Professional Documents
Culture Documents
201–207
doi:10.1093/rpd/nch554
The implementation of film replacement digital radiographic imaging systems throughout Europe is now gathering
momentum. Such systems create the foundations for totally digital departments of radiology, since radiographic examinations
constitute the most prevalent modality. Although this type of development will lead to improvements in the delivery and
management of radiological service, such widespread implementation of new technology must be carefully monitored. The
implementation of effective QA tests on installation, at periodic intervals and as part of a routine programme will aid this
process. This paper presents the results of commissioning tests undertaken on a number of computed radiography imaging
systems provided by different manufacturers. The aim of these tests was not only to provide baseline performance measure-
ments against which subsequent measurements can be compared but also to explore any differences in performance,
which might exist between different units. Results of measurements will be presented for (1) monitor and laser printer set-
up; (2) imaging plates, including sensitivity, consistency and uniformity; (3) resolution and contrast detectability; and (4)
signal and noise performance. Results from the latter are analysed in relationship with both system and quantum noise
components.
ª The Author 2005. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oupjournals.org
P. CHARNOCK ET AL.
both the protocol and the CR performance. Meas- digital radiography (DDR) and so no analysis is
urements of SNR on these systems are also included presented here.
in the assessment.
Monitor and laser printer set-up
Test procedures—general considerations This involves the use of the SMPTE test pattern. On
The test protocol involves the assessment of the one unit the pattern generator was not included in
following when commissioning a system: the system. On another unit normal viewing condi-
tions on screen were not full size centred. It was
80% size and displaced 2 cm to the left.
monitor and laser printer set-up,
dark noise,
erasure cycle efficiency, Sensitivity index calibration
sensitivity index consistency, The protocol requires that tube factors are selected
uniformity, to provide an air kerma of 10 mGy to the plate. On
scaling errors, one particular system the output was very variable
blurring, for a slight adjustment in the set-up. Eventually an
limiting spatial resolution, air kerma of 12 mGy was produced. Although the
threshold contrast detectability, KCARE protocol recommends that no filtration is
laser beam function, employed using a Fuji system, a protocol obtained
Moiré patterns. from Philips, which appears to come from the Insti-
The details of the test methods are described in tute of Physics and Engineering in Medicine (IPEM)
the KCARE protocol. Once commissioning meas- protocol that recommends 1.5 mm Al.
urements have been undertaken it is recommended
that the following annual QA tests are performed: Sensitivity index consistency
For high-resolution plate the sensitivity was
monitor and laser printer set-up, 2.5 times higher for one system at least, when com-
sensitivity index consistency, pared with normal-resolution plates. Also on another
uniformity, system one plate recorded a sensitivity equivalent to
blurring and stitching artefacts, an unexposed plate even though it had been exposed.
limiting spatial resolution (45 only), Apparently this may occur if there is a small artefact
threshold contrast detectability. on the plate or the phosphor is not quite lined up.
The remedial levels are recommended in Table 1.
The commissioning tests have been performed on Limiting spatial resolution
a number of CR units representing three manu-
facturers, Agfa Gevaert, Kodak and Philips (Fuji). Problems were encountered in determining the pixel
A short paragraph on the different plate systems dimensions for a CR system. Local staff did not
is included in the Appendix section. Details of understand what this meant in relationship with the
only those tests relating to sensitivity index and CR system. When performing tests on a DDR sys-
threshold contrast detectability are presented here. tem, the pixel size was given on an image printout.
Also included are some measurements undertaken Pixel dimensions for other CR systems were not
with respect to signal-to-noise. Before presenting clearly understood and a proposal was made by the
these results it is worth noting that it is not always support staff to divide the image matrix by 3.2 mega
possible to access particular information at the pixels. Clarification is required on a number of issues
local level. It is also noted that measurements if field testing of the CR / DDR systems is to be
were only performed on the CR systems. At present, consistent.
only limited tests have been performed on direct
RESULTS AND DISCUSSION
Table 1. The recommended remedial levels for the annual Sensitivity index
QA test.
