You are on page 1of 9

Review Article

Advances in Microscopy Techniques


Daniel B. Schmolze, BS; Clive Standley, PhD; Kevin E. Fogarty, MS; Andrew H. Fischer, MD

NofContext.—Advances in microscopy enable visualization


a broad range of new morphologic features.
long been believed to be a limit to the resolution of light
microscopy. Fluorescence resonance energy transfer
Objective.—To review and illustrate advances in mi- (FRET) appears capable of visualizing diagnostically rele-
croscopy with relevance to pathologists. vant molecular events in living or fixed cells that are
Data Sources.—Literature review and new observations. immeasurable by other molecular techniques. We describe
Results.—Fluorescence microscopy enables multiantigen applications of 2-photon microscopy, FRET, structured
detection; allows novel optical-sectioning techniques, with illumination, and the subdiffraction techniques of near-
some advantages compared to paraffin sectioning; and field microscopy, photoactivated localization microscopy,
permits live-cell imaging. Live-cell imaging allows pathol- stochastic optical reconstruction microscopy, and stimu-
ogists to move from a period when all diagnostic expertise lated emission depletion microscopy.
was reliant on interpreting static images to a period when Conclusion.—New microscopy techniques present oppor-
cellular dynamics can play a role in diagnosis. New tunities for pathologists to develop improved diagnostic tests.
techniques have bypassed by about 100-fold what had (Arch Pathol Lab Med. 2011;135:255–263)

FLUORESCENCE MICROSCOPY detected; at the same time, a pseudo-H&E image of the


In traditional transmitted light microscopy, it is the same sample can be displayed by using 2 additional colors
removal of light from the image that provides information. in the spectrum. In a possible application, fine-needle
There is only so much light that can be removed to provide aspirations of lymph nodes could be immunofluores-
more information. In fluorescence microscopy, a virtually cently stained with multiple antibodies, to provide the
limitless addition of light to the image provides informa- information afforded by flow cytometry, but could still be
tion. The spectra of the light from fluorescent microscopy viewed digitally with standard, light microscope–like
can be separated (reviewed in Levenson et al1), giving morphology. Several commercial platforms exist for these
fluorescence microscopy many ‘‘dimensions’’ for convey- manipulations, as well as for extracting spectral informa-
ing information. It is a straightforward process to digitally tion and classifying objects (reviewed in Levenson et al1).
invert a fluorescent image (to have a white background) LIVE-CELL IMAGING
and to ‘‘pseudocolor’’ fluorescent nucleic acid and protein
stains (Figure 1, A and B) to closely resemble a standard Transmitted light microscopy requires use of dyes to
hematoxylin-eosin (H&E) section (Figure 1, C). visualize nuclear detail; however, no dyes are known to be
A possible application for this approach would be to suitable for live-cell imaging with transmitted light
perform multiple simultaneous immunofluorescent label- microscopy. Phase contrast and differential interference
ings on 1 section, while displaying a digital H&E image. contrast microscopy techniques cannot display subcellu-
With conventional light microscopy, a practical limit of lar morphology, particularly within thick tissue samples
about 3 to 4 colors can be usefully resolved, including any (Figure 2, B). Fluorescent dyes capable of penetrating
needed counterstain (eg, hematoxylin). Thus, with con- viable cells are available for staining nuclei or particular
ventional transmitted microscopy, 2 or possibly 3 color organelles, as are a wide variety of fluorescently labeled
immunohistochemical labels could be used, with only proteins that can be expressed in living cells through use
hematoxylin as a counterstain. However, colocalized of viral vectors (Figure 2, A and C). The fluorescently
antigens are difficult to separate in transmitted light stained tissue can subsequently be fixed and convention-
microscopy. On the other hand, if immunofluorescence is ally stained (Figure 2, D and E).
used, at least 3 colocalized antigens can be separately All of a pathologist’s criteria for cancer diagnosis are
still based on static images. If experience at other levels of
biology is a guide, it will be much easier to learn about the
Accepted for publication April 13, 2010.
From the University of Massachusetts Medical School, Worcester (Mr
abnormal physiology of pathologic tissue if dynamic
Schmolze); and the Departments of Physiology (Dr Standley and Mr images are studied. The dynamics of cancer cells will most
Fogarty) and Pathology (Dr Fischer), University of Massachusetts likely provide a fundamentally new set of diagnostic
Medical School, Worcester. features. The nuclear lamina irregularity, diagnostic of
The authors have no relevant financial interest in the products or papillary thyroid carcinoma, appears to be due to a
companies described in this article.
dynamic deformation of the nuclear lamina during
Reprints: Andrew H. Fischer, MD, Department of Pathology, University
of Massachusetts Medical School, Biotech Three, 1 Innovation Dr, interphase,2 and this dynamic property appears funda-
Worcester, MA 01605 (e-mail: andrew.fischer@umassmemorial.org). mentally different from the irregularity that develops
Arch Pathol Lab Med—Vol 135, February 2011 Advances in Microscopy Techniques—Schmolze et al 255
Figure 1. Fluorescence microscopy can provide the same information as conventional hematoxylin-eosin (H&E) microscopy. A conventional
paraffin section of an ampullary villous adenoma was deparaffinized and rehydrated to 50% alcohol, then simultaneously stained with 49,6-
diamidino-2-phenylindole (DAPI) (A) and eosin (B) without washing out any excess dye. The image was inverted and ‘‘pseudocolored’’ to match the
colors of an H&E section by using Photoshop (C). We have found that [Ru(bpy)3]Cl2 can be used like eosin and has a useful narrower emission
spectrum (original magnification 3100).

