Professional Documents
Culture Documents
PH 58905-30-35
PH 58905-30-35
Instruction Manual
This manual contains complete instructions for setting up and using
the 450 pH/ion meter. Applications information is also available.
The information contained in this manual was correct at the time of
going to print. However, we continue to improve products and reserve
the right to change specifications, equipment and maintenance
procedures at any time.
This manual is copyrighted, and all rights are reserved. No part of
this manual may be reproduced by any means or in any form without
prior consent in writing.
The power supply unit is classed as IEC Class II equipment
(equipment providing an adequate degree of protection against
electric shocks, in which additional safety precautions, for example,
double or reinforced insulation, are included). The 450 is intended for
use by persons knowledgeable in safe laboratory practices. If the 450
is not used in accordance with these instructions for use, the
protection provided by the equipment may be impaired.
The 450 is suitable for direct current only.
This equipment generates, uses and can radiate radio frequency
energy and if not installed and used in accordance with the
instruction manual, may cause interference with radio
communications. It has been tested and found to comply with the
limits for a Class A computing device pursuant to Subpart J of
Part 15 of FCC Rules, which are designed to provide reasonable
protection against such interference when operated in a commercial
environment. Operation of this equipment in a residential area is
likely to cause interference in which case the user at his own expense
will be required to take whatever measures may be required to
correct the interference.
There are no user replaceable parts in the 450 or power supply unit.
Do not remove the covers.
1 Set Up
1.1 Unpacking and Installation ................................................ 5
1.2 Input and Output Connections ........................................... 6
1.3 Display and Controls ......................................................... 7
1.4 Programming Menu Overview ........................................... 8
2 Operation
2.1 Calibrating a pH Electrode .............................................. 10
2.1.1 pH Measurements ........................................................... 11
2.2 mV Measurements .......................................................... 11
2.2.1 Absolute mV .................................................................... 11
2.2.2 Relative mV ..................................................................... 12
2.3 Calibrating an Ion Selective Electrode ............................ 12
2.3.1 Concentration Measurements ......................................... 13
2.4 Automatic Endpoint (Auto Read) ..................................... 14
2.5 Temperature Measurements ........................................... 15
2.6 Karl Fischer Titrations ..................................................... 16
2.7 Chart Recorder ................................................................ 16
2.8 Interfacing via the RS232 Output .................................... 17
3 Meter Programming
3.1 Calibration Menu ............................................................. 20
3.1.1 Calibration Point Setting .................................................. 20
3.1.2 Last Calibration Recall .................................................... 21
3.1.3 Isopotential Point Setting (pH) ........................................ 21
3.2 Display Menu ................................................................... 22
3.2.1 Channels to Display ......................................................... 22
3.2.2 Units for Concentration to Display ................................... 22
3.2.3 Resolution for pH to Display ............................................ 23
4 Support Information
4.1 Basic Theory .................................................................... 36
4.2 Operating Hints ................................................................ 37
4.3 Maintenance .................................................................... 38
4.3.1 Meter Maintenance .......................................................... 38
4.3.2 Electrode Maintenance .................................................... 38
4.4 Problem Solving .............................................................. 39
4.4.1 Meter Test ....................................................................... 39
4.4.2 Meter Error Codes ........................................................... 40
4.4.3 Electrode Problem Solving .............................................. 41
4.5 Meter Specifications ........................................................ 43
1
Meter with shorting (test) clips in the input sockets
Meter Dust Cover
Power Supply
This Instruction Manual
Memo Cards, English, French, German, Italian,
Spanish
Guide to ISE Measurements
Warranty Card
Calibration Certificate
electrode arm
arm base
end cap
screwdriver
fixing screw tension screw
4. Fill in and return the warranty card. For your record make
a note of the serial number, date of purchase and supplier
on the inside front cover of this manual.
1
socket for Karl Fischer
titrations pH - combination or pH electrode
sockets, Channel A & B
A B
1
mode Selects pH, mV, relative mV or concentration mode (for
Channel A or Channel B).
auto
read Selects automatic endpoint detection.
on/off Turns the display off, and places the meter in standby
mode. Wakes the meter up.
8
The 450 uses a menu system of software for the set up functions. There are eight main menus, these
are: calibration, display, system, temp, limits, datalog, GLP, diagnostics. This chart shows the eight
main menus and their sub routines.
prog
1 2 3 4 5 6 7 8
1 cal points channel time/date ATC type alarm on/off on/off self test
2 last cal conc units language manual set limits timer cal alarm reset
<
The numeric keypad can also be used as quick access keys. For example, to
enter change the display resolution to X.XX, press prog then 2 (to access menu 2),
<
< 3 (to access sub routine 3), 2 (to select option 2).
