from the second read does not match
the first read, the tube is ignored.
1.4.1.6.1 Functions
The rack is handled semi-automatically for ease of introduction into the system.
This rack handling includes:
• rack door opening and closing
• rack insertion and ejection
Refer to diagram "D-10, Rack Door System (internal view)" (Figure 1-9). The rack
door system includes:
• a post attached to the sample arm subassembly for closing the rack
insertion door
• a locking solenoid
• an open door sensor
• a closed door sensor
Locking
Solenoid
Closed door
sensor
Refer to diagram "D-10, Rack Insertion and Ejection System" (Figure 1-10).
The rack insertion and ejection system includes a bi-directional geared motor with
a rubber drive roller, the rack insertion sensor ("rack input sensor") and the rack
removal sensor ("rack output sensor").
The rack door is open only when the system is in Standby state. The door opens
as a result of the action of two return springs. The door closes when the post on
the sample arm subassembly contacts the plate on the door latch mechanism.
The springs of the mechanism are compressed by this contact. The door swings
closed by means of a connecting rod. The locking solenoid is energized in order
to enable the door to swing freely. At the end of the movement, the solenoid is
released, locking the door in the closed position. The door closed opto-sensor is
then uncovered, indicating that the door is in the closed position.
The control of the locking solenoid and the action of the return springs opens the
door. The post on the sample arm subassembly is in contact with the plate on the
door latch mechanism during this time, to ensure smooth movement of the door.
The solenoid is released at the end of the process. The door open opto-sensor is
then uncovered, indicating that the door is open. Once open, a rack can be
inserted into the system. The user starts to introduce the rack manually. When the
rack goes past the reflective sensor that detects rack insertion, the rack driving
Rack removal is only done when requested by the user. The motor starts up in
the rack removal direction. The rack moves past the removal sensor. Once the
rack passes the sensor, the motor stops. The rack insertion sensor detects when
the user completely removes the rack from the system.
The touch screen panel allows the user to make selections and enter information.
The results are printed using a thermal printer on 112mm wide paper. Printouts
include the chromatogram, the list of hemoglobin peaks, retention times, result
calculations and identification.
The motherboard is interfaced with the ¼ VGA touch screen through a piggyback
board on the PC board and a LCD and touch panel interface board located close to
the LCD display. The screen is connected to the LCD port of the board. An
RS232C serial port interfaces the touch panel. The motherboard is also interfaced
with the thermal printer through an RS232C serial port. All common terminals can
be connected to the PC board due to the presence of the following ports:
• Keyboard port
• Mouse port
• Ethernet port
• Serial RS232 port
• VGA port
Lastly, the PC board communicates with the MC processor via a serial RS422 link.
MC, HPPC and TC processors are installed on the interface board. The MC
processor is the master controller and is the processor responsible for automatic
device operations based on the commands from the PC board. These include:
• X, Z and TH axis motors and sensors
• motorized syringe
• high-pressure valve
• rack door
• rack introduction and ejection
• barcode reading
• detector reading and driving
• waste circuit (pinch valves, pump, vacuum sensor)
• degasser driving (vacuum pump, vacuum sensor, bubble detectors)
• front cover (solenoid and sensor)
• interface with HPPC and TC processors
The HPPC is the processor that controls the DC motor of the high-pressure pump
on the basis of the signals from the optical encoder, the cam signal and the signal
from the pressure sensor. It also drives the proportioning valve in synchronization
with the high-pressure pump. It works as a slave of the MC processor, which
communicates with it through a serial (SPI) link.
The TC is the processor that measures the temperature using the Pt100 ohm
sensor in the cartridge module. On the basis of the measured temperature and a
regulation algorithm, it controls heating to the programmed target temperature by
controlling the two Peltier elements connected in series in the PWM mode. A
thermal fuse in series with the Peltier modules prevents accidental overheating.
The processor works as a slave of the MC, which communicates with it through an
SPI link.
Two standard power supplies furnish power to all of the electrical components.
Power supply 2 delivers +24VDC for the stepper motors and the thermal printer.
The stepper motor supply is switched off when the front door is open. The device is
connected to the mains by means of an input module fitted with two protective
fuses and an anti-parasitic filter to ensure that it is EMC compatible. A 2-pole switch
is used to switch the device on and off.
The first document (Sheet 1) shows the electronic module, which includes the
following:
• power supplies 1 and 2
• PC motherboard
• all the peripheral connectors
• interface board
• connection cables internal to the module
The second document (Sheet 2) shows the interface board and all connections to
the connection boards present on the other modules. For each module, it details
the connection of all the electrical elements of the module.
These symbols are intended to draw your attention to particularly important information
and alert you to the presence of hazards as indicated.
Note: only a qualified person aware of the safety procedures is allowed to repair or
perform service procedures on this instrument.
Earth terminal
Power On
Power Off
2 SPECIFICATIONS
2.1.2 Weight
2.2.2 Throughput
Buffer 1: 2000mL
Buffer 2: 2000mL
Wash/Diluent: 2000mL
Internal Volume
Buffer 1 and 2: 26.5mL
Wash/Diluent: 5.3mL
Typical Vacuum Level: 0.5bar ±0.05bar; can be modified
Maximum Vacuum Level: 0.7bar
Vacuum Sensor: Piezoelectric sensor
Accuracy: ±0.05bar @ 0.5bar
Full Scale: 1bar
Vacuum Control: An external connection is provided to allow
connection of another instrument
Vacuum regulation: 0.05bar hysteresis; vacuum controlled by
micro-controller
2.2.13 Detector
Wavelength
Sample: 415nm
Zero Abs. Reference: 415nm, measured at the beginning of the
cycle
Filter: Interference filter, centered on 415nm,
bandwidth 10nm
Light Source
Type: LED, 430nm peak wavelength
Direct Current: 15mA typical
Current Control: Numeric potentiometer, 255 points,
0.1mA/point
Lifetime: 50,000h
2.2.17 Onboard PC
2.2.19 Printer
Type: Thermal
Paper: 112mm-width roll
3 SYSTEM DESCRIPTION
Only a qualified person aware of the safety procedures is allowed to repair or perform
Service procedures on this instrument.
3.2 Frame
3.3 Modules
3.3.1 Syringe Module
Figure 3-4: Syringe Module, Front Figure 3-5: Syringe Module, Rear
1 HP Pump
2 7-port, 3-position HP valve
3 Proportioning Valve
4 Pressure Sensor
5 HPLC Module Board
6 Cartridge Module
7 Detector Module
* these devices are not identified by a wire number and are interchangeable
** these front cover devices are connected to the HPLC Module Board via an
interlock switch (see 3.3.13)
1 Thermal Fuse
2 Pt100 Thermal Sensor
3 Peltier Modules
3.3.5.1 Description
3.3.5.2.1 Identification
Two versions of the detector module exist. The identification is done by the
revision number on the detector module label:
The basic difference between the two versions is the size of the reference
photodiode.
Increasing the active area induces a higher signal to noise (S/N) ratio.
The new versions are equipped with a 3-pin connector associated with a resistor
and a jumper.
The jumper is used to configure the detector board for each version of the
reference photodiode.
Only the new version of detector boards DAE011+6 is delivered as spare part
(P/N: 2200265).
Figure 3-31: Sample Arm Module, Front Figure 3-32: Sample Arm Module, Rear
1 Protective Grid
2 Power Supply 2 (+24VDC)
3 Power Supply 1 (+5VDC, +12VDC, +26VDC, -12VDC)
4 Interface Board
5 PC Motherboard
1 Keyboard Port
2 RS232 Port*
3 Rack Feeder Port
4 LAN Port*
5 VGA Port
6 Mouse Port
Power Supply 1:
• Location: FS1
• Type: F5A/250V
• Dimensions: 5mm x 20mm
Power Supply 2:
• Location: F1
• Type: F4A/250V
• Dimensions: 5mm x 20mm
Figure 3-49: Degasser Module, Front Figure 3-50: Degasser Module, Rear
Figure 3-57: Rack Drive Module, Front Figure 3-58: Rack Drive Module, Rear
1 Dilution/Wash Station
2 +24VDC Security Connector
3 Sample Rack (with Tubes)
4 Guide Rail
5 Rack Input Sensor
6 Rack Output Sensor
7 Rack Positioning Motor Assembly
Figure 3-62: Rack Door Module, Inside Figure 3-63: Rack Door Module, Outside
Figure 3-66: External Shell, Exploded View Figure 3-67: Front Cover Assembly
The electrical parts of the front cover assembly are connected to 2 interlock switches:
• one for the ON/OFF switch (+24VDC power supply of the stepper motors)
• one for the solenoid and the sensor
Depending on the chassis version, these interlock connectors can be located on two
places:
1. for the first chassis version: on the HLPC module (solenoid and sensor) and
under the rack drive module (ON/OFF switch) (see Figure 3-72)
2. for the last chassis version: on the chassis itself, near the main ON/OFF switch
(see Figure 3-71)
Figure 3-71: Interlock Connector Locations, Figure 3-72: Interlock Connector Locations,
Last Chassis Version First Chassis Version
The ON/OFF interlock switch is connected to the power cable coming from Power
Supply 1 (see 3.3.7).
The solenoid and sensor interlock switch is connected to the High Pressure Module
Board (see 3.3.2).
On the latest D-10 version, two backups of the 'instrument.ini' file are
available:
• a paper copy
• an electronic copy stored on a floppy disk
* all tubing with this part number included in package (with fittings)
** tubing included and attached to flow cell
3.5 Accessories
Sample Rack
Type: Sysmex #3
Equipment: Anti-Rotation Rubbers on the bottom
Number: 2
4 INSTALLATION PROCEDURE
• Choose a location for the system that is away from direct sunlight and relatively
dust-free.
• Room temperature should be between 15 and 30°C.
• The bench or table should have a flat, level surface that is free from vibrations and
is capable of supporting a weight of more than 35kg (77lbs.).
• The system requires the minimum benchtop space of 100cm (W) x 48cm (H) x
65cm (D) [ 39in. (W) x 19in. (H) x 26in. (D) ] to allow for the system plus proper
clearances.
• Maintain a minimum clearance of 20cm (8in.) on the left side and 40cm (16in.) on
the right side. Maintain clearance of 10cm (4in.) from the back side to ensure that
the power cord input is readily available to the operator.
• A grounded electrical receptacle should be within 1.8 meters (6 feet) of the system.
Input power requirements for the system are between 100VAC and 240VAC. The
maximum power consumption is 250W.
• The use of a circuit breaker before the main electrical receptacle is required. The
electrical installation must comply with locally applicable standards and
requirements.
4.2 Installation
4.2.1 Unpacking and Inspection
4.2.2 Set Up
4.3 De-installation
To de-install a D-10, the following steps must be performed in the order listed below
before shipment.
Warning: use caution when de-installing the unit and follow local regulations
regarding the treatment of spills and the disposal of biohazardous materials.
4.3.1 Decontamination
Wipe the outside of the sample probe with a cloth dampened with 5% hypochlorite.
1. Remove the analytical cartridge and replace it with the PEEK dummy cartridge.
Place the green end caps onto the cartridge and store it for re-use.
2. Allow the HP Pump to run for 5 minutes setting the flow rate at 2mL/min and the
percentage of buffer 2 to 0% (use the MAINTAIN screen).
3. Stop the HP Pump.
4. Remove the Buffer 1, Buffer 2 and Wash fluidic connections on the rear panel of
the D-10 and replace the 3 port plugs.
5. Empty and dry the external tubing from buffers and wash solution before placing
them in the accessories bag.
6. Remove the external Waste tube. Empty and wipe dry. Store it in a separate
plastic bag.
1. Move the sample arm to the right side of the instrument. Two solutions exist,
depending on the system status:
if the 'Dia.exe' application is running, go to the MAINTAIN screen, select
Service, then Access Wash st.
if the D-10 Service software is running, go to the D-10 Service Software
(2/2) screen, select SAMPLE ARM, then DILUT. /WASH STAT. REPLA.
The sample arm moves to the right side and the tube holder hides the probe tip.
As soon as the front cover solenoid is activated, open the front door.
4. Manually push the sample arm to the right side until the arm standoff comes in
contact with the thumbscrew. Moderately tighten the screw.
4.3.4 Packing
Caution: before packing, make sure that the system is well cleaned,
decontaminated and has storage fluid in the fluid pathway.
1. Remove the paper roll from the thermal printer unit and place it in the
accessories box.
2. Remove all the electrical connections on the rear panel.
3. Remove all the bottles from the reagent tray and remove the reagent tray itself.
1. Collect all the accessories listed below and pack in the accessories box:
sample racks
sample vial adapters
power cord(s)
spares fuses
bottle closures with tubes
buffer 1 tubing
buffer 2 tubing
wash tubing
waste tank tubing
50mL syringe
barcode labels (CAL1, CAL2, CTRH, CTRL, PRIM)
1. Empty the waste bottle according to local regulations for the disposal of
biohazardous waste.
2. Rinse the bottle with bleach solution.
3. Rinse the bottle with water.
4. Use the small box to pack the 10L waste tank.
5. Close the box.
1. Place the bottom section of the box, with packing foam, on the floor. Determine
the position of the instrument in the foam.
2. Two people should lift the D-10. Grasp the unit from the benchtop and place it
carefully in the bottom box, in the appropriate position.
3. Turn the reagent tray upside down and replace it in its original position.
4. Place the accessories box over the reagent tray.
5. Enclose the D-10 with the plastic bag.
6. Place the top packing foam on the top of the instrument.
7. Place the top section of the box.
8. Use the 4 plastic latches to secure the upper and lower parts of the box.
9. Place the waste tank box on top of the main box.
10. Secure the two boxes together with bands.
5 SOFTWARE OVERVIEW
The \HbA1c, \BThal and \Decont folders have the same structure and contain
subfolders, that include:
• \Calib, containing the calibration files
• \Param, containing the parameter files
• \Results, containing the results files
• \Extra, currently not used
Note: some *.ini files may be updated during the lifetime of the instrument.
Some *.ini files are used to initialize functional parameters of the instrument.
Information about some files is given below. All the parameters are not described, but
the main or the most important parameters are given below.
On the latest D-10 version, two backups of the 'instrument.ini' file are available:
• a paper copy
• an electronic copy stored on a floppy disk
These backups are located inside the left door (see 3.3.13).
The "Rule1","Rule2" and "Rule3" parameters are used for the calculation of the
pump speed in order to have a target flow rate. The values are obtained by a flow
rate calibration of the pump.
Examples:
• "Rule1=0,1,2400,0": for a flow rate between 0 and 1mL/min, the slope and the
intercept for the calculation of the motor speed are 2400 and 0.
The "TCCal_Gain" and the "TCCal_Offset" parameters correspond to the slope and
the intercept of the temperature measurement circuit on the Interface Board. They
are associated with the Interface Board.
The "TCCal_SensorOffset" is associated with the sensor and corresponds to the
sensor offset correction.
The "TCCal_GradientOffset" is associated with the cartridge module and
corresponds to the gradient correction between the position of the sensor (at the
back of the module) and the position of the cartridge (at the front of the module).
This file contains the majority of the information and parameters for running the D-10.
These parameters are assay-dependent (A1c, A2/F/A1c, Decont). The following
parameters are grouped according to their use:
• status and general information
• initialization movements
• syringe circuit priming
This file contains information and parameters that are assay-dependent. For A1c, a
'Profile2.ini' file is found in the \Param folder of the \HbA1c folder. For A2/F/A1c, a
'Profile2.ini' file is found in the \Param folder of the \BThal folder. The following
parameters are grouped according to their use:
• status section
• kit section, containing information associated with the kit
• inject section, containing information associated with sample preparation
• chroma section, containing information associated with the chromatogram
• priming and pump parameters sections
Information or errors about the LIS connection are logged into the 'Logevent.txt' file
located in the \Dia folder.
• *.csv files contain the sample and the LED reference measurements for the
test
• *.raw files contain the absorbance measurement of the test. The data analysis
module (DAM) processes this file.
• *.dam files contain the results of the data processing. This file is used to
generate result printouts. These files are listed in a 'datalist.txt' file.
• datalist.txt file
The update kit disk, contained in the reorder pack, will update the information in the
corresponding \Param folder.
All motors, valves and pumps are in standby at power up and only receive power at a
specified time in the initialization sequence. If power is supplied to these components
before the initialization sequence starts, it indicates a problem with the component or
any component associated with it. The power up sequence is as follows:
• The PC boots:
A short beep indicates the PC has booted after the internal self test
CPU identification
Memory test
IDE detection
Hardware check
• At the same time, the low-level processors MC, HPPC and TC boot. The
associated red LEDs are blinking, indicating that the boot is OK (frequency
between 1 and 2 Hz) (for the location of these LEDs, see Figure 3-42). The
degasser pump is switched on until the default degasser vacuum level is
reached. If the level is not reached after a time out, the pump is stopped.
• Windows NTe operating system is launched.
• The BIO-RAD logo and the D-10 icon are displayed.
• The 'Dia.exe' application is launched and the *.ini files are loaded:
Dia.ini for the preparation of the user interface
Profile2.ini for loading test names and kit parameters
instrument.ini for loading instrument parameters
DiaDAM.ini for loading data analysis parameters
• Data files are loaded (500 last samples of the selected test).
• The instrument parameters are transmitted to the low level processors (on the
Interface Board).
• The internal devices under control of the low level processors are initialized
(see hardware initialization below).
The RUN screen is displayed when the power up sequence is complete. If any
hardware component does not initialize, an error message is displayed.
Hardware Initialization
• If a test is running, wait until the complete rack is processed or select the
Stop button to discontinue processing after the current sample is finished.
• Eject the rack from the system after the sample is complete.
• Select Sleep in order to return to the Sleep state. Wait until the end gradient
is finished.
• Once the system is in Sleep state, select the Shut Down button.
• A message appears "Please wait while the system writes unsaved data to the
disk". If there is no external monitor, this message is not seen in its entirety.
• After several seconds a new message is displayed: "It is now safe to turn off
your computer". The main ON/OFF switch can now be switched to OFF.
This description refers to the DiaSer Service software version 2.00, which is installed
with Dia software version 3.0B. This description remains valid for future versions if no
additional documentation is provided.
5.10.1 Objective
5.10.2 Access
modification is done for test issues, the initial value of the parameter must
be restored.
• A third password (level 3) is defined for installation process (SETUP) from the
floppy. If this password is entered, the system goes directly to the SETUP
process defined on the floppy disk.
Passwords:
Level 1: D-10QC
Level 2: DIA904
Level 3: D-10SETUP
5.10.3 Exit
By selecting the EXIT button in one of the DiaSer main screens, you can quit the
Service software and return to the 'Dia.exe' program.
Warning: when the EXIT button is pushed, 'DiaSer.exe' is closed and 'Dia.exe'
is started automatically. During the boot sequence, the Dia icon is displayed on
the screen. DO NOT select this icon!!! If selected, 'Dia.exe' is started again, and
an error message is displayed.
Note: the executable software program started by selecting the EXIT button is given
in the 'DiaSer.ini' file: it is the name associated with the "User Software" parameter,
in the "Maintain" section. The default value is "Dia.exe".
The 'DiaSer.exe' software uses the 'DiaSer.ini' file containing some initialization
values. A view of the file is given in the Appendix A.
Note: depending on the software version and level selected, some options indicated
on the screens cannot be accessed. These options will be grayed out.
allow the user to modify the data directly. When a new value is edited, it can
be marked by an "N" (for "New") indication, which is removed when this new
value is saved.
• A field with a grey background displays data that can not be modified. Areas
with low contrast correspond to options that are not accessible in the context
or that are not implemented.
• When an action is ongoing, the user should wait until it is finished before
starting the next action. If an action is started before the end of the current
action, a message is displayed with the option of waiting for the current action
to finish or to stop the current action.
On the bottom of each screen, the current level in the Service software is given
(module identification), with the current screen number and the total number of
screens for the same level. Date and time are also displayed. The question mark can
be used to obtain additional help regarding use of the Service software. Information
on all screens may not always be available. A GRAPHIC button may be available. If
the corresponding option is implemented, it can be used to display the variation of
the selected electrical signal with time.
One of the two following screens opens when the user has to enter new data. If the
data is a numerical value, the numeric keyboard opens. When the data is a password
or a file name, the alphanumeric keyboard opens.
One of the two following screens opens during a file export to a floppy, or during a file
import from a floppy.
Figure 5-3A: File Selection, Save Figure 5-3B: File Selection, Open
On those screens, the directory, the files, and the file type can be selected.
• DONE: perform the desired action on the selected file.
• CANCEL: cancel the operation.
• DELETE: delete the selected *.csv file. Only a *.csv file can be deleted. A
confirmation screen is open to confirm the deletion.
• EDIT: allows for editing of the file name. The alphanumeric keyboard screen
is displayed.
Sensed Values
Figure 5-4:
Example of Sensed Information
On several screens which include control buttons for motors, pumps, valves… it is
possible to see if a device is on or off. The information is obtained by measuring the
amplified voltage across a small resistor installed on each device's electronic
controller. This information is given in the "xxx" sense field, with "xxx" identifying the
device. Refer to Section 5.10.26 for typical sense values.
The unit can be volts or points. It can be changed by clicking in the corresponding
field.
• volts indicates that the value is given in Volts.
• points indicates that the value is given in number of points of the Analog to
Digital Converter (scale: 0 to 255).
Graphic Screen
Figure 5-5:
Graphic Screen
By selecting the GRAPHIC button, the evolution with the time of electrical signals is
displayed. The sampling rate is approximately 1 sample per second. The list of the
signals available is given on the screen, by a pull-down menu. Typically, following
signals are available:
• Pressure: signal delivered by the high-pressure sensor.
• Absorbance: signal delivered by the sample photodiode.
• Reference: signal delivered by the reference photodiode.
Note: in the following section, some parts of the software contain multiple screens for
the component selected. The screen number is indicated along with the total number
of screens for that component. For example, Main Screen (1/2) indicates that this is
screen 1 of 2.
When the 'DiaSer.exe' application is started, the Service software access screen is
displayed. A level 1, level 2 or level 3 password can be entered by clicking in the
white field.
Figure 5-6:
Password Screen
The alphanumeric keyboard will appear. Once the password of level 1 or level 2 is
entered, the first of the two main screens is displayed. If a level 3 password is
entered, the system goes directly to the SETUP procedure.
These two main screens allow access to all modules of the D-10:
• SYRINGE: access to the Syringe Module
• DEGASSER: access to the Degasser Module
• DETECTOR: access to the Detector Module
• RACK DOOR: access to the Rack Door Module
• RACK INTRO.: access to the Rack Introduction Module
• FRONT COVER: access to the Front Cover Module
• HP: PUMP + VALVE: access to the HP Module: High-Pressure Pump and
Valve
• HP: SENSOR CALIB.: access to the HP Module: High-Pressure Sensor
Calibration
• HP: FLOW RATE CALIB.: access to the HP Module: Flow Rate Calibration
• THERMAL CONTROL: access to the Thermal Control Module
• SAMPLE ARM: access to the Sample Arm Module
Selecting the SYRINGE button in the Main Screen (1/2), opens a set of six screens
for testing the Syringe Module.
Figure 5-9:
D-10 Syringe module (1/6)
Figure 5-10:
D-10 Syringe module (2/6)
Figure 5-11:
D-10 Syringe module (3/6)
Figure 5-12:
D-10 Syringe module (3A/6)
This screen shows the result of the current waste bottle check and indicates the
waste bottle check parameters:
• High level time: indicates the time required for the vacuum to reach the High
vacuum level given in the parameters area. Unit: s.
• Dilution valve low level time: indicates the time required for the vacuum to
return to the atmospheric pressure when the Dilution Valve is switched off.
Unit: s.
• Wash valve low level time: indicates the time required for the vacuum to
return to the atmospheric pressure when the Wash Valve is switched off. Unit:
s.
• Zero vacuum level: indicates the offset of the vacuum electrical signal at the
atmospheric pressure. Unit: selected by clicking in the points area.
• points: unit of the vacuum. Options are: kPa, psi, volts (voltage delivered by
the sensor) and points (8-bit Analog to Digital Converter result).
• Result: indicates the waste check has passed (OK) or failed (FAILED).
• High vacuum level: level defined as the limit for a good internal vacuum
circuit. Unit: kPa, psi, volts or points.
• Offset for the low vacuum level: offset added to the measured Zero
vacuum level, to define the low vacuum level for the measurement of the
time required by the decreasing of the vacuum. Unit: points, kPa, psi, volts.
• High level time out: maximum time allowed to obtain the High vacuum level
during the test. Unit: s.
• Dilution valve low level time out: maximum time allowed for the vacuum to
return to the low vacuum level defined by Zero vacuum level + Offset for
the low vacuum level, on the Dilution valve circuit. Unit: s.
• Wash valve low level time out: maximum time allowed for the vacuum to
return to the low vacuum level defined by Zero vacuum level + Offset for
the low vacuum level, on the Wash valve circuit. Unit: s.
• RESTART: start a waste bottle check again.
• CLOSE: close the waste bottle check window.
Figure 5-13: D-10 Syringe module (4/6) Figure 5-14: D-10 Syringe module (5/6)
These two screens show, for the selected assay, the values of the most important
parameters of sample preparation. These values are found in the corresponding *.ini
files. These values can be modified and saved in the corresponding files only if the
user has entered the level 2 password.
Figure 5-15:
D-10 Syringe module (6/6)
This screen shows the values of the internal waste parameters. These values are in
the 'Profile.ini' file. These values can be modified and saved in the corresponding
files only if the user has entered the level 2 password.
• High vacuum level (points): level defined as the limit for a good internal
vacuum circuit. Unit: points.
• Offset for the low vacuum level (points): offset added to the measured
Zero vacuum level (see Syringe module (3A/6), to define the low vacuum
level for the measurement of the time required by the decreasing of the
vacuum. Unit: points.
• High level time out (0.1s): maximum time allowed to obtain the High
vacuum level during the test (see Syringe module (3A/6). Unit: 0.1s.
• Dilution valve low level time out (0.1s): maximum time allowed for the
vacuum to return to the low vacuum level defined by Zero vacuum level +
Offset for the low vacuum level, on the Dilution valve circuit. Unit: 0.1s.
• Wash valve low level time out (0.1s): maximum time allowed for the
vacuum to return to the low vacuum level defined by Zero vacuum level +
Offset for the low vacuum level, on the Wash valve circuit. Unit: 0.1s.
• DONE: save the modified parameters. A verification message is displayed to
confirm the modification. The DONE button is usable only if a modification is
made and if the level 2 password was entered.
• CANCEL: cancel the modifications. A verification screen is displayed to
confirm the cancellation. The CANCEL button is usable only if a modification
is made and if the level 2 password was entered.
Selecting the DEGASSER button in the Main Screen (1/2), opens a set of four
screens for testing the Degasser Module.
Figure 5-16:
D-10 Degasser module (1/4)
• START PUMP: start the degasser vacuum pump. The real time clock of the
MC processor continuously checks the vacuum. If the target value of the
vacuum (High vacuum, see Degasser module (3/4) is reached, the pump is
automatically switched off.
• STOP PUMP: stop the degasser vacuum pump. The real time clock of the
MC processor continuously checks the vacuum. If the target value of the
vacuum (Low vacuum, see Degasser module (3/4) is reached, the pump is
automatically switched on.
• Vacuum: current vacuum level of the degasser measured by the vacuum
sensor on the degasser module board. The unit of measure appears in the
box next to the value.
• points: unit of vacuum. Options are: kPa, psi, volts (voltage delivered by the
sensor) and points (8-bit Analog to Digital Converter result).
• Sense: indicates the electrical current intensity sensed through the vacuum
pump. Unit: points or volts.
• GRAPHIC: see 5.10.5.
• VACUUM CONTROL ENABLE (FOR TEST): if selected, the automatic
switch off/on is disabled. In this case, the threshold levels indicated on the
screen are used for this test.
• Thresholds for test use, Low and High: thresholds for the vacuum pump
control if the VACUUM CONTROL ENABLE (FOR TEST) is activated.
Figure 5-17:
D-10 Degasser module (2/4)
• BUFFER 1: start the automatic priming of the buffer 1 line through the
degasser. Approximate duration: 10 minutes.
• BUFFER 2: start the automatic priming of the buffer 2 line through the
degasser. Approximate duration: 10 minutes.
• WASH: start the automatic priming of the wash line through the degasser.
Approximate duration: 3 minutes.
• ALL: start the automatic priming of all the lines through the degasser.
Approximate duration: 23 minutes.
• Bubble detector signal for Buffer 1 line, Buffer 2 line, Wash line: indicates
the level of the bubble detector signal. Unit: points or volts. The bubble
detectors are currently not in use.
• GRAPHIC: see 5.10.5.
Figure 5-18:
D-10 Degasser module (3/4)
This screen shows the values of the parameters of the degasser control and the
bubble detection. These values are found in the *.ini files. These values can be
modified and saved in the corresponding files only if the user has entered the level 2
password.
• Low vacuum: when the degasser vacuum is less than this value the
degasser pump is switched on. The unit of measure appears in the box next
to the value. The parameter is saved in points.
• points: unit of the Low vacuum value. Options are: kPa, psi, volts (voltage
delivered by the sensor) and points (8-bit Analog to Digital Converter result).
• High vacuum: when the degasser vacuum is greater than this value, the
degasser pump is switched off. The unit of measure appears in the box next
to the value. The parameter is saved in points.
• Time out: maximum time allowed to obtain the vacuum level when the pump
is started. Unit: s.
• Wash bubble det., Buf. 1 bubble det., Buf. 2 bubble det.: threshold levels
for the bubble detection. This functionality is currently not used.
• DONE: save the modified parameters. A verification message is displayed to
confirm the modification. The DONE button is usable only if a modification is
made and if the level 2 password was entered.
• CANCEL: cancel the modifications. A verification screen is displayed to
confirm the cancellation. The CANCEL button is usable only if a modification
is made and if the level 2 password was entered.
Figure 5-19:
D-10 Degasser module (4/4)
This screen shows the values of the parameters used for the flush of the system.
These values are found in the *.ini files. These values can be modified and saved in
the corresponding files only if the user has entered the level 2 password.
• Buf 1/2, Wash Syr. Asp. Speed: index to define the speed of the aspiration
of buffer by the syringe. No unit.
• Buf 1/2, Wash Syr. Deli. Speed: index to define the speed of the delivery of
buffer by the syringe. No unit.
• Number of strokes, Buffer 1/2: number of syringe movements for the
priming of the buffer 1 and buffer 2 lines.
• Number of strokes, Wash: number of syringe movements for the priming of
the wash line.
Selecting the DETECTOR button in the Main Screen (1/2), opens a set of four
screens for testing the Detector Module.
Figure 5-20:
D-10 Detector module (1/4)
• Load gradient file: selection of the file (*.hpl) for a gradient, if the user wants
to run a gradient.
• Selected log file: selection of the file (*.csv) for the logging of the results of
the detector when the logging is started.
• Set Light: light intensity parameter, from 0 to 255. The default value is 150
and corresponds to a direct current in the LED of 15mA. When this value is
set to 0, the LED is switched off.
• Set Gain: numerical gain in the reference channel of the detector, from 0 to
255. The default value is 64. This gain is automatically adjusted, before each
sample, in order to adjust the baseline of the chromatogram to a given level
which is typically 0.020AU.
• Flow rate: manual: flow rate (ml/min) or speed of the pump (rpm), used for
the manual control of the pump.
• ml/min: unit of the Flow rate. The options are: ml/min or rpm.
• Flow rate: gradient: flow rate (ml/min) or speed of the pump (rpm), in the
selected gradient file.
• START GRAD. + LOG: start/stop the gradient selected into the Load
gradient file field (*.hpl) and start/stop the recording of the detector values
into the log file selected in the Selected log file field. The flow rate used in
the gradient file (*.hpl) is displayed in the Flow rate: gradient field.
• START LOG: start/stop the recording of the detector values into the log file
selected in the Selected log file field. The file is a *.csv file and can be
exported for analysis (see Detector module (2/4).
• START PUMP or STOP PUMP: start/stop the high-pressure pump for a flow
rate chosen in the Flow rate: manual field.
• START PROP. VALVE: start/stop the proportioning valve.
• GRAPHIC: see 5.10.5.
• Abs. Meas.: light measurement in the sample channel of the detector. The
result is given in 24-bit converter points (from 0 to 16,777,215).
• Ref. Meas.: light measurement in the reference channel of the detector. The
result is given in 24-bit converter points (from 0 to 16,777,215).
• Log.: it is equal to log10 (reference measurement/sample measurement). The
result is given in Absorbance Unit (AU).
• Pressure: current pressure in the high-pressure circuit. The unit of measure
appears in the box next to the value.
• psi: pressure unit. The options are psi or kg/cm2.
Figure 5-21:
D-10 Detector module (2/4)
• RESULT FILE EXPORT: export a result file to a floppy disk. The explorer
screen is shown for the selection of the file to export.
• Abs. Meas.: light measurement in the sample channel of the detector. The
result is given in 24-bit converter points (from 0 to 16,777,215).
• Ref. Meas.: light measurement in the reference channel of the detector. The
result is given in 24-bit converter points (from 0 to 16,777,215).
• Log.: it is equal to log10 (reference measurement/sample measurement). The
result is given in Absorbance Unit (AU).
• Pressure: current pressure in the high-pressure circuit. The unit of measure
appears in the box next to the value.
• psi: pressure unit. Options are psi or kg/cm2.
• Baseline adjustment sequence: the baseline adjustment process consists
of a software adjustment of the gain on the reference measurement channel
(numeric potentiometer) in order to obtain an absorbance defined by the
calculated value log10 (1 + BaseLinePercent). This adjustment is
Warning: in the DiaSer Service software version 2.00, the values displayed
in the result field are wrong. In later DiaSer versions, the displayed values
are correct.
Warning: the result of the adjustment is lost when coming back to the
Detector module (1/4) screen, because the Light and Gain values were
already sent to the detector components.
Figure 5-22:
D-10 Detector module (3/4)
This screen shows the values of some parameters used for the detector, based on
the selected assay. These values are found in the *.ini files. These values can be
modified and saved in the corresponding files only if the user has entered the level 2
password.
Figure 5-23:
D-10 Detector module (4/4)
This screen shows the values of some other parameters used for the detector based
on the selected assay. These values are found in the *.ini files. These values can be
modified and saved in the corresponding files only if the user has entered the level 2
password.
Selecting the RACK DOOR button in the Main Screen (1/2), opens a set of two
screens for testing the Rack Door Module.
Figure 5-24:
D-10 Rack Door module (1/2)
• START RACK DOOR SOLENOID: switch the rack door solenoid on/off to
lock/unlock the rack door. The rack door is locked when the solenoid is in
standby mode. When the solenoid is powered, it will be switched off
automatically after 10 seconds, to avoid overheating.
• Solenoid sense: indicates the electrical current intensity sensed through the
rack door solenoid. Unit: points or volts.
• GRAPHIC: see 5.10.5.
• Closed door sensor: status of the closed door detection sensor
obstructed: the door is in an intermediate position: the home flag is
in the slot of the optical sensor.
free, the optical slotted sensor is on (light is going from the emitter to
the receiver). The door is completely closed.
• Open door sensor: status of the open door detection sensor
obstructed: the door is in an intermediate position: the home flag is
in the slot of the optical sensor.
free: it indicates that the optical slotted sensor is on (light is going
from the emitter to the receiver). The door is completely open.
• Door status: current status of the rack door (open, closed or undefined).
• OPEN THE DOOR: launch the sequence to open the rack door.
• CLOSE THE DOOR: launch the sequence to close the rack door.
Figure 5-25:
D-10 Rack Door module (2/2)
This screen shows the values of the parameters used for the rack door control.
These values are found in the 'Profile.ini' file. These values can be modified and
saved in the corresponding file only if the user has entered the level 2 password.
• Run with rack door: for R&D or Manufacturing use only. It indicates if the
system considers that a rack door is mounted and runs well in the instrument
(yes) or not (no). The no option allows a run without rack door (the door is
always in the open position).
• Rack door monitor: indicates if the system continuously checks the rack
door during the run (yes) or not (no). If the rack door is monitored (yes), the
run is stopped when the rack door is accidentally opened.
• DONE: save the modified parameters. A verification message is displayed to
confirm the modification. The DONE button is usable only if a modification is
made and if the level 2 password was entered.
• CANCEL: cancel the modifications. A verification screen is displayed to
confirm the cancellation. The CANCEL button is usable only if a modification
is made and if the level 2 password was entered.
Selecting the RACK INTRO. button in the Main Screen (1/2), opens a screen for
testing the Rack Introduction Module.
Figure 5-26:
D-10 Rack Introduction module (1/1)
• Rack input sensor status: status of the rack input detection sensor
(reflective sensor)
reflection, a rack is in front of the sensor and reflects the infra-red
light.
no reflection, no rack in front of the sensor.
• Rack output sensor status: status of the rack output detection sensor
(reflective sensor)
reflection, a rack is in front of the sensor and reflects the infra-red
light.
no reflection, no rack in front of the sensor.
• Rack motor sense: indicates the electrical current intensity sensed through
the rack motor. Unit: points or volts.
• RACK MOTOR: IN START: switch the rack motor on/off in the "rack
introduction" direction.
• RACK MOTOR: OUT START: switch the rack motor on/off in the "rack
ejection" direction.
• LAUNCH RACK INTRODUCTION: start a rack introduction sequence.
• LAUNCH RACK EJECTION: start a rack ejection sequence.
• GRAPHIC: see 5.10.5.
Selecting the FRONT COVER button in the Main Screen (1/2), opens a set of two
screens for testing the Front Cover Module.
Figure 5-27:
D-10 Front Cover module (1/2)
• START FRONT COVER SOLENOID: switch the front cover solenoid on/off to
lock/unlock the front cover. The front cover is locked when the solenoid is in
standby mode. When the solenoid is powered, it will be switched off
automatically after 10 seconds, to avoid overheating.
• Solenoid sense: indicates the electrical current intensity sensed through the
front cover solenoid. Unit: points or volts. This function is currently not
available.
• GRAPHIC: not available on this screen.
• Front cover sensor status: current status of the front cover sensor
(reflective sensor)
reflection, the metal flag is in front of the sensor and reflects the
infra-red light. The front cover is closed.
no reflection, the front cover is open.
• Front cover status: current status of the front cover (open or closed).
Figure 5-28:
D-10 Front Cover module (2/2)
This screen shows the value of the parameter used for the front cover control. This
value is found in the 'Profile.ini' file. It can be modified and saved in the file only if the
user has entered the level 2 password.
• Front door monitor: indicates if the system continuously checks the front
door during the run (yes) or not (no). If the front door is monitored, the run is
stopped when the front door is opened.
Note: when the front door is open, the sample arm motors power supply is
switched off. If a run is in progress, it will be stopped by a motor error.
Selecting the ALL STATUS button in the Main Screen (1/2), opens a set of two
screens that show the following:
• operational status of subassemblies
• sensors status
• devices status (valves, solenoids…)
Figure 5-29:
D-10 All Status (1/2)
• X axis busy, Z axis busy, TH axis busy, Syr. axis busy: status of the X, Z,
TH and Syringe movements
busy, motor running.
free, motor in standby.
• X axis sensor, Z axis sensor, TH axis sen.: status of the X, Z, and TH axis
(position) sensors
obstructed: the optical sensor is "off". Infra-red light from the emitter
cannot reach the receiver because of the position of the slotted disk.
free: the optical sensor is "on". Infra-red light from the emitter reaches
the receiver because of the slot on the disk.
• X home sen., Z home sen., TH home sen.: status of the X, Z, and TH home
sensors
obstructed: the optical sensor is "off". Infra-red light from the emitter
cannot reach the receiver because the home flag is in the optical
sensor slot.
free: the optical sensor is "on". Infra-red light from the emitter reaches
the receiver.
• Hall effect: detection of a magnet in front of the sensor
magnet
no magnet
• Tube height: status of the tube height sensor
obstructed: the optical sensor is "off". Infra-red light from the emitter
cannot reach the receiver because the home flag is in the optical
sensor slot.
free: the optical sensor is "on". Infra-red light from the emitter reaches
the receiver.
• Syr. home: status of the syringe sensor
obstructed: the syringe is empty; the home flag is in the slot of the
syringe optical sensor (home position).
free: the optical slotted sensor is "on". Infra-red light from the emitter
reaches the receiver.
• HP valve: status of the high-pressure valve
load: valve in the Load position (home position).
inject: valve in the Inject position.
prime: valve in the Prime position.
not defined: valve in an intermediate, uncontrolled position.
• Open door: status of the open door detection sensor
obstructed: the door is in an intermediate position; the metal flag is in
the optical sensor slot.
free: the optical slotted sensor is "on". Infra-red light from the emitter
reaches the receiver. The door is completely open.
• Closed door: status of the closed door detection sensor
obstructed: the door is in an intermediate position; the metal flag is in
the slot of the optical sensor.
free: the optical sensor is "on". Infra-red light from the emitter reaches
the receiver. The door is completely closed.
Figure 5-30:
D-10 All Status (2/2)
Selecting this button, from the Main Screen (1/2), initializes all stepper motors:
• HP valve stepper motor (home position = Load position)
• Z motor (probe in home position)
• Tube Holder (home position)
• X motor (Sample Arm in the home position)
• Syringe motor (home position)
For the initialization of the syringe motor, the liquid pump is switched on, in order to
remove liquid from the dilution well.
Selecting this button, from the Main Screen (1/2), initializes the shutdown process:
• A message box opens for confirmation
if the user selects no, the shutdown process is aborted.
if the user selects yes, the shutdown process is started.
• A message appears "Please wait while the system writes unsaved data to the
disk". If there is no external monitor, this message is not seen in its entirety.
• After several seconds a new message is displayed: "It is now safe to turn off
your computer". The main ON/OFF switch can now be switched to OFF.
Selecting the HP: PUMP + VALVE button in the Main Screen (2/2), opens a set of
seventeen screens for testing the High-Pressure Pump and Valves.
Figure 5-31:
D-10 HP module: pump, valve (1/17)
• Flow rate/Pump speed: flow rate (ml/min) or its translation in pump speed
(rpm). The value can be modified only if the pump is switched off.
