NoVent Microfluidics
Installation Guide
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error-
free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might
result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
Software version:
Contents
• Module Description
• Connecting the NoVent Microfludics Module to the TRACE 1300/1310
• Installing the Mounting Bracket
• Preparing the NoVent Microfluidics Restrictor Tubing
• Attaching the Ferrule and Nut to the GC Column
• Attaching the New Tubing to the Transfer Line
• Connecting the Capillaries to the Microfluidics Splitter
• Configuring the Post-Column
• Using the Module
• Ordering Parts
Module Description
The NoVent Microfluidics is a solution for maintaining carrier gas flow into the mass
spectrometer without breaking vacuum during column replacement, injector maintenance, or
both. See Figure 1.
On/Off Switch
Bottom Fittings
The module includes the solenoid valve for starting and stopping the carrier gas flow, the
On/Off switch, the On/Off status LED (Light Emitting Diode), and the fittings for the
connection to the SilFlow™ device by using a 300 mm × 75 μm i.d. fused silica restrictor.
CAUTION Where a dummy module is installed, the gas connection is blocked by a plug.
1. Cool the oven, injector or injectors, transfer line, ion source, and any installed GC
detectors to room temperature and shut down the GC.
WARNING - BURN HAZARD: The injectors, detectors, oven, and transfer line may be hot.
Allow them to cool to room temperature before touching them.
WARNING - FIRE HAZARD: If you are using hydrogen, do NOT reach over the top of the
instrument to power it off. Instead, reach around the right side or go to the back of the
instrument and flip down the power switch.
O-rings
WARNING Make sure all four o-rings are correctly seated on the gas connection. Do not
install the module if the o-rings are missing.
Note This module requires a constant carrier gas pressure of 60 psig (410 kPa).
Note If you already have a mounting bracket installed in your GC with either the SSL or
PTV backflush kit, skip this step.
1. Remove the nut at the top end of the mounting bracket. This nut secures the two sections
of the mounting bracket.
Figure 5. Mounting Bracket Parts
8. Attach the Thermo Scientific™ microfluidics splitter to the mounting bracket. The
microfluidics splitter snaps into place. The small hole should be positioned at the top.
See Figure 7 for the correct orientation.
9. Position the holder for the microfluidics splitter upright and use two 7 mm wrenches to
secure it.
Note If you are using a 0.25 mm id column, use a 0.4 mm SilFlow ferrule and a 0.4 mm
SilFlow FingerTite jig. If you are using a 0.32 mm id column, use a 0.5 mm SilFlow
ferrule and a 0.5 mm SilFlow FingerTite jig.
Ferrule
3. Insert the ferrule and the tubing into the NoVent microfluidics module.
Figure 10. Securing the 30 cm Fused Silica Restrictor to the No-Vent Microfluidics Module
7. Use a scoring wafer to cut the tubing after inserting it through the ferrule. See Figure 12.
Then use the SilFlow pre-swage tool to secure the ferrule into position. When done
properly, the tubing will extend slightly past the tip of the ferrule. It is important to use
the SilFlow pre-swage tool in order to prevent crushing the tip of the fused silica.
8. Place the column and ferrule into SilFlow pre-swage tool until the tubing reaches the
bottom of the tool. When done properly, the tubing will extend slightly past the tip of the
ferrule. It is important to use the pre-swage tool in order to prevent crushing the tip of the
fused silica. See Figure 13.
Figure 13. Inserting the Column and Ferrule into the SilFlow Pre-Swage Tool
SilFlow
Pre-Swage Tool
9. Use the SilFlow FingerTite tool to swage the ferrule to the tubing. Be sure to keep the tip
of the fused silica bottomed out in the pre-swage tool. See Figure 14.
Figure 14. Swaging the Ferrule Using the SilFlow FingerTite Tool
SilFlow
FingerTite
Pre-Swage Tool
10.Remove the jig, and lay the tubing carefully on the bottom of the GC until you are ready
to connect it to the microfluidics splitter.
Figure 15. Positioning the Nut and Ferrule Correctly on the Column
3. Use a scoring wafer to cut the column after inserting it through the ferrule. Then use the
appropriate pre-swage tool to secure the ferrule into position. When done properly, the
tubing will extend slightly past the tip of the ferrule. It is important to use the pre-swage
tool in order to prevent crushing the tip of the fused silica. See Figure 16.
