You are on page 1of 122

SB.

3500
INSTRUMENT SERIES

INSTRUCTION MANUAL
Copyright © 1997, Daytronic Corporation. All rights reserved.

No part of this document may be reprinted, reproduced, or used in any form or by


any electronic, mechanical, or other means, including photocopying and recording,
or in any information storage and retrieval system, without permission in writing from
Daytronic Corporation. All specifications are subject to change without notice.
3500 Series Instruction Manual, v. SB.3
Pub. No. 3500M.3, Issued 1/97
Part No. 91668

3500
INSTRUMENT
SERIES
INSTRUCTION MANUAL

Daytronic Corporation
2211 Arbor Blvd. • Dayton, OH 45439-1521 • Tel (937) 293-2566 • Fax (937) 293-2586
www.daytronic.com
CONTENTS

1 INTRODUCTION
a. Using This Manual ........................................................................................................ 1.1
b. General Instrument Descriptions
1. Introduction ....................................................................................................................... 1.1
2. The Model 3510 Thermocouple Conditioner ....................................................... 1.2
3. The Model 3530 LVDT Conditioner .......................................................................... 1.2
4. The Model 3540 Frequency Input Conditioner .................................................... 1.2
5. The Model 3560 Voltage Conditioner ..................................................................... 1.3
6. The Model 3570 DC Strain Gage Conditioner ...................................................... 1.3
7. The Model 3578 AC Strain Gage Conditioner ...................................................... 1.3
c. Physical Layout .............................................................................................................. 1.4
d. Panel Mounting ............................................................................................................... 1.6
e. Summary of Setup Button Functions ............................................................ 1.7
f. Summary of Logic I/O Functions ..................................................................... 1.8
g. Mnemonic Commands ............................................................................................. 1.9

2 SETUP: CONNECTIONS AND POWER


a. Transducer Connections
1. The “Standard” Analog Input Connector ............................................................... 2.1
2. Connecting a Thermocouple to the Model 3510 ................................................ 2.1
3. Connecting an LVDT or Variable Reluctance Transducer
to the Model 3530 .......................................................................................................... 2.2
4. Connecting a Frequency Source to the Model 3540 ........................................ 2.4
5. Connecting a Voltage Source to the Model 3560 .............................................. 2.6
6. Connecting a DC Strain Gage Transducer to the Model 3570 ...................... 2.7
7. Connecting an AC Strain Gage Transducer to the Model 3578 .................... 2.8
b. RS-232 (“Single-Node”) Connections .......................................................... 2.10
c. RS-485 (“Multinode Network”) Connections ......................................... 2.12
d. Analog Output Connections .............................................................................. 2.15
e. Logic Input/Output Connections ................................................................... 2.16
f. Powerup ............................................................................................................................ 2.18

3 SETUP: INSTRUMENT CONFIGURATION


a. Configuring Through the Front Panel
1. Entering and Exiting SETUP MODE .......................................................................... 3.1
2. Security Code .................................................................................................................. 3.2
3. RS-232 Communications Parameters: COM Key .............................................. 3.3
4. RS-485 Communications Parameters: COM Key .............................................. 3.7
5. Input Range: RANGE Key ........................................................................................... 3.8
a. Setting the Model 3510’s TC Type and Scale ................................................ 3.8
b. Setting the Model 3530’s LVDT Input Range .................................................. 3.9
c. Setting the Model 3540’s Frequency Input Range and Sensitivity ...... 3.10
d. Setting the Model 3560’s Voltage Input Range .......................................... 3.10
e. Setting the Model 3570’s DC Strain Gage Input Range
and Excitation .......................................................................................................... 3.11
f. Setting the Model 3578’s AC Strain Gage Input Range ........................... 3.11

iv
CONTENTS
6. Filter: FILTER Key ......................................................................................................... 3.12
7. Analog Output: ANO Key .......................................................................................... 3.13
8. Limits: LIMIT Key .......................................................................................................... 3.14
9. Tare Offset: TARE Key ................................................................................................ 3.17
10. Print and Output Parameters: PRINT Key ........................................................... 3.18
b. Configuring Through the RS-232/485 Interface .................................. 3.20
1. Security Code ............................................................................................................... 3.20
2. Communications Parameters ................................................................................. 3.20
3. Thermocouple Type (Model 3510 ONLY) ........................................................... 3.21
4. Input Range or Scale .................................................................................................. 3.21
5. Excitation (Models 3570 and 3578 ONLY) .......................................................... 3.22
6. Sensitivity (Model 3540 ONLY) ............................................................................... 3.22
7. Filter .................................................................................................................................. 3.22
8. Analog Output .............................................................................................................. 3.23
9. Limits ................................................................................................................................ 3.23
10. Tare Offset ...................................................................................................................... 3.23
11. Print and Output Parameters ................................................................................... 3.23

4 SETUP: INSTRUMENT CALIBRATION


a. Introduction: Calibration Techniques ............................................................ 4.1
1. Absolute Calibration ...................................................................................................... 4.1
2. “Actual” Two-Point (Deadweight) Calibration ....................................................... 4.2
3. “Simulated” Two-Point (Deadweight) Calibration ............................................... 4.2
4. Internal 15-Segment Linearization ........................................................................... 4.3
5. Calculated Calibration .................................................................................................. 4.3
6. Calculated Calibration Via Mnemonic Command .............................................. 4.3
b. Phase and Symmetry Adjustment of the
Model 3578 AC Strain Gage Conditioner .................................................... 4.4
c. Calibrating Through the Front Panel
1. Using the CAL Key ......................................................................................................... 4.8
2. “Actual” Two-Point (Deadweight) Calibration ....................................................... 4.8
3. “Simulated” (Shunt) Calibration for a Strain Gage Conditioner
(Model 3570 or 3578) ................................................................................................. 4.10
4. 15-Segment Linearization
a. Introduction .............................................................................................................. 4.12
b. By “TABLE” ............................................................................................................... 4.13
c. By “FORCE” .............................................................................................................. 4.15
5. “Calculated” Calibration
a. For the Models 3510, 3530, 3540, and 3560 ............................................... 4.16
b. For the Model 3570 (ONLY) ............................................................................... 4.17
d. Calibrating Through the RS-232/485 Interface .................................... 4.18
1. Setting the Active Calibration Method .................................................................. 4.18
2. “Two-Point” Calibration .............................................................................................. 4.19
3. 15-Segment Linearization ........................................................................................ 4.19
4. “Calculated” Calibration ............................................................................................ 4.20

v
CONTENTS

5 RUN-TIME OPERATION
a. RS-232/485 Communications
1. RS-232/485 Outputs: CHN and DMP Commands ............................................ 5.1
2. Data-Transmission Format ......................................................................................... 5.1
3. “Opening” a Node to Receive Commands from the Computer ................... 5.2
b. Peak Capture ................................................................................................................... 5.3
c. Tare and Reset ................................................................................................................ 5.5
d. Initiating Hard-Copy Printouts ............................................................................. 5.6
e. Track/Hold Function .................................................................................................. 5.7

Appendix A 3500 SERIES SPECIFICATIONS


1. General Specifications ........................................................................................... A.1
2. Individual Conditioner Specifications
a. Model 3510 Thermocouple Conditioner .............................................................. A.2
b. Model 3530 LVDT Conditioner ................................................................................. A.3
c. Model 3540 Frequency Input Conditioner ........................................................... A.4
d. Model 3560 Voltage Conditioner ............................................................................. A.5
e. Model 3570 DC Strain Gage Conditioner ............................................................. A.6
f. Model 3578 AC Strain Gage Conditioner ............................................................. A.7

Appendix B COMMAND AND RESPONSE SYNTAX


1. Introduction: RS-232 and RS-485 Modes .................................................... B.1
2. The OPEN (OPN) Command ................................................................................ B.1
3. Response to “Invalid” Commands ................................................................. B.1
4. Setup Commands ....................................................................................................... B.2
5. Interrogation Commands ...................................................................................... B.2
6. Imperative Commands ............................................................................................ B.2
7. Commands That Initiate Data Transmissions ....................................... B.3
8. Table of Mnemonic Commands ....................................................................... B.3

Appendix C TABLE OF ERROR NUMBERS ...................................... C.1

Appendix D NETWORKING GUIDELINES ........................................... D.1

Appendix E RUN-TIME BUTTON FUNCTIONS ............................. E.1

Appendix F STANDARD LOGIC CONFIGURATION ................... F.1

vi
CONTENTS

Illustrations
1(a) Typical “Run-Time” Display ........................................................................................... 1.4
1(b) Typical “Setup” Display .................................................................................................. 1.4
2(a) Rear Panel for the Model 3510 Thermocouple Conditioner
and the Model 3578 AC Strain Gage Conditioner ................................................ 1.5
2(b) Rear Panel for All Other 3500 Models ...................................................................... 1.6
3 Panel Mounting ................................................................................................................. 1.6
4 Use of NUMERIC BUTTONS in SETUP MODE ...................................................... 1.8
5(a) Standard Logic INputs and Outputs ......................................................................... 1.8
5(b) Shunt-Control Logic Inputs for the Model 3570 DC Strain Gage
Conditioner (ONLY) .......................................................................................................... 1.9
6 Model 3510 Transducer Cabling ............................................................................... 2.2
7(a) Model 3530 Transducer Cabling: 5-Wire LVDT Cabling
(under 20 ft. in length) .................................................................................................... 2.3
7(b) Model 3530 Transducer Cabling: 7-Wire LVDT Cabling (20 ft. or longer) ... 2.3
7(c) Model 3530 Transducer Cabling: 3-Wire Variable Reluctance Cabling
(under 20 ft. in length) .................................................................................................... 2.3
7(d) Model 3530 Transducer Cabling: 5-Wire Variable Reluctance Cabling
(20 ft. or longer) ................................................................................................................ 2.4
8(a) Model 3540 Transducer Cabling: Differential (Floating) Frequency Input .. 2.5
8(b) Model 3540 Transducer Cabling: Single-Ended (Grounded) Frequency
Input ...................................................................................................................................... 2.5
8(c) Model 3540 Transducer Cabling: Input from a Zero-Velocity Sensor .......... 2.5
8(d) Model 3540 Transducer Cabling for Elimination of DC Offset ........................ 2.5
8(e) Model 3540 Transducer Cabling for Suppression of High-Frequency
Noise .................................................................................................................................... 2.5
9(a) Model 3560 Transducer Cabling: General Voltage Source ............................. 2.6
9(b) Model 3560 Transducer Cabling: External Potentiometer ............................... 2.6
9(c) Model 3560 Transducer Cabling: External DC-to-DC LVDT ............................. 2.6
10(a) Model 3570 Transducer Cabling: 4-Wire Cabling (under 20 ft. in length) ... 2.7
10(b) Model 3570 Transducer Cabling: 8-Wire Cabling (20 ft. or longer) .............. 2.7
11(a) Model 3578 Transducer Cabling: 4-Wire Cabling (under 20 ft. in length) ... 2.8
11(b) Model 3578 Transducer Cabling: 8-Wire Cabling (20 ft. or longer) .............. 2.8
11(c) Model 3578 Transducer Cabling: Installation of User’s External Shunt
Calibration Resistor ......................................................................................................... 2.9
11(d) Model 3578 Transducer Cabling: 8-Wire Cabling to LEBOW 1600
SERIES TRANSDUCER (ONLY) ................................................................................... 2.9
12 Suggested RS-232-C Interface Connections (to 25-Pin RS-232-C
Connector) ....................................................................................................................... 2.11
13 Suggested RS-232-C Interface Connections (to 9-Pin RS-232-C
Connector) ....................................................................................................................... 2.12
14(a) Connections for a Network of Three Instrument Nodes
(where the first is a 3500 Series Instrument) ...................................................... 2.13
14(b) RS-485 Cabling Between the Model 5E485 and the FIRST Network
Node (if it is a 3500 Series Instrument) ................................................................. 2.14
14(c) RS-485 Cabling Between Successive 3500 or 4000 Instrument Nodes .. 2.14
14(d) RS-485 Cabling Between a 3500 or 4000 Instrument Node and a
5000 Instrument Node ................................................................................................ 2.14
15 Analog Output Programming Pins .......................................................................... 2.15
16 Analog Output Connections (ALL Conditioners) ............................................... 2.16
(cont’d)

vii
CONTENTS

17(a) Logic I/O Connections: Input from External Switch .......................................... 2.17
17(b) Logic I/O Connections: External TTL Logic ......................................................... 2.17
17(c) Logic I/O Connections: External Controller ......................................................... 2.17
17(d) Logic I/O Connections: Output to External Relay .............................................. 2.18
18 Limit Zones ...................................................................................................................... 3.15
19 High and Low Hysteresis Windows ........................................................................ 3.16
20 Symmetry and Phase Adjustment Controls for the Model 3578 ................... 4.4
21 Typical Linearization Curve with Seven Segments .......................................... 4.12
22 Capture and Hold of Successively Higher-Valued Maxima ............................ 5.4
23 Capture and Hold of Successively Lower-Valued Maxima Using
Peak Reset ......................................................................................................................... 5.4
24 Tare Offset Operation .................................................................................................... 5.5
25 3500 Physical Dimensions .......................................................................................... A.1

viii
INTRODUCTION 1
1.a USING THIS MANUAL
THIS MANUAL TREATS ALL SIX STANDARD MODELS BELONGING TO THE
DAYTRONIC 3500 INSTRUMENT SERIES. THESE MODELS ARE AS FOLLOWS:
• the Model 3510 Thermocouple Conditioner
• the Model 3530 AC LVDT Conditioner
• the Model 3540 Frequency Input Conditioner
• the Model 3560 Voltage Conditioner
• the Model 3570 DC Strain Gage Conditioner
• the Model 3578 AC Strain Gage Conditioner

EXCEPT WHERE OTHERWISE STATED, THE INFORMATION IN THIS MANUAL


APPLIES EQUALLY TO ALL SIX MODELS.

DESCRIPTIONS, PROCEDURES, AND OPERATIONS SPECIFIC TO A GIVEN


MODEL OR MODELS WILL BE CLEARLY IDENTIFIED AS SUCH.

1.b GENERAL INSTRUMENT DESCRIPTIONS


1.b.1 INTRODUCTION
Every 3500 Series instrument can be completely set up and operated either
through the front-panel keypad or via simple mnemonic commands received
from an external computer or terminal through its RS-232/485 Interface Port.
The instrument can transmit data from this port in response to an interrogation
from an external computer, or can send it to an RS-232 serial printer when the
PRINT button is pressed (assuming that it is set to RS-232 mode). When in RS-
485 mode, a given 3500 instrument can represent one of up to 99 data-collection
"nodes" of a high-speed multidrop network.

Features common to all models include


• continuous dual-limit monitoring with front-panel annunciation, programmable
hysteresis windows, and TTL-level logic control outputs
• front-panel security code
• real-time positive peak capture or track/hold operation
• user-settable automatic tare offset
• internal 15-segment linearization for calibration of nonlinear inputs (except for
the Model 3510 Thermocouple Conditioner)
• selectable digital filtering
• selectable analog filtering (except for the Model 3510 Thermocouple Condi-
tioner)
• scalable analog output to drive strip-chart recorders or other devices (it may
also be used as feedback for a PID control loop)
• data-transmission formatting options, including "header" and "tailer" character
strings, node-number "echo," and limit-status indication

1.a USING THIS MANUAL 1.b GENERAL INSTRUMENT DESCRIPTIONS 1.1


1 INTRODUCTION

• logic-input control of peak capture, unlatching of latched limit conditions,


application of tare offset, and initiation of hard-copy printout

For complete specifications, see Appendix A.

1.b.2 THE MODEL 3510 THERMOCOUPLE CONDITIONER


This instrument accepts a temperature signal from a Type B, E, J, K, R, S, or T
Thermocouple. The sensor may be grounded or ungrounded. The conditioner
employs "absolute" calibration, which means that no calibration is required by the
user, once the proper "TC Type" is entered. During operation, appropriate refer-
ence-junction compensation, digital linearization, and engineering-unit scaling are
automatically applied. Detection of "open" thermocouples is also provided.

The 3510’s special rear connector assembly contains a thermistor for precise
measurement of the reference-junction temperature. Therefore, no external cold
junction is required—although the user may supply his own Controlled Ambient
Temperature Zone for reference-junction purposes, if desired.

1.b.3 THE MODEL 3530 LVDT CONDITIONER


This instrument measures displacement, force, pressure, and other parameters
obtained with a 5- or 7-wire linear variable differential transformer (LVDT)
capable of 3280-Hz operation and having primary impedance of 80 Ω or greater.
It can also be used with a 3- or 5-wire variable reluctance transducer. The
"normal" input range can be 0-150, 0-300, or 0-600 mV/V, full scale. Input provi-
sions also exist for "long-stroke" LVDT's (0-1, 0-2, or 0-4 V/V, full scale). Nominal 3
V-AC (rms) excitation is supplied.

1.b.4 THE MODEL 3540 FREQUENCY INPUT CONDITIONER


This instrument is used for measurement of flow, rpm, and other phenomena that
can be sensed by pulse transformer transducers with two-wire isolated windings
(tachometer pickups, turbine flowmeters, etc.), transistor or logic-circuit drivers,
"zero-velocity" (true digital output) sensors, and similar frequency-generating
transducers. It will accept any AC or unipolar pulse signal, floating or grounded,
irrespective of waveform. Input range is from 10% to 100% of 250, 500, 1000,
2000, 4000, 8000, 16000, or 32000 Hz.

The 3540’s analog-input threshold level is selectable to accommodate signals


from 100 mV to 100 V, thus guaranteeing reliable triggering when the input is at
the low end of the frequency range. Capacitive coupling of 0.1 µF is provided for
low-frequency inputs, to eliminate false triggering by signal noise or any DC offset
that exists for the frequency signal. For "zero-velocity" sensors, an excitation of
nominal ±5 V-DC ± 5% is supplied.

When you know the manufacturer-supplied full-scale rating of the frequency


source (or the highest frequency expected to be measured), the Model 3540’s
measurement channel can be quickly calibrated by issuing an "FRQ" command
through the RS-232/485 Interface Port.

1.2 1.b GENERAL INSTRUMENT DESCRIPTIONS


INTRODUCTION 1
1.b.5 THE MODEL 3560 VOLTAGE CONDITIONER
This is a general-purpose instrument for conditioning, displaying, and monitoring
the signal received from a DC-to-DC LVDT, potentiometer-type sensor, or other
external two-wire voltage source, either floating (differential) or grounded (sin-
gle-ended). The input signal may also represent output from some other instru-
ment system. Allowable full-scale voltage ranges are ±0.5, ±1.0, ±2.0, ±5.0, ±10.0,
and ±20.0 V-DC (up to ±100 V without damage). A ±12-V excitation is supplied for
sources that require it.

1.b.6 THE MODEL 3570 DC STRAIN GAGE CONDITIONER


This is a general-purpose instrument for input of pressure, force, torque, weight,
and other variables measured by conventional DC-excited strain gage trans-
ducers. It accepts a single input from any conventional 4-arm strain gage bridge,
nominal 120 ohms or higher, with a full-scale range of 0.75, 1.5, or 3.0 mV/V.
(User-supplied bridge-completion circuitry allows input from a 2-wire 1/4-bridge,
3-wire 1/4-bridge, or 1/2-bridge gage configuration.) The user may select a nomi-
nal excitation level of 2, 5, or 10 V-DC. Remote sensing of bridge voltage yields
consistently stable ratiometric measurement, unaffected by possible power-sup-
ply drift.

Simple two-point "zero and span" calibration is provided for the 3570’s input
channel. In addition, a 100-kΩ, 0.1% shunt resistor is supplied. You can use this
resistor—or one of your own—to apply an "equivalent input" for calibration purpos-
es, when the transducer's full-scale mV/V sensitivity is accurately known. The cal-
ibration shunt may be switched in and out for either a positive or negative
up-scale reading via simple commands issued to the RS-232/485 port or by
means of logic-level command signals through the rear Analog Input Connector.

A third calibration technique for the Model 3570 involves application of an "MVV"
command through the RS-232/485 Interface Port when both "mV/V" sensitivity
and corresponding full-scale rating of the transducer are known.

1.b.7 THE MODEL 3578 AC STRAIN GAGE CONDITIONER


This instrument is similar to the Model 3570 DC Strain Gage Conditioner, above,
but is of phase-sensitive carrier-amplifier design. Intended for applications involv-
ing transformer-coupling to the transducer bridge (as with rotary-transformer
torque sensors), it can also be used when high sensitivity is required or where
the electrical environment is especially noisy. Responding only to the modulated
carrier frequency, the 3578 rejects extraneous voltages that can cause errors in
DC systems, particularly when there is a need to "blow up" a portion of the trans-
ducer range.

The Model 3578 accepts input from a 4-arm bridge of nominal 90 ohms or higher,
and a full-scale range of 0.75, 1.50, or 3.00 mV/V. Excitation is fixed at 3 V-AC
(rms) at 3280 Hz. There are user-settable phase and symmetry controls. This
conditioner also offers the same shunt calibration provisions as the 3570, except
that shunt calibration of the AC Strain Gage Conditioner cannot be controlled via
logic-input commands.

1.b GENERAL INSTRUMENT DESCRIPTIONS 1.3


1 INTRODUCTION

1.c PHYSICAL LAYOUT


Study the following diagrams to acquaint yourself with the most important front
and rear elements.

Fig. 1(a) shows a typical "RUN-TIME" display, with "live" data and limit-status
annunciation. Note the four front-panel buttons that are active (but not necessari-
ly lit) during normal run-time operation. For a full description of run-time button
functions, see Appendix E.

Fig. 1(b) shows a typical SETUP display—specifically, the display that appears
after the security code has been entered and the unit is ready for any of the front-
panel setup procedures given in Sections 3.a and 4.a.

"Live" Data "Live" Limit


Fig. 1(a) Typical “Run-Time” Display Display Status Indicators

COM RANGE CAL FILTER ANO DEC LIMIT

HI

OK

LO

ENTER

Active Run-Time Buttons

"DEC" button lights when


Active Setup Setup message or decimal-point location of
Buttons are lit parameter value setup parameter can be
changed

COM RANGE CAL FILTER ANO DEC LIMIT

HI

OK

LO

PEAK TARE
SET UP PRINT ENTER
TRACK RESET

Fig. 1(b) Typical “Setup” Display Limit Status Indicators


updated with every change
of the display
1.4 1.c PHYSICAL L AYOUT
INTRODUCTION 1
NOTE: Your instrument is supplied with a large assortment of standard engineer-
ing unit legends on a 4" x 5 1/2" dry transfer sheet. The selected legend may be
rubbed directly onto the instrument's front-panel frame using the tip of a ball-point
pen or the blunt end of a stylus or other burnishing tool. DO NOT PRESS TOO
HARD. You can easily make up your own legends, since the sheet includes indi-
vidual numerals, upper- and lower-case letters, ampersand ("&"), and Greek "∆,"
"ø," "µ," and "π."

The two rear panel types are shown below. The Model 3510 Thermocouple
Conditioner and the Model 3578 AC Strain Gage Conditioner have the rear
panel illustrated in Fig. 2(a). Each of these instruments employs a special CONDI-
TIONER CONNECTOR that attaches directly to the rear edge of the unit's internal
Analog Input Board. In addition, the 3578 (only) has rear-panel Symmetry and
Phase Controls. All other models use a standard "clip-on" Analog Input Connec-
tor like that shown in Fig. 2(b).

Analog Output
Programming Pins*
(behind cover plate—see Fig. 15)
Conditioner Connector
Analog Output Symmetry and Logic I/O —attaches to Analog
Connector Phase Controls* Connector Input Board
(see Fig. 16) (see Fig. 20) (see Figs. 5, 17) (see Fig. 6 or 11)
+

AAA
A
AAA
AAAA
A
AAA
Panel-
ON AAA
AA
AAA
AAAAA
AA
Mount
Clamp OFF
Screw

Fuse RS-232/485 Inter-


ON-OFF AC Power (0.5-amp SLO- face Connector
Switch Connector BLO) (see Figs. 12 - 14)

* Not present on the Thermocouple Conditioner.

Fig. 2(a) Rear Panel for the Model 3510 Thermocouple


Conditioner and the Model 3578 AC Strain Gage Conditioner

1.c PHYSICAL LAYOUT 1.5


1 INTRODUCTION
Fig. 2(b) Rear Panel for All
Analog Output
Programming Pins* Other 3500 Series Models
(behind cover plate—see Fig. 15)
Analog Input
Analog Output Logic I/O Connector**
Connector Connector (see Fig. 7, 8, 9,
(see Fig. 16) (see Figs. 5, 17) or 10)
+

+ +

ON

AA
AA
Panel-
Mount
Clamp OFF
Screw

Fuse RS-232/485 Inter-


ON-OFF AC Power (0.5-amp SLO- face Connector
Switch Connector BLO) (see Figs. 12 - 14)

* Not present on the Frequency Conditioner.


** The number of connector terminals will vary with the conditioner type.

1.d PANEL MOUNTING


You can easily mount the instrument in your own precut panel. Cutout dimen-
sions for a panel-mounted unit are standard DIN (see Fig. 3); panel thickness
should not exceed 6 mm (0.24 in).

Simply unscrew the two rear-panel CLAMP SCREWS and slide the CLAMP
SLIDES rearwards out of their grooves (THE FRONT BEZEL NEED NOT BE
REMOVED). Insert the unit through the panel cutout, from the front of the panel (if
the unit has rubber feet, these will have to be removed). Then reinstall the
CLAMP SLIDES, and tighten the CLAMP SCREWS until the instrument is securely

AA
mounted.

CLAMP
SCREW

CLAMP SLIDE

Fig. 3 Panel Mounting

68 ± 0.7 mm
(2.68 ± 0.03 in) 138 ± 1.0 mm
(5.43 ± 0.04 in)

1.6 1.d PANEL MOUNTING


INTRODUCTION 1
1.e SUMMARY OF SETUP BUTTON FUNCTIONS
The use of the SETUP buttons is explained in detail in Sections 3.a and 4.a. The
following table summarizes the relevant functions:

Button Button Function (in SETUP MODE):

Used to exit current setup procedure ("COM," "RANGE," "CAL,"


SET UP
etc.) or to exit SETUP MODE.

TARE Used to enter a value into the TARE REGISTER.


RESET

Used to enter desired PRINT AND OUTPUT PARAMETERS:


PRINT
printing interval; node-number "echo" (ON/OFF); and limit-sta-
tus indication (ON/OFF).

Used to enter desired COMMUNICATIONS PARAMETERS:


COM
baud rate; number of data bits; number of stop bits; parity;
node number; INPUT TERMINATOR character; and OUTPUT
TERMINATOR character(s).

Used to enter desired INPUT RANGE or SCALE, plus other


RANGE
parameters, where applicable (EXCITATION, SENSITIVITY, etc.)

Used for CALIBRATION of the analog input ("Two-Point," "Lin-


CAL
earization," or "Calculated").

Used to set the ANALOG and DIGITAL FILTERS.


FILTER

Used to scale the ANALOG OUTPUT.


ANO

DEC Used to change DECIMAL-POINT LOCATION for certain setup


values.

Used to enter LIMIT PARAMETERS: high limit; high hysteresis;


LIMIT
high latch (ON/OFF); low limit; low hysteresis; low latch
(ON/OFF).

ENTER
Used to display the "existing value" of a setup parameter and
to finalize entry of a modified value.

In addition to the above SETUP buttons, the instrument provides two NUMERIC
BUTTONS for each displayed character of the LCD display, plus a polarity "sign"
button. These buttons let you change the value of a displayed number or step
forwards or backwards through a displayed series of allowed setup values. In
general, to increase by "1" the numeric value of a displayed digit (up to a maxi-
mum of "9"), press the TOP LCD SEGMENT of that digit, whether or not it is lit. To
decrease by "1" the numeric value of a displayed digit (down to a minimum of "0"),
press the BOTTOM LCD SEGMENT of that digit, whether or not it is lit. To change
the polarity of the displayed number, press the "minus-sign" segment at the
extreme left of the display (whether or not it is lit). The NUMERIC BUTTONS are
only active when the instrument is in SETUP MODE.

1.e SUMMARY OF SETUP BUTTON FUNCTIONS 1.7


1 INTRODUCTION
Press top segment to
increment digit (up to "9")
Fig. 4 Use of
NUMERIC
BUTTONS in
SETUP MODE

COM RAN FILTER ANO DEC LIMIT

HI

OK

LO

Press "–" SET UP


TARE
PRINT ENTER
RESET
segment to
change polarity

Press bottom segment to


decrement digit (down to "0")

1.f SUMMARY OF LOGIC I/O FUNCTIONS


The rear-panel LOGIC I/O CONNECTOR provides seven active logic input/output
bits in open-collector, negative-true form, where the "Logic 1" state is defined as
nominal 0 V-DC and "Logic 0" as nominal +5 V-DC. The standard logic I/O config-
uration is shown in Fig. 5(a), below, with reference to the 10-terminal Logic I/O
Connector on the rear of the unit. For recommended logic interconnections, see
Section 2.e. Each I/O function is fully described in Appendix F.

+
LOGIC LOW
TO ENABLE

* Logic Inputs:
UNLATCH
Fig. 5(a)
PEAK
Standard Logic
Inputs and Outputs
TARE

PRINT
* Ground connections to
be provided by user.
See also Fig. 17(a).
Logic Outputs:
"LIVE" DATA IN
"OK" ZONE
"LIVE" DATA IN
"LESS THAN" ZONE
"LIVE" DATA IN
"GREATER THAN" ZONE

PLEASE NOTE: THE MINIMUM TIME ALLOWED BETWEEN ACTIVATION AND


REACTIVATION OF ANY OF THE FOUR LOGIC CONTROL INPUTS IS 100 MIL-
LISECONDS.

1.8 1.f SUMMARY OF LOGIC I/O FUNCTIONS


INTRODUCTION 1
You can use the three logic control outputs to actuate solenoid valves, illuminate
panel displays, sound alarms, start and stop motors or pumps, initiate and control
safety shut-down sequences, and perform many other automation tasks that
require "intelligent" switching, even of substantial amounts of power.

In addition to the standard logic inputs and outputs provided by the LOGIC I/O
CONNECTOR, the Model 3570 DC Strain Gage Conditioner (ONLY) also
accepts two negative-true logic inputs at its rear-panel ANALOG INPUT CONNEC-
TOR. These two inputs, shown in Fig. 5(b), let the operator directly control the
SHUNT CALIBRATION process, which is explained in detail in Section 4. Again,
see Appendix F for a full description of these logic functions.

ANALOG INPUT CONNECTOR


LOGIC LOW
SHUNT CAL TO ENABLE
CONTROL
PWR – SIG
SHLD – + COM SIG COM
Fig. 5(b)
Shunt-Control Logic
Inputs for the Model
3570 DC Strain Gage
Conditioner (ONLY)
* Ground Logic Inputs:
connections to
be provided by *
user. See also + CALIBRATE
Fig. 17(a).
– CALIBRATE

1.g MNEMONIC COMMANDS


There are two ways to issue commands to a 3500 Series instrument. One way is
through the front-panel push buttons. These buttons allow you
• to enter all necessary SETUP COMMANDS, as explained in Sections 3.a and
4.a*; and also
• to enter RUN-TIME COMMANDS for control of positive peak capture, appli-
cation of tare offset, and initiation and halting of hard-copy transmissions (see
Sections 5.b, 5.c, and 5.d).

The second way to issue commands to the instrument is via the RS-232/485
Computer/Network Communications Interface. When set to the RS-232 (SINGLE-
NODE) mode, this interface will accept commands from a connected computer,
terminal, or other RS-232-C device. When the unit is used in the RS-485 (MULTIN-
ODE) mode, commands will normally originate from an application program in the
network's supervisory computer.

In either RS-232 or RS-485 mode, commands to the instrument must be transmit-


ted in a standard ASCII Command Syntax. This syntax uses simple three-letter

* As explained in Section 3.a.2, if a nonzero SECURITY CODE has been specified, the operator will
have to enter that code before any SETUP COMMANDS can be applied via the front-panel but-
tons.

1.g MNEMONIC COMMANDS 1.9


1 INTRODUCTION

English mnemonics, and includes SETUP ("WRITE"), INTERROGATION ("READ"),


and TRANSMISSION-INITIATING commands. In RS-485 (MULTINODE) operation,
every command received at the RS-232/485 Interface Port will evoke a response
from that port (either "ACKNOWLEDGED," "NOT ACKNOWLEDGED," or the re-
quested PARAMETER or DATA VALUE(S)). For a complete listing of mnemonic
commands and responses, see Appendix B.