The sensitivity index was measured for different val-
ues of entrance air kerma to the imaging plate for all
Test Remedial level
three manufacturers’ systems. The results are presen-
ted in Figure 1 where the calculated and measured
Sensitivity index consistency Baseline 20% values are shown. Note that all the manufacturers
Limiting resolution Baseline 20%
Threshold contrast detectability Baseline 30%
give the sensitivity a different label. Agfa calls it
‘lgM’, Kodak calls it ‘EI’ or exposure index and
202
EVALUATION AND TESTING OF CR SYSTEMS
4.00
3.50
3.00
Log [Sensitivity]
2.50
2.00
1.50
1.00
0.0 0.5 1.0 1.5 2.0 2.5
Log [Air Kerma (µGy)]
Kodak - Measured Kodak - Calculated
Philips - Measured Philips - Calculated
Agfa - Measured Agfa - Calculated
Fuji simply calls it ‘S’. The functional forms relating For the Kodak system the MPV and sensitivity
exposure to sensitivity are: values are identical. However, for the Agfa and
Fuji (Philips) system they are somewhat different.
Agfa exposure ¼ 9.33 10lgM 3.2766 The MPV data for the Agfa system now does show
Kodak exposure ¼ 8.7 10n, a square root dependence on air kerma. The Fuji
where n ¼ (EI 2000)/1000 (Philips) system shows an almost constant MPV
Fuji (Philips) exposure ¼ 1740/S for air kerma values >10 mGy.
The curves shown in Figure 1 demonstrate the wide
variations in sensitivity as a function of exposure
among the different manufacturers. Not only are
Signal-to-noise
the gradients different but also the functional form.
The curve for the Kodak system roles off markedly In order to explore the SNR performance of the
for very low exposures. CR systems the average pixel value in a given ROI
Previous measurements performed on Agfa sys- as well as the standard deviation was meas-
tems have indicated a square root dependence ured(12,13), where SNR equals the MPV divided
between log (sensitivity) and log (air kerma)(15), by the standard deviation. These measurements
whereas Figure 1 shows an almost direct proportion- could not be performed on the Agfa system since
ality. The slope of the Fuji (Philips) system shown in the image display facility would not provide a
Figure 1 was the negative reflection of the Agfa measure of standard deviation for an ROI. This
system. It must be noted that the sensitivity index again highlights the variations in system perform-
‘S’ given for the Fuji (Philips) system probably does ance between manufacturers. The results for the
not represent the actual pixel value employed. Fur- Kodak and Fuji (Philips) systems are shown in
ther work required to show what it does represent is Figure 3. Obviously, it is not possible to compare
in progress. the two systems directly when SNR has been
The results presented in Figure 1 were nominally expressed in this manner since both systems have
obtained for unprocessed data. However, once the completely different exposure characteristics, owing
CR plate has been read, the system assigns an MPV to the fact that there is no common relationship
to the histogram of pixel values obtained in an between the pixel value and the exposure. How-
image. This ensures that all relevant image data are ever, it should be possible to compare systems if
optimally displayed. Figure 2 shows the log(MPV) signal and noise values are expressed in terms of
plotted against log(incident air kerma). plate exposure(12).
203
P. CHARNOCK ET AL.
3.6
Kodak
Philips
Agfa
3.4
3.2
Log [MPV]
3.0
2.8
2.6
2.4
0.5 0.7 0.9 1.1 1.3 1.5 1.7 1.9 2.1
Log [Air Kerma (µGy)]
Figure 2. Log (MPV for 0.5 cm2 ROI) vs. log (air kerma).
650
Kodak
Philips
550
450
SnR
350
250
150
50
0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0
Log [Air Kerma (µGy)]
Figure 4 shows the SNR, expressed in exposure using the information shown in Figure 1 gave
terms, for the Kodak system. Unfortunately, a decreasing SNR value with exposure. As indic-
given the exposure characteristics expressed in ated previously the sensitivity value ‘S’ given
terms of sensitivity it was not possible to make a for the Fuji (Philips) system probably does not
sensible appraisal of SNR for the Fuji (Philips) correspond to the actual pixel value. The varia-
system. In fact values calculated for this system tion in MPV with exposure shown in Figure 2
204
EVALUATION AND TESTING OF CR SYSTEMS
for the Fuji (Philips) system does not clarify the Threshold contrast detectability
situation.
