Professional Documents
Culture Documents
Abstract
Antiepileptic drugs (AEDs) are a diverse group of pharmacological agents used in the treatment of epilep-
tic seizures. Over the past several decades some new AEDs, including lamotrigine (LTG), levetiracetam
(LVA), oxcarbazepine (OXC), topiramate (TOP), and zonisamide (ZNS), have become widely used. This
chapter describes a very simple and rapid liquid chromatography-tandem mass spectrometry method for
simultaneous quantitation of LVA, ZNS, LTG, TOP, and MHD in human serum. The method requires a
very small amount of serum (50 μL) for multiple drug measurements and has a total analysis time of 4 min
that makes it well suited for routine clinical analysis of several drugs simultaneously. The imprecision (CVs)
measured at various concentrations across the analytical measurement range (AMR) are less than 7 % for
all analytes. The AMR for each analyte is as follows: LVA (1–100 μg/mL), ZNS (0.8–80 μg/mL), TOP
(0.5–50 μg/mL), and 0.6–60 μg/mL for LTG and MHD.
1 Introduction
Uttam Garg (ed.), Clinical Applications of Mass Spectrometry in Drug Analysis: Methods and Protocols, Methods in Molecular
Biology, vol. 1383, DOI 10.1007/978-1-4939-3252-8_4, © Springer Science+Business Media New York 2016
29
30 Dean C. Carlow et al.
2 Materials
2.1 Samples Serum or plasma (heparin) samples are acceptable for this proce-
dure. Samples are stable 1 week when refrigerated or 4 weeks when
frozen at −20 °C.
Table 1
Calibrator concentrations (μg/mL)
2.4 Calibrators 1. Calibrators: Six calibrators are used when generating a stan-
and Controls dard calibration curve with concentrations depicted in Table 1
(see Notes 1 and 2).
2. High calibrator preparation: Begin by making the high calibra-
tor (calibrator 6) as follows: To three individual 13 × 100 mm
glass tubes add LVA (250 μL), ZNS (400 μL), LTG (200 μL),
MHD (300 μL), and TOP (250 μL) from their respective stan-
dard stock solutions. Evaporate the solvent of the high calibra-
tor to dryness at 37 °C in a TurboVap evaporator using a
stream of nitrogen. Reconstitute the compounds with 2 mL
drug-free serum per tube and vortex mix thoroughly. Combine
the three reconstituted serum tubes to generate 6 mL of cali-
brator 6. To produce calibrators 1–5 (see Table 2 and Note 3).
3. Check the new lot of standards by verifying five unknown
patient samples with the current lot of calibrators. The agree-
ment between the two calculated concentrations must be
within 10 %.
4. Controls: AED II Serum Toxicology Controls (Bi-Level) were
purchased from UTAK Laboratories. Reconstitute controls
with 5 mL of water. The reconstituted control is stable for 25
days at 2–8 °C (see Note 3).
Simultaneous Quantitation of Lamotrigine, Levetiracetam, 10-Hydroxycarbazepine… 33
Table 2
Calibrator preparation
2.5 Analytical 1. HPLC series 200 system (Perkin Elmer) coupled to an Applied
Equipment Biosystems API 4000 triple quadrupole mass spectrometer
and Supplies (AB Sciex).
2. Guard Column: C18, 2 × 4 mm (Phenomenex).
3. HPLC column: Gemini C18, 50 × 2.0 mm i.d., 3 μm particle
(Phenomenex).
4. Eppendorf 1.5 mL microcentrifuge tubes and National
Scientific 2 mL amber glass vials with inserts and pre-slit caps
or equivalent.
5. TurboVap LV evaporator (Biotage).
Table 3
HPLC method
Table 4
Analyte precursor and product ions (m/z)
3 Methods
3.1 Stepwise 1. Run a system suitability to confirm the system performance (see
Procedure Note 2).
2. To 1.5 mL microcentrifuge tubes pipette 25 μL sample (cali-
brators, controls, or patient specimen) (see Note 3).
3. Add 50 μL of the extraction solution (0.1 M zinc sulfate solu-
tion with 0.1 % formic acid).
Simultaneous Quantitation of Lamotrigine, Levetiracetam, 10-Hydroxycarbazepine… 35
3.2 Analysis 1. The data are analyzed using Analyst 1.4.1 software (AB Sciex).
2. Standard curves are based on a linear regression for all analytes.
Weighted linear regression models with weights inversely pro-
portional to the X values were used. The analysis compared I.S.
peak area to sample peak area (y-axis) versus analyte concentra-
tion (x-axis) using the quantifying ions indicated in Table 4.
3. Acceptability of each run is confirmed if the calculated control
concentrations fall within 2 standard deviations of the target
mean values. Target values are established as the mean of 20
runs. If any of the controls are greater than 3 standard devia-
tions the run cannot proceed and troubleshooting procedure
must commence.
4. Typical coefficients of correlation of the standard curve are
>0.99 (see Note 5).
5. Typical chromatograms for a calibrator and a patient sample
are shown in Figs. 1 and 2.
Fig. 1 LC-ESI-MS/MS ion chromatograms of LVA, ZNS, LTG, MHD, and clonaze-
pam-D4 (I.S.) primary product ions from a calibrator sample
36 Dean C. Carlow et al.
Fig. 2 LC-ESI-MS/MS ion chromatograms of LVA, ZNS, LTG, MHD, and clonaze-
pam-D4 (I.S.) primary product ions from a patient sample
4 Notes
References
1. Schachter SC (1999) Antiepileptic drug ther- 3. Patsalos PN et al (2008) Antiepileptic drugs—
apy: general treatment principles and applica- best practice guidelines for therapeutic drug
tion for special patient populations. Epilepsia monitoring: a position paper by the subcom-
40(s9):S20–S25 mission on therapeutic drug monitoring, ILAE
2. Neels HM et al (2004) Therapeutic drug moni- Commission on Therapeutic Strategies.
toring of old and newer anti-epileptic drugs. Epilepsia 49(7):1239–1276
Clin Chem Lab Med 42(11):1228–1255
Simultaneous Quantitation of Lamotrigine, Levetiracetam, 10-Hydroxycarbazepine… 37
4. Johannessen SI et al (2003) Therapeutic drug 6. Perucca E (2000) Is there a role for therapeutic
monitoring of the newer antiepileptic drugs. drug monitoring of new anticonvulsants. Clin
Ther Drug Monit 25(3):347–363 Pharmacokinet 38(3):191–204
5. Johannessen SI, Tomson T (2006) Pharmaco- 7. Krasowski MD (2010) Therapeutic drug moni-
kinetic variability of newer antiepileptic drugs. toring of the newer anti-epilepsy medications.
Clin Pharmacokinet 45(11):1061–1075 Pharmaceuticals 3(6):1909–1935