Professional Documents
Culture Documents
1 / 64
Chapter 1 Warnings and Cautions
This User Manual was published by Medsure Medical Technology Co., Ltd. and contains
information about the safe use of this product. Users of this product (including 4000D A,
4000D AP, 4000D B, 4000D BP) shall be trained to use the pump and read this User Manual,
including warnings and cautions before using it. Key aspects of the training included how to
install the syringe; Setting of parameters, such as infusion rate; Conventional troubleshooting
methods. In case fails to follow these warnings, cautions, and instructions may result in serious
injury or death.
Although the User Manual has been considered as thoroughly as possible in preparation
for publication, Medsure Medical Technology Co., Ltd. assumes no responsibility for any
inaccuracies that may occur.
The copyright of this User Manual belongs to Medsure Medical Technology Co., Ltd.
Without the approval of our company, any unit or individual is not allowed to copy, modify or
translate the contents speculated in this User Manual.
1.1 Warnings
5 / 64
33. The factors which could greatly reduce the performance of infusion accuracy could be
misusing of syringe, using incorrect syringe, air in pipeline or too much driving
resistance.
34. The pump use mains plug as the isolation means from supply mains. It should not be
positioned in a way that makes it difficult to disconnect from mains power if necessary.
Disconnect from supply mains by removing the mains plug from the wall outlet.
1.2 Cautions
6 / 64
Contents
Chapter 1 Warnings and Cautions..............................................................2
1.1 Warnings .......................................................................................2
1.2 Cautions.........................................................................................6
Chapter 2 Introduction .............................................................................10
2.1 Working principle and structure .................................................10
2.2 Syringe pump function ................................................................10
2.3 Basic product configuration ........................................................11
2.4 External characteristics of the syringe pump ..............................11
2.4.1 Front of syringe pump......................................................11
2.4.2 Rear of syringe pump .......................................................13
Chapter 3 Installation ...............................................................................14
3.1 Pole clamp installation ................................................................14
3.2 Fixation of the syringe pump.......................................................15
Chapter 4 Basic Operations ......................................................................17
4.1 Position of the operator and the syringe pump ..........................17
4.2 Operation workflow ....................................................................18
4.3 Operation steps ...........................................................................19
4.3.1 Power on ..........................................................................19
4.3.2 Install syringe ...................................................................19
4.3.3 Select syringe brand .........................................................21
4.3.4 Select infusion mode........................................................21
4.3.5 Priming .............................................................................22
4.3.6 Set infusion parameters ...................................................22
4.3.7 Start infusion ....................................................................22
4.3.8 Infusion completed ..........................................................23
4.3.9 Unload or replace syringe ................................................24
4.3.10 Power off .........................................................................24
Chapter 5 Features ...................................................................................25
5.1 Infusion mode .............................................................................25
5.1.1 Rate Mode .......................................................................25
5.1.2 Time Mode .......................................................................25
7 / 64
5.1.3 Weight Mode ...................................................................25
5.2 Bolus infusion ..............................................................................27
5.3 KVO (keep vein open) ..................................................................27
5.4 Sleep ............................................................................................28
5.5 Syringe calibration .......................................................................28
5.6 History Log...................................................................................29
Chapter 6 Functional setting interface .....................................................30
Chapter 7 Alarm System...........................................................................32
7.1 Alarm System Description ...........................................................32
7.2 Failure Analysis and Troubleshooting .........................................36
Chapter 8 Battery .....................................................................................38
8.1 Working with internal battery .....................................................38
8.2 Battery Performance ...................................................................39
8.3 Battery charging ..........................................................................39
8.4 Battery Replacement...................................................................40
8.4.1 Battery replacement frequency .......................................40
8.4.2 Battery Replacement Method .........................................40
8.5 Battery recycling..........................................................................41
Chapter 9 Cleaning and Maintenance ......................................................42
9.1 Cleaning .......................................................................................42
9.2 Maintenance ...............................................................................42
Chapter 10 Transportation and Storage.....................................................45
Chapter 11 Symbols Description ................................................................45
Chapter 12 Specifications ...........................................................................48
12.1 Product ........................................................................................48
12.2 Classification................................................................................49
12.3 Power supply ...............................................................................49
12.4 Infusion rate ................................................................................49
12.5 Maximum infusion rate and KVO rate .........................................50
12.6 Infusion pressure .........................................................................50
12.7 Infusion accuracy.........................................................................50
12.8 Occlusion – Occlusion level .........................................................51
12.9 Occlusion – Occlusion time .........................................................51
8 / 64
12.10 Occlusion release - Bolus volume ............................................52
12.11 Basic performance ...................................................................52
12.12 Supported syringe brands and sizes ........................................53
12.13 Curves ......................................................................................55
Chapter 13 Electromagnetic Compatibility Statement ..............................58
13.1 EMC performance .......................................................................58
13.2 Product use precautions .............................................................59
13.3 Electromagnetic environment requirements ..............................60
9 / 64
Chapter 2 Introduction
The external features of the 4000D series are similar. with only minor differences in the
characters used for the product model identification and part of the button pattern. This section
uses 4000D BP as an example to introduce the main components of the series of syringe
pumps, including the keys, operation interface and back of the syringe pump.