<
This is useful when operating the 450 remotely from a PC, see Section 2.8.
1
using both measurement channels (A and B), you have
to select which channel you want to set up).
< calibration display system >
<
Press
< and to move round the menus.You can move
either way, and the menu wraps round at
each end. Each time you press an arrow key
the next menu appears, e.g. press
<
< display system temp >
<
your required menu is highlighted,
press enter
Before you can enter or change a value the value must be
flashing, for example: cal 1 = 7.000
Press enter Change the flashing value using the keypad, then
press enter again.
If you have the 'auto read' feature (see Section 2.4) turned on,
the meter will automatically endpoint when the reading is stable.
To manually endpoint and freeze the display press
read
1 point calibration
Place the tip of the electrode in the first buffer and press cal
If you are displaying both channel A and B, select the one you
are calibrating and press enter
2 point calibration
Place the tip of the electrode in the second buffer and press cal
At the measurement endpoint a tick (✔) will show next to the
buffer and the electrode slope will display.
Rinse the tip of the electrode with distilled or deionized water
and goto pH Measurements (Section 2.1.1) or 3 point calibration.
2.1.1 pH Measurements
2
If you are using the 'auto read' function the display freezes when
a stable endpoint is reached (see Section 2.4). To manually freeze
2.2 mV Measurements
The millivolt capability of pH meters can be used to determine the
actual slope of pH electrodes but is most frequently used for doing
potentiometric readings or titrations with Platinum Oxidation-
Reduction electrodes.
2.2.1 Absolute mV
Put the tip of the electrode in the sample and press read The
meter will display absolute millivolts.
When the reading is stable, press read to freeze the display (or
use 'auto read', see Section 2.4).
2.2.2 Relative mV
Press mode until Rel mV shows at the top of the display.
Put the tip of the electrode in the reference solution and press
cal The meter will be displaying absolute millivolts.
When the reading is stable, press read to freeze the display (or
use 'auto read', see Section 2.4).
2 If you have the 'auto read' feature (see Section 2.4) turned on,
the meter will automatically endpoint when the reading is stable.
To calibrate an electrode:
Press cal If you are using both measurement channels (A
and B), select the one you want to calibrate
using and then␣ press enter
<
<
1 point calibration
Place the tip of the electrode in the first standard. At the
measurement endpoint a tick (✔) appears.
2 point calibration
To continue with a 2 point calibration:
Place the tip of the electrode in the second standard and
press cal
2
any order.
Use and to select the standard you want to
<
<
The display will show the concentration of the sample. The units
can be changed (see Section 3.2.2). If you are using a
temperature probe, temperature of the sample will be displayed.
If you are using 'Auto' endpoint the display will freeze
automatically (see Section 2.4). To endpoint manually
press read
The 'auto read' function checks the electrode output for stability
and automatically freezes the display when the reading appears
stable. It can be used during calibration and routine
measurements.
auto
Turn auto endpoint on by pressing read
auto
To cancel auto endpoint press read again.
2
Temperature should always be recorded along with the pH value.
For others to duplicate your readings they must know what the
sample temperature was when you read it. The pH of most
samples will change with temperature and very few will change
by exactly the same amount. To know what the pH of a hot or
cold sample is at room temperature, it must be measured at
room temperature.
1.679 4.000 4.008 4.010 6.860 6.865 7.000 7.413 9.180 9.210 10.010
0°C 1.666 4.000 4.003 4.010 6.980 6.948 7.120 7.534 9.464 9.520 10.320
5°C 1.668 4.000 3.999 4.010 6.950 6.951 7.090 7.500 9.395 9.450 10.250
10°C 1.670 4.000 3.998 4.000 6.920 6.923 7.060 7.472 9.332 9.380 10.180
15°C 1.672 4.000 3.999 4.000 6.900 6.900 7.040 7.448 9.276 9.320 10.120
20°C 1.675 4.000 4.002 4.000 6.880 6.881 7.020 7.429 9.225 9.260 10.060
25 °C 1.679 4.000 4.008 4.010 6.860 6.865 7.000 7.413 9.180 9.210 10.010
30°C 1.683 4.010 4.015 4.010 6.850 6.853 6.990 7.400 9.139 9.160 9.970
35°C 1.688 4.020 4.024 4.020 6.840 6.844 6.980 7.389 9.102 9.110 9.930
40°C 1.694 4.030 4.035 4.030 6.810 6.838 6.980 7.380 9.068 9.060 9.890
45°C 1.700 4.040 4.047 4.040 6.830 6.834 6.970 7.373 9.038 9.030 9.860
50°C 1.707 4.060 4.060 4.060 6.830 6.833 6.970 7.367 9.011 8.990 9.830
+ve
2
The REC socket (2.5 mm phone jack) provides a voltage
proportional to the displayed reading in pH, mV, relative mV and
concentration modes. The polarity of the output is the same as
the polarity of the input signal.