4. Place the column and ferrule into the SilFlow pre-swage tool until the column reaches the
bottom of the jig. See Figure 17.
Figure 17. Inserting the Column and Ferrule into the SilFlow Pre-Swage Tool
SilFlow
Pre-Swage Tool
5. Use the FingerTite tool to swage the ferrule to the column. Be sure to keep the tip of the
fused silica bottomed out in the pre-swage tool. See Figure 18.
Figure 18. Swaging the Ferrule Using the SilFlow FingerTite Tool
SilFlow
FingerTite Tool
6. Remove the pre-swage tool, and lay the column carefully on the bottom of the GC until
you are ready to connect it to the microfluidics splitter.
Note Use the 60 cm tubing if you are using helium as a carrier gas. Use the 120 cm length
of tubing if you are using hydrogen as a carrier gas.
When connecting the column to the transfer line, you may use either the regular transfer line
nut or the spring loaded transfer line nut with the graphite Vespel™ ferrule.
3. Unwind about one turn of the column from the column outlet end.
Note Wear clean, lint- and powder-free gloves when you handle the column and
transfer line ferrule.
4. Wipe approximately 300 mm (12 in.) of the column with a tissue soaked in methanol.
5. Choose an appropriate ferrule for the outer diameter of your column.
Note If the maximum oven temperature in your method is ≥ 290 °C (554 °F),
Thermo Fisher Scientific recommends using a spring loaded transfer line nut with a
graphite Vespel ferrule or a SilTite™ nut and ferrule. By cycling the oven at and above
this temperature, expansion and contraction of the graphite Vespel material can cause
leaks in the transfer line.
6. Insert the column through the transfer line nut and ferrule, entering through the tapered
end of the ferrule. Wipe the column again with a tissue soaked in methanol.
Figure 19. Transfer Line Nut and SilTite Ferrule Orientation
Flat on the
Transferline Nut
SilTite Ferrule
7. Insert the column into the column measuring tool (see Figure 20), which is in the ISQ
Toolkit, so that it is even with the lines at the end of the column. Figure 21 indicates
proper positioning of the column in the tool for accurate measuring.
8. Use a scoring wafer to score and break the column. Use a magnifying glass to check for an
even, flat cut. Repeat if necessary.
9. Use a 5/16 in. wrench to hold the column measuring tool steady.
Figure 20. Column Measuring Tool
10. While holding the column measuring tool steady, tighten the transfer line nut with a
1/4 in. wrench until the column just stops moving in the ferrule.
11. Turn the transfer line nut 1 flat backward so the column is able to move in the ferrule
with slight resistance.
12. Line up the outlet of the column with the arrows on the end of the column measuring
tool.
Figure 21. Lining Up the Column in the Column Measuring Tool
Column Outlet
13. Place a septum with a notch cut into it behind the transfer line nut. The septum marks
the place on the column where it should exit the nut.
Figure 22. Positioning the Septum
Transfer Line
Column Nut
Measuring Tool
Septum
14. Pull the column back from the transfer line nut. Do not move the septum from its
position on the column.
Figure 23. Pulling the Column Back from the Transfer Line Nut
Column
Column
Measuring Tool
Transfer Line
Nut Septum
15. Loosen the transfer line nut from the column measuring tool.
16. Remove the column, transfer line nut and ferrule from the column measuring tool,
making sure not to move the septum from its location on the column.
17. Insert the column into the transfer line.
Figure 24. Inserting the Column into the Transfer Line
Ferrule Transfer Line
Column
Nut
18. Tighten the transfer line nut until it is just secure enough so that you cannot move it.
19. Loosen the nut by turning it exactly 1 flat backward.
20. Position the column in the transfer line. Use the septum as a guide to measure the correct
length you should insert the column. Be careful not to change the location of the septum
on the column.
Septum
21. Tighten the nut 1 flat forward—back to where it is secure enough in the transfer line that
you cannot move it.
22. Tighten the nut 1 additional quarter turn.
23. Remove the cut septum.
24. Close the front door of the GC.
Note If you are using a SilTite ferrule, follow the instructions that come with SilTite
ferrules. If you are using a graphite Vespel ferrule, they require conditioning to ensure
a leak-tight seal. See the ISQ Spare Parts Guide for information about ordering these
ferrules.