Threshold contrast detectability measurements
The value of SNR vs. exposure presented in
were undertaken on the three CR systems, Agfa,
Figure 4 for the Kodak system is in reasonable
Kodak and Fuji (Philips), using the Leeds FAXIL
agreement with the values calculated by this method
test phantom TO20(16). This consists of 144 circu-
previously(12). The value of SNR expressed in this
lar details with 12 sizes and 12 contrasts. Detail
form corresponds to the number of independent
sizes range from 11 to 0.25 mm diameter and
noise levels in the dynamic (average signal) range of
contrast range from 0.97 to 49.7% for an 80 kV
the system at the appropriate exposure level. Such a
beam filtered by 1 mm Cu. Measurements were
measure should be related with the threshold
made when employing this phantom viewed under
contrast for the appropriate detail size and exposure
the normal local viewing conditions. No effort
conditions.
90
80
70
Air Kerma / Delta Air Kerma
60
50
40
30
20
10
0
0.0 0.5 1.0 1.5 2.0 2.5
Log [Air Kerma (µGy)]
1
3.9 µGy
12.7 µGy
Contrast (%)
0.1
0.01
0.1 1 10 100
Diameter (mm)
205
P. CHARNOCK ET AL.
was made to optimise the conditions since we were visual detection the threshold signal is give by:
interested in performing true field type measure-
ments. S ¼ k noise, ð1Þ
Viewing conditions varied from a small darkened where k is the threshold SNR.
room to a larger room with multiple VDUs (Agfa). Threshold contrast for a given detail size is given
No attempt was made to process or contrast enhance by:
the image.
S noise
Contrast-detail diagrams for each of the systems Cmin ¼ ¼k , ð2Þ
are shown in Figures 5–7. The Agfa and Fuji s average signal
(Philips) systems showed similar contrast-detail where s is the average signal. The value of noise/
behaviour across all detail sizes and exposure condi- average signal is the reciprocal of the SNR values
tions in Figures 6 and 7, respectively. The Kodak shown in Figure 4. For the Kodak system, the
system showed most exposure dependence and lower 0.5 cm2 ROI gave an SNR value at 3.9 mGy of 44.
threshold detectability for smaller detail sizes. For Assuming a threshold SNR value of 2 for circular
4.2 µGy
10.9 µGy
18.4 µGy
Contrast (%)
0.1
0.01
0.1 1 10 100
Diameter (mm)
2.9 µGy
4.9 µGy
12.14 µGy
Contrast (%)
0.1
0.01
0.1 1 10 100
Diameter (mm)
206
EVALUATION AND TESTING OF CR SYSTEMS
detail, the calculated threshold contrast value for a 8. Buhr, E., Gunther-Kohfahl, S. and Neitzel, U. Accur-
0.5 cm2 ROI on the Kodak system is 0.05 (5%). The acy of a simple method for deriving the pre-sampled
calculated value is roughly twice that observed in modulation transfer function of a digital radiographic
Figure 5 (2.3%) and perhaps indicates that with system from an edge image. Med. Phys. 30(9), 2323–
more careful field measurements simple SNR meas- 2330 (2003).
9. Illers, H., Buhr, E., Benjamin, D. and Hoeschen, C.
urements might usefully underpin threshold Measurement of the detective quantum efficiency (DQE )
contrast-detail measurements. of digital x-ray imaging device. According to the Forth-
coming Standard IEC 6220-1. In: Proceedings of the
SPIE Conference on Medical Imaging, San Diego
SUMMARY AND CONCLUSIONS (Bellingham: SPIE) (2004).