during a period of days in neutrophils, or possibly other predicted based on static images, and they point toward
benign cells that show nuclear irregularity in static limitless future diagnostic applications of the dynamic
images. Live-cell imaging in animal models has demon- properties of tumor cells.
strated a direct correlation between tumor cell migration
and seeding of the bloodstream.3 Further, dynamic studies THE DIFFRACTION LIMIT TO RESOLUTION IN
have shown an interaction between tumor cells and CONVENTIONAL MICROSCOPY
macrophages in the process of intravasation of tumor It has been known since the 19th century that the wave
cells into vessels.3 These interactions could not be directly nature of light imposes limits on optical resolution. The

Figure 2. Transmitted light microscopy of thick tissues does not permit visualization of cell structure, whereas fluorescent proteins expressed in
living cells are easy to resolve. Living normal human thyroid tissue was visualized after being in organ culture for 18 hours (A–C) or after being fixed
at 36 hours (D and E). Green fluorescent protein (GFP) bearing replication-incompetent adenovirus was added at the initiation of the organ culture.
The GFP signal is shown in (A). The GFP in this example is unconjugated and therefore diffuses throughout infected cells. Green fluorescent protein
fused to particular proteins, for example, lamins, permits visualization of subcellular detail (not shown). Barely discernable thyroid follicles are
evident by differential interference contrast microscopy in B. C shows the overlay of A and B. Histologically and cytologically, the tissue appears
normal after 36 hours (D) and GFP can be detected by immunohistochemistry in the paraffin-embedded tissue section (E) (original
magnifications 3200).

256 Arch Pathol Lab Med—Vol 135, February 2011 Advances in Microscopy Techniques—Schmolze et al
Figure 3. Airy disk and point spread func-
tion. Because of diffraction, a microscope
produces a blurry Airy disk (A) as the image of
a point source of light. The surface plot of this
disk defines the point spread function (B),
whose shape in turn determines resolution.