WARNING BRUN HAZARD The oven may be hot. Allow it to cool to room temperature
before opening it. The injector will still be hot, so do not touch it.
d. If you are using hydrogen as a carrier gas, replace the front panel screw.
e. Replace all remaining hydrogen safety screws if you are using hydrogen.
27. Power on the mass spectrometer.
WARNING - FIRE HAZARD: If you are using hydrogen, do NOT reach over the top of the
instrument to power it on. Instead, reach around the side or go to the back of the
instrument and flip up the power switch.
28. Once the instrument is pumped down and able to scan, click Air & Water / Tune on the
Dashboard view air water spectra and look for evidence of leaks with a large m/z 28 signal.
If you observe a leak, stop scanning and gently tighten the nut in small increments until
no leaks appear when scanning.
1. Place the ferrule connected to the original GC column into capillary B location on the
microfluidics splitter. See Figure . This corresponds to the bottom of the three holes on a
correctly installed splitter.
GC Column
Tip If a ferrule gets stuck in the microfluidics splitter, use a thumbtack or similar
pointed tool for removal. Insert the point of the tool between the ferrule and threads
and press so that the ferrule is forced off center. This will dislodge the ferrule.
2. Use the FingerTite tool to secure the nut you previously attached to the column to top of
the three holes on the microfluidics splitter.
3. Orient the capillary connected to the transfer line and the one attached to the no-vent
module as shown in Figure 27.
GC Column
Capillary Restrictor to No-Vent Module
1. On the home page of the TRACE 1310 touchscreen, click the Configuration icon.
Figure 28. Locating the Configuration Icon on the TRACE 1310 Touchscreen
Configuration Icon
2. The default GC Configuration (Oven) screen opens. Click the Back Column or Front
Column button, corresponding to where you installed the No-vent microfluidics.
3. The GC Configuration (Front or Back Column) screen opens. Click the down arrow to
access the post column settings.
4. Select Yes from the Post Column? menu. Enter the post column length and ID in the
boxes below.
Figure 29. Configuring the Post Column Settings
Home Icon
IMPORTANT For hydrogen, the Post Column Length is 1.2 m. For helium, the Post
Column Length is 0.6 m. The Post Column ID is 0.17 mm for both hydrogen and
helium.
Status Pane
Column
Setup
4. The Column Setup window opens. Check the Using Post-Column checkbox and enter
the post-column length and ID.
Post-column Settings
5. Click Apply.
6. Click Close.
Before replacing the column or performing maintenance on an injector module, flip the
switch to the On position. Ensure the mass spectrometer is in standby mode and the filament
is off. The status LED lights up as solid green indicating that the solenoid valve is activated.
The carrier gas flows into the mass spectrometer to avoid breaking vacuum. See Figure 33.
Status LED On
Switch on On Position
When you have completed your tasks, flip the switch to the Off position. The status LED
turns Off indicating the solenoid valve is deactivated.
Ordering Parts
Use Table 1 below to order new kits or spare parts for your NoVent Microfluidics system.
Table 1. Ordering information for new kits and spare parts
Description Part number
NoVent Microfluidics Installation Kit 19098025
NoVent SilTite FingerTite Column Connection Body 827310103
SilFlow FingerTite Tool 60201-401
SilFlow Nuts (10/pk) 290SF302
SilFlow FingerTite Ferrules 0.4 mm i.d. (10/pk) 29063466
SilFlow FingerTite Ferrules 0.5 mm i.d. (10/pk) 29063467
Blanking Ferrule (5/pk) 290ST414
3-Port SilFlow Replacement MCD 60201-398
170 μm Deactivated Tubing 0.363 mm OD 60 cm 60201-390
75 μm Untreated Fused Silica Tubing 0.363 OD 30 cm 60201-391
170 μm Deactivated Tubing 0.363 mm OD 120 cm 60201-394
75 μm Untreated Fused Silica Tubing 0.363 OD 80 cm 60201-395
SilTite FingerTite Slotted Nut (5/pk) 290ST130
SilTite FingerTite Ferrule 0.4 mm Hole (10/pk) 290S1132