1.10 1.g MNEMONIC COMMANDS


SETUP: CONNECTIONS AND POWERUP 2
2.a TRANSDUCER CONNECTIONS
2.a.1 THE “STANDARD” ANALOG INPUT CONNECTOR
All 3500 Series instruments except the Model 3510 Thermocouple Conditioner
and the Model 3578 AC Strain Gage Conditioner use the "standard" Analog
Input Connector. Shown in Fig. 2(b), this connector is located on the rear of the
unit. The number of terminals and the specific terminal assignments will depend
on the model itself.

Referring to the appropriate cabling diagram below, connect the wires of your
transducer cable to the corresponding screw terminals of the Analog Input Con-
nector. To facilitate cable connection, the front (screw-terminal) portion of the
connector may be removed from the rear (pin) portion, which is mounted on the
internal Analog Input Board. Press hard when reinserting the front portion, to
make sure it is fully engaged (the small clips should snap into place on the rear
portion).

The special CONDITIONER CONNECTORS used by the Model 3510 Thermocou-


ple Conditioner and the Model 3578 AC Strain Gage Conditioner are described in
the respective sections below.

PLEASE NOTE: CABLE SIGNAL WIRES OR TWISTED WIRE PAIRS SHOULD


ALWAYS BE PROPERLY SHIELDED, AS INDICATED IN THE CABLING DIAGRAMS.
THIS WILL MINIMIZE THE PRODUCTION OF UNWANTED ELECTRICAL NOISE
FROM CAPACITIVE AND INDUCTIVE EFFECTS.

2.a.2 CONNECTING A THERMOCOUPLE TO THE MODEL 3510


The Model 3510 Thermocouple Conditioner's rear Analog Input Board mates
with a special CONDITIONER CONNECTOR (shown in Fig. 2(a) and in Fig. 6,
below), which provides a precision thermistor for reference-junction compensa-
tion. This connector contains a screw-terminal pair labelled "+" and "–." The other
two terminals are not used.

Each TC lead should be directly attached to its corresponding screw terminal (it
should never be soldered). The connector itself is "keyed" by a small plastic
insert embedded between a certain terminal-pin pair, which matches a slot in the
rear Analog Input Board. This prevents the connector from being inadvertently
attached upside-down.

Open the connector housing by removing the four screws that hold it together
(two on each side). Be sure to put back the insulating foam block before
reassembling the connector. It's also a good idea to wrap each cable wire
around the respective strain-relief post.

The "shield" wire of the transducer cable should be soldered to the exposed ter-
minal of the L-shaped GROUND LUG located under the head of one of the con-
nector's two captive screws. This will ensure direct shield contact with the 3510’s
metal case.

Open Thermocouple Detection—In the event of a broken thermocouple wire or


other "open TC" condition, the 3510 will automatically report an indeterminate off-
scale reading—that is, a value well outside the normal range of the TC type for
which it has been set.

2.a TRANSDUCER CONNECTIONS: MODEL 3510 2.1


2 SETUP: CONNECTIONS AND POWERUP
Open TC Detection
Programming Jumpers
Fig. 6 Model 3510
Transducer Cabling
To
This terminal + + Thermocouple
NOT USED Conditioner
Analog Input
LOW HI Board

– SIGNAL

+ SIGNAL
– + This terminal
NOT USED
+ +
Strain
Relief
Post
Ground
SHIELD Lug

The Model 3510 is normally preset at the factory for positive off-scale "open TC"
indication. However, you may easily reset it for negative indication. Open the
connector housing and locate the "Open TC Detection Programming Jumpers"
(see Fig. 6). You will have to remove the solder drop connecting the middle termi-
nal pad to the "HI" (positive) pad, and to place a solder drop between the middle
pad and the "LOW" (negative) pad. Use a fine-point solder gun to heat the solder
drop to be removed, until it has melted sufficiently for you to wipe it off with a
clean rag. Make sure you remove all traces of solder from the jumper pads you
wish to disconnect.

2.a.3 CONNECTING AN LVDT OR VARIABLE


RELUCTANCE TRANSDUCER TO THE MODEL 3530
With regard to transducer cabling for the Model 3530 LVDT Conditioner, please
note the following:

a. 5-wire LVDT cabling (Fig. 7(a)) or 3-wire variable reluctance transducer


cabling (Fig. 7(c)) is to be used when the cable is under 20 feet in length. In
this case, the +SENSE and –SENSE lines are tied to the corresponding EXCI-
TATION lines at the CONDITIONER CONNECTOR.

7-wire LVDT cabling (Fig. 7(b)) or 5-wire variable reluctance transducer


cabling (Fig. 7(d)) is to be used when the cable is 20 feet or longer. In this
case, the +SENSE and –SENSE lines are tied to the corresponding EXCITA-
TION lines at the transducer.

b. When wiring an LVDT transducer to the 3530, you should connect both series-
opposed secondary coils to the terminal labelled "CENTER WIRE," as shown
in Figs. 7(a) and 7(b).

c. NOTE THAT THERE ARE SPECIAL +SIGNAL AND –SIGNAL CONNECTIONS


FOR USE WITH LONG-STROKE LVDT'S (FULL-SCALE INPUT OF 0-1, 0-2, OR
0-4 VOLTS/VOLT).

Thus, to allow for the larger input voltages produced by such a sensor, you
would connect its +SIGNAL line to the terminal labelled "HI +SIG," instead of to
the "+SIG" terminal. Similarly, you would connect the –SIGNAL line to the ter-
minal labelled "HI –SIG" instead of to the "–SIG" terminal.

2.2 2.a TRANSDUCER CONNECTIONS: MODEL 3530


SETUP: CONNECTIONS AND POWERUP 2
d. When wiring a variable reluctance transducer to the 3530, you must install a
10-kilohm "half-bridge completion" resistor between the –SIGNAL line and
each of the two EXCITATION lines, as shown in Figs. 7(c) and 7(d).

LVDT INPUT

HI – HI + – CNTR + – – + +
SHLD N/C N/C SIG SIG SIG WIRE SIG EX SENS SENS EX

–SENSE +SENSE

SHIELD
Fig. 7(a) Model 3530 –EXCITATION
For
Transducer Cabling: Long-Stroke
5-Wire LVDT Cabling +SIGNAL
LVDT's
CENTER
(under 20 ft. in length) WIRE Sec. 1
PRIMARY
SECONDARY COIL
Sec. 2
COILS +EXCITATION

–SIGNAL
LVDT INPUT

HI – HI + – CNTR + – – + +
SHLD N/C N/C SIG SIG SIG WIRE SIG EX SENS SENS EX

SHIELD
–EXCITATION
For Fig. 7(b) Model 3530
Long-Stroke
LVDT's
+SIGNAL Transducer Cabling:
CENTER 7-Wire LVDT Cabling
WIRE Sec. 1
PRIMARY –SENSE (20 ft. or longer)
SECONDARY COIL
Sec. 2
COILS
+SENSE
–SIGNAL +EXCITATION

LVDT INPUT

HI – HI + – CNTR + – – + +
SHLD N/C N/C SIG SIG SIG WIRE SIG EX SENS SENS EX

–SIGNAL
–SENSE +SENSE
10K

10K

SHIELD
Fig. 7(c) Model 3530
Transducer Cabling: +EXCITATION

3-Wire Variable
Reluctance Cabling +SIGNAL
(under 20 ft. in length)
–EXCITATION

2.a TRANSDUCER CONNECTIONS: MODEL 3530 2.3


2 SETUP: CONNECTIONS AND POWERUP
LVDT INPUT

HI – HI + – CNTR + – – + +
SHLD N/C N/C SIG SIG SIG WIRE SIG EX SENS SENS EX

–SIGNAL
10K

10K

SHIELD
+SENSE
Fig. 7(d) Model 3530
Transducer Cabling: 5-Wire Variable +EXCITATION

Reluctance Cabling (20 ft. or longer)


+SIGNAL

–EXCITATION

–SENSE

2.a.4 CONNECTING A FREQUENCY


SOURCE TO THE MODEL 3540
Fig. 8(a) shows recommended cabling for connecting the Model 3540 Frequen-
cy Conditioner to a pulse transformer transducer with two-wire isolated wind-
ings (tachometer, turbine flowmeter, etc.). Fig. 8(b) shows the connection to a
transistor or logic-circuit driver, while Fig. 8(c) shows the connection to a "zero-
velocity" (i.e., true digital output) sensor. For a single-ended input, note that the
–SIG terminal should be tied to the SIG COM terminal, as shown in Fig. 8(b).

Note too that, when used with an open-collector type sensor—such as a zero-
velocity sensor—a pull-up resistor of typically 10 kilohms is required between the
terminal to which the +SIGNAL line is tied and the +5V terminal, as shown in Fig.
8(c). For a zero-velocity sensor, the +SIGNAL line is to be tied to the +SIGA termi-
nal to eliminate any DC offset (see below).

The Model 3540’s input channel is equipped with a capacitive-coupled input (0.1
µF). This special input may be used with either floating or grounded configura-
tions if you require elimination of DC offset or suppression of high-frequency
noise. Figs. 8(d) and 8(e) show how these effects can be achieved.

Elimination of DC Offset—The 0.1-µF capacitive coupling can be used to elimi-


nate any positive or negative DC offset that exists for the frequency signal. Sim-
ply connect the +SIGNAL line from the frequency source to the +SIGA terminal,
instead of to the normal +SIG terminal (see Fig. 8(d)). The capacitor is here in
series with the +SIGNAL input and allows only AC to pass.

Suppression of High-Frequency Noise—False triggering can sometimes occur,


especially at the low-frequency input range, because of stray pickup of frequen-
cies outside the common-mode range. Capacitive coupling of the frequency
input to ground can in such cases serve to suppress unwanted signal noise.
Thus, if you find your frequency reading to be unacceptably unstable or "noisy,"
you should tie the +SIGA terminal to the SIG COM terminal, while maintaining the
normal +SIGNAL connection to the +SIG terminal (see Fig. 8(e)). With reference to
the suppression of high-frequency noise by grounding +SIGA, note that
• it is always recommended for magnetic-pickup sensors; and
• in general, it is NOT to be used with transducers that produce
active output (e.g., TTL logic drivers).

2.4 2.a TRANSDUCER CONNECTIONS: MODEL 3540


SETUP: CONNECTIONS AND POWERUP 2
EXCITATION SIGNAL INPUTS EXCITATION SIGNAL INPUTS
+ PWR – + + SIG – + PWR – + + SIG –
SHLD SHLD
5V COM 5V SIGA SIG COM SIG 5V COM 5V SIGA SIG COM SIG

SHIELD SHIELD

+SIGNAL +SIGNAL

+ +

FREQUENCY FREQUENCY
SOURCE SOURCE

– –
–SIGNAL –SIGNAL

Fig. 8(a) Model 3540 Transducer Fig. 8(b) Model 3540 Transducer
Cabling: Differential (Floating) Cabling: Single-Ended (Grounded)
Frequency Input Frequency Input

EXCITATION SIGNAL INPUTS


+ PWR – + + SIG –
SHLD
5V COM 5V SIGA SIG COM SIG

10K

SHIELD
Fig. 8(c) Model 3540
10K Pull-up Resistor (for Transducer Cabling:
use with open-collector
sensor) Input from a Zero-
Velocity Sensor
+EXCITATION

+SIGNAL

–EXCITATION

ZERO-VELOCITY SENSOR

EXCITATION SIGNAL INPUTS EXCITATION SIGNAL INPUTS


+ PWR – + + SIG – + PWR – + + SIG –
SHLD SHLD
5V COM 5V SIGA SIG COM SIG 5V COM 5V SIGA SIG COM SIG

+SIGNAL +SIGNAL

Fig. 8(d) Model 3540 Transducer Fig. 8(e) Model 3540 Transducer
Cabling for Elimination of DC Offset Cabling for Suppression of High-
Frequency Noise

2.a TRANSDUCER CONNECTIONS: MODEL 3540 2.5


2 SETUP: CONNECTIONS AND POWERUP

2.a.5 CONNECTING A VOLTAGE


SOURCE TO THE MODEL 3560
Fig. 9(a) gives standard cabling for connecting to the Model 3560 Voltage Con-
ditioner a general ANALOG SIGNAL SOURCE, floating or grounded, with its own
power supply (if required); Fig. 9(b), for connecting an EXTERNAL ZERO-TO-FULL-
SCALE POTENTIOMETER with a resistance from 2 to 10 kilohms, using the instru-
ment's ±12-V excitation; and Fig. 9(c), for connecting an EXTERNAL DC-TO-DC
LVDT, again using the ±12-V excitation. As shown in Fig. 9(a), a floating input is to
be grounded by tying the transducer's –SIGNAL line to the SIG COM terminal.
When a –SIGNAL line from the transducer is not available, the –SIG and SIG COM
terminals should be connected by a jumper wire (as in Figs. 9(b) and 9(c)).

POWER SIGNAL
OUT INPUT POWER SIGNAL
± 40 MA MAX OUT INPUT
+ SIG – ± 40 MA MAX
SHLD N/C +12V COM –12V
SIG COM SIG
+ SIG –
SHLD N/C +12V COM –12V
SIG COM SIG

SHIELD
SHIELD +EXCITATION
+SIGNAL
Reg. Power
Supply + Add wire for
(if required) floating input
ANALOG +SIGNAL
SIGNAL 2K to
SOURCE 10K
– –EXCITATION

–SIGNAL
Fig. 9(b) Model 3560 Transducer
Cabling: External Potentiometer
Fig. 9(a) Model 3560 Transducer
Cabling: General Voltage Source

POWER SIGNAL
OUT INPUT
± 40 MA MAX
+ SIG –
SHLD N/C +12V COM –12V
SIG COM SIG

SHIELD +EXCITATION

+SIGNAL
DC-to-DC
LVDT –SIGNAL

Tie the –SIG and SIG COM


terminals if a –SIGNAL wire from
–EXCITATION the transducer is not available

Fig. 9(c) Model 3560 Transducer


Cabling: External DC-to-DC LVDT

2.6 2.a TRANSDUCER CONNECTIONS: MODEL 3560


SETUP: CONNECTIONS AND POWERUP 2
2.a.6 CONNECTING A DC STRAIN GAGE
TRANSDUCER TO THE MODEL 3570
Four-wire strain gage cabling (Fig. 10(a)) is to be used with the Model 3570 DC
Strain Gage Conditioner when the cable is under 20 feet in length. In this case,
the +SENSE and –SENSE lines are tied to the corresponding EXCITATION lines,
and also the CALIBRATION SENSE line to the +SIGNAL line, at the CONDITIONER
CONNECTOR.

Eight-wire strain gage cabling (Fig. 10(b)) is to used when the cable is 20 feet or
longer. In this case, the +SENSE and –SENSE lines are tied to the corresponding
EXCITATION lines, and the CALIBRATION SENSE line to the +SIGNAL line, at the
transducer. Note also the wire connected to the –SIGNAL line at the transducer,
but left unconnected at the instrument. This wire is to be paired with the CAL
SENSE line, as shown, for shielding purposes.

SHUNT CAL STRAIN GAGE INPUT


CONTROL
PWR – SIG + CAL – – + +
SHLD – + COM SIG COM SIG SENS EX SENS SENS EX

See Fig. 5(b) CAL –SENSE +SENSE


SENSE

SHIELD
Fig. 10(a) Model 3570
Transducer Cabling: +EXCITATION
4-Wire Cabling
(under 20 ft. in length)
+SIGNAL
–SIGNAL

–EXCITATION

SHUNT CAL STRAIN GAGE INPUT


CONTROL
PWR – SIG + CAL – – + +
SHLD – + COM SIG COM SIG SENS EX SENS SENS EX

See Fig. 5(b)

SHIELD

–SIGNAL

+SENSE
+EXCITATION

Fig. 10(b) Model 3570


Transducer Cabling:
+SIGNAL
8-Wire Cabling
CAL SENSE (20 ft. or longer)

–EXCITATION
–SENSE

UNCONNECTED WIRE
(PAIRED WITH "CAL SENSE")
2.a TRANSDUCER CONNECTIONS: MODEL 3570 2.7
2 SETUP: CONNECTIONS AND POWERUP

2.a.7 CONNECTING AN AC STRAIN GAGE


TRANSDUCER TO THE MODEL 3578
As shown in Fig. 2(a), the Model 3578 AC Strain Gage Conditioner's Analog
Input Board mates with a special CONDITIONER CONNECTOR. This connector
allows direct solder-terminal attachment of cable leads. The connector's internal
solder terminals are labelled 1 through 10 and A through L. The connector is
"keyed" by small plastic inserts embedded between certain terminal-pin pairs,
each of which matches a slot in the conditioner's Analog Input Board. This pre-
vents the connector from being inadvertently attached upside-down.

Open the connector housing by removing the four screws that hold it together.
Secure the cable by means of one of the internal clamp bars.

The "shield" wire of the transducer cable should be soldered to the exposed ter-
minal of the L-shaped GROUND LUG located under the head of one of the con-
nector's two captive screws. This will ensure direct shield contact with the 3578’s
metal case.

CONDITIONER CONNECTOR

+EXCITATION
A 1

–SENSE B 2

C 3 +SENSE
–SIGNAL +SIGNAL
D 4

CAL SENSE E 5
F 6
–EXCITATION
H 7

J 8

K 9

Fig. 11(a) Model 3578 L 10

Transducer Cabling: 4-Wire


SHIELD
Cabling (under 20 ft. in length) Ground Lug
Connector pins shown as viewed
from rear (cable) side of
connector.

+SENSE CONDITIONER CONNECTOR

+EXCITATION

A 1

B 2
–SIGNAL +SIGNAL
C 3

D 4

CAL SENSE E 5
F 6
–EXCITATION
H 7
–SENSE J 8

K 9

L 10
Unconnected wire
(Paired with "CAL SENSE")
SHIELD
Ground Lug
Fig. 11(b) Model 3578 Connector pins shown as viewed
from rear (cable) side of
Transducer Cabling: 8-Wire connector.
Cabling (20 ft. or longer)

2.8 2.a TRANSDUCER CONNECTIONS: MODEL 3578


SETUP: CONNECTIONS AND POWERUP 2
CONDITIONER CONNECTOR

A 1

–SENSE B 2

C 3 +SENSE
+SIGNAL D 4

E 5
External 6
F
Fig. 11(c) Model 3578 SHUNT
CALIBRATION H 7
Transducer Cabling: RESISTOR
8
(User- J
Installation of User’s Supplied) K 9
External Shunt L 10
Calibration Resistor
SHIELD

+SENSE CONDITIONER CONNECTOR

+EXCITATION

A 1
B
B 2
–SIGNAL +SIGNAL
A D C 3

C
D 4
Fig. 11(d) Model 3578
E 5
Transducer Cabling: 8-Wire
F 6
–EXCITATION
7
Cabling to LEBOW 1600 SERIES
H
–SENSE J 8
TRANSDUCER (ONLY)
CAL LEBOW CAL K 9
(E) L 10

SHIELD
Unconnected wire Ground Lug
(Paired with "LEBOW CAL")
Connector pins shown as viewed
from rear (cable) side of
connector.

Four-wire strain gage cabling (Fig. 11(a)) is to be used when the cable is under 20
feet in length.

The Model 3578 is equipped with an internal 59-KΩ, 0.1% calibration resistor for
the standard "shunt" calibration technique described in Section 4. If you wish to
use your own external shunt resistor, it should be tied between Pin 5 of the CON-
DITIONER CONNECTOR and the transducer's +SIGNAL line, as shown in Fig.
11(c). In this case, "CAL SENSE" (Pin 4) is not used.

With cabling under 20 feet in length, the +SENSE and –SENSE lines are tied to the
corresponding EXCITATION lines at the CONDITIONER CONNECTOR. Also, the
CALIBRATION SENSE line from Pin 4—or the optional external SHUNT RESISTOR
line from Pin 5—is tied to the +SIGNAL line at the CONDITIONER CONNECTOR.

Eight-wire strain gage cabling (Fig. 11(b)) is to used when the cable is 20 feet or
longer.* As before, you can install your own shunt calibration resistor between
Pin 5 and the +SIGNAL line, in which case Pin 4 is not used.

With cabling of 20 feet or over, the +SENSE and –SENSE lines are tied to the cor-
responding EXCITATION lines at the transducer. Also, the CALIBRATION SENSE

* NOTE: This cabling is to be used when connecting the Model 3578 to a Lebow 1800 Series
Transducer, regardless of cable length.

2.a TRANSDUCER CONNECTIONS: MODEL 3578 2.9


2 SETUP: CONNECTIONS AND POWERUP

line from Pin 4—or the optional external SHUNT RESISTOR line from Pin 5—is tied
to the +SIGNAL line at the transducer. Note the wire connected to the –SIGNAL
line at the transducer, but left unconnected at the instrument. This wire is to be
paired with the CAL SENSE or SHUNT RESISTOR line for shielding purposes.

Special 8-wire cabling (shown in Fig. 11(d)) is required for connecting the 3578 to
a Lebow 1600 Series Transducer. The cable should be shielded in four pairs,
as shown in the figure, with the shield open at the transducer end. Also note that
• SENSE and EXCITATION lines should be tied at the transducer.
• The conditioner's Pin 5 ("LEBOW CAL") is to be connected to the "CAL" pin on
the Lebow sensor (Pin 4 is not used in this case).
• You should leave the last (extra) wire unconnected at both ends, and pair it
with the "LEBOW CAL" line for the fourth shield.
• THE MODEL 3578 MUST BE INTERNALLY SET TO "SIGNAL COMMON"
MODE. CONTACT THE FACTORY FOR PRECISE INSTRUCTIONS.

2.b RS-232 (“SINGLE-NODE”) CONNECTIONS


You should set the RS-232/485 Interface Port for RS-232 operation when you
want your 3500 Series instrument to communicate with a single computer, termi-
nal, buffered printer, or other RS-232-C device. To do so, you need only MAKE
SURE THAT THE INSTRUMENT'S CURRENT NODE NUMBER IS "0" (ZERO). You
will be shown in Section 3.a.3 how to set the node number and other necessary
RS-232 communications parameters via the front-panel buttons.

If you did not order a specific RS-232-C Interface Cable with your 3500 Series
instrument, you will have to provide your own connection. Fig. 12 shows suggest-
ed cabling between the instrument and a computer, terminal, printer, etc., that
uses a 25-Pin RS-232-C Connector. FOR MAXIMUM DATA-TRANSFER SPEED
AND ACCURACY, A "FULL HANDSHAKE" INTERCONNECTION IS GENERALLY
RECOMMENDED (Fig. 12(a)). However, cabling is also given for "INCOMING
HANDSHAKE ONLY" and "NO HANDSHAKE" situations (Figs. 12(b) and 12(c),
respectively). Following RS-232-C conventions, the device at each end of the
interface is seen as "DATA TERMINAL EQUIPMENT (DTE)."

Fig. 13 shows suggested cabling between a 3500 Series instrument and a com-
puter, terminal, printer, etc., that uses a 9-Pin D-Subminiature Connector for its RS-
232-C interface (such as an IBM PC/AT).

Please note that the cabling in Figs. 12 and 13 is by no means definitive. In all
cases, you should carefully study the literature accompanying the specific RS-
232-C device you wish to connect to your instrument, to determine the cable
arrangement that will create the "handshake" you need (if any). On some devices,
for example, the DATA TERMINAL READY (DTR) signal may have a different
name (such as NOT BUSY) and may even appear on a pin other than No. 20 or
No. 4.

2.10 2.b RS-232 (“SINGLE-NODE”) CONNECTIONS


SETUP: CONNECTIONS AND POWERUP 2
5 4 3 2 1 RS-232/485 Interface
Connector—Male
9 8 7 6 (see Fig. 2)
Fig. 12 Suggested RS-232-C
Interface Connections Pin No. Function
(to 25-Pin RS-232-C Connector) 1 DATA (RS-485)
2 RECEIVE (RS-232)
3 TRANSMIT (RS-232)
4 DATA TERMINAL READY (RS-232)
5 COMMON
6 DATA (RS-485)
7 +12 V (RS-485)
8 CLEAR TO SEND (RS-232)
9 SHIELD

3500 Series Computer


Instrument or other RS-232-C Device

Fig. 12(a) RECEIVE 2 2 TRANSMIT


TRANSMIT 3 3 RECEIVE
DTR 4 5 CTS
* 6 DSR
COMMON 5 7 COMMON
CTS 8 20 DTR

FULL HANDSHAKE 25-Pin RS-232-C


(RECOMMENDED) Connector
* Required for IBM and IBM-compatible computers.

3500 Series Computer


Instrument or other RS-232-C Device

RECEIVE 2 2 TRANSMIT Fig. 12(b)


TRANSMIT 3 3 RECEIVE
DTR 4 5 CTS
COMMON 5 7 COMMON
CTS 8 20 DTR

INCOMING HAND- 25-Pin RS-232-C


SHAKE ONLY Connector

3500 Series Computer


Instrument or other RS-232-C Device

Fig. 12(c) RECEIVE 2 2 TRANSMIT


TRANSMIT 3 3 RECEIVE
DTR 4 4 RTS
5 CTS
COMMON 5 7 COMMON
CTS 8 20 DTR

NO HANDSHAKE 25-Pin RS-232-C


Connector

2.b RS-232 (“SINGLE-NODE”) CONNECTIONS 2.11


2 SETUP: CONNECTIONS AND POWERUP
Computer
or other RS-232-C Device

RECEIVE 2 2 RECEIVE
TRANSMIT 3 3 TRANSMIT
DTR 4 4 DTR
COMMON 5 5 COMMON
CTS 8 6 DSR
8 CTS

9-Pin RS-232-C
Fig. 13 Suggested RS-232-C Interface
Connector
Connections (to 9-Pin RS-232-C Connector)

2.c RS-485 (“MULTINODE NETWORK”) CONNECTIONS


You can also set a 3500 Series instrument's RS-232/485 Interface Port for RS-485
intercommunications with a multidrop network of up to 99 independent Daytronic
signal conditioning instruments (3500 Series, 4000 Series with “N” Option, and/or
5000 Series)—all controlled by a supervisory computer with RS-232-C I/O. To do
so, YOU MUST ASSIGN THE 3500 INSTRUMENT A UNIQUE NONZERO NODE
NUMBER. You will be shown in Section 3.a.4 how to set the node number and
other necessary RS-485 communications parameters via the front-panel buttons.

IMPORTANT: BEFORE NETWORK INTERCONNECTIONS ARE ESTABLISHED,


YOU SHOULD SET UP EACH 3500 SERIES NETWORK "NODE" INDIVIDUALLY, BY
MEANS OF THE SETUP PROCEDURES GIVEN IN SECTIONS 3 AND 4.

While it is possible to make separate network "branches" issue from a single


node, such an arrangement can lead to less than optimum signal-to-noise ratio
because of unwanted reflections over interface lines. A strictly linear configura-
tion like that shown in Fig. 14(a) is therefore highly recommended.

For proper conversion of interface levels, you must attach a Model 5E485
RS232-to-RS485 Adaptor to the computer's RS-232-C port, via the Model 5E25
DB25 Male-to-Female Converter, as shown in Fig. 14(a).* The adaptor’s RS-
485 Interface Port should then connect directly to the RS-232/485 Interface Port
of the first network node—which should be a 3500 Series instrument—via the
cabling given in Fig. 14(b). Note that the +12-V supply of the first node (when it is
a 3500 instrument) is used to power the Model 5E485, as shown in Fig. 14(b).**
Fig. 14(c) shows the pin-to-pin cabling to be used between each pair of adjacent
3500 instrument nodes, or between a 3500 node and an adjacent 4000 Series
node. Interconnections between a 3500 or 4000 node and a separately powered
5000 node are shown in Fig. 14(d).

* The Model 5E25 is not necessary when the computer is equipped with a Model PC-HSICA
High-Speed Serial Interface Card. Also, you may use your own RS-232-to-RS-485 converter
in place of the Model 5E485, if desired. Converter connections will depend on whether there
are separate "XMIT" and "RCV" pairs or a single "485 DATA" pair. Contact the factory for
instructions.
** If the network contains one or more 5000 Series instrument nodes, a separate power source
of 10 to 40 V-DC (nominal 24 V-DC recommended) is required to power both the 5E485 and
the 5000 nodes—in which case the +12-V pin on every 3500 node should not be used. See any
5000 instrument instruction manual for details.

2.12 2.c RS-485 (“MULTINODE NETWORK”) CONNECTIONS


SETUP: CONNECTIONS AND POWERUP 2
Computer 25-Pin
RS-232-C Port
Model 5E485
Model 5E25 RS232-to-RS485
Male-to-Female Adaptor
Converter* 4-pin RS-485 Interface
& DC Power Port
Node RS-485 Interface
Port or Computer
Interface Port in RS-485
For cabling, Communications Mode
see Fig. 14(b)

Node 1
3500 Series
Instrument
For cabling,
see Fig. 14(c)

Node 2

* Not required with COMMON


Model PC-HSICA
High-Speed Serial SHIELD
Interface Card.
485 DATA
Node 3 0.0047 µf
485 DATA 100 Ω

RS-485 Terminator
(for chains over 500 ft.)

Fig. 14(a) Connections for a Network of Three Instru-


ment Nodes (where the first is a 3500 Series instrument)

For RS-485 communications, the relevant pins of the RS-232/485 Interface Port
are as follows:
Pin No. Function
1 485 DATA
5 COMMON (GND)
6 485 DATA
7 +12 V
9 SHIELD

RS-485 interconnections require Belden 8162 Datalene 100-Ω shielded cable (or
equivalent). THE INDICATED SHIELDING IS VERY IMPORTANT AND SHOULD BE
FOLLOWED CLOSELY.

For network chains of over 500 feet, the RS-232/485 Interface Port of the last
node in the sequence should be "terminated" by means of a 0.0047-µf capacitor
and a 100-Ω resistor across the two "DATA" terminals (again, see Fig. 14(a)).

2.c RS-485 (“MULTINODE NETWORK”) CONNECTIONS 2.13


2 SETUP: CONNECTIONS AND POWERUP
RS-485 Interface Port RS-232/485 Interface Port
(Model 5E485) (3500 Series Instrument)

+10 to 40 V +12 V
PWR COM COMMON
Shield
SHIELD
485 DATA 485 DATA
485 DATA 485 DATA

For RS-485, use Belden 8162 Datalene


100- Shielded Twisted Pairs or other
cable intended for EIA RS-422/485

Fig. 14(b) RS-485 Cabling Between the Model 5E485 and


the FIRST Network Node (if it is a 3500 Series instrument)

RS-232/485 Interface Port RS-232/485 Interface Port


(3500 or 4000 Series Instrument) (3500 or 4000 Series Instrument)

COMMON COMMON
Shield
SHIELD SHIELD
485 DATA 485 DATA
485 DATA 485 DATA

For RS-485, use Belden 8162 Datalene


100- Shielded Twisted Pairs or other
cable intended for EIA RS-422/485

Fig. 14(c) RS-485 Cabling Between


Successive 3500 or 4000 Instrument Nodes

RS-485 Interface Port


RS-232/485 Interface Port (5000 Series Instrument)
(3500 or 4000 Series Instrument)
+10 to 40 V
COMMON PWR COM
Shield
SHIELD
485 DATA 485 DATA
485 DATA 485 DATA

For RS-485, use Belden 8162


Datalene 100- Shielded
Twisted Pairs or other cable
intended for EIA RS-422/485
+
Power
See 5000 Instrument Supply
Instruction Manual –
Fig. 14(d) RS-485 Cabling Between a 3500 or
4000 Instrument Node and a 5000 Instrument Node

2.14 2.c RS-485 (“MULTINODE NETWORK”) CONNECTIONS


SETUP: CONNECTIONS AND POWERUP 2
2.d ANALOG OUTPUT CONNECTIONS
As indicated in Fig. 15, the ±5-V analog output signal of the Model 3530 LVDT
Conditioner, Model 3560 Voltage Conditioner, Model 3570 DC Strain Gage
Conditioner, or Model 3578 AC Strain Gage Conditioner may represent the
state of the conditioned input (A) after the FIXED ANALOG FILTER; (B) after analog
peak capture; or (C) after the SELECTABLE ANALOG FILTER (5/10/20 Hz for each
of these conditioners).* Each unit is initially set at the factory to source the analog
output after the FIXED ANALOG FILTER (Point A). To select a different source
point for any of these conditioners, you should
1. First turn OFF the unit and disconnect the power cord.
2. Remove the screw(s) holding the rear plate that covers the ANALOG OUTPUT
PROGRAMMING PINS (see Fig. 2(a) or 2(b)).**
3. Using needle-nose pliers, pull out the single Berg-Pin jumper and reposition it
on the pair of pins corresponding to the desired source (A, B, or C—again, see
Fig. 15). The pins labelled D are not currently used.
4. Replace the cover and reactivate the unit.