effect occurs as light waves cause interference on their defined by a different PSF, x 5 l/(n sin2a), and the
way through the optical system and is known as the practical Abbe limit is about 500 nm. Until the 1990s, the
diffraction limit. The practical consequence of this effect is Abbe limit was widely thought of as an absolute barrier to
that a point source of light will produce a fuzzy circle resolution by light microscopy.
when viewed through a microscope. This characteristic
pattern is known as an Airy disk (Figure 3, A) and consists DIFFRACTION-LIMITED MICROSCOPY TECHNIQUES
of a central bright dot surrounded by increasingly dimmer Before discussing newer optical methods that achieve
circles of light. The corresponding intensity plot is called molecular-level resolution, it is worth looking at some
the point spread function, or PSF (Figure 3, B). useful techniques that remain bound by the Abbe limit.
The shape of the PSF determines resolution. If the PSF is
very narrow or the separation between the points is wide, Confocal Microscopy
then 2 PSFs will add up in such a way that the peaks can In standard epifluorescence microscopy, used in most
still have a trough and the 2 images can be resolved. Ernst laboratories for immunofluorescence of renal and skin
Abbe addressed the physics and mathematics of the PSF biopsies, or fluorescence in situ hybridization of individ-
in the 19th century4 and defined an equation that yields ual chromosome probes, no mechanism exists to restrict
the smallest distance x over which 2 objects can be out-of-focus light from being imaged. At high magnifica-
resolved: tions or with thick sections, blurry light from outside the
focal plane contributes to the image, significantly reducing
x~l=2n sin a, ð1Þ contrast and perceived resolution. This effect occurs in a
where l is the wavelength of light, n is the refractive index different, less obtrusive manner in transmitted light
of the medium and a is half the angular aperture of the microscopy. A laser scanning confocal microscope focuses
lens. As shown in Figure 4, the angular aperture deter- exciting light onto a small area of the specimen, and it
mines how much of the wave front can be captured by the images the resulting emitted light through a pinhole that
lens. A material of high refractive index (for example only allows light from the focal plane to pass through
immersion oil, see Figure 4) will bend more light into the (Figure 5). As a result, an ‘‘optical section’’ of the
field of view of the objective lens, acting like an increase in specimen at the focal plane is obtained by scanning the
a to improve resolution. The Abbe equation shows that focused spot across the specimen. By changing the focal
resolution increases as the wavelength decreases. The plane, such sections can be imaged at different depths in
higher resolution of an electron microscope (less than the specimen, and a 3-D representation can be recon-
1 nm) relates to the very short wavelength of electrons. structed.
Compared to new microscopy techniques (described Confocal microscopy was first described in the 1950s by
below), whose resolution is starting to approach that of Minsky5 and is now a relatively mature technology with
electron microscopy, the technique of electron microscopy many commercial implementations. The price and com-
does not easily yield 3-dimensional information; more- plexity have steadily decreased, and recently a ‘‘bench-
over, immunolocalization by electron microscopy is top’’ model has been released.6 Most designs use a set of
difficult and live-cell imaging for structural dynamics is mirrors to scan a laser across the specimen. Although
impossible. suitable for some live-cell imaging applications, an
The denominator of equation (1) is known as the alternative ‘‘spinning disk’’ design collects the image
numerical aperture (NA) of the lens, so the equation can faster, allowing for multiple images per second. The
be rewritten as: spinning disk design uses a set of concentrically arranged
pinholes in a spinning disk, effectively allowing multiple
x~l=2NA ð2Þ confocal scans to take place simultaneously. Unfortunate-
ly, a relatively powerful illumination source is usually
With high refractive index media and high NA lenses, needed in all confocal designs, since the pinhole passes
the Abbe limit is roughly 200 nm (0.2 mm) in the X-Y plane. only a small amount of light, and this can be damaging to
For objects above and below the plane, the resolution is live cells.
Arch Pathol Lab Med—Vol 135, February 2011 Advances in Microscopy Techniques—Schmolze et al 257
Figure 4. Concept of numerical aperture.
Both a large value of a (half the angular
aperture of the lens) and a high refractive
index (n) will serve to increase the amount of
light from the specimen that is captured by
the lens.

Two-Photon Microscopy
In the usual case of fluorescence emission, a fluorophore
that absorbs a single photon of light at a certain
wavelength moves to a higher energy state. When the
fluorophore returns to its ground state, it emits a photon
with a longer wavelength than that of the photon
absorbed. Instead of exciting the fluorophore with 1
photon, 2 photons with half the energy (twice the
wavelength) can achieve the same excitation of the
fluorophore.
A 2-photon microscope operates by using this princi-
ple.7 It uses a pulsed infrared laser that is focused to a spot
in the specimen. This laser requires a high instantaneous
intensity at each pulse, since the absorption of 2 photons
simultaneously is a rare event, but the average intensity is
low. The illumination intensity falls off rapidly above and
below the focal point, and since the chance for absorption
of 2 photons increases with the square of the intensity, the
probability of absorption falls steeply (roughly to the sixth
power) above and below the focal plane. Thus, only
fluorophores in a thin volume are excited. When this spot
is scanned across the specimen, only light from this thin
excitation plane is collected and an optical section is
produced (Figure 6).
Whereas, in a confocal microscope, fluorophores out-
side the focal plane are excited but prevented from being
imaged by a pinhole, in a 2-photon microscope, absorp-
tion is confined to a small volume in the focal plane. The
absorption plane can be as thin as about half a micron.
Outside this plane, the sample is exposed only to a
relatively mild infrared beam with a low average
intensity. Thus, photobleaching and phototoxicity is
restricted to the plane of focus, and the sample as a whole
Figure 5. How a confocal microscope captures emitted light from
is exposed to much less potentially damaging radiation,
only 1 plane. The dichroic mirror serves to reflect the exciting light making the technique well suited to live-cell imaging.
while allowing the emitted fluorescence to pass through. The pinhole A second advantage of 2-photon imaging is that the
blocks light from outside the focal plane. infrared excitation source penetrates more deeply into
258 Arch Pathol Lab Med—Vol 135, February 2011 Advances in Microscopy Techniques—Schmolze et al
Figure 6. Two-photon microscopy. A pinhole is not needed, since the pulsed laser achieves the necessary high intensity to simultaneously excite
fluorophores with 2 photons at 1 plane.