pusher
handle
tubing
retaining clip
syringe jaw
mains power
indicator charging syringe clamp
indicator
11 / 64
Operation Interface Description
The description of the operation interface of this product is shown by above figure. The
two channels of the product can be operated and displayed independently.
Key Description
The following table is the main functional description of each key. The function of each
key can be executed according to the prompt of the functional interface.
Key symbol Description
1) Prime/Bolus key is used to prime out the air in the tubing before it
is connected to the patient.
2) During the infusion, it can also be used to quickly give additional
dose.
Pause key is used to pause the alarm audio for 2 minutes.
12 / 64
unlock.
Specification USB
battery cover label handle Pole Clamp interface
AC power interface
Name Description
Specification The back label contains product information, including product name,
label model number, service life, input power, etc. For more information, see
the back label of the product.
AC power The pump shall be connected to a suitable AC power source using a
interface power cord.
Pole Clamp Use the pole clamp to fix the pump to a vertical infusion pole or a
horizontal railing.
USB interface The pump has two USB ports, each channel has its own USB port. The
USB port is sealed off by a soft plastic cover. It is only used for
debugging by the manufacture.
13 / 64
Speaker The bottom of the syringe pump is equipped with a speaker that can
generate a self-test tone, a key pressed tone and an alarm audio.
Chapter 3 Installation
Each syringe pump has a pole clamp that is mounted on the pump. The pump can be
mounted on a horizontal or vertical railing via the pole clamp.
The safe load of the pole clamp is 13kg, and the total load is less than 3.0±0.5kg. The
safety factor of all supports is designed to be 4 times in terms of wear, corrosion, material
fatigue and aging without compromising the performance of the supports.
Vertical direction:
14 / 64
Pole clamp installation:
1. Unscrew the hexagon socket set screw with a hex wrench.
2. Place the pole clamp against the syringe pump in the desired direction so that the screw
holes in the pole clamp align with the holes on the back of the syringe pump.
3. Install the pole clamp onto the syringe pump with two socket head cap screws and tighten
the screws with an Allen key until the pole clamp is securely fastened.
15 / 64
Clipped to the infusion pole
16 / 64
Chapter 4 Basic Operations
17 / 64
4.2 Operation workflow
parameters
Connect to patient Connect tubing to patient
Disconnect with the patient disconnect the tubing from the patient
18 / 64
4.3 Operation steps
4.3.1 Power on
1. Take a syringe filled with solution, connect a compatible tubing correctly, and make sure
the tubing is securely connected.
2. Open the syringe clamp, press the pusher handle, and pull the pusher to the far end.
19 / 64
3. Install the syringe into the syringe holder and ensure that the syringe casing curls
into the crimp groove of the front casing of the syringe pump and close the syringe
clamp.
4. Press the pusher handle to fully open the syringe jaw and move the pusher to snap
the syringe jaw into the pusher.
5. Pass the tubing from bottom to top through the tubing retaining clip on the side of
the syringe pump. (as the picture shows)
20 / 64
4.3.3 Select syringe brand
1. After the pump is powered on and the syringe is properly installed, the brand selection
interface will appear on the screen of the corresponding channel.
2. Check that the automatically identified syringe size is correct and select the syringe brand
which is actually being used.
3. The title bar of the interface shows the brand and size of the currently selected syringe.
For details of the supported syringe, please refer to the “12.12 Supported Syringe Brands
and Sizes” section of this User Manual.
The syringe pump totally supports three infusion modes. They are rate mode, time mode
and body weight mode (Body weight mode is supported only by 4000D AP and 4000D BP).