Analog Output Analog Output


Programming Pins
(see Fig. 2)
Berg-Pin Jumper
(connect one pair only) A B C D

Analog Input

+PEAK A/D
A B C
Fixed Analog Filter: Selectable Analog Filter:
Model 3530 LVDT Conditioner: 100 Hz 5/10/20 Hz
Model 3560 Voltage Conditioner: 2000 Hz
Model 3570 DC Strain Gage Conditioner: 2000 Hz
Model 3578 AC Strain Gage Conditioner: 20 Hz

This diagram does not apply to the Models


3510 and 3540, which do not have Analog
Output Programming Pins.

Fig. 15 Analog Output Programming Pins

* For the Model 3540 Frequency Conditioner, the SELECTABLE ANALOG FILTER is 2.5/5/10 Hz;
for the Model 3510 Thermocouple Conditioner, there is no SELECTABLE ANALOG FILTER. The
3510’s ANALOG OUTPUT is taken directly from the DIGITAL/ANALOG CONVERTER; the 3540’s
ANALOG OUTPUT is always taken from souce "C" in the figure.
** The Model 3578 AC Strain Gage Conditioner has only one screw; the others have two.

2.d ANALOG OUTPUT CONNECTIONS 2.15


2 SETUP: CONNECTIONS AND POWERUP

Fig. 16 shows how an external device connects to the ANALOG OUTPUT CON-
NECTOR on the rear of the conditioner, including the Models 3510 and 3540. The
output is single-ended, and returns to "SIGNAL COMMON" (i.e., GROUND).

For offsetting and scaling of the ANALOG OUTPUT, see Section 3.a.7. The fre-
quency characteristics and step-response settling time of the output will depend
on which pair of programming pins are jumpered. For details, see Appendix A.

ANALOG
PWR
GND OUT +5V COM Fig. 16 Analog Output
Connections
(ALL Conditioners)

Shield
This terminal not
currently used
Signal

EXTERNAL
DEVICE
Signal –
Common

2.e LOGIC INPUT/OUTPUT CONNECTIONS


The rear LOGIC I/O CONNECTOR is shown in Figs. 2 and 5(a). It has eight
labelled TTL/CMOS-compatible I/O terminals (one of which is normally unused),
plus a LOGIC REFERENCE terminal (+5 V) and a GROUND terminal (COM). In the
standard logic configuration described in Section 1.f and Appendix F, the first
three I/O terminals (left to right) are logic outputs, while the last four are logic
inputs. For the Model 3570 DC Strain Gage Conditioner (only), two logic inputs
("+" and "–") are also provided on the Analog Input Connector, for control of shunt
calibration—see Fig. 5(b). For complete logic-signal specifications, see Appendix
A.

Fig. 17 shows how to wire


a. negative-true logic INPUT to a given terminal from external switch contacts;
b. negative-true logic INPUT to a given terminal from an active TTL logic system
and open-collector logic OUTPUT from a given terminal to an active TTL logic
system;
c. negative-true logic INPUT to a given terminal from an INPUT CONTROL
BLOCK and open-collector logic OUTPUT from a given terminal to an isolated
POWER CONTROL BLOCK.
d. open-collector logic OUTPUT from a given terminal to drive an external relay
or TRIAC controller.

2.16 2.e LOGIC INPUT/OUTPUT CONNECTIONS


SETUP: CONNECTIONS AND POWERUP 2
Logic I/O Connector
Fig. 17(a) Logic I/O
Connections: Input
from External Switch COM
+5 V OPEN = LOGIC 0
CLOSED = LOGIC 1

...
Logic Input*

...
* Also includes "+" and "–" Calibration
Inputs for the Model 3570 DC Strain Gage
Conditioner (ONLY)—see Fig. 5(b).

Fig. 17(b) Logic I/O Logic I/O Connector


Connections:
External TTL Logic COM
+5 V
TTL or CMOS

...
+5
Logic Input*

...
+5
Logic Output
...

* Also includes "+" and "–" Calibration


Inputs for the Model 3570 DC Strain Gage
Conditioner (ONLY)—see Fig. 5(b).

Fig. 17(c) Logic I/O Logic I/O Connector


Connections:
External Controller COM
+5 V
NOTE THAT THE
...

POWER CONTROL
Input BLOCK IS ENERGIZED
+ – IN THE TRUE STATE
Logic Input* Control
Block
...

+ Power –
Control
Logic Output Block
...

* Also includes "+" and "–" Calibration


Inputs for the Model 3570 DC Strain Gage
Conditioner (ONLY)—see Fig. 5(b).
* Also includes "+" and "–" Calibration Inputs for the Model
3570 DC Strain Gage Conditioner (ONLY)—see Fig. 5(b).

2.e LOGIC INPUT/OUTPUT CONNECTIONS 2.17


2 SETUP: CONNECTIONS AND POWERUP

Logic I/O Connector


Fig. 17(d) Logic
I/O Connections:
Output to COM
External Relay
+5 V

...
Logic Output
Coil

...
Resistance
> 500

2.f POWERUP
Accepting a line voltage from 90 to 265 V-AC (50-400 Hz; 10 W max), your 3500
Series instrument will automatically sense the power input level and adjust its
internal regulator accordingly. THEREFORE, YOU NEED NOT SET THE INSTRU-
MENT FOR NOMINAL 110-V OR 220-V OPERATION.

IN THE EVENT OF AN APPARENT POWER-SUPPLY FAILURE, FIRST CHECK THE


0.5-AMP (250-V) BUSS FUSE LOCATED ON THE REAR PANEL (see Fig. 2). WHEN
REPLACING A "BLOWN" FUSE, ALWAYS INVESTIGATE THE CAUSE OF OVER-
LOAD BEFORE REACTIVATING THE INSTRUMENT.

To change the fuse, use a screwdriver to turn the fuse slot counterclockwise, and
the fuse holder will spring out.

The 6-foot, three-conductor power cord supplied with the unit plugs into the rear
AC POWER CONNECTOR. The offset pin on the power connector is ground.
THE INSTRUMENT MUST BE PROPERLY GROUNDED. To safely operate from
a two-contact outlet, use a 3-prong-to-2-prong adaptor and connect the green
pigtail on the adaptor to earth ground.

Since the presence of electrical noise can affect the ultimate integrity of your
data, the noise level should be suppressed as much as possible. In particular,
care should be taken to avoid utility-line problems that can interfere with or possi-
bly even damage sensitive microprocessor-based equipment. Such noise can
also be generated by electrical motors, relays, and motor control devices.

While your instrument has internal circuitry to protect it from overvoltage tran-
sients and mild EMI, a clean line is still very desirable. No protection is provided
against dropout longer than 8 milliseconds or brownout below 90 volts. Depend-
ing on your line conditions, a number of protective devices are available (isola-
tors, regulators, uninterruptible power supplies, etc.).

The ON/OFF switch is also located on the rear panel (see Fig. 2). When you
powerup the unit, you will observe the following sequence of events:

2.18 2.f POWERUP


SETUP: CONNECTIONS AND POWERUP 2
• The entire LCD (buttons and digital display) will be turned momentarily ON
and OFF.
• The number "88888" will be displayed.
• The display will show either EPASS or EFAIL, following a checksum per-
formed on the EEPROM where program parameters are stored. If EFAIL
appears, it means that there is a difference between the calculated checksum
and the stored checksum. IN THE EVENT OF AN "EFAIL" MESSAGE, CON-
TACT THE FACTORY CUSTOMER SERVICE DEPARTMENT FOR INSTRUC-
TIONS.
• The display will show either rPASS or rFAIL, following a modified "walking bit"
test performed on the RAM chip. If rFAIL appears, it probably means either
that a RAM write did not take or that a write to one location affected another
byte of the RAM. IN THE EVENT OF AN "rFAIL" MESSAGE, CONTACT THE
FACTORY CUSTOMER SERVICE DEPARTMENT FOR INSTRUCTIONS.
• The display will show either cPASS or cFAIL, following a checksum per-
formed on the code PROM. If cFAIL appears, it means that the calculated
checksum was different from the stored checksum. IN THE EVENT OF A
"cFAIL" MESSAGE, CONTACT THE FACTORY CUSTOMER SERVICE DEPART-
MENT FOR INSTRUCTIONS.
• The instrument model number will be displayed.
• The instrument software version will be displayed.
• "Live" numeric data will be displayed.

2.f POWERUP 2.19


2.20 This page intentionally blank.
SETUP: INSTRUMENT CONFIGURATION 3
3.a CONFIGURING THROUGH THE FRONT PANEL
3.a.1 ENTERING AND EXITING SETUP MODE
Your 3500 Series instrument's "setup configuration" can be fully established via
the front-panel push buttons—including the NUMERIC BUTTONS described in
Section 1.e. For configuration setup by means of MNEMONIC COMMANDS
issued to the instrument through its RS-232/485 Interface, see Section 3.b, below.

When you press


SET UP

you will first be asked for the current SECURITY CODE if a nonzero code has
already been specified. If it has not, you will be given a chance to enter one (see
below for the full procedure).

Once the correct nonzero security code has been entered (if called for) and you
have either changed the code or kept it as it is, the "Setup" display shown in Fig.
1(b) will appear. Note that the limit status indicators are no longer "live." They will
be updated, however, each time the LCD display changes.

Once in the "Setup" display, you can press any of the illuminated buttons and fol-
low the steps given below to define the pertinent setup parameter(s). There is no
necessary setup sequence.

To exit SETUP MODE while the "Setup" display (Fig. 1(b)) appears, just press the
SETUP button again. If no setup parameters have been changed since you last
entered SETUP MODE, you will return immediately to the RUN-TIME ("live" data)
MODE. If, however, one or more setup parameters have been changed, the new
parameter value(s) will now be automatically saved. The unit will tell you this by
displaying the word "SAVING" for a few seconds:

You will then be returned to RUN-TIME ("live" data) mode with the
new setup configuration fully in effect.

If you press SETUP while within a given setup-button sequence ("COM,"


"RANGE," "CAL," etc.), you will return to the "Setup" display (Fig. 1(b)). You can
now either choose another setup button or exit SETUP MODE altogether by
pressing SETUP once again. If you haven't changed any setup values since you
last entered SETUP MODE, you will immediately return to a "live" data display. If
you have made changes, these will be saved before you return to "live" data.

In general, keeping your finger on a given setup button (COM, RANGE, CAL, FIL-
TER, etc.) will cycle you through all pertinent setup parameters. When you come
to the one you want to view and edit, just press ENTER to see that parameter's
existing value.

REMEMBER: AFTER CHANGING THE DISPLAYED VALUE OF A GIVEN SETUP


PARAMETER, YOU MUST PRESS

ENTER

3.a CONFIGURING THROUGH THE FRONT PANEL 3.1


3 SETUP: INSTRUMENT CONFIGURATION

IN ORDER TO REGISTER THE CHANGE. IF YOU ONLY PRESS SETUP AFTER


CHANGING A SETUP VALUE, YOU WILL RETURN TO THE "SETUP" DISPLAY
WITHOUT HAVING ACTUALLY "ENTERED" THE CHANGE.

ALSO PLEASE NOTE: IF YOU ATTEMPT TO ENTER AN OUT-OF-RANGE OR OTH-


ERWISE UNACCEPTABLE VALUE FOR A GIVEN SETUP PARAMETER, THE UNIT
WILL DISPLAY AN ERROR NUMBER, SUCH AS

If this occurs, refer to the ERROR-CODE TABLE in Appendix C. Note that the
ERROR-NUMBER will remain on display until you press the ENTER or SETUP but-
ton.

3.a.2 SECURITY CODE


If a nonzero SECURITY CODE has not already been entered, the operator will be
given the chance to do so as soon as the SETUP key is pressed. The unit will
ask:

If you do not want to enter a nonzero code at this time, just press SETUP again
and the "Setup" display (Fig. 1(b)) will appear. To enter a new code,

a. Press
ENTER

The current security code will be displayed (INITIALLY, THIS IS "0" FOR "NO
CODE").

b. Use the NUMERIC BUTTONS as discussed in Section 1.e, above, to display


any integral number from 1 through 98.

WARNING!
BE SURE TO WRITE DOWN THE NUMBER YOU SELECT AS YOUR SECURITY
CODE, SO THAT YOU DON'T FORGET IT. ONCE YOU PRESS ENTER, THERE IS
NO WAY TO INTERROGATE THE 3500 INSTRUMENT FOR THIS NUMBER,
EXCEPT BY A SECURITY CODE (SCD) COMMAND APPLIED THROUGH THE RS-
232/485 INTERFACE PORT (see Section 3.b, below). NOTE, HOWEVER, THAT
THE PROPER COMMUNICATIONS PROTOCOL MUST ALREADY BE ESTAB-
LISHED IN ORDER FOR THE UNIT TO RECOGNIZE AND RESPOND TO THIS COM-
MAND. UNTIL SUCH COMMUNICATIONS HAVE BEEN SET UP, THE ONLY
ACCESS YOU WILL HAVE TO THE EFFECTIVE SECURITY CODE WILL BE YOUR
OWN MEMORY—AND WITHOUT THIS CODE, YOU ARE PROHIBITED FROM MAK-
ING ANY SETUP CHANGES VIA THE FRONT PANEL, INCLUDING COMMUNICA-
TIONS PROTOCOL!

c. Press ENTER once again. The "Setup" display (Fig. 1(b)) will appear. From
now on, you will have to enter the number that has just been specified in order
to enter SETUP MODE. Once in SETUP MODE, however, you can easily
change the existing code, as explained below.

3.2 3.a CONFIGURING THROUGH THE FRONT PANEL


SETUP: INSTRUMENT CONFIGURATION 3
If a nonzero SECURITY CODE is already in effect, the operator will be asked to
enter it whenever the SETUP key is pressed:

The procedure for entering the existing code is as follows:

d. Press
ENTER

A zero will be displayed.

e. Use the NUMERIC BUTTONS to display the current security code.

f. Press ENTER again. If an incorrect code has been entered, you will simply
return to the "live" data display. To re-enter the (correct) code, you must again
press SETUP, followed by ENTER.

g. If the correct code has been entered, you will be given a chance at this time to
change it. The display will show Sc=?. If you want to change the code, follow
Steps a through c, above. If you don't want to change the code, press the
SETUP key to invoke the "Setup" display.

3.a.3 RS-232 COMMUNICATIONS PARAMETERS: COM KEY


PLEASE NOTE: FOR PROPER DATA INTERCHANGE BETWEEN THE 3500
INSTRUMENT AND A SINGLE CONNECTED RS-232-C COMPUTER, TERMINAL,
PRINTER, OR OTHER DEVICE TO OCCUR,
• THE INSTRUMENT SHOULD BE SET TO A NODE NUMBER OF "0," AND
• THE INSTRUMENT’S RS-232/485 INTERFACE MUST BE SET TO CONFORM
EXACTLY WITH THE PROTOCOL STIPULATED BY THE CONNECTED DEVICE.

To set the RS-232-C protocol,

a. Once you are in SETUP MODE, press the button labelled

COM

The unit will display the word "BAUD":

b. To display the current BAUD RATE setting, press

ENTER

If you don't want to display the baud rate, press COM again and proceed to
Step e.

c. To change the baud rate, you can cycle forwards or backwards through the
allowed settings (300 through 153.6K) by pressing the top or bottom segment,
respectively, of any digit display (even if it is blank).

3.a CONFIGURING THROUGH THE FRONT PANEL 3.3


3 SETUP: INSTRUMENT CONFIGURATION

d. When the desired baud rate appears, press ENTER to return to the "BAUD"
display. Then press COM. The unit will display "DATA" (for "NUMBER OF
DATA BITS"):

e. To display the current data-bits setting, press ENTER. If you don't want to dis-
play the number of data bits, press COM again and proceed to Step h.

f. To change the data-bits setting, you can toggle between "7" and "8"—the only
allowable values—by pressing the top or bottom segment of the displayed
digit.

g. When the desired number of data bits appears, press ENTER and then COM.
The unit will display "STOP" (for "NUMBER OF STOP BITS"):

h. To display the current stop-bits setting, press ENTER. If you don't want to dis-
play the number of stop bits, press COM again and proceed to Step k.

i. To change the stop-bits setting, you can toggle between "1" and "2"—the only
allowable values—by pressing the top or bottom segment of the displayed
digit.

j. When the desired number of stop bits appears, press ENTER and then COM.
The unit will display "PAR" (for PARITY):

k. To display the current parity setting, press ENTER. If you don't want to display
the parity, press the COM button again and proceed to Step n.

l. To change the parity setting, you can cycle forwards or backwards through
the allowed settings by pressing the top or bottom segment, respectively, of
any digit display (even if it is blank). The four parity settings are

— where the total number of "1's" in each transmitted charac-


ter is ODD
— where the total number of "1's" in each transmitted charac-
ter is EVEN
— where the parity bit is always present, but always ZERO

— for NO parity bit

m. When the desired parity setting appears, press ENTER to return to the "PAR"
display. Then press COM. The unit will display "NODE" (for "NODE NUM-
BER"):

3.4 3.a CONFIGURING THROUGH THE FRONT PANEL


SETUP: INSTRUMENT CONFIGURATION 3
n. To display the current node-number setting, press ENTER. If you don't want
to display the node number, press COM again and proceed to Step q.

IMPORTANT: FOR RS-232-C COMMUNICATIONS, THE NODE NUMBER MUST


REMAIN AT THE INITIAL SETTING OF ZERO ("00"). For configuration of RS-485
(MULTINODE NETWORK) communications, see Section 3.a.4, below.

IMPORTANT: IF YOU HAVE SWITCHED THE INSTRUMENT FROM "RS-232" TO


"RS-485" MODE OR VICE VERSA—I.E., IF YOU HAVE CHANGED THE NODE NUM-
BER FROM "0" TO A NONZERO VALUE OR VICE VERSA—YOU WILL HAVE TO
RECYCLE INSTRUMENT POWER BEFORE THE NEW MODE SETTING CAN GO
INTO EFFECT.

o. If the node number is already "00," you should leave it there by pressing COM
and proceeding to Step q. If the current node number is not "0," you should
change it to zero by pressing the bottom segment of each nonzero digit until
that digit is "0."

p. When the node number is "0," press ENTER. Then press COM. The unit will
display the word "INPUT":

q. Press either COM or ENTER to display the word "TER" (for "TERMINATOR"):

The INPUT TERMINATOR—also called "COMMAND TERMINATOR (CMT)"—is


the single ASCII CONTROL CHARACTER from [00] through [1F] to be recog-
nized as the termination for all mnemonic commands received by the unit
through its RS-232/485 Interface Port (in either RS-232 or RS-485 mode). The
INPUT TERMINATOR cannot be the ASCII [Esc] character.

r. If you don't want to display the INPUT TERMINATOR, press the COM button
again and proceed to Step u. To display the current INPUT TERMINATOR,
press ENTER. The unit will show this parameter as a two-character hexadeci-
mal word in square brackets. For example, if an INPUT TERMINATOR of CAR-
RIAGE RETURN ([CR]) is in effect, the unit will display

s. To change the INPUT TERMINATOR, you can cycle forwards or backwards


through allowed values of each active digit by pressing its top or bottom seg-
ment, respectively. NOTE: The instrument will not accept an INPUT TERMINA-
TOR entry greater that "1F," even though it is possible to increment the most
significant digit up to "7."

t. When the desired INPUT TERMINATOR appears, press ENTER to return to


the "INPUT" display. Then press the COM button twice. The unit will display
the word "OUTPUT":

3.a CONFIGURING THROUGH THE FRONT PANEL 3.5


3 SETUP: INSTRUMENT CONFIGURATION

u. Press either COM or ENTER to display the word "TER" (for "TERMINATOR").
The OUTPUT TERMINATOR—also called "END-OF-TRANSMISSION TERMINA-
TOR (EOT)"—is a string of up to four ASCII CONTROL CHARACTERS from [00]
through [1F] to be transmitted at the end of each complete transmission from
the RS-232/485 Interface Port (in either RS-232 or RS-485 mode)—including
every response to a received MNEMONIC COMMAND (see Appendix B).

v. If you don't want to display the OUTPUT TERMINATOR, press the COM button
again and proceed to Step a', below. To display the first character of the OUT-
PUT TERMINATOR, press ENTER. The unit will display

w. Press ENTER again.* The unit will show a two-character hexadecimal word in
square brackets. For example, if the first character of the OUTPUT TERMINA-
TOR is currently LINE FEED ([LF]), the unit will display

x. To change the first character of the OUTPUT TERMINATOR, you can cycle for-
wards or backwards through allowed values of each active digit by pressing
its top or bottom segment, respectively. NOTE: The instrument will not accept
an OUTPUT TERMINATOR character entry greater that "1F," even though it is
possible to increment the most significant digit up to "7."**

y. When the desired character appears, press ENTER. The unit will display
Eot2. Press ENTER to display the second character of the OUTPUT TERMI-
NATOR. Change it as you did for the first character, and press ENTER. The
unit will display EOT3. Proceed as above to change the third and fourth char-
acters of the OUTPUT TERMINATOR, if required.

z. When the final character of the OUTPUT TERMINATOR has been entered, you
will return to the "OUTPUT" display. Press the COM button twice. The unit will
display the word "BAUD." It has now cycled through all the communications
parameters.

a'. Press SETUP to return to the "Setup" display, from which you can proceed to
enter more setup values or exit SETUP MODE altogether.

NOTE: IF YOU ONLY HAVE TO CHANGE ONE COMMUNICATIONS SETUP PARA-


METER, YOU CAN QUICKLY STEP THROUGH THE SERIES BY KEEPING YOUR
FINGER ON THE COM BUTTON UNTIL THE PARAMETER OF INTEREST
APPEARS. THEN RELEASE THE COM BUTTON AND PRESS ENTER TO DISPLAY
THAT PARAMETER'S CURRENT VALUE.

* Pressing the COM button at this point would step to the display of Eot2, etc.
** For the significance of an "EOT" character of [00], see the EOT command, Section 3.b.

3.6 3.a CONFIGURING THROUGH THE FRONT PANEL


SETUP: INSTRUMENT CONFIGURATION 3
3.a.4 RS-485 COMMUNICATIONS PARAMETERS: COM KEY
You can use the front-panel buttons to set up RS-485 communications for every
instrument "NODE" within a MULTINODE NETWORK, exactly as explained in Sec-
tion 3.a.3, above, for RS-232 linkage to a single device.

NOTE, HOWEVER, THE FOLLOWING IMPORTANT POINTS:

a. For the instrument to function properly in a multinode network, IT MUST BE


ASSIGNED A UNIQUE NODE NUMBER FROM 1 THROUGH 99. This number
will be used to identify this particular node within the network. A node number
of "0" is used (as above) ONLY when the instrument is communicating directly
with a single device using the RS-232-C interface standard. Node-number
assignments need not be in any particular order. Just make sure that no two
nodes have been given the same number.

b. EVERY NETWORK NODE MUST BE SET TO RECOGNIZE THE EXACT SERIAL


INTERFACE PROTOCOL STIPULATED BY THE SUPERVISORY COMPUTER
(BAUD RATE, NUMBER OF STOP BITS, NUMBER OF DATA BITS, AND PARITY).
For the setting of these communications parameters, see Section 3.a.3,
above.

c. When setting the BAUD RATE, keep in mind that the maximum speed at
which network interchanges can take place will depend on (1) the capacity of
your computer's input buffer; (2) the network setup configuration; and (3) the
efficiency of your computer's data-input routine.

Specifically, since there is no hardware handshake for RS-485 communica-


tions, it is necessary that the computer have an input buffer capable of receiv-
ing each entire answer at the desired baud rate, without interruption.

Refer to Appendix D for some guidelines on maximizing network speed. A


sample data-collection program is included in this appendix.

d. THE INPUT TERMINATOR (or "CMT") MUST BE THE SAME FOR ALL NET-
WORK NODES.

e. THE OUTPUT TERMINATOR (or "EOT") MUST BE THE SAME FOR ALL NET-
WORK NODES. IN ADDITION, IN ANY NETWORK OF MORE THAN ONE
NODE, THE LAST CHARACTER OF THE OUTPUT TERMINATOR MUST BE THE
SAME AS THE INPUT TERMINATOR.

f. The front-panel PRINT button and the automatic print function are disabled
when the unit is in RS-485 MODE. To send formatted data only once to a con-
nected printer, you can always use the CHANNEL (CHN) or DUMP (DMP)
command as explained in Section 5.a.

3.a CONFIGURING THROUGH THE FRONT PANEL 3.7


3 SETUP: INSTRUMENT CONFIGURATION

3.a.5 INPUT RANGE: RANGE KEY


Once you are in SETUP MODE, press the button labelled

RANGE

The unit will display the word "RANGE":

—unless it is a Model 3510 Thermocouple Conditioner, in which case it will dis-


play the word "TYPE":

NOW REFER TO THE SECTION BELOW THAT APPLIES TO YOUR SPECIFIC CON-
DITIONER MODEL.

3.a.5.a SETTING THE MODEL 3510’S TC T YPE AND SCALE

The Model 3510 Thermocouple Conditioner must be set to match the thermo-
couple type to which it is connected.

1. If, after pressing the RANGE button, you want to display the current TC Type
setting, press ENTER. Otherwise, press RANGE again and proceed to Step
4. If, for example, the conditioner is set for a "Type E" thermocouple, after you
press ENTER it will display

2. To change the TC Type setting, you can cycle forwards or backwards through
the allowed settings ("E," "J," "K," "R," "S," "T," and "B") by pressing the top or
bottom segment, respectively, of any digit display (even if it is blank).

3. When the desired TC Type appears, press ENTER. The unit will display the
word "SCALE":

4. If you want to display the current temperature scale setting, press ENTER.
Otherwise, press RANGE again and proceed to Step 7. If, for example, the
conditioner is set to read temperature measurements in degrees Fahrenheit,
when you press ENTER it will display

5. To change the temperature scale, you can cycle forwards or backwards


through the allowed settings ("degrees F" and "degrees C") by pressing the
top or bottom segment, respectively, of any digit display (even if it is blank).

3.8 3.a CONFIGURING THROUGH THE FRONT PANEL


SETUP: INSTRUMENT CONFIGURATION 3
6. When the desired temperature scale appears, press ENTER. The conditioner
will now tell you whether its internal reference-junction compensation is cur-
rently enabled. If it is, you may now disable it, if you intend to provide your
own Controlled Ambient Temperature Zone for reference-junction purposes.
(Note that if an external cold junction is being used, you must supply the
Model 3510 with an already 32°-compensated TC signal on copper wire.)

If the internal cold-junction compensation is in effect, the conditioner will dis-


play "CLD J":

If it is NOT presently in effect, "NO CJ" will be displayed:

7. To toggle between "CLD J" and "NO CJ," press the top or bottom segment of
any digit display (even if it is blank).

8. When the desired state appears, press ENTER to return to the "TYPE" display.
Then press SETUP to return to the "Setup" display, and proceed to Section
3.a.6.

3.a.5.b SETTING THE MODEL 3530’S LVDT INPUT RANGE

The Model 3530 LVDT Conditioner must be set to match the full-scale input
range of the displacement transducer.

1. After pressing the RANGE button, press ENTER. The number "5," "10," or
"20" will now appear, depending on the current input range setting. The fol-
lowing table shows how each number corresponds both to a "normal" input
range and to a "long-stroke" input range (for "normal" vs. "long-stroke" cabling,
see Section 2.a.3):
Range "Normal" "Long-Stroke"
Number Input Range (f.s.) Input Range (f.s.)
5 0 - 150 mV/V 0 - 1 V/V
10 0 - 300 mV/V 0 - 2 V/V
20 0 - 600 mV/V 0 - 4 V/V

2. To change the input range, you can cycle forwards or backwards through the
allowed settings by pressing the top or bottom segment, respectively, of any
digit display (even if it is blank).

3. When the desired range number appears, press ENTER to return to the
"RANGE" display. Then press SETUP to return to the "Setup" display, and pro-
ceed to Section 3.a.6.

3.a CONFIGURING THROUGH THE FRONT PANEL 3.9


3 SETUP: INSTRUMENT CONFIGURATION

3.a.5.c SETTING THE MODEL 3540’S FREQUENCY


INPUT RANGE AND SENSITIVITY

The Model 3540 Frequency Conditioner must be set to match the full-scale
input range of the frequency source.

1. If, after pressing the RANGE button, you want to display the current input-
range setting, press ENTER. Otherwise, press RANGE again and proceed to
Step 5. If you press ENTER, the unit will display a number representing the
current input-range setting (in Hz): "250," "500," "1000," "2000," "4000," "8000,"
"16000," or "32000."

2. To change the input range, you can cycle forwards or backwards through the
allowed numbers by pressing the top or bottom segment, respectively, of any
digit display (even if it is blank).

3. When the desired frequency range appears, press ENTER. The unit will dis-
play "SENST" (for "SENSITIVITY"):

The "input sensitivity" is the minimum amplitude of AC input that will trigger an
output—that is, the lowest-valued input signal for which a valid reading is
required.

4. If you want to display the current sensitivity setting, press ENTER. Otherwise,
press RANGE again and proceed to Step 6. If, for example, the conditioner is
set for a sensitivity of 1 V to 20 V, when you press ENTER it will display "1."

5. To change the input sensitivity, you can cycle forwards or backwards through
the allowed settings (".1" for .1 V to 5 V; "1" for 1 V to 20 V; and "10" for 10 V to
100 V) by pressing the top or bottom segment, respectively, of any digit dis-
play (even if it is blank).

6. When the desired sensitivity appears, press ENTER to return to the "RANGE"
display. Then press SETUP to return to the "Setup" display, and proceed to
Section 3.a.6.

3.a.5.d SETTING THE MODEL 3560’S VOLTAGE INPUT RANGE

The Model 3560 Voltage Conditioner must be set to match the full-scale input
range of the voltage source.

1. After pressing the RANGE button, press ENTER. The current range setting
will now appear. If, for example, the conditioner is set for a full-scale input of
±0.5 volts, the display will show

2. To change the input range, you can cycle forwards or backwards through the
allowed settings (.5, 1, 2, 5, 10, and 20 volts) by pressing the top or bottom
segment, respectively, of any digit display (even if it is blank).

3.10 3.a CONFIGURING THROUGH THE FRONT PANEL


SETUP: INSTRUMENT CONFIGURATION 3
3. When the desired voltage range appears, press ENTER to return to the
"RANGE" display. Then press SETUP to return to the "Setup" display, and pro-
ceed to Section 3.a.6.

3.a.5.e SETTING THE MODEL 3570’S DC STRAIN GAGE


INPUT RANGE AND EXCITATION

The Model 3570 DC Strain Gage Conditioner must be set to match the full-
scale range of the strain gage transducer.

1. If, after pressing the RANGE button, you want to display the current input-
range setting, press ENTER. Otherwise, press RANGE again and proceed to
Step 5. If you press ENTER, the unit will display a number representing the
current input-range setting (in mV/V): ".75," "1.5," or "3.0."

2. To change the input range, you can cycle forwards or backwards through the
allowed numbers by pressing the top or bottom segment, respectively, of any
digit display (even if it is blank).

4. When the desired mV/V range appears, press ENTER. The unit will display
"ECITE" (for "EXCITATION"):

5. If you want to display the current excitation setting, press ENTER. Otherwise,
press RANGE again and proceed to Step 7. If, for example, the conditioner is
set for an excitation of 5 V-DC, when you press ENTER it will display "5."

6. To change the excitation level, you can cycle forwards or backwards through
the allowed settings (2, 5, and 10 V) by pressing the top or bottom segment,
respectively, of any digit display (even if it is blank).

7. When the desired excitation appears, press ENTER to return to the "RANGE"
display. Then press SETUP to return to the "Setup" display, and proceed to
Section 3.a.6.

3.a.5.f SETTING THE MODEL 3578’S AC STRAIN GAGE INPUT RANGE

The Model 3578 AC Strain Gage Conditioner must be set to match the full-
scale range of the strain gage transducer.

1. After pressing the RANGE button, press ENTER. The current range setting
will now appear. If, for example, the conditioner is set for a full-scale input of
1.5 mV/V, the display will show the number "1.5."

