You are on page 1of 8

Received: 11 April 2019 Accepted: 19 May 2019

DOI: 10.1111/jerd.12494

REVIEW ARTICLE

Mechanism of and factors associated with the loosening of the


implant abutment screw: A review

Yiqing Huang DDS, MDS1,2 | Jiawei Wang DDS, PhD1

1
Department of Prosthodontics, Hubei-MOST
KLOS & KLOBM, School and Hospital of Abstract
Stomatology, Wuhan University, Wuhan, Objective: While the loosening of the abutment screw is one of the most common
China
2 complications in implant-supported restorations, there is a lack of comprehensive lit-
Department of Stomatology, The Central
Hospital of Wuhan, Tongji Medical College, eratures for mechanism of and factors associated with the loosening of the implant
Huazhong University of Science and
abutment screw. The review was to summarize the mechanism of and factors associ-
Technology, Wuhan, China
ated with the loosening of the implant abutment screw.
Correspondence
Overview: A total of 99 relevant articles were included in the literature review. The
Jiawei Wang, Department of Prosthodontics,
School and Hospital of Stomatology, Wuhan mechanism of the abutment screw loosening was explained. The factors contributed
University, 237 Luoyu Road, Wuhan 430079,
to abutment screw loosening were divided into five aspects and then expounded
China.
Email: wb000238@whu.edu.cn respectively.
Conclusions: The internal connection and abutments with anti-rotational and conical
Funding information
National Natural Science Foundation of China, designs have better resistance to screw loosening. Cantilevers increase the risk of
Grant/Award Numbers: 81570956, 81870744
screw loosening. The effect of surface treatment of the abutment screw is unsure.
Clinicians need to tighten the abutment screw to the recommended torque while
avoiding repeated tightening and loosening, and increase the frequency of follow-ups
to retighten the loosened screws in time.
Clinical Significance: While the loosening of the abutment screw is one of the most
common complications in implant-supported restorations, there is a lack of compre-
hensive literatures for mechanism of and factors associated with the loosening of the
implant abutment screw. The review was to summarize the mechanism of and factors
associated with the loosening of the implant abutment screw, so that clinicians may
make better choices in clinical practice.

KEYWORDS
abutment screw, implant, mechanical complication, screw loosening

1 | I N T RO D UC T I O N However, implant restoration is not perfect. In addition to surgical


trauma, high cost, and long period of treatment, the implant-
Since the end of the last century, the rapid development of oral supported restorations associated biological and mechanical complica-
implant technology has opened up a new era of prosthodontics. tions are troublesome in clinical practice.1-7 These complications
Compared with traditional restorations, the introduction of implant mainly include peri-mucositis, peri-implantitis, loosening or fracturing
technology has significantly improved the esthetic and functional out- of the abutment screw, fracturing of the abutment or superstructure,
comes of definitive restoration, thereby greatly improving the quality crown loosening, or porcelain cracking.8-12 Among these, the loosen-
of life of the patients. It has also provided more choices for clinicians ing of the abutment screw is one of the most common mechanical
and patients. complications.11,13,14 The incidence rate of abutment screw loosening

J Esthet Restor Dent. 2019;1–8. wileyonlinelibrary.com/journal/jerd © 2019 Wiley Periodicals, Inc. 1


2 HUANG AND WANG

reaches 5.3% in the first year after loading,15 and 5.8%-12.7% after the ideal preload is usually 60%-80% of the yield strength of the
the 5-year follow-up.14,16-18 The loosening of the abutment screw is material.25,40-42
likely to deteriorate into fracture of the screw or even fracture of the Only 10% of the torque is converted into the preload, whereas
implant without timely management. The instability of the prosthesis, the remaining 90% is used to overcome the friction between the
which is caused by the loosening of the abutment screw, changes the interfaces of the joint's components.40,42,43 The loss of preload was
distribution of the occlusal force during function, thus accelerating observed 2-3 minutes26,33,44,45 or 15 hours29 after tightening, even

the progress of the abutment screw loosening.19,20 Moreover, the without external force. It might have been caused by the irregular sur-

loosening of the abutment screw increases the micro-motion and face morphology of the contact interfaces of the joint at the micro-