The user can select the desired mode by the mode key. See “5.1 Infusion Mode” for the
detailed description of each mode.
21 / 64
4.3.5 Priming
The air remaining in the tubing shall be primed out and the mechanical clearance between
the pusher handler and the push head of syringe shall be eliminated before starting infusion to
ensure that the syringe pump begins the infusion at its best.
1. Ensure the tubing is not connected to the patient, if there is a tubing clip, loose the tubing
clip first.
2. In stop state, press the prime/bolus key and set the desired prime rate of the interface. If
pressing and holding the prime/bolus key, the syringe pump will prime the air in the tubing
at the set rate.
3. The screen displays the priming interface. Press and hold the prime/bolus key until the air
has been primed out and the liquid begins to drip from the end of the tubing.
4. After the air in the tubing is primed out, connect it to the patient and prepare to begin the
infusion.
Note 1: The maximum priming volume of each time is about 2mL. After 2mL volume has been
primed out, the syringe pump stops priming and triggers the alarm “Purge vol. to limit.!” to
notify the user that the priming volume has been to its limitation.
Note 2: The priming volume is not included in the total volume of infusion.
In each infusion mode, the user can set the infusion parameters via the number setting
keys and selection key. See “5.1 Infusion Mode” for the detailed programming instructions.
22 / 64
5. For 4000D AP and 4000D BP models, while infusing the flow rate can be changed
directly (Don’t need to stop infusion firstly).
simultaneously to lock all keys of corresponding channel except the Power key, and the
lock symbol will be displayed on the screen. Or after 2 minutes of infusion, the system
will automatically lock the keys. When locking, the keys of corresponding channel will
not work except the Power key to prevent accidental operation, press and
simultaneously to unlock.
BD Preci. 50mL 1
24 / 64
Chapter 5 Features
Rate mode
Rate ---.-
---.- mL/h
VTBI ---.- mL
Accum. vol. 0.0 mL
Set Sel
Time mode
VTBI ---.-
---.- mL
Infusion time 00 hr 00
Rate ---.- mL/h
Accum. vol. 0.0 mL
Set Sel
Note: The infusion rate is automatically calculated according to the VTBI and the infusion
time. The infusion rate = VTBI / infusion time.
25 / 64
Weight mode Weight mode
Programmable Descriptions
parameters
Rate Please refer to the maximum infusion rate and KVO rate
(<1000mL/h, the minimum step is 0.1mL/h, ≥1000mL/h , the
minimum step is 1mL/h)
Accumulated Volume Non editable, 0.0mL~9999mL, press the select key to set it to zero
when it is selected.
26 / 64
5.2 Bolus infusion
1. While infusing, press the prime/bolus key to enter bolus infusion setting interface and can
edit the bolus rate and the bolus volume. Press the select key to select parameters.
27 / 64
5.4 Sleep
While sleeping, the display is off. Except , and , other keys are disable.
When a channel of the syringe pump is in stop state and except the AC disconnected
alarm there isn’t other alarm, press and key of the corresponding channel at the same
time to manually set the channel to sleep. The same keys can also be used to exit sleeping.
28 / 64
5.6 History Log
If the device has been used to infuse, some critical infusion data will be saved in the
history log. The pump can store up to 5000 history logs. In the case the stored history log to the
limitation, the first one will be overwritten. For each log, the pump only displays infusion time,
rate and infused volume, and the latest log is displayed at the top line. For the details of the log,
the biomed or service engineers can use the compatible PC application to retrieve them from
the pump by communication cable.
Press the menu key to view the history log, and navigate other logs by pressing up or
down key.
29 / 64
Chapter 6 Functional setting interface
The following parameters are the functional parameters of the 4000D BP model syringe
pump. Some parameters are not applicable to other models. See the descriptions in the
parameter list section for details. The default values for the Menu parameters are shown below:
Menu
Menu
Alarm volume 6
Alarm type type1
Date 2017 - 01 - 01
Time 00 : 00 : 00
Syringe cali.
Sel Ok Exit
Menu
Night Mode Off
18 : 00 - 07 : 00
Recommended brands
Alarm reset
Version 4D BP V1.0.0.R
Sel OK Exit
Press the menu key to display the menu setting interface, press the number setting key to
navigate parameters, and press the selection key to select the parameter. If the parameter is
selected, press the number setting key to modify its value, then press the select key to save the
setting or press the mode key to give up the setting.