Work has been undertaken in order to evaluate the 10. Illers, H., Buhr, E. and Vandenbracke, D. Measure-
ment of correlated noise in images of computed radio-
KCARE test protocol for the CR and DDR imaging
graphy systems and its influence on the detective
systems. Results indicate that the protocol is quantum efficiency. In: Proceedings of the SPIE Con-
extremely useful for those wishing to gain a more ference on Medical Imaging, San Diego (Bellingham:
detailed knowledge of the performance of these sys- SPIE) (2004).
tems. However, our experience has shown that sup- 11. Cowen, A. R., Workman, A. and Price, J. S. Physical
port and knowledge on these new imaging systems is aspects of photostimulable phosphor computed radio-
extremely varied. Information which is important graphy. Br. J. Radiol. 66, 332–345 (1993).
for the assessment of performance is not available. 12. Meier, N. and Fiebich, M. Standard tools for analysis
Equally, information that is available is not easily of image quality in digital imaging. Radiat. Prot.
Dosim. 57(1–4), 141–143 (1995).
understood in terms of its relevance to the assess-
13. Wagner, R. F. and Brown, D. G. Unified SNR analysis
ment of imaging performance. Preliminary measure- of medical imaging systems. Phys. Med. Biol. 30, 489–
ments of SNR indicate that, given the right technical 578 (1985).
information, the simple SNR assessment methods 14. KCARE. Protocol for the QA of computed radiography
could provide useful information on imaging per- systems—commissioning and annual QA tests. Draft
formance and help to underpin threshold contrast- 3.0, King’s College Hospital, London (2003).
detail measurements. 15. Ter-Pogossian, M. M. The Physical Aspects of Dia-
gnostic Radiology (London: Harper and Row) (1967).
16. Cowen, A. R., Haywood, J. M., Workman, A. and
ACKNOWLEDGEMENT Clarke, O. F. A set of x-ray test objects for image
This work was undertaken with support provided by quality control in digital subtraction fluorography. I.
Design considerations. II. Application and interpretation
the Community of European Communities under of results. Br. J. Radiol. 60, 1001–1009, 1010–1018
grant no. FIGM-CT-2000-0036. (1987).
17. KCARE, MDA. Computed Radiography (CR)
System. Agfa HealthCare ADC Solo/Compact. Evalu-
REFERENCES ation Report No. MDA02062 (2002).
1. Kamm, K. F. The future of digital imaging. Br. J. 18. KCARE, MDA. Computed Radiography (CR)
Radiol. 70(Special issue), S145–S152 (1997). System. Fuji Medical Imaging FCR 5000. Evaluation
2. Rowlands, J. A. The physics of computed radiology. Report No. MDA02062 (MHRA) (2003).
Phys. Med. Biol. 47(23), R123–R166 (2002).
3. Doi, K. Present status and future horizons for computer
aided diagnostic in radiology. In: Physics and Engineer- APPENDIX
ing in Medicine in the New Millennium. Sharp, P. F. All three systems encountered are based on the fol-
and Perkins, A. C., Eds. (York: IPEM) pp. 84–87 (2000). lowing design; plates, cassettes, ID system, processor
4. Dobbins, J. T. and Godfrey, D. J. Digital X-ray tomo-
synthesis: current state of the art and clinical potential. and acquisition workstation(17–18). They all have
Phys. Med. Biol. 48(19), R65–R106 (2003). various plate/cassette sizes but this work only
5. Dobbins, J. T. Image quality metric for digital systems. encountered high resolution plates with the Kodak
In: Handbook of Medical Imaging. Vol. 1. Physics system. Agfa plates can be read using a different
and Psychophysics. Beutel, J., Kundel, H. L. and Van speed class, which refers to the acceptable exposure
Meter, R. L., Eds. (Bellingham: SPIE) pp. 161–222 range. Speed class ranges from 50 to 400, which
(2000). corresponds to a saturation exposure of 35 to
6. Samei, E., Flynn, M. J. and Reimann, D. A. A method 290 mGy respectively(17). The Philips (Fuji) system
for measuring the pre-sampled MTF of digital radio- encountered here appears to consist of a Fuji / Philips
graphic systems using an edge test device. Med. Phys.
25, 102–113 (1998). collaboration. This may not be the case with all Fuji
7. Workman, A., Cowen, A. R. and Brettle, D. S. sites. The plates themselves have a wide dynamic
Physical evaluation of computed radiography as a range, which may account for a lack of high resolu-
mammographic X-ray imaging systems. Br. J. Radiol. tion plates(18). Kodak have both normal and high
67, 988–996 (1994). resolution plates.
207