You are on page 1of 1

93-3/4" (2381 mm)

9-5/8" (244 mm) 77" (1956 mm) Full Range of Chair Movement
Headrest Extension Chair in Supine Position
Ceiling Monitor Mount or Ceiling 12-1/2" (318 mm)
Mount Light Positioning Point
S2 16" - 18"
Forward Chair
4" - 6" (see monitor mount table) (406 mm - 457 mm)
Travel (from base
(102 mm - 153 mm) down to base up)

Track Light
Position
Ceiling Monitor Mount ELECTRICAL
Positioning Point DRAIN * 16-5/8"
(see monitor mount table) (422 mm)
C2

UMBILICAL ENTRANCE
A Chair Swivel WATER

CL
26-7/8" Center Point
(683 mm) Support Link 60°
Ceiling Monitor Mount C1 Pivot Point Swivel Rotation AIR
Optional
2" DATA CONDUIT
Positioning Point
*
(Locate as Needed)

8" (204 mm) VACUUM


24-7/8" (see monitor mount table)
(632 mm) 34-3/4" (883 mm)
Length of Baseplate
13-3/4" [349 mm]
(head of chair baseplate to
incoming power location) * Tubing
VACUUM: 5/8" O.D. (16 mm)
For Central Vacuum Only
* DRAIN: 1-1/2" (38 mm) Nominal
Pipe For Gravity Drain (when
(see general utility notes) required)

4" - 6"
(102 mm - 153 mm) Recommended Monitor Mount Placement
Ceiling Monitor Mount or Ceiling
S1 The table and diagram below show recommended locations to place the A-dec Ceiling Monitor
Mount Light Positioning Point 16" - 18"
Mount. Primary recommendations minimize light and monitor arm interference.
(see monitor mount table) (406 mm - 457 mm)
IF THE SYSTEM HAS INSTALL THE MONITOR MOUNT HERE:
THIS TYPE OF LIGHT: Primary Recommendation Secondary Recommendation

A-dec 511 Plumbing Elevations A-dec Track Light


CAUTION:
CL Please see
note below.
5/8" (16 mm) O.D. tubing for wet systems. 7-1/2" 77-1/2"
A-dec Headquarters 40 mm O.D. tubing for dry/semi-dry systems (191 mm) (1969 mm)
2601 Crestview Drive Newberg, Oregon 97132 USA Central vacuum only. (see general utility notes) 58"
(1473 mm)
Tel: 1.800.547.1883 Within USA /Canada
Fax: 1.503.538.0276 Within USA /Canada 27-3/4" - 31-1/2"
(705 mm-800 mm)
1-1/2" (38 mm) nominal
15"-39"
Tel: +1.503.538.7478 Outside USA /Canada pipe for gravity drain
(381 mm-991 mm)
(when required)
www.a-dec.com
7-1/2" (191 mm) from track light to
2" (51 mm) centerline of mounting post.
5" (127 mm) Positioning ensures adequate space

A-dec 511 1" (25 mm)

Floor Surface
1" (25 mm) max.height

Floor Surface
for both mounting plate and
mounting plate cover.

25"
A-dec 577L Light

Dental Chair 1/2" (13 mm) nominal pipe


compression fitting with right 1/2" (13 mm) conduit for
(635 mm)

angle manual shut-off valves for electrical. (supplied by contractor;

Mini Template air & water. check amperage requirements)


NOTE: If the track opposite the transformer is closer than 9-1/2″ (241 mm) from
A-dec 6300 Light

Copyright 2017 A-dec Inc. a wall, the trolley and post assembly must be installed before the track is
PLEASE NOTE: This drawing is for reference purposes ONLY. It is NOT TO SCALE. installed on the ceiling.
86.0905.01 Rev E All rights reserved.