Professional Documents
Culture Documents
Features
1 Designed for anesthesia and critical care 1 Burst suppression detection, BSR
1 Shares the easy-to-use S/5 user-interface
™
1 Artifact detection
1 Referential or bipolar measurement possible 1 EMG from one EEG channel
1 Impedance check automatically at user defined AEP
intervals 1 Auditory evoked potentials (AEP) for up to 2
1 User-defined measurement montages can be saved channels simultaneously with EEG, using the same
1 Preconfigured leadsets for the most frequently used AEP simulator
montages, automatic recognition of the leadset 1 Single or continuous (moving average) measure-
EEG ment
1 4 channels of EEG 1 Displays averaged evoked response
1 Raw EEG waveform displayed with user-defined 1 Latencies and amplitudes can be measured and
scaling saved for printing out at the end of the case
1 Spectral analysis by Fast Fourier Transform (FFT) 1 Possibility to save a reference evoked response and
1 Graphical trends of spectral parameters display it together with the current response
1 Compressed Spectral Array (CSA) display and
printout
S/5 Modules
Technical specifications
Direct function keys AEP
EP Start/Stop Starts or stops AEP measurement Stimulation
Imp. Check Starts checking electrode impedance Click (condensating) Duration 100 µs
Dimensions and weight Frequency 1.1 to 9.1 Hz
Intensity 10 to 90 dB nHL, 10 dB steps
Module size (WxDxH) 37 x 180 x 112 mm/1.5 x 7.1 x 4.4 in
Headbox size (WxDxH) 97 x 174 x 34 mm/ 3.8 x 6.9 x 1.3 in
Measurement
Module weight 0.3 kg/0.7 lbs
Sampling frequency 2400 Hz for MLAEP / 4800 Hz for BAEP
Headbox weight 0.4 kg/0.9 lbs (incl. 3 m / 118 in cable)
Frequency range 0.5 to 1000 Hz
EEG Highpass filter Off / 10 / 30 / 50 / 75 / 100 / 150 Hz
Measurement Single or continuous averaging possibility.
Sampling frequency 100 Hz per channel Single average
Range ± 400 µV Number of averaged responses 100 to 2000
Frequency range 0.5 to 30 Hz Moving average
Resolution 60 nV Gross average 100 to 2000 stimuli
Input impedance 8 MΩ Updateinterval After every 100 stimuli (200, when gross
Noise level < 0.5 µV rms from 0.5 Hz to 30 Hz average is 2000)
CMRR > 100 dB EMG
Analysis
Frequency range 60 to 300 Hz
Parameters from power spectrum SEF, median frequency, relative power in
Amplitude Root Mean Square (RMS)
frequency bands
Burst suppression BSR
Datex®, Ohmeda® and other trademarks are property of Instrumentarium Corp. or its subsidiaries. All other product and company names are property of their respective owners.
Ordering information
Description Order code
S/5 EEG Module M-EEG
S/5 EEG Headbox N-EEG
8005241/0203 © 2003 Datex-Ohmeda Division Instrumentarium Corp. All right reserved. Subject to change without notice.
catalog.