• ml/min: unit of Flow rate/Pump speed. The options are ml/min or rpm.
• START PUMP or STOP PUMP: start/stop the high-pressure pump.
Figure 5-32:
D-10 HP module: pump, valve (2/17)
Figure 5-33:
D-10 HP module: pump, valve (3/17)
This screen shows the HPPC status and gives low level information. It is mainly for
R&D use.
• HP pump sense: indicates the electrical current intensity sensed through the
high-pressure pump. Unit: points.
• Prop. Valve sense: indicates the electrical current intensity sensed through
the proportioning valve. Unit: points.
• Encoder signal: logical level signal from the pump motor encoder: A and B.
• Cam signal: logical level signal from the Cam flag sensor.
Figure 5-34:
D-10 HP module: pump, valve (4/17)
• HOME MVT (RESET): rotate the valve from any position to the home (Load)
position.
• TO LOAD POSITION: rotate the valve to the Load position, when it is in one
of the two other working positions (Inject or Prime).
• TO INJECT POSITION: rotate the valve to the Inject position, when it is in the
Load position.
• TO PRIME POSITION: rotate the valve to the Prime position, when it is in the
Load position.
Note: in order to go from Prime to Inject, the valve MUST return to Load position
first.
Figure 5-35:
D-10 HP module: pump, valve (5/17)
This screen shows the values of the parameters used for the High-Pressure Valve
control based on the selected assay. These values are found in the 'Profile2.ini' file.
These values can be modified and saved in the corresponding files only if the user
has entered the level 2 password.
Figure 5-36:
D-10 HP module: pump, valve (6/17)
This screen shows the values of the parameters used for the piston flush. These
values are found in the 'Profile.ini' file. These values can be modified and saved in
the corresponding files only if the user has entered the level 2 password.
Figure 5-37:
D-10 HP module: pump, valve (7/17)
Figure 5-38:
D-10 HP module: pump, valve (8/17)
This screen shows the values of the parameters of the High-Pressure Pump. These
values are found in the 'instrument.ini' file. These values can be modified and saved
in the corresponding files only if the user has entered the level 2 password.
• WC (Waiting Count): phase shift between the cam flag signal and the filling
phase start. Unit: ° (angle).
• FC (Filling Count): filling phase angle. Unit: ° (angle).
• Test selection: select the assay (A1c or A2/F/A1c or Decont).
• Safety High Pressure limit: limit for the HPPC pump controller. If the
instantaneous pressure is higher than this limit, the pump is stopped.
• psi: pressure unit. The options are psi or kg/cm2.
• Run High Pressure limit: high limit of the pressure during the run. If the
mean value of the pressure is higher than this limit, the run is stopped.
• Meas. Number for High limit average: indicates the number of pressure
values used for the calculation of the mean value, which is compared to the
Run High Pressure limit. The sample rate is approximately 2Hz.
• Run Low Pressure limit: low limit of the pressure during the run. If the mean
value of the pressure is lower than this limit, the run is stopped.
• Meas. Number for Low limit average: indicates the number of pressure
values used for the calculation of the mean value, which is compared to the
Run Low Pressure limit. The sample rate is approximately 2Hz.
Figure 5-39:
D-10 HP module: pump, valve (9/17)
• Load gradient file: selection of the file (*.hpl) for a gradient. The selected file
is loaded into the HPPC controller.
• START TEST or STOP TEST: start/stop the test by running the selected
gradient.
• START PUMP or STOP PUMP: start/stop the High-Pressure Pump, for the
displayed flow rate.
• START PROP. VALVE: start/stop the proportioning valve.
• AUTOMATIC TESTS: launch the automatic WC/FC determination (see
5.10.16.1).
• WC (Waiting Count): value used for the determination. Unit: ° (angle).
• FC (Filling Count): value used for the determination. Unit: ° (angle).
Note: when these WC/FC values are modified, the Load gradient file field is
erased and a gradient file must be selected again.
• Flow rate: flow rate (ml/min) or its translation in pump speed (rpm). The
value is found in the gradient file.
• ml/min: flow rate unit. The options are ml/min or rpm.
• Pressure: current pressure in the high-pressure circuit. The unit of measure
appears in the box next to the value.
• psi: pressure unit. The options are psi or kg/cm2.
• Abs. Meas.: light measurement in the sample channel of the detector. The
result is given in 24-bit converter points (from 0 to 16,777,215).
• Ref. Meas.: light measurement in the reference channel of the detector. The
result is given in 24-bit converter points (from 0 to 16,777,215).
• Log.: it is equal to log10 (reference measurement/sample measurement). The
result is given in Absorbance Unit (AU).
• GRAPHIC: see 5.10.5.
Figure 5-40:
D-10 HP module: pump, valve (10/17)
This screen is used to measure the absorbance of the buffer 2 dye solution in static
conditions (proportioning valve continuously on).
• Selected log file: selection of the file (*.csv) for the logging of the results of
the detector.
• START TEST or STOP TEST: start/stop the test by running the selected
gradient.
• START PUMP or STOP PUMP: start/stop the High-Pressure Pump, for the
displayed flow rate.
• RESULT FILE EXPORT: export a *.csv file to the floppy disk.
• START CALCULATION: start the calculation of the buffer 1 and buffer 2 dye
absorbance, and of the relative absorbance of buffer 2. The calculation is
done on the selected *.csv file.
• buffer 1: mean value for buffer 1 absorbance.
• buffer 2: mean value for buffer 2 dye absorbance.
• Rel.: relative absorbance of buffer 2 vs. buffer 1.
• Test remaining time: remaining time before test completion.
• Flow rate: flow rate (ml/min) or its translation in pump speed (rpm). The
value is found in the gradient file.
• ml/min: flow rate unit. The options are ml/min or rpm.
• Pressure: current pressure in the high-pressure circuit. The unit of measure
appears in the box next to the value.
• psi: pressure unit. The options are psi or kg/cm2.
• Abs. Meas.: light measurement in the sample channel of the detector. The
result is given in 24-bit converter points (from 0 to 16,777,215).
• Ref. Meas.: light measurement in the reference channel of the detector. The
result is given in 24-bit converter points (from 0 to 16,777,215).
• Log.: it is equal to log10 (reference measurement/sample measurement). The
result is given in Absorbance Unit (AU).
• GRAPHIC: see 5.10.5.
Figure 5-41:
D-10 HP module: pump, valve (11/17)
This screen is used to measure the absorbance for 50% and 100 % steps of buffer 2
dye solution (Dilamex gradient) or for 20%, 50% and 100% of buffer 2 dye solution
(Dilamex+ gradient) in dynamic conditions (proportioning valve synchronized with the
HP pump).
• Load gradient file: selection of the file (*.hpl) for a gradient. The selected file
is loaded into the HPPC controller.
• Selected log file: selection of the file (*.csv) for the logging of the results of
the detector.
• START TEST or STOP TEST: start/stop the test by running the selected
gradient.
• WC (Waiting Count): value used for the test. Unit: ° (angle).
• FC (Filling Count): value used for the test. Unit: ° (angle).
Note: when these WC/FC values are modified, the Load gradient file field is
erased and a gradient file must be selected again.
• START PUMP or STOP PUMP: start/stop the High-Pressure Pump, for the
displayed flow rate.
• RESULT FILE EXPORT: export a *.csv file to the floppy disk.
• Test remaining time: remaining time before test completion.
• Flow rate: flow rate (ml/min) or its translation in pump speed (rpm). The
value is found in the gradient file.
• ml/min: flow rate unit. The options are ml/min or rpm.
• Pressure: current pressure in the high-pressure circuit. The unit of measure
appears in the box next to the value.
• psi: pressure unit. The options are psi or kg/cm2.
• Abs. Meas.: light measurement in the sample channel of the detector. The
result is given in 24-bit converter points (from 0 to 16,777,215).
• Ref. Meas.: light measurement in the reference channel of the detector. The
result is given in 24-bit converter points (from 0 to 16,777,215).
Figure 5-42:
D-10 HP module: pump, valve (12/17)
• Selected log file: selection of the file (*.csv) for the results calculation.
• START CALCULATION: start the calculation of the buffer 1, 20% buffer 2
dye, 50% buffer 2 dye and 100% buffer 2 dye absorbance. The calculation is
done on the selected *.csv file.
• Dilamex+: if the box is checked, the Dilamex+ test (0%, 20%, 50% and 100%
steps) is used for the calculations. In the other case, the Dilamex test (0%,
50%, 100% steps) is used.
• Mean Val. (AU): mean value for each step, in Absorbance Unit (AU).
• Rela. Val. (AU): mean value for each step, in relative Absorbance Unit (AU).
• Rela. Val. (%): mean value for each step, in % of the relative Absorbance
Unit.
• 0%: values for buffer 1 (0% buffer 2 dye) absorbance.
• 20%: mean values for the 20% buffer 2 dye absorbance (if the Dilamex+ test
is selected).
• 50%: mean values for the 50% buffer 2 dye absorbance.
• 100%: mean values for the 100% buffer 2 dye absorbance.
• The relative value (in AU) of the 100% buffer 2 dye absorbance is compared
to the 100% buffer 2 relative value obtained under static conditions (see
Figure 5-40). The result of the comparison is given next to the 100% buffer 2
relative value field:
√ indicates that the difference is within tolerance
X indicates that the difference is out of range
? indicates that the static value is not available
• If the Dilamex+ test is selected, the relative value (in %) of the 20% of buffer 2
dye absorbance is compared to the theoretical 20% value. The result of the
comparison is given next to the 20% buffer 2 relative value field:
√ indicates that the difference is within the tolerance
Figure 5-43:
D-10 HP module: pump, valve (13/17)
This screen is used to measure the absorbance for 10% to 100% of buffer 2 dye
solution, by steps of 10% (Step gradient) in dynamic conditions (proportioning valve
synchronized with the HP pump).
• Load gradient file: selection of the file (*.hpl) for a gradient. The selected file
is loaded into the HPPC controller.
• Selected log file: selection of the file (*.csv) for the logging of the results of
the detector.
• START TEST or STOP TEST: start/stop the test by running the selected
gradient.
• WC (Waiting Count): value used for the test. Unit: ° (angle).
• FC (Filling Count): value used for the test. Unit: ° (angle).
Note: when these WC/FC values are modified, the Load gradient file field is
erased and a gradient file must be selected again.
• START PUMP or STOP PUMP: start/stop the High-Pressure Pump, for the
displayed flow rate.
• RESULT FILE EXPORT: export a *.csv file to the floppy disk.
• Test remaining time: remaining time before test completion.
• Flow rate: flow rate (ml/min) or its translation in pump speed (rpm). The
value is found in the gradient file.
• ml/min: flow rate unit. The options are ml/min or rpm.
Figure 5-44:
D-10 HP module: pump, valve (14/17)
• Selected log file: selection of the file (*.csv) for the results calculation.
• START CALCULATION: start the calculation of the Step gradient results.
The calculation is done on the selected *.csv file.
• buffer 1: mean value for buffer 1 relative absorbance.
• 50% buffer 2: mean value for the 50% buffer 2 dye relative absorbance.
• 100% buffer 2: mean value for the 100% buffer 2 dye relative absorbance.
The relative value (in AU) of the 100% buffer 2 dye absorbance is compared
to the 100% buffer 2 relative value obtained under static conditions (see
Figure 5-40). The result of the comparison is given next to the 100% buffer 2
relative value field:
√ indicates that the difference is within tolerance
X indicates that the difference is out of range
? indicates that the static value is not available
• Calc. %: calculated relative % for each step. The relative value (in %) of each
step is compared to the tolerance window. The result of the comparison is
given next to each % value:
√ indicates that the difference is within tolerance
X indicates that the difference is out of range
Figure 5-45:
D-10 HP module: pump, valve (15/17)
This screen is used to measure the absorbance of the buffer 2 dye solution for a
continuous gradient from 0% to 100% of buffer 2 dye in dynamic conditions
(proportioning valve synchronized with the HP pump).
• Load gradient file: selection of the file (*.hpl) for a gradient. The selected file
is loaded into the HPPC controller.
• Selected log file: selection of the file (*.csv) for the logging of the results of
the detector.
• START TEST or STOP TEST: start/stop the test by running the selected
gradient.
• WC (Waiting Count): value used for the test. Unit: ° (angle).
• FC (Filling Count): value used for the test. Unit: ° (angle).
Note: when these WC/FC values are modified, the Load gradient file field is
erased and a gradient file must be selected again.
• START PUMP or STOP PUMP: start/stop the High-Pressure Pump, for the
displayed flow rate.
• RESULT FILE EXPORT: export a *.csv file to the floppy disk.
• Test remaining time: remaining time before test completion.
• Flow rate: flow rate (ml/min) or its translation in pump speed (rpm). The
value is found in the gradient file.
• ml/min: flow rate unit. The options are ml/min or rpm.
• Pressure: current pressure in the high-pressure circuit. The unit of measure
appears in the box next to the value.
• psi: pressure unit. The options are psi or kg/cm2.
• Abs. Meas.: light measurement in the sample channel of the detector. The
result is given in 24-bit converter points (from 0 to 16,777,215).
• Ref. Meas.: light measurement in the reference channel of the detector. The
result is given in 24-bit converter points (from 0 to 16,777,215).
Figure 5-46:
D-10 HP module: pump, valve (16/17)
• Selected log file: selection of the file (*.csv) for the results calculation.
• START CALCULATION: display the Step gradient results. The calculation is
done on the selected *.csv file.
• Visualization of the Continuous Gradient test: Continuous gradient result.
Figure 5-47:
D-10 HP module: pump, valve (17/17)
This screen is used to visualize graphically data recorded in *.log files during
previous tests.
Selecting the AUTOMATIC TESTS button in the HP Module: Pump, Valve (9/17)
screen opens a set of two screens used for an automatic WC/FC determination.
Figure 5-48:
D-10 Automatic Tests (1/2)
This screen is used to link automatically the Static, Dilamex(+), Step and
Continuous gradient tests.
• Static: selection of the Static test in the workflow (first test). The number of
repeats can be entered in the white field on the right side (on the same
horizontal line). If 0 is entered, the test is ignored.
• Dilamex: selection of the Dilamex test. By clicking in the next field, the *.hpl
file can be selected (Dilamex or Dilamex+ test). The number of repeats can
also be entered.
• Step gradient: selection of the Step gradient test. By clicking in the next field,
the *.hpl file can be selected. The number of repeats can also be entered.
• A Step gradient and a Continuous gradient are done for the selected WC/FC
couple.
• The test ends with a 15-minute flush sequence using buffer 1.
The result *.csv files produced during each step of the determination are identified
in the following way:
YYYYMMDD_hhmmss_xxxx_WC-FC.csv
Where:
• "YYYYMMDD" is the file creation date (year, month, day format).
• "hhmmss" is the file creation time (hour, minute, second format).
• "xxxx" is the test name ("stat", "dila+", "step" or "cont").
• "WC-FC" are the WC and FC values used for the test.
Figure 5-49:
D-10 Automatic Tests (2/2)
• WC Min, WC Max, WC Current: Waiting Count. Unit: ° (angle). Min and Max
define the range of WC used for the automatic determination. Current
indicates the WC value currently being tested.
• FC Min, FC Max, FC Current: Filling Count. Unit: ° (angle). Min and Max
define the range of FC used for the automatic determination. Current
indicates the FC value currently being tested.
• SETTING: open the screen for the test settings (see 5.10.16.2).
• START TEST: launch the automatic WC/FC determination process.
• RESULTS: open the screen with the test results for each step of the process
(see 5.10.16.3).
• 100% buf.2 (AU) Static: indicates the relative value (in AU) of the 100%
buffer 2 dye absorbance measured during the Static test.
• 100% buf.2 (AU) Dilamex: indicates the relative value (in AU) of the 100%
buffer 2 dye absorbance measured during the Dilamex+ test.
• Dilamex values (%) 20%: indicates the relative value (in %) of the 20%
theoretical buffer 2 dye step, measured during the Dilamex+ test.
• Dilamex values (%) 50%: indicates the relative value (in %) of the 50%
theoretical buffer 2 dye step, measured during the Dilamex+ test.
• Grad. name: indicates the running gradient file.
• Grad. remaining time: remaining time before the end of the running gradient.
• Abs. Meas.: light measurement in the sample channel of the detector. The
result is given in 24-bit converter points (from 0 to 16,777,215).
• Ref. Meas.: light measurement in the reference channel of the detector. The
result is given in 24-bit converter points (from 0 to 16,777,215).
• Log.: it is equal to log10 (reference measurement/sample measurement). The
result is given in Absorbance Unit (AU).
• Pressure: current pressure in the high-pressure circuit. The unit of measure
appears in the box next to the value.
• psi: pressure unit. The options are psi or kg/cm2.
• GRAPHIC: see 5.10.5.
Figure 5-50:
D-10 WC/FC Determination test setting
This screen is used to set selections for the automatic WC/FC determination.
This screen displays the test results for each step of the WC/FC determination
process, taken from the 'YYYYMMDD_hhmmss_allresults.csv' file.
This screen displays the Static test results, the Dilamex+ results for each WC/FC
couple, with the Status, and the Step gradient results with the Status. It can be
used for the visualisation of any 'YYYYMMDD_hhmmss_allresults.csv' file.
Selecting the HP: SENSOR CALIB. button in the Main Screen (2/2), opens a set of
two screens for calibrating the High-Pressure Sensor.
Figure 5-52:
D-10 HP module: sensor calibration (1/2)
This screen shows the values of the High-Pressure Sensor parameters. These
values are found in the 'instrument.ini' file. These values can be modified and saved
in the corresponding files only if the user has entered the level 2 password.
Figure 5-53:
D-10 HP module: sensor calibration (2/2)
Selecting the HP: FLOW CALIB. button in the Main Screen (2/2), opens a set of four
screens for calibrating the flow rate.
Figure 5-54:
D-10 HP module: flow rate calib. (1/4)
This screen shows the flow rate calibration values. These values are found in the
'instrument.ini' file. These values can be modified and saved in the corresponding
files only if the user has entered the level 2 password.
• rule 1, rule 2, rule 3: correspond to 3 linear segments used for the flow rate
calibration.
• Limit 1: low limit of each linear segment. Unit: ml/min.
• Limit 2: high limit of each linear segment. Unit: ml/min.
• Slope: slope of each linear segment. Unit: rpm.
• Intercept: intercept for each linear segment. Unit: rpm.
• DONE: save the modified parameters. A verification message is displayed to
confirm the modification. The DONE button is usable only if a modification is
made and if the level 2 password was entered.
• CANCEL: cancel the modifications. A verification screen is displayed to
confirm the cancellation. The CANCEL button is usable only if a modification
is made and if the level 2 password was entered.
HP Module: Flow Rate Calibration (2/4): New Calibration, Control and Measurement
Figure 5-55A:
D-10 HP module: flow rate calib. (2/4)
• Flow rate: select the theoretical flow rate (1 ml/min, 1.5 ml/min or 2 ml/min).
The corresponding theoretical speed of the high-pressure pump motor is
indicated in rpm.
• Duration: select the test duration. When the test is launched, the high-
pressure pump is started and is stopped automatically after the selected
duration. Unit: min.
• START PUMP or STOP PUMP: start/stop the high-pressure pump.
• START TEST or STOP TEST: start/stop the test. When the test is started, the
high-pressure pump is switched on, runs at the theoretical selected Flow
rate, and the timer starts. When the test is stopped, the high-pressure pump
is switched off and the timer is frozen.
• Pressure: current pressure in the high-pressure circuit. The unit of measure
appears in the box next to the value.
• psi: pressure unit. The options are psi or kg/cm2.
• GRAPHIC: see 5.10.5.
• Remaining time: remaining time before the end of the test. This counter is
initialized with the Duration value. Unit: min.
• TIMER RESET: set the timer to the Duration value.
• Data entry for calc.: these fields are used to enter the data for the flow rate
calculation
Time: this field indicates the delivery time since the test start. This
value can be edited. If it is edited, the time should be indicated in
seconds.
Volume: the measured volume of the liquid delivered should be
entered here. Unit: ml.
• LAUNCH CALC.: launch the flow rate calculation, based on the Time and
Volume values.
• Data calculation results: ml/min: these 3 fields indicate the calculated flow
rates compared to the theoretical flow rate. Unit: ml/min.
• 1.0: calculated flow rate for the theoretical 1.0ml/min.
• 1.5: calculated flow rate for the theoretical 1.5ml/min.
• 2.0: calculated flow rate for the theoretical 2.0ml/min.
Weight Method
Figure 5-55B:
D-10 HP module: flow rate calib. (2/4)
• Density: indicates the density of the liquid used for the test. This value can be
edited.
• Flow rate: select the theoretical flow rate (1 ml/min, 1.5 ml/min or 2 ml/min).
The corresponding theoretical speed of the high-pressure pump motor is
indicated in rpm.
• Duration: select the test duration. When the test is launched, the high-
pressure pump is started and is stopped automatically after the selected
duration. Unit: min.
• START PUMP or STOP PUMP: start/stop the high-pressure pump.
• START TEST or STOP TEST: start/stop the test. When the test is started, the
high-pressure pump is switched on, runs at the theoretical selected Flow
rate, and the timer starts. When the test is stopped, the high-pressure pump
is switched off and the timer is frozen.
• Pressure: current pressure in the high-pressure circuit. The unit of measure
appears in the box next to the value.
• psi: pressure unit. The options are psi or kg/cm2.
• GRAPHIC: see 5.10.5.
• Remaining time: remaining time before the end of the test. This counter is
initialized with the Duration value. Unit: min.
• TIMER RESET: set the timer to the Duration value.
• Data entry for calc.: these fields are used to enter the data for the flow rate
calculation
Time: this field indicates the delivery time since the test start. This
value can be edited. If it is edited, the time should be indicated in
seconds.
Weight: the measured weight of the liquid delivered should be
entered here. Unit: g.
• LAUNCH CALC.: launch the flow rate calculation, based on the Time and
Weight values.
• Data calculation results: ml/min: these 3 fields indicate the calculated flow
rates compared to the theoretical flow rate. Unit: ml/min.
• 1.0: calculated flow rate for the theoretical 1.0ml/min.
Figure 5-56:
D-10 HP module: flow rate calib. (3/4)
This screen summarizes the measurements done with the previous screen and
calculates the slope and intercept. All the flow rate values can be edited. After
changing of a value, the calculation is automatically done and the new results are
displayed.
• rule 1, rule 2, rule 3: correspond to 3 linear segments used for the flow rate
calibration.
• Theo. flow rate: theoretical flow rate. Can be modified by entering a new
value.
• Meas. flow rate: measured flow rate. Can be modified by entering a new
value.
• Limit 1: low limit of each linear segment. Unit: ml/min.
• Limit 2: high limit of each linear segment. Unit: ml/min.
• Slope: slope of each linear segment. Unit: rpm/ml/min.
• Intercept: intercept for each linear segment. Unit: rpm.
• SAVE RESULTS: store the calibration results in the system ('instrument.ini'
file), and replaces the previous values. This button is available only if a
calculation is done and if the level 2 password was entered.
Figure 5-57A:
D-10 HP module: flow rate calib. (4/4)
• Flow rate: select the target flow rate. The corresponding theoretical speed of
the high-pressure pump motor is indicated in rpm.
• Duration: select the test duration. When the test is launched, the high-
pressure pump is started and is stopped automatically after the selected
duration. Unit: min.
• START PUMP or STOP PUMP: start/stop the high-pressure pump.
• START TEST or STOP TEST: start/stop the test. When the test is started, the
high-pressure pump is switched on, runs at the theoretical selected Flow
rate, and the timer starts. When the test is stopped, the high-pressure pump
is switched off and the timer is frozen.
• Pressure: current pressure in the high-pressure circuit. The unit of measure
appears in the box next to the value.
• psi: pressure unit. The options are psi or kg/cm2.
• GRAPHIC: see 5.10.5.
• Remaining time: remaining time before the end of the test. This counter is
initialized with the Duration value. Unit: min.
• TIMER RESET: set the timer to the Duration value.
• Data entry for calc.: these fields are used to enter the data for the flow rate
calculation
Time: this field indicates the delivery time since the test start. This
value can be edited. If it is edited, the time should be indicated in
seconds.
Volume: the measured volume of the liquid delivered should be
entered here. Unit: ml.
Weight Method
Figure 5-57B:
D-10 HP module: flow rate calib. (4/4)
• Density: indicates the density of the liquid used for the test. This value can be
edited.
• Flow rate: select the target flow rate. The corresponding theoretical speed of
the high-pressure pump motor is indicated in rpm.
• Duration: select the test duration. When the test is launched, the high-
pressure pump is started and is stopped automatically after the selected
duration. Unit: min.
• START PUMP or STOP PUMP: start/stop the high-pressure pump.
• START TEST or STOP TEST: start/stop the test. When the test is started, the
high-pressure pump is switched on, runs at the theoretical selected Flow
rate, and the timer starts. When the test is stopped, the high-pressure pump
is switched off and the timer is frozen.
• Pressure: current pressure in the high-pressure circuit. The unit of measure
appears in the box next to the value.
• psi: pressure unit. The options are psi or kg/cm2.
• GRAPHIC: see 5.10.5.
• Remaining time: remaining time before the end of the test. This counter is
initialized with the Duration value. Unit: min.
• TIMER RESET: set the timer to the Duration value.
• Data entry for calc.: these fields are used to enter the data for the flow rate
calculation
Time: this field indicates the delivery time since the test start. This
value can be edited. If it is edited, the time should be indicated in
seconds.
Selecting the THERMAL CONTROL button in the Main Screen (2/2), opens a set of
seven screens for testing the thermal module.
Figure 5-58:
D-10 Thermal control module (1/7)
Figure 5-59:
D-10 Thermal control module (2/7)
• Peltier module sense: indicates the electrical current intensity that is sensed
through the Peltier effect modules. Unit: points. This function is currently not
available.
• GRAPHIC: see 5.10.5.
Figure 5-60:
D-10 Thermal control module (3/7)
• PWM value: selection of the Pulse Width Modulation value to apply to the
Peltier modules.
• SET PELTIER 100% PWM: apply a 100% Pulse Width Modulation to the
Peltier modules.
• RESET PELTIER 0% PWM: apply a 0% Pulse Width Modulation to the Peltier
modules (modules switched off).
• Peltier module sense: indicates the electrical current intensity that is sensed
through the Peltier effect modules. Unit: points. This function is currently not
available.
• GRAPHIC: see 5.10.5.
Figure 5-61:
D-10 Thermal control module (4/7)
This screen shows the values of the parameters of the thermal module. These values
are found in the 'instrument.ini' file. These values can be modified and saved in the
corresponding files only if the user has entered the level 2 password.
Figure 5-62:
D-10 Thermal control module (5/7)
Figure 5-63:
D-10 Thermal control module (6/7)
Figure 5-64:
D-10 Thermal control module (7/7)
Selecting the SAMPLE ARM button in the Main Screen (2/2), opens a screen for
testing the sample arm module.
Figure 5-65:
D-10 Sample arm module
This screen gives access to different functions of the sample arm module:
• TH MOVE: test the TH movement
• Z MOVE: test the Z movement
• X MOVE: test the X movement
• BARCODE: test the barcode reader
• ALL STATUS: visualize the status of all components
• ARM STATUS: visualize the status of all sample arm components
• PROBE REPLACEMENT: replace the sample probe
• DILUT./WASH STAT. REPLA.: replace the dilution/wash station
• SAMPLE ARM INITIALIZATION: move the sample arm to its home position
• HALL SENSOR TEST: test the Hall effect adapter sensor
• TUBE IDENTIFICATION: test the tube height sensing
Selecting the TH MOVE button in the Sample Arm Module screen opens a set of
four screens for testing the TH movement.
Figure 5-66:
D-10 Sample arm module: TH (1/4)
• Home movement: if checked, the tube holder will move to its home position.
Figure 5-67:
D-10 Sample arm module: TH (2/4)
Figure 5-68:
D-10 Sample arm module: TH (3/4)
This screen shows the value of the TH movement parameter. This value is found in
the 'instrument.ini' file. It can be modified and saved in the 'instrument.ini' file only if
the user has entered the level 2 password.
• All movements speed index: TH movement speed selection for service use.
Index from 0 (low speed) to 9 (high speed).
• TH adjustment offset: mechanical offset for the TH movement. Unit: steps.
Figure 5-69:
D-10 Sample arm module: TH (4/4)
This screen shows the value of the TH movement speed parameter for several
sequences in the 'Dia.exe' software. These values are found in different *.ini files.
They can be modified and saved in the corresponding *.ini file only if the user has
entered the level 2 password. The speed is defined by an index from 0 (low speed)
to 9 (high speed).
Selecting the Z MOVE button in the Sample Arm Module screen opens a set of four
screens for testing the Z movement.
Figure 5-70:
D-10 Sample arm module: Z (1/4)
• Home movement: if checked, the probe will move to its home position.
• To sample aspiration position (tube): if checked, the probe will move to the
primary tube sampling position for the selected assay and the selected tube
number.
• To sample aspiration position (microvial): if checked, the probe will move
to the microvial sampling position for the selected assay and the selected
tube number.
• START: start the selected movement.
• Test selection: select the assay (A1c or A2/F/A1c or Decont). This selection
defines the parameters for the probe movement.
• Tube position: define the position of the selected movement (from 1 to 10).
• Z step counter: Z step counter. Reset when a home movement is done.
• Home sensor: status of the Z home sensor
obstructed: the optical sensor is "off". Infra-red light from the emitter
cannot reach the receiver because the home flag is in the optical
sensor slot.
free: the optical sensor is "on". Infra-red light from the emitter reaches
the receiver.
• Axis sensor: status of the Z axis (position) sensor
obstructed: the optical sensor is "off". Infra-red light from the emitter
cannot reach the receiver because of the position of the slotted disk.
free: the optical sensor is "on". Infra-red light from the emitter reaches
the receiver because of the slot on the disk.
Figure 5-71:
D-10 Sample arm module: Z (2/4)
Figure 5-72:
D-10 Sample arm module: Z (3/4)
This screen shows the value of the Z movement parameter. This value is found in
the 'instrument.ini' file. It can be modified and saved in the 'instrument.ini' file only if
the user has entered the level 2 password.
• All movements speed index: Z movement speed selection for service use.
Index from 0 (low speed) to 9 (high speed).
• Z adjustment offset: mechanical offset for the Z movement. Unit: steps.
Figure 5-73:
D-10 Sample arm module: Z (4/4)
This screen shows the value of the Z movement speed parameter for several
sequences in the 'Dia.exe' software. These values are found in different *.ini files.
They can be modified and saved in the corresponding *.ini file only if the user has
entered the level 2 password. The speed is defined by an index from 0 (low speed)
to 9 (high speed).
Selecting the X MOVE button in the Sample Arm Module screen opens a set of four
screens for testing the X movement.
Figure 5-74:
D-10 Sample arm module: X (1/4)
• Home movement: if checked, the sample arm will move to its home position.
• To probe wash position: if checked, the sample arm will move to the probe
wash position.
• To tube: if checked, the sample arm will move to the selected tube number.
• To rack door opening position: if checked, the sample arm will move to the
position used to open the rack door.
• START: start the selected movement.
• X step counter: X step counter. Reset when a home movement is done.
• Home sensor: status of the X home sensor
obstructed: the optical sensor is "off". Infra-red light from the emitter
cannot reach the receiver because the home flag is in the optical
sensor slot.
free: the optical sensor is "on". Infra-red light from the emitter reaches
the receiver.
• Axis sensor: status of the X axis (position) sensor
obstructed: the optical sensor is "off". Infra-red light from the emitter
cannot reach the receiver because of the position of the slotted disk.
free: the optical sensor is "on". Infra-red light from the emitter reaches
the receiver because of the slot on the disk.
• Hall effect sensor: detection of a magnet in front of the sensor
magnet
no magnet
Figure 5-75:
D-10 Sample arm module: X (2/4)
• Direction: select the movement direction (from left to right or from right to
left).
• Movement select: select the kind of movement
steps: move for the selected number of steps.
strips: move for the selected number of strips (see 2.2.7).
sensors: move until the sensor configuration defined on the screen is
met.
• START: start the selected movement.
• selection: when the sensors movement is selected, it defines the sensor(s)
to be used as final position.
• target: when the sensors movement is selected, it defines the final status for
each selected sensor:
obstructed: the optical sensor is "off". Infra-red light from the emitter
cannot reach the receiver because the home flag is in the optical
sensor slot.
free: the optical sensor is "on". Infra-red light from the emitter reaches
the receiver.
• current: indicates the current status of each sensor:
obstructed: the optical sensor is "off". Infra-red light from the emitter
cannot reach the receiver because the home flag is in the optical
sensor slot.
free: the optical sensor is "on". Infra-red light from the emitter reaches
the receiver.
• Home sensor, Axis sensor: sensors used for the X movement.
• Hall effect sensor: current status of the magnet detection in front of the
sensor
magnet
no magnet
• Power on motor: select the motor supply
run (full power)
hold (low power)
free (no supply)
• APPLY: apply the selected power to the motor (only if the motor is standby
state).
Figure 5-76:
D-10 Sample arm module: X (3/4)
• All movements speed index: X movement speed selection for service use.
Index from 0 (low speed) to 9 (high speed).
Figure 5-77:
D-10 Sample arm module: X (4/4)
This screen shows the value of the X movement speed parameter for several
sequences in the 'Dia.exe' software. These values are found in different *.ini files.
They can be modified and saved in the corresponding *.ini file only if the user has
entered the level 2 password. The speed is defined by an index from 0 (low speed)
to 9 (high speed).
Selecting the BARCODE button in the Sample Arm Module screen opens a set of
two screens for testing the barcode reader.
Figure 5-78:
D-10 Sample arm module: barcode (1/2)
• Tube position: select the tube number for the barcode reading (from 1 to 10).
• GO TO TUBE: move the sample arm to the selected Tube position.
• BARCODE LED ON or BARCODE LED OFF: switch on/off the red LED
source of the barcode reader.
• Last code read: indicates the last barcode read by the barcode reader.
• CLEAR LAST CODE: clear the Last code read field.
• RACK INTRODUCTION: shortcut to the Rack Introduction (1/1) screen.
This screen is used to test the barcode reading on several sample tubes in a rack.
• Result of reading: indicates, for each tube, the barcode read during the test.
• X pos.: indicates the exact X position (in steps from the home position) at
which the barcode was read.
• Tube nb: the entire part of the number indicates the tube number identified by
the software. The decimal part of the number indicates the range for each
tube. The nominal range is from 0.2 to 0.8. Next to the field, there is an
indication of the result:
✔ indicates that the result is in the standard tube range
✘ indicates that the result is out of range
• START or STOP: start/stop the reading of the rack previously introduced in
the instrument.
• CLEAR: clear the Result of reading table.
Selecting the SAMPLE ARM STATUS button in the Sample Arm Module screen
opens a set of two screens that show the sample arm components status. In
addition to the sensors status, the screens indicate the logical level at the input of
the MC.
Figure 5-80:
D-10 Sample arm module: status (1/2)
Figure 5-81:
D-10 Sample arm module: status (2/2)
• TH step counter, Z step counter, X step counter: last values of the step
counters. Reset when a home movement is done.
• TH adjustment offset: mechanical offset for the TH movement. Unit: steps.
• Z adjustment offset: mechanical offset for the Z movement. Unit: steps.
Selecting the PROBE REPLACEMENT button in the Sample Arm Module screen
starts the probe replacement sequence. See D-10 Operation Manual, Section 6.3.
Selecting the DILUT./WASH STAT. REPLA. button in the Sample Arm Module
screen starts the dilution/wash station replacement sequence. See D-10 Operation
Manual, Section 5.2.
Selecting the SAMPLE ARM INITIALIZATION button, in the Sample Arm Module
screen starts the Z, TH and X home movements.
Selecting the HALL SENSOR TEST button in the Sample Arm Module screen
opens a screen for testing the Hall effect adapter sensor.
Figure 5-82:
D-10 Sample arm module: Hall effect (1/1)
• First tube: select the position of the first tube to be tested (from 1 to 9).
• Last tube: select the position of the last tube to be tested (from 1 to 9).
• START: start the Hall sensor test. The sample arm moves close to the slot
which is just before the slot of the First tube (-90 position in the first column),
checks the Hall effect sensor and displays the test result. It then moves 10
steps towards the right (-80 position in the first column), checks again and
displays the result. This sequence is repeated until the slot corresponding to
the First tube (0 position in the second column). The arm then moves to the
slot of the Last tube (0 position in the third column), checks the Hall effect
sensor and displays the test result. It then moves 10 steps towards the right
(+10 position in the third column), checks again and displays the result. This
sequence is repeated until 90 steps after the slot corresponding to the Last
tube (+90 position in the fourth column). The arm then goes back to its home
position.
• magnet: indicates the test results
no: no magnet is detected.
yes: a magnet is detected.
Selecting the TUBE IDENTIFICATION button in the Sample Arm Module screen
opens a set of three screens for testing the tube identification.
Figure 5-83:
D-10 Sample arm mod.: Tube Ident. (1/3)
This screen shows the values of the parameters used for the tube identification.
Some parameters are based on the selected assay. These values are found in the
*.ini files. These values can be modified and saved in the corresponding files only if
the user has entered the level 2 password.
Figure 5-84:
D-10 Sample arm mod.: Tube Ident. (2/3)
• START: start the TH offset measurement process. The arm moves from its
home position to tube position 1 and senses the tube height for an empty
rack. The result is displayed when the arm comes back to its home position.
The same sequence is then repeated for tube positions 5 and 10.
• TH result position 1, TH result position 5, TH result position 10: indicates
the result of the tube height sensing. Unit: steps.
• Calculated mean value: mean value of the three tube height measurements.
Unit: steps.
• Theoretical TH result: theoretical value of the tube height sensing for an
empty rack (7775 steps).
• TH Offset: current TH offset value and new calculated value. Unit: steps.
• SAVE RESULTS: store the result of the measurement in the system
('instrument.ini' file), and replaces the previous value. This button is available
only if a new determination is made and if the level 2 password was entered.
• tube1, tube2, tube3… tube10: each tube number can be selected for the
test by checking the corresponding box.
• all: automatically select all tube numbers (from 1 to 10).
• START: start the tube identification test. The arm moves from its home
position to each selected tube position and senses the tube height. The
measured value is corrected with the TH offset. The results for each tube are
displayed when the arm comes back to its home position.
• A1c or A2/F/A1c or Decont: select the assay. The tube height sensing speed
is defined for each assay.
• Template: select a *.tpl (template) file containing the awaited test result.
Three *.tpl files are available in the \Dia\Service\Template\VnV folder. The
'Rack1_VnV.tpl' file corresponds to the TH movement adjustment tool (P/N:
2200432). Two other files 'bad_VnV.tpl' and 'Rack2_VnV.tpl' are available as
examples.
• TH result: indicates the result of the tube height sensing. Unit: steps.
• corr. result: indicates the result of the tube height sensing corrected with the
TH offset.
• magnet: indicates if a magnet was detected (yes) or not (no).
• tube: indicates the tube identification result:
no tube
microvial
primary tube
Next to the field, there is an indication of the final tube identification result
(comparison with the awaited result in the selected *.tpl file – when a *.tpl file
is selected):
✔ indicates a correct result.
✘ indicates a difference.
Selecting the MISC. button in the Main Screen (2/2), opens a set of 2 screens for
additional functions.
Miscellaneous (1/2)
Figure 5-86:
D-10 Miscellaneous (1/2)
• FILE IMPORT: open an IMPORT screen (see Figure 5-2B) in order to import
a file from a floppy disk. The file import is only possible if the user has entered
the level 2 password.
• FILE EXPORT: open an EXPORT screen (see Figure 5-2A) in order to export
a file to a floppy disk.
• SETUP: launch an application from a floppy disk (firmware upload, parameter
update…). It is only possible if the level 2 password was entered (see
5.10.21.1).
• LANGUAGE: set up the language parameters (see 5.10.21.2).
• LIS: set up the LIS parameters (see 5.10.21.3).
• RS232 TEST: test the RS232 port.
• COUNTERS: visualize the injection counter (see 5.10.21.4).
• REPORT: set up the patient report (see 5.10.21.5).
Miscellaneous (2/2)
Figure 5-87:
D-10 Miscellaneous (2/2)
YYYYMMDDhhmmss.bmp or YYYYMMDDhhmmss.jpg
Where:
"YYYYMMDD" is the file creation date (year, month, day format)
"hhmmss" is the file creation time (hour, minute, second format)
All of these files are stored in the \Dia\Service\ScreenShot folder.
Selecting the SETUP button in the Miscellaneous (1/2) screen opens a screen for
launching an application from a floppy disk.
Figure 5-88:
D-10 Setup
Selecting the LANGUAGE button in the Miscellaneous (1/2) screen opens a screen
for setting up the language parameters.
Figure 5-89:
D-10 Language
• User software (DIA) language: select the language used in the 'Dia.exe'
application. Currently available options include English, Japanese, and
Français.
• Setting of the languages: set up three additional parameters linked to each
language (Pressure unit, Date format and Time format).
• Language: select the language for which the parameters will be edited.
• Pressure unit: choose the pressure unit. Four options are available: Kg/cm2,
Psi, Mpa or Bar.
• Date format: choose the date format. Three options are available:
Month/Day/Year, Year/Month/Day or Day/Month/Year.
• Time format: choose the time format. Two options are available: 0-12h
AM/PM or 0-24h.
Selecting the LIS button in the Miscellaneous (1/2) screen opens a screen for
setting up the LIS parameters.
Figure 5-90:
D-10 LIS
Selecting the COUNTERS button in the Miscellaneous (1/2) screen opens a screen
for visualizing the injection counter.
Figure 5-91:
D-10 Counters (1/1)
• Injection counter: indicates the injection number since the beginning of the
instrument "life". This value cannot be edited.
Selecting the REPORT button in the Miscellaneous (1/2) screen opens a screen for
setting up the patient report.
Figure 5-92:
D-10 Report
• Add Reference values to the printout of the patient report (only for the
A1c test): print/don't print the reference values and the associated
information (found in the 'Profile2.ini' file) on the patient test report.
• DONE: save the modified parameter. A verification message is displayed to
confirm the modification. The DONE button is usable only if a modification is
made and if the level 2 password was entered.