the micro-gap of the implant-abutment interface, thereby aggravating scopic level. The high points tend to be flattened during the
installation process, thereby consuming some of the torque. The phe-
the micro-leakage at the implant-abutment interface and causing bio-
nomenon of the preload loss due to installation is called the settling
logical complications in the end.10,21,22
effect.42,46 Some researchers suggested that the screw should be
retightened at an interval of 10 minutes after the initial tighten-
2 | THE MECHANISM OF THE ABUTMENT ing.42,47 As Siamos et al42 demonstrated, retightening could reduce
SCREW LOOSENING torque loss by 17%-19%. However, in most studies, the preload was
believed to keep decreasing with the repeated tightening of the
The abutment screw is an integral part of the implant-abutment con- screw.38,39,48 Cardoso et al38 found that the removal torque value con-
nection in most implant systems. The screw presents elastic deforma- tinues decreasing with the same tightening torque along the cyclic tight-
tion and is elongated when tightened, thereby creating a tensile force ening and loosening of the abutment screw. After 10 cycles of
called preload.23-25 Another definition of the preload is the axial force tightening and loosening, the resistance to screw loosening could not be
generated between the threads of the abutment screw and the inner increased even if the abutment was changed to a new one. However,
part of the implant along the direction of the long axis of the some studies suggested that the preload would increase after repeated
implant.26-29 The screw acts like a tensioned spring, and the clamping tightening and loosening.41,49 This difference might be explained by the
force it induces between the implant and the abutment holds the two differences in methods and materials used.38
parts together. The preload determines the clamping force to a large
extent and is a crucial factor in the stability of the implant-abutment
connection.24,30-33
3 | T H E F A C TOR S C O N T RI B U T ED T O
Bickford34 divided the process of screw loosening into two stages.
A B U T M E N T S C R E W L O O S E N I NG
The initial tensile deformation of the screw is decreased under the
3.1 | Types of implant-abutment connection
impact of external force primarily, thereby leading to a decrease in the
clamping force. In the second stage, with further decrease in clamping The stability of the implant-abutment connection is closely related to
force, the micro-motion of the implant-abutment interface intensifies, its connection type, which has been confirmed by a large number of
and the instability of the connection finally leads to screw loosening. studies.50,51 There are two types of implant-abutment connection in
The loss of the preload by external force causes the loosening of the two-piece implant system, namely, external connection and internal
abutment screw. connection. The protruded structure of a hexagon or an octagon with
The implant-supported restoration is subjected to external force 1-2 mm height on the top of the implant connecting to the abutment
in complex ways when functioning in the mouth. The lateral loads of forms the external connection. The external connection has its advan-
25N in average are applied to the implant-supported restoration dur- tage in orientation of the abutment and anti-rotation. However, due
ing mastication, according to in vivo studies.35 The stress distribution to the limited height and diameter of the contact parts of the joint, it
of the assembly under lateral force is more uneven than that of the is vulnerable to the micro-motion of the implant-abutment interface
axial force; thus, it is harder to maintain the stability of the joint. As under external force, especially lateral force, and leads to instability of
many in vitro studies have reported, the friction between the abut- the joint. According to Binon et al,52 in the external connection, the
ment screw and the inner of the implant decreases after cyclic loading, external force is transmitted directly to the abutment screw and the
and so does the torque.19,36,37 A higher preload means a stronger top of the implant at the joint, thereby causing stress concentration in
resistance of the screw to loosening, and this slows down the process the areas and possibly resulting in the loosening or even fracturing of
38,39
of screw loosening. the abutment screw. The internal connection is the connection
The preload is positively correlated with the value of the screw formed by the abutment, with a structure (4-6 mm) into the implant.
tightening torque. The abutment screw functions in its status of elas- This design increases the contact area between the abutment and the
tic deformation; thus, larger tightening torque and preload does not implant and also changes the stress distribution when loaded. Cibirka
mean better. When the preload exceeds the yield limit of the abut- et al53 have found in the internal connection that the inner part of
ment screw's material, the screw is permanently deformed and loses implant that is in contact with the abutment will withstand most of
its function, thereby loosening or even fracturing screw. Therefore, the external force, thereby greatly reducing the stress at the screw,
HUANG AND WANG 3