30 / 64
Parameters Attributes
Occlusion There are low, medium and high levels. It is recommended to choose the
level appropriate occlusion level according to the actual requirements of the clinic.
Change rate There are on and off options. If the setting is on, the infusion rate can be changed
directly while infusing and don’t need to stop infusion firstly. Otherwise the
infusion rate cannot be changed directly. (only for 4000D AP and 4000D BP)
Nearly The time for setting the near complete alarm or syringe near empty alarm. It is
empty time adjustable from 1 to15 minutes. For 4000D A and 4000D AP models, it is a fixed
value, 3 minutes.
KVO Setting the KVO infusion rate. It is used to keep the patient line open.
Dose unit The unit of Dose in body weight mode is mcg/kg/min, mg/kg/h, which can be
selected according to actual requirements. (only for 4000D AP and 4000D BP)
Alarm Setting the audio volume for alarm. It shall be adjusted only by biomed technician.
volume While changing the alarm volume, if the auditory alarm signal sound pressure
levels are less than ambient levels, it will impede the operator to recognize the
alarm condition and the potential risk could occur. The biomed technician shall
avoid this dangerous case.
It is adjustable from 1 to 10. 1 is the lowest alarm volume, 10 is the highest alarm
volume. For 4000D A and 4000D B models, it is fixed to 10. This parameter can be
modified only after inputting the authorization code 123.
Syringe It is used to calibrate the syringes of custom 1~5 to ensure the accuracy of infusion
calibration and meet the requirements of clinical use. (only for 4000D B and 4000D BP)
Night mode There are on and off options. If the setting is on, the night mode is enabled. In night
mode the display backlight is automatically turned off and the alarm volume is
automatically set to 1. Otherwise the night mode is disable. (only for 4000D AP
and 4000D BP)
Night time Setting the night time. The first part is for the start time, and the last part is for the
end time. (only for 4000D AP, 4000D BP)
31 / 64
Parameters Attributes
Recommend Configuring the commonly used syringe brands. Only the selected syringe brand
ed brands may be displayed in the syringe selection interface while the syringe is installed.
Alarm reset Restore the alarm-related values to the default state. The alarm-related values
contain “Occlusion Level”, “Nearly empty time”, “Alarm volume”.
version A version that identifies the current software used by the syringe pump.
Description:
1. For the KVO adjustable range, please refer to the maximum infusion rate and KVO rate.
2. Night time can only be adjusted while the setting of night mode is on.
3. The alarm type has the following features:
32 / 64
Alarm
Complete!!!
Accum. Vol. 1.0mL
Alarm duration 00h 00min
KVO infusion 0.2mL/h
Turn off
In alarm state the corresponding alarm indicator will be light on. If the alarm is a high
priority alarm, the red indicator will flash. If the alarm is a low priority alarm, the yellow
indicator will be light on constantly. The characteristics of the alarm visual indicator signal is
as below.
Alarm Alarm Sound signal Lighting flashing duty ratio
level light period frequency
High Red 15s~17s 1.67Hz 33.30%
The alarm visual indicator signal and audio signal comply with the requirements of IEC
60601-1-8:2012 and IEC 60601-2-24:2012. When multiple alarms occur simultaneously,
each individual alarm information will be automatically alternate and circularly displayed on
the screen. All alarms of this alarm system are technical alarms. The alarm system will be in
the technical alarm state after any alarm is triggered.
If there is nonsystematic alarm and the Start/Stop key is pressed, the alarm will be cleared
(the sound and indicator of alarm will be close) and exit from the alarm interface.
The list of all alarms is as follows:
33 / 64
the pressed state.