• CANCEL: cancel the modification. A verification screen is displayed to
confirm the cancellation. The CANCEL button is usable only if a modification
is made and if the level 2 password was entered.
Selecting the TOUCH SCREEN CALIB. button in the Miscellaneous (1/2) screen
opens a screen for calibrating the touch screen.
Figure 5-93:
D-10 Touch screen calibration: warning
message
Figure 5-94:
D-10 Touch screen calibration: first step
• DONE: record the positions of the 9 crosses at the end of the process. If this
button is used before clicking on the last cross, the process is aborted and
returns to the Miscellaneous (1/2) screen.
• CANCEL: abort the touch screen calibration process and return to the
Miscellaneous (1/2) screen.
Figure 5-95:
D-10 Touch screen calibration: final step
• DONE: record the new values of the touch screen calibration and return to the
Miscellaneous (1/2) screen.
• CANCEL: abort the touch screen calibration process and return to the
Miscellaneous (1/2) screen.
Selecting the MACRO button in the Main Screen (2/2), opens a set of 2 screens for
testing several high level functions.
Macro (1/2)
Figure 5-96:
D-10 Macro functions (1/2)
Note: the field near the bottom of the screen is used to display information about
the current step of the running sequence.
Macro (2/2)
Figure 5-97:
D-10 Macro functions (2/2)
Selecting the VERSIONS button in the Main Screen (2/2), opens a screen showing
the installed versions of software and firmware.
Figure 5-98:
D-10 Service software version (1/1)
Selecting the ERRORS button in the Main Screen (2/2), opens a screen showing the
last errors on the instrument. The information is found in the 'Dialog.txt' file.
Figure 5-99:
D-10 Last Errors (1/1)
The following table gives the default values of the most important process
parameters, which were previously described. These values refer to Dia software
version 3.0B.
The following table gives the typical values of the sensed information (see 5.10.5)
available for motors, solenoids…
6 REPLACEMENT PROCEDURES
Unless otherwise specified, the level 2 password should always be used to enter the
Service software in the following procedures.
Passwords:
Level 1: D-10QC
Level 2: DIA904
Level 3: D-10SETUP
Warning: when using the 50mL syringe to prime the high-pressure circuit, do not
pull excessively on the plunger as it may damage the proportioning valve.
For a detailed view of the Service software screens used in the following procedures,
please refer to Section 5.
1) Shut down the D-10, turn off the power and unplug the external power cord.
2) Unscrew the 2 ¼ turn fasteners which lock the left door and open
it.
6) Disconnect the 2 security cables and completely remove the old front cover assembly.
7) Connect the 2 security cables coming from the new front cover assembly.
8) Install the new front cover assembly and secure it with the 4 screws and 4 nylon washers.
9) Close the side doors and tighten the 2 ¼ turn fasteners that lock the left door.
1) Shut down the D-10, turn off the power and unplug the external power cord.
6) Connect the 2 plugs to the new thermal printer and install it in the plastic holder.
12) Click on YES to start the printing. If the test page isn't
printed correctly or the text isn't legible, go back to step
3) with a new thermal printer (P/N: 2200225).
7) Replace the metal bottle support and lock the ¼ turn fastener.
8) Slide the metal top cover back on and lock the 2 ¼ turn fasteners.
10) Replace the 3 reagent bottles. The correct positioning order for the bottles is indicated on the
metal top cover:
• 1 = buffer 1
• 2 = buffer 2
• W = wash/diluent
2) Shut down the D-10, turn off the power and unplug the external power cord.
10) Connect the flat cable to the new syringe module board.
11) Install the new syringe module and secure it using the 4 screws, 4 flat washers and 4 spring
lock washers. Replace the door brace if it was previously removed.
14) Re-connect the blue knurled fitting to the 3-way valve (tube coming from the degasser).
20) Check for leaks on the new syringe module while the pump is running. Make sure that liquid
is flowing through the waste circuit. After approximately 5 minutes, click on Stop Pump.
1) Shut down the D-10, turn off the power and unplug the external power cord.
3) Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
11) Disconnect all cables from the thermal printer. Remove the printer.
14) Hold the electronic module with one hand and remove
the 4 screws and 4 spring lock washers that secure the
module to the frame.
15) Remove the old electronic module from the left side of the instrument.
Each interface board has a sticker on it, giving the board's specific parameters. Note these
values; they must be entered in the 'instrument.ini' file once the replacement is done.
16) Install the new electronic module and secure it with the 4 screws and 4 spring lock washers.
17) Replace the printer and re-connect the mini-jack and 6-pin DIN plugs.
18) Re-connect the 2 flat cables between the PC motherboard, the floppy drive and the hard disk.
20) Re-connect all cables and flat cables to the interface board. Note that there is only one
possible connection for each cable and flat cable.
21) Close the left door and tighten the 2 ¼ turn fasteners.
22) Replace the grid and secure it with the 2 screws, 2 spring lock washers and 2 flat washers.
5) Open the cartridge module by pulling the black metal cover all the way to the left.
Each spare cartridge module has a sticker on it, giving the module's specific parameter.
Note this value; it must be entered in the 'instrument.ini' file once the replacement is done.
10) Install the new cartridge module and secure it using the 4 screws and 4 spring lock washers.
Make sure that the ground connection is in place.
11) Connect the thermal sensor (labeled '4') to the HPLC module board (CN4).
12) Connect the 2 Peltier modules and the thermal fuse to the HPLC module board. Note that
these 3 plugs are fully interchangeable.
13) Re-connect the tube coming from the detector (beige PEEK fitting).
16) On the HP pump head, remove the plug and re-connect the beige PEEK fitting.
17) Push the HPLC drawer back in and tighten the 2 captive screws.
22) Check for leaks while the pump is running. After approximately 5 minutes, click on STOP
PUMP.
1) Replace the probe according to the procedure in Section 5 of the Operation Manual. Make
sure the old probe is decontaminated prior to removal.
1) Verify that the system is in Sleep state. If the system is in StandBy, place it in Sleep state
using the RUN screen.
3) Open the upper front door. If the door is not opened within 10 seconds, it will automatically
lock again.
5) With the lever still depressed, lift the dilution/wash station a few mm and unscrew the 2 LUER
fittings.
7) Connect the 2 silicon tubings to the new dilution/wash station, matching the colors.
8) Pull the new dilution/wash station down until you hear a 'click' coming from the lever. Try to
lift the dilution/wash station again, in order to make sure that it is secured properly.
9) Close the upper front door. The sample arm will automatically return to its home position.
To release the cables coming from the motorized syringe, it may be necessary to remove
the syringe module board. (Section 6.51)
8) Install the new 3-way valve and secure it with the 2 screws and 2 spring lock washers.
10) Replace the syringe support and secure it with the 4 screws and 4 spring lock washers.
13) Re-connect the tube coming from the motorized syringe to the bottom port of the 3-way
valve.
14) Connect the new 3-way valve supply cable (labeled '5') to the syringe module board (CN5).
15) Re-connect the 2 cables coming from the motorized syringe (unlabeled) to the syringe
module board (CN2 and CN3).
16) Put the 3 cables in the metal cable holder and close it.
To release the cables coming from the motorized syringe, it may be necessary to remove
the syringe module board. (Section 6.51)
6) Pass the 2 cables of the new motorized syringe through the hole in the support and through
the grommet.
7) Install the new motorized syringe and secure it with the 2 screws and 2 spring lock washers.
9) Re-connect the tube coming from the 3-way valve to the top of the motorized syringe (black
knurled fitting).
10) On the rear side of the syringe module, connect the 2 cables coming from the motorized
syringe (unlabeled) to the syringe module board (CN2 and CN3).
11) Put the 3 cables in the metal cable holder and close it.
4) Apply thermal grease on the inner surface of the housing and push the new thermal fuse into
the housing.
5) Put the cable in the metal cable holder and close it.
1) Shut down the D-10, turn off the power and unplug the external power cord.
11) Connect the flat cable to the new degasser module board.
12) Install the new degasser and secure the captive screw.
13) Replace the 4 screws, 4 spring lock washers and 4 flat washers at the rear.
14) Re-connect the 3 color-coded fittings at the degasser. The correct order is (from top to
bottom):
• green
• white
• blue
15) Replace the metal plate and the 2 white plastic knurled nuts.
16) Connect the new silicon tubing (coming from the vacuum pump) to the air output.
19) Re-connect the white, green and blue fittings at the rear (bottom left corner). Be careful to
MATCH THE COLORS.
7) On the old pump, remove the remaining silicon tubing and connect it to the corresponding
nozzle of the new vacuum pump.
8) Install the new pump on the bracket and secure it with the 2 screws.
10) Slide the plastic cable tie along the vacuum pump back to its original position.
To protect electronic cards against damage from static electricity, several precautions must
be taken:
• use a grounded wrist strap before handling the components. If one is not
available, first touch both of your hands to a safely grounded or metal object
• hold components by the edges and try not to touch the IC chips, leads,
connectors, or other components
• place components on a grounded antistatic pad or on the bag that came with the
components whenever they are separated from the system
1) Shut down the D-10, turn off the power and unplug the external power cord.
2) Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
6) Install these EPROMs on the new interface board. Be careful NOT TO SWAP THE 2
EPROMs.
Each spare interface board has a sticker on it, giving the board's specific parameters. Note
these values; they must be entered in the 'instrument.ini' file once the replacement is done.
9) Re-connect all cables and flat cables. Note that there is only one possible connection for
each cable and flat cable.
10) Close the left door and tighten the 2 ¼ turn fasteners.
3) Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
8) Remove the rack door module. Be mindful of the cables, as there are several metal cable
holders on their way to the X board.
9) Install the new rack door module and secure it to the frame using the 4 screws and 3 spring
lock washers.
10) Lead the 3 cables to the X board and connect them at CN2, CN7 and CN12. Re-connect the
flat cable to the board if it was previously removed.
13) Close the side doors and tighten the 2 ¼ turn fasteners that lock the left door.
15) Go to the RACK DOOR (1/2) screen to check that the new
rack door is working properly:
• test the solenoid
• open and close the rack door
• monitor the door and the sensors' status
9) Install the new valve and secure it with the 2 screws and 2 spring lock washers.
10) Connect the new proportioning valve cable to the HPLC module board (CN15).
11) Re-connect the black, white and green fittings. Be careful to MATCH THE COLORS.
12) Push the HPLC drawer back in and tighten the 2 captive screws.
8) Put the new HP pump cables in position; one of them should pass through the grommet.
Each spare HP pump has a sticker on it, giving the pump's specific parameters (Waiting
Count and Filling Count). Note these values; they must be entered in the 'instrument.ini' file
once the replacement is done.
9) Install the new pump and secure it with the 4 screws, 4 spring lock washers and 4 flat
washers.
10) Connect the 3 pump cables to the HPLC module board (CN8, CN9, CN16).
11) Re-connect all of the tubing to the pump head. Be careful to connect the tube coming from
the proportioning valve ONLY AT THE END.
12) Push the HPLC drawer back in and tighten the 2 captive screws.
7) Pass the new HP valve cables through the hole in the bracket.
8) Install the new HP valve and secure it with the 4 screws and 4 spring lock washers.
9) Re-connect all of the tubing to the stator (making sure that the fittings are in their proper
position numbers) and the HP pump head.
11) Push the HPLC drawer back in and tighten the 2 captive screws.
17) Check for leaks while the pump is running. After approximately 5 minutes, click on STOP
PUMP.
18) Go back to the HP: PUMP + VALVE (4/17) screen and move
the valve to inject position.
20) Check for leaks while the pump is running. After approximately 5 minutes, click on STOP
PUMP.
21) Go to the HP: PUMP + VALVE (4/17) screen and move the
valve to load position.
The proportioning valve and the pressure sensor are mounted on a small bracket, which is
secured to the HPLC drawer by 4 Nylock nuts.
11) Connect the 2 beige PEEK fittings to the new pressure sensor:
• the tube coming from the HP valve should be connected at the top (beveled face)
• the tube coming from the HP pump head should be connected at the bottom
Each spare pressure sensor has a sticker on it, giving the sensor's specific parameters.
Note values; they must be entered in the 'instrument.ini' file once the replacement is done.
12) Install the new sensor and secure it with the 2 screws, 2 spring lock washers and 2 flat
washers. Be careful NOT TO MOUNT THE SENSOR UPSIDE DOWN. The beveled face
must be oriented upwards.
13) Pass the new sensor cable through the grommet and connect it to the module board (CN6).
15) Re-connect the black, white and green fittings. Be careful to MATCH THE COLORS.
16) Push the HPLC drawer back in and tighten the 2 captive screws.
5) Grasp the 2 standoffs and pull out the column heater coil
support, along with the fixed cartridge support. Slightly
moving the column heater coil from left to right and from
top to bottom, while pulling it out, will facilitate removal.
Use caution, as the 2 parts are still connected together.
8) Replace the fixed cartridge support and the column heater coil support and secure them
using the 3 screws and 3 spring lock washers. There's no need to add thermal grease in the
housing.
9) Replace the black metal cover back and secure it with the 4 screws and 4 spring lock
washers.
11) On the HP valve, connect the stainless steel tubing coming from the cartridge module to port
3 (stainless steel fitting).
12) On the HP pump head, re-connect the beige PEEK fitting (tube coming from the pressure
sensor).
13) Push the HPLC drawer back in and tighten the 2 captive screws.
17) Check for possible leaks while the pump is running. After approximately 5 minutes, click on
Stop Pump.
1) Shut down the D-10, turn off the power and unplug the external power cord.
7) Hold the Z motor with one hand and remove the 2 screws, 2 spring lock washers and 2 flat
washers.
9) On the Z motor:
• unlock – but do not remove – the 2 screws, 2 spring
lock washers and 2 flat washers that secure the
metal plate
• swivel the plate to have access to the sensor
• remove the old Z position sensor
11) Install the new Z position sensor on the metal plate and secure it with the 2 nuts and 2
serrated lock washers.
12) Swivel the metal plate back in its original position and tighten the 2 screws.
13) Connect the new Z position sensor cable to the Z board (CN13).
14) Tilt the Z motor and replace it on its support. During the positioning of the motor, don't forget
to slide the shaft into the pulley.
15) Keep the Z motor in position with the 2 screws, 2 spring lock washers and 2 flat washers –
but do not tighten the screws.
17) Tighten the pulley screw. For the exact position of the pulley on the shaft, follow the example
of the TH motor.
18) Adjust the tension of the belt by sliding the Z motor in the oblong holes and tighten the 2 Z
motor screws. Follow the example of the TH motor.
22) Use the Z movement control tool (P/N: 2200431) to verify the
setting of the Z movement. If it is OK, go to step 25). If it isn't,
continue with step 23).
Place the Z movement control tool in contact with the TH movement adjustment tool. Verify
that the probe support is between the 1st stage (included) and the 3rd stage (excluded). The
ideal situation is when the probe support touches the 2nd stage.
24) Verify the setting of the Z movement again, as described in steps 21) and 22).
8) Replace the LCD display and secure it with the 4 screws and 4 spring lock washers.
3) Remove the old power supply 2 from the electronic module (2 plastic standoffs in the bottom
left and top right corners).
4) Install the new power supply 2 on the 2 standoffs and secure it with the 2 screws, 2 spring
lock washers and 2 flat washers.
6) Remove the old power supply 1 from the electronic module (2 plastic standoffs in the bottom
left and top right corners).
7) Install the new power supply 1 on the 2 standoffs and secure it with the 2 screws, 2 spring
lock washers and 2 flat washers.
8) Replace the electronic module and secure it with the 4 screws and 4 spring lock washers.
10) Pull out the thermal printer and re-connect the mini-jack and 6-pin DIN plugs.
12) Re-connect the 2 flat cables going from the PC motherboard to the floppy drive and the hard
disk.
14) Re-connect all cables and flat cables to the interface board. Note that there is only one
possible connection for each cable and flat cable.
15) Close the left door and tighten the 2 ¼ turn fasteners.
Before being able to use the new power supply 1 (bottom), it is necessary to set its output
voltage to 5.15 V. The test points are located on the interface board:
5.15 VDC
GND
17) Replace the protective grid and secure it with the 2 screws, 2 spring lock washers and 2 flat
washers.
To protect electronic cards against damage from static electricity, several precautions must
be taken:
• use a grounded wrist strap before handling the components. If one is not
available, first touch both of your hands to a safely grounded or metal object
• hold components by the edges and try not to touch the IC chips, leads,
connectors, or other components
• place components on a grounded antistatic pad or on the bag that came with the
components whenever they are separated from the system
9) Disconnect:
• the VGA connector link cable
• the RS422 & RS232 connectors link
cable
10) Disconnect:
• the hard disk flat cable
• the floppy drive ribbon cable
• the mouse & keyboard connectors
link cable
11) Remove the 11 screws and 11 flat washers that secure the motherboard to the module.
CAREFULLY remove the old motherboard.
12) CAREFULLY remove the Dual Inline Memory Module (DIMM) from the old motherboard and
insert it on the new one. Because the number of pins is different on either side of the break,
the module will only fit in one orientation.
15) Remove the processor from the old motherboard and install it on the new board.
Locate the socket and open it by pulling the lever upwards. The socket lever must be fully
open (90° to 100°) when removing and installing the processor. Insert the processor with
the correct orientation, knowing that there is a notched corner. DO NOT EXERT TOO MUCH
FORCE, but press the processor down gently into the slot.
Once completely inserted, close the socket lever while holding down the processor.
16) Install the heatsink on the new motherboard. It is not absolutely necessary to apply thermal
grease on the processor.
20) Install the new motherboard on the electronic module and secure it with the 11 screws and 11
flat washers. DO NOT OVERTIGHTEN.
21) CAREFULLY replace the LCD internal board and secure it with the screw, flat washer and
spacer. DO NOT OVERTIGHTEN.
22) Re-connect all cables and flat cables to the motherboard. Note that there is only one possible
connection for each cable and flat cable.
23) Re-connect the 2 flat cables and the supply cable to the LCD internal board.
24) Re-connect the link cable between the motherboard and the LAN connector.
25) Replace the interface board support and secure the 2 captive screws.
26) Re-connect the flat cable located on the top end of the interface board.
To protect electronic cards against damage from static electricity, several precautions must
be taken:
• use a grounded wrist strap before handling the components. If one is not
available, first touch both of your hands to a safely grounded or metal object
• hold components by the edges and try not to touch the IC chips, leads,
connectors, or other components
• place components on a grounded antistatic pad or on the bag that came with the
components whenever they are separated from the system
3) Remove the 4 screws, 4 spring lock washers and 4 flat washers that secure the LCD
interface board to the user interface module. Remove the board.
4) Connect all cables and flat cables to the new LCD interface board.
5) Install the new board on the module and secure it with the 4 screws, 4 spring lock washers
and 4 flat washers.
4) Connect the backlight supply cable of the new LCD display and the 2 flat cables.
5) Install the new LCD display and secure it with the 4 screws and 4 spring lock washers.
1) Shut down the D-10, turn off the power and unplug the external power cord.
7) Secure the new manifold using the 2 thumbscrews and 2 nylon washers. The screws should
be finger-tight ONLY. DO NOT OVERTIGHTEN.
1) Shut down the D-10, turn off the power and unplug the external power cord.
6) Install the new waste manifold and secure it using the 2 screws, 2 spring lock washers and 2
flat washers.
11) Check for leaks on the new waste manifold while the pump is running. Make sure that liquid
is flowing through the waste circuit. After approximately 5 minutes, click on Stop Pump.
1) Shut down the D-10, turn off the power and unplug the external power cord.
8) Unscrew the 2 ¼ turn fasteners which lock the left door and open
it.
10) On the right side of the printer module, pull the flat
cable out of the plastic cable holder.
13) Install the new hard disk and secure it with the 4 screws and 4 serrated lock washers.
Confirm that the new hard disk IDE jumper is set to 'Master'.
14) Replace the printer module and secure it using the 4 screws, 4 spring lock washers and 4 flat
washers.
15) On the right side of the printer module, replace the flat cable in the plastic cable holder.
16) Re-connect the hard disk and floppy drive supply cables.
17) Replace the cables in the 2 plastic cable holders and close them.
18) Re-connect the 2 flat cables going from the PC board to the floppy drive and the hard disk.
21) Close the left door and tighten the 2 ¼ turn fasteners.
7) Turn off the power and unplug the external power cord.
17) Install the new barcode reader on the bracket and replace the 2 screws and 2 flat washers –
but do not tighten the screws.
19) Connect the cable coming from the new barcode reader (unlabeled) to the Z board.
Place the barcode reader positioning jig in contact with the sample arm, as indicated by the
unbroken line below. Push the barcode reader against the jig, as indicated by the dotted
line below. Securely tighten the 2 screws.
21) Replace the sample probe and re-connect the red fitting.
23) Slide the metal top cover back on and lock the 2 ¼ turn fasteners.
25) Replace the 3 reagent bottles. The correct positioning order for the bottles is indicated on the
metal top cover:
• 1 = buffer 1
• 2 = buffer 2
• W = wash/diluent
9) Install the new detector module and lock the captive screw.
10) Remove the thumbscrew and black cover from the new module.
13) Connect:
• the blue PEEK tubing to the cartridge module (beige PEEK fitting)
• the orange PEEK tubing to the white union (grey knurled fitting)
14) Push the HPLC drawer back in and tighten the 2 captive screws.
1) Shut down the D-10, turn off the power and unplug the external power cord.
3) Unscrew the 2 ¼ turn fasteners which lock the left door and open
it.
8) Remove the plate/sensor assembly from the rack drive module. Pay attention to the sensor
cable as there are several cable holder on its way to the X board.
9) Remove the old rack input sensor from the metal plate.
//
11) Secure the plate/sensor assembly to the rack drive module using the 2 screws, 2 spring lock
washers and 2 flat washers. The plate should be pushed to the far right (towards the rack
door) prior to securing the screws.
12) Lead the cable to the X board and connect it at CN9. Re-connect the flat cable to the board if
it was previously removed.
14) On the front side of the D-10, connect the front cover
sensor simulator (P/N: 2200415) and the front cover
switch simulator (P/N: 2200416).
WARNING: Use extreme caution when using these tools, as all moving parts can be
engaged.
Move the rack manually and insert the rack Change the position of the gauge and
sensor adjustment gauge between the rack monitor again the Rack input sensor status.
and the rack drive module. Move the This time, the status should be no reflection.
plate/sensor assembly and monitor the Rack If the result is not correct, move the
input sensor status on the RACK INTRO. (1/1) plate/sensor assembly again. Repeat both
screen. With the gauge fully inserted, the tests until the results are OK. Tighten the
status should be reflection. screws when done.
26) Unplug the front cover sensor simulator and the front cover switch simulator.
1) Shut down the D-10, turn off the power and unplug the external power cord.
3) Unscrew the 2 ¼ turn fasteners which lock the left door and open
it.
8) Remove the plate/sensor assembly from the rack drive module. Pay attention to the sensor
cable as there are several cable holder on its way to the X board.
9) Remove the old rack output sensor from the metal plate.
//
12) Lead the cable to the X board and connect at CN9. Re-connect the flat cable to the board if it
was previously removed.
14) Push the HPLC drawer back in and tighten the 2 captive screws.
17) Go to the RACK INTRO. (1/1) screen to check that the new
sensor is working well:
• the default status should be no reflection
• click on LAUNCH RACK INTRODUCTION and
introduce a rack
• the sensor's status should change to reflection
• click on LAUNCH RACK EJECTION
• the status should come back to no reflection
1) Shut down the D-10, turn off the power and unplug the external power cord.
6) Manually move the sample arm to the middle of the rack drive module.
10) Remove the user interface module. Pay attention to the contrast adjustment potentiometer
located on the right side.
11) Install the new user interface module and secure it with the 4 screws and 4 spring lock
washers. Pay attention to the contrast adjustment knob located on the right side.
12) Replace the flat cable in the plastic cable holder and re-connect it to the LCD interface board.
15) Slide the metal top cover back on and lock the 2 ¼ turn fasteners.
17) Replace the 3 reagent bottles. The correct positioning order for the bottles is indicated on the
metal top cover.
6.36 Arm
1) Shut down the D-10, turn off the power and unplug the external power cord.
5) Disconnect the red knurled fitting at the top of the sample probe. CAREFULLY remove the
probe. Use extreme caution, as the probe is a biohazard.
10) Manually move the sample arm to the middle of the rack
drive module.
12) Remove the 4 screws, 4 spring lock washers and 4 flat washers, which secure the old arm to
the frame. Be careful to HOLD THE ARM with one hand.
14) CAREFULLY install the new arm in the frame – but do not tighten the screws.
Make sure that the sample arm rests on the 2 pins indicated by the unbroken arrows. Then,
push (or pull) the arm so that it comes in contact with the 4 other pins (indicated by the
dotted arrows). Securely tighten the 4 screws.
18) On the rear side of the sample arm, slide in the grommet and close the metal cable holders.
20) Replace the thumbscrew and small aluminum bracket, in order to fasten the probe tubing.
21) Install the dummy probe (P/N: 2200424) and secure it with the red knurled fitting.
23) On the front side of the D-10, connect the front cover
sensor simulator (P/N: 2200415) and the front cover
switch simulator (P/N: 2200416).
WARNING: Use extreme caution when using these tools, as all moving parts can be
engaged.
30)
Manually move the arm towards the right, in order to reveal
the second set screw of the X pulley. Loosen this screw.
34) Manually move the X pulley around the X motor shaft, in order to center the dummy probe in
the cone of the X movement adjustment tool.
38) Manually move the arm towards the right, in order to reveal the second set screw of the X
pulley. Tighten this screw.
42) Move the X home sensor and repeat steps 39) to 41) until the X position sensor is between 2
slots. At the end, check the probe position as described in steps 26) to 28).
Place the Z movement control tool in contact with the TH movement adjustment tool. Verify
that the probe support is between the 1st stage (included) and the 3rd stage (excluded). The
ideal situation is when the probe support touches the 2nd stage.
51) Make sure that the Z position sensor is still between 2 slots; if needed, move the slotted disk.
52) Verify the setting of the Z movement again, as described in steps 47) and 48).
55) Perform the procedure from Section 7.13 Hall Effect Sensor Adjustment.
60) Remove the X movement adjustment tool and replace the dilution/wash station.
61) Remove the dummy probe and replace the sample probe.
62) Unplug the front cover sensor simulator and the front cover switch simulator.
1) Shut down the D-10, turn off the power and unplug the external power cord.
7) Manually move the sample arm to the middle of the rack drive module.
10) Remove the 8 (or 4) screws, along with the spring lock washers and flat washers, that secure
the old rack drive module to the frame. Be careful to HOLD THE MODULE with one hand.
12) CAREFULLY install the new rack drive module in the frame – but do not tighten the screws.
Make sure that the rack drive module is in contact with all 6 pins, then tighten the 8 (or 4)
screws.
16) On the syringe module, connect the green and yellow LUER fittings.
17) Connect the 3 cables coming from the new rack drive module to the X board at CN2, CN7
and CN12.
19) On the front side of the D-10, connect the front cover
sensor simulator (P/N: 2200415) and the front cover
switch simulator (P/N: 2200416).
WARNING: Use extreme caution when using these tools, as all moving parts can be
engaged.
23) Remove the probe and replace it with the dummy probe (P/N: 2200424). WARNING: the
probe is a biohazard. Use EXTREME CAUTION when removing.
29) Remove the X movement adjustment tool and replace the dilution/wash station.
30) Remove the dummy probe and replace the sample probe.
31) Unplug the front cover sensor simulator and the front cover switch simulator.
32) Push the HPLC drawer back in and tighten the 2 captive screws.
3) Alternate the remaining 2 screws, loosening them a few turns at a time until the head is free.
6) Install the new pump head on the follower assembly and secure the 2 center mounting
screws. Alternate between the 2 screws, giving a few turns at a time until the pump head is
fastened.
3) Remove the old piston straight out from the piston guide.
5) Replace the piston assembly in the follower assembly. The flat faces on the piston guide
should be in the vertical position.
4) Fill the syringe about half full of de-ionized water (preferably degassed).
5) Point the tip up and push the plunger in until most of the air is removed from the syringe.
6) Place a finger over the end of the syringe and pull the plunger back rapidly, in order to create
some vacuum in the barrel, and shake vigorously. Repeat this operation at least 5 times.
7) Install the new seals in the pump head with the piston seal removal tool. This may be done
under water to avoid air introduction.
WARNING: Use extreme caution when using these tools, as all moving parts can be
engaged.
5) Disconnect the red knurled fitting at the top of the sample probe. CAREFULLY remove the
probe. Use extreme caution, as the probe is a biohazard.
8) Turn off the power and unplug the external power cord.
18) Replace the old TH tube height sensor with the new one.
20) Fold the end of the new flex cable, following the example of the old one.
21) Fasten the plate/sensor assembly to the arm using the 2 screws, 2 spring lock washers and 2
flat washers – but do not tighten the screws.
Position the TH home flag positioning jig on the sample arm. Make sure the surfaces are in
contact. Match the surfaces indicated with the dotted and unbroken arrows.
Move the TH home flag so that it comes in contact (on both surfaces) with the jig. Tighten
the screws when done.
23) Route the flex cable to the Z board and plug it in.
25) Fasten the TH tube height flag to the arm using the 2 screws, 2 spring lock washers and 2 flat
washers – but do not tighten the screws.
Place the TH tube height flag positioning jig on the new TH tube height sensor. The
surfaces indicated with the unbroken arrows must come in contact.
Move the TH tube height flag so that it comes in contact with the 2nd stage of the jig.
Tighten the screws.
Verify that the TH tube height flag doesn't touch the 1st stage, but is placed between the 2nd
stage (included) and the 3rd stage (excluded). If this is not accomplished, repeat the
adjustment.
34) Turn off the power and unplug the external power cord.
35) Replace the sample probe and secure the red knurled fitting.
36) Unplug the front cover sensor simulator and the front cover switch simulator.
38) Slide the metal top cover back on and lock the 2 ¼ turn fasteners.
40) Replace the 3 reagent bottles. The correct positioning order for the bottles is indicated on the
metal top cover.
The figure below shows the components of a typical 7-Port, 3-Position HP valve.
4) Remove the stator and stator face assembly from the stator ring.
8) Replace the stator ring so that the 2 short pins on the ring are placed into the corresponding
holes in the body.
10) Replace the stator and stator face assembly on the valve so that the pin on the stator ring is
in the hole on the stator.
11) Replace the 3 stator screws. Be careful to TIGHTEN THEM A FEW TURNS AT A TIME.
12) Re-connect all of the tubing to the stator, making sure that the fittings are in their proper
position numbers.
13) Push the HPLC drawer back in and tighten the 2 captive screws.
18) Check for leaks while the pump is running. After approximately 5 minutes, click on STOP
PUMP.
19) Go back to the HP: PUMP + VALVE (4/17) screen and move
the valve to inject position.
21) Check for leaks while the pump is running. After approximately 5 minutes, click on STOP
PUMP.
22) Go to the HP: PUMP + VALVE (4/17) screen and move the
valve to load position.
4) CAREFULLY pull out the lid and the 3 internal Teflon tubings.
6) Install the new buffer tubings on the lid. MANUALLY tighten the black plastic fittings. Stop
when the parts are in contact.
9) Replace the lid on the body. Be very careful NOT TO PINCH any tubing between 2 parts.
10) Secure the lid with the 6 screws and 6 spring lock washers, tightening them a few turns at a
time.
3) CAREFULLY pull out the lid and the 3 internal Teflon tubings.
5) Install the new wash tubing on the lid. MANUALLY tighten the black plastic fittings. Stop
when the parts are in contact.
8) Replace the lid on the body. Be very careful NOT TO PINCH any tubing between 2 parts.
9) Secure the lid with the 6 screws and 6 spring lock washers, tightening them a few turns at a
time.
To protect electronic cards against damage from static electricity, several precautions must
be taken:
• use a grounded wrist strap before handling the components. If one is not
available, first touch both of your hands to a safely grounded or metal object
• hold components by the edges and try not to touch the IC chips, leads,
connectors, or other components
• place components on a grounded antistatic pad or on the bag that came with the
components whenever they are separated from the system
1) Shut down the D-10, turn off the power and unplug the external power cord.
2) Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
7) Replace the interface board support and tighten the 2 captive screws.
8) Re-connect all cables and flat cables to the interface board. Note that there is only one
possible connection for each cable and flat cable.
To protect electronic cards against damage from static electricity, several precautions must
be taken:
• use a grounded wrist strap before handling the components. If one is not
available, first touch both of your hands to a safely grounded or metal object
• hold components by the edges and try not to touch the IC chips, leads,
connectors, or other components
• place components on a grounded antistatic pad or on the bag that came with the
components whenever they are separated from the system
10) CAREFULLY install the new LCD internal board and secure it with the screw, flat washer and
spacer. DO NOT OVERTIGHTEN.
11) Re-connect the 2 flat cables and the supply cable to the LCD internal board.
12) Re-connect the link cable between the motherboard and the LAN connector.
13) Replace the interface board support and tighten the 2 captive screws.
14) Re-connect the flat cable located on the top end of the interface board.
6) Operating from the right side of the D-10, locate the white union and unscrew the grey
knurled fitting (orange PEEK tubing coming from the detector).
14) Re-connect the ground cable to the bottom of the new drawer.
Each spare HPLC module is delivered with a Specifications sheet, giving the module's
specific parameters. All these values must be entered in the 'instrument.ini' file once the
replacement is done.
15) Re-connect the 2 ribbon cables to the HPLC module board. Re-connect the 2 plugs of the
front cover solenoid link cable if they were previously removed.
18) Re-connect the silicon tubing coming from the piston flush circuit to the waste manifold 2.
19) Re-connect the orange PEEK tubing coming from the detector to the white union.
20) On the HP valve, re-connect the red knurled fitting (tube coming from the probe – port 5) and
the blue knurled fitting (tube coming from the syringe module – port 7).
21) Re-connect the white and green fittings to the proportioning valve. Be careful to MATCH THE
COLORS.
23) Push the HPLC drawer back in and tighten the 2 captive screws.
30) Check for leaks on the drawer while the pump is running.
3) Replace the old HPLC module board with the new one (4 plastic standoffs).
4) Re-connect all
cables to the HPLC
module board.
6.49 Z Board
1) Shut down the D-10, turn off the power and unplug the external power cord.
6) Manually move the sample arm to the middle of the rack drive module.
11) Install the new Z board and secure it with the 3 screws, 3 spring lock washers and 3 flat
washers.
12) Re-connect the 8-pin and 12-pin flex cables. Secure them with a new cable tie. DO NOT
OVERTIGHTEN.
15) Slide the metal top cover back on and lock the 2 ¼ turn fasteners.
17) Replace the 3 reagent bottles. The correct positioning order for the bottles is indicated on the
metal top cover
19) Place a barcoded microvial adapter into position #7 of a sample rack. Place an empty sample
vial into the adapter in the rack. Make sure the cap is snapped completely in place.
6.50 X Board
1) Shut down the D-10, turn off the power and unplug the external power cord.
3) Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
7) Remove the 4 screws, 4 spring lock washers and 4 flat washers that secure the X board to
the sample arm. Remove the board.
8) Install the new X board and secure it with the 4 screws, 4 spring lock washers and 4 flat
washers.
11) Close the left door and tighten the 2 ¼ turn fasteners.
3) Replace the old syringe module board with the new one (4 plastic standoffs).
11) Open the upper front door. If the door is not opened within 10 seconds, it will automatically
lock again.
12) Replace the dilution/wash station with the clear dilution/wash station (P/N: 2200427).
WARNING: Use extreme caution when using this tool, as all moving parts can be engaged.
15) Remove the upper front door simulator and close the upper front door.
16) Use the Access Wash st. button to remove the clear
dilution/wash station.
3) Replace the old degasser module board with the new one (4 plastic standoffs).
4) Re-connect all cables to the module board. Be careful, as the cables coming from the bubble
detectors (labeled '1', '3' and '6') are NOT INTERCHANGEABLE.
1) Shut down the D-10, turn off the power and unplug the external power cord.
7) CAREFULLY remove the old rack positioning motor assembly. Be mindful of the cable, as
there are several metal cable holders on its way to the X board.
First, loosen the locking counter-nut. Then, tighten the screw until there is a small
clearance between the mounting screw of the rubber disk and the rack drive module.
12) Push the HPLC drawer back in and tighten the 2 captive screws.
15) Go to the RACK INTRO. (1/1) screen to check that the new
rack positioning motor assembly is working well:
• test the motor in both directions
• click on LAUNCH RACK INTRODUCTION and
introduce a rack
• click on LAUNCH RACK EJECTION
2) Unlock the screw and slide out the old rubber disk.
3) Install the new rubber disk on the shaft and tighten the screw.
4) Apply thermal grease on the inner surface of the housing and push the new temperature
sensor into the housing.
Each spare temperature sensor has a sticker on it, giving the sensor's specific parameter.
Note this value; it must be entered in the 'instrument.ini' file once the replacement is done.
5) Replace the cable in the metal cable holder and close it.
11) Replace the 4 cables in the metal cable holders and close them.
4) Open the upper front door. If the door is not opened within 10 seconds, it will automatically
lock again.
6) Turn off the power and unplug the external power cord.
10) Install the new upper front door and secure it with the 4 screws and 4 flat washers.
11) Fasten the gas spring to the new upper front door.
4) Install the new lower front door and secure it with the 4 screws.
6.60 Micro-Switch
6) Install the new micro-switch on the bracket and secure it with the 2 screws and 2 spring lock
washers.
8) Replace the micro-switch assembly and secure it with the nut and flat washer.
80mm
7) Open the metal cable holder and remove the old front cover solenoid/sensor.
11) Secure the sensor with a drop of tamperproofing coating (P/N: 2200451).
14) Replace the micro-switch assembly and secure it with the nut and flat washer.
110mm
8) Install the cable on the new front cover, using the screw and flat washer.
7) Manually open and close the rack door to make sure that the new solenoid is correctly
positioned.
8) Fasten the solenoid cable (along with the open door and closed door sensors cables) to the
rack door assembly with new tie-wraps.
11) Go to the RACK DOOR (1/2) screen to check that the new
rack door solenoid is working properly:
• click on START RACK DOOR SOLENOID
• open and close the rack door
6) Install the new open door sensor on the bracket and replace the 2 screws and 2 serrated lock
washers – but do not tighten the screws.
Place the rack door sensor positioning jig in contact with the rack door assembly, as
indicated by the unbroken line below. Push the open door sensor against the jig, as
indicated by the dotted line below. Securely tighten the 2 screws.
10) Fasten the sensor cable (along with the solenoid and closed door sensor cables) to the rack
door assembly with a new tie-wrap.
13) Go to the RACK DOOR (1/2) screen to check that the new
open door sensor is working properly:
• open and close the rack door
• monitor the sensor status and the door status
6) Install the new closed door sensor on the bracket and replace the 2 screws and 2 serrated
lock washers – but do not tighten the screws.
Place the rack door sensor positioning jig in contact with the rack door assembly, as
indicated by the unbroken line below. Push the closed door sensor against the jig, as
indicated by the dotted line below. Securely tighten the 2 screws.
8) Fasten the sensor cable (along with the solenoid and open door sensor cables) to the rack
door assembly with a new tie-wrap.
11) Go to the RACK DOOR (1/2) screen to check that the new
closed door sensor is working properly:
• open and close the rack door
• monitor the sensor status and the door status
5) Re-assemble the 3 blocks of the pump head and secure them with the 4 screws. DO NOT
OVERTIGHTEN.
11) Perform the procedure from Section 7.7 Flow Rate Verification.
2) Manually move the sample arm to the right side of the D-10.
7) Install the new X home sensor on the sample arm module and replace the 2 screws and 2
serrated lock washers – but do not tighten the screws.
Place the X home sensor positioning jig in contact with the frame, as indicated by the
unbroken line below. Push the X home sensor against the jig, as indicated by the dotted
line below. Securely tighten the 2 screws.
11) Replace the sample probe with the dummy probe (P/N: 2200424).
12) On the front side of the D-10, connect the front cover
sensor simulator (P/N: 2200415) and the front cover
switch simulator (P/N: 2200416).
WARNING: Use extreme caution when using these tools, as all moving parts can be
engaged.
18) Manually move the X home sensor towards the center of the instrument.
26) Move the X home sensor and repeat steps 23) to 25) until the X position sensor is between 2
slots. At the end, check the probe position as described in steps 15) to 17).
30) Remove the X movement adjustment tool and replace the dilution/wash station.
31) Remove the dummy probe and replace the sample probe.
32) Unplug the front cover sensor simulator and the front cover switch simulator.
1) Shut down the D-10, turn off the power and unplug the external power cord.
2) Unscrew the 2 ¼ turn fasteners which lock the left door and open
it.
5) On the X motor:
• loosen – but do not remove – the 2 screws, 2
spring lock washers and 2 flat washers that
secure the metal plate
• swivel the plate to have access to the sensor
• remove the old X position sensor
6) Install the new X position sensor on the metal plate and secure it with the 2 nuts and 2
serrated lock washers.
7) Swivel the metal plate back in its original position and tighten the 2 screws.
8) Connect the cable coming from the new X position sensor to the X board (CN10).
14) Open the upper front door. If the door is not opened within 10 seconds, it will automatically
lock again.
15) Replace the sample probe with the dummy probe (P/N: 2200424).
WARNING: Use extreme caution when using this tool, as all moving parts can be engaged.
29) Remove the X movement adjustment tool and replace the dilution/wash station.
30) Remove the dummy probe and replace the sample probe.
31) Remove the upper front door simulator and close the upper front door.
1) Turn off the power and unplug the external power cord.
4) Manually move the sample arm to the middle of the rack drive module.
5) Manually turn the TH and Z motors, in order to lower the tube holder assembly and the probe
support.
8) Remove the silicon rubber insulator from the old sensor connector and install it on the new
one.
Make sure that the upper face of the TH home sensor is flush with the upper sample arm
plate. Push the sensor against the small metal bracket or use a flat tool (such as the upper
front door simulator).
10) Route the TH home sensor cable to the Z board and connect it at CN12.
16) Make sure the TH position sensor is still between 2 slots; if needed, move the slotted disk.
1) Shut down the D-10, turn off the power and unplug the external power cord.