tissue than the visible or ultraviolet light needed for advantage is that the plane of imaging can be precisely
confocal or conventional epifluorescence microscopy. chosen without wasting tissue. Applications would
Although the emitted light is still scattered by tissue, it appear to be particularly useful for small biopsies in
appears feasible to image to a depth of about 200 mm into which a tangential optical section on the various sides of
the tissue and still display diagnostic cellular-level the biopsy specimen could allow diagnosis. Diagnostic
morphology with the 2-photon technique. Confocal features for many common problems (eg, determining the
optical sectioning is reasonably crisp only to a depth of presence of invasion in breast biopsy specimens) typically
about 100 mm in whole tissue. span a diameter of only 200 mm.11
The concept of 2-photon microscopy was first described
in 1931.8 Costs and complexity have significantly declined 4Pi Microscopy
in recent years, but microscopes remain expensive (largely The Abbe limit refers to resolution in the X-Y plane.
owing to the cost of the pulsed laser). Nevertheless, 2- Resolution in the Z plane is much lower for all microscope
photon imaging is being explored for various clinical designs previously described. Confocal and 2-photon
applications, for example in phototherapy for cancer and microscopy are limited to about 500 nm of resolution in
macular degeneration.9,10 the Z axis. The Abbe equation shows that an objective lens
Two-photon imaging can allow optical sections to be with a large angular aperture will produce a higher-
cut, analogously to a paraffin section, on unfixed (living) resolution image than one with a smaller angular
or fixed unembedded tissues (Figure 7). Fluorescent dyes aperture. 4Pi microscopy places 2 identical objective
are available that can be taken up by living cells to display lenses on either side of the specimen, thereby effectively
nuclear morphology (eg, 49,6-diamidino-2-phenylindole doubling the angular aperture and thus the resolution in
[DAPI]). The thickness of the optical section can be as thin the Z plane of the 2 lenses, to about 200 nm (Figure 8).12 4Pi
as about 1 mm, comparing favorably to a paraffin section. microscopy remains diffraction limited but can be paired
Compared to confocal microscopy, it should be possible to with subdiffraction techniques such as stimulated emis-
cut such optical sections more deeply into the tissue sion depletion, discussed below.
because of the higher penetration of the exciting infrared
light. The advantage of optical sectioning compared to Spatially Modulated Illumination Microscopy
paraffin sectioning is the decreased technical steps When 2 densely patterned images are combined, an
required for imaging (there is no paraffinization or interference effect called a Moiré pattern can result. This
physical sectioning, and the staining itself can take place effect is often undesirable (for example, showing up in
in a liquid fixative or tissue culture media without the scanned magazine images). The Moiré effect is exploited
need for washing out the fluorescent dyes). Another major in spatially modulated illumination (SMI) by using a
Arch Pathol Lab Med—Vol 135, February 2011 Advances in Microscopy Techniques—Schmolze et al 259
Figure 7. Demonstration of 2-photon microscopy optical sectioning
of ovarian cancer in a liquid fixative. Two-photon image, about 5 mm in Figure 8. 4Pi microscopy principle. Two lenses are placed on
thickness, of a cluster of ovarian cancer cells from ascites, stained with opposing sides of the specimen, thereby effectively doubling the
49,6-diamidino-2-phenylindole (DAPI) and eosin in CytoRich Red angular aperture and the resolution.
fixative (Thermo Scientific, Hudson, New Hampshire), without washing
out any excess dye. Pseudo hematoxylin-eosin staining was accom-
plished as in Figure 1 (original magnification, approximately 31000).
equation limits optical resolution to about 200 nm. For a
long time, this limit was thought to be unsurpassable, but
pattern of fine lines as the illumination source.13,14 When recently several microscopy methods have been devel-