System Error Travel sensor system High 100 Restart the
03!!! disconnected milliseconds pump
System Error Syringe size recognition High 100 Restart the
04!!! sensor system milliseconds pump
disconnected
System Error Pressure sensor abnormal High 100 Restart the
05!!! milliseconds pump
System Error Syringe plunger detection High 100 Restart the
06!!! system disconnected milliseconds pump
System Error Transmission monitoring High 100 Restart the
07!!! system disconnected milliseconds pump
System Error Speed sensor system High 100 Restart the
08!!! disconnected milliseconds pump
System Error While infusing High 120 Restart the
09!!! transmission system failed milliseconds pump
System Error While travel sensor failed High 100 Restart the
10!!! milliseconds pump
System Error While syringe size High 100 Restart the
11!!! detection sensor failed milliseconds pump
System Error While the external flash High 2 seconds Restart the
12!!! module failed pump
System Error While there isn’t kick High 3 seconds Restart the
13!!! signal from another pump
channel
System Error While there isn’t kick High 3 seconds Restart the
14!!! signal to another channel pump
System Error While the communication High 3 seconds Restart the
15!!! between these two pump
channels cannot work
normally
Battery empty!!! Battery Depleted High 1 second Connect to the
34 / 64
AC power
Drive error!!! Abnormal motor speed High 9 minutes Press the
"Stop" key to
clear this
alarm
Occlusion!!! While infusing, if the High See 12.9 Press the
tubing or patient line is Occlusion "Stop" key to
blocked, the pressure in sensoring- clear this
the tubing will exceed the Time to alarm
set occlusion threshold. Occlusion
Complete!!! Infusion complete High 200 Press the stop
milliseconds key to clear
the alarm
Syringe In the case the volume to High 200 Press the stop
empty!!! be infused is more than the milliseconds key to clear
actual volume in syringe. the alarm
Syringe loaded The syringe push head is High 100 Press the stop
error!!! not installed or disengaged milliseconds key to clear
while infusing the alarm
Syringe Syringe clamp is removed High 100 Press the stop
unloaded!!! while infusing milliseconds key to clear
the alarm
Transmission The clutch handle is High 100 Press the stop
error!!! opened while infusing milliseconds key to clear
the alarm
Syringe is Syringe is moved while High 100 Press the stop
moved!!! infusing milliseconds key to clear
the alarm
Purge vol. to The priming volume of low 200 Press the stop
limit.! single time is to 2mL milliseconds key to clear
the alarm
Near In the case the remaining low 200 Press the stop
35 / 64
complete! infusion time is no more milliseconds key to clear
than the set near complete the alarm
time
Syringe near While the set volume to be low 200 Press the stop
empty! infused is more than the milliseconds key to clear
actual volume in syringe, the alarm
in the case the remaining
infusion time is no more
than the set nearly
complete time.
Low Battery! Low battery low 1 second Connect to the
AC power
Operation In stop state during 2 low 2 minutes Press any key
reminders! minutes there isn’t any
operation.
AC AC disconnected low 1 second Connect to the
disconnected AC power
Note 1: If there are two or more alarms, the alarm signal is issued according to its sequence of
triggering.
Note 2: The high priority alarm flashing frequency is 1.67Hz and the duty ratio is 33.3%.
Notice: If the alarm indicator is not lighted on, there is no alarm tone or the alarm information
is wrong (including no alarm text message), it means alarm system is abnormal, and the pump
shall be sent to a qualified technician for repairing.
36 / 64
Medsure Medical Technology Co., Ltd. can provide the documents listed in 7.9.3 IEC 60601-
1:2012 to the designated qualified technicians.
37 / 64
The casing or other parts of Please contact the after-sales service department of
the syringe pump are Medsure Medical Technology Co., Ltd.
damaged
1. Verify that the syringe pump infusion indicator is
No flow after starting
flashing normally, otherwise refer to the instructions
infusion
in the user manual for infusion settings and initiating
an infusion.
2. If the infusion indicator indicates that the syringe
pump is running and the tube is connected to the
patient, clamp the tube clip to disconnect the tubing
from the patient.
3. Make sure the tubing is not blocked. If the tubing is
clogged, stop the syringe pump, check the occlusion
position and clear it, and press the push rod release
handle to release the pressure from the tubing, then
release the tubing clip and restart the syringe pump.
4. Please contact with the after-sales service department
of Medsure Medical Technology Co., Ltd.
Chapter 8 Battery
38 / 64
If the power supply is not connected when the syringe pump is turned on, the syringe
pump will automatically trigger AC disconnected alarm after the self-test, and automatically
switch to the internal backup battery. After the syringe pump is reconnected to mains power
supply or mains power supply resumes, it will automatically switch back to mains power
supply.
39 / 64
While pump is being used, if the battery is depleted, the alarm audio and alarm indicator
is activated, please connect mains power supply in time to charge, so as to avoid the battery
being over-discharged and damaged.
If the pump will not be used for a long time, do not connect to mains power supply, but it
shall be charged every three months and stored in a fully charged state to prevent the battery
from being damaged due to long-term non-use.