2. To change the input range, you can cycle forwards or backwards through the
allowed settings (.75, 1.5, and 3.0 mV/V) by pressing the top or bottom seg-
ment, respectively, of any digit display (even if it is blank).

3. When the desired mV/V range appears, press ENTER to return to the
"RANGE" display. Then press SETUP to return to the "Setup" display.

3.a CONFIGURING THROUGH THE FRONT PANEL 3.11


3 SETUP: INSTRUMENT CONFIGURATION

3.a.6 FILTER: FILTER KEY


For the Model 3530 LVDT Conditioner, Model 3560 Voltage Conditioner,
Model 3570 DC Strain Gage Conditioner, and Model 3578 AC Strain Gage
Conditioner, you can select a low-pass corner frequency of 5, 10, or 20 Hz for
the instrument's analog filter. For the Model 3540 Frequency Conditioner, you
can select a frequency of 2.5, 5, or 10 Hz for this filter. There is NO selectable
analog filtering for the Model 3510 Thermocouple Conditioner.

For ALL conditioners, you can set the magnitude of the smoothing constant for
the digital filter. The effect of the digital filter is to remove small unwanted dynam-
ic signal components, while allowing large-scale fluctuations to pass unaffected.
To set either or both filters,

a. Once you are in SETUP MODE, press the button labelled

FILTER

The unit will display "FILTR" (for "FILTER"):

b. Press either the FILTER or the ENTER button. Unless your unit is a Model
3510 Thermocouple Conditioner, it will display "ANALG" (for "ANALOG"):

If your unit is a Model 3510, it will display "DGTAL" (for "DIGITAL"); proceed to
Step f, below.

c. If you don't want to display the current ANALOG FILTER setting, press FILTER
again and proceed to Step f. To display the analog filter setting, press

ENTER

If, for example, the current filter "response" is 5 Hz, the unit would display

d. To change the analog filter response, you can cycle forwards or backwards
through the allowed settings by pressing the top or bottom segment, respec-
tively, of any digit display (even if it is blank). Allowed "RS" values for the
Model 3540 Frequency Conditioner are "2" (for "2.5"), "5," and "10"; for all
other conditioners with selectable analog filtering, they are "5," "10," and "20."
For filter frequency characteristics and step-response settling times, see
Appendix A.

e. When the desired filter setting appears, press ENTER. The unit will then dis-
play "DGTAL" (for "DIGITAL"):

3.12 3.a CONFIGURING THROUGH THE FRONT PANEL


SETUP: INSTRUMENT CONFIGURATION 3
f. If you don't want to display the current DIGITAL FILTER setting, press FILTER
to return to the "FILTER" display. To display the digital filter setting, press
ENTER. A number from 0 through 9 will be displayed. These "filter constants"
signify increasing amounts of automatic digital filtering, where "0" is NO FIL-
TER and "9" is MAXIMUM FILTER. If, for example, the current digital filter num-
ber is "1," the unit will display

g. To change the digital filter response, increase or decrease the number to the
right of the equals sign by pressing its top or bottom segment, respectively
(the other NUMERIC BUTTONS are inactive).

h. When the desired filter setting appears, press ENTER to return to the "FILTER"
display. Then press SETUP to return to the "Setup" display.

3.a.7 ANALOG OUTPUT: ANO KEY


Following the steps given below, you can specify an offset for the ±5-V analog
output between -25.5% and 25.5% of the nominal input, in 0.1% increments. You
may also scale the analog output to between 74.5% and 125.5% of nominal input,
in 0.1% increments.* Note that while they are set in the same way, the "ANALOG
OFFSET" and "ANALOG AMPLIFICATION (GAIN)" parameters function differently
for the Model 3510 Thermocouple Conditioner than they do for all the other
conditioners. See the explanation at the end of this section.

a. Once you are in SETUP MODE, press the button labelled

ANO

The unit will display "ANOFF" (for "ANALOG OFFSET"):

b. To display the current analog-output offset (in %), press

ENTER

If you don't want to display the offset, press ANO again and proceed to Step
e.

c. To change the offset value, use the NUMERIC BUTTONS as explained in Sec-
tion 1.E to modify the displayed number (you cannot change its "0.1" resolu-
tion). NOTE: The instrument will not accept a value outside the range of ±25.5
(%).

d. When the desired offset value appears, press ENTER to return to the
"ANOFF" display. Then press ANO again. The unit will display "AN A" (for
"ANALOG AMPLIFICATION"):

* For complete analog-output specifications, see Appendix A. Note that the following procedure
is the only way to impose an offset on the analog output, which is unaffected by the instru-
ment's current ZERO OFFSET ("b") and TARE REGISTER values.

3.a CONFIGURING THROUGH THE FRONT PANEL 3.13


3 SETUP: INSTRUMENT CONFIGURATION

e. To display the current analog-output gain value, press ENTER. If you don't
want to display the gain, press SETUP to return to the "Setup" display.

f. To change the amplification value, use the NUMERIC BUTTONS as explained


in Section 1.e to modify the displayed number (again, you cannot change its
"0.1" resolution). NOTE: The instrument will not accept a value outside the
range of 74.5 through 125.5 (%).

g. When the desired amplification setting appears, press ENTER to return to the
"AN A" display. Then press SETUP to return to the "Setup" display.

ANALOG OFFSET and ANALOG AMPLIFICATION (GAIN) for the Model 3510
Thermocouple Conditioner—For the 3510, the user should enter for the "ANA-
LOG OFFSET" value the temperature that is to be represented by an analog out-
put value of 0 volts. For the "ANALOG OUTPUT AMPLIFICATION" value, the user
should enter the temperature that is to be represented by an analog output value
of 5 volts.

Suppose, for example, that an ANALOG OFFSET of "0" and an ANALOG OUTPUT
AMPLIFICATION of "500" are entered for the Model 3510. When "0°" is on display
(°C or °F, any TC type, with or without internal cold junction), the analog output will
equal 0 V; when "250°" is on display, it will be 2.5 V; when "500°" is on display, it will
be 5 V; and when "-100°" is on display, it will be -1 V. If, however, an ANALOG OFF-
SET of "200" is entered, and an ANALOG OUTPUT AMPLIFICATION of "400," then
a reading of "200°" will produce an analog output of 0 V; a reading of "400°," 5 V; a
reading of "300°," 2.5 V; a reading of "100°," -2.5 V; and a reading of "0°," -5 V.*

IMPORTANT: YOU MUST NOT ENTER FOR THE MODEL 3510 AN ANALOG
AMPLIFICATION VALUE GREATER THAN 6 TIMES THE DIFFERENCE BETWEEN
THE ANALOG AMPLIFICATION VALUE AND THE ANALOG OFFSET VALUE. THAT
IS,
AOG
≤ 6
(AOG - AOO)

where "AOG" is the ANALOG AMPLIFICATION value and "AOO" is the ANALOG
OFFSET value. For example, an "AOO" of 2500 and an "AOG" of 2600 are not
allowed, since 2600/100 = 260. An "AOO" of 2500 and an "AOG" of 3500 will work,
however, since 3500/1000 = 3.5 (≤ 6).

3.a.8 LIMITS: LIMIT KEY


By specifying two independent limit setpoints ("HIL" and "LOL"), you define three
distinct LIMIT ZONES (Fig. 18). Your 3500 instrument will continuously monitor
the "live" data reading for its status with respect to these zones. When a "HI" or
"LO" limit violation is detected, the corresponding logic output will be issued and
the corresponding front-panel limit status indicator will light.

* You can even make the ANALOG OFFSET for a Model 3510 greater than the ANALOG OUTPUT
AMPLIFICATION, in order to make the slope of the analog output negative compared to the
input.

3.14 3.a CONFIGURING THROUGH THE FRONT PANEL


SETUP: INSTRUMENT CONFIGURATION 3

HI Greater Than Zone


HIL
OK Between ("OK") Zone
LOL
Fig. 18 Limit Zones LO Less Than Zone

As explained below, when you define each limit value, you will indicate whether
you want the limit to be LATCHING or NONLATCHING. You will also be able to
specify a HYSTERESIS "WINDOW" for any NONLATCHING limit.

a. Once you are in SETUP MODE, press the button labelled

LIMIT

The unit will display "HIL" (for "HIGH LIMIT"):

b. To display the unit's current HIGH LIMIT value, press

ENTER

If you don't want to display the "HIL" value, press LIMIT again and proceed to
Step e.

c. To change the HIGH LIMIT, use the NUMERIC BUTTONS as explained in Sec-
tion 1.e to modify the displayed number. NOTE: You cannot change the preci-
sion (i.e., decimal-point location) of either limit value. This precision is
automatically set to match that of the SCALING FACTOR ("m" coefficient) that
currently applies to the data reading (see Section 4). Note too that you can-
not set a HIGH LIMIT that is lower than the existing LOW LIMIT—although the
two values may be equal—nor can either value exceed the unit's inherent
count range of ±32700.

d. When the desired HIGH LIMIT value appears, press ENTER to return to the
"HIL" display. Then press LIMIT again. The unit will display "HIHYS" (for "HIGH
HYSTERESIS"):

e. To display the current HIGH HYSTERESIS percentage, press ENTER. If you


don't want to display the HIGH HYSTERESIS, press LIMIT again and proceed
to Step h.

f. The number shown for the HIGH HYSTERESIS parameter is the percentage of
the current SCALING FACTOR ("m" coefficient) which, when subtracted from
the HIGH LIMIT value, determines the lower threshold of the HIGH HYSTERE-
SIS WINDOW. The window extends upwards (in the positive direction) from
this threshold to the HIGH LIMIT value itself, as illustrated in Fig. 19. For front-
panel display of the current "m" value, see Section 4.a.4.

3.a CONFIGURING THROUGH THE FRONT PANEL 3.15


3 SETUP: INSTRUMENT CONFIGURATION

Greater Than Zone


t1 t2
HIL HI Hysteresis
Window
t3
OK Zone

"Live" Data Reading

LO Hysteresis
Window
LOL
Less Than Zone

Fig. 19 High and Low Hysteresis Windows

The reason for setting up a HIGH HYSTERESIS WINDOW is to prevent low-


level signal noise from toggling control outputs on and off when the data
reading is in the neighborhood of the HIGH LIMIT value.

A "HIGH VIOLATION" will be triggered only when the data reading enters the
"GREATER THAN ZONE" by exceeding the current "HIL" value. This is what
happens at time t1 in the above figure. At time t2, the data falls back to the "OK
ZONE." If there were no hysteresis window—and assuming that this is a NON-
LATCHING limit—the "HIGH VIOLATION" would cease to occur as soon as the
live data reading became less than "HIL" at time t2. In this case, however, the
limit status continues to be evaluated as "HIGH VIOLATION" until such time as
the data reading becomes less than the lower threshold of the HIGH HYS-
TERESIS WINDOW—which is what happens at time t3. At this point, the limit
status is evaluated as "OK."

To change the HIGH HYSTERESIS value, use the NUMERIC BUTTONS as


explained in Section 1.e to modify the displayed number. REMEMBER: This
number expresses a PERCENTAGE OF THE EXISTING "m" COEFFICIENT. If
you don't need a hysteresis window, just enter a value of "0."

g. When the desired HIGH HYSTERESIS value appears, press ENTER to return
to the "HIHYS" display. Then press LIMIT again. The unit will display "HILAT"
(for "HIGH LATCH"):

h. To display the current HIGH LATCH status, press ENTER. If you don't want to
display the HIGH LATCH, press LIMIT again and proceed to Step k.

i. If the HIGH LATCH status is OFF, it means that the limit is NONLATCHING: the
"HIGH VIOLATION" status will cease to occur as soon as the data reading
leaves the "GREATER THAN ZONE." If the HIGH LATCH status is ON, it means
that the limit is LATCHING: once a "HIGH VIOLATION" is triggered, this condi-
tion will remain in effect until it is reset by means of an "UNLATCH" command

3.16 3.a CONFIGURING THROUGH THE FRONT PANEL


SETUP: INSTRUMENT CONFIGURATION 3
at the rear LOGIC I/O CONNECTOR or by a RELEASE (RLS) command
received through the RS-232/485 Interface (see Appendices B and F).

To change the HIGH LATCH status, you can toggle between ON and OFF by
pressing the top or bottom segment of any display character.

j. When the desired HIGH LATCH status appears, press ENTER to return to the
"HILAT" display. Then press LIMIT again. The unit will display "LOL" (for "LOW
LIMIT").

k. Proceed to set the LOW LIMIT ("LOL"), LOW HYSTERESIS ("LOHYS"), and LOW
LATCH ("LOLAT") just as you did for the HIGH LIMIT. The LOW LIMIT value
cannot be greater than the existing HIGH LIMIT, and, like the HIGH LIMIT, auto-
matically assumes the decimal-point precision of the existing SCALING FAC-
TOR ("m"). The LOW HYSTERESIS WINDOW is just like the HIGH HYSTERESIS
WINDOW, except that it works in the opposite direction and with reference to
the "LOW LIMIT VIOLATION."

l. After you have entered the desired LOW LATCH status by pressing ENTER,
press SETUP to return to the "Setup" display.

NOTE: You can include LIMIT-ZONE INDICATION in the data transmissions. See
Section 3.a.10, below, for instructions.

3.a.9 TARE OFFSET: TARE KEY


By means of the TARE/RESET button in SETUP MODE, you can establish a tare
offset to be continuously subtracted from the "live" data reading whenever the
unit is in "TARE" MODE. To do so,

a. Once you are in SETUP MODE, press the button labelled


TARE
RESET
The unit will display the word "TARE":

b. To display the contents of the unit's TARE REGISTER, press

ENTER

The number in the TARE REGISTER represents the net data reading you want
to see (for the current value of input loading) when the tare offset has been
subtracted. It should be expressed in the desired engineering units and will
automatically reflect the existing precision of the SCALING FACTOR ("m").
When the unit is placed in "TARE" MODE,* the displayed data reading will be
brought to the existing TARE REGISTER value. From this point, the displayed
data will continue to track the "live" input, but now with the continuous applica-
tion of the offset determined by the difference between the reading that exist-
ed when "TARE" MODE was initiated and the last-entered TARE REGISTER
value (see Fig. 24, Section 5.c).

* By the TARE/RESET button or TARE (TAR) command (see Section 5.c).

3.a CONFIGURING THROUGH THE FRONT PANEL 3.17


3 SETUP: INSTRUMENT CONFIGURATION

c. To change the current TARE REGISTER value, use the NUMERIC BUTTONS as
explained in Section 1.e to modify the displayed number.

d. When the desired TARE REGISTER value appears, press ENTER to return to
the "TARE" display. Then press SETUP to return to the "Setup" display.

3.a.10 PRINT AND OUTPUT PARAMETERS: PRINT KEY


When in SETUP MODE, the PRINT button lets you indicate (1) the interval at which
you want hard-copy output to be issued automatically from the RS-232/485 Inter-
face (WHEN IT IS SET FOR RS-232-C COMMUNICATIONS BY VIRTUE OF A NODE
NUMBER OF "0"—see Section 3.a.3); (2) whether or not the NODE NUMBER is to
be included in the RS-232/485 data transmissions; and (3) whether or not LIMIT-
STATUS INDICATION is to be included in all data transmissions from the RS-
232/485 Interface.

a. Once you are in SETUP MODE, press the button labelled

PRINT

The unit will display the word "PRINT":

b. Press either the PRINT or the ENTER button to display "INT" (for "INTERVAL"):

c. To display the current PRINT INTERVAL number, press

ENTER

If you don't want to display the interval number, press PRINT again and pro-
ceed to Step f.

d. The actual PRINT INTERVAL is approximately equal to the number now being
displayed multiplied by 0.125 second. Thus, for example, an interval number
of "16" denotes an actual time interval of about 2 seconds between succes-
sive hard-copy transmissions. To change the interval number, use the
NUMERIC BUTTONS as explained in Section 1.e to modify the displayed (inte-
gral) number, which can be as large as "32700."

e. When the desired PRINT INTERVAL number appears, press ENTER. The unit
will display "ECO" (for "NODE-NUMBER ECHO"):

f. To display the current "ECHO" status, press ENTER. If you don't want to dis-
play the "ECHO" status, press PRINT again and proceed to Step i.

g. If the "ECHO" status is ON, it means that the unit's NODE NUMBER will be
included in all responses to a CHANNEL (CHN) command, even if that num-

3.18 3.a CONFIGURING THROUGH THE FRONT PANEL


SETUP: INSTRUMENT CONFIGURATION 3
ber is "0" (as it usually is in the single-node case). For the precise response
format, see Section 5.a.2. If the "ECHO" status is OFF, it means that the node-
number "echo" will not appear in any CHN transmission.

To change the "ECHO" status, you can toggle between ON and OFF by press-
ing the top or bottom segment of any display character.

h. When the desired "ECHO" status appears, press ENTER. The unit will display
"STATS" (for "LIMIT-STATUS INDICATION"):

i. To display the current "LIMIT-STATUS INDICATION" status, press ENTER. If


you do not want to display this status, press PRINT again and proceed to Step
k.

j. If the "LIMIT-STATUS INDICATION" is ON, it means that one of three numbers


("-1," "0," or "1") will be included in all responses to a CHANNEL (CHN) com-
mand, indicating the LIMIT ZONE which the "live" data reading currently occu-
pies: -1 for "LESS THAN ZONE"; 0 for "OK ZONE"; and 1 for "GREATER THAN
ZONE." If the "LIMIT-STATUS INDICATION" status is OFF, it means that limit-
zone indication will not appear in any CHN transmission.

To change the "LIMIT-STATUS INDICATION" status, you can toggle between


ON and OFF by pressing the top or bottom segment of any display character.

k. When the desired "LIMIT-STATUS INDICATION" status appears, press ENTER


to return to the "PRINT" display. Press SETUP to return to the "Setup" display.

Note that there are two mnemonic commands by which you can further format
your data outputs. By means of the LABEL (LBL) command, you can specify a
string of up to 8 characters to serve as a "HEADER" prefix for each transmission,
while the ENGINEERING UNIT STRING (EUS) command lets you specify a ter-
minating "TAILER" of up to 8 characters for each transmission. For a full discus-
sion of possible transmission formats, see Section 5.a.2.

3.a CONFIGURING THROUGH THE FRONT PANEL 3.19


3 SETUP: INSTRUMENT CONFIGURATION

3.b CONFIGURING THROUGH THE RS-232/485 INTERFACE


When proper RS-232/485 communications have been established with an exter-
nal computer or terminal, it is possible for a 3500 Series instrument to receive
and respond to SETUP COMMANDS issued directly to it through this interface.
There is a specific SETUP COMMAND for each of the front-panel button proce-
dures described in the previous section, with the exception of the node-number
assignment (see the NODE (NOD) command, below). These commands are list-
ed below. For a full discussion of "Command and Response Syntax," along with a
list of all valid mnemonic commands, see Appendix B of this manual.

Except for NODE (NOD), each of the following SETUP COMMANDS can take a
"WRITE" or "READ" form. Only the "WRITE" form is given here. Again, for full
details, see Appendix B. In the following command expressions, [CMT] means
"COMMAND TERMINATOR" (that is, the INPUT TERMINATOR character which the
unit has been set to recognize—see the procedure for setting the INPUT TERMI-
NATOR via the COM button in Section 3.a.3, above).

IMPORTANT: WHEN CONFIGURING YOUR UNIT "REMOTELY" THROUGH THE


RS-232/485 INTERFACE, DON'T FORGET TO ISSUE A COMMAND OF

SAV=ON [CMT]

WHEN YOU'RE FINISHED. THIS COMMAND WILL SAVE THE EXISTING SETUP
CONFIGURATION TO NONVOLATILE EEPROM.* IF NOT SAVED, THE SETUP VAL-
UES YOU HAVE JUST ENTERED WILL BE LOST ON INTERRUPTION OF POWER.

3.b.1 SECURITY CODE


SECURITY CODE (SCD): SCD=c [CMT]
Sets the security code to "c" (where 0 ≤ c ≤ 98).

3.b.2 COMMUNICATIONS PARAMETERS


BAUD RATE (BAU): BAU=b,s,d,p [CMT]
Sets the baud rate ("b"), number of stop bits ("s"), number of data bits ("d"),
and parity ("p"). Refer to Appendix B for the values these variables can take.

NODE (NOD): NOD [CMT]


Reads the node number of the currently "open" node (for "opening" a node,
see Section 5.a.3). THERE IS NO "WRITE" FORM OF THIS COMMAND; A
NODE NUMBER CAN BE ASSIGNED TO A UNIT ONLY THROUGH THE
FRONT-PANEL COM BUTTON SEQUENCE DESCRIBED IN SECTION 3.a.3.

* Note that an interrogation of the form SAV [CMT] will return "ON" or "OFF," depending on
whether or not the setup is in the process of being automatically saved as a result of pressing
the front-panel SETUP button (Section 3.a.1). Note too that the SETUP COMMAND SAV=OFF
[CMT] has no effect.

3.20 3.b CONFIGURING THROUGH THE RS-232/485 INTERFACE


SETUP: INSTRUMENT CONFIGURATION 3
COMMAND TERMINATOR (CMT): CMT=c [CMT]
Sets the COMMAND (or "INPUT") TERMINATOR to "c" (where c is a single
ASCII character entered as a hexadecimal byte in square brackets, [01]
through [1F]).

END-OF-TRANSMISSION TERMINATOR (EOT): EOT=$ [CMT]


Sets the END-OF-TRANSMISSION (or "OUTPUT") TERMINATOR to "$" (where
$ is a string of up to four ASCII characters, each entered as a hexadecimal
byte in square brackets, [01] through [1F]). In a multinode network, the last
EOT character must be the same as the CMT character.
NOTE: A working EOT cannot include a character of "[00]," except as a default
value for unused characters. If, for example, you specify a two-character EOT
of "[0A][0D]" and then use the front-panel procedure given in Section 3.a to
interrogate for the current EOT characters, you will be told that "Eot1" = [0A];
"Eot2" = [0D]; "Eot3" = [00]; and "Eot4" = [00]. In fact, the instrument will stop
transmitting as soon as it perceives an EOT character of [00]. Thus, if you
were to enter an EOT of "[00][0D] ...," no EOT at all would be transmitted.

3.b.3 THERMOCOUPLE T YPE (MODEL 3510 ONLY)


TYPE (TYP): TYP=y [CMT] or TYP=y,N [CMT]
The first version of the command sets the Thermocouple Type to "y" (where
"y" = E, J, K, R, S, T, or B) and enables the internal cold-junction reference.
The second version sets the Thermocouple Type to "y" and disables the inter-
nal cold-junction.

3.b.4 INPUT RANGE OR SCALE


RANGE (RNG): RNG=r [CMT]
Sets the input range or scale to "r," as given by the following table:

Conditioner "r" Actual Input


Model value Range or Scale
3510 TC Conditioner CEN degrees C
FAR degrees F
3530 LVDT Conditioner 5 0-150 mV/V "normal";
0-1 V/V "long-stroke"
10 0-300 mV/V "normal";
0-2 V/V "long-stroke"
20 0-600 mV/V "normal";
0-4 V/V "long-stroke
3540 Frequency Conditioner 250 250 Hz
500 500 Hz
1000 1000 Hz
2000 2000 Hz
4000 4000 Hz
8000 8000 Hz
16000 16000 Hz
32000 32000 Hz
(cont’d)

3.b CONFIGURING THROUGH THE RS-232/485 INTERFACE 3.21


3 SETUP: INSTRUMENT CONFIGURATION

Conditioner "r" Actual Input


Model value Range or Scale
3560 Voltage Conditioner 0.5V ±0.5 V-DC
1V ±1.0 V-DC
2V ±2.0 V-DC
5V ±5.0 V-DC
10V ±10.0 V-DC
20V ±20.0 V-DC
3570 DC Strain Gage Conditioner .75 0.75 mV/V
1.5 1.5 mV/V
3.0 3.0 mV/V
3578 AC Strain Gage Conditioner .75 0.75 mV/V
1.5 1.50 mV/V
3.0 3.00 mV/V

3.b.5 EXCITATION (MODELS 3570 AND 3578 ONLY)


EXCITATION (EXC): EXC=v [CMT]
Sets the excitation level to "v," as given by the following table:

Excitation No. ("v") Excitation


2 2 V (±1 V)
5 5 V (±2.5 V)
10 10 V (±5 V)

3.b.6 SENSITIVITY (MODEL 3540 ONLY)


SENSITIVITY (SEN): SEN=s [CMT]
Sets the input sensitivity range to "s," as given by the following table:

Sensitivity No. ("s") Sensitivity Range


.1 0.1 to 5 V
1 1 to 20 V
10 10 to 100 V

3.b.7 FILTER
RESPONSE (RES): RES=f [CMT]
Sets the analog filter corner to "f" Hz. For the Model 3540 Frequency Condi-
tioner, "f" = 2 (for 2.5), 5, or 10; for all other conditioners except the Thermo-
couple Conditioner, "f" = 5, 10, or 20. For the Model 3510 Thermocouple
Conditioner, the RES command is inactive— although an interrogation of
RES [CMT] will always answer "20."

FILTER (FIL): FIL=f [CMT]


Sets the digital filtering constant to "f" (where 0 ≤ f ≤ 9).

3.22 3.b CONFIGURING THROUGH THE RS-232/485 INTERFACE


SETUP: INSTRUMENT CONFIGURATION 3
3.b.8 ANALOG OUTPUT
ANALOG OUTPUT OFFSET (AOO): AOO=v [CMT]
Sets the analog output offset to "v" (where v is a percentage of full scale to the
nearest tenth of a percent between -25.5% and 25.5%).

ANALOG OUTPUT GAIN (AOG): AOG=v [CMT]


Sets the analog output gain to "v" (where v is a percentage of full scale to the
nearest tenth of a percent between 74.5% and 125.5%).

See Section 3.a.7 for an explanation of the special use of the AOO and AOG para-
meters for the Model 3510 Thermocouple Conditioner.

3.b.9 LIMITS
HIGH LIMIT (HIL): HIL=h [CMT]
Sets the HIGH LIMIT value to "h."

HIGH HYSTERESIS (HHY): HHY=p [CMT]


Sets the HIGH HYSTERESIS window depth to "p" (where p is a percentage of
the existing SCALING FACTOR ("m") value to the nearest tenth of a percent).

HIGH LATCH (HLA): HLA=ON [CMT] or HLA=OFF [CMT]


Sets the HIGH-LIMIT LATCH status to "ON" or "OFF."

LOW LIMIT (LOL): LOL=L [CMT]


Sets the LOW LIMIT value to "L."

LOW HYSTERESIS (LHY): LHY=p [CMT]


Sets the LOW HYSTERESIS window depth to "p" (where p is a percentage of
the existing SCALING FACTOR ("m") value to the nearest tenth of a percent).

LOW LATCH (LLA): LLA=ON [CMT] or LLA=OFF [CMT]


Sets the LOW-LIMIT LATCH status to "ON" or "OFF."

3.b.10 TARE OFFSET


TARE REGISTER (TRR): TRR=v [CMT]
Sets the TARE REGISTER value to "v."

3.b.11 PRINT AND OUTPUT PARAMETERS


PRINT INTERVAL (PRI): PRI=t [CMT]
Sets the automatic print interval to approximately (t x 0.125) second(s), where
0 ≤ t ≤ 32700.

3.b CONFIGURING THROUGH THE RS-232/485 INTERFACE 3.23


3 SETUP: INSTRUMENT CONFIGURATION

ECHO (ECO): ECO=ON [CMT] or ECO=OFF [CMT]


Sets the node-number "ECHO" to "ON" or "OFF."

LIMITS (LIM): LIM=ON [CMT] or LIM=OFF [CMT]


Sets the "LIMIT-STATUS INDICATION" to "ON" or "OFF."

(See also the LABEL (LBL) and ENGINEERING UNIT STRING (EUS) commands
described in Section 5.a.2.)

3.24 3.b CONFIGURING THROUGH THE RS-232/485 INTERFACE


SETUP: INSTRUMENT CALIBRATION 4
4.a INTRODUCTION: CALIBRATION TECHNIQUES
The calibration methods that apply to each 3500 Series conditioner model are
summarized in the following table.

Calibration Techniques

Calculated via Mnemonic Command


Absolute (NO USER CALIBRATION)

Internal 15-Segment Linearization


Two-Point (Simulated)
Two-Point (Actual)

Calculated
Conditioner
Model

3510 TC
X X X
Conditioner

3530 LVDT
X X X
Conditioner

3540 Frequency
X X X X
Conditioner

3560 Voltage
X X X
Conditioner

3570 DC Strain
X X X X X
Gage Conditioner

3578 AC Strain
X X X
Gage Conditioner

For each instrument, the applicable calibration procedures can be performed


either via the front-panel SETUP buttons or via mnemonic commands received
through the RS-232/485 Interface Port. For the mnemonic commands relating to
each calibration technique, see Section 4.d.

4.a.1 ABSOLUTE CALIBRATION


THE MODEL 3510 THERMOCOUPLE CONDITIONER IS THE ONLY ONE WHICH
EMPLOYS STRICTLY "ABSOLUTE" CALIBRATION. With this kind of calibration, the
zero and scaling characteristics of all allowable thermocouples are fixed and
known. Since calibration values for the TC input are prestored in the 3510’s
processor memory, there is no need for any further calibration procedures, as

4.a INTRODUCTION: CALIBRATION TECHNIQUES 4.1


4 SETUP: INSTRUMENT CALIBRATION

soon as the "TC TYPE" and desired measurement units are specified. For the
front-panel entry of TC type and scale, see Section 3.a.5(a). For use of the TYPE
(TYP) and RANGE (RNG) commands to accomplish the same thing, see Section
3.a.B.

NOTE: While "absolute" calibration is sufficient for almost all applications of the
Model 3510, the "Two-Point" technique described below can also be used for
applications where it is desirable to force multiple TC readings to the same exact-
ly known temperature value (in heat balance applications, for example).* In such
cases, the two-point method can be used to improve on the "absolute" calibration
inherent in the 3510.

Also note that the internal 15-SEGMENT LINEARIZATION procedure described in


Section 4.c.4 does not apply to the Model 3510 Thermocouple Conditioner.

4.a.2 “ACTUAL” T WO-POINT (DEADWEIGHT) CALIBRATION


This conventional "zero and span" procedure can be applied to all 3500 Series
conditioner models. It should be used when the transducer signal is relatively lin-
ear and when there are at least two independently and accurately known calibra-
tion values.

With "TWO-POINT" calibration, you will command the instrument's micro-


processor to compute and store two constant values: a ZERO OFFSET term ("b")
and a SCALING FACTOR coefficient ("m," also called "multiplier" or "gain" factor).
Automatically and continuously applied to all subsequent data readings, these
two calibration constants define the linear proportionality expressed by the equa-
tion "y = mx + b."**

4.a.3 “SIMULATED” T WO-POINT (DEADWEIGHT)


CALIBRATION
This calibration method applies to the Model 3570 DC Strain Gage Conditioner
and the Model 3578 AC Strain Gage Conditioner. In "actual" two-point calibra-
tion (above), the second ("span") calibration input is produced by actually loading
the transducer. When this kind of "deadweighting" is impossible or inconvenient,
the same effect can be achieved for either of the two Strain Gage Conditioners by
shunting a resistor of known magnitude across one arm of the strain-gage bridge,
thereby "simulating" a particular up-scale value of mechanical input.

* Note, however, that for a TC input, two-point calibration is of no advantage unless you can gen-
erate temperatures that are known to an accuracy exceeding that given in Appendix A for the
Model 3510.
** Where "y" is the reported measurement value and "x" is the ratio of the actual voltage of the
analog input signal to the positive full-scale voltage of that input (based on the selected input
range). As such, "x" is a unitless number operated upon by the ("slope") coefficient "m" and the
offset term "b" to yield a true analog measurement in the appropriate engineering units.

4.2 4.a INTRODUCTION: CALIBRATION TECHNIQUES


SETUP: INSTRUMENT CALIBRATION 4
4.a.4 INTERNAL 15-SEGMENT LINEARIZATION
For an input signal derived from a nonlinear transducer, an internal "look-up" table
can be created.* This table will be automatically reloaded on every powerup. The
purpose of such a table is to linearize the signal in question, to "bend" its charac-
teristic curves to achieve at least approximately "straight-line" performance.