this illumination pattern (Figure 9, B) interferes with the oped that achieve far higher resolution. In fact, the
detail emitted from or transmitted by the fine structure of diffraction limit was never a ‘‘hard’’ limit, and several
the sample (Figure 9, A), a new pattern results (Figure 9, caveats exist that point the way toward subdiffraction
C). imaging. First, equation (2) only applies in the so-called far
The Moiré pattern is always more coarsely patterned field, where the light rays are allowed to travel a certain
than the sample, and so it may be resolvable even if the distance before being imaged (roughly, longer than the
fine structure of the sample is not. The Moiré pattern does wavelength of light). This is because very close to a source
not necessarily actually resemble the sample, but by (in the ‘‘near field’’) the action of so-called evanescent
collecting a series of images in which the illumination nonpropagating waves predominates, and these waves
pattern has been slightly altered (eg, by simply shifting its are not accounted for by the usual rules of interference.17
position), the underlying structure can be computed. The Secondly, molecular interactions between different fluo-
interference effect only occurs at the focal plane, and as the rophores can be visualized. Finally, the Abbe equation
illumination pattern is changed, the out-of-focus light only applies if 2 objects with an identical interaction with
remains invariant and can thus be subtracted from the light are simultaneously visualized. If there can be a
final image. In this way, efficient optical sectioning is distinction between the 2 objects in the spectra of their
possible. Although the technique requires the collection of emitting photons, or their temporal release of photons, the
multiple images, the computation can be performed Abbe limit does not apply, and in fact there is no theoretic
quickly, enabling live-cell imaging in conjunction with limit to light microscopy resolution, even in living cells!
fast cameras. Spatially modulated illumination is com-
mercially available as an add-on to most epifluorescence Near-Field Scanning Optical Microscope
microscopes at a cost that is considerably cheaper than Microscopy techniques have been developed that
that of confocal microscopy.15 exploit all of these caveats to break the diffraction limit.
Spatially modulated illumination is limited by diffrac- The near-field scanning optical microscope scans a tiny
tion in that there is a limit to the detail of the illumination probe across the surface of a specimen at distances over
pattern. Like 4Pi microscopy, SMI can be combined with which near-field effects come into play. The resolution is
other methods to yield further resolution enhancement. limited only by the size of the probe, which can be as small
One technique merges SMI and stimulated emission as 10 nm.18 A similar diffraction-limited technique called
depletion to yield subdiffraction imaging with a method total internal reflection fluorescence microscopy generates
termed nonlinear structured illumination.16 a near-field wave at the surface of a specimen.19 This wave
then excites fluorophores over a very narrow distance
SUBDIFFRACTION TECHNIQUES (Figure 10). Both techniques have been successfully used
With lenses of the largest possible angular aperture, in to visualize dynamic events, in the range of 10 nm,
media with the highest possible refractive index, the Abbe occurring at the plasma membrane of living cells.
260 Arch Pathol Lab Med—Vol 135, February 2011 Advances in Microscopy Techniques—Schmolze et al
Figure 9. The Moiré effect in structured illumination microscopy. A sample containing fine detail (A) when illuminated with a regularly patterned
image (B) gives rise to a coarsely patterned Moiré image (C). The original fine detail can be computed through collection of several Moiré images. An
important effect is that the out-of-plane component of the image is able to be computationally subtracted. Reprinted with permission from
Gustafsson.15 Copyright 2005, National Academy of Sciences, United States.