Before returning to use, the syringe pump shall be connected to mains power supply until
the battery is fully charged.
The service life of the battery depends on the using frequency and times. The capacity of
the battery will decrease as the increase of the number of charge and discharge. If the battery is
properly maintained and stored, the service life of the lithium battery is about not less than 300
full charge and discharge. Improper use of the battery will shorten its life or enter a state of
failure. It is recommended to replace the lithium battery every 3 years.
If the battery is depleted and has not been charged for more than 2 months, the battery will
fail. Do not charge the failed battery and replace the failed battery as soon as possible.
The battery of this device is not removable, and the replacement of the battery must be
performed by a maintenance personnel authorized by the company.
40 / 64
Steps to replace the battery:
1. Use a screwdriver to remove the battery cover fixing screw.
2. Remove the battery cover;
3. Remove the battery cushion;
4. Remove the battery and disconnect the battery outlet from the battery inlet.
5. Connect the replaced battery output to the battery inlet and place the battery in the battery
compartment;
6. Install the battery cushion and battery cover to the battery compartment in turn, then tighten
the screws.
41 / 64
Chapter 9 Cleaning and Maintenance
9.1 Cleaning
The syringe pump can prevent accidental spatter. Please wipe the pump's casing and
surface with a rag, or if necessary, with a rag coated with 75% medical alcohol.
If a spray cleaner is used, please spray it on a rag instead of spraying it on the syringe
pump directly.
If you need to store the syringe pump for a long time, please clean it first, then store it in a
clean and dry environment at room temperature. If possible, use the original packaging when
storing to protect the syringe pump.
Cleaning the syringe pump periodically by following steps:
1. Disconnect the syringe pump from the mains power supply.
2. Wipe the external surface of the syringe pump, pay attention to the syringe clamp and the
edge groove of the syringe.
3. After cleaning, disinfect the syringe pump with appropriate disinfectant and wipe it with a
clean damp cloth to remove any residual disinfectant.
9.2 Maintenance
Hospitals or medical institutions that use this equipment shall establish a comprehensive
maintenance plan, otherwise it may cause equipment failure and unpredictable consequences,
and may endanger personal safety.
In addition to cleaning and disinfection, this equipment requires the following inspections
and maintenance before it can be used in the ward.
42 / 64
Inspection/maintenanc
Maintenance content Frequency
e project
Environment and power supply
meet the requirements.
The power cord has no wear and
good insulation performance.
After continuous used for
No mechanical damage to
Comprehensive one year, repaired or
equipment.
inspection upgraded,check by
The alarm system is functioning
biomed or service engineer.
normally.
Battery performance is normal.
Self-test function and infusion
function are normal.
Every two years, or
suspected that the occlusion
alarm is abnormal, the
infusion flow is not
Preventive Pressure calibration, sensor
accurate, and the syringe is
maintenance calibration
recognized incorrectly,
calibrate only by service
engineer according to the
service manual.
According to the description of
chapter 7 alarm system, check
whether it can work normally or Every two years, either
Alarm system not, during the inspection, after repaired or upgraded,
inspection occlusion and syringe empty check by biomed or service
alarms shall be triggered correctly engineer.
at least. See “Note4” for
examples.
43 / 64
Every two years, or after
Safety inspection the replacement of the
according to IEC board or the syringe pump,
60601-1:2012 it needs to be done, check
by service engineer.
Note:
1. The above maintenance items should be carried out by technicians or professional
maintenance personnel approved by the company. If any damage or abnormality is found,
please do not use it and contact the company immediately.
2. The maintenance of the syringe pump does not require lubrication. In fact, any lubricant
containing organic solvents can damage the pump housing plastic.
3. Take it out every three months. Charge the battery, turn on the syringe pump and perform
a functional test to make sure it can work normally when needed.
4. The examples of checking alarm system are as below:
44 / 64
Chapter 10 Transportation and Storage
1. Do not throw during loading and unloading, avoid strong vibration during transportation;
2. Prevent heavy pressure, without direct sunlight, rain and snow;
3. It shall be stored in an environment where the relative humidity is not higher than 95%,
non-corrosive gas, cool, dry, well ventilated and clean.
Rechargeable battery
Type CF equipment
Date of manufacture
Manufacturer
Serial Number.
45 / 64
Authorized representative in the European Community.