effectively decreasing the torque loss, and maintaining the stability of abutment in contact with the inner implant is also associated with tor-
the implant-abutment connection. que loss. Rotational freedom between the implant and the abutment
The internal connection is reported to have advantages over is a critical factor related to the stability of the implant-abutment con-
external connection in the stability of the implant-abutment connec- nection. The implant-abutment connection is most stable when the
tion and has better resistance of torque loss and screw loosening.54-59 degree of rotational freedom is less than 2 ,66 and significant torque
According to a recent systematic review of Vetromilla et al, 60
abut- loss occurs when the degree is greater than 5 .67-70 The anti-rotation
ment screw loosening and abutment fracture occurred more often in design of the abutment can limit the rotational freedom, thereby
external hexagon connection comparing to internal hexagon or Morse maintaining well the stability of the joint. In the experiment of Yao
taper connections in clinical studies. There are also studies stating that et al,71 all the abutments without anti-rotation design were rotated
the difference in connection types does not have a significant influ- after cyclic loading, and fatigue fracture was observed. In addition, the
ence on screw loosening. Vigolo et al61 compared the two connection anti-rotation design can also effectively reduce the incidence of abut-
types by a retrospective study. In a follow-up period of at least 5 years ment screw loosening by increasing the preload.72 The abutment with
involving 2010 implants in 1159 patients, no significant difference a taper design possibly increases the stability of the implant-abutment
was found in the incidence of screw loosening between two connec- connection. The tapered contact interface creates friction-locking
tions, which was similar to the result of Chae et al.62 However, many mechanism and has a wedge effect that could enhance the friction-
other factors may also affect the results in clinical practice. Moreover, locking mechanism.73 in vitro studies proved that the connection with
the difference in the loosening of abutment screw may not be obvi- a tapered abutment has higher resistance to torque loss than other
ous, even if there was a significant difference in torque loss. connections before and after cyclic loading.74,75 The taper design has
also better adaptation between the abutment and the implant than
other abutment designs, thereby reducing the micro-gap and micro-
3.2 | The geometrical morphology, material used,
motion at the implant-abutment interface and decreasing the torque
and manufacturing method of the abutment
loss.51
As the abutment plays an important role in the implant-abutment con- Abutment material could also influence the stability of the implant-
nection, the difference of abutments may change the stability of the abutment connection. Jo et al76 compared the stability of the joint of
joint. The relevant factors of the abutment mainly include its geomet- three abutments made of commercially pure grade 3 titanium (group
rical morphology, material used, and manufacturing method. T3), commercially pure grade 4 titanium (group T4), or Ti-6Al-4V (group
The different geometric designs of the abutment are very likely to TA). It was found that preload and compressive bending strength values
alter the stress distribution at the implant-abutment connection and were significantly higher in group TA in contrast to groups T3 and T4.
have an effect on torque loss. In the study of Moris et al,63 with the The result might be explained by higher strength and lower frictional
only difference being the diameter of the upper part of the abutment, coefficient of Ti-6Al-4V.
which was connected with the superstructure, the torque loss would The use of the porcelain abutment effectively overcame the draw-
be different. Although there was no significant difference in the tor- back of the titanium abutment in esthetic outcome during the restora-
que loss between the two groups before loading, the torque loss of tion of upper anterior teeth. However, compared with the conventional
the 3.8 mm-diameter group was significantly higher than that of the titanium abutment, both zirconia and alumina abutments had greater
4.8-mm group after cyclic loading. brittleness and higher modulus of elasticity, thereby influencing the pre-
It is often necessary to select the collar height of the abutment load and the torque loss. With more similar elastic modulus of the mate-
depending on the thickness of the gingiva to achieve better esthetic rial, the stress distribution of the joint under external force would be
outcome. Siadat et al64 studied the influence of different abutment more even. In the implant-abutment connection, the modulus of elastic-
collar heights on torque loss. There was no significant difference in ity of the porcelain abutment is higher than that of the implant and the
torque loss among the different abutment collar heights (1.5, 3.5, and abutment screw. Thus, the stress would concentrate more on the
5.5 mm) before cyclic loading. However, after cyclic loading, the abut- screw and the implant under external force, thereby increasing the
ment with higher collar height was more vulnerable to torque loss. An risk of screw loosening and implant fracture.77 Zirconia has higher
increase in the collar height of the abutment increased the length of flexural strength and lower modulus of elasticity than alumina,78
the axial cantilever, which may explain this result. making it a better choice for manufacturing porcelain abutments.
Angled abutment is often used when the implant is not placed par- According to Dhingra et al,32 the torque loss of the zirconia abut-
allel to the axial force. Though provided convenience in prosthodontic ment was higher than that of the titanium abutment after cyclic
restoration, angled abutment is more prone to develop transverse loading. However, there was no significant difference. Debris found
force under loads, thus resulting in off-axis forces. El-Sheikh et al65 between the zirconia abutment and the screw, and between the
have proved that increasing abutment angulation as well as collar abutment and the implant, might contribute to retard further torque
height had significant effect on removal torque value after dynamic loss and keep the joint stable. Therefore, zirconia abutment might
cyclic loading, so that screw loosening would increase. still be a choice for clinical applications.
In addition to the geometry of the upper part of the abutment Except for the difference that is possibly found after cyclic load-
connecting with the superstructure, the geometry of the lower part of ing, the micro-leakage and torque loss of the zirconia abutment was
4 HUANG AND WANG