2
3 RS232 Rx (signal receive)
4 not used
5 protective ground (0V)
6 CTS (clear to send)
7 not used
8 output 1 (0V within limit, 5V below low limit)
9 output 2 (0V within limit, 5V above high limit)
9 6
CTS
low limit output
high limit output
------------------------
GLP ON GLP function selected
23/11/93 10:35:52 Date and time
ID NUMBER 09 ID number (user enterable)
MODEL M450 pH meter model number
ELECTRODE ID.......... Electrode ID (e.g. serial number)
CHANNEL A Measurement channel identifier
TEMP COMP MTC Temperature mode (manual)
CAL 1 2 5 . 0° C 7.000pH Cal point, temperature and result
2
GLP OFF GLP function deselected
To move around the software menu easily you can use the
numeric keypad as quick access keys. See Section 1.4 for a menu
overview, and the relevant ‘quick keys’. The 450 can also be
controlled by a PC with terminal software.
To turn on the remote function send: TON
To return to a known position in the software menu send: S (for
standby) H (for home). This returns the 450 to pH mode, channel
A measurement only.
To turn the remote function off send: X
NOTE Allow a 1 second delay between sending characters.
RS232 Keyboard Definitions
Key RS232 Key RS232 Key RS232
Command Command Command
M
<
mode U 5 5
prog P D 6 6
2
<
auto
read
A
<<
< 7 7
on/off S > 8 8
cal C 1 1 9 9
2 2 0 0
read R
3 3 - –
enter E
4 4
Press enter or
< when the cal points sub routine is
highlighted.
If you enter the cal points sub routine from pH mode, the cal
points for pH calibration are set up. If you enter from
concentration mode, the concentration calibration points are set
up, and the standard set as the first calibration point is the
isopotential point.
The display flashes showing that the value can be changed. Enter
the required value using the keypad and press enter
3
buffers (see Section 2.5).
The next cal point is highlighted. If you want to change this next
point, press enter and continue as before.
<
change,
OR
press
< or prog to leave the sub routine without any further
changes.
Press enter or
< when last cal sub routine is highlighted.
press
< or prog to leave the sub routine.
Press enter or
< when iso sub routine is highlighted.
Press enter the display flashes showing that the value can be
changed.
3
Enter the required value using the keypad and press enter
Press enter or
< when the channel sub routine is
highlighted.
<
Press enter or
< when the conc units sub routine is
highlighted.
3
Use and to select the units of measurement you want
<
<
NOTE Calibrate the 450 in the units you want to use for
sample measurement. Changing the units does not
change the reading; for example, the 450 will not convert
from ppm to mmol/L.
Press enter or
< when the pH resolution sub routine is
highlighted.
3
109120-1 Rev. A, 11/96 23
3.3 System Menu
Press enter or
< when the time/date sub routine is
highlighted.
options.
Press enter the display flashes showing that the value can be
changed.
Enter the time or date using the keypad and
< and press enter
3.3.2 Language
3
Default language is English.
Press enter or
< when language sub routine is highlighted.
Press enter or
< when the baud sub routine is highlighted.
Use and to select the baud rate you want to use and
<
<
press enter
Press enter or
< when the beeper sub routine is
highlighted.
3
109120-1 Rev. A, 11/96 25
3.4 Temperature Menu
Press enter or
< when the ATC type sub routine is
highlighted.
press␣ enter
Press enter or
< when the manual sub routine is
highlighted.
3
can be changed.
Enter the required temperature using the keypad and press enter
Press enter or
< when the alarm sub routine is highlighted.
Press enter or
< when the set limits sub routine is
3
highlighted.
Enter the required limit using the keypad and press enter
The maximum limits that can be set are: high 19.999 and low
–1.999.
The high/low limits for pH are available at the DATA socket. See
Section 2.8 for further information.
Press enter or
< when the temp sub routine is highlighted.
<
Enter the required limit using the keypad and press enter
The maximum limits that can be set are: high 130.0 and low
–30.0.
Repeat for other limit.
3
28 109120-1 Rev. A, 11/96
3.6 Datalog Menu
Press enter or
< when the on/off sub routine is highlighted.
press enter
NOTE You can use the on/off function to temporarily stop data
logging, without cancelling the timer and other options.