Fluorescence/Förster Resonance Energy Transfer corresponding increase in acceptor fluorescence. Donor


Fluorescence/Förster resonance energy transfer (FRET) and acceptor probes can be attached to 2 different
exploits a different physical principle to obtain high molecules of interest, or to 2 different domains of a single
spatial precision.20 Two fluorophores interact in propor- protein, making FRET a useful tool for measuring protein-
tion to the sixth power of their intermolecular separation. protein interactions or conformational changes at the
Within a few diameters of fluorescent molecules, one molecular level (Figure 11). Fluorescence/Förster reso-
excited molecule (the donor) can transfer its excited state nance energy transfer is used in a nonmorphologic context
to a different nonexcited fluorescent molecule (the as a read-out for human papilloma virus (HPV) detection
acceptor). No emission of light is involved in this transfer, in the Third Wave Technologies of Hologic Inc (Bedford,
and the acceptor will proceed to fluoresce at its charac- Massachusetts). In this case, 2 adjacent FRET fluorophores
teristic emission wavelength. The degree of separation of on a nucleic acid probe become separated through the
the 2 fluorochromes (in the range of 1 nm) can be action of a specific nuclease that cleaves a site between
estimated from the decrease in donor fluorescence and a them.
Fluorescence/Förster resonance energy transfer also
has morphologic applications in anatomic pathology, with
the potential to characterize diagnostic molecular events
that are invisible to existing light microscopy or existing
nonmorphologic molecular assays. An important example
is provided by the signaling pathway from receptor
tyrosine kinases to RAS, RAF, and, ultimately, MAPK.
There is strong evidence for activation of this pathway
(particularly at points upstream and downstream of RAS)
in papillary thyroid carcinomas.21–23 Paradoxically, activa-
tion of RAS is common in follicular-type thyroid tumors, a
tumor with distinctly different morphologic and clinical
features.23 Immunohistochemistry and Western blotting
have not been able to distinguish papillary thyroid
carcinomas with tyrosine kinase or BRAF activations from
follicular-type tumors or normal thyroid.23,24 The solution
to the paradox of how the same signaling pathway to
MAPK can cause 2 different responses (apparently in the
same cell of origin) could be the existence of differences in
the precise temporal and/or spatial aspects of signaling
(reviewed in Murphy and Blenis26). The power of FRET is
its ability to combine cellular-level localization with
dynamic information about molecular interactions, and
FRET applications are beginning to dissect out the precise
spatial and temporal involvement of events in the MAPK
Figure 10. Near-field microscopy with the total internal reflection
fluorescence microscope. The exciting light is angled such that it is
pathway that accounts for paradoxes such as this.27,28 By
completely reflected when it reaches the sample. A near-field wave is analogy to the findings in PC12 neuroblastoma cells, one
generated, which excites fluorophores in a very narrow range. The may expect papillary thyroid carcinomas to have a
resulting fluorescence is imaged as usual. sustained low-level activation of MAPK, whereas follicu-
Arch Pathol Lab Med—Vol 135, February 2011 Advances in Microscopy Techniques—Schmolze et al 261
Figure 12. Photoactivated localization microscopy and stochastic
optical reconstruction microscopy. Within a large population of
Figure 11. Fluorescence resonance energy transfer. When 2 fluoro- fluorophores (A), an initial pulse of light (B) causes a sparse subset of
chromes remain separated by more than a few molecular distances, the the fluorophores to become ‘‘activated’’ (C). A second pulse of light
exciting light causes only the first fluorochrome to fluoresce. As the then causes the activated fluorophores to actually fluoresce (D). The
distance is decreased, the excited state is transferred to the ‘‘acceptor’’ process is repeated until the location of every fluorophore in the sample
molecule, which proceeds to fluoresce at a different wavelength. is individually recorded.

lar-type tumors with RAS mutations would be expected to to fluoresce. The first pulse is weak enough to activate
have a transient high level activation of MAPK.26 It is only about 1 fluorophore molecule in the area of a laser
conceivable that FRET-based assays could find use some excitation (an area about 300 nm in diameter). By
day in ‘‘diagnosis’’ of specific oncogenic signals or following the first low-intensity ‘‘activation’’ pulse with
molecular interactions that are otherwise ‘‘invisible’’ by a second excitation pulse (Figure 12, D), individual
mutational analysis or immunohistochemistry. Fluores- molecules within the 300-nm area fluoresce a population
cence/Förster resonance energy transfer applications on of photons. The population of photons scatters according
viable cells provide dynamic, temporal information and to a point spread function, and the center of the function
would typically require some in vitro manipulations (eg, can be estimated. The process is repeated through
adenoviral-induced expression of green fluorescent pro- successive 300-nm areas until the location of every
tein fusions, see Figure 2), but static FRET can also be fluorescent molecule in the sample is multiply sampled.
performed on fixed material after labeling with informa- A high-resolution image is then computed. The process
tive FRET pairs.29 can take many hours to complete, and is thus not yet
suitable for studying cellular dynamics, but resolution of
Photoactivated Localization Microscopy and Stochastic about 15 nm has been achieved.31
Optical Reconstruction Microscopy
Even if a point source of light is converted by an optical Stimulated Emission Depletion
system into a fuzzy Airy disk, repeated sampling of an Stimulated emission depletion32 is notable in that a
individual source of light can statistically show the commercial product has recently been announced.33
location of the center of its Airy disk. The statistical Stimulated emission depletion makes use of 2 separate
sampling defines localization precision, a concept distinct geometries and wavelengths of pulsed light. The first
from optical resolution. Localization precision is not pulse of light is focused to the smallest possible spot
limited by the Abbe limit, and statistical
pffiffiffiffi sampling (about 300 nm in diameter) and excites all of the
improves resolution by a factor of 1 N, where N is the fluorochromes within this spot, just like a conventional
number of sampled photons. Photoactivated localization confocal microscope. A second pulse of light is focused
microscopy30 and stochastic optical reconstruction micros- into a donut-shaped image, with the hole of the donut
copy31 are related applications of this concept. Within a centered on the first pulse (Figure 13). The second pulse
large population of fluorophores (Figure 12, A), a low- has a wavelength that quenches fluorescence induced by
intensity laser excitation first ‘‘activates’’ rare fluoro- the first pulse. The subdiffraction resolution of stimulated
phores to become susceptible to a second excitation laser emission depletion is due to nonlinear optical effects34 that
pulse (Figure 12, B and C). The first pulse does not come into play as the intensity of the quenching beam
actually induce fluorescence; it merely brings the fluo- increases. The end result is a point-by-point scan through
rophore to a state where it can become repeatedly excited the sample at a much higher resolution than conventional
262 Arch Pathol Lab Med—Vol 135, February 2011 Advances in Microscopy Techniques—Schmolze et al
Figure 13. Stimulated emission depletion
microscopy. An initial excitation beam is
followed closely by a second beam shaped
like a donut, which serves to quench the
fluorescence of the first beam. The result is an
excitation spot smaller than diameter that can
be achieved by diffraction of a single source
of light.

