WIN XRF WINDOWS

PUT THE HARDWARE LOCK

1ST Step

PRODUCTION

UNIQUANT

PREPARE TASK

CREATE (for Example mill scale 07)

OK
Cassette no 1
Sample no mill scale
Sample no 07
Uniquant identification - njob.004 (this is very important do not change this name, but
you can change the no)

CLIK APPEND

PUT THE RESULT FILE NAME

Now the program will display

OK

EXIT

Go to the UNIQUANT

Measure intensities

Select the program (Mill scale 07)

OK
Now the sample will go for analysis so you have to keep the sample in the cassette no 1
as you have created in the step 1 .this analysis will take 20 min .after completion of this
analysis . Now Minimize the WIN XRF Software.

DESKTOP WINDOWS

ON THE DESKTOP TRY TO OPEN THE OPTIQUANT -5 ICON BY DOUBLE CLIK


J -JOB MODE –CLIK

I – IMPORT INTENSITIES –CLIK

I-IMPORT CLIK


List of Program will display and the program near the yellow colours box
represents the current program, Right clik on the yellow box

END OF IMPORTING

OK


BACK

J-JOB CLICK

INSERT NO OF THE SAMPLE (004)



ENTER

C-CAL Click
S-SAMPLE –CLICK

NOW IF THE RESULT IS ELEMENT FORM ,TO CHANGE TO OXIDE


FROM ,GO TO THE “G” GENERAL DATA ,SELECT THE ELEMENT FORM
CHANGE TO OXIDE

NOW YOU WILL GET THE CONCENTRATION OF THE MILL SCALE 004

NOTE:- The Results received from this optiquant method is only approximate.