Reselect an ‘on’ option to restart data logging.
Press enter or
< when the timer sub routine is highlighted.
Press enter the display flashes showing that the interval can be
press 0
changed.
Enter the timer interval using the keypad and
1
< 3 0
< 3
< and
press enter
0 enter
3
109120-1 Rev. A, 11/96 29
3.6 Datalog Menu (cont)
NOTES Use the following checklist to make sure you have set
datalogging up correctly:
check manual endpoint facility is selected, not
automatic (see Section 2.4).
check 'datalogging on' is selected (print only or store
& print - see Section 3.6.1).
check a timer interval is entered (Section 3.6.2).
check the decimal point is flashing (i.e. the 450 is
measuring) when you return to measurement mode.
3 Press enter or
< when the recall data sub routine is
highlighted.
If you recall the data to the external printer, it will print all the
data logged (up to 100 results).
Press
< to leave the sub routine.
Press enter or
< when the clear data sub routine is
highlighted.
3
109120-1 Rev. A, 11/96 31
3.7 GLP (Good Laboratory Practice) Menu
Press enter or
< when the GLP sub routine is highlighted.
samples.
Press enter the display flashes showing that the value can be
changed. The time interval can be set between
1 second and 23 hours 59 minutes and 59 seconds;
the number of samples can be set between 0 and 99.
Press enter or
< when the ID no. sub routine is highlighted.
Press enter the display flashes showing that the number can be
changed.
3
109120-1 Rev. A, 11/96 33
3.8 Diagnostics Menu
Press enter or
< when the self test sub routine is
highlighted.
The 450 displays the keyboard test. Press each control and check
that the highlighting on the display goes off. When all the
controls have been checked the 450 will test the ROM (read only
memory) and then test the display.
Watch the display test, and check for any missing segments.
The 450 will then send the following characters to the DATA
socket (RS232)
0123456789ABCDEFGHIJKLMN
3
check the printout on the external printer.
The 450 finally tests the RAM (random access memory). There
will be a short delay while the RAM is tested.
At the end of the self test routine press read to return directly
to measurement.
Press enter or
< when the reset sub routine is highlighted.
NOTE This will also clear any data stored during data logging.
The 450 will reset all values to default except baud rate, and
reinitialize the display.
Press
< to leave the sub routine without resetting data.
3
109120-1 Rev. A, 11/96 35
4.1 Basic Theory
4
hydrogen ion concentration:
pH = –log[H+]
pH 0 is very acidic, pH 14 is very alkaline and pH 7 is neutral.
This section gives some brief operating hints and good laboratory
practices to help assure trouble free pH measurements.
1.
2.
New electrodes (or electrodes that have been in storage)
should be conditioned in one inch (25 mm) of pH 7.0 or 4.0
buffer for several hours prior to their use. Make sure that the
wetting cap has been removed from the tip of the electrode.
Always remove the wetting cap and fill hole plug during
calibration and measurements. Replace the fill hole plug
when done.
4
3. Calibrate the electrode daily. Though a one point
calibration may be suitable for some applications, we
recommend that a two point calibration (that brackets the
expected pH range) be performed.
4. Be aware that if you are measuring hot or cold samples
without the use of an ATC probe, the values displayed are
not accurate. You should be calibrating and measuring at
room temperature.
5. When transferring from one sample to another, always
rinse the electrode with distilled water and blot dry. Handle
the electrode carefully and do not use it as a stirring rod.
6. Do not use calibration buffers after the expiration date
printed on the package. Keep the bottles tightly capped and
stored according to the manufacturer's instructions. Never
put used buffer back into the bottle of new buffer.
7. For small sample volumes, make sure that both the pH bulb
and the reference junction are in contact with the sample.
8. Response time is a function of both the electrode and the
solution. Some solutions have very fast response while
others, particularly those with low ionic strength, may take
several minutes.
9. Samples must be in solution (water). You cannot measure
the pH of a dry sample.
Refer to the electrode manufacturer's instructions for proper
electrode care and maintenance. See Section 4.3.2 for general
electrode maintenance and Section 4.4.3 for problem solving.
4
The Model 450 Meter needs no maintenance except for an
occasional wipe with a damp cloth. The front panel is made of
polyester and is not affected by most solvents. Polyester is known
to be affected by some organic solvents, including toluene, xylene
and methyl-ethyl-ketone. It is good laboratory practice to wipe
away any spills as soon as they occur.
4
electrode using a cotton swab soaked with alcohol or
acetone. The tip of Glass body electrodes can be put
directly into organic cleaners but do not put plastic or
epoxy body electrodes into organic solvents. Rinse the
electrode and soak in pH 7.0 buffer until stable.