You will set up a linearization table for the analog input by entering a ZERO ("MINI-
MUM") POINT and up to 15 subsequent LINEARIZATION POINTS. These points
will define up to 15 segments of the ideal linear output for that measurement sig-
nal. The more segments you specify, the greater the overall linearity. You can
load numeric points directly into the linearization table, or you can set up the table
by "forcing" a sequence of inputs to read desired measurement values ("out-
puts").

4.a.5 CALCULATED CALIBRATION


A conditioner can also be calibrated by directly entering an appropriate ZERO
OFFSET ("b" term) and an appropriate SCALING FACTOR ("m" coefficient) of the
desired precision (decimal-point location).

This technique is useful for some conditioners, but not for others. With the Model
3570 DC Strain Gage Conditioner, for example, it is a very convenient method
when both the "mV/V" sensitivity and the corresponding full-scale rating of the
transducer are accurately known. For the Model 3578 AC Strain Gage Condi-
tioner, on the other hand, it SHOULD NOT BE USED AT ALL.

For the Model 3540 Frequency Conditioner or Model 3560 Voltage Condi-
tioner, you can use "calculated" calibration if the unit is being used to measure
frequency or voltage itself. Here, all you need to do is to enter an appropriate
SCALING FACTOR of the desired precision.

For the Model 3510 Thermocouple Conditioner or Model 3530 LVDT Condi-
tioner, "calculated" calibration is only useful if you want to transfer the calibration
of one unit to another unit that has the same transducer/cabling system.

When calibrating a conditioner in this way, you should keep in mind that the
resulting accuracy is limited to either the stated "initial offset" accuracy or the stat-
ed "gain accuracy" for that conditioner, whichever represents a greater error
value (see Appendix A), and that the maximum attainable unit resolution is
±32700, fixed decimal.

4.a.6 CALCULATED CALIBRATION VIA


MNEMONIC COMMAND
"Calculated" calibration of the Model 3540 Frequency Conditioner can be per-
formed by issuing the FREQUENCY CALIBRATION (FRQ) command through
the RS-232/485 Interface Port. Similarly, the Model 3570 DC Strain Gage Con-
ditioner can be calibrated by applying the MILLIVOLT/VOLT CALIBRATION
(MVV) command. These commands are described in Section 4.d.

* Except for the Model 3510 Thermocouple Conditioner, to which this technique does not
apply.

4.a INTRODUCTION: CALIBRATION TECHNIQUES 4.3


4 SETUP: INSTRUMENT CALIBRATION

4.b PHASE AND SYMMETRY ADJUSTMENT OF THE


MODEL 3578 AC STRAIN GAGE CONDITIONER
PLEASE NOTE: THIS PROCEDURE APPLIES ONLY TO THE MODEL 3578 AC
STRAIN GAGE CONDITIONER. FOR ALL OTHER CONDITIONERS, YOU MAY
IGNORE THIS MANUAL SECTION.

Before you calibrate your Model 3578 for the first time, you should perform the follow-
ing PHASE AND SYMMETRY ADJUSTMENT procedure. When using the 3578 with a
Lebow 1800 Series Transducer, you should follow the special procedure outlined at
the end of this section.

ONCE SET FOR YOUR TRANSDUCER, THIS ADJUSTMENT NEED NOT BE REPEATED
UNLESS A SIGNIFICANT CHANGE IN CABLE LENGTH OR CAPACITANCE IS
REQUIRED.

1. Make sure your transducer connection is good (see Section 2.a.7). With the 3578
in "run" mode, load the transducer in the positive direction with a convenient
"deadweight" value which is greater than one-half of full scale.

2. Locate the PHASE ADJUSTMENT CONTROL on the rear of the 3578 (see Fig. 20).
Adjust the PHASE control until a maximum reading is obtained on the display.

Symmetry Phase
Adjustment Adjustment
Control Control
+

Fig. 20 Symmetry and Phase Adjustment Controls for the Model 3578

3. Remove the transducer load.

4. Enter SETUP MODE as explained in Section 3.a.1 and press the

CAL

button. If the unit does not now display

continue to press the CAL button until it does.

5. Press the ENTER button. The unit will display "OFSET" (for "ZERO OFFSET"):

4.4 4.b PHASE AND SYMMETRY ADJUSTMENT OF THE MODEL 3578


SETUP: INSTRUMENT CALIBRATION 4
6. Press ENTER once again. This will return you temporarily to a display of the "live"
analog input.

7. Press ANY KEY EXCEPT CAL OR SETUP. This will freeze the displayed data
value.

8. Using the NUMERIC BUTTONS as explained in Section 1.e, change the displayed
value to "0" (zero).

9. Press ENTER. The display will return to "OFSET." Now press CAL. The unit will
display the word "FORCE":

10. Press ENTER. The unit will display the mnemonic for "SHUNT CALIBRATION—
POSITIVE":

11. Press ENTER. The display will show the status of the positive shunt: "ON" means
that the shunt is connected; "OFF," that it is not. If the positive shunt is currently
OFF, turn it ON by pressing the top or bottom segment of any displayed charac-
ter, followed by ENTER. The unit will display the mnemonic for "SHUNT CALIBRA-
TION—NEGATIVE":

12. Press ENTER. The display will show the status of the negative shunt (it should be
OFF, since you just turned ON the positive shunt). Press ENTER. This will return
you temporarily to a display of the "live" analog input.

13. Record the currently displayed data reading.

14. Press ENTER. This will freeze the displayed data reading.

15. Press ENTER twice to return to "SHP."

16. Press ENTER and turn OFF the positive shunt by pressing the top or bottom seg-
ment of any displayed character, followed by ENTER.

17. Press ENTER once more and turn ON the negative shunt by pressing the top or
bottom segment of any displayed character, followed by ENTER. This will return
you temporarily to a display of the "live" analog input.

18. Now adjust the rear-panel SYMMETRY control until the negative value of the read-
ing you recorded in Step 13 appears.

19. Press CAL. The display will return to "OFSET." YOU ARE NOW READY TO CALI-
BRATE THE 3578, AS EXPLAINED IN SECTION 4.c. Note that the negative shunt
was automatically turned OFF when you pressed the CAL button.

4.b PHASE AND SYMMETRY ADJUSTMENT OF THE MODEL 3578 4.5


4 SETUP: INSTRUMENT CALIBRATION

PHASE/SYMMETRY ADJUSTMENT FOR A


LEBOW 1800 SERIES TRANSDUCER

1. WHEN USING THE MODEL 3578 WITH A LEBOW 1800 SERIES TRANSDUCER,
YOU SHOULD FIRST INSTALL THE SHUNT RESISTOR SUPPLIED WITH THE
TRANSDUCER. Refer to Fig. 11(c) and connect the resistor between Pin 5 of the
CONDITIONER CONNECTOR and the transducer's +SIGNAL line (you will not use
the "CAL SENSE" pin (Pin 4) in this case).

2. Locate the "CAL/RUN" Switch in the cable harness of the 1800 Series transducer.
Place this switch in the "CAL" position.

3. Establish a zero input by removing all load from the 1800 Series transducer.

4. Enter SETUP MODE (see Section 3.a.1), and press the CAL button. If the 3578 unit
does not now display "2 Pt," continue to press CAL until it does.

5. Press the ENTER button. The unit will display "OFSET" (for "ZERO OFFSET").

6. Press the CAL button. The unit will display "FORCE."

7. Press ENTER. The unit will display "SHP" (for "SHUNT CALIBRATION—POSITIVE").

8. Press ENTER again. The display will show the status of the positive shunt: "ON"
means that the shunt is connected; "OFF," that it is not. If the positive shunt is cur-
rently OFF, turn it ON by pressing the top or bottom segment of any displayed
character, followed by ENTER. The unit will display "SHN" (for "SHUNT CALIBRA-
TION—NEGATIVE").

9. Press ENTER. The display will show the status of the negative shunt (it should be
OFF, since you just turned ON the positive shunt). Press ENTER. This will return
you temporarily to a display of the "live" analog input.

10. Now adjust the rear-panel PHASE ADJUSTMENT CONTROL (see Fig. 20) until a
maximum reading is obtained.

11. Press the CAL button. The display will return to "OFSET" (the positive shunt has
been automatically turned OFF).

12. Press ENTER. This will return you temporarily to a display of the "live" analog
input.

13. Press ANY KEY EXCEPT CAL OR SETUP. This will freeze the displayed data
value.

14. Using the NUMERIC BUTTONS as explained in Section 1.e, change the displayed
value to "0" (zero).

15. Press ENTER. The display will return to "OFSET." Now press CAL. The unit will
again display "FORCE."

16. Repeat Steps 7 through 9.

17. Press ANY KEY EXCEPT CAL OR SETUP. This will freeze the displayed data
value.

4.6 4.b PHASE AND SYMMETRY ADJUSTMENT OF THE MODEL 3578


SETUP: INSTRUMENT CALIBRATION 4
18. Using the NUMERIC BUTTONS, change the displayed value to the "EQUIVALENT
INPUT" VALUE GIVEN BY THE TRANSDUCER MANUFACTURER FOR THE CALI-
BRATION RESISTOR YOU HAVE INSTALLED (for a discussion of "equivalent input,"
see Section 4.c.3, below). This value should be expressed in the desired engineer-
ing units and with the desired measurement precision.

19. Press ENTER twice to return to "SHP."

20. Press ENTER and turn OFF the positive shunt by pressing the top or bottom seg-
ment of any displayed character, followed by ENTER.

21. Press ENTER once more and turn ON the negative shunt by pressing the top or
bottom segment of any displayed character, followed by ENTER. This will return
you temporarily to a display of the "live" analog input.

22. Now adjust the rear-panel SYMMETRY control to produce a reading equal to the
negative value of the "equivalent input" you entered in Step 18 (or some other spe-
cific negative engineering-unit value, if such a value is given by the transducer
manufacturer for the calibration resistor).

23. Press the CAL button. The display will return to "OFSET" (the negative shunt has
been automatically turned OFF).

24. Move the transducer's "CAL/RUN" Switch to the "RUN" position.

25. With the transducer load still at zero, press ENTER. This will return you temporari-
ly to a display of the "live" analog input.

26. Press ANY KEY EXCEPT CAL OR SETUP. This will freeze the displayed data value.

27. Using the NUMERIC BUTTONS, change the displayed value to "0" (zero). Then
press ENTER. The display will return to "OFSET."

28. Now press CAL. The unit will again display "FORCE."

29. To exit the CAL button sequence and return to the "Setup" display, press SETUP.

THE LEBOW 1800 / DAYTRONIC 3578 SYSTEM IS NOW FULLY CALIBRATED. YOU
NEED NOT PERFORM A SUBSEQUENT "DEADWEIGHT" OR "SIMULATED" CALIBRA-
TION.

4.b PHASE AND SYMMETRY ADJUSTMENT OF THE MODEL 3578 4.7


4 SETUP: INSTRUMENT CALIBRATION

4.c CALIBRATING THROUGH THE FRONT PANEL


4.c.1 USING THE CAL KEY
When you are in SETUP MODE and press the

CAL

button, the unit will display the name of the calibration technique that is currently
in effect. Thus, if your 3500 instrument was last set up for "TWO-POINT" calibra-
tion—either "actual" or "simulated"—the display will read

If you wish to use "TWO-POINT" calibration, press ENTER and follow the instruc-
tions given in Section 4.c.2, below.

If you don't want to use "TWO-POINT" calibration, press CAL again. The unit will
display the next calibration method in the sequence—which, in this case, is "LIN"
(for "LINEARIZATION"):

Note that the Model 3510 Thermocouple Conditioner does not offer this cali-
bration method.

If you wish to use the 15-segment internal linearization, press ENTER and pro-
ceed to Section 4.c.4, below. If not, press CAL again. The unit will display "CALC"
(for "CALCULATED" calibration):

If you wish to use "CALCULATED" calibration, press ENTER and proceed to Sec-
tion 4.c.5, below. Pressing CAL again at this point will return you to the "2 PT" dis-
play. To exit the CAL button sequence and return to the "Setup" display, press
SETUP.

NOTE: The calibration method that is on display when you press the SETUP but-
ton becomes the subsequently active method, and will be the one that therefore
first appears when you next press the CAL button. You can change the currently
active calibration method from the RS-232/485 Port by means of the CALIBRA-
TION (CAL) command, as explained in Section 4.d.

4.c.2 “ACTUAL” T WO-POINT (DEADWEIGHT) CALIBRATION


a. Once you have selected "TWO-POINT" calibration as explained above, the unit
will display "OFSET" (for "ZERO OFFSET"):

4.8 4.c CALIBRATING THROUGH THE FRONT PANEL


SETUP: INSTRUMENT CALIBRATION 4
b. Once again, press

ENTER

This will return you temporarily to a display of the "live" analog input.

c. Apply an accurately known value of input loading to the source transducer—a


value less than 50% of the nominal full-scale range. If it is possible to set this
"zero point" load equal to the minimum value of the transducer's rated operat-
ing range, you should do so. For a displacement sensor, for example, such a
load value is obtained when the sensor is in its "NULL" position.

d. Now press ANY KEY EXCEPT CAL OR SETUP. This will freeze the displayed
data value, which should be the value of input loading at your first ("zero") cali-
bration point.

NOTE: THOUGH THE DISPLAYED INPUT VALUE IS FROZEN, THE INPUT


ITSELF IS NOT. "ZERO" CALIBRATION WILL BE BASED ON WHATEVER
ACTUAL INPUT EXISTS WHEN THAT CALIBRATION POINT IS ENTERED.

e. Using the NUMERIC BUTTONS as explained in Section 1.e, change the dis-
played value to the desired measurement reading for the known input,
expressed in appropriate engineering units and with appropriate polarity.
Note that the button labelled

DEC

is now lit. This button enables you to change the DECIMAL-POINT LOCATION
of the displayed number during the "FORCE" procedure, below. Pressing
DEC will move the decimal point one digit to the left—or, when the decimal
point is at the leftmost position, to remove it (thus returning the number to an
integral value). Remember, too, that your offset entry cannot exceed the
inherent number count of ±32700.

f. Press ENTER to establish your first calibration point. The display will return to
"OFSET." Now press CAL. The unit will display the word "FORCE":

g. Press ENTER. If your unit is a Model 3510, 3530, 3540, or 3560, this will
return you temporarily to a display of the "live" analog input. Proceed directly
to Step h, below.

If, however, your unit is a Strain Gage Conditioner (Model 3570 or 3578), it
will display the mnemonic for "SHUNT CALIBRATION—POSITIVE":

You now have the opportunity to connect the internal shunt resistor for a posi-
tive up-scale reading. For the complete SHUNT CALIBRATION procedure, go
directly to the following section (3.c.3). If you are performing "actual" two-
point calibration, just press the CAL button here. The unit will display the
mnemonic for "SHUNT CALIBRATION—NEGATIVE":

4.c CALIBRATING THROUGH THE FRONT PANEL 4.9


4 SETUP: INSTRUMENT CALIBRATION

Press the CAL button again. This will return you temporarily to a display of the
"live" analog input.

h. Apply a second accurately known value of input loading—a value from 80% to
100% of the transducer's nominal full-scale rating.

i. Now press ANY KEY EXCEPT CAL OR SETUP. This will freeze the displayed
data value, which should be the value of input loading at your second ("force")
calibration point.

NOTE: THOUGH THE DISPLAYED INPUT VALUE IS FROZEN, THE INPUT


ITSELF IS NOT. "FORCE" CALIBRATION WILL BE BASED ON WHATEVER
ACTUAL INPUT EXISTS WHEN THAT CALIBRATION POINT IS ENTERED.

j. Again using the NUMERIC BUTTONS, change the displayed value to the
desired measurement reading for the known input, in the desired engineering
units, and with appropriate polarity. This "forces" the data reading to equal
this value, thereby determining the SCALING FACTOR to be applied to all sub-
sequent readings. Like the offset entry, above, the "force" entry cannot
exceed the inherent number count of ±32700. Also, if you try to "force" a num-
ber that is less than 50% of full scale, you will get an "INPUT INSUFFICIENT"
error.

NOTE: The second calibration entry also sets the desired measurement pre-
cision. If, for example, you're measuring "millimeters," and enter a "FORCE"
value of "10," then all subsequent readings will be rounded to the nearest mil-
limeter. If you enter "10.0," then all readings will be rounded to the nearest
tenth of a millimeter. This same precision will be automatically reflected in the
LIMIT parameters ("HIL" and "LOL"—see Section 3.a.8).

k. Press ENTER to establish your second calibration point. The display will
return to "FORCE." Press SETUP to return to the "Setup" display.

4.c.3 “SIMULATED” (SHUNT) CALIBRATION FOR A STRAIN


GAGE CONDITIONER (MODEL 3570 OR 3578)
This is an easier though generally less accurate technique than "actual" two-point
calibration. It is useful, however, when overall "deadweighting" is impossible or
inconvenient, and is good for an accuracy of about 0.2% (depending, of course,
on the accuracy of the specified EQUIVALENT INPUT, and on the resistor/ bridge
tolerance and temperature). The known calibration input is not produced by load-
ing the transducer, as in the "deadweight" method, but by shunting a resistor of
known magnitude across one arm of the strain-gage bridge, thereby "simulating"
a particular up-scale value of mechanical input. This known EQUIVALENT INPUT
then serves to determine the SCALING FACTOR for the channel.

Every Model 3570 or 3578 is equipped with a 100-kΩ, 0.1% calibration resistor
which you may, if you wish, replace with a resistor of some other value (strain-
gage transducer manufacturers often supply such resistors with their instru-
ments, along with the exact values of EQUIVALENT INPUT thereby produced).
See Fig. 11(c) for resistor installation.

4.10 4.c CALIBRATING THROUGH THE FRONT PANEL


SETUP: INSTRUMENT CALIBRATION 4
EQUIVALENT INPUT can be approximated from a knowledge of the Shunt Cali-
bration Resistance (R); the transducer's Bridge Resistance (B); and the transduc-
er's Full-Scale Sensitivity (K, in mV/V, full scale).

To determine the EQUIVALENT INPUT (X) as an approximate percentage of full-


scale output, you may use the following equation:

25000(mV/V)B
X= %
KR
a. Perform Steps a through f of the "Actual" two-point calibration procedure
given above.

b. Press ENTER. The unit will display "SHP" (for SHUNT CALIBRATION—POSI-
TIVE).

c. If you want to switch in the shunt resistor for a positive up-scale reading, press
ENTER once again and proceed to Step d. If you want to switch in the shunt
resistor for a negative up-scale reading, press the CAL button to display
"SHN" (for "SHUNT CALIBRATION—NEGATIVE"), and press ENTER.

d. After you press ENTER, the 3570/3578 will show the status of the selected
shunt: "ON" means that the shunt is connected; "OFF," that it is not. If the
shunt is currently OFF, turn it ON by pressing the top or bottom segment of
any displayed character, followed by ENTER. This will return you temporarily
to a display of the "live" analog input.

NOTE: By turning ON either of the two shunts, you will automatically turn OFF the
other one (if it is presently ON).

ALSO NOTE: Toggling the "SHP" or "SHN" display to ON is equivalent to entering a


command of SHP=ON [CMT] or SHN=ON [CMT], respectively, through the RS-
232/485 Interface. Toggling a shunt to OFF is equivalent to entering a command
of SHP=OFF [CMT] or SHN=OFF [CMT]. As explained in Section 1.f, the state of
the instrument's positive and negative calibration shunts can also be controlled
by means of logic inputs to the rear-panel Analog Input Connector. However, if
the unit receives contradictory shunt commands from the keypad/interface and
from a shunt-control logic input, it will return to normal "live" measurement.

e. Now freeze the displayed data value by pressing ANY KEY EXCEPT CAL OR
SETUP.

f. Using the NUMERIC BUTTONS, change the displayed value to the known
EQUIVALENT INPUT produced by the shunt, with appropriate polarity. As
before, this second calibration entry also sets the desired measurement pre-
cision.

g. Press ENTER to establish your second calibration point. The display will
return to "FORCE."

h. Press the CAL button. The display will return to "OFSET." Note that pressing
the CAL button will automatically turn OFF whichever shunt is presently ON,
so that normal measurement will resume as soon as you return the condition-
er to "RUN" mode.

i. To exit the CAL button sequence and return to the "Setup" display, press
SETUP.

4.c CALIBRATING THROUGH THE FRONT PANEL 4.11


4 SETUP: INSTRUMENT CALIBRATION

4.c.4 15-SEGMENT LINEARIZATION


NOTE: THIS CALIBRATION TECHNIQUE DOES NOT APPLY TO THE THERMO-
COUPLE CONDITIONER.

4.c.4.a INTRODUCTION

Whether you set up your internal linearization by directly entering input/output


values or by "forcing" each output value based on a corresponding value of actual
input loading, you should first divide your transducer's rated operating range into
a number of approximately equal segments (up to a total of fifteen). The more
nonlinear curve portions, however, may require shorter and more numerous lin-
earization segments, as with segments (f5 to f6) and (f6 to f7) in Fig. 21. Note also
that the final (or "highest") segment effectively extends in a straight line past the
last defined "endpoint" all the way to the end of the present full-scale range.

f7
f6 Segment No. 7
f5
A Segment No. 6
Displayed Output

f4 AA Segment No. 5

A
Segment No. 4
f3
f2
Segment No. 3
AA Does not apply to
Thermocouple

AA
Segment No. 2
f1 Conditioner
Segment No. 1
z i
Physical Input
Fig. 21 Typical Linearization Curve with Seven Segments

Once you have selected the "LINEARIZATION" calibration method as explained in


Section 4.c.1, above, the unit will display the word "RESET":

You now have the option of resetting the present linearization table. WHEN YOU
"RESET" THE TABLE, YOU ARE IN EFFECT INSTRUCTING THE 3500 INSTRUMENT
TO SET THE NUMBER OF ACTIVE LINEARIZATION SEGMENTS TO ZERO.

Note, however, that the endpoint definitions for each line segment in the existing
table are neither cancelled nor altered when the table is "reset." They will be
changed only when new input/output values are entered for the segment in ques-
tion— either directly or by "force." Until then, the segment's existing definition
remains in memory as it was prior to the table "reset," though it no longer has any
effect on calibration. As such, it is still accessible through the "READ" form of the
LINEARIZE (LIN) command, but not through the front panel. Thus, if you acciden-
tally reset the linearization table, any and all segment definitions that have not

4.12 4.c CALIBRATING THROUGH THE FRONT PANEL


SETUP: INSTRUMENT CALIBRATION 4
been changed since the reset can be recovered through the RS-232/485 Inter-
face (see the LIN command, Section 4.d).

You should reset the linearization table when you are linearizing your 3500 instru-
ment for the first time, or whenever you want to relinearize it by the "FORCE" pro-
cedure described in Section 4.c.4(c), below. If you only want to view the existing
table, or to edit one or more existing line-segment definitions, YOU SHOULD NOT
RESET THE TABLE.

To reset the linearization table, press

ENTER

following the display of "RESET." The word "YES" will appear:

Confirm your intention to reset by again pressing ENTER. You will now see the
word "TABLE":

If your knowledge of the desired input/output relationship allows, you can use the
procedure given in Section 4.c.4(b), below, to load points directly into the table.
Press ENTER once more and proceed to that section. If, however, you prefer not
to enter input values directly, but to "force" existing "live" inputs to desired output
values, press the

CAL

button and proceed to Section 4.c.4(c).

To view the existing linearization table without resetting it, or to select linearization
"by force," press ANY KEY OTHER THAN ENTER, following the display of "RESET."
The word "TABLE" will appear. To view the existing table, press ENTER again, and
proceed to the following section.

4.c.4.b BY “TABLE”

1. Once you have pressed ENTER, following the display of "TABLE," the unit will
display "SEG[MENT] 0":

If you have reset the linearization table or are linearizing your 3500 instrument
for the first time, "SEG 0" will be displayed, telling you that you are now going
to enter the "zero" point "z" of the linearization curve (see Fig. 21). Proceed to
Step 2, below.

If you have NOT reset the existing linearization table, "SEG 0" will still be dis-
played, but you can now advance to any line segment in the table in order to
view—and, if desired, to change—that segment's current definition. Follow the
procedure given in Step 11, below, for editing an already defined and active
segment.

4.c CALIBRATING THROUGH THE FRONT PANEL 4.13


4 SETUP: INSTRUMENT CALIBRATION

Pressing the CAL button at this point will take you immediately to the highest
segment in the existing table, which is one more than the last defined seg-
ment, unless the last defined segment is Segment No. 15. If, for example, the
table already contains five fully defined line segments, pressing CAL will dis-
play "SEG 6," which represents the line connecting the last defined endpoint
(the endpoint of Segment No. 5) to the end of the full-scale range. You can
now proceed to define a specific endpoint for Segment 6, if desired, specify-
ing INPUT/OUTPUT values as explained below.

2. Press ENTER again. The unit will display the word "INPUT":

3. Press ENTER again. A number will appear that represents the current con-
tents of the INPUT VOLTAGE REGISTER (in millivolts). Use the NUMERIC BUT-
TONS as explained in Section 1.e to change this number to the value of actual
physical input (again, in millivolts) that will generate the first "output" point—the
"zero" point—on the desired linearization curve.

4. Press ENTER again.The unit will display the word "OUTPUT":

5. Press ENTER again. A number will appear that represents the current con-
tents of the OUTPUT REGISTER. Use the NUMERIC BUTTONS to change this
number to the value you want to be displayed—in the desired engineering
units— when the physical input is the millivolt value you specified in Step 3.

6. Press ENTER again. The unit will again display "SEG 0." To enter the next
point on the linearization curve (the endpoint of the first actual segment),
press the top of the displayed "0" to change it to a "1."

7. Press ENTER again to display "INPUT." Press ENTER and set the number to
the INPUT VOLTAGE (in mV) for the second point.

8. Press ENTER again to display "OUTPUT." Press ENTER and set the number
to the desired OUTPUT (in engineering units) for the second point.

9. Press ENTER again to display "SEG 0." Now increment the "0" to "2," in order
to enter the next point on the linearization curve (the endpoint of Segment No.
2).

10. Continue to enter INPUT/OUTPUT numbers for each segment endpoint of


your linearization curve. It can take up to 16 segments in all.

11. NOTE: If you want to go back and change the INPUT and/or OUTPUT value of
a point that has already been entered, you may easily do so. When the "SEG
0" display appears, just change the "0" to the number of the segment end-
point you want to modify.* Press ENTER twice to display the existing INPUT
VOLTAGE for the point. Use the NUMERIC BUTTONS to change it, if desired.
Press ENTER twice to display the existing OUTPUT for the point. Use the

* Remember that the highest value the displayed segment number can take at any time is one
more than the number of the highest defined line segment.

4.14 4.c CALIBRATING THROUGH THE FRONT PANEL


SETUP: INSTRUMENT CALIBRATION 4
NUMERIC BUTTONS to change it, if desired. Press ENTER to return to the
"SEG 0" display. Now you can add more points, if desired, or quit.

12. After you have defined the last segment of your curve, press SETUP to return
to the "Setup" display.

4.c.4.c BY “FORCE”

NOTE: When you choose this linearization method, you will have to be able to
continuously vary the magnitude of the physical input, or "load," on the source
transducer, and to know the resulting output accurately in the desired engineer-
ing units.

1. Once you have pressed CAL after the display of "TABLE," the unit will display
the word "FORCE":

2. Press ENTER. The unit will display the word "SET" and a number from 0
through 15—for example:

If the linearization table has been reset or if you are linearizing your unit for
the first time, the displayed "SET" number will be "0," and you should proceed
to Step 3, below.

If, however, the existing table has NOT been reset, the displayed number will
be that of the line segment within whose input range the instrument's present
"live" input lies. If in Fig. 21, for example, the present input value (i) lies
between the values f4 and f5, "SET 5" will be displayed, since this input range
corresponds to the endpoints of Segment No. 5. The thing to remember is
that, with linearization "by force," the displayed "SET" number is "live," and will
change as the instrument's "live" input changes (this is in fact the only way to
move from segment to segment, since the "SET" display digits are not
NUMERIC BUTTONS). Also note that if the present "live" input is less than the
existing "zero" point "z," the display will read "SET 0." To "reforce" an existing
line segment, see the procedure given in Step 11, below.

3. Press ENTER. This will return you temporarily to a display of the "live" analog
input.

4. Apply an accurately known MINIMUM value of input loading to the source


transducer (corresponding to the "zero" point "z" of the linearization curve).

5. Now press ANY KEY EXCEPT CAL OR SETUP. This will freeze the displayed
data value. NOTE: THOUGH THE DISPLAYED INPUT VALUE IS FROZEN, THE
INPUT ITSELF IS NOT. LINEARIZATION WILL BE BASED ON WHATEVER
ACTUAL INPUT EXISTS WHEN THAT POINT IS ENTERED.

6. Using the NUMERIC BUTTONS as explained in Section 1.e, change the dis-
played value to the desired measurement reading for the known input, in the
desired engineering units. As in Section 4.c.2, above, you can use the DEC

4.c CALIBRATING THROUGH THE FRONT PANEL 4.15


4 SETUP: INSTRUMENT CALIBRATION

button to set the DECIMAL-POINT LOCATION of the displayed number. SINCE


THE PRECISION OF THE MEASUREMENTS REPORTED BY THE 3500 INSTRU-
MENT WILL ALWAYS MATCH THAT OF THE LINEARIZATION POINTS, IT IS
IMPORTANT THAT ALL "FORCE" ENTRIES BE EXPRESSED TO THE SAME
PRECISION (DECIMAL-POINT LOCATION).

7. Press ENTER. The unit will display "SET 1." You are now ready to establish
the endpoint of the first actual segment. Press ENTER once again to display
the "live" data reading.

8. Adjust the transducer load to equal a second known value (which will be "f1,"
the endpoint of the first linearization segment). Repeat Steps 5 through 8.

9. Continue to "force" as many linearization segments as desired (up to a total of


15).

10. If you want to modify an existing linearization table by "reforcing" a given line-
segment endpoint, you can do so by adjusting the "live" input until you reach
the line segment in question—i.e., until its number is displayed with the word
"SET." Then repeat Steps 3 through 6 and press ENTER.

11. After you have defined the last segment of your curve, press SETUP to return
to the "Setup" display.

NOTE: At any point during the "BY FORCE" procedure, you can shift to the "BY
TABLE" procedure described in Section 4.c.4(b), above, to verify the actual
numerical INPUT/OUTPUT values stored for each "forced" line-segment end-
point—and even to modify these values, if the "ideal" numbers are known. Before
moving to "TABLE" from "FORCE," however, you should first exit SETUP MODE
altogether in order to save the linearization points you have entered so far.

4.c.5 “CALCULATED” CALIBRATION


Remember that in "TWO-POINT" calibration (Sections 4.c.2 and 4.c.3.), the SCAL-
ING FACTOR ("m" coefficient) and ZERO OFFSET ("b" term) were automatically
computed as the two calibration points were established. The "CALCULATED"
calibration technique allows you to enter these two constants directly.*

4.c.5.a FOR THE MODELS 3510, 3530, 3540, AND 3560

As mentioned earlier, for the Model 3510 Thermocouple Conditioner or the


Model 3530 LVDT Conditioner, this calibration method is useful only if you want
to transfer the calibration of one instrument to another instrument that uses the
same transducer/cabling system.

For the Model 3540 Frequency Conditioner or the Model 3560 Voltage Con-
ditioner, it is generally convenient only when the instrument is used to measure
frequency or voltage itself (respectively).** In this case, you would calibrate the

* When calibrating a conditioner in this way, keep in mind that the resulting accuracy is limited to
either the stated "initial offset" accuracy or the stated "gain accuracy" of the conditioner,
whichever represents a greater error value (see Appendix A), and that the maximum attainable
unit resolution is ±32700, fixed decimal.
** "Calculated" calibration of the Model 3540 can be accomplished through the RS-232/485
Interface by means of the FRQ command, described in the next section.

4.16 4.c CALIBRATING THROUGH THE FRONT PANEL


SETUP: INSTRUMENT CALIBRATION 4
instrument simply by loading a SCALING FACTOR equal to the full-scale range to
which the instrument has been set, expressed in the desired engineering units
and to the desired measurement precision.

1. Once you have selected "CALCULATED" calibration as explained in Section


4.c.1, the unit will display "CALC":

2. Press ENTER. The unit will display "OFSET":

3. Press ENTER again. The unit will display the ZERO OFFSET ("b" term) current-
ly being applied to the data reading.

4. Use the NUMERIC BUTTONS as explained in Section 1.e to change the dis-
played number to the desired "b" value, if it does not already have this value.
A TARE OFFSET can be added to the measurement reading later, if required
(see Section 5.c).