significantly higher than that of the titanium abutment in the leakage the uncoated screws after cyclic loading. The coating of the screw on
tests of Sahin et al,10 which might be explained by the poorer fitness the one hand has increased the fit of the screw to the internal implant,
between the zirconia abutment and the implant. The different proper- and on the other hand, it has reduced the effect of wear. However,
ties of the materials caused the differences between the manufactur- Elias et al90 have found that the joint with Teflon coated abutment
ing processes of the porcelain abutments and the titanium abutments, screw was not as stable as that with uncoated screw after cyclic load-
thereby leading to differences in manufacturing precision and fit dis- ing. Although the coating increased the preload at a certain inserting
crepancy between the two. As Vigolo et al79 have pointed out, the por- torque by reducing the coefficient of friction between the screw and
celain abutment manufactured by computer-aided design/computer- the implant, the removal torque was correspondingly reduced due to
aided manufacturing (CAD/CAM) has lower precision than the titanium smaller coefficient of friction. This result is in accordance with the
abutment and has larger degree of rotational freedom, although without result of Wu.91 Wu et al treated the surface of the abutment screw
significant difference and the rotational freedom of all the specimens with different lubricants. The use of lubricant reduced the removal
was less than 3 . Comparing the prefabricated titanium abutment with torque before and after cyclic loading, and the lubricated screw was
the customized zirconia abutment, the latter is inferior to the former in more prone to loosening than the screw without lubricant.
terms of the precision, degree of rotational freedom, and torque loss; The taper design is now widely used in the manufacturing of the
hence, it is more prone to the instability of the implant-abutment abutment. With the same mechanism, the taper design can also be
connection.80,81 used for the screw head. Although the vertical dimension of taper and
Along with the entrance of CAD/CAM and third-party prosthetic the angle of taper vary at a limited axial distance, the conical-head
components into the clinic, concern about the non-original abutments screw was experimentally confirmed to resist torque loss.34,92
82
is raised. Park et al compared the torque loss of four abutment
groups. The results showed that the torque loss of the original abut-
3.4 | The design of the implant
ments was significantly lower than that of the copy abutments. In
another in vitro study, Alonso-Pérez et al83 came to a similar conclu- It is quite common to use angled abutment to correct the angulation
sion, verifying that the internal fit of the connection was crucial to the of implant-supported restorations for a long time. While nowadays,
mechanical behavior of the assembly, and so that the original compo- new angulation-correcting technique which the angled part is built
nents excelled. over the implant is available. Straight and 12 implants were tested in
the study of Hotinski et al.93 Dynamic cyclic loading at 18 to the long
axis of the implant was applied, and the torque loss of the angulation-
3.3 | The design and material of the abutment screw
correcting implant group was significantly lower than that of the
As the most important part of the implant-abutment connection, the straight group. It was suggested that the angulation-correcting implant
abutment screw allows the preload to tightly connect the implant and had better resistance to screw loosening because of the reduced angle
the abutment and maintains the stability of the implant-abutment of abutment screw loading.
connection under external force. The characteristics of the material
used to make the screw, such as yield strength, modulus of elastic-
3.5 | Types of the restoration
ity, and fatigue life, are important factors that affect the stability of
the implant-abutment connection.84 According to the research of Different types of restorations have different stress distributions
Shinohara et al,85 the screws made of grade 4 pure titanium led to when loading. The implant-supported single crown is more prone to
significantly more loosening than the screws made of titanium alloy screw loosening than the implant-supported fixed bridge, when the
Ti-6Al-4V. The friction coefficient of the interface between the retention modes of the superstructure are the same. The latter has
screw and the internal surface of the implant is one of the main fac- the incidence of screw loosening of 5.6% in the 5-year follow-up,
tors determining the preload value. With the same settling torque, a whereas the former has reached 12.7%.16 It is speculated that the
smaller friction coefficient results in higher preload.86,87 When the components of the implant-supported fixed bridge may bear more
screw is being installed or subjected to external force, there is a rel- evenly under external force and can act as a whole to work better in
ative sliding of the surfaces of the internal implant and the screw in anti-rotation. However, in the single crown, the stress concentrates
dry contact, and the metal with relatively low strength is worn on on the implant-abutment connection part, especially on the abutment
the surface.88 screw, thereby leading to a higher incidence of screw loosening.94
48
In the study of Lee et al, the surface of the screw was contami- Cantilevers are also common in restorations, including both the
nated with titanium nanoparticles to simulate the rough surface of the cantilever in the fixed partial denture and eccentric position of the
screw with larger friction coefficient. The contaminated screws implant in the single crown. The cantilever would work as a lever that
exhibited a more significant torque loss than the untreated ones after amplifies the effect of external force and produces a torque.
repeated insertion and removal. The increased friction coefficient of Shinohara et al85 compared the loosening torque of two different
the screw surface made it more susceptible to settling effect, thereby loading place after cyclic loading: loading at the eccentric point (9 mm
resulting in greater loss of preload. As Anchieta et al89 have similarly from screw center) and at the centric point (3 mm from screw center).
shown that the coated screws had significantly less torque loss than The results showed that eccentric point loading led to more loosening
HUANG AND WANG 5

than centric point loading with significant difference. In an in vitro 3. The original components are more stable than the non-original
experiment of Goiato et al,95 photoelastic analysis was used to evalu- ones in general.
ate the stress distribution of implant-supported restorations. A greater 4. The effect of the surface treatment of the abutment screw on
concentration of force on the cervical part of the implant was improvement of the stability of implant-abutment connection is
observed on the models with two crowns in the cantilever than on controversial.
the models with one crown in the cantilever, showing that the cantile- 5. Cantilevers lead to stress concentration and increase the risk of
ver length did have a direct influence on stress distribution. Suedam screw loosening.
96
et al drew a similar conclusion through their in vitro analysis, affir- 6. The abutment screw needs to be tightened to the recommended

ming the greater length of cantilever could lead to greater concentra- torque by the manufacturer during insertion. Repeated tightening

tion of stress. The findings of the in vitro studies might explain the and loosening of the screw need to be avoided. Retorquing of the
screw after the initial torquing with a time interval may be useful
results of clinical studies. In the systematic review and meta-analysis
in reducing the preload loss.
of Freitas da Silva et al,97 although cantilevers were not proved to
7. Individual differences of patients need to be paid attention to,
have negative influence on the marginal bone loss or the survival rate
especially in terms of the parafunctions, such as bruxism, exces-
of the implant-supported restorations, restorations with short cantile-
sive occlusal force, and bad chewing habits.
vers or without cantilevers did have minor complications like abut-
ment screw loosening or screw fracture.
Moreover, the clinicians need to increase the frequency of follow-
ups, retighten the loosened screws in time, or replace the screws with
3.6 | Human-associated factors new ones to prevent further loosening or even fracture.

Human factors are inevitable in clinical practice. The tightening torque


and insertion technique clinicians used determine the preload value. ACKNOWLEDG MENTS
A suitable preload can better maintain the stability of the implant-
This study was financially supported by grants from the National Nat-
abutment connection. Excessive preload would lead to overload of
ural Science Foundation of China (81570956 and 81870744). The
the connection part, thereby causing the fracturing of components.
authors would like to thank Key Laboratory for Molecular Diagnosis
The insufficient preload would increase the rotation freedom, thereby
of Hubei Province, The Central Hospital of Wuhan, Tongji Medical
resulting in instability of the assemblies.3,98 The different insertion
College, Huazhong University of Science and Technology.
techniques used might influence the settling effect and change the
preload value. Although the insertion technique of retightening after
the first tightening before loading has been supported by some CONFLIC T OF INT ER E ST