Temperature limitation
Numerical display upper and lower temperature limits (°C)
Humidity limitation
The numerical value shows the upper and lower humidity
limits (%)
Atmospheric pressure limitation
Numerical values show upper and lower atmospheric
pressure (kPa)
46 / 64
Keep away from sunlight
Alarm symbol
47 / 64
Chapter 12 Specifications
12.1 Product
Direction Horizontal
Temperature for
Transportation and -40℃~+55℃
Storage
Relative Humidity for
Transportation and Not higher than 95%
Storage
Atmospheric Pressure for
Transportation and 50kPa~106kPa(500hPa~1060hPa)
Storage
48 / 64
12.2 Classification
Class I, internal power supply, suitable for direct application to the CF type of the heart,
continuous operation, cannot be used in the flammable anesthetic gas mixed with air or
flammable anesthetic environment combined with oxygen or nitrous oxide, This device is a
portable device and is not recommended for use as a portable device.
Normal Infusion rate 0.1mL / h ~ 1500mL / h (4000D A is 0.1mL / h ~ 1200mL / h), the
maximum rate depends on the size of the syringe
Priming rate The default rate is 1000mL/h, but does not exceed the maximum
rate supported by the syringe.
User can set, set the range of 100 mL / h ~ the maximum infusion
rate of the syringe
49 / 64
Bolus rate The default rate is 1000mL/h, but does not exceed the maximum
rate supported by the syringe.
User can set, see "5.2 Bolus Infusion" for the range of settings.
51 / 64
Note 3: The response time of occlusion alarm may be affected by the ambient temperature and
the length of the tubing between the blocked part and the syringe interface. The above test
results are tested at the temperature of 23±2℃ and the tubing of 1 m.
Name Description
Infusion accuracy The actual infusion rate and the set infusion rate error do
not exceed ± 2%
Occlusion alarm Alarm when the pressure in the pipeline reaches the set
value
Infusion completed alarm or An alarm that occurs after the infusion of the drug
Syringe empty alarm solution in the syringe tube is completed;
Or set the VTBI of alarms that occur after the infusion is
completed.
52 / 64
12.12 Supported syringe brands and sizes
The syringe required to be used is a syringe that has been registered with the FDA and
obtained a registration certificate, and the syringes supported by the three infusion modes are
identical. Among them, the syringes of Romsons Junior 50mL, NIPRO 50mL, BD Precise
20mL and B Braun 50mL have passed the accuracy test according to the "IEC 60601-2-
24:2012 medical electrical equipment Part 2-24: Particular requirements for the basic safety
and essential performance of infusion pump and controller ".
Code Syringe brand 10mL 20mL 30mL 50mL
01 AYSET -- √ -- √
02 Fresenius -- -- -- √
Injectomat®
03 BD Plastipak™ -- √ √ √
04 Romsons Junior -- -- -- √
05 B Braun -- -- -- √
Omnifix®
06 NIPRO -- √-- √ √
07 Polfa -- √ -- --
08 TERUMO® -- √ √ √
09 Apexmed -- √ -- √
10 Unolok -- √ -- √
11 BD Precise -- √ -- √
12 Dispovan -- √ -- --
13 Custom 1 √ √ √ √
14 Custom 2 √ √ √ √
15 Custom 3 √ √ √ √
16 Custom 4 √ √ √ √
17 Custom 5 √ √ √ √
Note 1: Only 4000D B, 4000D BP supports custom brand calibration. For 4000D A and 4000D
AP, only syringes with the specified brand and specification in the table above can be
used.
53 / 64
Note 2: The 4000D B and 4000D BP custom brands can be used after the calibration function.
Only the syringes that can be supported by the 4000D B and 4000D BP syringe pumps
can be calibrated. If the accuracy is over the defined specification, the user can
calibrate the syringe before using it.
Note 3: The infusion accuracy of the calibrated brand may be over ±2%.
Note 4: This product is recommended to use BD Plastipak syringes.
Note 5: While replacing syringe, please make sure to select the syringe whose brand and
specification is same as the actual installed one, otherwise forbid to use it.
54 / 64
12.13 Curves
According to the "IEC 60601-2-24:2012 medical electrical equipment Part 2-24:
Particular requirements for the basic safety and essential performance of infusion pump and
controller " in the 201.12.1 Accuracy of controls and instruments to test. The curve is drawn
according to the data collected during the two-hour measurement period.
