Professional Documents
Culture Documents
Francis Und Carter - 1976 - A Pressure Vessel For Corrosion Testing at High Te
Francis Und Carter - 1976 - A Pressure Vessel For Corrosion Testing at High Te
Related content
- Thin film Mossbauer spectroscopy
A pressure vessel for corrosion testing at high possibilities with a commercially available
spectrometer
temperatures G A Fatseas
- Mössbauer spectroscopy
T E Cranshaw
To cite this article: P E Francis and C W Carter 1976 J. Phys. E: Sci. Instrum. 9 1067
- A furnace to reach 1100 degrees C in
NMR cross coil probes
G A Styles and T B Sheffield
1067
Apparatus and techniques
Band heaters
PTFE blov+out
4
f
Pressurized Hydraulic pressure
t e s t solution ( v i a distilled w a t e r )
(01 I61
Figure 1 Pressure vessel with palladium/gold liner to Figure 3 (a) Arrangement to allow a flow of test solution.
operate up to 250°C (b) Arrangement to isolate the test solution
hot zone is enclosed by two PTFE cylinders. Two types of would attack 316 stainless steel used in the pressurizing circuit.
cylinder assembly have been used. Type 1 (figure 3(a)) is The test solution is trapped between the O-ring seals on the
normally used. The cylinders are a loose fit inside the pressure cylinders. The remainder of the system is filled with pressurized
vessel to allow for their thermal expansion and a flow of test distilled water. When the temperature is raised the lower
solution. The lower cylinder is slotted to hold the specimen. piston moves downwards to accommodate the expansion of
During refreshment the test solution flows in a thin layer the test solution. The lower cylinder incorporates a PTFE blow-
between the wall of the pressure vessel and the lower cylinder out disc which will operate and bypass the O-ring seal to
so that efficient heat transfer can occur. The solution reaches prevent excessive pressure from developing between the
final thermal equilibrium in the reservoir near the end of the cylinders if the piston reaches the end of its travel. This safety
lower piston before entering the working zone. To leave the device would be required in the event of excessive hydrogen
hot zone the solution flows between the upper cylinder and the evolution from the specimen.
wall of the pressure vessel, through the O-ring bypass, which
prevents convective flow, and is cooled by the upper water 5 Operating experience
jacket before finally leaving the vessel. The apparatus has been used to determine the corrosion
Type 2 cylinder assembly (figure 3(b)), which allows experi- behaviour of various alloys (Incoloy 825, Hastelloy C and
ments with static solutions only, is used when the test solution MP35N (35Ni, 35Co,20Cr, 10Mo)) in a range of salt and acid
solutions at 2 5 0 T (Beynon et a1 1974). The experiments were
from 3 to 100 h duration, depending on the aggressivity of the
Presswe gauge solution towards the alloy under test. Using the refreshed
mode of operation experiments have been carried out with the
Flow control following solutions: 0.01 M NaOH, 0.2 M MgS04, 0.5 M
Pressure valve V3
H3P04, 0.2 M Has04 and 0.1 M HCI. Hydrolysis of MgS04
occurred in the hot zone of the apparatus and when a 2 M
solution was used the white hydrolysis product blocked the
flow of solution. Using the static mode of operation a wider
range of solutions was used as follows: 2 M CaC12, 2 M
MgS04, 2 M NaCl, 2 M Na2S04, 2 M MgC12, 1 M NaOH,
1 M H3P04,0.5 M HN03,l M HnS04 and 1 M HCl. Although
the salt solutions caused only slight corrosion of the alloys
(metal losses were typically 10mg dm-* in a 24 h period)
Test solution
many of the solutions were exceedingly corrosive at 250°C; for
Figure 2 Ancillary pressurizing circuit example Hastelloy C lost 1100 mg dm-2 during a 3 h test in
0.5 M H&04 solution.
1068
Apparatus and techniques
The apparatus has been used for corrosion tests at 250°C
but the apparatus has been designed to operate at 350°C. The
PTFE cylinders, however, must be made from a more tempera-
ture-resistant material; alumina, or PTFE loaded with 50%
zirconia, could be used.
6 Conclusions
Conventional autoclaves which are only partly filled with
solution have certain disadvantages. An apparatus which
overcomes these disadvantages has been constructed and used
to measure corrosion rates in solutions at 250°C. The com-
position of the solutions was maintained by refreshing the test
solution and contamination by extraneous ions was prevented
by a noble-metal liner. Tests with extremely corrosive liquids
were successfully carried out. These tests were necessarily of
short duration because of rapid attack on the alloys being
tested and were only possible because of the quick rates of
heating and cooling of the apparatus. The apparatus could be
easily modified for use up to 350°C.
References
Beynon J G, Butler G and Francis P E 1974 NationalPhysical
Laboratory Internal Report DMA 166
McKie A S 1973 Proc. ConJ on High-temperature High-
pressure Electrochemistry in Aqueous Solutions, University of
Surrey
Voukalovitch M P 1958 Thermodynamic Properties of Water
and Steam, 6th edn (Berlin: Veb. Verlag. Technik)
1069