studies,42,47 there is still no consensus on a recommended insertion The authors do not have any financial interest in the companies
technique in clinical practice. whose materials are included in this article.
Moreover, in terms of the individual differences of patients, it has
been found in a retrospective comparative study that the bruxers
OR CID
group had a significantly higher prevalence of mechanical complica-
tions including but not limited to screw loosening and screw fracture Jiawei Wang https://orcid.org/0000-0003-1405-4811
in comparison with the non-bruxers group.99 So parafunctions such as
bruxism, excessive occlusal force, bad chewing habits, and others
RE FE RE NCE S
which might be risk factors of screw loosening, should cause concern.
1. Sahin S, Cehrli MC. The significance of passive framework fit in
implant prosthodentics: current status. Implant Dent. 2001;10:85-92.
4 | CO NC LUSIO N 2. Kim SK, Koak JY, Heo SJ, et al. Screw loosening with interchangeable
abutments in internally connected implants after cyclic loading. Int J
Oral Maxillofac Implants. 2012;27:42-47.
Screw loosening is one of the most common mechanical complications
3. Eckert SE, Meraw SJ, Cal E, Ow RK. Analysis of incidence and associ-
of implant-supported restoration. Knowing the causes and relevant ated factors with fractured implants: a retrospective study. Int J Oral
factors can help clinicians make better choices in clinical practice. Maxillofac Implants. 2000;15:662-667.
According to this review, the following conclusions can be obtained: 4. Nissan J, Gross M, Shifman A, Assif D. Stress levels for well-fitting
implant superstructures as a function of tightening force levels, tight-
ening sequence, and different operators. J Prosthet Dent. 2001;86:
1. The internal connection might have better resistance to torque 20-23.
loss and screw loosening than external connection. 5. Tsuge T, Hagiwara Y. Influence of lateral-oblique cyclic loading on
abutment screw loosening of internal and external hexagon implants.
2. The geometrical morphology of the abutment including anti-
Dent Mater J. 2009;28:373-381.
rotational and conical designs affects the stability of the implant- 6. Chiapasco M, Romeo E, Coggiola A, Brusati R. Long-term outcome of
abutment connection. dental implants placed in revascularized fibula free flaps used for the
6 HUANG AND WANG