9) Holding the TH motor with one hand, remove the 2 screws, 2 spring lock washers and 2 flat
washers. Remove the TH motor.
11) Remove the silicon rubber insulator from the old sensor connector and install it on the new
one.
12) Install the new TH position sensor on the metal plate and secure it with the 2 nuts and 2
serrated lock washers.
13) Swivel the metal plate back in its original position and tighten the 2 screws.
14) Install the TH motor on its support with the 2 screws, 2 spring lock washers and 2 flat
washers – but do not tighten the screws.
15) Replace the pulley on the shaft and tighten the screw. For the exact position of the pulley on
the shaft, follow the example of the Z motor.
17) Adjust the tension of the belt by sliding the TH motor in the oblong holes and tighten the 2 TH
motor screws. Follow the example of the Z motor.
19) Fasten the TH position sensor cable (along with the TH motor cable) to the sample arm
module with a new tie-wrap.
1) Turn off the power and unplug the external power cord.
4) Manually move the sample arm to the middle of the rack drive module.
5) Manually turn the TH and Z motors, in order to lower the tube holder assembly and the probe
support.
8) Install the new Z home sensor on the sample arm module and replace the 2 screws and 2
serrated lock washers – but do not tighten the screws.
Position the Z home sensor positioning jig on the sample arm. Make sure the surfaces are
in contact. Match the surfaces indicated with the dotted and unbroken arrows.
Move the Z home sensor so that it comes in contact with the jig. Tighten the screws when
done.
10) Route the Z home sensor cable to the Z board and connect it at CN2.
16) Make sure the Z position sensor is still between 2 slots; if needed, move the slotted disk.
20) Use the Z movement control tool (P/N: 2200431) to verify the
setting of the Z movement. If it is OK, go to step 25). If it isn't,
continue with step 21).
Place the Z movement control tool in contact with the TH movement adjustment tool. Verify
that the probe support is between the 1st stage (included) and the 3rd stage (excluded). The
ideal situation is when the probe support touches the 2nd stage.
23) Make sure the Z position sensor is still between 2 slots; if needed, move the slotted disk.
24) Verify the setting of the Z movement again, as described in steps 19) and 20).
This procedure also applies to the 30-tooth pulley mounted on the Z motor.
5) Install the new 30-tooth pulley on the shaft and tighten the mounting screw. For the exact
position of the pulley on the shaft, follow the example of the Z motor.
7) Adjust the tension of the belt by sliding the TH motor in the oblong holes and tighten the 2 TH
motor screws. Follow the example of the Z motor.
For the 30-tooth pulley mounted on the TH motor, continue with step 9a) and stop. Steps
9b) to 13) apply ONLY to the pulley mounted on the Z motor.
10) Use the Z movement control tool (P/N: 2200431) to test the Z movement.
Place the Z movement control tool in contact with the TH movement adjustment tool. Verify
that the probe support is between the 1st stage (included) and the 3rd stage (excluded). The
ideal situation is when the probe support touches the 2nd stage.
This procedure also applies to the 15-tooth pulley mounted on the Z lead screw.
5) Install the new 15-tooth pulley on the TH lead screw and tighten the mounting screw. Make
sure the lead screw is pushed all the way up prior to securing the screw. Don't forget to put
the flat washer between the lead screw bearing and the 15-tooth pulley.
7) Adjust the tension of the belt by sliding the TH motor in the oblong holes and tighten the 2 TH
motor screws. Follow the example of the Z motor.
For the 15-tooth pulley mounted on the TH lead screw, continue with step 9a) and stop.
Steps 9b) to 13) apply ONLY to the pulley mounted on the Z lead screw.
10) Use the Z movement control tool (P/N: 2200431) to test the Z movement.
Place the Z movement control tool in contact with the TH movement adjustment tool. Verify
that the probe support is between the 1st stage (included) and the 3rd stage (excluded). The
ideal situation is when the probe support touches the 2nd stage.
5) Adjust the tension of the belt by sliding the Z motor in the oblong holes and tighten the 2 Z
motor screws. Follow the example of the TH motor.
8) Use the Z movement control tool (P/N: 2200431) to test the Z movement.
Place the Z movement control tool in contact with the TH movement adjustment tool. Verify
that the probe support is between the 1st stage (included) and the 3rd stage (excluded). The
ideal situation is when the probe support touches the 2nd stage.
5) Adjust the tension of the belt by sliding the TH motor in the oblong holes and tighten the 2 TH
motor screws. Follow the example of the Z motor.
4) Disconnect the power supply cable from the liquid pump. Observe polarity. Remove the old
supply cable.
6) Connect the other end of the cable to the syringe module board (CN6).
7) Put all cables in the metal cable holder and close it.
5) Connect the tubings to the new liquid pump head. Be careful NOT TO SWAP THE 2
TUBINGS.
6) Secure the new pump to the syringe module with the 2 screws and 2 spring lock washers.
7) Re-connect the supply cable to the pump. Be very careful to RESPECT POLARITY.
1) Verify that the system is in Sleep state. If the system is in Standby, place it in Sleep state
using the RUN screen.
3) Remove the internal waste bottle from the C-clip by gently pulling.
5) Connect the tubings to the new internal waste bottle. Be careful to MATCH THE COLORS.
Make sure that the cap is screwed on tightly and that the LUER fittings on the tubings are
connected tightly.
This procedure applies ONLY to the old internal waste bottle cap design.
1) Clean the internal waste bottle according to the procedure in Section 5 of the Operation
Manual.
6) Install the 3 new LUER fittings on the cap and secure them with the plastic nuts. DO NOT
OVERTIGHTEN, as there is a risk of damaging the seals.
7) Replace the piece of silicon tubing on the LUER fitting which is marked in red.
6) Connect the other end of the supply cables to the syringe module board (CN7 and CN8).
7) Put all cables in the metal cable holder and close it.
1) Verify that the system is in Sleep state. If the system is in Standby, place it in Sleep state
using the RUN screen.
6) CAREFULLY insert the tubing in the pinch valve head (push up).
6) Install the new pinch valve on the syringe module and secure it with the 2 screws, 2 spring
lock washers and 2 flat washers.
4) Install the new detector board on the module and secure it with the 2 screws, 2 spring lock
washers and 2 flat washers.
5) Re-connect the 3 cables to the new detector board. Note that there is only one possible
connection for each cable.
8) Move the jumper located on the bottom end of the detector board, according to the type of
reference photodetector previously identified.
With a SMALL photodetector, the jumper With a LARGE photodetector, the jumper
should be in the UPPER position. should be in the LOWER position.
When replacing the reference board, it may prove to be necessary to replace the detector
board also – see step 10).
Prior to installing the new reference board, make sure that the photodiode is clean. If the
window needs to be cleaned, use ethyl alcohol and wipe off the window gently. Avoid using
any other organic solvents, as they may cause deterioration of the device's resin coating.
When using tweezers or other hard tools, be careful NOT TO SCRATCH the window surface.
7) Pass the cable coming from the new reference board through the hole in the detector base.
11) Connect the new reference board cable to the detector board.
Prior to installing the new sample board, make sure that the filter is clean. If the filter needs
to be cleaned, use ethyl alcohol and wipe off the filter gently. Avoid using any other organic
solvents, as they may cause deterioration of the device's resin coating. When using
tweezers or other hard tools, be careful NOT TO SCRATCH the filter surface.
8) Pass the cable coming from the new sample board through the hole in the detector base.
10) Install the new sample board on the black nylon standoffs and secure it with the 2 screws and
2 spring lock washers.
11) Replace the black metal bracket on the detector base and secure it with the 2 screws and 2
spring lock washers.
12) Re-connect the 3 cables to the detector board. Note that there is only one possible
connection for each cable.
8) Pass the cable coming from the new LED board through the hole in the detector base.
10) Install the new LED board on the black Delrin standoffs and replace the 2 screws and 2
spring lock washers – but don't tighten the screws.
Move the LED board in order to center the blue halo on the flow cell lens. If necessary, do
not hesitate to add 1 (or more) flat washer(s) between the LED board and the black Delrin
standoffs.
17) Disconnect the ribbon cable from the detector module board.
18) Unplug the front cover sensor simulator and the front cover switch simulator.
19) Replace the black metal bracket on the detector base and secure it with the 2 screws and 2
spring lock washers.
20) Connect the 2 remaining cables to the detector board. Note that there is only one possible
connection for each cable.
6) Remove the 2 screws and the small metal bracket, which secure the flow cell to the detector
assembly.
BLUE ORANGE
10) Install the new flow cell and secure it with the
small metal bracket and the 2 screws. Make
sure it is in contact with the support and the 2
plastic-coated pins prior to securing the 2
screws.
11) Tighten the 2 screws and 2 spring lock washer that secure the sample board.
chamfered end
1) Shut down the D-10, turn off the power and unplug the external power cord.
8) Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
10) On the right side of the printer module, pull the flat
cable out of the plastic cable holder.
15) Connect the 2 flat cables going from the PC board to the floppy drive and the hard disk.
17) Close the left door and tighten the 2 ¼ turn fasteners.
1) Shut down the D-10, turn off the power and unplug the external power cord.
5) Unscrew the 2 ¼ turn fasteners which lock the left door and open
it.
9) Connect the new floppy drive supply cable to the interface board.
10) Connect the other end of the cable to the floppy drive.
11) Replace the supply cable (along with all the other cables) in the 3 plastic cable holders and
close them.
12) Close the left door and tighten the 2 ¼ turn fasteners.
1) Shut down the D-10, turn off the power and unplug the external power cord.
5) Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
9) Connect the new hard disk supply cable to the interface board.
10) Connect the other end of the cable to the hard disk.
11) Replace the supply cable (along with all the other cables) in the 3 plastic cable holders and
close them.
12) Close the left door and tighten the 2 ¼ turn fasteners.
1) Shut down the D-10, turn off the power and unplug the external power cord.
7) Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
13) Slide the metal top cover back on and lock the 2 ¼ turn fasteners.
15) Replace the 3 reagent bottles. The correct positioning order for the bottles is indicated on the
metal top cover:
• 1 = buffer 1
• 2 = buffer 2
• W = wash/diluent
16) Replace the supply cable (along with all the other cables) in the plastic cable holder and
close it.
17) Close the left door and tighten the 2 ¼ turn fasteners.
1) Shut down the D-10, turn off the power and unplug the external power cord.
3) Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
9) Replace the cable (along with all the other cables) in the plastic cable holder and close it.
10) Close the left door and tighten the 2 ¼ turn fasteners.
12) Replace the power cord and turn on the D-10. Make sure that air is EXTRACTED from the
unit.
To protect electronic cards against damage from static electricity, several precautions must
be taken:
• use a grounded wrist strap before handling the components. If one is not
available, first touch both of your hands to a safely grounded or metal object
• hold components by the edges and try not to touch the IC chips, leads,
connectors, or other components
• place components on a grounded antistatic pad or on the bag that came with the
components whenever they are separated from the system
6) Install the new link cable. To make sure that it won't interfere with the rotation of the
processor fan, it is recommended to CAREFULLY twist the cable.
7) Replace the interface board support and tighten the 2 captive screws.
8) Re-connect the flat cable located on the top end of the interface board.
To protect electronic cards against damage from static electricity, several precautions must
be taken:
• use a grounded wrist strap before handling the components. If one is not
available, first touch both of your hands to a safely grounded or metal object
• hold components by the edges and try not to touch the IC chips, leads,
connectors, or other components
• place components on a grounded antistatic pad or on the bag that came with the
components whenever they are separated from the system
2) Remove the link cable between the motherboard and the LAN connector. (Section 6.93)
5) Fold the new ribbon cable, following the example of the old one.
8) Replace the link cable between the motherboard and the LAN connector.
To protect electronic cards against damage from static electricity, several precautions must
be taken:
• use a grounded wrist strap before handling the components. If one is not
available, first touch both of your hands to a safely grounded or metal object
• hold components by the edges and try not to touch the IC chips, leads,
connectors, or other components
• place components on a grounded antistatic pad or on the bag that came with the
components whenever they are separated from the system
2) Remove the link cable between the motherboard and the LAN connector. (Section 6.93)
7) Fold the new flat cable, following the example of the old one.
12) Replace the link cable between the motherboard and the LAN connector.
To protect electronic cards against damage from static electricity, several precautions must
be taken:
• use a grounded wrist strap before handling the components. If one is not
available, first touch both of your hands to a safely grounded or metal object
• hold components by the edges and try not to touch the IC chips, leads,
connectors, or other components
• place components on a grounded antistatic pad or on the bag that came with the
components whenever they are separated from the system
KB
To protect electronic cards against damage from static electricity, several precautions must
be taken:
• use a grounded wrist strap before handling the components. If one is not
available, first touch both of your hands to a safely grounded or metal object
• hold components by the edges and try not to touch the IC chips, leads,
connectors, or other components
• place components on a grounded antistatic pad or on the bag that came with the
components whenever they are separated from the system
5) Fold the new flat cable, following the example of the old one.
8) Push the new rack feeder port link cable in the plastic cable holder.
To protect electronic cards against damage from static electricity, several precautions must
be taken:
• use a grounded wrist strap before handling the components. If one is not
available, first touch both of your hands to a safely grounded or metal object
• hold components by the edges and try not to touch the IC chips, leads,
connectors, or other components
• place components on a grounded antistatic pad or on the bag that came with the
components whenever they are separated from the system
1) Remove the link cable between the motherboard and the LAN connector. (Section 6.93)
4) Install the new heatsink on the motherboard. It is not absolutely necessary to apply thermal
grease on the processor.
6) Replace the link cable between the motherboard and the LAN connector.
To protect electronic cards against damage from static electricity, several precautions must
be taken:
• use a grounded wrist strap before handling the components. If one is not
available, first touch both of your hands to a safely grounded or metal object
• hold components by the edges and try not to touch the IC chips, leads,
connectors, or other components
• place components on a grounded antistatic pad or on the bag that came with the
components whenever they are separated from the system
2) Remove the old processor and replace it with the new one.
Locate the socket and open it by pulling the lever upwards. The socket lever must be fully
open (90° to 100°) when removing and installing the processor. Insert the processor with
the correct orientation, knowing that there is a notched corner. DO NOT EXERT TOO MUCH
FORCE, but press the processor down gently into the slot.
Once completely inserted, close the socket lever while holding down the processor.
3) Replace the heatsink. It is not absolutely necessary to apply thermal grease on the new
processor.
5) Replace the link cable between the motherboard and the LAN connector.
To protect electronic cards against damage from static electricity, several precautions must
be taken:
• use a grounded wrist strap before handling the components. If one is not
available, first touch both of your hands to a safely grounded or metal object
• hold components by the edges and try not to touch the IC chips, leads,
connectors, or other components
• place components on a grounded antistatic pad or on the bag that came with the
components whenever they are separated from the system
1) Shut down the D-10, turn off the power and unplug the external power cord.
6) Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
12) Fold the new flat cable, following the example of the old one.
17) Replace the interface board support and tighten the 2 captive screws.
18) Re-connect all cables and flat cables to the interface board. Note that there is only one
possible connection for each cable and flat cable.
19) Close the left door and tighten the 2 ¼ turn fasteners.
To protect electronic cards against damage from static electricity, several precautions must
be taken:
• use a grounded wrist strap before handling the components. If one is not
available, first touch both of your hands to a safely grounded or metal object
• hold components by the edges and try not to touch the IC chips, leads,
connectors, or other components
• place components on a grounded antistatic pad or on the bag that came with the
components whenever they are separated from the system
1) Shut down the D-10, turn off the power and unplug the external power cord.
6) Unscrew the 2 ¼ turn fasteners which lock the left door and open
it.
11) Fold the new flat cable, following the example of the old one.
12) Connect the new floppy drive flat cable to the motherboard.
13) Connect the other end of the cable to the floppy drive.
16) Replace the interface board support and tighten the 2 captive screws.
17) Re-connect all cables and flat cables to the interface board. Note that there is only one
possible connection for each cable and flat cable.
18) Close the left door and tighten the 2 ¼ turn fasteners.
6.102 Flat Cable: LCD Internal Board > LCD Interface Board
To protect electronic cards against damage from static electricity, several precautions must
be taken:
• use a grounded wrist strap before handling the components. If one is not
available, first touch both of your hands to a safely grounded or metal object
• hold components by the edges and try not to touch the IC chips, leads,
connectors, or other components
• place components on a grounded antistatic pad or on the bag that came with the
components whenever they are separated from the system
1) Shut down the D-10, turn off the power and unplug the external power cord.
12) Fold the new flat cable, following the example of the old one.
13) Connect the new flat cable to the LCD internal board.
14) Replace the interface board support and tighten the 2 captive screws.
15) Re-connect all cables and flat cables to the interface board. Note that there is only one
possible connection for each cable and flat cable.
16) Install the new flat cable in all the plastic cable holders.
17) Connect the new flat cable to the LCD interface board.
19) Slide the metal top cover back on and lock the 2 ¼ turn fasteners.
21) Replace the 3 reagent bottles. The correct positioning order for the bottles is indicated on the
metal top cover:
• 1 = buffer 1
• 2 = buffer 2
• W = wash/diluent
7) Turn off the power and unplug the external power cord.
13) Cut and remove the tie-wraps, which fasten the link
cable to the sample arm.
16) Connect the new link cable to the Hall effect adapter sensor.
18) Connect the new link cable (labeled '1') to the Z board.
19) Replace the sample probe and re-connect the red fitting.
21) Slide the metal top cover back on and lock the 2 ¼ turn fasteners.
23) Replace the 3 reagent bottles. The correct positioning order for the bottles is indicated on the
metal top cover:
• 1 = buffer 1
• 2 = buffer 2
• W = wash/diluent
2) Turn off the power and unplug the external power cord.
7) Install the new Hall effect adapter sensor on the sample arm and secure it with the screw,
spring lock washer and flat washer.
11) Perform the procedure from Section 7.13 Hall Effect Sensor Adjustment.
1) Shut down the D-10, turn off the power and unplug the external power cord.
2) Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
7) Clean off the residue of adhesive on the inner side of the frame.
BLUE
BROWN
12) Glue the new plastic protection on the power entry module.
13) Replace all 3 cables in the 2 plastic cable holders and close them.
14) Close the left door and tighten the 2 ¼ turn fasteners.
1) Shut down the D-10, turn off the power and unplug the external power cord.
5) Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
BLUE
10) Connect the new cable (along with the
cable coming from the power boards) to
the main switch. Prior to connecting the 4
plugs, slip the new piece of heat-
shrinkable tubing on both cables.
BROWN
BLUE
BROWN
13) Replace all 3 cables in the 2 plastic cable holders and close them.
14) Push the HPLC drawer back in and tighten the 2 captive screws.
17) Close the left door and tighten the 2 ¼ turn fasteners.
1) Shut down the D-10, turn off the power and unplug the external power cord.
4) On the front side of the D-10, open the lower front door.
7) Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
15) Install the new cable in the 2 plastic cable holders and close them.
BROWN
18) Replace all 3 cables in the 2 plastic cable holders and close them.
19) Close the left door and tighten the 2 ¼ turn fasteners.
20) Push the HPLC drawer back in and tighten the 2 captive screws.
23) Replace the protective grid and secure it with the 2 screws, 2 spring lock washers and 2 flat
washers.
To protect electronic cards against damage from static electricity, several precautions must
be taken:
• use a grounded wrist strap before handling the components. If one is not
available, first touch both of your hands to a safely grounded or metal object
• hold components by the edges and try not to touch the IC chips, leads,
connectors, or other components
• place components on a grounded antistatic pad or on the bag that came with the
components whenever they are separated from the system
1) Shut down the D-10, turn off the power and unplug the external power cord.
2) Unscrew the 2 ¼ turn fasteners which lock the left door and open
it.
10) Replace the interface board support and tighten the 2 captive screws.
11) Re-connect all cables and flat cables to the interface board. Note that there is only one
possible connection for each cable and flat cable.
12) Close the left door and tighten the 2 ¼ turn fasteners.
1) Shut down the D-10, turn off the power and unplug the external power cord.
4) Unscrew the 2 ¼ turn fasteners which lock the left door and open
it.
12) Connect the other end of the cable to the interface board.
15) Replace the thermal printer supply cable (along with all the other cables) in the plastic cable
holder and close it.
16) Close the left door and tighten the 2 ¼ turn fasteners.
17) Replace the protective grid and secure it with the 2 screws, 2 spring lock washers and 2 flat
washers.
8) Install the new LCD contrast potentiometer on the user interface module and secure it with
the nut and serrated lock washer.
9) CAREFULLY pass the new potentiometer cable through the hole in the user interface
module.
12) Fasten the potentiometer cable to the user interface module using new tie-wraps.
6.111 Flat Cable: Interface Board > HPLC Module Board (24-pin)
Flat Cable: Interface Board > HPLC Module Board (34-pin)
1) Shut down the D-10, turn off the power and unplug the external power cord.
3) Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
FLAT CABLE:
INTERFACE
BOARD > HPLC MODULE
BOARD (34-pin)
FLAT CABLE:
INTERFACE
BOARD > HPLC MODULE
BOARD (24-pin)
5) Fold the new flat cable, following the example of the old one.
1) Shut down the D-10, turn off the power and unplug the external power cord.
2) Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
FLAT CABLE:
INTERFACE
BOARD > X BOARD
4) Fold the new flat cable, following the example of the old one.
1) Shut down the D-10, turn off the power and unplug the external power cord.
3) Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
FLAT CABLE:
INTERFACE
BOARD > DEGASSER
MODULE BOARD
FLAT CABLE:
INTERFACE
BOARD > DETECTOR
BOARD
5) Fold the new flat cable, following the example of the old one.
1) Shut down the D-10, turn off the power and unplug the external power cord.
4) Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
FLAT CABLE:
INTERFACE
BOARD > SYRINGE
MODULE BOARD
6) Fold the new flat cable, following the example of the old one.
10) Close the left door and tighten the 2 ¼ turn fasteners.
5) Connect the new supply cable to the vacuum pump. Be very careful to RESPECT
POLARITY.
6) Connect the other end of the cable to the degasser module board (CN8). Make sure that the
cable passes underneath the board.
7) Fasten the supply cable (along with the silicon tubing) to the vacuum pump using a new tie-
wrap. Be careful NOT TO PINCH the tubing.
chamfered end
chamfered end
2.45 ± 0.05mm
12) Install the new 23µL sample loop on the HP valve. Make sure the loop is oriented the same
way it was in step 10):
• 1st end on port 4
• 2nd end on port 1
14) Open the cartridge module, by pulling the black metal cover all the way to the left.
15) Remove the cartridge holder and replace it with the dummy PEEK cartridge holder (P/N:
2200435). Unscrew the LUER fittings from the dummy PEEK cartridge holder prior to
installing it.
18) Put the free end of the back pressure regulator into an empty vessel.
22) Check for leaks on the HP valve while the pump is running. After approximately 5 minutes,
click on STOP PUMP.
23) Go back to the HP: PUMP + VALVE (4/17) screen and click
on HOME MVT (RESET).
24) Turn off the power and unplug the external power cord.
26) Remove the dummy PEEK cartridge holder and replace the analytical cartridge.
27) Replace the rubber plug in the cartridge module door and close the door.
28) Push the HPLC drawer back in and tighten the 2 captive screws.
1
3) Remove the 2 Nylock nuts which secure 6
each bubble detector to the degasser
module. Remove the old bubble detectors.
5) Connect the new bubble detectors to the degasser module board (CN1, CN3 and CN6).
1) Shut down the D-10, turn off the power and unplug the external power cord.
5) Unscrew the 2 ¼ turn fasteners which lock the left door and open
it.
6) Manually move the sample arm to the middle of the rack drive module.
9) Fold the new flex cable, following the example of the old one.
11) Connect the new flex cable to the Z board and secure it with a new tie-wrap. DO NOT
OVERTIGHTEN.
12) Close the left door and tighten the 2 ¼ turn fasteners.
13) Slide the metal top cover back on and lock the 2 ¼ turn fasteners.
15) Replace the 3 reagent bottles. The correct positioning order for the bottles is indicated on the
metal top cover:
• 1 = buffer 1
• 2 = buffer 2
• W = wash/diluent
17) Place a barcoded microvial adapter into position #7 of a sample rack. Place an empty sample
vial into the adapter in the rack. Make sure the cap is snapped completely in place.
4) Fold the new flex cable, following the example of the old one.
7) Replace the 12-pin flex cable and secure it with a new tie-wrap. DO NOT OVERTIGHTEN.
9) Slide the metal top cover back on and lock the 2 ¼ turn fasteners.
11) Replace the 3 reagent bottles. The correct positioning order for the bottles is indicated on the
metal top cover:
• 1 = buffer 1
• 2 = buffer 2
• W = wash/diluent
13) Place a barcoded microvial adapter into position #7 of a sample rack. Place an empty sample
vial into the adapter in the rack. Make sure the cap is snapped completely in place.
1) Shut down the D-10, turn off the power and unplug the external power cord.
6) Manually move the sample arm to the middle of the rack drive module.
9) Push (or pull) the X motor towards the right, in order to loosen the X gear belt.
10) Remove the 4 screws, 4 spring lock washers and 4 flat washers that secure the X motor to
the sample arm. CAREFULLY remove the X motor.
≈ 1mm
15) Tighten the X gear belt and tighten 2 diagonally-opposed X motor mounting screws.
16) Vibrate the X gear belt and measure the frequency with the gear belt tension adjustment tool
(P/N: 2200409). The frequency should be between 62 Hz and 80 Hz. If it is OK, continue with
step 17). If it isn't, go back to step 15).
20) Replace the probe with the dummy probe (P/N: 2200424).
WARNING: Use extreme caution when using these tools, as all moving parts can be
engaged.
29)
Manually move the arm towards the right, in order to reveal
the second set screw of the X pulley. Loosen this screw.
33) Manually move the X pulley around the X motor shaft, in order to center the dummy probe in
the cone of the X movement adjustment tool.
37) Manually move the arm towards the right, in order to reveal the second set screw of the X
pulley. Tighten this screw.
45) Remove the X movement adjustment tool and replace the dilution/wash station.
46) Remove the dummy probe and replace the sample probe.
47) Unplug the front cover sensor simulator and the front cover switch simulator.
1) Shut down the D-10, turn off the power and unplug the external power cord.
6) Manually move the sample arm to the right side of the instrument.
9) Push (or pull) the X motor towards the right, in order to loosen the X gear belt.
10) Remove the 4 screws, 4 spring lock washers and 4 flat washers that secure the X motor to
the sample arm. CAREFULLY remove the X motor.
12) Install the new X gear belt on the sample arm and secure it with the 2 screws, 2 spring lock
washers and small metal plate. DO NOT OVERTIGHTEN.
15) Tighten the X gear belt and tighten 2 diagonally-opposed X motor mounting screws.
16) Vibrate the new X gear belt and measure the frequency with the gear belt tension adjustment
tool (P/N: 2200409). The frequency should be between 62 Hz and 80 Hz. If it is OK, continue
with step 17). If it isn't, go back to step 15).
6.122 Link Cable: HPLC Module Board > Front Cover Solenoid
Check the serial number of the D-10. For instruments DA4C324721, DA4C324730 or with a
serial number ≥DA4D337101, go to step 11).
4) Connect the 2 plugs of the new link cable to the HPLC module board.
Steps 11) to 21) apply ONLY to instruments DA4C324721, DA4C324730 or with a serial
number ≥DA4D337101.
12) Unlock the 2 captive screws and completely pull out the HPLC drawer.
15) Connect the 2 plugs of the new link cable to the HPLC module board.
17) Push the HPLC drawer back in and tighten the 2 captive screws.
Check the serial number of the D-10. For instruments DA4C324721, DA4C324730 or with a
serial number ≥DA4D337101, go to step 15).
6) Remove the +24VDC security cable. Be careful, as there might be an additional cable holder
underneath the rack drive module.
7) Connect the first end of the new +24VDC security cable to the supply cable between the
power boards and the interface board.
8) Put the new security cable (along with the 2 other cables) in the 2 plastic cable holders and
close them.
9) Route the cable to the rack drive module and install it on the bracket.
10) Push the HPLC drawer back in and tighten the 2 captive screws.
Steps 15) to 27) apply ONLY to instruments DA4C324721, DA4C324730 or with a serial
number ≥DA4D337101.
15) Unlock the 2 ¼ turn fasteners which lock the left door and open it.
20) Connect the first end of the new +24VDC security cable to the supply cable between the
power boards and the interface board.
22) Put the new security cable (along with the 2 other cables) in the 2 plastic cable holders and
close them.
The procedure is exactly the same with the new type of security cable. The additional cable
length must be secured with the 2 plastic cable holders also.
24) Close the left door and tighten the 2 ¼ turn fasteners.
6.124 Reference
Section 6.42: RheBuild Kit Package Insert; Rheodyne, Rohnert Park, California.
7 ADJUSTMENTS
Warning: when using the 50mL syringe to prime the high-pressure circuit, do not
pull excessively on the plunger as it may damage the proportioning valve.
Set the offset and gain on each channel (sample measurement and reference
measurement).
Tools/Materials
7.2.1 Preparation
Figure 7-1: Detector Board: Adjustment Trimmers and Test Point Locations
7.2.3 Procedure
3. Connect the front cover sensor simulator (P/N: 2200415) and the front cover switch
simulator (P/N: 2200416).
4. Remove the detector board cover. Replace the
screw into the hole of the detector to avoid any
external light.
8. Check the LED source electrical intensity. Measure the voltage between J test point
(+) and N test point (-). The value should be 150mV (±10mV):
• if the measured value is out of tolerance, check the LED source (ohmmeter,
low impedance in one direction)
• if the LED is OK, replace the detector board (Section 6.83)
9. On the DETECTOR (1/4) screen, set the Flow
rate to 1.5ml/min and click on START PUMP.
Note: because of the possible saturation of the first stage, the value may increase
for a short time during the rotation. Keep on turning until the minimum value is
reached.
12. Rotate the Reference Gain Trimmer counter-clockwise until the minimum value is
reached.
Note: because of the possible saturation of the first stage, the value may increase
for a short time during the rotation. Keep on turning until the minimum value is
reached.
13. Rotate the Sample Gain Trimmer clockwise until the displayed value is above
12,079,000 (-0, +300,000).
If the target value cannot be reached because the current value is lower than the
target, adjust to the maximum value. If the maximum value is lower than 10,000,000
the LED source must be replaced. (Section 6.86)
If the target value cannot be reached because the current value is higher than the
target, the LED source has a higher intensity than usual. Adjust to the minimum
value. Check if the value is lower than the saturation value (16,777,215).
14. Rotate the Reference Gain Trimmer clockwise until the displayed value of the
absorbance (Log.) is equal to 0.020AU (±0.002AU):
• if it is OK, go to step 16.
• if it isn't, continue with step 15.
15. If the Log. adjustment could not be made, rotate the Reference Gain Trimmer to the
maximum value.
If the maximum value is lower than 10,000,000 the LED source must be replaced.
(Section 6.86)
If the maximum value is higher than 10,000,000 rotate the Sample Gain Trimmer
until the displayed value of the absorbance (Log.) is equal to 0.020AU (±0.002AU).
16. Check the Offsets:
• enter 0 for Set Light (LED switched off)
• rotate the Sample Offset Trimmer until the displayed Abs. Meas. value is
between 1000 and 7000
• rotate the Reference Offset Trimmer until the displayed Ref. Meas. value is
between 1000 and 7000
17. Stop the high-pressure pump.
18. Exit the Service software.
19. Replace the detector board cover.
20. Unplug the front cover sensor simulator and the front cover switch simulator.
21. Push the HPLC drawer back in and tighten the 2 captive screws.
22. Replace the front cover assembly.
Determine the Waiting Count (WC) and Filling Count (FC) of the high-pressure
pump, with the help of an automated process.
The result *.csv files produced during each step of the determination are identified in
the following way:
YYYYMMDD_hhmmss_xxxx_WC-FC.csv
Where:
• "YYYYMMDD" is the file creation date (year, month, day format).
Tools/Materials
♦ WC-FC auto determination kit (P/N: 2200447). This kit consists of:
• a reagent bottle dip tube.
• a bottle cap and reagent tubing.
• a Teflon tubing coil (similar to the degasser buffer tubing).
• a reagent tubing to the proportioning valve NC inlet.
• a dye kit (P/N: 2200446) for the first use.
Except for the reagent bottle dip tube, these parts are pre-assembled in the kit and
sealed at each end. These parts are used to connect the dye solution bottle to the
proportioning valve through external tubing. The Teflon tubing coil is used to
provide similar hydrostatic pressure to that inside the degasser.
7.3.1 Preparation
P 1 Installation
P 1.0 The goal is to have an installation as
described on the picture.
P 1.1 Remove the buffer 2 bottle from the reagent tray and place it at bench level. It is not
necessary to disconnect the line.
P 1.2 Install the dye bottle on the reagent tray.
P 1.3 Remove the ¼" plug from the reagent tubing bottle cap and connect the reagent
bottle dip tube.
P 1.4 Plunge the dip tube in the dye bottle and tighten the cap.
P 1.5 Remove the ¼" cap from the other end of the
reagent line.
P 1.6 Open the cartridge module, by pulling the black metal cover all the way to the left.
P 1.7 Remove the cartridge holder and replace it with the dummy PEEK cartridge holder
(P/N: 2200435).
P 1.8 Remove the rubber plug from the cartridge module door and close the door.
P 1.9 Connect the backpressure coil between the
inlet and outlet of the dummy PEEK cartridge
holder.
P 1.12 Click on START PROP. VALVE and prime the buffer 2 line (dye solution) with the
50mL syringe. Stop the proportioning valve when the syringe is filled (40mL
minimum) and empty the syringe.
P 1.13 Leaving the Flow rate unchanged, click on START PROP. VALVE and then on
START PUMP. Monitor the pressure for approximately 5 minutes; if it fluctuates by
more than ±5%, manually prime the pump using the 50mL syringe and re-check the
pressure.
P 1.14 At the end, click on STOP PROP. VALVE.
P 1.15 Aspirate an additional 10mL of buffer 1, in order to remove the dye solution from the
Teflon tube between the proportioning valve outlet and the HP pump inlet.
P 2 Detector Check
P 2.1 Go to the DETECTOR (1/4) screen:
• set the Flow rate to 1.5ml/min
• click on START PUMP
• monitor the pressure for approximately
5 minutes; if it fluctuates by more than
±5%, manually prime the pump using
the 50mL syringe and re-check the
pressure
P 2.2 Even if already displayed, enter the following values to set the detector to its
standard conditions:
• 150 for Set Light
• 64 for Set Gain
P 2.3 If the Abs. Meas. is not stable (average reading fluctuating ±2000 or less is typical
with the pump running), remove the bubbles from the detector (see 7.2.3).
P 2.4 If necessary, modify the Set Gain value to obtain a Log. value between 0.010AU
and 0.035AU. Verify that the Log. value stops drifting before any test is started. The
drift should be less than 0.4mAU per minute (~10 to 15 minutes LED warmup
required).
P 2.5 Stop the high-pressure pump.
P 3 Leak Check
P 3.1 Go to the HP: PUMP + VALVE (1/17) screen and
make sure that the HP pump and proportioning
valve are switched off.
P 3.2 Unscrew the cap from the dye solution bottle and remove the dip tube from the
bottle.
P 3.3 Observe the meniscus in the dip tube. Check that there is no movement of the
meniscus under the following conditions:
• PROP. VALVE off and PUMP off
• PROP. VALVE off and PUMP on
If the meniscus moves down the tube, it means that the proportioning valve or buffer
tubing is leaking. Replace the leaking part.
P 3.4 Place the dip tube into the bottle and tighten the cap. Then stop the HP pump.
P 3.5 Check for air bubbles in the line. Prime again if necessary.
P 3.6 Click on START PROP. VALVE.
P 3.7 Unscrew the cap from the buffer 1 bottle and remove the dip tube from the bottle.
P 3.8 Observe the meniscus in the dip tube. Check that there is no movement of the
meniscus under the following conditions:
• PROP. VALVE on and PUMP off
• PROP. VALVE on and PUMP on
If the meniscus moves down the tube, it means that the proportioning valve or buffer
1 tubing is leaking. Replace the leaking part.
P 3.9 Place the dip tube into the bottle and tighten the cap. Then stop the HP pump and
stop the proportioning valve.
P 3.10 Check for air bubbles in the line. Prime again if necessary.
Click on CLOSE.
2 Result Memorization
2.1 The test ends:
• after the Continuous gradient test is done with the BEST WC/FC couple
• after testing all WC/FC couples in the selected range (if the specifications
are not met)
In the second case, try to find the reason for the abnormal results and make the
appropriate corrections before repeating the automatic test. For example, check the
pressure by selecting the CURVE button of the WC/FC DETERMINATION TEST RESULTS
screen.
If all is OK, modify the ranges for the WC and/or the FC and re-do the automatic test
(WC: Min = 149; FC: Max =114 for example).
2.2 Check the step gradient results on the WC/FC
DETERMINATION TEST RESULTS screen:
• if the results are not OK, select
another WC/FC couple which was OK
and re-do the Step gradient test
• if the results are not OK and no other
WC/FC couple is not available, go
back to step 2.1
• if the results are OK, record the final
WC/FC values
1.5 Click on STOP PROP. VALVE and aspirate an additional 10mL of buffer 1, in order
to remove the dye solution from the Teflon tube between the proportioning valve
outlet and the HP pump inlet.
1.6 Remove the bottle from the reagent tray.
1.7 Unscrew the bottle cap and remove it.
1.8 Disconnect the dip tube from the bottle cap and replace the ¼" plug on the cap.
1.9 Empty and dry the dip tube.
1.10 Remove the ¼" cap from the buffer 2 reagent tubing.
1.11 Disconnect the fitting at the proportioning valve (NC inlet) and place the ¼" cap on
the fitting.
2.4 Leaving the Flow rate unchanged, click on START PROP. VALVE and then on
START PUMP. Monitor the pressure for approximately 5 minutes; if it fluctuates by
more than ±5%, manually prime the pump using the 50mL syringe and re-check the
pressure.
2.5 At the end, click on STOP PROP. VALVE.
2.6 Disconnect and remove the backpressure coil.
2.7 Remove the dummy PEEK cartridge holder and replace the analytical cartridge.
2.8 Replace the rubber plug in the cartridge module door and close the door.
2.9 Go to the DETECTOR (1/4) screen:
• set the Flow rate to 1.5ml/min
• click on START PUMP to remove the
air bubbles introduced in the flow cell
• if the Abs. Meas. is not stable
(average reading fluctuating ±2000 or
less is typical with the pump running),
remove the bubbles from the detector
(see 7.2.3)
Verify the Waiting Count (WC) and Filling Count (FC) parameters found in the
'instrument.ini' file.
Process Description
The Static and Step gradient tests are mandatory. The Dilamex+ and the Continuous
gradient tests are optional.
Tools/Materials
Except for the reagent bottle dip tube, these parts are pre-assembled in the kit and
sealed at each end. These parts are used to connect the dye solution bottle to the
proportioning valve through external tubing. The Teflon tubing coil is used to
provide similar hydrostatic pressure to that inside the degasser.
7.4.1 Preparation
7.4.2 Procedures
Objective
Determine the relative absorbance of the dye solution used instead of buffer 2,
when the proportioning valve is continuously on.
1 Measurement
1.1 Go to the HP: PUMP + VALVE (10/17) screen:
• set the Flow rate to 1.5ml/min
• click on START PUMP
1.2 Click in the white field next to Selected log file and select a file or create a new file.
1.3 Click on START TEST.
1.4 At the end of the test (Test remaining time at 0), click on STOP PUMP.
2 Calculation
2.1 On the HP: PUMP + VALVE (10/17) screen, click
on START CALCULATION.
Objectives
• Measure the theoretical 20%, 50% and 100% of dye solution dynamic mixing.
• Adapt the WC and FC values in order to have a 100% dynamic buffer 2 close
to the static value, and to have values for the 20% and 50% steps within the
defined tolerances. Typical tolerances are:
20%: from 13 to 14%
50%: from 49.5 to 50.5%
1 Measurement
1.1 Go to the HP: PUMP + VALVE (11/17) screen:
• set the Flow rate to 1.5ml/min
• enter the WC and FC values to be
tested
1.2 Click in the white field next to Load gradient file and load the 'dilamex+_grad.hpl'
file.
1.3 Click on START PUMP.
1.4 Click in the white field next to Selected log file and select a file or create a new file.
1.5 Click on START TEST.
1.6 At the end of the test (Test remaining time at 0), click on STOP PUMP.
2 Calculation
2.1 Go to the HP: PUMP + VALVE (12/17) screen:
• click in the white field next to Selected
log file
• select the *.log file corresponding to
the Dilamex+ gradient test
2.3 Click on START CALCULATION. The results are displayed on the screen.
3 Result Examination
The relative value (in AU) of the 100% buffer 2 absorbance is compared to the 100%
buffer 2 relative value obtained under static conditions (see 7.4.2.1). The difference
must be less than 8mAU. The result of the comparison is given next to the 100%
buffer 2 relative value field:
√ indicates that the difference is within the tolerances; go to step 3.2
X indicates that the difference is out of range; go to step 3.4
? indicates that the static value is not available; perform the Static
Absorption Measurement procedure from Section 7.4.2.1
3.2 The relative value (in %) of the 50% of buffer 2 absorbance is compared to the
theoretical 50% value. Tolerance: from 49.5 to 50.5%. The result of the comparison
is given next to the 50% buffer 2 relative value field:
√ indicates that the difference is within the tolerances; go to step 3.3
X indicates that the difference is out of range; go to step 3.4
3.3 The relative value (in %) of the 20% of buffer 2 absorbance is compared to the
theoretical 20% value. Tolerance: from 13 to 14%. The result of the comparison is
given next to the 20% buffer 2 relative value field:
√ indicates that the difference is within the tolerances; go to step 3.5
X indicates that the difference is out of range; go to step 3.4
3.4 If one result is out of range, the WC/FC values should be modified. Refer to Section
7.3 Automatic WC/FC Determination.
3.5 If the results are OK, note down the WC and FC values and introduce them in the
'instrument.ini' file. (Section 7.4.2.5)
Objectives
• Measure the response curve of the buffer mixing for the WC and FC values.