All others - Soak tip of the electrode in 0.1 M
Hydrochloric acid for one hour. Rinse the electrode and
soak in pH 7.0 buffer until stable.
1. Press enter or
< when the self test sub routine is
highlighted, see Section 3.8.1.
The 450 tests the ROM and RAM, and allows the keyboard
controls, RS232 output and display to be checked.
2. Disconnect the electrode from the pH socket, and connect
the shorting clip (or shorting plug). Select mV mode. The
450 displays 0 ±4 mV. Disconnect the shorting clip (or
shorting plug), and the reading changes. Test both
channels.
3. Disconnect the temperature probe – the display shows
MTC. Reconnect the temperature probe (or electrode
incorporating ATC); the display shows ATC or PT1000 and
the ambient temperature.
4. Reconnect electrode.
4
related. Refer to Section 4.4.3 for electrode troubleshooting.
If the meter fails the above test call the number on the back of
this manual for warranty and service information.
4
the calibration to the theoretical slope defined by the Nernst
equation. This value is displayed as a percent of theoretical and
is called electrode efficiency (see Section 4.1 for Basic Theory).
Generally the slope value can be interpreted as follows:
4
a number of things. Those listed below assume that everything
was fine in the calibration buffers.
4
mV ±1999.9 Temperature Compensation
Rel. mV ±1999.9 –5 to 105°C, automatic and
Temp. –30 to 130°C manual
Conc. 0.0001 to 19999
Display
Resolution Dot matrix LCD display
pH 0.001 / 0.01 / 0.1
mV 0.1 Outputs
Rel. mV 0.1 RS232
Recorder (Channel A only)
Temp. 0.1°C
Karl Fischer
Relative Accuracy*
Input Conditions
pH ±0.001
Impedance greater than 1012 ohms
mV ±0.2
Rel. mV ±0.2 Operating Conditions
Temp. Operating temperature: 5 to 40°C
NTC -30 - 0°C ±1% Operating humidity: max 80% at
0 - 50°C ±0.5°C 35°C (non condensing)
50 - 130°C ±1% Installation category 2
PT1000 -30 - 0°C ±0.4°C Pollution category degree 2
0 - 100°C ±0.2°C Size / Weight
100 - 130°C ±0.4°C 10 x 8 x 4 inches (255 x 200 x
Conc. ±0.5% 95 mm) / 2.9 lbs (1.3 kg)
* ±1 least significant digit
Power Requirements
Time The 450 is supplied with an
0 to 23 hours 59 minutes appropriate power supply unit, e.g.
Timer USA/Japan 100V - 120V
50/60Hz, 9VA
User selectable
0 seconds to 11 hours 59 mins Europe 230V 50Hz, 9.7VA
59 secs Output from PSU 9V DC
Memory NOTE The 450 should only be
used with the power
100 memories
supply unit provided.
Isopotential Point 450 Power Rating 4.5VA
7.000 pH, user adjustable
4 Regulatory Compliance
The 450 is manufactured in a FDA (Food & Drug
Administration) and ISO 9001 approved plant for Corning, and
complies with the following regulatory standards: UL1262,
CSA151, IEC1010. The 450 also complies with the European
EMC Directives and therefore carries the CE mark.
Environmental Compliance
The casework components of the 450 are marked with the
appropriate recycling identification symbol. The packaging is
manufactured using recycled cardboard, and printed with water
based ink. The packaging is recyclable. The manual is printed on
environmentally friendly paper.
Warranty:
Corning warrants this product to be free from defects in materials and workmanship. The
warranty period for the meter is two (2) years from the date of purchase and the probe is
six (6) months from the date of purchase.
THIS WARRANTY IS MADE IN LIEU OF ALL OTHER WARRANTIES EXPRESSED OR
IMPLIED INCLUDING THE WARRANTIES OF MERCHANTABILITY AND FITNESS FOR
A PARTICULAR PURPOSE. CORNING SHALL NOT BE LIABLE FOR ANY LOSS OR
DAMAGES ARISING FROM THE USE OF THESE PRODUCTS NOR FOR
CONSEQUENTIAL DAMAGES OF ANY KIND.
In the event that a meter or probe fails under normal laboratory conditions within the
specified period because of a defect in material or workmanship, Corning will, at its
option, repair or replace the product. Contact Corning Customer Service for return
authorization and shipping instructions at 1-800-222-7740.
© 1992. All rights reserved.
Corning Incorporated
Science Products Division
Corning, New York 14831
USA
Fax: 1-607-974-0345