reconstruction of maxillo-mandibular defects due to extreme atrophy. 27. Wang RF, Kang B, Lang LA, Razzoog ME. The dynamic natures of
Clin Oral Implants Res. 2011;22:83-91. implant loading. J Prosthet Dent. 2009;101:359-371.
7. Gbara A, Darwich K, Li L, Schmelzle R, Blake F. Long-term results of 28. Burguete RL, Johns RB, King T, Patterson EA. Tightening characteris-
jaw reconstruction with microsurgical fibula grafts and dental tics for screwed joints in osseointegrated dental implants. J Prosthet
implants. J Oral Maxillofac Surg. 2007;65:1005-1009. Dent. 1994;71:592-599.
8. Wittneben JG, Millen C, Brägger U. Clinical performance of screw- 29. Cantwell A, Hobkirk JA. Preload loss in gold prosthesis-retaining
versus cement-retained fixed implant-supported reconstructions: a screws as a function of time. Int J Oral Maxillofac Implants. 2004;19:
systematic review. Int J Oral Maxillofac Implants. 2014;29:84-98. 124-132.
9. Jung RE, Zembic A, Pjetursson BE, Zwahlen M, Thoma DS. Systematic 30. Jaarda MJ, Razzoog ME, Gratton DG. Geometric comparison of five
review of the survival rate and the incidence of biological, technical, interchangeable implant prosthetic retaining screws. J Prosthet Dent.
and esthetic complications of single crowns on implants reported in 1995;74:373-379.
longitudinal studies with a mean follow-up of 5 years. Clin Oral 31. Lang LA, Kang B, Wang RF, Lang BR. Finite element analysis to deter-
Implants Res. 2012;23:2-21. mine implant preload. J Prosthet Dent. 2003;90:539-546.
10. Sahin C, Ayyildiz S. Correlation between microleakage and screw 32. Dhingra A, Weiner S, Luke AC, Ricci JL. Analysis of dimensional
loosening at implant-abutment connection. J Adv Prosthodont. 2014; changes in the screw and the surface topography at the interface of a
6:35-38. titanium screw and a zirconia abutment under cyclic loading: an
11. Goodacre CJ, Bernal G, Rungcharassaeng K, Kan JY. Clinical complica- in vitro study. Int J Oral Maxillofac Implants. 2013;28:661-669.
tions with implants and implant prostheses. J Prosthet Dent. 2003;90: 33. Kano SC, Binon P, Bonfante G, Curtis DA. Effect of casting proce-
121-132. dures on screw loosening in UCLA-type abutments. J Prosthodont.
12. Zembic A, Kim S, Zwahlen M, Kelly JR. Systematic review of the sur- 2006;15:77-81.
vival rate and incidence of biologic, technical, and esthetic complica- 34. Bickford JH. An introduction to the design and behavior of bolted
tions of single implant abutments supporting fixed prostheses. Int J joints. J Test Eval. 1995;19:94-96.
Oral Maxillofac Implants. 2014;29:99-116. 35. Morneburg TR, Pröschel PA. In vivo forces on implants influenced by
13. Assenza B, Scarano A, Leghissa G, et al. Screw- vs cement-implant- occlusal scheme and food consistency. Int J Prosthodont. 2003;16:
retained restorations: an experimental study in the beagle. Part 481-486.
1. Screw and abutment loosening. J Oral Implantol. 2005;31:242-246. 36. Pardal-Peláez B, Montero J. Preload loss of abutment screws after
14. Assunção WG, Delben JA, Tabata LF, Barão VA, Gomes EA, Garcia IR dynamic fatigue in single implant-supported restorations. A system-
Jr. Preload evaluation of different screws in external hexagon joint. atic review. J Clin Exp Dent. 2017;9:e1355-e1361.
Implant Dent. 2012;21:46-50. 37. Katsuta Y, Watanabe F. Abutment screw loosening of endosseous
15. Pjetursson BE, Thoma D, Jung R, et al. A systematic review of the sur- dental implant body/abutment joint by cyclic torsional loading test at
vival and complication rates of implant-supported fixed dental pros- the initial stage. Dent Mater J. 2015;34:896-902.
theses (FDPs) after a mean observation period of at least 5 years. Clin 38. Cardoso M, Torres MF, Lourenço EJ, et al. Torque removal evaluation
Oral Implants Res. 2012;23:22-38. of prosthetic screws after tightening and loosening cycles: an in vitro
16. Pjetursson BE, Brägger U, Lang NP, Zwahlen M. Comparison of sur- study. Clin Oral Implants Res. 2012;23:475-480.
vival and complication rates of tooth-supported fixed dental prosthe- 39. Ferreira MB, Delben JA, Barão VA, et al. Evaluation of torque mainte-
ses (FDPs) and implant-supported FDPs and single crowns (SCs). Clin nance of abutment and cylinder screws with Morse taper implants.
Oral Implants Res. 2007;18:97-113. J Craniofac Surg. 2012;23:e631-e634.
17. Jung RE, Pjetursson BE, Glauser R, et al. A systematic review of the 40. Piermatti J, Yousef H, Luke A, et al. An in vitro analysis of implant
5-year survival and complication rates of implant-supported single screw torque loss with external hex and internal connection implant
crowns. Clin Oral Implants Res. 2008;19:119-130. systems. Implant Dent. 2006;15:427-435.
18. Feitosa PCP, de Lima APB, Silva-Concílio LR, et al. Stability of external 41. Haack JE, Sakaguchi RL, Sun T, Coffey JP. Elongation and preload
and internal implant connections after a fatigue test. Eur J Dent. stress in dental implant abutment screws. Int J Oral Maxillofac
2013;7:267-271. Implants. 1995;10:529-536.
19. Xia D, Lin H, Yuan S, et al. Dynamic fatigue performance of implant- 42. Siamos G, Winkler S, Boberick KG. The relationship between implant
abutment assemblies with different tightening torque values. Biomed preload and screw loosening on implant-supported prostheses. J Oral
Mater Eng. 2014;24:2143-2149. Implantol. 2002;28:67-73.
20. Fu JH, Hsu YT, Wang HL. Identifying occlusal overload and how to 43. Park JK, Choi JU, Jeon YC, et al. Effects of abutment screw coating
deal with it to avoid marginal bone loss around implants. Eur J Oral on implant preload. J Prosthodont. 2010;19:458-464.
Implantol. 2012;5:S91-S103. 44. Delben JA, Gomes EA, Barao VA, Assunção WG. Evaluation of the
21. Broggini N, Mcmanus LM, Hermann JS, et al. Persistent acute inflam- effect of retightening and mechanical cycling on preload maintenance
mation at the implant-abutment interface. J Dent Res. 2003;82: of retention screws. Int J Oral Maxillofac Implants. 2011;26:251-256.
232-237. 45. Assunção WG, Santos PH, Delben JA, et al. Effect of misfit on preload
22. Park IS, Won SY, Bae TS, et al. Fatigue characteristics of five types of maintenance of retention screws of implant-supported prostheses.
implant-abutment joint designs. Met Mater Int. 2008;14:133-138. J Mater Eng Perform. 2009;18:935-938.
23. Winkler S, Ring K, Ring JD, Boberick KG. Implant screw mechanics 46. Khraisat A, Abu-Hammad O, Al-Kayed AM, Dar-Odeh N. Stability of
and the settling effect: overview. J Oral Implantol. 2003;29:242. the implant/abutment joint in a single-tooth external-hexagon
24. Byrne D, Jacobs S, O'Connell B, et al. Preloads generated with implant system: clinical and mechanical review. Clin Implant Dent Relat
repeated tightening in three types of screws used in dental implant Res. 2004;6:222-229.
assemblies. J Prosthodont. 2006;15:164-171. 47. Shin HM, Huh JB, Yun MJ, et al. Influence of the implant-abutment
25. McGlumphy EA, Mendel DA, Holloway JA. Implant screw mechanics. connection design and diameter on the screw joint stability. J Adv
Dent Clin North Am. 1998;42:71-89. Prosthodont. 2014;6:126-132.
26. Assunção WG, Barão VA, Delben JA, et al. Effect of unilateral misfit 48. Lee HW, Alkumru H, Ganss B, et al. The effect of contamination of
on preload of retention screws of implant-supported prostheses sub- implant screws on reverse torque. Int J Oral Maxillofac Implants. 2015;
mitted to mechanical cycling. J Prosthodont Res. 2011;55:12-18. 30:1054-1060.
HUANG AND WANG 7