• Check the theoretical 20% and 50% points.
1 Measurement
1.1 Go to the HP: PUMP + VALVE (13/17) screen:
• set the Flow rate to 1.5ml/min
• enter the WC and FC values to be
tested
1.2 Click in the white field next to Load gradient file and load the 'step_grad.hpl' file.
2 Calculation
2.1 Go to the HP: PUMP + VALVE (14/17) screen:
• click in the white field next to Selected
log file
• select the *.log file corresponding to
the Step gradient test
2.2 Click on START CALCULATION. The results are displayed on the screen.
3 Result Examination
3.1 The relative value (in AU) of the 100% buffer 2 absorbance is compared to the 100%
buffer 2 relative value obtained under static conditions (see 7.4.2.1). The difference
must be less than 8mAU. The result of the comparison is given next to the 100%
buffer 2 relative value field:
√ indicates that the difference is within the tolerances; go to step 3.2
X indicates that the difference is out of range; go to step 3.4
? indicates that the static value is not available; perform the Static
Absorption Measurement procedure from Section 7.4.2.1
3.2 The relative value (in %) of the 50% of buffer 2 absorbance is compared to the
theoretical 50% value. Tolerance: from 49.5 to 50.5%. The result of the comparison
is given next to the 50% buffer 2 relative value field:
√ indicates that the difference is within the tolerances; go to step 3.3
X indicates that the difference is out of range; go to step 3.4
3.3 The relative value (in %) of each step (10 to 90%) is compared to a tolerance
window. Typical tolerance ranges are:
10%: 3 to 10% 60%: 60 to 68%
20%: 13 to 14% 70%: 70 to 79%
30%: 21 to 30% 80%: 80 to 89%
40%: 32 to 40% 90%: 90 to 97%
3.5 If the results are OK, note down the WC and FC values and introduce them in the
'instrument.ini' file. (Section 7.4.2.5)
Objectives
1 Measurement
1.1 Go to the HP: PUMP + VALVE (15/17) screen:
• set the Flow rate to 1.5ml/min
• enter the WC and FC values to be
tested
1.2 Click in the white field next to Load gradient file and load the 'cont_grad.hpl' file.
1.3 Click on START PUMP.
1.4 Click in the white field next to Selected log file and select a file or create a new file.
1.5 Click on START TEST.
1.6 At the end of the test (Test remaining time at 0), click on STOP PUMP.
2 Calculation
2.1 Go to the HP: PUMP + VALVE (16/17) screen:
• click in the white field next to Selected
log file
• select the *.log file corresponding to
the Continuous gradient test
3 Result Examination
3.1 Verify that the curve has the appropriate shape (see the example given in step 2.1):
• if it is OK, go to step 3.3
• if it isn't, go to step 3.2
3.2 If the curve has not the appropriate shape, the WC/FC values should be modified.
Refer to Section 7.3 Automatic WC/FC Determination.
3.3 If the results are OK, note down the WC and FC values and introduce them in the
'instrument.ini' file. (Section 7.4.2.5)
7.4.2.5 Entering the WC and FC Values
Process Description
The Static and Step gradient tests are mandatory. The Dilamex+ and the Continuous
gradient tests are optional.
1 Measurement
1.1 Go to the HP: PUMP + VALVE (9/17) screen and
click on AUTOMATIC TESTS.
In the next field, enter the number of repetition(s) for each test: 1.
1.5 Launch the sequence by clicking on the START TEST button.
1.6 The sequence ends when the Total remaining time is at 0.
Tools/Materials
♦ dummy PEEK cartridge holder (P/N: 2200435). Unscrew the LUER fittings prior to
using it.
7.5.1 Preparation
7.5.2 Procedure
1 Measurement
1.1 Go to the HP: PUMP + VALVE (1/17) screen:
• select psi as unit for the Pressure
reading
• set the Flow rate to 1.5ml/min
• click on START PUMP
1.2 Adjust the backpressure regulator until the pressure sensor verification tool reads
approximately 500psi.
1.3 After stabilization, check the value displayed on the D-10 screen and compare it to
the value on the pressure sensor verification tool. Tolerance: less than ±5%.
1.4 At the end, click on STOP PUMP.
2 Result Examination
2.1 If the result is not within tolerance, perform the Pressure Sensor Calibration
procedure from Section 7.6.
If the result is acceptable, go to step 2.2.
2.2 Disconnect the pressure sensor verification tool and 1500psi backpressure regulator
from the dummy PEEK cartridge holder.
2.3 Remove the dummy PEEK cartridge holder and replace the analytical cartridge.
2.4 Replace the rubber plug in the cartridge module door and close the door.
3 Detector Check
3.1 Go to the DETECTOR (1/4) screen:
• set the Flow rate to 1.5ml/min
• click on START PUMP
3.2 Even if already displayed, enter the following values to set the detector to its
standard conditions:
• 150 for Set Light
• 64 for Set Gain
3.3 If the Abs. Meas. is not stable (average reading fluctuating ±2000 or less is typical
with the pump running), remove the bubbles from the detector (see 7.2.3).
3.4 Stop the high-pressure pump.
3.5 Select Set Light and enter 0 to switch off the LED source.
Tools/Materials
♦ 1500psi backpressure regulator with fittings – identical to the HP pump flow rate
calibration/ verification tool (P/N: 2200439).
♦ dummy PEEK cartridge holder (P/N: 2200435). Unscrew the LUER fittings prior to
using it.
♦ pulse dampener (P/N: 1957226).
7.6.1 Preparation
7.6.2 Procedure
1 Measurements
1.1 Go to the HP: PUMP + VALVE (1/17) screen:
• set the Flow rate to 1.5ml/min
• click on START PUMP
Note: you must leave and re-enter the Service software for the new parameters to
take effect.
2.3 Go back to the HP: PUMP + VALVE (1/17) screen
and click on STOP PUMP.
3 Verification
3.1 Disconnect and remove the pulse dampener from the circuit.
3.2 Re-connect the orange PEEK tubing from the HP valve to the outlet of the pressure
sensor.
3.3 Connect the pressure sensor verification tool to the inlet of the dummy PEEK
cartridge holder.
3.4 Perform the Pressure Sensor Verification procedure from Section 7.5.
Check the HP pump flow rate at 1.5mL/min and 500psi. The measurement uses either
the volumetric method (see 7.7.2) or the gravimetric method (see 7.7.3).
This procedure must be performed after replacement of any pump components where
faults (pressure leakage) did not accompany insufficient flow.
Tools/Materials
7.7.1 Preparation
1 Measurement
1.1 Go to the HP: FLOW RATE CALIB. (4/4) screen
and select volume from the pull-down menu
next to Method selection.
1.10 When the meniscus reaches the 5mL mark, select STOP TEST. The Remaining
time is indicated on the screen. It should be approximately 40 seconds.
Note: if the Remaining time is 0, the test is automatically stopped. If the value 0 is
obtained before the 5mL vessel is filled, there is a problem with the delivery by the
HP pump or the selected duration was not 4 minutes.
1.11 Click in the white field next to Volume and enter 5. The Time field is automatically
updated and is calculated from the time of START TEST was selected.
1.12 Click on the LAUNCH CALC. button. The result is displayed in the Calculated field.
The calculated value is compared to the theoretical value. Tolerance: 5%.The result
of the comparison is given next to the Calculated field:
√ indicates that the difference is within the tolerances
X indicates that the difference is out of range; perform the Flow Rate
Calibration procedure from Section 7.8
3 Detector Check
3.1 Go to the DETECTOR (1/4) screen:
• set the Flow rate to 1.5ml/min
• click on START PUMP
3.2 Even if already displayed, enter the following values to set the detector to its
standard conditions:
• 150 for Set Light
• 64 for Set Gain
3.3 If the Abs. Meas. is not stable (average reading fluctuating ±2000 or less is typical
with the pump running), remove the bubbles from the detector (see 7.2.3).
3.4 Stop the high-pressure pump.
3.5 Select Set Light and enter 0 to switch off the LED source.
1 Measurement
1.1 Go to the HP: FLOW RATE CALIB. (4/4) screen
and select weight from the pull-down menu
next to Method selection.
1.2 Enter the density of the liquid to be used in the Density field:
• buffer 1: typically 1.001g/ml
• distilled or deionized water: 1.000g/ml
1.3 Set the Flow rate to 1.5ml/min.
1.4 Click in the white field next to Duration and enter 10. If the value indicated in the
Remaining time field is different from 10 minutes, reset this value by clicking on the
TIMER RESET button.
1.5 Check the backpressure at this flow rate before beginning the test:
• click on START PUMP
• if necessary, adjust the backpressure regulator to a pressure of 500psi
(mean value)
1.6 Check for leaks in the circuit.
1.7 Click on STOP PUMP.
1.8 Empty the vessel and dry it thoroughly.
1.9 Weigh the empty vessel and note down this value.
1.10 Replace the tubing from the backpressure regulator into the vessel.
1.11 Click on START TEST.
1.12 At the end of the test, weigh the vessel, being careful NOT TO SPILL the liquid.
1.13 With the help of the value measured at step 1.9, calculate the weight of the liquid
delivered by the HP pump (in g).
1.14 Click in the white field next to weight and enter the weight of the liquid delivered by
the HP pump (in g). The Time field is automatically updated.
1.15 Click on the LAUNCH CALC. button. The result is displayed in the Calculated field.
The calculated value is compared to the theoretical value. Tolerance: 5%.The result
of the comparison is given next to the Calculated field:
√ indicates that the difference is within the tolerances
X indicates that the difference is out of range; perform the Flow Rate
Calibration procedure from Section 7.8
2.2 Disconnect the pressure sensor verification tool and 1500psi backpressure regulator
from the dummy PEEK cartridge holder.
2.3 Remove the dummy PEEK cartridge holder and replace the analytical cartridge.
2.4 Replace the rubber plug in the cartridge module door and close the door.
3 Detector Check
3.1 Go to the DETECTOR (1/4) screen:
• set the Flow rate to 1.5ml/min
• click on START PUMP
3.2 Even if already displayed, enter the following values to set the detector to its
standard conditions:
• 150 for Set Light
• 64 for Set Gain
3.3 If the Abs. Meas. is not stable (average reading fluctuating ±2000 or less is typical
with the pump running), remove the bubbles from the detector (see 7.2.3).
3.4 Stop the high-pressure pump.
3.5 Select Set Light and enter 0 to switch off the LED source.
• Measure the flow rate of the HP pump with the theoretical parameters.
• Calculate the correction.
• Check the result of the correction.
Tools/Materials
7.8.1 Preparation
Objective
Measure the time it takes to fill a volumetric vessel at a defined flow rate and
backpressure. A calculation is performed using the theoretical and measured flow
rates.
Note: if the Remaining time is 0, the test is automatically stopped. If the value 0 is
obtained before the 5mL vessel is filled, there is a problem with the delivery by the
HP pump or the selected duration was not 6 minutes.
1.11 Click in the white field next to Volume and enter 5. The Time field is automatically
updated and is calculated from the time of START TEST was selected.
1.12 Click on the LAUNCH CALC. button. The result is displayed in the 1.0 field. Check
the calculated value. If the value differs from the set flow rate by more than 15%,
there is an error in the measurement and the test must be repeated (go back to step
1.1).
1.13 Empty the vessel and dry it thoroughly.
2.3 Check the backpressure at this flow rate before beginning the test:
• click on START PUMP
• if necessary, adjust the backpressure regulator to a pressure of 500psi
(mean value)
2.4 Check for leaks in the circuit.
2.5 Click on STOP PUMP.
2.6 Empty the vessel and dry it thoroughly.
2.7 Replace the tubing from the backpressure regulator into the vessel.
2.8 Click on START TEST.
2.9 When the meniscus reaches the 5mL mark, select STOP TEST. The Remaining
time is indicated on the screen. It should be approximately 40 seconds.
Note: if the Remaining time is 0, the test is automatically stopped. If the value 0 is
obtained before the 5mL vessel is filled, there is a problem with the delivery by the
HP pump or the selected duration was not 4 minutes.
2.10 Click in the white field next to Volume and enter 5. The Time field is automatically
updated and is calculated from the time of START TEST was selected.
2.11 Click on the LAUNCH CALC. button. The result is displayed in the 1.5 field. Check
the calculated value. If the value differs from the set flow rate by more than 15%,
there is an error in the measurement and the test must be repeated (go back to step
1.1).
2.12 Empty the vessel and dry it thoroughly.
Note: if the Remaining time is 0, the test is automatically stopped. If the value 0 is
obtained before the 5mL vessel is filled, there is a problem with the delivery by the
HP pump or the selected duration was not 3 minutes.
3.10 Click in the white field next to Volume and enter 5. The Time field is automatically
updated and is calculated from the time of START TEST was selected.
3.11 Click on the LAUNCH CALC. button. The result is displayed in the 2.0 field. Check
the calculated value. If the value differs from the set flow rate by more than 15%,
there is an error in the measurement and the test must be repeated (go back to step
1.1).
3.12 Empty the vessel and dry it thoroughly.
Objective
Measure the weight of a liquid delivered by the pump at three different flow rates for
a defined time and backpressure. A calculation is performed using the theoretical and
measured flow rates.
1.2 Enter the density of the liquid to be used in the Density field:
• buffer 1: typically 1.001g/ml
• distilled or deionized water: 1.000g/ml
2.13 Click on the LAUNCH CALC. button. The result is displayed in the 1.5 field. Check
the calculated value. If the value differs from the set flow rate by more than 15%,
there is an error in the measurement and the test must be repeated (go back to step
1.1).
2.14 Empty the vessel and dry it thoroughly.
Tools/Materials
7.9.1 Preparation
P 1. Open the cartridge module, by pulling the black metal cover all the way to the left.
P 2. Remove the cartridge holder and replace it with the cartridge module temperature
control tool (P/N: 2200438).
P 3. Remove the rubber plug from the cartridge module door and close the door.
P 4. Connect the sensor wire to the Fluke™ multimeter.
P 5. Close the lower front door.
7.9.2 Procedure
1 Measurement
1.1 Go to the THERMAL CONTROL (1/7) screen.
1.2 Compare the values displayed on the D-10 screen and the Fluke™ multimeter.
Tolerance: ±1.6°C.
If the values are out of range, perform the Thermal Module Calibration procedure
from Section 7.10.
3 Detector Check
3.1 Go to the DETECTOR (1/4) screen:
• set the Flow rate to 1.5ml/min
• click on START PUMP
3.2 Even if already displayed, enter the following values to set the detector to its
standard conditions:
• 150 for Set Light
• 64 for Set Gain
3.3 If the Abs. Meas. is not stable (average reading fluctuating ±2000 or less is typical
with the pump running), remove the bubbles from the detector (see 7.2.3).
3.4 Stop the high-pressure pump.
3.5 Select Set Light and enter 0 to switch off the LED source.
Tools/Materials
7.10.1 Preparation
P 3. Connect the front cover sensor simulator (P/N: 2200415) and the front cover switch
simulator (P/N: 2200416).
P 4. Switch on the thermometer and set the range to "-199.9 TO 199.9°C" (one decimal
digit).
Objective
Determine the gain and offset of the electronic interface (slope and intercept).
1. Connect the low temperature control set (blue color) to the thermometer using the
adapter wire and note down the LT EXT value (in °C).
2. Connect the high temperature control set (red color) to the thermometer using the
adapter wire and note down the HT EXT value (in °C).
3. To avoid overheating during the calibration, open the Peltier module circuit by
removing the plug at CN10 on the HPLC module board (see Figure 7-4).
4. Disconnect the temperature sensor (labeled '4') from the HPLC module board (see
Figure 7-4).
5. Connect the low temperature control set (blue color) to the HPLC module board, at
CN4.
6. Go to the THERMAL CONTROL (5/7) screen. The
system should display a Current calc. temp.
of approximately 20 - 22°C.
Note: the new parameters are immediately taken in account by the system.
18. Remove the high temperature control set from the HPLC module board and re-
connect the temperature sensor at CN4.
19. Re-connect the plug at CN10 on the HPLC module board.
20. Perform the Sensor Offset Correction Measurement procedure from Section
7.10.3.
Objective
Objective
1. Open the cartridge module, by pulling the black metal cover all the way to the left.
2. Remove the cartridge holder and replace it with the cartridge holder equipped with
the temperature sensor – available in the cartridge module temperature calibration
kit (P/N: 2200441).
3. Remove the rubber plug from the cartridge module door and close the door.
4. Connect the sensor wire to the Hanna™ thermometer.
5. Close the lower front door.
6. Go to the THERMAL CONTROL (7/7) screen.
7.10.5 Verification
Objective
2 Measurement
2.1 Go to the THERMAL CONTROL (1/7) screen.
2.2. Compare the values displayed on the D-10 screen and the Hanna™ thermometer;
they should be the same. Tolerance: ±0.2°C.
If it is not OK, repeat the Thermal Module Calibration procedure. (Section 7.10)
4 Detector Check
4.1 Go to the DETECTOR (1/4) screen:
• set the Flow rate to 1.5ml/min
• click on START PUMP
4.2 Even if already displayed, enter the following values to set the detector to its
standard conditions:
• 150 for Set Light
• 64 for Set Gain
4.3 If the Abs. Meas. is not stable (average reading fluctuating ±2000 or less is typical
with the pump running), remove the bubbles from the detector (see 7.2.3).
4.4 Stop the high-pressure pump.
4.5 Select Set Light and enter 0 to switch off the LED source.
• Check the performance of the internal waste bottle vacuum system and
compare to the standard performance.
• Check the tubing from the dilution/wash station to the internal waste bottle and
compare to the standard performance.
Tool/Material
Objective
Check the vacuum level, in order to have an indication on the liquid pump
performance and the airtightness of the internal waste circuit.
Note: in the 'Dia.exe' software, the threshold of the waste bottle check is typically set
to 29kPa.
Objectives
• Check the vacuum rise time to have an indication on the airtightness and
cleanliness of the internal waste circuit.
• Check the vacuum fall time of the dilution well circuit to detect tubing blockage
between the dilution well and the internal waste bottle.
• Check the vacuum fall time of the probe wash well circuit to detect tubing
blockage between the wash well and the internal waste bottle.
Typically the High level time should be lower than 6s (High level time out). If this
performance is not reached, there is a problem on the liquid pump. This issue should
have been detected in the previous check (see 7.11.1).
The fall time of each circuit should be less than 0.5s. If this performance is not
reached, the corresponding circuit is totally or partially pinched or blocked. Check or
replace the tubings on this circuit.
Tools/Materials
1.3 Click on the LAUNCH RACK INTRODUCTION button and introduce the sample
rack for the test.
If the results are not acceptable, perform the Hall Effect Sensor Adjustment
procedure from Section 7.13.
2.2 Remove the microvial adapter from position 7. Keep the other adapters in positions
2, 3, 4, 5, 6 and 8.
2.3 Click on the LAUNCH RACK INTRODUCTION button and re-introduce the sample
rack for the test.
2.5 The results of the test are displayed on the screen. Varying positions for the First
tube and Last tube that range from -90 to +90 are displayed. A yes at any of these
positions indicates that a magnet was detected, while no indicates that no magnet
was detected.
2.6 Expected results:
• results of the First tube should show no beginning at least at the -20
position. Results are acceptable if no appears in any position prior to -20.
• results of the Last tube should show no at least until +20. Results are
acceptable if no appears in any position after +20.
If the results are not acceptable, perform the Hall Effect Sensor Adjustment
procedure from Section 7.13.
Tools/Materials
6. If the First tube position is not acceptable, shift the Hall effect sensor towards the
right side of the instrument:
• CAREFULLY modify the sensor orientation with a pair of pliers
• modify the position of the whole board
If the Last tube position is not acceptable, shift the Hall effect sensor towards the
left side of the instrument, as described above.
Tools/Materials
4. The result of the test is displayed in the Calculated field. Check the value.
Tolerance: ±200.
5. If the Calculated value is within range, click on the SAVE RESULTS button.
If the Calculated value is out of range, adjust the tube height sensor position as
described in Section 6.41.
6. Perform the Tube Identification Check from Section 7.15.
Tools/Materials
Tools/Materials
1. Insert the barcoded primary tubes in all positions of the sample rack.
If it is impossible to have 10 tubes at the same time, repeat the procedure until all
tube positions are checked.
Note the barcode label values for each position.
2. Go to the RACK INTRO. (1/1) screen.
3. Click on the LAUNCH RACK INTRODUCTION button and introduce the sample
rack for the test.
4. Go to the BARCODE (2/2) screen and click on
START.
5. The X pos. field indicates the exact X position (in steps from the home position) at
which the barcode was read.
The entire part of the Tube nb indicates the tube number identified by the software.
The decimal part of the Tube nb indicates the range for each tube. The nominal
range is from 0.2 to 0.8.
The results of the test are given at the end of each line:
✔ indicates that the result is in the standard tube range
✘ indicates that the result is out of range
6. Compare the barcode label values from step 1. with the Result of the reading field.
7. If all tubes give the expected result, go to step 8.
If at least one result is out of range, adjust the barcode reader position as described
in Section 6.31.
Tools/Materials
Objective
Modify the interface board to increase the sensitivity of the front cover sensor signal
interface circuit.
This procedure must be performed if an error appears on the front cover sensor: "front
door open" message while the door is closed.
Tools/Materials
1. Shut down the D-10, turn off the power and unplug the external power cord.
2. Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
5. Change the version number indicated on the label. Add an "A" with a pencil:
• from "DAE010+2" to "DAE010+2A"
• from "DAE010+3" to "DAE010+3A"
6. Close the left door and tighten the 2 ¼ turn fasteners.
7. Replace the external power cord.
8. Switch on the D-10 and launch the Service
software:
• go to the MAINTAIN screen
• click on the Service button
• click on Exit to Service software
Backup the 'instrument.ini' file, which is specific for each D-10 unit.
This procedure must be performed after any modification of the 'instrument.ini' file, or
as a preventive maintenance.
Process Description
These backups are located inside the left door (see 3.3.13). The floppy disk is
identified with a sticker:
D-10: backup of
INSTRUMENT.INI
S/N: DAxXxxxxxx
Tools/Materials
1.5 A confirmation screen opens. If the confirmation is given, the copy begins.
1.6 If the copy was successful, the instrument.ini
file should appear in the Files in Floppy field.
2.4 A confirmation screen opens. If the confirmation is given, the printing begins.
2.5 At the end of the printing, it is recommended to make a photocopy printout on
standard paper.
Tools/Materials
7.20.1 Preparation
P1 Perform the 'instrument.ini' File Backup procedure from Section 7.19 (if possible).
7.20.2 Procedure
1 Imaging the Hard Disk with Image Boot Disk Version 1.0
1.1 If necessary, replace the hard disk. (Section 6.30)
1.2 Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
1.3 Connect the CD-ROM drive to the hard
disk flat cable and the spare power
supply cable. You may use the CD-ROM
flat cable extension (P/N: 2200402) for
easier connection. Make sure that the
drive is configured as 'slave' device prior
to connecting it.
1.4 Connect a keyboard and a VGA screen to the D-10 rear panel.
1.5 Insert the image boot disk (P/N: 2702214) in the floppy drive.
1.6 Switch on the D-10. It will automatically boot from the floppy.
1.7 Select 1 on the keyboard for Restore a
Partition or Drive.
1.14 Remove the image boot disk and the 2.22 image CD from their respective drives.
1.15 Switch off the D-10.
1.16 Disconnect the keyboard and VGA screen.
1.17 Disconnect the CD-ROM drive and the CD-ROM flat cable extension (if used).
2 Imaging the Hard Disk with Image Boot Disk Version 1.2 or later
2.1 If necessary, replace the hard disk. (Section 6.30)
2.2 Unscrew the 2 ¼ turn fasteners which lock the left door and open it.
2.3 Connect the CD-ROM drive to the hard
disk flat cable and the spare power
supply cable. You may use the CD-ROM
flat cable extension (P/N: 2200402) for
easier connection. Make sure that the
drive is configured as 'slave' device prior
to connecting it.
Note: the image boot disk will convert back to a V2 style disk on its own after 1 use.
2.6 Insert the image boot disk in the floppy drive.
2.7 Switch on the D-10. It will automatically boot from the floppy.
2.8 Insert the 2.22 image CD (P/N: 2200443) into the CD-ROM drive. Imaging will take
from 1 to 2 minutes.
2.9 Remove the image boot disk and the 2.22 image CD from their respective drives.
2.10 Switch off the D-10.
2.11 Disconnect the CD-ROM drive and the CD-ROM flat cable extension (if used).
3.6 At the end of the copy, remove the disk from the floppy drive and replace it in the left
door.
4.8 Take the "D-10 MC Firmware (1/1)" disk in the 3.0B upgrade kit (P/N: 2200445) and
insert it in the floppy drive.
4.14 Disconnect the keyboard and VGA screen from the D-10.
4.15 Close the left door and tighten the 2 ¼ turn fasteners.
Process Description
The Laboratory Information System (LIS) stores and manages information regarding
patients, tests ordered for patients, and the results of the tests which have already
been performed. When a patient sample is collected or when it arrives at the
laboratory, it is assigned a sample identification. The sample ID is associated with the
patient, with the tests to be performed, and with the results of the sample. The
connection between the D-10 and the LIS allows the results of the sample IDs to be
transferred from the D-10 to the LIS. This transfer of results can be initiated by the LIS
or by the D-10.
To establish a connection between the D-10 and the LIS, both systems must have
compatible hardware. A null modem cable must connect the systems through RS232
ports.
To reduce the risk of incompatibility between the systems, the NCCLS standards for
the protocol transfers were followed. The low level protocol is specified in NCCLS
standard LIS1-A, Specification for Low Level Protocol to Transfer Messages Between
Clinical Laboratory Instruments and Computer Systems. The high level protocol is
specified in NCCLS standard LIS2-A, Specification for Transferring Information
Between Clinical Instruments and Computer Systems.
Tools/Materials
Export Activated: YES or NO. YES must be selected for the LIS communication.
Export to LIS: allows the setting of the automatic export to LIS after each run, after each
injection or manual export only.
LIS ID #: allows an LIS ID to be defined (0 to 99).
8 TROUBLESHOOTING
4. Prediluted sample not mixed 4. Re-make the prediluted sample, mix and re-run.
Low Total Area
5. Leak at dilution chamber 5. Call Technical Service.
Check wash/diluent bottle volume and verify wash line is
6. Air in sample lines or diluent syringe 6.
properly placed in bottle. Perform system flush.
Uptake line clogged with undissolved
7. 7. Flush the sample uptake lines.
sample particulate
Manually dilute sample at higher dilution ratio (1:400) and
Check the sensors in the Service software: axis sensor and home
reached.
sensor. If a sensor does not give the right answer, check it by
5. connecting it to the plug of another sensor (see 8.4). You may use
the 5-pin extension lead (P/N: 2200417). Replace the sensor(s) if
necessary. (Sections 6.67 & 6.68)
Check if the X motor can close (high effort) and open (low effort)
the rack door. If the motor cannot open the door, check the rack
6.
door movement. If the rack door movement is stiff, replace the
rack door module. (Section 6.16)
Check if the X movement is stiff by manually moving the sample
7. arm. If yes, the sample arm module should be replaced. (Section
6.36)
Go to the Service software. Select Sample Arm, Z Move. Select
1. the speed index 9 for the service movement (highest speed). This
Mechanical error: the Z
is the worst case for the motor (lowest torque).
position (axis) sensor did not
Check if the Z movement is not blocked in the upper position:
(E012) Z motor error sense any movement or the 2.
manually turn the pulley.
L20015501
Check the sensors in the Service software: axis sensor and home
sensor. If a sensor does not give the right answer, check it by
7. connecting it to the plug of another sensor (see 8.4). You may use
the 5-pin extension lead (P/N: 2200417). Replace the sensor(s) if
necessary. (Sections 6.22 & 6.71)
8. Check if the Z movement is noisy. If yes, lubricate the lead screw.
9. Check if the Z movement is stiff. If yes, adjust the belt tension.
board.
8 - 10
board and between the syringe module board and the interface
board. If the motor does not move, inspect the syringe module
2.
board (connectors, soldered points and wires). Replace the board
if necessary. (Section 6.51) If the syringe module board is OK,
(E014) Syringe motor error Mechanical error.
replace the motorized syringe. (Section 6.11) If the problem
persists, replace the interface board. (Section 6.15)
Check the syringe home sensor in the Service software. If the
sensor fails, inspect the syringe module board (connectors,
3. soldered points and wires). Replace the board if necessary.
(Section 6.51) If the syringe module board is OK, replace the
motorized syringe. (Section 6.11)
1. Go to the Service software. Select HP: Pump + Valve (4/17).
Check the HP valve stepper motor movement. If the motor does
not move or if it moves with a bad noise, check the electrical
connections on the HPLC module board and between the HPLC
module board and the interface board. If the motor does not
(E015) Inj. Valve motor error Mechanical error.
2. move, inspect the HPLC module board (connectors, soldered
L20015501
(E039) Error:Pressure too high over 800psi for more than 6 blockage by opening the high-pressure circuit at different spots.
seconds during the run. Check the pressure sensor. Perform the procedure from Section
3.
7.5 Pressure Sensor Verification.
The adjustment of the LED Check the flat cable going from interface board to the detector
1.
source was impossible: the board.
maximum current intensity Check the connection of the reference sensor to the detector
2.
on the LED command circuit board.
(E040) Replace Light source !!! was reached, without a If all connections are OK, replace the detector module. (Section
correct light intensity. 3.
6.32)
4. Perform the procedure from Section 9.2.5 Flow Cell Cleaning.
The flat cable going to the
detector board is broken. 5. Perform the procedure from Section 7.2 Detector Adjustment.
L20015501
L20015501
7. Dirty flow cell. 7. "Clean/Decontaminate Sampling Fluid Path" on page 5-2 of the
Operation Manual.
8 - 16
error corruption. Warning: never re-install over an existing faulty hard disk as
Analysis : Not Initialised error this will erase any hint of the actual problem. Return the
Analysis : Log Error faulty disk to Bio-Rad for investigation.
Disk or file system
Analysis : Write Init Failed
corruption.
Analysis : Not loaded
L20015501
D-10 Hemoglobin Testing System
The following table lists sensor compatibility. All sensors with the same identification
can be connected alternately for testing.
Sensor Connector
Designation Connection board Identification
Type Number
Syringe home Syringe module
Optical switch CN2 A
sensor board
CN1
Analog Degasser module
Bubble detectors CN3 B
opto-sensors board
CN6
Reflective
Front cover sensor HPLC module board CN1 C
sensor
Reflective
Rack input sensor X board CN8 C
sensor
Reflective
Rack output sensor X board CN9 C
sensor
Cam flag sensor Optical switch HPLC module board CN9 D
Closed door sensor
Optical switch X board CN7 D
(rack door)
Open door sensor
Optical switch X board CN12 D
(rack door)
X home sensor Optical switch X board CN11 D
X position sensor Optical switch X board CN10 D
Z home sensor Optical switch Z board CN2 D
Z position sensor Optical switch Z board CN13 D
TH home sensor Optical switch Z board CN12 D
TH position sensor Optical switch Z board CN5 D
HP pump encoder HP encoder HPLC module board CN8 E
Optical
HP valve sensors HPLC module board CN7 F
switches
Tube height sensor Optical switch Z board CN8 G
reversed polarity
• the thermal fuse should have a nominal short
circuited value (0Ω) for both polarities
If one value is abnormal, replace the failed component.
(Sections 6.12 & 6.56)
Check the temperature sensor (Pt100 sensor). Disconnect
Temperature check
it from the HPLC module board (connector CN4). Check
failed
• System Offset: -4 to +4
• Sensor Offset: -2 to +2
Temperature check
failed (continued)
D. Temperature interface failure. • Gradient Offset: -2 to +2
Check the parameters:
3.
• the System Gain and System Offset are indicated on
the interface board
• the Sensor Offset is indicated on a sticker on the
temperature sensor wire
• the Gradient Offset is indicated on a sticker on the
cartridge module
If the values are out of range, perform the procedure from
Section 7.10 Thermal Module Calibration.
If the temperature sensor is in the nominal range
(impedance), check the temperature measurement
interface by examining the value displayed on the Thermal
Control (1/7). Check the Calculated temperature. If not
L20015501
4.
OK, perform the procedure from Section 7.10 Thermal
Module Calibration and check if the interface board is
responding. If not OK, replace the interface board.
(Section 6.15)
L20015501
9 MAINTENANCE
This section contains the minimum requirements for performing annual Preventive
Maintenance on the D-10. Any additional maintenance items can be performed based
on regional requirements. Refer to the D-10 Operation Manual for daily, monthly and
periodic maintenance procedures.
9.1 Tools/Materials
* all these parts are included in the preventive maintenance kit (P/N: 2200320)
1. The sample probe must be removed before performing any maintenance on the
D-10.
2. The external power cord must be removed when the D-10 is switched off and
the bypass tools are in place.
9.2.2 Preparation
1. Clean and decontaminate the sample fluid path. Follow the steps in
"Clean/Decontaminate Sampling Fluid Path" on page 5-2 of the Operation
Manual.
2. Remove the front cover assembly. (Section 6.2)
3. Install the front cover sensor simulator (P/N: 2200415) and front cover switch
simulator (P/N: 2200416). Refer to Section 3.3.13 for details.
The system is fully functional with these bypass tools in place. Use extreme
caution when working with the instrument in this state.
Use extreme caution when removing the sample probe as it is very sharp and a
potential biohazard.
7. Select Exit to Service software in the MAINTAIN screen. Enter the appropriate
password.
8. Check the current firmware and software versions in the VERSIONS (1/1) screen.
Update the software and/or firmware if needed (see 7.20.2).
1. Select SAMPLE ARM from the main (2/2) screen. Select the DILUT./WASH
STAT. REPLA. button to move the arm for easy access to the dilution/wash
station.
2. Press the lever on the left side of the dilution/wash station. Slowly lift it and
disconnect the 2 LUER fittings. Remove the dilution/wash station completely.
3. Clean the dilution/wash station with bleach and rinse thoroughly with deionized
water. Verify that the drain holes are cleared of debris.
4. Check the tubing (S4 and S5) from the dilution/wash station to the pinch valves
on the syringe module. Open and/or remove the side doors if needed. Replace
the tubing if damaged (P/N: 2200317).
5. Install the dilution/wash station. Connect the color-coded LUER fittings before
pushing the dilution/wash station completely into place. A click from the lever will
be heard when it is secured properly. Try to lift the dilution/wash station again, in
order to make sure that it is secured properly.
6. Check evacuation of the drain lines. Select SYRINGE from the main (1/2)
screen.
7. Select START DILUTION VALVE and START WASH VALVE from the SYRINGE
(1/6) screen.
8. Using a squirt bottle filled with deionized water, fill the dilution well and wash well
with about 1mL of water each.
9. Select START LIQUID PUMP.
10. Open the dilution valve by selecting STOP DILUTION VALVE and verify that the
water is evacuated properly. Close the dilution valve again by selecting START
DILUTION VALVE.
11. Open the wash valve by selecting STOP WASH VALVE and verify that the
water is evacuated properly.
12. Open the dilution valve by selecting STOP DILUTION VALVE.
13. Select STOP LIQUID PUMP.
The tube holder is biohazardous. Use disposable gloves and extreme caution
when handling.
1. Select SAMPLE ARM from the main (2/2) screen. Select the X MOVE button.
2. On the X MOVE (1/4) screen (see Figure 9-1):
select To tube
set the tube number to 3
Figure 9-1:
D-10 Sample arm module: X (1/4)
Figure 9-2:
D-10 Sample arm module: TH (2/4)
Figure 9-3:
D-10 Sample arm module
1. Select SAMPLE ARM from the main (2/2) screen. Select the Z MOVE button.
2. On the Z MOVE (1/4) screen (see Figure 9-4):
select To sample aspiration position (tube)
set Test selection to A1c
set Tube position to 2
Figure 9-4:
D-10 Sample arm module: Z (1/4)
3. Select START to move the arm to its final position for lubrication.
4. Place an absorbent towel under the bottom of the lead screws to collect excess
oil.
5. Place a drop of lubricant oil in the upper part of the nut for both the Z and TH
movements.
6. Select Home movement and START to begin the movement (see Figure 9-4).
7. Go to the Z MOVE (2/4) screen (see Figure 9-5):
select Direction to from up to down
set Movement select to 230 strips
Figure 9-5:
D-10 Sample arm module: Z (2/4)
Sections 9.2.6 through 9.2.13 require the system to be shut down, turned off and
the external power cord removed.
1. Shut down the D-10, turn off the power and unplug the external power cord.
2. Make a dilution of the flow cell cleaning solution using one part cleaning solution
to two parts deionized water. Total volume should be at least 10mL.
3. Remove the tubing at the detector in the following order to avoid twisting the
backpressure tube:
disconnect the detector outlet tubing (orange PEEK) from the union
remove the union from the backpressure tube and connect it to the
detector outlet PEEK tube
connect the Teflon tube from the flow cell cleaning kit (P/N: 1957263) to
the other end of the union using the appropriate nut and ferrule
place the open end of the Teflon tube into the cleaning solution
4. Open the cartridge module and remove the cartridge holder with the analytical
cartridge. Remove the cartridge and cap the ends.
5. Install the dummy PEEK cartridge holder (P/N: 2200435). Place an absorbent
towel under the outlet to catch any liquid.
6. Remove the rubber plug from the cartridge module door and close the door.
7. Connect the 10mL syringe of the flow cell cleaning kit to the LUER connector at
the outlet of the dummy PEEK cartridge holder.
8. Gently pull back on the syringe plunger until 1mL of cleaning solution has been
aspirated. Soak for 10 minutes.
9. Repeat step 8.
10. Remove the Teflon tube from the cleaning solution and place it into a container
of deionized water.
11. Gently pull back on the syringe plunger until 2mL of deionized water has been
aspirated. Disconnect the syringe and empty it.
12. Reconnect the syringe to the dummy PEEK cartridge holder and pull an
additional 6mL, 2mL at a time. Remove the syringe.
13. Remove the union at the outlet of the detector.
14. Remove the union from the Teflon tube, connect it to the backpressure tube and
to the outlet of the detector.
15. Remove the dummy PEEK cartridge holder.
16. Wash the internal surface of the cartridge module, near the outlet to the flow cell,
with deionized water. Replace the analytical cartridge.
17. Replace the rubber plug in the cartridge module door and close the door.
18. After the system has been powered up, flush the fluid line with the HP pump
(see 9.3.1).
1. Verify that the system is switched off and the external power cord is removed.
2. Inspect all internal tubing for signs of leakage. Tighten or replace fittings and
tubing as needed.
3. Clean the bottom of the HPLC module. A small leak behind the pump head, at
the piston flush circuit, is not unusual.
4. Clean all other internal and external surfaces.
1. Verify that the system is switched off and the external power cord is removed.
2. Place an absorbent towel under the cartridge module.
3. Open the cartridge module. Check the movement of the cartridge module door
as it is opened.
4. Remove the cartridge holder.
5. Clean the internal surface of the cartridge module.
6. Re-install the cartridge holder, checking the installation.
7. Close the cartridge module and dispose of the towel.
The internal waste bottle is biohazardous. Use disposable gloves and extreme
caution when handling.
1. Verify that the system is switched off and the external power cord is removed.
2. Inspect all internal tubing for signs of leakage. Tighten or replace fittings and
tubing as needed.
9.2.7.4 Wiring
1. Verify that the system is switched off and the external power cord is removed.
2. Check the connection of the flex cables to the Z board (see Figure 9-6).
Figure 9-6:
Flex Cables on Z Board
Figure 9-7:
Flex Cables on X Board
4. Check the tube height sensor flex cable and its connection to the Z board.
1. Verify that the system is switched off and the external power cord is removed.
2. Clean the barcode reader window using a soft cloth dampened with deionized
water.
1. Verify that the system is switched off and the external power cord is removed.
2. Check the external waste tubing. Replace if needed (P/N: 2200321).
1. Verify that the system is switched off and the external power cord is removed.
2. Remove the external grid and clean it.
3. Clean the dust off the blades of the fan.
4. Check the free rotation of the motor.
5. Replace the external grid.
1. Verify that the system is switched off and the external power cord is removed.
2. Clean the external surface of the rack track.
3. Clean the dust off the external surface of the rack reflective sensors.
4. Check the position of the rubber disk mounted on the rack positioning motor
assembly.
If the internal waste bottle is equipped with a filter, the liquid pump cleaning is not
necessary.
The liquid pump is biohazardous. Use disposable gloves and extreme caution
when handling.
1. Verify that the system is switched off and the external power cord is removed.
2. Loosen – but don't remove – the 4 screws which secure the pump head (see
Figure 9-8).
Figure 9-8:
Liquid Pump
3. The pump head consists of two green plastic parts; remove only the upper part
(see Figure 9-9).
Figure 9-9:
Liquid Pump Components
4. Clean the upper part, making sure there are no rubber particles trapped in the
nozzles.
5. Remove the two black rubber parts (see Figure 9-9) from the pump, being
careful not to tear them. Clean the pump head. DO NOT remove the green
plastic part from the pump body.
Liquid pump replacement may be performed in case the liquid pump is damaged.
(Section 6.77)
6. Re-install the two black rubber parts and re-assemble the pump. When re-
installing the two green plastic parts, be careful to respect the alignment of the
two lines etched on one side of the pump head (see Figure 9-10).
Figure 9-10:
Correct Pump Head Orientation
1. Verify that the system is switched off and the external power cord is removed.
2. Disconnect the ends of each pinch valve tube from the LUER fitting and the
stainless steel tube on the waste manifold 1 (see Figure 9-11).
Figure 9-11:
Pinch Valve Tubing Connections
1. Verify that the latest firmware is on the MC controller, the HPPC controller and
the TC controller (see Figure 3-42).
The sample probe is very sharp and a potential biohazard. Use extreme caution
when handling.
1. Dispose of the old sample probe according to local requirements for the disposal
of sharps.