techniques—about 20 nm. Stimulated emission depletion 11. Istvanic S, Fischer AH, Banner BF, et al. Cell blocks of breast FNAs
frequently allow diagnosis of invasion or histological classification of proliferative
can be applied to green fluorescent proteins and their changes. Diagn Cytopathol. 2007;35(5):263–269.
derivatives, enabling live-cell imaging, but it is currently 12. Hell S, Stelzer E. Properties of a 4Pi confocal fluorescence microscope. J
too slow for visualizing live-cell dynamics. Opt Soc Am. 1992;A9(12):2159–2166.
13. Neil MAA, Juskaitis R, Wilson T. Method of obtaining optical sectioning by
using structured light in a conventional microscope. Opt Lett. 1997;22(24):1905–
IMPLICATIONS FOR PATHOLOGY 1907.
At the finest levels of biology it is easy to appreciate 14. Gustafsson MG. Surpassing the lateral resolution limit by a factor of two
using structured illumination microscopy. J Microsc. 2000;198(pt 2):82–87.
how a change in protein structure affects the function of a 15. Qioptiq LINOS, Inc. Qioptiq Web site. http://www.qioptiqlinos.com/.
protein, and it is easy at the tissue level or organ level to Accessed February 4, 2010
appreciate how changes in structure lead to changes in 16. Gustafsson MG. Nonlinear structured-illumination microscopy: wide-field
physiology. At the cellular level, the relation between fluorescence imaging with theoretically unlimited resolution. Proc Natl Acad Sci
U S A. 2005;102(37):13081–13086.
structure and function is less obvious. For example, it is 17. Girard C, Dereux A. Near-field optics theories. Rep Prog Phys. 1996;59(5):
not obvious why nuclear structure should differ between 657–699.
papillary thyroid carcinoma and follicular-type thyroid 18. Dürig U, Pohl DW, Rohner F. Near-field optical-scanning microscopy. J
Appl Phys. 1986;59(10):3318–3327.
tumors, or why nucleolar enlargement would be charac- 19. Fish KN. Total internal reflection fluorescence (TIRF) microscopy. Curr
teristic of prostatic intraepithelial neoplasia. The cellular Protoc Cytom. 2009;chapter 12:unit 12.18.
level is the bottleneck for understanding mechanisms of 20. Clegg RM. Fluorescence resonance energy transfer. Curr Opin Biotechnol.
1995;6(1):103–110.
disease. We can know the gene that is responsible for 21. Fagin JA. How thyroid tumors start and why it matters: kinase mutants as
certain diseases (eg, Huntington disease) and we can targets for solid cancer pharmacotherapy. J Endocrinol. 2004;183(2):249–256.
know the tissue-level structural changes that result from 22. Michaloglou C, Vredeveld LC, Mooi WJ, Peeper DS. BRAF(E600) in benign
the gene, but precise understanding of disease mecha- and malignant human tumours. Oncogene. 2008;27(7):877–895.
23. Nikiforova MN, Lynch RA, Biddinger PW, et al. RAS point mutations and
nisms is lost at the cellular level. Advances in imaging at PAX8-PPAR gamma rearrangement in thyroid tumors: evidence for distinct
the cellular to molecular level, especially techniques that molecular pathways in thyroid follicular carcinoma. J Clin Endocrinol Metab.
allow dynamic imaging of living cells, should enable 2003;88(5):2318–2326.
24. Pomerance M, Quillard J, Chantoux F, Young J, Blondeau JP. High-level