49. Tzenakis GK, Nagy WW, Fournelle RA, Dhuru VB. The effect of 69. Junqueira MC, Silva TE, Ribeiro RF, et al. Abutment rotational free-
repeated torque and salivary contamination on the preload of slotted dom evaluation of external hexagon single-implant restorations after
gold implant prosthetic screws. J Prosthet Dent. 2002;88:183-191. mechanical cycling. Clin Implant Dent Relat Res. 2013;15:927-933.
50. Kitagawa T, Tanimoto Y, Odaki M, et al. Influence of implant/abutment 70. Binon PP. Evaluation of three slip fit hexagonal implants. Implant
joint designs on abutment screw loosening in a dental implant system. Dent. 1996;5:235-248.
J Biomed Mater Res B Appl Biomater. 2005;75:457-463. 71. Yao KT, Kao HC, Cheng CK, et al. The potential risk of conical
51. Freitas-Junior AC, Almeida EO, Bonfante EA, et al. Reliability and fail- implant-abutment connections: the antirotational ability of Cowell
ure modes of internal conical dental implant connections. Clin Oral implant system. Clin Implant Dent Relat Res. 2015;17:1208-1216.
Implants Res. 2012;24:197-202. 72. Aboyoussef H, Weiner S, Ehrenberg D. Effect of an antirotation resis-
52. Binon PP. Implants and components: entering the new millennium. Int tance form on screw loosening for single implant-supported crowns.
J Oral Maxillofac Implants. 2000;15:76-94. J Prosthet Dent. 2000;83:450-455.
53. Cibirka RM, Nelson SK, Lang BR, Rueggeberg FA. Examination of the 73. Coppedê AR, de Mattos Mda G, Rodrigues RC, Ribeiro RF. Effect of
implant-abutment interface after fatigue testing. J Prosthet Dent. repeated torque/mechanical loading cycles on two different abut-
2001;85:268-275. ment types in implants with internal tapered connections: an in vitro
54. Sakamoto K, Homma S, Takanashi T, et al. Influence of eccentric cyclic study. Clin Oral Implants Res. 2009;20:624-632.
loading on implant components: comparison between external joint 74. Coppedê AR, Bersani E, de Mattos Mda G, et al. Fracture resistance
system and internal joint system. Dent Mater J. 2016;35:929-937. of the implant-abutment connection in implants with internal hex and
55. Zeno HA, Buitrago RL, Sternberger SS, et al. The effect of tissue internal conical connections under oblique compressive loading: an
entrapment on screw loosening at the implant/abutment interface of in vitro study. Int J Prosthodont. 2009;22:283-286.
external- and internal-connection implants: an in vitro study. 75. Norton MR. An in vitro evaluation of the strength of an internal coni-
J Prosthodont. 2016;25:216-223. cal interface compared to a butt joint interface in implant design. Clin
56. Gracis S, Michalakis K, Vigolo P, et al. Internal vs. external connec- Oral Implants Res. 1997;8:290-298.
tions for abutments/reconstructions: a systematic review. Clin Oral 76. Jo JY, Yang DS, Huh JB, et al. Influence of abutment materials on the
Implants Res. 2012;23:202-216. implant-abutment joint stability in internal conical connection type
57. Wang JH, Judge R, Bailey D. A 5-year retrospective assay of implant implant systems. J Adv Prosthodont. 2014;6:491-497.
treatments and complications in private practice: the restorative com- 77. Piconi C, Maccauro G. Zirconia as a ceramic biomaterial. Biomaterials.
plications of single and short-span implant-supported fixed prosthe- 1999;20:1-25.
ses. Int J Prosthodont. 2016;29:435-444. 78. Garine WN, Funkenbusch PD, Ercoli C, et al. Measurement of the
58. Balik A, Karatas MO, Keskin H. Effects of different abutment connection rotational misfit and implant-abutment gap of all-ceramic abutments.
designs on the stress distribution around five different implants: a Int J Oral Maxillofac Implants. 2007;22:928-938.
3-dimensional finite element analysis. J Oral Implantol. 2012;38:491-496. 79. Vigolo P, Fonzi F, Majzoub Z, Cordioli G. An in vitro evaluation of tita-
59. Michalakis KX, Calvani PL, Muftu S, et al. The effect of different nium, zirconia, and alumina procera abutments with hexagonal con-
implant-abutment connections on screw joint stability. J Oral nection. Int J Oral Maxillofac Implants. 2006;21:575-580.
Implantol. 2014;40:146-152. 80. Alikhasi M, Monzavi A, Bassir SH, et al. A comparison of precision of
60. Vetromilla BM, Brondani LP, Pereira-Cenci T, Bergoli CD. Influence of fit, rotational freedom, and torque loss with copy-milled zirconia and
different implant-abutment connection designs on the mechanical and prefabricated titanium abutments. Int J Oral Maxillofac Implants.
biological behavior of single-tooth implants in the maxillary esthetic 2013;28:996-1002.
zone: a systematic review. J Prosthet Dent. 2019;121:398-403. 81. Siadat H, Beyabanaki E, Mousavi N, Alikhasi M. Comparison of fit
61. Vigolo P, Gracis S, Carboncini F, Mutinelli S. AIOP (Italian Academy of accuracy and torque maintenance of zirconia and titanium abutments
Prosthetic Dentistry) clinical research group internal- vs external- for internal tri-channel and external-hex implant connections. J Adv
connection single implants: a retrospective study in an Italian popula- Prosthodont. 2017;9:271-277.
tion treated by certified prosthodontists. Int J Oral Maxillofac Implants. 82. Park JM, Baek CH, Heo SJ, et al. An in vitro evaluation of the loosen-
2016;31:1385-1396. ing of different interchangeable abutments in internal-connection-
62. Chae SW, Kim YS, Lee YM, et al. Complication incidence of two type implants. Int J Oral Maxillofac Implants. 2017;32:350-355.
implant systems up to six years: a comparison between internal and 83. Alonso-Pérez R, Bartolomé JF, Ferreiroa A, et al. Original vs. non-
external connection implants. J Periodontal Implant Sci. 2015;45: original abutments for screw-retained single implant crowns: an
23-29. in vitro evaluation of internal fit, mechanical behaviour and screw
63. Moris IC, Faria AC, Ribeiro RF, Rodrigues RC. Torque loss of different loosening. Clin Oral Implants Res. 2018;29:1230-1238.
abutment sizes before and after cyclic loading. Int J Oral Maxillofac 84. Yousef H, Luke A, Ricci J, Weiner S. Analysis of changes in implant
Implants. 2015;30:1256-1262. screws subject to occlusal loading: a preliminary analysis. Implant
64. Siadat H, Pirmoazen S, Beyabanaki E, Alikhasi M. Does abutment col- Dent. 2005;14:378-382.
lar length affect abutment screw loosening after cyclic loading. J Oral 85. Shinohara R, Ueda K, Watanabe F. Influence of the difference
Implantol. 2015;41:346-351. between implant body and screw materials on abutment screw loos-
65. El-Sheikh MAY, Mostafa TMN, El-Sheikh MM. Effect of different ening. Dent Mater J. 2019;38:150-156.
angulations and collar lengths of conical hybrid implant abutment on 86. Stüker RA, Teixeira ER, Beck JC, da Costa NP. Preload and torque
screw loosening after dynamic cyclic loading. Int J Implant Dent. removal evaluation of three different abutment screws for single
2018;4:39. standing implant restorations. J Appl Oral Sci. 2008;16:55-58.
66. Binon PP. The effect of implant/abutment hexagonal misfit on screw 87. Martin WC, Woody RD, Miller BH, Miller AW. Implant abutment
joint stability. Int J Prosthodont. 1996;9:149-160. screw rotations and preloads for four different screw materials and
67. Kwon JH, Han CH, Kim SJ, Chang JS. The change of rotational free- surfaces. J Prosthet Dent. 2001;86:24-32.
dom following different insertion torques in three implant systems 88. Vitos L, Larsson K, Johansson B, et al. An atomistic approach to the
with implant driver. J Adv Prosthodont. 2009;1:37-40. initiation mechanism of galling. Comput Mater Sci. 2006;37:193-197.
68. Aramouni P, Vigolo P, Majzoub Z, et al. Evaluation of rotational free- 89. Anchieta RB, Machado LS, Bonfante EA, et al. Effect of abutment
dom of In-Ceram ceramic blank abutments before and after infiltra- screw surface treatment on reliability of implant-supported crowns.
tion. Implant Dent. 2010;19:50-56. Int J Oral Maxillofac Implants. 2014;29:585-592.
8 HUANG AND WANG