5. Press ENTER. The display will now show the "live" data reading with the new
offset in effect.

6. Press CAL. The unit will display the word "SCALE":

7. Press ENTER to display the SCALING FACTOR currently being applied to the
data reading.

8. Use the NUMERIC BUTTONS as explained in Section 1.E to change the dis-
played value to the desired (full-scale) gain value.

9. Press ENTER. The display will now show the "live" data reading with the new
scaling in effect.

10. Press SETUP to return to the "Setup" display.

4.c.5.b FOR THE MODEL 3570 (ONLY)

For the Model 3570 DC Strain Gage Conditioner, this is a convenient method
when factory calibration data is known for your strain gage transducer. When
such is the case, "calculated" calibration is a faster and inherently more accurate
alternative to either "actual" or "simulated" two-point calibration—though the final
accuracy of calibration will depend, of course, on the accuracy of the transducer
manufacturer's specifications. As mentioned earlier, this technique should NOT
be used to calibrate the Model 3578 AC Strain Gage Conditioner.

1. Perform Steps 1 through 7 of the procedure given in Section 4.c.5(a), above.

2. Use the following equation to calculate an appropriate value of the SCALING


FACTOR ("m"):

4.c CALIBRATING THROUGH THE FRONT PANEL 4.17


4 SETUP: INSTRUMENT CALIBRATION

r u
m= x
0.75 i

where "i" is the manufacturer-supplied transducer sensitivity rating in "mV/V,


full scale"; "u" is the transducer's nominal full-scale rating in whatever engi-
neering units are desired; and "r" is the selected full-scale input range (0.75,
1.5, or 3.0 mV/V—see Section 3.a.5).

IMPORTANT: The value "u" in the above formula represents the transducer's
nominal full-scale rating only when the unit is set to an EXCITATION level of 10
V-DC. When the excitation is 5 V-DC, it is necessary to set "u" equal to twice
the nominal full-scale rating. When the excitation is 2 V-DC, you must set "u"
equal to five times the nominal full-scale rating.

3. Use the NUMERIC BUTTONS as explained in Section 1.e to change the dis-
played number to the "m" value calculated in Step 2.

4. Press ENTER. The display will now show the "live" data reading with the new
scaling in effect.

5. Press SETUP to return to the "Setup" display.

NOTE: The above calibration procedure can be accomplished through the RS-
232/485 Interface by means of the MV/V CALIBRATION (MVV) command,
described in the next section.

4.d CALIBRATING THROUGH THE RS-485 INTERFACE


When proper RS-232/485 communications have been established with an exter-
nal computer or terminal, SETUP commands can be issued to a 3500 Series
instrument for remotely controlled "TWO-POINT (DEADWEIGHT)" calibration
(either "actual" or "simulated"), 15-SEGMENT LINEARIZATION, and "CALCULAT-
ED" calibration.

For a full discussion of "Command and Response Syntax," along with a list of all
valid mnemonic commands, see Appendix B of this manual.

Only the "WRITE" form is given here for commands that have both a "READ" and a
"WRITE" form (there are a number of exceptions). Again, for full details, see
Appendix B. In the following command expressions, [CMT] is the "COMMAND
TERMINATOR" (that is, the INPUT TERMINATOR character which the unit has
been set to recognize—see the procedure for setting the INPUT TERMINATOR via
the COM button in Section 3.a.3, above).

4.d.1 SETTING THE ACTIVE CALIBRATION METHOD


CALIBRATION (CAL):
CAL=MXB [CMT] or CAL=LIN [CMT]
Sets the active calibration method to "MXB" (i.e., "TWO-POINT" or "CALCU-
LATED") or "LIN" ("15-SEGMENT LINEARIZATION"). This command is not rec-
ognized by the Model 3510 Thermocouple Conditioner, which is always in
the "MXB" mode.

4.18 4.d CALIBRATING THROUGH THE RS-485 INTERFACE


SETUP: INSTRUMENT CALIBRATION 4
4.d.2 “T WO-POINT” CALIBRATION
ZERO (ZRO): ZRO=z [CMT]
Sets the ZERO OFFSET ("b" term) that is to be used whenever "MXB" calibra-
tion is in effect such that the existing input yields a reading of "z" (full range is
±32700). When used in "TWO-POINT" calibration, must precede the FORCE
(FRC) command.
IMPORTANT: THIS COMMAND SHOULD ONLY BE APPLIED WHEN THE 3500
INSTRUMENT IS SET FOR "TWO-POINT" OR "CALCULATED" CALIBRATION—
THAT IS, WHEN THE CAL=MXB [CMT] COMMAND IS IN EFFECT. IF
ENTERED WHEN CAL=LIN [CMT], UNPREDICTABLE READINGS COULD
OCCUR. The "READ" form of the ZRO command will not be recognized.

FORCE (FRC): FRC=z [CMT]


Sets the SCALING FACTOR ("m" coefficient) that is to be used whenever
"MXB" calibration is in effect such that the existing input yields a reading of "z"
(full range is ±32700). Also sets desired precision (decimal-point location).
When used in "TWO-POINT" calibration, must be preceded by ZERO (ZRO)
command.
IMPORTANT: THIS COMMAND SHOULD ONLY BE APPLIED WHEN THE 3500
INSTRUMENT IS SET FOR "TWO-POINT" OR "CALCULATED" CALIBRATION—
THAT IS, WHEN THE CAL=MXB [CMT] COMMAND IS IN EFFECT. IF
ENTERED WHEN CAL=LIN [CMT], UNPREDICTABLE READINGS COULD
OCCUR. The "READ" form of the FRC command will not be recognized.

Note: The two following commands apply only to the Model 3570 DC STRAIN
GAGE CONDITIONER and the Model 3578 AC STRAIN GAGE CONDITIONER.

SHUNT CALIBRATE—POSITIVE:
SHP=ON [CMT] or SHP=OFF [CMT]
Closes or opens the instrument’s positive calibration shunt. SHP=ON [CMT]
should be preceded by a ZERO (ZRO) command (above) and should be fol-
lowed by a command of FRC=X [CMT], where "X" is the EQUIVALENT INPUT
produced by the shunt (to calculate "X," see Section 4.c.3). SHP=OFF [CMT]
should then be applied to resume normal measurement. See also Appendix
F for the interaction of SHP with the logic-input shunt commands.

SHUNT CALIBRATE—NEGATIVE:
SHN=ON [CMT] or SHN=OFF [CMT]
Closes or opens the instrument’s negative calibration shunt. SHN=ON [CMT]
should be preceded by a ZERO (ZRO) command (above) and should be fol-
lowed by a command of FRC=X [CMT], where "X" is the EQUIVALENT INPUT
produced by the shunt (to calculate "X," see Section 4.c.3). SHN=OFF [CMT]
should then be applied to resume normal measurement. See also Appendix
F for the interaction of SHN with the logic-input shunt commands.

4.d.3 15-SEGMENT LINEARIZATION


Your 3500 instrument need NOT be set to CAL=LIN [CMT] in order for the follow-
ing linearization-related commands to be effective. Commands that change the
linearization table will not affect present calibration as long as CAL=MXB [CMT].
The indicated changes will go into effect, however, as soon as the instrument is

4.d CALIBRATING THROUGH THE RS-485 INTERFACE 4.19


4 SETUP: INSTRUMENT CALIBRATION

placed in "LIN" calibration mode. Note too that these linearization commands do
not apply to the Model 3510 Thermocouple Conditioner.

LINEARIZER RESET (LNR): LNR [CMT]


An "IMPERATIVE" command that resets the internal linearization table. For the
effects of resetting the table, see Section 4.c.4(a), above. NOTE: Applying this
command will automatically set the active calibration mode to "MXB" (see
above).

LINEARIZATION SEGMENTS (LNS): LNS [CMT]


Reads the number of currently active linearization segments. Resetting the
linearization table sets the "LNS" number to zero— even though the definition
of each segment (now "inactive") that existed when the table was reset
remains in storage and can be retrieved by the "READ" form of the LIN-
EARIZE (LIN) command, below, until overwritten by subsequent linearization
entries.*

LINEARIZE (LIN): LIN=n,i,o [CMT]


Sets the effective endpoint of linearization Segment No. n, where "i" is the
INPUT value in millivolts, and "o" is the desired OUTPUT in desired engineer-
ing units. Replaces the INPUT/OUTPUT data currently in the linearization
table for Segment No. n with the specified "i" and "o" values.

NOTE: The "READ" form of the LINEARIZE (LIN) command requires that you
specify the segment in question, which need not be an "active" segment (see
the LNS command, above):
LINn [CMT]
where 0 ≤ n ≤ 15. This command will return the current INPUT/OUTPUT data
for Segment No. n in the format i, o [EOT].

LINEARIZATION FORCE (LFC): LFC=v [CMT]


Sets to "v" the OUTPUT value for the linearization segment corresponding to
the existing "live" INPUT (i.e., "forces" the existing input to read "v"). FOR
ERROR MESSAGES THAT CAN OCCUR WITH REGARD TO THE LFC COM-
MAND, SEE APPENDIX C. The "READ" form of the LFC command will not be
recognized.

4.d.4 “CALCULATED” CALIBRATION


IMPORTANT: THE BEE, EMM, FRQ, AND MVV COMMANDS SHOULD ONLY BE
APPLIED WHEN THE 3500 INSTRUMENT IS SET FOR "TWO-POINT" OR "CALCU-
LATED" CALIBRATION—THAT IS, WHEN THE CAL=MXB [CMT] COMMAND IS IN
EFFECT. IF ENTERED WHEN CAL=LIN [CMT], UNPREDICTABLE READINGS
COULD OCCUR.

ZERO OFFSET (BEE): BEE=b [CMT]


Sets to "b" the ZERO OFFSET ("b" term) that is to be used whenever "MXB"
calibration is in effect (full range is ±32700).

* Asking LNS [CMT] when the active calibration mode is "MXB" will tell you how many segments
are in the last-entered (now inactive) linearization table.

4.20 4.d CALIBRATING THROUGH THE RS-485 INTERFACE


SETUP: INSTRUMENT CALIBRATION 4
SCALING FACTOR (EMM): EMM=m [CMT]
Sets to "m" the SCALING FACTOR ("m" coefficient) that is to be used whenev-
er "MXB" calibration is in effect (full range is ±32700). Also sets desired preci-
sion (decimal-point location).

Note: The following command applies to the Model 3540 FREQUENCY CONDI-
TIONER only.

FREQUENCY CALIBRATION (FRQ) FRQ=i,u [CMT]


Sets an appropriate SCALING FACTOR for the Model 3540 Frequency Condi-
tioner when transducer calibration data is known; "i" is the manufacturer-sup-
plied full-scale rating of the frequency source (or the highest frequency
expected to be measured), in Hertz; "u" is the corresponding value of the
measured phenomenon, expressed in the desired engineering units and
measurement precision. You need not zero the conditioner in this case.
Note that in the answer returned in response to an FRQ [CMT] interrogation,
the displayed "i" and "u" values will have been automatically rescaled to reflect
the current input frequency range. Thus, for example, if the "8000 Hz" range is
in effect and the command FRQ=10000,8000 [CMT] has been entered, an
answer of "8000,6400" will be given to a subsequent query of FRQ [CMT]
(since 8000/10000 = 6400/8000).

Note: The following command applies to the Model 3570 DC STRAIN GAGE
CONDITIONER only.

MV/V CALIBRATION (MVV) MVV=i,u [CMT]


Sets an appropriate SCALING FACTOR for the Model 3570 DC Strain Gage
Conditioner when transducer calibration data is known; "i" is the transducer
sensitivity rating in "mV/V, full scale"; "u" is the nominal full-scale rating
expressed in desired engineering units and measurement precision, if the
instrument is set to an EXCITATION level of 10 V-DC. When the excitation is 5
V-DC, it is necessary to set "u" equal to twice the nominal full-scale rating.
When the excitation is 2 V-DC, you must set "u" equal to five times the nominal
full-scale rating.
Also note that in the answer returned in response to an MVV [CMT] interro-
gation, the displayed "i" and "u" values will have been automatically rescaled
to reflect the current "mV/V range" setting. Thus, for example, if the "3 mV/V"
range is in effect and the command MVV=2.500,5000 [CMT] has been
entered, an answer of "3.000, 6000" will be given to a subsequent query of
MVV [CMT] (since 3.000/2.500 = 6000/5000).

4.d CALIBRATING THROUGH THE RS-485 INTERFACE 4.21


4.22 This page intentionally blank.
RUN-TIME OPERATION 5
5.a RS-232/485 COMMUNICATIONS
5.a.1 RS-232/485 OUTPUTS:
CHN AND DMP COMMANDS
A 3500 Series instrument will issue output from its RS-232/485 Interface Port in
response to
• a command to "PRINT" from the front-panel PRINT button or the rear logic I/O
connector (when the interface is in RS-232 mode);
• a CHANNEL (CHN) or DUMP (DMP) command received through the Inter-
face Port (in either RS-232 or RS-485 mode);
• a valid interrogation command received through that port (in RS-232 mode),
or
• any valid or invalid command received through that port (in RS-485 mode).

For the initiation of RS-232 PRINT transmissions, see Section 5.d, below. For pos-
sible responses to mnemonic commands, see Appendix B.

The general form of the CHANNEL (CHN) command is


CHN [CMT]
where [CMT] is the "COMMAND TERMINATOR" (that is, the INPUT TERMINATOR
character which the 3500 instrument has been set to recognize—see Section
3.a.3). This command instructs the unit to transmit the current data for its single
measurement channel from the RS-232/485 Interface Port, in the format selected
by the user.

The general form of the DUMP (DMP) command is


DMP [CMT]
This command is functionally identical to the CHN command.

5.a.2 DATA-TRANSMISSION FORMAT


The data transmitted when the 3500 instrument is instructed to "PRINT" (in RS-232
mode) or when a CHANNEL (CHN) or DUMP (DMP) command has been
received through the RS-232/485 Interface Port (in either RS-232 or RS-485
mode) can include

a. An optional "HEADER" string defined by a LABEL (LBL) command of


LBL=$ [CMT]
where "$" is a string of up to eight characters, including spaces.* The data
"label" is typically a channel name or description (e.g., CHAN 4, INPUT 7, TEST
R, etc.). It has no numeric effect on measurement scaling.

b. An optional NODE-NUMBER "ECHO," if so indicated by the user during the


PRINT button setup procedure (Section 3.a.10) or by application of an ECHO
(ECO) command of
ECO=ON [CMT]

* To remove an existing LBL string from the transmission, just command LBL=N/A [CMT].

5.a RS-232/485 COMMUNICATIONS 5.1


5 RUN-TIME OPERATION

c. An optional LIMIT-STATUS INDICATION number of "-1," "0," or "1," if so indicat-


ed by the user during the PRINT button setup procedure (Section 3.a.10) or
by application of a LIMITS (LIM) command of
LIM=ON [CMT]

d. An optional "TAILER" string defined by an ENGINEERING UNIT STRING


(EUS) command of
EUS=$ [CMT]
where "$" is a string of up to eight characters, including spaces.* For the unit
string, you could have, for example, INCHES, FT-LBS, etc. Like LBL, it has no
numeric effect on measurement scaling.

With all optional elements included, the transmission format is


[HEADER]n,w,s[TAILER]
where "n" is the optional NODE NUMBER (0-99), "w" is the data value, and "s" is
the optional LIMIT-STATUS INDICATION number ("-1," "0," or "1"). Notice that there
are no spaces separating the optional LBL-defined HEADER prefix and EUS-
defined TAILER suffix from the rest of the output.

5.a.3 “OPENING” A NODE TO RECEIVE


COMMANDS FROM THE COMPUTER
The RS-232/485 Interface Port will recognize any of the standard mnemonic com-
mands listed in Appendix B. However, WHEN SET TO THE RS-485 ("MULTINODE")
MODE, an individual node will receive and respond to computer-issued com-
mands ONLY WHEN IT IS THE CURRENTLY "OPEN" NETWORK NODE. The com-
puter can "open" any selected node by issuing an OPEN (OPN) command to the
entire network. This command has the general form
OPNn [CMT]
where "n" is the nonzero NODE NUMBER that has been assigned to the node the
computer wishes to address.

IMPORTANT: With regard to the OPN command, please note these two rules:
a. There must be NO SPACE between the letters "OPN" and the node number
"n."
b. The node number "n" must have NO LEADING ZERO(S)—e.g., the node num-
ber "1" cannot be entered as "01."

Only one network node can be "open" at any time. While "opening" Node No. "n"
for communication with the computer, the above command at the same time
"closes" (that is, prohibits communication with) all other nodes. Only Node No. n
will respond to subsequent commands issued to the network, until another node
is "opened."

* To remove an existing EUS string from the transmission, just command EUS=N/A [CMT].

5.2 5.a RS-232/485 COMMUNICATIONS


RUN-TIME OPERATION 5
5.b PEAK CAPTURE
During setup, a 3500 Series instrument can be placed in either PEAK CAPTURE
or TRACK/HOLD mode, as follows:

1. Enter SETUP MODE by pressing the SETUP button and entering the current
security code, if necessary (as explained in Section 3.a.1).

2. Press the PEAK/TRACK button. The current button mode will be displayed:
either “P” (for PEAK CAPTURE) or “H” (for TRACK/ HOLD). To toggle between
the two settings, simply press the top or bottom segment of the displayed let-
ter.

3. When the desired mode letter is displayed, you can exit SETUP MODE by
pressing the SETUP button once again. For a description of 3500
TRACK/HOLD operation, see Section 5.e.

When a 3500 instrument is in PEAK CAPTURE mode, you can control the peak-
capture function in any one of three ways:

1. By the front-panel PEAK/TRACK button (WHEN THE "PEAK" TERMINAL OF


THE REAR LOGIC I/O CONNECTOR IS AT "LOGIC 0"). Pressing this button
once will place the unit in "+PEAK" MODE, if the "PEAK" terminal is disconnect-
ed (i.e., at "Logic 0").* Pressing the button once more will return the unit to
"TRACK" MODE.

2. By a PEAK (PEK) command applied via the RS-232/485 Interface (WHEN THE
"PEAK" TERMINAL OF THE REAR LOGIC I/O CONNECTOR IS AT "LOGIC 0"). A
command of
PEK=ON [CMT]
will place the unit in "+PEAK" MODE, if the "PEAK" terminal is disconnected
(i.e., at "Logic 0").* A command of
PEK=OFF [CMT]
will return the unit to "TRACK" MODE.

3. By a LOGIC-STATE TRANSITION at the "PEAK" terminal of the rear Logic I/O


Connector. Thus, a transition from Logic 0 to Logic 1 will place the unit in
"+PEAK" MODE, while a transition from Logic 1 to Logic 0 will return it to
"TRACK" MODE, regardless of the last-entered PEAK (PEK) command or the
current state of the PEAK/TRACK button.

Fig. 22 shows the capture and hold of successively higher-valued signal maxima,
after entering the "+PEAK" MODE at time t1. Until time t1, the "live" data reading
had been continuously "tracking" the analog input. After time t1, it continuously
reports the highest input-signal value perceived since "+PEAK" last began. From
time t1 to time t2, the input is continuously rising, and so the reading appears to be
continuing to track it. At time t2, however, the input signal reaches its first true
maximum since time t1. The instrument "captures" this "positive" peak (P1), hold-
ing it as a constant until time t3, when a yet higher input value is detected, and the
reading begins once more to track the input upwards to a yet higher peak (P2).

* When the "PEAK" terminal is connected to GROUND (i.e., at the "Logic 1" level), both the
PEAK/TRACK button and the PEAK (PEK) command are disabled.

5.b PEAK CAPTURE 5.3


5 RUN-TIME OPERATION

P2
+PEAK
READING

P1

ANALOG INPUT

TRACK + PEAK
t0 t1 t2 t3 t4

Fig. 22 Capture and Hold of Successively Higher-Valued Maxima

ANALOG
INPUT +PEAK READING
P1 P3

P2
RESET

+ PEAK + PEAK
t0 t1 t2 t3 t4

Fig. 23 Capture and Hold of Successively


Lower-Valued Maxima Using Peak Reset

Fig. 23 shows the capture of successively lower-valued signal maxima. Here, it is


necessary to reset the "+PEAK" reading—to get it "back on track," so to speak—
somewhere along the rise of the input toward the second, lower-valued peak (P2).
This is done by returning the unit momentarily to "TRACK" MODE at time t2.

5.4 5.b PEAK CAPTURE


RUN-TIME OPERATION 5
5.c TARE AND RESET
You can control a 3500 instrument's tare function in any one of three ways:

1. By the front-panel TARE/RESET button (WHEN THE "TARE" TERMINAL OF


THE REAR LOGIC I/O CONNECTOR IS AT "LOGIC 0"). Pressing this button
once will place the unit in "TARE" MODE, if the "TARE" terminal is disconnected
(i.e., at "Logic 0").* Pressing the button once more will return the unit to a dis-
play of "untared" data.

2. By a TARE (TAR) command applied via the RS-232/485 Interface (WHEN THE
"TARE" TERMINAL OF THE REAR LOGIC I/O CONNECTOR IS AT "LOGIC 0"). A
command of
TAR=ON [CMT]
will place the unit in "TARE" MODE, if the "TARE" terminal is disconnected (i.e.,
at "Logic 0").* A command of
TAR=OFF [CMT]
will return the unit to a display of "untared" data.

3. By a LOGIC-STATE TRANSITION at the "TARE" terminal of the rear Logic I/O


Connector. Thus, a transition from Logic 0 to Logic 1 will place the unit in
"TARE" MODE, while a transition from Logic 1 to Logic 0 will return it to an
"untared" data reading, regardless of the last-entered TARE (TAR) command
or the current state of the TARE/RESET button.

Fig. 24 shows the effect of "TARE" MODE on the displayed data reading. Until
time t1, this reading is "untared" (there is no constant offset). At time t1, the read-
ing is brought down to the current value of the TARE REGISTER. From this point, it
continues to "track" the analog input, but now with the continuous application of
the constant TARE OFFSET determined by the difference between the input value
at time t1 and the TARE REGISTER in effect at that time. At time t2, the instrument
returns to its normal "untared" data reading.

ANALOG INPUT Fig. 24 Tare


Offset Operation

TARE TARED DATA READING


OFFSET
VALUE

TARE
REGISTER
VALUE
TAR =
TAR = OFF TAR = ON OFF
t0 t1 t2

* When the "TARE" terminal is connected to GROUND (i.e., at the "Logic 1" level), both the
TARE/RESET button and the TARE (TAR) command are disabled.

5.c TARE AND RESET 5.5


5 RUN-TIME OPERATION

5.d INITIATING HARD-COPY PRINTOUTS


There are two ways you can cause your 3500 instrument to output formatted data
from its RS-232/485 Interface Port according to the time interval specified by the
operator (Section 3.a.10), WHEN THAT PORT IS SET TO RS-232 MODE*:

1. By the front-panel PRINT button WHEN THE "PRINT" TERMINAL OF THE REAR
LOGIC I/O CONNECTOR IS AT "LOGIC 0" AND THE FOLLOWING PRINT
ENABLE (PRN) COMMAND IS IN EFFECT:
PRN=ON [CMT]
Pressing this button once will initiate RS-232 transmissions at the specified
interval. Pressing it once more will halt any transmission in progress at the
end of the line currently being sent.

2. By a LOGIC-STATE TRANSITION at the "PRINT" terminal of the rear Logic I/O


Connector WHEN THE PRN=ON [CMT] COMMAND IS IN EFFECT. Thus, a
transition from Logic 0 to Logic 1 at the "PRINT" terminal will initiate RS-232
transmissions at the specified interval, while a transition from Logic 1 to Logic
0 will halt any transmission in progress.

When the print transmissions are started by setting the "PRINT" terminal to Logic
1, the PRINT button is disabled, and a transmission can only be stopped by reset-
ting the "PRINT" terminal to Logic 0 or by issuing a command of
PRN=OFF [CMT]

As you can see, the PRINT ENABLE (PRN) command serves only to enable (or
disable) the automatic print function. It does not itself initiate or halt data trans-
missions, and has no effect on the operation of the CHANNEL (CHN) or DUMP
(DMP) commands.

NOTE: The PRN setting is not saved to EEPROM. THIS IS STRICTLY A "RUN-
TIME" COMMAND, THE POWERUP DEFAULT BEING ALWAYS PRN=ON [CMT].
Therefore, the unit is always ready to print as soon as it is turned on, and you can
enable (or re-enable) the print function at any time just by turning the unit off and
on.

* The front-panel PRINT button and the automatic print function are disabled when the unit is
in RS-485 ("MULTINODE") MODE. To send formatted data only once to a connected printer,
you can always use the CHANNEL (CHN) or DUMP (DMP) command as explained in Sec-
tion 5.a.

5.6 5.d INITIATING HARD-COPY PRINTOUTS


RUN-TIME OPERATION 5
5.e TRACK/HOLD FUNCTION
See Section 5.b, above, for instructions on setting a 3500 Series instrument to
TRACK/HOLD mode. When a 3500 instrument is in TRACK/HOLD mode, it can be
commanded to freeze the analog input signal in any of three ways:

1. By the front-panel PEAK/TRACK button (WHEN THE "PEAK" TERMINAL OF


THE REAR LOGIC I/O CONNECTOR IS AT "LOGIC 0"). Pressing this button
once will place the unit in "HOLD" MODE, if the "PEAK" terminal is disconnect-
ed (i.e., at "Logic 0").* The existing analog input reading will be held without
decay, regardless of the subsequent behavior of the input signal. Pressing
the button once more will return the unit to "TRACK" MODE.

2. By a PEAK (PEK) command applied via the RS-232/485 Interface (WHEN THE
"PEAK" TERMINAL OF THE REAR LOGIC I/O CONNECTOR IS AT "LOGIC 0"). A
command of
PEK=ON [CMT]
will place the unit in "HOLD" MODE, if the "PEAK" terminal is disconnected (i.e.,
at "Logic 0").* A command of
PEK=OFF [CMT]
will return the unit to "TRACK" MODE.

3. By a LOGIC-STATE TRANSITION at the "PEAK" terminal of the rear Logic I/O


Connector. Thus, a transition from Logic 0 to Logic 1 will place the unit in
"HOLD" MODE, while a transition from Logic 1 to Logic 0 will return it to
"TRACK" MODE, regardless of the last-entered PEAK (PEK) command or the
current state of the PEAK/TRACK button.

* When the "PEAK" terminal is connected to GROUND (i.e., at the "Logic 1" level), both the
PEAK/TRACK button and the PEAK (PEK) command are disabled.

5.e TRACK/HOLD FUNCTION 5.7


5.8 This page intentionally blank.
3500 SERIES SPECIFICATIONS A

Appendix A
3500 Series Specifications

A.1 GENERAL SPECIFICATIONS


(These specifications apply equally to all 3500 Series models.)

Physical: DIN package outline of extruded metal, with splash-resistant front


panel; secure rear connections via screw terminals. Instrument weight approx-
imately 3.25 lb (1.47 kg). For dimensions, see Fig. 25, below.
Power Requirements:
Input Voltage Range: 90-265 V-AC
Frequency Range: 50-400 Hz
Power In (Max): 10 W
Environmental:
ESD Immunity: Meets Class 3 Industrial Standard (8 kV) everywhere on the
case and on every line
Operating Temperature Range: 0° C to +50° C (+32° F to +122° F)
Storage Temperature Range: -40° C to +80° C (-40° F to +176° F)
Operating Humidity Range: 10 to 95% max, noncondensing
Display: 5-digit LCD reflective, non-light-emitting
A/D Conversion: 16-bit; approximately 1-kHz conversion rate; transparent auto-
calibration every 2 seconds
Digital Filtering selectable from front panel
Limits: Three limit zones ("HI," "OK," "LO") with front-panel annunciation and cor-
responding TTL-level logic outputs (outputs for HIGH and LOW limit conditions
can be latching or nonlatching). Each limit setpoint has a user-settable hys-
teresis window to prevent toggling of the limit output from signal noise.
Serial Communications: 9-pin RS-232/RS-485 Port (interface mode selectable
by node-number setting); standard baud rates up to 153.6K for both RS-232

.46 .09
1.17 .23 Allow 2.5 in.
(6.4 cm) in
rear for cable
bend
COM RANGE CAL FILTER ANO DEC LIMIT

HI
2.84
OK
7.21
LO

PEAK TARE
SET UP PRINT ENTER
TRACK RESET

DIMENSIONS
IN INCHES/CM
5.68 6.60
14.43 16.76
.13 .09
.33 .23

Fig. 25 3500 Physical Dimensions

A.1 GENERAL SPECIFICATIONS A.1


A 3500 SERIES SPECIFICATIONS

and RS-485. RS-485 configuration allows operation as an individual data-col-


lection "node" within a computer-controlled network. NOTE: RS-485 conver-
sion is required at the computer's I/O port.
Logic Inputs and Outputs: TTL- and CMOS-compatible; isolated (±1500 V) from
power and communication ports. Negative-true outputs for limit monitoring
(isolated +5 V provides up to 50 mA for driving external devices directly); nega-
tive-true inputs for control of PRINT, TARE, PEAK, and UNLATCH functions.
Inputs*: Logic Volts Without Instru-
Level (Operating) ment Damage

0 0.9 V (max) –0.5 V (min)


1 2.5 V (min) 5.5 V (max)
Outputs: Logic
Level Voltage at Current

0 0.45 V (max) @ 10 mA (sink)


1 2.6 V (min) @ 0.5 mA (source

A.2 INDIVIDUAL CONDITIONER SPECIFICATIONS


A.2.a MODEL 3510 THERMOCOUPLE CONDITIONER
Transducer Types: E, J, K, R, S ,T, and B Thermocouples (see table)
TC Display Typical Maximum
Type Range Resolution Accuracy** Error**

E -130° C to +1000° C 0.1° C ±0.6° C ±0.8° C


J -130° C to +750° C 0.1° C ±0.8° C ±1.3° C
K -75° C to +1350° C 0.1° C ±1.1° C ±1.5° C
R -20° C to +1750° C 0.1° C ±1.3° C ±2.6° C
S -20° C to +1750° C 0.1° C ±1.3° C ±2.6° C
T -130° C to +400° C 0.1° C ±0.7° C ±1.1° C
B +480° C to +1820° C 0.1° C ±1.3° C ±2.6° C
E -200° F to +1800° F 0.2° F ±1.0° F ±1.4° F
J -200° F to +1400° F 0.2° F ±1.4° F ±2.2° F
K -100° F to +2500° F 0.2° F ±1.8° F ±2.6° F
R 0° F to +3200° F 0.2° F ±2.2° F ±4.6° F
S 0° F to +3200° F 0.2° F ±2.2° F ±4.6° F
T -200° F to +750° F 0.2° F ±1.2° F ±1.8° F
B +900° F to +3300° F 0.2° F ±2.2° F ±4.6° F

Linearization: Digital look-up table, ±0.05%


Reference-Junction Compensation: At connector block, using built-in precision
thermistor
Thermocouple Break Detection: Off-scale positive or negative, user-selectable
Normal-Mode Range: ±80 mV operating, ±5 V without instrument damage
Common-Mode Range: ±40 V operating, ±50 V without instrument damage

* The minimum time allowed between activation and reactivation of any logic input is 100 mil-
liseconds.
** Including ± one count of least significant digit displayed (i.e., ± 0.2 or 0.1). Can be readily
improved by control of instrument temperature, calibrating at known temperatures, etc.