pathologists to play a central role in uncovering the expression, activation, and subcellular localization of p38-MAP kinase in thyroid
elusive structure-function relationships that underlie and neoplasms [comment in J Pathol. 2006;210(1):133–134]. J Pathol. 2006;209(3):
define diseases. 298–306.
25. Zuo H, Nakamura Y, Yasuoka H, et al. Lack of association between BRAF
References V600E mutation and mitogen-activated protein kinase activation in papillary
1. Levenson RM, Fornari A, Loda M. Multispectral imaging and pathology: thyroid carcinoma. Pathol Int. 2007;57(1):12–20.
seeing and doing more. Expert Opin Med Diagn. 2008;2:1067–1081. 26. Murphy LO, Blenis J. MAPK signal specificity: the right place at the right
2. Fischer AH, Taysavang P, Jhiang SM. Nuclear envelope irregularity is time. Trends Biochem Sci. 2006;31(5):268–275.
induced by RET/PTC during interphase. Am J Pathol. 2003;163(3):1091–1100. 27. Abankwa D, Hanzal-Bayer M, Ariotti N, et al. A novel switch region
3. Wyckoff JB, Wang Y, Lin EY, et al. Direct visualization of macrophage- regulates H-ras membrane orientation and signal output. EMBO J. 2008;27(5):
assisted tumor cell intravasation in mammary tumors. Cancer Res. 2007;67(6): 727–735.
2649–2656. 28. Moreto J, Llado A, Vidal-Quadras M, et al. Calmodulin modulates H-Ras
4. Abbe EC. Neue Apparate zur Bestimmung des Brechungs - und mediated Raf-1 activation. Cell Signal. 2008;20(6):1092–1103.
Zerstreuungsvermögens fester und fluüssiger Körper. Mauke’s Verlag. 1874. 29. Periasamy A, Wallrabe H, Chen Y, Barroso M. Chapter 22—quantitation of
5. Minsky M. Memoir on inventing the confocal scanning microscope. protein-protein interactions: confocal FRET microscopy. Methods Cell Biol.
Scanning. 1988;10:128–138. 2008;89569–89598.
6. Olympus. New FluoView FV10i Confocal Laser Scanning Microscope from 30. Betzig E, Patterson G, Sougrat R, et al. Imaging intracellular fluorescent
Olympus. http://www.nanowerk.com/news/newsid513565.php. Accessed Feb- proteins at nanometer resolution. Science. 2006;313(5793):1642–1645.
ruary 4, 2010. 31. Rust M, Bates M, Zhuang X. Sub-diffraction-limit imaging by stochastic
7. Diaspro A, Chirico G, Collini M. Two-photon fluorescence excitation and optical reconstruction microscopy (STORM). Nat Methods. 2006;3(10):793–795.
related techniques in biological microscopy. Q Rev Biophys. 2005;38(2):97–166. 32. Hell S, Wichmann J. Breaking the diffraction resolution limit by stimulated
8. Göppert-Mayer M. Über Elementarakte mit zwei Quantensprüngen. Ann emission: stimulated-emission-depletion fluorescence microscopy. Opt Lett.
Phys. 1931;401(3):273–294. 1994;19(11):780–782.
9. Bhawalkar JD, Kumar ND, Zhao CF, Prasad PN. Two-photon photodynamic 33. Leica Microsystems. Leica TCS STED—confocal microscopes: Leica
therapy. J Clin Laser Med Surg. 1997;15(5):201–204. Microsystems. http://www.leica-microsystems.com/products/confocal-microscopes/
10. Ogawa K, Kobuke Y. Recent advances in two-photon photodynamic details/product/leica-tcs-sted. Accessed February 4, 2010.
therapy. Anticancer Agents Med Chem. 2008;8(3):269–279. 34. Boyd R. Nonlinear Optics. 3rd ed. New York, NY: Academic Press; 2008.

Arch Pathol Lab Med—Vol 135, February 2011 Advances in Microscopy Techniques—Schmolze et al 263

You might also like