Professional Documents
Culture Documents
Data Sheet
DC Transducer /Isolating Amplifier
Product Features
Regulations
Electric Isolation
Electrically isolated analog output prevents interference voltage Electromagnetic Compatibility Acc. to IEC 61326 - 1
and current. Solves grounding problem in meshed signal networks. Protection For Housing : IP40
High electric isolation between input and output 3.2 kV, and Terminals : IP20
power supply versus all other circuits 5.2 kV. Electrical standards Acc. to IEC 61010 -1 /
EN 61 010 -1
Pollution degree 2
Function Over voltage category III for power supply.
Simple dc isolator serves to electrically isolate input dc signal in the II for measuring input and
range 0 – 20 mA or 4-20 mA or 0-10V or 2-10V is then converted to measuring output.
signal 0 – 20 mA or 4-20 mA or 0-10V or 2-10V finds its application
for isolate input and output. Double Insulation - Power supply versus all other
circuit.
- Measuring input versus
Technical Specifications measuring output.
Measuring inputs : Test Voltage Power supply versus :
DC current standard ranges 1) 0...20mA -All 5.2 kV, DC 1 min
2) 4...20mA Measuring inputs versus :
3) 1...5mA -Measuring output 3.2 kV, DC
1min
DC voltage standard ranges 1) 0...10V
Ambient Temperature
2) 2...10V
Climatic rating Climate case 3Z acc. to
3) 1...5V
VDI /VDE 3540
4) 0...300V
Nominal Range of use 0 ºC to 45 ºC (Usage Group II)
Measuring outputs :
Storage temperature -40 ºC to 70 ºC
DC current standard ranges 1) 0...20mA
2) 4...20mA Annual mean relative humidity < 75% standard Climatic rating.
Burden voltage 12V Installation Data
Mechanical Housing Lexan 940 (polycarbonate)
External Resistance Rext max. [k Ohm] = 12V/ IAN [mA] Flammability Class V-0 acc.
I AN =Output circuit full scale to UL 94 self extinguishing,
value non dripping, free of halogen.
DC voltage standard ranges 1) 0...10V Mounting position Rail mounting / wall mounting
2) 2...10V Weight Approx. 0.15kg
Burden Rext min. [kOhm] = UAN [V]/ 5 mA Connection Terminal
UAN =Output circuit full scale Connection Element Conventional Screw type
value terminal with indirect wire
pressure
Current limiter at Rext =0 < 30mA for voltage output Permissible cross section
2
of the connection lead 4.0 mm single wire or 2 x 2.5mm
Voltage limiter at Rext =∞. < 17V for current output
Fine wire.
Residual ripple in Output current < 0.4% p.p. Permissible Vibrations 2 G acc. to EN 60 068-2-6
Shocks 3 x 50 g 2 shocks each in 6
Response time < 50 ms
directions
Common mode voltage 100V Acc. to EN 60 068-2-27
Pollution degree 2 Output characteristics
3 4 0…..80 15 0…....40 28
0….100 16 0.2…...1 29
0.2….1 17 1…......5 30
1…....5 18 4…....20 31
2…..10 19 0…....48 32
0...20 1 0...10 3
4...20 2 2...10 4
5 6
1 2
Input + - Input 4)
~
+ ~
- Auxiliary supply Ordering Code
60 …300V AC/DC H
Auxiliary Supply ~, + 5
~, - 6