A.2 A.2 INDIVIDUAL CONDITIONER SPECIFICATIONS: MODEL 3510


3500 SERIES SPECIFICATIONS A
Common-Mode Rejection Ratio: -116 dB at DC; -120 dB at 60 Hz
Input Impedance: Differential: 10 megohms; Common-Mode: 0.5 megohm
Offset: Initial: ±0.005 mV; vs. Temperature: ±0.0001 mV/°C; vs. Time: ±0.001
mV/month
Gain Accuracy: ±0.05% of absolute mV input range of -10 to +80 mV
Gain Stability: Vs. Temperature: ±50 ppm/°C; vs. Time: ±20 ppm/month
Analog Filtering for Displayed Reading: 2-pole modified Butterworth; 3 dB down
at 1 Hz; 60 dB down at 50 Hz
Step Response Settling Time (Full-Scale Output)
To 1% of final value: 1.0 second
To 0.1% of final value: 2 seconds
To 0.02% of final value: 6 seconds
Total System Accuracy (typical): See table, above
Digital Peak Capture: Displays the most positive measurement reading, which is
digitally held for indefinite display; PEAK and TRACK modes controlled by front-
panel button, rear-panel logic input, or computer-port command
Analog Output: ±5 V, microprocessor driven and scaled to user-specified range;
±1 mV resolution
Accuracy: 0.1% of current voltage reading ±2 mV
Output Bandwidth: 1 Hz maximum*
Frequency Characteristics and Step-Response Settling Times: See "Analog
Filtering for Displayed Reading" for the Model 3510, above

A.2.b MODEL 3530 LVDT CONDITIONER


Full-Scale Input Ranges: "Normal": 0-150, 0-300, and 0-600 mV/V; "Long-stroke":
0-1, 0-2, and 0-4 V/V
Excitation: Nominal 3 V-AC (rms) at 3280 Hz; 40 mA max
Common-Mode Range: ±5 V operating, ±12 V without instrument damage
Common-Mode Rejection Ratio: Infinite at DC and 60 Hz; -60 dB at 3 kHz
Input Impedance: Normal-Mode: 400 kΩ; Common-Mode: 100 kΩ
Offset: Initial: ±0.25% of full scale; vs. Temperature: ±20 ppm/°C; vs. Time: ±10
ppm/month
Gain Accuracy: ±0.02% of full scale ± 1 count LSD, typical, following calibration
Gain Stability: Vs. Temperature: ±50 ppm/°C; vs. Time: ±20 ppm/month
Analog Filtering for Displayed Reading: 5-pole filter with selectable low-pass
corner frequency of 5, 10, or 20 Hz
Corner Frequency
Response at . . . 5 Hz 10 Hz 20 Hz
- 3 dB 5 Hz 10 Hz 20 Hz
-60 dB 32 Hz 65 Hz 125 Hz
Step Response Settling Time
(Full-Scale Output) 5 Hz 10 Hz 20 Hz
To 1% of final value 250 msec 125 msec 60 msec
To 0.1% of final value 400 msec 200 msec 100 msec
To 0.02% of final value 600 msec 300 msec 170 msec

* Due to analog input bandwidth, settling times are input-limited also.

A.2 INDIVIDUAL CONDITIONER SPECIFICATIONS: MODEL 3530 A.3


A 3500 SERIES SPECIFICATIONS

Analog Peak Capture: Positive analog peak with minimum full-scale input pulse
duration of 15 msec (to 1% of full-scale accuracy), 25 msec (to 0.1% of full-scale
accuracy), and 35 msec (to 0.02% of full-scale accuracy); digitally held for indefi-
nite display; PEAK and TRACK modes controlled by front-panel button, rear-panel
logic input, or computer-port command
Internal 15-Segment Linearization, programmable from front panel
Analog Output: Range scalable in 0.1% increments between 74.5% and 125.5% of
nominal input. May be sourced by conditioned input (A) after the FIXED ANALOG
FILTER; (B) after PEAK CAPTURE; or (C) after the SELECTABLE ANALOG FILTER
(see Fig. 15, Section 2.d)
Frequency Characteristics and Step-Response Settling Times:
AFTER THE FIXED ANALOG FILTER and AFTER +PEAK IN "PEAK" or
"TRACK" MODE: 3 dB down at 100 Hz; 60 dB down at 750 Hz. Settling Time
to 1% of final value: 15 msec; to 0.1% of final value: 25 msec; to 0.02% of final
value: 35 msec
AFTER THE SELECTABLE ANALOG FILTER: See "Analog Filtering for Dis-
played Reading" for the Model 3530, above
Configuration: Single-ended, return to SYSTEM COMMON
Full-Scale Range: ±5 V nominal, ±8 V maximum
Offset Range Adjustment: ±25.5% in 0.1% increments
Allowable Loading: 5 mA, max.
Offset Accuracy: 0.1% max
Span Accuracy: 0.2% max
Offset and Span Drift: ±50 ppm/°C; ±20 ppm/month

A.2.c MODEL 3540 FREQUENCY INPUT CONDITIONER


Input Type: Any AC or unipolar pulse signal, grounded or floating, irrespective of
waveform
Frequency Range: From 10% to 100% of 250, 500, 1000, 2000, 4000, 8000, 16000,
or 32000 Hz
Threshold Level: Selectable to accommodate signals from 100 mV to 100 V
Excitation (for "Zero-Velocity" sensors): Nominal 10 (i.e., ±5) V-DC ± 5%; ±50 mA
maximum
Normal-Mode Range: ±100 V operating, ±150 V without instrument damage
Common-Mode Range: ±100 V operating and without instrument damage
Common-Mode Rejection Ratio: -120 dB at 60 Hz; -60 dB at 1 kHz
Input Impedance: Normal-Mode: 200 kΩ; Common-Mode: 50 kΩ
Offset: Initial: ±0.05% of full scale; vs. Temperature: ±25 ppm/°C; vs. Time: ±20
ppm/month
Gain Accuracy: ±0.02% of full scale
Gain Stability: Vs. Temperature: ±40 ppm/°C; vs. Time: ±20 ppm/month
Analog Filtering for Displayed Reading: 5-pole filter with selectable low-pass
corner frequency of 2.5, 5, or 10 Hz

A.4 A.2 INDIVIDUAL CONDITIONER SPECIFICATIONS: MODEL 3540


3500 SERIES SPECIFICATIONS A
Corner Frequency
Response at . . . 2.5 Hz 5 Hz 10 Hz
- 3 dB 2.5 Hz 5 Hz 10 Hz
-60 dB 15 Hz 32 Hz 65 Hz
Step Response Settling Time
(Full-Scale Output) 2.5 Hz 5 Hz 10 Hz
To 1% of final value 450 msec 275 msec 150 msec
To 0.1% of final value 850 msec 425 msec 225 msec
To 0.02% of final value 1.3 sec 650 msec 350 msec

Digital Peak Capture: Displays the most positive measurement reading, which is
digitally held for indefinite display; PEAK and TRACK modes controlled by front-
panel button, rear-panel logic input, or computer-port command
Internal 15-Segment Linearization, programmable from front panel
Analog Output: Range scalable in 0.1% increments between 74.5% and 125.5% of
nominal input
Frequency Characteristics and Step-Response Settling Times: See "Analog
Filtering for Displayed Reading" for the Model 3540, above
Configuration, Full-Scale Range, Offset Range Adjustment, Allowable
Loading, Offset Accuracy, Span Accuracy, Offset and Span Drift: See the
Model 3530 LVDT Conditioner, above

A.2.d MODEL 3560 VOLTAGE CONDITIONER


Full-Scale Input Ranges: ±0.5, ±1.0, ±2.0, ±5.0, ±10.0, and ±20.0 V-DC
Excitation: ±12 V-DC ± 5%, ±40 mA maximum
Common-Mode Range: ±50 V operating, ±100 V without instrument damage
Common-Mode Rejection Ratio: -70 dB at DC; -90 dB at 60 Hz and 3 kHz; -80 dB
a 1 kHz
Input Impedance: Differential: 200 kΩ; Common-Mode: 50 kΩ
Offset: Initial: ±0.02% of full scale; vs. Temperature: ±25 ppm/°C; vs. Time: ±10
ppm/month
Gain Accuracy*: ±0.02% of full scale ± 1 count LSD, typical, following calibration
Gain Stability: Vs. Temperature: ±50 ppm/°C; vs. Time: ±20 ppm/month
Analog Filtering for Displayed Reading: See the Model 3530 LVDT Conditioner,
above
Analog Peak Capture: Positive analog peak with minimum full-scale input pulse
duration of 6 msec (to 1% of full-scale accuracy), 12 msec (to 0.1% of full-scale
accuracy), and 20 msec (to 0.02% of full-scale accuracy); digitally held for indefi-
nite display; PEAK and TRACK modes controlled by front-panel button, rear-panel
logic input, or computer-port command
Internal 15-Segment Linearization, programmable from front panel
Analog Output: Range scalable in 0.1% increments between 74.5% and 125.5% of
nominal input. May be sourced by conditioned input (A) after the FIXED ANALOG
FILTER; (B) after PEAK CAPTURE; or (C) after the SELECTABLE ANALOG FILTER
(see Fig. 15, Section 2.d)

* Initial (uncalibrated) inaccuracy may be as great as ±0.04% of full scale. Maximum error that
could occur upon replacement of the 3560 not followed by calibration is ±0.08% of full scale.

A.2 INDIVIDUAL CONDITIONER SPECIFICATIONS: MODEL 3560 A.5


A 3500 SERIES SPECIFICATIONS

Frequency Characteristics and Step-Response Settling Times:


AFTER THE FIXED ANALOG FILTER or AFTER +PEAK IN "TRACK" MODE:
3 dB down at 2 KHz; 60 dB down at 16 kHz. Settling Time to 1% of final value:
0.6 msec; to 0.1% of final value: 0.75 msec; to 0.02% of final value: 1.25 msec
AFTER +PEAK IN "PEAK" MODE: Settling Time to 1% of final value: 6 msec; to
0.1% of final value: 12 msec; to 0.02% of final value: 20 msec
AFTER THE SELECTABLE ANALOG FILTER: See "Analog Filtering for Dis-
played Reading" for the Model 3560, above
Configuration, Full-Scale Range, Offset Range Adjustment, Allowable
Loading, Offset Accuracy, Span Accuracy, Offset and Span Drift: See the
Model 3530 LVDT Conditioner, above

A.2.e MODEL 3570 DC STRAIN GAGE CONDITIONER


Transducer Types and Ranges: Conventional 4-arm strain gage bridge, nominal
120 Ω or higher, with a full-scale range of 0.75, 1.5, or 3.0 mV/V. 1/4-, 1/2-, or full-
bridge gage configurations can be accommodated by appropriate external
bridge-completion circuitry supplied by the user. Since channel zeroing is by dig-
ital techniques, no input balance control is provided. The allowable input range,
therefore, must include any initial unbalance (which, in commercially produced
strain gage transducers, is negligible). Other transducers may have to be exter-
nally trimmed to be used with the instrument, if zero unbalance exceeds 20% of
full scale.
Actual Absolute Input at which
Normal-Mode Maximum Display Reading
Nominal Input-Signal Overrange Corresponds
Input Range Voltage (Max) (50%) to "m" Term
3.0 mV/V 45 mV 4.5 mV/V 4.0 mV/V
1.5 mV/V 22.5 mV 2.25 mV/V 2.0 mV/V
0.75 mV/V 11.25 mV 1.125 mV/V 1.0 mV/V

DC Excitation: User-selectable 10 V (±5 V), 5 V (±2.5 V), or 2 V (±1 V), 85 mA maxi-


mum
Normal-Mode Range: ±45 V operating, ±8 V without instrument damage
Common-Mode Range: ±0.50 V operating, ±8 V without instrument damage
Common-Mode Rejection Ratio: -60 dB at DC; -90 dB at 60 Hz, 1 kHz, and 3 kHz
Input Impedance (Differential and Common-Mode): Greater than 100 MΩ
Offset: Initial: ±0.025% of full scale; vs. Temperature: ±25 ppm/°C; vs. Time: ±10
ppm/month
Gain Accuracy*: ±0.02% of full scale ± 1 count LSD, typical, following calibration
Gain Stability: Vs. Temperature: ±50 ppm/°C; vs. Time: ±20 ppm/month
Analog Filtering for Displayed Reading: See the Model 3530 LVDT Conditioner,
above
Analog Peak Capture: See the Model 3560 Voltage Conditioner, above
Internal 15-Segment Linearization, programmable from front panel

* Initial (uncalibrated) inaccuracy may be as great as ±0.05% of full scale. Maximum error that
could occur upon replacement of the 3570 not followed by calibration is ±0.1% of full scale.

A.6 A.2 INDIVIDUAL CONDITIONER SPECIFICATIONS: MODEL 3570


3500 SERIES SPECIFICATIONS A
Analog Output: Range scalable in 0.1% increments between 74.5% and 125.5% of
nominal input.* May be sourced by conditioned input (A) after the FIXED ANA-
LOG FILTER; (B) after PEAK CAPTURE; or (C) after the SELECTABLE ANALOG FIL-
TER (see Fig. 15, Section 2.d)
Frequency Characteristics and Step-Response Settling Times:
AFTER THE FIXED ANALOG FILTER or AFTER +PEAK IN "TRACK" MODE:
3 dB down at 2 KHz; 60 dB down at 16 kHz. Settling Time to 1% of final value:
0.6 msec; to 0.1% of final value: 0.75 msec; to 0.02% of final value: 1.25 msec
AFTER +PEAK IN "PEAK" MODE: Settling Time to 1% of final value: 6 msec; to
0.1% of final value: 12 msec; to 0.02% of final value: 20 msec
AFTER THE SELECTABLE ANALOG FILTER: See "Analog Filtering for Dis-
played Reading" for the Model 3570, above
Configuration, Full-Scale Range, Offset Range Adjustment, Allowable
Loading, Offset Accuracy, Span Accuracy, Offset and Span Drift: See the
Model 3530 LVDT Conditioner, above

A.2.f MODEL 3578 AC STRAIN GAGE CONDITIONER


Transducer Types and Ranges: Conventional 4-arm strain gage bridge, nominal
90 Ω or higher, with a full-scale range of 0.75, 1.50, or 3.00 mV/V. 1/4-, 1/2-, or full-
bridge gage configurations can be accommodated by appropriate external
bridge-completion circuitry supplied by the user. Since channel zeroing is by dig-
ital techniques, no input balance control is provided. The allowable input range,
therefore, must include any initial unbalance (which, in commercially produced
strain gage transducers, is negligible). Other transducers may have to be exter-
nally trimmed to be used with the instrument, if zero unbalance exceeds 20% of
full scale.
Actual Absolute Input at which
Normal-Mode Maximum Display Reading
Nominal Input-Signal Overrange Corresponds
Input Range Voltage (Max) (33%) to "m" Term
3.00 mV/V 13.5 mV 4.0 mV/V 4.0 mV/V
1.50 mV/V 6.75 mV 2.0 mV/V 2.0 mV/V
0.75 mV/V 3.375 mV 1.0 mV/V 1.0 mV/V

AC Excitation: Regulated 3 V-AC (rms) at 3280 Hz; 50 mA (rms), maximum


Amplifier: AC-coupled demodulator with user-settable phase and symmetry con-
trols
Normal-Mode Range: ±12 mV rms, operating; ±8 V-DC without instrument dam-
age
Common-Mode Range: ±0.8 V-DC operating, ±8 V-DC without instrument damage
Common-Mode Rejection Ratio: infinite at DC; -90 dB at 60 Hz; -80 dB at 1 kHz;
-60 dB at 3 kHz
Input Impedance (Differential and Common-Mode): 10 MΩ
Offset: Initial: ±3% of full scale; vs. Temperature: ±0.005% f.s./°C; vs. Time: ±0.002%
f.s./month

* When (1) the analog output scale is 100.0%, (2) the analog output offset is 0.0%, and (3) the
excitation is 10V, then any of the following inputs will result in an analog output of 5 V: an input
of 4 mV/V on the 3.0-mV/V range; an input of 2 mV/V on the 1.5-mV/V range; or an input of 1
mV/V on the 0.75-mV/V range.

A.2 INDIVIDUAL CONDITIONER SPECIFICATIONS: MODEL 3578 A.7


A 3500 SERIES SPECIFICATIONS

Gain Accuracy*: ±0.025% of full scale,typical, following calibration


Gain Stability: Vs. Temperature: ±50 ppm/°C; vs. Time: ±20 ppm/month
Analog Filtering for Displayed Reading: 5-pole filter with selectable low-pass
corner frequency of 5, 10, or 20 Hz
Corner Frequency
Response at . . . 5 Hz 10 Hz 20 Hz
- 3 dB 5 Hz 10 Hz 20 Hz
-60 dB 32 Hz 65 Hz 125 Hz
Step Response Settling Time
(Full-Scale Output) 5 Hz 10 Hz 20 Hz
To 1% of final value 250 msec 125 msec 100 msec
To 0.1% of final value 400 msec 200 msec 150 msec
To 0.02% of final value 600 msec 300 msec 225 msec

Analog Peak Capture: Positive analog peak with minimum full-scale input pulse
duration of 100 msec (to 1% of full-scale accuracy), 150 msec (to 0.1% of full-
scale accuracy), and 225 msec (to 0.02% of full-scale accuracy); digitally held for
indefinite display; PEAK and TRACK modes controlled by front-panel button, rear-
panel logic input, or computer-port command
Internal 15-Segment Linearization, programmable from front panel
Analog Output: Range scalable in 0.1% increments between 74.5% and 125.5% of
nominal input.** May be sourced by conditioned input (A) after the FIXED ANA-
LOG FILTER; (B) after PEAK CAPTURE; or (C) after the SELECTABLE ANALOG FIL-
TER (see Fig. 15, Section 2.d)
Frequency Characteristics and Step-Response Settling Times:
AFTER THE FIXED ANALOG FILTER and AFTER +PEAK IN "PEAK" or
"TRACK" MODE: 3 dB down at 20 Hz; 60 dB down at 250 Hz. Settling Time
to 1% of final value: 100 msec; to 0.1% of final value: 150 msec; to 0.02% of
final value: 225 msec
AFTER THE SELECTABLE ANALOG FILTER: See "Analog Filtering for Dis-
played Reading” for the Model 3578, above.
Configuration, Full-Scale Range, Offset Range Adjustment, Allowable
Loading, Offset Accuracy, Span Accuracy, Offset and Span Drift: See the
Model 3530 LVDT Conditioner, above

* Initial (uncalibrated) inaccuracy may be as great as ±3% of full scale. Maximum error that
could occur upon replacement of the 3578 not followed by calibration is ±6% of full scale.
** When the analog output scale is 100.0% and the analog output offset is 0.0%, then any of the
following inputs will result in an analog output of 5 V: an input of 4 mV/V on the 3.0-mV/V range;
an input of 2 mV/V on the 1.5-mV/V range; or an input of 1 mV/V on the 0.75-mV/V range.

A.8 A.2 INDIVIDUAL CONDITIONER SPECIFICATIONS: MODEL 3578


COMMAND AND RESPONSE SYNTAX B

Appendix B
Command and Response Syntax

B.1 INTRODUCTION: RS-232 AND RS-485 MODES


WHEN A 3500 INSTRUMENT'S INTERFACE PORT IS SET TO RS-232 MODE, every
valid INTERROGATION command received at that port will elicit an appropriate
RESPONSE, as will the special transmission-initiating commands CHANNEL (CHN)
and DUMP (DMP), described in Section 5.a. Suppose, for example, that an interro-
gation of HIL [CMT] is received, where [CMT] is the INPUT (or "COMMAND") TERMI-
NATOR which the instrument has been instructed to recognize (see Section 3.a.3).
An answer of h [EOT] will be returned, where "h" is the current HIGH LIMIT setting
and [EOT] is the OUTPUT (or "END-OF-TRANSMISSION") TERMINATOR which the
instrument has been instructed to issue at the end of every transmission from its
RS-232/485 Port (again, see Section 3.a.3).

Note that NO response is given to a SETUP COMMAND such as HIL = h [CMT] or to


an IMPERATIVE COMMAND such as RLS [CMT] when the interface is in the RS-232
mode.

HOWEVER, WHEN THE INTERFACE PORT IS SET TO RS-485 MODE, EVERY


MNEMONIC COMMAND—OF ANY TYPE, "VALID" OR "INVALID"—will elicit an appro-
priate response, as explained below.

B.2 THE OPEN (OPN) COMMAND


As mentioned in Section 5.a.3, OPEN (OPN) is the only computer-issued command
directed to a multinode network as a whole (each node having been set to RS-485
communications). The "opened" node (only) will return to the computer an "acknowl-
edgement" of
ACK [EOT]

ALL OTHER MNEMONIC COMMANDS ISSUED TO THE NETWORK WILL BE DIRECT-


ED EXCLUSIVELY TO THE CURRENTLY "OPEN" NODE.

B.3 RESPONSE TO “INVALID” COMMANDS


WHEN IN RS-485 MODE, RECEIPT BY A 3500 INSTRUMENT OF AN "INVALID" COM-
MAND OF ANY KIND WILL PRODUCE A RESPONSE OF "NOT ACKNOWLEDGED":
NAK [EOT]
An "invalid" command may be one which does not appear on the list below, or which
contains an unacceptable range or other element. An invalid command will elicit NO
RESPONSE when the interface is set to RS-232 mode.

All other commands fall into four basic categories:

B.1 INTRODUCTION B.2 THE OPEN COMMAND B.3 RESPONSE TO “INVALID” COMMANDS B.1
B COMMAND AND RESPONSE SYNTAX

B.4 SETUP COMMANDS


A SETUP (or "WRITE") COMMAND instructs the 3500 instrument to give a particular
value to a particular setup parameter. It will have the general form
[MNEMONIC] = [value] [CMT]
Examples of setup commands are EMM=3000 [CMT], FIL=5 [CMT], ECO=OFF
[CMT], and SAV=ON [CMT].

Upon receipt of a setup command, the instrument (in RS-485 mode only) will issue a
response of either "ACKNOWLEDGED" or "NOT ACKNOWLEDGED"—i.e., of either
ACK [EOT] or NAK [EOT]
depending on whether or not the command has been recognized as "valid." As men-
tioned above, a SETUP COMMAND received by the instrument in RS-232 mode will
elicit NO RESPONSE.

Keep in mind that a response of "ACK" does not prove that the "acknowledged" com-
mand has been executed, only that it has been received. Its purpose is to free the
network controller for other communications. To prove execution of a SETUP COM-
MAND, an appropriate INTERROGATION would have to be made.

B.5 INTERROGATION COMMANDS


An INTERROGATION (or "READ") COMMAND asks the 3500 instrument for the cur-
rent value of a particular setup parameter. It will have the general form
[MNEMONIC] [CMT]
Examples of interrogation commands are EMM [CMT], FIL [CMT], ECO [CMT], and
SAV [CMT].

Upon receipt of a "valid" interrogation command, the instrument (in either RS-232 or
RS-485 mode) will issue a response of
[value] [EOT]
For example, the response to a command of ECO [CMT] will be either ON [EOT] or
OFF [EOT], while a command of FIL [CMT] will return the current digital filter con-
stant (0-9).

And as always, if it is an "invalid" command received by the interface in RS-485 mode,


the only response will be NAK [EOT].

B.6 IMPERATIVE COMMANDS


An IMPERATIVE COMMAND does not impart or request information, but rather tells
the 3500 instrument to do something. The general form is like that of an interrogation
command (see above). Upon receipt of an imperative command, the instrument (in
RS-485 mode only) will issue a response of either ACK [EOT] or NAK [EOT],
depending on whether or not the command has been recognized as "valid." Note
that the only strictly IMPERATIVE commands presently in the command set are RLS
and LNR.

B.2 B.4 SETUP COMMANDS B.5 INTERROGATION COMMANDS B.6 IMPERATIVE COMMANDS
COMMAND AND RESPONSE SYNTAX B
B.7 COMMANDS THAT INITIATE DATA TRANSMISSIONS
The CHANNEL (CHN) and DUMP (DMP) commands, discussed in Section 5.a.1, are
the only ones that belong to this category—with the possible addition of VERSION
(VER), which causes a transmission of instrument "version" information to be issued
from the RS-23/485 Interface Port.

B.8 TABLE OF 3500 MNEMONIC COMMANDS


NOTE: Unless otherwise indicated, every command in the following list
applies equally to all 3500 Series conditioner models.

For the use of CONFIGURATION SETUP commands, see also Section 3.b. For
the use of CALIBRATION SETUP commands, see also Section 4.d. For the use
of RUN-TIME commands, see also Section 5.

AOG ANALOG OUTPUT GAIN


AOG=v [CMT]
Sets to v the ANALOG OUTPUT GAIN, where v is a percentage of full scale to the near-
est tenth of a percent between 74.4% and 125.6%. For RS-485 (only), returns ACK
[EOT].
AOG [CMT]
Reads current ANALOG OUTPUT GAIN. Returns v [EOT].

AOO ANALOG OUTPUT OFFSET


AOO=v [CMT]
Sets to v the ANALOG OUTPUT OFFSET, where v is a percentage of full scale to the
nearest tenth of a percent between -25.6% and 25.6%. For RS-485 (only), returns ACK
[EOT].
AOO [CMT]
Reads current ANALOG OUTPUT OFFSET. Returns v [EOT].

BAU BAUD RATE


BAU=b,s,d,p [CMT]
Sets BAUD RATE and other protocols for the RS-232/485 Interface Port: b = Baud rate:
1 = 300, 2 = 600, 3 = 1200, 4 = 2400, 5 = 4800, 6 = 9600, 7 = 19.2K, 8 = 38.4K, 9 = 153.6K;
s = No. of Stop Bits: 1 or 2; d = No. of Data Bits: 7 or 8; p = Parity: E = EVEN, O = ODD,
N = NONE, Z = ZERO. For RS-485 (only), returns ACK [EOT].
BAU [CMT]
Reads current protocol values for RS-232/485 Interface Port. Returns b,s,d,p [EOT].

BEE ZERO OFFSET


BEE=b [CMT]
Sets to b the ZERO OFFSET to be applied to the data reading, where -32700 ≤ b ≤
32700. For RS-485 (only), returns ACK [EOT].
IMPORTANT: THIS COMMAND SHOULD ONLY BE APPLIED WHEN THE 3500
INSTRUMENT IS SET FOR "TWO-POINT" OR "CALCULATED" CALIBRATION—THAT IS,
WHEN THE CAL=MXB [CMT] COMMAND IS IN EFFECT. IF ENTERED WHEN
CAL=LIN [CMT], UNPREDICTABLE READINGS COULD OCCUR.
BEE [CMT]
Reads current ZERO OFFSET. Returns b [EOT].

B.7 COMMANDS THAT INITIATE DATA TRANSMISSIONS B.8 3500 MNEMONIC COMMANDS B.3
B COMMAND AND RESPONSE SYNTAX

CAL CALIBRATION
CAL=MXB [CMT]
Sets the active calibration method to "MXB" (i.e., "TWO-POINT" or "CALCULATED" cali-
bration using the "y = mx+b" linear equation). For RS-485 (only), returns ACK [EOT].
CAL=LIN [CMT]
Sets the active calibration method to "15-SEGMENT LINEARIZATION." Not recognized
by the Model 3510 Thermocouple Conditioner. For RS-485 (only), returns ACK
[EOT].
CAL [CMT]
Reads the currently active calibration method. Returns MXB [EOT] or LIN [EOT],
except for the Model 3510 Thermocouple Conditioner, which can only return MXB
[EOT].

CHN CHANNEL
CHN [CMT]
Reads the current measurement data. Outputs LBLn,w,sEUS [EOT], where LBL =
header string (optional—see LABEL (LBL) command); n = node number (optional—
see ECHO (ECO) command); w = data value; s = limit-status indicator number (option-
al—see LIMITS (LIM) command); EUS = tailer string (optional—see ENGINEERING
UNIT STRING (EUS) command).

CMT COMMAND TERMINATOR


CMT=c [CMT]
Specifies the single ASCII CONTROL CHARACTER c to be recognized as the INPUT
(or "COMMAND") TERMINATOR for all commands received via the RS-232/485 Inter-
face Port; c must be entered as a hexadecimal byte in square brackets ([01] through
[1F]). CMT must be the same for all network nodes, and cannot be the ASCII [ESC]
character. For RS-485 (only), returns ACK [EOT].
CMT [CMT]
Reads the currently effective COMMAND TERMINATOR. Returns c [EOT].

DMP DUMP
DMP [CMT]
Same as CHANNEL (CHN) command, above.

ECO ECHO
ECO=ON [CMT]
Causes the 3500 instrument's NODE NUMBER to be included in all transmissions from
the RS-232/485 Interface Port in response to a CHANNEL (CHN) or DUMP (DMP)
command. For RS-485 (only), returns ACK [EOT].
ECO=OFF [CMT]
Cancels Node-Number "echo" for CHN and DMP transmissions. For RS-485 (only),
returns ACK [EOT].
ECO [CMT]
Reads current Node-Number "echo" status. Returns ON [EOT] or OFF [EOT].

EMM SCALING FACTOR


EMM=m [CMT]
Sets to m the SCALING FACTOR to be applied to the data reading, where -32700 ≤ m ≤
32700. Also sets desired precision (decimal-point location). For RS-485 (only),
returns ACK [EOT].
IMPORTANT: THIS COMMAND SHOULD ONLY BE APPLIED WHEN THE 3500
INSTRUMENT IS SET FOR "TWO-POINT" OR "CALCULATED" CALIBRATION— THAT IS,
WHEN THE CAL=MXB [CMT] COMMAND IS IN EFFECT. IF ENTERED WHEN
CAL=LIN [CMT], UNPREDICTABLE READINGS COULD OCCUR.
EMM [CMT]
Reads current SCALING FACTOR. Returns m [EOT].

B.4 B.8 3500 MNEMONIC COMMANDS


COMMAND AND RESPONSE SYNTAX B
EOT END-OF-TRANSMISSION TERMINATOR
EOT=$ [CMT]
Specifies an OUTPUT (or "END-OF-TRANSMISSION") TERMINATOR of $, where $ is a
string of up to four ASCII characters, each entered as a hexadecimal byte in square
brackets ([01] through [1F]—the current transmission will be instantly halted if the
instrument perceives an EOT character of [00]). EOT must be the same for all net-
work nodes. In a multinode network, the last EOT character must be the same as the
COMMAND TERMINATOR (see the CMT command). For RS-485 (only), returns ACK
[EOT].
EOT [CMT]
Reads the currently effective END-OF-TRANSMISSION TERMINATOR. Returns $
[EOT].

EUS ENGINEERING UNIT STRING


EUS=$ [CMT]
Specifies "tailer" string $ for all transmissions from the RS-232/485 Interface Port in
response to a CHANNEL (CHN) or DUMP (DMP) command ($ is a string of up to 8
characters, including spaces). For RS-485 (only), returns ACK [EOT].
EUS=N/A [CMT]
Clears all characters from the existing "tailer" string. For RS-485 (only), returns ACK
[EOT].
EUS [CMT]
Reads current "tailer" string. Returns $ [EOT] or N/A [EOT].

EXC EXCITATION — Model 3570 DC Strain Gage Conditioner only


EXC=v [CMT]
Sets to v volts (DC) the excitation level of the Model 3570 DC Strain Gage Conditioner
(v = 2, 5, or 10). For RS-485 (only), returns ACK [EOT].
EXC [CMT]
Reads current excitation level. Returns v [EOT].

FIL FILTER
FIL=f [CMT]
Sets to f the digital filtering constant (0 ≤ f ≤ 9). For RS-485 (only), returns ACK [EOT].
FIL [CMT]
Reads current digital filtering constant. Returns f [EOT].

FRC FORCE
FRC=v [CMT]
Sets the SCALING FACTOR ("m" coefficient) so that the existing input yields a reading
equal to v (-32700 ≤ v ≤ 32700). Also sets desired precision (decimal-point location).
In "TWO-POINT" calibration, must be preceded by a ZERO (ZRO) command. For RS-
485 (only), returns ACK [EOT].
IMPORTANT: THIS COMMAND SHOULD ONLY BE APPLIED WHEN THE 3500
INSTRUMENT IS SET FOR "TWO-POINT" OR "CALCULATED" CALIBRATION—THAT IS,
WHEN THE CAL=MXB [CMT] COMMAND IS IN EFFECT. IF ENTERED WHEN
CAL=LIN [CMT], UNPREDICTABLE READINGS COULD OCCUR.
[NOTE: There is no valid interrogation form for the FRC command.]

FRQ FREQUENCY CALIBRATION —


Model 3540 Frequency Conditioner only
FRQ=i,u [CMT]
Sets an appropriate SCALING FACTOR for the Model 3540 Frequency Conditioner
based on "i" (the nominal full-scale rating of the frequency source—or the highest fre-
quency expected to be measured—in Hertz) and "u" (the corresponding value of the

B.8 3500 MNEMONIC COMMANDS B.5


B COMMAND AND RESPONSE SYNTAX

measured phenomenon, expressed in the desired engineering units and precision).


For RS-485 (only), returns ACK [EOT].
IMPORTANT: THIS COMMAND SHOULD ONLY BE APPLIED WHEN THE 3540 IS SET
FOR "TWO-POINT" OR "CALCULATED" CALIBRATION—THAT IS, WHEN THE
CAL=MXB [CMT] COMMAND IS IN EFFECT. IF ENTERED WHEN CAL=LIN [CMT],
UNPREDICTABLE READINGS COULD OCCUR.
FRQ [CMT]
Reads current "FRQ" calibration constants. Returns i,u [EOT]. See Section 4.d.4
regarding the answer returned in response to an FRQ [CMT] interrogation.