2. Install the new probe and connect the sample tubing.
Comments
D-10 Hemoglobin Testing System
10 SPARE PARTS
Bio-Rad
Part Description Picture Comments
P/N
LCD display
(with touch screen)
2200221
Reagent tray
2200222
Hard disk
Floppy drive
Thermal printer
Bio-Rad
Part Description Picture Comments
P/N
complete assembly
2200226
(with hinges)
complete assembly
2200227
(with hinges)
Gas spring
Arm
(w/ barcode reader)
2200231
Bio-Rad
Part Description Picture Comments
P/N
X gear belt
2200232
Z gear belt
2200233
TH gear belt
2200234
2200235
Z home sensor
2200236
Bio-Rad
Part Description Picture Comments
P/N
TH position sensor
2200237
TH home sensor
2200239
X position sensor
2200240
X home sensor
2200241
Bio-Rad
Part Description Picture Comments
P/N
Z board
No. DAE001
Rev 3
2200242
compatible with all previous revisions
(U.S.) BaaN Description:
D10 PCB Z MOVEMENT DAE001 R3
X board
No. DAE002
Rev 3
2200243
compatible with all previous revisions
(U.S.) BaaN Description:
D10 PCB X MOVEMENT DAE002 R2
2200244
2200245
Barcode reader
Bio-Rad
Part Description Picture Comments
P/N
Sample probe
Syringe module
2200248
Pinch valve
2200249
2200250
Liquid pump
2200251
Bio-Rad
Part Description Picture Comments
P/N
2200253
2200255
HPLC module
Bio-Rad
Part Description Picture Comments
P/N
2200257
Proportioning valve
2200258
Pressure sensor
2200259
Detector module
2200261
Bio-Rad
Part Description Picture Comments
P/N
LED board
2200262
Reference board
2200263
Sample board
2200264
Detector board
No. DAE011
Rev 4
2200265
compatible with all previous revisions
(U.S.) BaaN Description:
D10 PCB: DETECTOR DAE011 R1
Flow cell
Bio-Rad
Part Description Picture Comments
P/N
Cartridge module
Temperature sensor
2200268
Peltier module
2200269
Thermal fuse
2200270
Cartridge holder
2200271
Bio-Rad
Part Description Picture Comments
P/N
2200272
not compatible
2200273
with Variant II injection loop
Electronic module
Power supply 1
delivered voltages:
+5VDC
2200275 +12VDC
+26VDC
(U.S.) BaaN Description: (-12VDC)
D10 POWER SUPPLY, 5 VOLT
Power supply 2
delivered voltage:
2200276
+24VDC
(U.S.) BaaN Description:
D10 POWER SUPPLY, 24 VOLT
Bio-Rad
Part Description Picture Comments
P/N
2200277
Motherboard (PCM-5896)
2200278
2200279
2200280
Interface board
No. DAE010
Rev 6
2200281
compatible with all previous revisions
(U.S.) BaaN Description:
D10 PCB: INTERFACE DAE010 R3
Bio-Rad
Part Description Picture Comments
P/N
2200283
Rubber disk
2200285
2200286
Bio-Rad
Part Description Picture Comments
P/N
2200287
Dilution/wash station
2200288
Degasser module
2200289
Bubble detector 1
Bio-Rad
Part Description Picture Comments
P/N
Bubble detector 2
Bubble detector 3
Vacuum pump
2200294
2200295
2200296
Bio-Rad
Part Description Picture Comments
P/N
Microvial adapters
2200297
Front cover
Left door
Right door
2200301
Bio-Rad
Part Description Picture Comments
P/N
Sysmex rack
2200302
Sysmex adapters 12
2200303
Sysmex adapters 13
2200304
Sysmex adapters 14
2200305
Main fan
2200306
Bio-Rad
Part Description Picture Comments
P/N
2200307
2200308
2200309
Z position sensor
2200310
2200311
Bio-Rad
Part Description Picture Comments
P/N
External tubing
contents:
T1: 1/8" x 1/16" x 700mm + T2: 1/8" x
1/16" x 700mm + T5: 1/8" x 1/16" x
2200314
700mm + 3 bottle caps + 3x T15: 1/8" x
1/16" x 207mm + S13: 6mm x 3mm x
(U.S.) BaaN Description: 2500mm
D10 TUBING,BUFFER:EXTERNAL,SET
contents:
T3: 1/8" x 1/16" x 570mm + T4: 1/8" x
Teflon internal tubing 1/16" x 650mm + T6: 1/8" x 1/16" x
530mm
+ T7: 1/16" x 0.75mm x 115mm + T9:
2200315 1/16" x 0.75mm x 1250mm + T10: 1/16"
x 0.75mm x 210mm + T11: 1/16" x
0.75mm x 530mm + T12: 1/16" x
(U.S.) BaaN Description:
0.75mm x 150mm + T13: 1/16" x
D10 TUBING TEFLON INTERNAL,SET
0.75mm x 110mm + T14: 1/16" x
0.25mm x 1200mm
PEEK tubing
contents:
2200316 P1: 1/16" x 0.5mm x 170mm + P2: 1/16"
x 0.5mm x 120mm
Bio-Rad
Part Description Picture Comments
P/N
contents:
S4: 6mm x 3mm x 550mm + S5: 6mm x
Silicon internal tubing
3mm x 550mm + S8: 6mm x 3mm x
140mm + S9: 6mm x 3mm x 270mm +
S10: 6mm x 3mm x 160mm + S12: 6mm
2200317
x 3mm x 330mm + S14: 6mm x 3mm x
30mm + S15: 6mm x 3mm x 120mm +
(U.S.) BaaN Description: S16: 6mm x 3mm x 35mm + S17: 6mm x
D10 TUBING, SILICON, SET 3mm x 60mm + S18: 6mm x 3mm x
390mm
Service manual
2200318
Waste bottle
Bio-Rad
Part Description Picture Comments
P/N
2200323
2200324
Waste manifold 1
(large)
2200325
Waste manifold 2
(small)
2200326
with seals
2200327
for the old internal waste bottle cap only
(U.S.) BaaN Description:
D10 FITTING,LUER:INT WASTE SET
Bio-Rad
Part Description Picture Comments
P/N
Backpressure tubing
2200331
Bio-Rad
Part Description Picture Comments
P/N
2200333
2200335
2200336
2200337
2200339
Bio-Rad
Part Description Picture Comments
P/N
2200340
2200341
2200342
2200343
2200344
Bio-Rad
Part Description Picture Comments
P/N
2200345
2200346
2200347
2200348
2200349
Bio-Rad
Part Description Picture Comments
P/N
2200350
2200351
2200352
2200353
2200354
Bio-Rad
Part Description Picture Comments
P/N
2200355
2200356
2200357
2200358
2200359
Bio-Rad
Part Description Picture Comments
P/N
2200360
2200361
2200362
2200363
2200364
Bio-Rad
Part Description Picture Comments
P/N
2200365
2200366
Flange nut
2200367
Lead screw
2200369
Bio-Rad
Part Description Picture Comments
P/N
Slide bearing
Belt pulley
(16-tooth)
driving pulley
of the X movement
2200372
with mounting screw
(U.S.) BaaN Description:
D10 BELT PULLEY (16 TEETH)
Belt pulley
(15-tooth)
driving pulley
of the Z and TH movements
2200373
with mounting screw
(U.S.) BaaN Description:
D10 BELT PULLEY (15 TEETH)
Belt pulley
(30-tooth)
led pulley
of the Z and TH movements
2200374
with mounting screw
(U.S.) BaaN Description:
D10 BELT PULLEY (30 TEETH)
Bio-Rad
Part Description Picture Comments
P/N
Printer paper
CCFL backlight
2200376
2200377
2200378
2200379
Bio-Rad
Part Description Picture Comments
P/N
2200380 2x A5 sheets
2200381
2200382
2200383
2200384
Bio-Rad
Part Description Picture Comments
P/N
2200385
2200386
2200387
2200388
2200389
Bio-Rad
Part Description Picture Comments
P/N
2200390
2200391
2200392
2200393
2200394
Bio-Rad
Part Description Picture Comments
P/N
2200395
2200396
2200398
Accessories box
2200399
Bio-Rad
Part Description Picture Comments
P/N
Accessories kit
2200400
50mL syringe
2200401
2200402
Fuses
2200403
2200404
Bio-Rad
Part Description Picture Comments
P/N
2200405
Lubricant oil
Cyanoacrylate glue
Anaerobic glue
2200409
Bio-Rad
Part Description Picture Comments
P/N
Torque driver
Balance
2200411
Mini-screwdriver
Bio-Rad
Part Description Picture Comments
P/N
2200417
2200418
Bio-Rad
Part Description Picture Comments
P/N
2200420
Micro-switch
Dummy probe
2200424
Bio-Rad
Part Description Picture Comments
P/N
2200427
Bio-Rad
Part Description Picture Comments
P/N
2200430
2200431
2200432
Bio-Rad
Part Description Picture Comments
P/N
M3 standoffs
2200436
¼" caps
Cartridge module
temperature control tool
Bio-Rad
Part Description Picture Comments
P/N
Cartridge module
temperature calibration kit
Hanna digital thermometer + adapter
cable + temperature sensor inserted in
2200441
an analytical cartridge + 2 reference
resistors
(U.S.) BaaN Description:
D10 TOOL: TEMP CALIB KIT
2.22 image CD
2200443
Fluidic diagram
2200444
Bio-Rad
Part Description Picture Comments
P/N
Dye kit
125mL plastic bottle, with a mark at
100mL, and containing 0.22g of 'Acid
2200446
Yellow 42' powder + 10mL syringe +
technical note
(U.S.) BaaN Description:
D10 DYE KIT
2200449
Bio-Rad
Part Description Picture Comments
P/N
Tamperproofing coating
9600414
Table of Contents:
Dia.ini A-1
instrument.ini A-3
Profile.ini A-4
Profile2.ini A-6
DiaSer.ini A-9
DiaLIS.ini A-13
*.csv (partial) from Service software A-14
'Dia.ini' File:
[Status]
Version=2.29
[State]
CurrentInjNumber=130
DateTag=202
WPistonFlushTag=1
CurrentWasteVol=60400
[Language]
TotalLanguages=3
DirectoryName0=English
DisplayName0=English
DirectoryName1=Japanese
DisplayName1=Japanese
DirectoryName2=French
DisplayName2=Francais
[Settings]
Language=0
AutoWarmUpTime=27000
AutoShutDownDuration=1800
AutoShutDownDurationMin=1800
AutoShutDownDurationMax=5400
BarCodePresent=1
ArchiveToFloppy=0
ArchiveToSerial=1
ArchiveToNetwork=1
ReportCalError=1
ReportCtrlError=1
PressureUnit=1
WarmUpTimeOut=900
MaxNSRRuns=18
DisableAutoWarmUp=1
ReagentWarningLevel=5
CartWarningLevel=5
WasteWarningLevel=5
TotalWasteVol=100000
[Lis]
LisID=1
ExportActivated=0
LisAfterEachRun=0
LisAfterEachInjection=0
LisManualOnly=1
[Print]
AutoPrintOut=1
DailyLog=1
[Data]
ExportLis=0
ExportFloppy=0
ExportSelectedSample=1
ExportBetweenDate=0
DataOrientation=1
[Assay]
CurrentAssay=0
CurrentReagentLot=AB06256
CurrentControlLot=33560
CurrentCalibC1=0.053
CurrentCalibC2=0.119
CurrentControlLL=0.051
CurrentControlLM=0.056
CurrentControlLU=0.061
CurrentControlHL=0.093
CurrentControlHM=0.097
CurrentControlHU=0.101
MaintainBPercent=50
[Maintain]
MaintainFlowRate=2.2
MaintainBPercent=50
ServiceSoftware=DiaSer.exe
[HPPump]
EPISTONVOL=380.000000
EENCRES=1024
EMOTORREDUCTION=20
EVRESPTIME=10
EVALVEHIGHTIME=20
EVRESPTIMEP=50
EVRESETOFFSET=0
[Barcode]
Calib1Header=CAL1
Calib2Header=CAL2
ControlHighHeader=CTRH
ControlLowHeader=CTRL
PrimerHeader=PRIME
'instrument.ini' File:
[Status]
Version=2.20
[Instrument]
SerialNum=DA4F357306
[HPPump]
Rule1=0.0,1.0,2414.47,0.00
Rule2=1.0,1.5,2438.47,-23.86
Rule3=1.5,99.0,2344.28,116.13
HPSensor_Offset=20.5556
HPSense_Gain=12.6699
ECAMFILLSHIFT1=155
EFILLDURATION=114
[Arm]
wTHOffset=5
wZOffset=0
[BubbleDetect]
Bubble_A_1=0
Bubble_A_2=255
Bubble_B_1=0
Bubble_B_2=255
Bubble_W_1=0
Bubble_W_2=255
[TC]
TCCal_Gain=1.0046
TCCal_Offset=0.3028
TCCal_SensorOffset=-0.1000
TCCal_GradientOffset=-0.7000
[XXADILXX]
XXADIL1XX=243
'Profile.ini' File:
[Status]
Version=2.29
[General]
StandbyFlow=0.2
[Kit]
Kit0=0
Kit1=0
Kit2=0
Kit3=0
Kit4=0
[Init]
XSpeed=9
ZSpeed=2
THSpeed=2
SSpeed=2
VSpeed=0
InitType=1
NeedlePrime=1
wExhaustDelay=10
[LowPrime]
FillSyringeSpeed=7
EmptySyringeSpeed=4
XSpeed=6
THSpeed=5
ZSpeed=5
ExhaustChamberDelay=50
StrokeCount=3
[HighPrime]
FillSyringeSpeed=5
EmptySyringeSpeed=4
XSpeed=1
THSpeed=5
ZSpeed=5
WashStrokeCount=8
StrokeCount=38
[HpFlush]
FillSyringeSpeed=7
EmptySyringeSpeed=4
XSpeed=1
THSpeed=5
ZSpeed=5
ExhaustChamberDelay=50
StrokeCount=3
wPumpFlow=300
[SystemTable]
bFakeRead=0
bMXOverrun=3
wMXWarnSteps=2500
wMXWarnNbr=0
wMZWarnSteps=13001
wMZWarnNbr=0
wMTHWarnSteps=11002
wMTHWarnNbr=0
wMSWarnSteps=5200
wMSWarnNbr=0
wMVWarnSteps=2400
wMVWarnNbr=0
DegasLowPress=118
DegasHighPress=125
DegasTimeOut=120
bRackDoorMonitor=1
bFrontDoorMonitor=1
bRackDoorPresent=1
bWasteGoodVacuum=76
bWasteCheckTimeOut=30
bWasteReleaseOffset=5
bWasteRelease1TO=5
bWasteRelease2TO=5
[Misc]
SleepLightSourceLevel=0
StandbyLightsourceLevel=150
WarmUpPressureMonitoringPercent=60
[PumpParameters]
wHPAnalogPeriod=250
PumpCurrentMax=181
wPumpCurrentDefaultMax=64
wPumpSpeedMax=1000
wPumpSpeedDefaultMax=2
fProfileSpeedSteps_fudge=400.0
fPSI_fudge=2500.0
[Print]
AutoPrintOut=1
'Profile2.ini' File:
[Status]
Version=2.29
[Kit]
Name=Hb A1c
SName=Hb A1c
StName=A1c
TestFile=a1c.hpl
PrimeFile=primer.hpl
WarmUpGradient=1
WarmUpGradientFile=warmupa1c.hpl
BeginRun=1
BeginRunFile=begina1c.hpl
EndRun=1
EndRunFile=enda1c.hpl
RealTotalBuffer1Vol=20000
RealTotalBuffer2Vol=10000
RealTotalWashVol=16000
TotalBuffer1Vol=40000
TotalBuffer2Vol=10000
TotalWashVol=16000
TotalCartridgeInj=400
Buffer1InjConsuption=100
Buffer2InjConsuption=25
WashInjConsuption=40
WasteInjVolume=165
DualKit=1
CurrentBuffer1Vol=18847
CurrentBuffer2Vol=9722
CurrentWashVol=15560
CurrentCartridgeInj=389
[ReferenceValues]
Enabled=1
DisplayedDecimal=0
Up=8
Median=7
Low=6
[Inject]
wSampleVolume=20
wDiluantVolume=480
wSecondDilutionSampleVolume=20
wSecondDilutionDiluantVolume=680
wDilutionValveDelay=10
wAirBubbleVolume=60
wAntiDropBubbleVolume=0
wTHSenseMicrovialHigh=6675
wTHSenseMicrovialLow=7675
wTHSenseMin=0
wZsamplingPosition=230
wZMicrovialSampling=180
wZantiDropOffset=105
wMicroVialAntiDropVolume=25
wFromDilutionToValve=745
wChamberCleaningVolume=1000
bFillSyringeSpeed=9
bEmptySyringeSpeed=9
bAirGapSyringeSpeed=1
bAspSampleSyringeSpeed=0
bDspSampleSyringeSpeed=0
bLoopFillSyringeSpeed=0
bXSpeed=2
bTHSpeed=6
bTHTubeHoldSpeed=2
bZSpeed=6
bZPiercingSpeed=1
bSyringeValveDelay=10
bExhaustChamberDelay=50
bBCReadFlag=1
wMicroVialSampleVolume=250
wMicroFromDilutionToValve=460
wMicroVialAirBubbleVolume=10
[Chroma]
wSampleRate=36
wSampleDuration=180
wInjectionDelay=20
bRefAdjust=1
bLightSourceIntensity=150
bIOGainValue=64
dAbsMax=16106127
dAbsMin=8388608
dAbsRef=8808038
bLightSourceIntensityMax=240
wBaseLinePercent=470
wInjectionInterval=180
dMaxRefDrift=13421772
bVSpeed=4
bLightIncrement=10
bLightStartTempo=60
bLightChangeTempo=15
[Prime]
bVSpeed2=4
bVSpeed=4
wLoadDuration1=120
wInjectDuration1=12
wLoadDuration2=150
wInjectDuration2=18
wLoadDuration3=60
wInjectDuration3=60
wLoadDuration4=60
wInjectDuration4=210
wLoadDuration5=30
[PumpParameters]
wMaintainHighPRessureAlarm=2000
wRunHighPressureAlarm=1067
RunHighAverageMeasNumber=12
wRunLowPressureAlarm=130
RunLowAverageMeasNumber=60
'DiaSer.ini' File:
[Status]
Version=1.29-1A
[Maintain]
UserSoftware=Dia.exe
LoadProfile=0
CheckBeforeExit=1
AL1=81d80815851d73ee717f27e5a2ce6851
AL2=97b6efa84a8ecd3851b76273e7a6e6b8
AL3=e8f95b17eea00d1f2bf5febb3e77877e
AL4=103d20ecc62edfd70f1b198485eb2122
NonStopRunSoftware=dia_nsr.exe
TouchScreenCalib=C:\drivers-tools\V10L10\Calwin.exe
[SerialPort]
Port=4
Speed=19200
[State]
CurrentInjNumber=2
MaxInjNumber=1000
DateTag=202
[Test]
Number=3
Test1=HBA1C
Test2=BTHAL
Test3=DECONT
[Sequence]
Injection=InjectionSequence.csv
DiluChambWashing=DiluChambWashingSequence.csv
ProbeWashing=ProbeWashingSequence.csv
PickUpDilutionLoadTube=PickUpDilutionAndLoadOfASampleFromATube.csv
PickUpDilutionTube=PickUpAndDilutionOfASampleFromATube.csv
LoadDiluChamb=LoadOfASampleFromTheDilutionChamber.csv
LoadMicrovial=LoadOfASampleFromAMicrovial.csv
InjectionOfASample=InjectionOfASample.csv
DiluChambProbeWashing=DiluChambAndProbeWashingSeq.csv
[Detector]
bLightSourceIntensity=150
bIOGainValue=64
LogonTime=200
[RackDoor]
AgingPause=6
[Motor]
XStepCourseMax=2200
XStripCourseMax=44
XAllMoveSpeedDefault=5
ZStepCourseMax=12000
ZStripCourseMax=240
ZAllMoveSpeedDefault=5
THStepCourseMax=8000
THStripCourseMax=160
THAllMoveSpeedDefault=5
ZStepDilutionChamber=3500
ZStripDilutionChamber=70
ZStepWashChamber=10200
ZStripWashChamber=204
THOffsetTheo=7775
[HPPump]
FlowRateMax=30
FlowRateDefault=15
FlowRateCalibDefault=1,1.5,2,99
FlowRateTolerance=5
DensityDefault=1.001
WCMin=140
WCMax=165
FCMin=100
FCMax=130
[PropValve]
AgingDurationMax=60
[ThermalMod]
TempMin=30
TempMax=50
[WCFCDeterTest]
WCMin=144
WCMax=161
WCMinDef=153
WCMaxDef=157
FCMin=106
FCMax=120
FCMinDef=111
FCMaxDef=113
FlushFlowRate=30
FlushTimeDila=240
FlushTimeCont=480
[StaticGradTest]
GradDuration1=300
GradDuration2=300
GradDuration3=480
NbWin=2
START1=150
START2=510
END1=300
END2=600
[DilamexGradTest]
NbWin=3
START1=45
START2=240
START3=510
END1=70
END2=360
END3=600
RAU100=0.005
MIN50=49.3
MAX50=50.7
MINCOR50=48
MAXCOR50=52
[Dilamex+GradTest]
NbWin=4
START1=60
START2=295
START3=475
START4=655
END1=120
END2=310
END3=490
END4=670
RAU100=0.008
AVE50=50
MIN20=13
MIN50=49.5
MAX20=14
MAX50=50.5
MINCOR50=48
MAXCOR50=52
[StepGradTest]
NbWin=11
START1=60
START2=265
START3=445
START4=625
START5=805
START6=985
START7=1165
START8=1345
START9=1525
START10=1705
START11=1885
END1=120
END2=310
END3=490
END4=670
END5=850
END6=1030
END7=1210
END8=1390
END9=1570
END10=1750
END11=1930
RAU100=0.005
MIN10=3
MIN20=12.5
MIN30=21
MIN40=32
MIN50=49.3
MIN60=60
MIN70=70
MIN80=80
MIN90=90
MAX10=10
MAX20=14
MAX30=30
MAX40=40
MAX50=50.7
MAX60=68
MAX70=79
MAX80=89
MAX90=97
MINCOR10=8
MINCOR20=18
MINCOR30=28
MINCOR40=38
MINCOR50=48
MINCOR60=58
MINCOR70=68
MINCOR80=78
MINCOR90=88
MAXCOR10=12
MAXCOR20=22
MAXCOR30=32
MAXCOR40=42
MAXCOR50=52
MAXCOR60=62
MAXCOR70=72
MAXCOR80=82
MAXCOR90=92
'DiaLIS.ini' File:
[Com]
Port=3
BaudRate=9600
DataBits=8
StopBits=1
Parity=0
[LisBehaviour]
Version=1.20
ActUponQuery=1
QRespMaxNbRes=1
PushedResFN=PushRes.tsv
[LisMap]
HBA1C=4
BTHAL=1
InstName=D10
Version=3.00
Appendix B: Diagrams
Table of Contents:
Tubings: T7
dilution chamber
piston
from HP pump to the cartridge
and sensor module G Y
2 pinch valves
check valve
T14 P4 waste manifold 1
from detector
B B outlet
(backpressure tubing)
S13 S12 S9
vacuum
S8 sensor
R R
CPC fast
connector S15 S14 S10
Tubings: T7
dilution chamber
piston
from HP pump to the cartridge
and sensor module G Y
2 pinch valves
check valve
T14 P4 waste manifold 1
from detector
B B outlet
(backpressure tubing)
S13 S12 S9
vacuum
S8 sensor
R R
CPC fast
connector S15 S14 S10
B B
to the 3-way valve pressure sensor
B T5 T6 encoder
DT1 3 bubble
detectors
M
T1 G DT2 G T3 G
NO T12 P1 P2
IN OUT
T15
T2 W W W
piston flush
G W T4 NC
circuit
IN OUT
DT3 proportioning valve
cam sensor
wash/diluent
S1 S2
S17
T15 T15
vacuum
pump S3 T9
buffer 1 buffer 2
vacuum
Y R
7-port,
sensor 3-position
6 5
T11 HP valve
Tubings: (load position)
to the 1mL syringe
B B
to the 3-way valve pressure sensor
B T5 T6 encoder
DT1 3 bubble
detectors
M
T1 G DT2 G T3 G
NO T12 P1 P2
IN OUT
T15
T2 W W W
piston flush
G W T4 NC
circuit
IN OUT
DT3 proportioning valve
cam sensor
wash/diluent
S1 S2
S17
T15 T15
vacuum
pump S3 T9
buffer 1 buffer 2
vacuum
Y R
7-port,
sensor 3-position
6 5
T11 HP valve
Tubings: (inject position)
to the 1mL syringe
T5 T6
B NC B NO Y R 7-port,
check valve
S1 S2 3-position
3-way T11 6 5
HP valve G
valve (load position)
buffer 1 buffer 2
B
1 7 4
vacuum T13 B
pump
vacuum
sensor sample rack
2 3
Y
T15 S18 G
S3 1mL syringe S5
S4
G Y
SS2
wash/diluent SS1
sample loop
NO NO
B B dilution probe wash
P4 P3 valve valve
S6 S7
S13 S12 heating
T14 analytical
R flow
CPC fast (backpressure cell waste
connector tubing) manifold 1
cartridge loop vacuum
sensor
waste manifold 2 Hb detector cartridge module
S8
R S9
G Green 6 5 6 5 6 5
plastic barbed fitting
LUER fitting
1 7 4 1 7 4 1 7 4
W White or plastic knurled fitting stainless steel tube
D-10:
ELECTRONIC MODULE Electrical Block Diagram
sheet 1/2
LCD 320 x 240
with CCFL backlight RS232 rack
keyboard mouse
external LAN port VGA port feeder
touch screen port port
port port
ON/OFF unipolar see sheet 2/2 see sheet 2/2 see sheet 2/2 see sheet 2/2 see sheet 2/2
switch electronic board
(front cover)
sub assembly
(X movement)
module X stepper motor
sample arm
X board CN4
pinch valve (dilution) board
module
CN7
solenoid A16 CN11 X home sensor
pinch valve (probe wash) CN8 drivers (5) w/ interface board
syringe module
current CN10 X axis sensor
pinch valve 3 CN9 sensing
stepper motor
vacuum sensor driver w/
syringe stepper motor CN3
interface current sensing
CN3 rack positioning motor
rack drive
CN4
A13
module
syringe home sensor CN2 optosensor
rack input sensor
CN1
CN8
stepper motor interface (6)
syringe 3-way valve CN5 driver
(microsteps) w/ Hall effect CN9 rack output sensor
vacuum current sensing sensor interface
internal waste bottle sensor
optosensor solenoid and CN2 rack door solenoid
rack door
DC motor
module
interface (1) CN5
drivers w/
vacuum sensor vacuum sensor current sensing CN6
CN7 closed door sensor
degasser module
A11
CN4
+ ROM stepper motor
bubble detector 1 (buffer 1) CN1 driver + I/O CN4 CN1 Hall effect adapter sensor
driver w/
A15
bubble detector 2 (buffer 2) CN3 degasser current sensing
CN10
module
optosensor CN11 Z stepper motor
interface (3) optosensor
bubble detector 3 (wash/dil.) CN6 board
(Z and TH movements)
interface (2) CN2 Z home sensor
proportioning valve CN15 DC motor Z board CN6 barcode reader (RS232 link)
driver w/
HP valve (V) stepper motor CN14 RS232 port
current sensing
TH stepper motor
CN12
CN7
Peltier module 1 CN13 solenoid driver stepper motor
(2 levels) driver w/
w/ current sensing current sensing CN12 TH home sensor
Peltier module 2 CN11
HPLC module
stepper motor
TH axis sensor
A10
thermal fuse CN10 optosensor CN5
driver w/ current
interface (3)
sensing
cam flag sensor CN9 CN8 tube height sensor
Peltier module
driver w/ current
HP pump encoder CN8
sensing
CN5
A14
I/O interface to
w/ current rack feeder unit to optional rack feeder board rear panel
pressure sensor CN6 sensing
A12
detector
CN4
module
HPLC pressure
CN1 photodetector: sample
front cover solenoid module sensor interface
cover
front
sensor TC HPPC
µ-controller for µ-controller
heater and + RAM
electronic board temperature + ROM
sensor interface + I/O
D-10:
motor or solenoid Electrical Block Diagram
sheet 2/2
sub assembly
Table of Contents:
L1 L
P
PT10
VDD IC9 PT13
M R46 IC8 D1
8 1 1 2
R48 IDC ISWC VIN VOUT 3
PT11 7 R44 VOUT 6
IPK VOUT 5VANA
ADJ
5 7
6 COMP VOUT
3 VCC
TCAP
4
+ R47 +
4
+ 2 GND R45 C21 C24 C23 C27
ISWE
C26 C25 C22
R43
DG ND DG ND DG ND DG ND DGND N
PT12
AGND
CN4
5VANA
1 DADY
2 5VANA
3 /RESET C18
K
4 SD1
5 AGND
R41 P9 8
6 SD0
R40 IC5B
7 5 +
8 SCLK Vref
7
9 CS0 V0
6
10 CS1 -
IC7 R38
11 CS2
12 CS3 4
C19
13 DSYNC
14 DSYNC
AGND AGND AGND
15 VDD
AGND
16
DGND
Title
DETECTOR BOARD
V0
V0
B +
R1
CN1 R8 PT1 C2
IC2A
1 3 +
2 1 AGND AGND
3 2
-
ABS
R5 R2
5VANA
R3 R11
C3
V0
V0 8 C
R14 IC2B
5 + PT2
7
6 +
-
D
4 C5 AIN0
Vref PT3 R4
V0
R13
C10 R7
A
AGND
PT0
V0
V0
E
CN2
R25 PT4
1 IC3A
2 3 +
3 1
4 2
-
REF
R20 R16
G
PT6
Jumper 5VANA
1-2 connected R17 R26
C11
1
R50
P7 2
V0
V0 8 IC .3 A2 IC .4 W2
R29 IC2B
3
5 + R18
7
6
-
H
4 C12 +
Vref PT7
V0
R28
AIN1
R30 R27 AGND R23 R19
V0
C13
-
C15 R24 V0
AGND
V0
Title
DETECTOR BOARD
F
IC4
PT5
7
PDWN VDD
1
11 DV 2 C14
AI N1 +
-
12 Ain0
Ain1
DGND
DGND
8
C9 DGND
13 3
AI N0 +
-
14 Ain2
Ain3
Xin
Y1 C8
9 4
Vref Vref+ Xout
10
16 Vref- 6
AGND AGND DSYNC DSYNC
15 21
V0 Aincom Dout SD0
17 20
5VANA AVdd Din SD1
22
SCLK SCLK
5
RST /RESET
19
CS CS1
DRDY
Ain3
BUF
R31
R21
18
24
23
P6 VDD
DG ND
R33
P5
VDD
P4 VDD R32
P3 DG ND
R34
P2 IC1
P1 VDD 7 R22
PDWN VDD
1
11 DV 2 C4
AI N0 +
-
12 Ain0
Ain1
DGND
DGND
8
DGND
C1 13 3
14 Ain2 Xin
AGND Ain3
9 4
Vref Vref+ Xout VDD
10
16 Vref- 6
AGND AGND DSYNC
15 21
V0 Aincom Dout
17 20
5VANA AVdd Din 22 R9
SCLK 5
RST 19
CS CS0
DRDY
Ain3
BUF
C7
AGND
18
24
23
R10
DADY
Title
DETECTOR BOARD
CN3
Vref
1
AGND 2
R
5VANA 3
4
PT14
5
LED
R35
IC5A
A1
3
3 + R37 R49
W1
1 2 Q1
B1
2
-
J
R39
1
PT8
R36
C16 C17
AGND
AGND AGND
VDD
C20
DGND
16
IC6
23
VCC
A1 A1
22
W1 W1
24
B1 B1
12 21
SD1 SDI ADGN1
13
SD0 SDO
14 3
SCLK CLK A2 IC .3 A2
4
W2 IC .4 W2
2
B2 1
ADGN2
11 19
CS2 CS A3
15 18
/RESET RS W3 20
B3 17
ADGN3
7
R42 A4 8
10 W4 6
VDD SHDN B4 5
ADGN4
GND
VDD
AD8403AR-10 AGND
9
DGND
Title
DETECTOR BOARD
C1 1 CONDENSATEUR 0805 100nF 50V 100nF 50V 100nF 50V 100nF 50V
TANT TANT TANT TANT
C2 1 CONDENSATEUR
10µF 10V 10µF 10V 10µF 10V 10µF 10V
C3 1 CONDENSATEUR 0805 100nF 50V 100nF 50V 100nF 50V 100nF 50V
C4 1 CONDENSATEUR 0805 100nF 50V 100nF 50V 100nF 50V 100nF 50V
C5 1 CONDENSATEUR 0805 100nF 50V 100nF 50V 100nF 50V 100nF 50V
C6 1 CONDENSATEUR 0805 33nF 50V 33nF 50V 33nF 50V 33nF 50V
C7 1 CONDENSATEUR 0805 100nF 50V 100nF 50V 100nF 50V 100nF 50V
C8 1 CONDENSATEUR 0805 10pF 10pF 10pF 10pF
C9 1 CONDENSATEUR 0805 10pF 10pF 10pF 10pF
C10 1 CONDENSATEUR 0805 100nF 50V 100nF 50V 100nF 50V 100nF 50V
C11 1 CONDENSATEUR 0805 100nF 50V 100nF 50V 100nF 50V 100nF 50V
C12 1 CONDENSATEUR 0805 100nF 50V 100nF 50V 100nF 50V 100nF 50V
C13 1 CONDENSATEUR 0805 68nF 68nF 68nF 68nF
C14 1 CONDENSATEUR 0805 100nF 50V 100nF 50V 100nF 50V 100nF 50V
C15 1 CONDENSATEUR 0805 100nF 50V 100nF 50V 100nF 50V 100nF 50V
C16 1 CONDENSATEUR 0805 100nF 50V 100nF 50V 100nF 50V 100nF 50V
C17 1 CONDENSATEUR 0805 100pF 100pF 100pF 100pF
C18 1 CONDENSATEUR 0805 100nF 50V 100nF 50V 100nF 50V 100nF 50V
C19 1 CONDENSATEUR 0805 100nF 50V 100nF 50V 100nF 50V 100nF 50V
C20 1 CONDENSATEUR 0805 100nF 50V 100nF 50V 100nF 50V 100nF 50V
47µF 16V 47µF 16V 47µF 16V 47µF 16V
C21 1 CONDENSATEUR
BOITIER / D BOITIER / D BOITIER / D BOITIER / D
C22 1 CONDENSATEUR 0805 470pF 50V 470pF 50V 470pF 50V 470pF 50V
C23 1 CONDENSATEUR 0805 100nF 50V 100nF 50V 100nF 50V 100nF 50V
C24 1 CONDENSATEUR 0805 100nF 50V 100nF 50V 100nF 50V 100nF 50V
47µF 16V 47µF 16V 47µF 16V 47µF 16V
C25 1 CONDENSATEUR
BOITIER / D BOITIER / D BOITIER / D BOITIER / D
C26 1 CONDENSATEUR 0805 100nF 50V 100nF 50V 100nF 50V 100nF 50V
22µF 6V3 22µF 6V3 22µF 6V3 22µF 6V3
C27 1 CONDENSATEUR
BOITIER / C BOITIER / C BOITIER / C BOITIER / C
C28 1 CONDENSATEUR 1nF 50V 100nF 50V 100nF 50V 100nF 50V
IC1 1 CIRCUIT INTEGRE ADS 1240 E ADS 1240 E ADS 1240 E ADS 1240 E
IC2 1 CIRCUIT INTEGRE AD 8572 AR AD 8572 AR AD 8572 AR AD 8572 AR
IC3 1 CIRCUIT INTEGRE AD 8572 AR AD 8572 AR AD 8572 AR AD 8572 AR
IC4 1 CIRCUIT INTEGRE ADS 1240 E ADS 1240 E ADS 1240 E ADS 1240 E
V CC
C21 C17 C8 C6
100pF 100pF 100pF 100pF
CN11 Moteur Z
CN7 Moteur TH 2A
2A OUT1A _Z 2A 1
R10 R11 C26 C25
hall se nsor is moun t OUT1A _TH 2A 1 OUT2A _Z 2
2.2K 150 100pF 100pF
OUT2A _TH 2 3
3 2A 4
AD 4 SENSE1_Z 5
CN1 2A 2A
1.2mm SENSE1_TH 5 SENSE2_Z 6
adapter sensor 2A
1 (Hall e ffect). SENSE2_TH 6
1
C9 C13 JST B6B-EH-A
2 C22 C23 100pF 100pF
JST B6B-EH-A
3 100pF 100pF
4
A DAPTATOR_SENSOR 5
C24 JST B5B-PH-K- S
100pF
V CC
V CC
NC
NC R4 R1 C1 C2
R17 R16 C33 C34 if hall sensor is m ount 2.2K 150 100pF 100pF
CN4 if hall sensor is m ount 2.2K 150 100pF 100pF
2A
1 OUT1A _Z ZP
2A CN13 Z posit ion sensor
2 OUT2A _Z THP 1.2mm
2A CN5
3 SENSE1_Z 1.2mm 1
2A TH posi tion sensor
1
4 2A SENSE2_Z 1 2
1
5 2A OUT1A _TH 2 3
6 2A OUT2A _TH 3 4
7 2A SENSE1_TH 4 Z_POS_SENSOR 5
8 SENSE2_TH V CC TH_POS_SENSOR 5
1.2mm C7 JST B5B-PH-K- S
HE14 8cts C35 JST B5B-PH-K- S 100pF
1
100pF
CN10
1A V CC
1
2 Z_POS_SENSOR
V CC
3 Z_HOME_SENSOR
GND C36 V CC
4 TH_POS_SENSOR
1.2mm + 10µF/10V C37
5 TH_HOME_SENSOR
100nF
6 TH_HIGHT_ SENSOR NC
1
1
HE14 12cts
2 2
3 3
4 4
TH_HOME_SENSOR 5 Z_HOME_SENSOR 5
C3 JST B5B-PH-K- S JST B5B-PH-K- S
100pF C27
100pF
V CC
V CC NC NC NC NC
R14 R15 C32 C31
if hall sensor is m ount 2.2K 150 100pF 100pF
NC
ZR
R7 R8 C16 C15 1.2mm CN3 NC
if hall sensor is m ount 2.2K 150 100pF 100pF
1
1
THT 2
1.2mm CN8
TH high t sensor 3
1 4
1
2 Z_RES_SENSOR NC 5
3 C30 JST B5B-PH-K- S
4 100pF
TH_HIGHT_ SENSOR 5
C14 JST 05FE-ST- V K-N
100pF
V CC V CC
TX
1.2mm
1
1 3 1.2mm CN9 NC
BA RCODE_Tx D OUT IN 1 TH sens or
2 1
1
reserve d
3 2
4 3 Modification :
L1
3 1 5 4 Ajout du condensateur C36 de 10µF/10 V en boitier B.