Startup curve
Rate : 1mL/h
Syringe brand : BD Plastipak Syringe size : 50mL
Sampling equipment : 3 sets Sampling syringes : 3 pcs
Sampling interval : △t = 0.5 min Experimental period : 120min
Rate : 5mL/h
Syringe brand : BD Plastipak Syringe size : 50mL
Sampling equipment : 3 sets Sampling syringes : 3 pcs
Sampling interval : △t = 0.5 min Experimental period : 120min
55 / 64
Trumpet curve
Rate 1.0 mL/h (Overall percentage error is 0.60%)
Syringe brand : BD Plastipak Syringe size : 50mL
Sampling equipment : 3 sets Sampling syringes : 3 pcs
Sampling interval : △t = 0.5 min
Duration of observation window: p△t = 2, 5, 11, 19, 31 min
56 / 64
Rate 5.0mL/h (Overall percentage error is 0.28%)
Syringe brand : BD Plastipak Syringe size : 50mL
Sampling equipment : 3 sets Sampling syringes : 3 pcs
Sampling interval : △t = 0.5 min
Duration of observation window: p△t = 2, 5, 11, 19, 31 min
57 / 64
Chapter 13 Electromagnetic Compatibility
Statement
This product has passed the electromagnetic compatibility test to meet the requirements of
the IEC 60601-1-2:2014 standard for medical device equipment. These limits provide
reasonable protection against harmful interference in a typical medical installation.
59 / 64
13.3 Electromagnetic environment requirements
This product is intended to be used in the electromagnetic environment specified below,
and the purchaser and user of this product should ensure that it is used in this electromagnetic
environment——Professional healthcare facility environment.
Table 1 Guidance and Manufacturer's Declaration - Electromagnetic Emissions
This product is intended to be used in the electromagnetic environment specified below, and
the purchaser or user should ensure that it is used in this electromagnetic environment.
Emission test Compliance Electromagnetic environment - guide
RF emissions Group 1 This product uses RF energy only for its internal
CISPR 11 functions. Therefore, its RF emissions are low and
there is little chance of interference with nearby
electronic equipment.
RF emissions Class A
This product is suitable for use in all facilities that are
CISPR 11
not directly connected to the home and to the public
Harmonic Not
low-voltage power supply network of the home.
emissions applicable
IEC 61000-3-2
Voltage Not
fluctuations/ applicable
flicker
emissions
IEC 61000-3-3
60 / 64
Electrostatic discharge
±8kV contact discharge ±8kV contact discharge
(ESD)
±15kV air discharge ±15kV air discharge
IEC 61000-4-2
Electrical fast
2 kV , 1KV 2 kV , 1kV
transient/burst
100kHz repetition frequency 100kHz repetition frequency
IEC 61000-4-4
Voltage interruptions
0 % UT; 250/300 cycle 0 % UT; 250/300 cycle
IEC 61000-4-11
Power frequency
30A/m 30A/m
(50Hz/60Hz) magnetic
50Hz/60Hz 50Hz/60Hz
Field IEC 61000-4-8
61 / 64
Table 3 Guidance and Manufacturer's Declaration - Electromagnetic Immunity
3Vrms
0.15MHz-80MHz
Conducted RF 3Vrms
6Vrms in ISM bands between 0.15 MHz and
IEC 61000-4-6
80MHz
6Vrms
80% AM at 1 kHz
Radiated RF 3V/m 3V/m
IEC 61000-4-3 80MHz-2.7GHz
80% AM at 1 kHz
NOTE 1: At 80 MHz the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
62 / 64
745 787 13, modulation b)
780 17 217 Hz
810 800- GSM Pulse 2 0,3 28
870 960 800/900 modulation b)
930 TETRA 18 Hz
800
iDEN 820
CDMA
850
LTE Band
5
1720 1700- GSM 1800 Pulse 2 0,3 28
1990 CDMA modulation b)
1845
1970 1900 217 Hz
GSM 1900
DECT,
LTE Band
1, 3,
4,25;
UMTS
2450 2400- Bluetooth, Pulse 2 0,3 28
2570 WLAN, modulation b)
802.11 217 Hz
b/g/n
RFID 2450
LTE Band
7
5240 5100- WLAN Pulse 0,2 0,3 9
5500 5800 802.11 modulation b)
5785 a/n 217 Hz
63 / 64
Medsure Medical Technology Co.,Ltd.
EC REP
Luxus Lebenswelt GmbH Kochstr. 1,47877, Willich, Germany
64 / 64