HHY HIGH HYSTERESIS


HHY=p [CMT]
Sets the HIGH HYSTERESIS window depth to p (where p is a percentage of the exist-
ing SCALING FACTOR ("m") value to the nearest tenth of a percent). For RS-485 (only),
returns ACK [EOT].
HHY [CMT]
Reads current HIGH HYSTERESIS window depth. Returns p [EOT].

HIL HIGH LIMIT


HIL=h [CMT]
Sets the HIGH LIMIT value to h (-32700 ≤ h ≤ 32700; h must be greater than or equal to
the existing LOW LIMIT (LOL) value). For RS-485 (only), returns ACK [EOT].
HIL [CMT]
Reads current HIGH LIMIT value. Returns h [EOT].

HLA HIGH LATCH


HLA=ON [CMT]
Sets the HIGH LIMIT to LATCHING status. For RS-485 (only), returns ACK [EOT].
HLA=OFF [CMT]
Sets the HIGH LIMIT to NONLATCHING status. For RS-485 (only), returns ACK [EOT].
HLA [CMT]
Reads current HIGH-LIMIT LATCH status. Returns ON [EOT] or OFF [EOT].

LBL LABEL
LBL=$ [CMT]
Specifies "header" string $ for all transmissions from the RS-232/485 Interface Port in
response to a CHANNEL (CHN) or DUMP (DMP) command ($ is a string of up to 8
characters, including spaces). For RS-485 (only), returns ACK [EOT].
LBL=N/A [CMT]
Clears all characters from the existing "header" string. For RS-485 (only), returns ACK
[EOT].
LBL [CMT]
Reads current "header" string. Returns $ [EOT] or N/A [EOT].

LFC LINEARIZATION FORCE


LFC=V [CMT]
Sets to v the OUTPUT value for the linearization segment corresponding to the exist-
ing "live" INPUT. Not recognized by the Model 3510 Thermocouple Conditioner.
For RS-485 (only), returns ACK [EOT].
[NOTE: There is no valid interrogation form for the LFC command.]

LHY LOW HYSTERESIS


LHY=p [CMT]
Sets the LOW HYSTERESIS window depth to p (where p is a percentage of the exist-
ing SCALING FACTOR ("m") value to the nearest tenth of a percent). For RS-485 (only),
returns ACK [EOT].

B.6 B.8 3500 MNEMONIC COMMANDS


COMMAND AND RESPONSE SYNTAX B
LHY [CMT]
Reads current LOW HYSTERESIS window depth. Returns p [EOT].

LIM LIMITS
LIM=ON [CMT]
Includes LIMIT-STATUS Indicator Number in all transmissions from the RS-232/485
Interface Port in response to a CHANNEL (CHN) or DUMP (DMP) command. For
RS-485 (only), returns ACK [EOT].
LIM=OFF [CMT]
Cancels Limit-Status indication for CHN and DMP transmissions. For RS-485 (only),
returns ACK [EOT].
LIM [CMT]
Reads current Limit-Status indication status. Returns ON [EOT] or OFF [EOT].

LIN LINEARIZE
LIN=n,i,o [CMT]
Sets the effective endpoint of linearization Segment No. n, where "i" is the INPUT value
in millivolts, and "o" is the desired OUTPUT in desired engineering units. Not recog-
nized by the Model 3510 Thermocouple Conditioner. For RS-485 (only), returns
ACK [EOT].
LINn [CMT]
Reads the current INPUT/OUTPUT data for Segment No. n (0 ≤ n ≤ 15; it need not be
an "active" segment). Not recognized by the Model 3510 Thermocouple Condition-
er. Returns i,o [EOT].

LLA LOW LATCH


LLA=ON [CMT]
Sets the LOW LIMIT to LATCHING status. For RS-485 (only), returns ACK [EOT].
LLA=OFF [CMT]
Sets the LOW LIMIT to NONLATCHING status. For RS-485 (only), returns ACK [EOT].
LLA [CMT]
Reads current LOW-LIMIT LATCH status. Returns ON [EOT] or OFF [EOT].

LNR LINEARIZER RESET


LNR [CMT]
Resets the internal linearization table (see Section 4.c.4(a)); sets the active calibration
mode to "MXB" (see the CAL command). Not recognized by the Model 3510 Ther-
mocouple Conditioner. For RS-485 (only), returns ACK [EOT].

LNS LINEARIZATION SEGMENTS


LNS [CMT]
Reads the number of currently active linearization segments. Returns n [EOT], where
0 ≤ n ≤ 15, except for the Model 3510 Thermocouple Conditioner, which always
returns 0 [EOT].

LOL LOW LIMIT


LOL=L [CMT]
Sets the LOW LIMIT value to l (-32700 ≤ L ≤ 32700; L must be less than or equal to the
existing HIGH LIMIT (HIL) value). For RS-485 (only), returns ACK [EOT].
LOL [CMT]
Reads current LOW LIMIT value. Returns L [EOT].

B.8 3500 MNEMONIC COMMANDS B.7


B COMMAND AND RESPONSE SYNTAX

MVV MV/V CALIBRATION —


Model 3570 DC Strain Gage Conditioner only
MVV=i,u [CMT]
Sets an appropriate SCALING FACTOR for the Model 3570 DC Strain Gage Condition-
er based on "i" (the transducer sensitivity rating in mV/V, full scale) and "u" (the nomi-
nal full-scale rating of the transducer expressed in the desired engineering units and
precision), if excitation = 10 V; when excitation = 5 V, "u" = twice the full-scale rating;
when excitation = 2 V, "u" = five times the full-scale rating). For RS-485 (only), returns
ACK [EOT].
IMPORTANT: THIS COMMAND SHOULD ONLY BE APPLIED WHEN THE 3570 IS SET
FOR "TWO-POINT" OR "CALCULATED" CALIBRATION—THAT IS, WHEN THE
CAL=MXB [CMT] COMMAND IS IN EFFECT. IF ENTERED WHEN CAL=LIN [CMT],
UNPREDICTABLE READINGS COULD OCCUR.
MVV [CMT]
Reads current "MVV" calibration constants. Returns i,u [EOT]. See Section 4.d.4
regarding the answer returned in response to an MVV [CMT] interrogation.

NOD NODE
NOD [CMT]
Reads the NODE NUMBER ("n") of the currently "open" node. Returns n [EOT], where
0 ≤ n ≤ 99. A Node Number can be assigned only through the COM button procedure
given in Section 3.a.3.

OPN OPEN
OPNn [CMT]
"Opens" Network Node No. n (only) for RS-485 communications with the supervisory
computer. There must be no space between "OPN" and the node number n; also, the
node number must not contain a leading zero. For RS-485 (only), returns ACK [EOT].

PEK PEAK
PEK=ON [CMT]
Places the unit in "+PEAK" mode, if Terminal 6 of the rear Logic I/O Connector is at
Logic 0. Disabled when Terminal 6 is at Logic 1. For RS-485 (only), returns ACK
[EOT].
PEK=OFF [CMT]
Places the unit in "TRACK" mode, if Terminal 6 of the rear Logic I/O Connector is at
Logic 0. Disabled when Terminal 6 is at Logic 1. For RS-485 (only), returns ACK
[EOT].
PEK [CMT]
Reads current "+PEAK" status. Returns ON [EOT] or OFF [EOT].

PRI PRINT INTERVAL


PRI=t [CMT]
Sets the automatic print interval to approximately (t x 0.125) second(s), where 0 ≤ t ≤
32700. For no interval, enter a t of "0." For RS-485 (only), returns ACK [EOT].
PRI [CMT]
Reads current automatic print interval setting. Returns t [EOT].

PRN PRINT ENABLE


PRN=ON [CMT]
Enables automatic RS-232 print transmission for the instrument (to be controlled by
front-panel PRINT button or rear logic input when the instrument is in RS-232 mode).
Automatically in effect on powerup. Has no effect on CHN or DMP command. For
RS-485 (only), returns ACK [EOT].
PRN=OFF [CMT]
Disables automatic RS-232 print transmission. Has no effect on CHN or DMP com-
mand. For RS-485 (only), returns ACK [EOT].

B.8 B.8 3500 MNEMONIC COMMANDS


COMMAND AND RESPONSE SYNTAX B
PRN [CMT]
Reads current PRINT ENABLE status. Returns ON [EOT] or OFF [EOT].

RES RESPONSE
RES=f [CMT]
Sets the selectable analog filter corner to f Hz, where f = 5, 10, or 20 for the Model
3530 LVDT Conditioner, Model 3560 Voltage Conditioner, Model 3570 DC Strain
Gage Conditioner, and Model 3578 AC Strain Gage Conditioner. For the Model
3540 Frequency Conditioner, f = 2 (for 2.5), 5, or 10. Not recognized by the Model
3510 Thermocouple Conditioner. For RS-485 (only), returns ACK [EOT].
RES [CMT]
Reads analog filter setting. Returns f [EOT], except for the Model 3510 Thermocou-
ple Conditioner, which always returns 20 [EOT].

RLS RELEASE
RLS [CMT]
Simultaneously releases both the HIGH-LIMIT LATCH and the LOW-LIMIT LATCH,
returning to a "live" state the logic output and front-panel indicator light associated
with each limit (see the HIGH LATCH (HLA) and LOW LATCH (LLA) commands).

RNG RANGE
RNG=r [CMT]
Sets the input range (or unit scale) to r. For allowable "r" values for each conditioner
model, see the table in Section 3.b.4. For RS-485 (only), returns ACK [EOT].
RNG [CMT]
Reads current input-range setting. Returns r [EOT].

SAV SAVE
SAV=ON [CMT]
Saves the existing setup configuration (including calibration data) to nonvolatile
EEPROM. For RS-485 (only), returns ACK [EOT].
[NOTE: The SAV=OFF [CMT] command has no effect.]
SAV [CMT]
Reads "SAVE" status. Returns ON [EOT] when a "save" is currently in progress; other-
wise returns OFF [EOT].

SCD SECURITY CODE


SCD=c [CMT]
Sets the security code to c (where 0 ≤ c ≤ 98). For RS-485 (only), returns ACK [EOT].
SCD [CMT]
Reads current security code number. Returns c [EOT].

SEN SENSITIVITY — Model 3540 Frequency Conditioner only


SEN=s [CMT]
Sets the input sensitivity of the Model 3540 Frequency Conditioner to "s," where s = .1
(for 0.1 to 5 V), 1 (for 1 to 20 V), or 10 (for 10 to 100 V). For RS-485 (only), returns ACK
[EOT].
SEN [CMT]
Reads current input sensitivity setting. Returns s [EOT].

B.8 3500 MNEMONIC COMMANDS B.9


B COMMAND AND RESPONSE SYNTAX

SHN SHUNT CALIBRATE—NEGATIVE —


Model 3570 DC Strain Gage Conditioner and
Model 3578 AC Strain Gage Conditioner only
SHN=ON [CMT]
Closes the DC or AC Strain Gage Conditioner's shunt-calibration switch for a negative
up-scale reading. Opens the positive shunt-calibration switch if it is currently closed.
For RS-485 (only), returns ACK [EOT].
SHN=OFF [CMT]
Opens the negative shunt-calibration switch to resume normal measurement. For RS-
485 (only), returns ACK [EOT].
SHN [CMT]
Reads current status of the negative shunt-calibration switch. Returns ON [EOT] or
OFF [EOT].

SHP SHUNT CALIBRATE—POSITIVE —


Model 3570 DC Strain Gage Conditioner and
Model 3578 AC Strain Gage Conditioner only
SHP=ON [CMT]
Closes the DC or AC Strain Gage Conditioner's shunt-calibration switch for a positive
up-scale reading. Opens the negative shunt-calibration switch if it is currently closed.
For RS-485 (only), returns ACK [EOT].
SHP=OFF [CMT]
Opens the positive shunt-calibration switch to resume normal measurement. For RS-
485 (only), returns ACK [EOT].
SHP [CMT]
Reads current status of the positive shunt-calibration switch. Returns ON [EOT] or
OFF [EOT].

TAR TARE
TAR=ON [CMT]
Places the 3500 instrument in "TARE" mode, if Terminal 5 of the rear Logic I/O Con-
nector is at Logic 0. Disabled when Terminal 5 is at Logic 1. For RS-485 (only), returns
ACK [EOT].
TAR=OFF [CMT]
Returns the 3500 instrument to display of "untared" data, if Terminal 5 of the rear Logic
I/O Connector is at Logic 0. Disabled when Terminal 5 is at Logic 1. For RS-485 (only),
returns ACK [EOT].
TAR [CMT]
Reads current "TARE" status. Returns ON [EOT] or OFF [EOT].

TRR TARE REGISTER


TRR=v [CMT]
Sets the TARE REGISTER value to v (-32700 ≤ v ≤ 32700). For RS-485 (only), returns
ACK [EOT].
TRR [CMT]
Reads current TARE REGISTER value. Returns v [EOT].

TYP TYPE — Model 3510 Thermocouple Conditioner only


TYP=y [CMT]
Sets the THERMOCOUPLE TYPE to "y" (where y = E, J, K, R, S, T, or B) and enables the
Model 3510 Thermocouple Conditioner's internal cold-junction reference. For RS-485
(only), returns ACK [EOT].
TYP=y,N [CMT]
Sets the TC TYPE to "y" (where y = E, J, K, R, S, T, or B) and disables the internal cold-
junction reference. For RS-485 (only), returns ACK [EOT].
TYP [CMT]
Reads current TC Type and cold-junction status. Returns y [EOT] or y,N [EOT].

B.10 B.8 3500 MNEMONIC COMMANDS


COMMAND AND RESPONSE SYNTAX B
VER VERSION
VER [CMT]
Causes "version" information (Model No. and Software Version No.) to be transmitted
from the RS-232/485 Interface Port, followed by [EOT].

ZRO ZERO
ZRO=v [CMT]
Sets the ZERO OFFSET ("b" term) so that the existing input yields a reading equal to v
(-32700 ≤ v ≤ 32700). In "TWO-POINT" calibration, must precede the FORCE (FRC)
command. For RS-485 (only), returns ACK [EOT].
IMPORTANT: THIS COMMAND SHOULD ONLY BE APPLIED WHEN THE 3500
INSTRUMENT IS SET FOR "TWO-POINT" OR "CALCULATED" CALIBRATION— THAT IS,
WHEN THE CAL=MXB [CMT] COMMAND IS IN EFFECT. IF ENTERED WHEN
CAL=LIN [CMT], UNPREDICTABLE READINGS COULD OCCUR.
[NOTE: There is no valid interrogation form for the ZRO command.]

B.8 3500 MNEMONIC COMMANDS B.11


B.12 This page intentionally blank.
TABLE OF ERROR NUMBERS C

Appendix C
Table of Error Numbers
Most of the following ERROR CODES refer to the INTERNAL LINEARIZATION pro-
cedure discussed in Section 4.c.4.

Error Display Meaning

Err1 Insufficient input error—i.e., the desired display resolution


is too great for the input change.
Err2 Table entry error; the entered "input" is within the end-
points of another segment.
Err3 While calculating the "b" term of the segment, an arith-
metic overrange has occurred in justifying the decimal
points.
Err4 The current range is not legal for this instrument. Re-
enter the range.
Err5 The engineering-unit "output" entered for the current seg-
ment is equal to that of the previous segment (resulting in
a slope of "0").
Err6 The "input" of the current segment is equal to that of the
previous segment.
Err7 The entered "input" has an incorrect precision.

TABLE OF ERROR NUMBERS C.1


C.2 This page intentionally blank.
NETWORKING GUIDELINES D

Appendix D
Networking Guidelines
As noted in Section 3.a.4, the maximum baud rate at which network inter-
changes can take place will depend on
• the capacity of the supervisory computer's input buffer;
• the actual network setup configuration; and
• the efficiency of the computer's data-input routine.

Since there is no hardware handshake, it is necessary that the computer be


equipped with an input buffer capable of receiving the entire answer to each and
every network interrogation at the desired baud rate, without interruption.

To prevent buffer overruns under conditions of peak loading—where, for exam-


ple, the computer program requires frequent writing of data to disk, or where
"Terminate and Stay Resident" programs attach to the timer interrupts—care
must be taken in setting up the program for requesting data from the network.
See the sample GW-BASIC inputting routine given below.

To ensure proper network communications, you should always

1. Assign a unique node number to each network node.


2. Make sure that every OPEN (OPN) command observes the proper syntax.
See Section 5.a.3 and Line 3030 of the GW-BASIC program below.
3. Set all network nodes to the same baud rate, number of data bits, and parity
setting.
4. Set all network nodes to 2 stop bits.

To maximize network speed, you should also observe the following rule:

5. For each network node, the COMMAND TERMINATOR ("CMT") and END-OF-
TRANSMISSION TERMINATOR ("EOT") should both be set to CARRIAGE
RETURN (= hexadecimal [0D]).

SAMPLE PROGRAM

100 '-----------------------------------------------------------
110 TERM$=CHR$(13) 'Terminator is carriage return.
120 TIMEOUT%=1! 'Timeout is 1 second.
130 OPEN "COM1:9600,N,8,2,CS,DS" AS #1
132 INPUT "Is this a network (RS 485) or a single (RS 232) device?
(N/S) ";ANS$
134 IF ANS$ = "S" OR ANS$ = "s" GOTO 170 'Bypass opening a node.
136 IF ANS$ <> "N" AND ANS$ <> "n" GO TO 132 'Ask again.
140 INPUT "Enter Node Number ";NODE%
150 GOSUB 3000 'Open node
160 IF ERRCODE% = 2 THEN PRINT "Unable to open node"NODE% : GOTO 250

NETWORKING GUIDELINES D.1


D NETWORKING GUIDELINES

170 PRINT #1,"CHN" 'Send command to read a channel.


190 GOSUB 2000 'Get response.
200 IF ERRCODE% <> 0 THEN PRINT "No response.", : GOTO 250
210 PRINT RESPONSE$
250 PRINT
260 GOTO 134 'Do it again.
270 '-----------------------------------------------------------
1000 'These subroutines assume that the following variables have
1010 'been defined in the application program's main module:
1020 '
1030 ' TIMEOUT - A single-precision floating-point variable
1040 ' equal to the timeout value (in seconds) to be
1050 ' used on reads.
1060 ' TERM$ - A string equal to a carriage return
1070 '
1080 'The GOSUB 2000 and GOSUB 3000 routines return the
1090 'variable ERRCODE% which indicates an error condition
1100 'was detected during communication with the
1110 'network.
1120 'The error codes are as follows:
1130 '
1140 ' ERRCODE% = 0 Successful completion of subroutine call.
1150 ' 1 Instrument responded with 'NAK'.
1160 ' 2 Instrument did not successfully open after three
tries.
1170 ' 3 Timeout error -- Instrument did not respond.
1180 '
1190 '------------------------------------------------------
2000 'Routine to read a response from a command.
2010 RESPONSE$="" : CHARIN$=""
2020 ELAPSED=0 : ERRCODE%=0
2030 OLDTIME=TIMER
2040 WHILE (ELAPSED < TIMEOUT%) AND (CHARIN$ <> TERM$)
2050 WHILE (LOC(1) <> 0) AND (CHARIN$ <> TERM$)
2060 CHARIN$=INPUT$(1,1)
2070 IF CHARIN$ <> TERM$ THEN RESPONSE$=RESPONSE$+CHARIN$
2080 WEND
2090 ELAPSED=TIMER-OLDTIME
2100 IF ELAPSED < 0 THEN ELAPSED=ELAPSED+86400!
2110 WEND
2120 IF CHARIN$ <> TERM$ THEN ERRCODE%=3
2130 ELSE IF RESPONSE$ = "NAK" THEN ERRCODE%=1
2140 RETURN
2150 '
2160 '------------------------------------------------------
3000 'Routine to open a node on a network.
3010 TRIES%=0 : ERRCODE%=-1
3020 WHILE (TRIES% < 3) AND ERRCODE%
3030 PRINT #1,"OPN"+MID$(STR$(NODE%),2) : GOSUB 2000
3040 TRIES%=TRIES%+1
3050 WEND
3060 IF ERRCODE% = 3 THEN ERRCODE%=2
3070 RETURN
3080 '
3090 '------------------------------------------------------

D.2 NETWORKING GUIDELINES


RUN-TIME BUTTON FUNCTIONS E

Appendix E
Run-Time Button Functions
The following front-panel buttons are active during normal instrument operation.
For full details on all standard run-time procedures—including peak capture, tare,
and printout—see Section 5.

SET UP

When this button is pressed while the 3500 instrument is in RUN-TIME MODE,
the operator will be asked to enter the current SECURITY CODE (or to enter a
new code, if it is currently "0"). Once the correct security code has been entered,
the instrument will be placed in SETUP MODE, and the display shown in Fig. 1(b)
will appear. The button will remain lit as long as the instrument is in the SETUP
MODE. While in SETUP MODE, the SETUP button is used to exit the current
setup procedure or to exit SETUP MODE altogether.

PEAK
TRACK
When the "PEAK" logic input terminal is disconnected (i.e., at the "Logic 0" level),
pressing the PEAK/TRACK button once will place the 3500 instrument in the
"+PEAK" MODE. It will then continuously display the most positive value experi-
enced since the button was pressed (see Figs. 22 and 23, Section 5.b).* The but-
ton will remain lit as long as the instrument is in the "+PEAK" MODE. Each
captured maximum value will remain in digital memory until occurrence of a sub-
sequent higher-valued maximum or until the instrument is returned to "TRACK"
MODE. Pressing the button a second time will cause the instrument to return to
"TRACK" MODE.

When the "PEAK" logic input terminal is connected to GROUND (i.e., at the "Logic
1" level), the PEAK/TRACK button is DISABLED.**

NOTE: The PEAK/TRACK button is not active during SETUP MODE.

* For all models except the Model 3510 Thermocouple Conditioner and the Model 3540 Fre-
quency Conditioner, the captured "+PEAK" can be issued as the instrument's ANALOG OUT-
PUT, when the appropriate jumper connections have been made (see Section 2.d). Note also
that when the "PEAK" logic input terminal is at "Logic 0," a command of PEK=ON [CMT]
applied via the RS-232/485 Interface Port has the same effect as pressing the PEAK/TRACK
button the first time, while PEK=OFF [CMT] has the same effect as pressing the button the
second time (see Section 5.b for the use of these commands).

** Changing the logic level of the "PEAK" logic input terminal from "Logic 0" to "Logic 1" has the
same effect as pressing the PEAK/TRACK button the first time (i.e., it places the 3500 instru-
ment in "+PEAK" MODE), while changing the "PEAK" logic input terminal from "Logic 1" to
"Logic 0" has the same effect as pressing the button the second time (i.e., it returns the instru-
ment to "TRACK" MODE). A Logic 1 level at Logic Input 6 also disables the PEK=ON [CMT]
and PEK=OFF [CMT] mnemonic commands as well as the PEAK/TRACK button (see Section
5.b).

RUN-TIME BUTTON FUNCTIONS E.1


E RUN-TIME BUTTON FUNCTIONS

TARE
RESET
When the "TARE" logic input terminal is disconnected (i.e., at the "Logic 0" level),
pressing the TARE/RESET button once will instantly bring the displayed data
reading to the value currently contained in the 3500 instrument's TARE REGIS-
TER. From this point, the displayed data will continue to track the "live" input, but
now with the continuous application of the constant TARE OFFSET determined
by the difference between the reading that existed when the button was first
pressed and the last-entered TARE-REGISTER value (see Fig. 24, Section 5.c).1
The button will remain lit as long as the displayed data represents a "tared" read-
ing. Pressing the button a second time will reset the data display to the normal
"untared" reading.

When the "TARE" logic input terminal is connected to GROUND (i.e., at the "Logic
1" level), the TARE/RESET button is DISABLED.2

PRINT

When the "PRINT" logic input terminal is disconnected (i.e., at the "Logic 0" level)
AND the PRN=ON [CMT] command is in effect, pressing the PRINT button once
will initiate the transmission of formatted hard-copy output from the RS-232/485
Interface Port according to the time interval specified by the user (see Section
3.a.10).3 The button will light only when a print transmission is actually occurring
and the instrument updates the LCD display during that time. Pressing the but-
ton a second time will halt any print transmission in progress at the end of the
line currently being sent.4

1 The operator can load the TARE REGISTER via the TARE/RESET button in SETUP MODE, or
via the TARE REGISTER (TRR) command. Note also that when the "TARE" logic input terminal
is at "Logic 0," a command of TAR=ON [CMT] applied via the RS-232/485 Interface Port has
the same effect as pressing the TARE/RESET button the first time, while TAR=OFF [CMT] has
the same effect as pressing the button the second time (see Section 5.c for the use of these
commands).
2 Changing the logic level of the "TARE" logic input terminal from "Logic 0" to "Logic 1" has the
same effect as pressing the TARE/RESET button the first time (i.e., it tares the "live" data read-
ing), while changing the "TARE" logic input terminal from "Logic 1" to "Logic 0" has the same
effect as pressing the button the second time (i.e., it resets the unit to untared data display). A
Logic 1 level at Logic Input 5 also disables the TAR=ON [CMT] and TAR=OFF [CMT] mnemon-
ic commands as well as the TARE/RESET button (see Section 5.c).
3 Assuming, that is, that the 3500 instrument's NODE NUMBER is currently "0." If the NODE
NUMBER is not "0," the instrument will not produce hard-copy output. Note too that the
PRN=ON [CMT] command is automatically in effect on every instrument powerup.
4 Changing the logic level of the "PRINT" logic input terminal from "Logic 0" to "Logic 1" when
PRN=ON [CMT] has the same effect as pressing the PRINT button the first time (i.e., it initiates
hard-copy printout). However, it also disables the PRINT button, which cannot now be used to
stop the printout. The only way to halt printing is to issue to the 3500 instrument a command of
PRN=OFF [CMT] or to change the "PRINT" logic input terminal from "Logic 1" to "Logic 0"
when PRN=ON [CMT]. See Section 5.d for a full discussion of the PRINT ENABLE (PRN)
command.

E.2 RUN-TIME BUTTON FUNCTIONS


STANDARD LOGIC CONFIGURATION F

Appendix F
Standard Logic Configuration
I/O Terminal Logic Function

HI "HIGH" VIOLATION OUTPUT


The "HI" logic output terminal will go to a Logic 1 (0 V-DC) level
whenever the "live" data reading exceeds the current "HIL" limit
value (see Section 3.a.8 for defining limit zones). If LATCHING is
NOT in effect for HIL, the "HI" terminal will remain at Logic 1 and
the front-panel red "HI" indicator will remain lit until the "live"
data reading becomes less than the HIGH HYSTERESIS thresh-
old specified by the user (again, see Section 3.a.8). If LATCH-
ING is in effect for HIL, the "HI" terminal will remain at Logic 1
and the front-panel red "HI" indicator will remain lit regardless of
the subsequent behavior of the "live" data reading, until the
"UNL" logic input terminal ("UNLATCH TRIGGER") has been set
(momentarily) to Logic 0, or until application of a RELEASE
(RLS) command through the RS-232/485 Interface Port.

LO "LOW" VIOLATION OUTPUT


The "LO" logic output terminal will go to a Logic 1 (0 V-DC) level
whenever the "live" data reading becomes less than the current
"LOL" limit value. If LATCHING is NOT in effect for LOL, the "LO"
terminal will remain at Logic 1 and the front-panel red "LO" indi-
cator will remain lit until the "live" data reading becomes greater
than the LOW HYSTERESIS threshold specified by the user. If
LATCHING is in effect for LOL, the "LO" terminal will remain at
Logic 1 and the front-panel red "LO" indicator will remain lit
regardless of the subsequent behavior of the "live" data read-
ing, until the "UNL" logic input terminal ("UNLATCH TRIGGER")
has been set (momentarily) to Logic 0, or until application of a
RELEASE (RLS) command through the RS-232/485 Interface
Port.

2 "OK" (NO VIOLATION) OUTPUT


The "OK" logic output terminal will be at a Logic 1 (0 V-DC) level
whenever the "live" data reading is greater than or equal to the
current "LOL" limit value and less than or equal to the current
"HIL" limit value.

[No. 3] [IN THE STANDARD LOGIC CONFIGURATION, THIS LOGIC I/O


TERMINAL HAS NO ASSIGNED FUNCTION.]

PRINT When the "PRINT" logic input terminal is at Logic 0 (+5 V-DC)
AND the PRN=ON [CMT] command is in effect, the front-panel
PRINT button will be enabled (see the above section). A transi-
tion at the "PRINT" terminal from a Logic 0 (+5 V-DC) to a Logic 1
(0 V-DC) state when PRN=ON [CMT] will initiate the transmis-

STANDARD LOGIC CONFIGURATION F.1


F STANDARD LOGIC CONFIGURATION

sion of formatted hard-copy output from the RS-232/485 Inter-


face Port according to the time interval specified by the user
(see Section 3.a.10).* A transition at the "PRINT" terminal from a
Logic 1 (0 V-DC) to a Logic 0 (+5 V-DC) state when PRN=ON
[CMT] will halt any print transmission currently in progress. If
the PRN=OFF [CMT] is in effect, the "PRINT" terminal will be
inactive.

TARE A transition at the "TARE" logic input terminal from a Logic 0 (+5
V-DC) to a Logic 1 (0 V-DC) state will tare the "live" data reading
(see above). A transition at the "TARE" terminal from a Logic 1
(0 V-DC) to a Logic 0 (+5 V-DC) state will reset the data display
to the normal "untared" reading. When the "TARE" terminal is at
Logic 1 (0 V-DC), both the front-panel TARE/RESET button and
the mnemonic TAR command are disabled.

PEAK A transition at the "PEAK" logic input terminal from a Logic 0 (+5
V-DC) to a Logic 1 (0 V-DC) state will place the unit in the
"+PEAK" MODE (see above). A transition at the "PEAK" terminal
from a Logic 1 (0 V-DC) to a Logic 0 (+5 V-DC) state will return
the unit to its normal "TRACK" MODE. When the "PEAK" termi-
nal is at Logic 1 (0 V-DC), both the front-panel PEAK/TRACK
button and the mnemonic PEK command are disabled.

UNL A transition at the "UNL" logic input terminal from a Logic 0 (+5
V-DC) to a Logic 1 (0 V-DC) state will unlatch a latched HIGH
and/or LOW limit, returning each limit's logic output and front-
panel indicator to a "live" condition. Note that keeping the
"UNL" terminal at Logic 1 will not prevent the HIGH or LOW limit
from latching on its next violation, if you have set that limit up as
a latching limit (see Section 3.a.8).

For the Model 3570 DC Strain Gage Conditioner (ONLY), the shunt-control
inputs shown in Fig. 5(b) are also negative true, where the "Logic 1" state is
defined as nominal 0 V-DC and "Logic 0" as nominal +5 V-DC. Thus, when the "+"
terminal of the Analog Input Terminal is at Logic 0 (+5 V-DC), the internal shunt
calibration resistor is switched in for a positive up-scale reading (this is equiva-
lent to the application of a SHP=ON [CMT] command from the RS-232/485 Inter-
face Port or to the closing of the positive shunt via the CAL button sequence—
see Section 4.d). When the "–" terminal is at Logic 0, the shunt is switched in for
a negative up-scale reading (equivalent to a SHN=ON [CMT] command or to the
closing of the negative shunt via the CAL button sequence).

Note that if a positive or negative shunt is in place as a result of the correspond-


ing logic input, and a command is then entered via RS-232/485 Interface or front-
panel button that instructs the 3500 instrument to switch in the opposite shunt,
the instrument will simply return to normal "live" measurement. Similarly, if a
shunt has been closed via mnemonic or front-panel command, and a logic input
is issued to switch in the opposite shunt, the instrument will again return to nor-
mal "live" measurement.

* Assuming, that is, that the NODE NUMBER is currently "0." If the NODE NUMBER is not "0," the
3500 instrument will not produce hard-copy output. Logic 1 at the "PRINT" terminal will also
disable the front-panel PRINT button. Note too that the PRN=ON [CMT] command is automat-
ically in effect on every instrument powerup.

F.2 STANDARD LOGIC CONFIGURATION


This page intentionally blank. F.3
Daytronic Corporation
2211 Arbor Blvd. • Dayton, OH 45439-1521• (800) 668-4745
Tel (937) 293-2566 • Fax (937) 293-2586 • www.daytronic.com

You might also like