GND
10 Title
1
VCC
1
9 2A SENSE1_X 2 X posit ion sensor C30 100pF
10 2A SENSE1_X 3
11 2A OUT2A_X 4 SENSE1_Z
12 2A OUT2A_X X_POS_SENSOR 5 C28 100pF
13 2A SENSE2_X
C1 JST B5B-PH-K- S
14 2A SENSE2_X SENSE2_Z
TPA 100pF
15 2A OUT1A_Z
1.2mm C26 100pF
16 2A OUT1A_Z
VCC
1
17 2A SENSE1_Z OUT1A_TH
SENSE1_Z U1F
18 2A C24 100pF
19 2A OUT2A_Z
12 13
20 2A OUT2A_Z ADAPTATOR_SENSOR_TRIGGERED A DAPTATOR_SENSOR NC OUT2A_TH
21 2A SENSE2_Z
R2 R5 C8 C5 C22 100pF
22 2A SENSE2_Z
74HC14A /SO 2.2K 150R 100pF 100pF
23 2A OUT1A_TH SENSE1_TH
24 2A OUT1A_TH
XH CN11 C20 100pF
25 2A SENSE1_TH
V CC 1.2mm
26 2A SENSE1_TH 1 SENSE2_TH
1
1.2mm
27 2A OUT2A_TH 2 X home sensor
1
U1E C37 100pF
28 2A OUT2A_TH 3 V CC
29 2A SENSE2_TH 4
10 11
30 1A SENSE2_TH X_HOME_SENSOR 5
VCC C36 100pF
31 1A C2 CN5
JST B5B-PH-K- S Z_POS_SENSOR
32 100pF 2A
X_POS_SENSOR 74HC14A /SO OUT1A_Z
33 + C48 C49 1 2A C34 100pF
34 X_HOME_SENSOR 10µF/10V 2 2A OUT2A_Z
100nF VCC
35 OPEN_DOOR_SENSOR 3 2A SENSE1_Z Z_HOME_SENSOR
36 Z_POS_SENSOR 4 2A SENSE2_Z
C32 100pF
37 1A Z_HOME_SENSOR 5 2A OUT1A_TH
38 6 2A OUT2A_TH TH_POS_SENSOR
39 TH_POS_SENSOR 7 2A SENSE1_TH
R3 R6 C9 C6 C31 100pF
40 TH_HOME_SENSOR 8 SENSE2_TH
2.2K 150R 100pF 100pF
41 TH_HEIGHT_SENSOR TH_HOME_SENSOR
HE14 8cts
42 1A ADAPTATOR_SENSOR_TRIGGERED
C29 100pF
43 DL CN12
44 BA RCODE_RxD TH_HEIGHT_SENSOR
GND 1.2mm
45 1A BA RCODE_Tx D 1
1.2mm C27 100pF
1
46 2 Door lo cking
1
1
2 9 BARCODE_Tx D
C38 C39 JST B6B-EH-A U1C positio n
100pF 100pF 3 10
sensor
6 5 4 11 R17 NC
CLOSED_DOOR_SENSOR 5 12 RACK_OUT_SENSOR
RACK_IN_SENSOR
C10 R10 NC JST B5B-PH-K- S HE14 12cts
2A 74HC14A /SO 100pF 4.7K 0R0
R16 NC
C42 VCC
100pF
CN2 0R0
SOL_POS
C44 1 Solenoi d for door l ocking
D1 47nF (2220 ) p lastique 2 R8 R14 C17 C14
SMBJ33A 3 2.2K 150R 100pF 100pF
SOL_NEG JST B3B-PH-K-S
RI CN8
C40 1.2mm
100pF 2A 1
1
2 Rack ou tput
U1B
3 sensor RACK_POS_SENSOR replaced by CLOSED_DOOR_SENSOR
4 3 4
RACK_OUT_SENSOR_TRIGGERED 5 ADIL INSTRUMENTS
C12 R12 NC JST B5B-PH-K- S
74HC14A /SO 100pF 4.7K Title
DIA 904 SAMPLING ARM CONNECTION BOARD X
RACK_OUT_SENSOR Size Document Number Rev
A3 DA E 002 E/3 3
1
PT100_HEAT_S2 2 PT100_EXT_S2 2
C19 PT100_HEAT_S3 3 PT100_EXT_S3 3
R2 R1 C17 C22
NC 2.2K 150 100pF 100pF 4 4
VCC
C39 C41 C44 JST B4B-EH-A C38 C40 C43 JST B4B-EH-A
+
CN5 CAMO CAMI 100pF 100pF 100pF HE AT S 100pF 100pF 100pF EX T S
10µF/10V 1.2mm 1.2mm CN9
1
1
2 1
3 CAM_FLAG 2
U1A
4 3
5 TACKY_A 4
2 1
6 CAM_FLAG 5 TAO TAI
7 TACKY_B
C24 JST B5B-PH-K-S 1.2mm 1.2mm C20
8
1
74HC14A /SO 100pF CA M S C21 100pF
9 SENSOR_0
U1B 100pF CN8
10 SENSOR_1
11 SENSOR_2 VCC 1
4 3
12 FRONT_DOOR_LOCK_SENSOR TACKY_A 2
13 3
HP_SENSOR_SUPPLY U1C
14 VCC C23 4
74HC14A /SO
15 6 5 100pF JST B4B-PH-K-S
16 HP_SENSOR_POS TACKY_B HP P E
17
18 HP_SENSOR_NEG
74HC14A /SO
19 R10 R9 C45 C48
1
20 HP_SENSOR_GND
20K 150 100pF 100pF
21 1.2mm 1.2mm
22 PT100_HEAT_S1
TBO TBI
23 FDL CN1
24 PT100_HEAT_S2
1.2mm
25 1
1
26 PT100_HEAT_S3 2 C31 C27
27 3 100pF R3 100pF
28 PT100_EXT_S1 4 0R0 CN6
29 FRONT_DOOR_LOCK_SENSOR 5
2
L1
30 PT100_EXT_S2 HP_SENSOR_SUPPLY 1
C47 JST B5B-PH-K-S
GND
31 100pF F DOOR S 1 3 2
32 PT100_EXT_S3 HP_SENSOR_NEG OUT IN 3
33 NFM40R01C101 4
34 VCC 5
NFM40R01C101 6
HE10 34 CTS
1 3 JST B6B-PH-K-S
GND
HP_SENSOR_POS OUT IN HP S
VCC
HP_SENSOR_GND
L2 C34 C37
C18 100pF 100pF
2
S0I S1I S2I
VCC
S0O 1.2mm 1.2mm 1.2mm C26
1
1.2mm 100nF R4 R5 R6 R7 100pF
1
74HC14A /SO
5
U1E
6
10 11 R8 Molex 70543
SENSOR_1 HP V S
S2O 0R0 C25 C29 C32 C35 C28 C30 C33 C36
1.2mm NC 100nf 100nF 100nF 100nF 100pF 100pf 100pf 100pf
1
74HC14A /SO
U1F
12 13
SENSOR_2
74HC14A /SO
FRONT_DOOR_LOCK_SENSOR
C5 C3
100pF 100pF HP VAL VE
CN14
2A C13 C9
OUT1A_HPV 8 100pF 100pF
2A 7 CN11 1A
CN12 OUT2A_HPV MIXING_VALVE_POS
6 2A CN15
1 HP_PUMP_POS 5 HEATER_POS 1 C12
2 HP_PUMP_POS SENSE1_HPV 2A 4 C15 2 1
D2 47nF (2220)
3 HP_PUMP_POS 3 2
Molex 5566-02 SMBJ33A
4 HP_PUMP_POS SENSE2_HPV 2A 2
HP_PUMP_NEG MIXING_VALVE_NEG JST B2B-PH-K-S
5 1 100pF 2A 1A LP M V
6 HP_PUMP_NEG HP V M
C8 C11 C10
7 HP_PUMP_NEG
100pF 100pF Molex 6373 8cts C14 100pF
8 HP_PUMP_NEG
9 MIXING_VALVE_POS
CN10
10 MIXING_VALVE_NEG
100pF
11 OUT1A_HPV
C2 1 Modificat ion :
12 OUT1A_HPV 2 5 La résistance R8 n'est plus montée
100pF C16
13 SENSE1_HPV
SENSE1_HPV Molex 5566-02
14 Im ax = 4A HP PUM P
15 OUT2A_HPV HP_PUMP_POS
100pF C46 4 Remplacement de la valeur de la rési stance R10 elle passe
16 OUT2A_HPV
17 SENSE2_HPV CN16 2A 100pF de 2.2K a 20K
D1 C4 1 2 1A CN2
18 SENSE2_HPV + + C6 SOL_B_POS
SMBJ33A 47nF (2220) 3 4
19 HEATER_POS + + 1 3 Retournement du connecteur CN14, Les sorti es sont sur
C42
20 HEATER_POS 2
Molex 5566-04 CN13 D3 47nF (2220) les broches 1, 3,5,7. La mise en série des bobines se fait sur
21 HEATER_NEG HP P M 3
100pF SMBJ33A
22 HEATER_NEG HP_PUMP_NEG Im ax = 4A 1
JST B3B-PH-K-S les broches 2 et 6, 4 et 8.
23 SOL_B_POS HEATER_NEG 2A 2 SOL_B_NEG 1A SO L
C1
24 SOL_B_NEG
100pF C7 Molex 5566-02 C49
HE10 24 CTS 100pF 100pF ADIL INSTRUMENTS
Title
DIA904 HIGH PRESSURE CONNECTION BOARD
HE10 26cts
1 OUT1A_SYR
C14 C24 C18
2 OUT1A_SYR
100pF
3 SENSE1_SY R
100pF 100pF
4 SENSE1_SY R 1A
OUT2A_SYR SY RINGE_VA LVE_POS CN3
5 2A
OUT2A_SYR CN5 OUT1A_SYR
6 2A 1
7 SENSE2_SY R 1 SENSE1_SY R 2A 2 SYR M
D5 C16
8 SENSE2_SY R 2 SENSE2_SY R 3
SMBJ33A 47nF (2220 ) p lastique
9 SYRINGE_VA LV E_POS OUT2A_SYR 4
SYRINGE_VA LV E_NEG JST B2B-PH-K- S
10 3W V 2A
LIQUID_PUMP_POS Panduit idc 4 cts
11 1A C20 C22
12 LIQUID_PUMP_NEG SY RINGE_VA LVE_NEG
13 PINCH_VA LV E_1_POS
V CC C13 100pF 100pF
14 PINCH_VA LV E_1_NEG
1.2mm 100pF
15 PINCH_VA LV E_2_POS
1
16 PINCH_VA LV E_2_NEG
17 PINCH_VA LV E_3_POS
18 PINCH_VA LV E_3_NEG
19 V CC
20
21 SYR_HOME_SENSOR
C4
22 100pF
23 L_VA C_SENSOR_SUPPLY
A GND_V AC_SENSOR
1
24 GND 1A
25 L_VA C_SIGNAL PINCH_VA LVE_3_POS
1.2mm
26 C9
CN4 100pF CN9
D2 C6
1A SMBJ33A 47nF (2220) p lastique 1
LIQUID_PUMP_POS 2
JST B2B-PH-K-S
CN6 RED PV
D3 C11 1A
1 PINCH_VA LVE_3_NEG
SMBJ33A 47nF (2220 ) p lastique
2 C2
JST B2B-PH-K- S 100pF
LIQ P
1A
LIQUID_PUMP_NEG
C7
100pF
C3
100pF
1A
PINCH_VALVE_2_POS
CN8
D1 C5 1A
SMBJ33A 47nF (2220 ) plastique 1A 1
2
JST B2B-PH-K- S
YELL PV
1A
PINCH_VALVE_2_NEG
C1
V CC 100pF
SY R CN2 C10
1.2mm 100pF
1 Syringe home
1
2 sensor 1A
3 PINCH_VALV E_1_POS
4
SY R_HOME_SENSOR 5
CN7
C19 Panduit idc 5 cts D4 C12
100pF SYR M SMBJ33A 47nF (2220 ) plastique 1
2
JST B2B-PH-K- S
GREEN PV
1A
PINCH_VALV E_1_NEG
C8
100pF
A GND VA C
1.2mm 1.2mm
1
VACUUM SENSOR
U1
L_VA C_SIGNA L 1
AGND_VA C_SENSOR 2
L_V A C_SENSOR_SUPPLY 3
C17 4
5
6
1
C15 CN1
1 ADIL INSTRUMENTS
2 Title
3 DIA 904 SYRINGE CONNECTION BOARD
4
C26 C25 C23 JST B4B-PH-K-S Size Document Number Rev
100pF 100pF 100pF VAC S A3 DAE 004 E/2 2
C12 VCC
1.2mm
+
VCC
1
CN4 10µF/10V
1 D_VA C_ SENSOR_SUPPLY
2 D_VA C_ SIGNAL2
3 D_VA C_ SIGNAL1
R7 R9 C20 C18
4 AGND_V AC_SENSOR NC 2.2K 825R 100pF 100pF
5 VCC
6 BD1
7 BUB_DET_1 Bubble detector 1
1.2mm Grain de caf e CN1
8 BUB_DET_2
1
9 BUB_DET_3 1
1
10 JP3 2 2
11 D_VA CUUM_PUMP_POS 3
12 4
13 D_VA CUUM_PUMP_NEG BUB_DET_1 5
3
14 C16 JST B5B-PH-K-S
100pF R8
HE10 14 cts 1K NC
D_VAC_SENSOR_SUPPLY
1
GND
1.2mm
VCC
U1
1
2
3 R4 R6 C14 C13
4 NC 2.2K 825R 100pF 100pF
5
AGND VA C1 6 BD2
1.2mm 1.2mm MPX5100DP CASE 867C-0 5 issue F 1.2mm Grain de caf e CN3 Bubble detector 2
1
1
1
JP2 2 2
3
CN2
4
VA CUUM_SIGNAL_1 1 BUB_DET_2 5
3
AGND_VA C_SENSOR 2 C11 JST B5B-PH-K-S
D_VAC_SENSOR_SUPPLY 3 100pF R5
C15 4 1K NC
D_VAC_SENSOR_SUPPLY
JST B4B-PH-K- S
C21 C19 C17
100pF 100pF 100pF 100nF
1
1.2mm
VDD
VCC
U2
R1 R3 C10 C9
1 NC 2.2K 825R 100pF 100pF
2
3 BD3
4 1.2mm Grain de caf e CN6 Bubble detector 3
5
1
6 1
1
VA C2 JP1 2 2
MPX5100DP CASE 867C-0 5 issue F
1.2mm 3
4
1
BUB_DET_3 5
3
C8 JST B5B-PH-K-S
100pF R2
CN5 1K NC
VA CUUM_SIGNAL_2 1 D_VAC_SENSOR_SUPPLY
AGND_VA C_SENSOR 2
D_VA C_SENSOR_SUPPLY 3
C7 4
JST B4B-PH-K- S
C3 C2 C1 100nF
100pF 100pF 100pF
R27
E
PSW0
10K VC CAFF
PSW1 VCCAFF VC CAFF
PSW2 R28 B Q2
PSW3 10K BC 857B
NFM41P11C204T1
15 CL2(SHFCLK)
+12LCD R34 D2 D3
F3 B1 PSW4 R21 VC CAFF
+ C11 20K BAT54S BAT54S
FLM(FLM)
E
+12LCD 1
C
C1 10uFCTT16V 10K
CL1(LP)
RXD
TXD
2 100nF R10 R14 GND1
GND GND B2 ANL0 B Q1
GND 3 8K2 1K C39 R22 10K BC 857B
D0
D1
D2
15 D3
M
GND 4 NFM40R01C101T1 1 4,7nF NC
GND 680pF
+5LCD 5 +5LCD 2
12
12
DS90C031TM RXD F4 C17 C21 MOLEX 52043-0410
4
6 3 GND1 GND1 GND1
C
F2 NFM41P11C204T1 + C12
C
U2 VC C R11 R15 B7
7 C2 10uFCTT16V 4
TXD ANL1 R23 GND1 TC1
GND 8 5 1
EN
EN
8K2 1K
DI1
DI2
DI3
DI4
EN1
DI1
DI2
DI3
DI4
EN1
D3 100nF NC B Q4 TC2
9 C3 C4 NFM40R01C101T1 F5 6 10K 2
D2 680pF 4,7nF C37 BC 847B TC3
10 7 C22 R24 3
D1 GND C18 VCCAFF TC4
DO1+
DO2+
DO3+
DO4+
DO1+
DO2+
DO3+
DO4+
100nF 100nF
DO1-
DO2-
DO3-
DO4-
DO1-
DO2-
DO3-
DO4-
11 8 GND1 GND1 10K GND1 4
E
D0 R12 R16 GND1
E
12 GND 9 10K R29
U1 GND ANL2 VC CAFF VC CAFF
13 10 8K2 1K R30 Q3
DS90C031TM B
14 NFM41P11C204T1 4,7nF C40 10K BC 857B
VCCPC HE10-10C 680pF
15
2
3
6
5
10
11
14
13
2
3
6
5
10
11
14
13
F6 C19 C23 NC D4
16 GND1 GND1 BAT54S
C
B5 R13 R17
17
C
+ C13 ANL3 R25
18 C5 10uFCTT16V 1 8K2 1K NC Q5 D1
B
19 100nF 2 4,7nF C38 10K BC 847B BAT54S
680pF
20 JST B 2B-PH-SM3-TB C24 GND1 R26
C20
21 GND1 GND1 10K GND1
GND B005
E
22
23 1
GND1 GND1 B007
RX
EEPCK
TX
EEPSO
EEPCS
24 2 GND1
EEPSI
TC1
25 NC JP3 JP4 1
TC2
26 R31 NC NC 2
TC3
27 VC CAFF R32 10K 3
VCCAFF VC CAFF TC4
28 10K GND1 4
29
NC
30
64
63
62
61
60
59
58
57
56
55
54
53
52
51
50
49
31 VC CAFF
U11
32
8
33
UC KI
UC KO
RXD
EROMSO
PnPCEL
VC C
EROMCS
V SS
EROMSI
TXD
EROMCK
OS0
OS1
NC
NC
NC
VC C U12
34
CL2(SHFCLK) +12LCD U5
35 +5LCD
VCC
FLM(FLM) NC 1 48 EEPCS 1
36 NC NC CS
M 2 47 EEPCK 2
NC U4 37 GND VC CAFF S/P NC SK
CL1(LP) U6 ANL0 3 46 EEPSO 3
38 NC VC CPC EMSR0 NC DI
U3 ANL1 4 45 EEPSI 4
GND
NC GND 39 EMSR1 NC DO
ANL2 5 44
40 EMSR2 NC
GND ANL3 6 43
41 B3 EMSR3 NC
7 42 PSW0 NC
42 NC PSW0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
70246-2621 8 41
43 NC NC
5
9 40
44 NC NC GND1
10 39
45 NC NC
11 38
46 VC CAFF AVCC NC VCCAFF
12 37
47 AVREF NC
C25 GND1 13 36
2
B4
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
48 GND1 AVSS NC
1
2
3
4
5
6
7
8
9
70247-2601 14 35 U13
49 VC CAFF GND1 VC CAFF CSEL NC
VC CAFF 15 34
50 PINT NC
16
VCC
100nF RX 16 33 EEPCS 3
RXINT NC CS
HARWIN M22-6122522 VC CAFF EEPCK 4
PnPINT1
PnPINT0
R1 SK
MOD0
MOD1
VCCAFF 18 EEPSO 5
PSW4
PSW3
PSW2
PSW1
ON-OFF DI
14
GND
RST
VSS
0 EEPSI 6
NC
NC
NC
NC
X0
X1
GND
VC CAFF DO
U18A U18B U18C GND1 9 10 RX
D7 +12LCD1 R2IN R2OUT
1 2 3 4 5 6 14 13 CTS MB89635R1120
BAS21LT1 GND1 R1IN R1OUT
74HC 14 74HC 14 74HC14 S29390FAJ
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
C41 GND1 8 11 TX
U18D T2OUT T2IN
7
2.2uF 9 8 15 12
T1OUT T1IN GND1
7
PSW4
PSW3
PSW2
PSW1
GND1 74HC14 GND1
GND1 GND1 6 C31 GND1 GND1
CTS
RES
C2- 100nF X1
C29 7
R2 R3 R4 R5 R6 R7 R8 R9 GND1 100nF -9VOUT
5 8MHz 4HSMX VC CAFF VC CAFF
L2 C2+
4
C1-
100
100
100
100
100
100
100
100
VC CAFF C28 NC VEEAFF C30 3 C34 22uFCTT35V 22uFCTT35V
VC CAFF +9VOUT
VCC
100nF 2 C32 C33 22pF 22pF C35 + C36 +
C1+ C7 C15
L1 D6 100nF
NC 100nF 100nF
LQH3C100K34 220nF STPS340U
JP6
U15 GND1 GND1
SP310ECT INV2
5
17
D5 R36 3 GND1 GND1
STPS340U VC CAFF IN+
330K 1% 1
ON/OFF
VIN
VSW
2
GND
4 3
SHDN FB NC
U7
GND
GND
C27 VC CAFF NC INV1 B9
10
14
15
10
14
15
+ C26 22uFCTT35V U8 1 F10 2 1 3
2
1
6
7
2
1
6
7
RI2-
RI3-
RI4-
RI1-
RI2-
RI3-
RI4-
2
3
RI1+
RI2+
RI3+
RI4+
RI1+
RI2+
RI3+
RI4+
RO2
RO3
RO4
RO1
RO2
RO3
RO4
+ C16
EN1
EN1
EN
C10
B8 GND1 MAX809_EUR T 1 5
IN+ RTHV
1 R19 VC CAFF RES VC CAFF 100nF 10uFCTT16V OUT1
3
2 4
2 82K VC CAFF GND OUT2
3
11
13
4
12
11
13
4
12
D
1K INVTDKCXAL10L
1
NC 4 10K
GND GND GND1 CL1(LP)A GND1 U17
1
U18E
GND
JP2 JST B4B-PH-SM3-TB VOAFF
11 10 D3A CL2(SHFCLK)A CLR R37 VC CAFF
1 F11
VC C
NC D8 74HC14 D2A RX 2 7 ICSPData BKLGT G Q6 2 B6
BAS21LT1 GP5 GP0 100 IRLR014 12
GND1 GND1 2 D1A TX 3 6 ICSPCLK NC
GP4 GP1 1
DISPOFF
GND
GND
GP3 GP2 2
3
PCB NC GND1 GND1
CI14 NC CI18 NC CI22 NC VCCAFF
P1 330K VOAFF NC NC
VC CAFF
PCB PCB PCB NC
GND1
C42 13
U18F
12 D0A
GND1 PROG VCCAFF NC
1
12
2
F13 NC
NC
3
PCB PCB PCB R18 10uFCTT16V ICSPCLK GND1 GND1
3K9 5 C8 C9
CI16 NC CI20 NC CI24 NC ICSPData
CI1 NC CI3 NC CI5 NC CI7 NC 4
PCB PCB PCB 100nF 100nF GND1
3 GND1
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A13 A12
OUT1A_SYR SPI_IN PE0MC PE0MC C169
1 1 GND
SPI_OUT 100pF
2 GND 2
SENSE1_SYR SPI_CLK RES RES C178
3 3
SPI_OUT 100pF
4 4
OUT2A_SYR Micro MC Commande Moteurs Commande solenoides SPI_OUT C170
5 GND 5
DIA904-3.Sch DIA904-1.Sch DIA904-2.Sch SPI_IN NC
6 6
SENSE2_SYR CSCANHP OUT1A_X LIQUID PUMP LIQUID PUMP+ SPI_IN C179
7 CSCANHP OUT1A_X LIQUID PUMP LIQUID PUMP+ GND 7
SENSE1_X PINCH VALVE1 LIQUID PUMP- SPI_CLK NC
8 SENSE1_X PINCH VALVE1 LIQUID PUMP- 8
SYRINGE_VALVE+ IO1CPLD OUT2A_X PINCH VALVE2 CS0 SPI_CLK C188
SPI_IN
SPI_CLK
SPI_CLK
SPI_OUT
SPI_OUT
9 IO1CPLD PH_1_X PH_1_X OUT2A_X PINCH VALVE2 9
MC D3 adr: 285 SYRINGE_VALVE- FAN_S SENSE2_X D_VACUUM_PUMP PINCH VALVE1+ CS1 NC
10 FAN_S PH_2_X PH_2_X SENSE2_X D_VACUUM_PUMP PINCH VALVE1+ 10
LIQUID PUMP+ SYRINGE_VALVE PINCH VALVE1- CS2 CS0 C187
11 DISABLE_X DISABLE_X SYRINGE_VALVE PINCH VALVE1- 11
MC PT3 LIQUID PUMP- BARCODE RXD VREF_X RESERVE_VALVE CS3 100pF
12 BARCODE RXD I0_1_X I0_1_X VREF_X RESERVE_VALVE 12
PINCH VALVE1+ BARCODE TXD PINCH VALVE3 PINCH VALVE2+ DSYNC CS1 C195
13 BARCODE TXD I1_1_X I1_1_X PINCH VALVE3 PINCH VALVE2+ 13
MC D0 adr: 285 PINCH VALVE1- TXDTCRS SOL PINCH VALVE2- CS5 100pF
14 TXDTCRS I2_1_X I2_1_X SOL PINCH VALVE2- 14
PINCH VALVE2+ RXD MOT1 CS2 C206
15 RXD I3_1_X I3_1_X MOT1 VCC 15
MC D1 adr: 285 PINCH VALVE2- TXD0_HP MOT2 D_VACUUM_PUMP+ 100pF
16 TXD0_HP I0_2_X I0_2_X MOT2 D_VACUUM_PUMP+ GND 16
PINCH VALVE3+ RTS_HP LP MIXING VALVE1 D_VACUUM_PUMP- CS3 C207
17 RTS_HP I1_2_X I1_2_X LP MIXING VALVE1 D_VACUUM_PUMP-
MC D5 adr: 285 PINCH VALVE3- TXD_TC LP MIXING VALVE2 HE10-16 100pF
18 TXD_TC I2_2_X I2_2_X LP MIXING VALVE2
RTS_TC HP PUMP SYRINGE_VALVE+ A15 DSYNC C212
19 VCC RTS_TC I3_2_X I3_2_X HP PUMP SYRINGE_VALVE+
FAN SYRINGE_VALVE- SOL+ 100pF
20 GND CSPOT_X CSPOT_X FAN SYRINGE_VALVE- 1
MC PD3 SYR_HOME_SENSOR R+ HEATER1 CS5 C210
21 R+ HEATER1 2
R- OUT1A_Z HEATER2 RESERVE_VALVE+ SOL- MC D6 adr: 285 100pF
22 GND R- OUT1A_Z HEATER2 RESERVE_VALVE+ 3
LV PT1 T+ SENSE1_Z RESERVE_VALVE-
23 VDDA T+ PH_1_Z PH_1_Z SENSE1_Z RESERVE_VALVE- 4
CAV1 T- OUT2A_Z LIQUID PUMP_S MOT+ MC D1 adr: 286 -> 0:+ ,1:-
24 GNDA T- PH_2_Z PH_2_Z OUT2A_Z LIQUID PUMP_S 5
MC PAD3 WASTE_VAC_SIGNAL SENSE2_Z PINCH VALVE1_S PINCH VALVE3+ MOT- MC D2 adr: 286 -> 0:OFF ,1:ON
25 DISABLE_Z DISABLE_Z SENSE2_Z PINCH VALVE1_S PINCH VALVE3+ 6
CSHPPC PINCH VALVE2_S PINCH VALVE3- OUT1A_X
26 GND CSHPPC I0_1_Z I0_1_Z PINCH VALVE2_S PINCH VALVE3- 7
CSTC D_VACUUM_PUMP_S
CSTC I1_1_Z I1_1_Z D_VACUUM_PUMP_S 8
HE10-26 CSRACK SYRINGE_VALVE_S SOL+ SENSE1_X
CSRACK I2_1_Z I2_1_Z SYRINGE_VALVE_S SOL+ 9
A10 CS0 RESERVE_VALVE_S SOL-
CS0 I3_1_Z I3_1_Z RESERVE_VALVE_S SOL- 10
HP PUMP+ CS1 PINCH VALVE3_S OUT2A_X
1 CS1 I0_2_Z I0_2_Z PINCH VALVE3_S 11
CS2 SOL_S MOT+
2 CS2 I1_2_Z I1_2_Z SOL_S MOT+ 12
CS3 MOT_S MOT- SENSE2_X
3 CS3 I2_2_Z I2_2_Z MOT_S MOT- 13
HPPC PT3 DSYNC LP MIXING VALVE_S
4 DSYNC I3_2_Z I3_2_Z LP MIXING VALVE_S 14
HP PUMP- CS5 HP PUMP_S LP MIXING VALVE+ OUT1A_Z
5 CS5 HP PUMP_S LP MIXING VALVE+ 15
DOOR_LOCK_SENSOR OUT1A_TH FAN_S LP MIXING VALVE-
6 DOOR_LOCK_SENSOR OUT1A_TH FAN_S LP MIXING VALVE- 16
PE0MC SENSE1_TH HEATER_S SENSE1_Z
7 PE0MC PH_1_TH PH_1_TH SENSE1_TH HEATER_S 17
PE1MC OUT2A_TH HP PUMP+
8 PE1MC PH_2_TH PH_2_TH OUT2A_TH HP PUMP+ 18
HPPC PD0 (12v) LP MIXING VALVE+ PE3MC SENSE2_TH HP PUMP- OUT2A_Z
9 PE3MC DISABLE_TH DISABLE_TH SENSE2_TH HP PUMP- 19
HPPC PD1 (26v) LP MIXING VALVE- XTAL_MC
10 XTAL_MC I0_1_TH I0_1_TH 20
OUT1A_HPV RES SOLB1+ SENSE2_Z
11 RES I1_1_TH I1_1_TH SOLB1+ 21
VREF_X SOLB-
12 VREF_X I2_1_TH I2_1_TH SOLB- 22
SENSE1_HPV VAC_SIGNAL OUT1A_TH
13 VAC_SIGNAL I3_1_TH I3_1_TH 23
AUX_DEG_VAC_SIG HEATER+
14 AUX_DEG_VAC_SIG I0_2_TH I0_2_TH HEATER+ 24
OUT2A_HPV WASTE_VAC_SIGNAL HEATER- SENSE1_TH
15 WASTE_VAC_SIGNAL I1_2_TH I1_2_TH HEATER- 25
VREF_SYR
16 VREF_SYR I2_2_TH I2_2_TH 26
SENSE2_HPV LIQUID PUMP_S OUT2A_TH
17 LIQUID PUMP_S I3_2_TH I3_2_TH 27
PINCH VALVE1_S
18 PINCH VALVE1_S 28
HEATER+ PINCH VALVE2_S OUT1A_HPV SENSE2_TH
19 PINCH VALVE2_S OUT1A_HPV 29
TC RC1 D_VACUUM_PUMP_S SENSE1_HPV SPI_IN C236
20 D_VACUUM_PUMP_S PH_1_HPV PH_1_HPV SENSE1_HPV 30 GND
TC RC2 HEATER- SYRINGE_VALVE_S OUT2A_HPV 100pF
21 SYRINGE_VALVE_S PH_2_HPV PH_2_HPV OUT2A_HPV VCC 31
RESERVE_VALVE_S SENSE2_HPV Micro HPPC SPI_OUT C237
22 RESERVE_VALVE_S DISABLE_HPV DISABLE_HPV SENSE2_HPV GND 32
SOLB1+ PINCH VALVE3_S DIA904-4.Sch X_POS_SENSOR MC PD0 100pF
23 PINCH VALVE3_S I0_1_HPV I0_1_HPV 33
MC D4 adr: 285 SOLB- SOL_S X_HOME_SENSOR MC PD1 SPI_CLK C239
24 SOL_S I1_1_HPV I1_1_HPV 34
MOT_S CSHPPC HP PUMP DOOR_LOCK_SENSOR MC PE7 100pF
MOT_S I2_1_HPV I2_1_HPV CSHPPC HP PUMP 35
HE10-24 SPI_CLK Z_POS_SENSOR MC PD4 CSRACK C240
I3_1_HPV I3_1_HPV SPI_CLK 36
A9 HP_SENSOR SPI_IN LP MIXING VALVE1 Z_HOME_SENSOR MC PD5 100pF
HP_SENSOR I0_2_HPV I0_2_HPV SPI_IN LP MIXING VALVE1 37
LIQUID PUMP SPI_OUT LP MIXING VALVE2 TXDTCRS C241
1 VCC LIQUID PUMP I1_2_HPV I1_2_HPV SPI_OUT LP MIXING VALVE2 GND 38
PINCH VALVE1 TH_POS_SENSOR MC PAD4 100pF
2 GND PINCH VALVE1 I2_2_HPV I2_2_HPV 39
MC PT0 CAM FLAG IN PINCH VALVE2 XTAL_MC TH_HOME_SENSOR MC PAD5 PE1MC C242
3 PINCH VALVE2 I3_2_HPV I3_2_HPV XTAL_MC 40
D_VACUUM_PUMP RES TH_HIGHT_SENSOR MC PAD6 100pF
4 GND D_VACUUM_PUMP RES 41
TACHY A IN SYRINGE_VALVE OUT1A_SYR RXD ADAPTOR_SENSOR MC PD6 IO1CPLD C243
5 SYRINGE_VALVE OUT1A_SYR RXD 42
RESERVE_VALVE SENSE1_SYR TXD0_HP 100pF
6 GND RESERVE_VALVE PH_1_SYR PH_1_SYR SENSE1_SYR TXD0_HP NFM40R100pF GND 43 F1
TACHY B IN PINCH VALVE3 OUT2A_SYR PE3MC RTS_HP BARCODE RXD IO2CPLD C244
7 PINCH VALVE3 PH_2_SYR PH_2_SYR OUT2A_SYR PE3MC RTS_HP 44 F2
SOL SENSE2_SYR BARCODE TXD NFM40R100pF 100pF
8 GND SOL DISABLE_SYR DISABLE_SYR SENSE2_SYR 45
MC PT4 SENSOR IN 0 MOT1 IO3CPLD C245
9 MOT1 I0_1_SYR I0_1_SYR GND 46
MC PT5 SENSOR IN 1 MOT2 VREF_SYR HP_SENSOR RACK_POS_SENSOR MC PD2 100pF
10 MOT2 I1_1_SYR I1_1_SYR VREF_SYR HP_SENSOR 47
MC PT6 SENSOR IN 2 X_POS_SENSOR LP MIXING VALVE_S RACK_IN_SENSOR MC PD7
11 X_POS_SENSOR I2_1_SYR I2_1_SYR LP MIXING VALVE_S 48
MC PT7 FR D LOCK SENSOR X_HOME_SENSOR HP PUMP_S RACK_OUT_SENSOR MC PAD7
12 X_HOME_SENSOR I3_1_SYR I3_1_SYR HP PUMP_S 49
RACK_POS_SENSOR
13 GND RACK_POS_SENSOR I0_2_SYR I0_2_SYR 50
SYR_HOME_SENSOR CAM FLAG
14 VRef SYR_HOME_SENSOR I1_2_SYR I1_2_SYR CAM FLAG C233 C234 C235 C238
Z_POS_SENSOR TACHY A GND HE10-50
15 GND Z_POS_SENSOR I2_2_SYR I2_2_SYR TACHY A 100pF 100pF 100pF 100pF
HP_SENSOR+ Z_HOME_SENSOR TACHY B A14
16 Z_HOME_SENSOR I3_2_SYR I3_2_SYR TACHY B
ADAPTOR_SENSOR R+
17 GND ADAPTOR_SENSOR CSPOT_SYR CSPOT_SYR 1
HP_SENSOR- RACK_IN_SENSOR R-
18 RACK_IN_SENSOR 2
TH_POS_SENSOR T+
19 GND TH_POS_SENSOR U1A 3 CI7 CI11
TH_HOME_SENSOR C265 Micro TC T-
20 GNDA TH_HOME_SENSOR 4
TH_HIGHT_SENSOR 1 2 CAM FLAG DIA904-5.Sch SPI_IN D I V E RS D I V E RS
21 GND TH_HIGHT_SENSOR 5
PT100HEAT1 RACK_OUT_SENSOR 74HC14SO R66 NC NC
22 RACK_OUT_SENSOR GND 6
SENSOR 0 10nF CAM FLAG IN CSTC HEATER1 SPI_OUT
23 GND SENSOR 0 U1F 10K CSTC HEATER1 7 CI8 CI10
PT100HEAT2 SENSOR 1 SPI_IN HEATER2
24 SENSOR 1 SPI_IN HEATER2 GND 8
SENSOR 2 13 12 TACHY A SPI_OUT FAN SPI_CLK D I V E RS D I V E RS
25 GND SENSOR 2 SPI_OUT FAN 9
FR D LOCK SENSOR 74HC14SO SPI_CLK NC NC
26 2V5 FR D LOCK SENSOR SPI_CLK GND 10
IO2CPLD TACHY A IN CSRACK
27 GND IO2CPLD U1B 11 CI6 CI12
PT100EXT1 IO3CPLD RXD PT100HEAT1 TXDTCRS
28 IO3CPLD RXD PT100HEAT1 12
GEN FAN_S 3 4 TACHY B TXD_TC PT100HEAT2 PE1MC MC PE1 D I V E RS D I V E RS
29 GND GEN FAN_S TXD_TC PT100HEAT2 13
PT100EXT2 BUB1 74HC14SO RTS_TC NC NC
30 BUB1 RTS_TC GND 14
BUB2 TACHY B IN PT100EXT1 IO1CPLD
31 GND BUB2 U1E PT100EXT1 15 CI9 CI13
BUB3 C266 XTAL_MC PT100EXT2
32 2V5 BUB3 XTAL_MC PT100EXT2 GND 16
CSPBUB1 11 10 SENSOR 0 RES IO2CPLD D I V E RS D I V E RS
33 GND CSPBUB1 RES 17
CSPBUB2 74HC14SO R165 NC NC
34 VCC CSPBUB2 GND 18
VCC +12 +24 +26 CSPBUB3 10nF SENSOR IN 0 FAN_S IO3CPLD
PT7 PT8 PT9 PT10 HE10-34 CSPBUB3 U1C 10K FAN_S 19 CI15 CI16
BUB_DET_1 C267 HEATER_S
BUB_DET_1 HEATER_S GND 20
NC 5V NC +12V NC 24V NC 26V A1 BUB_DET_2 5 6 SENSOR 1 D I V E RS D I V E RS
BUB_DET_2
+ + + + BUB_DET_3 74HC14SO R166 HE10-20 NC NC
1 GND BUB_DET_3
10nF SENSOR IN 1 13K pour INA122UA A16
2 U1D 10K 4K75 1% CI2 CI18
C268 GEN FAN- R-
3 1
9 8 SENSOR 2 R7 R9 D I V E RS D I V E RS
4 2
GND GND GND GND X_POS_SENSOR C262 74HC14SO R167 C1 T+ NC NC
C11 C10 C8 C9 5 GND GND 1 GND 3 CI3 CI17
100pF GND10nF SENSOR IN 2 100pF U4
ECEV1VA101UP ECEV1VA101UP 6 10K 4
8 1 R+ D I V E RS D I V E RS
ECEV1VA101UP ECEV1VA101UP 7 RG+ RG- 5
X_HOME_SENSOR C263 C232 WASTE_VAC_SIGNAL 7 2 HP_SENSOR- NC NC
8 GND GND VDDA VCC IN- 6
100pF C2 100pF 6 3 HP_SENSOR+ T-
9 +24 OUT IN+ 7 CI1 CI19
5 4
10 VCC GND 2V5 REF VEE 8
RACK_POS_SENSOR C92 C88 DOOR_LOCK_SENSOR D I V E RS D I V E RS
DE1 11 +26 GND GND GND 9
R1 100pF 100nF 100pF MAX4194ESA GNDA NC NC
12 -12 10
HP
GND 470 VCC 13 +12 CI20 CI4
SYR_HOME_SENSOR C230 C95 CAM FLAG IN PT2 C18 HE10-10
14 GND GND
CL150UR3000 100pF 100pF R12 CAV1 GND GNDA A6 D I V E RS D I V E RS
15 VDDA GNDA
16 VCC PT3 PT4 GND 1 NC NC
DE4 Z_POS_SENSOR C264 C98 TACHY A IN 12K1 1% 100nF
R3 GND GND CAV1 CAV1 2
MOLEX 39-28-1163 100pF 100pF CI21 CI14
GND 2K7 +24 GNDA 3
D I V E RS D I V E RS
+24 4
11
CL150UR3000 Z_HOME_SENSOR C86 C99 TACHY B IN 10K GND GNDA
GND GND R10 R11 +12 5 NC NC
100pF 100pF C12
DE5 100nF VCC 6
R4 A2 2 GND
2V5 10K 1% GNDA C61
ADAPTOR_SENSOR C89 C108 HP_SENSOR+ R15 1 HP_SENSOR MOLEX 39-28-1063
GND 3K3 +26 1 GND GND GND 10K 100pF
100pF 100pF 3 R14
2 10K 1% U5A
CL150UR3000 R20 R51 A18
3 C17
RACK_IN_SENSOR C93 C110 HP_SENSOR- TLV2374ID VCC
DE2 4 GND GND 2V5 1M 100nF VCC 150 1
R2 100pF Détecteurs de bulles 100pF C19 A17
5 +12 GND 2
4
DIA904-6.Sch R19 RESERVE_VALVE+
GND 1K2 +12 6 20K 1% GNDA VRef 1 R46 C59VCC 3
TH_POS_SENSOR C90 BUB_DET1 GNDA RESERVE_VALVE-
7 GND BUB_DET1 2 2K2 100pF GND 4
CL150UR3000 100pF BUB_DET2 VDDA 100nF V1 V2
8 VCC BUB_DET2 5
PT11 BUB_DET3 U6 PT5 PT6 NC
DE3 BUB_DET3 GND C54
R5 NC MOLEX 39-28-1083 TH_HOME_SENSOR C87 BUB1 CAV1 CAV1 A11 FR D LOCK SENSOR NC
GND BUB1 100pF
A3 A4 100pF BUB2 GNDA
GND NC -12 BUB2 VDDA 1
BUB3 CSCANHP 1 2 VAC_SIGNAL
1 GND 1 GND BUB3 CS CH0 2
NC -12V TH_HIGHT_SENSOR C91 SPI_CLK 7 3 AUX_DEG_VAC_SIG AUX_DEG_VAC_SIG C146
2 2 GND CLK CH1 2V5 3 GND
100pF CSPBUB1 100pF
3 +12 3 +12 CSPBUB1 GNDA 4
CSPBUB2 SPI_IN 6 8 VAC_SIGNAL C147 GND
4 VCC 4 VCC CSPBUB2 DOUT VCC Vref VCC 5
RACK_OUT_SENSOR C94 CSPBUB3 SPI_OUT 5 4 100pF
GND CSPBUB3 DIN GND GNDA GND 6
MOLEX 39-28-1043 MOLEX 39-28-1043 100pF BUB_DET1 BUB_DET1 C155
7
A5 SPI_CLK BUB_DET2 100pF
SPI_CLK 8
SENSOR IN 0 C101 SPI_OUT BUB_DET3 BUB_DET2 C156
1 GND GND SPI_OUT GND 9
100pF LTC1288CS8 100pF