90. Elias CN, Figueira DC, Rios PR. Influence of the coating material on area of cantilevered implant-supported fixed partial dentures. J Appl
the loosing of dental implant abutment screw joints. Mater Sci Eng C. Oral Sci. 2016;24:114-120.
2006;26:1361-1366. 97. Freitas da Silva EV, Dos Santos DM, Sonego MV, et al. Does the pres-
91. Wu T, Fan H, Ma R, et al. Effect of lubricant on the reliability of dental ence of a cantilever influence the survival and success of partial
implant abutment screw joint: an in vitro laboratory and three- implant-supported dental prostheses? Systematic review and meta-
dimension finite element analysis. Mater Sci Eng C Mater Biol Appl. analysis. Int J Oral Maxillofac Implants. 2018;33:815-823.
2017;75:297-304. 98. Goheen KL, Vermilyea SG, Vossoughi J, Agar JR. Torque generated
92. Coppedê AR, Faria AC, de Mattos Mda G, et al. Mechanical compari- by handheld screwdrivers and mechanical torquing devices for
son of experimental conical-head abutment screws with conventional osseointegrated implants. Int J Oral Maxillofac Implants. 1994;9:
flat-head abutment screws for external-hex and internal tri-channel 149-155.
implant connections: an in vitro evaluation of loosening torque. Int J 99. Chrcanovic BR, Kisch J, Albrektsson T, Wennerberg A. Bruxism and
Oral Maxillofac Implants. 2013;28:e321-e329. dental implant treatment complications: a retrospective comparative
93. Hotinski E, Dudley J. Abutment screw loosening in angulation-correcting study of 98 bruxer patients and a matched group. Clin Oral Implants
implants: an in vitro study. J Prosthet Dent. 2019;121:151-155. Res. 2017;28:e1-e9.
94. Sailer I, Mühlemann S, Zwahlen M, et al. Cemented and screw-
retained implant reconstructions: a systematic review of the survival
and complication rates. Clin Oral Implants Res. 2012;23:163-201.
95. Goiato MC, Shibayama R, Gennari Filho H, et al. Stress distribution in How to cite this article: Huang Y, Wang J. Mechanism of and
implant-supported prostheses using different connection systems factors associated with the loosening of the implant abutment
and cantilever lengths: digital photoelasticity. J Med Eng Technol.
screw: A review. J Esthet Restor Dent. 2019;1–8. https://doi.
2016;40:35-42.
96. Suedam V, Moretti Neto RT, Sousa EA, Rubo JH. Effect of cantilever
org/10.1111/jerd.12494
length and alloy framework on the stress distribution in peri-implant

You might also like