2 27K GND 10
BUB_DET_1 R45 D_VACUUM_PUMP+ BUB_DET3 C158
3 +12 BUB_DET_1 11
SENSOR IN 1 C102 BUB_DET_2 100pF
4 +24 GND BUB_DET_2 GND 12
CI5 100pF BUB_DET_3 D_VACUUM_PUMP-
5 BUB_DET_3 U21C 13
NC R39
6 VCC GND 14
SENSOR IN 2 C106 9 C52
+24 GND 100pF 1K GNDA
MOLEX 39-28-1063 100pF A9a GEN FAN_S 8 HE10-14
10 R6
1 VRef
A7 A8 FR D LOCK SENSOR C107 HP_SENSOR+ GEN FAN-
GND 2 1K
100pF HP_SENSOR- TLV2374ID
1 -12 1 +12 3 C3
GEN FAN- +
2 +12 2 4 GNDA 1uFCTT16V Title
3 GND PT15 5 GND INTERFA CE BOA RD
JST B2B-PH-SM3-TB NC GNDA
NC NC
Size Document Number Rev
A3 DA E 010 E/3 3
C208
GND VCC
100nF
8
R147
10 21 PU1 SENSE2_X
VCP1 OUT2B 1.2 SENSE2_X
VCC
C196
10nF C253
11 20 100pF
VCP2 SENSE2 GND
C197 12 C251
GND VBOOT 100pF
100nF 14 OUT2A_X
OUT2A OUT2A_X
C111
GND VCC
C209 7 35 SENSE1_X 100nF
GND TRI_CAP SENSE1 SENSE1_X
1nF
8
C252
R85
2 34 PU2 100pF 10 21 PU7 SENSE2_HPV
PH_1_X PH_1 OUT1B 1.2 GND VCP1 OUT2B 1.2 SENSE2_HPV
VCC
4 C26
I0_1_X I0_1 R134
3 C249 10nF C154
I1_1_X I1_1 100pF 100pF
32 5 OUT1A_X 11 20
I2_1_X I2_1 OUT1A OUT1A_X VCP2 SENSE2 GND
33
I3_1_X I3_1 C151
17 13 C114 12
PH_2_X PH_2 VS GND VBOOT 100pF
15 31 100nF 14 OUT2A_HPV
I0_2_X I0_2 VS +24 OUT2A OUT2A_HPV
16 VCCR62
I1_2_X I1_2
23 28
I2_2_X I2_2 VREF1 NC
22 26 VREF_X C23 7 35 SENSE1_HPV
I3_2_X I3_2 VREF2 VREF_X GND TRI_CAP SENSE1 SENSE1_HPV
6 1nF
EA_OUT1
EA_OUT2
DISABLE_X DISABLE R63
R59 C144
EA_IN1
EA_IN2
C171 NC 2 34 PU8 100pF
R157 100nF 51K PH_1_HPV PH_1 OUT1B 1.2 GND
U43 4 R78
2K2 I0_1_HPV I0_1 C145
L6258 U37 3
I1_1_HPV I1_1 100pF
GND 1 8 32 5 OUT1A_HPV
C67 L H I2_1_HPV I2_1 OUT1A OUT1A_HPV
7 GND 33
VCC W I3_1_HPV I3_1 VCC
29
30
25
24
10nF GND 3 6 17 13
CS VDD VCC PH_2_HPV PH_2 VS R23
4 5 15 31
DIN SCLK I0_2_HPV I0_2 VS +24 12K
16
C70 I1_2_HPV I1_2
MAX5400EKA-T 23 28
R145 R146 100nF I2_2_HPV I2_2 VREF1
22 26
330K 330K I3_2_HPV I3_2 VREF2
6
EA_OUT1
EA_OUT2
C189 C190 DISABLE_HPV DISABLE R24
GND
EA_IN1
EA_IN2
C22 10K
R17 U13 100nF
470pF 470pF 2K2 L6258
GND
CSPOT_X C29 GND
SPI_OUT VCC
29
30
25
24
10nF
SPI_CLK
C199
GND VCC
100nF
R79 R81
8
330K 330K
R150
10 21 PU3 SENSE2_Z C109 C112
VCP1 OUT2B 1.2 SENSE2_Z
VCC
C200
10nF C257
11 20 100pF 470pF 470pF
VCP2 SENSE2 GND
C201 12 C255
GND VBOOT 100pF
100nF 14 OUT2A_Z
OUT2A OUT2A_Z
C198 7 35 SENSE1_Z
GND TRI_CAP SENSE1 SENSE1_Z
1nF
C256
2 34 PU4 100pF
PH_1_Z PH_1 OUT1B 1.2 GND
4 R135
I0_1_Z I0_1 C254
3
I1_1_Z I1_1 100pF
32 5 OUT1A_Z
I2_1_Z I2_1 OUT1A OUT1A_Z
33
I3_1_Z I3_1 VCC
17 13 C182
PH_2_Z PH_2 VS R64 GND VCC
15 31 100nF
I0_2_Z I0_2 VS +24 12K
16
8
I1_2_Z I1_2
23 28 R123
I2_2_Z I2_2 VREF1
22 26 10 21 PU9 SENSE2_SYR
I3_2_Z I3_2 VREF2 VCP1 OUT2B 1.2 SENSE2_SYR
VCC
6 C184
EA_OUT1
EA_OUT2
DISABLE_Z DISABLE R60 10nF C220
EA_IN1
EA_IN2
30
25
24
10nF
C183 7 35 SENSE1_SYR
GND TRI_CAP SENSE1 SENSE1_SYR
1nF
C219
R148 R149 2 34 PU10 100pF
330K 330K PH_1_SYR PH_1 OUT1B 1.2 GND
4 R122
C191 C192 I0_1_SYR I0_1 C217
3
I1_1_SYR I1_1 100pF
32 5 OUT1A_SYR
I2_1_SYR I2_1 OUT1A OUT1A_SYR
33
470pF 470pF I3_1_SYR I3_1
17 13
PH_2_SYR PH_2 VS
15 31
I0_2_SYR I0_2 VS +24
16 VCCR55
I1_2_SYR I1_2
23 28
I2_2_SYR I2_2 VREF1 NC
22 26 VREF_SYR
I3_2_SYR I3_2 VREF2 VREF_SYR
C203 6
EA_OUT1
EA_OUT2
GND VCC DISABLE_SYR DISABLE R58
100nF R56
EA_IN1
EA_IN2
C168 NC
8
C204 GND 1 8
10nF C261 C63 L H
7 GND
VCC W
29
30
25
24
11 20 100pF 10nF GND 3 6
VCP2 SENSE2 GND CS VDD VCC
4 5
C259 DIN SCLK
C205 12
GND VBOOT 100pF C66
100nF 14 OUT2A_TH MAX5400EKA-T
OUT2A OUT2A_TH R124 R125 100nF
330K 330K
C202 7 35 SENSE1_TH C173 C174 GND
GND TRI_CAP SENSE1 SENSE1_TH
1nF
C260
2 34 PU6 100pF 470pF 470pF
PH_1_TH PH_1 OUT1B 1.2 GND
4 R151
I0_1_TH I0_1 C258 CSPOT_SYR
3
I1_1_TH I1_1 100pF
32 5 OUT1A_TH
I2_1_TH I2_1 OUT1A OUT1A_TH
33
I3_1_TH I3_1 VCC
17 13
PH_2_TH PH_2 VS R65
15 31
I0_2_TH I0_2 VS +24 12K
16
I1_2_TH I1_2
23 28
I2_2_TH I2_2 VREF1
22 26
I3_2_TH I3_2 VREF2
6
EA_OUT1
EA_OUT2
EA_IN2
C172 10K
R156 U45 100nF
2K2 L6258
GND
C69 GND
VCC
29
30
25
24
10nF
R152 R153
330K 330K
C193 C194
470pF 470pF
Title
INTERFA CE BOA RD
C181
VDDA
27K GND +26
R129 NC
100nF D15 C223
0
C74 C44 GND GND
U42 100pF
U38D 100nF 100nF PINCH VALVE3 10 1 PINCH VALVE3+ LIQUID PUMP+
R128 PINCH VALVE3 IN1 OUT1 LIQUID PUMP+
VCC
13 C175 9 2 LIQUID PUMP-
100pF 1K GNDA GND IN2 OUT2 LIQUID PUMP-
LIQUID PUMP_S 14 GNDA SOL 8 3 SOL+ R159
LIQUID PUMP_S SOL IN3 OUT3
12 R69 7 4 C224 1 a la place de 0.47
GND IN4 OUT4 GND 1 GND
LIQUID PUMP- 100pF
10K 6
TLV2374ID DIAG NC
+ C79 5 D16 C225
1uFCTT16V GND GND GND GND
100pF
VN330SP PINCH VALVE1+
15K GNDA PINCH VALVE1+
R139 PINCH VALVE1-
PINCH VALVE1-
C180 R160
C226
U38C GND +26 GND 1 GND
R140 100pF
9 C186 100nF
0
1K GNDA NC
PINCH VALVE1_S 8 100pF U41 D17 C227
PINCH VALVE1_S GND GND
10 R70 SYRINGE_VALVE 10 1 SYRINGE_VALVE+ 100pF
SYRINGE_VALVE IN1 OUT1
VCC
PINCH VALVE1- LIQUID PUMP 9 2 LIQUID PUMP+ PINCH VALVE2+
1K LIQUID PUMP IN2 OUT2 PINCH VALVE2+
TLV2374ID PINCH VALVE1 8 3 PINCH VALVE1+ PINCH VALVE2-
C80 PINCH VALVE1 IN3 OUT3 PINCH VALVE2-
+ PINCH VALVE2 7 4 PINCH VALVE2+ R161
1uFCTT16V PINCH VALVE2 IN4 OUT4
C228
GND 1 GND
6 100pF
15K GNDA DIAG
R130 5 NC
GNDA GND GND D4 C163
GND GND
11
VN330SP 100pF
U39A D_VACUUM_PUMP+
R131 D_VACUUM_PUMP+
2 C72 D_VACUUM_PUMP-
100pF 1K GNDA D_VACUUM_PUMP-
PINCH VALVE2_S 1 C128 R121
PINCH VALVE2_S
3 R71 C164 1 a la place de 0.47
GND +26 GND 1 GND
PINCH VALVE2- 100pF
1K 100nF
0
TLV2374ID NC
+ C81 U20 D3 D14 C221
GND GND
4
1uFCTT16V RESERVE_VALVE 10 1 SMBYW02-200 RESERVE_VALVE+ 100pF
VDDA RESERVE_VALVE IN1 OUT1
VCC
LP MIXING VALVE1 9 2 LPM LP MIXING VALVE+ SYRINGE_VALVE+
15K GNDA LP MIXING VALVE1 IN2 OUT2 SYRINGE_VALVE+
R88 8 3 SYRINGE_VALVE-
GNDA IN3 OUT3 SYRINGE_VALVE-
D_VACUUM_PUMP 7 4 D_VACUUM_PUMP+ R158
D_VACUUM_PUMP IN4 OUT4
11
C222
U17A GND 1 GND
R89 6 100pF
DIAG
2 C40 5 SM6T30A
100pF 1K GNDA GND GND D6
D_VACUUM_PUMP_S 1 C141
D_VACUUM_PUMP_S D1 GND GND
3 R52 VN330SP 100pF
D_VACUUM_PUMP- C32 SMBYW02-200 RESERVE_VALVE+
10K RESERVE_VALVE+
TLV2374ID RESERVE_VALVE-
LPMV2+
C53 GND +12 RESERVE_VALVE-
+ R105
4
1uFCTT16V 100nF C142
0
VDDA GND 0.33 GND
U16 100pF
15K GNDA
R126 LP MIXING VALVE2 10 1 NC
GNDA LP MIXING VALVE2 IN1 OUT1
VCC
FAN 9 2 SOLB+ D18 C231
FAN IN2 OUT2 GND GND
11
8 3 100pF
U38A IN3 OUT3
R127 7 4 PINCH VALVE3+
GND IN4 OUT4 PINCH VALVE3+
2 C71 PINCH VALVE3-
100pF 1K GNDA PINCH VALVE3-
SYRINGE_VALVE_S 1 6 R162
SYRINGE_VALVE_S DIAG
3 R68 5 C229
GND GND GND GND 1 GND
SYRINGE_VALVE- 100pF
1K C271 VN330SP
TLV2374ID NC
+ C78 220nF D13 C246
GND GND
4
1uFCTT16V MOT2 100pF
VDDA MOT2
MOT- SOL+
10K GNDA SOL+
R137 U32 SOL-
C270 SOL-
20 1 SMOT R163
U22D 22nF VREF SENSE
19 2 C247
U38B BOOT2 ENABLE GND 0.33 GND
R138 9 8 18 100pF
IN2
6 C75 17 4 R164
100pF 1K GNDA GND GND 1
RESERVE_VALVE_S 7 74HC14SO 16 5 C250
RESERVE_VALVE_S GND GND GND
5 R43 15 6 100pF
GND GND GND GND
RESERVE_VALVE- 14 7 MOT+
1K GND GND MOT+
TLV2374ID MOT1 13
C55 MOT1 IN1
+ 12 9 GND
1uFCTT16V BOOT1 OUT2
11 10 MOT-
C269 OUT1 VS +26 MOT-
15K GNDA 22nF C64
R141 L6201
100nF GND
C248
U39B GND MOT+ 100pF
R142
6 C76
100pF 1K GNDA
PINCH VALVE3_S 7 +12 C49 NC
PINCH VALVE3_S D5
5 R72 C139
GND GND GND
PINCH VALVE3- 100pF
1K R42 100nF LP MIXING VALVE+
TLV2374ID LP MIXING VALVE+
+ C82 1K LP MIXING VALVE-
1uFCTT16V LP MIXING VALVE-
S
R101
G C140
10K GNDA Q2 GND 1 GND
R143 100pF
R41 IRFR9014
220
U39C HP PUMP+
R144
D
9 C77 NC
100pF 1K GNDA D2
C
SOL_S 8 C136
SOL_S GND GND
10 R67 R36 100pF
SOL- HP PUMP B HP PUMP+
1K HP PUMP 1K Q1 HP PUMP+
TLV2374ID HP PUMP-
C83 R37 BC847B HP PUMP-
+ R103
1uFCTT16V 10K C138
GND GND
E
0.01 100pF
GNDA
27K
R132 NC
GND D12 C161
GND GND
100pF
U39D SOLB1+
R133 SOLB1+
13 C73 SOLB1+ SOLB-
100pF 1K GNDA SOLB-
MOT_S 14 R120
MOT_S
12 R73 CI23 C162
GND 0.33 GND
SMOT R91 NC SOLB+ 100pF
10K
TLV2374ID
+ C84 100K
1uFCTT16V CI22
R97
NC RESERVE_VALVE+
15K U17D
GNDA R90 D7 D8
13 C41
100pF 1K GNDA
HEATER_S 14
U17B HEATER_S +12 GND
R98 12 R35
6 C43 PU11 SM6T15A SM6T15A
100pF 1K GNDA 1K
LP MIXING VALVE_S 7 TLV2374ID
LP MIXING VALVE_S C46
5 R95 + C157
1uFCTT16V GND
LP MIXING VALVE- 100pF
1K
TLV2374ID GNDA
+ C37 HEATER+ HEATER+
1uFCTT16V HEATER+
R99 HEATER- HEATER-
HEATER-
100K GNDA
+12 U19 C159
C124 GND
0 20 100pF
GND GND 22nF
1
U17C C122 GND D9
R100 GND
9 C45 100nF GND 4 17
100pF 1K GNDA OUT2 ENABLE +12 GND
HP PUMP_S 8 5 16 C123
HP PUMP_S +12 VS SENSE D11
10 R94 GND 6 15 SM6T15A
OUT1 VREF GND
HP PUMP- 7 14 SM6T15A
10K BOOT1 BOOT2
TLV2374ID 8 13 220nF
C42 IN1 IN2
+
1uFCTT16V C39 10 11 R96
GND GND 0.01
R44 22nF
GNDA
GND L6201PS GND 0.01 a la place de 0.022
10K U22B
3 4
HEATER1
GND
U21D U21B 74HC14SO
R38
13 6 C50
100pF 1K GNDA HEATER2
14 FAN_S 7
FAN_S
12 5 R54
GNDA
SOLB-
1K
TLV2374ID TLV2374ID
+ C65
1uFCTT16V
Title
GNDA INTERFA CE BOA RD
VDDA
LIQUID PUMP
4
U21A LIQUID PUMP
PINCH VALVE1
PINCH VALVE1
3 R102 C131 A16_MC 2 PINCH VALVE2
A16 SU1 PINCH VALVE2
1 MUX XFC_MC XTAL_MC A17_MC 30 12 A0_MC D_VACUUM_PUMP
100 GND GND A17 A0 D_VACUUM_PUMP
2 A18_MC 1 11 A1_MC SUP SYRINGE_VALVE
IO3CPLD
I1_2_TH
I0_2_TH
I3_2_TH
I2_2_TH
I2_1_TH
I3_1_TH
I0_1_TH
I1_1_TH
R31 X2 A18 A1 SYRINGE_VALVE
C132 100nF C34 10 A2_MC NC RESERVE_VALVE
I1_2_Z
I0_2_Z
I3_2_Z
I2_2_Z
I2_1_Z
I3_1_Z
I0_1_Z
CLK_MC
RST_MC
A2 RESERVE_VALVE
10M
22pF 16MHz L108E (IQD) D8_MC 13 9 A3_MC PINCH VALVE3
TXD
TLV2374ID
RTS
GND DQ0 A3 PINCH VALVE3
D9_MC 14 8 A4_MC SOL
DQ1 A4 SOL
11
100nF EXTAL_MC D10_MC 15 7 A5_MC SU23 MOT1
GNDA GND DQ2 A5 MOT1
C35 D11_MC 17 6 A6_MC SU P MOT2
DQ3 A6 MOT2
22pF D12_MC 18 5 A7_MC 822498-1
C47 DQ4 A7
D13_MC 19 27 A8_MC PH_1_X
100
DQ5 A8 PH_1_X
99
98
97
96
94
93
92
90
89
87
85
84
83
81
80
79
78
77
76
100pF D14_MC 20 26 A9_MC I0_1_X
DQ6 A9 I0_1_X
U34 U25 D15_MC 21 23 A10_MC U33 I1_1_X
DQ7 A10 I1_1_X
BUB1 13 3 25 A11_MC I2_1_X
BUB1 X0 X A11 I2_1_X
I
I
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
BUB2 14 X_POS_SENSOR 20 52 A0_MC CSROM_MC 22 4 A12_MC I3_1_X
BUB2 X1 DATA0/KWD0/PD0 ADDR0/ PB0 CE A12 I3_1_X
VREF_SYR 15 X_HOME_SENSOR 21 53 A1_MC OE_MC 24 28 A13_MC PH_2_X
VREF_SYR X2 DATA1/KWD1/PD1 ADDR1/ PB1 OE A13 PH_2_X
BUB3 12 RACK_POS_SENSOR 22 54 A2_MC RW_MC 31 29 A14_MC RES 1 75 I0_2_X
BUB3 X3 DATA2/KWD2/PD2 ADDR2/ PB2 WE A14 I/O I/O I0_2_X
SYRINGE_VALVE_S 1 SYR_HOME_SENSOR 23 55 A3_MC U23 3 A15_MC 2 I1_2_X
SYRINGE_VALVE_S X4 DATA3/KWD3/PD3 ADDR3/ PB3 A15 I/O I1_2_X
PINCH VALVE1_S 5 VDDA Z_POS_SENSOR 24 56 A4_MC MBM29F040C-90JC 73 TDO I2_2_X
PINCH VALVE1_S X5 DATA4/KWD4/PD4 ADDR4/ PB4 TDO I2_2_X
RESERVE_VALVE_S 2 16 Z_HOME_SENSOR 25 57 A5_MC TDI 4 72 I1_1_Z I3_2_X
RESERVE_VALVE_S X6 VDD DATA5/KWD5/PD5 PB5/ADDR5 TDI I/O I3_2_X
LIQUID PUMP_S 4 ADAPTOR_SENSOR 26 58 A6_MC D8_MC 11 10 A0_MC D8_MC 5 71 I0_1_X DISABLE_X
LIQUID PUMP_S X7 DATA6/KWD6/PD6 PB6/ADDR6 O0 A0 I/O I/O DISABLE_X
8 RACK_IN_SENSOR 27 59 A7_MC D9_MC 12 9 A1_MC D9_MC 6 70 I3_2_X PH_1_Z
VSS DATA7/KWD7/PD7 PB7/ADDR7 O1 A1 I/O I/O PH_1_Z
6 D10_MC 13 8 A2_MC D10_MC 7 69 I1_1_X I0_1_Z
INH O2 A2 I/O I/O I0_1_Z
VDDA 11 7 D8_MC 28 60 A8_MC D11_MC 15 7 A3_MC D11_MC 8 68 I0_2_X I1_1_Z
A VEE DATA8/PC0 PA0/ADDR8 O3 A3 I/O I/O I1_1_Z
10 D9_MC 29 61 A9_MC D12_MC 16 6 A4_MC D12_MC 9 67 I3_1_X I2_1_Z
B DATA9/ PC1 PA1/ADDR9 O4 A4 I/O I/O I2_1_Z
C166 9 GNDA D10_MC 30 62 A10_MC D13_MC 17 5 A5_MC D13_MC 10 I3_1_Z
C DATA10/ PC2 PA2/ADDR10 O5 A5 I/O I3_1_Z
D11_MC 31 63 A11_MC D14_MC 18 4 A6_MC 65 I2_1_X PH_2_Z
DATA11/ PC3 PA3/ADDR11 O6 A6 I/O PH_2_Z
HEF4051BM1 D12_MC 32 64 A12_MC D15_MC 19 3 A7_MC D14_MC 12 64 I1_2_X I0_2_Z
DATA12/ PC4 PA4/ADDR12 O7 A7 I/O I/O I0_2_Z
100nF IO_A2 D13_MC 33 65 A13_MC 25 A8_MC D15_MC 13 63 I2_2_X I1_2_Z
DATA13/ PC5 PA5/ADDR13 A8 I/O I/O I1_2_Z
C165 IO_A1 D14_MC 34 66 A14_MC U24 24 A9_MC TXD0_MC 14 62 TCK I2_2_Z
DATA14/ PC6 PA6/ADDR14 A9 I/O TCK I2_2_Z
IO_A0 D15_MC 35 67 A15_MC CY62256L-70SNC 21 A10_MC TMS 15 61 I3_1_SYR I3_2_Z
DATA15/ PC7 PA7/ADDR15 A10 TMS I/O I3_2_Z
CS_MUX1 23 A11_MC IO1CPLD 16 60 I2_1_SYR DISABLE_Z
A11 I/O I/O DISABLE_Z
100nF CS_MUX2 BKGD_MC 19 11 A16_MC 2 A12_MC A0_MC 17 PH_1_TH
BKGD / TAGHI ADDR16/PG0 A12 I/O PH_1_TH
C51 12 A17_MC 26 A13_MC 58 I2_2_SYR I0_1_TH
ADDR17/PG1 A13 I/O I0_1_TH
U35 RST_MC 40 13 A18_MC VCC 1 A14_MC A1_MC 19 57 I3_2_SYR I1_1_TH
RST ADDR18/PG2 A14 I/O I/O I1_1_TH
9 VCC 16 28 A2_MC 20 56 I0_2_SYR I2_1_TH
C ADDR19/PG3 C129 VCC I/O I/O I2_1_TH
100nF 10 GNDA XTAL_MC 47 17 CS_MUX1 20 CSRAM_MC RW_MC 21 55 I1_2_SYR I3_1_TH
B XTAL ADDR20/PG4 100nF CS I/O I/O I3_1_TH
11 7 EXTAL_MC 46 18 CS_MUX2 22 OE_MC CSPLD 22 54 I0_1_SYR PH_2_TH
A VEE R104 EXTAL ADDR21/PG5 OE I/O I/O PH_2_TH
GNDA 6 14 27 RW_MC IO2CPLD 23 53 I1_1_SYR I0_2_TH
INH 10K GND GND WE I/O I/O I0_2_TH
8 XFC_MC 44 RTS_HP 24 52 MOT2 I1_2_TH
VSS XFC I/O I/O I1_2_TH
SOL_S 4 87 MUX TXD0_HP 25 I2_2_TH
SOL_S X7 PAD0/ AN0 I/O I2_2_TH
PINCH VALVE3_S 2 16 PE0MC PE0MC 36 88 VAC_SIGNAL I3_2_TH
PINCH VALVE3_S X6 VDD XIRQ /PE0 PAD1/ AN1 I3_2_TH
MOT_S 5 PE1MC 37 89 AUX_DEG_VAC_SIG DISABLE_TH
MOT_S X5 IRQ /VPP/ PE1 PAD2/ AN2 DISABLE_TH
PINCH VALVE2_S 1 VDDA RW_MC 38 90 WASTE_VAC_SIGNAL VCC PH_1_HPV
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
PINCH VALVE2_S X4 R/ W/ PE2 PAD3/ AN3 PH_1_HPV
VREF_X 12 PE3MC 39 91 TH_POS_SENSOR I0_1_HPV
VREF_X X3 LSTRB /TAGLO/ PE3 PAD4/ AN4 I0_1_HPV
D_VACUUM_PUMP_S 15 CLK_MC 48 92 TH_HOME_SENSOR I1_1_HPV
D_VACUUM_PUMP_S X2 ECLK/ PE4 PAD5/ AN5 C167 C153 C150 C160 I1_1_HPV
GEN FAN_S 14 MODA_MC 49 93 TH_HIGHT_SENSOR I2_1_HPV
GEN FAN_S X1 MODA/ IPIPE0/ PE5 PAD6/ AN6 I2_1_HPV
27
28
29
30
31
32
33
35
36
37
40
41
42
44
SYRINGE_VALVE45
46
PINCH VALVE147
PINCH VALVE248
PINCH VALVE3 49
50
D_VACUUM_PUMP
FAN_S 13 3 MODB_MC 50 94 RACK_OUT_SENSOR 100nF 100nF 100nF 100nF EPM7128STC100 I3_1_HPV
RESERVE_VALVE
FAN_S X0 X MODB/ IPIPE1/ PE6 PAD7/ AN7 I3_1_HPV
DOOR_LOCK_SENSOR 51 PH_2_HPV
ARST/ PE7 PH_2_HPV
I2_2_HPV
I1_2_HPV
I0_2_HPV
I3_1_HPV
I2_1_HPV
I1_1_HPV
I0_1_HPV
I3_2_HPV
HEF4051BM1 SPI_CLK 74 CS_SPI2 GND I0_2_HPV
SOL
TXD_TC
SPI_CLK PF6/CSP1 I0_2_HPV
RTS_TC
LEDMC
VTXTC
SPI_OUT 73 CSROM_MC I1_2_HPV
MOT1
SPI_OUT PF5/CSP0 U22E I1_2_HPV
SPI_IN RXD 97 72 CSRAM_MC I2_2_HPV
SPI_IN PS0/ RxD0 PF4/CSD I2_2_HPV
U18 TXD0_MC 98 71 CSPLD RW_MC 11 10 OE_MC I3_2_HPV
PS1/ TxD0 PF3/CS3 I3_2_HPV
CSPOT_X 15 1 IO_A0 70 IO_A2 DISABLE_HPV
CSPOT_X Y0 A PF2/CS2 DISABLE_HPV
CSPBUB1 14 2 IO_A1 RXD1_MC 99 69 IO_A1 74HC14SO PH_1_SYR
CSPBUB1 Y1 B PS2/ RxD1 PF1/CS1 PH_1_SYR
CSPBUB2 13 3 IO_A2 TXD1_MC 100 68 IO_A0 I0_1_SYR
CSPBUB2 Y2 C PS3/ TxD1 PF0/CS0 I0_1_SYR
CSPBUB3 12 B4 I1_1_SYR
CSPBUB3 Y3 I1_1_SYR
CSPOT_SYR 11 VCC SS_MC 104 75 DISABLE_X TCK I2_1_SYR
CSPOT_SYR Y4 PS7/ SS PH0/KWH0 1 2 GND J I2_1_SYR
CSHPPC 10 6 SPI_CLK 103 76 DISABLE_Z TDO I3_1_SYR
CSHPPC Y5 G1 PS6/ SCK PH1/KWH1 U22C DE7 3 4 VCC I3_1_SYR
CSTC 9 4 CS_SPI1 SPI_OUT 102 77 DISABLE_TH R40 VCC TMS PH_2_SYR
CSTC Y6 G2A PS5/ SDO/ MOSI PH2/KWH2 5 6 PH_2_SYR
CSRACK 7 5 SS_MC SPI_IN 101 78 DISABLE_HPV LEDMC 5 6 I0_2_SYR
CSRACK Y7 G2B PS4/ SDI/ MISO PH3/KWH3 470 7 8 I0_2_SYR
81 DISABLE_SYR TDI I1_2_SYR
PH4/KWH4 9 10 I1_2_SYR
C120 74HC138SO 82 PH_1_X 74HC14SO CL150UR3000 I2_2_SYR
VCC GND R47 PH5/KWH5 I2_2_SYR
22nF 85 83 PH_2_X CAV10 GND I3_2_SYR
10K VRef V RH PH6/KWH6 I3_2_SYR
U46 86 84 BUB_DET_3 DISABLE_SYR
GNDA V RL PH7/KWH7 DISABLE_SYR
CS5 15 1 IO_A0
CS5 Y0 A
DSYNC 14 2 IO_A1 42 105 BUB_DET_1 R115
DSYNC Y1 B VCC VDDX1 PT0/ IOC0
CS3 13 3 IO_A2 41 106 BUB_DET_2 VTXTC
CS3 Y2 C GND VSSX1 PT1/ IOC1 10K VCC
CS2 12 VCC 107 CS_SPI1
CS2 Y3 PT2/ IOC2
CS1 11 VCC 79 108 LIQUID PUMP JP4
CS1 Y4 VCC VDDX2 PT3/ IOC3
CS0 10 6 80 109 SENSOR 0 CAV2 C85
CS0 Y5 G1 GND VSSX2 PT4/ IOC4
CSCANHP 9 4 CS_SPI2 110 SENSOR 1 XTAL_MC
CSCANHP Y6 G2A PT5/ IOC5 VCC GND XTAL_MC
7 5 SS_MC 2 111 SENSOR 2 RES
Y7 G2B VCC VDDX3 PT6/ IOC6 RES
1 112 FR D LOCK SENSOR GND
GND VSSX3 PT7/ IOC7/ PAI 100nF
C211 74HC138SO VRef
VCC GND
22nF 14 3 PH_1_Z U47
VCC VDD KWJ0/PJ0
15 4 PH_2_Z 18
PE0MC C125 GND VSS KWJ1/PJ1
5 PH_1_TH
VCC ON-OFF
PE0MC 100nF KWJ2/PJ2
PE1MC 43 6 PH_2_TH RXD1_MC 10 9 BARCODE TXD
PE1MC GNDA VCC VDDPLL KWJ3/PJ3 R2OUT R2IN BARCODE TXD
PE3MC 45 7 PH_1_HPV 13 14
PE3MC GND VSSPLL KWJ4/PJ4 R1OUT R1IN GND
IO1CPLD VDDA 8 PH_2_HPV
IO1CPLD KWJ5/PJ5
95 9 PH_1_SYR TXD1_MC 11 8 BARCODE RXD
VDDA KWJ6/PJ6 T2IN T2OUT BARCODE RXD
96 10 PH_2_SYR TXD_TC 12 15 TXDTCRS
GNDA VSSA KWJ7/PJ7 T1IN T1OUT TXDTCRS
+
TXD0_HP C130 C215 6 C216
TXD0_HP C33 100nF C2-
RTS_HP 7
RTS_HP 1uFCTT16V 100nF -9VOUT 100nF GND
68HC812A4PV8 5
C2+
TXD_TC GNDA GNDA 4 C214
TXD_TC VCC C1- 100nF
RTS_TC C213 3
RTS_TC C143 C133 C119 +9VOUT GND
2
100nF 100nF 100nF C1+
RXD
RXD GND 100nF ST232CW
VCC
IO2CPLD
IO2CPLD
IO3CPLD
IO3CPLD
BUB_DET_1
BUB_DET_1
BUB_DET_2
BUB_DET_2
BUB_DET_3
BUB_DET_3
Title
INTERFA CE BOA RD
TACHY B
HP PUMP
RES
CLK HP
R77 C103
XFC_HP
100 GND
IT2
GND
C104 100nF
100
GND
99
98
97
96
94
93
92
90
89
87
85
84
83
81
80
79
78
77
76
100nF U9
I
I
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
U12
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
I/O
R82 XIRQ /PE0 PAD1/ AN1
IRQHP 37 89 LP MIXING VALVE_S
IRQ /VPP/ PE1 PAD2/ AN2
RW_HP 38 90 HP PUMP_S
R/ W/ PE2 PAD3/ AN3
39 91 C137
LSTRB /TAGLO/ PE3 PAD4/ AN4
27
28
29
30
31
32
33
35
36
37
40
41
42
44
45
46
47
48
49
50
CLK HP 48 92 SU2 100nF EPM7128STC100
ECLK/ PE4 PAD5/ AN5 SU7
MODA_HP 49 93 SU P
MODA/ IPIPE0/ PE5 PAD6/ AN6 VCC GND
D15_HP
D14_HP
D13_HP
D12_HP
D10_HP
D11_HP
TACHY
MODB_HP 50 94 NC SU P
A3_HP
A2_HP
A1_HP
A0_HP
D8_HP
D9_HP
MODB/ IPIPE1/ PE6 PAD7/ AN7
51 822498-1
ARST/ PE7
XTAL_MC 74
XTAL_MC PF6/CSP1
RES GND 73 CSROM_HP A16_HP 2
RES PF5/CSP0 A16
RXD 97 72 CSRAM_HP A17_HP 30 12 A0_HP
PS0/ RxD0 PF4/CSD A17 A0
CSHPPC TXD0_HP 98 71 CSPLDHP A18_HP 1 11 A1_HP
CSHPPC PS1/ TxD0 PF3/CS3 A18 A1
SPI_CLK 70 10 A2_HP
SPI_CLK PF2/CS2 A2
SPI_IN 99 69 D8_HP 13 9 A3_HP
SPI_IN PS2/ RxD1 PF1/CS1 DQ0 A3
SPI_OUT 100 68 D9_HP 14 8 A4_HP B2
SPI_OUT PS3/ TxD1 PF0/CS0 DQ1 A4
D10_HP 15 7 A5_HP TCKHP
DQ2 A5 1 2 GND JTAG
CSHPPC 104 75 D11_HP 17 6 A6_HP TDOHP
PS7/ SS PH0/KWH0 DQ3 A6 3 4 VCC
SPI_CLK 103 76 D12_HP 18 5 A7_HP TMSHP
PS6/ SCK PH1/KWH1 DQ4 A7 5 6
SPI_OUT 102 77 D13_HP 19 27 A8_HP
PS5/ SDO/ MOSI PH2/KWH2 DQ5 A8 7 8
SPI_IN 101 78 D14_HP 20 26 A9_HP TDIHP
PS4/ SDI/ MISO PH3/KWH3 DQ6 A9 9 10
81 D15_HP 21 23 A10_HP
PH4/KWH4 DQ7 A10
CAM FLAG 82 25 A11_HP CAV10 GND
CAM FLAG PH5/KWH5 A11
TACHY A 85 83 CSROM_HP 22 4 A12_HP
TACHY A VRef V RH PH6/KWH6 CE A12
TACHY B 86 84 OE_HP 24 28 A13_HP
TACHY B GNDA V RL PH7/KWH7 OE A13
RW_HP 31 29 A14_HP HP PUMP
WE A14 HP PUMP
42 105 CAM FLAG U7 3 A15_HP LP MIXING VALVE1
VCC VDDX1 PT0/ IOC0 A15 LP MIXING VALVE1
41 106 TACHY A MBM29F040C-90JC LP MIXING VALVE2
GND VSSX1 PT1/ IOC1 LP MIXING VALVE2
107 TACHY B
PT2/ IOC2
HP_SENSOR 79 108 HP PUMPb D8_HP 11 10 A0_HP
HP_SENSOR VCC VDDX2 PT3/ IOC3 O0 A0 U22F DE6
LP MIXING VALVE_S 80 109 IT1 D9_HP 12 9 A1_HP R18 VCC
LP MIXING VALVE_S GND VSSX2 PT4/ IOC4 O1 A1
HP PUMP_S 110 TACHYAXORB D10_HP 13 8 A2_HP LEDHPPC 13 12
HP PUMP_S PT5/ IOC5 O2 A2 470
2 111 IT2 D11_HP 15 7 A3_HP
VCC VDDX3 PT6/ IOC6 O3 A3
PE3MC 1 112 TACHY D12_HP 16 6 A4_HP 74HC14SO CL150UR3000
PE3MC GND VSSX3 PT7/ IOC7/ PAI O4 A4
VRef D13_HP 17 5 A5_HP R74
O5 A5
14 3 PE3MC D14_HP 18 4 A6_HP MODA_HP
VCC VDD KWJ0/PJ0 O6 A6 10K VCC
15 4 DIR D15_HP 19 3 A7_HP R75
C96 GND VSS KWJ1/PJ1 O7 A7
5 LEDHPPC 25 A8_HP MODB_HP
100nF KWJ2/PJ2 A8 100K
43 6 U8 24 A9_HP R76
VCC VDDPLL KWJ3/PJ3 A9
45 7 CY62256L-70SNC 21 A10_HP
GND VSSPLL KWJ4/PJ4 A10 4K7 GND
GNDA 8 23 A11_HP R83
KWJ5/PJ5 A11
VDDA 95 9 2 A12_HP
VDDA VDDA KWJ6/PJ6 A12 10K VCC
96 10 26 A13_HP
GNDA VSSA KWJ7/PJ7 A13
VCC 1 A14_HP B1
A14
+ BKGD_HP
C16 C100 1 BDM
20 CSRAM_HP
C13 100nF 100nF CS GND 2
68HC812A4PV8 22 OE_HP
1uFCTT16V OE 3
27 RW_HP RSTHP
WE 4
GNDA GNDA
VCC 5
GND
C116 C117 C105 C97 U22A VCC 6
100nF 100nF 100nF 100nF RW_HP 1 2 OE_HP CAV6
GND
74HC14SO
Title
INTERFA CE BOARD
L3 VDDA
LQH3C100K34 U3 PT12
L78M05CDT VDDA NC
1 3
+26 IN OUT
GND
L1 +
LQH3C100K34 +
C24 C21
XTAL_MC C20 10uFCTT35V 100nF C25
2
JP3 100nF 10uFCTT35V
NC EXTAL_TC
GNDA
GND VRef
GND L2 U2 PT13
C148 X3 LQH3C100K34 LT1019CS8-5
NC NC
NC 2 6
+26 IN OUT VRef
GND
XTAL_TC
GND C7
C149 + + 560 pour INA122UA
NC 10uFCTT35V C15 C14
R21
100nF 100nF
4
B5 150 1%
NC
CLR GNDA C6 R25 R84
1 10uFCTT16V 10K 0.1% 113 0.1%
GND U10
2 VCC
8 1
3 GND RG+ RG-
ICSPData R22 7 2 GND
4 VDDA VCC IN-
ICSPCLK 6 3
5 100K OUT IN+ Vref
5 4
2V5 REF VEE C115
C27
2.2uF MAX4194ESA GNDA R80 100pF PT100HEAT2
10K 0.1% PT100HEAT2
C152
GNDA C38
VCC GND
PT100HEAT1
VDDA GNDA PT100HEAT1
100nF
100nF C113
U30 U11 100pF
R27
EXTAL_TC 9 21 RTS_TC 560 pour INA122UA
OSC1 RB0 150 1%
22 CSCAN GND
RB1
23 CKCAN CSCAN 1 2 E1
RB2 CS CH0 R32 R87
24 OUTCAN CKCAN 7 3 E2
RB3 CLK CH1 10K 0.1% 113 0.1%
XTAL_TC 10 25 INCAN U14
OSC2 RB4
26 RB5 OUTCAN 6 8 8 1
RB5 DOUT VCC Vref RG+ RG-
FAN_S 2 27 ICSPCLK INCAN 5 4 R26 7 2 GND
RA0 RB6 DIN GND GNDA VDDA VCC IN-
HEATER_S 3 28 ICSPData 6 3
RA1 RB7 100K OUT IN+ Vref
XTAL_MC LEDTC 4 5 4
XTAL_MC RA2 2V5 REF VEE C121
RES RA3 5 11 HEATER1 C30
RES RA3 RC0 GND R86 100pF
RA4 6 12 HEATER2 LTC1288CS8 2.2uF MAX4194ESA GNDA PT100EXT2
RA4 RC1 10K 0.1% PT100EXT2
CSTC 7 13 FAN C28
RA5 RC2
CSTC 14 SPI_CLK GNDA C31
CSTC RC3 GNDA VRef
SPI_CLK CLR 1 15 SPI_OUT PT100EXT1
SPI_CLK MCLR RC4 VDDA GNDA PT100EXT1
SPI_IN 16 SPI_IN 100nF
SPI_IN VCC RC5 C118
SPI_OUT 17 TXD_TC 100nF
SPI_OUT D10 RC6 100pF
18 RXD
BAS21LT1 RC7
RXD
RXD
TXD_TC R53 16F876-20SP GND
TXD_TC
RTS_TC RES
RTS_TC 1K SU36
SU P
110-97-328-41-001
FAN_S U36
FAN_S
HEATER_S EXTAL_TC 9 21 RTS_TC
HEATER_S OSC1 RB0
22 CSCAN HEATER1
RB1 HEATER1
23 CKCAN HEATER2
RB2 HEATER2
24 OUTCAN
RB3
XTAL_TC 10 25 INCAN FAN VRef
OSC2 RB4 FAN
26 RB5
RB5
FAN_S 2 27 ICSPCLK 2V5
RA0 RB6 R13
HEATER_S 3 28 ICSPData
RA1 RB7 10K 1%
LEDTC 4
RA2
RA3 5 11 HEATER1 2V5
RA3 RC0
RA4 6 12 HEATER2 13
RA4 RC1 PT14
CSTC 7 13 FAN 14
RA5 RC2
14 SPI_CLK 12 NC
RC3 U5D
CLR 1 15 SPI_OUT
MCLR RC4
16 SPI_IN TLV2374ID
RC5 R16
17 TXD_TC
RC6 C176 10K 1%
18 RXD
RC7 100nF
NC
GNDA
DE8 VCC 6
R57
LEDTC 7
470 5
GNDA U5B
CL150UR3000
TLV2374ID
9
8
10
GNDA U5C Title
INTERFACE BOA RD
TLV2374ID
Size Document Number Rev
A3 DAE010E+3 3
VDDA R108
6
7 BUB1
BUB1 R110 10M
5
BUB_DET1 U48B NC VDDA
4
TLV2374ID
R49 C60 U49A
4K7 100nF R107
BUB1 3
51K
1 BUB_DET_1
BUB_DET_1
2
GNDA
GNDA TLV2374ID
11
R109
11
NC
2
1 BUB2 GNDA
BUB2
3
BUB_DET2 U48A U27 VDDA GNDA
TLV2374ID 1 8
R48 C58 GNDA L H
7
W
4
4K7 100nF CSPBUB1 3 6
VDDA CS VDD VCC
SPI_OUT 4 5
DIN SCLK
MAX5400EKA-T C57
GNDA SPI_CLK 100nF
GND
13
14 BUB3 VDDA R112
BUB3
12
BUB_DET3 U48D
R111 10M
TLV2374ID
R50 C62 NC
4K7 100nF
U49B
R106
BUB2 5
51K
7 BUB_DET_2
BUB_DET_2
GNDA 6
9
8 TLV2374ID
10 R113
GNDA U48C NC
TLV2374ID
R118 10M
NC
U49C
R116
BUB3 10
51K
8 BUB_DET_3
BUB_DET_3
9
TLV2374ID
R117
NC
U49D
12
GNDA
14
13
TLV2374ID
Title
INTERFACE BOA RD
U1 2SS52M
CN1
1 3
GND VCC 1
2
OUT
3
JST B3B PH K S
C1
2
100nF
ADIL INSTRUMENTS
Title
ADAPTATOR SENSOR BOARD
CN1
2
D1 D2 D3
1 LED S 1 227-66BR S 1 227-33BR
2
1
A GND K
ADIL INSTRUMENTS
Title
PHOTODETECTOR A N LED BOARD